Documente Academic
Documente Profesional
Documente Cultură
INTRODUCTION
1. BACKGROUND
Ammonia Assay Kit (Colorimetric) (ab83360) provides a rapid, simple,
sensitive, and reliable assay suitable for research and high throughput
assay of Ammonia or Ammonium. In the assay, ammonia or ammonium
is converted to a product that reacts with the OxiRed probe to generate
color (max = 570 nm) which can be easily quantified by plate reader. The
kit can detect 1 nmol (~20 M) of ammonia or ammonium, much more
sensitive than measuring NADPH based ammonia assay.
2. ASSAY SUMMARY
Sample preparation
GENERAL INFORMATION
3. PRECAUTIONS
Please read these instructions carefully prior to beginning the
assay.
All kit components have been formulated and quality control tested
to function successfully as a kit.
We understand that, occasionally, experimental protocols might
need to be modified to meet unique experimental circumstances.
However, we cannot guarantee the performance of the product
outside the conditions detailed in this protocol booklet.
Reagents should be treated as possible mutagens and should be
handle with care and disposed of properly. Please review the Safety
Datasheet (SDS) provided with the product for information on the
specific components.
Observe good laboratory practices. Gloves, lab coat, and protective
eyewear should always be worn. Never pipet by mouth. Do not eat,
drink or smoke in the laboratory areas.
All biological materials should be treated as potentially hazardous
and handled as such. They should be disposed of in accordance
with established safety procedures.
5. LIMITATIONS
Assay kit intended for research use only. Not for use in diagnostic
procedures.
Do not mix or substitute reagents or materials from other kit lots or
vendors. Kits are QC tested as a set of components and
performance cannot be guaranteed if utilized separately or
substituted.
6. MATERIALS SUPPLIED
Item Amount Storage Storage
Condition Condition
(Before (After
Preparation) Preparation)
Ammonia Assay Buffer 25 mL -20C -20C
OxiRed Probe 200 L -20C -20C
Developer 1 vial -20C -20C
Enzyme Mix 1 vial -20C -20C
(Lyophilized)
Converting Enzyme 1 vial -20C -20C
(Lyophilized)
NH4Cl Standard (10 mM) 100 L -20C -20C
8. TECHNICAL HINTS
This kit is sold based on number of tests. A test simply refers
to a single assay well. The number of wells that contain sample,
control or standard will vary by product. Review the protocol
completely to confirm this kit meets your requirements. Please
contact our Technical Support staff with any questions.
Selected components in this kit are supplied in surplus amount to
account for additional dilutions, evaporation, or instrumentation
settings where higher volumes are required. They should be
disposed of in accordance with established safety procedures.
Avoid foaming or bubbles when mixing or reconstituting
components.
Avoid cross contamination of samples or reagents by changing tips
between sample, standard and reagent additions.
Ensure plates are properly sealed or covered during incubation
steps.
Ensure all reagents and solutions are at the appropriate temperature
before starting the assay.
Samples which generate values that are greater than the most
concentrated standard should be further diluted in the appropriate
sample dilution buffer.
Make sure you have the right type of plate for your detection method
of choice.
Make sure all necessary equipment is switched on and set at the
appropriate temperature.
ASSAY PREPARATION
9. REAGENT PREPARATION
Briefly centrifuge small vials at low speed prior to opening
9.1. Ammonia Assay Buffer:
Ready to use as supplied. Equilibrate to room temperature before
use. Store at -20C.
9.2. OxiRed Probe:
Ready to use as supplied. Warm by placing in a 37C bath for 1- 5
minutes to thaw the DMSO solution before use. NOTE: DMSO
tends to be solid when stored at -20C, even when left at room
temperature, so it needs to melt for few minutes at 37C. Store at
- 20C protected from light. Once the probe is thawed, use with
two months.
9.3. Developer:
Reconstitute in 220 L Ammonia Assay Buffer. Aliquot developer
so that you have enough volume to perform the desired number of
assays Store at -20C. Once reconstituted, use within 2 months.
9.4. Enzyme Mix:
Reconstitute in 220 L Ammonia Assay Buffer. Aliquot enzyme so
that you have enough volume to perform the desired number of
assays. Store at -20C. Once reconstituted, use within 2 months.
9.5. Converting Enzyme:
Reconstitute in 220 L Ammonia Assay Buffer. Aliquot converting
enzyme so that you have enough volume to perform the desired
number of assays. Store at -20C. Once reconstituted, use within
2 months.
