Sunteți pe pagina 1din 8

ARTICLE

Toxic anterior segment syndrome: Common


causes
Carolee M. Cutler Peck, MD, MPH, Jacob Brubaker, MD, Sue Clouser, RN, MSN, CRNO,
Chris Danford, Henry E. Edelhauser, PhD, Nick Mamalis, MD

PURPOSE: To identify the most common risk factors associated with toxic anterior segment
syndrome (TASS).
SETTING: Ophthalmic surgical centers in the United States, Argentina, Brazil, Italy, Mexico, Spain,
and Romania.
METHODS: A TASS questionnaire on instrument cleaning and reprocessing and extraocular and in-
traocular products used during cataract surgery was placed on the American Society of Cataract and
Refractive Surgery web site. A retrospective analysis of questionnaires submitted by surgical cen-
ters reporting cases of TASS was performed between June 1, 2007, and May 31, 2009, to identify
commonly held practices that could cause TASS. Members of the TASS Task Force made site visits
between October 1, 2005, and May 31, 2009, and the findings were evaluated.
RESULTS: Data from 77 questionnaires and 54 site visits were analyzed. The reporting centers
performed 50 114 cataract surgeries and reported 909 cases of TASS. From January 1, 2006, to
date, the 54 centers reported 367 cases in 143 919 procedures; 61% occurred in early 2006. Com-
mon practices associated with TASS included inadequate flushing of phaco and irrigation/aspiration
handpieces, use of enzymatic cleansers, detergents at the wrong concentration, ultrasonic bath,
antibiotic agents in balanced salt solution, preserved epinephrine, inappropriate agents for skin
prep, and powdered gloves. Reuse of single-use products and poor instrument maintenance and
processing were other risk factors.
CONCLUSIONS: The survey identified commonly held practices associated with TASS. Understand-
ing these findings and the safe alternatives will allow surgical center personnel to change their
practices as needed to prevent TASS.
Financial Disclosure: No author has a financial or proprietary interest in any material or method
mentioned.
J Cataract Refract Surg 2010; 36:10731080 Q 2010 ASCRS and ESCRS

Toxic anterior segment syndrome (TASS) is a sterile in- however, in severe cases there may be lasting sequelae,
flammatory reaction of unknown incidence that can oc- such as permanent corneal edema, glaucoma, and
cur after anterior segment surgery. It typically presents other effects of chronic inflammation. Various entities
within 12 to 48 hours of surgery. The most common have been shown to cause TASS. These include, but
finding is diffuse limbus-to-limbus corneal edema are not limited to, endotoxin; denatured ophthalmic
(Figure 1) secondary to damage from a toxic insult to viscosurgical devices (OVDs); preservatives such as
the endothelial cell layer. Widespread breakdown of benzalkonium chloride, bisulfites, and metabisulfites;
the bloodaqueous barrier is another hallmark of this heavy-metal residue, fine-matter particulates, and
condition, with fibrin in the anterior chamber and hypo- substances introduced into the anterior chamber that
pyon present in 75% of cases (Figure 2). Damage to the are at a pH or concentration that is toxic to the sensitive
iris may cause the pupil to dilate or become slightly endothelial cells. In addition, residue of materials used
irregular, and glaucoma secondary to trabecular mesh- in the cleaning and sterilization of ophthalmic instru-
work damage may also occur. ments are an increasingly important source of TASS.
Treatment with intense topical steroidal agents will Members of industry and the American Society of
eventually lead to resolution of the inflammation; Cataract and Refractive Surgery (ASCRS) have joined

Q 2010 ASCRS and ESCRS 0886-3350/$dsee front matter 1073


Published by Elsevier Inc. doi:10.1016/j.jcrs.2010.01.030
1074 COMMON CAUSES OF TASS

to create a TASS task force. The goal of this task force is


to educate anterior segment surgeons on the causes,
symptoms, and treatment of TASS and to help investi-
gate outbreaks of TASS.
Questionnaires are available on the ASCRS web
siteA to facilitate the reporting of TASS cases. The in-
formation entered into the questionnaires is reviewed
by members of the TASS Task Force with the goal of
identifying potential causes of TASS. Recommenda-
tions on preventative measures to stop recurrences
are then made. This study will focus on common
practices that may lead to TASS as well as alternative
measures that will help protect patients having
anterior segment surgery.

