Troubleshooting
Technical Publication
Ver.3.2
Troubleshooting
EVA-HF325/525 generator
0
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
ADVISORY SYMBOLS
1
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
CONTENTS
1. INTRODUCTION ------------------------------------3
6. WAVEFORM DATA--------------------------------- 13
7. JOB CODE---------------------------------------- 16
8. X-RAY CHECK------------------------------------ 40
2
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
1. INTRODUCTION
3
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
2. ERROR CODES
System error codes are displayed on the kVp display LCD(condition display LCD)
Each code is preceded by the letter E(i.e. E16). Error codes are described in below
tables.
4
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
5
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
3. FLOW CHART
Power ON
Text 1
LED 9 ON Check 1
LED10,11,12 ON
Text 2
Text 3
LED 1 OR 2 ON
LED 3 OR 4 OR 5 ON Check 3
Text 4
Contactor 1
Contactor 2 Check 4
Text 5
LED 4 ON
Initialize OK
6
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
TEXT 1
Led 9 ON is meaning of system power input OK.
Led 10,11,12 ON is meaning of power supply +5V,+12V,-12V OK.
That Leds located on Relay power board.
TEXT 2
Led 7 ON is meaning of charge of capacitor ON.
Led 8 ON is meaning of
Filament power input(AC60) ON.
Bucky power input(AC110) ON.
Lock power input(DC30) ON.
Collimeter power input(AC20) ON.
And relay power(DC24V) in relay power board
That Leds located on Relay power board.
TEXT 3
Led 1 ON is meaning of selected large focus filament.
Led 2 ON is meaning of selected small focus filament.
Leds ON will select Led 1 or Led 2.
Led 3 ON is meaning of selected bucky 1.
Led 4 ON is meaning of selected bucky 2.
Led 5 ON is meaning of selected bucky 3.
Led 3,4,5 all OFF is meaning of selected NO bucky and That leds ON will selected
Led 3 or Led 4 or Led 5.
That Leds located on Relay Power Board.
TEXT 4
Contactor 1, 2 is meaning of Main contactor and charge contactor ON.
TEXT 5
That meaning is System initializing OK.
Led 4 located on HT contol cpu board.
7
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
4. ERROR CHECK
CHECK 1
- Line input power check(see JOB CODE 001)
- Communication cable check(see JOB CODE 002)
- Membrane key check(see JOB CODE 003)
- Bridge diode in Relay power board(see JOB CODE 004)
- AC12V input check(see JOB CODE 005)
CHECK 2
- Power(+5V,+12V,-12V) in Relay power board(see JOB CODE 006)
- Power ON signal check(see JOB CODE 007)
CHECK 3
- Check ERROR CODE TABLE
CHECK 4
- Check charge/discharge monitor board(see JOB CODE 008)
- Check charge contactor(see JOB CODE 009)
- Check charge signal in HT control cpu board(see JOB CODE 010)
8
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
5. DATA TABLE
The Calibration data and some technique data stored in the U7-UV_PROM
And APR data stored in the U23-EEPROM of the Console display Board.
5.1 CALIBRATION DATA TABLE IN UV_PROM
9
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
10
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
6. WAVEFORM DATA
11
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
5. Ready preheating (at low mA data) 6. KV data output (at low kV data)
12
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
13
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
13. TL594 output (at low kV data) 14. TL594 output (at high kV data)
14
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
15
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
7. JOB CODE
Check 1. input power of line power cable L1 and L2 voltage is 215 ~230VAC
Check 2. connection of AC connection wire are tight to each terminals
16
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
J1 J2 J2 J1
Pic1.view in mother board Pic2. wiew in console
J1 J2
+5V(HC) +KV
GND(HC) -KV
TXD DC24V-2(HC)
RXD TUBE2 READY
HX-RAY TUBE2X-RAY
HREADY FLUORO MODE
HINHIBIT POWER OFF
HTRANSFER POWER ON
CREADY POWER COM
CX-RAY
CTRANSFER
CINHIBIT C+5V
+12V(CH) C+12V
+12V(HC) C GND
17
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
18
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
SW49
SW50
misprint silkscreen
membrane key
19
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
R40
1.2K D16
RY9
1N4001 SW
LD9
J1-21,43(AC220V)
LED J1-19,41(AC220V)
J1-20,42(AC220V)
- + R34 R35 Q5
1K 1K 2N2222A
D9
DF06 + C12 R37 Q6
2N2222A R36
1K
220uF/35V 4.7K
R38
4.7K
to console control
J1-1(S/W ON))
J1-2,24(S/W COM)
J1-23(S/W OFF)
J1-22(AC12V)
J1-44(AC12V)
LED9 ON
SWITCH ON/COM/OFF LINE IS CONNECTED WITH COMMUNICATION AND CONSOLE.
