Sunteți pe pagina 1din 67

Instruction Manual

C 200 Series
Multiparameter Bench
Photometers

These Instruments are in


Manufacturers since 1978 Compliance with the CE Directives
Dear Customer,
PRELIMINARY EXAMINATION
Thank you for choosing a Hanna product. Please read this instruction
manual carefully before using the meter. This manual will provide you Remove the instrument from the packing material and examine it
with the necessary information for the correct use of the instrument. If carefully to make sure that no damage has occurred during ship-
you need additional technical information, do not hesitate to e-mail ment. If there is any damage, notify your Dealer.
us at tech@hannainst.com. These instruments are in compliance Each Meter is supplied complete with:
with directives EN 50081-1 and EN 50082-1. Two Sample Cuvets and Caps*
One Transport Cap
TABLE OF CONTENTS 9 V Battery
PRELIMINARY EXAMINATION ............. 3 TOTAL HARDNESS MEDIUM RANGE .. 70 12 VDC Transformer (HI 710005 or HI 710006)
GENERAL DESCRIPTION ..................... 3 TOTAL HARDNESS LOW RANGE ......... 73 Note: Save all packing material until you are sure that the instru-
PRINCIPLE OF OPERATION ................ 4 HYDRAZINE ................................... 76 ment functions correctly. Any defective item must be returned
GUIDE TO DISPLAY CODES ................. 6 IODINE ......................................... 78 in their original packing with the supplied accessories.
SPECIFICATIONS .............................. 7 IRON HIGH RANGE ......................... 80 * C200, C206, C207, C209, C210 and C213 are supplied with 3 cuvets and caps

FUNCTIONAL DESCRIPTION ............... 8 IRON LOW RANGE .......................... 82


TIPS FOR AN ACCURATE MEASURE- MANGANESE HIGH RANGE .............. 85 GENERAL DESCRIPTION
MENT ............................................. 9 MANGANESE LOW RANGE ............... 87
PARAMETERS REFERENCE TABLES ... 10 MOLYBDENUM .............................. 90 C 200 Series is a line of 14 different bench, microprocessor-based
OPERATIONAL GUIDE ...................... 16 NICKEL HIGH RANGE ...................... 93 photometers that measure up to 40 parameters in water and
ALUMINUM ................................... 18 NICKEL LOW RANGE ....................... 95 wastewater. These multipurpose meters are manufactured to measure
AMMONIA HIGH RANGE .................. 20 NITRATE ........................................ 98 the most important parameters of the application they have been
AMMONIA MEDIUM RANGE ............. 22 NITRITE HIGH RANGE .................... 100 especially designed:
AMMONIA LOW RANGE ................... 24 NITRITE LOW RANGE ..................... 102 C 200 Laboratories C 203 Aquaculture
BROMINE ..................................... 26 DISSOLVED OXYGEN ..................... 104 C 205 Boilers & Cooling Towers C 206 Environmental Testing
FREE CHLORINE ............................ 28 pH ............................................. 107 C 207 Industrial Wastewater C 208 Water Conditioning
FREE CHLORINE HIGH RANGE .......... 31 PHOSPHATE HIGH RANGE ............. 109 C 209 Education C 210 Pulp & Paper Mills
TOTAL CHLORINE ........................... 34 PHOSPHATE LOW RANGE .............. 111
C 211 Chemical Manufacturers C 212 Power Plant Utilities
TOTAL CHLORINE HIGH RANGE ......... 37 PHOSPHORUS ............................ 113
CHLORINE DIOXIDE ........................ 40 C 213 Municipal Wastewater C 215 Nutrient Analyses
SILICA ........................................ 115
CHROMIUM VI HIGH RANGE ............ 43 SILVER ....................................... 118 C 216 Swimming Pool C 218 Environmental Application
CHROMIUM VI LOW RANGE ............. 45 ZINC ........................................... 121 Application
COLOR OF WATER .......................... 47 INTERFACE WITH PC ..................... 124 All meters use an exclusive positive-locking system to ensure that the cuvet
COPPER HIGH RANGE ..................... 49 DROPS/mL CROSS REFERENCE ...... 125 is in the same place every time it is placed into the measurement cell.
COPPER LOW RANGE ...................... 51 STANDARD METHODS ................... 126 The reagents are in liquid or powder form and are supplied in bottles
CYANIDE ....................................... 53 BATTERY REPLACEMENT ................ 127 or in packets. The amount of reagent is precisely dosed to ensure the
CYANURIC ACID ............................. 56 ACCESSORIES ............................. 128 maximum repeatability.
FLUORIDE ..................................... 59 WARRANTY ................................. 130 Display codes aid the user in routine operations.
CALCIUM HARDNESS ...................... 61 CE DECLARATION The meters have an auto-shut off feature, turning the unit off after
MAGNESIUM HARDNESS ................ 64 OF CONFORMITY ......................... 131 10 minutes of non-use.
TOTAL HARDNESS HIGH RANGE ........ 67 C 200 Series allows user to connect the meter to a personal computer
All rights are reserved. Reproduction in whole or in part is through three-wire RS 232. In such a way the meter can be fully
prohibited without the written consent of the copyright owner, controlled from personal computer via RS 232 under the supervision of
Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA. user's software.

2 3
A special tungsten lamp emits radiation which is first optically
PRINCIPLE OF OPERATION conditioned, then spectrally filtered to obtain a light beam of intensity
Io. This is characterized by a narrow spectral bandwidth, which is
Absorption of Light is a typical phenomenon of interaction between
beamed to the sample contained in the cuvet.
Electromagnetic Radiation and Matter. When a light beam crosses a
substance, some of the radiation may be absorbed by atoms, The optical path is fixed by the diameter of the cuvet.
molecules or crystal lattices. The photoelectric cell collects the radiation I that is not absorbed by
If pure absorption occurs, the fraction of light absorbed depends both the sample and converts it into an electric current, producing a
on the optical path length through the matter and on the physico- potential in the mV range.
chemical characteristics of substance according to the Lambert-Beer The microprocessor uses this potential to convert the incoming value
Law: into the desired measuring unit and to display it on the LCD.
-log I/Io = c d The measurement process is carried out in two phases: first the meter
or is zeroed and then the actual measurement is performed.
A = c d The cuvet has a very important role because it is an optical element
and thus requires particular attention. It is important that both the
Where -log I/Io = Absorbance (A) measurement and the calibration (zeroing) cuvets are optically
Io = intensity of incident light beam identical to provide the same measurement conditions. Whenever
I = intensity of light beam after absorption possible use the same cuvet for both.
= molar extinction coefficient at wavelength It is also necessary that the surface of the cuvet is clean and not
c = molar concentration of the substance scratched. This is to avoid measurement interference due to unwanted
d = optical path through the substance reflection and absorption of light. It is recommended not to touch the
Therefore, the concentration "c" can be calculated from the absorbance cuvet walls with hands.
of the substance as the other factors are known. Furthermore, in order to maintain the same conditions during the
Photometric chemical analysis is based on the possibility to develop an zeroing and the measuring phases, it is necessary to close the cuvet
absorbing compound from a specific chemical reaction between sample to prevent any contamination.
and reagents.
Given that the absorption of a compound strictly depends on the
wavelength of the incident light beam, a narrow spectral bandwidth
should be selected as well as a proper central wavelength to optimize
measurements.
Optical system of Hanna C 200 multiparameter photometers is based
on special subminiature tungsten lamps and narrow-band interfer-
ence filters to guarantee both high performance and reliable results.
Four measuring channels are available which, working at four
different wavelengths, allow a wide range of tests.

C 200 BLOCK DIAGRAM (OPTICAL LAYOUT)

4 5
Light under range. The zero sample is too
GUIDE TO DISPLAY CODES dark for proper zeroing. If this is not the
Note: The secondary LCD below shows a generic "P ", whereas the case, contact your dealer or the nearest
meter will indicate the exact program number (e.g. "P1" for Hanna Customer Service Center.
Aluminum). Light over range. The zero sample is too
This indicates that the meter is in a ready light for proper zeroing. If this is not the
state and zeroing can be performed. case, contact your dealer or the nearest
Hanna Customer Service Center.
Sampling in Progress. This prompt appears The blinking "LOBAT", while a measurement
each time the meter is performing a measure- is in progress, indicates that the battery
ment. voltage is getting low and the battery needs
to be replaced.
This indicates that the battery is dead and
This indicates that the meter is in a zeroed must be replaced.
state and measurement can be performed.

This indicates that the meter has lost its


This indicates that the zeroing procedure failed configuration. Contact your dealer or the
due to a low signal-to-noise ratio. In this case nearest Hanna Customer Service Center.
press ZERO again.

A zero reading was not taken. Follow the


instructions described in the measurement pro-
cedures for zeroing the meter.

Under range. A blinking "0.00" indicates


that the sample absorbs less light than the
zero reference. Check the procedure and make
sure you use the same cuvet for reference SPECIFICATIONS
(zero) and measurement.
Over range. A flashing value higher than the Light Life Life of the instrument
maximum concentration (see specifications) in- Light Detector Silicon Photocell
dicates that the sample absorbs too much
light, meaning the concentration is too high. Environment 0 to 50C (32 to 122F);
Dilute the sample. max 95% RH non-condensing
A flashing value lower than the maximum Power Supply 1 x 9 V battery with life of approximately 40
concentration indicates that the measurement hours of continuous use / 12 to 20 VDC through
has been performed in a low signal-to-noise voltage adapter
ratio condition. In this case accuracy of the
result is not guaranteed. Repeat the reading Auto-Shut off After 10' of non-use
procedure. Dimensions 230 x 165 x 70 mm (9.0 x 6.5 x 2.8")
Weight 640 g (22.6 oz.)
For specifications related to each single parameter (e.g. range,
accuracy, etc.), refer to the related measurement section.

6 7
FUNCTIONAL DESCRIPTION TIPS FOR AN ACCURATE MEASUREMENT
FRONT PANEL The instruction listed below should be carefully followed during testing
to ensure best accuracy.
Do not let the test sample stand too long after reagent is added
or accuracy will be lost.
Whenever the cuvet is placed into
the measurement cell, it must be
completely free of fingerprints, oil
or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior
to insertion.
It is important that the sample
does not contain any debris. This
would corrupt the readings.
Each time the cuvet is used, the cap must be tightened to the
1) Cuvet Holder same degree.
2) Dual Level Liquid Crystal Display It is possible to take multiple readings in a row, but it is
3) Programs List recommended that a zero reading be taken for each sample and
4) READ DIRECT, to perform measurement suddenly that the same cuvet is used for zeroing and measurement.
5) TIMER, to perform measurements after a preprogrammed It is important to discard the sample immediately after the
countdown reading is taken because the glass might become permanently
6) ZERO, to zero the meter prior to measurement stained.
7) Program and , to select the desired parameter Shaking the cuvet can generate bubbles in the sample, causing
8) ON/OFF, to turn the meter on and off higher readings. To obtain accurate measurements, remove such
REAR PANEL bubbles by swirling or by gently tapping the vial.
All the reaction times reported in this manual are referred to
20C (68F). As a general rule of thumb, they should be doubled
at 10C (50F) and halved at 30C (86F).

1) Power Socket 12V to 20V DC


2) RS 232 Socket
3) Battery Compartment

8 9
C 205 - BOILERS & COOLING TOWERS
PARAMETERS REFERENCE TABLES
Code Parameter Page Code Parameter Page
C 200 - LABORATORIES 1 Aluminum 18 13 Iron HR 80
Code Parameter Page Code Parameter Page 2 Ammonia MR 22 14 Iron LR 82
3 Ammonia LR 24 15 Molybdenum 90
1 Aluminum 18 19 Iodine 78
4 Bromine 26 16 Nitrate 98
2 Ammonia MR 22 20 Iron HR 80
5 Free Chlorine 28 17 Nitrite HR 100
3 Ammonia LR 24 21 Iron LR 82
6 Total Chlorine 34 18 Nitrite LR 102
4 Bromine 26 22 Manganese HR 85
7 Chlorine Dioxide 40 19 Dissolved Oxygen 104
5 Free Chlorine 28 23 Manganese LR 87
8 Chromium VI HR 43 20 pH 107
6 Total Chlorine 34 24 Molybdenum 90
9 Chromium VI LR 45 21 Phosphate HR 109
7 Chlorine Dioxide 40 25 Nickel HR 93
10 Copper HR 49 22 Phosphate LR 111
8 Chromium VI HR 43 26 Nitrate 98
11 Copper LR 51 23 Silica 115
9 Chromium VI LR 45 27 Nitrite HR 100
12 Hydrazine 76 24 Zinc 121
10 Color of Water 47 28 Nitrite LR 102
11 Copper HR 49 29 Dissolved Oxygen 104
12 Copper LR 51 30 pH 107
13 Cyanide 53 31 Phosphate HR 109 C 206 - ENVIRONMENTAL TESTING
14 Cyanuric Acid 56 32 Phosphate LR 111
Code Parameter Page Code Parameter Page
15 Fluoride 59 33 Phosphorus 113
1 Ammonia MR 22 13 Nickel LR 95
16 Ca Hardness 61 34 Silica 115
2 Ammonia LR 24 14 Nitrate 98
17 Mg Hardness 64 35 Silver 118
3 Free Chlorine 28 15 Nitrite HR 100
18 Hydrazine 76 36 Zinc 121
4 Total Chlorine 34 16 Nitrite LR 102
5 Chromium VI HR 43 17 Dissolved Oxygen 104
C 203 - AQUACULTURE 6 Chromium VI LR 45 18 pH 107
Code Parameter Page Code Parameter Page 7 Color of Water 47 19 Phosphate HR 109
1 Ammonia MR 22 8 Nitrite HR 100 8 Copper HR 49 20 Phosphate LR 111
2 Ammonia LR 24 9 Nitrite LR 102 9 Copper LR 51 21 Phosphorus 113
3 Free Chlorine 28 10 Dissolved Oxygen 104 10 Cyanuric Acid 56 22 Silica 115
4 Total Chlorine 34 11 pH 107 11 Molybdenum 90 23 Silver 118
5 Copper HR 49 12 Phosphate HR 109 12 Nickel HR 93 24 Zinc 121
6 Copper LR 51 13 Phosphate LR 111
7 Nitrate 98

10 11
C 207 - INDUSTRIAL WASTEWATER C 209 - EDUCATION
Code Parameter Page Code Parameter Page Code Parameter Page Code Parameter Page
1 Aluminum 18 11 Nickel HR 93 1 Ammonia MR 22 11 Nitrite HR 100
2 Free Chlorine 28 12 Nickel LR 95 2 Ammonia LR 24 12 Nitrite LR 102
3 Total Chlorine 34 13 Nitrate 98 3 Free Chlorine 28 13 Dissolved Oxygen 104
4 Color of Water 47 14 Dissolved Oxygen 104 4 Total Chlorine 34 14 pH 107
5 Copper HR 49 15 pH 107 5 Chromium VI HR 43 15 Phosphate HR 109
6 Copper LR 51 16 Phosphate HR 109 6 Chromium VI LR 45 16 Phosphate LR 111
7 Fluoride 59 17 Phosphate LR 111 7 Color of Water 47 17 Phosphorus 113
8 Manganese HR 85 18 Phosphorus 113 8 Copper HR 49 18 Silica 115
9 Manganese LR 87 19 Silver 118 9 Copper LR 51 19 Silver 118
10 Molybdenum 90 20 Zinc 121 10 Nitrate 98 20 Zinc 121

