Friedrich Hofmann

Principles of Electromagnetic Flow Measurement


As head of development, Friedrich
Hofmann (DipEng) was responsible
for the first generations of EMFs
featuring pulsed direct current
fields.
In his more than 30 years with
KROHNE, he had a significant
influence on the development of
electromagnetic flowmeters
from expensive, high-maintenance
specialised devices to reliably
functioning, maintenance-free
standard measuring devices.

Principles of Electromagnetic Flow Measurement

KROHNE Messtechnik GmbH www.krohne.com


Ludwig-Krohne-Str. 5
47058 Duisburg
Deutschland
Tel.: +49 203 301 0
Fax: +49 203 301 103 89
info@krohne.de
www.krohne.com
Principles of Electromagnetic Flow Measurement

Friedrich Hofmann 2011


Imprint
2. Edition 2011

Copyright
KROHNE Messtechnik GmbH
Ludwig-Krohne-Str. 5
47058 Duisburg
info@krohne.de
www.krohne.com

Printed in Germany
Foreword

In many industries, measuring devices However, when it comes to the plan-


and sensors are an important part of ning, actual selection and use of the
process control technology and in equip- devices and materials, the respective
ping automated plants. The significance device manufacturer's information and
of flow measurement is reflected in the documentation is always binding.
variety of measuring methods specifi-
cally designed for this purpose.
Friedrich Hofmann
Because of its many advantages, the
principle of electromagnetic flow meas-
urement is especially important.

This book is intended to serve as a gen-


eral source of information when it comes
to the use and technology of electro-
magnetic flowmeters (EMF). To start off,
the reader is introduced to the basics
of the electromagnetic flow measur-
ing principle, including its advantages
and limitations. Further explanations
regarding the theory follow.

This book is also beneficial because of


the wealth of information it provides re-
garding the selection and use of elec-
tromagnetic flowmeters.

Many of the recommendations made are


based on the author's years of practical
experience as well as on the feedback of
customers in the KROHNE Group.

V
Contents

List of figures....................................................................................................... X

List of tables..................................................................................................... XIV

KROHNE - The Company................................................................................. XVII

KROHNE EMF Highlights, a world's first.......................................................XVIII

Chapter 1............................................................................................................ 21
The fundamentals of flow measurement

Chapter 2............................................................................................................ 25
The fundamentals of electromagnetic flowmeters
2.1. Introduction................................................................................................. 26
2.2. Measuring principle.................................................................................... 27
2.3. Advantages.................................................................................................. 29
2.4. Primary head............................................................................................... 30
2.5. Signal converter.......................................................................................... 34

Chapter 3............................................................................................................ 37
EMFs for fully filled pipelines
3.1. Advantages.................................................................................................. 40
3.2. Limits.......................................................................................................... 44
3.3. Maintenance................................................................................................ 44
3.4. Testing......................................................................................................... 46
3.4.1. Testing accuracy test with a calibration rig.......................................... 46
3.4.2. Testing and documentation with testing equipment............................ 46
3.4.3. Intelligent self-diagnostics................................................................... 48

Chapter 4............................................................................................................ 53
EMFs in special areas of applications
4.1. Explosion protected versions...................................................................... 55

VI
2 Contents

4.2. EMFs for custody transfer.......................................................................... 59


4.3. EMFs in environments with strong magnetic fields................................... 60
4.4. Measuring pulsating flows using EMFs...................................................... 62
4.5. EMFs for products with high solids content............................................... 64
4.6. EMFs without electrodes with capacitive signal pick-up........................... 70
4.7. EMFs for partially filled pipelines............................................................... 72
4.8. Battery-operated EMFs for remote measuring points............................... 74
4.9. Two-wire EMFs............................................................................................ 79
4.10. EMFs for rapid volumetric filling.............................................................. 83
4.11. EMF probes............................................................................................... 91

Chapter 5............................................................................................................ 95
Electromagnetic flowmeters planning
5.1. Analysing the range of application and the operating conditions.............. 96
5.2. Device selection........................................................................................ 103
5.3. Installation................................................................................................ 110
5.3.1. Primary head....................................................................................... 112
5.3.2. Signal converter.................................................................................. 115
5.3.3. Signal cable......................................................................................... 117
5.4. Grounding . ............................................................................................... 118
5.5. Surge Protection....................................................................................... 119

Chapter 6.......................................................................................................... 121


Quality assurance at KROHNE
6.1. KROHNE burn-in test................................................................................ 122
6.2. KROHNE calibration.................................................................................. 124

Chapter 7.......................................................................................................... 129


The fundamental principles of fluid mechanics
7.1. Flow measurement................................................................................... 130

VII
Contents

7.2. Basic laws of fluid mechanics.................................................................. 131


7.2.1. Continuity equation............................................................................. 131
7.2.2. Principle of energy conservation........................................................ 132
7.3. Flows......................................................................................................... 134
7.3.1. Laminar flows..................................................................................... 134
7.3.2. Turbulent flows................................................................................... 136
7.4. The Reynolds number Re.......................................................................... 137
7.5. Pressure loss in incompressible flows..................................................... 138

Chapter 8.......................................................................................................... 143


The theory of electromagnetic flowmeters
8.1. Measuring principle.................................................................................. 144
8.2. Physical background................................................................................. 146
8.3. Signal generation and processing............................................................ 150
8.3.1. Internal resistance.............................................................................. 150
8.3.2. AC and DC field .................................................................................. 150
8.3.3. Signal cables with bootstrap.............................................................. 154
8.4. Signal converter in detail.......................................................................... 157
8.5. Flow profiles.............................................................................................. 160
8.6. Empty pipe detection................................................................................ 167
8.7. Grounding . ............................................................................................... 170
8.7.1. Classic grounding methods grounding the process liquid.............. 170
8.7.2. Newer grounding methods without grounding the process liquid.... 172
8.8. Surge Protection....................................................................................... 175
8.9. Norms and standards............................................................................... 179

Chapter 9.......................................................................................................... 183


Summary and outlook

References....................................................................................................... 186

Index................................................................................................................. 194

VIII
List of figures

Fig. 1: Compact EMF sandwich version (KROHNE, OPTIFLUX 1300C)....... 26


Fig. 2: Measuring principle of electromagnetic flowmeters.......................... 27
Fig. 3: Structure of an EMF............................................................................ 30
Fig. 4: World's first EMF with 3 m nominal size (KROHNE M960, 1983)....... 31
Fig. 5: EMF sandwich version (KROHNE, OPTIFLUX 1000)......................... 32
Fig. 6: EMF hygienic version with weld-in connections
(KROHNE, OPTIFLUX 6000)................................................................. 33
Fig. 7: L-shaped gasket with trapezoidal sealing lip..................................... 33
Fig. 8: KROHNE compact EMF, flanged version,
with IFC 300 (OPTIFLUX 4300C) signal converter............................... 35
Fig. 9: KROHNE compact EMF, flanged version,
with IFC 100 (OPTIFLUX 2100C) signal converter............................... 35
Fig. 10: KROHNE's modular EMF product line, signal converters.................. 38
Fig. 11: KROHNE's modular EMF product line, primary heads....................... 39
Fig. 12: EMF limits of error at reference conditions
(KROHNE, OPTIFLUX 5300)................................................................. 40
Fig. 13: EMF measuring tube featuring a stainless steel reinforced PFA liner
(KROHNE, OPTIFLUX 4000) ................................................................ 41
Fig. 14: Effects of the flow profile on EMF measuring accuracy,
here using a WIB-SIREP-E-1747-S-94 test with a
KROHNE EMF with a nominal size of DN 50....................................... 42
Fig. 15: EMF test equipment (KROHNE, MagCheck)....................................... 46
Fig. 16: Monitoring of general error modes
using the measuring electrodes.......................................................... 49
Fig. 17: Magnetic field with normal field coil polarity
for flow measurement......................................................................... 50
Fig. 18: Measuring the flow profile by reversing the polarity
of the upper and lower field coils........................................................ 50
Fig. 19: The KROHNE specialists for special applications . ............................ 54
Fig. 20: XY metrology marking......................................................................... 60

X
List of figures

Fig. 21: Example of the measurement of a pulsating flow,


pictured here at the outlet of a diaphragm pump............................... 62
Fig. 22: Wear on pipe connections during hydraulic transport
and edges protruding into the pipeline............................................... 67
Fig. 23: Protective ring to protect the liner
of the EMF during hydraulic transport................................................ 67
Fig. 24: Flow and concentration profile when
hydraulically transported in a horizontal pipeline.............................. 68
Fig. 25: Flow patterns during hydraulic transport in long,
perpendicular pipelines....................................................................... 69
Fig. 26: Structure of an EMF with capacitive signal pick-up........................... 70
Fig. 27: KROHNE TIDALFLUX 4110PF, the worlds first EMF for partially
filled pipelines featuring non-wetted level measurement................... 72
Fig. 28: Measuring principle of an electromagnetic flowmeter
for partially filled pipelines (KROHNE, TIDALFLUX 4110PF).............. 73
Fig. 29: KROHNE WATERFLUX 3070C.............................................................. 75
Fig. 30: WATERFLUX 3070C measuring tube................................................... 75
Fig. 31: WATERFLUX 3070C battery................................................................. 77
Fig. 32: Error limits for water meters in accordance
with measuring instrument directive 2004/22/EC,
Appendix MI-001 and OIML R49.......................................................... 78
Fig. 33: 070C Signal converter for the WATERFLUX 3000................................ 79
Fig. 34: Functional principle of multiwire EMFs.............................................. 80
Fig. 35: Functional principle of two-wire EMFs............................................... 81
Fig. 36: KROHNE BATCHFLUX 5015C............................................................... 83
Fig. 37: Volumetric filling with KROHNE BATCHFLUX EMF............................ 84
Fig. 38: KROHNE BATCHCONTROL 5014C....................................................... 85
Fig. 39: Comparison of the wiring involved with a filling machine
with a classic EMF and with one BATCHCONTROL............................. 86
Fig. 40: Record of a filling process "valve open"........................................... 88

XI
List of figures

Fig. 41: Results of an independent test investigating the long-term stability


of EMFs with PTFE and ceramic measuring tubes, size DN 15,
after continuous stress (e.g. 600 changes 18C / 80C +
hot acid 80C + superheated steam 140C)........................................ 90
Fig. 42: Electromagnetic flow probe (KROHNE, DWM 2000)........................... 92
Fig. 43: Reduction of the gas volume by increasing pressure,
relative gas-volume content standardized to 1 at pabs = 1 bar.......... 100
Fig. 44: Reducer upstream and downstream of the EMF.............................. 103
Fig. 45: Compact version of a signal converter (KROHNE, IFC 300C)........... 109
Fig. 46: Wall-mount version of a signal converter (KROHNE, IFC 300W) . ... 109
Fig. 47: Possible measurement errors when inlet runs too short................ 112
Fig. 48: EMF in compact version (KROHNE, OPTIFLUX 4300C)..................... 115
Fig. 49: EMF remote version (KROHNE, OPTIFLUX 4300W).......................... 115
Fig. 50: Service life of electronics depending on the temperature................ 116
Fig. 51: Burn-in test using the KROHNE IFC 300........................................... 123
Fig. 52: Setup of KROHNE Altometer calibration rig for EMFs
up to DN 3000.................................................................................... 125
Fig. 53: Calibration certificate (KROHNE, OPTIFLUX 2000)........................... 126
Fig. 54: Calibration rig (KROHNE, Netherlands)............................................ 127
Fig. 55: Explanations regarding the basic laws of
fluid mechanics................................................................................. 133
Fig. 56: Example of layers of fluids................................................................ 135
Fig. 57: Laminar flow...................................................................................... 136
Fig. 58: Turbulent flow.................................................................................... 136
Fig. 59: Flow resistance.................................................................................. 138
Fig. 60: Measuring principle of electromagnetic flowmeters........................ 145
Fig. 61: x-component of valence vector W according to Shercliff
on the electrode level........................................................................ 148
Fig. 62: Typical time characteristic of field current and electrode
voltage in an EMF with pulsed dc field.............................................. 153

XII
List of figures

Fig. 63: Line capacitance and signal distortion in longer signal cables
without single conductor shields...................................................... 154
Fig. 64: Line capacitance and signal distortion for signal cable with
shielding (Bootstrap)......................................................................... 155
Fig. 65: Block diagram of an EMF signal converter (KROHNE, IFC 300)....... 158
Fig. 66: Smoothing the flow profile by reducing
the measuring cross section (PTB Berlin, 1991).............................. 160
Fig. 67: Measurement errors downstream of a partially open
valve in inlet runs of 1D and 5D......................................................... 161
Fig. 68: Distorted flow profile downstream of an elbow................................ 163
Fig. 69: Distorted flow profile downstream of a sudden expansion............... 163
Fig. 70: Swirl flow caused by an out-of-plane double bend........................... 164
Fig. 71: Grounding the process liquid in metal pipelines
not coated on the inside.................................................................... 170
Fig. 72: Grounding the process liquid in pipelines
with electrically insulated walls........................................................ 171
Fig. 73: Virtual grounding via the device electronics when installing
the EMF in pipelines with insulated walls......................................... 173
Fig. 74: Surge Protection in a compact EMF version..................................... 177
Fig. 75: Surge Protection in a remote system EMF version . ........................ 178

XIII
List of tables

Table 1: Diagnostic requirements for EMFs as per


VDI / VDE 2650, Sheet 3................................................................... 47
Table 2: KROHNEs IFC 300 diagnostic tools which exceed
VDI / VDE 2650 requirements.......................................................... 48
Table 3: Preconditions for an explosion........................................................ 56
Table 4: Equipment categories for gas and dust atmospheres.................... 57
Table 5: Classification of hazardous operating areas................................... 58
Table 6: Recommendations for selecting the sizes
of BATCHFLUX and BATCHCONTROL
depending on filling volume and flowrate....................................... 89
Table 7: Examples of process liquids in different industries....................... 96
Table 8: Selection criteria for the design of EMF devices.......................... 105
Table 9: Overview of measuring tube and liner materials for EMFs.......... 106
Table 10: Suitable electrode materials depending on process liquid.......... 108
Table 11: Criteria for selecting signal converter version............................. 110
Table 12: Suggestions for selecting a suitable
installation site for the primary head . ......................................... 111
Table 13: Determining the pipe friction coefficient
in turbulent flows.......................................................................... 140
Table 14: Smoothing effect of a reducer
(KROHNE, OPTIFLUX 5000)........................................................... 162

XIV
KROHNE The Company

Ludwig Krohne & Sohn in Duisburg From an early stage, KROHNE focused
started manufacturing variable area on measuring devices for industrial use.
flowmeters in 1921. Both the company history and being situ-
ated in one of the most industrialised
The variable area flowmeters that came
regions in Europe justified KROHNE's
later made it possible to adjust the gas/
choice of industrial measurement tech-
air mixture for optimal combustion.
nology as its core business. It all started
The small family business survived the with variable area flow measurement.
world financial crisis from 1928 to 1932
KROHNE was the first manufacturer to
and the Second World War.
successfully calculate and form cylindri-
Kristian Rademacher-Dubbick, a nephew cal glass tubes into tapered measuring
of company founder Ludwig Krohne, tubes. The tolerances were sufficiently
took over the family business in 1948. reduced to make the measuring tubes
At the time the company had only 12 interchangeable. The calculation method
employees. for glass cones developed by KROHNE
went on to form the basis for the VDI/
Subsequent years saw intensive develop-
VDE 3513, Sheet 2 directive. This method
ment in research and sales strategies.
allowed the flow display to be converted
The once small family business grew into
as the properties of the process liquid
one of the world's leading manufactur-
changed.
ers of measuring devices:
In 1962, KROHNE took over the company
KROHNE Messtechnik.
"Alto", previously known as "Tobi", which
was based in Rotterdam and renamed it
KROHNE Altometer. At the time Tobi or
Alto was the world's first manufacturer
of electromagnetic flowmeters (EMF)
for industrial use.

XVI
KROHNE The Company

Over time, KROHNE expanded into many Level measurement


other areas of industrial measurement In 1989, KROHNE introduced the world's
technology including: first non-contact radar level measure-
ment system (FMCW) for process tanks.
Ultrasonic flow measurement
The BM 70 made it possible to measure
EMFs require electrically conductive
the precise level in the presence of tur-
liquids. In 1983, KROHNE developed
bulent liquid surfaces, stirring devices
the world's first ultrasonic flowmeter
and built-in components in the tank as
(UFM) for non-conducting media featuring
well as when dust, steam or mist are in
two parallel measuring beams. These
the atmosphere.
devices exhibit smaller measurement
error during changes in the flow profile Expanding the product range
than single beam UFM. In 1999, with the Over recent years, KROHNE has expanded
5-beam UFM ALTOSONIC V, KROHNE its range of products to include sensors
attained the world's first approval for for pressure, temperature and analy-
a UFM suitable for the custody transfer sis.
of all liquids other than water. This ap-
KROHNE Oil & Gas
proval guarantees a measurement error
KROHNE Oil & Gas develops and sup-
of less than 0.2% across a measuring
plies products, systems and engineering
span of 10:1.
services for bulk measurement, man-
Coriolis mass flow measurement agement and monitoring.
KROHNE has been setting the bench-
KROHNE Water Solutions
mark with Coriolis mass flowmeters
KROHNE Water Solutions specialises
(CFM) as well. In 1986 the CORIMASS
in complete measurement solutions for
S was the world's first CFM to feature
the water and wastewater industry.
a single loop measuring tube and boast
highly precise digital signal processing KROHNE Skarpenord
supported by a microprocessor. In 1994, KROHNE Skarpenord offers complete
KROHNE introduced the world's first solutions including cargo handling and
CFM featuring a single straight tube. ballast monitoring systems for seagoing
vessels.

XVII
KROHNE EMF Highlights, a world's first

1967 The short model EMF features a trodes, allows use with chemicals
modified inhomogeneous mag- and provides superior stability,
netic field, is more reasonably even with small nominal sizes.
priced and easier to install.
1983 The world's largest EMF with a
1968 EMFs for hazardous areas can nominal size of DN 3000 / 120"
also be used in the chemical in- and a measuring range of over
dustry. 100,000 m3/h for drinking water
and wastewater is built.
1974 The introduction of the pulsed dc
field with automatic zero correc- 1984 The first P signal converter fea-
tion means that for the first time turing full digital signal processing
ever, EMFs are stable, mainte- is launched.
nance-free precision measuring
1985 The first P signal converter in
devices.
a flameproof enclosure (Ex-d)
1978 The first EMF with a proper, fully can be adjusted and replaced in
digital measuring range setting hazardous areas.
is developed.
1991 KROHNE becomes the world's first
1979 The largest calibration rig for EMF manufacturer to obtain ISO
nominal sizes up to DN 3000 / 9001 certification.
120" with a measuring uncertainty
1992 The first EMF with an optimised
of less than 0.05% is approved for
flow profile measuring tube is
official calibration.
developed. The high degree of
1980 For the first time, EMFs with pulsed measurement accuracy convinced
dc field replace high-maintenance calibration bodies around the
ac field EMFs even in the pulp and world to install it.
paper industry.
1993 For the first time, an EMF with
1982 The first sandwich EMF, featuring aseptic and modular connection
an Al2O3 ceramic measuring tube adaptors is introduced into the
and fused-in-place platinum elec- food and beverage industry. It is

XVIII
KROHNE EMF Highlights, a world's first

simple, flexible and cost-efficient nominal sizes up to DN 3000 /


to use. 120" is accredited according to
EN 17 025 and also approved for
1994 First EMF with a ceramic measur-
official calibrations.
ing tube for volumetric filling. It
is equipped with fused-in-place 2001 The first 2-wire EMF featuring in-
CERMET electrodes and exhibits telligent energy optimisation and
outstanding test results on filling integrated noise filters. Ex-i, Ex-e,
machines. and Ex-d approvals make the use
of 2-wire technology possible in
1995 EMFs with pulsed dc field make
almost any EMF application.
it possible to measure rapid pul-
sating flows without a pulsation 2004 The new IFC 300 signal converter
dampener. exceeds NAMUR NE 107 and VDI/
VDE 2650 diagnostic directives,
1996 The first EMF for partially-filled
complies with Ex-e, Ex-d, Ex-i
pipelines with capacitive level
and has a measuring accuracy of
measurement is less sensitive
0.15%. The IFC 300 is compat-
to coating.
ible with almost all primary heads
1997 The first EMF with vibration resist- manufactured by both KROHNE
ant capacitive signal pick-up for and other manufacturers since
the measurement of liquids with 1974.
minimal electrical conductivity (up
2005 The introduction of flanged ver-
to 0.05 S/cm).
sions for ceramic EMFs simplifies
1998 Flexible, modular Ex-i design for installation.
all inputs and outputs, including
2006 The new, patented virtual ground-
PROFIBUS and FOUNDATION
ing makes it possible to eliminate
Fieldbus.
expensive grounding rings made
1999 The new, large calibration rig of special materials, thus making
featuring a measuring uncer- installation easier.
tainty of 0.04% and suitable for

XIX
1. The fundamentals of flow measurement
1 The fundamentals of flow measurement

Superior measuring technology is now


a prerequisite for accurate processes, (2) V = t2 qv(t) dt
t1

e.g., when dosing or metering liquid or


gaseous products.

In addition to temperature and pres- (3) m = t1 t2 qm(t) dt


sure, flow is one of the most important
measurements in many processes.
The mass m is independent from tem-
Flow measurement also involves meas- perature and pressure. However, the
uring volume V and mass m. These two volume V and the density of liquids
measured values are dependent on the are slightly dependent on temperature.
density : The effect of temperature and pressure
is even greater on gases. Here the flow
values are referred to a standard state,
(1) Density = Mass =m
Volume
V e.g. at 20C and 1.013 bar.

For a short introduction to the funda-


In most processes, instantaneous values
mental principles of fluid mechanics
are required, i.e., the flow as volume or
see Chapter 7.
mass per unit of time, e.g., in m3/h or in
kg/min. Thus, qV represents the volume
flow and qm the mass flow.

Many flowmeters are equipped with total-


isers to control filling processes. These
can be programmed as preset counters,
for example, to control filling processes.
When totalising the quantity, the flow q
is integrated from time t1 to t2:

22
2. The fundamentals of electromagnetic flowmeters
2 The fundamentals of electromagnetic flowmeters

2.1. Introduction Special designs also make flow meas-


urement possible in partially filled
Electromagnetic flowmeters (EMF) can
pipelines.
measure the volume flow of liquids,
slurries, sludges and pastes in almost Typical process liquids include:
any industry. The only prerequisite to Drinking water and wastewater;
use an EMF is that the product being Beer, milk, beverages;
measured must have minimum electri- Acids and caustics, also in high
cal conductivity. concentrations;
Sludge from sewage treatment
EMF devices such as those pictured in
plants;
Fig. 1 are used in fully filled pipelines.
Pulp suspensions, pulp;
Ore and dredger waste with solid
particles and rocks.
Fig. 1: Compact EMF
sandwich version
(KROHNE, OPTIFLUX
1300C)

26
The fundamentals of electromagnetic flowmeters 2

2.2. Measuring principle The process liquid in the tube with di-
ameter D flows through a magnetic field
Electromagnetic flow measurement is
positioned perpendicular to the direc-
based on Faraday's Law of Induction.
tion of flow and with a strength of B.
The law states that voltage is induced
The movement through the magnetic
across a conductor moving through a
field causes an electrical voltage to be
magnetic field. The functional principle
induced in the process liquid. This in-
of electromagnetic measuring devices
duced voltage U is proportional to the
is also based on this law of nature.
flow velocity
v and thus also to the vol-
When an electrically conducting liquid ume throughput.
flows through the magnetic field of an
EMF, voltage is induced as shown in
Fig. 2.

Fig. 2: Measuring principle of electromagnetic


flowmeters
1 B = induction (magnetic field strength)
2 D = tube diameter
3 v = mean flow velocity
4 U = voltage = k x B x
vxD
k = meter constant

1 U
4
2
3

27
2 The fundamentals of electromagnetic flowmeters

The following applies (simplified) for at its outputs for process control, e.g.
the voltage U: active output current of 420 mA.

The signal converter supplies the two


(4) U=kB
vD field coils with an active square wave
current and the field coils then generate
the primary head magnetic field. This
where
current supplies alternating positive and
k device constant
negative values. Alternating positive and
B magnetic field strength
negative flow-proportional signal volt-
v mean flow velocity

ages Ui are created by the alternating


D tube diameter
magnetic field strength B. The signal
For a circular tube cross section the converter subtracts these positive and
following applies: negative signal voltages present at the
electrodes from one another. This process
2 always occurs when the field current
(5) q=
vD /4
has reached its stationary value, sup-
pressing the induced interfering volt-
The indicated volume flowrate qi can ages and slowly changing (compared to
thus be determined as follows: the measuring cycle) external or noisy
voltages.

(6) qi = U D For a thorough introduction to the
4kB
theory of the electromagnetic flow prin-
ciple, refer to Chapter 8.
The induced voltage signal is picked up
using two electrodes that are in contact
with the process liquid and then supplied
to a signal converter. This signal converter
then eliminates interfering signals and
amplifies the measured value to make
suitable measurement signals available

28
The fundamentals of electromagnetic flowmeters 2

2.3. Advantages Because of the many modular interfaces


that can be combined, such as
The principle behind electromagnetic
flow measurement offers a variety of local display of flow and quantity
advantages which are briefly explained (totalising);
below. mA, pulse and status outputs, as well as
control inputs;
EMF devices enable the linear display
HART, FOUNDATION Fieldbus,
and totalising of volume flow in both flow
PROFIBUS , etc.

directions. The measurement is virtually
unaffected by pressure, density, tem- it makes it possible to easily integrate
perature, viscosity and flow profile. the measurement into system environ-
ments.
Because the magnetic field occupies
the entire cross section of the measur- The unobstructed tube cross section pre-
ing tube that the process liquid flows vents any additional loss of pressure. Even
through, each of the moving volume ele- when the process liquid contains solid
ments adds to the signal voltage U. This particles, the risk of damage through
means that the flow profile is averaged abrasion is limited because with EMFs
over the entire cross section of the tube. there are no parts protruding into the
This in turn means that only short inlet tube. When it comes to abrasive or even
and outlet runs are required. aggressive products, there is a broad
range of materials available for the
EMFs also feature very high measuring
wetted parts.
accuracy. Depending on the version, the
error under reference conditions can be Since EMFs do not have any moving
less than 0.15% of the measured value. mechanical parts, they are virtually
Because of the high degree of repeat- maintenance free. Consequently, they
ability a standard deviation of less than are extremely reliable, have a long ser
0.03% of the measured value is achieved. vice life and boast superior long-term
Measuring ranges of more than 100 : 1 stability and availability.
can be achieved with errors of less than
1% of the measured value.

