Sunteți pe pagina 1din 17

C H A P T E R

1
Milestones in the Development
of Gas Chromatography
Walter G. Jennings, Colin F. Poole

O U T L I N E

1.1. Introduction 2 1.5. Interfacing Glass Capillary Columns


to Injectors and Detectors 7
1.2. The Invention of Gas Chromatography 2
1.6. The Hindelang Conferences and the
1.3. Early Instrumentation 2
Fused-Silica Column 8
1.3.1. Early Commercial Instruments 3
1.7. Increasing Sophistication of
1.4. Early Column Developments 4
Instrumentation 10
1.4.1. Do-it-Yourself Glass Capillary
1.7.1. Column Heating 11
Columns 4
1.7.2. Sample Introduction 12
1.4.2. The Positive Results of Patent
1.7.3. Detectors 13
Enforcement 5
1.7.4. Data Handling 14
1.4.3. Mileposts in Coating WCOT
1.7.5. Comprehensive Gas
Capillary Columns 5
Chromatography 15
1.4.4. Commercial Column Manufacturers 6
1.4.5. Bonded, Crosslinked, and/or 1.8. Decline in the Expertise of the Average
Immobilized Stationary Phases 6 Gas Chromatographer 16
1.4.6. Further Improvements
in Stationary Phases 7

Gas Chromatography DOI: 10.1016/B978-0-12-385540-4.00001-8 1 Copyright 2012 Elsevier Inc. All rights reserved.
2 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

1.1. INTRODUCTION was first conceived by Erika Cremer, an


Austrian scientist at the University of Inns-
This article was started by the senior author bruck, Austria, in the late 1940s during the
Walter Jennings who was unable to complete it period of the Second World War. As Ettre points
due to poor health. Sections 1.2e1.6 are out, this was a period when women, especially
a personal account of the early days of gas chro- in Germanic countries, were expected to confine
matography seen through the eyes of one of the their activities to children, church, and
major pioneers and innovators in this field. kitchen. In spite of her superb Ph.D. thesis
With only minor editorial changes made by work she had great difficulty in finding a posi-
the junior author, these are presented as Walter tion. Her opportunity came in 1940 when the
intended. As the junior author I am responsible war started, and university teachers were
for Sections 1.7 and 1.8. These sections extend drafted. She obtained an academic position at
Walters comments on the early days of gas the University of Innsbruck, Austria (then
chromatography to the present day. a part of Germany) in the Institute of Physical
Chemistry. It was here that she and her students
(with major credit to Fritz Prior) constructed the
1.2. THE INVENTION OF GAS first prototype of a gas chromatograph
CHROMATOGRAPHY (Figure 1.1) and, after a long delay that was
probably attributable to the war, published the
In 1952 A.J.P. Martin and R.L.M. Synge were results of their research in 1951 [3,4]. At this
both awarded Nobel Prizes for their work in the time she was promoted to Professor and, some
field of liquid/solid chromatography. Martin, in twenty years later, to director of the Universitys
his award address, suggested it might be possible Institute of Physical Chemistry. Professor Dr.
to use a vapor as the mobile phase. Some years Cremer, by all accounts a brilliant woman scien-
later, James and Martin used ethyl acetate vapor tist, died in 1996.
to desorb a mixture of fatty acids that had been
affixed to an adsorbent, and placed in a tube.
The vapor stream eluting from that tube was 1.3. EARLY INSTRUMENTATION
directed to an automated titration apparatus,
resulting in a graph showing a series of steps By 1953, several petroleum companies,
that reflected the sequential additions of base as primarily in Great Britain and the Netherlands,
each eluted acid was neutralized by automated were exploring this new analytical technique,
titration [1]. Many practitioners have for far too and in 1954, a few flavor chemists (including
long considered this as the starting point of gas this author) were building crude chromato-
chromatography. graphs, many of them based on an article by
In 2008 Leslie Ettre published an article in N. H. Ray [5]. Ray inserted thermal conduc-
which he stated, . the activities of Professor tivity cells into a Wheatstone bridge, whose
Erika Cremer and her students at the University outlet connected to a strip chart recorder, thus
of Innsbruck, Austria, in the years following the generating a Gaussian peak for each eluting
Second World War, represented the true start of solute; this was (to my knowledge) the first
their continuous involvement in gas chromatog- gas chromatogram as we know them today.
raphy [2]. After exploring Ettres arguments The schematics and chromatograms published
and conducting some research myself, I fully by Ray encouraged a number of readers
agree with his conclusions. I now believe that (including the author) to build similar chro-
the theoretical basis for gas chromatography matographs. Almost every part of these crude
1.3. EARLY INSTRUMENTATION 3

FIGURE 1.1 The gas chromatographic system used in Prior work, in 1945e1947. A adsorbent for purification of the
carrier gas (hydrogen); B sample inlet system; C buret containing mercury with niveau glass for sample introduction;
D Dewar flask; E separation column (containing silica gel on activated carbon); and F thermal conductivity detector.
Source: From ref. [4] copyright Friedr. Vieweg & Sohn.

instruments had to be self-designed and self- and George (London) and Metropolitan
made, but this introduced the days of the Vickers Electric (Manchester), but the U.S.
packed column; supports, such as granules of instrument companies had a greater impact
diatomaceous earth, were coated with a variety on the development of this new area [7]. Per-
of high boiling fluids (e.g. diethylene glycol kin Elmer was one of the first companies to
succinate, DEGS) and, in our earliest efforts, market a gas chromatograph; in May of
packed into copper columns, typically 3e6 m 1954, they introduced their Model 154 Vapor
long with internal diameters of about 6 mm, Fractometer. The temperature of the column
and (usually) coiled. It was soon recognized oven was adjustable from room temperature
that copper columns were quite active and to 150  C, and it offered a flash vaporizer
most workers switched, first to stainless steel, with a rubber septum permitting syringe
and then to coiled glass tubing of similar injection into the carrier gas stream. The
dimensions. The reminiscences of many of the detector was a thermal conductivity cell. The
early pioneers in gas chromatographic instru- instrument was a great success and sold
mentation are summarized in [6]. widely [8]. In early 1956, Perkin Elmer fol-
lowed up with their Model 154-B, with
a temperature range from room temperature
1.3.1. Early Commercial Instruments to 225  C and could be fitted with an optional
The first companies to manufacture gas rotary valve offering a variety of sample loops
chromatographs (GCs) in Europe were Griffin for the injection of gas samples.
4 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

