8.912.8001EN / 2016-11-11
Metrohm AG
CH-9100 Herisau
Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01
info@metrohm.com
www.metrohm.com
pH Meter/Conductometer
912 Conductometer | 913 pH Meter |
914 pH/Conductometer
Manual
Table of contents
1 Introduction 1
1.1 Instrument description ......................................................... 1
1.1.1 Instrument and sales versions .................................................. 2
1.1.2 Power supply ........................................................................... 2
1.1.3 Interfaces ................................................................................ 3
1.1.4 Sensors .................................................................................... 3
1.2 Intended use ......................................................................... 3
1.3 About the documentation ................................................... 3
1.3.1 Symbols and conventions ........................................................ 3
1.4 Safety instructions ................................................................ 4
1.4.1 General notes on safety ........................................................... 4
1.4.2 Electrical safety ........................................................................ 5
1.4.3 Flammable solvents and chemicals ........................................... 5
1.4.4 Recycling and disposal ............................................................. 6
3 Installation 12
3.1 Unpacking and inspecting the instrument ....................... 12
3.1.1 Packaging .............................................................................. 12
3.1.2 Checks .................................................................................. 12
3.1.3 Application area .................................................................... 12
3.2 Power supply ...................................................................... 12
3.2.1 Charging the accumulator ..................................................... 13
3.2.2 Operation with power supply unit ......................................... 14
3.2.3 Operation via USB connector (PC) .......................................... 15
3.3 Connecting sensors ............................................................ 15
3.4 Connecting a printer .......................................................... 16
3.5 Initial configuration ............................................................ 17
3.5.1 Setting the Language ............................................................. 17
3.5.2 Setting the date and time ...................................................... 18
pH Meter/Conductometer III
Table of contents
4 Operation 20
4.1 Switching the instrument on and off ............................... 20
4.2 Displays ............................................................................... 20
4.3 Status displays .................................................................... 23
4.3.1 Accumulator status ................................................................ 23
4.3.2 User rights ............................................................................. 23
4.3.3 Sensor quality for pH electrodes ............................................ 24
4.4 Control keys ........................................................................ 24
4.5 Basic operation ................................................................... 25
4.5.1 Main dialog with two measuring channels ............................. 25
4.5.2 Main dialog with one measuring channel ............................... 26
4.5.3 Operation in the main dialog ................................................. 27
4.5.4 Menu dialog .......................................................................... 28
4.5.5 Editing dialog ........................................................................ 29
4.5.6 Selection dialog ..................................................................... 31
4.5.7 Changing the user ................................................................. 32
4.6 Menu structures .................................................................. 33
4.6.1 912 Conductometer .............................................................. 34
4.6.2 913 pH Meter ........................................................................ 35
4.6.3 914 pH/Conductometer IS ..................................................... 36
4.6.4 914 pH/Conductometer ......................................................... 37
4.7 Menu dialogs ...................................................................... 38
4.7.1 Parameters pH/U/T and Parameters pH/U/T IS ........................ 38
4.7.2 Parameters /TDS/Sal//T ...................................................... 39
4.7.3 Measured values .................................................................... 40
4.7.4 Sensors .................................................................................. 42
4.7.5 Report ................................................................................... 47
4.7.6 Configuration ........................................................................ 48
4.7.7 User ...................................................................................... 49
4.8 pH measurement ................................................................ 50
4.8.1 pH electrode calibration ......................................................... 50
4.8.2 Measurement ........................................................................ 52
4.9 Conductivity measurement ................................................ 53
4.9.1 Determination of the cell constant (calibration) ...................... 53
4.9.2 Measurement ........................................................................ 54
4.10 Issuing reports/measured values ....................................... 56
4.10.1 Printing out ........................................................................... 56
4.10.2 PC/LIMS and CSV data transfer .............................................. 56
IV pH Meter/Conductometer
Table of contents
6 Troubleshooting 61
6.1 General ................................................................................ 61
6.2 Problems ............................................................................. 62
6.2.1 Troubleshooting .................................................................... 62
6.3 Restarting/resetting the instrument ................................. 64
6.3.1 Instrument reset .................................................................... 64
6.3.2 Resetting the instrument to factory settings ........................... 64
6.4 Messages ............................................................................. 65
7 Appendix 66
7.1 Saved buffer series ............................................................. 66
7.1.1 Metrohm ............................................................................... 67
7.1.2 NIST (according to DIN standard 19266, 2000) ...................... 68
7.1.3 DIN (according to DIN standard 19267, 2012) ....................... 69
7.1.4 Fisher .................................................................................... 70
7.1.5 Mettler Toledo ....................................................................... 71
7.1.6 Merck CertiPUR 20 / Titrisol ................................................... 72
7.1.7 Merck CertiPUR 25 ................................................................ 73
7.1.8 Beckmann ............................................................................. 74