29
2 The fundamentals of electromagnetic flowmeters

2.4. Primary heads magnetic field comes from the signal


converter.
The following describes the structure
and the most important components of Measuring tube
the primary head in an electromagnetic The magnetic field generated by the
flowmeter. Its most important compo- field coils located on the measuring tube
nents are shown in Fig. 3. must be able to penetrate the process
liquid through the wall of the tube. For
this reason, the measuring tube may
not be made of a ferromagnetic mate-
4
rial. Neither may the inner wall of the
tube be electrically conducting, so that
the signal voltage induced in the pro
cess liquid cannot be short-circuited.
The measuring tube is thus normally
made out of ceramic, plastic or non-
2
3
ferromagnetic stainless steel with an
U
electrically insulating liner or internal
coating. The interior wall of the tube
1 comes in contact with the process liquid
and must thus be corrosion-resistant
to the process liquid.
Fig. 3: Structure of an EMF
1 Field coils
2 Measuring tube For an overview of common materials
3 Electrodes
4 Signal converter and their fields of application, see
Table 9 in Section 5.2.
Field coils
The field coils are protected within the
housing of the primary head. They are
diametrically fastened to the measur-
ing tube and perpendicular to the elec-
trode axis. The supply that generates the

30
The fundamentals of electromagnetic flowmeters 2

Electrodes according to EN, ASME, ISO, JIS, etc.


The electrodes are in direct contact with standards. Fig. 4 shows the possible
the process liquid and must therefore nominal sizes.
be corrosion resistant while ensuring
good electrical transfer to the process
liquid.

For common materials and their field


of application, refer to Table 10 in
Section 5.2.

One exception is the capacitive electrodes


that measure without liquid contact.
For more information on this topic see
Section 4.6. Fig. 4: World's first EMF with 3 m nominal size
(KROHNE M960, 1983)

Designs
With the flanged version, the EMF
The primary heads for electromagnetic
measuring tube is usually made of non-
flowmeters are available in a variety of
ferromagnetic stainless steel. Typical
versions. They include:
materials for the liner include hard
Flanged version; rubber, polypropylene, polyurethane
Sandwich version; and fluoropolymers such as PFA, PTFE
Hygienic version. and ETFE. Ceramic measuring tubes
are also possible up to a nominal size
See below for a more in-depth look at
of 300 mm or 12".
the designs, properties and range of
applications.

Flanged version
This is the most common version for
installing the EMF into a process line.
The flanges are available in a variety
of nominal sizes and pressure ratings

31
2 The fundamentals of electromagnetic flowmeters

Sandwich version and low thermal expansion coefficient


With this version, pictured in Fig. 5, of ceramics guarantee high measuring
bolts are used to clamp the EMF be- accuracy and long-term stability, even
tween the pipeline flanges, which is when the process liquids are at a high
why this type of EMF is often referred to temperature. In addition, the tapered
as the "sandwich" or "wafer" version. form of the ceramic measuring tube
smoothes out severely distorted flow
profiles, thus significantly reducing
measurement errors in unfavourable
installation conditions.

Hygienic version
In the food and beverage industry as
well as in pharmaceuticals, all fittings
must be free of crevices to ensure that
no bacteria can grow and multiply. It
must therefore be possible to clean and
sterilise EMFs using chemical processes
(CIP) and steam (SIP).

The materials used may not release any


foreign particles into the process liquid
Fig. 5: EMF - sandwich version and must be, for example, FDA approved
(KROHNE, OPTIFLUX 1000)
and suitable for the food industry. For
In principle, all liner and measuring this reason, PFA and PTFE are gener-
tube materials used for the flanged ally used as liner materials. KROHNE's
versions can also be used. This may ceramic measuring tubes are also FDA
include, for example, stainless steel compliant.
measuring tubes with a PFA liner (OP-
TIFLUX 1000) or measuring tubes made
of non-porous zirconium oxide ceramic
(OPTIFLUX 5000). The highly stable form

32
The fundamentals of electromagnetic flowmeters 2

Depending on plant standards, the pro


cess connections used must comply with
a variety of standards. This may include,
for example, dairy or SMS connections,
clamps, hygienic flanges, etc. Weld-in
connections are also possible. Fig. 6
shows the OPTIFLUX 6000, a hygienic
version from KROHNE.

Fig. 7: L-shaped gasket with trapezoidal sealing lip

The OPTIFLUX 6000 was redesigned in


2003 to conform to the current EHEDG
guidelines. Both the hygienic design and
the cleanability of the measuring device
were tested. These tests ensure that the
EMF is suitable for hygienic applications.
Those areas where product residue can
be deposited or where microbes can
build up during production are tested
in particular.

Fig. 6: EMF - hygienic version with weld-in connec-


tions (KROHNE, OPTIFLUX 6000)

The EMF connection adaptors are


welded directly into the open pipe-
line. The sealing between the tube
liner and the process pipe is achieved
with a gap-free L-shaped gasket
with a trapezoidal lip as illustrated in
Fig. 7.
33
2 The fundamentals of electromagnetic flowmeters

2.5. Signal converter outputs. In addition, upper range values,


units and many other parameters for each
One of the tasks of a signal converter
output can be individually scaled.
is to supply the current required by the
primary head field coils to generate the Alternatively, or in addition, the signal
magnetic field. The signal converter is converter can transmit measuring or
also responsible for the P-controlled counting values and detailed self-diag-
processing of the flow-proportional nostic information via interfaces such as
signal voltage. HART, FOUNDATION Fieldbus and
PROFIBUS. These analogue and dig-
This signal voltage U is in the V to
ital interfaces are modular and can be
mV range and is subject to noise and
combined with one another, making it
interference voltages. The converter's
easy to integrate into system environ-
signal processing converts this voltage
ments.
into digital values and digitally filters the
noise and interference signals out. From KROHNE currently has two signal con-
this filtered signal, the flow velocity
v verters for use with EMFs. One is the
and the volume flow qV are calculated IFC300 which has been on the market
using scaling of the calibration constant since 2004, see Fig. 8. The IFC 100 has
and the nominal size of the primary been available since 2008 and is an option
head. with a very good priceperformance ratio.
It is an economical measuring solution
The signal converter can also provide
featuring a high standard of technology,
a variety of diagnostic data including
see Fig. 9. Both of these converters are
the conductivity of the process liq-
compatible with almost any primary head
uid, indication of the flow profile, etc.
produced both by KROHNE and other
The output of the data measured can
manufacturers since 1974.
then be visually displayed on the local
display or in analogue form via the mA
current output as well as in binary form
by way of frequency, pulse and status

34
The fundamentals of electromagnetic flowmeters 2

Fig. 8: KROHNE Com-


pact EMF, flanged
version, with IFC 300
(OPTIFLUX 4300C)
signal converter

The IFC 300 exceeds the NAMUR NE 107


and VDI/VDE 2650 diagnostic guidelines.
It is approved for use in hazardous areas
with optional flammable protection type
Ex-e as well as Ex-d and Ex-i for the I/O
connections. Its measuring inaccuracy
is less than 0.15%.

Despite the fact that it looks quite


different, the IFC 100 features many of
the same functions as its big brother,
the IFC 300. Conductivity measurement,
the major diagnostic functions and the Fig. 9: KROHNE Compact EMF, flanged version,
with IFC 100 (OPTIFLUX 2100C) signal converter
extremely convenient programmable
menus to name a few.

35
2
3. EMFs for fully filled pipelines
3 EMFs for fully filled pipelines

EMFs feature properties which clearly set pipelines. Fig. 10 and Fig. 11 contains an
them apart from all other flowmeters. overview of KROHNE's comprehensive
This chapter describes the properties range of EMF products.
of conventional EMFs for fully filled

1 2

3 4 5 6

Fig. 10: KROHNE's modular EMF product line,


signal converters pictured here
1 IFC 100 W
2 IFC 100 C
3 IFC 300 R
4 IFC 300 W
5 IFC 300 F
6 IFC 300 C

38
EMFs for fully filled pipelines 3

1 2 3

4 5 6

Fig. 11: KROHNE's modular EMF product line, 4 OPTIFLUX 5000 Sandwich
primary heads pictured here Maximum chemical resistance, abrasion sta-
1 OPTIFLUX 1000 bility and accuracy thanks to high performance
The economical solution with standard func- ceramics
tionality for simple applications 5 OPTIFLUX 5000 flange
2 OPTIFLUX 2000 Maximum chemical resistance, abrasion sta-
The first choice for the water and wastewater bility and accuracy thanks to high performance
industry ceramics
3 OPTIFLUX 4000 6 OPTIFLUX 6000
The standard solution for the process industry The device for the food and pharmaceutical
industry

39
3 EMFs for fully filled pipelines

3.1. Advantages the linear measuring principle;


the high measuring accuracy:
It is not without reason that EMFs are
the unobstructed cross-sectional
the favourite flowmeters in many in-
area of the tube;
dustries. Electromagnetic flowmeters
no mechanically moving parts;
feature advantages that make them
largely independent of the viscosity
indispensable for many applications.
and density of the process liquid;
The main advantages include:
largely independent of the flow profile;
measurement in both directions
of flow

1.0%

0.8%

0.6%
1

0.4%

0.2%
2

0.0%

0 1 2 3 4 5 6 7 8 9 10 11 12
3

Fig. 12: EMF limits of error at reference condi- 1 Measurement error (% of the measured value)
tions (KROHNE, OPTIFLUX 5300) 2 Limit of error (0.15% +1 mm/s)
3 Flow velocity v [m/s]

40
EMFs for fully filled pipelines 3

Linear measuring principle This advantage makes it possible to use


The linear relationship between the flow them in gravity feed applications, e.g.
v and the signal voltage U
velocity in wastewater technology, without in-
allows for linear signal processing. This creasing back pressure. Pump capacity
also enables simple but comprehen- can be reduced in high pressure pipes
sive settings of the measuring ranges to save energy. The unobstructed cross
as well as accurate measurement of section of the EMF tube also makes it
pulsating flows. possible to measure process liquids
with high solids content, e.g. ore or
High measuring accuracy
dredging slurries and suspensions,
EMFs generally exhibit minimal meas-
without risk of blockage or abrasion.
urement error. Fig. 12 illustrates an
Fig. 13 illustrates the unobstructed EMF
example for the specification of error
measuring tube.
limits under reference conditions.

Because of the accuracy and linear meas-


uring principle used, the latest EMFs can
handle measuring ranges of more that
100:1 at measurement errors below 1%
of the measured value. In addition, the
excellent repeatability is down to 0.03%
of the measured value.

Unobstructed tube cross section


The size of the EMF is usually the same
as the nominal size of the process piping.
The unobstructed tube cross section pre-
vents any additional loss of pressure.
Fig. 13: EMF measuring tube featuring a
stainless steel reinforced PFA liner (KROHNE,
OPTIFLUX 4000)

41
3 EMFs for fully filled pipelines

Independent of flow profile When the size of the EMF is the same
When flow changes from turbulent to as the nominal size of the pipeline, a
laminar, minor variations in measure- high degree of measuring accuracy can
ment accuracy may occur, depending be guaranteed through a straight and
on the design of the primary head, see unobstructed inlet run, the length of
Section 7.3. which is about five times the internal
diameter of the tube.
Independent tests confirm the minimal
influence of asymmetrically distorted flow This kind of inlet run can lead to in-
profiles and swirl flow on EMF measuring stallation-related problems in isolat-
accuracy, as shown in Fig. 14.

2
0 1 2 3 4 5 6
1.0

0.8

0.6

0.4
4
0.2

0.0
1
-0.2 5

-0.4

-0.6

-0.8

-1.0

0 100 200 300 400 500 600 700 800


3

Fig. 14: Effect of the flow profile on 2 Average flow velocity


v (m/s)
EMF measuring accuracy, here using a 3 Flowrate (l/min)
WIB-SIREP-E-1747-S-94 test with a 4 Swirl, two 90 elbows 5D upstream of EMF
KROHNE EMF with a nominal size of DN 50 5 Elbows 5D upstream of EMF electrode position
1 Measuring error (% of measured value)

42
EMFs for fully filled pipelines 3

ed cases, see Section 8.5. Problems Independent of viscosity, density and


could arise, for example, if the length flow profile
of the inlet run for an EMF with a nomi- EMFs measure the average flow velocity
nal size of DN 1000 was already 5 m. v of the process liquid in the pipeline.

In such cases, the inlet run can be halved The velocity depends only on the volume
by using reducers and an EMF with a flow rate q and on the inner diameter
smaller nominal size, see Section 5.2. of the pipe D but not on the physical
However, in so doing, an additional pres- properties of the process liquid. Due
sure drop, albeit a minimal one, should to the minimal influence of the flow
be taken into account. profile on the measuring accuracy of
the EMF, see Section 7.3, the transition
No mechanically moving parts
from a laminar to turbulent flow profile,
EMFs have no mechanically moving parts.
for example, caused by a change in the
This means that they are practically wear
viscosity, causes only minimal measure-
free and robust. For this reason, EMFs
ment error.
have an extremely long service life and
are largely maintenance free, even in Measurement in both flow
difficult application areas. directions
An EMF can measure with equal accuracy
Due to the lack of mechanical compo-
in both directions of flow provided that
nents, EMFs have no start-up threshold.
there are sufficiently long inlets on both
Thus, an EMF can measure extremely low
sides of the measuring device.
flow velocities, close to zero. For certain
applications still requiring a start-up
threshold, this can be set manually in the
signal converter. A start-up threshold may
prove useful as a low-flow cut-off to rule
out, for example, the effect of thermally
induced residual flows on volume flow
counting during downtimes.

43
3 EMFs for fully filled pipelines

3.2. Limits 3.3. Maintenance

In addition to their advantages, these EMFs function without any mechani-


standard EMFs also have limitations, cally moving parts and thus do not wear.
as explained below. EMFs in many applications can work
for decades without maintenance. Main-
Minimum conductivity of
tenance work thus primarily involves
process liquid
checking for potential damage that may
Due to the nature of the measuring prin-
be caused by external influences, deposits
ciple, the process liquid must exhibit a
or through abrasion in the measuring
minimum electrical conductivity. Depend-
tube. For this reason, regular checks
ing on the process liquid, the device ver-
should be performed.
sion and the measuring application, this
is between 0.05 S/cm and 50 S/cm. Modern EMFs use integrated diagnostic
EMFs are thus not suitable for the functions to continuously monitor their
measurement of hydrocarbons such own functioning as well as process pa-
as gasoline, oil, etc. and also not for rameters that could negatively affect their
gases. Process measurement problems function. However, the checks outlined
may occur below the specified minimum below are still important to guarantee
conductivity e.g. measuring errors or the service life of the device as well as
stable readings. Higher conductivities plant reliability.
have no influence on the measuring
Checking the flange seals
accuracy.
Leaks in the flange seals can lead to
Completely filling the measuring corrosion damage to the flanges or to
tube the process liquid penetrating into the
To measure accurately with conven- inside of the EMF, causing measurement
tional EMFs, the measuring tube must failure. Faulty flange seals must always
be completely full. There are special be replaced. Observe the manufacturer's
EMFs for partially filled pipelines, see recommended tightening torque when
Section 4.7. reinstalling.

44
EMFs for fully filled pipelines 3

Checking the terminal Inspecting the measuring tube for


compartments deposits and abrasion
Moisture in the terminal compartments Some process liquids have a tendency to
of the primary heads and signal con- form deposits or incrustations. In such
verter can result in measuring errors cases, the measuring tube should be
or complete failure. Moist terminal regularly cleaned. Cleaning processes
compartments must thus be dried out and cleaning intervals will depend on the
prior to closing. An electrical test as process and on the type of deposit.
described in Section 3.4. must then be
Abrasive action increases the inside
carried out.
diameter of the measuring tube. The
Testing the ground connections for resultant measurement errors can be
corrosion corrected by recalibrating the device. If
Ground connections can corrode. The the liner wears thin, an imminent failure
connections at the EMF and the connec- of the EMF can be expected. EMFs that
tions of the ground conductors should be have sustained such abrasion damage
regularly tested through to the central must be relined by the manufacturer
grounding electrode. or replaced.

Checking for vibrations and


water hammers
Heavy vibrations and water hammers
in the pipeline can occur downstream
of positive-displacement pumps or if
quick-closing valves are installed. This
can cause damage to compact EMFs in
particular. During heavy vibration, the
pipeline should be supported on both
sides of the EMF.

45
3 EMFs for fully filled pipelines

3.4. Testing However, this process is labour-, time-


and cost-intensive. In addition, removing
The periodic testing of flowmeters is
the device requires an interruption in
now required and set out by a company's
operation and can only be carried out
internal regulations (such as ISO 9001:
during plant shutdown.
measuring equipment monitoring) as well
as legal regulations (e.g. wastewater). 3.4.2. Testing and documentation
The accuracy of flowmeters can be tested with testing equipment
in a variety of ways. These include:
External testing equipment makes it pos-
periodic removal of devices for testing sible to test devices without removing
on external calibration rigs; the EMFs from the pipeline. They thus
periodic testing using special external allow statements regarding the functional
measuring equipment; accuracy of the device to be made with
the self-diagnostic function of an EMF, minimal effort.
which monitors the accuracy and all
process parameters during operation
or upon request, eliminating the need
for additional testing

3.4.1. Accuracy testing on a


calibration rig

The accuracy of an EMF can be tested


by the manufacturer or via a flow cali-
bration laboratory. To do this, the EMF
must first be removed from the pipeline.
Fig. 15: EMF test equipment (KROHNE, MagCheck)
This method allows you to check and
clean the electrodes and the lining at
the same time.

46
EMFs for fully filled pipelines 3

The latest generation of external testing all outputs.


equipment uses the device bus to read all
The measured data are first stored in
of the EMF's set operating parameters
the test control unit. Using KROHNE's
and then stores them. A test program
MagCheck as seen in Fig. 15, up to 70
then runs automatically to test the elec-
records can be collected without battery,
trode resistance, field coil resistance and
storage battery or external power supply
insulation, the accuracy and linearity
connection during the device testing.
of the primary signal processing and

Application-specific errors and Possible Possible causes


faults according to VDI / VDE 2650 consequences
Entrained gas in the process liquid Measurement error Cavitation
Unsteady reading Negative pressure
Faulty installation
Electrode corrosion Unsteady reading Incorrect material
selection
Corrosion Change in the
Leakage of the composition of
process liquid the process liquid
Conductivity too low Reading too low Process liqiud has
changed
Unsteady reading Incorrect choice of
measuring device
Liner damage a small % Incorrect material
measurement error selection
to complete Abrasion caused by
solids
Destruction of the EMF High temperature and
vacuum
Leakage of the process Faulty installation
liquid
Deposits on electrodes Measurement error Depositing of oils
and grease on the
electrodes
Unsteady reading Electrode passivation
due to incorrect
Measurement failure material selection
External magnetic fields Measurement error Electric furnaces
Fluctuating reading Electrolysis plants
Electrode short circuit Reading approaches Metallic deposits
zero in measuring tube
Partial filling Measurement error Faulty installation
Fluctuating reading

Table 1: Diagnostic requirements for EMFs as per


VDI / VDE 2650, Sheet 3

47
3 EMFs for fully filled pipelines

This data is then subsequently down- calibration in terms of its electrical and
loaded to a PC. Using the PC software electronic function.
included with delivery, these measure-
3.4.3. Intelligent self-diagnostics
ment results are evaluated in relation
to specified limit values and compared In order to avoid any interruption in
to the current setting data in relation to operation to perform checks, users
the measuring results of the last test. and manufacturers formed a joint
Based on this information, the software NAMUR working group to compile
generates trend analyses of all data for the most frequent error modes in the
each measuring station and advises when VDI/VDE 2650 directive. They are listed
changes are too large. If the changes in Table 1 (see previous page).
are within the tolerance range, a test
The IFC 300 by KROHNE complies with
certificate is issued, certifying that the
and exceeds this directive. It achieves
EMF exhibits a measurement error of
this by using diagnostic tools that go
less than 1% compared to the initial
above and beyond the requirements. For

Application-specific errors and Possible Possible causes


faults consequences
Inaccuracy of signal processing Measurement error Component drift
Coil temperature too high Reduced service Process liquid
life temperature too
Electronics temperature too high Reduced service High process liquid
life temperature
not permitted
High probability of Solar radiation
failure
Drift, measurement Faulty components
Non-linearity of magnetic circuit and Measurement error External magnetic fields
Signal processing Fault with electronics
Faulty field current value Measurement error Fault with electronics
Coil interruption
Faulty current output Measurement error Load too high
Interruption
Flow profile Measurement error Installation, e.g.
Measurement failure downstream of valve

Table 2: KROHNE's IFC 300 diagnostic tools which


exceed VDI / VDE 2650 requirements

48
EMFs for fully filled pipelines 3

examples of these tools, please refer to The following error modes can be ob-
Table 2. tained:
Electrode contamination;
Self-diagnostic functions via the
Short circuit or interruption in the
measuring electrodes
electrode connection cable (important
During self-diagnostics, the signal con
in "remote version" EMFs).
verter induces an alternating current IEP
into the process liquid via the electrodes. The measured resistance REI ena-
This current creates a drop in voltage bles the indication of the conductiv-
URE, which is dependent on the resist- ity with the following statements:
ance RE, in other words the electrical conductivity outside of permissible
conductivity of the process liquid, see limit of EMF or process liquid;
Fig. 16. Quality of cleaning processes (e.g.
wastewater);
Change in process liquid (e.g. transition
from process to cleaning liquid or vice
versa with CIP processes).
2
3 3
The status outputs with their adjust-
able switching points or bus connec-
tions signal that the preset conductivity
switching point has been exceeded or
4 1
not attained.

Testing the flow profile


Another diagnostic option which can be
used to discover potential error sources is
to measure and evaluate the flow profile
Fig. 16: Monitoring of general
error modes using the measuring electrodes in the measuring tube.
1 Coil
2 URE = (RE, )
3 IEP
4 REI =f()

49
3 EMFs for fully filled pipelines

Fig. 17 illustrates how deposits in the Partly filled measuring tube;


EMF, for example, can lead to incorrect deposits at the bottom of the
measurement results. measuring tube;
faulty liner;
Poor installation, e.g. a gasket
1
protrudes into the measuring tube or
2 the inlet/outlet runs are not sufficiently
3 long.
4
5 The symmetry of the flow profile is tested
by reversing the polarity of the upper
and lower field coils, as illustrated in
Fig. 18.

Fig. 17: Magnetic field with normal field coil


polarity for flow measurement
1 Ferromagnetic yoke
2 Field coils
3 Magnetic field with induction B
4 Measuring electrode
5 Measuring tube

A flow-dependent voltage is induced in a


flowing process liquid. In coatings where
v = 0 the voltage is zero. The sum of
these two voltage parts at the electrode
gives a faulty reading of the flow.

With flow profile measurement, reliable Fig. 18: Measuring the flow profile by reversing
the polarity of the upper and lower field coils
statements can be made about a variety
of states and errors during measure- This means that flow-dependent voltages
ment: with opposite polarities are induced in
the upper and lower half of the meas-
uring tube.

50
EMFs for fully filled pipelines 3

In the case of an undisturbed symmetrical


flow profile, the flow values in the up-
per and lower halves of the measuring
tube are the same. This means that the
induced voltages both above and below
are of the same size. Due to the reversal
in polarity of the magnetic field, they
have opposite polarities and add up to
zero at the electrodes.

With a disturbed flow profile, however,


the values of the signal voltages at both
electrodes are not the same. The signal
converter can determine the degree of
the disruption of the flow profile and the
thickness of the coating from the differ-
ence between the signal voltages.

For more details on flow profiles, refer


to Section 8.5.

51
4. EMFs in special areas of application
4 EMFs in special areas of application

This chapter introduces additional device Fig. 19 shows a small selection of special
versions for special areas of application solutions that KROHNE can provide in
for electromagnetic flowmeters. these cases.

1 2 3

4 5 6

Fig. 19: KROHNE specialists for 4 TIDAFLUX 4110 PF


special applications For partially filled pipelines
1 OPTIFLUX 4040 5 BATCHFLUX/BATCHCONTROL
A 2-wire device For volumetric filling systems in the beverage
2 OPTIFLUX 7080 FLANGE industry using CANopen-Bus
For the highest requirements using capacitive 6 WATERFLUX
electrodes and ceramic measuring tubes For remote measuring stations with no
3 CAPAFLUX 5080 SANDWICH power supply
For the highest requirements using capacitive
electrodes and ceramic measuring tubes

54
EMFs in special areas of application 4

EMF properties are so universal that 4.1. Explosion protected versions


these measuring devices are often used
Explosion protection is generally a safety
as the basis for solving very specific
topic and is thus subject to many legal
problems in flow measurement.
and technical rules. For this reason, there
Often, all that is necessary to find the will be no detailed handling of explo-
answer to a difficult question is a cer- sion protection in industrial measuring
tain amount of application experience, devices here. The aim here is rather to
commitment on the part of the EMF illustrate the background to the topic
manufacturer and a reliable EMF. In of explosion protection.
other cases, properties are required
For more detailed information, please
that vary so widely from any standards
refer to technical regulations and in-
that a completely new development is
formation provided by the device and
necessary. Both cases can be seen in
component manufacturer.
the following examples.
Risk of explosion and hazardous area
In the chemical and petrochemical
industry, in crude oil and natural gas
exploration and in many other indus-
tries, gases and vapours may escape
under certain circumstances during
the production and processing as well
as the transport and storage of flam-
mable substances. Many production
processes e.g. in the food industry or
in mining, may produce flammable
dust. These flammable gases, va-
pours, and dust can form a hazardous
atmosphere when combined with
the oxygen in the air. In such cases,
there is a risk of explosion. The area

55
4 EMFs in special areas of application

in which such risks occur is known as and the resulting dangers to life and
an "Ex-area". health. The requirements for operational
explosion protection are regulated by the
To cause an explosion, an effective ignition
European directive 1999/92/EC (ATEX 137,
source must be available. Sometimes
formerly ATEX 118 a) and in Germany by
all it takes is a small spark or a surface
the Ordinance on Industrial Safety and
that is too hot.
Health (BetrSichV). From 1 July 2003
Preconditions for an explosion only equipment (devices, components)
In order for a fire or explosion to take that complies with EC directive 94/9/EC
place, three conditions must be in place (ATEX 95, previously ATEX 100a) may be
at the same time. They are: a flammable used in hazardous areas.
material, a source of oxygen and potential
This directive regulates the design and
ignition sources that could trigger an
testing of explosion-protected systems,
explosion or fire. With the help of the
devices and components. It applies in
examples given in Table 3, it is easy to
the CENELEC (countries in the Euro-
determine when or if these precondi-
pean Community and the EFTA). Unlike
tions are present in a plant or not.
previous explosion protection directives
Explosion protection and ordinances, ATEX for the first time
The explosion protection contains pre- includes mechanical devices and com-
cautionary measures to avoid explosions ponents.