1.4. EARLY COLUMN granted patents on the open-tubular concept,


DEVELOPMENTS essentially worldwide. I purchased two Perkin
Elmer wall-coated open-tubular (WCOT)
All of the early instruments utilized packed columns at different times, only to find that
columns, some coiled, some U-shaped. Packed the columns available from them produced
columns all have one thing in common: they abominable results. Perkin Elmer apparently
possess a high resistance to gas flow, and this recognized this fact, because they essentially
limits the practical length of the column, usually abandoned their research efforts on WCOT
to a few meters. In much later days, some columns, outsourced their production, and re-
packed columns approached an efficiency directed their efforts to columns whose interiors
equivalent to our current 0.32 mm internal were first coated with a support material (e.g.
diameter capillary columns (ca. 3000 plates per diatomaceous earth) and then with the
meter), but because of their length limitations, stationary phase. These were dubbed support
they could achieve perhaps 35,000 theoretical coated open-tubular (SCOT) columns. Perkin
plates while a 30 m open-tubular 0.25 mm Elmer continued to rigorously enforce the
internal diameter column should be capable of patent on WCOT (and SCOT) columns, but
three times that efficiency, merely because it is under considerable pressure (especially from
three times longer. the applicant) they were forced to issue one
It was at the 1958 Second International license, to Hansjurg Jaeggi, a former assistant
Symposium on Gas Chromatography in of Kurt Grob in Switzerland. Under Swiss law,
Amsterdam that Marcel Golay, who was then if a patent bars a Swiss from conducting his or
a consultant to Perkin Elmer, introduced the her business, a license must be issued. Jaeggi
theory of open-tubular capillary columns and had been and still was making excellent glass
demonstrated their superiority to packed WCOT columns. The high quality of his
columns [9]. (There is no connection between the columns led Perkin Elmer to later propose that
old Perkin Elmer referred to here and the Perkin they collaborate, but, probably because of the
Elmer of today, they are entirely different compa- acrimonious battle he had gone through to
nies.) These columns demanded much smaller obtain his license, he wanted nothing more to
sample injections, necessitating more sensitive do with Perkin Elmer, and refused their offer.
detectors than the thermal conductivity detec- His obituary, written by Konrad Grob, makes
tors that were in common use at the time. Fortu- interesting reading [15].
nately, James E. Lovelock had invented an
electron-capture detector in 1957 [10], and in
1.4.1. Do-it-Yourself Glass Capillary
1958 the flame ionization detector appeared;
some credit this to Harley, Nel, and Pretorious
Columns
[11], and others to McWilliams and Dewer In 1965, Desty et al. invented an elegantly
[12,13]; these increased detector sensitivities by simple machine for drawing long lengths of
ca. 103e106. The invention and development of coiled glass capillaries [16]. Besides the fact
the flame ionization detector is discussed in that his was much less expensive tubing, it
more detail in [14]. Early capillary columns also had a smoother interior and it was trans-
were of plastic and copper tubing; the former parent. For the first time, it was possible to scru-
had serious temperature limitations and the tinize the layer of stationary phase as it existed
latter was active; this led to stainless steel on the column wall; soon it was obvious that
tubing. Perkin Elmer had filed for and been when a new unused column exhibiting a thin
1.4. EARLY COLUMN DEVELOPMENTS 5
uniform film of stationary phase was exposed to column could be coiled only after coating,
higher temperatures, the stationary phase a distinct drawback. The method was improved
collected into beads that were randomly scat- by Ilkova and Mistrykov [19], who coiled, and
tered over the inner surface. Almost immedi- then filled the column and sealed one end. The
ately scores of scientists realized that many of open end of the filled column was fed into
the low viscosity fluids that worked well in a heated oven by supporting the column on
packed columns were unsatisfactory for a rotating rod fitted with a drive roller at the
WCOT columns and should be replaced with entrance to the column oven, thus literally
high viscosity materials that would retain their screwing the rest of the column into the oven.
high viscosity even at higher temperatures. Up until just a few years ago, all chemistry
Low-viscosity silicone fluids (e.g. OV 101) departments at the multicampus Universities
were replaced with high viscosity silicones of California system frowned on applied
(e.g. OV 30, a viscous paste-like silicone), and research, and chemistry department faculties
experimenters soon began producing much drifting away from pure organic chemistry or
more stable columns. pure physical chemistry rarely survived to
tenure. Analytical chemistry was essentially
1.4.2. The Positive Results of Patent forbidden. There were many faculty members
scattered through various other departments
Enforcement
who were analytical chemists, and eventually
The low quality of the Perkin Elmer WCOT they formed a Group in Analytical and Envi-
columns and enforcement of their patent led ronmental Chemistry, open to anyone in any
many scientists to begin making their own department who had interests in the analytical
columns for their own use. This caused investi- side. At one time I chaired that group for
gators in many other fields, who would have several years and it still exists. At this time,
preferred to purchase usable columns from an my title was Professor of Food Science and
outside source and confine their research to Technology and Chemist in the Experiment
some other field, to now have to make columns; Station. This had several advantages: for one
scores of scientists were now studying and thing, those with just the academic title worked
publishing on methods of pretreating, deacti- nine months per year, while the Experiment
vating, and coating columns. Their combined Station operated eleven months a year. My
results were responsible for many of the department chairman tolerated what he called
advances in column improvements, and they my dabbling in gas chromatography, but
soon surpassed the results that had been gener- insisted that I should also be working on
ated by Perkin Elmer [17]. subjects more aligned with the food indus-
tries; he suggested circulation cleaning
1.4.3. Mileposts in Coating WCOT which was just emerging. Swallowing what I
wanted to say, I assigned a new Ph.D. student
Capillary Columns
(Malcolm Bourne) to the project, built a minia-
Golay had coated his original glass open- ture circulation pipeline, and helped him bake
tubular columns by completely filling them radio-labeled tristearin onto glass microscope
with a dilute solution of stationary phase in slides and strips of stainless steel of similar
a low-boiling solvent, sealing one end, and width and length. These were inserted into
then drawing the column, open end first, the rapid circulating system, and the levels of
through an oven [18]. As used by Golay, the radioactivity measured every two minutes.
6 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

