Specification Sheets
Submittal Package
as of 2/24/2012
www.tektone.com
Copyright 2012 TekTone Sound & Signal Mfg., Inc., All rights reserved.
Phone:
(828) 524-9967
Toll-Free: (800) 327-8466
Sales:
Option 2
Tech Support: Option 3
No part of these publications may be copied without the express written permission
of TekTone Sound & Signal Mfg., Inc. The contents of these specification sheets are
furnished for informational use only, are subject to change without notice, and should not
be construed as a commitment by TekTone Sound & Signal Mfg., Inc. TekTone Sound
& Signal Mfg., Inc. assumes no responsibility or liability for any errors or inaccuracies
that may appear in this documentation.
IL770
Section C
Rev. 2 - 10/2009
www.tektone.com
FEATURES
The NC401TS Master Station with Touchscreen is a primary
component of the Tek-CARE400 Nurse Call System.
The NC401TS is the user interface for responding to calls,
issuing staff requests, sending custom pages, viewing staff
locations, and altering room programming to accommodate
differing patient needs.
The master station is completely self-contained, including a
built-in 10.4" color touchscreen LCD display, quick-access
function buttons, and a pull-out keyboard drawer. The
NC401TS displays up to 20 incoming calls simultaneously,
allowing staff to select or scroll to any active call using
just the touchscreen. Each call displays a unique timer that
increments until that call is reset. Calls and staff locations
may also be displayed as room icons arranged by user
preference into lockable positions on the touchscreen.
The master stations programming information and power
are provided by the NC450 Central Equipment. Therefore,
any NC401TS master station on the system can be replaced
with another NC401TS without any reprogramming.
The NC401TS master station supports full-duplex audio
to the patient stations, and allows staff to adjust incoming
and outgoing volume for each station. Tone volume can be
adjusted to increase the volume during the day and reduce
the volume at night. Routine-level calls may be silenced
using the master stations tone- off button to further reduce
noise at the nurses station.
The membrane keypad provides these function buttons:
Button
alpha
Alpha/Numeric
Arrow keys
enter
next
view
page
monitor
priority
zone
volume
tone silence
menu
stat
l1
l2
l3
Function
Toggle from alpha to numeric key functions
AZ, SPACE, 19
up, down, right, left for menu navigation
Selects dialed item
Cycles between beds on a single station
View a selected item
Audio page to selected zones
Monitor selected zones
Cycle priority for selected station
Select zone from list
Access master station volume dialog
Disable audible call annunciation*
Access drop-down menu
Upgrade/place staff emergency call** to
selected station
Level 1 service request
Level 2 service request
Level 3 service request
SPECIFICATIONS
Dimensions:
Capacity:
Environments:
13.2"9"5.5" (335mm229mm140mm)
Address 640 stations in up to 256 zones
26C, relative humidity should not
exceed 80%
Terminations:
Plug-in connector to CE
System Protection: CE power limited by self-resetting current
limiting device. ESD protected to 8kV.
IL771
Section C
Rev. 1 - 09/2010
www.tektone.com
FEATURES
The NC401VFD Master Station is a primary component of the
Tek-CARE400 Nurse Call System. The NC401VFD is the
user interface for responding to calls, issuing staff requests,
sending custom pages, viewing staff locations, and altering
room programming to accommodate differing patient needs.
The master station is completely self-contained, including
a 440 Vacuum Fluorescent Display, 20 programmable
function buttons, alphanumeric and navigational keys, and a
pull-out drawer with quick reference user instructions. The
NC401VFD displays up to 4 incoming calls simultaneously
and allows staff to select or scroll to any active call using
navigational keys on the keypad. Each call displays a unique
timer that increments until that call is reset. Staff locations and
system faults are displayed on separate dedicated screen views
with pending status indications visible on the main call screen.
The master stations programming information and power
are provided by the NC450 Central Equipment. Therefore,
any NC401VFD master station on the system can be replaced
with another NC401VFD without any reprogramming.
The NC401VFD master station supports full-duplex audio
to the patient stations, and allows staff to adjust incoming
and outgoing volume for each station. Tone volume can be
adjusted to increase the volume during the day and reduce
the volume at night. Routine-level calls may be silenced
using a programmable button assigned to the tone-off
function to further reduce noise at the nurses station.
Description
Level 1 service request to selected station
Level 2 service request to selected station
Level 3 service request to selected station
Upgrade/place Staff Emergency call to
selected station
tone silence Disable audible call annunciation
monitor
Monitor audio in selected zones
attend
Display staff at selected station
follow
Send incoming calls to selected station
hold
Returns current call to the call list
The membrane keypad also provides these standard keys:
Button
Function
Toggle from alpha to numeric key functions
Alpha/Numeric AZ, SPACE, 19
Arrow keys up, down, right, left for menu navigation
#
Selects bed when dialing a station
enter
Selects dialed item; cycles between beds on
a single station
menu
Cycles between Call Screen, Fault Screen,
Staff Screen and User Screen
reset
Resets an audio connection or clears a
selection
talk
Establishes an audio connection
alpha
SPECIFICATIONS
Dimensions:
Capacity:
Environments:
13.2"9"3.5"(335mm229mm89mm)
Address 640 stations in up to 256 zones
26C, relative humidity should not
exceed 80%
Terminations:
Plug-in connector to CE
System Protection: CE power limited by self-resetting current
limiting device. ESD protected to 8kV.