MATERIALS AND METHODS


The TASS questionnaires were posted on the ASCRS web Figure 1. Toxic anterior segment syndrome after phakic IOL surgery.
site in 2007. A retrospective analysis was performed on ques- Note the limbus-to-limbus corneal edema.
tionnaires submitted between June 1, 2007, and May 31,
2009. The questionnaires address instrument cleaning and
reprocessing practices and surgical protocols as well as were excluded from analysis. Of those excluded, 5
substances and techniques used for cleaning phaco and were duplicates, 2 did not report any TASS cases,
irrigation/aspiration (I/A) handpieces. They also address and 2 were reporting cases of endophthalmitis. Of
products used extraocularly and intraocularly, such as med-
ications, and the reuse of cannulas and other disposable the centers included, 62 were in the United States; 6 in-
items. Responses were analyzed, and commonly held prac- ternational centers responded, with 1 each from Ar-
tices that lead to cases of TASS are presented here. gentina, Brazil, Italy, Mexico, Spain, and Romania.
In addition, members of the TASS Task Force made site Members of the TASS Task Force made 54 site visits,
visits at the request of centers with TASS cases. The visits all in the United States. Overall, 909 cases of TASS
occurred between October 1, 2005, and May 31, 2009. During
the visits, task force members observed preoperative, intra- were reported in 50 114 cataract surgeries performed
operative, and instrument-processing procedures to identify concurrently at the reporting centers that submitted
practices that could be associated with TASS. The data questionnaires. From January 1, 2006, to date, the 54
gathered from these visits are presented. centers visited by a TASS Task Force member reported
367 cases of TASS in 143 919 procedures performed;
RESULTS 61% of them occurred in 2006.
During the study period, 77 TASS questionnaires were Table 1 shows the cleaning procedures associated
submitted through the ASCRS web site, 9 of which with TASS. Inadequate flushing of the phaco and
I/A handpieces after surgery was the most commonly
observed and reported behavior that can lead to TASS.
Submitted: December 15, 2009.
Final revision submitted: January 25, 2010.
Accepted: January 26, 2010.
From the John A. Moran Eye Center (Cutler Peck, Brubaker,
Mamalis), University of Utah, Salt Lake City, Utah; UT Southwestern
Medical Center (Clouser), Dallas, and Alcon Laboratories, Inc.
(Danford), Fort Worth, Texas; and Emory University (Edelhauser),
Atlanta, Georgia, USA.
Supported in part by unrestricted grants from Research to Prevent
Blindness, Inc., New York, New York, USA, to the Department of
Ophthalmology & Visual Sciences, University of Utah.
Presented at the ASCRS Symposium on Cataract, IOL and
Refractive Surgery, San Francisco, California, USA, April 2009.
Corresponding author: Nick Mamalis, MD, John A. Moran Eye
Center, University of Utah, 65 Mario Capecchi Drive, Salt Lake Figure 2. Hypopyon is seen in the anterior chamber of this patient
City, Utah 84132, USA. E-mail: nick.mamalis@hsc.utah.edu. with TASS.