Inevitably, semi-log plots of our cleaning 1.4.4. Commercial Column


curves showed a rapidly dropping curve that Manufacturers
terminated in a straight line with a slight
downward slope. All at once Bourne realized At this point, one of my Ph.D. students,
that he was looking at two first-order reactions Robert Wohleb, suggested that we had a market-
that were occurring simultaneously. The baked able product and should establish a column
film of tristearin existed only at one of two production company. When I broached this
energy levels, a loosely bound system, and suggestion with my department chairman, he
a much more tightly bound system e nothing called in the dean. After much discussion, they
in between. Spatial estimates indicated that agreed that I could engage in this activity only
the tightly bound form was not a monolayer, if I agreed to several restrictions: (1) as long as
but could be up to at least twenty molecular I was a full time professor, I could receive no
thicknesses [20e22]. He also discovered that remuneration from this venture; (2) there could
by manipulating time and temperature, he be no connection between my university
could change the ratio of the two species. research and activities of the company; (3) I
Bourne went on to study the kinetics of clean- could not get involved in the day-to-day activi-
ing in much greater depth, and made quite ties of the company. I could, on my own time,
a name for himself. I went back to my answer trouble-shooting questions and engage
dabbling, then slowly realized that my newly in educational activities. J&W Scientific was
gained knowledge on removal of thin films was founded in 1974; I constructed two drawing
just the reverse of achieving more stable coat- machines in my machine shop at home, and
ings of stationary phases to the column surface. left them with Wohleb as I departed on my
This may have guided me as I attempted to second sabbatical leave, this time in Karlsruhe,
modify our column coating apparatus. After Germany. Sandy Lipsky had launched his
replacing the opaque lid of the oven with a glass company, Quadrex, slightly ahead of J&W.
lid, it was obvious that the solution was super-
heated and evaporated in a series of minor 1.4.5. Bonded, Crosslinked, and/or
explosions, leaving blotches of stationary phase
Immobilized Stationary Phases
randomly scattered over the surface of the
column. I decided to introduce the column In 1975, J&W noticed a sudden drop in
into the oven through a preheater made from column sales from several regular customers;
a short curved length of 1/8 inch stainless steel fortunately, we had been using (and saving)
tubing lagged with an electrically heated wire bar codes to trace every step each column
and heated to 150  C: the columns passage went through. The bar codes on all of the
through the curved preheater was not smooth; columns those customers had been buying
indeed, it scraped the walls and vibrated. To showed they had all gone through the same
avoid breakage, I introduced graphite powder coating machine in the same general time
into the loop. The column continued to vibrate, period. I called several of the buyers and asked
but more gently. In retrospect, that vibration at if they were having problems with our
this point on its passage into the oven would columns, and was shocked when they replied,
also discourage super-heating. It may have no, the columns are great e they never wear
been dumb luck, but from then on, the evapora- out. This was serious, we were putting
tion of the solvent occurred smoothly. We were ourselves out of business. In checking that
routinely producing very stable columns with coating machine, it was discovered that the
uniformly thin coatings. thermocouple on the preheater through which
1.5. INTERFACING GLASS CAPILLARY COLUMNS TO INJECTORS AND DETECTORS 7
the column entered the oven had failed some succinate (DEGS). As more and more scientists
time ago and was giving much lower readings. entered the field, the range of stationary phases
Over the past few weeks, the technician on the exploded to over 200 different phases, most of
machine had simply compensated for the low which were gradually discarded. The polysilox-
reading by boosting the voltage. Instead of ane stationary phases are now widely used and
150  C, that preheater was close to 300  C. The have been reviewed by Blomberg [23] and
stationary phase we were using (SE-54) con- Haken [24].
tained 1% vinyl. Our R&D head, Rand Jenkins,
realized that serendipity had smiled on us! He
immediately ordered materials and began add-
ing vinyl-containing compounds, to our 1.5. INTERFACING GLASS
coating solutions, hoping that the phase would CAPILLARY COLUMNS TO
not only crosslink but also connect with some INJECTORS AND DETECTORS
surface hydroxyls. With some pretty harsh
testing, we found that the columns exposed to Attaching these relatively fragile glass capil-
these higher preheater temperatures were laries to injectors and detectors via leak-proof
much more rugged, and the deposited phase connections soon became a problem; some
could not be removed with solvent rinsing. At workers punched holes in small disks of silicone
the next Advances in Chromatography rubber which were then substituted for the
meeting in Houston, we announced the ferrules supplied with 1/16 inch Swagelock
Bonded Phase Column. If the technician on fittings, but all too frequently, as the nut on the
that machine had reported the failing thermo- Swagelock fitting was tightened to the point
couple, we would simply have replaced it and that there were no leaks, the compressive strain
none of this would have happened. It turned on the glass caused breakage at that point.
out that Kurt Grob in Switzerland had (inde- Others tried lead disks, with similar effects.
pendently) been working along these same I was on sabbatical leave at the nuclear research
lines. Checking on the dates that both parties institute in Karlsruhe, Germany, when a clerk in
had ordered vinyl chemicals, Rand was just the supply room called my attention to a sheet of
about 5 days ahead. plastic heavily infused with graphite. After
sliding Swagelock caps onto both column ends,
I wound small strips of this around both ends
1.4.6. Further Improvements
of a glass capillary, removed the stock ferules
in Stationary Phases from the Swagelock connectors to the inlet and
In the late 50s and extending into the 60s, gas detector, and connected the column. As I tight-
chromatographers dealt with a plethora of ened the connections to a point where there
stationary phases. Petroleum chemists, dealing was no leakage and found that the column was
with nonpolar products, favored phases such still intact, I was delighted. The company that
as squalane, a fully saturated hydrocarbon made these graphitized sheets was just a few
(C30H62) on the basis of like-attracts-like. The miles outside of Karlsruhe, and I paid them
upper temperature limit for Squalane is 125 , a visit. They were very cordial and gave me
and a higher temperature paraffin, Apolane 87 a number of samples. On my return to the U.S.,
(C87H176), was sometimes used instead. Chem- J&W ordered several steel dies dimensioned on
ists dealing with more polar products needed the inner measurements of a 1/16 inch Swage-
more polar stationary phases, and they turned lock fitting and began pressing ferrules. These
to high boiling esters such as diethylene glycol went on the market in 1975; they were a great
8 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