IL772
Section C
Rev. 1 - 09/2010
FEATURES
The NC450 Central Equipment consists of a PM450
Backplane Assembly installed in an IH450 Housing,
equipped with a PM457 Power Entry Module and PM466
AC Power Bus Module. The NC450 also includes the
LS450 Config Tool Programming Software.
All add-on hubs, modules and power supplies use keyed
plug-in connections for easy insertion/removal and to
prevent connection errors. Each power supply is UL
and cUL 2601 approved, and supplies 28VDC and 100
watts of power.
The NC450 includes mounting brackets and dedicated
connections for:
one PM451 Hub Control Module,
one PM452 CE Communications Module,
seven PK450 Power Supplies, and
seven BA450K Battery Backup Kits;
SPECIFICATIONS
Dimensions:
Input Power:
Temperature:
Humidity:
REQUIRED COMPONENTS
www.tektone.com
IL844
Section C
Rev. 1 - 09/2010
FEATURES
The PK450 is a U-channel mounted 120W power supply.
It has a power output of 90W with convection cooling,
and an output voltage of 28VDC at 3.75 amps. It will
operate with an input from 90264VAC at 4763Hz. It is
UL 1069 Listed with class B EMI filtration. It offers
an MTBF of at least 250,000 hours.
SPECIFICATIONS
Dimensions:
Input Power:
Temperature:
Humidity:
REQUIRED COMPONENTS
Tek-CARE400
NC450
www.tektone.com
IL776
Section C
Rev. 1 - 09/2010
FEATURES
The BA450K Battery Backup Kit provides temporary
uninterrupted DC power to the nurse call system in the
event of primary power failure. Switchover from primary
power to the BA450K is automaticas is recharging
the systems batteries upon restoration of normal power.
Backup power is supplied through the use of two sealed,
lead-acid batteries, each rated at 12 volts for at least two
amp-hours. Batteries are individually replaceable, with
connections made via male quick-disconnect terminals.
SPECIFICATIONS
Dimensions:
(per battery)
Battery:
Connections:
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
www.tektone.com
IL845
Section C
Rev. 1 - 09/2010
FEATURES
The PM451 Hub Control Module is a 10/100BaseT
switching hub with 3 status LEDs for each of its 8 ports.
The PM451 provides intercommunication between various
central equipment (CE) modules, and is required for any
system that has more than one PM453 Master & Station
Module, PM455 P5 Master & Station Module, and/or
PM456 NC300II Head End Module. One of the hubs ports
is isolated with a standard RJ45 ethernet connection for
ancillary devices such as the NC470 Tek-BRIDGE PC, or a
laptop computer running the LS450 Config Tool. All ports
are isolated to help shield the CE from externally induced
damage and to help prevent external voltage differentials
from affecting the nurse call system.
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
REQUIRED COMPONENTS
Tek-CARE400
NC450
www.tektone.com
IL846
Section C
Rev. 1 - 09/2010
FEATURES
The PM452 CE Communications Module enables reliable
networking of two or more NC450 central equipment (CE)
cabinets using a single pair of #24 twisted pair wiring over
distances of up to 2000' (457 m). Multiple CEsup to 15
totalmay be linked together, each up to 1500' (457m)
from its neighbor in the network. This allows systems to
span many thousands of feet, while maintaining both data
and audio integrity.
PM452
CE Communications
Module
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
REQUIRED COMPONENTS
Tek-CARE400
NC450
www.tektone.com
IL848
Section C
Rev. 1 - 09/2010
FEATURES
The PM454 Pager Module has an isolated RS232 serial port
designated for a compatible paging transmitter. The port is
isolated to help shield the NC450 Central Equipment from
externally induced damage and to help prevent external
voltage differentials from affecting the nurse call system.
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
NC450 Central Equipment
NC369 Tek-PAGING Paging Transmitter
NC399P Alphanumeric Pager
www.tektone.com
IL909
Section C
Rev. 0 - 09/2010
FEATURES
The PM455 P5 Master & Station Module connects two
master stations and 64 P5-series patient stations to the
NC450 Central Equipment (CE). Up to five PM455
modules may be installed per CE, allowing up to 10 master
stations and 320 P5-series patient stations per CE.
The PM455 P5 Master & Station Module can drive up to
1000' (305 m) of cable per connector pair. Master station
configuration data is held within the module, allowing
master stations to be replaced without reprogramming.
The PM455 P5 Master & Station Module is UL 1069
Listed and cUL Listed to CSA C22.2 No. 205.
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
Height: 10.6"
(269 mm)
Width:
2.376"
(60 mm)
Depth:
4.082"
(102 mm)
070C.
595%, non-condensing
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
NC450 Central Equipment
IR400P5-series Stations
www.tektone.com
IL861
Section C
Rev. 1 - 09/2010
FEATURES
The PM456 NC300II Head End Module enables connection
of two Tek-CARE400 master stations and 64 TekCARENC300II patient stations to the NC450 central
equipment (CE), without rewiring the station runs. Up to
five PM456 modules may be installed per CE, allowing
up to 10 master stations and 320 patient stations per CE.
The NC300II Head End Module can drive up to 1000'
(305m) of cable per port.
PM456 NC300II
Head End Module
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
REQUIRED COMPONENTS
Tek-CARE400
NC450
www.tektone.com
IL862
Section C
Rev. 0 - 10/2009
FEATURES
The PM464 Telephone Interface Module provides an input
for an analog telephone line through which an operator
may access the nurse call system remotely to perform
normal functions for patients.