J CATARACT REFRACT SURG - VOL 36, JULY 2010


COMMON CAUSES OF TASS 1075

Of the 68 centers that filled out questionnaires, more


Table 2. Observations during site visits at centers with diag-
than 60% used less than the recommended 120 cc per nosed cases of TASS.
port (range 2 to 100 cc per port). Other factors were
occluded I/A tips during surgery, a symptom of Frequency of Observation,
inadequate flushing. Observation Category n (%) (N Z 54)
Table 2 shows the observations made during site Inadequate flushing of phaco and 48 (89)
visits at centers with diagnosed cases of TASS. The I/A handpieces and cannulated
top observations were inadequate flushing of phaco equipment
and I/A handpieces and cannulated equipment, use Use of enzymatic cleaners and 43 (80)
of enzymatic cleaners and detergents, use of reusable detergents
cannulas, and inadequate or no manual cleaning of Use of reusable cannulas 37 (69)
instruments. Cleaned instruments were often left on Inadequate or no manual cleaning 35 (65)
towels that were not lint-free, and many centers did of instruments
Use of preserved epinephrine 28 (52)
not train personnel regarding TASS and proper
Reuse of single use 28 (52)
cleaning practices. devices, blades, tips,
Medications given intracamerally or added to a bal- sleeves, etc.
anced salt solution irrigant were another potential Use of tap water with no sterile 27 (50)
source of TASS. Table 3 shows the products associated water final rinse
Inadequate personnel or trays to 23 (43)
allow proper preparation of
Table 1. Cleaning procedures associated with TASS.
instruments
Centers Reporting No immediate cleaning allowing 20 (37)
Practice, n (%) OVD and surgical solutions to
Reported Practice (N Z 68) dry on devices
Use of preserved medicines in the 18 (33)
Amount of flush used for handpieces eye
120 cc or more 18 (26) Reuse of tubing for flushing, latex 20 (37)
!120 cc (range 2100 cc) 42 (62) bulb for irrigation, no training,
Unknown 3 (4) no term sterilization,
Not specified 5 (7) instruments stored on towels
Occluded I/A tips Touching of IOL or patient contact 18 (33)
Yes 27 (40) areas of instruments with
No 41 (60) gloved hands
Final rinse water Off-label use of lidocaine gel 17 (31)
Deionized/distilled 30 (44) Poor instrument maintenance, 15 (28)
Sterile/pyrogen free 22 (32) residual autoclave residue and
Tap 12 (18) instrument mild, rust,
Unknown 2 (3) particulates, lint, etc.
Not specified 2 (3) Use of powdered gloves 15 (28)
Use of enzymatic cleaners Additives added to balanced salt 15 (28)
Yes 36 (53) solution against DFU
No 28 (41) Improper use of prep solutions, 14 (26)
Unknown 4 (6) detergents, and cleaners
Use of US Bath Failure to follow manufacturers 11 (20)
Yes 43 (63) DFU, including
No 25 (37) no air flush, use of unapproved
Frequency of cleaning US bath enzymatics
After each use 8 (19) Use of postoperative ointment in 11 (20)
More than once daily 31 (72) clear cornea cases
Unknown 3 (7) Povidoneiodine placed in eye at 9 (17)
No cleaning protocol 1 (2) end of procedure
Ophthalmic instruments separated for Incorrect concentration of 4 (7)
cleaning detergent and enzymatic
Yes 55 (81) cleaners used
No 9 (13) Lack of routine cleaning of US 2 (4)
Dont know 4 (6)
DFU Z directions for use; I/A Z irrigation/ aspiration; IOL Z intraocu-
I/A Z irrigation/aspiration; US Z ultrasonic lar lens; OVD Z ophthalmic viscosurgical device; US Z ultrasound