success, and I have always regretted my failure I explained that the first step was his idea of
to patent that product. feeding a filled column, open end first, into
a heated oven. Ilkova and Mistrykovs idea of
screwing the coiled column into the oven
1.6. THE HINDELANG showed real ingenuity: my preheater contribu-
CONFERENCES AND THE tion would have been proposed by anyone
FUSED-SILICA COLUMN that could actually see the operation taking
place. Now that I have reached (and probably
The first symposium restricted to open- surpassed) Golays age at the time, I believe I
tubular (capillary) columns was organized by can now understand his curiosity, and I hope
Rudolf Kaiser and chaired by Dennis Desty in he realized that another of his brilliant ideas
1975; this was the first of the four Hindelang had borne fruit. He departed courteously, and
conferences and was held in an isolated village I was left to continue my attempts to visit every
in the Bavarian Alps. Ettre visited Kaiser, first table and meet every participant. I did meet
to voice his opposition to holding the meeting, virtually all the experts, absorbed a great deal
and then proposed a delay, arguing that it was of knowledge, and left filled with new ideas
being held far too early, but many other and a desire to get home quickly where I could
workers in the field were enthused by the get back to work in my own lab.
idea. I still regard the first Hindelang Confer- There are many types of glass, but we will
ence as the most exciting meeting I have ever restrict our interests to those studied for their
attended. It attracted nearly everybody that chromatographic interest by Dandeneau and
was working with glass capillary columns. We Zerenner: soda lime, borosilicate, uranium,
listened to presentations from our associates in potash soda lead, and fused silica (Table 1.1)
the morning, and spent afternoons and evenings [25]. At the third Hindelang conference, I pre-
gathered around tables seating perhaps eight sented a paper on multiple short pass capillary
participants, drinking strong German brews chromatography, in which one of my Ph.D.
and exchanging views on chromatography, students (James A. Settlage) had recycled
with the emphasis on WCOT columns. It was a butane injection sixteen times through two
then that Golay approached me; he was feeble 7.5 m columns to achieve 1170 theoretical plates
at the time and steadied by Ettre. He wanted per second, and generated 850,000 theoretical
to talk about J&Ws method of coating columns. plates in those sixteen passes in less than twelve

TABLE 1.1 Approximate Compositions for Glass Materials Used for Capillary Column Fabrication

Metal oxide percent composition (w/w)

Glass Type SiO2 Al2O3 Na2O K2O CaO MgO B2O3 PbO U2O3 BaO

Soda lime 68 3 15 6 4 2 2
Borosilicate 81 2 4 13
Uranium glass 76 3 4 2 14 1

Potash soda lead 56 2 4 9 29


Fused silica* 100

* Typically contains less than 1 ppm total metals.


1.6. THE HINDELANG CONFERENCES AND THE FUSED-SILICA COLUMN 9
minutes [26]; as I finished, several members a polyimide; this seals surface flaws in the
of the audience rushed to the podium with tubing. It is usually applied in several thin coats
congratulations; at a final synopsis of the during the drawing process.
meeting, Desty cited also this paper, but I Not everyone recognized the advantages of
knew they were all wrong. The most significant the fused-silica column; indeed, in spite of my
paper was that of Dandeneau and Zerenner vigorous protests, my partners at J&W dismissed
[25], releasing the fused-silica column. When it as a gimmick. Two weeks later, it was impos-
Dandeneau presented his paper, very few sible to give, let alone sell, a glass capillary
people realized its importance. Sandy Lipsky, column, and J&W made the conversion to fused
who had started Quadrex about the same time silica. At this point, things became very
that J&W was founded, was one, and I was touchy. The vast majority of scientists working
another. Both Lipsky and I had been trying to in gas chromatography had careers based on
convert industrial analysts to WCOT glass capil- glass capillary columns e drawing, deactivation,
laries, with but scant success. Industrial analysts coating, etc. e now those careers were passe.
realized the superiority of results generated by I had a seminar scheduled for Milan, Italy, where
these columns, but they also recognized their I was warmly received e until my first slide was
fragility. Downtime is rarely critical in shown. It was a fused-silica column. Several
academia, but in industry it can be very serious; attendees left the room, and the atmosphere
glass capillaries break easily, packed columns took on a chill. Very few in this audience wanted
lasted longer. However, this fused-silica column to hear how their futures might be affected.
changed everything! Here was a capillary Shortly thereafter, Georges Guiochon organized
column that was both strong and flexible, but a chromatography Symposium in Cannes,
because it was both labor intensive and mate- France. The GC column portion of the meeting
rials intensive, it was much more expensive. was chaired by Gerhardt Schomburg. As he
Fused-silica columns have immense tensile called the meeting to order, he announced that
strength, permitting an extremely thin column there would be no discussion of fused-silica
wall, which in turn makes a flexible column. columns. As the meeting proceeded, several
This permitted drawing long lengths of straight attendees expressed their objections to the
tubing, which can then be coiled. Interfacing opening restriction and announced that they
these to detectors and injectors was now had come to the meeting just to learn more about
a simple task; no longer must we flame the fused-silica column. With obvious distaste,
straighten the ends of the column, it was already Schomburg turned to me and said, tell them
inherently straight! But the outer surface of something about fused silica. At this time, I
drawn silica requires protection. Like all silica- had been exploring on-column injections with
based glasses, fused silica is subject to flaws at fused silica, sometimes trapping samples within
its surface, and flaws grow at a rate determined the fused-silica tubing by folding the flexible
by stress e and coiling this long straight piece of tube into a U shape in a Dewar flask filled
fused silica places it under stress e as the flaw with liquid nitrogen, then withdrawing the U
grows larger, the column snaps at that point. and re-inserting it into a flask of heated oil for
To protect the outer surface, Hewlett Packard injection. This approach demanded a flexible
first coated the columns with a coating of poly- tube and would have been impossible with glass
siloxane, but this ended up as a very sticky capillary tubing. An obviously uncomfortable
column. DuPont then came forward with a poly- Schomburg interrupted me with the observation
amide coating that was applied during the that flexibility appeals only to Americans, and
drawing process. Later they changed to only because they are too clumsy to handle
10 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

