Sunteți pe pagina 1din 18

Re

pub
li
cofS
out
hAf
ri
ca
EDI
CTOFGOVERNMENT
Inor
dertopromotepubl
iceduca
ti
onandpubl
ics
afet
y,equ
aljus
ti
cefora
ll
,
abet
te
ri n
for
me dcit
iz
enry,th
eruleo
flaw,worl
dtra
deandworldpea
ce,
t
hisl
egaldoc
ume n
tisherebymadeav
ail
abl
eonan onc
ommerci
albas
is
,asit
i
stheri
ghtofallhumanstoknowandspe
akthela
wsthatgov
ernthe
m.

SANS 5011 (2005) (English): Water quality -


Determination of pH
ICS 13.060.50
SANS 5011:2005
ISBN 0-626-16592-X Edition 5

ISO 10523:1994
Edition 1

SOUTH AFRICAN NATIONAL STANDARD

Water quality Determination of pH

This national standard is the identical implementation of ISO 10523:1994 and is


adopted with the permission of the International Organization for Standardization.

Published by Standards South Africa


1 dr lategan road groenkloof private bag x191 pretoria 0001
tel: 012 428 7911 fax: 012 344 1568 international code + 27 12
www.stansa.co.za
Standards South Africa
SANS 5011:2005
Edition 5
ISO 10523:1994
Edition 1

Table of changes
Change No. Date Scope

Abstract
Specifies a method for the determination of pH in all types of water and waste water samples in the
range from pH 3 to pH 10.

Keywords
determination, pH, physicochemical properties, quality, reference data, tests, water.

National foreword
This South African standard was approved by National Committee StanSA SC 5140.19A, Water
Water sampling and analysis, in accordance with procedures of Standards South Africa, in
compliance with annex 3 of the WTO/TBT agreement.

This edition cancels and replaces the fourth edition (SABS SM 11:1990).
INTERNATIONAL ISO
STANDARD 10523
First edition
1994-08-01

Water quality - Determination of pH

Qualit6 de Ieau - Determination du pH

Reference number
ISO 10523:1994(E)
ISO 10523:1994(E)

Foreword

ISO (the International Organization for Standardization) is a worldwide


federation of national Standards bodies (ISO member bodies). The work
of preparing International Standards is normally carried out through ISO
technical committees. Esch member body interested in a subject for
which a technical committee has been established has the right to be
represented on that committee. International organizations, governmental
and non-governmental, in liaison with ISO, also take part in the work. ISO
collaborates closely with the International Electrotechnical Commission
(1EC) on all matters of electrotechnical standardization.

Draft International Standards adopted by the technical committees are


circulated to the member bodies for voting. Publication as an International
Standard requires approval by at least 75 % of the member bodies casting
a vote.

International Standard ISO 10523 was prepared by Technical Committee


lSO/TC 147, Water quality, Subcommittee SC 2, Physical, Chemical,
biochemical methods.

Annexes A, B, C, D and E of this International Standard are for information


only.

0 ISO 1994
All rights reserved. Unless otherwise specified, no patt of this publication may be reproduced
or utilized in any form or by any means, electronie or mechanical, including photocopying and
microfilm, without Permission in writing from the publishar.
International Organkation for Standardization
Case Postale 56 l CH-l 211 Geneve 20 l Switzerland
Printed in Switzerland
0 ISO ISO 10523:1994(E)

Introduction
The measurement of the pH-value of water is of great importante for
many types of Sample. High and low pH-values are toxic for aquatic
organisms, either directly or indirectly. lt is the most essential Parameter
used in assessing the corrosive properties of an aquatic environment.
Also, it is important for the effective Operation 0% water treatment pro-
cesses and their control (e.g. flocculation and chlorine disinfection), control
of plumbosolvency of drinking waters and biological treatment of sewage
and sewage discharges.
Various methods of determination, ranging from simple methods using
indicator Paper to sophisticated methods using a pH-meter are used. The
determination of pH may be broadly classified into two classes,
colorimetric and electrometric methods (see [l] and [2] in annex E).
Colorimetric methods employ indicators which develop a range of colours
at different pH. Their accuracy is restricted and they are only satisfactory
for use in a field test.
The electrometric method is based on measuring the electromotive forte
of an electrochemical cell, consisting of the Sample, a glass electrode and
a reference electrode. A Standard deviation of Apl-i = 0,05 or less tan be
obtained using this method. If the Sample is low in ionic strength, i.e. less
than 5 mS/m electrolytic conductivity, special analytical equipment and
procedures are necessary (see 7.3).
This page intentionaliy left blank
INTERNATIONAL STANDARD 0 ISO ISO 10523:1994(E)

Water quality - Determination of pH

1 Scope cent edition of the Standard indicated below. Mem-


bers of IEC and ISO maintain registers of currently
valid International Standards.
1.1 Field of application
ISO 5667-3: -l), Water quality - Sampling - Part 3:
This method is applicable to all types of water and Guidance on the preservation and handling of
waste water samples in the range from pH 3 to samples.
pH 10.