PM464
Telephone Interface
Module
SPECIFICATIONS
Dimensions:
Temperature:
Humidity:
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System, including:
NC450 Central Equipment
NC464 Central Control Module (CCM)
PM454 Pager Module or NC470 Tek-BRIDGE PC
PM464 Telephone Interface Module
Wireless Telephone System (by others)
PBX System (by others)
www.tektone.com
IL872
Section C
Rev. 0 - 10/2009
FEATURES
The NC464 Central Control Module (CCM) is a component
of TekTone s telephone interface system for the
Tek-CARE400 nurse call system. The NC464 CCM
adds the ability to initiate voice communication between
the Tek-CARE400 nurse call system and the wireless
telephone system. It also sends text messages from TekCARE400 master stations and patient stations via a PBX
system to a wireless telephone system.
SPECIFICATIONS
NC464
Power Supply:
Dimensions:
Operating
Temperature:
Weight:
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System, including:
NC450 Central Equipment
NC464 Central Control Module (CCM)
PM454 Pager Module or NC470 Tek-BRIDGE PC
PM464 Telephone Interface Module
Wireless Telephone System (by others)
PBX System (by others)
www.tektone.com
IL773
Section C
Rev. 1 - 09/2010
FEATURES
The NC455 Chassis Short Monitoring Device monitors
a singleNC450 Central Equipment for any short circuit
of 1M ohms or less between system power and chassis
ground. The device has a test button as well as a tone
silence button that silences the audible alarm without
extinguishing the alarm LED.
SPECIFICATIONS
Dimensions:
Mounting:
Connections:
Height: 4.5"
(114 mm)
Width:
2.8"
(71 mm)
Depth:
1.45" (37 mm)
Standard single-gang electrical box
Plug-in type
REQUIRED COMPONENTS
Tek-CARE400
www.tektone.com
IL774
Section C
Rev. 2 - 01/2012
FEATURES
The NC460 Power Fault Monitor provides supervision
of a master port power line. The device includes
integrated battery backup via a replaceable 9V battery,
for annunciation in the event of power failure. The device
includes both a test button and a test-okay LED to verify
its condition. It includes a silence button that stops the
audible alarm without extinguishing the alarm LED.
Install one NC460 per Tek-CARE400 network to ensure
visual and audible indication of single-fault power losses.
Install the NC460 on the power output of a PM453, PM455
or PM456 master station port where an active master
station is also attached.
SPECIFICATIONS
ARCHITECTS AND ENGINEERS
SPECIFICATION
Power Fault Monitor shall be TekTones NC460 or
approved equal. The module shall sound an alarm in the
event of a power failure to the nurse call system, and
shall use a replaceable 9V battery to provide power for
annunciation in the event of power failure. The module
shall have a button to manually test the alarm mechanism,
a button to silence the audible alarm, and an indicator light
to signal an okay condition during testing. The module
shall fit into a single-gang electrical enclosure. To ensure
visual and audible indication for single-fault power losses,
one NC460 shall be installed per Tek-CARE400 network.
The NC460 shall be installed on the power output of a
PM453, PM455 or PM456 master station port where an
active master station is also attached.
Dimensions:
Mounting:
Connections:
Height: 4.5"
(114 mm)
Width:
2.8"
(71 mm)
Depth:
1.45" (37 mm)
Standard single-gang electrical box
Plug-in type
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
Plus, one or more of these modules:
PM453 Master & Station Module
PM455 P5 Master & Station Module
PM456 NC300II Head End Module
www.tektone.com
IL856
Section C
NC470 Tek-BRIDGE PC
Specification Sheet
Rev. 1 - 05/2010
FEATURES
NC470
COMPONENTS
www.tektone.com
IL894
Section C
Rev. 2 - 06/2011
FEATURES
The LS451 Versus Tracking Interface Software runs on
the NC470 Tek-BRIDGE PC, which is networked to
the Tek-CARE400 Nurse Call System and the facilitys
Versus Information System (VIS).
Versus Tracking Interface Software receives tracking badge
location and button-press information from the VIS, and
displays it on the Tek-CARE400 Nurse Call System.
Staff and equipment locations are displayed in the list
view of NC401TS Touchscreen Nurse Master Stations,
and a search function is available via the menu button.
Programming is via an online configuration tool, and can
configure the nurse call system to:
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
NC470 Tek-BRIDGE PC
LS451 Versus Tracking Interface Software
Versus Information System & Versus Badges
NC401TS Touchscreen Nurse Master Station
www.tektone.com
IL876
Section C
Rev. 1 - 06/2010
FEATURES
The LS452 ADT Interface Software using the HL7 Standard
runs on the NC470 Tek-BRIDGE PC, which is networked
to the Tek-CARE400 Nurse Call System and the facilitys
hospital information system (HIS).
The LS452 software receives patient information from
the facilitys Admit, Discharge, Transfer (ADT) system
including patient name, room number, bed number, and
doctor nameand inserts it into the corresponding fields in
the Tek-CARE400 Nurse Call Systems patient database.
Information in these fields is then available for display at
NC401TS Touchscreen Nurse Master Stations.
The LS452 software is pre-installed on the NC470 TekBRIDGE PC. The software is activated using a software
component license (purchased separately), and enabled via
the LS450 Config Tool programming software.
