Sunteți pe pagina 1din 59

KOHTECT AVV-701

LASER SHAFT ALIGNMENT SYSTEM


OPERATING INSTRUCTIONS MANUAL

KOHTECT(c) 2013
CONTENT

1. General and introduction ......................................................................................................................... 5


1.1 Laser Safety Precautions .................................................................................................................... 5
2. Technical Description .............................................................................................................................. 6
2.1. Designation ....................................................................................................................................... 6
2.2. Specification and Features ................................................................................................................ 6
2.3. System Package. ............................................................................................................................... 7
2.4 Misalignment Parameters .................................................................................................................. 8
3. Machine Alignment ................................................................................................................................. 9
3.1 Input Measurement Data .................................................................................................................. 10
3.2 Set-up the Device ............................................................................................................................. 10
3.3 Connecting the Measuring Units ..................................................................................................... 10
3.4 Input of Dimension .......................................................................................................................... 10
3.5 Rough Alignment Procedure ........................................................................................................... 12
4. Getting Started ....................................................................................................................................... 14
4.3 Horizontal Machine Alignment ....................................................................................................... 14
Input dimensions ................................................................................................................................ 14
Change parameters ............................................................................................................................. 15
Biaxial and uniaxial adjustment modes.......................................................................................... 16
Collect data ........................................................................................................................................ 16
View and edit the points .................................................................................................................... 17
To call collect data popup menu ........................................................................................................ 18
Calculate alignment results ................................................................................................................ 19
Shim simulation ................................................................................................................................. 20
Thermal growth.................................................................................................................................. 20
Setup tolerances ................................................................................................................................. 21
4.3.2 To start correction for the misalignment................................................................................... 21
4.3.3 To save alignment measurement data ....................................................................................... 22
Quit program ..................................................................................................................................... 22
4.4 Vertical Machine Alignment ........................................................................................................... 23
Input dimensions ................................................................................................................................ 23
Change parameters ............................................................................................................................. 24
Collect data ........................................................................................................................................ 24
Calculate alignment results ................................................................................................................ 25
4.4.1 To start correction for the vertical misalignment..................................................................... 25
4.4.2 To save alignment measurement data ....................................................................................... 26
4.5 Program for Soft Foot check and correction .................................................................................... 27
Input dimensions ................................................................................................................................ 27
Take measurement ............................................................................................................................. 27
To save soft foot result....................................................................................................................... 28
4.6 Bore Center Line Program ............................................................................................................... 29
Units configuration ........................................................................................................................... 29
Measurement preparation .................................................................................................................. 29
To start bore centerline program ........................................................................................................ 29
Set parameters .................................................................................................................................... 29
Take measurement ............................................................................................................................. 30
Evaluate result.................................................................................................................................... 31
Live adjustment.................................................................................................................................. 31
KOHTECT(c) 2013 2
Save result .......................................................................................................................................... 32
4.7 Flatness Program.............................................................................................................................. 33
Select grid shape ................................................................................................................................ 33
Setup grid parameters ........................................................................................................................ 34
Select data source.............................................................................................................................. 35
Taking measurements ........................................................................................................................ 35
The result display ............................................................................................................................... 36
Circle shaped plane ............................................................................................................................ 37
Setup circular grid parameters ........................................................................................................... 37
Taking measurements ........................................................................................................................ 38
The result display ............................................................................................................................... 38
To save data ....................................................................................................................................... 39
Quit flatness program......................................................................................................................... 39
4.8 Squareness Program ......................................................................................................................... 40
Set parameters .................................................................................................................................... 40
Taking measurements ........................................................................................................................ 41
Measurement result ............................................................................................................................ 42
To save data ....................................................................................................................................... 42
Quit squareness program ................................................................................................................... 43
4.9 Straightness Program ....................................................................................................................... 44
Straightness measurement.................................................................................................................. 44
Setup parameters ................................................................................................................................ 44
Set measurement points number ........................................................................................................ 44
Set dimensions ................................................................................................................................... 44
Define a splice ................................................................................................................................... 45
Switch between manual data and data from detector ........................................................................ 45
Laser beam adjustment principle ....................................................................................................... 45
Carrying measurements ..................................................................................................................... 45
Evaluate measurement results............................................................................................................ 46
Splice concept .................................................................................................................................... 46
Step by step instruction to start measure new splice ......................................................................... 47
The Live move mode ......................................................................................................................... 48
To save data ....................................................................................................................................... 48
4.10 Cardan program ............................................................................................................................. 49
Input dimensions ................................................................................................................................ 51
Collect data ........................................................................................................................................ 52
Calculate alignment results ................................................................................................................ 53
To start corrections for the misalignment .......................................................................................... 53
To save data ....................................................................................................................................... 53
Quit cardan program .......................................................................................................................... 55
4.11 Spindle program ............................................................................................................................. 56
Input distances ................................................................................................................................... 56
Taking measurement .......................................................................................................................... 56
Evaluate results .................................................................................................................................. 57
To save data ....................................................................................................................................... 58
5.0 Standards Tolerance of Shaft Misalignment ........................................................................................ 59
6.0 Delivery Set ......................................................................................................................................... 59

KOHTECT(c) 2013 3
KOHTECT(c) 2013 4
1. General and introduction

1.1 Laser Safety Precautions

The KOHTECT AVV-701 alignment system is the class II laser device at typical wavelength of
670nm, delivered output power of less than 1 mW and maximum radiant energy per pulse of
0.1 mJ. The Class II laser comply with requirement outlined by USAs FDA as well as
international ANSI, BS 4803 and IEC 825 standard.

Be sure to follow the following safety precautions to avoid personal injuries and damage to the
system

Do not look directly into laser the beam at any time!

Do not direct laser beam on to the peoples eyes!

ATTENTION!
Do not try open / dismantle measuring units and the display unit this can damage the system,
and your after-sales service warranty will come void.

Warning!
Be sure the machines to be measured, cannot be started unintentionally as this can cause
injuries. For this purpose, before the mounting of equipment, either block the power switch in
the Off position or remove the safety fuses. These precautionary rules must be followed until
the measuring system is dismantled from the measured machine.

INJURY RESPONSIBILITY DISCLAIMER

Neither the NPP KOHTECT enterprise nor our authorized dealers are liable for the damages
caused to machinery or equipment by use of the AVV-701 system. We carefully check text of
this manual to eliminate errors, nonetheless there may be mistakes or inaccuracy involved. We
will be grateful for your reporting to us about any error, and we will be able to correct them in
the subsequent editions of the manual.

KOHTECT(c) 2013 5
2. Technical Description
2.1. Designation

AVV-701 alignment system (C911 sub-system) (further as System) is designed for:


checking of shaft alignment of mechanisms;
estimation of the surface flatness.