7.1.9 Radiometer Analytical ............................................................ 75
7.1.10 Baker ..................................................................................... 76
7.1.11 Hamilton DURACAL ............................................................... 77
7.1.12 Fluka ..................................................................................... 78
8 Technical specifications 79
8.1 Measuring inputs ................................................................ 79
8.2 Measured value memory ................................................... 80
8.3 TFT display .......................................................................... 80
8.4 Interfaces ............................................................................. 80
8.5 Power supply ...................................................................... 80
8.6 Charging time ..................................................................... 81
8.7 Runtime with accumulator ................................................ 81
8.8 Housing specification ......................................................... 82
8.9 Safety specification ............................................................ 82
8.10 Electromagnetic compatibility (EMC) ................................ 82
8.11 Ambient temperature ......................................................... 83
8.12 Reference conditions .......................................................... 83
8.13 Dimensions/material ........................................................... 83
pH Meter/Conductometer V
Table of contents
9 Accessories 84
Glossary 85
Index 86
VI pH Meter/Conductometer
Table of figures
Table of figures
Figure 1 pH Meter/Conductometer - front ...................................................... 7
Figure 2 912 Conductometer - connectors ...................................................... 8
Figure 3 913 pH Meter - connectors ............................................................... 8
Figure 4 914 pH/Conductometer (digital) - connectors .................................... 9
Figure 5 914 pH/Conductometer (analog) - connectors ................................... 9
Figure 6 pH Meter/Conductometer in a laboratory setup ............................... 10
Figure 7 pH Meter/Conductometer for mobile application ............................. 11
Figure 8 USB Y cable ..................................................................................... 17
Figure 9 View - main dialog .......................................................................... 21
Figure 10 View - menu dialog ......................................................................... 21
Figure 11 View - editing dialog ....................................................................... 22
Figure 12 View - selection dialog .................................................................... 22
Figure 13 Operation - main dialog, two-channel ............................................. 26
Figure 14 Operation - main dialog, one-channel ............................................. 27
Figure 15 Operation - menu dialog ................................................................. 29
Figure 16 Operation - editing dialog ............................................................... 30
Figure 17 Operation - selection dialog ............................................................. 31
Figure 18 Example of a message ..................................................................... 65
pH Meter/Conductometer VII
1 Introduction
1 Introduction
This manual gives you a comprehensive overview of the installation, func-
tioning and operation of pH Meter/Conductometer.
NOTE
913 pH Meter With an analog and a digital measuring channel each for measuring pH,
potential and temperature.
914 pH/Conducto- With a digital measuring channel for measuring pH, potential and temper-
meter IS ature and a measuring channel for measuring conductivity, TDS, salinity
and temperature.
914 pH/Conducto- With an analog measuring channel for measuring pH, potential and tem-
meter perature and a measuring channel for measuring conductivity, TDS, salin-
ity and temperature.
pH Meter/Conductometer 1
1.1 Instrument description
NOTE
2 pH Meter/Conductometer
1 Introduction
1.1.3 Interfaces
You can connect a printer or establish a connection with a PC for data
transfer (PC/LIMS report and CSV format) using the USB interface.
1.1.4 Sensors
Metrohm offers various sensors for specific measurements.
NOTE
CAUTION
pH Meter/Conductometer 3
1.4 Safety instructions
WARNING
This instrument has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
instrument, the following instructions must be observed carefully.
4 pH Meter/Conductometer
1 Introduction
WARNING
WARNING
WARNING
Only use the power supply unit for its intended purpose. Inappropriate
use or use of non-approved or incompatible power supply units may
cause fires and result in the revocation of the guarantee or warranty.
If you think that the accumulator or the power supply unit has been
damaged, have it checked by a service center. Do not use damaged
accumulators or power supply units.
Do not use the power supply unit outdoors.
WARNING
pH Meter/Conductometer 5
1.4 Safety instructions
6 pH Meter/Conductometer
2 Overview of the instrument
3 Display 4 Keypad
For measured values and menus (see Chap- With a total of eight keys to operate the
ter 4.2, page 20). instrument (see Chapter 4.4, page 24).
pH Meter/Conductometer 7
2.1 Instrument connectors
8 pH Meter/Conductometer
2 Overview of the instrument
pH Meter/Conductometer 9
2.2 Application environment
10 pH Meter/Conductometer
2 Overview of the instrument
pH Meter/Conductometer 11
3.1 Unpacking and inspecting the instrument
3 Installation
3.1.2 Checks
Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.
CAUTION
NOTE
12 pH Meter/Conductometer
3 Installation
CAUTION
Unauthorized manipulations
The instrument may be damaged as a result of unauthorized manipula-
tions.
For charging, only use the supplied power supply unit (6.2166.100)
or the optional car charger adapter (6.2166.500), which have been
approved as accessories for use with this instrument.
Your instrument's accumulator cannot be removed.
Do not attempt to remove the accumulator from the instrument. To
replace the accumulator, take the instrument to your nearest author-
ized Metrohm Service.
Unauthorized replacement of the accumulator may result in a loss of
the warranty.
NOTE
NOTE
Charge accumulator
The accumulator is charged only when the instrument is switched on.
Please observe the following procedure:
1. First, connect the USB cable to the supplied power supply unit, then
the power supply unit to the socket, and then finally the mini USB
connector of the cable to your instrument.
2. Switch on the instrument with the .
3. When the accumulator is fully charged, disconnect the power supply
unit first from the instrument and then from the socket.
or
pH Meter/Conductometer 13
3.2 Power supply
If no power socket is available, you can also charge your instrument via
the USB interface of a computer (PC) or a car charger adapter.
NOTE
Charging capacity
The charging process requires a minimum capacity of 500 mAh.
Use a USB hub with external power supply or a suitable USB connec-
tor on the computer (PC).
For mobile charging of the accumulator, use a car charger adapter
from the Metrohm accessories (6.2166.500).
1. Connect the USB cable first to the computer and then to your instru-
ment.
2. Then switch on the instrument.
3. As soon as the accumulator is fully charged, disconnect the USB cable
first from the instrument and then from the computer.
