General GC1000
Specifications Ethernet Analyzer Bus System
GS 11B06A01-01E
Hub
Optical Fiber
Interface Unit
(ASIU) Twisted Pair
Wire
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Jun. 01,2006-00
<<Contents>> <<Index>> 3
2.3 Analyzer Server Engineering Terminal 2.5 Analyzer Bus Gateway Unit
Realized by engineering terminal software (ASET) Established by Yokogawas FCJ Autonomous
installed on PC. Controller and analyzer server gateway unit software
PC specifications (ASGW).
Type: IBM PC compatible desktop computer Interfaces with DCS or other host system as a Modbus
Hard disk drive: 10 GB or more free space RTU slave.
Display: SVGA (1024x768) Instrument specifications
OS: Windows 2000SP4, WindowsXP Analyzer bus connection:
Professional Edition SP2 Max. number of connected GC1000 Mark II and
Japanese edition/English edition interface units: 31 sets
Ethernet port: 1 port (100Base-TX or 100Base-FX) DCS connection
Others: CD drive Serial connection: Modbus RTU slave, 2 ports available
Analyzer server engineering terminal software (ASET) Hardware specifications
specifications
Serial communication:
Function: Display of the status of GC1000 Mark II, Function: transmit and receive data.
Analyzer bus interface unit (FCN) and Type: RS232-C
network. Number of port: 2 ports
Display of the results and alarms stored on Protocol: Modbus
the servers (PCAS). Transmission rate:300, 1200, 2400, 4800, 9600,
Operation of GC1000 Mark II and analyzer 14400, 19200, 28800, 38400, 57600,
bus interface unit. 115200 bps
Screen display: a) Overview Data length: 8 bits
b) Analyzer Operation Parity bit: selectable
c) Analysis Result Stop bit: 1 bit
d) Chromatogram Compliant standards
e) Alarm Status
f) LCD Emulator (EtherLCD) EMC standards: CE Mark
Max. number of EtherLCD is 4. Standard for hazardous location equipment:
Analyzer server connection: Non-Incendive
Max. number of ASET connected to one
PCAS: 4 sets Refer to the specifications of FCJ.
ASET is connectable to PCAS on the same Hardware
PC. Model Name Item Remarks
One ASET is activated on one PC.
ASET is not activated with GCET on the NFJT100-S10x FCJ
same PC. T02.EPS
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Jun. 01,2006-00
<<Contents>> <<Index>> 4
GC1000 Mark II through mapping: Specification Item Remarks
- Run command
- Stop command NFBU200-Sxx Back board
- Stream sequence assign NFCP100-S0x CPU
- Calibration (validation) command NFPW441-1x Power card 100-120 VAC
- Stream (continuous) assign NFPW442-1x 220-240 VAC
NFPW444-1x
Interface unit through mapping NFDV151-P10/B5S00 DI card 32 points, 24 VDC
- DO on command NFDV551-P10/D5S00 DO card 32 points, 24 VDC
- DO off command NFDR541-P00/C4S70 Relay output 16 points, 24-110 VDC
(b) Status detection of card /100-240 VAC
GC1000 Mark II and interface unit in common NFAI135 AI card 4-20 mA 8 points
- In operation NFAI143 AI card 4-20 mA 16 points
- Communication error NFFCV01 Dummy cover for I/O card
- Write error NFFCV00 Dummy cover for power card
T05.EPS
GC1000 Mark II
- Analyzer normality/failure Software license
- Change of analyzer alarm status
License Code Item Remarks
- Measurement, stop, or maintenance status
- Progress of stream sequence NT711AJ-LS03E FCN/FCJ basic Required for each
- Rejection of request of stream (continuous) software for ASIU
- Rejection of request of calibration/validation single CPU
- Data update excluding JAVA
- Calibration coefficient update NT8035J-LW11A Modbus Required for each
Portfolio ASIU
- Data validity T06.EPS
- Each alarm condition
Concentration alarm of each peak, retention time Software media
alarm, variation coefficient alarm, tailing coefficient CR-ROM Code Item Remarks
alarm
Interface unit NT203AJ-PC11E Software Loading a license for
- Unit normality/failure media installation
Modification of IP
- Main power supply failure, 24 V DC failure, IO address
module failure T04.EPS
- DI data
- DO data Analyzer server interface unit software (ASIU)
- AI IOP Capable of commanding, detecting status, and reading
IOP occurs when value exceeds 106.3 % of input data as follows.
range or is below -6.3 %. If IOP occurs, the (a) Command
previous value will be held as AI data. - DO on/off command
(c) Data of (b) Status detection
GC1000 Mark II - Unit normality/failure
- Stream number - Main power supply failure, 24 V AC failure,
- First peak number IO module failure
- Number of peak - DI data
- Sampling time - DO data
- Analysis result - AI IOP
- Retention time IOP occurs when value exceeds 106.3 % of input
- Calibration coefficient range or is below -6.3 %. If IOP occurs, the
Interface unit previous value will be held as AI data.
- AI read data (c) Data
(Actual number in 2 words: Range 0.0-1.0) - AI read data
When multiple requests are received in update period, (Actual number in 2 words: Range 0.0-1.0)
the last request is executed.