Preconditions for an Examples


explosion:
Ignition sources Hot surfaces, open flames, electric sparks, mechanical friction
and impact sparks, electromagnetic radiation, electrostatic dis-
charges, etc.
Oxygen sources Air, pure oxygen, chemical compounds giving off oxygen
Flammable substances Gases, vapours, mist, dusts that come from flammable liquids or
solids and are present in the right concentration for ignition

Table 3: Preconditions for an explosion

56
EMFs in special areas of application 4

Laws, ordinances and standards Equipment categories in hazardous


A wide variety of groups and institutions areas (excluding mines)
are involved with "explosion protection" All equipment, including measuring de-
from lawmakers to standards com- vices, is divided into three categories
mittees, testing and approval bodies, depending on the degree of safety for
monitoring, accident prevention and su- areas with explosion hazards arising
pervisory organisations to trade unions, from gas and dust, see Table 4.
insurance providers, manufacturers,
Gas Dust Required level of
installers and plant operators. safety
II 1G II 1D Very high
For more information about standards
II 2G II 2D High
dealing with explosion protection, go to:
II 3G II 3D Normal

www.explosionsschutz.ptb.de Table 4: Equipment categories for gas and


dust atmospheres
www.newapproach.org
Measuring devices which fall under
In countries outside of the CENELEC area,
equipment category 1 or 2 must first
other standards and regulations are in
be approved by a recognised testing
place. For example, the USA, Canada,
body. In accordance with ISO 9001:2000,
China, Japan, Australia, CIS, Hungary,
companies that manufacture equipment
Brazil and South Africa require their own
or components for these devices must
national approvals.
possess a special QA certification in com-
pliance with the 94/9/EC directive. Only
then may they display the CE marking
on the data plate.

57
4 EMFs in special areas of application

Installation in hazardous areas Gas Dust Danger


EMFs must be installed in accordance
Zone 0 Zone 20 Constant
with the Ex instructions of the equip-
Zone 1 Zone 21 Occasional
ment manufacturer. These Ex instruc- Zone 2 Zone 22 Rare
tions make up part of the Ex approval. An
Table 5: Classification of hazardous operating
EMF may only be used in the operational areas

conditions that fall within the limits of the determine which areas are hazardous
information given, e.g. range of applica- and select the equipment suitable for
tion, equipment category, temperature use in this area. The operator alone un-
class and any additional conditions. dertakes to meet all organisational and
technical measures to protect against
Non-compliance with the information
explosion and in particular to carry out or
contained in the installation guidelines
have carried out the required testing in a
results in termination of the approval for
timely manner. In addition, all equipment
the measuring device installed.
in hazardous areas that may contain an
Operation in hazardous areas ignition source must be marked as such in
Areas in which a hazardous, potentially accordance with directive 94/9/EC (ATEX).
explosive atmosphere may occur are The plant operator must then issue an
divided into zones depending on the prob- explosion protection document.
ability of the occurrence of this explosive
atmosphere, as shown in Table 5.

Atmospheres with explosion hazards


arising from gas are classified as Zone
0, Zone 1 and Zone 2. Atmospheres with
explosion hazards arising from dust fall
into Zone 20, Zone 21 and Zone 22.

In accordance with the BetrSichV [Ordi-


nance on Industrial Safety and Health], the
operator of the plant is solely responsible
for the safe operation of the plant. He must

58
EMFs in special areas of application 4

4.2. EMFs for custody transfer EMFs can be used, for example, as wa-
ter meters, or as a component of heat
There is an obligation to have measuring
meters or measuring systems for all
devices and systems officially calibrated
quantities of liquids other than water.
if that equipment is used to measure
If the EMF measurement results are
quantities commercially between inde-
used for billing purposes, the EMFs
pendent partners and where the result
are subject to mandatory calibration.
of the reading affects the amount of an
In other words, they must be officially
invoice.
calibrated by a notified body.
The legal basis for these devices is the
The prerequisite for this official calibra-
"Directive 2004/22/EC of the European
tion is previous approval of the device
Parliament and the Council of 31st March
design type for custody transfer. This
2004", also known as the "Measuring In-
approval for custody transfer consists
struments Directive" (MID). This directive
of a type examination, also known as
supersedes any previous, national and
conformity assessment. The conformity
EEC directives. It regulates the require-
assessment is a test of how the device,
ments that the device under test must
its accompanying documentation and
comply with during the type certification
the manufacturer's QA system conform
test, from the manufacturer and from
to the requirements of EU Directive
the notified bodies. Approvals based on
2004 /22/EC of 31 March 2004. The con-
older directives expire on 29.03.2016
formity assessment is performed by
at the latest. In the area of liquid flow
an accredited body (e.g. the National
measurement, the directive applies to
Metrology Institute) on a sample device
the following devices or systems:
and the accompanying documents. This
Water meter (Appendix MI-001);
approval describes or limits the range of
Heat meter (Appendix MI-004);
application, as well as the conditions of
Measuring systems for the continuous
use and installation. Here, for example,
and dynamic measurement of quantities
a longer unobstructed inlet run can be
of liquids other than water (Appendix
prescribed for use in custody transfer.
MI-005).

59
4 EMFs in special areas of application

Once the test has been successfully com- 4.3. EMFs in environments with
pleted, the approval for custody trans- strong magnetic fields
fer is granted for the approved design.
EMFs can also be used, for example, in the
An EMF may not be calibrated without
vicinity of electrode feeder lines for elec-
this approval for custody transfer.
tric furnaces and in electrolysis plants in
The first, easily recognisable sign that which current strengths well in excess of
a device has been calibrated is the CE 10 kA generate strong magnetic fields.
and metrology markings, as pictured The flow indication of an EMF can be
in Fig. 20. affected by strong dc or ac field cur-
rents.

Strong dc fields
If the external magnetic field is caused
1 2 3 by very strong direct currents, such as
in electrolysis plants, the following may
Fig. 20: XY metrology marking
1 CE marking occur.
2 Metrology marking
3 Number of notified body
EMFs have a ferromagnetic magnetic
field feedback circuit. This return circuit
affects the field strength of the magnetic
field generated by the EMF field coils.
If the external interference field has a
strong effect on the magnetic circuit of
the EMF, its magnetic resistance in-
creases. The magnetic field strength
B and the signal voltage U of the EMF
become weaker. This can result in large
measurement errors.

60
EMFs in special areas of application 4

In principle, this effect can also occur of an electrolysis plant still has a ripple
when magnetic fields are generated by that is expressed as an alternating cur-
alternating current. But in this case this rent component. In the case of electric
would only happen if the currents were arc furnaces, additional disruptions may
much higher. occur if the electrodes short circuit via
the melted goods or the electric arc is
Strong external alternating fields
interrupted. As a result, current fluctua-
With external magnetic fields generated
tions of a few kA can be caused which in
by strong alternating currents, in alter-
turn induce currents and voltages which
nating current furnaces for example, a
can then cause interference.
further effect can be seen.
The suitability of the EMF for this ap-
These magnetic fields can induce such
plication thus depends on the manufac-
high currents in the pipelines in the vicin-
turer and the type. The prerequisite for
ity of the transformers or the electrode
smooth functioning is first and foremost
feeders for the furnace, that ground
the correct choice of EMF type. To make
conductors and ground connections
the correct choice, a variety of informa-
from the EMF to the pipeline can melt.
tion is required in the planning stages.
The pipelines and grounding connec-
This includes the nominal size, type of
tions function like secondary windings
pipeline, strength of the magnetic field
of a transformer. The currents induced
at the installation site of the EMF and the
therein and the coupled voltages are
distance to the cables carrying the cur-
transferred via the process liquid to the
rent and their current strength, as well
EMF electrodes and can noticeably dis-
as the type of current (direct, alternating
rupt measurement.
or three-phase), etc., see Section 5.
Effects of strong magnetic field
fluctuations
Both effects (interference with the mag-
netic circuit and induced currents and
voltages) can be superimposed on one
another. For example, the direct current

61
4 EMFs in special areas of application

4.4. Measuring pulsating Fig. 21 illustrates the digitised flow values


flows using EMFs when measuring pulsating flows on the
internal device bus in a KROHNE signal
There are only a few flowmeters that can
converter after the first digital filter.
measure pulsating flows without pulsa-
tion dampeners being installed directly Pulsating flow can be recognised by
downstream of the positive-displacement peak values which can be more than 3
pump. KROHNE EMFs are always able times higher than the average flowrate.
to measure pulsating flows. For precise measurement, the peak

700
3
600

500

400

300
1 4
200

100

-100
0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400
2

Fig. 21: Example of the measurement of 1 q [l/h]


pulsating flow, pictured here at the outlet of 2 t [ms]
a diaphragm pump 3 q [t] [l/h]
4 qi [t] [l/h] T = 10s

62
EMFs in special areas of application 4

values must be processed quickly and


accurately in a linear manner without
saturation and overmodulation. The
sampling frequency, i.e. the number of
sampling pulses from the signal con-
verter per time unit, must be at least 10
times higher than the highest occurring
pulsation frequency.

The time constant of the signal converter


must be set sufficiently high enough to
average the pulsating flow sufficiently
and achieve a steady reading. The suit-
ability of the EMF for this application
thus depends on the manufacturer and
the type. For this reason, users should
be well informed and seek advice before
selecting a device.

With pulsating flows, manufacturers


generally have special setting regula-
tions which must be taken into account.
Strong vibrations may occur in the vi-
cinity of positive-displacement pumps.
KROHNE thus recommends the use of
remote version electronics for the EMFs.
The signal converter must be firmly
mounted to a vibration-free support.
For compact EMFs, the pipeline must
be supported on both sides.

63
4 EMFs in special areas of application

4.5. EMFs for products with high Use on dredging vessels


solids content The task of an EMF when used on a
dredging vessel is to record the meas-
EMFs have been successfully used in
urement and to control the height of the
hydraulic transport for approx. 50 years.
suction pipe above ground for optimum
Due to the number of variables that apply
capacity.
here, it is not possible to deal extensively
with the topic of "hydraulic transport" The transport medium in this case is
at this time. This chapter is therefore sea water. Solids generally include sand,
limited to a few practical examples and gravel and large rocks. The solids content
tips to be used in practice. can reach 50% by weight for flow veloci-
ties of up to more than 10 m/s. When
The completely unobstructed cross-sec-
used in this application, EMFs usually
tional tube area and the measuring prin-
have a nominal size of approx. 300 mm
ciple that determines the full volumetric
to 1400 mm and usually feature thick
flow, including solids, make it possible
polyurethane or ceramic liners.
for EMFs to measure liquids with a high
solids content. For this reason, EMFs are Coal mining
frequently combined with radiometric Here, EMFs are used in the coal wash, in
density meters to determine mass flow. other words, when separating the coal
This combination also monitors the mass and the rocks. The medium here is then
flow at a sufficiently high flow velocity water with coal and rocks.
and to ensure the maximum permis-
In individual cases, EMFs have been
sible concentration of solids in order to
used to transport the coal/rock mixture
prevent blockages in the pipeline.
hydraulically out of the mine from
Below are a few examples of application the bottom to the surface in order to
areas and process liquids: measure production quantities (at
transport distances of approx. 1000 m
from the bottom to the surface).

64
EMFs in special areas of application 4

Ore mining right down to the head box on the pa-


From gold mines to copper mines, EMFs per machine. EMFs can also be used
measure the hydraulically transport- at other stages, e.g. during bleaching
ed quantities of ore and rock on their or colouring. Extremely fine solid sus-
way from the conveyor point to the pensions with TiO2 and other fillers and
ore processing plant. For this reason, coatings are required to give the paper
EMFs with a soft rubber liner are often a fine, smooth surface. The mixture of
used. these suspensions is also controlled
using EMFs.
When it comes to the chemical pro-
cesses involved in treating the ore, EMFs Cement
measure acids and other reagents used The supports on oil drilling platforms,
to separate the metal from the rock. for example, are partially filled with ce-
ment to prevent the platform from float-
Aluminium industry
ing up. EMFs measure the compressed
Here, EMFs are used for measurements
quantity of cement to ensure that the
in all process stages in which liquid media
right amount is filled.
are found, even those with high solids
content e.g. abrasive bauxite slurry at Special requirements for EMFs in
temperatures up to 140C. hydraulic transport
Solid particles can often be abrasive.
Pulp and paper industry
The wetted materials and the liner must
In this industry, EMFs measure, for ex-
thus be abrasion-resistant. For help in
ample, the flow of wood pulp or dissolved
selecting the right device, see Table 9
recovered paper in the de-inking indus-
in Section 5.2. High pressure is also
try as well as highly abrasive tailings.
present in many applications, which
Solids content of 15% downstream of
further limits selection.
the refiners pose no problem for EMFs
with pulsed dc fields. The quantities
are measured and controlled by EMFs,

65
4 EMFs in special areas of application

Selecting an EMF for applications in as on the operating conditions and the


hydraulic transport installation position.
These examples illustrate how varied the
In horizontally positioned pipelines, some
requirements that are placed on EMFs are
of the solid particles always rub along
when it comes to the respective meas-
the bottom of the tube, which can lead
uring tasks in hydraulic transport.
to abrasion.
The following points are important when
Centrifugal forces act on the solid parti-
selecting the EMF and its materials for
cles when there is a change in direction,
this application:
for example, as a result of elbows in
Solids content (in per cent by weight or the pipe. The coarser particles are then
volume); pushed outwards and rub against the
Size, density, hardness and shape of walls of the tube for long stretches.
the solid particles;
In this instance, high demands are placed
Flow velocity;
on the EMF liner and the EMF should
Installation position (horizontal, vertical or
be installed at least 10xDN downstream
at a slope);
of an elbow.
Temperature, pressure of the process
liquid; Edges that protrude into the flow are
Effects on flow profile, e.g. elbows also subject to intense abrasion. These
in front of the EMF. edges could be the result of different
pipe diameters or unevenly matched
This information will help in the selec-
pipelines, as shown in Fig. 22. This situ-
tion of suitable materials.
ation is especially critical in the case of
Installation guidelines for EMFs in conventional EMFs featuring plastic liners
hydraulic transport as they usually have a somewhat smaller
The topic of abrasion is a top priority diameter than standard pipelines. In this
when it comes to products with high case, there are two possible options.
solids content. EMF abrasion depends
on the properties of the solid as well

66
EMFs in special areas of application 4

1 2

dR2
dR1

Fig. 22: Wear on pipe connections during 1 Wear in the case of varying pipe diameters
hydraulic transport and edges protruding into the (dR2 < dR1)
pipeline 2 Wear in the case of unevenly matched pipelines
3 Zones attacked by abrasion

The most common solution is to install


1 a protective ring on the edge of the inlet
side of the EMF as depicted in Fig. 23.

A different and more elaborate solution


2
is to match the inside diameter of the
pipeline to that of the EMF. Depending
on the nominal size and liner material,
the inside diameter of the EMF can be
adapted closely enough to result in a
3
step-free transition.

Flow profiles in products with


high solids content
When the density of the solid is higher
Fig. 23: Protective ring to protect the liner
of the EMF during hydraulic transport than that of the carrier liquid, the solid
1 Gasket
2 EMF liner tends to sink. The speed at which the
3 Protective ring
particle sinks depends on the difference
in density as well as the shape and size
of the particle.

67
4 EMFs in special areas of application

In a horizontal pipeline, the solids sink stay in motion. This kind of distorted
towards the bottom of the pipe. The flow flow profile does not, however, notice-
keeps the particles in motion. The speed ably affect the measuring accuracy of
of the particles is slower at the bottom the EMF and rarely becomes a problem
of the pipe, resulting in an increase in practice.
in concentration of solids towards
When the pipeline is in a vertical position,
the bottom of the tube, as shown in
the behaviour is different, see Fig. 25.
Fig. 24.
When the flow velocity is relatively low,
a constant flow in which the solids are
evenly distributed develops, as shown
1 in Fig. 25 on the left. However, when the
flow velocity is high and the inlet run is
sufficiently long and unobstructed, the
solid particles concentrate more in the
middle of the pipe. This results in the car-
2 rier medium flowing around the "solids
column" in a circular pattern towards
Fig. 24: Flow and concentration profile when hy-
draulically transported in a horizontal pipeline the top. Fig. 25 on the right illustrates
1 c : Concentration of solids
2 v : Flow velocity just such a flow. The dark core in the
middle marks the flow velocity of the
The result is distortion of the normal
solids, whereas the light area represents
symmetrical flow profile. When the flow
the flow profile of the carrier liquid in
is low, the solids settle on the bottom.
the ring between the solids column and
The unobstructed cross-sectional area
the wall of the pipe.
above the layer of deposits then becomes
smaller. This happens until the flow veloc-
ity increases enough for all of the other
particles above the layer of deposits to

68
EMFs in special areas of application 4

1 2

Fig. 25: Flow patterns during hydraulic 1 Flow profile at low flow velocity
transport in long, perpendicular pipelines 2 Flow profile at high flow velocity

Extremely high flow velocities can occur transport the product. Increasing the
in the ring between the solids column flow velocity does not further increase
and the wall of the pipe. This results the solids transport.
in a flow profile featuring a slow solids
EMFs register higher flow velocity in the
column in the middle of the pipe (e.g.
vicinity of the electrodes, which means
vS = 5 m/s) and a much faster ring of the
that such extremely distorted flow pro-
carrier liquid (e.g. vF 15 m/s). In this
files result in a flow value indication that
case, only the sheer stress between the
is too high.
liquid and the solids column is used to

69
4 EMFs in special areas of application

It makes, therefore, more sense to in- 4.6. EMFs without electrodes with
stall the EMF in a horizontal section of capacitive signal pick-up
the pipeline or at a distance of 10xDN
For process liquids with extremely low
downstream from an elbow where a
electrical conductivity and for those that
solids suspension is not yet completely
form insulating deposits on the wall of the
separated.
tube, interrupting the contact between
EMFs have also been successful in ex- the process liquid and the electrode,
tremely harsh applications in hydraulic there is an EMF featuring non-contacting
transport. To ensure smooth function- capacitive signal pick-up.
ing and a long service life, a detailed
Fig. 26 shows the structure of an EMF
discussion with the EMF supplier is
with capacitive signal pick-up.
necessary.

It is also extremely important to use


tried and tested solutions and materi-
als, even if at first glance there are other
alternatives that may seem to be more 3

cost-effective.

Fig. 26: Structure of an EMF with capacitive


signal pick-up
1 Field coils
2 Measuring tube
3 Signal converter
4 Capacitive electrodes

70
EMFs in special areas of application 4

With these EMFs, the electrodes are at- EMFs with capacitive signal pick-up
tached as large area condenser plates provides outstanding performance with
to the outside of the liner or to the non- products that deposit insulating layers
conducting measuring tube. The signal on the walls of the pipe e.g. latex or
voltage from the process liquid is capaci- bitumen suspensions. These layers can
tively picked up and without contact to insulate the electrodes of conventional
the liquid. These signals pick-ups and EMFs, resulting in measurement failure.
the associated projecting surfaces are With capacitive EMFs, such thin layers
fused-in-place to the outside of the ce- do not lead to failure. Instead, the lay-
ramic measuring tube by using methods ers have only a minimal influence on
in microsystem technology. This ensures measuring accuracy.
that vibrations and impacts cannot cause
The flow indicator of a capacitive EMF is
any interference voltages through piezo-
steadier than that of the classic EMF with
electric effects and that the EMF display
inhomogeneous products. This applies,
remains steady.
for example, to products containing a high
EMFs with capacitive signal pick-up offer degree of solids content, products that
a variety of advantages compared to con- have been poorly mixed or that exhibit as
ventional EMFs with classic electrodes yet incomplete chemical reactions, e.g.
in contact with the product. downstream from a neutralising point.
The capacitive signal pick-ups integrate
Thus, a minimum conductivity of just
spatial interferences across their large
0.05 S/cm is sufficient for measure-
electrode area.
ment. However, for conductivities less
than 0.3 S/cm, consult the EMF manu- With conventional EMFs it is possible
facturer prior to selecting a device as that there is no electrode material that is
some products prove to be completely compatible with all of the various process
non-conductive in the actual process liquids that flow through a pipeline. With
even if they show a laboratory value of, capacitive EMFs, on the other hand, it
e.g., 0.2 S/cm. is enough that the material of the pipe
is compatible.

71
4 EMFs in special areas of application

EMFs featuring non-wetted, capacitive 4.7. EMFs for partially filled


signal pick-up have proven successful pipelines
in practice and are well suited to pro
The pipelines in sewage and wastewater
cess liquids
systems are generally not completely
with low electrical conductivity;
filled, so classic EMFs had to be installed
for which it is either difficult or
in a so-called sluice underpass to ensure
impossible to select metallic
that the pipeline was completely full.
electrodes;
That cost money and required periodic
which lead to insulating deposits in
cleaning to avoid blockages. With the
the measuring tube and on the
introduction of EMFs for partially filled
electrodes, e.g. latex or bitumen
pipelines, such problems are a thing of
suspensions;
the past, see Fig. 27.
with a high degree of solids content;
with intense fluctuations in terms of
chemical properties, e.g. poor
mixing or as yet incomplete
chemical reactions (neutralisation).

EMFs with capacitive signal pick-up offer


a variety of advantages in practice.

Fig. 27: KROHNE TIDALFLUX 4110PF, the world's


first EMF for partially filled pipelines featuring
non-wetted level measurement

The design of such EMFs is very simi-


lar to that of EMFs for fully filled pipe-
lines. However, EMFs for partially filled
pipelines have at least one additional,
electrode pair mounted near the bot-
tom. This makes it possible to measure

72
EMFs in special areas of application 4

the signal voltage U and thus the flow tube is known. Using the average flow
velocity
v even when the level in the v and the flow cross section A,
velocity
pipeline is very low. In addition, these the volume flow q can be calculated:
KROHNE EMFs feature patented capaci-
tive level sensors embedded into the
(7) q=
vA
wall of the tube. This setup allows the
filling height h to be measured in the
measuring tube itself. Based on the filling Fig. 28 illustrates once again the measur-
height, the flow cross section A can be ing principle of EMFs for partially filled
calculated because the diameter of the pipelines.

2
1

Fig. 28: Measuring principle of an electromagnetic 1 Capacitive, non-wetted level measurement


flowmeter for partially filled pipelines (KROHNE, (embedded in the liner)
TIDALFLUX 4110PF) 2 Field coil
3 Measuring electrodes
4 Filling height h

73
4 EMFs in special areas of application

Prior to using an EMF for partially filled 4.8. Battery-operated EMFs for
pipelines, note that these EMFs start remote measuring points
measuring only when the minimum fill-
In many drinking water, untreated water
ing level has been reached. This level
and tap water systems, as well as in
generally corresponds to 10% of the
agricultural irrigation systems, water
tube diameter. The minimum flowrate
meters are installed in lines located in
at which measurement begins depends
remote measuring points. There is of-
on the pipe gradient and roughness of
ten no available power supply in these
the pipeline wall but is typically ap-
areas and the meter must work without
prox. 2.5% of the flowrate in a fully filled
power.
measuring tube.
Mechanical water meters have always
The inflow conditions for EMFs for par-
been used in such situations. However,
tially filled pipelines in open channels
their measuring accuracy frequently
are much more difficult to control than
suffered due to incrustations occur-
in fully filled pipelines. This is mainly
ring with the use of hard water as well
because the exposed surface of the
as contamination from untreated water.
flow, depending on the kind of distur-
The use of filters makes these measur-
bance, becomes calm only after very
ing points expensive and maintenance
long stretches. In such cases, the user
intensive.
should seek advice from the appropriate
manufacturer. While EMFs are not plagued with these
disadvantages, they do require a power
supply. Vandalism and theft made
the use of solar panels impractical. In
battery mode, the high power demand
of conventional EMFs reduces long
battery life.

The WATERFLUX 3070 EMF was


developed by KROHNE as a battery-
operated electromagnetic water meter

74
EMFs in special areas of application 4

for standard applications in the water measuring points with large nominal
and wastewater industry, see Fig. 29. sizes. Any additional costs for spare
filters, flow straighteners as well as
maintenance and cleaning work are
eliminated.

The key feature of the WATERFLUX 3070


is its measuring tube. The tube design
and material contribute significantly to
both the function and accuracy of the
WATERFLUX 3070. The shape of the
measuring cross sectional area was
designed in such a way that the electrode
distance D was larger than the height
of the measuring tube, i.e. a smaller
distance between the coils. The result
is the almost rectangular measuring
Fig. 29: KROHNE WATERFLUX 3070C cross section shown in Fig. 30. The rec-
tangular shape of the measuring tube
The WATERFLUX 3070 is thus superior
has a positive effect on the measuring
to mechanical water meters, mainly be-
accuracy and the battery life.
cause of its measuring principle without
any moving parts. No moving parts means
that there is no rubbing, wear or block-
ages caused by dirt and incrustations. The
surface of the lining is smooth and largely
corrosion resistant, thus reducing the
tendency towards deposits. This results
in excellent long-term stability.

The WATERFLUX 3070 boasts signifi-


cant cost advantages when it comes to Fig. 30: WATERFLUX 3070C measuring tube

75
4 EMFs in special areas of application

When investigating the influence of the The large electrode spacing D, the strong
flow profile, J. A. Shercliff discovered magnetic field B and the increased
the valence function for circular cross flow velocity
v result in a higher signal
sections, see Section 8.2. He also dem- voltage, even in the presence of a low
onstrated that the accuracy of EMFs with flowrate. The optimised wall thickness
a rectangular measuring tube cross sec- of the measuring tube minimises elec-
tion, homogeneous magnetic field and trical eddy currents.
surface electrodes are barely affected,
The measuring tube itself is cast from
even with a distorted flow profile. That
aluminium alloy using patented rein-
means that the WATERFLUX 3070 re-
forcements to ensure maximum form
quires only short inlet and outlet runs
and pressure stability. The tube is fitted
and can be installed in very narrow
with a smooth, low-friction coating made
standpipes and shafts.
of Rilsan. This polyamide coating has
The minimal height of the measuring been successfully used for over 30 years,
tube decreases the distance between especially in the water supply industry.
the field coils, thereby increasing the Rilsan is created from biomass and has
field strength B of the magnetic field. In been approved by the world's leading
addition, the flow velocity
v increases industrial countries for use with drink-
significantly due to the rectangular ing water. Rilsan can withstand all of
reducer. the chemicals generally used to treat
drinking water including ozone, chlorine
Note:
compounds and flocculants and lasts
The following applies to signal voltage
longer than, for example, some high
U:
quality stainless steels in sea water.

(8) U=KB
vD

76
EMFs in special areas of application 4

The power consumption of the As shown in Fig. 32, the


WATERFLUX 3070 is 5000 times lower WATERFLUX 3070 complies with the
than that of conventional EMFs. Power can requirements of custody transfer as per
thus be supplied using one or two bat- OIML R-49 as well as the measurement
teries. The battery is shown in Fig. 31. instruments directive 2004/22/EC,
Appendix MI-001.

There is usually no cable connection to


the control system at remote measuring
points. Users are thus dependent on re-
mote data transmission when it comes to
remote water sources, wells and network
transitional points. In cases like these the
signal converter on the WATERFLUX 3070
can be equipped using a compact module
with a datalogger and a GSM transmit-
Fig. 31: WATERFLUX 3070C battery
ter, see Fig. 33.
The WATERFLUX 3070 boasts an as yet
The datalogger saves all the flow data,
unrivalled battery life of 15 years, making
including the peak values and low flow-
this EMF particularly suited to measur-
rates measured during the night. The
ing points where connecting to a power
data is then transmitted via the GSM
supply is not a straightforward option.
network to the next base station and
The design of the WATERFLUX 3070 from there, if necessary, via satellite to
ensures very good overall measuring a server that the user can then access.
accuracy of 0.2% of the measured value The battery status and error messages,
despite the low field current and the like the detection of an empty pipe, are
extremely low battery load. transmitted as well.

