Sunteți pe pagina 1din 18

ETH-256: Two Channel Combination are three AC-coupled, high pass filter settings: 0.

03,
Bridge/Biopotential Amplifier 0.3, and 3 Hz. For ECG and EEG and EOG, 0.3 Hz is
normally used; for EMG, 3 Hz is commonly used. The
Overview result of using a high-pass filter with a higher value is
the creation of a more stabile recording baseline; more
The ETH-256 is a combination transducer and biopo-
signal contributions from body movement or breathing
tential amplifier. Its controls, filter settings and gains
are removed with higher high-pass filter settings.
were selected with these applications in mind, thereby
making it well suited for the student physiology and the
Input Offset
research lab. The ETH-256 is powered by a wall trans-
former (12 Volts AC 1000mA). It connects to the ETH- The position of the recording baseline, in either
256 with a standard 3.0mm phono plug through a Bridge (DC) or Bio (0.03, 0.3 and 3Hz) mode, can be
receptacle on the rear panel. The ETH-256 uses an moved up (positive direction on Y-axis) or down
optically isolated pod and cable (C-ISO-255) for (negative direction on Y-axis) by using the Input
recording biopotentials (ECG, EMG, EEG, EOG) from Offset knob. This control is very important in Bridge
human subjects. mode where the output of a transducer may need to be
zeroed, or the maximum height of the recording display
area may need to be set.
Note: Under no circumstances should human subjects be
connected to the ETH-256 without employing a C-ISO-255
iolated pod and cable. Gain
The gain of each channel on the ETH-256 can be
With the C-ISO-255 pod and cable and the ETH-256 adjusted independently. This allows very small signals
any of the twelve ECG lead positions can be recorded, to be amplified before presentation to the display
one at a time by manually changing the lead wires to device. The gain control allows you to select among
the appropriate connectors. eight preset multipliers: x1, x5, x10, x50, x100, x500,
x1k, and x5k.

Low-Pass Filter (Hz)


The low-pass filter control sets the upper limit of
frequencies measurable by the ETH-256. The setting
labeled 10k offers a maximum frequency response of
about 10kHz. Other settings may be chosen to limit
Figure 1-1: ETH-256 Front Panel noise. For example, the 50 filter removes all signals
above 50 Hz. Powerline noise at 60 Hz, the most
Front Panel and Controls common source of electrical interference, would be
The ETH-256 is divided into two units or channels filtered from the recording along with any other
with identical functions that are capable of making low frequencies above 50 Hz with the use of this low-pass
noise, AC-coupled recordings of bioelectric potentials filter setting. For very slow (low frequency) signals,
or DC-coupled transducer recordings (Figure 1-1 on such as the output of a force or pressure transducer,
page 1). Each channel has three push-buttons and one the 5Hz setting will provide the quietest recording.
knob for controlling the functions described below. The
DIN8 and BNC input connectors for each channel are Input
also found on the front panel.
The input connectors are: either, BNC for single-
ended devices; or, DIN 8 for transducers, isolation
High-Pass Filter pods, head stages, and current to voltage adapters.
The high-pass filter has four positions. Each time the
button is pushed, the filter advances, in step-wise
fashion, to the next position as indicated by a LED.
The DC position directly couples the input to the
amplifier and is used for measuring from wheatstone
bridge-style transducers or other ground-referenced
sensors requiring impedance conversion, such as pH
electrodes or ion-selective electrodes (ISEs). There Figure 1-2: ETH-256 Rear Panel

iWorx Amplifiers 1
Rear Panel and Connectors The output of the transducer can now be viewed (at
the selected gain) from the BNC connector of the output
The BNC output connectors for the ETH-256 are
on the rear panel. Adjust the offset as necessary.
located on the rear panel. It can be cabled to your data
acquisition system using a common BNC-BNC cable. Using a strain gauge transducer with the ETH-256:
The power transformer input and the power switch are
1 After the transducer has been correctly fitted with an eight-
also found on the rear panel of the ETH-256. pin DIN connector and attached to the appropriate input
socket on the amplifier, connect the output BNC on the
Operation in Bridge Mode ETH-256 to the input of your recording device.

2 Set the ETH-256 gain to x1, the low-pass filter to 5Hz, and
Input the high pass filter to DC.
On the ETH-256, the input for transducers, pods, and 3 Observe the output of the ETH-256 on your recording
adapters is a standard eight-pin DIN socket located on device. Confirm that the baseline position can be affected
the front panel of the unit. This socket provides a rapid by turning the input offset knob for this channel. Adjust the
and reliable connection of these devices to the amplifier. baseline to zero Volts.
The connector on the transducer or other device that will 4 If you cannot see the baseline move when you turn the input
mate with the DIN8 socket of the amplifier is a male offset knob, try increasing the gain. A word about increasing
eight-pin DIN plug available from Preh (p/n 71408). The gain its possible that your recorder will also amplify and/
or position a signal. To avoid confusion, use only one set of
pin configuration for the DIN8 connectors used on iWorx
controls to adjust gain or offset once the signal is
devices is illustrated in Figure 1-3 on page 2. A list of positioned.
wiring schemes for common transducer types is found in
the User Area of the iWorx website: www.iworx.com. 5 Continue to increase the gain until you can easily position
the line with the input offset control. If you cannot see a
signal on your recorder or if the signal is pegged all the
way to the top or bottom of the recorder range, disconnect
the transducer immediately and refer to the troubleshooting
section of this manual.