NOTE
Accumulator condition
The accumulator performance may deteriorate over time. If the operat-
ing times are much shorter than usual, take the instrument to the clos-
est Metrohm Service to have the accumulator replaced.
CAUTION
14 pH Meter/Conductometer
3 Installation
NOTE
NOTE
NOTE
Parameter setting
Please note that if you change the sensor, the sensor either has to be
selected in the menu dialog Menu Parameters X Measuring
parameters Sensor name, or a new sensor has to be entered in
the sensor list.
pH Meter/Conductometer 15
3.4 Connecting a printer
NOTE
NOTE
Printer function
The connected printer will only work if the
pH Meter/Conductometer is connected to the power supply unit.
NOTE
16 pH Meter/Conductometer
3 Installation
3 USB type B
Printer connector.
NOTE
You can access the menu structures via the Menu item (see Figure 13,
page 26) on the main screen.
1
Use the or arrow key to select the entry point.
pH Meter/Conductometer 17
3.5 Initial configuration
2
Use the button to change to the menu structures.
3
Use the arrow key to select the Configuration menu struc-
4
Use the arrow key to select the Language menu dialog and
5
Use the or arrow key to select the required lan-
NOTE
You can access the menu structures via the Menu item (see Figure 13,
page 26) on the main screen.
1
Use the or arrow key to select the entry point.
2
Use the button to change to the menu structures.
3
Use the arrow key to select the Configuration menu struc-
18 pH Meter/Conductometer
3 Installation
4
Use the arrow key to select the Date or Time menu dialog
6
Confirm the final value with the editing element and the
key.
pH Meter/Conductometer 19
4.1 Switching the instrument on and off
4 Operation
1
Press the key.
The instrument is initialized and a system test is performed. This pro-
cess takes some time.
A starting image is displayed during start-up.
Then the main dialog is displayed. Now the instrument is ready.
1
Press the key.
The 912-129 Shut down message appears, the instrument saves
the data and switches off.
4.2 Displays
The pH Meter/Conductometer has a total of four display types contain-
ing specific displays and/or operating functions.
Main dialog
Menu dialog
Editing dialog
Selection dialog
NOTE
20 pH Meter/Conductometer
4 Operation
Main dialog The main dialog (example: both measuring channels displayed) is the nor-
mal status after the instrument has been switched on.
Menu dialog The menu dialog is used for navigating through the functional structures.
Menu lines with an arrow contain another, deeper structure with further
dialogs.
Editing dialog Editing dialogs are used in general for data entry and editing.
Depending on the data type, a different set of possible characters is avail-
able.
pH Meter/Conductometer 21
4.2 Displays
NOTE
You can insert capital letters and special characters by holding the
key pressed down.
Selection dialog Selection dialogs offer default values for selection in corresponding data
fields.
22 pH Meter/Conductometer
4 Operation
NOTE
Flash icon
The flash icon indicates that the instrument is connected to a power
source for charging.
If the key icon (at the top in the main dialog) is displayed, then the
user menu is limited to the functions for Routine users.
pH Meter/Conductometer 23
4.4 Control keys
24 pH Meter/Conductometer
4 Operation
NOTE
Calibration
A sensor can be calibrated only in the corresponding one-channel main
dialog.
The BACK key causes the entry to be accepted and/or exits the dialog.
The LEFT/RIGHT arrow keys are used for navigating in the text and
number editor for selecting characters, or for toggling between the mea-
suring channel displays in the main dialog.
The UP/DOWN arrow keys are used for navigating the selection bar one
line up or down, or for selecting characters in the text editor.
NOTE
This does not apply for the 912 Conductometer, as this instrument
has only one measuring channel.
NOTE
Temperature display
The temperature displays on the two measuring channels can only be
compared in the same medium.
As a result of temperature sensor and instrument tolerances, the dis-
played values might deviate from each other.
pH Meter/Conductometer 25
4.5 Basic operation
NOTE
The procedure for functions of the main dialog with two measuring
channels is the same as for the main dialog with one measuring chan-
nel:
(see "Accessing the menu structures", page 27).
(see "Toggling from one-channel to two-channel view", page 28).
26 pH Meter/Conductometer
4 Operation
7 ID1 8 ID2
Input option for sample designation/identifi- Input option for sample designation/identifi-
cation (e.g. name, number, etc.). cation (e.g. batch number, lot number, etc.).
You can access the menu structures via the Menu item (14-2) in the main
dialog.
1
Select the entry point with the or arrow keys.
pH Meter/Conductometer 27
4.5 Basic operation
2
Use the key to change to the menu structures.
28 pH Meter/Conductometer
4 Operation
3 Arrow icon
The arrow icon indicates that there are fur-
ther substructures.
You can access the menu structures via the Menu item (14-2) in the main
dialog.
1
Select the desired menu line with the or arrow
keys.
2
Change into the next substructure with the key.
3
Change back to the higher structure with the key.
pH Meter/Conductometer 29
4.5 Basic operation
5 Function elements
The editing dialog is closed without changing the existing data value with
the Cancel editing element.
The entire contents of the data field are deleted and a new data value can
be entered with the Delete all editing element.
The character to the left of the cursor is deleted in the data field with the
Backspace editing element.
The cursor moves one space to the left in the data field with the One
space to the left editing element.
The cursor moves one space to the right in the data field with the One
space to the right editing element.
NOTE
30 pH Meter/Conductometer
4 Operation
2
Select the required data value with the or arrow
keys.