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
NC470 Tek-BRIDGE PC, including LS452 ADT
Interface Software using the HL7 Standard
LS452 Software Component License
Connection to the facilitys network
NC401TS Touchscreen Nurse Master Station
Admit, Discharge, Transfer (ADT) system using the
HL7 Standard, connected to the facilitys network
A port to listen for messages from the ADT system
www.tektone.com
IL790
Section C
Rev. 1 - 05/2010
FEATURES
TekTones LS486 Tek-MMARS 400 Management
Monitoring And Reporting Software provides realtime nurse/patient communications monitoring. TekMMARS400 allows nurse call system activity to be
monitored on the Tek-BRIDGE PC and stored for followup review and evaluation. All reports may be printed to
hard copy, as required. When connected to a facilitys
LAN, the Tek-BRIDGE PC provides these services on
networked computers via a web browser.
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
NC470 Tek-BRIDGE PC
LS486 Tek-MMARS400 Software
www.tektone.com
IL900
Section C
Rev. 0 - 09/2010
FEATURES
The IR400P5 Multipurpose Station provides audible,
visual and digital communications to and from the TekCARE400 central equipment (CE) when receiving or
sending signals to remotely connected devices.
The IR400P5 includes inputs for contact closures and all
types of emergency call switches for these systems. It also
provides an output to a speaker or paging amplifier, and
a corridor lamp. It does not have the controls, LEDs or
speakers normally associated with patient, staff or duty
stations.
Some practical applications for the IR400P5 are to provide:
input for code emergency call stations in an ICU; a PA
amplifier interface for the nurse call master; a control point
for exit/entry doors and narcotics cabinets; input for bath
emergency call stations in public bath areas; or input to a
remote door intercom station for access control.
SPECIFICATIONS
Dimensions:
Construction:
Connections:
Housing:
Wiring:
Height: 4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL 94 HB) plastic
Plug-in connectors for peripherals
RACO 692, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry boxes
must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
IL901
Section C
Rev. 0 - 09/2010
FEATURES
The IR410P5 Staff Station provides reliable full-duplex,
hands-free communication between the nurse master
station and utility room, examination rooms, kitchen, or any
location where staff members may need to originate calls
or where call cords are not required. Calls are placed using
either the staff emergency button or the programmable staff
call button. Multicolored LEDs on the panel indicate that a
call has been placed and that the intercom is in use. As an
additional feature, the staff station is field programmable
to function as a patient station, if desired.
SPECIFICATIONS
ARCHITECTS AND ENGINEERS
SPECIFICATION
Staff Stations shall be TekTone IR410P5 or approved
equal. Wiring from the central equipment (CE) to stations
shall be fully supervised. All station control and power
will be supplied over CAT5 cable. Staff stations shall
include three programmable push buttons (call, reset, and
emergency by default) and two multicolored call-placed/
status LEDs. The station shall have a 2.5" 2.5" speaker and
a separate microphone for full-duplex audio. The station
shall be field programmable to emulate a patient station,
if desired. All station components shall be mounted on
an attractive, flame-retardant (UL 94 HB) plastic panel.
Wiring connections shall be plug-in.
The IR410P5 shall be UL 1069 Listed and cUL Listed
to CSA C22.2 No. 205.
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height: 4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL 94 HB) plastic
2.5" 2.5" (64 mm 64 mm)
Plug-in connectors for peripherals
RACO 692, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry boxes
must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
IL902
Section C
Rev. 0 - 09/2010
FEATURES
IR411P5
Patient
Station
IR412P5
Patient
Station
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height: 4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL 94 HB) plastic
2.5" 2.5" (64 mm 64 mm)
Plug-in connectors for peripherals
RACO 692 (RACO 697 when PM421P5
or PM422P5 is present), or Steel City
#H3BD with Steel City #3GC plaster
ring, or exact equal. Conduit access for
masonry boxes must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
IL908
Section C
PM421P5, PM422P5
Pillow Speaker Modules
Specification Sheet
Rev. 1 - 11/2011
FEATURES
The PM421P5 and PM422P5 Pillow Speaker Modules
provide intercom through SF41ZDL and SF41ZKL pillow
speakers, plus on/off control of two lamp sources
generally one room lamp and one reading lamp. The
PM421P5 is designed to plug into IR411P5 Single Patient
Stations; the PM422P5 is designed to plug into IR412P5
Dual Patient Stations. If desired, SF401A Call Cords may
be used without removing the pillow speaker modules.
SPECIFICATIONS
Dimensions:
Power:
Construction:
Height: 2.56"
(65 mm)
Width:
5.02"
(128 mm)
Depth:
0.59"
(15 mm)
2428 VDC, 120 mA
PC board with quality electronic
components.
Station Housing: RACO 697, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry boxes
must be through the top.
Connections:
Plugs into IR411P5 or IR412P5 patient
station using keyed dual row header on
edge of board.
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
IR411P5 or IR412P5 Patient Station
SF41ZDL or SF41ZKL Pillow Speaker
www.tektone.com
IL903
Section C
Rev. 1 - 07/2011
FEATURES
The IR415P5 Duty Station provides reliable full-duplex,
hands-free communication between the nurse master
station and utility rooms, examination rooms, kitchen, or
any location where a nurse may be away from the master
station. When a call from a duty station is answered at
the master station, the LED indicator lights red and the
nurse may then carry on a conversation from the remote
location. A patient station call is indicated audibly and
visually at the duty station to which it is assigned. Station
and peripheral-switch calls are indicated by one of the
four colored LEDs, with a field-programmable flash rate.
Corridor light indication is also provided. All calls initiated
from stations other than the duty station can be reset only
at the master station or point of origin.