The checking of shaft alignment means adjustment of the relative position of two coupled
machines (e.g. motor and pump) so that the centre line of the axis will be concentric when the
machines are running during normal working conditions.

2.2. Specification and Features

2.2.1. Separation distance between measuring transducer units, up to 10m


2.2.2. Display control operating temperature range, -10..+55 degree C
2.2.3. Measurement accuracy, 1%+0.01
2.2.4. Laser type: Visible red 635-670 nm, <1 mW
2.2.5. Detector type: Positional-sensitive photodiodes, 1010 mm
2.2.6. Display resolution, 0.01 or 0.001 mm, (1 or 0.1 mil )
2.2.7. Measuring resolution, 0.001mm
2.2.8. Electronic inclinometer resolution, 0.1 degree
2.2.9. Power supply: Rechargeable NiMH battery
2.2.10. Gross weight, 7.5 kg
2.2.11. Built-in application programs and options:
horizontal shaft alignment at any 90 shaft position;
vertical shaft alignment; Setup options;
horizontal shaft alignment with rotation angle less than 120;
Soft foot;
thermal growth;
selection of shimming simulator to calculate for expected alignment;

KOHTECT(c) 2013 6
2.3. System Package.

The System includes (Fig. 1):


AVV701 display control unit
two measuring transducer units
universal chain brackets for mounting of the
measuring units
measuring tape, mm/inch
220 Volts AC charger
connecting cables;
CD-ROM PC software
USB PC communication cable
Operating instructions manual in CD
Carrying case with form-inserted

* Top positions to be faced up while setting up both the transducer.


Datum lines for measuring of dimension input

KOHTECT(c) 2013 7
2.4 Misalignment Parameters
Misalignment of any rotating machine is expressed in parallel (Offset) and angular (Gap) of the
shafts. Most frequently in practice, both of them are present simultaneously. Different kinds of
misalignment of axes are shown in Fig. 2.

Parallel misalignment of axes Offset (displacement)

Angular misalignment of axes Gap

Parallel and angular misalignment of axes (Offset + Gap)

Fig 2

The parallel (Offset) and angular (Gap) misalignment of axes is determined in two mutually
perpendicular planes. For the purpose of elimination of the parallel and angular misalignment
of axes, in each of the planes a correction of position of the movable machine (M) will be done.
For the horizontal mounted machine the movable machine (M) position is adjusted in the
horizontal and vertical planes.
For the vertical mounted machine, operator determines arrangement of the correction planes,
basing on considerations of the convenience and technological effectiveness of moving of the
movable (M) machine.
Stationary machine (S) - in the process of eliminating of the axes misalignment the position of
this machine stay static, i.e. it does not move.
Movable machine (M) the machine, which position is adjusted for eliminating of the parallel
and angular misalignment of axes.
The measurement system calculates the values of the angular and parallel misalignment of
axes in the plane of the coupling (in two mutually perpendicular planes), and the adjustment
values for the machine feet on the movable (M) machine, that is necessary for elimination of
this misalignment of axes. Fig. 3 shows misalignment of axes and the values for its correction
just for vertical plane.

Fig 3.

KOHTECT(c) 2013 8
3. Machine Alignment

Mount the Measuring transducer units on the shafts of the (S) and (M) machines
Select program according to the application of machine to be measured
Input the distances between the (S) and (M) units, the coupling and the movable
machine feet
Press to record readings from the measuring units at three different positions of
the shafts
Adjust the machine feet position of the movable machine in accordance with
correction results of the calculated value on control screen
Save the measured result into the file

While making the measurement, it is may be necessary to observe and understand the
orientation and rotation direction of the shafts with the (S) and (M) measuring transducer units
with regard to the relative position of the (S) and (M) machines as in Fig.4.

Fig 4.

Fig. 4 shows the view of (S) machine from the (M) end view, at the 12:00 oclock position. The
measuring transducer units have marking (S) and (M) on the top of each unit, should be
mounted with brackets onto the shafts of the (S) stationary and (M) movable machine
respectively.

KOHTECT(c) 2013 9
3.1 Input Measurement Data

AVV-701 system function is based on the measurement of the laser beam movement on the
detector receivers window during the turning of shafts with the measuring units mounted.
To enable the System to carry out the shaft alignment calculation, it is necessary to record the
measurement data in at least three positions of the shafts rotation. Turning the shafts in the
range of 180 is sufficient, but for the best results turn as wide as possible recommended.
If design features of the machines do not make it possible to carry out rotation of the shafts
with the measuring units mounted up to 180, the instrument also provides smaller shaft
rotation angle measurement mode that allows possibility to enter the measuring data when the
shafts rotation angle constitutes as low as 60, note turn as wide as possible always
recommended. Systems advanced multipoint mode allows shafts rotation in any direction and
up to 36 points.

3.2 Set-up the Device

Before the beginning of the work, check the battery voltage and charge it if necessary.
Battery voltage is indicated on the display of the device in the main menu.
The device is automatically turned off when the voltage is lower than 4.6 V.
Check and clean if necessary the surface of detectors and aperture of laser.
Use soft tampons, moistened with alcohol, for the cleaning. Solvents must not be used!
Check and setup if necessary, date and time of the system clock.

3.3 Connecting the Measuring Units


There are two connectors in the Control Display unit and each Measuring transducer units.
Connection to the Control Display unit is arbitrary, i.e. the units can be connected either in
parallel or in series (Fig. 5) with any of the two cables contained in the set, to any of the
connectors in the Control Display unit and the Measuring transducer units.

Connecting the measuring units in parallel

Connecting the measuring units in series


Fig. 5

3.4 Input of Dimension

To enable the AVV-701 to carry out accurate calculations it is necessary to input the distances
between the measuring transducer units, the coupling and the machine feet.

KOHTECT(c) 2013 10
Fig. 6 shows the dimensions input for the horizontal plane alignment.
Fig. 7 shows the dimensions input for the alignment of vertical flange mount machine.

Fig 6.

S-M distance between measuring transducer units.


S-C distance between S and center of coupling.
S-F1 distance between stationary detector (S) and the feet pair 1 (F1).
S-F2 distance between S and F2 (must be longer than S-F1). If the machine has three
pairs of feet, you can change this distance after finished measurement, and then
repeat the calculation and get a new adjustment value for this pair.

KOHTECT(c) 2013 11
Fig 7.

S-M distance between measuring transducer units.


S-C distance between S and center of coupling.
S-F1 distance between stationary detector (S) and the plane of alignment (F1).

