Sunteți pe pagina 1din 2

Pre-deployment Setup For a direct connection between a TitanSMA and a

computer
5) Review and change the configuration As you tilt the device, watch the Mean values of the
three channels on the Waveform page. Use the XYZ
This section will guide you through the first-time To configure the TitanSMA, log in using the admin user
Use the link-local IP address of 169.254.30.30 with marker on the top of the device for orientation and
setup of your TitanSMA. Use this section to help you account, open the Configuration menu, and select
a subnet mask of 255.255.255.0 review the table below to understand the approximate
familiarize yourself with the device and its Web Configuration.
values that you should see in the various orientations.
interface, and to configure it for field deployment. For a network connection Default admin user name admin
Connect the TitanSMA to a DHCP-enabled network Default admin password admin
Ready to install your TitanSMA? and allow it to automatically assign an IP address
Configuration menu
These are not field deployment instructions. Refer to the TitanSMA to the device
Installation Quick Reference or click Help in the TitanSMA Web
interface for installation instructions. Use Disco, a Nanometrics command-line computer Review the configuration prior to deployment to
application, to search the LAN for Nanometrics ensure that the device is properly configured for your
instruments and applications application. The table below outlines some features
1) What you need you may want to take advantage of in your deployment.
Note the IP address of the TitanSMA
You will need the following items to set up your The Help will provide instructions and guidance.
TitanSMA in a lab or office environment: Disco must be run on the same subnet as the devices you
want to find. Action Configuration
TitanSMA
Disco is part of Apollo Toolkit. Visit the Nanometrics Record in the Accelerometer: the default accelerometer
Power supply and cable Support Site at http://support.nanometrics.ca to desired mode mode is 4 g
download Apollo Toolkit and its user guide.
Ethernet cable Change the Digitizer: the default sample rate is 200 Hz
sample rate (200 samples per second)
GPS antenna and cable
Automatically Digitizer: enable triggers for channels that
Computer detect events should detect events
Orientation X (g) Y (g) Z (g)
Events: tailor settings to capture events
X/Y arrows horizontal (upright, on its feet) 0 0 0
according to your specifications
2) Connect the cables X arrow up (device on left side) +1 0 -1
Trigger Sharing: enable trigger sharing if
Connect the cables in the following order: event detection involves sending votes to and X arrow down (device on right side) -1 0 -1
receiving votes from other devices Y arrow down (device on media door) 0 -1 -1
1. GPS antenna cable
Archive events to Events Data Archive: enable data archiving of
2. Ethernet cable removable media events
3. Power cable (ensure that the power cable is Hardware: insert SD media card in media bay
already connected to the power source) Stream data to Data Streaming: create and configure a
Apollo Server streamer of the desired type

6) Perform a tilt test


4) Insert a media card A tilt test involves rotating the TitanSMA while watching
the acceleration in real-time on the Waveform page of
If you plan to archive events or record continuous data
the Web interface.
on a removable SD media card, you must insert the
card into the SD card slot behind the media bay door To avoid clipping, the device should be in 2 g or 4 g mode
(see image below). When the card is properly inserted when performing a tilt test.
and ready for use, the Media eject LED will turn solid If you want to perform more detailed testing, you can
red for 10 minutes and then turn off. If you want to perform a levelled and aligned freestanding installation
using the supplied TitanSMA fixed foot. See the TitanSMA
remove the card, push the Media eject button and
Installation Quick Reference or the Help for instructions.
wait for the Media eject LED to become solid green,
indicating that it is safe to remove the card.

3) Open the Web interface Understanding the external LED states (unless otherwise stated, the LED default state is blinking)
USB port
In a standard Web browser, go to the IP address of
Power Green Power OK Media Off Data archiving disabled (default)
the TitanSMA (http://TitanSMA IP) to open its Web USB eject
button
Solid red Processor boot up or shut down Green SD card present and recording
interface.
Solid yellow Application boot up or shut down Yellow Checking SD card status
Use the Web interface for
Blinking red Force power on; in override mode Red Card missing, full, corrupt, or not mountable
USB eject LED
State-of-health and waveform monitoring Media card slot Off No power Event Off No SD card (default) / No recorded events
(SD card)
Listing and downloading events Link Green Ethernet link established Yellow Checking SD card for events
(Ethernet)
Calibration, configuration, and maintenance Yellow Data Rx/Tx activity Green At least one event recorded to SD card
Off No Ethernet link Status Green Device operating within expected parameters
Force Power On button
Check the Help! Time Green GPS lock, time synch, and position OK Yellow Device booting
Media eject button (GPS)
Click Help in the Web interface to find instructions on installing the (SD card) Media
Yellow GPS initializing and attempting lock Red Device error: check Health page
device, using the Web interface, and reference material such as bay
Red No GPS lock; time uncertainty >100 s
Media eject LED
specifications, transfer function, and noise floor. (SD card)
Connector Pinouts Technical Support Factory Configuration
Power If you need technical support, please submit a request Each TitanSMA ships from the factory with the following
on the Nanometrics technical support site or by email configuration settings:
Type: or fax. Include a full explanation of the problem and
Digitization
3-pin, shell size 8 related information such as log files.
Sample rate 200 sps
MIL-C_26482G Series 1 Email: techsupport@nanometrics.ca
C A Channel subsampling Disabled
Recommended mate: Support site: http://support.nanometrics.ca
PWR_RTN PWR
Offset removal 0g
MS3116J8-3S