C 208 - WATER CONDITIONING


C 210 - PULP & PAPER MILLS
Code Parameter Page Code Parameter Page
Code Parameter Page Code Parameter Page
1 Ammonia MR 22 13 Nickel HR 93
1 Aluminum 18 7 pH 107
2 Ammonia LR 24 14 Nickel LR 95
2 Free Chlorine 28 8 Phosphate HR 109
3 Free Chlorine 28 15 Nitrate 98
3 Total Chlorine 34 9 Phosphate LR 111
4 Total Chlorine 34 16 Dissolved Oxygen 104
4 Chlorine Dioxide 40 10 Silica 115
5 Copper HR 49 17 pH 107
5 Color of Water 47 11 Silver 118
6 Copper LR 51 18 Phosphate HR 109
6 Dissolved Oxygen 104 12 Zinc 121
7 Fluoride 59 19 Phosphate LR 111
8 Iron HR 80 20 Phosphorus 113
9 Iron LR 82 21 Silica 115
10 Manganese HR 85 22 Silver 118
11 Manganese LR 87 23 Zinc 121
12 Molybdenum 90

12 13
C 211 - CHEMICAL MANUFACTURERS C 213 - MUNICIPAL WASTEWATER
Code Parameter Page Code Parameter Page Code Parameter Page Code Parameter Page
1 Aluminum 18 12 Molybdenum 90 1 Aluminum 18 13 Nickel HR 93
2 Ammonia MR 22 13 Nickel HR 93 2 Ammonia MR 22 14 Nickel LR 95
3 Ammonia LR 24 14 Nickel LR 95 3 Ammonia LR 24 15 Nitrate 98
4 Chromium VI HR 43 15 pH 107 4 Bromine 26 16 Nitrite HR 100
5 Chromium VI LR 45 16 Phosphate HR 109 5 Free Chlorine 28 17 Nitrite LR 102
6 Copper HR 49 17 Phosphate LR 111 6 Total Chlorine 34 18 Dissolved Oxygen 104
7 Copper LR 51 18 Phosphorus 113 7 Chromium VI HR 43 19 pH 107
8 Cyanuric Acid 56 19 Silica 115 8 Chromium VI LR 45 20 Phosphate HR 109
9 Iodine 78 20 Silver 118 9 Color of Water 47 21 Phosphate LR 111
10 Iron HR 80 21 Zinc 121 10 Copper HR 49 22 Phosphorus 113
11 Iron LR 82 11 Copper LR 51 23 Silver 118
12 Iodine 78 24 Zinc 121

C 216 - SWIMMING POOL

Code Parameter Page Code Parameter Page


C 212 - POWER PLANT UTILITIES
1 Free Chlorine HR 31 5 pH 107
Code Parameter Page Code Parameter Page
2 Total Chlorine HR 37 6 Total Hardness LR 73
1 Ammonia MR 22 8 Molybdenum 90
3 Cyanuric Acid 56 7 Total Hardness MR 70
2 Ammonia LR 24 9 Phosphate HR 109
4 Bromine 26 8 Total Hardness HR 67
3 Free Chlorine 28 10 Phosphate LR 111
4 Total Chlorine 34 11 Phosphorus 113
5 Copper HR 49 12 Silica 115 C 218 - ENVIRONMENTAL APPLICATION
6 Copper LR 51 13 Silver 118
7 Hydrazine 76 Code Parameter Page Code Parameter Page
1 Ammonia MR 22 6 Phosphorus 113
2 Ammonia HR 20 7 Nitrite HR 100
3 Cyanide 53 8 Nitrite LR 102
4 Chromium VI HR 43 9 Nitrate 98
5 Chromium VI LR 45

14 15
After the desired program number appears on the secondary
OPERATIONAL GUIDE display, follow the measurement procedure depicted in the related
chapter.
POWER CONNECTION
Plug the 12VDC adapter (HI 710005 - 110VDC, or HI 710006 - Note: in any moment it is possible to change the parameter
220VDC) into the DC socket. Plug the adapter into the outlet. measurement procedure by pressing the PROGRAM ! and
PROGRAM keys.
Alternatively, remove the battery cover on the back of the meter;
attach a fresh 9V battery and replace the cover.
Note: Insure the main line is surge protected.
Note: Always turn the meter off before unplugging it to insure no
data is lost.
MEASUREMENT PROCEDURE Note: in the following measurement sections,
a generic "P " will be placed on the
secondary LCD instead of the exact re-
Turn the meter on by pressing ON/OFF.
lated message (e.g. in C 200, "P1" for
Aluminum).

The meter will first perform an LCD self


diagnostic test by displaying a full set of
figures.
Then it will show a scrolling "c 2xy Hanna Inst" message.

When the LCD displays ----, the meter is


ready. On the secondary LCD "P1" will
appear to inform that the first parameter
measurement procedure (Aluminum) can
be performed.
Press the PROGRAM ! and PROGRAM keys to select the
desired parameter.

For the program number, see the tables on page 10 or look at


the list printed on the mask of the meter.

16 17
Fill two cuvets with 10 mL of sample 10 mL 10 mL
ALUMINUM each (up to the mark).
SPECIFICATIONS
Range 0.00 to 1.00 mg/L
Resolution 0.01 mg/L Add the content of one HI 93712C
Accuracy 0.02 mg/L 4% of reading bleaching powder packet to one
Typical EMC 0.01 mg/L of the two cuvets. Replace the
Deviation cap and shake gently. This is the
Light Source Tungsten lamp with narrow band interference blank.
filter @ 525 nm
Method Adaptation of the aluminon method. The reac- Place the blank into the holder and
tion between aluminum and reagents causes a
reddish tint in the sample. ensure that the notch on the cap is
positioned securely into the groove.
REQUIRED REAGENTS Press TIMER and the display will show the countdown prior to
Code Description Quantity zeroing the blank. Alternatively wait for 15 minutes and then
HI 93712A-0 Ascorbic acid 1 packet press ZERO. "SIP" will blink during zeroing.
HI 93712B-0 Aluminum reagent 1 packet
HI 93712C-0 Bleaching powder 1 packet
REAGENT SETS
HI 93712-01 Reagents for 100 tests
HI 93712-03 Reagents for 300 tests
For other accessories see page 128. Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
MEASUREMENT PROCEDURE
Select the program number corresponding to Aluminum on the
secondary LCD by pressing PROGRAM ! and . Remove the blank and insert the other
cuvet into the instrument.

Fill a graduated beaker with 50 mL of sample. Press READ DIRECT. "SIP" will blink during measurement.

Add the content of one packet of HI 93712A


ascorbic acid reagent and mix until dissolution is
complete. The instrument directly displays concentration in mg/L of alumi-
num on the Liquid Crystal Display.
INTERFERENCES
Add the content of one packet of HI 93712B Interference may be caused by:
aluminum reagent and mix until dissolution is Iron above 20 mg/L Alkalinity above 1000 mg/L
complete. This is the sample.
Phosphate above 50 mg/L Fluoride will interfere at all levels.

Aluminum 18 19 Aluminum
Remove the cuvet. 1 mL
AMMONIA HIGH RANGE of
sample
SPECIFICATIONS
Range 0.0 to 50.0 mg/L Fill another cuvet with 1 mL of
9 mL
Resolution 0.1 mg/L unreacted sample, by means of
Accuracy 0.05 mg/L 5% of reading the syringe.
Typical EMC 0.01 mg/L
Deviation Add 9 mL of HI 93733B Ammonia Reagent,
Light Source Tungsten lamp with narrow band interference by means of the 3 mL plastic pipette.
filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and Add 4 drops of HI 93733A
reagents causes a yellow tint in the sample. Nessler Reagent. Replace
REQUIRED REAGENTS the cap and swirl the solu- x4
Code Description Quantity tion.
HI 93733A-0 Nessler Reagent 4 drops (in fresh and seawater)
HI 93733B-0 Ammonia Reagent 9 mL (in fresh and seawater)
REAGENT SETS Reinsert the cuvet into the instrument.
HI 93733-01 Reagents for 100 tests
HI 93733-03 Reagents for 300 tests Press TIMER and the display will show the countdown prior to
For other accessories see page 128. measurement or, alternatively, wait for 3 minutes and 30 seconds
and then press READ DIRECT. In both cases the display will show
MEASUREMENT PROCEDURE SIP during measurement.
Select the program number corresponding to Ammonia HR on the
secondary LCD by pressing PROGRAM ! and .
Fill the cuvet with 10 mL of sample (up
to the mark) and replace the cap. 10 mL

Place the cuvet into the holder and The instrument directly displays concentration in mg/L of Ammo-
ensure that the notch on the cap is nium ion (NH4+) on the Liquid Crystal Display.
positioned securely into the groove. To convert the reading to mg/L of Ammonia (NH3), multiply by
Press ZERO and SIP will blink on the display. the factor 0.944. To convert the reading to ammonia nitrogen
(NH3-N), multiply by the factor 0.778.

INTERFERENCES
Interference may be caused by:
Wait for a few seconds and the display will acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron,
show -0.0-. Now the meter is zeroed organic chloramines, sulfide, various aliphatic and aromatic amines.
and ready for measurement.

Ammonia HR 20 21 Ammonia HR
Remove the cuvet.
AMMONIA MEDIUM RANGE
SPECIFICATIONS
Range 0.00 to 10.00 mg/L Add 4 drops of the first reagent
Resolution 0.01 mg/L (6 drops in case of seawater
Accuracy 0.05 mg/L 5% of reading analysis). Replace the cap and
Typical EMC 0.01 mg/L swirl the solution.
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Add 4 drops of the second re-
Method Adaptation of the ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler agent (10 drops in case of
method. The reaction between ammonia and seawater analysis). Replace the
reagents causes a yellow tint in the sample. cap and swirl the solution.
REQUIRED REAGENTS
Code Description Quantity
HI 93715A-0 Second Reagent 4 drops (10 drops in seawater)
HI 93715B-0 First Reagent 4 drops (6 drops in seawater) Reinsert the cuvet into the instrument.
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests Press TIMER and the display will show the countdown prior to the
For other accessories see page 128. measurement or, alternatively, wait for 3 minutes and 30 seconds
and press READ DIRECT. In both cases SIP will blink during
MEASUREMENT PROCEDURE measurement.
Select the program number corresponding to Ammonia MR on the
secondary LCD by pressing PROGRAM ! and .
10 mL
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.

Place the cuvet into the holder and The instrument directly displays concentration in mg/L of ammo-
ensure that the notch on the cap is nia nitrogen (NH3-N) on the display.
positioned securely into the groove.
To convert the reading to mg/L of ammonia (NH3), multiply by
Press ZERO and SIP will blink on the display. the factor 1.214.
INTERFERENCES
Interference may be caused by:
Acetone Alcohols
Aldehydes Glycine
Wait for a few seconds and the display will Hardness above 1 g/L Iron
show -0.0-. Now the meter is zeroed Organic chloramines Sulfide
and ready for measurement. Various aliphatic and aromatic amines.

Ammonia MR
22 23 Ammonia MR
Remove the cuvet.
AMMONIA LOW RANGE
SPECIFICATIONS Add 4 drops of the first reagent
Range 0.00 to 3.00 mg/L (6 drops in case of seawater
Resolution 0.01 mg/L analysis). Replace the cap and
Accuracy 0.04 mg/L 4% of reading swirl the solution.
Typical EMC 0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Add 4 drops of the second re-
Method Adaptation of the ASTM Manual of Water and
agent (10 drops in case of
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and seawater analysis). Replace the
reagents causes a yellow tint in the sample. cap and swirl the solution.
REQUIRED REAGENTS
Code Description Quantity
HI 93700A-0 Second Reagent 4 drops (10 drops in seawater)
Reinsert the cuvet into the instru-
HI 93700B-0 First Reagent 4 drops (6 drops in seawater)
ment.
REAGENT SETS
HI 93700-01 Reagents for 100 tests Press TIMER and the display will show the countdown prior to the
HI 93700-03 Reagents for 300 tests measurement or, alternatively, wait for 3 minutes and 30 seconds
For other accessories see page 128. and press READ DIRECT. In both cases SIP will blink during
MEASUREMENT PROCEDURE measurement.
Select the program number corresponding to Ammonia LR on the
secondary LCD by pressing PROGRAM ! and . 10 mL

Fill the cuvet with 10 mL of unreacted sample, up


to the mark, and replace the cap.
The instrument directly displays concentration in mg/L of ammo-
Place the cuvet into the holder and nia nitrogen (NH3-N) on the display.
ensure that the notch on the cap is To convert the reading to mg/L of ammonia (NH3), multiply the
positioned securely into the groove. display by a factor of 1.214.
Press ZERO and SIP will blink on the display. INTERFERENCES
Interference may be caused by:
Acetone Alcohols
Aldehydes Glycine
Hardness above 1 g/L Iron
Wait for a few seconds and the display will Organic chloramines Sulfide
show -0.0-. Now the meter is zeroed
Various aliphatic and aromatic amines.
and ready for measurement.

Ammonia LR 24 25 Ammonia LR
Wait for a few seconds and the display will
BROMINE show -0.0-. Now the meter is zeroed
and ready for measurement.
SPECIFICATIONS
Range 0.00 to 8.00 mg/L
Resolution 0.01 mg/L Remove the cuvet and add the
Accuracy 0.08 mg/L 3% of reading content of one packet of HI 93716
Typical EMC 0.01 mg/L reagent. Replace the cap and shake
Deviation gently.
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18 t h
edition, DPD method. The reaction between bro- Reinsert the cuvet into the instrument.
mine and the reagent causes a pink tint in the
sample.
REQUIRED REAGENTS Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and 30 seconds
Code Description Quantity
and press READ DIRECT. In both cases "SIP" will blink during
HI 93716-0 DPD Reagent 1 packet
measurement.
REAGENT SETS
HI 93716-01 Reagents for 100 tests
HI 93716-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
The instrument directly displays concentration in mg/L of bromine
Select the program number corresponding to Bromine on the
on the Liquid Crystal Display.
secondary LCD by pressing PROGRAM ! and .
INTERFERENCES
10 mL
Interference may be caused by:
Fill the cuvet with 10 mL of unreacted sample (up Chlorine
to the mark) and replace the cap. Iodine
Ozone
Place the cuvet into the holder and Oxidized forms of chromium and manganese
ensure that the notch on the cap is Alkalinity above 300 mg/L and acidity above 150 mg/L as CaCO3
positioned securely into the groove. require neutralization since the color could instantly fade. To resolve
this, neutralize the sample with diluted HCl or NaOH.
Press ZERO and SIP will blink on the display. In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 1 minute after adding the reagent.