3.5 Rough Alignment Procedure


Rough alignment should be applied only when the alignment is extremely poor, the laser
beams may travel outside the detectors during rotation of the shafts with the Measuring units
mounted. If this happen it is necessary to do a rough alignment first.

Rough alignment procedure (variant 1), (Fig. 8):


Turn shafts with measuring units to the 9 oclock position. Aim the laser beams at the centre
of the closed detectors.
Turn shafts with measuring units to the 3 oclock position.
Check where the laser hits, then using the laser adjustment screws, adjust the beam half
the travel in direction to the centre of the target (Fig. 8).
Adjust the movable machine so that the laser beam hits the centers of both the targets, (S)
and (M).
Follow the regular procedure to continue.

KOHTECT(c) 2013 12
Fig 8.

Rough alignment procedure (Variant 2):


Turn the shafts with measuring units to the 9 oclock. Apply the targets in the view of scaled
paper sized 50 x 50 mm to the detector surface. Aim the laser beams at the centre of these
targets.
Switch the display unit into the mode of manual data input.
Turn the shafts with measuring units to the 12 oclock position and enter the values of laser
beam at the 12 oclock position, then turn the shafts to the 3 oclock and enter the values of
laser beams at the target marking.
The system will calculate roughly the offset value and adjustment value for movable
machine (M).
Adjust the movable machine according to the results of calculations.
Follow the regular procedure to continue.
While entering manually the values of the position of laser beam at the target, take into account
the sign (Fig. 9)

View of a target (M) at the 12 oclock position

View of a target (S) at the 12 oclock position

Fig 9.

KOHTECT(c) 2013 13
4. Getting Started
To start AVV-701:
In Main Menu by using the arrow keys move the selection bar to Alignment then press

.
The display shows alignment menu screen with programs option
as in Fig.10.
The lower section display the day, date and time of system clock,
and the battery charge power within the device.

4.3 Horizontal Machine Alignment

Mount the Measuring transducer marked with (S) on the shafts of the stationary machine and
Measuring transducer marked (M) on the moveable machine. Connect the cables as per
section 3.3, between the Measuring transducer units and Control Display unit.

Select Horizontal to enter horizontal alignment program, then press .

Press to start new alignment task or press to resume saved task.

Important! During alignment measurement:


DO NOT change the positions of measuring units at the time of interruption of the work;
DO NOT move the movable machine when the device is turned on.

Input dimensions
Dimensions between the two transducers(S-M), between S unit and coupling center
(S-C), between the S unit and the front feet of moveable machine (S-F1) and between the S
unit and the rear feet of moveable machine(S-F2) should be entered.
KOHTECT(c) 2013 14
measure units

MA means manual
angle IA means use
inclinometer

MD means manual data


LD means use
transducers data

2D means biaxial
adjustment mode 1D
means uniaxial mode

Use or to move from dimension to dimension to input or reconfirm all dimension


data.
To start input dimension use key, when the blinking cursor appeared use numeric key pad
to enter desired value then press to confirm.

Change parameters

To change parameters call popup menu by press of key


When popup menu appears use to select between inclinometer and manually entered
angle, use to select between transducers data and manually entered data, use to
select between biaxial and uniaxial alignment mode.
MA means manual
angle IA means use
inclinometer

MD means manual data


LD means use
transducers data

2D means biaxial
adjustment mode 1D
means uniaxial mode

KOHTECT(c) 2013 15
Biaxial and uniaxial adjustment modes
When biaxial mode used both vertical and horizontal transducers data used in live
adjustment mode to perform live adjustment. Shafts rotation not needed and strictly prohibited
to adjust moveable machine in vertical and horizontal planes.
uniaxial means traditional live adjustment mode where you should turn shafts to predefined
transducers position to perform vertical or horizontal adjustment. The position should be 12 or
6 oclock for vertical adjustment and 3 or 9 oclock for horizontal adjustment.

Collect data
AVV-701 system does collect data in advanced multipoint mode. Up to 36 points using
any shafts rotation direction can be stored for the best results. At least three points should be
stored.

To start collect data press The laser beams turned on now.


Screen of wait connection to transducers appears. Note that Bluetooth connection need more
time for connection to be established.
If connection cannot be established during more than 90 seconds check if pin codes entered or
check transducers.

Transducers serial numbers become


available when connection established

After the connection has been established data collect screen appears
Using fine adjusting screws on the Measuring transducers, adjust the laser beams to the
centers of detectors shutter cover one at a time, refer Fig. 1a & 1b. Once laser is centered on
both transducer, open the shutters. The screen starts display detectors coordinate position
readings for both transducers (S) & (M)
Face the stationary machine from the movable machine.

KOHTECT(c) 2013 16
Stored or manually
Main clockface
entered Y values
with stored points
showed

Stored or manually
Selected point
entered angle
index

R means we Current angle


currently view
already stored
point Current Y values

To store current point press


Blinking outer circle means
minimum turn violation

If minimum turn requirement not


satisfied then hint shows
minimum turn required

The minimum turn between points(>10) should be respected.

Up to 36 points can be stored


NOTE: to achieve acceptable results the whole shafts turn must not be less than 60
degrees.

View and edit the points


Each of the points can be viewed, replaced or deleted. To increase selected point index
press and to decrease selected point index press You can edit point only when R
sign blinks at the left bottom corner of the screen.

To delete selected point press (you should not be in manual value edit state)

Delete confirmation dialog will pop up. To confirm point deletion press

To replace selected point in the manual data mode enter new manual data and confirm by
press Replace confirmation dialog will pop up. To confirm value replacement press

KOHTECT(c) 2013 17
Replace confirmation dialog

press to confirm

To replace selected point when transducers data used press and


confirm replacement.

To call collect data popup menu

press . Then
To change display resolution between 2 and 3 decimal digits press
To enable or disable input data filter press you can change filter averaging number in the

round robin fashion by press . To reset filter press .


If manual angle used press to enter angle value press to confirm. When manual
angle entered and manual data mode is choosed then cursor switched to enter M and S Y
values automatically. So to enter all manual values press then enter angle then press
to confirm then enter M Unit value then press to confirm then enter S Unit value then press
to confirm.
If manual data used press to enter M unit value then press to confirm and then press
to enter M unit value then press to confirm.
Filter averaging value

KOHTECT(c) 2013 18
Calculate alignment results

When all of the desired points gathered press to go to view calculated results and to make
live adjustment

Note: Screen message show Do not turn shafts while


aligning

The alignment measurement result of calculations shows the values of the angular Gap and
parallel Offset misalignment in the plane of the coupling (in Horizontal and Vertical planes) as
well as the correction values for the machine feet F1 (machine front feet) & F2 (machine rear
feet) on the movable (M) machine that are necessary for eliminate the horizontal and vertical
misalignment.
STOP means live
adjustment now off

use for change

! indicates tolerances
violation

Note:
For the purpose of clarity the values of parallel Offset and angular Gap misalignment in the
plane of the coupling are shown in the view of the symbols of half-couplings.
The adjustment values of the position of the feet F1 and F2 of the machine (M) in the horizontal
plane indicate the value of horizontal shift. The positive values mean that the feet must be
pushed, the negative values the feet must be pulled.
The adjustment values of the position of the front feet F1 and F2 of the machine (M) in the
vertical plane indicate the value of vertical shift. The positive values mean that the feet must be
lifted, the negative values the feet must be lower.