77
4 EMFs in special areas of application

5.0
Q1 Q2

4.0
1

3.0 Q1 Q2
4 2
3
2.0 Q3 Q4

1.0 Q3 Q4

0.0
0.1 1 10 100 5 1000

Fig. 32: Error limits for water meters in ac- 3 Error curve KROHNE WATERFLUX 3070,
cordance with measuring instrument directive DN 100, Example
2004/22/EC, Appendix MI-001 and OIML R49 4 Absolute error (%)
1 Directive 2004/22/EC MI-001, Q3 / Q1 = 400 5 Flowrate (m3/h)
2 OIML R49 Class 1, Span Q3 / Q1 = 250

The new primary head for the WATER- The WATERFLUX 3070 is available in
FLUX 3070 is also compatible with the all common variants. This includes the
line-operated signal converters by compact version as well as the remote
KROHNE, the IFC 100 and the IFC 300. version with a primary head featuring
Therefore, even those industries that a protection category of IP68 or even a
usually have a power supply at every version in which the primary head can
measuring point can still benefit from be installed in the ground.
the advantages provided by the unique
design.

78
EMFs in special areas of application 4

4.9. Two-wire EMFs

The fundamental difference between two-


wire and multiwire measuring devices
is that the multiwire measuring devices
have a power connection which can be
supplied from a separate power source
to cater for any power requirements
(e.g. 20 Watt).

Classic multiwire EMF


Multiwire EMFs have a separate power
Fig. 33: 070C signal converter for the
WATERFLUX 3000 connection, which allows them virtually
unlimited power input. Fig. 34 shows
The lengths of the WATERFLUX 3070 also
the functional principle of multiwire
comply with the DVGW W420 and DIN ISO
EMFs.
13359 (WP short). This makes it simple
to replace a mechanical water meter They usually have user-friendly output
with the WATERFLUX 3070 EMF. options, such as a number of active out-
puts which feed a passive mA indicator
This electromagnetic water meter thus
without additional power (e.g. the classic
either meets or exceeds all industry
moving-coil instrument) or actuate total-
requirements.
izers with current or voltage pulses.

Multiwire EMFs typically power the field


coils with approximately 0.4 W to 1.5 W
(and up to 15 W in special cases). This
creates comparatively strong magnetic
fields and high signal voltages.

79
4 EMFs in special areas of application

However, fluctuating electrochemical Two-wire EMF


voltages can in turn cause interference Two-wire EMFs arose from a need for
voltages. This can happen, for example, simple and low-cost wiring, as has long
with inhomogeneous media, chemical been used, for example, with differential-
reactions still in progress, noise due to pressure sensors and orifice plates. With
low conductivity, solids, gas bubbles or EMFs, this requirement was much more
ambient electrical disturbances. These difficult to meet, even if the principle of
interference voltages can be reliably sup- the two-wire EMF bears a striking re-
pressed using digital signal processing semblance to that of the multiwire EMF.
methods. For the functional principle of two-wire
EMFs, see Fig. 35.

1
2
3 4
+ IF

- IF
6
v

+ U
i
- 7
5

Ui

Fig. 34: Functional principle of multiwire EMFs 1 Signal converter


2 Primary head
3 Sensor supply
4 Mains supply
5 Signal processing
6 Power supply e.g. 230 V, 20 VA
7 Output signal 420 mA

80
EMFs in special areas of application 4

Nowadays, electromagnetic flowmeters At low flowrates, the two-wire EMF


featuring two-wire connection technol- has only a few mW available to feed
ogy, such as the KROHNE OPTIFLUX 4040, the primary head and thus to gener-
have the dynamics, reliability and accu- ate the measuring signal. This output
racy of conventional four-wire EMFs. is still 10 to 40 times lower than that of
classic multiwire EMFs.
They can be used at an electrical
conductivity as low as 5 S/cm. With an This greatly limited the range of appli-
auxiliary voltage of 15 V, for example, cation for earlier two-wire EMFs. This
the two-wire 420 mA connection only is evident in earlier two-wire EMFs in
makes 0.060.3 W available. the requirement for a conductivity of at
least 50 S/cm.

1
2
3 4
+ IF

- IF

+ U
i
-
5 6

Ui

Fig. 35: Functional principle of two-wire EMFs 1 Signal converter


2 Primary head
3 Sensor supply
4 Mains supply
5 Signal processing
6 Output signal (supply) 420 mA

81
4 EMFs in special areas of application

However, modern two-wire EMFs by Simple integration into systems with


KROHNE offer the same application intrinsically safe "Ex" concept;
limits as standard EMFs at least 5 S/ Simple replacement of, for example,
cm. This is achieved by using state-of- differential pressure flowmeters by
the-art electronics, new digital filtering EMFs without the need for rewiring;
techniques for noise suppression and Low operating costs, practically
other innovative technologies including, maintenance-free, with diagnostic
for example, intelligent power units that functions;
optimally utilise all available energy for High measuring accuracy: 0.5% of the
sensor supply and noise-free signals. measured value at v > 1 m/s.

For particularly difficult applications, Modern two-wire EMFs from KROHNE


additional application-specific digital thus offer an outstanding signal to noise
filters can be activated via the user ratio, enabling the same application range
menu on the KROHNE two-wire EMF as standard EMFs despite the low power
to block out noise. available to the primary head.

The major benefits of two-wire EMFs


include:

Simple, low-cost wiring (savings poten-


tial of up to approx. 1,800 per meas-
uring point as compared to multiwire
EMFs);
Easy to incorporate into "Ex" concepts,
KROHNE two-wire EMFs allow the
user to select freely between the
"i", "e" and "d" types of ignition
protection;

82
EMFs in special areas of application 4

4.10. EMFs for rapid BATCHFLUX


volumetric filling Fig. 36 shows a BATCHFLUX 5015C
from KROHNE with an installation
In the beverage industry, the PET bottle
width of just 50 mm.
continues to gain popularity due to its
clear advantages including its shatter- KROHNE's BATCHFLUX was specially
proof nature, low weight and minimal developed for volumetric filling and
transport costs. features the following points:

However, PET bottles require a special the sturdy, only 50 mm wide stainless
volumetric filling technique because steel housing produced by way of in-
they expand differently when being vestment casting;
filled with carbonated beverages un- the inherently stable, vapour diffusion
der pressure. resistant ceramic measuring tube;
CIP and SIP resistance;
The shape of reusable PET bottles
the internal device bus;
changes over time due to mechani-
low energy consumption;
cal and thermal loads such as those
placed on the bottles during cleaning
processes. When this happens, there
is no longer a clear relation between
the filling height and the filled volume,
meaning that the accuracy of the filled
volume can no longer be guaranteed
using the classic fill level method with
PET bottles.

This is how volumetric filling with special


EMFs came about.

Fig. 36: KROHNE BATCHFLUX 5015C

83
4 EMFs in special areas of application

the possibility of recording filling pro the outside, fully approved for use in the
cesses; food and beverage industry as well as
maximum repeatability. the fulfillment of additional customer
requirements.
A BATCHFLUX EMF measures the volume
flow at every filling point. This process
continues throughout the entire filling
process.
1
9 4
The EMFs transmit volume pulses (e.g.
t
10 pulses for every millilitre that flows
10
through) to a batch controller in the cen-
tral computer of the filling machine. The 2

controller then integrates the volume 11


pulses and closes the valve as soon as
the preset number of pulses has been 3

reached and the desired quantity has t


6
been filled. This process is shown in 12
7 5 8
Fig. 37.

With filling times of a few seconds, this Fig. 37: Volumetric filling with
KROHNE BATCHFLUX EMF
application requires repeatability of, for 1 Start command, when container is ready to be
filled, Batch Controller (Batch Computer)
example, 0.2%. This includes the flow- 2 Valve control signal
3 External communication (volume preselection,
meter uncertainty, the time response statistics)
4 Pulses / volume e.g. 10 pulses per ml
of the valves and other uncertainties in 5 Filling time tF
6 On / Off
the filling process. 7 Valve opens
8 Valve closes
9 Storage vessel (pressurized if
Additionally, there are further require- liquid contains CO2)
10 Flowmeter BATCHFLUX
ments such as full CIP and SIP capability 11 Valve
12 Container Bottle, can, keg, paper or plastic
(e.g. superheated steam cleaning fol- container (pressurized if liquid contains CO2)

lowed by cold rinse), hermetic imperme-


ability, resistance to all cleaning agents
and procedures even when cleaned from

84
EMFs in special areas of application 4

BATCHCONTROL pressure valve, bleeder valve and purge


The BATCHCONTROL EMF is similar valve. The function of each of the contacts
to the established BATCHFLUX. How- can be set using the CAN bus, depending
ever, instead of the pulse output, the on the filling volume and time. This bus
BATCHCONTROL features an integrated can also be used to control each circuit
and intelligent batch controller which while filling is in progress, depending on
can simultaneously control up to five the position at the time. The CAN bus
valves. interface can be used to centrally adjust
the target volume and other parameters
while filling is in progress.

Two of these outputs can also be config-


ured as control inputs for the following
functions:

Filling start;
Binary input for the bus;
Emergency off;
Off (no function).

The most important difference between


classic EMFs with pulse output and the
BATCHCONTROL is shown in Fig. 39.
Fig. 38: KROHNE BATCHCONTROL 5014C
The left side of the picture shows eight
This integrated batch controller can in- filling stations on a carousel filler. The
dividually control the valves and valve flowmeters are only equipped with one
circuits at a filling point using five con- pulse output.
tact outputs. The BATCHCONTROL offers
contacts, e.g. for the filling valve, for the
fine adjustment contact for a slow filling
start and slow filling end with precise
metering as well as a contact for the

85
4 EMFs in special areas of application

Each one of these outputs and valve cir- To avoid having to guide all these con-
cuits is connected to the correspond- nections via loops from the rotating car-
ing interfaces of the process control ousel to a stationary process computer,
system. the computers with the interfaces are
usually mounted directly on the rotat-
This means that in the case of valves with
ing carousel.
five controllable channels, the number
of control lines can be increased by a
factor of 5.

1
8

3
3

7
5

Fig. 39: A comparison of the wiring involved with


a filling machine with a classic EMF and one with 5 Valve control cables
BATCHCONTROL 6 Interface: One Batch controller and one
1 Flowmeter with impulse output control output per valve
2 Package 7 Batchcontrol flowmeter with integrated batch
3 Valves controller
4 Counting pulse lines 8 CAN-Bus

86
EMFs in special areas of application 4

The right side of the figure also shows BATCHFLUX and BATCHCONTROL
eight filling stations. In this case, All measuring and operating data are
however, the filling stations feature transferred to the BATCHFLUX and
BATCHCONTROL EMFs equipped with BATCHCONTROL via the internal de-
integrated batch computers and CAN- vice bus BatchMon.
bus. This makes it possible to install
From this information, precise documen-
the computer in a stationary position
tation regarding device settings and all of
next to the carousel filler and to con-
the filling processes is created on external
nect the data bus to the computer in a
PCs. This gives the manufacturer of the
different way. The benefit to this is the
filling machine information about, for
considerable reduction in the amount
example, the dynamic behaviour of the
of wiring required and the elimination
valves, the flowrate during filling, the
of interfaces.
machine output and the repeatability of
The integrated batch controller automati- the filling quantity. The filling process
cally performs the overfill correction. This can then be optimised. Fig. 40 shows an
process is made possible by a special, example for this recording option.
patented start procedure when a new
At KROHNE, the filling machine manu-
device, machine or a new package size
facturer's optimum setting parameters
is introduced for the first time. This way,
can then be taken and the EMF can be
the first packages are already within the
set to these parameters prior to being
required tolerances. This is an advantage
delivered.
that prevents filling losses that occur
when using other EMFs during the first
incorrect filling processes.

87
4 EMFs in special areas of application

Application carbonation, to milk, pharmaceutical


BATCHFLUX and BATCHCONTROL are products and latex paint and cement.
available in sizes DN 2.5 to DN 40. This
With continued development, we may
range of sizes enables optimum adjust-
even see faster filling times, smaller
ment to filling quantities of less than
filling quantities and improved repeat-
20 ml to well over 50 l.
ability in future.
These application ranges can be moved
For a wide selection of suitable sizes
up or down following appropriate test-
see Table 6.
ing. Applications have previously ranged
from filling beverages with or without

1
1.80E-04
2
1.60E-04

1.40E-04

1.20E-04

1.00E-04

8.00E-05

6.00E-05

4.00E-05

2.00E-05

0.00E+00
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 3

Fig. 40: Record of a filling process "valve open" 1 Q [m3/s]


2 Valve opens, rising edge of flow
3 t (20ms / tick mark)

88
EMFs in special areas of application 4

Advantages The ZrO2-ceramic measuring tube fea-


BATCHFLUX and BATCHCONTOL focus tures a surface roughness Ra of just
exclusively on the requirements of filling 0.8 m and the platinum electrodes are
machines and the operating and ambient fused-in-place resulting in no gaps or
conditions. The two EMFs for volumetric cavities and are resistant to, for example,
filling is a result of experience in thou- sodium hydroxide, nitric acid, all CIP
sands of applications, including ones with liquids and superheated steam.
aseptic fillers. The following describes
Unlike tubes with plastic liners, the
the basic benefits of the features of both
measuring tube dimensions remain
EMFs.
stable, even at high temperatures,
Both EMFs feature a fully welded, ex- high pressure and after many cleaning
tremely sturdy stainless steel housing cycles using superheated steam, see
in an investment casting with M12x1 Fig. 41. This contributes significantly to
connections. The thin and narrow de- the high degree of measuring accuracy
sign, with a width of just 50 mm, means and the long-term stability of the BATCH-
that installation does not take up much FLUX and BATCHCONTROL EMFs.
space.

Size [mm] Recommended Recommended filling quantities


flowrate when filling from
DN 2.5 310 ml/s 10 ml
DN 4 1030 ml/s 20 ml
DN 6 2060 ml/s 40 ml
DN 10 60200 ml/s 100 ml
DN 50 150500 ml/s 200 ml
DN 25 4001200 ml/s 600 ml
DN 32 7002000 ml/s 1000 ml
DN 40 10003000 ml/s 1500 ml

Table 6: Recommendations for selecting the size


of BATCHFLUX and BATCHCONTROL depending on
filling volumes and flowrates

89
4 EMFs in special areas of application

In addition, BATCHFLUX and BATCH- These properties allow filling machines


CONTROL feature a power consump- with outstanding performance data to be
tion of just 3 W, which corresponds to designed in a cost-effective manner.
approx. one third of what other EMFs in
Quality assurance
this application field require, resulting
BATCHFLUX and BATCHCONTROL are
in additional savings.
subject to the same quality assurance
They also allow for viewing of the filling measures as any other KROHNE EMF,
pattern of both the valve and the machine, see Section 6.1.
making optimisation possible.

1 2
0.30%

0.25%

4
0.20% 7
322 l/h
644 l/h
0.15% 1307 l/h

0.10%
4

0.05%
5
6
0.00%
3 F E D C B A

Fig. 41: Results of an independent test 3 Maximum difference (% of the measured value)
investigating the long-term stability of EMFs with of 10 measurements with each device for 1 h
PTFE and ceramic measuring tubes, size DN 15, each at the specified flowrate
after continuous stress (e.g. 600 changes 18C / 4 Average of the maximum error at 322 l/h
80C + hot acid 80C + superheated steam 140C) 5 Average of the maximum error at 644 l/h
1 EMF with PFA liner 6 Average of the maximum error at 1307 l/h
2 EMF with ceramic measuring tube 7 Flowrate AF: EMF flowmeters - test setup

90
EMFs in special areas of application 4

KROHNE uses cyclical temperature tests 4.11. EMF probes


to reduce the probability of failure of
Given sufficiently long inlet runs of more
these EMFs to a minimum. During these
than 20xDN and Newtonian fluids with
tests, all electronic units must prove
undisturbed radial-symmetrical flow
their reliability and accuracy in a test
profiles, it is possible to calculate the
cabinet at cyclical temperature changes
total flowrate from the measured flow
of -20C to +60C prior to installation.
velocity.
This process eliminates early failures
and ensures smooth and accurate op- Electromagnetic flow probes are sens-
eration. ing probes in a insertion design, which
project into the pipeline and measure the
In addition, each and every BATCHFLUX
flow velocity at its tip, see Fig. 42.
and BATCHCONTROL EMF is individu-
ally calibrated by way of a direct com- The advantage that electromagnetic
parison of volume. The calibration rigs flow probes have over Pitot tubes and
used have a measuring uncertainty of vane sensors is that they measure with
less than 0.02%. This property guaran- no mechanically moving parts, making
tees a maximum measurement error of them largely immune to contamination
0.2% for EMFs for volumetric filling and insensitive to overload conditions.
processes. This high accuracy allows Also, sporadically entrained solid par-
a device to be replaced in case of error ticles have a negligible influence on
without the filling station in question measurement.
having to be immediately recalibrated
Application
or corrected.
Electromagnetic flow probes can be used
with sufficient conductivity and process
liquids with relatively low solids content.
For this reason, they are frequently used
as limit switches, e.g. for pump protec-
tion or as control units with continuous
flow display.

91
4 EMFs in special areas of application

Installation in pipeline
To keep the Reynolds number depend-
ence, even at turbulent flow profiles,
to a minimum, it is recommended to
position the probe tips at a distance of
an eighth of the inside diameter of the
tube from the inside wall of the tube.

Markings on the welding supports help


with these installation instructions.

Fig. 42: Electromagnetic flow probe


(KROHNE, DWM 2000)

92
5. Electromagnetic flowmeters planning
5 Electromagnetic flowmeters planning

The use of EMFs should be planned in 5.1. Analysing the range of applica-
detail in advance. Both the field of ap- tion and the operating conditions
plication and the operating conditions
Process liquid
for the EMF should be analysed prior
EMFs are used in many industries.
to selecting a device. Important aspects
There is a wide spectrum of liquid media
including EMF installation, grounding
whose flow parameters can be measured
and surge protection as well as impor-
using electromagnetic flowmeters.
tant standards that must be taken into
Table 7 illustrates a few examples of
consideration when using EMFs must
possible process liquids in a variety of
be considered.
industries.

Outlined below are some important points


to consider when analysing the process
liquid prior to selecting the right EMF.

Industry Typical process liquids

Food industry Milk, liquid dough, mineral water, liquid egg,


ice-cream, ketchup, yogurt,
Beverage industry Beer, lemonades, wine, fruit juice, mineral water,
fruit mixes,
Chemical industry Acids, alkalis, suspensions,
Water supply Raw water, potable water, chlorine dioxide solution,
flocculants,
Wastewater treatment Raw sewage, sludges up to 30% suspended solids content,
Power supply Gypsum suspensions in flue gas desulphurisation systems,
cooling water, measurement of thermal energy,
Mining Ore slurries, drilling fluids,
Building industry Plaster, screed, concrete,
Pulp and paper High-consistency pulps, papermaking stock, filler,
bleaching agents, screenings,

Table 7: Examples of process liquids in different


industries

96
Electromagnetic flowmeters planning 5

Minimum electrical conductivity interference voltages in the process


In practice, the required electrical con- liquid and on the electrodes. For this
ductivity of a process liquid is usually reason an EMF should always be in-
more than 1 S/cm. When using capaci- stalled in front of and never behind a
tive signal pick-up, minimum conduc- mixing, injection or neutralisation point,
tivities of 0.05 S/cm are possible, see see Section 5.3.
Section 4.6.
Solids content
Depending on the EMF version and the Due to the unobstructed cross sectional
application, a minimum conductivity of at tube area and the fact that solids do not
least 0.05 S/cm to 50 S/cm is required. influence the structure of the magnetic
The conductivity of diluted solutions is field, an EMF is the ideal flowmeter for
generally greater than 5S/cm. the hydraulic transport of solids.

Temperature Applications with ferromagnetic sus-


The selection of a resistant material for pensions e.g. with nickel or magnetite
the measuring tube depends first and must first be discussed with the EMF
foremost on the temperature of the pro manufacturer.
cess liquid. When using common PTFE
When used in hydraulic transport with
and PFA liners as well as ceramic meas-
abrasive solids content in the process
uring tubes, the maximum permissible
liquid, sufficiently abrasion-resistant
temperature of the process liquid can
measuring tubes or liner materials must
be up to 180C , for more information
be selected.
see Section 5.2.
When the flow velocity of the suspension
Homogeneity
is low, but always remaining above its
Poorly mixed process liquids with a spa-
critical velocity, the optimal solution is to
tially inhomogeneously distributed elec-
install in a vertically ascending pipeline,
trical conductivity can lead to meas-
see Section 5.3. If the flow velocity is so
urement errors. Incomplete chemical
high that a phase separation occurs,
reactions and the associated electro-
installation in a short vertical pipeline is
chemical processes can also lead to
possible. When the flow velocity is high

97
5 Electromagnetic flowmeters planning

enough to prevent any solid deposits, Air can be drawn in if the suction pieces
installation in a horizontal pipeline is in a container or tank are too close to
also possible. However, when selecting the surface;
the material, the heavy wear placed on Stirring devices do not only mix the
the bottom of the tube by the abrasive product, they also draw air down at its
particles must be taken into account. shafts and the air enters the pipeline
along with the product and gets into
Gas content
the EMF;
EMFs are first and foremost volume
Gases can escape from the process
measuring devices, see Chapter 2. Un-
liquid when the pressure is not suffi-
wanted gas content is thus one of the
ciently high or in the case of cavitation,
most common causes of measurement
as with carbonated beverages, beer,
error. This topic is discussed in more
wastewater;
detail below.
In the case of biosolids, ridges of welded
As with almost all volumetric flowmeters, seams or the edges of fittings can cause
EMFs include any gas dispersed in the outgassing;
liquid as part of the volume measure- Air can remain on fibres and be
ment. The EMF flow indicator displays transported into the liquids. An
a value that is too high by exactly the example: In the pulp and paper industry,
same value as the gas volume. The display the air sticks to the pulp fibres intro-
may also become very unstable. duced. At standard pressure, the volume
content of the air is often as high as
The entrainment of gas in liquids is usu-
the fibre content. e.g. the air content
ally unintentional. It can happen for a
can be as much as 10% at standard
variety of reasons. Here are some ex-
pressure with pulp containing 10%
amples:
suspended solids.
Air can be drawn in through pump shaft
seals;

98
Electromagnetic flowmeters planning 5

Usually, when using EMFs, only the In addition, if an EMF must be installed
volumetric flow of the liquid phase is in the vicinity of a pump, it should be
measured. EMFs, however, measure installed on the delivery side and not on
the volumetric flow of the total of the the suction side. The entrained gas is
process liquid including the entrained heavily compressed on the delivery side,
gas volume. So, the reading is too high causing the real volumetric proportion of
when compared to the expected meas- the gas and the difference between the
ured value of the volumetric flow of the measured and the actual transported
gas-free liquid or the volume of water liquid volume of the liquid to be very
including solids. The EMF thus displays small. When an EMF is installed on the
a positive measurement deviation. suction side of a pump, the pressure in the
product is low, even causing a vacuum.
If the diameter of the entrained gas reach-
The gas volume in the process liquid, its
es a size similar to that of the diameter
total volume and thus also the indicated
of the EMF electrodes, the flow indicator
error is accordingly higher.
may stop working for a short period of
time or become extremely unsteady. In Fig. 43 illustrates this effect. When a
horizontal pipelines, the gas can collect pump is initially started the pressure
in the top of the pipe, which can in turn on the suction side can decrease from
lead to measurement errors. 1 bar to 0.1 bar and the gas volume can
increase by a factor of 10. The amount
Such effects can be avoided or at least
of entrained gas becomes much larger.
reduced by using gas separators, bleeder
The EMF indicator does not only indi-
valves or vents installed upstream of
cate a seemingly significantly greater
the EMF. It is important to ensure that
error, it can also become very unsteady
the minimum filling height in tanks or
as the large amount of entrained gas
containers is sufficiently high so that no
briefly isolates the electrodes from the
air can be drawn in through discharge
liquid.
connections or the blades of the stirring
unit, see Section 5.3.

99
5 Electromagnetic flowmeters planning

If, however, an EMF is installed in the However, measurement error caused


vicinity of the delivery side of a pump, by gas content in the process liquid
the entrained gas is heavily compressed. is largely independent of the installa-
Due to the limited gas volume fraction, tion position of the EMF. When installed
the measurement can be carried out horizontally, the entrained gas mostly
with a high degree of accuracy. gathers at the top of the tube. For this
reason, always ensure that the electrode
For this reason we emphasise that EMFs
axis is almost horizontal when the EMF is
should be installed where the pressure
installed. Otherwise, the gas may insulate
is at its highest. This does not only ap-
the electrode at the top of the tube and
ply to EMFs.
the measurement will then fail.

10

1
5

3
1

0.5

0.2
2

0.1
0.1 0.2 0.5 1 2 5 10

Fig. 43: Reduction of the gas volume by 1 Relative gas-volume content dependent on
increasing pressure, relative gas-volume con- operating pressure, gas content
tent standardised to 1 at pabs = 1 bar 2 Absolute pressure (bar)
3 Relative gas-volume content

100
Electromagnetic flowmeters planning 5

With vertical installation, the entrained contains gas, the sonic velocity and
gas may briefly interrupt its connection possibly also the signal strength will
to the process liquid when passing the be significantly lower.
electrodes. This made lead to a fluctu-
Diagnosis via signal converter:
ating reading.
Entrained gas flowing through an EMF,
Methods to test for gas content in causes irregular fluctuations of the
media without solids electrode voltages, so-called electrical
A variety of methods are suitable for noise. The EMF signal converters IFC
testing for gas content in media without 300 and 100 by KROHNE can measure
solids. this noise and output a value for the
noise level as well as indicate if the
Listen to the pipeline:
adjustable limit value is exceeded.
Hold the tip of a screwdriver on the
pipeline to serve as an acoustic coupler. Methods to test for gas content in
A trained ear can hear the noise and products containing solids
recognise whether or not there is gas With fibrous solids as e.g. pulp and sheep's
in the liquid. wool, the air is attached so densely to
the fibres that it is difficult to separate
Using clamp-on ultrasonic
them. In such cases, determining the
flowmeters:
gas content in suspensions or pulps is
Clamp-on ultrasonic flowmeters such
not that easy. Electrical noise caused by
as KROHNE's OPTISONIC 6300 enable
the gas content disappears in the heavy
the detection of gas content in homo-
electrical noise of such suspensions due
geneous liquids. This gas content can
to the zeta potential of the fibres and
lead to a significant reduction in the swirl
added additives. The noise measurement
velocity and the strength of the signal
in the diagnosis range of the KROHNE
received. A process liquid free of gas,
IFC 300 and 100 signal converters can-
e.g. after a long period of downtime,
not provide clear information here.
will display a high sonic velocity and
signal strength. When the medium

101
5 Electromagnetic flowmeters planning

So-called anti-foaming agents or de- When dealing with products that have a
foamers have been successful in sup- tendency towards outgassing, a reducer
pressing gases and helping to prevent (which otherwise offers many improve-
the measurement errors caused by those ments to measuring accuracy) can actu-
gases. They are added to the suspension ally cause outgassing.
at the same time as the pulp is added.
Adhering to these tips can help avoid
Observing the mass balance in the pro
faulty measurements caused by high
cess both before and after indicates if
gas content.
and how much gas was removed by using
the additive. Nominal sizes
The range of application of conventional
However, the entrapment of air in the
EMFs now goes from DN 1 with an inte-
process liquids can be prevented at an
rior diameter of 1 mm to over DN 3000
earlier stage. Some points can be taken
with an inside diameter of more than
into account as early as the plant plan-
3 m. This means flowrates from approx.
ning stages.
1 l/h to well over 100,000 m3/h can be
The most important thing to take into measured.
consideration is the correct installation
Pressure ratings
of the EMF on the delivery side of the
Pressure ratings are based on the flange
pump.
standards. The standard used is the most
Next, the suction intake points on con- common pressure rating for each nomi-
tainers, tanks or even lakes must be nal size, for example, nominal pressure
located far beneath the surface so that PN 40 for nominal sizes up to DN 50.
no air can be drawn into the supply line.
Special EMF versions with nominal pres-
Stirring units must be in a deep enough
sures of over 1,500 bar are possible.
position so that no air is drawn into the
shaft of the unit and down into the sup-
ply line.