6 Now that you have your signal positioned, deflect your


transducer by an amount appropriate to your experiment.
For example, if you have connected a blood pressure trans-
ducer, apply approximately 250mmHg, or if you have
connected a force transducer, hang a weight from it which is
approximately equal to the maximum force you expect to
see.

7 Does the trace go off the screen or paper? If so, reduce the
gain. If the deflection on the screen appears small (less
than 20% of the full screen or paper), increase the gain.
Your ETH/transducer is now ready to calibrate. Refer to the
manual that came with your recorder for the best procedure
Figure 1-3: Pin configuration for DIN8 socket on front of ETH amps and
to accomplish this.
iWorx A/D.

Operation of devices without resistive bridges, or


excitation voltages
Offset
The high input impedance of the ETH-256 in bridge
Use the 10-turn Offset knob on the front panel of the mode makes it useful for transducer types other than
ETH-256 to control the precise position of the baseline, resistive bridges. Piezoelectric, pH or any device with a
while allowing inherent offsets of the transducer to be signal output less than 5V can be applied to the inputs
nulled. in differential or single-ended modes. Be careful to
shield the cable and source when using high impedance
Operation of devices with excitation voltages devices.
Once transducers, pods, or adapters are wired for the Using a pH sensor with ETH-256:
correct configuration of the input of the ETH-256, these
devices will be supplied with excitation voltages (+5V), Set the ETH-256 gain switch to x1 and the low pass filter
setting to 50Hz.
inputs (+, -) for the differential amplifier, and a circuit
Put the electrode in pH 7.0 buffer. Observe the output of the
ground. ETH-256 on your recording device. Use the input offset to
adjust the output to zero.

iWorx Amplifiers 2
Use a pH 10 buffer to do the second measurement. Select a Operation
gain on the ETH-256 that allows the deflection from zero
produced by the change of buffer to be seen. Refer to the With the described settings, you should now see the
manual that came with your recorder for the best procedure to trace slowly approach zero as the recording begins. This
accomplish this.
event can take 10 to 15 seconds if the amplifier circuit
Operation in Bio Mode has been open for any length of time. Be patient and let
the trace settle before you attempt to adjust offset. If the
With the LED of the 0.03, 0.3, or 3 Hz high-pass filter
trace is noisy or wanders, see the troubleshooting
illuminated, the amplifier is ready to record biopotentials.
section of this manual. Assuming the trace is stable and
Any of these three settings will AC-couple the input to
quiet, increase the gain until you can see an acceptable
the amplifier. AC-coupling selectively removes the DC
signal. When used with the isolated pod, a gain of x10
component of signals being recorded. For example,
set on the front panel will produce an electrocardiogram
when recording ECG, the electrodes placed on the
of between 0.5 and 1 Volt.
surface of the skin will generate a relatively small
galvanic (steady state) potential between them. This The ETH-256 can be used to record signals such as
small galvanic potential is many times the amplitude of ECG, EMG, EOG and EEG. It can also be used to
the ECG signal being recorded. In DC-coupled mode, measure potentials directly from muscles and nerves. It
the gain applied to the ECG signal is also applied to the is ideally suited for extracellular recordings from animal
galvanic potential. In most cases in DC mode, a gain tissue with hook, pin, or suction electrodes.
large enough to see the ECG will saturate the amplifier Recording an ECG with the ETH-256:
with the signal of the galvanic potential.
1 Plug the supplied isolated pod and cable into the input
If the input is AC-coupled to the amplifier, the galvanic socket for channel one.
potential is selectively removed from the complete signal
2 Using an alcohol prep or alcohol soaked cotton ball, lightly
and the ECG signal remains. All the gain required to
abrade the skin on the underside of each forearm and on
record the ECG can now be used. In addition to DC the back of the calf near the ankle. After the skin dries,
signals, the 0.03, 0.3, or 3Hz filters will remove low attach a disposable, adhesive-backed electrode to each of
frequency AC signals, like the artifacts due to breathing the sites.
or movement. 3 Connect the green lead (ground) to the electrode on the
right leg. Connect the red lead (+) to the right arm and the
Control/Setup black lead (-) to the left arm.

Advance the high pass selection so that the 0.3 Hz 4 Start recording. The output seen on Channel 1 is a Lead 1"
LED is illuminated and set the gain to x10. As with the ECG. Increase the gain until the form of the ECG can be
bridge mode, it is a good idea to use the gain or the clearly seen. Other ECG lead configurations are given at
the end of this chapter.
offset adjustments of the ETH-256 to adjust the size and
position of the signal appearing on the screen. Select Recording an EMG with the ETH-256:
the 150Hz low pass filter if you are recording ECG, or 5 Plug the supplied isolated pod and cable into the input
10kHz if you are recording EMG. socket for channel one.

6 Using an alcohol prep or alcohol soaked cotton ball, lightly


Isolated Pod and Patient Cable abrade the skin on the underside of each forearm at the
The ETH-256 comes with a 3-lead isolated pod and locations indicated in Figure 1-4 on page 4. After the skin
dries, attach a disposable, adhesive-backed electrode to
cable (C-ISO-255). This cable may be used on either
each of the sites.
channel of the ETH-256. The C-ISO-255 contains a x50
amplifier and an optically isolated output stage. When 7 Connect the green lead (ground) to the electrode on the
wrist. Connect the red lead (+) to the upper portion of the
this pod is connected to the ETH-256, the gains set on
forearm with the greatest diameter and the black lead (-) to
the front panel are multiplied by 50. the middle of the forearm.
Attache the three color-coded lead wires to the 8 Have the subject sit quietly in a chair with his or her
isolation block on the pod. forearm resting on a table with palm turned upward
and finger flexor muscles relaxed,
Connecting the Leads 9 Start recording. After five seconds, the subject should
Attach surface electrodes to the subject. These leads gently contract his or her finger flexor muscles to
should have a snap connector to be used with the make a fist. The subject should hold this clench for
two seconds, then relax again for five seconds.
supplied lead wires.