Press to freeze the alignment result on screen to temporary jot down of coupling results,
Fig 19a, while prepare for live adjustment.
To resume live measuring mode, Fig 19b, on machine, press again and start shimming
and adjust horizontal movement with under the live adjustment mode.

KOHTECT(c) 2013 19
Note: To view option features(pop up) in Fig.19e, press and hold
For prompts screen message on the function key, press for
show following info:
to toggle resolution 0.01mm or 0.001mm
Shim simulator to check the shimming which size is
different from the results of calculation (Fig.19c).
Thermal growth to enter compensation values to thermal
growth (Fig.19d ).
to save the result. Enter My Documents. You may save this file in the old folder or
create a new folder.

Shim simulation

The shim simulation allows easy and convenient way to evaluate rest misalignment when shim
added or removed. To open shim simulation press .

Values in brackets shows calculated


results need to compensate
misalignment

Press to enter shim value for the front foot


Press to enter shim value for the rear foot

Use to confirm entered value

See rest
misalignment here

Thermal growth

Most machines develop a certain amount of heat while running. In the best
case both the driving and the driven machine are affected equally requiring
no input of compensation values. But in several applications the driven
machine is either hotter, i.e. a pump for hot liquid, or cooler than the driving
machine. The machine manufacturers define the thermal growth of the
machines differently, but in most cases you will find it as a factor of
deliberately misalignment expressed in parallel offset and angular error

KOHTECT(c) 2013 20
To enter thermal growth press .

Press to enter horizontal offset value


Press to enter horizontal gap value
Press to enter vertical offset value
Press to enter vertical gap value

Use to confirm entered value

Setup tolerances

To setup tolerances call options menu by press and then press

Use for cyclically change rpm, include set to OFF state

4.3.2 To start correction for the misalignment


To correct the horizontal and vertical plane in biaxial as per obtained coupling result in, press
to switch frozen result screen to LIVE mode with shaft remain at last recording clock
position. Loosen moveable machine feet and start adjust the feet correction values (add or
remove shims) according, and at the same time adjust the machine horizontal movement at
feet (push or pull using jacking bolt if any ) according to LIVE horizontal correction reading.
There are some differences when uniaxial mode used. For live adjustment in uniaxial
adjustment mode: freeze result screen if required(use key). Turn shafts to the 12 oclock
transducers position. Switch to LIVE mode and perform machine vertical movement(add or
remove shims). Do not move machine in the horizontal plane at the same time. When machine
vertical movement done press to freeze result screen.
Turn shafts to the 3 oclock transducers position. Switch to LIVE mode and perform machine
horizontal movement(push or pull using jacking bolt if any). Do not move machine in vertical
plane at the same time.
When machine horizontal movement done press to freeze results screen.

KOHTECT(c) 2013 21
4.3.3 To save alignment measurement data

When measurement finish, press to enter file save display.


Press again to save the alignment result main directory or
to make new directory or select the existing directory to
save.

Quit program

When quit the program:

- press to backup changes in measured data.


- press to discard changes.

KOHTECT(c) 2013 22
4.4 Vertical Machine Alignment
The vertical machine alignment program shares many similarities with horizontal machine
alignment. See also: horizontal machine alignment

Select Vertical to enter vertical alignment program, then press .

Press to start new alignment task or press to resume


saved task.

Mount the Measuring transducer units on the shafts of the machines (S) and (M) as shown in
Fig. 20a.
Note: Mark the clock positions 9-12-3-6 at the flange.
Using the connecting cables, connect Measuring transducer units and Control Display

Input dimensions

Input diameter here

Input bolts number


here

KOHTECT(c) 2013 23
Input the first dimension between the two transducer units, Fig.20a. Press. to start input
dimension then input the new dimension follow by to confirm input data.
Use or to move to 2nd dimension input, distance from Measuring transducer (S)
unit to coupling center, Fig 20b.
Use or to move to 3nd dimension input, distance from Measuring transducer (S)
unit to flange face, Fig 20c.
Use or to move to 4th dimension input, flange bolt center distance.
Use or to move to last dimension input, number of bolts.

Change parameters

Same lake in horizontal alignment program, to change program parameters press to call
popup menu
Note: inclinometers cannot be used in the vertical machine alignment

Press to switch between transducers data (LD) and


manual data(MD)
Press to switch between biaxial (2D) and
uniaxial(1D) live adjustment

Press when you enter all dimensions and set parameters.

Collect data

Like in the horizontal alignment wait connection screen appears

KOHTECT(c) 2013 24
The display shows information of the serial number on both Measuring transducer units to
identify device is connected properly.
When connection established system switch to collect data.

Refer to horizontal collect data as the procedure mostly the same,


except that the angle always entered manually(inclinometers not
used). All important things specified for the horizontal alignment
should be preserved

Calculate alignment results

When all of the desired points gathered press to go to view calculated results and to make
live adjustment

Vertical alignment results shares many similarities with horizontal


alignment results, except that there are no vertical or horizontal
planes here, but plane 9-3and plane 6-12 defined instead.
Also flange shim results can be viewed using
All important things specified for the horizontal alignment should be
preserved
As inclinometers not used care must be taken in uniaxial
adjustment mode about that shafts must be settled in proper position
before start of adjustment.

4.4.1 To start correction for the vertical misalignment

Press Screen message show Do not turn shaft while aligning before display shimming
values required for each individual bolts and make angular adjustment by adding in or remove
shims as per display values.
To freeze the result screen and LIVE adjustment mode, press to
toggle. Hand tight all bolts when shims adjustment done.
Press to toggle between offset adjustment (in 9-3 and 6-12
planes) and shims adjustment screen. Move the (M) machine along
9-3 and 6-12 direction as per offset adjustment values display on
screen.

KOHTECT(c) 2013 25
Press to repeat the entire vertical measurement task again to confirm machine alignment
has been done.

4.4.2 To save alignment measurement data

When measurement finish, press to enter file save display. Press


again to save the alignment result main directory or press to
make new directory or select the existing directory to save.