Bromine 26 27 Bromine
Press ZERO and SIP will blink on the display.
FREE CHLORINE

SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L Wait for a few seconds and the display will
Accuracy 0.03 mg/L 3% of reading show -0.0-. Now the meter is zeroed
Typical EMC 0.01 mg/L and ready for measurement.
Deviation
Remove the cuvet.
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the EPA recommended DPD method Powder reagents procedure
330.5. The reaction between free chlorine and the Add the content of one packet of
DPD reagent causes a pink tint in the sample. HI 93701 DPD reagent. Replace
REQUIRED REAGENTS the cap and shake gently.
POWDER:
Code Description Quantity Wait for a few seconds, until all air
HI 93701-0 DPD 1 packet bubbles have completely dissolved and
reinsert the cuvet into the instrument.
LIQUID:
Code Description Quantity Press READ DIRECT and the display
HI 93701A-F DPD1 Indicator 3 drops will show SIP during measurement.
HI 93701B-F DPD1 Buffer 3 drops

REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder) The instrument directly displays concentration in mg/L of free
HI 93701-03 Reagents for 300 tests (powder) chlorine on the Liquid Crystal Display.
For other accessories see page 128. Liquid reagents procedure
MEASUREMENT PROCEDURE Add 3 drops of HI 93701A-F DPD1 indicator and 3 drops of
Select the program number corresponding to Free Chlorine on the HI 93701B-F DPD1 buffer to another cuvet. Shake gently before
secondary LCD by pressing PROGRAM ! and . adding 10 mL of unreacted sample. Replace the cap and shake
gently again.
10 mL

Fill the cuvet with 10 mL of unreacted sample (up


to the mark) and replace the cap.

Place the cuvet into the holder and x3 x3


ensure that the notch on the cap is
positioned securely into the groove.

28 29 Free Chlorine
Free Chlorine
FREE CHLORINE HIGH RANGE
Reinsert the cuvet into the instrument. SPECIFICATIONS
Range 0.00 to 10.00 mg/L
Resolution 0.01 mg/L
Wait for approximately 30 seconds and then press READ DIRECT; Accuracy 0.03 mg/L 3% of reading
the display will show SIP during measurement. Typical EMC 0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the EPA recommended DPD method
330.5. The reaction between free chlorine and the
The instrument directly displays concentration in mg/L of free DPD reagent causes a pink tint in the sample.
chlorine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: REQUIRED REAGENTS
Code Description Quantity
Oxidizing agents:
HI 93734A-0 Free & Total Chlorine Reagent 3 drops
Bromine
HI 93734B-0 Free & Total Chlorine Reagent 5 mL
Iodine
Fluorine
Ozone REAGENT SETS
Oxidized Manganese HI 93734-01 Reagents for 100 tests
Oxidized Chromium HI 93734-03 Reagents for 300 tests
In case of water with hardness greater than 500 mg/L CaCO3, shake For other accessories see page 128.
the sample for approximately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO3 or
acidity greater than 150 mg/L CaCO3, the color of the sample could MEASUREMENT PROCEDURE
disappear or develop only partially. To resolve this, neutralize the Select the program number corresponding to Free Chlorine HR on
sample with diluted HCl or NaOH. the secondary LCD by pressing PROGRAM ! and .

Fill the cuvet with 10 mL of unreacted 10 mL


sample (up to the mark) and replace
the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Free Chlorine 30 31 Free Chlorine HR


Press ZERO and SIP will blink on the display. Replace the cap
and shake gently
again.

Reinsert the cuvet into the instrument.


Wait for a few seconds and the display
will show -0.0-. Now the meter is
zeroed and ready for measurement. Wait for approximately 30 seconds and then press READ DIRECT.
The display will show SIP during measurement.

Remove the cuvet. 5 mL

Add to the other cuvet 3 drops x3 The instrument directly displays concentration in mg/L of free
of HI 93734A and 5 mL of chlorine on the Liquid Crystal Display.
HI 93734B, by means of the
syringe and shake gently.
Note: rinse the 3 mL plastic pipette 2 or 3 times with sample before
Note: To measure exactly 5 mL of reagent with the syringe, push the adding it to the cuvet with reagent.
plunger completely into the syringe and insert the tip into
HI 93734B reagent bottle. Pull the plunger out until the
lower edge of the seal is on the 5 mL mark of the syringe. INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Fluorine
Ozone
Oxidized manganese and Chromium
Alkalinity above 250 mg/L or acidity above 150 mg/L will not
reliably develop the full amount of color or it may rapidly fade. To
probable level of resolve this, neutralize the sample with diluted HCl or NaOH.
liquid taken up by In case of water with hardness greater than 500 mg/L CaCO3, shake
syringe the sample for approximately 1 minute after adding the reagents.
5 mL
of sample

Add 5 mL of unreacted sample, up to


the 10 mL mark of the cuvet, using the
3 mL plastic pipette. 10 mL

Free Chlorine HR 32 33 Free Chlorine HR


Press ZERO and SIP will blink on the display.
TOTAL C HLOR I N E
SPECIFICATIONS
Range 0.00 to 3.50 mg/L
Resolution 0.01 mg/L
Accuracy 0.03 mg/L 3% of reading Wait for a few seconds and the display will
Typical EMC 0.01 mg/L show -0.0-. Now the meter is zeroed
Deviation and ready for measurement.
Light Source Tungsten lamp with narrow band interference Remove the cuvet.
filter @ 525 nm
Powder reagents procedure
Method Adaptation of the EPA recommended DPD method
330.5. The reaction between the chlorine and the Add 1 packet of HI 93711-0.
DPD reagent causes a pink tint in the sample. Replace the cap and shake gen-
tly.
REQUIRED REAGENTS
Wait for a few seconds, until all air
POWDER:
Code Description Quantity bubbles have completely dissolved and
HI 93711-0 DPD 1 packet reinsert the cuvet into the instrument.
LIQUID: Press TIMER and the display will show the countdown prior to the
Code Description Quantity measurement or, alternatively, wait for 2 minutes and 30 seconds
HI 93701A-T DPD1 indicator 3 drops and press READ DIRECT. In both cases SIP will blink during
HI 93701B-T DPD1 buffer 3 drops measurement.
HI 93701C DPD3 solution 1 drop
REAGENT SETS
HI 93711-01 Reagents for 100 total chlorine tests (powder)
HI 93711-03 Reagents for 300 total chlorine tests (powder)
HI 93701-T Reagents for 300 total chlorine tests (liquid)
For other accessories see page 128. The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.

MEASUREMENT PROCEDURE Liquid reagents procedure


Select the program number corresponding to Total Chlorine on Add 3 drops of HI 93701A-T DPD1 indicator and 3 drops of
the secondary LCD by pressing PROGRAM ! and . HI 93701B-T DPD1 buffer to another cuvet. Add 1 drop of HI 93701C.
10 mL Shake gently before adding 10 mL of unreacted sample. Replace
the cap and shake gently again.
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.

Place the cuvet into the holder and x3 x3 x1


ensure that the notch on the cap is
positioned securely into the groove.

34 35 Total Chlorine
Total Chlorine
TOTAL CHLORINE HIGH RANGE
Reinsert the cuvet into the instrument. SPECIFICATIONS
Range 0.00 to 10.00 mg/L
Resolution 0.01 mg/L
Wait for approximately 30 seconds and then press READ DIRECT, Accuracy 0.03 mg/L 3% of reading
the display will show SIP during measurement. Typical EMC 0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the EPA recommended DPD method
330.5. The reaction between total chlorine and
the DPD reagent causes a pink tint in the
The instrument directly displays concentration in mg/L of total sample.
chlorine on the Liquid Crystal Display.
REQUIRED REAGENTS
Note: Free and total chlorine have to be measured separately with Code Description Quantity
fresh unreacted samples following the related procedure if both HI 93734A-0 Free & Total Chlorine Reagent 3 drops
values are requested. HI 93734B-0 Free & Total Chlorine Reagent 5 mL
INTERFERENCES HI 93734C-0 Total Chlorine Reagent 1 drop
Interference may be caused by: REAGENT SETS
Bromine HI 93734-01 Reagents for 100 tests
Iodine HI 93734-03 Reagents for 300 tests
Fluorine For other accessories see page 128.
Ozone MEASUREMENT PROCEDURE
Oxidized manganese and Chromium Select the program number corresponding to Total Chlorine HR on
Alkalinity above 250 mg/L or acidity above 150 mg/L will not the secondary LCD by pressing PROGRAM ! and .
reliably develop the full amount of color or it may rapidly fade. To Fill the cuvet with 10 mL of unreacted
resolve this, neutralize the sample with diluted HCl or NaOH. sample (up to the mark) and replace
In case of water with hardness greater than 500 mg/L CaCO3, shake the cap. 10 mL
the sample for approximately 1 minute after adding the reagent.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

36 37
Total Chlorine Total Chlorine HR
Wait for a few seconds and the display will
show -0.0-. Now the meter is zeroed Reinsert the cuvet into the instrument.
and ready for measurement.
Remove the cuvet.
Add to the other cuvet 3 drops of HI 93734A, 5 mL of
HI 93734B, by means of the 5 mL syringe, and 1 drop of HI Wait for approximately 30 seconds and then press READ DIRECT,
93734C. Shake gently for 10 seconds. the display will show SIP during measurement.

5 mL

x1
x3
The instrument directly displays concentration in mg/L of total
chlorine on the Liquid Crystal Display.

Note: To measure exactly 5 mL of reagent with the syringe, push the Notes: free and total chlorine have to be measured separately with
plunger completely into the syringe and insert the tip into fresh unreacted samples following the related procedure if both
HI 93734B reagent bottle. Pull the plunger out until the values are requested.
lower edge of the seal is on the 5 mL mark of the syringe. Rinse the 3 mL plastic pipette 2 or 3 times with sample before
adding it to the cuvet with reagent.

INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Fluorine
probable level of liquid Ozone
taken up by syringe
Oxidized manganese and Chromium
5 mL Alkalinity above 250 mg/L or acidity above 150 mg/L will not
Add 5 mL of unreacted sample, up to the of sample reliably develop the full amount of color or it may rapidly fade. To
10 mL mark, by means of the 3 mL resolve this, neutralize the sample with diluted HCl or NaOH.
plastic pipette. In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 1 minute after adding the reagents.
10 mL

Replace the cap and shake


gently again.

Total Chlorine HR 38 39 Total Chlorine HR


Add the content of one packet of HI 93738B dechlorinating
CHLORINE DIOXIDE reagent to one only of the two cylinders (#1), close it and invert
several times to mix. This is the blank.
SPECIFICATIONS
Range 0.00 to 2.00 mg/L

B
Resolution 0.01 mg/L
Accuracy 0.1 mg/L 5% of reading
Typical EMC 0.01 mg/L
Deviation #1
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the Standard Methods for the Add precisely 0.5 mL of HI 93738C chlorine dioxide reagent to
Examination of Water and Wastewater, 18 t h each cylinder (#1 & #2), close them and invert several times to
edition, Chlorophenol Red method. The reaction mix.
between chlorine dioxide and reagents causes a
purple to greenish-yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93738A-0 Reagent A 1 mL
HI 93738B-0 Dechlorinating Reagent B 1 packet C

HI 93738C-0 Reagent C 1 mL
HI 93738D-0 Reagent D 1 mL
Add 0.5 mL of HI 93738D chlorine dioxide reagent to each
REAGENT SETS cylinder (#1 & #2), close them and invert several times to mix.
HI 93738-01 Reagents for 100 tests Cylinder #2 is the sample.
HI 93738-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Chlorine Dioxide on
the secondary LCD by pressing PROGRAM ! and .
Fill two graduated D
mixing cylinders
(#1 & #2) up to #1
the 25 mL mark with
the sample.
Add 0.5 mL of HI 93738A chlorine dioxide reagent to each Fill a cuvet with 10 mL of the 10 mL

cylinder (#1 & #2), close them and invert several times to mix. blank (#1) up to the mark and
replace the cap.

Place the blank (#1) into the holder


and ensure that the notch on the cap
is positioned securely into the groove.
A

Chlorine Dioxide 40 41 Chlorine Dioxide


Press ZERO and "SIP" will blink on the display.
CHROMIUM VI HIGH RANGE
SPECIFICATIONS
Range 0 to 1000 g/L
Resolution 1 g/L
Wait for a few seconds and the display will Accuracy 5 g/L 4% of reading
show "-0.0-". Now the meter is zeroed and Typical EMC 1 g/L
ready for measurement. Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
#2 Method Adaptation of the ASTM Manual of Water and
Environmental Technology, D1687-92,
Fill a cuvet with 10 mL of the diphenylcarbohydrazide method. The reaction be-
reacted sample (#2) up to 10 mL tween chromium VI and the reagent causes a
the mark and replace the cap. purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
Insert the sample into the instrument. HI 93723-0 Powder reagent 1 packet
REAGENT SETS
Press READ DIRECT and "SIP" will blink HI 93723-01 Reagents for 100 tests
during measurement. HI 93723-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Chromium VI HR on
the secondary LCD by pressing PROGRAM ! and .
The instrument directly displays concentration in mg/L of chlorine Fill the cuvet up to the mark with 10 mL of
10 mL

dioxide on the Liquid Crystal Display. unreacted sample and replace the cap.
SAMPLING PROCEDURE Place the cuvet into the holder and
It is recommended to analyze chlorine dioxide samples immediately ensure that the notch on the cap is
upon collection. Chlorine dioxide samples must be stored in dark, glass positioned securely into the groove.
stoppered bottles, with minimal head space. Excessive heat (greater
than 25C/78F), agitation and exposure to light must be avoided. Press ZERO and SIP will blink on the display.

INTERFERENCES
Interferences may be caused by strong oxidants.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Chlorine Dioxide 42 43 Chromium VI HR


CHROMIUM VI LOW RANGE
Remove the cuvet and add the
content of one packet of HI 93723 SPECIFICATIONS
reagent. Replace the cap and Range 0 to 300 g/L
shake vigorously. Resolution 1 g/L
Accuracy 1 g/L 4% of reading
Reinsert the cuvet into the instru- Typical EMC 1 g/L
ment. Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Press TIMER and the display will show the countdown prior to the Method Adaptation of the ASTM Manual of Water and
measurement or, alternatively, wait for 6 minutes and press READ Environmental Technology, D1687-92, diphenyl-
DIRECT. In both cases "SIP" will blink during measurement. carbohydrazide method. The reaction between
chromium VI and the reagent causes a purple
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93749-0 Powder reagent 1 packet
The instrument directly displays concentration in g/L of chro- REAGENT SETS
mium VI on the Liquid Crystal Display. HI 93749-01 Reagents for 100 tests
INTERFERENCES HI 93749-03 Reagents for 300 tests
Interference may be caused by: For other accessories see page 128.
Vanadium above 1 ppm, however, 10' after the reaction the
interference is removed
MEASUREMENT PROCEDURE
Iron above 1 ppm Select the program number corresponding to Chromium VI LR on
Mercurous and mercuric ions, slight inhibition of the reaction. the secondary LCD by pressing PROGRAM ! and .
Fill the cuvet with 10 mL of unreacted sample 10 mL

(up to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Chromium VI HR
44 45 Chromium VI LR
Remove the cuvet and add the COLOR OF WATER
content of one packet of HI 93749 SPECIFICATIONS
reagent. Replace the cap and
Range 0 to 500 PCU (Platinum Cobalt Units)
shake vigorously.
Resolution 1 PCU
Accuracy 10 PCU 5% of reading
Typical EMC 1 PCU
Reinsert the cuvet into the instru- Deviation
ment. Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Method Adaptation of the Colorimetric Platinum Cobalt
Press TIMER and the display will show the countdown prior to the method.
measurement or, alternatively, wait for 6 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement. REQUIRED ACCESSORIES
0.45 m membrane for true color measurement.
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Color of Water on
the secondary LCD by pressing PROGRAM ! and .
The instrument directly displays concentration in g/L of chro-
mium VI on the Liquid Crystal Display. Fill one cuvet up to the mark with deionized 10 mL

water and replace the cap. This is the


INTERFERENCES blank.
Interference may be caused by:
Vanadium above 1 ppm, however, 10' after the reaction the Fill another cuvet up to the mark with 10 mL

interference is removed unfiltered sample and replace the cap. This is


Iron above 1 ppm the apparent color.
Mercurous and mercuric ions, slight inhibition of the reaction.