glass. A few months later J&W was selling TABLE 1.2 Important Advances in Technology that
Impacted Gas Chromatography (contd)
Schomburg fused-silica columns.
1980e1990 Period of technical advances
Gum and immobilized phases developed
1.7. INCREASING
SOPHISTICATION OF Columns with thick-film stationary
phases developed
INSTRUMENTATION
Wide-bore open-tubular columns
Table 1.2 provides a timeline for important developed
developments in instrumentation. The essen- Fundamental basis for injection
tial elements of instruments were developed mechanisms developed
On-column and PTV injection developed
TABLE 1.2 Important Advances in Technology that
Impacted Gas Chromatography Large-volume injection described
Autosamplers for unattended injection
1955 First commercial instrument (thermal
developed
conductivity detection)
1990e2000 Period of consolidation and refinement
1955e1960 Rapid growth in technology
Keyboard instrumentation (PC control of
Invention of ionization detectors
operation and data handling)
Interfaces for coupling to mass
Electronic pneumatic control
spectrometry
Selectable elemental detection (SED)
Microsyringes produced commercially
Pulsed flame photometric detector
Temperature programming introduced
developed (FPD)
1960e1970 Period of technical advances
Electron-capture detector with pulsed
First open-tubular columns introduced discharge source developed (ECD)

Transistors replace vacuum tubes More versatile and affordable


spectroscopic detectors (MS, FTIR)
Improvements in detector technology
Solid-phase microextraction affords a new
Stable rubidium sources introduced for approach to sampling and sample
TID introduction
Improved FPD (several designs) First microfabricated gas chromatograph
Pulsed ECD introduced on a chip introduced but fails
commercially
1970e1980 Period of consolidation and refinement
Field portable instruments move gas
Microprocessor-based instruments chromatography out of the laboratory
introduced
2000e Slowing down in the growth of technology
Self-made glass open-tubular columns present
increasingly used Micro-GC instruments for fast gas
chromatography developed
Computing integrators introduced for
data handling Capillary columns with ionic liquid
stationary phases introduced
Fused-silica open-tubular columns
introduced in 1979 and catalyzes further Gas generators start to replace
changes in column and instrument pressurized gas cylinders
technology Programmable robotic systems used to
(Continued) automate sample preparation
1.7. INCREASING SOPHISTICATION OF INSTRUMENTATION 11
by the early 1960s [8], with further develop- accompanying the introduction of fused-silica
ments occurring in short bursts of innovation open-tubular columns meant significant
and advances in technology followed by longer changes were needed as capillary gas chroma-
periods of evolutionary changes and consolida- tography quickly became normal practice.
tion. Many advances were catalyzed by Apart from the use of liquid thermostats in the
advances in column technology or electronics. early days of gas chromatography, the column
For example, the advent of the microprocessor heater in a gas chromatograph is typically
brought about a radical change in instrument a forced-circulation air oven, the temperature
design and use. From this point onward, of which can be changed in a controlled manner
instrument functions were monitored and with time for temperature programmed separa-
controlled by networks of circuits communi- tions. Good temperature control is essential to
cating with each other and with a central obtain reproducible retention times and to avoid
controller. This paved the way for the emer- peak distortions associated with temporal and
gence of the keyboard instrument controlled spatial oven temperature gradients. A low
by software running on a personal computer, thermal mass for the oven is also important
which dominated instruments for laboratory since it allows rapid cooling after temperature
use by the early 1990s. A significant milestone programming. Typically, forced air-circulation
in achieving full automation of instrument ovens can maintain a higher rate of heating at
operation was the introduction of electronic lower temperatures than higher temperatures
pressure control in the early 1980s. This due to the greater amount of heat lost to the
allowed carrier gas and support gas pressure surrounding environment as the temperature
and flow rates to be set and monitored by elec- ramp progresses. For fast gas chromatography,
tromechanical devices communicating with some manufacturers have addressed this
a central processor. With the introduction of problem with the oven inside an oven concept
robotic autosamplers at about the same time, or more directly using resistive heating
the gas chromatograph could now operate achieved through the use of a metallic heating
without human intervention, 24 h operation element to transfer heat to the column by
became standard practice for routine analysis conduction. Fast gas chromatography makes
in high sample throughput environments, and use of short columns of small internal diameter,
gas chromatographs were deployed to remote thin film columns, higher carrier gas flow rates,
locations and monitored electronically with and fast temperature program rates (Table 1.3)
only occasional visits for service and routine [28]. For the fastest separations the limiting
maintenance. The complex functions of gas instrument conditions become the available
chromatography had been reduced to those of column inlet pressure, maximum oven tempera-
a black box analyzer 50 years after its invention, ture program rate, maximum detector sampling
but still evolutionary changes continue in tech- rate, detector sensitivity and noise level, and
nology with the view of minimizing the impor- sample introduction (initial band width).
tance of the skill and knowledge of the operator Typical laboratory instruments can usually
in the production of data [27]. meet the requirements for fast gas chromatog-
raphy (first entry in Table 1.3), but special-
purpose instruments are required for very fast
1.7.1. Column Heating
gas chromatography (second entry in Table
The increasing interest in capillary columns 1.3). The entries in Table 1.3 reflect what is
demonstrated the inadequacies of packed within the capability of commercial instruments
column instruments, and the paradigm shift with little modification (first entry) and the
12 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