1.2 Interferences 3 Definition

The temperature, some gases and organic materials For the purposes of this International Standard, the
interfere with the pH-measurement. Suspended ma- following definition applies.
terials in the Sample may Cause significant errors
(Suspension effect). Wait for Sedimentation and only 3.1 pH: The negative decadic logarithm of the nu-
insert the electrodes in the clear fraction. Alterna- merical value of hydrogen ion activity expressed in
tively, an Ultrafiltrate tan be taken. When measuring moles per litre. Because of ion interactions, the ac-
sewage and some surface waters, there is a particu- tivity of the hydrogen ions is somewhat lower than
larly high risk of smearing the electrodes or contami- their concentration.
nating the membranes and diaphragms with Oil,
The practical pH-scale is specified in terms of the pH
grease or other contaminants.
of a series of primary Standard reference solutions
In the case of reference electrodes, this tan be pre- (see annex B).
vented (see 5.4, note 2). If precipitation takes place in
the diaphragm, for example silver sulfide or Protein
4 Reagents
flocs, it will be necessary to form an electrolyte bridge
between the Sample and the reference electrode with
inert electrolytes, e.g. c(KN03) = 1 mol/l. 4.1 Water which is free from carbon dioxide
Regular cleaning of the electrodes is necessaty (see Prepare this water from distilled or deionized water
9 .2) . by boiling or by equilibration with nitrogen. This water
is required to dilute the buffer solutions for the double
calibration (see 9.1).
2 Normative reference
The following Standard contains provisions which, 4.2 Standard buff er solutions
through reference in this text, constitute provisions
of this International Standard. At the time of publi- Use solutions B, C, D, F and 1, which are given in an-
cation, the edition indicated was valid. All Standards nex B, or commercially available primaty buffer sol-
are subject to revision, and Parties to agreements utions that are not affected by the growth of
based on this International Standard are encouraged microorganisms. If the solutions are not sterilized,
to investigate the possibility of applying the most re- they are stable for about 6 weeks. Carbon dioxide

1) To be published. (Revision of ISO 56673:1985)


ISO 10523:1994(E) (QISO

from the atmosphere affects buffers with pH-values The cell tan also be a commercially availab Ie combi-
of more than 4. nation e Iectrode (i .e. a mono-rod mea chain).

Buffer solutions which are of greater ionic strength NOTES


than the buffer solutions in annex B are not suitable
3 Calomel electrodes contain mercury and its salts. In the
for measurements in water with a low salt concen-
event of a breakage, the heavy toxic contents enter the
tration (see [9] in annex E).
medium, therefore in general and especially in water used
for drinking or swimming, a silver/silver chloride reference
NOTES
electrode should be used. To prevent clogging the
Commercially available buffer solutions containing mer- diaphragm with silver chloride, a reference electrolyte of
CU ry salts as a preservative should be avoided. c(KCI) = 1,0 mol/1 is recommended.

2 For other buffer solutions, see [6], [IO] and [l l] - in an- 4 Reference electrodes with a flowrate of 0,l ml to 2 ml
nex E. per day are recommended. All new electrodes should be
tested before use and then at regular intervals (see 9.1). To
prevent contamination of the diaphragm, an excess
4.3 Reference electrolyte hydrostatic pressure in the reference electrolyte, equivalent
to a differente in water level of 2 cm or more, is necessary.
Electrolyte Solution to refill the reference electrode
according to the instructions of the manufacturer, e.g. 5 Glass electrode for water with a low conductivity (see
7.3.1).
potassium chloride solutions with the following con-
centrations: 6 Test procedures of electronie units and electrodes (see
[8] in annex E).
1: c(KCl) = 3,5 mol/1 [p(KCI) = 261 g/l]