Filter 50 mL of sample through a filter with a


0.45 m membrane.

Fill the third cuvet up to the 10 mL mark with 10 mL

the filtered sample and replace the cap. This is


the true color.

Place the blank into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Chromium VI LR
46 47 Color of Water
Press ZERO and "SIP" will blink on the display.
COPPER HIGH RANGE

SPECIFICATIONS
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Wait for a few seconds and the display will Accuracy 0.02 mg/L 4% of reading
show "-0.0-". Now the meter is zeroed and Typical EMC 0.01 mg/L
ready for measurement. Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Remove the blank, insert the appar- Method Adaptation of the EPA approved method. The
ent color cuvet into the instrument reaction between copper and the bicinchoninate
and ensure that the notch on the cap reagent causes a purple tint in the sample.
is positioned securely into the groove. REQUIRED REAGENTS
Code Description Quantity
Press READ DIRECT and "SIP" will blink on the display. HI 93702-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
The meter directly displays the value of apparent color. For other accessories see page 128.
Remove the cuvet, insert the true color cuvet into the instrument MEASUREMENT PROCEDURE
and ensure that the notch on the cap is positioned securely into Select the program number corresponding to Copper HR on the
the groove. secondary LCD by pressing PROGRAM ! and .
Press READ DIRECT and "SIP" will blink on the display.
Fill the cuvet with 10 mL of unreacted sample 10 mL

(up to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.
The meter directly displays concentration in color units of true color
on the Liquid Crystal Display. Press ZERO and SIP will blink on the display.
INTERFERENCES
There are no known interferences for this method.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.
Remove the cuvet.

Color of Water 48 49 Copper HR


COPPER LOW RANGE
Add the content of one packet
of HI 93702 reagent. Replace SPECIFICATIONS
the cap and shake gently. Range 0 to 1000 g/L
Resolution 1 g/L
Accuracy 10 g/L 5% of reading
Reinsert the cuvet into the instrument. Typical EMC 1 g/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Press TIMER and the display will show the countdown prior to the Method Adaptation of the EPA approved method. The
measurement or, alternatively, wait for 45 seconds and press reaction between copper and the bicinchoninate
READ DIRECT. In both cases SIP will blink during measurement. reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93747-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93747-01 Reagents for 100 tests
INTERFERENCES HI 93747-03 Reagents for 300 tests
Interference may be caused by: For other accessories see page 128.
Silver
Cyanide MEASUREMENT PROCEDURE
For samples overcoming buffering capacity of reagent (around pH 6.8), Select the program number corresponding to Copper LR on the
pH should be adjusted between 6 and 8. secondary LCD by pressing PROGRAM ! and .
Fill the cuvet with 10 mL of unreacted sample 10 mL

(up to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0-. Now the meter is zeroed and
ready for measurement.

50 51 Copper LR
Copper HR
Remove the cuvet.
Add the content of one packet of CYANIDE
HI 93747 reagent. Replace the SPECIFICATIONS
cap and shake gently until the Range 0.000 to 0.200 mg/L
reagent is dissolved. Resolution 0.001 mg/L
Accuracy 0.005 mg/L 3% of reading
Reinsert the cuvet into the instru- Typical EMC Dev. 0.001 mg/L
ment. Light Source Tungsten lamp with narrow band interference
filter @ 610 nm
Method Adaptation of the Standard Methods for the
Press TIMER and the display will show the countdown prior to the Examination of Water and Wastewater, 18th
measurement or, alternatively, wait for 45 seconds and press edition, Pyridine-Pyrazalone method. The reac-
READ DIRECT. In both cases "SIP" will blink during measurement. tion between cyanide and reagents causes an
orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93714A-0 Reagent A 1 spoon
HI 93714B-0 Reagent B 1 packet
HI 93714C-0 Reagent C 1 packet
The instrument directly displays concentration in g/L of copper REAGENT SETS
on the Liquid Crystal Display. HI 93714-01 Reagents for 100 tests
INTERFERENCES HI 93714-03 Reagents for 300 tests
Interference may be caused by: For other accessories see page 128.
Silver MEASUREMENT PROCEDURE
Cyanide Select the program number corresponding to Cyanide on the
For samples overcoming buffering capacity of reagent (around pH 6.8),
secondary LCD by pressing PROGRAM ! and .
pH should be adjusted between 6 and 8. 10 mL
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

52 53 Cyanide
Copper LR
The instrument directly displays concentration in mg/L of cyanide
Remove the cuvet and add 1 level on the Liquid Crystal Display. To convert the result in ppm of
spoon of HI 93714A Cyanide Re- Potassium Cyanide (KCN) multiply it by 2.5.
agent. Remember to close the Note: Shake the cuvet 4 or 5 times vigorously during the first 20
reagent bottle immediately after use. minutes of the countdown prior to the measurement. Accuracy
Note: Pay attention to the way the is not affected by undissolved reagent powder.
spoon is filled:
- do not overfill it; INTERFERENCES
Interference may be caused by large amounts of turbidity that will
cause high readings.
- do not press the powder. Oxidizing (like chlorine) or reducing agents (such as sulfide or sulfur
Replace the cap immediately, to dioxide) are known to interfere with the measurement. Distillation will
prevent the escape of chlorine gas remove these.
which is developed during the reac- Samples with high pH values should be adjusted to approximately
tion, and shake gently for 30 seconds. pH 7 before testing.

Wait for 30 seconds leaving the

B
CAUTION: cyanides and their solutions and the hydrogen cyanide
cuvet tightly capped and undis-
liberated by acids are very poisonous.
turbed, then add the content of
10"
one packet of HI 93714B reagent 12

and shake gently for 10 seconds.


9 3

6
C

Immediately add the content of


one packet of HI 93714C re-
agent, replace the cap and shake
vigorously for 20 seconds.

Reinsert the cuvet into the instru-


ment.

Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 25 minutes and press
READ DIRECT. In both cases "SIP" will blink during measurement.

Cyanide 54 55 Cyanide
Fill a beaker up to the
CYANURIC ACID 25 mL mark with the
sample, add the con-
SPECIFICATIONS
tent of one packet of
Range 0 to 80 mg/L
HI 93722 reagent
Resolution 1 mg/L
and stir gently to mix.
Accuracy 1 mg/L 15% of reading
Typical EMC 1 mg/L 10 mL

Deviation Fill a second cuvet with 10 mL of the reacted


Light Source Tungsten lamp with narrow band interference sample up to the mark. Replace the cap.
filter @ 525 nm
Method Adaptation of the turbidimetric method. The reac-
tion between cyanuric acid and the reagent
causes a white suspension in the sample. Reinsert the cuvet into the instru-
ment.
REQUIRED REAGENTS
Code Description Quantity
Press TIMER and the display will show the countdown prior to the
HI 93722-0 Powder reagent 1 packet
measurement or, alternatively, wait for 45 seconds and press
REAGENT SETS READ DIRECT. In both cases "SIP" will blink during measurement.
HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Cyanuric Acid on the
secondary LCD by pressing PROGRAM ! and . 10 mL The instrument directly displays concentration in mg/L of cyanuric
Fill the cuvet with 10 mL of unreacted sample (up acid on the Liquid Crystal Display.
to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0-. Now the meter is zeroed and
ready for measurement.

Cyanuric Acid 56 57 Cyanuric Acid


Place the cuvet with the reacted
FLUORIDE distilled water into the holder and
ensure that the notch on the cap is
SPECIFICATIONS
positioned securely into the groove.
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy 5% of reading Press ZERO and "SIP" will blink on the display.
Typical EMC 0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the SPADNS method. The reaction
between fluoride and the liquid reagent causes a Wait for a few seconds and the display will
red tint in the sample. show "-0.0-". Now the meter is zeroed and
REQUIRED REAGENT ready for measurement.
Code Description Quantity
HI 93729-0 SPADNS Reagent 4 mL
REAGENT SETS Insert the other cuvet with the reacted
HI 93729-01 Reagents for 100 tests sample into the instrument.
HI 93729-03 Reagents for 300 tests
For other accessories see page 128.
Press READ DIRECT and "SIP" will blink on the LCD during
MEASUREMENT PROCEDURE measurement.
Select the program number corresponding to Fluoride on the
secondary LCD by pressing PROGRAM ! and .
10 mL

Fill one of the cuvets with 10 mL of distilled


water (up to the mark).

10 mL
The instrument directly displays concentration in mg/L of fluoride
Fill the other cuvet with 10 mL of sample (up to on the Liquid Crystal Display.
the mark).
Note: With wastewater and seawater samples, before performing any
measurement a distillation of the sample is needed.

Add to each cuvet 2 mL of HI 93729


SPADNS reagent, replace the caps and
invert several times to mix. INTERFERENCES
Negative interferences may be caused by:
12
1 Alkalinity (as CaCO3) above 5000 mg/L
Aluminum above 0.1 mg/L
Wait for one minute. 9 3
Iron, ferric above 10 mg/L
6

58 59 Fluoride
Fluoride
Positive interferences may be caused by:
Chloride above 700 mg/L CALCIUM HARDNESS
Phosphate, ortho above 16 mg/L
SPECIFICATIONS
Sodium hexametaphosphate above 1.0 mg/L
Range 0.00 to 2.70 mg/L
Sulfate above 200 mg/L
Resolution 0.01 mg/L
Highly colored and turbid samples may require distillation
Accuracy 0.11 mg/L 5% of reading
Highly alkaline samples can be neutralized with nitric acid.
Typical EMC 0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18th
edition, Calmagite method. The reaction between
calcium and reagents causes a red tint in the
sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93720A-0 Ca & Mg indicator 0.5 mL
HI 93720B-0 Alkali solution 0.5 mL
HI 93720C-0 EGTA solution 1 drop
REAGENT SETS
HI 93720-01 Reagents for 100 tests
HI 93720-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Hardness Ca on the
secondary LCD by pressing PROGRAM ! and .
Fill a graduated bea-
ker to the 50 mL mark
with the sample.

Add 0.5 mL of HI 93720A Cal-


cium indicator solution and swirl
to mix.

Add 0.5 mL of Alkali solution for


HI 93720B Calcium and Magnesium
and swirl to mix.
10 mL 10 mL

Fill two cuvets up to the mark with


10 mL of sample each.

Fluoride 60 61 Hardness Ca
SAMPLE DILUTION
This meter is designed to determine low levels of hardness, typically
Add 1 drop of HI 93720C EGTA solution found in water purification systems.
to one cuvet, replace the cap and swirl When testing some other sources of water, it is not uncommon to come
the solution. This is the blank. across levels of hardness that are greater than the range of this
meter.
Place the blank into the holder and This problem can be overcome through dilution. Dilutions must be
ensure that the notch on the cap is performed with hardness-free water or the readings will be erroneous.
positioned securely into the groove. A dilution to reduce the level of hardness by a factor of one hundred
is performed as follows:
Fill a 1 mL syringe with the sample.
Press ZERO and "SIP" will blink on the display. Place the syringe in a 50 mL beaker, making sure that the
beaker is clean and empty, and inject 0.5 mL into the beaker.
Fill the beaker up to the 50 mL mark with hardness-free water.
Now, follow normal measurement procedure. The true value of the
sample is the reading obtained multiplied by a factor of one hundred
(the dilution factor).
Wait for a few seconds and the display For your reference, factors to convert readings in mg/L to French
will show "-0.0-". Now the meter is ze- degrees (FD), German degrees (DD) and English degrees (ED) of
roed and ready for measurement. hardness are as follows:
1 mg/L = 0.1 FD = 0.056 DD = 0.07 ED.
INTERFERENCES
Remove the buffered blank and insert
the second cuvet into the instrument. Interference may be caused by excessive amounts of heavy metals.

Press READ DIRECT. "SIP" will blink during measurement.

The instrument directly displays concentration in mg/L (ppm) of


calcium, as CaCO 3, on the Liquid Crystal Display.
To convert the result to mg/L Ca, multiply by 0.4.
Note: The test will detect any calcium or magnesium contamination
in the mixing cylinders, measuring droppers or sample cells. To
test cleanliness, repeat the test multiple times until you obtain
consistent results.
Note: For better accuracy wash glassware with HCl 6N.

Hardness Ca
62 63 Hardness Ca
10 mL 10 mL
MAGNESIUM HARDNESS Fill two cuvets up to the mark
SPECIFICATIONS with 10 mL of sample each.
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy 0.11 mg/L 5% of reading Add 1 drop of HI 93719C EDTA
Typical EMC 0.02 mg/L solution to one cuvet, replace the
Deviation cap and swirl the solution. This is
Light Source Tungsten lamp with narrow band interference the blank.
filter @ 525 nm Add 1 drop of HI 93719D EGTA
Method Adaptation of the Standard Methods for the solution to the second cuvet, replace
Examination of Water and Wastewater, 18th
the cap and swirl the solution. This
edition, EDTA colorimetric method. The reaction D

between magnesium and reagents causes a violet is the sample.


tint in the sample.
REQUIRED REAGENTS Place the blank into the holder and
ensure that the notch on the cap is
Code Description Quantity positioned securely into the groove.
HI 93719A-0 Ca & Mg indicator 0.5 mL
HI 93719B-0 Alkali solution 0.5 mL
HI 93719C-0 EDTA solution 1 drop Press ZERO and "SIP" will blink on the display.
HI 93719D-0 EGTA solution 1 drop
REAGENT SETS
HI 93719-01 Reagents for 100 tests
HI 93719-03 Reagents for 300 tests
For other accessories see page 128. Wait for a few seconds and the display
MEASUREMENT PROCEDURE will show "-0.0-". Now the meter is ze-
Select the program number corresponding to Hardness Mg on the roed and ready for measurement.
secondary LCD by pressing PROGRAM ! and .
Remove the blank and insert the sample
Fill a graduated beaker
into the instrument, making sure that
to the 50 mL mark with
the notch on the cap is positioned
the sample.
securely into the groove.
Press TIMER and the display will show the countdown prior to the
Add 0.5 mL of HI 93719A Cal- measurement. Alternatively wait for exactly 30 seconds and press
cium and Magnesium indicator READ DIRECT. In both cases SIP will blink during measurement.
solution, then swirl to mix.