J CATARACT REFRACT SURG - VOL 36, JULY 2010


1076 COMMON CAUSES OF TASS

assistance when cases of TASS occur, in an effort to


Table 3. Products associated with TASS and their use.
prevent future cases, not all cases of TASS may be rec-
Centers Reporting ognized or centers may choose not to report cases if
Product Used Use, n (%) only a small percentage of their cataract surgeries re-
Antibiotic added to BSS
sult in TASS and a problem is not perceived. Because
Vancomycin 16 (24) we do not have a tally of all TASS cases that occur in
Gentamicin 1 (1) the United States, and because those reporting cases
None 50 (74) all reported various timeframes in which the reported
Unknown 1 (1) cases occurred for each center, we cannot calculate the
Intracameral antibiotic agents incidence of TASS from our data.
Vancomycin 7 (10) The most frequently identified practice associated
Cefuroxime 3 (5) with TASS is inadequate cleaning of phaco and I/A
Moxifloxacin 3 (5) handpieces after use. To be most effective, the manu-
Cefazolin sodium 1 (1)
facturers directions for use should be strictly followed
None 54 (79)
and handpieces should be wiped with a lint-free cloth,
Preserved epinephrine in BSS
Yes 18 (36)
immediately immersed in sterile water until they are
No 32 (64)
flushed, or both.13 This was not done at 89% of centers
Preserved intracameral anesthetic visited. This practice is imperative to prevent residual
agent OVD and debris from drying within the handpiece
Yes 3 (6) and tips before they are flushed. This is especially
No 44 (94) important because dried OVD denatures after sterili-
zation at high temperatures and if this or other
BSS Z balanced salt solution
particles or debris are irrigated into the anterior cham-
ber, TASS may result.4 While instruments are still
moist, they must be flushed with 120 mL of fluid per
with TASS. Antibiotic agents were the most common port, or more if indicated in the manufacturers
additives to balanced salt solution; some centers directions for use. This can be performed with tap
added epinephrine with preservatives or stabilizers. water as deionized water. However, the final flush
Slightly more than 20% of centers used intracameral must then consist of sterile distilled or sterile deionized
antibiotic agents; 17 centers (31%) used preserved water. If cleaned instruments are stored on towels, the
topical lidocaine gel preoperatively. towels should be lint free.3,5,6
Poor instrument maintenance was also pervasive. In addition to addressing the problem of inadequate
Autoclaves were found to have autoclave residue, flushing, it is important to address some of the reasons
instrument milk, rust, particulates, or lint at more it occurs. Even if an adequate cleaning protocol is in
than a quarter of the sites. Other sites failed to follow place, if surgical centers are understaffed or there are
manufacturers directions for use when cleaning too few surgical trays to allow proper cleaning of
instruments; this included omitting an air flush of instruments between cases (as observed at 43% of cen-
cannulated equipment, although this had been recom- ters visited), time constraints will lead to insufficient
mended, or using enzymatics during processing when flushing and patients will be placed at risk for TASS.
this was not approved. The reuse of tubing for flushing and latex bulbs for
In centers in which non-ophthalmologic surgeries irrigation is also a potential issue. These items are
were performed, ophthalmic instruments were pro- difficult to clean and hard to sterilize, which can lead
cessed along with instruments from other types of to improper processing between cases.
surgeries at 9 (13%) centers responding by question- It is imperative that surgical centers ensure that
naire; 4 (6%) respondents did not know whether there are adequate numbers of personnel and surgical
instruments were separated at the time of processing. trays to allow all steps of cleaning and sterilization to
be completed between uses of ophthalmic instru-
ments. It is also crucial that personnel are trained
DISCUSSION sufficiently in the causes of TASS and the proper steps
The data in this analysis were self-reported in the case of cleaning and sterilization.2,3,79
of the questionnaires or the result of self-referrals for Eighty percent of surgical centers visited and 53% of
site inspection in relation to TASS cases. The data those responding via questionnaire were using enzy-
from the site visits were taken from written reports matic detergents in the processing of their ophthalmic
submitted at the end of the visit. Although there is mo- instruments. The benefit of using enzymatic deter-
tivation on the part of surgical centers to obtain gents to clean ophthalmologic instruments has not