102
Electromagnetic flowmeters planning 5

5.2. Device selection The process liquids have a tendency


to form deposits (The flow velocity v
Selecting the right device means se-
should then be more than 3 m/s);
lecting the right nominal size, the de-
sign appropriate to the intended use, Inlet conditions are unfavourable (e.g.
suitable materials for the measuring when an unimpeded, straight inlet run
tube and electrodes as well as the right as in Section 3.1 is not possible);
protection category and signal converter
A very large measuring span is required
version.
(e.g. with DN 50 a span of 0.212 m/s
Nominal size or 2.284 m3/h).
The nominal size (DN) of an EMF is gen-
If the nominal size of the EMF selected
erally the same as the nominal size of
is smaller than that of the pipeline, re-
the pipeline into which the EMF is in-
ducers and expanders must be installed
stalled. A smaller nominal size for the
in the pipeline both downstream and
EMF may be selected if installation is to
upstream of the EMF, as shown in
take place between reducers. However,
Fig. 44. Reducers can be considered a
the nominal size of the EMF may never
part of the inlet run.
be larger than that of the pipeline. It is
necessary to select a smaller size if:

V1 V2 V1

/2 < 8

d1 d2 d2 d1

Fig. 44: Reducer upstream and downstream of EMF

103
5 Electromagnetic flowmeters planning

This increases the flow velocity in the Additional information:


EMF, thus decreasing the risk of de-
The prescribed straight inlet runs must
posits. Depending on the process liquid,
be provided on the inlet side between
the flow velocity v should be more than
excentric reducers and the EMF;
3 m/s.
Special rules for inlet runs are indicated
In addition, a concentric reducer with
in the approval for EMFs to be used in
/2 < 8 installed downstream of the
custody transfer;
EMF smooths out the flow profile. This
allows the required inlet run to be reduced Check whether the usually small pres-
from a previous length of five times the sure loss is acceptable. In addition,
tube diameter d1 to approx. 3 to 4 times for media with a tendency to outgas,
the inside diameter of the EMF d2 with it should be determined at what flow
no additional noticeable measurement rate in a reducer this occurs when
errors, see Section 8.5. the discharge pressure drops, see
Section 8.2.
The measurement error of an EMF
increases as the flow velocity to be Design
measured decreases, see Fig. 12. If a When selecting the design, it is important
large measuring span is required, the to know whether a flanged or sandwich
flow velocity can be increased by using a EMF is to be used, see Section 2.4.
reducer. At low flowrates, the measure-
Sandwich version EMFs are less expen-
ment error then decreases accordingly.
sive than flanged version EMFs.
The reducer can be configured so that
the flow velocity v is approx. 10 or 12 For a brief overview of the major se-
m/s at maximum flow. lection criteria for the two designs see
Table 8.

104
Electromagnetic flowmeters planning 5

Measuring tube Examples of such materials include ETFE


Chemical resistance alone is not enough as well as natural types of rubber or
when selecting the material for the synthetic rubber. Measuring tubes made
measuring tube. Good dimensional of extremely dense Al2O3 or ZrO2 ceram-
stability and vacuum resistance along ics exhibit the best properties.
with sufficient abrasion resistance can
Abrasion resistance
be equally crucial to the service life and
When selecting materials for use with
long-term stability of the EMF.
abrasive products, the selection depends
Dimensional stability and vacuum on the density, size and hardness of the
resistance solid particles as well as the flow veloc-
For injected plastic liners such as PFA ity and installation site. There are also
and PP, good dimensional stability and cases in which soft rubber liners can
vacuum resistance is achieved by means achieve a long service life. In extremely
of stainless steel reinforcement elements difficult cases, large nominal sizes (e.g.
(coated with plastic) that are welded into DN 1000) are lined with thick ceramic
the measuring tube. Selecting plastics tiles to achieve an economically optimal
or natural and synthetic elastomers that service life.
stick to the tube wall can also be help-
ful.

Selection criteria Sandwich Flange

Costs More favourable Less favourable


(assuming otherwise identical
configuration)

Installation weight Lighter Heavier


Compatibility with flange standards
Universal Depends on standard
(e.g. DIN, ASME, JIS)
Availability of precision ceramics DN 2.5DN 100 DN 25DN 300
or or
1/104 112

Metrological properties Equally good

Table 8: Selection criteria for the design of EMF devices

105
5 Electromagnetic flowmeters planning

Table 9 offers an overview of suitable For specific applications, consult the


measuring tube materials. It should technical data sheets and recommended
be considered as a summarised guide information regarding materials and their
only. resistance.

Material Max. Chemical Abrasion Permanent Avail- Vacuum


Process resistance resist- deforma- able DN resistant
liquid ance tion under ranges
tempera- pressure /
ture tempera-
ture
Non-reinforced up to from hot negligible pro- 10600 poor
fluoroplastics, 180C alkaline solutions to nounced to to
e.g. PTFE to hot concen- good negligible negligible
trated acids
PFA with up to from hot good minimum 2.5150 very good
stainless steel 180C alkaline solutions
reinforcement to hot concen-
trated acids
ETFE up to from hot very good minimum 2002000 very good
120C alkaline solutions
to hot concen-
trated acids
Hard rubber up to 90C low minimum negligible 253000 very good
concentration
Polypropylene up to 90C low good minimum 25150 very good
with stainless concentration
steel
Soft rubber, up to 60C low good minimum 253000 very good
Neoprene concentration
Polyurethane up to 60C low excellent minimum 501800 very good
concentration
Aluminium up to from warm material maximum 2.5300 very good
oxide and 180C alkaline solutions with maxi- long-term
zirconium (medium concen- mum stability,
oxide trations) to highly abrasion even with
ceramics concentrated resistance small DN
acids below 100C

Table 9: Overview of measuring tube and liner


materials for EMFs

106
Electromagnetic flowmeters planning 5

Electrode If an EMF with conventional electrodes


Before selecting electrodes, first decide is selected for an application, both the
on the EMF configuration suitable for chemical resistance as well as the elec-
the measuring application. The avail- trical contact between the surface of the
able configurations include EMFs with electrodes and the process liquid must
conventional wetted electrodes and EMFs be ensured. Selecting a material for the
with capacitive, non-wetted electrodes, electrodes which does not feature suf-
see Section 2.4 and Section 4.6. The ficient chemical resistance can result in
choice between these two configura- total loss of the measuring device. The
tions depends primarily on the process formation of insulating layers on the
liquid. surface of the electrodes, e.g. through
passivation or electrolytically generated
EMFs with capacitive signal pick-up
gas films, can also lead to measurement
are best suited to applications in which
failure, resulting in expensive produc-
the conductivity of the process liquid
tion loss.
is less than 1 S/cm and a minimum
of 0.05S/cm. For an overview of suitable electrode
materials for particular process liquids,
In addition, process liquids that are ei-
see Table 10. The materials are listed
ther highly abrasive or that tend to coat,
as examples only and are not meant
which would insulate the electrodes of
to be used without prior testing for the
conventional EMFs, as is the case with
application in question.
latex or bitumen suspensions etc., can
limit the functionality of conventional
EMF electrodes. Even with chemically
inhomogeneous products with a high de-
gree of solid content or ongoing chemical
reactions, e.g. following neutralisation
or the injection of additives, capacitive
EMFs are recommended. In all other
cases, an EMF with conventional elec-
trodes is the first choice.

107
5 Electromagnetic flowmeters planning

Medium Proven For measuring systems in locations at


electrode materials
risk of submersion (max. 30 min. up
Water, Stainless steel, Hastelloy
wastewater C 4, to a hydrostatic head of 1 m), the IP 67
Acids Tantalum, platinum, protection category is sufficient, provided
platinum-cermet
the diameter of the electrical cables has
Alkalis Hastelloy C 4, platinum,
partially platinum-Cermet been properly selected and the cable
entries have been professionally and
Table 10: Suitable electrode materials depending
on process liquid meticulously sealed.

In all cases where the user does not If the risk of longer lasting or higher
have adequate experience, the chemi- submersion cannot be excluded, a
cal resistance of the electrode material remote version with a primary head
and its contact reliability to the pro- featuring protection category IP 68 (pro-
cess liquid must be tested with respect tection against lasting submersion in
to the composition, concentration and water) must be selected. With this type
temperature of the process liquid. Most of protection, manufacturers and users
EMF suppliers will provide assistance must agree on the typical submersion
in this respect. depth and guaranteed submersion time
on a case by case basis. In this case, the
Protection category
signal converter must be installed above
Modern EMF primary heads and compact
the highest possible submersion level,
devices generally comply with protec-
otherwise there is a risk of irreparable
tion category IP 67 for protection against
damage to the device.
temporary submersion (max. 30 min.
up to a hydrostatic head of 1m). 19 rack signal converters which gen-
erally only have IP 20 protection, are
Remote signal converters comply with
protected from dust and water because
protection category IP 65 (protection
they are installed in central cabinets with
against water spray) or IP 67, depend-
higher protection categories.
ing on the housing design.

108
Electromagnetic flowmeters planning 5

EMFs with remote and compact signal Fig. 46 shows the same IFC 300 signal
converters converter by KROHNE, this time in the
Most EMFs are available in either remote version and in protection category
remote or compact versions. Both ver- IP 65. An IP 67 version is also available
sions often have the same primary head (see Fig. 11, IFC 300F).
and signal converter electronics, in other
Always select a remote version if the use
words, the same metrological properties.
of the compact version is not recom-
However, different operating conditions
mended, even if only according to one
and limits of use must be taken into
of the operating conditions outlined in
account when selecting. Fig. 45 shows
Table 11.
KROHNE's OPTIFLUX 4000 with the
IFC 300 signal converter in the compact
version.

Fig. 45: Compact version of a signal converter Fig. 46: Wall-mount version of a signal converter
(KROHNE, IFC 300C) (KROHNE, IFC 300W)

109
5 Electromagnetic flowmeters planning

Operating conditions at Compact Remote


the measuring point version version

High process temperature Not appropriate Appropriate

High ambient temperature Not appropriate Appropriate

Severe vibrations Not appropriate Appropriate

Risk of submersion Appropriate


Not appropriate
(for protection category

Corrosive atmosphere Not appropriate Appropriate

Good accessibility Appropriate Not appropriate

Table 11: Criteria for selecting signal converter


version

5.3. Installation of the respective manufacturer of the


electromagnetic flowmeters.
An explanation of how EMFs are to be
installed in a measuring system in order Only the essential points to be observed
to achieve an accurate measuring result when installing the primary head are
is provided below. addressed here.

When installing the remote version of Table 12 illustrates some typical installa-
an EMF, note the difference between the tion situations. This includes suggestions
installation instructions for the primary as to the selection of the installation
head and for the signal converter. site for the primary head that have been
proven as optimal in practice.
For details and required regulations
to be observed when selecting the in-
stallation site for the primary head as
well as the installation process in the
pipeline, consult the planning guidelines

110
Electromagnetic flowmeters planning 5

Highest point of pipe run


Avoid installation!
(Air bubbles may form in the meas-
uring tube, leading to faulty meas-
urement.)

Downpipe
Avoid installation!
(Open discharge.)

Long pipeline
Always install control and shutoff
valves downstream of the flowmeter.

Flow in a downpipe over 5 m in 1


length
Install air valve downstream
of the flowmeter.
2

Horizontal pipe run


Install flowmeter in slightly
ascending pipe section.

Open feed or discharge


Install flowmeter in lowest section
of pipe.

Pumps
Do not install flowmeter on suction
side of pump.

Table 12: Suggestions for selecting a suitable in- 1 Automatic vent and air valve
stallation site for the primary head 2 >5m

111
5 Electromagnetic flowmeters planning

5.3.1. Primary head Inlet and outlet runs


To avoid measurement errors due to
Observe the following points when se-
distorted flow profiles, most EMF manu-
lecting the installation site for the EMF
facturers recommend having an unim-
primary head for fully filled pipelines:
peded inlet run with a length of at least
Inlet and outlet runs; five times the nominal size downstream
Ambient electrical conditions; of elbows and T-pieces. Control devices
Accessibility; such as slide valves and gate valves may
Installation process; only ever be installed downstream of an
EMF, never upstream. If the inlet runs
The applicable requirements are out-
are shorter or the control devices im-
lined below.
properly positioned, additional errors
may occur, as shown in Fig. 47.

10

1 5

0
0 1 2 3 4 5
2

Fig. 47: Possible measurement errors when inlet 1 Measurement error in % of measured value
runs are too short 2 Distance of the interfering element from the
EMF in nominal tube sizes (DN)

112
Electromagnetic flowmeters planning 5

By using a reducer upstream of the EMF, Ambient electrical conditions


the inlet run can be shortened from five Nowadays, EMFs can be installed in the
times the nominal size to three times vicinity of electric furnaces with power
the nominal size without any additional capacities in the megawatt range as
measurement error during operation, well as in chlorine electrolysis plants
see Section 5.2. only a few metres away from bus bars
carrying several thousand amperes. To-
Partially closed control valves or open
day's EMFs are generally immune to
covers and similar disruptive elements
electrical and magnetic disturbances.
upstream of the EMF can cause sig-
However, if an EMF is to be used in such
nificantly greater measurement errors
an area, the type selection and installa-
despite the greater upstream distance,
tion site should first be discussed with
see Section 8.5. That is why control ele-
the relevant manufacturer just to make
ments and shutoff devices such as valves,
sure.
slide valves and gate valves may only be
installed downstream of an EMF and Accessibility
never upstream. Such elements often It is now seldom necessary to remove
cause cavitation which not only causes an EMF. Nevertheless, free accessibil-
damage to the control and shutoff valves, ity to the measuring device should be
but it can also cause additional meas- ensured. Compact EMFs in particular
urement errors and potentially unsteady should be accessible in the event that
flow readings due to the formation of parameter settings (e.g. changing the
gas bubbles. measuring range end value) need to be
changed or the electronic unit has to be
Control processes that rapidly reduce
replaced (e.g. to upgrade or in the case
the flowrate lead to a short-term but
of a failure).
considerable pressure drop and even
vacuum shocks which can damage the
EMF liner, see Table 9.

113
5 Electromagnetic flowmeters planning

Installing an EMF In order to facilitate the installation and


When installing the EMF in a horizontal removal of an EMF with a nominal size
pipeline, the electrode axis must be nearly from DN 200, as well as to compensate
horizontal. Otherwise, entrained gas may for low pipeline tolerances, it is recom-
collect on the top electrode or deposits mended that adaptor slip sections is
may settle on the lower electrode. Both installed downstream of the EMF. Install-
situations can lead to measurement er- ing upstream of the EMF may result in
rors or measurement failure. eddy shedding, which would negatively
affect the measuring accuracy.
With EMFs featuring PTFE, PFA, neo-
prene or polyurethane liners, the liner Following their installation in the pipeline,
usually covers the raised faces of the the sealing stoppers in the cable glands
EMF flange. If this is the case, additional may only be removed once the cables
gaskets are not necessary as these liners need to be inserted.
are sufficiently elastic. This is provided
The cable diameter must lie within
that the mating flanges are positioned
the clamping range of the cable gland,
exactly parallel to one another.
e.g. with M20x1.5 cable glands, between
The torque values for the flange bolts 7 and 13 mm. The cap nuts must be
specified in the EMF manufacturer's tightened very carefully. Any open con-
installation instructions, or delivered nections are to be sealed with a plug.
separately, should be strictly followed Otherwise, leaks may occur and the
otherwise the liner in contact with the protection category is no longer valid.
raised faces of the flange or gasket Humidity penetrating the inside can also
may be damaged and this may lead cause the EMF to fail.
to leakage.
Natural and synthetic elastomer liners
and the display of the signal converter
must be protected from prolonged ex-
posure to strong UV rays.

114
Electromagnetic flowmeters planning 5

5.3.2. Signal converter

When selecting a signal converter,


the choice is between a compact and a
remote version.

With a compact version, the primary head


and the signal converter form one unit,
3 2 1
as shown in Fig. 48.

Fig. 49: EMF remote version


(KROHNE, OPTIFLUX 4300W)
1 Output signal e.g. 420 mA
2 Power supply
3 Field current + electrode signal

Remote versions are appropriate in the


Fig. 48: EMF in compact version following situations.
(KROHNE, OPTIFLUX 4300C)
1 Power supply
2 Output signal e.g. 420 mA Protection from submersion
If submersion for more than 30 min-
With the remote version, the signal con-
utes or a hydrostatic head of more
verter is installed away from the primary
than 1 m is expected, protection cat-
head. They are connected to each other
egory IP67 is no longer sufficient. In
via signal and field current cables, as
this case, a primary head with protec-
shown in Fig. 49.
tion category IP 68 must be used, see
Section 5.2.

115
5 Electromagnetic flowmeters planning

Remote signal converters (including ability of failure then increases by a factor


those in protection category IP67) must of approx. 30, as shown in Fig. 50.
then be installed above the area at risk
For this reason, signal converters must
of submersion.
be protected from sunlight and other
Protection from excessive tempera- heat sources such as furnaces.
tures
With compact EMFs, on the other hand,
As the temperature increases, so too
high process temperatures can also heat
does the failure rate of the electronic
up the signal converter housing and
components.
thus the electronics contained within.
For example, direct sunlight can heat up In such cases, a remote version EMF is
a signal converter to over 80C. The prob- recommended.

200 40

150 30

1 100 20 2

50 10

0 0
20C 30C 40C 50C 60C 70C 80C
3

Fig. 50: Service life of electronics depending on 1 MTBF (average service life in years)
the temperature 2 Number of failures per year for 100 devices
3 Temperature of the signal converter housing

116
Electromagnetic flowmeters planning 5

Protection against vibration Despite such conditions, it must be pos-


Vibrations can damage the signal convert- sible to lay the cables without breaking
er electronics. To avoid damage, signal in the cold and the cable should retain its
converters should be firmly attached to insulation value for years. They should
a vibration-free mounting support. With be flame resistant and have multiple
compact variants, the pipeline on either shields to protect against stray pick-
side of the EMF must be supported if ups from external fields. In addition,
there is a risk of vibration or shock. the specific capacity must be small to
enable great lengths in the intrinsically
Accessibility
safe electrode circuit.
Regardless of any protective measures,
good accessibility to the EMF must al- EMF manufacturers like KROHNE rec-
ways be guaranteed. The signal con- ommend the use of type-tested cables
verter must also be freely accessible. manufactured to specifications that
remote version signal converters with take into account the relevant operat-
local display should be installed at eye ing conditions.
level, if possible.
The signal cable must be permanently
5.3.3. Signal cable wired in a location free of vibration. The
cable preparation and connection must
The electrode signal cable of remote
take place according to the EMF manu-
version EMFs is a component whose
facturer's instructions.
influence on the function of the EMF
is often underestimated. Signal cables Cable entries must be professionally
are intended to transfer the measuring sealed, otherwise the protection cat-
signals with millivolt amplitudes with an egory of the EMF is not maintained.
accuracy in the microvolt range.

Signal cables are often exposed to harsh


conditions in the process. They may lie
outside for many years in the rain, on
wet ground, in snow at -40C or exposed
to the UV rays of the sun in Australia.

117
5 Electromagnetic flowmeters planning

In addition, signal cables should not be 5.4. Grounding


routed next to high voltage cables.
Like all electrical equipment, EMFs must
For more information on routing be grounded according to the applicable
signal cables, especially when cov- safety regulations. Some examples of
ering great distances between the this are protective grounding and equi-
signal converter and the primary potential bonding during use in hazard-
head, see Section 8.3.3. ous areas.

Differences in potential between the


process liquid and the signal converter
can interfere with the measuring signal.
The signal voltage is typically in the mV
range and the signal converter can only
process such small signals without inter-
ference and in high resolution when the
difference between the potential of the
process liquid and that of the reference
in the first signal processing stages is
not too great.

There are several possible methods


available to ensure equipotential bond-
ing, including the classic methods of
grounding the process liquid. More re-
cent grounding methods, on the other
hand, do not require the process liquid
to be grounded. For more details on the
different grounding methods, refer to
Section 8.7.

118
Electromagnetic flowmeters planning 5

5.5. Surge Protection

When using EMFs in regions at risk of


lightning, measures to protect against
power surges are recommended.

This applies in particular to cases where


the EMFs are installed in the open air
and where their network and inlet/out-
let pipes are connected to plants that
exhibit a different grounding potential,
such as in underground pipelines and
sewage treatment plants.

Protective methods and elements may


vary according to the manufacturer, type
of device, installation site and the number
of inputs and outputs used. For more
details refer to Section 8.8.

119
6. Quality assurance at KROHNE
6 Quality assurance at KROHNE

KROHNE EMFs are rugged industrial 6.1. KROHNE burn-in test


measurement devices that, after more
The burn-in test is conducted as stand-
than 50 years of industrial experience,
ard at KROHNE on every device prior to
boast high measuring accuracy, reliabil
delivery and contributes to high reli-
ity and stability. These features are the
ability and thus high plant availability
result of continuous further development
at the operator.
in technology, production and quality
assurance. During this testing procedure, all signal
converters are placed in a climatic expo-
KROHNE was the first EMF manufacturer
sure test cabinet where they are exposed
to receive the ISO-9001certification, back
to cyclical temperature changes from
in 1991. This is an indication of the im-
-20C to +60C. Fig. 51 illustrates the
portance KROHNE has always attached
principle of this testing method.
to quality assurance.
During the test, a computer documents,
For quality assurance, at KROHNE, each
evaluates and monitors the output signals
measuring device must pass a number
and other measurements for compliance
of inspections and tests. Not only do
to specified quality characteristics in-
these tests ensure compliance with the
cluding temperature coefficients, drifts,
specified technical data but they also
and failures, etc. are monitored.
guarantee precision and reliability, even
under harsh conditions. The stress caused by this temperature
change accelerates early failures of weak
The burn-in test and EMF calibration
components at this in-house stage. These
are among the testing methods used
failures then no longer occur in opera-
at KROHNE. Both tests are performed
tion, significantly increasing operational
as routine tests on each device prior
reliability.
to delivery.

122
Quality assurance at KROHNE 6

Fig. 51: Burn-in test


2 3 using the KROHNE
IFC 300 as an
example
1 Climatic exposure
test cabinet
2 Input signals
3 Output signals

Evaluating premature failures gives


KROHNE insight into the quality of the
parts, components and production meth-
ods used. This test enables corrective
measures to be implemented early on and
ensures adherence to the high degree of
measuring accuracy even with changing
ambient temperatures, including smooth
operation outside during cold winters
and hot summers.

123
6 Quality assurance at KROHNE

6.2. KROHNE calibration The difference between the exact known


calibration volume and the volume meas-
Before they leave the factory, every single
ured by the EMF is described as the
flowmeter is wet calibrated by KROHNE
measurement error and is documented
on test rigs that are traceable to national
in the calibration certificate, see Fig. 53,
and international standards.
which is included with every KROHNE
EMF calibration at KROHNE is done as a EMF. This means that its calibration
direct comparison of volume a method is recorded against an internationally
used at KROHNE for more than 40 years. traceable standard.
The low uncertainty (0.02% or 0.04%
At KROHNE, calibration rigs for EMFs
depending on the system) guarantees
and ultrasonic flowmeters as shown in
a very small tolerance for the calibra-
Fig. 54 are accredited by RvA, the Dutch
tion data.
Accreditation Council, to ISO 17025. All
The EMF is installed into the measuring KROHNE EMFs are calibrated under ref-
line of the calibration tower for calibra- erence conditions similar to the EN 29
tion, as shown in Fig. 52. The calibration 104 standard.
tower is then filled. Once the valve has
The methods and standards applied to
been opened, the water flows from the
each calibration are monitored by NMi
calibration tower through the EMF. The
(the Dutch Metrology Institute) and are
calibration tower is divided into officially
traceable to international standards.
calibrated and precisely documented
This is the basis for the traceability of
partial volumes by way of 24 precision
the calibration results in the calibration
level switches.
certificate.
As soon as the water level has passed
the upper level switch, it sends the EMF
volume pulses through to the pulse
counter. When the water level passes
the preselected lower switch set by the
calibration program, volume counting
is shut off.

124
Quality assurance at KROHNE 6

1 7
6

12
3 9

10
5

11 13
4

Fig. 52: Setup of KROHNE Altometer calibration 7 Computer


rig for EMFs to DN 3000 8 Calibration vol. true value * 200,000.0 Litre
1 Calibration volume 9 Data output calibration certificate
2 Start calibration 10 Control valve to maintain constant flow
3 End calibration 11 EMF (meter under test)
4 Calibration tower with precision level switches 12 Volume pulses from EMF, e.g. 1 pulse / Litre
Height = 43 m/volume = 500 m3 13 Collecting tank
5 Flow straightener
6 Pulse counter measured vol. 200,000.0 Litre

The flow calibration system at KROHNE The uncertainty is less than 0.04% at
Altometer is the largest and most ac- flowrates of 18 m3/h to 18,000 m3/h. The
curate in the world. Here, EMFs and uncertainty of the KROHNE calibration
ultrasonic flowmeters up to DN 3000 or rigs is considerably less than the error
120 can be calibrated. The maximum limits of the measuring devices to be
3
flowrate is more than 30,000 m /h tested.
3
(approx. 10 m /s).