Add 0.5 mL of HI 93719B Alkali


solution for Calcium and Magne-
sium, then swirl to mix.

Hardness Mg
64 65 Hardness Mg
The instrument directly displays concentration in mg/L (ppm) of
magnesium hardness, as CaCO3, on the Liquid Crystal Display. TOTAL HARDNESS HIGH RANGE
To convert the result to mg/L Mg, multiply the result by 0.243.
SPECIFICATIONS
Note: The test will detect any calcium or magnesium contamination
Range 400 to 750 mg/L
in the mixing cylinders, measuring droppers or sample cells. To
Resolution 5 mg/L
test cleanliness, repeat the test multiple times until you obtain
Accuracy 10 mg/L or 2% of reading,
consistent results.
whichever is greater
SAMPLE DILUTION Typical EMC 5 mg/L
This meter is designed to determine hardness typically found in water Deviation
purification systems. In order to measure samples with high hardness, Light Source Tungsten lamp with narrow band interference
follow dilution procedure explained on page 63 (Ca Hardness). filter @ 466 nm
Method Adaptation of the EPA recommended method
130.1. The reaction between calcium, magnesium
For your reference, factors to convert readings in mg/L to French and reagents causes a red-violet tint in the
degrees (FD), German degrees (DD) and English degrees (ED) of sample.
hardness are as follows: REQUIRED REAGENTS
1 mg/L = 0.1 FD = 0.056 DD = 0.07 ED. Code Description Quantity
INTERFERENCES HI 93735A-HR Hardness Indicator reagent HR 9.5 mL
HI 93735B-0 Hardness Buffer reagent 2 drops
Interference may be caused by excessive amounts of heavy metals. HI 93735C-0 Fixing reagent 1 packet
REAGENT SETS
HI 93735HR Reagents for 100 tests High Range
(400 to 750 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Total
Hardness HR on the secondary LCD by pressing
PROGRAM ! and .
Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
Note: for accurate results, fill the syringe up to the
1 mL mark, then transfer 0.5 mL to the
cuvet.
With the plastic pipette fill the cuvet up to the
10 mL mark adding HI 93735A-HR Indicator
reagent.
Note: the liquid in the cuvet forms a
convexity on the top; the bottom
of this convexity must be at the 10 mL
same level of the 10 mL mark.

66 67 Total Hardness HR
Hardness Mg
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
Add 2 drops of HI 93735B
Buffer reagent. x2

Replace the cap and shake gently to mix.


The instrument directly displays the hardness in ppm of CaCO3, on
the Liquid Crystal Display.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove. Use the conversion factors to obtain the reading in f, D and E

1 ppm = 0.1 f = 0.05556 D = 0.07 E


Press ZERO and "SIP" will blink on the display.

INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
Note: if the sample is very acidic, some extra drops of HI 93735B
Buffer reagent may be added.
Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.

Remove the cuvet and add the content


of 1 packet of HI 93735C Fixing re-
agent.

Replace the cap and


shake gently to mix.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

68 69 Total Hardness HR
Total Hardness HR
TOTAL HARDNESS MEDIUM RANGE
SPECIFICATIONS Add 2 drops of HI 93735B
Range 200 to 500 mg/L Buffer reagent x2
Resolution 5 mg/L
Accuracy 7 mg/L or 3% of reading,
whichever is greater
Typical EMC 5 mg/L
Deviation Replace the cap and shake gently to mix.
Light Source Tungsten lamp with narrow band interference
filter @ 466 nm
Method Adaptation of the EPA recommended method
130.1. The reaction between calcium, magnesium Place the cuvet into the holder and
and reagents causes a red-violet tint in the ensure that the notch on the cap is
sample. positioned securely into the groove.
REQUIRED REAGENTS
Code Description Quantity Press ZERO and "SIP" will blink on the display.
HI 93735A-MR Hardness Indicator reagent MR 9.5 mL
HI 93735B-0 Hardness Buffer reagent 2 drops
HI 93735C-0 Fixing reagent 1 packet
REAGENT SETS
HI 93735MR Reagents for 100 tests Medium Range
(200 to 500 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L) Wait for a few seconds and the display
For other accessories see page 128. will show "-0.0-". Now the meter is
zeroed and ready for measurement.
MEASUREMENT PROCEDURE
Select the program number corresponding to Total
Hardness MR on the secondary LCD by pressing
PROGRAM ! and .
Using the syringe, add exactly 0.5 mL of Remove the cuvet and add the con-
unreacted sample to the cuvet. tent of 1 packet of HI 93735C Fixing
Note: for accurate results, fill the syringe up to the reagent.
1 mL mark, then transfer 0.5 mL to the
cuvet.
With the plastic pipette fill the cuvet up to the Replace the cap and
10 mL mark adding HI 93735A-MR Indicator shake gently to mix.
reagent.
Note: the liquid in the cuvet forms a
convexity on the top; the bottom Place the cuvet into the holder and
of this convexity must be at the 10 mL ensure that the notch on the cap is
same level of the 10 mL mark. positioned securely into the groove.

Total Hardness MR 70 71 Total Hardness MR


Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ TOTAL HARDNESS LOW RANGE
DIRECT. In both cases "SIP" will blink during measurement. SPECIFICATIONS
Range 0 to 250 mg/L
Resolution 1 mg/L from 0 to 100 mg/L
5 mg/L from 100 to 250 mg/L
Accuracy 5 mg/L or 4% of reading,
whichever is greater
Typical EMC 5 mg/L
Deviation
The instrument directly displays the hardness in ppm of CaCO3, on Light Source Tungsten lamp with narrow band interference
the Liquid Crystal Display. filter @ 466 nm
Method Adaptation of the EPA recommended method
130.1. The reaction between calcium, magnesium
Use the conversion factors to obtain the reading in f, D and E and reagents causes a red-violet tint in the
sample.
1 ppm = 0.1 f = 0.05556 D = 0.07 E REQUIRED REAGENTS
Code Description Quantity
HI 93735A-LR Hardness Indicator reagent LR 9.5 mL
INTERFERENCES HI 93735B-0 Hardness Buffer reagent 2 drops
Interference may be caused by excessive amounts of heavy metals. HI 93735C-0 Fixing reagent 1 packet
Note: if the sample is very acidic, some extra drops of HI 93735B REAGENT SETS
Buffer reagent may be added. HI 93735LR Reagents for 100 tests Low Range (0 to 250 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Total
Hardness LR on the secondary LCD by pressing
PROGRAM ! and .
Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
Note: for accurate results, fill the syringe up to the
1 mL mark, then transfer 0.5 mL to the
cuvet.
With the plastic pipette fill the cuvet up to the
10 mL mark adding HI 93735A-LR Indicator
reagent.
Note: the liquid in the cuvet forms a
convexity on the top; the bottom
of this convexity must be at the 10 mL
same level of the 10 mL mark.

Total Hardness MR 72 73 Total Hardness LR


Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 2 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
Add 2 drops of HI 93735B-0
Buffer reagent. x2

Replace the cap and shake gently to mix. The instrument directly displays the hardness in ppm of CaCO3, on
the Liquid Crystal Display.

Place the cuvet into the holder and


Use the conversion factors to obtain the reading in f, D and E
ensure that the notch on the cap is
positioned securely into the groove.
1 ppm = 0.1 f = 0.05556 D = 0.07 E

Press ZERO and "SIP" will blink on the display.


INTERFERENCES
Interference may be caused by excessive amounts of heavy metals.
Note: if the sample is very acidic, some extra drops of HI 93735B
Buffer reagent may be added.

Wait for a few seconds and the display


will show "-0.0-". Now the meter is
zeroed and ready for measurement.

Remove the cuvet and add the con-


tent of 1 packet of HI 93735C Fixing
reagent.

Replace the cap and shake gently to mix.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Total Hardness LR 74 75 Total Hardness LR


Place the blank into the holder and
HYDRAZINE ensure that the notch on the cap is
positioned securely into the groove.
SPECIFICATIONS
Range 0 to 400 g/L Press TIMER and the display will show the countdown prior to
Resolution 1 g/L zeroing the blank. Alternatively wait for 12 minutes and then
Accuracy 3% of full scale press ZERO. "SIP" will blink during zeroing.
Typical EMC 2 g/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Method Adaptation of the ASTM Manual of Water and
Environmental Technology, method D1385-88 for
natural and treated water. The reaction between Wait for a few seconds and the display will
hydrazine and the liquid reagent causes a yellow show -0.0-. Now the meter is zeroed
tint in the sample. and ready for measurement.
REQUIRED REAGENT Remove the blank.
Code Description Quantity
HI 93704-0 Liquid Reagent 12 drops Insert the cuvet with the reacted sample
REAGENT SETS into the instrument.
HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests Press READ DIRECT and SIP will blink during measurement.
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Hydrazine on the
secondary LCD by pressing PROGRAM ! and .
10 mL
The instrument directly displays concentration in g/L of hydra-
Fill one cuvet with 10 mL of unreacted sample (up zine on the Liquid Crystal Display.
to the mark) and replace the cap.
INTERFERENCES
10 mL
Fill a second cuvet up to the mark with 10 mL of Interference may be caused by:
distilled water (this is the blank). Highly colored samples
Highly turbid samples
Aromatic amines
Presence of oxidizing agents will reduce the concentration of hydra-
Add 12 drops of the HI 93704 zine.
reagent to each cuvet. Replace
the caps and swirl the solu- x 12
tions.

76 77 Hydrazine
Hydrazine
Remove the cap and add the con-
IODINE tent of one packet of HI 93718
reagent. Replace the cap and
SPECIFICATIONS
shake gently.
Range 0.0 to 12.5 mg/L
Resolution 0.1 mg/L
Accuracy 0.1 mg/L 5% of reading
Typical EMC 0.1 mg/L
Deviation Reinsert the cuvet into the instrument.
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18 t h Press TIMER and the display will show the countdown prior to the
edition, DPD method. The reaction between iodine measurement or, alternatively, wait for 2 minutes and 30 seconds
and the reagent causes a pink tint in the and press READ DIRECT. In both cases "SIP" will blink during
sample. measurement.
REQUIRED REAGENTS
Code Description Quantity
HI 93718-0 DPD Reagent 1 packet
REAGENT SETS
HI 93718-01 Reagents for 100 tests The instrument directly displays concentration in mg/L of iodine
HI 93718-03 Reagents for 300 tests on the Liquid Crystal Display.
For other accessories see page 128.
INTERFERENCES
MEASUREMENT PROCEDURE Interference may be caused by:
Select the program number corresponding to Iodine on the Bromine
secondary LCD by pressing PROGRAM ! and .
10 mL
Chlorine
Fill the cuvet with 10 mL of unreacted sample (up Ozone
to the mark) and replace the cap. Oxidized forms of chromium and manganese
Place the cuvet into the holder and Alkalinity above 300 mg/L and acidity above 150 mg/L as CaCO3
ensure that the notch on the cap is require neutralization since the color could instantly fade. To resolve
positioned securely into the groove. this, neutralize the sample with diluted HCl or NaOH.
Press ZERO and SIP will blink on the display. In case of water with hardness greater than 500 mg/L CaCO3, shake
the sample for approximately 1 minute after adding the reagent.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Iodine 78 79 Iodine
Remove the cuvet and add the con-
IRON HIGH RANGE tent of one packet of HI 93721
SPECIFICATIONS reagent. Replace the cap and agitate
until dissolution is complete.
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Accuracy 0.04 mg/L 2% of reading
Typical EMC 0.01 mg/L
Deviation Reinsert the cuvet into the instru-
Light Source Tungsten lamp with narrow band interference ment.
filter @ 525 nm
Method Adaptation of the phenantroline EPA recommended
method 315B for natural and treated waters.
The reaction between iron and reagents causes Press TIMER and the display will show the countdown prior to the
an orange tint in the sample. measurement or, alternatively, wait for 3 minutes and press READ
REQUIRED REAGENTS DIRECT. In both cases "SIP" will blink during measurement.
Code Description Quantity
HI 93721-0 Powder Reagent 1 packet
REAGENT SETS
HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
For other accessories see page 128.
The instrument directly displays concentration in mg/L of iron on
MEASUREMENT PROCEDURE the Liquid Crystal Display.
Select the program number corresponding to Iron HR on the
secondary LCD by pressing PROGRAM ! and . INTERFERENCES
Interference may be caused by:
Fill the cuvet with 10 mL of unreacted sample 10 mL
Molybdate Molybdenum above 50 ppm
(up to the mark) and replace the cap.
Calcium above 10,000 ppm (as CaCO3)
Magnesium above 100,000 ppm (as CaCO3)
Place the cuvet into the holder and Chloride above 185,000 ppm.
ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Iron HR
80 81 Iron HR
IRON LOW RANGE Place the cuvet into the holder and
ensure that the notch on the cap is
SPECIFICATIONS
positioned securely into the groove.
Range 0 to 400 g/L
Resolution 1 g/L
Accuracy 10 g/L 8% of reading Press ZERO and "SIP" will blink on the display.
Typical EMC 1 g/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the TPTZ Method. The reaction
between iron and the reagent causes a blue tint Wait for a few seconds and the display will
in the sample. show "-0.0-". Now the meter is zeroed and
REQUIRED REAGENTS ready for measurement.
Code Description Quantity
HI 93746-0 TPTZ Reagent 2 packets
REAGENT SETS Fill one graduated mixing cylinder up to the
HI 93746-01 Reagents for 50 tests 25 mL mark with the sample.
HI 93746-03 Reagents for 100 tests
For other accessories see page 128. Add the content of one packet of HI 93746 TPTZ reagent, close
MEASUREMENT PROCEDURE the cylinder and agitate for 30 seconds.
Select the program number corresponding to Iron LR on the
secondary LCD by pressing PROGRAM ! 12
30"
and . 9 3

Fill one graduated mixing cylinder up to the


6

25 mL mark with deionized water.


Add the content of one packet of HI 93746 TPTZ reagent, close
the cylinder and agitate for 30 seconds. This is the blank.
Fill a cuvet with 10 mL of the reacted
12
30" sample up to the mark and replace 10 mL
9 3 the cap. This is the sample.
6

Insert the sample into the instrument.


Fill a cuvet with 10 mL of the blank up 10 mL

to the mark and replace the cap.