TABLE 1.3 Instrumental Parameters for Fast Gas quantitative analysis of wide boiling point
Chromatography mixtures is difficult, and for samples present
(i) Fast gas chromatography (separation times <10 min) in dilute solution, detectability is limited by
the small amount of sample transferred to the
Peak widths 0.5e2 s
column. The splitless injection technique was
Columns 5e15 m with I.D. 0.25e0.1 mm devised (actually discovered accidentally [29])
Temperature program rates 20e60  C/min to overcome some of the deficiencies of split
injection for the analysis of mixtures in dilute
Column inlet pressure <15 bar
solution through the transfer of relatively large
Data acquisition rate 50 Hz sample volumes to the column. The gas flow
Mobile phase velocity 1.4  uopt through a splitless injector is relatively low,
and the sample is introduced into the column
(ii) Very fast gas chromatography (separation time in over a comparatively long time (30e60 s),
seconds) relying on cold trapping and/or solvent effects
Peak widths 50e200 ms to refocus the compounds at the head of the
column. The importance of these refocusing
Columns 2e10 m and I.D. 0.1e0.05 mm
mechanisms was not fully understood at first
Temperature program rates >60  C/min and much of our current knowledge of the injec-
Data acquisition rates 50e200 Hz tion mechanism owes a great deal to the exem-
plary studies of Konrad Grob and the
publication in 1986 of his classic book on split
and splitless injection [29].
The programmed-temperature vaporizer
limits of current technology requiring special (PTV) injector is perhaps the most versatile
purposing of instruments (second entry). From injector developed for gas chromatography. It
the perspective of separation speed, the capa- facilitates split and splitless injection as well as
bility of gas chromatographic instruments limits new approaches for large-volume injection
performance more so than column technology with solvent elimination [30]. The PTV injector
in contemporary practice. has a low thermal mass to allow rapid heating
and cooling. The sample is introduced at a rela-
tively low temperature and then raised ballisti-
1.7.2. Sample Introduction
cally to a temperature sufficient for rapid
The limited sample capacity and low carrier volatilization of the highest boiling sample
gas flow rates associated with open-tubular components. Slow sample introduction at
columns make sample introduction more diffi- a low temperature with solvent elimination
cult than for packed columns. A thermostated facilitates the injection of large sample volumes
flash vaporization chamber in which the evapo- (usually <1 ml) for trace analysis. The PTV
rated sample is mixed with carrier gas and injector was originally introduced in Europe in
divided between a stream entering the column the 1980s but it was over a decade later before
(carrier gas flow) and a stream vented to waste it gained traction in the USA. For much of this
(split flow) was the first practical solution to time it was an option but not heavily promoted
this problem. Split injection can be used to or supported by instrument companies based in
generate extremely small bandwidths for high the USA. This was probably due to a combina-
peak capacity separations but discriminates tion of commercial preferences and the not-
against less volatile compounds and the invented-here syndrome.
1.7. INCREASING SOPHISTICATION OF INSTRUMENTATION 13
The production of wide-bore fused-silica preparation and handling steps using a syringe
capillary columns coated with immobilized containing a retractable sorbent fiber compatible
stationary phases in the early 1980s allowed the with solvent-free sample introduction [33]. By
syringe needle to be introduced directly into the turn of the century, this had become one of
the column and eliminated the problem of the most popular sampling/sample introduction
removal of the stationary phase by the relatively techniques in gas chromatography. It can be seen
large volume of solvent released inside the as the forerunner of the liquid-phase microextrac-
column by the syringe [31]. These advances in tion techniques developed over the last decade.
column technology facilitated the development Together these microextraction formats are
of cold on-column injection where the sample is responsible for achieving a better scaling of
introduced as a liquid into the column inlet sample size requirements to sample utilization
and subsequently vaporized. Discrimination capabilities of capillary columns and are having
because of volatility differences was virtually a considerable impact on laboratory working
eliminated and the risk of sample decomposition practices.
minimized. With secondary cooling of the
injector, the oven temperature could be kept
1.7.3. Detectors
well above the boiling point of the solvent while
maintaining the column inlet at a much lower Gas chromatography is blessed by a number
temperature. This is important for using on- of robust and sensitive detectors based on gas-
column injection in high-temperature gas chro- phase ionization processes (e.g. flame ioniza-
matography. Dirty samples present a problem tion, thermionic ionization, electron capture,
owing to contamination of the sample introduc- and photoionization), bulk property (thermal
tion zone, which leads to poor chromatography conductivity), optical (flame photometric,
and unreliable quantification. Both on-column chemiluminescence, and atomic emission), elec-
and programmed-temperature vaporizer injec- trochemical (electrolytic conductivity), and
tion afford high accuracy and precision. These advanced spectroscopic (mass spectrometry
injectors also facilitated the direct coupling of and infrared spectroscopy) detection principles.
other chromatographic systems to gas chroma- Many of these detectors were introduced during
tography, such as liquid chromatography and the initial phase of the development of gas chro-
supercritical fluid chromatography [32]. In the matography and are still used today in a modi-
1980s, Carlo Erba manufactured an instrument fied form reflecting changes in enabling
for on-line liquid chromatographyegas chroma- technology. The exception is the (Hall) electro-
tography, but the uptake was poor and the lytic conductivity detector, which, because of
system was discontinued. its large detector volume and basis in wet chem-
The array of sample introduction methods for ical processes, is little used today. Interfacing of
gas chromatography and an understanding of modern detectors to capillary columns is not
the mechanistic details on which they rely were normally difficult except for fast gas chromatog-
complete by the end of the 1980s and modifica- raphy where the detector volume and data
tions since then have been evolutionary. The acquisition rates limit the use of some detectors.
most visible change is the shift from manual to The flame ionization detector facilitated many
automated sample introduction systems that of the early developments in capillary columns
accept samples in vials and can be programmed (Section 1.4). This detector has a low dead
for sequential analysis of each or selected vials volume, a high sensitivity for nearly all
in a batch. The introduction of solid-phase micro- carbon-containing compounds, and an
extraction in the early 1990s simplified sample extremely wide linear response range. Other
14 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