iWorx Amplifiers 3
10 Continue to record, as the subject alternates periods In the DC mode, when using the ETH-256 with data
of a progressively stronger contractions with relax- acquisition devices sampling at low speed, a slow
ation. Stop recording when the contraction reaches a sinusoidal drift appears. Sinusoidal drift observed when
using digital recording devices is almost always due to aliasing
maximum of higher frequency noise. The most likely culprit in this case is
. the mains frequency. Use the 50Hz low-pass filter on the ETH-
256 and any mains frequency or line filters that may be
available on your recorder to minimize or remove this artifact.
\When using a high pass filter other than DC, the signal
is noisy and wanders. Noise in the recording of biopotentials
is frequently from the power lines in the room. Normally the
differential properties of the amplifier can remove it, but this
becomes difficult if the electrode connection to the subject is of
high or variable resistance. Make sure that surface electrodes
are securely fastened to the subject by gently abrading the skin
surface where the electrode will be attached with some nylon
scrubbing material. Be sure to use enough electrode gel if you
have German silver (reusable) electrodes.
The offset works, you can see the trace, but there is no
signal. Make sure that you are using enough gain to see the
changes produced by your transducer or by an ECG/EMG.
Remember these signals can be just a few millivolts. If you use
the 10V scale you won't see much.
Check the obvious. Are the amplifier and recorder
connected to a working outlet? Are you turning the switches
and knobs on the correct channel? Are you sure about the
Figure 1-4: Placement of electrodes to record an EMG from the subjects connection to your recorder?
forearm.
Specifications
Number of Channels: 2

Operational Modes: Bridge (Transducer) or


Biopotential (ECG, EMG,
EEG, EOG, AP)

Gain: x1, x5, x10, x50, x100,


x500, x1k, x5k

Filters: High Pass (Hz): DC, 0.03,


0.3, 3.0

Low Pass (Hz): 5.0, 50,


150, 2K, 10K

Power Source Wall Adapter: 12 VAC,


1 Amp

Ground Isolation: Optical with C-ISO-255


Figure 1-5: Electromyogram (EMG) from the muscle of the forearm as the
isolated cable
subject is clinching his fist. Lower trace is the absolute integral of the EMG.
Input Impedance: 10 GOhm
Troubleshooting Output Impedance: 100 Ohm
In DC mode, trace remains maxed to one side, offset has Input Connectors: DIN 8 or BNC
no effect. Check that the transducer is not overloaded and that
the gain is not set too high the ETH-256 and your recorder Output Connectors: BNC
may be accurate. If your transducer is not reacting to a large
pressure or force, there are only two explanations for a large Offset Range: 5V
offset. The first is that one of the inputs to the differential
amplifier on this channel is not connected. The second is that Excitation Current: 50 mA per channel
one of the power leads that supply the transducer may have
shorted to the input or directly to ground. In either case, open Excitation Voltage: 5V
the connector on the transducer and confirm that the wiring is
correct and that there are no shorts. CMR 85dB up to 200 Hz

iWorx Amplifiers 4
ETH-401: Four Channel Transducer Rear Panel
Interface

Overview
The ETH-401is a dedicated bridge amplifier that
provides four channels of excitation power, gain, and
noise filtration. It exhibits low system noise and offset,
making it suitable for high resolution data acquisition Figure 1-7: Rear panel of the ETH-401.
systems. The ETH-401 is an ideal solution for interfacing
low bandwidth transducers, such as those used to Line Power Selection
measure force, blood pressure, and displacement to The ETH-401hardware can be powered from either 120
data acquisition systems such as the IWX-118. or 220 VAC by switching the Voltage Selector switch on
Gain control on the ETH-401 is very straightforward. A the rear panel. Either 50 or 60-Hz power is acceptable.
programming resistor in the DIN-8 connector of the The switch itself displays the selected voltage. By
transducer is used to set the gain of the input amplifier. selection of an appropriate input power cable, it is
Transducers from iWorx set the gain of the ETH-401 to possible to power the ETH-401 from virtually any supply
the optimal value for that transducer. This does not limit worldwide. It is critical that the Voltage Selector switch
the ETH-401 to use with iWorx transducers. Many non- be in the position appropriate to the local mains power.
iWorx transducers can be modified to work with the ETH- This should be done before powering the device to avoid
401. damaging it.

The ETH-401 is an inherently low offset device. This


Fuse Replacement
fact, together with the low offsets produced by bridge-
style transducers and the high resolution of the iWorx The ETH-401 is fused with a 100 mA, 250 VAC,
118 data acquisition system, dramatically minimizes the 5x20mm style, slow-blow fuse. The fuse is the only user
need for offset control. Therefore, the ETH-401 user serviceable part on the ETH-401and should only be
experience is simplified. replaced with the line power cable unplugged. Replace-
ments may be found at many electronic suppliers
The ETH-401 is packaged in a low profile aluminum
(LittleFuse Inc. P/N: 218.100 or equivalent) or contact
enclosure that provides both durability and increased
the sales representative if a replacement is required.
noise immunity. The device has a small foot print and
Because the ETH-401 is line powered, it presents a
multiple units are easily stacked to save bench space.
safety hazard if opened. If your service problem is not
resolved by a fuse replacement, contact the service
department at iWorx or a qualified service technician.
Setting Up
1 Ensure the Line Power Selection Switch on the rear panel is
set properly. Excitation Power
2 Connect the power cable and switch the hardware On. Each channel on the ETH-401can deliver up to 100mA
of 5 V excitation to power a transducer. Each channel
3 Connect the BNC output cable to the data acquisition input.
provides separate excitation power so that problems on
4 Connect the transducer input to the ETH-401 DIN8 input. one channel will not affect the operation of other
channels. If the transducer on a given channel draws in
excess of 100 mA the fault is detected thermally, and
Front Panel power is shut down to protect both the transducer, and
the ETH-401. Typically, this will cause a slow pulsing of
the ETH-401 output as the channels excitation power
regulator heats, is turned off, cools to its reset point,
then begins heating again.