AC20V
J5(11P)
J5
DC24V-2
GND
DC24V-1
AC220V(N)
AC220V(N)
J9(6P) IPM AC220V
IPM AC220V
J9 AC12V
AC220V AC12V
AC220V
(To mother board) AC220V
AC12V AC220V
(From
AC12V
sub-transformer)
AC20V
AC20V
21
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
LED10 ON : +5V
LED11 ON : +12V
LED12 ON : -12V
R10
R33 R25
22
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
Relay 3 and D5
20A
RY3
D5
+24V POWER ON
1N4007
23
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
J2-13,35(DC24V-1)
J1-4(DC24V-2)
R39
2.2K J1-12,34(DC24V-2 OUTPUT)
to 1pin of J11and 2pin of J2 in
D17 mother board and
LD8 RY8
LED 1N4001
J2-26(POWER ON)
To Relay Power
4. Check input power DC24V-2, DC24V-1(All power is from Main power board)
24
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
P2(charge sense)
P3(discharge signal)
R3- R9(33K) this resistor endure at 230VAC
Must check input power voltage(210 230VAC)
D1 Q1
IN4764 IRFP450
LED1
D3
D2 1N4733
IN4764 U1 P3
1 6 R4
1
3.3M 2 To 3,4pin of J10 in
DIS CHARGE Mother Board
2 5
P2
1 To 5,6pin of J10 in
3 4 2 Mother Board
3
CHARGE SEN
TLP531
AC 380/3Phase -> 9K
25
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
main
charge
precharge resistor
U19 R39-46 10
1 18
I1 Q1 J2-27(MAIN MAG)
2 17
I2 Q2 J2-5(POWER ON)
3 16
I3 Q3 J2-28(PRECHARGE)
4 15
I4 Q4 J2-6(CHARGE MAG)
5 14
I5 Q5 J2-7(TUBE SEL)
6 13
I6 Q6 J2-29(220/60)
7 12
I7 Q7 J2-30(HIGH SPEED RTON)
8 11
I8 Q8 J2-8(ROTOR ON)
9 10
GND COM 1K
ULN2803
26
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
If power not charge in storge capacitor will displayed E28 in console display.
27
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
If need change Calibration data with new data saved in Serial-ROM. You will read Serial-ROM
with Rom_writer and re_save data in same address. if you save abnormal data, it will be
displayed ERROR message in Console Display LCD.
If APR data brocken by noise. You can re_save with pushing set switch(save switch) in APR
MODE.
Pic. Serial-ROM and EEPROM location(Console display board(operating_panel_console board)
U7(M27C256B)
U25(25256)
28
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
KVp reference
data at 200mA
mA reference
data at 25mA
29
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
30
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
J10
SSR
CT
J7(To Interface Terminal)
SP(spark killer)
J10
1
2
CT1
0VAC
DC24V-1 ROTOR COM/MAIN/PHASE
SP1
SSR1
220VAC
SSR D4
1N4007
J7-1
J7-2
J7-3
J7-4
J7-5
C4
24uF
To ROTOR of TUBE
TEMP
G(PHASE)
B(MAIN)
31
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
W(COM)
32
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
U19 R39-46 10
1 18
I1 Q1 J2-27(MAIN MAG)
2 17
I2 Q2 J2-5(POWER ON)
3 16
I3 Q3 J2-28(PRECHARGE)
4 15
I4 Q4 J2-6(CHARGE MAG)
5 14
I5 Q5 J2-7(TUBE SEL)
6 13
I6 Q6 J2-29(220/60)
7 12
I7 Q7 J2-30(HIGH SPEED RTON)
8 11
I8 Q8 J2-8(ROTOR ON)
9 10
GND COM 1K
ULN2803
If did not output sensing signal of ROTOR sense, it will display E26/27 in Console
display LCD.
33
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
+12V C49
TP11
104
0.6V
8
10k
C48
104 +12V
R64
4.7K
8
U12B D25
5
+
VR9 7 5 12
6 U6
- TL082A J1_34(FILHIGH)
ULN2004
10k 1N4148
4
TP10
C50
3.3V -12V
104
If you want LOW limit change, use with VR8. and HIGH limit is VR9.
If filament sensing value is lower than TP11, will display E29 in Console Display
LCD.
And higher than TP10, will display E30 in Console Display LCD.