3
Apply the data value and exit the selection window with the
key.
pH Meter/Conductometer 31
4.5 Basic operation
Routine
Expert
4 Enter the Password set on the instrument and confirm with the
icon.
The instrument's functions are fully accessible and the key icon
is no longer displayed in the main menu.
32 pH Meter/Conductometer
4 Operation
NOTE
Menu dialogs
The menu dialogs and the corresponding menu lines are described in
more detail in the next chapter (see Chapter 4.7, page 38).
pH Meter/Conductometer 33
4.6 Menu structures
34 pH Meter/Conductometer
4 Operation
pH Meter/Conductometer 35
4.6 Menu structures
36 pH Meter/Conductometer
4 Operation
pH Meter/Conductometer 37
4.7 Menu dialogs
Measuring parame-
ters Menu dialog for the Measuring parameters.
Sensor name Selection dialog for selecting a sensor from the sensor list.
In instruments with iTrodes, this is only a display field.
Temperature Editing dialog for the manual entry of the measuring temperature.
Default value: 25.0 C / input range: 999.9 - +999.9 C
Does not apply for instruments with iTrodes.
Temperature Editing dialog for the manual entry of the calibration temperature.
Default value: 25.0 C / input range: 0.0 - 99.9 C
Number of buf- Selection dialog for the Number of buffers that are used for calibration.
fers
Default value: 2 / input range: 1 - 5
38 pH Meter/Conductometer
4 Operation
Measuring parame-
ters Menu dialog for the Measuring parameters.
Sensor name Selection dialog for a sensor from the sensor list.
Order number Display field for the Order number of the selected sensor.
Serial number Display field for the Serial number of the selected sensor.
Reference temp. Editing dialog for entering the reference temperature of the calibration stan-
dard.
Default value: 25.0C / input range: 0 - 99.9C
Temp. compens. Editing dialog for entering the temperature compensation value.
Default value: 2.00%/C / input range: 0.00 - 9.99%/C
If no temperature compensation is to be applied, enter 0.0%/C.
DIN
Permanently saved function for temperature compensation for natural
groundwater, well water or surface water in accordance with DIN 38404-C8.
TDS factor Editing dialog for entering the factor value for the TDS calculation.
pH Meter/Conductometer 39
4.7 Menu dialogs
Reference temp. Editing dialog for entering the reference temperature of the calibration stan-
dard.
Default value: 25.0C / input range: 0 - 99.9C
Stand. conduct. Editing dialog for entering the calibration standard value.
Default value: 12.870 mS/cm / input range: 0.0000 - 2,000.0 mS/cm
Values 1. Selection dialog for viewing and deleting the Values on the instrument.
2. Selection dialog to indicate how the Values are to be saved on the instru-
ment for output.
view
The measured values are shown individually on the display and can be tog-
gled individually with the arrow keys.
In addition, you can navigate in the measured value list as follows using the
key:
If you push the key briefly, then the last measured value will be dis-
played.
If you hold the key for longer, then the first measured value will be
displayed.
delete all
All measured values on the instrument will be irreversibly deleted.
delete last
The latest (newest) measured value will be irreversibly deleted.
save as CSV
The currently saved measured values will be saved on the instrument as a
CSV file (see "CSV file", page 57).
40 pH Meter/Conductometer
4 Operation
save as PC/LIMS
The currently saved measured values will be saved on the instrument as PC/
LIMS file (see "PC/LIMS report", page 57).
Data Selection dialog to indicate whether the Data are to be printed and/or saved.
print:
save:
print+save:
Criterion Selection dialog to indicate when the Measured values are applied during the
measurement.
immediately
The displayed measured value will be immediately applied.
time-dependent
The measured value will be applied during a Time interval that can be set.
For the PC/LIMS report, the individual measured values are grouped in a data
group.
The interval ends after the Stop time has been achieved.
The parameters are as follows:
Time interval in seconds
Default value: 4 s / input range: 1 - 999,999 s
Stop time in seconds
Default value: off / input range: 1 - 999,999 s
when changed
The subsequent measured value will only be applied automatically if the devi-
ation from the previous measured value is greater than the value
delta pH/T/mV/ defined here.
Delta pH
Default value: 0.50 pH / input range: 0.10 - 16.00 pH
Delta T(pH)
Default value: 0.5 C / input range: 0.1 - 100.0 C
Delta mV
Default value: 30.0 mV / input range: 0.1 - 999.9 mV
Delta
Default value: 0.1 mS/cm / input range: 0.0001 - 10 mS/cm
Delta T()
Default value: 0.5 C / input range: 0.1 - 100.0 C
Stop time
Default value: off / input range: 1 - 999,999 s
Primary channel
Selection dialog for selecting the measuring channel that has to fulfill
the change criterion.
pH Meter/Conductometer 41
4.7 Menu dialogs
drift-dependent
The measured value will be applied when the value is stable according to the
drift criterion.
The drift thresholds are preset and cannot be changed:
pH measurement: 0.028 pH/min
Potential measurement U/mV: 1.875 mV/min
Temperature measurement T/C: 0.974 C/min
For the conductivity, various drift thresholds are stored (depending on the
measuring range):
0.005 mS/cm/min in the measuring range up to 16 S/cm
0.5 mS/cm/min in the measuring range from 16 S/cm to 1 mS/cm
10 mS/cm/min in the measuring range from 1 mS/cm
For instruments with two measuring channels, the primary measuring chan-
nel for the fulfillment of the drift criterion has to be selected.