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height: 4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL 94 HB) plastic
2.5" 2.5" (64 mm 64 mm)
Plug-in connectors
RACO 692, or Steel City #H3BD with
Steel City #3GC plaster ring, or exact
equal. Conduit access for masonry boxes
must be through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
www.tektone.com
IL904
Section C
Rev. 1 - 03/2011
FEATURES
The IR418P5 Resident Station provides reliable two-way
hands-free communication between patient and staff, and
provides for programmable privacy, paging and patient
priority from the master station. Calling the nurse is
accomplished by pulling on the end of the pull cord. An
LED on the panel indicates the call has been placed and
the intercom is in use. Calls are canceled by pushing the
reset button.
SPECIFICATIONS
ARCHITECTS AND ENGINEERS
SPECIFICATION
Resident Station shall be TekTone IR418P5 or approved
equal. Wiring from the central equipment (CE) to stations
shall be fully supervised. All station control and power will
be supplied over CAT5 cable. The resident station shall
include a reset button, a 6' washable TPU plastic pull cord
for call placement, and a multicolored call-placed/status
LED. Call type shall be programmable. The station shall
use a 2.5" 2.5" speaker/microphone for accurate voice
reproduction. All station components shall be mounted on
an attractive, flame-retardant (UL 94 HB) plastic panel.
Wiring connections shall be plug-in.
The IR418P5 shall be UL 1069 Listed and cUL Listed
to CSA C22.2 No. 205.
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height: 4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL 94 HB) plastic
2.5" 2.5" (64 mm 64 mm)
Plug-in connectors for peripherals
Steel City #H3BD with Steel City #3GC
plaster ring, or exact equal. Conduit
access for masonry boxes must be
through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
IL905
Section C
Rev. 1 - 03/2011
FEATURES
The IR445P5 Resident Station with Check-In provides
reliable two-way hands-free communication between
resident and staff, plus programmable privacy, paging
and patient priority from the master station. The check-in
function eliminates up and about checksthe resident
pushes the stations reset/check-in button, which removes
the residents apartment number from the master stations
not checked-in list. Multiple check-in times can be
used. Calling the master station is accomplished either by
pulling the 6' long help-call cord mounted on the panel,
or by pushing the routine call button. LEDs on the panel
indicate a call has been placed and the intercom is in use.
Calls are canceled by pushing the reset/check-in button.
SPECIFICATIONS
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Wiring:
Height: 4.5"
(114 mm)
Width:
7.0625" (179 mm)
Depth:
up to 2.5" (64 mm)
Projects: 0.5"
(12.7 mm)
Flame-retardant (UL 94 HB) plastic
2.5" 2.5" (64 mm 64 mm)
Plug-in connectors for peripherals
Steel City #H3BD with Steel City #3GC
plaster ring, or exact equal. Conduit
access for masonry boxes must be
through the top.
Control and power over CAT5 cable
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
PM455 Master & Station Module
SF300-series peripheral switches
www.tektone.com
IL767
Section E
Rev. 4- 08/2010
FEATURES
The TekTone SF401A single call cord is for use with
the Tek-CARENC300II nurse call system, where 8-pin
DIN jacks are used. The SF401A is an oxygen safe call
cord that is certified safe for use by all patients receiving
oxygen therapy. This waterproof and impact resistant call
cord has a completely sealed push button.
UL 1069 Listed
Oxygen safe
Impact resistant and waterproof
Small, lightweight pendant fits comfortably in hand
Bed clips included
SF401A
SPECIFICATIONS
Connection:
Construction:
Dimensions:
REQUIRED COMPONENTS
ARCHITECTS AND ENGINEERS
SPECIFICATION
Tek-CARENC300II
IR319B or IR320B
www.tektone.com
IL824
Section E
SF41ZKL (standard)
FEATURES
SF41-series Pillow Speakers are used with Tek-CARE400
IR411- and IR412-series, IR411P5 and IR412P5 patient
stations, and Tek-CARE NC300II IR319- and IR320series patient stations. The pillow speakers provide call
placement to the master station, as well as remote control
for hospital-grade digital TVs (Zenith is standard; RCA
and Philips are available by special order). When used in
conjunction with the appropriate TekTone pillow speaker
modules, the pillow speakers also supply two-way, handsfree communication with the master station, and automatic
silencing of TV audio while the intercom is in use.
Push buttons are provided to call the nurse, plus TV power,
channel selection up/down, Closed Captioning, mute and
two light controls. Buttons are also provided for volume
up/down (analog volume controls are available by special
order), and for direct channel access (in some models).
SF41ZDL (standard)
Rev. 6 - 09/2010
UL 1069 Listed
Oxygen safe
Fully sealed for easy cleaning
Flame-retardant (UL 94V-0) high-impact plastic
Long lasting switches surpass UL1069 requirements
8-foot cord with DIN plug and security clip
Provides pillow speaker intercom and light control
(when used with the appropriate modules)
Special Order: RCA or Philips TV controls, analog
volume control, 10-foot cord, earphone jack
REQUIRED COMPONENTS
Hospital-grade Zenith Digital TV
for Tek-CARENC300II Nurse Call Systems:
IR319- or IR320-series Patient Station
PM311C or PM312C Pillow Speaker Intercom Module
PM321B or PM322B Light Control Module
for Tek-CARE400 Nurse Call Systems:
IR411- or IR412-series Patient Station
PM421 or PM422
Pillow Speaker & Bed Interface
Module
for Tek-CARE400 P5 Nurse Call Systems:
IR411P5 or IR412P5 Patient Station
PM421P5 or PM422P5 Pillow Speaker Module
www.tektone.com
IL829
Section E
Rev. 1- 08/2010
FEATURES
TekTone SF401G Call Cord is for use with the TekCARENC300II Nurse Call System, where 8-pin DIN jacks
are used. Patient stations include IR319/A and IR320/A.