34
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
Temp(from tube)
If TUBE temperature signal is open, will display E32 in Console display LCD.
J7 J7(10P)
BUCKY1
BUCKY1
BUCKY2
BUCKY2
BUCKY3
BUCKY3
TUBE1 TEMP
TUBE1 TEMP
TUBE2 TEMP
35
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
TUBE2 TEMP
From HV tank
36
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
HCPL4503
VPC(GND1)
J1-4
PC1
8 2 VP1(+15VOUTPUT1)
J1-1
C5
6 0.1
3 FPO(-FAULT1)
J1-5
5
HCPL4503
Table. Check Point(All power was setted originally in factory. Dont need contrast)
Checking point Value Accuracy limit
J1 1,4 VDC 14.9 ~ 15.1VDC
J1 2,4 VDC 4.95 ~ 5.05VDC
J2 1,4 VDC 14.9 ~ 15.1VDC
J2 2,4 VDC 4.95 ~ 5.05VDC
Pic.IPM and IPM drive board
IPM
IPM drive board
J2
J1
J3
+5V
GND
KDR1
J1(J2) KDR2
FAULT FAULT
GND AC220V
-KV DR AC220V
+5V
+15V
37
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
+mA
+kV
GND
When measure +kV wih OSC, real kV is 10kV per 1V.
Also, real mA is 20mA per 1V.
Case 1
38
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
U27
U6 13 18
D0 VDD
6
12 17
1 16 5 12 11 D1 RS
I1 O1 CT VCC 11 D2
2 15
RT
C1 10
I2 O2 8 9 D3
3 14 16 C2 9 D4
4 I3 O3 13 +IN1 E1 8 3
15 10 7 D5 VREF 14
5 I4 O4 12 3 -IN1 E2 14 D6 CS
I5 O5 FB REF 6 16
6 11 1 13 D7 LDAC 4
I6 O6 2 +IN2 CTL 4 AGND
7 10 -IN2 DTC DGND
5
8 I7 O7 9
7
GND 1
GND COM 15 VSS 2
WR VOUT
U3
ULN2004 5. TL594
HV AD7224
transformer bad
change HV transformer.
Case 2
2. Aanother factor.
1. If IPM driver board broken, always display E00(power on state ,not exposure state)
change IPM driver board
check IPM drive input power(220VAC) and signal connection.
2. If IPM was damaged, sound tingand display E00 at high kVp exposing
change IPM
39
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
8. X-RAY CHECK
1. FILAMENT DRIVE BOARD CHECK
Tools & Equipment required : DVM,OSC
Pic. Filament drive board
filament enable signal (high low)
DC0V - 100V
+12V
VIN
GND
V+
1K 1K
U2 Q10 473 C16 C19
C26 ZD10
LM 3524
2
16
R43
10K 10uF
R48 1K
D22 D18
20K
R44 1N4148
1N4148
D21 1K C33
C35 1N4148 D20 102
+12V
-12V
0.1 1N4148 J 1-4,5,26,27(FIL OUTA)
C37
+
8
U5B U5A
6 2
- - J 1-6,7,28,29(FIL OUTB)
100uF/35V 7 1
5 3
+ +
TL082 TL082
J2-19,20,41,42(-12V) C36
4
-12V
0.1
40
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
Q1 R1
BUX48A 0.5/3W
J2-15,37(EARTH)
R11 + C1 + C2
Q4 R4 R10
47K/3W DC0 - 100V
470uF/200V 470uF/200V BUX48A 0.5/3W 0.2/5W
+12V
C7
Q5 R5 R12 R14
TIP42 1K/3W 104 12K 12K
U6B
8
D1 LM358 6 R16
R6 -
R24 7
47K
5.2K 5
2.2K/1/2W +
R13 R15
D3 1N4148 R17
Q6 390 820
2N3440 47K
4
+12V C4
1N4148
C5
J2-21,22,43,44(GND)
0.1 0.1
R7 R8 U6A
+ C17 2.4K 2.7K 8
J2-17,18,39,40(+12V) D2 LM358 R20
2
1 -
100uF/35V 47K
3
+
1N4148 R21
47K
4
C3
R54 0.1
10K
J2-31(FOUT) C8
+12V
8
R58
-12V
8
10K R56 6
104 -
7
100K
5 U4B
+
R26 C6
TL082 R25
4.7K 0.1
C10 10K
4
+12V
8
104
6
8
-
7 2
J1-13,24,35(FSEN) -
U1B 5 1
+
U1A 3
TL082 + FSENSE
C11 TL082
4
-12V
4
104
VR1
41
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
All mA step condition will change uniformly with VR1 contrasting FOUT
signal.