125
6 Quality assurance at KROHNE

Fig. 53: Calibration certificate (KROHNE, OPTIFLUX 2000)

126
Quality assurance at KROHNE 6

Fig. 54: Calibration rig (KROHNE, Netherlands)

127
7. The fundamental principles of fluid mechanics
7 The fundamental principles of fluid mechanics

Chapter 1 contained a brief introduction 7.1. Flow measurement


to the major parameters in flow meas-
Superior measuring technology is now
urement. They are repeated here.
a prerequisite for exact processes, e.g.
Based on these parameters, what fol- when producing and metering any liquid
lows is a closer look at the fundamental or gaseous product.
principles of fluid mechanics.
In addition to temperature and pres-
First of all, the basic laws of fluid me- sure, flow is one of the most important
chanics such as the continuity equation measurements in many processes.
and the principle of energy conservation
Flow measurement also involves meas-
are explained.
uring both volume V and mass m. These
Then, the different types of flow that may two measured values are related via the
occur in a pipeline are indicated. They density :
range from laminar to turbulent.
m
An explanation of the Reynolds number (9) Density = Mass =
Volume V
Re, an important measure of whether a
flow is still laminar or already turbulent,
With most processes, instantaneous
follows. This number is also required
values are required, in other words
when calculating pressure loss in the
the flowrate as volume or mass per
case of turbulent flows.
unit of time, e.g. in m3/h or in kg/min.
Finally, there is an example of how the qV represents the volumetric flow and
pressure loss in a pipeline can be calcu- qm the mass flow.
lated with different types of flow, from
laminar to turbulent.

130
The fundamental principles of fluid mechanics 7

Many flowmeters are equipped with 7.2. Basic laws of fluid mechanics
totalisers. These can be set as preset
The basic laws of fluid mechanics ap-
counters in order to automatically con-
ply to flow in pipelines. However, they
trol filling processes. For volume flow
only apply under the assumption that
counting, the flowrate q is integrated
we are dealing with the ideal process
from time t1 to t2:
liquid, which

is incompressible;
(10) V = t2 qv(t) dt
t1
has a constant density;
creates no friction losses.

7.2.1. Continuity equation


(11) m = t1 t2 qm(t) dt
The flow law, also known as the continu-
ity equation, states that in a closed loop
where
any change in the cross section always
V Volume results in a change in the flow velocity.
m Mass Cross section A and flow velocity v are
qV Volumetric flow thus inversely proportional. The flow
qm Mass flow volume qV always remains constant.
The continuity equation states:
The mass m is not affected by tempera-
ture or pressure. However, the volume

V and the density of liquids are slightly (12) qV = A1v1 = = Anvn = const.

affected by the temperature. The impact
of temperature and pressure is even
greater on gases. The flow values here
are based on a standard state, e.g. at
20C and 1.013 bar.

131
7 The fundamental principles of fluid mechanics

7.2.2. Principle of energy conservation Based on the law of flow from Section
7.2.1, the process liquid is acceler-
Two types of pressure exist in every flow.
ated when the cross section narrows
They are known as static pressure Ps and
and slowed when the cross section ex-
dynamic pressure Pd. Static pressure Ps
pands. For this, working pressure W is
acts in all directions. Dynamic pressure
required. Working pressure is generally
Pd, also known as back pressure, acts
defined as:
only in the direction of flow.

As the flow velocity v increases, so too
(14) W = Fs = PAs = PV
does the dynamic pressure Pd, while the
static pressure Ps drops. The principle
of energy conservation, also known as where
Bernoulli's Law, applies to the total
F Force
pressure P. This law states that in a
P Pressure
stationary flow, the sum of the static and
A Cross-sectional area
dynamic pressure remains constant at
s Path
any point. The principle of energy con-
V Volume
servation looks like this:
Following a change in the cross section
between, for example, pos. 1 and pos. 2 in
(13) P = Ps + Pd = const.
Fig. 55, the working pressure is thus:

Fig. 55 illustrates the effects these basic


(15) W = W1-W2 = (Ps1-Ps2)V
laws of fluid mechanics have on a flow
in a closed pipeline with a change in
cross section. Assuming an ideal process liquid, this
working pressure W is completely con-
verted into acceleration energy, i.e. into
kinetic energy Wkin.

132
The fundamental principles of fluid mechanics 7

Pd1 Pd2 Pd3

Ps1 Ps3
Ps2

A1 A2 A3
v1 v2 v3

Fig. 55: Explanations regarding the basic laws of


fluid mechanics

The acceleration energy is generally Merging equations (15) and (18)


defined as: results in:

1 2 1
2 1 2 2
(16) Wkin = /2mv = /2Vv (19) (Ps1-Ps2)V = /2 V (v2 - v1 )


Following a change in the cross section, and thus:

for instance, between pos. 1 and pos. 2
in Fig. 55, the addition of kinetic energy 1 2 1
2
(20) Ps1+ /2 v1 = Ps2+ /2 v2
results in:


This means that the total pressure at
(17)
Wkin = Wkin2-Wkin1
both positions always remains con-
stant. Bernoulli's Law can thus be taken
and thus to: from equation (13) and formulated as
follows:
1 2 2
(18) Wkin = /2 V (v2 - v1 )
1 2
(21) P = Ps+ /2 v = constant

133
7 The fundamental principles of fluid mechanics

The component PS stands for the static 7.3. Flows


pressure component created by the move-
The flow in a pipeline can be either lami-
ment of tiny particles in the process liquid,
nar or turbulent. Both types of flow are
which acts equally in all directions.
explained in more detail below. Math-
/2 v stands for the dynamic pres-
1 2
ematical principles can only be analyti-
sure component which, accordingly, only cally determined with laminar flows.
acts in the direction of flow.
7.3.1. Laminar flows
Equation (20) shows that the static pres-
Flow is considered to be laminar when
sure downstream of a reducer is lower
no vortex is created. An internal fric-
than it is upstream:
tion is created in the process liquid as
a result of the action of force between
1 2 2
(22) Ps2 = Ps1 - /2 (v2 - v1 ) the molecules. This force, also known

as viscosity, is especially great when the
molecules are difficult to move.

The example in Fig. 56 illustrates the


laminar movement of adjacent layers as
a result of internal friction. This prop-
erty is termed dynamic viscosity . The
example assumes that the fluid adheres
to both of the grey plates, i. e. layer v0
rests, and to the permanent plate too
(v = 0 m/s), and the layer v moves with
the velocity v of the moving plate.

Between the two parallel grey plates


with the surface A, there is fluid at a
distance of x, shown here as a model in
6 layers. Ideally, the velocity decreases

134
The fundamental principles of fluid mechanics 7

1
v0
(23) F ~ A; F ~ v; F ~ 1
v1 x
v2
v3 x
v4 The following is valid:
v

2 v

(24) F ~ A v
Fig. 56: Example for layers of fluids x
1 permanent plate v = 0 m/s
2 moving plate

Accordingly, the force F is:


linearly from the moving to the station-
ary plate. The lowest layer v exercises

tangential force on the layer v4 , which (25) F = A v
x
then moves on with the velocity v4. This
fluid layer acts on the one above it and
Which means, the dynamic
results in velocity v3. In this way, each
velocity :
layer accelerates the next and is then
in turn slowed down in accordance with

the reaction principle. Thus, in order (26) = F . x
A v
to move the moving plate, a force F is
necessary. This force is proportional
The result is the shear stress :
to surface A of the plate, its velocity v
and to the distance x between the two

plates: (27) = F = v
A x

Dividing by the density results in


the dynamic viscosity :

v
(28) =
x

135
7 The fundamental principles of fluid mechanics

The dynamic viscosity of liquids decreases 7.3.2. Turbulent flows


as temperature increases.
A flow is turbulent as soon as a vortex
The consequence of these laws for pipe- is created. In the process, great flow
lines is that the process liquid molecules resistances occur as well as forces that
move in layers parallel to the axis when work against the direction of move-
the flow is laminar. The velocity v changes ment of the process liquid, thus slow-
with the radius r of the pipeline, that is ing it down. Flow resistance increases
at the same distance ri to the tube axis, with the square of the flow velocity.
the velocity vi is the same. The velocity is Fig. 58 shows the flow profile and the ve-
greatest in the middle of the pipeline and locity distribution in a turbulent flow.
at the walls of the tube it is v = 0 m/s.
Fig. 57 illustrates the flow profile and
velocity distribution in a laminar flow.

v Fig. 58: Turbulent flow

Fig. 57: Laminar flow

136
The fundamental principles of fluid mechanics 7

7.4. The Reynolds number Re transition from laminar to turbulent flow


takes place. Thus, if Re > 2320, the flow
The Reynolds number Re is a dimension
is turbulent. In practice, this transition
less number that gives a measure of the
point is dependent on a variety of basic
ratio of inertial forces to viscous forces.
conditions, e.g. upstream disturbances
This number allows comparisons be-
or vibrations in the pipeline.
tween process liquids and also allows
statements to be made about the pattern
of the flow profile. The Reynolds number
Re is also an important parameter when
it comes to calculating pressure losses
in pipelines.

To maintain a constant Reynolds number


with a constant pipeline diameter, the
flow velocity v , for example, must be
increased if the dynamic viscosity in-
creases.

The following applies to pipelines:


(29) Re = v D

where

the density of the process liquid


v the mean flow velocity
D the diameter of the pipeline
the dynamic viscosity

As long as Re < 2320 the flow is laminar.


This is the critical value at which the

137
7 The fundamental principles of fluid mechanics

7.5. Pressure loss in incompressible The following applies to a pressure loss


flows P between two points in a straight,
closed pipeline:
Pressure losses caused by flow and fric-
tion resistances always occur in flows.
The extent of these losses in pressure (30) P = P1 - P2
is determined by, amongst others, the
type of flow in the pipeline. The laws
The flow resistance R is generally de-
applicable to pressure losses in both
fined as the pressure loss P per flow
laminar and turbulent flows are de-
volume qV:
scribed below.

Pressure loss and flow resistance in


(31) R = P
laminar flows qV

Any pressure loss in a laminar flow is


primarily caused by flow resistance, as According to Hagen-Poiseuille, the fol-
shown in Fig. 59. lowing simplified equation applies:


r (32) R = 8 l
. r2

This results in a pressure loss P:


(33) P = R qV = 8 l
P1 P2


.r 2

Fig. 59: Flow resistance



The pressure loss P is thus proportional
to the length of the pipeline l.

138
The fundamental principles of fluid mechanics 7

Pressure loss and pipe friction co This results in a pressure loss coefficient
efficient in turbulent flows , which can be defined as:
Mathematical laws cannot be analytically
determined in a turbulent flow. Of par-
(35) = k Re k
7/8

ticular note here, among other things, is D D


the temporally non-constant mean flow
velocity as well as the surface property The starting values for this pressure loss
or roughness of the pipeline walls. coefficient in turbulent flows amount
to approximately 0.22.
The flow is laminar in a thin boundary
layer near the wall. Here, a distinction Table 13 contains a few empirically de-
is made between hydraulically smooth termined formulas which can be used to
and hydraulically rough as well as a calculate the pipe friction coefficient
transition area. for both a hydraulically smooth as well
as a hydraulically rough area and for
In addition to the Reynolds number Re,
the transition area.
the pipe friction coefficient is also an
important parameter for which the fol-
lowing generally applies:


k
(34) = f (Re, )
D

where

k mean height of all wall unevenness


D diameter of the pipeline
Re Reynolds number

139
7 The fundamental principles of fluid mechanics

Hydraulically smooth Transition area Hydraulically rough

Pipeline coefficient :

= f (Re) = f (Re, k/D) = f (k/D)

for Re < 2320 (according to


(according to Colebrook): (according to Nikuradse):
Prandtl):

-2 k 18.7 D -2
= [2lg (Re ) - 0.8] = [1.74 - 2lg (2 + )] = [2lg ( ) + 1.138]
D Re . k
5
for 2320 < Re <10 (according to Pham):
(according to Blasius):

0.3164 k 4.52 7 k
= = [2lg ( - lg + . )]
4
Re 3.7 . D Re Re 7D

5 8
for 10 <Re <10
(according to Herrmann):

-0.237
= 0.0032 + 0.221 Re

Pressure loss coefficient :

5 < < 225 > 225


<5

Table 13: Determining the pipe friction coefficient


in turbulent flows

The pressure loss P can then be ap-


proximately calculated according to:

2
(36) P = l v
D 2

with l length of the pipeline

140
8. The theory of electromagnetic flowmeters
8 The theory of electromagnetic flowmeters

As Chapter 2 described the fundamentals 8.1. Measuring principle


of electromagnetic flow measurement
Electromagnetic flow measurement is
in detail, this chapter offers a more
based on Faraday's law of induction. Ac-
thorough look into the scientific theory
cording to this law, a voltage is induced
behind the EMF.
in a conductor when it moves through a
First, we will review the function of the magnetic field. The functional principle
EMF measuring principle. of electromagnetic measuring devices
is also based on this law of nature.
Then, we will go into the physical back-
ground of the basics of the electromag- A voltage is also induced when a con-
netic measuring principle. ducting fluid flows through the magnetic
field of an EMF, as shown in Fig. 2.
This is followed by an introduction to
the creation and processing of the EMF In a tube with a diameter D, the process
measuring signal, making clear how liquid flows through a magnetic field
EMFs developed from expensive and created perpendicular to the direction of
sensitive individual devices to sturdy, flow with a strength equivalent to B. Due
maintenance-free process and preci- to its movement through the magnetic
sion measuring devices. field, an electrical voltage is induced
in the process liquid. The induced volt-
Finally, the theory surrounding signal
age U is thus proportional to the flow
converters, frequently occurring flow
velocity v and thus also to the volume
profiles, empty pipe detection, ground-
throughput.
ing and surge protection are delved into
in detail.

144
The theory of electromagnetic flowmeters 8

Fig. 60: Measuring principle of


electromagnetic flowmeters
1 B = induction (magnetic field strength)
2 D = tube diameter
3 v = mean flow velocity
4 U = voltage = k x B x
vxD
k = device constant

1 U
4
2
3

With k as the dimensionless device con- The following applies to a circular tube
stant, the voltage U is: cross section:

2
(37) U = k B
vD (38) q =
v D / 4

where Making the displayed volume flow q:

k device constant

B magnetic field strength (39) q = U D
4.k.B
v mean flow velocity

D tube diameter
The induced voltage signal is then picked
up via two electrodes in conducting con-
tact with the process liquid and supplied
to a signal converter.

145
8 The theory of electromagnetic flowmeters

The signal converter eliminates interfer- 8.2. Physical background


ing signals and amplifies the measured
Faraday discovered the law of induction
value to make suitable measuring signals
in 1832. This law describes a voltage
available at its outputs for process control
U induced in a conductor as it moves
e.g., an active current of 420 mA.
through a magnetic field:
The magnetic field in the primary head
is generated by two field coils which
(40) U = (v B) L
are supplied with an almost rectangular
current from the signal converter. This
current accepts alternating positive and where
negative values. Alternating positive and
U induced voltage (vector)
negative flow-proportional signal voltages
B magnetic field strength (vector)
Ui are created by the flow-proportional
L length of the conductor moved
magnetic field strength B. The signal
v velocity of the conductor moved
converter subtracts these positive and
(vector)
negative signal voltages present at the
electrodes from one another. This proc- Following this discovery, in the same year,
ess always occurs when the field current Faraday attempted to measure the flow
has reached its stationary value, sup- velocity of the Thames by determining
pressing the induced interfering volt- the voltage induced in the flowing water
ages and slowly changing (compared to by the earth's magnetic field.
the measuring cycle) external or noisy
B. Thrlemann and J. A. Shercliff were
voltages.
the first to investigate the properties of
electromagnetic flowmeters.

146
The theory of electromagnetic flowmeters 8

For a theoretical model with an infinitely


long homogeneous magnetic field and (42) U = (W B) v dx dy dz
point electrodes, it was established that x, y, z

the measuring voltage is independent of


This space integral describes the area
the flow profile in the measuring tube,
of the measuring tube permeated by
provided the flow profile is radially
the magnetic field. This integral cannot
symmetrical. On these assumptions,
be solved for general purposes but a
we obtain the flow-proportional signal
solution was found for the theoretical
voltage U as:
model and for rotation symmetrical flow
profiles.

(41) U = k B
vD
The valence vector W determines the
contribution of the finite elements of flow
where towards the signal voltage as a function
of their location in the measuring tube.
k device constant
According to this, the total signal voltage
B magnetic field strength
U can be approximated as the sum of
v mean flow velocity

the contributions of the finite elements


D tube diameter
of flow between the electrodes. Fig. 61
Shercliff recognized that the contribu- shows the components of the valence
tion of the finite elements of flow in vector W in the direction of the electrode
the measuring tube towards the total axis on the electrode level.
signal voltage is weighted as a factor
of their location in the measuring tube,
and created the term valence vector.
Proceeding from Maxwells equations,
he showed that the following applies to
the electrode signal voltage U:

147
8 The theory of electromagnetic flowmeters

0.5

0.65

0.85

2.0 1.2 1.0 1.2 2.0

0.85

0.65

Fig. 61: x-component


0.5
of the valence vector W
according to Shercliff
on the electrode level

Designs following this Shercliff theory In order to simulate the basic condition
have the advantage that the rotationally of having an "infinitely long and homo-
symmetrical changes of the flow profile geneous magnetic field", extremely long
such as the transition from a laminar to field coils were used. These were simi-
turbulent flow profile, have no influence lar in design to the windings of large
on the measuring accuracy. electric motors.

Interest in the EMF grew in the industry in However, in some respects this design
the years following 1960. The basic condi- proved impractical for industrial use.
tions of the Shercliff theory determined These included the extremely long
the practical design of electromagnetic length of the EMF needed for large
flowmeters until approx. 1967. nominal sizes (more than five times

148
The theory of electromagnetic flowmeters 8

the inside diameter of the tube), the Many of the literary sources contain in-
great weight and especially the high vestigations into inhomogeneous mag-
manufacturing costs which increased netic field patterns with the objective of
significantly for the larger nominal sizes reducing the effect of asymmetrical flow
needed for this field of application. In profiles on the measuring accuracy of
addition, this version exhibited noticeable the EMF. They also offer a more detailed
measurement errors in the asymmetri- overview of the theory of electromagnetic
cally distorted flow profiles due to the flowmeters and its practical versions.
high valence near the electrodes. The
EMFs with inhomogeneous magnetic
considerably higher sensitivity close to
fields have been on the market since
the electrodes is the result of the basic
around 1967 but work on the theory
conditions of the Shercliff theory.
and testing of properties continued for
Inhomogeneous magnetic field many years.
The idea of using inhomogeneous mag-
netic fields to reduce the induction B near
the electrodes was one thing that led to
this development. The result is a smaller
term near the electrodes, reducing the
influence of asymmetrically distorted
flow profiles on measuring accuracy.

Another contributing factor was that


shorter field coils could be used. This
made it possible to reduce the length of
large EMFs from what had typically been
five nominal sizes to just one nominal
size in the case of very large DNs.

These were the first steps towards rea-


sonably priced EMFs for a wide range
of applications.

149
8 The theory of electromagnetic flowmeters

8.3. Signal generation and process- 8.3.2. AC and DC fields


ing
One way to distinguish between the inter-
The flow-proportional signal voltage at ference dc voltage and the signal voltage
the electrodes amounts to only a few is to deliberately vary the signal voltage
mV, possibly even only a few V when over time by modulating the strength of
flow is very low. the magnetic field or induction B, i.e.
of the field current in the coils of the
8.3.1. Internal resistance
primary head.
The available power of this signaling
With the induction B = 0, there is a signal
circuit is generally some 1018 W to
voltage U = 0. If the field current through
1012 W. For reliable and interference-
the coils is increased, induction B and
free generation and transmission of such
thus signal voltage U increase accord-
small signals, special measures are re-
ingly. When the coil current is reversed, in
quired such as shielding of the signal
other words B is inverted, U will likewise
cable and grounding of the sensor.
reverse the sign. This effect is exploited
The signal voltage picked up at the in various forms in EMFs to release the
electrodes can be superimposed by signal voltage from the electrochemical
electrochemical interference voltages interference dc voltage.
that are formed at the interface between
EMF with sinusoidal ac field
electrodes and the process liquid. These
The first industrial EMFs had field coils
interference voltages can be over 100
simply connected to the line ac voltage.
mV and are exponentially greater than
the signal voltage to be evaluated. Ad-
ditionally, line-frequency interference
voltages are often superimposed on the
signal voltage.

150
The theory of electromagnetic flowmeters 8

The line-frequency and sinusoidal field the signal voltage U, making it erroneous.
current generates a line-frequency and This results in faulty measuring results.
sinusoidal magnetic field which means The signal converter cannot distinguish
that the following applies: between the interference voltage and
the signal voltage because both voltages
have the same frequency and waveform
(43) B(t) = B sin( t)
but no rigid phase relation.

The eddy currents in the wall of the meas-


Accordingly, the following applies to the
uring tube have an additional negative
induced signal voltage U :
effect. They generate their own magnetic
fields which oppose the signal field of
(44) U(t) = k B sin( t)
vD the coils, thus weakening it. The strength

of the magnetic fields depends on the
electrical conductivity of the measuring
This signal ac voltage U(t) is very easy
tube which, in turn, is heavily dependent
to distinguish from the electrochemical
on the temperature when it comes to
dc voltage and can be further processed
stainless steel tubes. This causes ad-
without being affected by it.
ditional temperature coefficients and
However, interfering side effects oc- thus measurement errors.
cur with ac field EMFs. These cannot
EMFs still operated with line-frequency
be completely eliminated, thus meriting
ac fields are becoming rare and obvi-
mention here.
ously are no longer available for new
The sinusoidal magnetic field induces plants. They are very sensitive to all line-
eddy currents and thus interfering volt- frequency field currents and external
ages in all of the electrically conducting fields both on and near the pipeline. The
parts of the primary head e.g. in the wall signal converter cannot fully discriminate
of the measuring tube, in the magnetic between the line-frequency interference
plate, in the process liquid as well as on voltages and the line-frequency signal
the measuring electrodes. The interfer- voltages.
ence voltage is often superimposed on

151
8 The theory of electromagnetic flowmeters

This thus leads to errors of the display voltages on the measuring accuracy is
and makes it periodically necessary eliminated.
to calibrate the zero point with line-
Line-frequency interference voltages
frequency ac field EMFs. To do this,
are easier to suppress because the
the flow must be shut off.
field and signal frequencies of EMFs
AC field EMFs are now rarely used in with pulsed dc fields have been defined
new plants. They have been replaced as deliberately deviating from the line
almost completely by EMFs with pulsed frequency. The signal converter's signal
dc fields. processing system can therefore readily
distinguish the signal voltage with its
EMF with pulsed dc field
other frequency from the line-frequency
It was only with the introduction of the
interferences.
pulsed dc field in 1973 that EMFs became
sturdy, maintenance-free process and The electrochemical interference dc
precision measuring devices. voltage can be suppressed by using
a high pass capacitor coupling or by
For EMFs with pulsed dc fields, the field
calculating the difference between a
coils of the primary head, as shown in
succession of sampled values or by
Fig. 62, are supplied with a precisely
using more complex techniques such
controlled current that has an ap-
as the interpolation method introduced
proximate trapezoidal waveform. The
by KROHNE in 1973.
interference voltage peaks that occur
briefly due to changeover of the field This completely avoids side effects such
current are simply suppressed. This is as those experienced with EMFs with
described below. ac fields.

The signal converter does not accept the


electrode voltage until the interference
voltage peaks have decreased sufficiently.
This is the case when the field current,
and induction, are constant (as with a dc
field). The influence of these interference

152
The theory of electromagnetic flowmeters 8

1
2

0 1
0 10 20 30 40 50 60 70 80 90 100

-1

2 4

1
3

0 1
0 10 20 30 40 50 60 70 80 90 100

-1

-2

1
5

0 1
0 10 20 30 40 50 60 70 80 90 100

-1

Fig. 62: Typical time characteristic of 1 t [ms]


field current and electrode voltage in an EMF 2 Field current, induction
with a pulsed dc field 3 Sampling interval
4 Electrode voltage
5 Sampled signal voltage
U = k B v- D

153
8 The theory of electromagnetic flowmeters

8.3.3. Signal cables with bootstrap Therefore, if the signal cable is very long,
the line capacitances C1, C2 are in turn
The electrode circuit of an EMF, as shown
very great. The almost rectangular sig-
in Fig. 63, provides signal voltages with
nal voltage Ui must periodically reload
the source impedance Zi. The size of Zi
these line capacitances via the source
can be roughly estimated at:
impedance Zi (e.g. the resistance of the
liquid and also that of the deposits on
6
(45) Zi 10 the electrodes). The shape of the signal
.d
voltage Ui is rounded in the process. The
rounded voltage US then arrives at the
where
signal converter. This can lead to a no-
Zi source impedance [Ohm]
ticeable loss in measuring accuracy.
conductivity [S/cm]
d electrode diameter [cm]

2
+Us/2 +
C1
Z/2 Z/2
Ui C2 ZE
C1
-Us/2 -

(S/cm) 3

Ui Us

t t

Fig. 63: Line capacitance and signal distortion in 2 Input amplifier


long signal cables without single conductor shields 3 Signal converter
1 Primary head 4 Sampling interval

154
The theory of electromagnetic flowmeters 8

For this reason, most EMF manufacturers When there is a great distance between
provide detailed information about per- the primary head and the signal converter,
missible cable lengths depending on the when conductivity is low or when conduc-
conductivity of the process liquid as well tivity decreases strongly or in the event
as on the type of device and cable. of high-impedance electrode contamina-
tion, the so-called "bootstrap" technique
In the scope of its diagnostics, the
must be used, see Fig. 64.
KROHNE IFC 300 signal converter moni-
tors whether the signal voltage in the
signal converter has properly engaged.
If not, an error message is output.

1 2

C1 +1.0
3
C2 +
+Us/2
Z/2 Z/2
Ui ZE
-Us/2
C1 -
C2 +1.0
(S/cm)

Ui Us

t t

Fig. 64: Line capacitance and signal distortion for 3 Input amplifier
signal cable with shielding (Bootstrap) 4 Signal converter
1 Primary head 5 Sampling interval
2 Impedance converter

155
8 The theory of electromagnetic flowmeters

This also applies to remote version EMFs This "bootstrap" wiring costs more than
with very small nominal sizes. The elec- wiring with a common shield for both
trodes on an EMF with a DN 2.5 mm signal conductors. However, the boot-
have a diameter of approx. 1 mm. The strap method ensures high measurement
source impedance of the signal circuit stability and measuring accuracy, even
is thus naturally high. with unfavourable process liquid condi-
tions and with high-resistance electrode
For this reason, signal cables as depicted
contamination.
in Fig. 64 are recommended for these
nominal sizes. The additional total outer
shield is not shown here. Each signal con-
ductor has its own shield that is brought
to the potential of its signal conductor
using an impedance converter at low
resistance with the amplification V= 1.
Because there is no longer a voltage dif-
ference between the signal conductors
and their shields, no current flows via the
line capacitance C1 between conductors
and shields. These capacitances are then
virtually zero. The currents that flow
via C2 to the total shield have no effect
because they do not have any retroactive
effect on the low resistance outputs of
the impedance converter.

The result is shown in Fig. 64 below.


The signal voltage US at the input of the
signal converter is now an exact copy of
the unstressed induced voltage Ui.