KOHTECT(c) 2013 26
4.5 Program for Soft Foot check and correction

Check the machine for the soft foot before fulfilling any alignment adjustments. Form the
alignment menu select Soft foot and press

Press to start new soft foot task or press to resume


saved task.

Input dimensions

You must check/input all the required machine dimensions before continue. See horizontal
alignment input dimensions
Note: Dont select manual input of data or angle

Take measurement

Wait until connection to sensors has been established

Screen message: Wait shaft should be in pos. 12 O cl to


required shafts to turn to 12 oclock position to start soft foot
measurement.

KOHTECT(c) 2013 27
Follow screen prompts message, release 1st bolt and wait
approximately 5 second for soft foot value to measure. Then press
key to record the value.
Tighten the bolt as screen prompts message, and press to move
to next bolt and repeat procedure as per prompts message on the
display until all the four individual feet soft foot readings are taken on
display. The result shows the difference between the released and
tightened bolts of the foot. Set maximum shimming values.

Choose to repeat entire soft foot check after remove the soft foot
as per measured result.

To save soft foot result

Press to enter data save screen and press again to save


soft foot reading or press to make new directory or select the
existing directory to save.

Press to quit soft foot program.

KOHTECT(c) 2013 28
4.6 Bore Center Line Program
The AVV-701 laser alignment system uses a visible laser beam to determine the relative the
relative centerline of a series of bores. The laser transmitter(emitter of the laser beam) is
mounted a one end of the machine and is aimed at the laser receiver(detector of the laser
beam), which is mounted near the center of each bore in turn. The receiver is mounted on a
special bracket which allow the freely rotation of receiver within the bore.

Units configuration
S unit laser transmitter (fixed position)
M unit receiver (on bore bracket)
M label should be on top side at 6h oclock pos.

Measurement preparation
Thoroughly clean the bores to be measured
Ensure that the boring machine(or bores) are readily moveable horizontally and vertically

To start bore centerline program

To start BoreCenterLine program press .


Press to start new soft foot task or press to resume
saved task.

Set parameters

Press to enter bores number from 3 to150

Press to switch between all distances between bores equal or


not
Use to select bores pair for enter distance between
them
Press to start enter distance, use numeric keypad to enter value and confirm input by
press Note: when all distances equal selected AVV-701 automatically set all distances to
entered value, so enter distance only once

KOHTECT(c) 2013 29
Press to change measurement configuration number of receiver positions where data
will be taken. In case of 3 points configuration it is recommended to enter bores diameters
(3: Diam.) it improves result precision. However if diameters are not entered, result calculation
is still possible.
In case of 8 points result is calculated with higher precision due to averaging.

Press to set Horizontal or Vertical bores orientation. In case of Vertical inclinometer will
not work. One should check measurement position manually.
Press to switch between use sensor data or manual data input

Take measurement

Press to begin measurement.


Sensor data

Plane number
Measurement
position data
Inclinometer
data
Plane
If sensor position differs from number
proper measurement position
warning start blinking. (If Bore center
inclinometer is ON) displacement

Measurement position indicated on screen by radial line. Current receiver (M) inclination is
indicated by blinking radial line.
Place receiver into measurement position and press to take
readings for current position. When readings for all position will
be taken bore center displacement will be calculated.

KOHTECT(c) 2013 30
Use to select next next position or to return to the
previous position. Plane number changed automatically in regard
to measurement mode

Evaluate result

When all bore planes are calculated, press result will display
Press to switch between table or H, V plot.

Move bore # you are going


to adjust into first line by

arrow key. Then press


to go live mode

Press to start enter reference points number, enter numbers for both points
confirmed by

Live adjustment

Press to go to live adjustment mode.


Sensor
live data
When in bore position adjustment mode (live
mode) receiver pin should be in vertical (6h
oclock) position:
Calculated bore
1) Fix receiver on adjusted bore check
center displacement
receiver position (relative to laser before adjustment
transmitter beam) by receiver data on
screen. Common beam should be
close to center of receivers window.

KOHTECT(c) 2013 31
2) When sensor is fixed in order to move together with

adjusted bore press to enable live bore


position data.

Live bore center


displacement during
adjustment

Adjust bore position to minimize displacement from common


centerline, when adjustment is finished press and stored
result update prompt will appear:
Press to save changes in bore position.

Save result

Press to enter data save screen and press again to save soft foot reading or press
to make new directory or select the existing directory to save.

KOHTECT(c) 2013 32
4.7 Flatness Program

This program for measurement of plane flatness relative to reference plane formed by laser
beam. The measurement points may be in the plane of circular or rectangular shape with
regular or variable grid step. The plane may contain cutouts. Up to 1600 measurement points
can be handled. The measurement values can be treated as absolute, recalculated to best fit
plane or three points can be set as reference plane.
Procedure: Plan the measurement and mark the points where the detector will be placed. Level
the laser within 0.5 mm [20 mils] in both X- and Y-direction. Start the Flatness program. Use S
or M unit as detector.
The detector unit should be installed with the label on top.

Start Flatness program move cursor and press

Choose option to resume interrupted work press . In this


case interrupted work will be opened for further data collecting,
editing, evaluating, saving on the flash drive.

To start new task press .

Select grid shape

When new task started the menu screen to choose plane


shape appeared.
Press for taking measurements on rectangular grid
Press for taking measurements on circular grid

KOHTECT(c) 2013 33
Setup grid parameters

Menu screen for setting up the grid parameters:


- Press to set grid step regular or variable
- Points across X 240
- Points across Y 240
- Grid size across X (for regular grid step)
- Grid size across Y (for regular grid step)
To set these parameters move blinking cursor by arrow keys

then press , enter the value, confirm by .

In case of irregular grid step (Grid Var option):


distances from first column to every next column across X can be
settled individually press to access menu
distances from first row to every next row across Y can be settled
individually press to access menu

To set distances across X choose


column by arrow keys then

press , enter the value, confirm by

To set distances across Y choose row


by arrow keys then press

, enter the value, confirm by .

KOHTECT(c) 2013 34
Measured plane can contain some cutouts of arbitrary shape those nodes that are
within cutouts can be just skipped out during taking of measurement. The flatness
program correctly treats such cutouts.
Row and column index numbering starts from 0. E.g. first row is Y0.

Select data source

When in grid setup use to switch between data from M Unit and manually entered data.

Taking measurements
Beam X position
indicator
Active Beam Y
node data

Place detector unit mounted on magnetic stand in Detector


inclination and
the node of measured plane. Center the laser beam of temperature
rotation laser transmitter AVV-RL20 using the fine
adjust knob on the center of detector window. Use live
X, Y data on the display for beam centering.
The AVV-RL20 laser must not be moved after
measurement has commenced.
First node(indexed [1;1]) is in the left bottom
corner by convention.
The order of taking data for nodes is not important.