Iron LR 82 83 Iron LR
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 30 seconds and press MANGANESE HIGH RANGE
READ DIRECT. In both cases "SIP" will blink during measurement. SPECIFICATIONS
Range 0.0 to 20.0 mg/L
Resolution 0.1 mg/L
Accuracy 0.2 mg/L 3% of reading
Typical EMC 0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
The instrument directly displays concentration in g/L of iron on filter @ 575 nm
the Liquid Crystal Display. Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18th edi-
tion, persulfate method. The reaction between
INTERFERENCES manganese and reagents causes a violet tint in
Interference may be caused by: the sample.
Cadmium above 4.0 mg/L
Chromium3+ above 0.25 mg/L REQUIRED REAGENTS
Chromium6+ above 1.2 mg/L Code Description Quantity
Cobalt above 0.05 mg/L HI 93709A-0 Citrate 1 packet
Copper above 0.6 mg/L HI 93709B-0 Sodium periodate 1 packet
Cyanide above 2.8 mg/L
Manganese above 50.0 mg/L REAGENT SETS
Mercury above 0.4 mg/L HI 93709-01 Reagents for 100 tests
Molybdenum above 4.0 mg/L HI 93709-03 Reagents for 300 tests
Nickel above 1.0 mg/L
Nitrite ion above 0.8 mg/L For other accessories see page 128.
Sample pH should be between 3 and 4 to avoid developed color to MEASUREMENT PROCEDURE
fade or turbidity.
Select the program number corresponding to Manganese HR on
the secondary LCD by pressing PROGRAM ! and .
Fill the cuvet with 10 mL of unreacted sample 10 mL
(up to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

Iron LR 84 85 Manganese HR
Wait for a few seconds and the display will
show -0.0-. Now the meter is zeroed MANGANESE LOW RANGE
and ready for measurement.
SPECIFICATIONS
Remove the cuvet. Range 0 to 300 g/L
Resolution 1 g/L

A
Add the content of one packet of Accuracy 2 g/L 3% of reading
HI 93709A citrate reagent. Re- Typical EMC 1 g/L
place the cap and shake gently. Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the PAN Method. The reaction
between manganese and the reagents causes an

B
Add the content of one packet of orange tint in the sample.
HI 93709B sodium periodate re-
agent. Replace the cap and shake REQUIRED REAGENT
gently. Code Description Quantity
HI 93748A-0 Ascorbic acid 2 packets
HI 93748B-0 Alkaline-cyanide sol. 0.40 mL
HI 93748C-0 0.1% PAN indicator 2 mL
Reinsert the cuvet into the instrument. HI 93703-51 Dispersing Agent 4-6 drops (only when
necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for 50 tests
Press TIMER and the display will show the countdown prior to the HI 93748-03 Reagents for 150 tests
measurement or, alternatively, wait for 1 minute and 30 seconds For other accessories see page 128.
and press READ DIRECT. In both cases SIP will blink during
measurement.
MEASUREMENT PROCEDURE
Select the program number corresponding to Manganese LR on
the secondary LCD by pressing PROGRAM ! and .
10 mL

Fill one cuvet with 10 mL of deionized water


The instrument directly displays concentration in mg/L of manga- (up to the mark).
10 mL
nese on the Liquid Crystal Display.
Fill a second cuvet with 10 mL of sample (up to
INTERFERENCES the mark).
Interference may be caused by:

HI93748A
Calcium above 700 mg/L
Add the content of one packet of
Chloride above 70000 mg/L
HI 93748A ascorbic acid to each
Iron above 5 mg/L cuvet, replace the caps and shake
Magnesium above 100000 mg/L. gently.

Manganese HR 86 87 Manganese LR
The instrument directly displays concentration in g/L of manga-
Add 0.2 mL of the nese on the display.
HI 93748B alkaline-cyanide Note: a temperature above 30C may cause turbidity. In this case,
reagent solution to each cu- before zeroing and taking readings, add 2-3 drops of Dispersing
vet, replace the caps and B
Agent (HI 93703-51) to each cuvet and swirl until complete dissolu-
shake gently. tion of the turbidity.
INTERFERENCES
Interference may be caused by:
Add 1 mL of the HI 93748C
Aluminum above 20 mg/L
0.1% PAN indicator solution
Cadmium above 10 mg/L
to each cuvet, replace the C
Calcium above 200 mg/L as CaCO3
caps and shake gently.
Cobalt above 20 mg/L
Copper above 50 mg/L
Place the cuvet with the reacted deion- Iron above 10 mg/L
ized water (blank) into the holder and Lead above 0.5 mg/L
ensure that the notch on the cap is Magnesium above 300 mg/L as CaCO3
positioned securely into the groove. Nickel above 40 mg/L
Zinc above 15 mg/L.
Press TIMER and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 2 minutes and then
press ZERO. "SIP" will blink during zeroing.

Wait for a few seconds and the display will


show "-0.0-". Now the meter is zeroed and
ready for measurement.

Insert the second cuvet with the re-


acted sample into the instrument.

Press READ DIRECT. "SIP" will blink during measurement.

88 89
Manganese LR Manganese LR
Fill one graduated mixing cylinder up to
MOLYBDENUM the 25 mL mark with the sample.
SPECIFICATIONS
Range 0.0 to 40.0 mg/L Add the content of one packet of HI 93730A molybdenum
Resolution 0.1 mg/L reagent, close the cylinder and invert several times until the
Accuracy 0.3 mg/L 5% of reading powder is completely dissolved.
Typical EMC 0.1 mg/L

A
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm
Method Adaptation of the mercaptoacetic acid method.
The reaction between molybdenum and the re-
agents causes a yellow tint in the sample.
REQUIRED REAGENT Add the content of one packet of HI 93730B molybdenum
Code Description Quantity reagent to the cylinder, close it and invert several times until the
HI 93730A-0 Reagent A 1 packet powder is completely dissolved.
HI 93730B-0 Reagent B 1 packet
HI 93730C-0 Reagent C 1 packet

B
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE Add the content of one packet of HI 93730C molybdenum
Select the program number corresponding to Molybdenum on the reagent to the cylinder, close it and shake for 15" until powder
secondary LCD by pressing PROGRAM ! and . has dissolved.
10 mL

C
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to the
Press ZERO and SIP will blink on the display. measurement or, alternatively, wait for 5 minutes and press READ
DIRECT. During the 5 minutes waiting time, fill an empty cuvet
with 10 mL of the reacted sample up to the mark, replace the

Wait for a few seconds and the display will


10 mL
show -0.0-. Now the meter is zeroed
and ready for measurement.

Molybdenum
90 91 Molybdenum
cap and reinsert the cuvet into the instrument. "SIP" will blink
during measurement. NICKEL HIGH RANGE
SPECIFICATIONS
Range 0.00 to 7.00 g/L
Resolution 0.01 g/L
Accuracy 4% of reading
Typical EMC 0.02 g/L
The instrument directly displays concentration in mg/L of molyb- Deviation
denum on the Liquid Crystal Display. Light Source Tungsten lamp with narrow band interference
INTERFERENCES filter @ 575 nm
Interference may be caused by: Method Adaptation of the photometric method. The reac-
tion between nickel and the reagent causes a
Aluminum above 50 mg/L blue tint in the sample.
Chromium above 1000 mg/L
Copper above 10 mg/L REQUIRED REAGENTS
Iron above 50 mg/L Code Description Quantity
Nickel above 50 mg/L HI 93726-0 Powder reagent 1 packet
Nitrite, as NO2
REAGENT SETS
Sulfate above 200 mg/L
HI 93726-01 Reagents for 100 tests
Highly buffered samples or with extreme pH may exceed the buffering
HI 93726-03 Reagents for 300 tests
capacity of the reagents.
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Nickel HR on the
secondary LCD by pressing PROGRAM ! and .
10 mL
Fill the cuvet up to the mark with 10 mL of
unreacted sample and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Molybdenum 92 93 Nickel HR
Remove the cuvet and add the NICKEL LOW RANGE
content of one packet of HI 93726 SPECIFICATIONS
reagent. Replace the cap and Range 0.000 to 1.000 mg/L
shake gently. Resolution 0.001 mg/L
Accuracy 0.01 mg/L 7% of reading
Reinsert the cuvet into the instru- Typical EMC 0.001 mg/L
ment. Deviation
Light Source Tungsten lamp with narrow band interference
Press TIMER and the display will show the countdown prior to the filter @ 575 nm
measurement or, alternatively, wait for 1 minute and press READ Method Adaptation of the 1-(2-pyridylazo)-2-naphtol
DIRECT. In both cases "SIP" will blink during measurement. (PAN) method. The reaction between nickel and
the reagent causes an orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93740A-0 Phthalate phosphate 2 packets
HI 93740B-0 0.3% PAN indicator 2 mL
The instrument directly displays concentration in g/L of nickel on HI 93740C-0 EDTA 2 packets
the Liquid Crystal Display. HI 93703-51 Dispersing Agent 4-6 drops (only when
necessary, see note)
INTERFERENCES
Interference may be caused by copper. REAGENT SETS
HI 93740-01 Reagents for 50 tests
HI 93740-03 Reagents for 150 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Nickel LR on the
secondary LCD by pressing PROGRAM ! and .
Note: for best results perform your tests between 20-24C.

Fill one graduated beaker with 25 mL of


deionized water (blank) and another one with
25 mL of sample.

Add the content of one packet of HI 93740A


phthalate-phosphate reagent to each beaker.
Cap and shake gently until the reagent is
dissolved.
Note: If sample contains iron (Fe3+), it is impor-
tant that all powder be dissolved completely
before continuing with following step.

94 95
Nickel HR Nickel LR
Press READ DIRECT and "SIP" will appear during measurement.

Add 1 mL of HI 93740B 0.3% PAN


solution to each beaker, swirl to mix.

The instrument directly displays concentration in mg/L of nickel on


Press TIMER and the display will show the countdown prior to
the Liquid Crystal Display.
adding reagent C or, alternatively, wait for 15 minutes. Then, in
both cases, add one packet of HI 93740C EDTA reagent to each Note: a temperature above 30C may cause turbidity. In this case,
beaker, cap and swirl to mix until complete dissolution. before zeroing and taking readings, add 2-3 drops of Dispersing
Agent (HI 93703-51) to each cuvet and swirl until complete dissolu-
tion of the turbidity.
INTERFERENCES
10 mL Interference may be caused by:
Co2+ must not be present
Fill one cuvet up to the mark with 10 mL of the
blank. Fe2+ must not be present
Al3+ above 32 mg/L
Place the cuvet into the holder and Ca2+ above 1000 mg/L (as CaCO3)
ensure that the notch on the cap is Cd2+ above 20 mg/L
positioned securely into the groove. Cl above 8000 mg/L
Cr3+ above 20 mg/L
Press ZERO and "SIP" will appear on the display. Cr6+ above 40 mg/L
Cu2+ above 15 mg/L
F above 20 mg/L
Fe3+ above 10 mg/L
K+ above 500 mg/L
Mg2+ above 400 mg/L
Wait for a few seconds and the display will
Mn2+ above 25 mg/L
show "-0.0-". Now the meter is zeroed and
Mo6+ above 60 mg/L
ready for measurement.
Na+ above 5000 mg/L
Pb2+ above 20 mg/L
Fill a second cuvet up to the mark with 10 mL of 10 mL
Zn2+ above 30 mg/L
the reacted sample.

Insert the second cuvet into the


instrument.

96 97 Nickel LR
Nickel LR
Remove the cuvet and add the content of one
NITRATE packet of HI 93728 reagent.
SPECIFICATIONS
Range 0.0 to 30.0 mg/L Replace the cap and im-
Resolution 0.1 mg/L mediately shake vigorously
Accuracy 0.5 mg/L 10% of reading for exactly 10 seconds by
moving the cuvet up and
Typical EMC 0.1 mg/L
down. Continue to mix by
Deviation
inverting the cuvet gently
Light Source Tungsten lamp with narrow band interference and slowly for 50 seconds,
filter @ 525 nm while taking care not to
Method Adaptation of the cadmium reduction method. induce air bubbles. A de-
The reaction between nitrate-nitrogen and the posit could remain but it
reagent causes an amber tint in the sample. does not affect the measurement. Time and way of shaking could
REQUIRED REAGENTS sensitively affect the measurement.
Code Description Quantity
HI 93728-0 Powder reagent 1 packet Reinsert the cuvet into the instru-
REAGENT SETS ment, taking care not to shake it.
HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 128. Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 4 minutes and 30 seconds
MEASUREMENT PROCEDURE
and press READ DIRECT. In both cases "SIP" will blink during
Select the program number corresponding to Nitrate on the measurement.
secondary LCD by pressing PROGRAM ! and . 10 mL
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is The instrument directly displays concentration in mg/L of nitrate-
positioned securely into the groove. nitrogen on the Liquid Crystal Display.
To convert the reading to mg/L of nitrate (NO3), multiply by a
Press ZERO and SIP will blink on the display. factor of 4.43.

INTERFERENCES
Interference may be caused by:
Ammonia and amines, as urea and primary aliphatic amines
Wait for a few seconds and the display Chloride above 100 ppm (negative interference)
will show -0.0-. Now the meter is ze- Chlorine above 2 ppm (positive interference)
roed and ready for measurement.

Nitrate 98 99 Nitrate
Remove the cuvet.
NITRITE HIGH RANGE
Add the content of one packet of
SPECIFICATIONS HI 93708 reagent. Replace the
Range 0 to 150 mg/L cap and shake gently.
Resolution 1 mg/L
Accuracy 4 mg/L 4% of reading
Typical EMC 1 mg/L
Deviation Reinsert the cuvet into the instrument.
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the Ferrous Sulfate method. The
reaction between nitrite and the reagent causes a Press TIMER and the display will show the countdown prior to the
greenish-brown tint in the sample. measurement or, alternatively, wait for 10 minutes and press
READ DIRECT. In both cases SIP will blink during measurement.
REQUIRED REAGENTS
Code Description Quantity
HI 93708-0 Powder reagent 1 packet
REAGENT SETS
HI 93708-01 Reagents for 100 tests
HI 93708-03 Reagents for 300 tests
For other accessories see page 128. The instrument directly displays concentration in mg/L of nitrite
on the Liquid Crystal Display.
MEASUREMENT PROCEDURE To convert the NO2 ion concentration to NO2-N concentration,
Select the program number corresponding to Nitrite HR on the divide the reading by a factor of 3.29.
secondary LCD by pressing PROGRAM ! and . To convert the NO2 ion to sodium nitrite NaNO2 concentration,
Fill the cuvet up to the mark with 10 mL of 10 mL multiply the reading by a factor of 1.5.
unreacted sample, and replace the cap. INTERFERENCES
C 200 Series will not recognize any glycol based samples.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Nitrite HR 100 101 Nitrite HR


Remove the cuvet.
NITRITE LOW RANGE

SPECIFICATIONS Add the content of one packet of


Range 0.00 to 0.35 mg/L HI 93707 reagent. Replace the
Resolution 0.01 mg/L cap and shake gently.
Accuracy 0.02 mg/L 4% of reading
Typical EMC 0.01 mg/L
Deviation Reinsert the cuvet into the instru-
Light Source Tungsten lamp with narrow band interference ment.
filter @ 525 nm
Method Adaptation of the EPA method 354.1. The reac-
tion between nitrite and the reagent causes a Press TIMER and the display will show the countdown prior to the
pink tint in the sample. measurement or, alternatively, wait for 6 minutes and press READ
DIRECT. In both cases SIP will blink during measurement.
REQUIRED REAGENTS
Code Description Quantity
HI 93707-0 Powder reagent 1 packet
REAGENT SETS
HI 93707-01 Reagents for 100 tests
HI 93707-03 Reagents for 300 tests The instrument directly displays concentration in mg/L of nitrite
For other accessories see page 128. nitrogen.
MEASUREMENT PROCEDURE To convert the NO2-N concentration to the nitrite ion concentra-
Select the program number corresponding to Nitrite LR on the tion (NO2), multiply the reading by factor of 3.29.
secondary LCD by pressing PROGRAM ! and . To convert the NO2-N concentration to sodium nitrite concentra-
10 mL tion (NaNO2), multiply the reading by factor of 4.93.
Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap. INTERFERENCES
Interference may be caused by the following ions:
Place the cuvet into the holder and ferrous, ferric, cupric, mercurous, silver, antimonious, bismuth, auric,
ensure that the notch on the cap is lead, metavanadate and chloroplatinate.
positioned securely into the groove. Strongly reducing and oxidizing reagents.
High levels of nitrate (above 100 mg/L) could yield falsely high
Press ZERO and SIP will blink on the display. readings due to a minute amount of reduction to nitrite that could
occur at these levels.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Nitrite LR 102 103 Nitrite LR