156
The theory of electromagnetic flowmeters 8

8.4. Signal converter in detail span of the mA output, etc.) and converts
the scaled digital values into suitable
When it comes to electromagnetic flow
standard signals for the process (e.g.
measurement technology, signal convert-
420 mA, pulses scaled in volumetric
ers have different functions to perform,
units for volume flow counting or also
as outlined below.
digital values directly transferable via
For EMFs with bipolar pulsed dc fields, computer interfaces to process control
the signal converter also functions as a systems). These values can also be shown
supply device to generate the magnetic on the local display.
field in the field coils through an active
Since 1995, signal converters with an
current, see Section 8.3.
internal device bus (such as IMoCom =
However, the main function of the signal Internal Modular Communication) have
converter is to process the signal volt- been used, see Fig. 65. Following conver-
age. It is the signal converter's task to sion, the induced signal voltage is digit-
amplify the signal voltage without feed- ally filtered by the first microprocessor
back. The input amplifier of the signal into digital values, scaled in accordance
converter must thus be of extremely with the set operating parameters and
high impedance so that the internal transferred to an internal device bus
resistance of the electrode circuit has with different output units.
no effect on the measuring accuracy.
The amplified electrode voltage is then
converted into digital values. Filters are
then used to free the signal voltage of
any superimposed interference which
may be exponentially higher than the
measured value itself. Complex digital
filter techniques are used to do this.
The signal converter then scales the
digital values in accordance with the
specified operating parameters (such
as full scale range, size of the sensor,
157
8 The theory of electromagnetic flowmeters

1
4 3

19

8D
9C
10 B
2 6
A
7

Block diagram of the IFC 300 signal converter


(simplified)
1 OPTIFLUX primary head
2 IFC 300 signal converter 11
3 Internal KROHNE device bus
4 Field current supply
5 Primary signal processing
6 I / O, outputs / inputs:
Standard I / O: outputs / inputs:
7 A Current output / HART 13
8 B: Status / limit value / control input
9 C: Status / limit value
10 D: Frequency / pulse / status / limit value 12
16
Alternative: modular I/O with up to four output
options, such as bus, analogue and binary out- 14
17
puts freely selectable. 15
Alternative: modular intrinsically safe (Ex-i)
I/O with up to four four output options, such 18
as bus, analogue and binary outputs freely
selectable.
11 Backplane: backup of all device data
12 Display and operating unit
13 Graphic LCD display
14 Optical buttons:
Operation while housing is closed
15 Optical device bus interface
for local communication with the
signal converter when housing is closed
16 Device bus interface for local
communication with the signal converter
17 Data manager
18 Bus manager
19 Galvanically separated

Fig. 65: Block diagram of an EMF signal converter


(KROHNE, IFC 300)

158
The theory of electromagnetic flowmeters 8

The central communication of the sig- The P filters the measurement and
nal converter is the internal device bus. diagnostic values and scales them
Supported by bus and data managers, it based on the calibration data which,
connects the functional units and com- like all other settings, is also stored
municates settings, calibration data, on a backplane.
measurement and diagnostic values,
The device bus transfers all of the meas-
etc. All of the data can also be called
urement and diagnostic data to the display
up, recorded and set online via the two
on the operating unit as well as to the
device bus interfaces.
selected outputs of the I/O unit.
The microprocessor (P) of the primary
The display and operating unit also
signal processing system controls the
serves to set the measuring ranges
switching of the field current supply. It
as well as measuring and diagnostic
supplies the field coils of the primary
functions in other areas. Settings can
head with an active, periodically pulsed
be made using the menu and optical
dc current.
keys or with a PC and adaptor via the
The signal voltage induced in the proc- device bus interfaces or via HART and
ess liquid and the voltages generated fieldbus systems. During operation, the
through diagnostic functions are trans- graphics display indicates measuring
ferred via a shielded signal cable from and diagnostic data as well as totaliser
the electrodes of the primary head to values.
the input amplifier in the signal process-
Due to its modular design, the I/O unit
ing system. This instrumentation am-
allows for simple integration into opera-
plifier with extremely high-impedance
tional infrastructures through flexible
inputs amplifies the electrode voltage
selectable features and functions of the
and supplies it to the A/D converter. This
inputs and outputs down to bus interfaces
converter then samples the electrode
and intrinsically safe Ex-i outputs.
voltage synchronously to the field cycle
and converts it into digital values.

159
8 The theory of electromagnetic flowmeters

8.5. Flow profiles But no user would want to use an EMF


with a nominal size of DN 1000 or larger
Section 7.3 covered the laminar and
if the faulty installation of that EMF would
turbulent flow profile for Newtonian
only be noticed some time after start-
fluids with constant viscosity in an un-
ing up the system and where, to that
disturbed pattern. However, additional
point, everything seemed to be running
or different conditions often occur in
smoothly.
actual plants.
Reducers offer a way around this where
Due to space constraints, electromag-
the inlet run recommended by the manu-
netic flowmeters must sometimes be
facturer cannot be complied with. Fig. 66
installed with inlet runs that are too
shows the effect of an integrated reducer
short. In some cases, EMFs must also
as used with a KROHNE OPTIFLUX 5000.
be used downstream of gate or slide
The figure shows LDV measurements in
valves or covers which is not advisable
an earlier EMF installed at a distance of
for a variety of reasons, see Section 5.3.
5xD downstream of the gate valve.

Fig. 66: Smoothing the flow profile by reducing


the measuring cross section (PTB Berlin, 1991)

160
The theory of electromagnetic flowmeters 8

Below are a few examples of situations with a nominal size of DN 500 down-
where distorted flow profiles occur. stream of a partially open valve. It was
installed at a distance of just one tube
Flow profiles downstream of partially
diameter (i.e. 500 mm) upstream of the
opened gate valves
inlet flange and then at a distance of
The flow profile of the process liquid
five diameters (i.e. 5xDN or 2500 mm)
can be severely distorted by a partially
downstream of the EMF. The results are
open valve.
shown in Fig. 67.
As early as 1980, the SIREP WIB took
profile influence measurements of EMFs

%12%

%10%
3
% 8%

% 6%

1 % 4%

% 2%
5
% 0%
6
%-2% 4

%-4%
0% 25% 50% 75% 100%
2

Fig. 67: Measurement errors downstream of a 1 Measurement error [% of measured value]


partially open valve in inlet runs of 1D and 5D 2 Valve opening [%]
3 Valve 1D upstream of EMF flange, valve axis at
45 to electrode axis
4 Valve 1D upstream of EMF flange, valve axis
perpendicular
5 Valve 5D upstream of EMF flange, valve axis
perpendicular
6 Valve 5D upstream of EMF flange, valve axis at
45 to electrode axis

161
8 The theory of electromagnetic flowmeters

Here it can be clearly seen that the valve Table 14 documents the smoothing effect
must be at least 5xDN away from the inlet of a reducer on the flow profile using
flange so that the measurement remains laser Doppler measurements. In this test,
accurate even when there is intense flow the peak velocity vmax at the nose of the
restriction through the valve. profile downstream of a valve was some
200% higher than the mean velocity v.
In 1991, PTB Berlin conducted a test
The extreme values of the flow velocity
in cooperation with KROHNE to show
in the reduced measuring cross section
how well a flow profile distorted by a
of the EMF deviate only approx. 4%
partially open valve can be smoothed
from the mean value. This corresponds
by a reducer (see Fig. 66). The following
to smoothing by a factor of approx. 50.
test conditions applied:
For this reason, KROHNE recommends
3
Water at 20C, DN 100, 85 m /h installing an EMF in a reducer when-
(v = 3 m/s); ever there is not enough space for the
1st Flow profile measurement 3xDN specified unimpeded straight inlet run,
downstream of 25% open valve; which must be equal to about five times
2nd Flow profile measurement 5xDN the nominal size.
downstream of this valve in the meas-
uring cross section of the EMF with a
reduction from DN 100 to DN 80

Flow Measurement at a distance Measurement within the


velocity of 3xDN downstream of the gate reducer of the EMF
Mean value v 3.01 m/s 4.71 m/s
Peak value vmax 9.50 m/s 4.90 m/s
vmax/v 3.15 1.04

Table 14: Smoothing effect of a reducer (KROHNE,


OPTIFLUX 5000)

162
The theory of electromagnetic flowmeters 8

Flow profiles downstream of elbows


and expanders
Distorted flow profiles also occur after
elbows. For this reason, an unimpeded
inlet run of approximately five times
the nominal size is recommended for
EMFs. The measurement error caused
by the elbow is then less than 0.2% of Fig. 68: Distorted flow profile downstream of
an elbow
the measured value (see Section 3.1,
Fig. 14).

The situation becomes more problematic


after a sudden tube expansion. Here,
the distorted flow profile features an
increased peak flow in the middle of the
Fig. 69: Distorted flow profile downstream of a
tube which is caused by the reverse flow sudden expansion

near the tube wall. When the distance


Additional reasons include the low pres-
to the expansion is greater, the reverse
sure in an expanded tube as well as the
flow near the tube wall disappears but
eddy shedding at the edge. Both these
the "long nose" of the profile remains
factors increase the risk of outgassing
as a whole over longer sections.
and unsteady EMF readings. But the
This problem leads to noticeable EMF distorted flow profile alone is enough
measurement errors and is the main to noticeably affect the measuring ac-
reason why EMFs should not be installed curacy of the EMF installed downstream
downstream of expansions and that the of a tube expansion.
EMF should never be larger than the
Annular orifices have an effect on the
nominal size of the pipeline.
flow profile similar to that of a sudden
expansion in the diameter upstream of
the EMF.

163
8 The theory of electromagnetic flowmeters

This is an opportune time to refer once


again to the diagnostic capabilities of
the OPTIFLUX IFC 300 signal converter
from KROHNE. The integrated flow profile
test shows whether and how severely
the profile in the measuring tube is
distorted. At the touch of a button, the
display indicates whether installation
was done properly or if, for example,
a gasket is offset and thus protruding
into the flow. The integrated diagnostic
functions of the IFC 300 signal converter
play a role in localising such causes of
errors quickly and without interfering in
the process, resulting in accurate and Fig. 70: Swirl flow caused by
an out-of-plane double bend
efficient troubleshooting.
which the welded seams protrude to the
Flow profiles during swirl flow
inside can also cause a swirl flow.
A swirl flow is an additional tangential
velocity component in a tube flow. It is A swirl flow dissipates very slowly. De-
usually caused by several subsequent pending on the diameter of the tube, the
changes in direction in a pipeline. Fig.70 flow velocity and the liquid parameters,
illustrates a well-known example of an dissipation may require a section that is
out-of-plane double bend (i.e. more than approx. 50 to 150 times the nominal size
one elbow on different planes). of the pipeline. The larger the nominal
size of the pipe and the flow velocity, the
In practice, swirling occurs in all pipe-
greater the probability of a swirl flow
lines in which two or more fittings such
occurring. The accuracy of EMFs is not
as pumps, throttled actuators, elbows,
as severely impacted by swirl flows as
T-pieces, etc. cause a change in the di-
is generally assumed (see Section 3.1,
rection of flow. Spiral welded pipes in
Fig. 14).

164
The theory of electromagnetic flowmeters 8

However, a swirl flow may be extremely Flow profiles for non-Newtonian


disruptive if the medium is inhomo fluids:
geneous, in other words if it consists There are fluids whose viscosity is not
of two components not completely mixed constant. Instead, it changes in relation
or cured. In this case, first one and then to the shear strain or the local velocity
the other component swirls periodically distribution and the duration and previ-
past one and then past the other measur- ous history of the mechanical load. Such
ing electrode, generating considerable fluids are referred to as "non-Newtonian
spikes in the electrochemical voltage. fluids" and they include a number of
This situation can be seen in the periodic fluids whose flow is measured by EMFs.
fluctuations in the flow display. A static Examples include:
mixer may be helpful in such cases.
Activated sludge in wastewater;
The most effective way to reduce swirling Tomato ketchup, toothpaste, liquid
is to use flow straighteners. When it comes soaps (soft soaps);
to swirling, honeycomb straighteners are Mortar, cement sludge, latex paint;
generally recommended. Pulp suspensions, fillers (e.g. kaolin)
and coating material in the paper
industry.

One exceptional example can be found


in kaolin suspensions with a large por-
tion of finely ground kaolin particles.
These suspensions cause flow profiles
with very low flow velocity near the EMF
electrodes and thus significant meas-
urement errors.

165
8 The theory of electromagnetic flowmeters

Nowadays, it is usually Coriolis mass When selecting an electromagnetic flow-


flowmeters such as the OPTIMASS series meter and during subsequent opera-
from KROHNE that are successfully used tion, it is generally not known whether
for these shear thickening fluids. swirl is present, whether the profile is
distorted or whether there is enough
Non-Newtonian fluids feature a fluctuat-
space for an unimpeded inlet run. Un-
ing but very high viscosity. So, once gas or
fortunately, EMFs are often incorrectly
air has entered the fluid, it is very difficult
installed, downstream and too close to.
for it to escape. As a result, the risk of
Other than a very few exceptions, they
faulty flow indicator values due to the
still usually function well when it comes
amount of gas is very high. This risk can
to the accuracy and reliability required
only be counteracted through relevant
in normal industrial applications.
methods when filling and transferring,
when mixing and by adding additives.

If fluids have a tendency to outgas, this


must be prevented using measures such
as high pressure, smooth and step-free
pipe transitions well before the meas-
uring station.

166
The theory of electromagnetic flowmeters 8

8.6. Empty pipe detection The following methods are common:

When the measuring tube of a mechani- Empty pipe detection via the
cal flowmeter is empty, its flow indicator measuring electrodes;
indicates the value "0" and flow totalis- Full tube detection via a full tube
ing stops. electrode;
Full tube detection via the flow
This is not the case with an EMF. The
profile test;
electrodes are no longer in contact
Empty pipe shut off via external
with the process liquid and are open.
control signals.
The electrode circuit has an extremely
high impedance when the tube is empty. Empty pipe detection via the
Electrical interferences and couplings measuring electrodes and the
from the surroundings can then lead to electrode resistance
error flow readings and totalised values. When the measuring tube is completely
This is another reason why the instal- filled, the measuring electrodes are
lation site of an EMF must generally be connected to each other and against
selected so that the measuring tube is the reference point of the EMF via the
always completely filled with the process process liquid.
liquid, even when the flowrate is "zero",
The resistance between the measur-
refer to Section 5.3.
ing electrodes or from one electrode
If the EMF measuring tube is still empty, to a reference point is a factor of the
the flow outputs and indicator as well as conductivity of the process liquid. This
volume totalising via functions such as resistance is lower with a completely
empty pipe detection, empty pipe shut filled measuring tube than with an empty
off and full pipe detection must be set measuring tube. So, if the resistance is
to zero to avoid errors. low, the empty pipe detection indicates:
"pipe filled" and when the resistance is
high it indicates "pipe empty".

167
8 The theory of electromagnetic flowmeters

The switching point must be set in such Full pipe detection via a
a way that even at the lowest conductivity full pipe electrode
occurring in the process, in other words This additional electrode is attached in
at the highest occurring resistance with the top of the measuring tube. When the
a full pipe, the message "pipe filled" is EMF is installed in a horizontal pipe-
definitely indicated. Otherwise, the sig- line, this electrode indicates partial
nal converter may indicate "pipe empty" filling even if the process liquid level
when the measuring tube is still filled decreases only slightly. Here too, the
and then wrongly set the outputs and switching point must be set below the
the display to "0" and stop the totalising. lowest electrical conductivity that can
For this reason, proceed with caution occur in the process.
when setting this option to avoid incor-
The advantage of an additional full pipe
rect messages.
electrode is the early indication of a par-
The advantage of this method is that no tially filled measuring tube.
additional hardware expense is neces-
When filling or emptying the pipeline,
sary for the primary head.
large volumes of the process liquid are
The disadvantage, however, is that the not measured and totalised before the
switching process only takes place when pipeline has been fully filled or emptied
the liquid level drops below the electrode again. In addition, the full pipe electrode
axis. Therefore, when the measuring does not respond at all or only very slowly
tube becomes coated, this method no in the case of highly viscous process liq-
longer functions reliably. It also becomes uids, incrustations and coatings sludge.
problematic with vertical and long pipe- The full pipe electrode indicates "pipe
lines in which the process liquid runs not full" even if only minimal gas content
down the tube walls for an extended has collected at the top of the tube.
period of time.
In the case of long, vertical pipelines
in which the process liquid runs down
the pipe walls for an extended period

168
The theory of electromagnetic flowmeters 8

of time even when the pipe is empty, Empty pipe shutoff via external control
the full pipe detection function may still signals
indicate "pipe filled". This is the simplest and most reliable
method of empty pipe shutoff. It uses
Full pipe detection via the
information and control signals for pumps
flow profile test
and valves which exist in most plants.
When a pipeline is partially filled, the
flow profile is asymmetrical. There is Almost all EMF signal converters feature
less process liquid flowing in the upper binary control inputs. This allows the
part of the measuring tube than in the outputs to be set to "0" and the counter
lower part. to be stopped. Only the control signals
of the pumps or valves must be applied
The KROHNE IFC 300 signal converter
to the control input of the signal con-
can monitor the measuring tube for
verter.
complete filling using its diagnostic
function "flow profile test", regardless This method is simple and reasonable.
of electrical conductivity, viscosity and In addition, the empty pipe shutoff is
incrustations. Generally, the flow profile guaranteed regardless of the process
test only responds when the liquid level liquid properties such as conductiv-
has sunk below a value of approx. 75%. ity, viscosity and contamination in the
The empty pipe detector must then be measuring tube.
activated as well.
However, there is the risk that any vol-
ume that comes after the pumps have
been switched off, or while the valve was
closed, may not be measured. In addition,
the empty pipe detection must be slightly
delayed in relation to the starting of the
pump and opening of the valve so that
the pipeline is sufficiently filled before
measurement can start again.

169
8 The theory of electromagnetic flowmeters

8.7. Grounding tive inside, e.g. in the case of non-coated


steel or stainless steel electrodes, the
Section 5.4 alluded to the fact that there
fluid in the tube always has the same
is more than one possible method to
potential as the grounded pipeline. The
ensure equipotential bonding between
signal voltage at the electrodes thus has
the process liquid and the reference po-
a fixed reference potential. Fig. 71 illus-
tential of the signal processing in the
trates the simplest case for grounding
signal converter. Section 8.7.1 covers
the process liquid.
the classical grounding methods used to
achieve equipotential bonding by ground- 2
ing the process liquid. Newer methods
1
to achieve equipotential bonding without
having to ground the process liquid are
6
introduced in Section 8.7.2.

8.7.1. Classical grounding methods


4
grounding the process liquid

Classical methods for grounding the 3


5
process liquid include
FE

Grounding in pipelines that are electri-


Fig. 71: Grounding the process liquid in metal
cally conductive inside; pipelines not coated on the inside
1 Primary head
Grounding using grounding rings or 2 Terminal box or signal converter
3 Flowmeter flanges
discs; 4 Pipeline
5 Flange of the pipeline flowmeter
Grounding using a grounding elec- 6 Interconnecting cables
FE Functional ground
trode.
Grounding the process liquid using
The classical grounding methods are
grounding rings
described below.
In pipelines made of plastic or concrete
Grounding the process liquid in a pipe- or those which have an insulating lining
line that is electrically conductive inside or coating inside, additional measures
In pipelines that are electrically conduc- must first be used to bring the process

170
The theory of electromagnetic flowmeters 8

liquid to a known fixed potential. To do This method is technically reliable and has
this, metal grounding rings or grounding been successfully used for decades.
discs where the inside face is in con-
The disadvantage to this method is that
tact with the process liquid are usually
costs are higher, for example, if special
used. These grounding rings are gener-
materials are required for aggressive
ally fitted between the pipeline and the
process liquids or when grounding rings
EMF flanges. Then, they are grounded
for extremely large sizes must be used.
along with the EMF sensor, as shown in
With larger potential differences between
Fig. 72.
the process liquid and earth in a system,
2 equalising currents run via the grounding
rings and earthing conductor.
1
Grounding the process liquid using
6 a grounding electrode
In this case, a grounding electrode lo-
cated in the base of the tube is directly
4
connected to the grounded housing of
the primary head.
3
5
7 The advantage of this method is that
FE
the grounding electrode generally costs
Fig. 72: Grounding the process liquid in pipelines less than the grounding ring.
with electrically insulated walls

1 Primary head However, it is particularly disadvanta-


2 Terminal box or signal converter
3 Flowmeter flanges geous when there are differences in
4 Pipeline
5 Flange of the pipeline flowmeter potential in the system of more than
6 Interconnecting wires
7 Grounding rings 0.2 Volt, as the grounding electrodes can
FE Functional ground
be irreparably damaged by electrolytic
action. In addition, abrasive solids in a
horizontal pipeline can quickly destroy
these grounding electrodes on the base
at the tube. In both cases, complete

171
8 The theory of electromagnetic flowmeters

destruction of the EMF primary head is The crucial advantage of this method
to be expected. In some circumstances, over the classical grounding electrode is
the process liquid may even leak out. that the reference electrode is no longer
exposed to electrolytic destruction by
8.7.2. Newer grounding methods way of potential differences in the sys-
without grounding the process liquid tem. In addition, with this method it is
also possible to use ungrounded EMFs
Newer methods for grounding process
in systems where voltages and currents
liquids include
are present in the pipelines. This is the
Floating grounding electrode to transfer case with, for example, electrolysis and
the reference potential of the process electroplating plants and systems with
liquid; cathodic protection.
Virtual grounding.
Virtual grounding
These newer grounding methods are The term "virtual grounding" may sound
described below. Of particular note is like this patented method does not re-
the virtual grounding method developed ally involve grounding. That is why the
by KROHNE. more technically precise term of "virtual
reference" is often used. Problems with
Floating grounding electrode
the conventional grounding of EMFs are
With this method, the grounding electrode
what triggered the development of virtual
in the base of the tube is no longer in
grounding by KROHNE. These problems
direct contact with the grounded housing
can be summarised as follows:
on the functional grounding of the primary
head. In this case, a floating electrode When it comes to extremely aggressive
transfers the potential of the process fluids, the grounding rings used with
liquid as reference potential to a high- conventional methods must usually be
impedance input at the signal converter. manufactured using expensive special
No measurable current runs through materials.
the floating reference electrode. In extreme cases, such as grounding
rings made of tantalum, costs may
equal those of the EMF itself.

172
The theory of electromagnetic flowmeters 8

Grounding electrodes, the low-cost KROHNE developed a simple, cost-effec-


alternative to grounding rings, can be tive solution for this. "Virtual reference"
destroyed through electrolytic action or "virtual grounding" can be done without
in the face of minimal differences in the use of grounding rings or grounding
potential of e.g. only 0.2 Volt in the electrodes, as shown in Fig. 73.
system.
In the case of virtual grounding, the
In systems with cathodic protection the
primary head of the EMF is built into
risk is that the cathodic protection will
pipelines with electrically insulated walls,
be impacted via the grounding methods
without grounding rings or grounding
on the EMF. This applies analogously
electrodes. The measuring electrodes
to electrolysis systems.
are then the only metallic elements of
2 the EMF left in contact with the pro
cess liquid.
1
The input amplifier of the IFC 300 signal
converter measures the potential of the
EMF measuring electrodes and, using a
method patented by KROHNE, generates
4
a voltage that corresponds to the potential
of the ungrounded process liquid. This
3 voltage is then used as a reference po-
5
FE tential for signal processing. Thus, during
signal processing there is no longer a
Fig. 73: Virtual grounding via the device electron-
ics when installing the EMF in pipelines with disruptive potential difference between
insulated walls
1 Primary head the reference potential and the voltage
2 Terminal box or signal converter
3 Flowmeter flanges at the measuring electrodes.
4 Pipeline
5 Flange of the pipeline flowmeter
FE Functional ground The use of virtual grounding boasts
several advantages over classical
grounding.

173
8 The theory of electromagnetic flowmeters

For one thing, virtual grounding elimi- So, getting rid of grounding rings com-
nates the need for additional wetted pletely and instead using EMFs with vir-
grounding equipment such as grounding tual grounding is the most economical
rings and grounding electrodes. This solution in this case.
results in lower costs and an EMF that
EMFs with virtual grounding can also be
is much easier to install. This advan-
installed ungrounded in systems with
tage is not to be undervalued. It is often
cathodic protection, without running the
faulty or a lack of grounding altogether
risk of degrading the cathodic protection
that is the cause of errors when start-
through error currents.
ing up EMFs.
Virtual grounding facilitates and makes
In addition, potential differences in sys-
installing an EMF cheaper. The need
tems with virtual grounding do not run
for additional gaskets between ground-
the risk of destruction due to electrolytic
ing rings and flanges, as required with
action, as is often the case with ground-
the classic installation using grounding
ing electrodes. No equalising current
rings, is eliminated. The risk of leakage
runs through process liquid and ground
is also smaller.
conductor.
As a general rule, virtual grounding is
EMFs with virtual grounding can also
possible starting at DN 10 and at a con-
be used in systems in which voltages
ductivity of more than 200 S/cm. The
and currents are present in the pipeline,
conductivity of aggressive process liquids
as with, for example, electrolysis and
such as inorganic acids and caustics is
galvanic plants. Otherwise, due to the
exponentially higher.
highly aggressive media, special materi-
als such as tantalum, nickel or titanium EMFs with virtual grounding in the
must be used for the grounding rings or KROHNE IFC 300 signal converter can
grounding electrodes in such systems. be used in almost all systems in which
Grounding rings made of these special classical grounding with grounding rings
materials result in considerably higher is problematic in terms of the techno
costs for the measuring system. logy or cost.