And, If you want each mA step condition change, use Calibration mode
42
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
43
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
R38 R37
R9
4.7K 4.7K U4B
10K
U6 15 14
7 10 16 D S
IN
U1A J2_26(R/F)
TL082A R8 ADG201A
KV
8
ZD3,ZD4 1N4744 1N4746 2001 8 29 2 ULN2004 U4A
C8 - 1M
1 2 3
R19 J1_1(KSEN) D S
3 1
100P + IN
J1_27(-MASEN) R15 1M
J1_5(+MASEN) C4 ADG201A
154K
4
C21 R16 C9 C10 103
-KV
+12V -M A
104 10K 273 100P
8
R10 R12 6
8
- U2B R17 R18 R1 R3 U1B
7 6
10K 1M -
5 7
+ 34K 100K 1K 100K
C5 ZD3 D5 D6 5
TL082A +
C1 ZD1 D1 D2
TL082A R6
102 1N4744 1N4148 1N4148 C22
4
4
T2
J1_4(+KVSEN) +12V -12V 104
100uH T1
U4A +12V -12V
C7 VR2 100uH
C6 ZD4 D7 D8 3 2
S D 1 C2 ZD2 D3 D4 C3
100P 20K IN
102 1N4744 1N4148 1N4148 R7
103
R11 R13 R14 ADG201A 104 1N4752 1N4148 1N4148 10K
U4B R2 R4 R5 J1_23(M SEN)
10K 1M 154K VR1
14 15
S D 16 1K 100K 10K
+KV 10K IN
+MA
ADG201A R28 +12V
U5B
1K
14 15 TP2
S D 16 C23
IN C15
6
104 U6
C18 5 12 TP3 TP4
ADG201A CT VCC R33-R36
11 1 16
RT
C1 I1 O1 J1_7(KDR1)
8 2 15
8
10K 4.99K
ADG201A 104
U5
U5D ADG201A
7 6 4 C27
8 D S
IN 104
ADG201A
1 kV output data voltage from 2 of U27 in Ht control cpu board 1(when exposured)
kV low (40kV see waveform 6)
kV high(70kV see waveform 7)
2 exposure/kV enable signal
high low (see waveform 8)
3 TP 2 is control waveform(see waveform 9)
44
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
45
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
J11(5P) J10(6P)
46
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
CONSOLE
DISPLAY AND
MEMBRANE
KEY
CONTROLLER
CONSOLE CPU
BOARD COMMUNICATION CABLE HT CONTROL
CPU BOARD
KOUT
Feedback
HT CONTROL IPM
CONTROL BOARD driver IPM
e
FILAMENT DRIVEFOUT
+ TUBE BOARD
HV cable
RELAY POWER
BOARD
HV transformer
+ MAIN POWER
- terminal BOARD
Filament power
T1 and T2
ROTOR power
47
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
kV mA Time Bucky
Console display CPU give kV data,mA data,time data,bucky select data to HT control CPU.
And HT control CPU re-give same data to Console display CPU.
So, Console display CPU check data error or good. If error detected, display EROOR code.
Ex) KV data error(E16,E17), mA data error(E18,E19), time data(E20)
TP1(Txd), TP2(Rxd)
48
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
TP3(Txd)
TP4(Rxd)
mA , time, bucky select signal not transfer. Only kV data data transfer
cf . kV data 8bit = 2 8 =
P P
256 information
49
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
kV data(138+1 = 139)
KV-SET data 139 in Console display cpu board give to HT control cpu board
And, HT control cpu give data to AD7224
When communicate HT control cpu board with console display cpu
board, sometimes error detected by electrical noise.
1. Check ground state.
2. Check shield state of communication cable.
3. Check noise of Power line.
]
50
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
51
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
52
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
AD7224
If you want to see waveform TP5,TP7(when push x-ray s/w) see Chap.6 6 and7 waveform.
BECAREFUL WHEN YOU CHECKING FOUT SIGNAL. IF ELECTRICAL
SHORT HAPPEN, U25(AD7224) IN HT CONTROL CPU BOARD WILL BE
BROCKEN.
1 2 3 4
1. KOUT signal 2. Feedback signal
53
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
TL594(494) receive two signal (KOUT(-ini) and feedback signal(+ini))and give output
signal(E1,E2) to IPM drive.
If KOUT(analog data) is much, width of TL594 output signal would be widen.
54
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
55
COMED Medical Systems Co., Ltd.
EVA-HF325/525
Troubleshooting
56
COMED Medical Systems Co., Ltd.