Primary channel
Selection dialog for selecting the measuring channel that has to fulfill
the drift criterion.
pH/mV or Cond.
Output date/time Selection dialog for selecting whether a time stamp is to be assigned to a mea-
sured value.
on: Measured values contain a time stamp on the report.
off: Measured values do not contain a time stamp on the report.
Default value: off
Output headers Selection dialog for selecting how the headers are output.
once
always: is the default setting
off
Calibration data Selection dialog for selecting whether the main Calibration data is assigned for
the output of the measured values.
on
off: is the default setting
4.7.4 Sensors
The Sensors menu dialog is shown below with the structure and the
description.
42 pH Meter/Conductometer
4 Operation
NOTE
NOTE
iTrodes
iTrodes sensors contain their own data in the data memory which
becomes available directly in the sensor data when the sensors are con-
nected to the instrument.
Some of this data cannot be edited.
Sensor list An available sensor can be selected in the Sensor list. The individual menu lines
are then also available corresponding to the selected sensor.
Selection Selection dialog for selecting an identified sensor for editing and displaying the
individual data.
k default
metal def.
pH default
temp default
etc.
Additional sensors entered by the user.
pH Meter/Conductometer 43
4.7 Menu dialogs
Calibration
temp. Display field indicating the temperature in C from the last calibration.
K Temp. compens. Display field indicating the value for temperature compensation of the last cali-
bration.
Default value: 2.07%/C / input range: 0.00 - 9.99%/C
Temp. calibra- Display field indicating the measurement method for temperature measurement
tion of the last calibration.
Calibration Editing dialog for entering time in days for the Calibration interval.
interval
Default value: off / input range: 1 - 999 d
off disables the Calibration interval.
Temp. sensor Selection dialog to indicate the temperature sensor type for the respective sen-
sor.
Pt1000: is the default value
NTC
R(25C)
Editing dialog
Default value: 30,000 / input range: 10,000 - 100,000
B value
Editing dialog
Default value: 4,100 / input range: 1,000 - 9,999
pH Slope limit value Selection dialog for selecting whether the limit value is to be applied.
44 pH Meter/Conductometer
4 Operation
on
off: is the default value
pH Slope lower limit Editing dialog for entering the lower limit value.
Default value: 95.00 / input range: 1.0 - 999.9%
Slope upper Editing dialog for entering the upper limit value.
pH limit
Default value: 103.00 / input range: 1.0 - 999.9%
Limit value Selection dialog for selecting whether the limit value is to be applied.
pH pH(0)
on
off: is the default value
pH pH(0) lower limit Editing dialog for entering the lower limit value.
Default value: 6.40 / input range: 0.00 - 99.99
pH(0) upper Editing dialog for entering the upper limit value.
pH limit
Default value: 8.00 / input range: 0.00 - 99.99
K Limit value c Selection dialog for selecting whether the limit value is to be applied.
on
off: is the default value
K c lower limit Editing dialog for entering the lower limit value.
Default value: 0.400/cm / input range: 0.001 - 500/cm
K c upper limit Editing dialog for entering the upper limit value.
Default value: 0.550/cm / input range: 0.001 - 500/cm
New sensor Menu dialog with the individual menu lines for entering a new sensor.
pH Meter/Conductometer 45
4.7 Menu dialogs
Calibration Editing dialog for entering time in days for the Calibration interval.
interval
Default value: off / input range: 1 - 999 d
off disables the Calibration interval.
Temp. sensor Selection dialog to indicate the temperature sensor type for the respective sen-
sor.
Pt1000
NTC
R(25C)
Editing dialog
Default value: 30,000 / input range: 10,000 - 100,000
B value
Editing dialog
Default value: 4,100 / input range: 1,000 - 9,999
pH Slope limit value Selection dialog for selecting whether the limit value is to be applied.
on
off: is the default value
pH Slope lower limit Editing dialog for entering the lower limit value.
Default value: 95.00 / input range: 1.0 - 999.9%
Slope upper Editing dialog for entering the upper limit value.
pH limit
Default value: 103.00 / input range: 1.0 - 999.9%
Limit value Selection dialog for selecting whether the limit value is to be applied.
pH pH(0)
on
off: is the default value
pH pH(0) lower limit Editing dialog for entering the lower limit value.
Default value: 6.40 / input range: 0.00 - 99.99
pH(0) upper Editing dialog for entering the upper limit value.
pH limit
Default value: 8.00 / input range: 0.00 - 99.99
K Limit value c Selection dialog for selecting whether the limit value is to be applied.
on
46 pH Meter/Conductometer
4 Operation
K c lower limit Editing dialog for entering the lower limit value.
Default value: 0.400/cm / input range: 0.001 - 500/cm
K c upper limit Editing dialog for entering the upper limit value.
Default value: 0.550/cm / input range: 0.001 - 500/cm
4.7.5 Report
The Report menu dialog is shown below with the structure and the
description.
NOTE
Printer
pH Meter/Conductometer instruments support various printer types for
report output. If your printer is not listed, please use the printer
Universal (ESC-POS), which has appropriate setting parameters.
Line feed Editing dialog to indicate the lines to be inserted at the end of the report.