II: C( KCI) = 3,0 mol/1 [p(KCI) = 224 g/l] 6 Sampling and samples
I11:c(KCI) = 1,O mol/1 [p( KCI) = 74,6 g/l] The pH-value may Change rapidly as a result of
Chemical, physical or biological processes in the water
Dissolve the specified mass of potassium chloride in Sample. For this reason, measure the pH as quickly
water and dilute to 1 litre. If Saturation with silver as possible but not later than 6 h after sampling (see
chloride is recommended, add a few millilitres of sil- ISO 5667-3). If, in particular cases, this is not possible
ver nitrate Solution [c(AgN03) a 1 mol/l]. or not necessary, pass a flexible tube from the sam-
pling tap to the bottom of a sampling bottle and fill the
5 Apparatus bottle to overflowing. Alternatively, rinse the bottle
with the water Sample and immerse it in the Sample.
Fill the bottle, avoiding turbulente if possible. Remove
5.1 Sample bottle, of minimum capacity 500 ml,
all air bubbles from the Sample by gentle shaking and
flat bottomed and made of low alkaline glass, e.g.
stopper the bottle. Analyse as soon as possible, but
borosilicate glass. Plastics bottles shall be imper-
not later than 24 h after sampling. Avoid changes in
meable to gases.
temperature and gas exchange with the atmosphere.

5.2 Thermometer, with a 0,5 C scale.


7 Procedure
5.3 pH-meter, with an input impedance greater than
lO* SZ, and with temperature compensation facilities 7.1 Preparation
and slope correction in millivolts per pH-value. The
pH-meter shall be capable of being read to a dis- Measure the temperature of the buffer solutions (see
crimination of a pH of 0,Ol or better. [l4] in annex E). Set the temperature control on the
pH-meter. In accordance with the instruction of the
5.4 Glass electrode and reference electrode. The manufacturer, consider the isopotential Point. If poss-
glass electrode measuring System shall have a zero ible, buffers and samples should have the same tem-
voltage between pH = 6,5 and pH = 7,5, using a refer- perature.
ence electrode similar to the reference electrode in-
side the glass membrane (symmetrical electrode 7.2 Calibrating and direct measurement
System) (see [12] and 6131 in annex E). The mem-
brane resistance of the glass electrode shall be Measure the temperature of the Sample (see [14] in
Pi,< 1ogc2. annex E) and reset the temperature control on the
0 ISO ISO 10523:1994(E)

pH-meter in accordance with the manufacturers in- 8 Calculation of pH for various


structions.
temperatures
Wash the electrode with water and with Sample and
immerse it in the Sample. Swirl the Solution and read The pH shall be referred at 25 C. If the value is
the pH-value without stirring. measured at another temperature, this shall be stated.
lf required, express the pH for a temperature other
Rinse the electrode with water and immerse it in than that measured, referring to figure 1 and using the
water to remove traces of Sample or buffer solution. following calculation:

Wash the electrode with water and immerse it in @-b, = PH,, + ApH,,
buffer D (see 4.2). Swirl the Solution gently about the
electrode and allow the Solution to come to rest where
(reading without stirring). Set the zero Point of the in- is the pH at 25 C;
PH25
strument at the pH-reading of buffer D. The zero Point
compensation shall be apH < 0,5 (see clause 5). If PH tm is the pH at the measured temperature;
this is not the case, then one of the electrodes is de-
fective. Rinse the electrode with water and immerse APHtm is the pH-deviation with reference to
in buffer C or F (see 4.2). Swirl the Solution and take 25 C for the measured temperature
the pH-reading without stirring. Adjust the slope factor (see figure 1).
control until the reading that matches the buffer value
NOTE 8 The calculation is valid for waters with a buffer
at the measured temperature is obtained. (For the
capacity mainly due to hydrogen carbonate ions.
permissible range of slope factors, see tableA.1).