Assigned node is indicated on the screen by blinking cursor. Use


arrow keys or alternative arrow keys to
move cursor on the node where detector is placed at. Node data
recorded

Press to record node data.


Repeat for each node in the plane. Nodes within cutouts can be
skipped. Node data not
recorded yet or
The display cursor assigns which node in the plane to place the node is skipped
detector at.

KOHTECT(c) 2013 35
The result display

When data recorded:


press to shift to result display.
use arrow keys or to move cursor
and review node data.
Hold to view plane data and key legend.
Sequentially press then - to setup tolerance value.
Available plane data max, min node value, spread, average,
SD standard deviation.
Press to switch between node data representation - , all
positive and all negative.

Press to switch between plane data representation


absolute, best fit, 3 points reference plane.

BF best fit
Assigning 3 points of reference plane: ABS absolute
REF 3 ref. points
- press to choose REF option.

- press to switch reference point number


(order of switching point #1, #2, #3, quit menu)
- to assign move cursor to node that will be

assigned as reference and press ,


repeat for points #1, #2, #3.

Three of the measurement nodes can be selected as


reference points, which will set them to zero. The
measurement value of the rest of the nodes will then
be recalculated. Reference points can be re-assigned
as required.

KOHTECT(c) 2013 36
Circle shaped plane

Start Flatness program move cursor and press

Choose option to resume interrupted work press . In this


case interrupted work will be opened for further data collecting,
editing, evaluating, saving on the flash drive.

To start new task press .

When new task started the menu screen to choose plane shape

appeared. Press for setting up circular grid

Setup circular grid parameters

- number of rings press , enter number of circles (1 to


10), confirm by
- number of points press , enter number of points (3 to
40), confirm by
- diameters of circles - use arrow keys to select

circle, press , enter diameter, confirm by .

KOHTECT(c) 2013 37
Taking measurements Active Beam X position Beam Y
node indicator data

Place detector unit mounted on magnetic stand in the Detector


node of measured plane. Center the laser beam of inclination and
temperature
rotation laser transmitter AVV-RL20 using the fine adjust
knob on the center of detector window. Use live X, Y
data on the display for beam centering.
The AVV-RL20 laser must not be moved after
measurement has commenced.
First node(indexed [1;1]) is on the top of outer
circle by convention.
The order of taking data for nodes is not
important.

Assigned node is indicated on the screen by Node data Node data not
recorded recorded yet or
blinking cursor. Use arrow keys or node is skipped

alternative arrow keys to move


cursor on the node where detector is placed at.

Press to record node data.


Repeat for each node in the plane. Nodes within cutouts may be
skipped.
The display cursor assigns which node in the plane to
place the detector at.

The result display

When data recorded:


press to shift to result display.
use arrow keys or to move cursor and
review node data.

Hold to view plane data and key legend.


Sequentially press then - to setup tolerance value.
Available plane data max, min node value, spread, average,
SD standard deviation.
Press to switch between node data representation - , all
positive, all negative.

KOHTECT(c) 2013 38
Press to switch between plane data representation absolute, best fit, 3 points reference
plane.

Assigning 3 points of reference plane: BF best fit


ABS absolute
REF 3 ref. points
- press to choose REF option.

- press to switch reference point number


(order of switching point #1, #2, #3, quit menu)
- to assign move cursor to node that will be

assigned as reference and press ,


repeat for points #1, #2, #3.

Three of the measurement nodes can be selected as reference points, which will set
them to zero. The measurement value of the rest of the nodes will then be
recalculated. Reference points can be re-assigned as required.

To save data

Measured data can be saved on the flash drive:


- in the result display menu press and select an existing
folder or create a new one, then press one more time.
Data saved in the file with default name created with date/time
stamp.
Data files can be transferred to PC software for report printing, or
can be opened in the AVV for review.

Quit flatness program

When quit the program:

- press to backup changes in measured data.


- press to discard changes.

KOHTECT(c) 2013 39
4.8 Squareness Program

This programs purpose is to measure perpendicularity of two surfaces. Rotating laser


transmitter RL-20 can deflect laser beam precisely by 90 degree by the means of a pentaprism.
These two perpendicular laser beams are used as a reference. Both the transducers S and M
can be used as detector.
Four readings should be taken two on the one of the surfaces, then laser beam direction
should be switched and another two readings taken on the other
surface.

Run Squareness program move cursor and press

Press to resume an interrupted measurement or

Press to start a new measurement

Set parameters

Enter the distances between the 1st, 2nd and 3rd, 4th

measurement positions. Use arrow keys to move

cursor between distances, then press and enter the

distance in mm confirmed by .

KOHTECT(c) 2013 40
Press to call popup menu where:

Key to set resolution to 2 or 3 decimals

Key to turn ON/OFF data filter

Key to set the tolerance value

Key to go to My document menu

Turn ON the data filter in noisy environment conditions such as


severe vibration, or when laser beam passes through air with
thermal gradients. Such conditions lead to unstable readings. Data
filter improves the stability of readings.

Use key to change number of data blocks (between 2


and 128) used to average the readings. Take into account
that greater value means slower response - more time needed to update readings.

Taking measurements

Place detector in the 1st position. Press to record the


reading

Place detector in the 2nd position. Press to record the


reading Blinking cursor indicates
active measurement
position.
Here 1st position

RL-20 Laser beam


direction for 1st and 2nd
measurement positions

KOHTECT(c) 2013 41
Place detector in the 3rd position. Press to record the
reading

Place detector in the 4th position. Press to record the


reading

The AVV-RL20 laser must not be moved after measurement


RL-20 Laser beam
has commenced. Switch laser beam direction carefully! direction for 3rd and 4th
measurement positions
The order of taking data is not important.

Use arrow keys to choose active


measurement position.

Measurement result

When readings for all four positions are recorded, press key to
go to the result display.

If calculated tilt is out of tolerance value second surface tilt


is displayed graphically.

To save data

Measured data can be saved on the flash drive:

in the result display menu press and select an existing folder or

create a new one, then press one more time.


Data saved in the file with default name created with date/time
stamp.

KOHTECT(c) 2013 42
Data files can be transferred to PC software for report printing, or can be opened in the AVV for
review.

Quit squareness program

When quitting the program:

- press to backup changes in measured data.

- press to discard changes.

KOHTECT(c) 2013 43
4.9 Straightness Program

Press to resume saved task, or press to start new task.