J CATARACT REFRACT SURG - VOL 36, JULY 2010


COMMON CAUSES OF TASS 1077

been established and may actually be prohibited in the completely removed from instruments before they
manufacturers directions for use for specific products. are immersed in bath water. Then, after each use,
Furthermore, the use of detergents mixed at the wrong and if in accordance with the manufacturers direc-
concentration has been linked to TASS outbreaks.10 tions for use, the bath water should be emptied and
Seven percent of centers visited were using the wrong the tub cleaned with an Environmental Protection
concentration of detergents, putting their patients at Agencyregistered facility-approved disinfectant.
risk for TASS as a result of residual detergents and in- This should be followed with a rinse using volumes
complete rinsing. of sterile or tap water adequate to remove the cleaning
Enzymatic detergents often have the exotoxin sub- agent completely. Then, 70% to 90% ethyl or isopropyl
tilisin or a-amylase enzymes as their active ingredi- alcohol should be used to clean the tub if this is in
ents. These are only deactivated at temperatures accordance with the manufacturers directions for
higher than 140 C, and most autoclaves do not reach use and not associated with a risk of fire. Endotoxin
temperatures higher than 120 C to 130 C. Thus, it is can be removed from the walls of the bath by wiping
likely that residue from enzymatic detergents will the walls with ethyl or isopropyl alcohol.1,14 The tub
build up on reused instruments.11 Human and rabbit should then be dried completely with a lint-free cloth
studies evaluating the effect of enzymatic detergents and refilled immediately before use.
on the anterior chamber showed dose-related corneal This process of cleaning must be followed after each
swelling; ultrastructural damage to the endothelial use to prevent endotoxin buildup. However, only 19%
layer, leading to increased corneal permeability; and of centers responding to the questionnaire cleaned the
an increased inflammatory response in the iris.12 ultrasonic bath after each use as recommended and 4%
The purpose behind the use of enzymatic detergents of centers visited had no protocol for routine cleaning
is to rid ophthalmic instruments of debris. If instru- of the ultrasonic bath.
ments are kept moist immediately after use before Of centers reporting via questionnaire, 25% added
flushing and if proper flushing with an adequate antibiotic agents to the balanced salt solution irrigant
volume is performed, there should be no adherent and 21% used intracameral antibiotic agents. The use
debris and thus no need for enzymatic detergents. If of antibiotic agents may be associated with toxicity
detergents are used, it is imperative that strict when they are included in anterior chamber irrigant
attention is paid to the dilution and expiration date. and when injected intracamerally at the end of
Furthermore, instruments processed with detergents a case.7 If antibiotic agents are improperly mixed, the
must be rinsed with copious amounts of fluid accord- concentration may be too high or the pH incorrect,
ing to the manufacturers directions for use. Recom- both of which can prove toxic to the anterior chamber
mended volumes should be considered a minimum tissues.
volume, and the final rinse should be performed The use of vancomycin and gentamicin sulfate in
with sterile distilled or sterile deionized water.1 anterior segment surgery has been described for
Sixty-three percent of reporting facilities used an prophylaxis against endophthalmitis.15 However, the
ultrasonic bath as part of the processing of their use of these products is associated with concerns
ophthalmologic instruments. This has been associated over vancomycin-resistant organisms as well as
with the accumulation of heat-stable endotoxins aminoglycoside-related macular toxicity, respectively.
produced by bacteria in the bath water.13 Endotoxin Furthermore, the concentration used in irrigating
remaining on instruments after cleaning and steriliza- solution and the time of contact with a possible
tion can induce the inflammatory reaction of TASS. contaminant is inadequate for their bacteriostatic or
As with enzymatic detergents, the purpose of the bacteriocidal properties to function.
ultrasonic bath is to dislodge dried debris from instru- Studies to evaluate intraocular cefotaxime for endo-
ments, particularly OVDs. Again, if instruments are thelial toxicity have been performed. A prospective
kept moist after use and then properly flushed with randomized masked study of 66 patients by
an adequate volume of water, there should be no Kramann16 found no toxicity with 0.4 mL of 0.25%
adherent debris and thus no need for an ultrasonic cefotaxime in the anterior chamber. Other studies
bath. evaluating the use of cefuroxime17,18 found no toxicity
If an ultrasonic bath is used, the manufacturers and a role for the agent in endophthalmitis prevention.
directions for use for instruments should be verified These findings were strengthened with the results in
because some instruments should not be processed a prospective randomized partially masked study in
in this manner. In addition, the ultrasonic bath should which 1.0 mg cefuroxime in 0.1 mL normal saline
be designated for medical use only and the manufac- injected into the anterior chamber after surgery de-
turers directions for maintenance should be strictly creased the risk for endophthalmitis (0.05% incidence
followed. Furthermore, all gross material should be rate when counting all endophthalmitis cases)