Contact Nanometrics Accelerometer


B

Getting Started
CHGND
Mode (peak acceleration) 4g
Nanometrics Inc.
Power (advanced)
250 Herzberg Road
Pin Name Function Low voltage disconnect 8.5 V
Kanata, Ontario, Canada K2K 2A1
A PWR Raw (battery) power in 9 V to 36 V DC Low voltage shutdown 9V
B CHGND Internal connection to chassis ground Voltage reconnect 10 V
Tel +1 613-592-6776
C PWR_RTN Raw power return
Fax +1 613-592-5929 Communications
Ethernet communications DHCP
Ethernet Email: info@nanometrics.ca Link-local (169.254.30.30)
Type: Web: www.nanometrics.ca
4-socket, shell size 8
A
MIL-C_26482G Series 1 Power On and Power Off
D
TX+

B
Insert position W About Nanometrics The media bay (see image in step 4 of the Pre-
RX+ TX- 10/100 Base-T Nanometrics Inc. is a world-class provider of precision deployment Setup section of this guide) provides
C Recommended mate: instrumentation, network technology and software buttons to power off or to force power on. There is also
RX- MS3116J8-4PW applications for seismological and environmental the option to shut down via the Web interface.
research. With state-of-the-art manufacturing facilities
in Ottawa, Ontario; Nanometrics is an award winning Typical power on
Pin Name Function Canadian exporter located in the heart of Canadas The TitanSMA powers on when power is applied and
A TX+ Ethernet MTL-3 Transmit + technology sector. the supply voltage is at or above the voltage reconnect
B TX- Ethernet MTL-3 Transmit - threshold (see Factory Configuration, above). Based on
C RX- Ethernet MTL-3 Receive - these factory settings, the device will require at least
D RX+ Ethernet MTL-3 Receive +
10 V the first time it is powered on, and each time
thereafter, unless the power settings are changed.
About This Guide
GPS This guide provides a high-level overview of the Force power on
Type: TitanSMA. Visit the Help for detailed information and If the supply voltage is at least 9 V but below
TNC female instructions. Open the Help from the TitanSMA Web the voltage reconnect threshold, the device can
A
GPS_ANT
Recommended mate: interface. Self-contained accelerometer be forced to power on by pressing and holding
TNC male The information contained in this guide has been and data recorder the Force Power On button in the media bay until the
Power LED on the top is lit. The device is then in
carefully reviewed and is believed to be reliable.
Nanometrics Inc. reserves the right to make changes at override mode. To disable override mode, configure the
Pin Name Function Exceptional dynamic range and low hysteresis
any time without notice to improve the reliability and power thresholds in the device configuration to meet
A GPS_ANT GPS antenna signal and power function of the product. Low power consumption the available power and reboot the TitanSMA.
Shell GPS ground, connected to chassis Force balance triaxial accelerometer
No part of this publication may be reproduced, stored
in a retrieval system, or transmitted, in any form or
Power off / shutdown
Three-channel 24-bit digitizer
USB by any means, electronic, mechanical, photocopying, To power the device off using the buttons in the media
GPS receiver and system clock
Type: recording, or otherwise, without the prior written bay, press and hold the Media eject ( ) and USB eject
USB 2.0 type A, female permission of Nanometrics. Inc. Removable data storage ( ) buttons for 6 seconds. The Media eject and USB
1 2 3 4 Recommended mate: Remote communication options eject LEDs will blink red as the device shuts down.
USB 2.0 type A, male Use as a component in a real-time data -OR-
acquisition network, or as a stand-alone time To power the device off from the Web interface,
Pin Name Function series data recorder click Shutdown on the Maintenance page of the Web
2012 Nanometrics Inc. All Rights Reserved.
4 GND Ground interface.
TitanSMA Getting Started Guide
3 D+ Data +
17203R3 2012-05-07
In either case, when all LEDs on the device are off, it is
2 D- Data - safe to disconnect power.
1 VBUS +5 V Applies to
Model number 17197 (TitanSMA) Pre-deployment setup instructions inside...