Documente Academic
Documente Profesional
Documente Cultură
scattered radiation), one can find out how much energy from the incident
radiation transfers into the scattering material, providing a glimpse into what
occurs on a microscopic level in these materials when subject to high energy
radiation [4-5]. In addition to verifying whether or not elastic scattering is
occurring, one can also compare the values of the change in energy to a
function of the angle and measure the slope to determine off what object in
the material scattering occurs [11]. If Comptons theory is correct and the
incident radiation scatters off electrons, the slope measured should be
inversely proportional to the electron mass [2]. Furthermore, our scattering
measurements of energy versus intensity allow us to calculate the
differential cross section of the scattering material through each angle,
providing a value against which we can compare theoretical results [1, 6]. In
comparing our data to these results, one can show that quantum mechanical
derivations of cross section provide more accurate results than classically
derived derivations.
II. BACKGROUND INFORMATION AND THEORY
The attenuation length of material determines the intensity loss
detected count, r0 the background count, x the thickness of our material, and
is the attenuation length [8-10]. Since the mean free path l through a
x
l
[8-10],
it becomes clear that attenuation length is the inverse of the mean free path
and one can obtain an equation:
x
1
=l=
r0
ln ( )
r
(1)
1
nl
(2)
[9,10].
One can also calculate the theoretical absorption cross section of an electron
from its classical radius using the Thomson cross section:
T , e=
8
e
=6.661029 m2
3 4 o me c 2
8 2
T=
r r=r e
3
(3)
1cos
h
=
mc
where
(4)
the speed of light, and h is Plancks constant [2, 4-7]. Using our scattering
data, we can calculate the differential cross section of the scattering rod
using:
d
R
=
d Lnd
(5)
and integrating over our measured count vs. energy peak [11]. In equation
(5), R represents count rate, n represents target particle number density, L
represents beam luminosity (obtained by taking a measurement without the
rod at 0 degrees), and is the solid angle. Additionally, we can compare the
measured cross section to the fully quantum mechanical Klein Nishina cross
section, defined differentially as:
2 ( 1cos )2
1+ (1cos )
1+cos 2
d kn r 2
1
=
d 2 [ 1+ ( 1cos ) ] 2
+
(6)
mc
(7)
where r is the radius of the particle scattering the wave, m is the mass, and
is the angle at which the scattering occurs [3, 11]. In the case of <<1,
(a)
(b)
(8)
where E is the energy in MeV and c is the input channel, from 1 to 16384.
without the scattering rod, for only 10 minutes each time. This
measurement, used to filter out both background radiation and the energy
going straight from the source to the detector without scattering (or
scattering off the lead-lined box), can be scaled up to match the time period
of the overall collection without significantly impeding the accuracy of our
results.
IV. RESULTS AND DISCUSSION
IV.A. Absorption Results
First, to calculate our mean free paths in lead, we averaged the
number of measured counts minus background radiation over our three
experimental runs, keeping track of our standard deviation. We also
averaged the number of counts measured with no blocker, taking the
standard deviation as our error. We then calculated the mean free paths in
lead and aluminum using equation (1), carrying through the error from the
averages of our blocker and background counts. We then averaged the
results across the varying thicknesses of each sample of material in order to
provide more accurate final results for that material. Ultimately, we obtained
a mean free path of 0.750 2.80 * 10-3 cm in lead and 5.60 1.41 * 10-3 cm
in Aluminum.
We then calculated the absorption cross sections of the materials using
the known densities of our samples and equation (2) to obtain 3.74 * 10-27
1.17 * 10-29 m2 for lead atoms and 3.13 * 10-28 8.21 * 10-32 m2 for aluminum
10
atoms [Refer to APPENDIX C for results for each sample]. The lead atomic
cross section differs by 5520% from the Thomson electron cross section,
while the aluminum atomic cross section differs by 370%. Calculating the
electron cross section by multiplying our number density by the number of
electrons per atom (82 and 13, respectively [12]), however, brings us much
closer to the Thomson cross section, with electron cross sections of 4.56 * 10 29
1.43 * 10-31 m2 for lead and 2.41 * 10-29 6.32 * 10-33 m2 for aluminum
(differing by 31.8% and 64% respectively). Our data lies even closer to the
total Klein-Nishina cross section, 3.67 * 10-29 m2 (found by integrating over
equation (6)), with only 24.3% and 34.3% errors respectively. This suggests
that, not only are electrons responsible for absorption of 662keV gamma rays
in these materials, but that the Klein-Nishina cross section provides a more
accurate result than the classical Thomson cross section.
Additionally, our measured atomic absorption cross sections differ from
the physical cross sections of the atoms, which are 7.45 * 10-20 m2 and 4.37 *
10-20 m2 [12], by over a factor of 107. This further implies that the atoms are
not absorbing the photons as a whole.
IV.B. Scattering Results
In order to analyze our scattering data, we first found the energy vs.
the intensity of our measurements by using our calibration fit in equation (8).