J CATARACT REFRACT SURG - VOL 36, JULY 2010


1078 COMMON CAUSES OF TASS

compared with a placebo or perioperative treatment sleeves, not be reused. Fifty-two percent of centers vis-
with a third-generation fluoroquinolone drop.19 How- ited were reusing these items. The reprocessing of
ever, an endophthalmitis incidence of 0.056% was these items is regulated and enforced by the U.S.
found with the perioperative use of a fourth- Food and Drug Administration.B
generation fluoroquinolone drop.20 Poor autoclave maintenance, including autoclave
At this time, because there are effective alternatives residue, instrument milk, rust, particulates, and lint
and because there is a risk for TASS from toxicity due within the autoclave, was documented at 28% of cen-
to the wrong concentration or pH of an improperly ters visited. Any such contaminants would be suspect
mixed antibiotic agent, use of intracameral antibiotic for causing TASS. One documented example of poor
agents must be viewed with caution.7 maintenance leading to heavy-metal contamination
Of surgical centers visited, 52% were using preserved of instruments showed carryover of sulfate, silica,
epinephrine in balanced salt solution and 37% were copper, zinc, and nickel from feedwater into steam
using other preserved medications intracamerally. In condensates associated with TASS outbreaks.26 This
addition, at least 36% of questionnaire respondents was found to be the result of changing from flushing
were adding preserved epinephrine to balanced salt so- and draining the steam generator once a week to
lution. Although bisulfites are not technically a preser- once every 4 to 8 weeks when maintenance was
vative but rather an antioxidant, they can also cause outsourced.
ocular toxicity. Benzalkonium chloride (BAC), a com- Surgical centers should make sure that preventive
monly used preservative, is highly toxic to ocular tis- maintenance, cleaning, and inspection of autoclaves
sues,21 and rabbit studies22 show a dose-response are performed as directed by the manufacturers direc-
pattern of conjunctivitis, flare, iritis, and corneal tions for use and also be documented. In addition, it
changes. In the rabbit study, a safety factor of 10 was should be verified that the sterilizer is functioning
used to set the safest intraocular concentration at properly on a daily basis where possible, or at least
0.001%, a level at which the preservative efficacy is weekly. Steam sterilization should be performed in
questionable. accordance with published guidelines, and instru-
An outbreak of TASS was associated with the use of ments should be sterilized according to directions for
Eye Stream rinse solution (Alcon Laboratories) pre- use provided by their manufacturer as well as the
served with 0.01% BAC, in which a concentration of sterilizers manufacturer. A terminal sterilization
0.013% was inadvertently irrigated into the anterior should be performed at the end of each surgical
chamber.23 Toxic anterior segment syndrome has day.3,5 Glutaraldehyde should not be used in the
also been reported after use of OVD preserved with sterilization process because it is highly toxic to the
BAC,24 and the use of lidocaine gel preserved with tissues of the anterior chamber and will cause TASS
BAC was recently implicated in a TASS outbreak if inadequately rinsed.27 Plasma gas sterilization has
(unpublished data). We recommend that no products also been shown to lead to degradation of brass to
with preservatives or additives (stabilizers), such as copper and zinc on cannulated surgical instruments
bisulfites and metabisulfites, be used in the anterior and should not be used because this has led to heavy
chamber.7,11 metalinduced TASS.28
Twenty percent of sites visited were using intraocular The use of powdered gloves can also be a source of
ointment after clear corneal cases. This practice was TASS because the powder can be toxic to the anterior
associated with a TASS outbreak in 8 cases. Of the chamber structures. Of sites visited, 28 were using
patients affected, 4 required penetrating keratoplasty powdered gloves during anterior chamber cases.
and glaucoma with no light perception developed in Powder-free gloves can also pose a risk for develop-
1 patient. Examination of each of these patients showed ing TASS. These gloves are made with a releasing com-
that petroleum-based ophthalmic ointments had pound that allows the glove to be easily removed from
gained access to the anterior chamber through a clear the mold after production. The releasing compound
corneal wound that was patched postoperatively.25 stays on the surface of the gloves and can cause ante-
Site visits showed that 69% of centers were using rior chamber toxicity if it comes in contact with IOLs
reusable cannulas. Cleaning cannulas is difficult to or any instrument entering the anterior chamber.C
perform and validate, and there are no manufacturers This can occur if the IOL or other instruments entering
directions for use. Furthermore, if the cannulas are the anterior chamber are touched during cataract
processed with enzymatic detergents, residue will surgery, a practice observed at 18 surgical centers
likely build up and then be irrigated into the anterior visited (33%).
chamber. Therefore, when possible, disposable cannu- Of questionnaire respondents, at least 13% indicated
las should be used.7,11 In addition, it is important that that ophthalmic instruments were processed for clean-
single-use devices, such as cannulas, blades, tips, and ing alongside dirty surgical trays from other,