2.7 Network Components
2.6 Analyzer Bus Interface Unit (a) Hub
Realized by Yokogawa's FCN Autonomous Controller Switching type 100Base hub is recommended.
and analyzer server interface unit software (ASIU).
example; MOXA made P/N:EDS-308-MM-SC
Function: Read and Write I/O interface data every (multi mode, wave length 1310 nm
200 ms. 100Base-FX (SC connector) 2 ports,
Hardware specifications TP (RJ45) 6 ports)
Max. number of contact inputs: 16 (Area classification: class 1 Div 2/Zone 2)
Max. number of contact outputs: 16
Max. number of analog inputs: 16
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Dec. 19,2007-00
<<Contents>> <<Index>> 5
(b) Cable Communication line of engineering terminal PC (ASET
or GCET) is not duplicated
Optical Fiber Cable
It is recommended to use optical fiber cable when
Engineering Analyzer Server
extending distance, wiring between separate buildings Terminal Gateway
or using the device in electrically noisy environments. Unit
Optical fiber cable does not require an Electric energy
blocking-off circuit for explosion proof.
The type of optical fiber is the multi mode which core
diameter is 50/125 mor 62.5/125 m.
GC1000 Mark II has port(s) with 100Base-FX SC
connector.
Twisted Pair Cable
GC1000 Mark II has port(s) with RJ-45 connector for
twisted pair output.
Interface
Use Category 5 or higher of ScTP (Screened twisted GC1000 MarkII GC1000 MarkII Unit F02.EPS
pair) or STP cable. CE Mark is declared on the
condition with ScTP or STP cable. (b) Restrictions on Total Distances of Network
Twisted pair cable requires an Electric energy According to the 100Base-TX standards, cables of up
blocking-off circuit for explosion proof. Refer to section to 100 meters can be used. Total distance can be
2.11, Explosion-proof. extended using multiple levels of switching hubs* in
There are two types of pair cables: straight-through or cascade connections.
cross-over. Generally, a straight-through cable is used *Refer to the switching hub.
to connect between a node and a hub, a cross-over
cable is used to directly connect between nodes. 2.9 Modbus Communication
Realized by the following methods.
(c) Media Converter
Since gateway interface unit (FCJ) and general PC do (a) Connection with analyzer bus gateway unit (FCJ)
not have optical interface. For connection to an optical Analyzer server gateway unit can interface as a
fiber network, media converter interface is required. Modbus RTU slave. Refer to section 2.5, Analyzer Bus
Gateway Unit.
Select the appropriate devices according to the
conditions of infrastructures. (b) Direct connection to GC1000 Mark II
example; MOXA made P/N:IMC-101-M-SC Ethernet communication of GC1000 Mark II and
(multi mode, wave length 1310 nm analyzer server interface unit support Modbus/TCP
100Base-FX (SC connector) 1 ports, client.
TP (RJ45) 1 ports) 2.10 OPC
(Area classification: class 1 Div 2/Zone 2) OPC requires analyzer bus gateway unit (FCJ) and
2.8 Analyzer Bus Network OPC server software.
(a) Redundancy Refer to FCN/FCJ OPC for Windows (GS 34P02Q61-
In a redundant network, if one of the paths fails, 01E).
communication line of analyzed data information is In addition, redundant network function license for FCN/
switched to the other automatically, minimizing the FCJ OPC Server software is required for connecting to
influence. the redundant network.
Redundant network requires analyzer bus gateway unit
2.11 Explosion-proof
(FCJ) and its software (ASGW).
The following conditions should be satisfied.
The following licenses and media are required to
duplicate server PC. (a) Process Gas Chromatograph (GC1000 Mark II)
License Code Refer to the General Specifications of GC1000 Mark II
(GS11B03A03-01E)
Licence Code Item Remarks
(b) Twisted Pair Wiring
NT783AJ-LM11A Redundant network Required for Twisted pair line should be disconnected when the
function license for each PC
explosion-proof status is not kept at the GC1000 Mark II
FCN/FCJ OPC Server
(media:NT203AJ) side.
T07.EPS A 12 V DC output of the analyzer should be used to
Software media turn on the power line of the hub. (Figure 3)
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Jun. 01,2006-00
<<Contents>> <<Index>> 6
Box
Hub
Power
Source
To Network
Relay
Explosion-proof enclosure
GC1000 MarkII
F03.EPS
Figure 3
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Jun. 01,2006-00
<<Contents>> <<Index>> 7
MODEL AND SUFFIX CODES 6. Analyzer Server Interface Unit Software
(ASIU)
1. Process Gas Chromatograph (GC1000)
Model Suffix Code Option Code Description
Any of the following Suffix Codes should be specified ASIU Software package
for Ethernet analyzer bus. Function -A01 Standard
Language E English
Model Suffix Code Option Code Description J Japanese
GC1000h Gas chromatograph -N Always -N
(Note) N Always N
Analyzer bus A Ethernet twisted pair T_ASIU.EPS
dual-channel
B Ethernet optical fiber
dual-channel
C Ethernet twisted pair
single-channel
D Ethernet optical fiber
single-channel
T_GC1000.EPS
Note: For specifications other than GC1000 Mark II
analyzer buss, consult our salesperson or
representative.
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Jun. 01,2006-00
<<Contents>> <<Index>> 8
TERMINAL ARRANGEMENT
1. Process Gas Chromatograph GC1000 Mark II
1.1 Ethernet (Twisted Pair Wiring)
Terminal Number
- Power off signal (A) 3 7 Power off signal (B) -
+ Power off signal (A) 4 8 Power off signal (B) +
F04.EPS
T T Transmit TX
R R Recieve RX
A B
(Lower) (Upper)
F05.EPS
All Rights Reserved. Copyright 2006, Yokogawa Electric Corporation GS 11B06A01-01E 1st Edition Jun. 01,2006-00
Subject to change without notice.