Straightness measurement
The laser alignment system use visible laser beam to measure the straightness of linear
objects. The laser transmitter (transducer that emits laser beam) is mounted at one end of the
object and then aimed in the receiver. Both the transmitter and the receiver are adjusted to
ensure receiver detects the laser. The receiver is moved along the object, and any changes of
the laser beam position in the receiver gives a measure of straightness. The laser beam is not
intended to be the absolute reference. The setup should be such that the laser beam always
strikes the detector throughout the entire length being measured.

Setup parameters

The setup parameters screen is the main screen of straightness program.

Splice number
You should define
overlap points quantity
when splice number
>1(2 or more)

Set measurement points number


To set measurement points number press 1 and enter desired value.

Set dimensions
To set dimensions press to select between all distances equal or different distances for
each segment. Then sequentially define distances use arrow keys to move to next or
previous distance, press and enter desired value confirmed by . Note: when all
distances equal selected system automatically set all distances to entered value, so enter
distance only once.
After all distances defined splice concept(see splice concept) should be followed if splice usage
planned.

KOHTECT(c) 2013 44
Define a splice
When splices planned to be used the splice should be defined. Press and then enter splice
number(the first splice is number 1). Else leave splice none.

Switch between manual data and data from detector

Press to switch between manual entered data and data collected from detector

Laser beam adjustment principle

Before measurement can begin the laser must be positioned such that the beam strikes the
approximate center of the receivers detector over the entire measurement path(whole splice or
whole object).

Adjustment procedure

Place the receiver at the closest(to transmitter) measurement position. Adjust beam position to
center the beam onto the detector. With the receiver at the far end of the measurement path,
adjust the horizontal and vertical angle of the beam to center onto the detector. Repeat above
steps as necessary until the beam remains centered on the detector in both the near and far
positions.

Carrying measurements

To switch to data collect screen press


Place the receiver at the closest measurement position for object(or splice)
Use (if necessary) to select the corresponding point number.

Point Splice number. In splice point


number if no splices then 0 index

If showed means
overlap point

Shows stored values when not in


live move mode, when in live
move mode shows difference
between desired and real object
position

Shows actual Shows actual tilt


values from the from the
transducer transducer

KOHTECT(c) 2013 45
Evaluate measurement results

To enter to results screen when you stands in data collect screen press
Measurement results screen appears
When reference points not selected and best fit
mode not choose measurement results is the plane
stored values

Reference points. - means not defined


Press 4 to enter the reference points, press 6 to
switch to the best fit mode

To select reference points press and then enter points numbers sequentially
(set value and confirm by for each of points) .
Note: overlap point should not be selected as reference!
To deselect reference points enter 255 value for both points
To switch between reference points/best fit plane mode press

When reference points is selected or best fit mode


is a choice measurement results is the difference
between ideal straight line(desired values) and the
real object position

- - means overlap point(appears if splices defined)


this point and its distance is excluded from the
results calculations

The choice is the best fit


mode Press to change
mode

Splice concept

Splices in straightness are partial overlapped parts of whole length of measured object.
Measurement divides by splices for the extension of the total length that can be measured,
where object length exceeds maximum allowed distance between transducers.

To start new task with splices


Press
Enter number of points and then define first splice(enter number 1 for Splice);
You should do it before starting measurement

KOHTECT(c) 2013 46
After you store measurements for all the points belonging to the current splice, you can switch
to the next splice.
To do it: while first point of next splice selected press to return to the straightness main
screen and increase splice number by one. If splice number is not 1 you should define number
of overlapped points too. Number of overlapped points should not be less than 2. To input

overlapped points number press then enter value, confirm by


You should sequentially increase splice numbers, gaps between numbers not allowed.
NOTE: distance between overlapping points must be equal to distances between
points that overlaps see overlapping example below

table s.1 Overlapping example


Overlapping example by Transmit- 1 2 3 4 5 6 7 8 9
table(2 overlapping points, ters first
overlapped segment position
designated by yellow back
color)
Absolute point number
Relative to splice point number X 1. 1. 1. 1. 1. 2. 2. 2. 2.
in format splice.point 1 2 3 4 5 1 2 3 4
Splice 1 Splice 2
Distances 1- 2- 3- 4- 5- 7- 8-
2 3 4 5 6 8 9
For Next splice transmitters X
position(transmitter should be
moved to here)
Distances equality rule: 5-6 equal to 3-4, 7-8 equal to 4-5

Step by step instruction to start measure new splice

Enter to collect data screen. Use for select(if required) the first point of new
splice(in example this is a point 6). Move receiver to its position(regards to currently selected
point; in example this is a point 6, note its distance on object is equal to point 4 distance).
Remember X,Y values, let me label them as V6). Move transmitter from its current position
to position, where point of last splice point minus overlapped points number minus 1 located.
In example this is a place of point 3. Do not touch receiver! Adjust transmitter so that X,Y
values on receiver became almost equal to the X,Y values indicated before the transmitter has
been moved. Move receiver to the closest overlapped point to the transmitter point(exact
position) (in example this is a point 6, note its distance on object is equal to point 4 distance).
Fine tune transmitters position, so that receivers X,Y values became almost equal to
receivers X,Y values that we previously labeled as V6. Repeat adjustment procedure if
needed. Move receiver to the closest overlapped point at the exactly point position (in
example point 6, note its physical place on object is equal to point 4 physical place, but point
number 6 should be selected on the screen). Store measurement data. Then move receiver to
the next overlapped point at the exactly point position (in example point 7, note its distance on
object is equal to point 5 distance, but point number 7 should be selected on the screen). Store
measurement data. Repeat measurements for the all overlapped points. Then continuously

KOHTECT(c) 2013 47
perform measurements for non-overlapped points of the current splice (in example there are
points 8-9).
About to how close should be values for the overlapped point pairs.
In example there are two point pairs 4-6 and 5-7, so how close should be X and Y values within
the pair. The answer is: as close as possible, but complete equality does not required.
AVV-701 system will automatically calculate and apply correction values for each splice, but
care should be taken on values difference, because if difference too big the beam can move
out of the target during measurement.

The Live move mode

After all measurements has been carried and results calculated, it may be necessary to make
corrections in the live move mode. The devices live mode make it possible to monitor these
corrections.
Before switch to the live mode reference points or best fit result mode should be selected.
If you move transmitter from its last position, two point adjustment procedure should be
performed(see adjustment procedure).
Place receiver at point where you intend to make corrections. Press to enter live mode.

String LIVE STOP appears at screens top right side. When you ready to move press to
enable live move detection. The LIVE LIVE start blinking at top right side of screen.
The V and H values at screens center now shows difference between desired and real object

positions. Move object toward zero of each of these values. To stop live correction press
again LIVE STOP appears at screens top right side.