The SF401G Call Cord is specifically designed for use
in geriatric and oxygen environments, but may also be
used in general applications. To place a call, the patient
need only apply slight pressure to the squeeze bulb using
any part of the body. The force of air through the vinyl
tubing closes a switch in the plug, thus placing a call.
An adjustable bed clip is supplied for the SF401G and is
attached to the call cord.
SF401G
SPECIFICATIONS
ARCHITECTS AND ENGINEERS
SPECIFICATION
The call cord shall be TekTone SF401G or approved
equal. Model SF401G shall include a single squeeze bulb
for call origination.
Switch:
Plug:
Cord:
Electrical:
Dimensions:
REQUIRED COMPONENTS
Tek-CARENC300II
IR319/A or IR320/A
www.tektone.com
IL886
Section E
Rev. 1 - 08/2010
FEATURES
The SF401EX Strain Relief Cable is used to increase the
life span of call cords, pillow speakers and room stations.
The cable is placed between the call cord or pillow speaker
and the room station. If the call cord or pillow speaker
is attached to a bed and then the bed is pulled away, the
strain relief cable disconnects. This prevents damage to
both the room station and the call cord or pillow speaker.
SF401EX
UL 1069 Listed
Molded connectors
Clean break-away action
Increases life of call cords, pillow speakers and room
stations
SPECIFICATIONS
Length:
Connections:
Construction:
REQUIRED COMPONENTS
for Tek-CARENC300II Nurse Call Systems:
IR319- or IR320-series Patient Stations
SF4-series
Call Cords or Pillow Speakers
for Tek-CARE400 Nurse Call Systems:
IR411- or IR412-series Patient Stations
SF4-series
Call Cords or Pillow Speakers
www.tektone.com
IL629
Section E
Rev. 11 - 03/2011
FEATURES
The SF337C Emergency Shower Switch is for any location
where activation of an emergency signal is required. The
design is such that it may be located anywhere in a shower,
bath or tub room, regardless of environmental conditions.
The combination of rubber-grommetted mounting screws,
rubber mounting gasket, and completely sealed front
faceplate make it waterproof (if properly installed).
Electronic components are mounted on a circuit board,
including an industry-proven reliable reed switch for circuit
initiation. The activating magnet is mounted on the large
pull-down lever, which is completely sealed in plastic.
The indicator LED is also completely sealed.
UL 1069 Listed
Waterproof if installed per manufacturers instructions
Call-placed indicator
Call cancel at point of origin
Attractive, flame-retardant (UL 94V-0) plastic faceplate
1.1" (28mm) wide pull-down lever
Washable TPU plastic pull cord kit is included
SPECIFICATIONS
Dimensions:
Height: 4.5" (114 mm)
Width:
2.8125" (72 mm)
Depth:
1.625" (41 mm)
Projects: 0.6875" (17 mm)
(including pull down lever)
Construction: Flame-retardant (UL 94V-0) plastic
Connections:
Plug-in type.
Housing:
Single gang ring with square box.
APPLICABLE SYSTEMS
Tek-CARE400 P5 Nurse Call System
Tek-CARENC300II Nurse Call System
Tek-CARE NC200, NC150, NC110 Nurse Call Systems
Tek-CARE500 Wireless Emergency Call System
www.tektone.com
IL744
Section E
Rev. 4 - 09/2010
FEATURES
The SF339 Emergency Switch includes a locking
mushroom-style switch, and is for use in any location
where activation of an emergency signal is required. All
electronic components are mounted on a circuit board,
including a reliable industry-proven DPDT switch for
circuit initiation. The DPDT switch and the indicator LED
are completely sealed in plastic.
The SF339 may be used as an emergency switch or as a
staff-assist emergency call switch. The large mushroomstyle button is easily activated, so that calls may be placed
using an elbow, shoulder, etc., when using the hands is
inconvenient or inappropriate. Calls are cancelled by
turning the mushroom-style button clockwise.
Dome light indication and system annunciation vary
according to the manner in which the SF339 is installed.
UL 1069 Listed
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame-retardant (UL 94V-0) plastic faceplate
0.9" diameter (23 mm) red mushroom-style button
SPECIFICATIONS
Dimensions:
Construction:
Connections:
Housing:
Height: 4.5"
(114.3 mm)
Width:
2.8125" (71.4 mm)
Depth:
1"
(25.4 mm)
Projects: 1"
(25.4 mm)
Flame-retardant (UL 94V-0) plastic
faceplate. DPDT switch and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square
back box (IH358) or exact equal.
APPLICABLE SYSTEMS
Tek-CARE400 P5 Nurse Call System
Tek-CARENC300II Nurse Call System
Tek-CARE NC200, NC150, NC110 Nurse Call Systems
Tek-CARE500 Wireless Emergency Call System
www.tektone.com
IL509
Section E
Rev. 16 - 03/2011
FEATURES
The SF340B Emergency Switch is for use in any location
where activation of an emergency signal is required. All
electronic components are mounted on a circuit board. The
specially mounted LED is completely sealed in plastic.
Due to the flexibility of its design, the SF340B may be used
as a pull cord emergency switch, a pull down activated
emergency switch, or a staff assist emergency call switch.