Figure 1-6: Front panel of the ETH-401.

iWorx Amplifiers 5
Gain Selection removing all energy above that frequency. Figure 2
shows the frequency response of an ideal filter--it has 0
Gain control of the ETH-401 is programmed by a gain
dB or unity gain for frequencies up to 35 Hertz, then cuts
set resistor in the input transducer. To minimize noise,
off all frequencies beyond that. Real filters roll off
the gain set resistor is housed in the transducer cable
around the cut-off frequency such that signals at the cut-
connector so that it is as close to the ETH-401 input as
off frequency are reduced to 70% of their input value,
possible. For Iworx transducers, gain values are preset,
and signals with greater frequency content are reduced
and are selected so that the ETH-401 output will be
in direct proportion with frequency.
large without exceeding 5 Volts to maintain compati-
bility with data acquisition systems with that input range.
For non-Iworx transducers, a gain set resistor can be Ideal Filter
inserted in the input connector. As a rule of thumb, to
determine an appropriate gain value, multiply the input 5.0000
voltage range of the device following the ETH--401 by

Magnitude(dB)
-15.0000
0.7, then divide the result by the transducers maximum
peak to peak output signal including its offset. If the -35.0000
input voltage range of the device following the ETH-401
is greater than 10 Volts, then substitute in 10 Volts to -55.0000
ensure that the ETH-401 amplifier is never saturated.
-75.0000
For simplicity, Table 1 lists resistor values used to 0.001 0.01 0.1 1 10 100
program common gain values. For gain values other Frequency
than those listed, choose the resistor by the equation:
Figure 1-8: Frequency response of an ideal filter

49.4k Filter characteristics are easily shown with Bode-plots,


R G ainSe t = -------------
G1 graphs that show amplitude, usually in dB, plotted on the
y-axis, against frequency, usually plotted logarithmi-
Tab l e 1-1 : Re s i st or Va l u es N ee d ed to S et cally, on the x-axis. Viewing a Bode-plot of a filter shows
N om i n a l G ai ns . the steepness of the filter roll-off, and that steepness is
directly related to the number of poles of the filter. Filters
are also characterized by their style. Some commonly
Gain (Nominal) Resistor(1%)
used filter implementations are Butterworth, Bessel, and
1 Open
2 49.9k Chebyshev. The filter implementation will also affect the
5 12.4k roll-off steepness.
10 5.49k The filter in the ETH-401 limits noise with frequency
20 2.61k content greater than 40 Hertz, while passing signals with
50 1.00k frequency content less than 40 Hertz. The Bessel style
100 499.00 filter is chosen to minimize ringing that results in other
200 249.00 filter configurations when the filter input has a step, or
500 100.00
sudden change, such as might occur when a
1000 49.90
displacement transducer is moved quickly, or when a
force-transducer measures a striated muscle twitch.

Figure XXX shows a Bode plot of the frequency


Frequency Response
response of a typical ETH-401 amplifier channel. There
Each ETH-401 channel contains an 8-pole, Besssel are three things to note about this graph. The first is that
style, low-pass filter with a fixed cut-off frequency set at the cut-off frequency at 35 Hertz. The second is that the
40 Hertz. graph ends at 335 Hertz, showing an attenuation of 81.6
In general, a filter is used to reduce noise content, dB, which is the limit of the resolution of the IWX-118
where noise content refers to any energy in frequency data acquisition system used to measure the ETH-401
bands outside of the signal of interest. A low-pass filter frequency response. The third is that 60 Hertz noise is
is used to remove energy content with frequency greater attenuated by about 7.5 dB. From a practical stand-
than the signal of interest. An ideal low-pass filter would point, 60 Hertz noise due to mains power supplies is the
pass all frequencies up to the cut-off frequency while most common and largest noise observed on bridge-