Default value: 2 lines / input range: 0 - 99 lines
pH Meter/Conductometer 47
4.7 Menu dialogs
Seiko (ESC-POS)
Roll printer with paper width 110 mm
Citizen ESC-POS)
Roll printer with paper width 80 mm
Custom (ESC-POS)
Roll printer with paper width 60 mm
Universal (ESC-POS)
Universal roll printer with variable settings:
Paper width
50 - 200 mm
Print resolution
100 - 600 dpi
Print type
Line or Matrix
4.7.6 Configuration
The Configuration menu dialog for all instrument settings is shown
below with the structure and the description.
Last decimal place Selection dialog for selecting whether the Last decimal place is displayed for
the pH measured values with three digits at most.
Drift value monitoring is not influenced by this setting.
on: Last decimal place is displayed.
off: Last decimal place is not displayed.
Power off after Editing dialog for entering the time for the function Power off after x
minutes. The instrument shuts down automatically after this time.
This function is disabled while the instrument is charged and during time-depen-
dent recording of measured values with a set time interval.
Default value: 15 / input range: 1 - 60, or
off for continuous operation.
Turn off LCD after Editing dialog for entering the time for the function Turn off LCD after x
minutes. The display turns off after this time and can be turned back on again
with any key other than the key.
Default value: 15 / input range: 1 - 60, or
auto for dimming after 20 seconds and switching off the display after
another 60 seconds.
off for continuous operation.
48 pH Meter/Conductometer
4 Operation
Change password Editing dialog for customizing the password for the Expert user rights.
The default setting ex works is Expert.
1. Old password
2. New password
3. Confirm
4.7.7 User
The User menu dialog for setting user restrictions and user data is shown
below with the description.
pH Meter/Conductometer 49
4.8 pH measurement
4.8 pH measurement
This chapter describes the required steps to carry out a simple pH mea-
surement with calibration. The description is limited to only the indispen-
sable steps and will enable you to carry out first measurements with the
instrument directly.
NOTE
pH calibration
By default, the calibration parameters are set for calibration with two
Metrohm buffer solutions (see Chapter 4.7.1, page 38). If you would like
to use other buffers, you have to select the corresponding buffer type and
the number of buffer solutions.
If the Report selection dialog is set to on in the Calibration param.
menu dialog, then the calibration data will be output immediately.
50 pH Meter/Conductometer
4 Operation
NOTE
Buffer exchange
If the buffer solution was not exchanged, then the message
912-181 Same buffer will appear.
Exchange the buffer solution and continue the calibration with the
key.
NOTE
key, or you can reject it with the key and use the exist-
ing calibration data.
pH Meter/Conductometer 51
4.8 pH measurement
4.8.2 Measurement
NOTE
key.
52 pH Meter/Conductometer
4 Operation
NOTE
NOTE
The calibration parameters are set to default values (see Chapter 4.7.2,
page 39).
If the Report selection dialog is set to on in the Calibration param.
menu dialog, then the calibration data will be output immediately.
pH Meter/Conductometer 53
4.9 Conductivity measurement
NOTE
4.9.2 Measurement
NOTE
54 pH Meter/Conductometer
4 Operation
key.
NOTE
pH Meter/Conductometer 55
4.10 Issuing reports/measured values
or button.
Printing out saved values in the following areas as reports:
Calibration
Sensors
Configuration
Parameter
Measured values
Report data can be printed using the Report selection dialog under:
Menu Report
NOTE
The value "dpH" indicates the difference between nominal value of the
buffer (by interpolating between two values from the buffer table) and
the pH value resulting from the measured voltage from the calibration
lines.
NOTE
USB cable
Data can be transferred to a PC connected using the supplied USB cable
(6.2151.110).
The optional USB Y cable (6.2151.140) cannot be used.
56 pH Meter/Conductometer
4 Operation
NOTE
Data output
For data output, the setting save or print+save is required when
recording measured values so that this data is saved on the instrument.
Data generation
The measured value data has to be re-generated before every data
transfer.
The data on the instrument memory can be saved in two data formats:
PC/LIMS report
Data in PC/LIMS format can be imported into and processed in the
Metrohm program tiBase for evaluation.
NOTE
Data collision
Transferring data from several instruments can lead to a data collision in
tiBase.
You should create an individual database for each measuring instru-
ment in tiBase.
CSV file
CSV data can be imported as text data in MS Excel and processed for
evaluation.
The following parameters are required for text conversion:
Data type = separated with a semicolon
Data source = Unicode (UTF-8)
Data format of the columns = standard
pH Meter/Conductometer 57
4.10 Issuing reports/measured values
NOTE
USB interface
If the report data is to be generated while the instrument is connected
to the PC, then the connection is briefly interrupted.
After the report data has been generated, the connection will be auto-
matically established again.
58 pH Meter/Conductometer
5 Operation and maintenance
CAUTION
NOTE
Sensor leaflet
Handling, care and storage are important factors for the correct and
accurate functioning of sensors.
Therefore, please note the specific information on the respective sensor
leaflets.
You can download the leaflets from the Internet at
http://www.metrohm.com.
pH Meter/Conductometer 59
5.2 Quality management and qualification with Metrohm
Qualification
Please contact your local Metrohm representative for support in qualifica-
tion of instruments and software. The Installation Qualification (IQ)
and Operational Qualification (OQ) are offered by Metrohm represen-
tatives as a service. They are carried out by trained employees using stan-
dardized qualification documents and in accordance with the currently
applicable requirements of the regulated industry.
Maintenance
The electronic and mechanical modules of Metrohm instruments can and
should be checked by specialist personnel from Metrohm as part of a reg-
ular preventive maintenance schedule. Please ask your local Metrohm rep-
resentative regarding the precise terms and conditions involved in con-
cluding a corresponding maintenance agreement.