The pull cord supplied may be used for emergency call
activation or, if desired, the large 1.1" (28mm) red pulldown lever may be used.
Dome light and system annunciation vary according to
the manner in which the SF340B is connected.
UL 1069 Listed
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame-retardant (UL 94V-0) plastic faceplate
1.1" (28 mm) wide pull-down lever
Washable TPU plastic pull cord kit is included
SPECIFICATIONS
Dimensions:
Height: 4.5" (114 mm)
Width:
2.8125" (72 mm)
Depth:
1.625" (41 mm)
Projects: 0.6875" (17 mm)
from mounting surface
(including pull-down lever)
Construction: Flame-retardant (UL 94V-0) plastic
faceplate. Switch and electronic
components are circuit board mounted.
Connections:
Plug-in connectors.
Housing:
Single gang ring (IH357) with square
box (IH358), or exact equal.
APPLICABLE SYSTEMS
Tek-CARE400 P5
Tek-CARENC300II
www.tektone.com
IL510
Section E
Rev. 14 - 09/2010
FEATURES
The SF341B Code Call Switch operates all system code
signals, and is designed for use in ICU/CCU, physical
therapy rooms, post-op recovery, or any location where
code call initiation may be required. The large 1.1" (28mm)
blue pull-down lever is used for code call activation. The
red call-placed LED indicator flashes to provide for call
assurance. Calls are registered on the master station (and
duty station, if used) via audible and visual signals pulsing
at a rate of twice that of a standard emergency signal.
Local indication on dome lights is also provided. Call
cancellation is possible only at the originating location.
UL 1069 Listed
Supervised
Registers audible and visual calls at master station
Indicates local call origination on dome lights
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame retardant (UL 94V-0) plastic faceplate
1.1" (28 mm) wide pull-down lever
SPECIFICATIONS
Dimensions:
Height: 4.50" (114 mm)
Width:
2.8125" (72 mm)
Depth:
1.625" (41 mm)
Projects: 0.6875" (17 mm)
from mounting surface
(including pull down lever)
Construction: Flame retardant (UL 94V-0) plastic
faceplate. Switch and electronic
components are circuit board mounted.
Connections:
Plug-in connectors.
Housing:
Single gang ring (IH357) with square
box (IH358), or exact equal.
APPLICABLE SYSTEMS
Tek-CARE400 P5
Tek-CARENC300II
www.tektone.com
IL407
Section E
Rev. 10 - 09/2010
FEATURES
The SF350B Presence Station provides push button
registration of nurse presence (green LED indication), or
aide presence (yellow LED indication) at remote locations.
The nurse/aide presence indicator unit can also be used
to transfer nurse follower feature from room to room
as nursing staff proceeds with rounds. Depressing the
nurse presence registration button automatically transfers
indication of all subsequent patient calls to the remote
location of the nurse, as well as to the master station.
Plug-in connectors provide for simple wiring.
UL 1069 Listed
Push button and green LED for nurse presence
indication
Push button and yellow LED for aide presence
indication
Automatic nurse follower transfer
Flame retardant (UL 94V-0) plastic faceplate
SF350B
SPECIFICATIONS
Dimensions:
Height: 4.50" (114 mm)
Width:
2.75" (70 mm)
Depth:
1.50" (38 mm)
Projects: 0.4375" (11 mm)
from mounting surface
Construction:
Flame retardant (UL 94V-0) plastic
faceplate. Switches and electronic
components are circuit board mounted.
Connections:
Plug-in connectors.
Housing:
Single gang ring (IH357) with square
box (IH358), or exact equal.
APPLICABLE SYSTEMS
Tek-CARE400 P5
Tek-CARENC300II
www.tektone.com
IL558
Section E
JK300A, JK300/2
Monitor Call Interface
Specification Sheet
Rev. 9 - 09/2010
FEATURES
JK300A
Easy installation
Mounted on a flame-retardant (UL 94V-0) plastic
panel
Simple, reliable operation
SPECIFICATIONS
JK300/2
Dimensions:
Height: 4.5" (114 mm)
Width:
2.75" (70 mm)
Depth:
1.5"
(38 mm)
Projects: 0.25" (6.5 mm)
from mounting surface
Construction:
Reliable components mounted on a
flame-retardant (UL 94V-0) plastic
panel
Connections:
Wire lead pigtails
Housing:
Single-gang electrical box or ring
REQUIRED COMPONENTS
Tek-CARE400 P5
IR400P5-series
-orTek-CARE400
PM423
IR400-series
-orTek-CARENC300II
IR300-series
www.tektone.com
IL906
Section C
Rev. 0 - 09/2010
FEATURES
LI484P5
Corridor Light
SPECIFICATIONS
Dimensions:
Power:
Construction:
Connections:
Housing:
Height: 5.0"
(127 mm)
Width:
4.75"
(121 mm)
Depth:
2.75"
(70 mm)
5 VDC, 250 mA
Base snaps to mounting plate
Plug-in connections
Standard single or dual gang ring.
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
IR4xxP5-series stations
www.tektone.com
IL907
Section C
Rev. 0 - 09/2010
FEATURES
The PM484ZP5 Zone Light Module, when added to
an LI484P5 Corridor Light, provides auxiliary visual
annunciation for call activity within the zones and call types
to which it is assigned. Patient priority and staff presence
are indicated by four programmable LEDs.