iWorx Amplifiers 6
style transducers, and most of it will be eliminated by the System Noise
common-mode rejection of the ETH-401input amplifier.
System noise is the noise attributed to the ETH-401
What remains will be more than cut in half by the filter.
amplifier itself, and is measured by shorting its inputs to
Consider a worst case example where a particular ground through 10 kOhm Resistors and measuring the
ETH-401 is configured for unity gain and has the output. The system noise is specified with reference to
guaranteed worst case common-mode rejection ratio of input (RTI), because this is a good indicator of the
80 dB. The transducer is powered by the ETH-401, and appropriateness of the amplifier to an application. The
thus cannot produce more than 5 Volts of noise at any ETH-401 is specified to have system noise that does not
frequency. Again assuming the worst case, this trans- exceed 10 uVolts, though it typically is less than 2 uV
ducer produces a 1 Volt signal at 10 Hertz, but has 5 RTI. The ETH-401 with appropriate gain, is a good
Volts of common mode noise at and above 60 Hertz. The choice for transducers that provide a full-scale output
common-mode rejection ratio of the amplifier attenuates signal of at least 100 uVolts, though typically it will
the common-mode noise by 80 dB, presenting the filter remain a good choice even for transducers with a full-
with noise of 5 mV at and above 60 Hertz. The 60 scale output as low as 20 uVolts.
Hertz noise is the least attenuated by the filter, but is
still cut in half, presenting the data acquisition system
with 2.5 mV of noise or less at and above 60 Hertz, Input Signal
and a signal of 1 Volt. This results in a signal to noise
The ETH-401 has differential inputs, meaning that
ratio of 400:1. Furthermore, when used with an IWX-118
each channel measures the difference between two
data acquisition system, any noise with magnitude less
signals. The inputs are the four DIN8 connectors located
than 1 mV will fall below the resolution of the analog to
on the front panel. The input range is 10 Volts,
digital converter. For this example, that means that noise
extending its useful range to signals other than those
with frequency content at or above 90 Hertz cannot be
derived from its excitation power. The input impedance
observed. Typically, the common-mode noise will be
of each channel is 10 GOhm so that high output
much less than the 5 Volts shown in this worst case
impedance transducers such as pH or ion-selective
example, meaning that almost all noise will fall below the
electrodes can be used without loading.
resolution of the data acquisition system.

ETH-401 Frequency Re spons e


Input Connector
10 Iworx transducers are configured to work with the ETH-
0 401, but other transducers can be built or modified to
-10 work with it as well. In most cases the modification will
-20 consist replacing the existing cable connector with a
-30
DIN8 male connector and inserting an appropriate gain
-40
set resistor in the connector shell. Figure 4 details the
-50
pin designations for the ETH-401 input connector--note
-60
-70 that the view is from the solder-side of the connector
-80 pins.
-90
0.00001 0.001 0.1 10 1000
F re que nc y

Magnitude -3dB

Figure 1-9: Bode plot of the frequency response of an ETH-401 amplifier


channel.

iWorx Amplifiers 7
Figure 1-10: Pin configurqation of male DIN 8 connector used for devices
comapatible with iWorx units.

Output Signal Output


The output impedance of the ETH-401 is typically less Number Of Outputs 4 Single Ended BNC Connector
than 1 Ohm, and is specified to be less than 10 Ohms at
Frequency Response 100Hz (+/- 10%) 8-Pole Low
all frequencies up to 50 Hertz.
Pass Bessel Filter

Gain 1 to 10k (Set By Transducer)


Specifications
Output Impedance <10 Ohm (Maximum) / < 1 Ohm
Input
(Maximum) (Typical)
Number of Inputs 4 Differential DIN8 Connector

Excitation Power 5 Volt (10%) Power Line-- 120/220 V AC, 60/50 Hz


(Switched)
Input Impedance 10 GOhm

Input Range 10 Volt

CMR

Gain =1 >80 dB (Minimum) / >92 dB (Typical)

Gain = 10 >95 dB (Minimum) / >110 dB (Typical)

Gain = 100 >110 dB (Minimum) / >120 dB (Typical)

Gain = 1K >112 dB (Minimum) / >135 dB (Typical)

Noise <10 uVolt (RTI) / <2uVolt (RTI)

iWorx Amplifiers 8
DHS-300: Differential Head Stage Figure 1-11: DHS-300 connected to iWorx 214 for recording.

Amplifier
Start the Software
Overview When using an iWorx data recording unit with DIN
The DHS-300 is a high CMR, low noise differential inputs (104, 114, 204, 214), or an IWorx data recording
amplifier designed especially for use with the ETH-256. unit (118) and an iWorx amplifier:
It can be placed close to the preparation, eliminating 1 Click the Windows Start menu, move the cursor to
long lengths of wire that can act as a pick-up antenna for Programs and then to the iWorx folder and select
line current frequencies and other sources of electrical LabScribe; or click on the LabScribe icon on the Desktop.
noise. This head stage is the best solution for the extra- 2 When the program opens, select Preferences from the Edit
cellular recording of biopotentials in situations where it menu.
is not possible to use the differential capability of the
3 Select the Channel preferences dialogue window. Name
amplifier to reduce common mode noise.
the channel to which the DHS-300 or the amplifier
supporting the DHS-300 is connected. Set the Mode and
Equipment Setup Function for this channel to Record and Raw Data,
respectively. Also, set the sampling rate and display time.
1 Plug the DIN8 connector of the DHS-300 into one of the DIN Click OK.
inputs of an iWorx data acquisition unit or amplifier (Figure
1-11 on page 9). 4 If you plan to use these settings again, click on the Settings
menu and select the Edit Group to add this setting file to
2 Attach the recording and ground electrodes to the inputs of your group of settings.
the DHS-300: positive (+) electrode to the red input socket;
negative (-) electrodes to the black input socket; and, the
ground electrode to the black alligator clip and lead wire
coming from the back of the head stage. Specifications
Input Impedance >10 GOhm

CMR 100dB @ 60 Hz

Noise (input shorted


at 1x gain) <2uV p-p, DC-100 Hz
<10uV p-p, DC-1KHz
(input shorted at 1x)

Gain at Head Stage Output x10

Input Mode Differential

Output Mode Single ended

iWorx Amplifiers 9
GSR-200: Galvanic Skin Response Adapter The GSR-200 presents a voltage of 200mV across the
resistance being measured, so that the Current (I)
Overview flowing across the skin from one electrode to the other is
equal to applied voltage divided by the resistance:
The GSR-200 applies a constant voltage across two
electrodes. The current flowing between the electrodes,
which is proportional to the skin conductance, can be
200mV
measured after the current is converted to voltage. It is I = ------------------ = 200mV G
R
the voltage seen at the output of the unit that can be
recorded and converted to units of conductance.
In the GSR amplifier, the Current (I) flows through a
1megOhm feedback resistor to produce the Output
Voltage (Vout ), so that:

1
V out = ----------------------- = I 1megoOhm = 200mV G 1megoOhm
R feedback

Solving for Conductance (G):

V out
G = -------------------------------------------------
0.200V 1meg O hm

Since 1 megOhm is the reciprocal of 1 Siemen, the


Conductance (G), in Siemens, is equal to 5 times the
Output Voltage:

G ( Siem e ns ) = 5 V out

Figure 1-12: Layout of GSR-200

Therefore, for the GSR-200, 1 Volt of output equals 5


Equipment Setup Siemens. This relationship along with the calibration
1 Plug one end of a male DIN-DIN cable into the GSR function of the recording device can be used to convert
amplifier. Connect the other end of the same cable to DIN the voltages recorded from the subject to Siemens of
input of an iWorx data acquisition unit or amplifier (Figure 1- conductance.
11 on page 9).

2 Affix the two electrodes attached to the input cable to the


first and second finger of the subjects hand. Changes in
skin conductance can now be measured as the subject
experiences various levels of stress.

Conductance and Output


The purpose of the GSR is to provide a measure of
skin conductance to record. Conductance (G), as
expressed in units known as Siemens, is the inverse of
Resistance (R):

1
G = ----
R
Figure 1-13: The equipment needed to record GSR.

iWorx Amplifiers 10
Start the Software 5 Select one person from your group to be the subject. Ask
the subject to go to the sink, wash his or her hands with
When using an iWorx data recording unit with DIN soap and water, and dry them thoroughly. Washing the
inputs (104, 114, 204, 214), or an IWorx data recording hands insures that surface oils or other substances, which
unit (118) and an iWorx amplifier: might lower skin conduction, are removed. Do not use
alcohol to clean the fingers, alcohol dehydrates the skin.
1 Click the Windows Start menu, move the cursor to
Programs and then to the iWorx folder and select 6 Connect the GSR electrodes to the BNC connector on the
LabScribe; or click on the LabScribe icon on the Desktop. GSR amplifier.

2 When the program opens, select Load Group from the 7 The subject will be sitting with his or her back to the
Settings menu. monitor, so use the hand closest to the iWorx
equipment. Attach the GSR electrodes to the palmar
3 When the dialog box appears, select IPLMV3.iws. Click surface of the distal finger segment on two non-adjacent
Load. fingers, usually the index finger and the ring finger. Attach
the electrodes with the Velcro straps so that the straps are
4 Click on the Settings menu again and select the GSR-A or
snug, but not overly tight.
GSR-B settings file.

5 After a short time, LabScribe will appear on the computer


screen as configured by the GSR-A or GSR-B settings.
Experiments
Calibration of GSR Amplifier Experiments using the GSR-200 amplifier and
electrodes can be found in the iWorx Physiology Lab
1 Pull down the Window menu on the LabScribe Main
window and select Preview. Click the arrows at the top of Manual (IPLM) which is available from the iWorx Users
Preview window to select the on-line display of the GSR website: www.iworx.com/users/default.htm.
channel (CH 4). Use the offset knob on the GSR amplifier to
move the observed output to zero. Click the OK button on
the Preview window.
Specifications
2 The GSR-200 amplifier is factory calibrated so that an
output of 1 Volt is equal to 5 microSiemens (S). Input impedance: 10 GOhm

3 To enter this conversion factor into the recording settings, Input connector: BNC
pull down the Edit menu in the Main window, and select
Working voltage: -200 mV
Preferences and go to the Channels window.
Output connector: DIN8
4 Under the Calibration section of the Channels window, the
following settings should be used: Output voltage swing: + 4.0 Volts
Calibration Bandwidth: DC to 3Hz
Raw Cal Unit Power: From DIN8 input of
1 => 5 S iWorx amplifier or A/D
0 => 0
Noise: 1.0mV p-p
These settings mean that 1 Volt (raw unit) is set equal
to 5 S (calibrated unit), and 0 Volt is set equal to 0 S.

iWorx Amplifiers 11
SI-100: Stimulus Isolator
Overview Control of SI-100
The SI-100 is an isolated constant current generator The SI-100 is controlled by any stimulator or DAC that
with 100 Volt compliance. It is specifically designed for is capable of delivering TTL input pulses. TTL input
safe transcutaneous stimulation of muscles. pulses are nominally 5V in amplitude and at least 100
microseconds (s) in duration. The frequency and
Front Panel duration of the stimulus pulses from the SI-100 are
designed by programming the TTL level pulses of the
The amplitude of the stimulating pulse from the SI-100
stimulator or DAC. When the output of the stimulator or
is controlled by a single knob on the front of the unit
DAC is connected to the Trigger input of the SI-100 and
(Figure 1-14 on page 12). This control is a ten turn
activated, the SI-100 will generate stimulus pulses of the
potentiometer which is attached to a dial counter (with
same frequency and duration as the TTL level pulses
integral lock) that shows the number of turns from 0.0 to
from the stimulator or DAC.
10.0. When turned fully clockwise, this Amplitude
control can command a current output of 10 milliam- The amplitude of the stimulus pulses from SI-100 is
peres. still under the control of the single knob on the front of
the SI-100.