J CATARACT REFRACT SURG - VOL 36, JULY 2010


COMMON CAUSES OF TASS 1079

nonophthalmic surgeries. However, ASCRS experts 11. Parikh CH, Edelhauser HF. Ocular surgical pharmacology:
recommend that all ophthalmic instruments be kept corneal endothelial safety and toxicity. Curr Opin Ophthalmol
2003; 14:178185
separate from other types of surgical instruments 12. Parikh CH, Sippy BD, Martin DF, Edelhauser JF. Effects of
during all steps of processing to avoid contamination enzymatic sterilization detergents on the corneal endothelium.
with bioburden or cleaning chemicals.1 Arch Ophthalmol 2002; 120:165172
In conclusion, analysis of data reported in question- 13. Kreisler KR, Martin SS, Young CW, Anderson CW, Mamalis N.
naires as well as in site visits from centers reporting Postoperative inflammation following cataract extraction caused
by bacterial contamination of the cleaning bath detergent.
TASS were analyzed carefully. The most common fac- J Cataract Refract Surg 1992; 18:106110
tors associated with TASS were inadequate cleaning 14. Franken KLMC, Hiemstra HS, van Meijgaarden KE, Subronto Y,
and flushing of ophthalmic instruments and handpie- den Hartigh J, Ottenhoff THM, Drijfhout JW. Purification of his-
ces. This included the use of detergents and ultra- tagged proteins by immobilized chelate affinity chromatography:
sound water baths. In addition, the use of intraocular the benefits from the use of organic solvent. Protein Expr Purif
2000; 18:9599
medications with preservatives or additives was 15. Gills JP. Filters and antibiotics in irrigating solution for cataract
associated with TASS in a significant number of cases. surgery [letter]. J Cataract Refract Surg 1991; 17:385
The proper cleaning and sterilization of ophthalmic in- 16. Kramann C, Pitz S, Schwenn O, Haber M, Hommel G, Pfeiffer N.
struments, as well as proper attention regarding the Effects of intraocular cefotaxime on the human corneal endothe-
use of any intraocular medications during surgery, lium. J Cataract Refract Surg 2001; 27:250255
17. Montan P, Wejde G, Setterquist H, Rylander M, Zetterstrom C.
can help prevent the occurrence TASS. These findings Prophylactic intracameral cefuroxime; evaluation of safety and
continue to validate the need to follow the recom- kinetics in cataract surgery. J Cataract Refract Surg 2002;
mendations detailed in the recommended practices 28:982987
document created by the TASS Task Force.1 18. Montan P, Wejde G, Koranyi G, Rylander M. Prophylactic in-
tracameral cefuroxime; efficacy in preventing endophthalmitis
REFERENCES after cataract surgery. J Cataract Refract Surg 2002;
1. American Society of Cataract and Refractive Surgery and 28:977981
the American Society of Ophthalmic Registered Nurses. Recom- 19. ESCRS Endophthalmitis Study Group. Prophylaxis of postoper-
mended practices for cleaning and sterilizing intraocular surgical ative endophthalmitis following cataract surgery: results of
instruments. J Cataract Refract Surg 2007; 33:10951100. the ESCRS multicenter study and identification of risk factors.
Available at: http://www.eyeworld.org/ewsupplementarticle. J Cataract Refract Surg 2007; 33:978988
php?idZ200. Accessed March 30, 2010 20. Jensen MK, Fiscella RG, Moshifar M, Mooney B. Third- and
2. AORN. Recommended practices for cleaning and caring for sur- fourth-generation fluoroquinolones: retrospective comparison
gical instrumentation and powered equipment. In: Association of of endophthalmitis after cataract surgery performed over
Operating Room Nurses. Standards, Recommended Practices 10 years. J Cataract Refract Surg 2008; 34:14601467
& Guidelines. Denver, CO, AORN, 2006; 555563 21. Swan KC. Reactivity of the ocular tissues to wetting agents.
3. Association for the Advancement of Medical Instrumentation. Am J Ophthalmol 1944; 27:11181122
Comprehensive Guide to Steam Sterilization and Sterility 22. Britton B, Hervey R, Kasten K, Gregg S, McDonald T. Intraocular
Assurance in Health Care Facilities. ANSI/AAMI ST79 2006; irritation evaluation of benzalkonium chloride in rabbits. Ophthal-
4. Kim JH. Intraocular inflammation of denatured viscoelastic mic Surg 1976; 7(3):4655
substance in cases of cataract extraction and lens implantation. 23. Liu H, Routley I, Teichmann KD. Toxic endothelial cell destruc-
J Cataract Refract Surg 1987; 13:537542 tion from intraocular benzalkonium chloride. J Cataract Refract
5. AORN. Recommended practices for sterilization in the perioper- Surg 2001; 27:17461750
ative practice setting. In: Association of Operating Room Nurses. 24. Eleftheriadis H, Cheong M, Sandeman S, Syam PP, Brittain P,
Standards, Recommended Practices & Guidelines. Denver, CO, Klintworth GK, Lloyd A, Liu C. Corneal toxicity secondary to in-
AORN, 2007; 673687. Available at: http://www.aorn.org/docs/ advertent use of benzalkonium chloride preserved viscoelastic
assets/3E41EE62-AB20-9320-21647715CD08CB26/Draft_RP_ material in cataract surgery. Br J Ophthalmol 2002; 86:299
Sterilization.pdf. Accessed March 30, 2010 305. Available at: http://www.ncbi.nlm.nih.gov/pmc/articles/
6. AORN. Recommended practices for the evaluation and selec- PMC1771062/pdf/bjo08600299.pdf. Accessed March 30, 2010
tion of products and medical devices used in perioperative 25. Werner L, Sher JH, Taylor JR, Mamalis N, Nash WA,
practice setting. AORN J 1998; 67:270272 Csordas JE, Green G, Maziarz EP, Liu XM. Toxic anterior
7. Mamalis N, Edelhauser HF, Dawson DG, Chew J, LeBoyer RM, segment syndrome and possible association with ointment in
Werner L. Toxic anterior segment syndrome. J Cataract Refract the anterior chamber following cataract surgery. J Cataract
Surg 2006; 32:324333 Refract Surg 2006; 32:227235
8. Burlew J. Care and Handling of Ophthalmic Microsurgical 26. Hellinger WC, Hasan SA, Bacalis LP, Thornblom DM,
Instruments 2nd ed.. San Francisco, CA, American Society of Beckmann SC, Blackmore C, Forster TS, Tirey JF, Ross MJ,
Ophthalmic Registered Nurses, 2006 Nilson CD, Mamalis N, Crook JE, Bendel RE, Shetty R,
9. Aorn. Explications for perioperative nursing. In: Association of Stewart MW, Bolling JP, Edelhauser HF. Outbreak of toxic
Operating Room Nurses. Standards, Recommended Practices anterior segment syndrome following cataract surgery associ-
& Guidelines. Denver, CO, AORN, 2006; 167197 ated with impurities in autoclave steam moisture. Infect Control
10. Breebaart AC, Nuyts RMMA, Pels E, Edelhauser HF, Hosp Epidemiol 2006; 27:294298
Verbraak FD. Toxic endothelial cell destruction of the cornea 27. U nal M, Yucel I, Akar Y, Oner A, Altn M. Outbreak of toxic anterior
after routine extracapsular cataract surgery. Arch Ophthalmol segment syndrome associated with glutaraldehyde after cataract
1990; 108:11211125 surgery. J Cataract Refract Surg 2006; 32:16961701

