Documente Academic
Documente Profesional
Documente Cultură
A. Graphical form
IEC 60751, 2008, ed. 2
Temperature (
C)
7
detectors
Extended range for wirewound detectors
Class C
Tolerance (C)
5
4
Class B
Class B
3
2
Class A
600
550
500
450
400
350
300
200
150
100
50
-50
-100
-150
250
Class
Class
AAAA
0
-200
Class A
Temperature (C)
B. Table
Tolerance class
AA
A
B
C
Tolerance values
C
(0.1 + 0.0017 | t |)
(0.15 + 0.002 | t |)
(0.3 + 0.005 | t |)
(0.6 + 0.01 | t |)
| t | = temperature in
C without regard to sign.
The temperature -196
C is close to the boiling point of nitrogen (N) which is used for calibration.
C. Comments
Compared to the earlier versions of this standard the present one has been adapted to cover more models of
Pt100 detectors (resistors). Where resistors are fulfilling extended or smaller ranges of validity or other
tolerance classes, this must be clearly stated between manufacturer and user.
The tolerance classes apply to all detector values (R 0) at 0
C, i. e. Pt100, Pt1000 and others.
The relationship temperature-resistance has not been changed compared to the previous standard.
2009-06-09