Then, we graphed the radiation detected and fit the energies with the
highest 10% of counts to a Gaussian. We needed to remove the bottom 90%
11
of points in order to get an accurate fit the filtered points were low energy
background noise that would have shifted our Gaussian away from the
proper peak [See FIG. 2]. The peak of our Gaussian fit provided us with a
shifted energy, which we then compared to the predicted energy at that
(a)
incident angle using equation (4) to determine that elastic scattering did
occur [See FIG. 3]. Additionally, this Gaussian fit gave us a standard
(b)
12
FIG. 2. A graph of our fit Gaussians and our measured counts at (a) 0 degrees
and (b) 10 degrees. Using only the top 10% of counts for our fit filtered out
the low energy background noise, especially the points below 450keV in
these images, which added extra data at energies below our peak. This noise
would have drastically shifted our fit away from the data we were actually
interested in. [R. Magee, unpublished, (2016)]
FIG. 3. Graph of scattering angle vs. scattered photon energy. Blue data
points represent our experimental data with error, while the red curve
represents the theoretical scattered photon energy. The data matches
13
closely, but begins to drift away from theoretical calculation at larger angles,
likely because of the increasingly low probability of backscattering as one
passes 90 degrees. [R. Magee, unpublished, (2016)]
Because elastic scattering occurred, by graphing the change in energy
divided by the incident and scattered energies versus 1-cos [see FIG. 4], we
found that our radiation scattered off an object with mass 415 13.1 keV/c2.
This value comes much closer to the accepted electron mass, 511 keV/c 2 [6],
than that of the much more massive proton or neutron mass, suggesting that
electrons scattered our gamma rays. However, our result still varies by
18.8% from the accepted value of the electron mass, likely as a result of our
consistently lower than expected scattered energies at larger angles.
14
FIG. 4. Graph of energy shift divided by the product of the incident and
scattered energies vs. 1-cos. By fitting a line to our graph and measuring
the slope, as suggested by equation (4) and the de Broglie wavelength, we
found that our gamma radiation elastically scattered off a particle with mass
415 13.1 keV/c2 (the inverse of our slope). [R. Magee, unpublished, (2016)]
Additionally, using the properties of our apparatus, such as detector
hole size, distance between the source and the detector, and source
luminosity, we calculated the differential scattering cross section of the
electrons in our scattering rod with equation (5) [Refer to APPENDIX D for
more details and results]. They matched the Klein-Nishina cross section,
calculated using equation (6), fairly accurately at most values, though with
what appears to be a systemic error [see FIG. 5]. However, at 0 and 10
degrees, our data varies drastically from the Klein-Nishina results. We believe
this occurs because at these small angles, some of our gamma radiation will
pass around the rod and enter the detector (because of the small size of the
rod diameter compared to our detector and source holes), rather than
through it or scattering off it, distorting our intensity measurements.
15
materials. For example, lead could be used to store materials and create
personal protective gear, while aluminum films could be used in situations
where one wishes to block physical particles (like air) without blocking much
radiation. Additionally, they tell us that, on an atomic level, particles
travelling through lead will collide with the particles constituting the material
much more often than in aluminum. Our absorption cross section results
further suggest that the electrons in the materials are responsible for the
majority of photon absorption our measured electron cross sections vary
only slightly from the two theoretical electron cross sections, while our
measured atomic cross sections vary from the physical atomic cross sections
by massive factors. On top of this, because our results lie closer to the KleinNishina cross section than the Thomson cross section, our data further
supports quantum theory over classical theory.
Our scattering results point to a couple of interesting conclusions.
Firstly, the fact that scattering results in an angle dependent energy
decrease explains why scattered radiation penetrates less deeply into
materials, solving a longstanding mystery in physics [4].
Secondly, at 662keV, the scattering interaction between gamma
radiation and aluminum is dominated by elastic scattering off electrons,
rather than protons or neutrons. However, our results still varied from the
mass of the electron by 18.8% error, suggesting that, in the future, longer
and more precise measurements, possibly taken at smaller angular intervals
for longer times, should be used to verify that electrons are doing the
17
18
(b)
20
21
FIG. 7. Graph of counts vs. voltage used to calibrate the power supply.
Counts appear to increase until about 0.8kV, where they roughly plateau
22
until 1.0kV. After 1.0kV, detected events decreases rapidly with between 1540% per 0.1kV of additional voltage supplied. [R. Magee, unpublished,
(2016)]
23
Area Density
2
(mg/cm )
6.35 * 10
-1
7435 8.04 * 10
-1
5.39 * 10
-3
3.64 * 10
-27
2.44 * 10
-29
3.17 * 10
-1
3632 7.15 * 10
-1
2.19 * 10
-3
4.19 * 10
-27
1.28 * 10
-29
1.62 * 10
-1
1890 7.17 * 10
-1
1.33 * 10
-3
4.12 * 10
-27
7.66 * 10
-30
8.12 * 10
-2
1230 7.60 * 10
-1
4.75 * 10
-4
2.98 * 10
-27
1.86 * 10
-30
-1
2.35 * 10
-3
3.73 * 10
-27
1.17 * 10
-29
24
d =
r2
d2
(D1)
where r was detector hole radius and d was the distance between our scattering rod
and detector hole [Refer to APPENDIX A for values].
25
R=
NN b
Time
(D2)
where N is the total count and comes from our data taken with the rod, Nb is
the background count comes from data taken without the rod, and time is
the total amount of time that we took data for.
We could then proceed to calculate the differential cross section with
equation (5) [see TABLE III], taking the error in our counts to be N0.5 using
Poisson statistics. Error in our other values arose from uncertainty in our
physical measurements. Using these values, we could then carry through the
error using standard error algebra.
26
27