174
The theory of electromagnetic flowmeters 8

A brief recap of the advantages of virtual 8.8. Surge Protection


grounding:
Measures to protect against surges in
Low cost: voltage are recommended when using
No grounding rings or grounding EMFs in regions at risk of electrical
electrodes required. Costs are thus storms. This is particularly applicable
lower. when EMFs are installed outdoors and
where the line and input/output cables are
No equalising currents:
wired to system components exhibiting
The reference potential is generated
a different grounding potential such as
in the IFC 300 signal converter and is
with underground pipelines or sewage
insulated against earth. No current
treatment plants.
runs through the pipeline, the pro-
cess liquid or the earth. There is thus The primary head of an EMF, or more
no equalising current in electrolysis precisely, its housing, its electrode cir-
or galvanic systems and no stress on cuit and the functional earth are at the
the cathodic protection. same potential as the functional earth
of the pipeline. The housings of the sig-
No additional risk of leakage:
nal converters are usually connected
Grounding rings with additional sealing
to the power supply protective earth.
points or grounding electrodes which run
Due to the galvanic separation, there
the risk of destruction through elec-
is no electrical connection between the
trolytic action are not necessary.
functional and the protective earth in
the signal converter. The same is true
for the signal converter outputs. Test
voltages for these galvanic separations
are usually between 0.5 kV and 1.5 kV. In
the event of a lightning strike, however,
considerably higher voltage differences
may occur between the functional and
protective earths, which could destroy
the affected devices and subsequent

175
8 The theory of electromagnetic flowmeters

instruments if appropriate additional For remote version EMFs, connect the


protective measures are not taken. protective earth connection of the signal
converter housing to the earth of the
In the case of EMFs, the best primary
primary head;
protection against surges in voltage is
Provide surge protection for all wires
the proper grounding of the pipeline,
on the input and outputs of the signal
primary head and signal converter to
converter;
a single point. Protective methods and
Surge protection also for L/N of power
elements may vary according to the
supply (not shown here);
EMF manufacturer, type, installation
Provide equipment for reliable surge
site and the number of inputs and out-
protection in close proximity to the
puts used.
EMF.
As seen in the following examples,
For larger systems, advice from a profes-
KROHNE offers fully installed equipment
sional service is recommended. Fig.74
to meet a wide variety of requirements
shows an example of voltage surge
when it comes to surge protection. Surge
protection measures for the KROHNE
protectors specific to your requirements
OPTIFLUX 2300C with integral signal
for all of the cables and system compo-
converter and with non-coated pipelines.
nents to be protected are selected.
For pipelines with insulated walls, the
For compact EMFs and remote versions additional appropriate grounding meth-
where the signal converter is installed in ods as presented in Section 8.7.2 are
close proximity to the primary head, the to be used.
following measures are sufficient:
If, in the case of remote version EMFs,
Proper grounding near the primary the signal converter is installed some
head (safety PE quality when the power distance from the primary head, signal
supply is more than 60 V); and field current cables as well as all
Do not apply PE from power supply connecting cables must be included in
cable to signal converter; the surge protection. When the distance
between the primary head and the signal

176
The theory of electromagnetic flowmeters 8

converter is great, the following addi- The protective devices should be


tional measures must be taken: installed as close as possible to the
signal converter in the same cabinet
Protect all of the wires for the signal
or shaft.
and field current cables;

Surge protective devices can influence


the measuring accuracy and should be
discussed with the manufacturer;

9 6

7 10
2
8

3 3

1
4

Fig. 74: Surge Protection in a 5 Power supply feed (disconnect PE!)


compact EMF version 6 Protected power supply line
1 EMF primary head 7 EMF signal output
2 EMF signal converter 8 Protected signal output line
3 Pipeline 9 Surge Protection for power supply
4 Lightning protection earth and protective 10 Surge Protection for signal output
earth (PE or FE) of the EMF

177
8 The theory of electromagnetic flowmeters

Fig. 75 shows an example of surge pro- For pipelines with insulated walls, the
tection measures using the remote ver- appropriate grounding methods outlined
sion of KROHNE's OPTIFLUX 2300W with in Section 8.7.2 are to be used.
pipelines with non-coated inside walls.
.

8 14 15
6
9 10
11

4
13 1

5
12
7

3 3

Fig. 75: Surge protection in a remote system 7 Signal output line


EMF version 8 Protected signal output line
1 EMF primary head 9 Surge protection for power supply
2 EMF signal converter 10 Surge protection elements for signal
3 Pipeline inputs and outputs of the EMF and field current
4 Lightning protection earth and protective 11 Assemly rail for the protective elements
earth (PE and FE) 12 Field current cable
5 Power supply line 13 Electrode signal cable
6 Protected power supply line 14 Protected field current cable
15 Protected electrode signal cable
178
The theory of electromagnetic flowmeters 8

8.9. Norms and standards VDI/VDE 2641 (no longer valid,


withdrawn 200001);
This is a list of only the most important
Electromagnetic flow measurement.
European standards applicable to elec-
tromagnetic flowmeters and no claim VDI/VDE 2641 Sheet 2 (no longer valid,
is made as to the completeness of the withdrawn 1996-11);
list. Electromagnetic flow measurement;
Installation lengths and flange con-
Specific standards for electromagnetic
nection dimensions of flowmeters with
flowmeters
flanges;
EN 29104:1993 (ISO 9104:1991) Instead, the following is recommended
Measurement of fluid flow in closed DIN ISO 13359(1995-11).
conduits;
Flange standards
Method for assessing the operating
performance of electromagnetic DIN EN 1092: Flanges and their
flowmeters for liquids; connections.
Substitute for: DIN 19200:1989-01.
General CE guidelines
DIN EN ISO 6817:1995
Pressure Equipment Directive
Measurement of conductive liquid flow
97/23/EC;
in closed conduits;
Method with electromagnetic flow Low Voltage Directive
meters. 2006/95/EC. 5. 3.;

DIN ISO 13359:1998-09 EMC Directive 2004/108/EC;


Measurement of conductive liquid flow
Machinery Directive 2006/42/EC.
in closed conduits;
Electromagnetic flowmeters with
flanges;
Installation lengths
(identical to DVGW W 420).

179
8 The theory of electromagnetic flowmeters

Directives for use in - OIML R117


hazardous areas (DIN 19217 Measuring systems for
liquids other than water);
Directive 94/9/EC (ATEX 100)
- OIML R75
SIL (safety integrity level)
Measuring thermal energy.
DIN EN/IEC 61508 / IEC 61511
Guidelines for surge protection
Housing protection categories
DIN VDE 0100-443;
(IP and other types of protection)
DIN EN 62305 and VDE 0185-305:
DIN 40050, depending on area of use also
These standards regarding lightning
DIN 40050-9;
protection provide information on all
DIN EN 60529 or IEC publication 529; aspects from "General objectives" to
"Risk management" right down to
NEMA Standard 250 -2003 (USA);
"Electrical and electronic systems
UL 50 (USA); in structural plants" and also con-
tain supplementary sheets covering
CSA-C22.2 No. 94-M91 (2006)
such content as "Lightning threat in
(Canada).
Germany".
Directives for custody transfer
VDI/VDE directives
Measurement Instruments Directive
VDI/VDE 2650:
2004/22/EC;
Requirements regarding self-monitoring
International metrological recommen-
and diagnosis in field instrumentation.
dations such as:

- OIML R49-1
(for both cold and hot water);

180
The theory of electromagnetic flowmeters 8

NAMUR guidelines NE 032 (08.01.03):


Data Retention in the Event of a Power
NAMUR is an international association
Failure in Field and Control Instruments
of automation technology users in the
with Microprocessors
process industry. NAMUR issues recom-
mendations and working sheets. NE 043 (03.02.03):
Standardization of the Signal Level
NAMUR Worksheets ("NA") provide as-
for the Failure Information of Digital
sistance in the form of checklists and
Transmitters
instructions to support member com-
panies in their practical work. NE 053 (04.02.03):
Software of Field Devices and Signal
NAMUR Recommendations ("NE") explain
Processing Devices with Digital Elec-
the state of the art and the regulations,
tronics
not only for member companies but also
for manufacturers, scientists and public NE 070 (26.01.06):
authorities. Electromagnetic Flowmeters (EMF)

Below is a selection of NAMUR work- NE 080 (14.04.03):


sheets and recommendations prepared The Application of the Pressure
specifically for electromagnetic flow- Equipment Directive to Process
meters or that are applicable to field Control Devices
devices in general.
NE 107 (12.06.06):
NA 101 (25.10.04): Self-Monitoring and Diagnosis of Field
The Calibration Requirements "in brief" Devices
for Flow Measuring Equipment
NE 131 (29.04.09):
NE 021 (22.08.07): NAMUR Standard Device Field Device
Electromagnetic Compatibility (EMC) Requirements for standard applica-
of Industrial Process and Laboratory tions
Control Equipment

181
9. Summary and outlook
9 Summary and outlook

In 2009, there were more than 3 million The wide range of use, high accuracy and
EMFs in use around the world. EMFs reliability of EMFs was made possible
play an important role, from water sup- by progress in a wide range of areas in
ply to the food and beverage industry technology.
right down to wastewater treatment.
One of these areas is, for example, coat-
The pipelines and sewer pipes of some
ing materials. The successful progress
mega-cities feature EMFs up to 3000
of EMFs in the chemical and food and
mm in size, measuring and totalising
beverage industries started when fluoro-
up to 100,000 m3/h.
plastics like PTFE and later PFA became
In steel mills, more than 100 EMFs available at reasonable prices. Only
monitor the cooling water circuits to when measuring tubes made of highly
furnaces and in strong magnetic fields dense engineered ceramics with their
to electric furnaces. appropriate stability became available
could EMFs be used in calibration rigs
Large chemical factories run several
and as flowmeters in volumetric filling
thousand EMFs around the clock. They
machines.
measure hot concentrated inorganic ac-
ids and caustics and are responsible Highly integrated switching circuits make
for mixing many products in the proper the P signal converters extremely stable,
proportions. user-friendly and provided them with a
variety of functions and interfaces.
In every pulp and paper factory, many
EMFs ensure that paper is produced in As with all measuring devices, KROHNE
an environmentally friendly manner, in will continue to take advantage of all
high quality and in sufficient quantity. of the technical trends as well as set
One large paper machine alone uses their own trends to constantly expand
up to 200 EMFs. the range of application of EMFs.

184
References

[1] Appel, E. (1979): [5] Baker, R.C. (1985):

Neue Techniken der Induktiven Theorie und Praxis der elektro-


Durchflumessung. magnetischen Durchfluss-
In: VFI Der Versuchs- und messung.
Forschungsingenieur, In: Technisches Messen, Vol. 52,
No. 3, pp. 2833. No. 1, pp. 412.

[2] Appel, E. (1985): [6] Baker, R. C. (1985):

Induktiver Durchflumesser mit The Effect on Electromagnetic


keramischer Auskleidung und Flowmeters of a Change in Fluid.
kapazitivem Signalabgriff. In: Measurement & Control,
In: Automatisierungstechnische No. 18, pp. 510.
Praxis, No. 4, pp. 199200.
[7] Bekkum, J. V.; Warffemius,
[3] Baker, R. C. (1970): M. M.; Oldenziel, D. M. (1989):

Effects of non-uniform conducti On The Hydrodynamical Stability


vity fluids in electromagnetic of Electromagnetic Flowmeters.
flowmeters. In: Harbrink, B.; Spencer, E. A.
In: Journal of Physics D: Applied (ed.): Flow Measurement FLO
Physics, No. 3, pp. 637639. MEKO, VDI, pp. 143150.

[4] Baker, R. C. (1982): [8] Bevir, M. K. (1970):

Electromagnetic Flowmeters. The Theory of Induced Voltage


In: Scott, R. W. W. (ed.): Develop- Electromagnetic Flowmeters.
ments in Flow Measurement In: Journal of Fluid Mechanics,
1, Applied Science Publishers, No. 43, pp. 577590.
London.

186
References

[9] Bevir, M. K. (1972): [13] Bonfig, K. W. (2002):

The effect of conducting pipe berblick ber die neueren


connections and surrounding Durchflussmessverfahren
liquid on the sensitivity of In: Bonfig, K. W. (Hrsg.): Praxis
electromagnetic flowmeters. der Durchflussmessung, Vol. 2,
In: Journal of Physics D: Applied b-Quadrat Buchverlag, Kreuztal.
Physics, No. 5, pp. 717729.
[14] Bonfig, K. W. (2002):
[10] Bonfig, K. W. (1970):
Magnetisch-induktive Durch-
Einflsse auf die Nutzspannung flussmessung.
bei der induktiven Durchflu- In: Bonfig, K. W. (Hrsg.): Praxis
messung. der Durchflussmessung, Vol. 2,
In: Archiv fr technisches b-Quadrat Buchverlag, Kreuztal.
Messen, Lfg. 414, pp. 145150.
[15] Brockhaus, H.; Klein, J. W.;
[11] Bonfig, K. W. ; Hofmann, F. van der Pol, R.; Schoth, U. (1997):
(1975):
Kapazitiver magnetisch-induktiver
Die Technik der magnetisch-in- Durchflussmesser unter Verwen-
duktiven Durchflussmessung. dung der Mikrosystemtechnik.
In: Archiv fr technisches In: Sensoren und Sensorsignal-
Messen, Lfg. 473, No. 6. verarbeitung, Vol. 8, p. 187.

[12] Bonfig, K. W., Feith, B. (1981): [16] Brockhaus, H.

Elektrochemische Strspan- Magnetisch-induktive Durch-


nungen bei der magnetisch-in- flussmessung mit kapazitiven
duktiven Durchflumessung in Verfahren
offenen Kanlen. In: Technisches Messen,
In: messen + prfen/automatik, No. 64, p. 5.
No. 1/2, pp. 3639.

187
References

[17] Brockhaus, H. ; Klein, J. W.; [21] Brockhaus, H.; Hofmann, F.


van der Pol, R.; Schoth, U. (1997): (2002):

Kapazitive magnetisch-induktive Spezielle Anwendungen der


Durchflussmessgerte unter magnetisch-induktiven Durch-
Verwendung der Mikrosystem- flussmessung.
technik. In: Bonfig, K. W. (Hrsg.): Praxis
In: Sensoren und Sensorsignal- der Durchflussmessung, Vol. 2,
verarbeitung, Vol. 8, p. 187. b-Quadrat Buchverlag, Kreuztal.

[18] Brockhaus, H. (2000): [22] Engl, W. L. (1970):

Magnetisch-induktiver Durch- Der induktive Durchflumesser


flussmesser, versorgt ber die mit inhomogenem Magnetfeld.
420 mA-Schnittstelle. Teil 1
In: Automatisierungstechnische In: Archiv fr Elektrotechnik,
Praxis, No. 11. No. 53, pp. 344359,

[19] Brockhaus, H. (2002): [23] Engl, W. L. (1972):

Mglichkeiten der Diagnose bei Der induktive Durchflumesser


Durchflussmessgerten. mit inhomogenem Magnetfeld.
In: Sensoren und Messdatenver- Teil 2
arbeitung, Vol. 4. In: Archiv fr Elektrotechnik,
No. 54 , pp. 269277.
[20] Brockhaus, H.; Hofmann, F.
(2002): [24] Faraday, M. (1832):

Magnetisch-induktiver Durch- Experimental Researches in


flussmesser als Zweileitergert. Electricity.
In: Sensoren und Messdaten- In: Philosophical Transactions of
verarbeitung , Vol. 4. the Royal Society, No. 15, p. 175.

188
References

[25] Hemp, J. (1975): [29] Hofmann, F.; Oehler, G. (1997):

Improved magnetic field for an Magnetisch-induktiver Durch-


electromagnetic flowmeter with flussmesser zur Abfllung
point electrodes. kleiner Volumina.
In: Journal of Physics D: Applied In: Sensoren und Sensorsignal-
Physics, No. 8, pp. 9831002. verarbeitung, Vol. 8, p. 180.

[26] Hofmann, F. (1978): [30] Hofmann, F. (1999):


Volumetric Filling Using Electro-
New Developments in Magnetic
magnetic Flowmeters.
Flow Measurement using Keyed
In: drink technology & marke-
D.C. Field.
ting, No. 9.
In: Dijstelbergen, H. H.; Spen-
cer, E. A. (ed.): Flow Measure-
[31] Hofmann, F. (1999):
ment of Fluids, North-Holland
Publishing, pp. 499502. Magnetisch-induktive Durch-
flussmessung fr teilgefllte
[27] Hofmann, F. (1978): Rohrleitungen.
In: Wasser, Luft und Boden,
Magnetic Flowmeter with Capa-
No. 10.
citive Signal Pick-Off.
In: Dijstelbergen, H. H.; Spen-
[32] Hofmann, F. (2000):
cer, E. A. (ed.): Flow Measure-
ment of Fluids, North-Holland Halbvoll und trotzdem genau
Publishing, pp. 493497. gemessen.
In: MSR-Magazin, No. 5.
[28] Hofmann, F. (1993):

Magnetisch-induktiver Durch-
flussmesser mit strmungsopti-
mierendem Messrohr.
In: Chemietechnik, No. 5, p. 142f.

189
References

[33] Hofmann, F. (2000): [37] Hofmann F. (2004):

Magnetisch-induktive Durch- Gleichfeld kontra Wechselfeld -


flussmesser aus Edelstahl: Magnetisch-induktive Durch-
Fr kompakte volumetrische flussmessung von Schlmmen
Fllanlagen. und Trben.
In: Deutsche Ernhrungs- In: P&A KOMPENDIUM, p. 140.
Industrie, No. 5.
[38] Horner, B.; Mesch, F. (1995):
[34] Hofmann, F. (2001):
An induction flowmeter insensi-
Vier sind oft zwei zuviel: Moder- tive to asymmetric flow profiles.
ne Durchflussmesstechnik mit Process tomography 1995 im-
zwei Leitern. plementation for industrial pro-
In: MSR Magazin, No. 3. cesses.
In: ECAPT Meeting "European
[35] Hofmann, F. (2004): Concerted Action on Process
Tomography", Bergen, Norway,
Magnetisch induktive Durch-
pp. 321-330.
flussmesser (MID).
In: Frh, K. F.; Meier, U. (ed.):
[39] Ketelsen, B.; Appel, E. (1975):
Handbuch der Prozess
automatisierung, 3rd edition, Die Meunsicherheit induktiver
Industrieverlag Mnchen, p. 419. Durchflumesser mit inhomoge-
nem Magnetfeld bei rotations-
[36] Hofmann, F. (2004): vernderten Strmungen Teil I.
In: Regelungstechnische Praxis,
Der Messstoff bestimmt die
No. 12, pp. 369375.
Technik Richtige Auswahl des
Durchflussmessers drosselt die
[40] Ketelsen, B.; Appel, E. (1976):
Folgekosten.
In: P&A, June 2004. Die Meunsicherheit induktiver
Durchflumesser mit inhomoge-

190
References

nem Magnetfeld bei rotations- [45] Polly, P.; Reinhold, I. (1974):


vernderten Strmungen Teil II.
Der Einfluss der Symmetrie-
In: Regelungstechnische Praxis,
Eigenschaften des Strmungs-
No. 1, pp. 1519.
profiles beim induktiven Durch-
flussmessgerten.
[41] Kiene, W. (1976): In: Archiv fr technisches Mes-
sen, Lfg. 464, No. 9, pp. 153158.
Magnetisch-induktive Durchflu-
messer.
[46] Reinhold I. (1975):
In: messen + prfen/automatik,
No. 6, pp. 316331. Magnetfeldformung und Me-
genauigkeit beim magnetisch-
[42] Kolin, A. (1936): induktiven Durchflumessgert.
In: Archiv fr technisches Mes-
An electromagnetic flowmeter.
sen, Lfg. 476, No. 9, p. 139144.
In: Proceedings of the Society for
Experimental Biology and Medi-
[47] Reinhold, I. (1978):
cine, Vol. 35, pp. 5357.
Velocity Profile Influence on
[43] Kolin, A. (1945): Electromagnetic Flowmeter
Accuracy.
An Alternating Field Induction
In: Dijstelbergen, H. H.; Spencer,
Flow Meter of High Sensitivity.
E. A. (Hrsg.): Flow Measurement
In: Review of Scientific Instru-
of Fluids, North-Holland Publi-
ments, Vol. 16, p. 109.
shing, pp. 181-185.

[44] Ludwig, D. (2002):


[48] Schommartz, G. (1974):
Volumetrische Flltechnik mit
Induktive Strmungsmessung.
Durchflussmessgerten auf
VEB Verlag Technik, Berlin.
magnetisch induktiver Basis.
In: VDI-Berichte 1652, Tagung
Langen, 4.3.2002, p. 31.

191
References

[49] Shercliff, J. A. (1954): [53] Timofeev, V. A.; Kirshtein,


G. K. (1983):
Relation between the velocity
profile and sensitivity of electro- Dependence of readings of elec-
magnetic flowmeters. tromagnetic flowmeters with
In: Journal of Physics D: Applied two-coil magnet structures on
Physics, Vol. 25, p. 817. velocity profile.
In: Magnetohydrodynamics,
[50] Shercliff, J. Vol. 19, No. 3, pp. 343348.

The theory of electromagnetic


[54] Trchtler, A.; Horner, B.;
flow-measurement.
Schupp, D. (1997):
In: Cambridge, University Press.
Tomographische Methoden bei
[51] Stern, K. (2002): der induktiven Durchflussmes-
sung zur Bestimmung von Str-
Magnetisch-induktive Durch-
mungsprofilen und parametern
flussmesser (MID) im Einsatz fr
In: Technisches Messen, No. 64,
Abfll- und Dosieraufgaben.
p. 10.
In: Bonfig, K. W. (Hrsg.):Praxis
der Durchflussmessung, Vol. 2,
b-Quadrat Buchverlag, Kreuztal.

[52] Thrlemann, B. (1941):

Methode zur elektrischen


Geschwindigkeitsmessung in
Flssigkeiten.
In: Helvetica Physicalica Acta,
No. 14, pp. 383419.

192
Alphabetical index

A C

Abrasion 41, 65, 105 cable (signal 80, 86, 154ff.


cable)
AC field 61, 150ff.
calibration, cali- 124ff.
Acceleration 132 bration rig
energy
capacitive fill 73
accuracy (error) 29, 40ff., 114, 124 level measure-
ambient 110, 122 ment
temperature capacitive signal 70
aseptic EMF 32 pick-up (no-
(hygienic) electrode EMF)

B chemical 105ff.
resistance
back pressure 132ff.
(dynamic pres- communication 29, 34, 86, 157ff.
sure) (interface)

Batch process 83ff. compact EMF 109, 113, 176


(volumetric conductivity, 26, 35, 44, 70ff.,
filling) electrical 80ff., 97
Battery powered 74ff. continuity equa- 131
Bernoullis Law 132 tion (Law of Flow)
(Principle of corrosion 30, 44
Energy Conser-
vation) counter (prese- 22, 74, 130
lection counter)
Bootstrap 154ff.
(shielding) Custody transfer 59ff.

Burn-in test 122 D

Bus (device bus) 157 device bus (bus) 157

diagnosis (test) 46ff.

194
Alphabetical index

Diameter (size, 31, 102ff. empty measuring 50, 167ff.


DN, nominal size) tube

DIN 19200 179 empty pipe 167ff.


detection
Direction of flow 29, 43
EN 29104 124, 179
DN (size, dia 31, 102ff. (= ISO 9104)
meter, nominal
size) EN standards 179

DVGW W420 79, 179 error (measure 29, 40ff., 114, 124
(= ISO 13359) ment error,
measuring
Dynamic 132ff. accuracy)
pressure (back
pressure) error modes 48

Dynamic viscosity 134ff. Ex-version, 55ff.


hazardous area,
E Explosion
Earth conductor 170ff. protection

earthing 118, 170ff. F

Earthing elec- 171ff. FE (Functional 170ff., 175ff.


trodes Earth)

earthing rings, 170ff. field coils 30, 50, 157


earthing discs field current 74, 159
electrical 26, 35, 44, 70ff., supply
conductivity 80ff., 97 Filling, volu 83ff.
(minimum) metric (batch
Electrode 107 process)
material flanged EMF 31, 104
Electrode voltage 150ff. flooding (protec- 108, 115
Electrodes 31, 107 tion category)

Flow probes 91

195
Alphabetical index

flow profiles 42, 50, 67, 112, IMoCom-Bus 157


134ff., 160ff., 169
Inlet/outlet runs 103, 112, 160ff.
flow rate (q) 22, 28, 130ff, 145.
inside diameter 31, 102ff.
flow resistance 138 (of EMF and
pipeline)
Flow velocity 27, 132ff.
installation site 112ff.
flow velocity (v) 27, 103, 145 (of primary head
fluid mechanics 130ff. and signal
converter)
Foundation 29, 34
Fieldbus (FF) interface 29, 34, 86, 157ff.
(communication)
four-wire EMF 80
internal resist- 150, 154, 157,
full pipe 167ff. ance (source
detection impedance)

full scale range 157 ISO 13359 79, 179


(Q100%) (= DVGW W420)

functional earth 170ff., 175ff. ISO 6817 179


FE
ISO 9104 124, 179
G (= EN 29104)

Gas content 98ff. K

H kinetic energy 133

HART 29, 34, 158 L

heat sources and 116 Laminar flow 134


sun
Law of Flow (con- 131
Hygienic EMF 32 tinuity equation)
(aseptic)
level measuring, 73
I capacitive

IEC standards 180

196
Alphabetical index

Lightning protec- 175 measuring 27, 144ff.


tion (overvoltage principle
protection)
measuring range 103
line (cable) 80, 86, 154ff.
minimum 26, 35, 44, 70ff.,
linear measuring 29, 40 conductivity, 80ff., 97
principle electrical

Liner 30, 50ff., 65ff., mounting loca- 112ff.


105 tion (of primary
head and signal
low flow cutoff 43 converter)
(starting
threshold) multiwire EMF 79

M N

MagCheck 46 NAMUR 181

magnetic fields, 47, 60ff. Newtonian fluids 91, 160


external
No-electrode 70
mains voltage 79, 115 EMF (capacitive
(power supply) signal pick-up)

maintenance 44 nominal size 31, 102ff.


(size, DN,
mass flow 22, 64 130, 166 diameter)
measurement
non-Newtonian 165
measured ranges 22, 130 fluids
measurement 29, 40ff., 114, 124 O
error (measuring
accuracy, OIML 78
measuring error)
P
measuring accu- 29, 40ff., 114, 124
racy (measure- partly filled pipes 72ff.
ment deviation, PE (Protective 170ff., 175ff..
measuring error) Earth)

197
Alphabetical index

pipe friction 139 Protective Earth


coefficient (PE)

pipeline, partially 72 pulsating flow 170ff., 175ff


filled
Pulsed direct 62
potential 118, 170ff. current (dc) field
difference
Q
Power supply 79, 115
(mains voltage) Q100% 157
(full scale range)
power surge pro- 175
tection (lightning R
protection) reduction (to a 103, 160
preselection 22, 74, 130 smaller nominal
counter (counter) size)

Pressure loss 137ff. Remote version 109, 115, 176


EMF
pressure ratings 31, 102
Resistance, 105ff.
primary head 30ff., 112ff. chemical

Principle (meas- 27, 144ff. Reynolds number 137


uring principle) Re

Principle of En- 132 S


ergy Conserva-
tion (Bernoullis Sandwich EMF 32
Law) Shielding 154ff.
process liquid 29, 47, 97, 106ff., (Bootstrap)
temperature 116 signal cable 80, 86, 154ff.
PROFIBUS 29, 34 (cable)
interface signal converter 38, 115ff., 157ff.
protection 108ff., 115, 180 Size (DN, 31, 102ff.
category nominal size,
diameter)

198
Alphabetical index

solid particles 29, 64ff. vibrations 45, 63, 110, 117

solids 64ff., 97 virtual earthing 172

source imped- 150, 154,157, viscosity 43, 134ff.


ance (internal
resistance) volume (flow) 22, 28, 130ff, 145

Standard 22, 131 volume counting 22, 130


condition volumetric filling 83ff.
Standards 179 (batch process)

start-up thresh- 43 W
old (low flow WIB-SIREP 42, 161
cutoff)

static pressure 132ff.

swirl flow 164

temperature 29, 47, 97, 106,


(process liquid 110, 116, 122
temperature
and ambient
temperature)

test (diagnosis) 46ff.

testing 46
equipment

turbulent flow 136

two-wire EMF 81ff.

vacuum 105, 113

VDI / VDE 179ff.

199