For detailed information on this topic, please visit www.metrohm.com.
60 pH Meter/Conductometer
6 Troubleshooting
6 Troubleshooting
6.1 General
If you experience problems during measurements, then you can check the
following aspects to eliminate them:
Application Difficult sample matrices or interfering influences may render accurate
measurements impossible (e.g. insufficient ionic strength, presence of
interfering ions, etc.).
Our Application Bulletins and Application Notes will support you in
choosing the appropriate analysis conditions and configuring the instru-
ment method.
Buffer solutions / The precision of the measurements mainly depends on the correct calibra-
standard solutions tion of the sensors. To do so, you should use clean and fresh buffer or
standard solutions.
A common cause of incorrect calibrations is, for example, the use of an
old pH 10 or pH 12 buffer. The pH value of such buffers may markedly
deviate from the certified pH value of a new buffer as a result of the intro-
duction of CO2 from the air.
Sensors The sensors are the most important component in the entire measuring
system.
For the correct handling of sensors, please read the corresponding leaf-
lets.
pH Meter/Conductometer 61
6.2 Problems
6.2 Problems
The following list describes some general problems that might occur dur-
ing measurements. Furthermore, the possible causes and solution
approaches are described.
NOTE
Sensor treatment
Follow the instructions given in the respective leaflets for sensors clean-
ing and maintenance.
6.2.1 Troubleshooting
Problem Cause Remedy
Measured value set- The glass membrane or the Clean the electrode following the instruc-
ting is sluggish. diaphragm is contami- tions in the leaflet.
nated.
The instrument does The instrument battery is Connect the instrument to the power sup-
not start. not charged. ply unit to charge it.
The battery is only charged when the
instrument is on.
(total charging time: approx. 9 hours)
The measured value The glass membrane or the Clean the electrode following the instruc-
drift criterion is not diaphragm is contami- tions in the leaflet.
fulfilled. nated.
62 pH Meter/Conductometer
6 Troubleshooting
Non-Metrohm power sup- Use only the supplied power supply unit
ply unit used for charging during measurement operation.
the battery.
The glass membrane or the Clean the membrane or the diaphragm fol-
diaphragm is contami- lowing the instructions in the correspond-
nated. ing leaflet.
The slope is insuffi- The glass membrane or the Clean the electrode following the instruc-
cient during calibra- diaphragm is contami- tions in the leaflet.
tion. nated.
pH Meter/Conductometer 63
6.3 Restarting/resetting the instrument
The instrument can be switched on again.
NOTE
Data storage
The currently measured data and modified settings cannot be saved if
the instrument is reset.
CAUTION
User data
The user data will be irreversibly deleted.
You can reset the instrument with the following key combination during
startup:
64 pH Meter/Conductometer
6 Troubleshooting
Afterwards, a message will be displayed saying that the user data has
been deleted.
6.4 Messages
The instruments notify you of possible errors or operation problems with
various specific messages. A message as shown in the following example
will appear on the current display.
NOTE
Message number
Each message contains a message number in the top left-hand corner.
Please indicate this number in the case of queries or complaints.
pH Meter/Conductometer 65
7.1 Saved buffer series
7 Appendix
CAUTION
Buffer quality
The precision of pH measurements mainly depends on the correct cali-
bration of the measuring chain. To do so, you should use clean and
fresh buffer solutions. A common cause of incorrect calibration is, for
example, the use of an old pH 10 or pH 12 buffer. The pH value of a
buffer solution may markedly deviate from the certified pH value of a
new buffer solution as a result of the introduction of CO2 from the air.
NOTE
pH values printed in bold are the values for the reference temperature
of the respective buffer set.
pH values highlighted in italics are interpolated or extrapolated values.
The other pH values correspond to the manufacturer's specifications.
66 pH Meter/Conductometer
7 Appendix
7.1.1 Metrohm
Table 6 Metrohm buffer solutions
Metrohm
Temp. pH pH pH
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
pH Meter/Conductometer 67
7.1 Saved buffer series
Temp. pH pH pH pH pH
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept as up to date as possible.
However, they may be changed by the respective manufacturers.
The NIST buffers are identical to the buffer solutions that are used in
Chinese pharmacopoeia.
68 pH Meter/Conductometer
7 Appendix
Temp. pH pH pH pH pH pH
95 - - - - - -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
pH Meter/Conductometer 69
7.1 Saved buffer series
7.1.4 Fisher
Table 9 Fisher buffer solutions
Fisher
Temp. pH pH pH pH
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
70 pH Meter/Conductometer
7 Appendix
Temp. pH pH pH pH pH
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
pH Meter/Conductometer 71
7.1 Saved buffer series
Temp. pH pH pH pH pH
95 - 4.000 7.020 - -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible. However, they may be
changed by the different manufacturers.
The CertiPUR 20 and Titrisol product lines have the identical values.