9.6. NH4Cl Standard:
Ready to use as supplied. Aliquot standard so that you have
enough volume to perform the desired number of assays. Store at
-20C. Keep on ice while in use.
ASSAY PROCEDURE
12.3. Add 50 L of Reaction Mix into each standard and sample wells.
12.4. Add 50 L Sample Control Reaction Mix to Sample Control wells.
12.5. Mix and incubate at 37C for 60 minutes protected from light.
12.6. Measure output immediately on a colorimetric microplate reader at
OD 570 nm.
DATA ANALYSIS
13. CALCULATIONS
Samples producing signals greater than that of the highest standard
should be further diluted in appropriate buffer and reanalyzed, then
multiply the concentration found by the appropriate dilution factor.
For statistical reasons, we recommend each sample should be
assayed with a minimum of two replicates (duplicates).
13.1. Average the duplicate reading for each standard and sample.
13.2. Subtract the mean absorbance value of the blank (Standard #1)
from all standard and sample readings. This is the corrected
absorbance.
13.3. Plot the corrected absorbance values for each standard as a
function of the final concentration of ammonia.
13.4. Draw the best smooth curve through these points to construct the
standard curve. Most plate reader software or Excel can plot
these values and curve fit. Calculate the trendline equation based
on your standard curve data (use the equation that provides the
most accurate fit).
13.5. Concentration of samples in the test samples is calculated as:
= ( )
Where:
Sa = amount of ammonia in the sample well calculated from
standard curve (nmol).
Sv = sample volume added into the wells (L).
D = Sample dilution factor.
Fig
ure 1: Typical ammonium standard calibration curve using colorimetric reading.
RESOURCES
16. TROUBLESHOOTING
Problem Cause Solution
Buffers must be at room
Use of ice-cold buffer
temperature
Assay not Plate read at Check the wavelength and filter
incorrect wavelength settings of instrument
working
Colorimetric: Clear plates
Use of a different 96-
Fluorometric: black wells/clear
well plate
bottom plate
Samples not
Use provided protocol for
deproteinized (if
deproteinization
indicated on protocol)
Cells/tissue samples
Use Dounce homogenizer,
not homogenized
increase number of strokes
completely
Sample with Samples used after
Aliquot and freeze samples if
erratic multiple free/ thaw
needed to use multiple times
readings cycles
Use of old or Use fresh samples or store at -
inappropriately stored 80C (after snap freeze in liquid
samples nitrogen) till use
Presence of
Check protocol for interfering
interfering substance
substances; deproteinize samples
in the sample
Improperly thawed Thaw all components completely
Lower/ components and mix gently before use
Higher Allowing reagents to
Always thaw and prepare fresh
readings in sit for extended times
reaction mix before use
samples and on ice
Standards Incorrect incubation Verify correct incubation times
times or temperatures and temperatures in protocol
17. INTERFERENCES
These chemicals or biological materials will cause interference in this
assay causing compromised results or complete failure:
Pyruvate if significant amount of pyruvate is suspected in your
sample, set up a Sample control as described in Section 12.
18. FAQs
What is the kit detecting: ammonium or ammonia?
This kit measures the total ammonia and total ammonium concentration
in your samples.
Will media containing phenol red interfere with the assay reading?
Phenol red could interfere if the amount of medium used is enough to
contribute to add red color to the sample. Typically, when diluted
medium is used and the well volume is made up to 50 L with buffer,
phenol red will not interfere with the assay. However, since the probe
emits at 570 nm and this is in the pink-orange-red zone of the spectrum,
red color from the medium can interfere if undiluted medium is used.
19. NOTES
Austria
Email: wissenschaftlicherdienst@abcam.com | Tel: 019-288-259
France
Email: supportscientifique@abcam.com | Tel: 01-46-94-62-96
Germany
Email: wissenschaftlicherdienst@abcam.com | Tel: 030-896-779-154
Spain
Email: soportecientifico@abcam.com | Tel: 911-146-554
Switzerland
Email: technical@abcam.com
Tel (Deutsch): 0435-016-424 | Tel (Franais): 0615-000-530
Canada
Email: ca.technical@abcam.com | Tel: 877-749-8807
Japan
Email: technical@abcam.co.jp | Tel: +81-(0)3-6231-0940
Copyright 2015 Abcam, All Rights Reserved. The Abcam logo is a registered trademark.