Swirl the solution. The sample becomes orange and a flocculant
D IS S OLV E D OX Y GEN agent will appear.
SPECIFICATIONS
Range 0.0 to 10.0 mg/L (ppm)
Resolution 0.1 mg/L
Accuracy 0.4 mg/L 3% of reading
Typical EMC 0.1 mg/L
Let the sample stand and the flocculant agent will
Deviation
start to settle.
Light Source Tungsten lamp with narrow band interference
filter @ 420 nm After approximately 2 minutes, when the upper
Method Adaptation of the Standard Methods for the half of the bottle becomes limpid, add 10 drops of
Examination of Water and Wastewater (18 t h HI 93732C.
edition), azide modified Winkler method. The
reaction between dissolved oxygen and the re- Replace the cap and swirl the solution. The sample is ready for
agent causes a yellow tint in the sample. measurement when it is yellow and completely limpid.
REQUIRED REAGENTS
Code Description Quantity
HI 93732A-0 Reagent A 5 drops
HI 93732B-0 Reagent B 5 drops
HI 93732C-0 Reagent C 10 drops
REAGENT SET
Fill the cuvet up to the mark with 10 mL of the 10 mL
HI 93732-01 Reagents for 100 tests
unreacted (original) sample, and replace the cap.
HI 93732-03 Reagents for 300 tests
This is the blank.
For other accessories see page 128.
MEASUREMENT PROCEDURE Place the cuvet into the holder and
Select the program number corresponding to Dissolved Oxygen on ensure that the notch on the cap is
the secondary LCD by pressing PROGRAM ! and . positioned securely into the groove.

Fill one 60 mL BOD bottle completely with the unreacted Press ZERO and SIP will blink on the display.
sample.

Replace the cap and ensure that a small part of the sample spills
over. This is to make sure that no air bubbles have been trapped
inside.
Wait for a few seconds and the display will
Remove the cap and add 5 drops of show -0.0-. Now the meter is zeroed
HI 93732A and 5 drops of and ready for measurement.
HI 93732B. Replace the cap again
and ensure that a part of the
sample spills over. Remove the cuvet and dispose of the blank.

Dissolved Oxygen 104 105 Dissolved Oxygen


Rinse the cuvet with some of the reacted
pH
sample. Then, fill it up to the mark with SPECIFICATIONS
10 mL of the reacted sample and replace Range 5.9 to 8.5 pH
the cap. Resolution 0.1 pH
Accuracy 0.1 pH
Typical EMC 0.2 pH
Reinsert the cuvet into the instru- Deviation
ment. Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Phenol red method. The reac-
Press READ DIRECT and "SIP" will blink during measurement. tion with the reagent causes a red tint in the
sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93710-0 Phenol red 5 drops
The instrument will then directly display the concentration of dis- REAGENT SETS
solved oxygen in mg/L. HI 93710-01 Reagents for 100 pH tests
INTERFERENCES HI 93710-03 Reagents for 300 pH tests
For other accessories see page 128.
Interferences may be caused by reducing and oxidizing materials.
MEASUREMENT PROCEDURE
Select the program number corresponding to pH on the secondary
LCD by pressing PROGRAM ! and . 10 mL

Fill the cuvet with 10 mL of unreacted sample (up


to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.

Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Dissolved Oxygen 106 107 pH


Remove the cuvet and add 5 PHOSPHATE HIGH RANGE
drops of HI 93710 Phenol Red SPECIFICATIONS
Indicator. Replace the cap and Range 0.0 to 30.0 mg/L
swirl the solution. x5
Resolution 0.1 mg/L
Accuracy 1 mg/L 4% of reading
Typical EMC Dev. 0.1 mg/L
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Reinsert the cuvet into the instrument. Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater (18 t h
edition), amino acid method. The reaction be-
Press the READ DIRECT key and SIP will blink on the display tween phosphate and reagents causes a blue
during measurement. tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93717A-0 Molybdate 10 drops
HI 93717B-0 Amino Acid 0.5 mL
HI 93717B-P Amino Acid powder 2 packets for 100 tests
The instrument directly displays the pH measured value on the
Liquid Crystal Display. REAGENT SETS
HI 93717-01 Reagents for 100 tests
HI 93717-03 Reagents for 300 tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Phosphate HR on
the secondary LCD by pressing PROGRAM ! and ".
Fill the cuvet with 10 mL of unreacted sample 10 mL

(up to the mark) and replace the cap.


Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and SIP will blink on
the display.

Wait for a few seconds and the display will


show "-0.0-". Now the meter is zeroed and
ready for measurement.

pH 108 109 Phosphate HR


Remove the cuvet.
PHOSPHATE LOW RANGE
Add the content of one packet of HI 93717B-P Amino
Acid powder to HI 93717B Amino Acid bottle when
you use it for the first time only. After you have SPECIFICATIONS
added the powder, replace the cap of the bottle Range 0.00 to 2.50 mg/L
and shake it for 1 minute to dissolve the powder. Resolution 0.01 mg/L
Leave the bottle undisturbed for 1 minute, then the Accuracy 0.04 mg/L 4% of reading
reagent is ready. Typical EMC 0.01 mg/L
Note: after adding the Amino Acid powder, the HI 93717B reagent Deviation
will last for one month. Light Source Tungsten lamp with narrow band interference
filter @ 610 nm
Method Adaptation of the ascorbic acid method. The
Add 10 drops of HI 93717A Molybdate reagent. reaction between phosphate and the reagent
x 10 causes a blue tint in the sample.
REQUIRED REAGENTS
Add to the cuvet 0.5 mL of Code Description Quantity
the solution prepared in the HI 93713-0 Powder reagent 1 packet
HI 93717B bottle. Replace the
cap and swirl the solution.
B
REAGENT SETS
HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
Reinsert the cuvet into the instrument. For other accessories see page 128.
MEASUREMENT PROCEDURE
Select the program number corresponding to Phosphate LR on
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 5 minutes and press READ the secondary LCD by pressing PROGRAM ! and .
DIRECT. In both cases "SIP" will blink during measurement. Fill the cuvet with 10 mL of unreacted sample (up 10 mL

to the mark) and replace the cap.

Place the cuvet into the holder and


ensure that the notch on the cap is
positioned securely into the groove.
The instrument directly displays concentration in mg/L of phos-
phate (PO4) on the Liquid Crystal Display. Press ZERO and SIP will blink on the display.
INTERFERENCES
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L as Cl
Calcium above 10000 mg/L as CaCO3
Wait for a few seconds and the display will
Magnesium above 40000 mg/L as CaCO3
show -0.0-. Now the meter is zeroed
Ferrous iron above 100 mg/L.
and ready for measurement.

Phosphate HR 110 111 Phosphate LR


Remove the cuvet and add the content
PHOSPHORUS
of one packet of HI 93713 reagent. SPECIFICATIONS
Replace the cap and shake gently. Range 0.0 to 15.0 mg/L
Resolution 0.1 mg/L
Accuracy 0.3 mg/L 4% of reading
Typical EMC dev. 0.2 mg/L
Reinsert the cuvet into the instru- Light Source Tungsten lamp with narrow band interference
ment. filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18th edi-
tion, amino acid method. The reaction between
Press TIMER and the display will show the countdown prior to the phosphorus and the reagents causes a blue tint
measurement or, alternatively, wait for 3 minutes and press READ in the sample.
DIRECT. In both cases "SIP" will blink during measurement.
REQUIRED REAGENTS
Code Description Quantity
HI 93706A-0 Molybdate 10 drops
HI 93706B-0 Amino Acid 0.5 mL
HI 93706B-P Amino Acid powder 2 packets for 100 tests
The instrument directly displays concentration in mg/L of phos- REAGENT SETS
phate on the Liquid Crystal Display. HI 93706-01 Reagents for 100 tests
To convert the PO43 ion concentration to P2O5 concentration, HI 93706-03 Reagents for 300 tests
multiply the reading by a factor of 0.747. For other accessories see page 128.
To convert the PO43 ion concentration to Phosphate P concentra-
MEASUREMENT PROCEDURE
tion, multiply the reading by a factor of 0.326.
Select the program number corresponding to Phosphorus on the
INTERFERENCES secondary LCD by pressing PROGRAM ! and . 10 mL
Interference may be caused by:
Fill the cuvet with 10 mL of unreacted sample (up
Iron above 50 mg/L to the mark) and replace the cap.
Silica above 50 mg/L Place the cuvet into the holder and
Silicate above 10 mg/L ensure that the notch on the cap is
Copper above 10 mg/L positioned securely into the groove.
Hydrogen sulfide, arsenate, turbid sample and highly buffered
samples also interfere. Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show "-0.0-". Now the meter is zeroed and
ready for measurement.

Phosphate LR
112 113 Phosphorus
Remove the cuvet.
Add the content of one packet of HI 93706B-P Amino SILICA
Acid powder to HI 93706B Amino Acid bottle when SPECIFICATIONS
you use it for the first time only. After you have Range 0.00 to 2.00 mg/L
added the powder, replace the cap of the bottle Resolution 0.01 mg/L
and shake it for 1 minute to dissolve the Accuracy 0.03 mg/L 3% of reading
powder. Leave the bottle undisturbed for 1 Typical EMC 0.01 mg/L
minute, then the reagent is ready. Deviation
Note: after adding the Amino Acid powder, the Light Source Tungsten lamp with narrow band interference
HI 93706B reagent will last for one month. x 10 filter @ 610 nm
Add 10 drops of HI 93706A molybdate re- Method Adaptation of the ASTM D859 method of the
agent. heteropoly blue method. The reaction between
silica and reagents causes a blue tint in the
sample.
Add to the cuvet 0.5 mL
REQUIRED REAGENTS
of the solution prepared
Code Description Quantity
in the HI 93706B bottle. B
HI 93705A-0 Molybdate 6 drops
Replace the cap and swirl
HI 93705B-0 Citric acid 1 packet
the solution.
HI 93705C-0 Amino acid 1 packet

Reinsert the cuvet into the instrument. REAGENT SETS


HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
For other accessories see page 128.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 5 minutes and press READ MEASUREMENT PROCEDURE
DIRECT. In both cases "SIP" will blink during measurement. Select the program number corresponding to Silica on the
secondary LCD by pressing PROGRAM ! and .
Fill the cuvet with 10 mL of unreacted sample 10 mL

(up to the mark).

The instrument directly displays concentration in mg/L of phos- Add 6 drops of HI 93705A mo-
phorus on the Liquid Crystal Display. lybdate reagent. Replace the cap x6
INTERFERENCES and swirl the solution.
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L as Cl Wait for 4 minutes, add the content
Calcium above 10000 mg/L as CaCO3 of one packet of HI 93705B citric
Magnesium above 40000 mg/L as CaCO3 acid reagent and shake until it is
Ferrous iron above 100 mg/L. completely dissolved.

114 115 Silica


Phosphorus
12
1 INTERFERENCES
Wait for 1 minute. This is the blank. 9 3 Interference may be caused by:
6 Phosphate above 60 mg/L (causes a 2% reduction in reading)
Phosphate above 75 mg/L (causes an 11% reduction in reading)
Place the cuvet into the holder and Sulfide and high concentration of iron
ensure that the notch on the cap is Eliminate color and turbidity interferences by zeroing the meter with
positioned securely into the groove. the original water sample.
Press ZERO and SIP will blink on the display.

Wait for a few seconds and the display will


show -0.0-. Now the meter is zeroed
and ready for measurement.

Remove the cuvet and add the con-


tent of one packet of HI 93705C
amino acid reagent and shake until it
has dissolved.

Reinsert the cuvet into the instrument.

Press TIMER and the display will show the countdown prior to the
measurement. Alternatively, wait for exactly 3 minutes and press
READ DIRECT. In both cases SIP will blink during measurement.

The instrument directly displays concentration in mg/L of silica


(SiO2) on the Liquid Crystal Display.

116 117 Silica


Silica
Add 1.0 mL of HI 93737B reagent to the other beaker (the
SILVER sample) and swirl gently to mix. Press TIMER and the display will
SPECIFICATIONS show the countdown prior to adding reagent C or, alternatively,
wait for 2 minutes. Then, in both cases, add 1.0 mL of HI 93737C
Range 0.000 to 1.000 mg/L (ppm)
reagent to each beaker and swirl.
Resolution 0.001 mg/L
Accuracy 0.005 mg/L 10% of reading
Typical EMC 0.001 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the 1-(2-pyridylazo)-2-naphtol PAN
method. The reaction between silver and reagents
causes an orange tint in the sample.
REQUIRED REAGENTS Press TIMER or, al-
Code Description Quantity ternatively, wait
HI 93737A-0 Buffer Reagent A 1 mL for 2 minutes.
HI 93737B-0 Buffer Reagent B 1 mL Then, in both
HI 93737C-0 Indicator Reagent C 2 mL cases, add 1.0 mL
HI 93737D-0 Fixing Reagent D 2 mL of HI 93737D re-
HI 93703-51 Dispersing Agent 4-6 drops agent to each
(only when necessary, see note) beaker and swirl.
REAGENT SETS Then press TIMER or, alterna-
HI 93737-01 Reagents for 50 tests tively, wait for 2 minutes to
HI 93737-03 Reagents for 150 tests allow reaction to complete.
For other accessories see page 128. 10 mL

MEASUREMENT PROCEDURE Fill one cuvet up to the mark with 10 mL of the


blank.
Select the program number corresponding to Silver on the
secondary LCD by pressing PROGRAM ! and .
Place the cuvet into the holder and
Note: for best results perform your tests between 20-24C.
ensure that the notch on the cap is
positioned securely into the groove.
Fill two graduated beakers with 25 mL of sample.

Press ZERO and "SIP" will blink on the display.