Power Setup
The SI-100 can draw power from either an AC or a DC
output wall transformer with an output greater than 9
Volts and less than 15Volts, at 1000mA. A wall trans-
former with an output of 12VAC and 1000ma is supplied
with the SI-100.
Figure 1-14: Front Panel.
Stimulation
The front panel of the SI-100 also contains two status Electrical stimuli are delivered by the SI-100 to the
lights; Power, and H.V. Ready. A glowing Power light preparation through a set of leads with either snaps or
indicates that the SI-100 unit is properly powered up, alligator clips. Each lead set connects to the SI-100
and the H.V. Ready light indicates that the high voltage through high voltage safety connectors. The safety
power supply is ready to deliver a stimulus. connectors on each lead are plugged into H.V. Output
The last items on the front panel of the SI-100 are the jacks on the front panel of the SI-100, so that the red
positive and negative jacks of the H.V. (high voltage) lead is plugged into the red high voltage safety jack
Output. The stimulus electrodes are connected to these labeled Positive and the black lead is plugged into the
outputs. black high voltage safety jack labeled Negative.

Rear Panel The stimulator or DAC driving the SI-100 determines


the durations and the frequencies of the stimuli delivered
to the preparation. However, the SI-100 limits the
duration of its stimulus outputs pulses to a maximum of
10ms and their frequency to a maximum of 50 Hz.

The SI-100 is a constant current device, which means


it will develop whatever voltage is necessary to maintain
the current flow set from the knob on the front of the SI-
100. The high voltage power supply in the SI-100 has a
Figure 1-15: Rear panel. 100 Volt compliance. This means that 100 Volts is the
maximum voltage that the SI-100 can develop while
The rear panel (Figure 1-15 on page 12) of the SI-100 trying to deliver the set current. In constant current
contains the Power on/off switch, the power receptacle, devices like the SI-100, more voltage is required to
and the BNC connector of the Trigger input. The Trigger deliver the set current to the subject, if the electrical
input is connected to a stimulator or a digital-to-analog
resistance increases where the stimulating electrodes
converter (DAC), which can provide TTL level trigger
connect to the subject.
pulses that drive the SI-100.

iWorx Amplifiers 12
Constant current devices differ from constant voltage If the A/D unit is an iWorx 114, 204, or 214, the BNC-BNC
devices. When presented with an increase in resis- cable needs to be connected to a BNC-double banana adapter
that is placed into the positive (red) and ground (green) banana
tance, like dehydration of the gel under the electrodes, a jacks on the stimulator of these units. This is the side of
constant current device increases its output voltage to adapter, embossed with the letters GND, goes into the ground
(green) jack of the stimulator.
maintain delivery of the set current to the subject. On the
other hand, a constant voltage device still delivers the
set voltage to the subject with an increase in the resis-
tance at the electrodes, but the current delivered to the
Additional Setup for use with SMT-100, Striated
subject decreases if it is limited by the power supply of
Muscle Transducer
the constant voltage device. 1 The subject should remove all jewelry from his/her right
hand and wrist.

Note: Although the high voltage supply can deliver up to 100 2 Obtain two disposable electrodes.
V, it is at very low current and then only for very brief periods
3 Peel the protective shield off one of the electrodes. Attach it
(fixed maximum of 10 milliseconds). The amount of power
slightly above the lateral edge of the right palm, about 3/8
delivered to the subject is not sufficient to cause injury or tissue onto the back of the hand and half way between the knuckle
damage. of the little finger and the wrist (Figure 1-17 on page 13).
This electrode is the negative stimulating electrode. It is
being placed where the stimulus usually elicits a large
Equipment Setup response from the little finger. Connect the black lead from
the Negative High Voltage Output of the SI-100 to this
1 Place the SI-100 on the bench near the subject. Plug the electrode.
cable of its power supply into the unit. Then, plug the power
supply into the outlet. Turn the SI-100 on.

2 Attach one end of a BNC-BNC cable to Trigger input on the


back panel of the iWorx SI-100 (Figure 1-15 on page 12).

Figure 1-17: A. Placement of the stimulating electrodes on the back of the


right hand. B. Placement of weight (10) across the inside of the hand and
sensor on the outer edge of little finger.

4 Peel the protective shield off the other electrode and attach
it in the center of the back of the right hand, half way
between the knuckle of the middle finger and the wrist
(Figure 1-17 on page 13). This is the positive stimulating
electrode and functions as a current sink or reference
electrode. Connect the red lead from the Positive High
Voltage Output of the SI-100 to this electrode.
Figure 1-16: Equipment used to measure muscle twitches from a subjects
5 Attach the twitch sensor along the side of the little finger.
little finger.
Use a velcro strap or piece of surgical tape to attach the tip
of the sensor to the end of the little finger. Use a longer
3 Connect the other end of the BNC-BNC cable to the output
velcro strap or piece of surgical tape to attach the base of
of the stimulator that will be used to drive the output of the
the sensor along the lateral edge of the palm of the hand.
SI-100:
Be sure the marked side of the transducer faces away from
if the A/D unit is an iWorx 104 or 118, the BNC-BNC cable the skin (Figure 1-17 on page 13).
can be connected directly to the stimulator output of the unit.

iWorx Amplifiers 13
Start the Software Experiments
1 Click the Windows Start menu, move the cursor to Exercises using the SI-100, Stimulus Isolator, and the
Programs and then to the iWorx folder and select SMT-100, Striated Muscle Transducer, can be found in
LabScribe; or click on the LabScribe icon on the Desktop. the iWorx Physiology Lab Manual (IPLM) which is
2 When the program opens, select Load Group from the available from the iWorx Users website:
Settings menu. www.iworx.com/users/default.htm.