detectors were developed in response to the measurements of peak heights, although it was
need for element-selective or structure-selective recognized that peak area measurement was
detection. These detectors allowed the analysis fundamentally better but more difficult. The
of target compounds at low concentrations in product of the peak height and the peak width
complex samples, and made an important at half height could be used as an approximate
contribution to the rapid acceptance of gas chro- area measurement but this was neither fast nor
matography in laboratories performing routine (usually) accurate, especially for narrow peaks.
analysis as well as research laboratories. Approaches based on planimetry and
A special mention should be made of the cuteand-weigh procedures could afford reason-
coupling of gas chromatography with mass spec- able accuracy but demanded considerable skill
trometry (GCeMS). This coupling dates almost to achieve acceptable results and were slow
from the beginning of gas chromatography, but and tedious. Electromechanical devices such as
in the early days the practical problems and the ball and disk integrators and integrating
high cost meant that the combination was amplifiers were early attempts at automating
confined to a few research laboratories. When this time-consuming process. These devices
packed columns dominated the practice of gas were rather limited in range and speed of
chromatography, separators were required to response. The advent of microprocessors in the
reduce typical column flow rates to the vacuum- early 1970s ushered in the computing integrator
handling capacity of the ion source of the mass which took over the labor-intensive process of
spectrometer. Separators based on a jet, glass data management, eventually resulting in
fritted tube, or diffusion membrane design devices that could record a chromatogram
allowed sample enrichment while simulta- simultaneously with data acquisition and,
neously reducing the gas flow to the ion source. when the separation was complete, instantly
Separators disappeared nearly completely with generate reports in tabular form of all sorts of
the introduction of fused-silica open-tubular descriptive information from the chromato-
columns, which could be routed from the column gram, including the calculation of peak areas.
oven through a heated transfer line directly into It was not long before mainframe computers
the ion source of the mass spectrometer. During were in use to handle data produced from
the time that instrumentation for gas chromatog- a number of instruments simultaneously. The
raphy was advancing, so were all aspects of mass development of the personal computer and the
spectrometry, which became more powerful, rapid growth in processor speed and memory,
affordable, reliable, and available to laboratories however, had a greater impact and resulted in
with low-skilled personnel. The great advantage data management being handled by software
of the mass spectrometer as a detector is its ability running on a dedicated personal computer
to provide structural information for identifica- with information being archived locally and/
tion using different ionization approaches and or centrally on a laboratory information
through techniques like tandem mass spectrom- network server. Today, the large amount of
etry. Another advantage is its high sensitivity data produced by capillary columns (especially
and selectivity as a detector in the selected ion when coupled to a mass spectrometer) can be
or high-resolution mode. handled by a personal computer which simulta-
neously functions as the system manager
controlling and monitoring all aspects of the
1.7.4. Data Handling
chromatograph. What has not changed is that
In the early days of gas chromatography peak the methods used to manipulate the chromato-
quantification was done by purely manual graphic data for display and calculation are
1.7. INCREASING SOPHISTICATION OF INSTRUMENTATION 15
considered proprietary information, requiring operation. It had to compete for preference
an act of faith that what is provided as a printout with a new generation of affordable and easy-
or on screen is exactly what happened in the to-use gas chromatographyemass spectrometry
column [34]. What is clear, however, is that it systems, however, and was not a commercial
would be nearly impossible to persuade scien- success [35]. In the early 1990s, Phillips intro-
tists to return to the status quo before the elec- duced comprehensive gas chromatography in
tronic age of data handling as laboratory which the whole separation on the first column
productivity would be severely limited. was transferred to a second column of different
selectivity, the two columns being connected
1.7.5. Comprehensive Gas through a modulation interface. In this case, the
chromatogram is recorded as a two-dimensional
Chromatography
contour plot with the planar axes consisting of
Sometimes the separating power of gas chro- the independent retention times for each
matography, even when augmented by the iden- compound on the two columns and the vertical
tification power of spectroscopic detectors such axis is related to the amount of each compound.
as mass spectrometry, is still insufficient to accu- This technique was not an instant success, but
rately determine the composition of a sample. after further refinements in modulator design,
This problem is tackled in contemporary studies a decade later it had entered the main stream of
using comprehensive gas chromatography. commerce. These instruments are at the cutting
Comprehensive separations have their origins edge of instrument development where capabil-
in multidimensional gas chromatography devel- ities are pushed to match column performance.
oped in the early 1960s after the introduction of The function of the modulator is to arrest,
the Deans switch for controlling the flow direc- concentrate, and launch a fraction of the material
tion of a gas stream at a T-piece using pressure contained within each peak of the chromatogram
changes. Multidimensional (typically two- from the first column and transfer them as
dimensional) gas chromatography employs a series of pulses of constant frequency to the
two independent columns connected to each second column. To avoid overlap of individual
other by an interface. In early versions a fraction separations on the second column, each chro-
of the chromatogram, a heartcut, from the first matogram must be complete in less time than
column was isolated by cold trapping and then the cycle time of the modulator. Thus, the
released by thermal desorption for separation second-column separations must be very fast
on the second column. The disadvantage of this with a total separation time measured in
approach is that the fraction collected at the inter- seconds. This requires incredibly small injection
face contains no information about the separa- bandwidths, short second columns of small
tion on the first column. In the 1980s, Siemens internal diameter, fast detector response times,
marketed a remarkable instrument for its time, and repetitious and fast changes in instrument
the SiChromat-2, which had two independently operating conditions. Comprehensive gas chro-
controlled air-circulation ovens, housing the matography generates a large amount of data,
two columns, connected by a T-piece to isolate, especially when mass spectrometry is used as
focus, and re-inject fractions from the first a detector, and challenges remain in how best
column to the second by live switching based to make these data available to the user in
on the Deans switching mechanism. This instru- a form convenient for analysis. In these instru-
ment likely contained the most complex ments, one can perhaps see a vision of the future
pneumatic system of any instrument of its gener- of single column gas chromatography (based on
ation and required considerable skill in its technology developed for the second-dimension
16 1. MILESTONES IN THE DEVELOPMENT OF GAS CHROMATOGRAPHY

