Sunteți pe pagina 1din 1

Glucose Iiquicolor Calculation of the Glucose Concentration

A sample
GOD-PAP Method c = 100 x [mg/dl] or
A [STD]
Enzymatic Colorimetric Test for Glucose
Method without Deproteinisation c = 5.55 x
A sample
[mmol/l]
Package Sizes A [STD]
[REF] 10260 4 x 100 ml Enzyme Reagent
Performance Characteristics
10121 1000 ml Enzyme Reagent
Linearity
10123 9 x 3 ml Standard
The test is Iinear up to a glucose concentration of 400 mg/dl or
[IVD] 22.2 mmol/I. Dilute the sample 1+2 with dest. water, if the glucose
concentration of the sample is over this limit and repeat the
Method 1 determination. Multiply the result by 3.
The glucose is determined after enzymatic oxidation in the presence of
Typical performance data can be found in the Verification Report,
glucose oxidase. The formed hydrogen peroxide reacts under catalysis of
accessible via:
peroxidase with phenol and 4-aminophenazone to a red-violet
quinoneimine dye as indicator. www.human.de/data/gb/vr/su-gllq.pdf or
www.human-de.com/data/gb/vr/su-gllq.pdf
Reaction Principle
GOD Normal Values 2
Glucose + O2 + H2O gluconic acid + H2O2 Serum, plasma (fasting): 75 - 115 mg/dl or 4.2-6.4 mmol/l
POD
2 H2O2 + 4-aminophenazone + phenol quinoneimine + 4 H2O Quality Control
All control sera with glucose values determined by this method can be
Contents used.
[REF] 10260 10121 10123 We recommend to use our quality control sera HUMATROL based on
[RGT] 4 x 100 ml 1 x 1000 ml animal serum or our SERODOS based on human serum.
[STD] 1 x 3 ml 1 x 3 ml 9 x 3 ml Automation
[RGT] Enzyme reagent Proposals to apply the reagents on analysers are available on request.
Phosphate buffer (pH 7.5) 0.1 mol/l Each laboratory has to validate the application in its own responsibility.
4-Aminophenazone 0.25 mmol/I
Phenol 0.75 mmol/l Notes
Glucose oxidase 15 KU/l Icteric sera interfere with the test and should not be used as a sample
Peroxidase 1.5 KU/l Triglycerides up to 2500 mg/dl, hemoglobin up to 500 mg/dl and ascorbic
Mutarotase 2.0 KU/l acid up to 20 mg/dl do not interfere with the test.
Stabilisers
References
[STD] Standard
Glucose 100 mg/dl or 5.55 mmol/l 1. Barham D., Trinder P., Analyst 97 (1972)
2. Teuscher A., Richterich P., Schweiz. med. Wschr. 101, 345 and 390
Reagent Preparation (1971)
[RGT] and [STD] are ready for use.

Reagent Stability SU-GLLQ2 INF 1026002 GB 07-2008-19 |


The reagents are stable even after opening up to the given expiry date
when stored at 2...8C. Contamination must be avoided.
At 15...25C [RGT] is stable for 2 weeks.

Specimen
Serum, plasma.
The glucose is stable for 24 hours at 2...8C, if serum or plasma is
prepared within 30 min. after collection.

Assay
Wavelength: 500 nm, Hg 546 nm
Optical path: 1 cm
Temperature: 20...25C or 37C
Measurement: Against reagent blank.
Only one reagent blank per series is required.

Pipetting Scheme
Macro Semi micro
Pipette into cu- [STD] or Reagent [STD] or Reagent
vettes sample blank sample blank
[STD]
or sample 20 l --- 10 l ---
[RGT] 2000 l 2000 l 1000 l 1000 l
Mix, incubate for 10 min. at 20...25C or 5 min. at 37C. Measure the
absorbance of [STD] and the sample against the reagent blank within
60 min. ( A).

Human Gesellschaft fr Biochemica und Diagnostica mbH


Max-Planck-Ring 21 65205 Wiesbaden Germany
Telefon +49 6122-9988-0 Telefax +49 6122-9988-100 e-Mail human@human.de