SPECIFICATIONS
Power:
Connections:
REQUIRED COMPONENTS
Tek-CARE400 Nurse Call System
LI484P5 Four-LED Corridor Light
www.tektone.com
IL928
Section E
CT400, CT400P5
Station Commons Bus Splitters
Specification Sheet
Rev. 0 - 09/2011
FEATURES
The CT400 and CT400P5 station commons bus splitters
are used to provide room drops (similar to T-Tapping) from
the station commons bus to addressable stations and zone
lamps. Typically, they are used when a station commons
bus is run down a hallway from corridor dome lamp to
corridor dome lamp, or centrally down the hallway. Each
splitter provides six jacksan in and an out, plus up to
4 room drops.
CT400
CT400P5
SPECIFICATIONS
Dimensions:
Connections:
REQUIRED COMPONENTS
Tek-CARE400
for CT400:
PM453
for CT400P5:
PM455
www.tektone.com
IL670
Section E
Rev. 9 - 08/2010
FEATURES
The NC369 UHF Paging Transmitter receives alphanumeric
messages and patient call information via serial interface
and transmits the information directly to pagers.
Information is sent to the transmitter at 9600 baud
to minimize bottlenecks. Call type and location are
determined by the system supplying information to the
transmitter. Up to 1000 pager IDs are supported. Battery
backup is also available. Programming is simple using
a PC or laptop and web based programming screens.
Dipswitches are also available on the unit for minimal
function onsite setup.
The systems range depends upon the site and the transmitters
location. To determine the exact coverage, conduct a site
survey. Estimate a range of up to one mile under ideal
conditions. The unit includes a -wave BNC-terminated
antenna to provide coverage that meets most requirements.
The optional CA133 Antenna is suited for installations that
require good building penetration and range. On average,
the CA133 Antenna adds about 25% to the transmission
footprint, and because it comes with approximately 10 feet
of cable, it can be mounted in more locations.
NC369
SPECIFICATIONS
Dimensions:
Construction:
REQUIRED COMPONENTS
NC399P Pagers
Tek-CARE400 Nurse Call System
-or- Tek-CARENC300II Nurse Call System
-or- Tek-CARENC110/150/200 Nurse Call System
(requires NC375 Transmitter Interface)
-or- Tek-CARE500 Wireless Emergency Call System
www.tektone.com
IL734
Section E
Rev. 6 - 08/2010
FEATURES
The NC399P alphanumeric pager is a radio pager receiver
with a large 2-line display, and is designed for use with
TekTones Tek-CARE500 Wireless Emergency Call
System, and TekTones Tek-CARE400, NC300II,
NC200, NC150 and NC110 Nurse Call Systems. The pager
stores up to 60 messages. Messages may be selectively
deleted or locked into memory. Messages not locked into
memory are automatically deleted as new messages arrive
(beyond the 60 message limit), with the oldest messages
being deleted first. The NC399P may be configured for
annunciation via vibration or audible tones, and has one
alarm for daily annunciation. It can be programmed to
turn on and off at preset times each day.
NC399P
Alphanumeric
2-line backlit display
Easy to use large 4-button design
Configurable for audible or vibration alerts
Field programmable
Includes holster belt clip and safety cord
SPECIFICATIONS
Dimensions:
Height: 1.9" (49 mm)
without holster belt clip
Width:
2.7"
(70 mm)
Depth:
0.7"
(19 mm)
Construction:
Electronic components surface mounted
to PCB enclosed in ABS plastic housing
Operating Freq.: 457.550 Mhz
Power supply: 1.5 vDC (1 AA cell)
APPLICABLE SYSTEMS
Tek-CARE500 Wireless Emergency Call System
Tek-CARE400, NC300II, NC200, NC150, NC110
Nurse call systems
NC365/A/AT
Transmitter/Control Unit
NC366A/AT
Transmitter/Control Unit
NC375
Transmitter Interface
www.tektone.com
IL890
Section E
Rev. 2 - 08/2010
FEATURES
The SI005 LED Message Sign is a radio paging receiver
with a large one-line alphanumeric display. It is designed for
use with TekTones Tek-CARE500 Wireless Emergency
Call and Tek-CARE400 Nurse Call Systems. The SI005
stores up to 24 messages. The messages are deleted after a
predetermined length of time. Messages are automatically
deleted as new messages arrive (beyond the 24 message
limit), with the oldest messages being deleted first. The
SI005 may be configured for annunciation via audible
tones. The SI005 has been pre-programmed to receive
call information from TekTone systems, and it can be
programmed for up to six CAP codes.
SI005
SPECIFICATIONS
ARCHITECTS AND ENGINEERS
SPECIFICATION
The alphanumeric LED message sign shall be TekTone
SI005. It shall have a large single-line display and shall be
capable of storing up to 24 messages. The unit shall provide
a choice of tone modes for page annunciation. The LED
message sign shall be pre-programmed to receive patient
call information from TekTones Tek-CARE500 Wireless
Emergency Call System and Tek-CARE400 Nurse Call
System via TekTones NC369 Paging Transmitter. The
LED message sign shall be programmable for up to six CAP
codes. The sign shall annunciate and display alphanumeric
patient call information as generated by the emergency
or nurse call system.
Dimensions:
Weight:
Construction:
APPLICABLE SYSTEMS
Tek-CARE400 Nurse Call System
Tek-CARE500 Wireless Emergency Call System
REQUIRED COMPONENT
NC369 Paging Transmitter
The LED message sign is an ancillary method of call annunciation,
and is not intended to replace the primary annunciation provided by
the nurse call system.