separation) while at the same time highlighting and the usefulness of gas chromatography to
the limitations of current technology for gas decision makers is called into question. When
chromatography. I first started to make my way in gas chromato-
graphy, I never thought that such a reversion
would occur so quickly, and as retirement
1.8. DECLINE IN THE EXPERTISE approaches, that intellectually I might leave
OF THE AVERAGE GAS the field in a poorer state than when I entered
CHROMATOGRAPHER it. There is no doubt that advances in tech-
nology have delivered better instruments and
At the present day the success in the columns but now in the hands of less qualified
commercial sector in gas chromatography has operators; this does not always lead to better-
resulted in a situation not all that different quality data.
from a narcotic dependency. To be classed as
an expert in gas chromatography, all that seems
to be necessary is the wherewithal to buy a gas
References
chromatograph and a column. Today, those [1] A.T. James, A.J.P. Martin, Gaseliquid partition chro-
tasked to operate a gas chromatograph do not matography: the separation and microestimation of
volatile fatty acids from formic to dodecanoic acid,
need to even make an injection, columns have Biochem J 50 (1952) 679e690.
become classified as laboratory supply items, [2] L.S. Ettre, The beginnings of gas adsorption chroma-
and data as something that appears on tography 60 years ago, LC-GC North America 26
a printout at the end of an experiment. When I (2008) 48e51.
was first introduced to gas chromatography in [3] F. Prior, Determination of adsorption heats of gases
and vapors by application of the chromatographic
the 1970s, it was necessary to make, modify, method in the gas phase. Doctoral theses. University
and continuously improve columns and instru- of Innsbruck: Innsbruck, Austria; 1947 (in German).
ments to meet the needs of each application. [4] O. Bobleter, Exhibition of the first gas chromato-
Expert chromatographers of this time were of graphic work of Erika Cremer and Fritz Prior, Chro-
necessity well versed in troubleshooting and matographia 43 (1996) 444e446.
[5] A.N.H. Ray, Gas chromatography, IeII, J Appl Chem
thoroughly knowledgeable of the principles 4 (1954). 21e25 and 82e85.
and practice at what often appeared to outsiders [6] L.S. Ettre, A. Zlatkis (Eds.), 75 years of chromatog-
as the dark arts of separations. On account of raphy e a historical dialogue, Elsevier, Amsterdam,
the immense powers of separations virtually 1979.
any unskilled individual can produce data [7] H.A. Laitinen, G.W. Ewing, A history of analytical
chemistry, division of analytical chemistry, American
today; even the village idiot can inject a sample Chemical Society, Washington, 1977, pp. 303e304.
and obtain peaks that are interpreted as usable [8] L.S. Ettre, The early development and rapid growth of
data. This has led to a continuing decline in the gas chromatographic instrumentation in the United
expertise of the average practicing chromatog- States, J Chromatogr Sci 40 (2002) 458e472.
rapher from the mid-1980s to the present time. [9] M.J.E. Golay, in: D.H. Desty (Ed.), Gas chromato-
graphy 1958 (Amsterdam symposium), Butterworths,
This can be perilous, because everything from London, 1958, pp. 139e143.
column selection to trouble-shooting skills is [10] J.E. Lovelock, S.R. Lipsky, Electron affinity spec-
based on a fundamental knowledge of chro- troscopy e a new method for the identification of
matographic principles, the absence of which functional groups in chemical compounds separated
degrades the quality and usefulness of the infor- by gas chromatography, J Am Chem Soc 82 (1960)
431e433.
mation acquired by these instruments. To [11] J. Harley, W. Nel, V. Pretorious, Flame ionization
address these problems requires a massive detector for gas chromatography, Nature (London)
educational effort before the knowledge is lost 181 (1958) 177e178.
REFERENCES 17
[12] I.G. McWilliam, R.A. Dewer, in: D.H. Desty (Ed.), Gas [24] J.K. Haken, Developments in polysiloxane stationary
chromatography 1958 (Amsterdam symposium), phases in gas chromatography, J Chromatogr 300
Butterworths, London, 1958, p. 142. (1983) 1e77.
[13] I.G. McWilliam, The origin of the flame ionization [25] R.D. Dandeneau, E.H. Zerenner, An investigation of
detector, Chromatographia 17 (1983) 241e243. glasses for capillary chromatography, J High Resolut
[14] L.S. Ettre, The invention, development and triumph Chromatogr 2 (1979) 351e356.
of the flame ionization detector, LC-GC (Europe) 15 [26] W. Jennings, J.A. Settlage, R.J. Miller, Multiple short
(2002) 364e373. pass glass capillary gas chromatography, J High Res-
[15] K. Grob, Hansjurg Jaeggi e Obituary, J High Resolut olut Chromatogr 2 (1979) 441e443.
Chromatogr 12 (1989) 459. [27] G.M. Gross, V.R. Reid, R.E. Synovec, Recent advances
[16] D.H. Desty, J.N. Haresnape, B.H.F. Whyman, in instrumentation for gas chromatography, Current
Construction of long lengths of coiled glass capillary, Anal Chem 1 (2005) 135e147.
Anal Chem 32 (1960) 302e304. [28] C. Cruz-Hernandez, F. Destaillates, Recent advances
[17] K. Grob, The role of column technology in capillary in fast gas chromatography: application to the sepa-
gas chromatography, J High Resolut Chromatogr 7 ration of fatty acid methyl esters, J Liq Chromatogr
(1984) 252e257. Rel Technol 82 (2009) 1672e1688.
[18] M.J.E. Golay, in: D.H. Desty (Ed.), Gas chromatog- [29] K. Grob, Classical split and splitless injection in
raphy 1958 (Amsterdam symposium), Butterworths, capillary gas chromatography, 2nd ed., Huethig,
London, 1958, pp. 36e55. Heidelberg, 1986. 1992.
[19] E.L. Ilkova, E.A. Mistrukov, A simple versatile [30] P. Sandra (Ed.), Sample introduction in capillary gas
method for coating glass capillary columns, J Chro- chromatography, Huethig, Heidelberg, 1985.
matogr Sci 9 (1971) 569e570. [31] K. Grob, On-column injection in capillary gas
[20] M.C. Bourne, W.G. Jennings, Existence of two soil chromatography, 2nd ed., Huethig, Heidelberg,
species in detergency investigations, Nature (London) 1987. 1991.
197 (1963) 1003. [32] K. Grob, On-line coupled LCeGC, Huethig, Heidel-
[21] M.C. Bourne, W.G. Jennings, Kinetic studies of berg, 1991.
detergency. I Analysis of cleaning curves, J Am Oil [33] J. Pawliszyn, Solid-phase microextraction: theory and
Chem Soc 40 (1963) 517e523. practice, Wiley-VCH, New York, 1997.
[22] M.C. Bourne, W.G. Jennings, Kinetic studies of [34] N. Dyson, Chromatographic integration methods, The
detergency. II Effect of age, temperature, and cleaning Royal Society of Chemistry, Cambridge, UK, 1998.
time on rates of soil removal, J Am Oil Chem Soc 40 [35] W. Bertsch, Two-dimensional gas chromatography
(1963) 523e530. concepts, instrumentation, and applications e part 2:
[23] L. Blomberg, Current aspects of the stationary phase comprehensive two-dimensional gas chromato-
in gas chromatography, J High Resolut Chromatogr graphy, J High Resolut Chromatogr 23 (2000)
5 (1982) 520e533. 167e181.

S-ar putea să vă placă și