72 pH Meter/Conductometer
7 Appendix
Temp. pH pH pH pH
0 - - - -
55 - - - -
60 - - - -
65 - - - -
70 - - - -
75 - - - -
80 - - - -
85 - - - -
90 - - - -
95 - - - -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible. However, they may be
changed by the different manufacturers.
pH Meter/Conductometer 73
7.1 Saved buffer series
7.1.8 Beckmann
Table 13 Beckmann buffer solutions
Beckmann
Temp. pH pH pH
55 4.08 6.98 -
60 4.09 6.98 -
65 4.11 6.99 -
70 4.12 6.99 -
75 4.14 7.00 -
80 4.16 7.00 -
85 4.18 7.01 -
90 4.19 7.02 -
95 4.21 7.03 -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
74 pH Meter/Conductometer
7 Appendix
Temp. pH pH pH pH
95 - 4.240 - -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
pH Meter/Conductometer 75
7.1 Saved buffer series
7.1.10 Baker
Table 15 Baker buffer solutions
Baker
Temp. pH pH pH pH
95 - - - -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
76 pH Meter/Conductometer
7 Appendix
Temp. pH pH pH pH
0 - - - -
55 - - - -
60 - - - -
65 - - - -
70 - - - -
75 - - - -
80 - - - -
85 - - - -
90 - - - -
95 - - - -
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
pH Meter/Conductometer 77
7.1 Saved buffer series
7.1.12 Fluka
Table 17 Fluka buffer solutions
Fluka
Temp. pH pH pH
NOTE
Update
The values of the individual buffers with the corresponding tempera-
tures are kept up to date as far as possible.
However, they may be changed by the respective manufacturers.
78 pH Meter/Conductometer
8 Technical specifications
8 Technical specifications
2.912.010 /TDS/Sal1)//T
Temperature:
0.5% at 16 S - 1 mS
1.0% at 1 mS - 500 mS
pH Meter/Conductometer 79
8.2 Measured value memory
8.4 Interfaces
USB connector Type A/B mini USB connector (USB 2.0) with the following functions:
Power supply
Data transmission
with USB cable (6.2151.110)
Print
with USB Y cable (6.2151.140)
80 pH Meter/Conductometer
8 Technical specifications
NOTE
Runtimes
The runtimes may vary according to the configuration used and the
usage habits.
The following values are based on operation under reference conditions
(see Chapter 8.12, page 83).
Uptime 8 hours
pH Meter/Conductometer 81
8.8 Housing specification
82 pH Meter/Conductometer
8 Technical specifications
8.13 Dimensions/material
Dimensions
Length 208 mm
Width 92 mm
Height 34 mm
pH Meter/Conductometer 83
9 Accessories
Up-to-date information on the scope of delivery and on optional accesso-
ries for your instrument can be found on the Internet. You can download
this information using the article number as follows:
NOTE
84 pH Meter/Conductometer
Glossary
Glossary
Display field
Display fields are menu lines with a designation and a displayed value.
Editing dialog
In editing dialogs, you can enter or edit values (see "Editing dialog", page
21).
IS
The abbreviation IS in instruments and menus stands for Intelligent
Sensor from the iTrode line of sensors.
A chip in the sensor head saves the data, which is automatically transmit-
ted when the sensor is connected with the instrument by means of the
854 iConnect. The measured data is transmitted digitally.
Main dialog
In the main dialog, measured values are displayed and you can trigger pri-
mary operations for measurements (see "Main dialog", page 21).
Menu dialog
Menu dialogs show an open menu structure with the corresponding
menu lines (see "Menu dialog", page 21).
Menu line
Menu lines are positions in the menu dialog that can be selected or that
display something.
Menu structure
The menu structure represents the navigation in the instrument through
the menus (see Chapter 4.6, page 33).
Selection dialog
In selection dialogs, you can select one option from a range of options
(see "Selection dialog", page 22).
dpH
Difference between nominal value of the buffer (by interpolating between
two values from the buffer table) and the pH value that results from the
measured voltage from the calibration lines (see Chapter 4.10, page 56).
pH Meter/Conductometer 85
Index
Index
A Instrument Power supply ........................ 2, 12
Accumulator ............................. 12 Switch off ........................... 20 Printer ...................................... 47
Switch on ..................... 13, 20 Printing .................................... 56
B Instrument version ...................... 2 Problems .................................. 61
Brightness ................................ 49 Interface ..................................... 3
Q
C L Quality management ................ 60
Calibration .......................... 50, 53 Language ................................. 49
Calibration param. .............. 38, 40 R
Conductivity ................... 1, 39, 53 M Report ................................ 47, 56
Conductometer .......................... 8 Maintenance agreement .......... 60 Resistance value ....................... 39
Configuration ........................... 48 Measured values ...................... 40
Connector .................................. 8 Measuring parameters ........ 38, 39 S
Control keys ............................. 24 Metrohm Service ...................... 59 Safety instructions ...................... 4
CSV file .................................... 57 Sales version ............................... 2
N Salinity ..................................... 39
D New sensor .............................. 45 Sensor ........................................ 3
Date ......................................... 48 Sensor list ................................. 43
Display ..................................... 20 O Sensors .................................... 42
Editing dialog ............... 21, 29 Operational qualification .......... 60 Service ....................................... 5
Main dialog ............ 21, 25, 26 Overview of the instrument ........ 7 Switch off ................................. 20
Menu dialog ................. 21, 28 Switch on ................................. 20
Selection dialog ............ 22, 31 P
Status ................................. 23 Password ............................ 32, 49 T
PC/LIMS report ......................... 57 Temperature ............................. 38
I pH value ................................... 38 Time ......................................... 48
Initial configuration .................. 17 pH/Conductometer Total Dissolved Solids ............... 39
Date/time ........................... 18 Analog ................................. 9
Language ........................... 17 Digital .................................. 9 U
Installation qualification ............ 60 Potential value .......................... 38 User ................................... 32, 49
86 pH Meter/Conductometer