Live move mode is ON

To save data

Measured data can be saved on the flash drive:

in the result screen press and select an existing folder or create


a new one, then press one more time or press to make
new directory or select the existing directory to save.

KOHTECT(c) 2013 48
Data saved in the file with default name created with date/time stamp.

4.10 Cardan program

Cardan Shafts, or offset mounted machines, need to be aligned just as much as an ordinary
mounted machines. The cardan shaft with its ball-joints admits an amount of offset mounting of
the machines, but it does not tolerate angular misalignment. To be able to perform an
alignment of an offset mounted machine you will need to use the optional brackets. See also
horizontal machine alignment.

Two machines joined via a cardan spacer shaft

Flange
Movable machine

Bracket rail

Swivel bracket

1. To compensate offset, mount the bracket rail on the shaft end of the stationary machine.
2. Mount the swivel bracket on the bracket rail. Mount S transducer on the swivel bracket
3. Mount the magnet fixture on the end of the shaft of the movable machine. Mount
M transducer on the fixture.
4. Connect the S- and M- transducer to the display unit and start the Cardan program to
switch on the lasers.

KOHTECT(c) 2013 49
5. Adjust the M transducer beam.

To adjust the laser beam: Turn the transducer into first position e.g. 9
oclock. Note the beam position on the opposite transducer (e.g. beam
hits position 1). Turn the unit half a turn (beam hits position 2). Adjust the
beam to the centre of rotation (beam hits position 3 in the middle between
point 1 and 2).

6. Adjust the brackets rail so the center of the S transducer is hit by the M transducer
laser.
7. Adjust the S transducer beam.

Adjust the bracket rail

KOHTECT(c) 2013 50
To start program press

Press to resume saved task, or press to start new task.

Input dimensions

As offset value is not used in cardan shafts only distance between transducers and distance
between front and rear feet of machine need to be entered
Use or to move from dimension to dimension to input or reconfirm all dimension
data.
To start input dimension use key,
when the blinking cursor appeared use
numeric key pad to enter desired value
then press to confirm.

KOHTECT(c) 2013 51
Collect data
Collect data for the cardan shaft shares many similarities with horizontal machine alignment,

see collect data. To start collect data press The laser beams turned on now.
Screen of wait connection to transducers appears. Note that Bluetooth connection need more
time for connection to be established.
If connection cannot be established during more than 90 seconds check if pin codes entered or
check transducers.

Transducers serial numbers become


available when connection established

After the connection has been established data collect screen appears.
Face the stationary machine from the movable machine.

Blinking outer circle means minimum


turn violation
Use to store point
Up to 36 points can be stored, minimum 3
points should be stored
Minimum turn requirements should be
respected(>10)
NOTE: to achieve acceptable results the
whole shafts turn must not be less than
120 degrees.

Stored point can be edited, or deleted, refer


to horizontal alignment view and edit the
points

KOHTECT(c) 2013 52
Calculate alignment results

When all of the desired points gathered press to go to view calculated results and to make
live adjustment
Note: Screen message shows Do not turn shafts
The alignment measurement result of calculations shows the values of the angular Gap
misalignment in the plane of the coupling (in Horizontal and Vertical planes) as well as the
correction values for the machine feet F1 (machine front feet) & F2 (machine rear feet) on the
movable (M) machine that are necessary for eliminate the horizontal and vertical angular
misalignment.

Sequentially press and to change tolerances, see also


horizontal alignment setup tolerances

To start corrections for the misalignment


To correct the horizontal and vertical plane in biaxial as per obtained coupling result in, press
to switch frozen result screen to LIVE mode with shaft remain at last recording clock
position.

Blinking LIVE at the top right of the screen indicates live adjustment
mode is on
Refer to horizontal alignment section corrections of the misalignment
Note that offset not used for the cardan shafts alignment
Use to switch between live and frozen mode

To save data

When measurement finish, press to enter file save display. Press


again to save the alignment result main directory or to make
new directory or select the existing directory to save.

KOHTECT(c) 2013 53
KOHTECT(c) 2013 54
Quit cardan program
When quitting the program:

- press to backup changes in measured data.

- press to discard changes.

KOHTECT(c) 2013 55
4.11 Spindle program

For spindle alignment S unit(transmitter) mounted in the spindle and the M unit(detector)
mounted at the part of the machine that can be moved along the working area of the machine.

To start program press

Input distances

Press to start input distance between first and second


detector position, use numeric keypad for entering value, confirm

by

Taking measurement

When distance entered press to go to tacking measurement

After the connection established measure points can be stored

KOHTECT(c) 2013 56
S unit Black shows S Unit
inclinometer data label position

M unit placed in first


position now
M unit
values

1 Place detector(M Unit) at the first position


1.1 Turn spindle so that the label of the S Unit positioned on top

Press to store point


1.2 Turn spindle so that the label of the S Unit positioned on
bottom(rotate it 180)

Press to store point


2 Move detector to the second position
2.1 Leave spindle in position so that the label of the S Unit positioned
on bottom

Press to store point


2.2 Turn spindle so that the label of the S Unit positioned on
top(rotate it 180)

Press to store point


Use to walk through measurement position if need

Evaluate results

After all measurements stored result screen appears

Angular error
showed numerically
and graphically

Use to go to remeasure positions if any

KOHTECT(c) 2013 57
To save data

When measurement finish, press to enter file save display.


Press again to save the alignment result main directory or
to make new directory or select the existing directory to save.

To quit program press

KOHTECT(c) 2013 58
5.0 Standards Tolerance of Shaft Misalignment
This chapter provides the standards alignment tolerance of misalignment for standard industrial
machinery with flexible coupling that can be used under condition only if existing in-house
standards or the machine or coupling OEM have not given any blinding values, and must not
be exceeded.

Speed, rpm Good Acceptable


Offset Angular (Gap) Offset Angular (Gap)
Up to 1000 0,08 0,07 0,12 0,10
Up to 2000 0,06 0,05 0,10 0,08
Up to 3000 0,04 0,04 0,07 0,07
Up to 4000 0,03 0,03 0,05 0,05
More than 4000 0,02 0,02 0,04 0,04

6.0 Delivery Set

Description Qty Note


1. Control Display Unit 1
2. Measuring Transducer Unit 2
3. Brackets Fame 2
4. Chains assembly 2
5. Supporting Rods 4
6. Connecting Cable 2
7. AC Charger, 220-230Volts 1
8. Tape Measure 2m 1
9. Carrying Case 1
10. Operating Instructions Manual 1
11. CD-ROM Software 1
12. USB PC Communication Cable 1

KOHTECT(c) 2013 59

S-ar putea să vă placă și