Repeat both Steps of the calibration until no further


improvements are obtained.
9 Checking of the accuracy and
NOTE 7 Buffer B, C, F or I should be Chosen so that the maintenance of the equipment
expected pH of the Sample lies within the range of the
buffers used.
9.1 Checking the equipment
7.3 Measurement in water with a low
Use the following test for screening the reference and
conductivity
glass electrodes for unsatisfactory Performance. lt
pH-measurements in water with low ionic strength shall be applied to new electrodes, to electrodes that
(conductivity less than 5 mS/m) and low buffer ca- have not been used for 2 weeks or more, and to
pacity need special care. Examples of such waters are electrodes in regular use at intervals of 1 or 2 weeks.
deionized condensate and rainwater.
Calibrate the electrodes using one of the primary
Standard solutions in table B.1. This value is PH(S).
7.3.1 Equipment Then dilute the same buffer with CO,-free water to
double the volume. The observed deviation ApH,
Use a commercially available glass electrode with a which exceeds the dilution effect Apl-l,,, given in
glass membrane of low solubility, i.e. a high alkali table B.1, is the liquid junction error ApHj (double cali-
electrode. These electrodes are recommended for all bration):
measurements in waters. lt is necessary to take a
reference electrode with a sleeve ground joint and a APHj = ApH - ApH,,,
reference electrolyte with c(KCI) = 1 mol/l. Avoid the
influence of potassium chloride on the glass electrode where
by slowly stirring the Solution. To avoid the influence
of air, allow patt of the Sample to flow through the ApH,,, is the pH of the diluted Standard Solution.
measuring cell. To eliminate electrostatic effects, the
Stir the Solution gently and, while still stirring, note
Sample shall be shielded by a Faraday tage and
the new stable PH,. The shift on stirring, ApH,, is
grounded by a metallic electrode in the Sample.
calculated as follows:

7.3.2 Calibration ApH, = pH, - PH(S) - ApHd

Check the calibration using one of the Standard sol- where PH(S) is the pH of the primary Standard sol-
utions of low buffer capacity given in annex C. ution.
ISO 10523:1994(E)

If the Signal fluctuates, estimate the differente, 11 Standard deviation


ApH,, between the maximum and minimum readings.
Use the list in table 1 for assessing the Performance
of the electrode.

Table 1 The Standard deviation of the repeatability obtained


from 10 replicate measurements with drinking water,
Performance performed in two laboratories, is Apl-i = 0,Ol.
characteristic
Systematic deviations up to ApH = 0,05 may be
Liquid junction error ApHj \< 0,03 < 0,05 caused by unavoidable interferences, e.g. in low con-
Shift on stirring r\pH, \< 0,005 < 0,02 ductivity water.
Noise apH, < 0,005 < 0,02
A collaborative test with synthetic samples of low
conductivity (C) Shows that a Standard deviation of
9.2 Maintenance of the electrodes ApH = 0,05 or less is achievable. The laboratories met
the target of ApH = 0,2 total error.
After an extensive series of measurements, it is
necessary to clean the electrode System (measuring NOTE 9 The relative error is: -Q AC/C = 10 ApH- 1. A
cell) when the pH of a Standard Solution requires a deviation of APH = 0,05 means a factor of 1,12 and an error
longer equilibrium period. Clean the electrodes regu- of 12 % in estimating the hydrogen ion activity.
larly (daily or weekly) by wiping carefully with cellulose
tissues, or, if the samples contained organic impuri-
ties, with ethanol (70 %), acetone or warm detergent
solution. Remove calcite with dilute hydrochloric acid.
Use organic solvents for a short time only, because
they decrease the sensitivity of glass electrodes by
dewatering the leached layer of the glass membrane.
If organic solvents are used, the electrode shall be
recharged by standing in water for several hours. Re- 12 Test report
calibration is essential.

Observe the manufacturers instructions regarding


storage and maintenance. Reference electrodes and
combined electrodes should be stored in the solutions The test report shall contain the following information:
which are also used as the reference electrolyte.
Glass electrodes should be stored in water. a) a reference to this International Standard;
Electrodes used for low conductivity water shall not b) precise identification of the Sample, Source lo-
be used for other samples, e.g. sewage. cality, time, sampling procedure and time elapsed
between sampling and measurement;
IO Expression of results
c) expression of the results according to clause 10;
Report the pH to two decimal figures.
d) any deviation from this method and a description
Report the temperature at which the measurement is
of any circumstances which may have influenced
carried out to the nearest degree. the results;
EXAMPLES
e) in addition, for rainwater: duration of Sample col-
pH 8,45, measured at IO,2 C. lection.

pH 7,62, measured at 16,4 C in the laboratory


2 h after sampling.
0 ISO ISO 10523:1994(E)

30 40
Temperature,'C

Figure 1 - Graphit temperature correction


OS0 10523:1994(E)

Annex A
(informative)

Calculation of the pH-value

The pH-value of the Sample is estimated as follows:


Table A.l - Variation of dope factor with
PH = PH(S) + (W-SI - U)(mcI,&q temperature
where

PHI31 is the pH-value of the Standard buffer


Solution;

WS1 is the electrode voltage of the glass


electrode cell in contact with the
Standard buffer Solution;

u is the voltage of the same electrode


cell in contact with the Sample;

F is the Faraday constant, in coulombs


per male F = 9,648 x 1O4 C/mol) ;
(
R is the molar gas constant, in joules per
mole kelvin [R = 8,314 J/(moLK)];

T is the temperature of the Sample, in


kelvins;

F/log,$T is the slope factor of an ideal


electrode and depends on tempera-
ture (see table A. 1).

The voltage U is the sum of the number of contribu-


ting Potentials, including those from the glass and
reference electrodes, a liquid junction and the
asymmetry potential, and depends on temperature.
Therefore, the samples and Standard buffer Solution
should preferably be at the same temperature.

The junction potential depends on the ionic content


of the Sample and the reference electrode.

Stirring the Sample changes the reading, the correct


value (without stirrinq) should be noted.
!

0 ISO ISO 10523:1994(E)

Annex B
(informative)

Primary pH Standard reference solutions

Table B.l - Values of PH(S) for primary Standard reference solutions (See [6] in annex E)
B C D F I
Temperature
Potassium Potassium Sodium
0
C hydrogen hydrogen Phosphate Borax carbonate/sodium
tartratel) phthaiate*b hydrogen carbonate
0 4,000 6,984 9,464 10,317
5 3,998 6,951 9,395 10,245
IO 3,997 6,923 9,332 10,179
15 3,998 6,900 9,276 10,118
20 4,001 6,881 9,225 10,062
25 3,557 4,005 6,865 9,180 10,012
30 3,552 4,011 6,835 9,139 9,966
35 3,549 4,018 6,844 9,102 9,926
40 3,547 4,027 6,838 9,068 9,889
45 3,547 4,038 6,834 9,038 9,856
50 3,549 4,050 6,833 9,011 9,828
c(mol/l) 0,027 0,016 0,029 0,020 0,029
APh /2 + 0,049 + 0,052 + 0,080 + 0,Ol + 0,079

1) The solubility of potassium hydrogen tartrate is lowered by cooling and cannot be used at a temperature lower than
25 "C. (See [15] in annex E.)
2) This Solution is called reference value pH Standard". (See [l l] in annex E.)

The primary Standard reference solutions should al- Dry the anhydrous salts at 120 C for 2 h. Dissolve
ways be prepared using water which is free from 4,45 g of disodium hydrogen Phosphate dihydrate
carbon dioxide (4.1). The following methods should (Na,HP0,.2H,O) (or 3,55 g Na,HPO,) and 3,40 g of
be used for solutions B, C, D, F and 1. potassium dihydrogen Phosphate in water at 25 C
and dilute to 1 litre with watet-. The anhydrous salts
Solution B: Potassium hydrogen tartrate, are dried at 120 C.
b (KHC4H403) = 0,214 mol/kg.
Solution F: Borax,
Add 30 g of potassium hydrogen tartrate to about b(Na2B407.1 OH,O) = 0,Ol mol/kg.
1 litre of water in a stoppered bottle and Shake for
20 min at 25 C & 1 C. Allow the bottle to stand and Dissolve 3,81 g of borax in water at 25 C and dilute
decant the clear Solution. Discard when growth ap- to 1 litre in a volumetric flask.
pears.
Solution 1: Sodium carbonate, w,CO,)
Solution C: Potassium hydrogen phthalate, = 0,025 mol/kg, and sodium hydrogen carbonate,
b[C6H&oCH)(CoCK)] = 0,05 mol/kg. b(NaHC0,) = 0,025 mol/kg.

Dissolve IO,21 g of potassium hydrogen phthalate, Dissolve 2,640 g of sodium carbonate, that has been
that has been dried for 2 h at 120 C, in water at dried for 90 min at 250 C, and 2,092 g of sodium
25 C and dilute to 1 litre in a volumetric flask. hydrogen carbonate, that has been dried for 2 d over
a molecular sieve, in water and dilute to 1 litre in a
Solution D: Disodium hydrogen Phosphate, volumetric flask.
&(Na,HPO,) = 0,025 mol/kg and potassium dihydro-
gen Phosphate, b(KH,PO,) = 0,025 mol/kg.
ISO 10523:1994(E)

Annex C
(informative)

Low buffer capacity Standard solutions

Cu-w4> = 0,05 mmol/l

This Solution has a pH of 4,005 at 0 C to 30 C.

c(NaHC0,) = 0,05 mmol/l, air equilibrated Cp(C02) = 316 mbar].