J CATARACT REFRACT SURG - VOL 36, JULY 2010


1080 COMMON CAUSES OF TASS

28. Duffy RE, Brown SE, Caldwell KL, Lubniewski A, Anderson N, B. U.S. Food and Drug Administration. Medical Devices. En-
Edelhauser H, Holley G, Tess A, Divan H, Helmy M, forcement Priorities for Single-Use Devices Reprocessed
Arduino M, Jarvis WR. An epidemic of corneal destruction by Third Parties and Hospitals. Available at: http://www.
caused by plasma gas sterilization; the Toxic Endothelial Cell fda.gov/MedicalDevices/DeviceRegulationandGuidance/Gui
Destruction syndrome Investigative Team. Arch Ophthalmol danceDocuments/ucm107164.htm. Accessed March 30,
2000; 118:11671176. Available at: http://archopht.ama-assn. 2010
org/cgi/reprint/118/9/1167?ckZnck. Accessed March 30, 2010 C. Lipner M, TASS: Pointing a Finger at Textured Gloves,
EyeWorld March 2008, page 56. Available at: http://www.
OTHER CITED MATERIAL eyeworld.org/article.php?sidZ4316&strictZ&morphologicZ&
A. www.ascrs.org queryZsurgeon. Accessed March 30, 2010

J CATARACT REFRACT SURG - VOL 36, JULY 2010

S-ar putea să vă placă și