3 When the dialog box appears, select IPLMV3.iws. Click


Load. Specifications
4 Click on the Settings menu again and select the Human- Output Current 0-10 mA
MuscleTwitch settings file. Compliance Voltage 100V

5 After a short time, LabScribe will appear on the computer Isolation Rated to 5300V RMS
screen as configured by the HumanMuscleTwitch settings. Power 9-15VAC/VDC 1000mA
Trigger input External TTL signal
Frequency Range Externally driven 0.1 to 50 Hz
Pulse Width Minimum 100uS
Pulse Width Maximum 10mS
Coupling Capacitance <=5pF

iWorx Amplifiers 14
Amplifier Tables
Use these tables to select the iWorx products that are
suitable for your applications.

Table 1: Cardiovascular System

Typical Parameter
Physiological Measurement Sensing Devices Appropriate Accessories/
Parameter Frequency
Parameters Required Used Amplifier Sensors
Amplitude (Hz)

Heart Potentials Electrocardiogram Surface Electrodes 2mV 0.05-80 ETH-256 Surface Elec-
trodes

Heart Electrodes 50mV 0.05-80

Vector cardiogram Surface Electrodes 2V 0.05-80 ETH-256 Surface Elec-


trodes

Fetal Electrocar- Surface Electrodes 10V 2-100 ETH-256 Surface Elec-


diogram (mother) trodes

Blood Pressure Direct Arterial Pressure 120mmHg DC-20 ETH-401 BP-100


Pressure at Bra- Transducer ETH-256
chial Artery

Direct Arterial Mercury Manometer 120mmHg DC-20 ETH-401 BP-100


Pressure at Fem- ETH-256
oral Artery

Direct Venous Pressure Transducer 9mmHg DC-20 ETH-401 BP-100


Pressure ETH-256

Indirect Arterial Sphygmomanometer 120/80mmHg ETH-401 Cuff


Pressure ETH-256 BP-100

Korotkoff 150mV 30-500 ETH-256


Microphone

Table 2: Respiratory System

Typical Parameter
Physiological Measurement Sensing Devices Appropriate Accessories/
Parameter Frequency
Parameter Required Used Amplifier Sensors
Amplitude (Hz)

Breathing Pneumogam Thermistor 500cc/breath 0.05-2 ETH-401


Pneumograph ETH-256

Respiratory Flow Pneumot- Pneumotachograph 20,000cc/mL DC-2 ETH-401 SP-304


achogram w/Pressure ETH-256 Flow Heads
Transducer

Respiratory Vol- Spirogram Spirometer 4,000cc DC-0.5 ETH-401 SP-304


ume ETH-256 Flow Heads

iWorx Amplifiers 15
Table 3: Brain Functions

Typical Parameter
Physiological Measurement Sensing Devices Appropriate Accessories/
Parameter Frequency
Parameter Required Used Amplifier Sensors
Amplitude (Hz)

Electrical Activity Electroencephalo- Scalp Electrodes 50V 0.5-100 ETH-256 Electrocap Mon-
graph tage

Intracranial Elec- 500V 0.5-100 ETH-256


trodes

Evoked Intracellular Microelectrodes 100mV 1-10,000 ETH-256


Responses Potentials

Extracellular Needle Electrodes 50V 1-1,000 ETH-256 DHS-300


Potentials

Eye Response Electroretinogram Contact Lens Elec- 100V 0.5-20 ETH-256


trode

Table 4: Muscular Functions

Typical Parameter
Physiological Measurement Sensing Devices Appropriate Accessories/
Parameter Frequency
Parameters Required Used Amplifier Sensors
Amplitude (Hz)

Muscle Strength Myogram Needle or Surface 1mV 10-5,000 ETH-256 Surface


Electrodes Electrodes

Muscle Electromyogram Needle or Surface 1mV 10-5,000 ETH-256


Potentials Electrodes

Electromyogram w/ ETH-256
Needle or Surface
Electrodes

Stimulation Stimulate w/Surface ETH-256


Electrodes

Nerve H Reflex Electromyograph w/ ETH-256


Conduction Response Reduced
Stimulation

Conduction Veloc- Electromyograph 10-50,000 ETH-256 NBC-200


ity

Smooth Muscle Electrogastro- Surface Electrodes 20mV 0.05-2 ETH-256


Activity gram

Organ Bath Stud- Force Transducer 5gms DC-4 ETH-401 FT-100


ies ETH-256 FT-302

iWorx Amplifiers 16
Table 5: Autonomic Nervous System

Typical Parameter
Physiological Measurement Sensing Devices Appropriate Accessories/
Parameter Frequency
Parameters Required Used Amplifier Sensors
Amplitude (Hz)

Sweat Gland Galvanic Skin Lead Surface Elec- 50kOhms DC-5 GSR-200
Activity Response trodes

Electrical Skin Lead Surface Elec- 50kOhms DC-5 GSR-200


Resistance trodes

Body Temperature Thermistor Probe 98 F DC-0.1 ETH-401 TM-100


Temperature ETH-256

Table 6: Oxygen Measurements

Typical Parameter
Physiological Measurement Sensing Devices Appropriate Accessories/
Parameter Frequency
Parameters Required Used Amplifier Sensors
Amplitude (Hz)

Dissolved Oxygen Change in Oxy- Clark Style Electrode 1-20nA DC-1 ETOX-M Oxygen
gen Concentration Electrode

iWorx Amplifiers 17
iWorx Amplifiers 18

S-ar putea să vă placă și