Temperature, C 10 20 30 40 50

PH 6,93 6,99 7,06 7,14 7,20

Aerate the buffer Solution by aeration with an aquarium Ventilation pump to achieve the correct partial pressure
of coz.
s

0 ISO ISO 10523:1994(E)

Annex D
(informative)

Synthetic rainwater

Table D.l - Synthetic rainwater produced by mixing samples of aqueous stock solutions (see [16] in
annex E)
Stock solution Control sample
Amount of
Reagent Mass concentration substance ml of stock solution in 5 litres of water
concentration

mg/1 m mol/1 Sample 1 Sample 2

H,SQa 50,o 2,500 7,500 *


CaCl,.ZH,O 920 6,26 0,30 1 ,oo
NaNO, 2 456 28,9 1 ,oo 2,00
NH,CI 1 497 28,0 1,500 0
KN03 650 6,43 1 ,oo 2,00
MgS04.7H20 1 026 4,16 1 ,oo 2,00
(NH412S04 9170 64,4 0 2,00
HNO, 100,o 0,050 5,000
HCI 50,o 0 2,500
NaF 588 l4,O 1 ,oo 2,00

pH (25 C ) 4,30 f 0,02 3,59 f 0,02

Conductivity, pS/cm 26 f 2 130 f 2


ISO 10523:1994(E)

Annex E
(informative)

Bibliography

[l] BATES, R. G. Determination of pH, J. Wiley & Cl01 COVINGTON,A. K., WHALLEY, P. D. and DAVISON,
Sons, New York (1973). W. Recommendations for the determination of
pH in low ionic strength fresh waters, (IUPAC)
[2] GALSTER, H. pH-Measurement, VCH-Verlag, Pure Appl. Chem. 57 (1985), pp. 877-886.
Weinheim (1991).
Cl11 COVINGTON,A. K., BATES, R. 6. and DURST, R. A.
[3] DIN 38404 Part 5:1984, German Standard Definition of pH scales Standard reference val-
methods for examination of water, waste water ues, measurement of pH and related termin-
and sludge; ph ysical and physical-Chemical par- ology, Pure & Appl. Chem. Vol. 57 (1985),
anietet-s (group C); determination of pH-value pp. 531-542.
(C5) .
Cl21 DIN 19263:1989, Glass electrodes for
[4] D IN 19260: 197 1, p H-measurement; definition p H-measuremen t.
of general terms.
cw DIN 19264: 1985, pH-measurement; reference
[S] DIN 19261:1971, pH-measurement; terms and electrodes.
definitions for methods of measuremen t using
galvanic cells. CW DIN 38404 Part 4:1976, German Standard meth-
ods for analysing wa ter, waste wa ter and
[6] DIN 19266:1979, pH-measurement; Standard sludge; ph ysical and ph ysical-Chemical par-
buffer solutions. ameters (Group C); determina tion sf tempera-
ture (C4).
[7] IEC 746-1 :1982, fxpression of Performance of
electrochemical analyzers. Part 1: General. m1 BAUCKE, F. G. K. Low Temperature Limit of NBS
(DIN) pH Standard Buffer Solution Potassium
[8] IEC 746.2:1982, Expression of Performance of Tetroxalate, Electrochim. Acta 24 (1979),
electrochemical analyzers. Part 2: pH value. pp. 95-97.

[9] DAVISON, W., COVINGTON,A. K. and WHALLEY, P. Cl61 VDI-Richtlinie 3870 Blatt 10 Entwurf Mai 1988:
D. Conventional residual liquid junction poten- Messen von Regeninhaltsstoffen; Messen des
tials in dilute solutions, Ana/. Chim. Acta 223 PH-Werts in Regenwasser. VDI-Kommission
(1989), pp. 441-447. Reinhaltung der Luft, Dsseldorf.

Standards South Africa

Copyright protected. This standard was downloaded on 2012-08-22 18:22 by Mr. Carl Malamud of Public.Resource.Org.
10 Single-user license only; copying and networking prohibited.

S-ar putea să vă placă și