Add 1.0 mL of HI 93737A reagent to


one beaker (the blank) and swirl
Wait for a few seconds and the display will
gently to mix.
show "-0.0.-". Now the meter is zeroed

Silver 118 119 Silver


and ready for measurement. 10 mL

ZINC
Fill a second cuvet up to the mark
SPECIFICATIONS
with 10 mL of the reacted sample.
Range 0.00 to 3.00 mg/L
Resolution 0.01 mg/L
Accuracy 0.03 mg/L 3% of reading
Insert the second cuvet into the
Typical EMC 0.01 mg/L
instrument. Deviation
Press READ DIRECT and the display will show "SIP" during Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18 t h
edition, Zincon method. The reaction between zinc
measurement. and the reagent causes a brownish-green to a
blue tint in the sample.
The instrument directly displays concentration in mg/L of silver on
the Liquid Crystal Display. REQUIRED REAGENT
Note: a temperature above 30C may cause turbidity. In this case, Code Description Quantity
before zeroing and taking readings, add 2-3 drops of Dispers- HI 93731A-0 Reagent A 1 packet
ing agent (HI 93703-51) to each cuvet and swirl until HI 93731B-0 Reagent B 0.5 mL
complete dissolution of turbidity. REAGENT SETS
HI 93731-01 Reagents for 100 tests
HI 93731-03 Reagents for 300 tests
INTERFERENCES For other accessories see page 128.
Interference may be caused by:
3+ 2+ MEASUREMENT PROCEDURE
Al above 30 mg/L Fe above 1.5 mg/L
Ca
2+
above 1000 mg/L Fe
3+
above 10 mg/L Select the program number corresponding to Zinc on the second-
as CaCO3 K
+
above 500 mg/L ary LCD by pressing PROGRAM ! and .
2+
Cd above 20 mg/L Mn2+ above 25 mg/L
-
Cl above 8000 mg/L Mg2+ above 1000 mg/L Fill one graduated mixing cylinder up to the
2+
Co above 1.5 mg/L as CaCO3 20 mL mark with the sample.
3+ +
Cr above 20 mg/L Na above 5000 mg/L
6+ 2+
Cr above 40 mg/L Ni above 1.5 mg/L Add the content of one packet of HI 93731A zinc reagent, close
2+ 2+
Cu above 15 mg/L Pb above 20 mg/L the cylinder and invert several times to mix. Take care that all of
- 2+
F above 20 mg/L Zn above 30 mg/L the reagent is dissolved.

Silver 120 121 Zinc


Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds
Fill two cuvets with 10 mL of the and press READ DIRECT. In both cases "SIP" will blink during
10 mL 10 mL
reacted sample up to the mark. measurement.

Replace the cap on one of the cuvets. This is the blank.


BLANK

Place the blank into the cell and ensure The instrument directly displays concentration in mg/L of zinc on
that the notch on the cap is positioned the Liquid Crystal Display.
securely into the groove.
INTERFERENCES
Interference may be caused by:
Press ZERO and SIP will blink on the display.
Aluminum above 6 mg/L
Cadmium above 0.5 mg/L
Copper above 5 mg/L
Iron above 7 mg/L
Manganese above 5 mg/L
Nickel above 5 mg/L
Wait for a few seconds and the display will
show -0.0-. Now the meter is zeroed
and ready for measurement.

Add 0.5 mL of HI 93731B cyclo-


hexanone to the other cuvet. This
will be the sample.
SAMPLE
B
Note: To prevent any contamination
from the polycarbonate cap,
prior to replacing it, close the
sample cuvet with the supplied
HDPE plastic stopper.

Replace the cap and swirl the sample for


SAMPLE

15 seconds.

SAMPLE

Remove the blank and insert the sample


into the instrument.

Zinc 122 123 Zinc


/PUP - Program Up
INTERFACE WITH PC /PDN - Program Down
To connect your meter to the PC use the optional HI 920010 /PTM - Turn Test Mode On
(available from your Hanna Dealer). Make sure that your meter is /Brx - Set the baud rate
switched off and plug the connectors, one into the meter RS 232C 1 - 300 2 - 600
3 - 1200 4 - 2400
socket, the other into the serial port of your PC.
/KBL - Lock Keyboard
Note: Cables other than HI 920010 may use a different configura-
/KBU - Unlock Keyboard
tion, in which case, communication between the meter and
the PC may not be possible. ?PR# - Send Current Program Number
?BRQ - Send current baud rate
SETTING THE BAUD RATE 1 - 150 2 - 300
The transmission speed (baud rate) of the meter and the external 3 - 600 4 - 1200
device must be identical. The meter is factory set to 2400. 5 - 2400
If you wish to change this value, please contact your nearest Hanna ?CNQ - Send Concentration (three bytes)
Center. conc
decimal point
SENDING COMMANDS FROM PC unit
With terminal programs such as, for example, Telix, Windows m - ppm b - ppb
t - ppt u - pcu
Terminal, it is possible to remotely control your Hanna Instruments
h - pH
bench meter. Use HI 920010 cable to connect the meter to the PC,
?ERR - send error / status information
start the terminal program and set the communication options as 0 - No error
follows: 8, N, 1, no flow control. 1 - CAP
Command Types 2 - HI
To send a command to the pH meter the scheme is: 3 - ZERO
4 - LO
<DLE> <command> <CR> 5 - IDLE
This line makes the computer send a Data Link Escape character (/ or 6 - ZERO DONE
?), the command expressed as a number or a 3-character sequence 7 - TIMED READ
and a CR character.
Note: Windows Terminal and all the other terminal programs that DROPS/
DROPS/mm L CROSS REFERENCE
support the ANSI escape sequence, represent the DLE character
by the string ^P and the CR character by the string In application where dropper bottles are supplied, please use the
^M. following chart as a guide to measurement.
Type of Commands mL # of Drops
/OFF - Turn the meter OFF 0.1 2
0.2 4
/PDR - Press Read Direct
0.25 5
/PTR - Press Timer 0.5 10
/PZR - Press Zero Note: # of drops represents an approximation.
Excel Copyright of "Microsoft Co."
Lotus 1-2-3 Copyright of "Lotus Co."
TELIX is registered Trademark of "Deltacomm"
Windows and Windows Terminal are registered Trademark of "Microsoft Co."

124 125
STANDARD METHODS BATTERY REPLACEMENT
Description Range Method
Aluminum 0.00 to 1.00 mg / L Aluminon Further to the 12 VDC supply, these meters can also be fed by one 9V
Ammonia, HR 0.0 to 50.0 mg/L Nessler
Ammonia, MR 0.00 to 10.00 mg / L Nessler battery.
Ammonia, LR 0.00 to 3.00 mg / L Nessler To prolong battery life, switch your meter off after use. However, the
Bromine 0.00 to 8.00 mg / L DPD
Chlorine, Free 0.00 to 2.50 mg / L DPD meter has an auto-shut off feature that will turn itself off after 10
Chlorine, Free HR 0.00 to 10.00 mg / L DPD minutes of non-use.
Chlorine, Total 0.00 to 3.50 mg / L DPD
Chlorine, Total HR 0.00 to 10.00 mg / L DPD A blinking "LOBAT" on the display, while a
Chlorine Dioxide 0.00 to 2.00 mg / L Chlorophenol Red measurement is in progress, indicates low
Chromium VI, HR 0 to 1000 g / L Diphenylcarbohydrazide
Chromium VI, LR 0 to 300 g / L Diphenylcarbohydrazide
voltage and the battery should be replaced.
Color of Water 0 to 500 PCU Colorimetric Platinum Cobalt If the battery is not replaced immediately, in
Copper, HR 0.00 to 5.00 mg / L Bicinchoninate
Copper, LR 0 to 1000 g / L Bicinchoninate
order to prevent erroneous readings due to
Cyanide 0.000 to 0.200 mg / L Pyridine-Pyrazalone low voltage, "-BA-" is displayed soon after-
Cyanuric Acid 0 to 80 mg / L Turbidimetric wards. At this point the battery must be
Fluoride 0.00 to 2.00 mg / L SPADNS
Hardness, Ca 0.00 to 2.70 mg / L Calmagite changed.
Hardness, Mg 0.00 to 2.00 mg / L Calmagite Battery replacement must only take place in a non-hazardous area
Hardness, Total HR 400 to 750 mg/L Calmagite
Hardness, Total MR 200 to 500 mg/L Calmagite using a 9V alkaline battery.
Hardness, Total LR 0 to 250 mg/L Calmagite
Hydrazine 0 to 400 g / L p-Dimethylaminobenzaldehyde
Simply slide off the battery cover on the
Iodine 0.0 to 12.5 mg / L DPD back of the meter. Detach the battery from
Iron, HR 0.00 to 5.00 mg / L Phenantroline the terminals and attach a fresh 9V battery
Iron, LR 0 to 400 g / L TPTZ
Manganese, HR 0.0 to 20.0 mg / L Periodate Oxidation while paying attention to the correct polar-
Manganese, LR 0 to 300 g / L PAN ity. Replace the battery and the cover.
Molybdenum 0.0 to 40.0 mg / L Mercaptoacetic Acid
Nickel, HR 0.00 to 7.00 g / L Photometric
Nickel, LR 0.000 to 1.000 mg / L PAN
Nitrate 0.0 to 30.0 mg / L Cadmium Reduction
Nitrite, HR 0 to 150 mg / L Ferrous Sulfate The meter will turn on automatically when a new
Nitrite, LR 0.00 to 0.35 mg / L Diazotization battery is connected. You can turn it off by
Oxygen, Dissolved 0.0 to 10.0 mg / L Winkler
pH 5.9 to 8.5 pH Phenol Red pressing ON/OFF.
Phosphate, HR 0.0 to 30.0 mg / L Amino Acid
Phosphate, LR 0.00 to 2.50 mg / L Ascorbic Acid
Phosphorus 0.0 to 15.0 mg / L Amino Acid
Silica 0.00 to 2.00 mg / L Heteropoly Blue
Silver 0.000 to 1.000 mg / L PAN
Zinc 0.00 to 3.00 mg / L Zincon

126 127
HI 93723-01 100 chromium VI HR tests
ACCESSORIES HI 93723-03 300 chromium VI HR tests
REAGENT SETS HI 93726-01 100 nickel HR tests
HI 93700-01 100 ammonia LR tests HI 93726-03 300 nickel HR tests
HI 93700-03 300 ammonia LR tests HI 93728-01 100 nitrate tests
HI 93701-01 100 free chlorine tests (powder) HI 93728-03 300 nitrate tests
HI 93701-03 300 free chlorine tests (powder) HI 93729-01 100 fluoride tests
HI 93701-F 300 free chlorine tests (liquid) HI 93729-03 300 fluoride tests
HI 93701-T 300 total chlorine tests (liquid) HI 93730-01 100 molybdenum tests
HI 93702-01 100 copper HR tests HI 93730-03 300 molybdenum tests
HI 93702-03 300 copper HR tests HI 93731-01 100 zinc tests
HI 93704-01 100 hydrazine tests HI 93731-03 300 zinc tests
HI 93704-03 300 hydrazine tests HI 93732-01 100 dissolved oxygen tests
HI 93705-01 100 silica tests HI 93732-03 300 dissolved oxygen tests
HI 93705-03 300 silica tests HI 93733-01 100 ammonia HR tests
HI 93706-01 100 phosphorus tests HI 93733-03 300 ammonia HR tests
HI 93706-03 300 phosphorus tests HI 93734-01 100 free & total chlorine HR tests
HI 93707-01 100 nitrite LR tests HI 93734-03 300 free & total chlorine HR tests
HI 93707-03 300 nitrite LR tests HI 93735-LR 100 total hardness LR tests
HI 93708-01 100 nitrite HR tests HI 93735-MR 100 total hardness MR tests
HI 93708-03 300 nitrite HR tests HI 93735-HR 100 total hardness HR tests
HI 93709-01 100 manganese HR tests HI 93735-0 300 total hardness tests
HI 93709-03 300 manganese HR tests HI 93737-01 50 silver tests
HI 93710-01 100 pH tests HI 93737-03 150 silver tests
HI 93710-03 300 pH tests HI 93738-01 100 chlorine dioxide tests
HI 93711-01 100 total chlorine tests (powder) HI 93738-03 300 chlorine dioxide tests
HI 93711-03 300 total chlorine tests (powder) HI 93740-01 50 nickel LR tests
HI 93712-01 100 aluminum tests HI 93740-03 150 nickel LR tests
HI 93712-03 300 aluminum tests HI 93746-01 50 iron LR tests
HI 93713-01 100 phosphate LR tests HI 93746-03 150 iron LR tests
HI 93713-03 300 phosphate LR tests HI 93747-01 100 copper LR tests
HI 93714-01 100 cyanide tests HI 93747-03 300 copper LR tests
HI 93714-03 300 cyanide tests HI 93748-01 50 manganese LR tests
HI 93715-01 100 ammonia MR tests HI 93748-03 150 manganese LR tests
HI 93715-03 300 ammonia MR tests HI 93749-01 100 chromium VI LR tests
HI 93716-01 100 bromine tests HI 93749-03 300 chromium VI LR tests
HI 93716-03 300 bromine tests OTHER ACCESSORIES
HI 93717-01 100 phosphate HR tests C115-00300 5 mL graduated syringe
HI 93717-03 300 phosphate HR tests HI 721310 9V battery (10 pcs)
HI 93718-01 100 iodine tests HI 731318 Tissue for wiping cuvets (4 pcs)
HI 93718-03 300 iodine tests HI 731321 Glass cuvets (4 pcs)
HI 93719-01 100 Mg hardness tests HI 740034 Cap for 100 mL beaker (6 pcs)
HI 93719-03 300 Mg hardness tests HI 740036 100 mL plastic beaker (6 pcs)
HI 93720-01 100 Ca hardness tests HI 740038 60 mL glass BOD bottle and stopper
HI 93720-03 300 Ca hardness tests HI 740142 1 mL graduated syringe
HI 93721-01 100 iron HR tests HI 740143 1 mL graduated syringe (6 pcs)
HI 93721-03 300 iron HR tests HI 740144 Pipette tip (6 pcs)
HI 93722-01 100 cyanuric acid tests HI 740157 Plastic refilling pipette (20 pcs)
HI 93722-03 300 cyanuric acid tests HI 93703-50 Cuvets cleaning solution (230 mL)

128 129
WARRANTY CE DECLARATION OF CONFORMITY
All Hanna Instruments meters are warranted for two years
against defects in workmanship and materials when used for their
intended purpose and maintained according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accidents, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure. If the repair
is not covered by the warranty, you will be notified of the charges
incurred.
If the instrument is to be returned to Hanna Instruments, first obtain
a Returned Goods Authorization Number from the Customer Service
department and then send it with shipment costs prepaid. When
shipping any instrument, make sure it is properly packaged for
complete protection.
To validate your warranty, fill out and return the enclosed warranty
card within 14 days from the date of purchase.
Hanna Instruments reserves the right to modify the design, construc-
tion and appearance of its products without advance notice. Recommendations for Users
Before using these products, make sure that they are entirely suitable for the environment in which
they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and
TV equipments, requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC
performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.

130 131
HANNA LITERATURE
Hanna publishes a wide range of catalogs and handbooks for
an equally wide range of applications. The reference literature
currently covers areas such as:

Water Treatment
Process
Swimming Pools
Agriculture
Food
Laboratory
Thermometry

and many others. New reference material is constantly being


added to the library.

PRINTED IN ITALY
For these and other catalogs, handbooks and leaflets contact
your dealer or the Hanna Customer Service Center nearest to
you. To find the Hanna Office in your vicinity, check our home
page at www.hannainst.com.

MANC200
05/00

W W W . H A N N A I N S T . C O M

S-ar putea să vă placă și