Polarographic Dissolved Oxygen Probes

Instruction Sheet
Probe/Meter Compatibility Chart
Cat. No. Use Meter Model Number Recommended Maintenance Period (See Note 3)
080510 Field, Lab 805, 810 1 to 2 months
081010 Field, Lab 810, 850 1 to 2 months
082010 Field 820 6 months
084010 Field 840 6 months
084025 Field 820, 840 6 months
084050 Field 820, 840 6 months
08405X Field 820, 840 6 months
086010 Lab 860 (see Note 1) 6 months
087010 Lab 870 (see Note 1) 12 months
088010 Lab 880 (see Note 2) 6 months

NOTES:
1. Meter is supplied with the probe
2. Discontinued
3. Maintenance period is dependent on use conditions
Thermo Orion dissolved oxygen probes are polarographic sensors designed for use with specific Thermo Orion 800 series Dissolved
Oxygen Meters. Accessory probes are often chosen to provide the convenience and flexibility of a specific cable length or
compatibility with accessories most appropriate for a particular application.

Use and Maintenance


All probes except the 081010 and 080510 are shipped with electrolyte, and are ready to use without further preparation. The 081010
and the 080510 are shipped dry and have to be filled with electrolyte solution before testing. The probe is continuously polarized
when connected to the instrument. Polarization for a new probe, or after the silver anode is cleaned, or when the probe has been
unplugged for more than an hour will take 25 to 50 minutes. Interrupted connections of less than an hour will take between 5 and 25
minutes
to repolarize.
Recommended maintenance intervals are shown on the chart above. Maintenance consists of cleaning the cathode and anode, and
replacing the electrolyte and membrane cap assembly.

Storage
For short term storage, overnight or between measurements, the probe should remain plugged into the meter, and kept in its
calibration sleeve, calibration beaker or instrument quiver with distilled water. For long-term storage, disconnect the probe from the
meter, remove the membrane cap, and clean and dry the probe.
Refer to your meter operation manual for detailed instructions regarding probe installation, operation, maintenance, and storage.

Thermo Orion 500 Cummings Center Beverly, MA 01915-6199 Tel: 978-232-6000 Fax: 978-232-6015 Int'l. Fax: 978-232-6031

A Thermo Electron business


Formerly Orion Research, Inc.
Dissolved Oxygen Accessories Instruction Sheet
Cat. No. Description Used with
080010 Membrane Cap All 820, 830, 840, 850, 860, 870, and 880 series probes
080013 Probe Service Kit Electrolyte, Cathode All 820, 830, 840, 850, 860, 870, and 880 series probes
cleaning paper, Anode cleaning solution
080113 Probe Service Kit 2 Bottles Electrolyte, All 805 and 810 series probes
Cathode Polishing Disk, 4 Membrane Caps
081003 3 Membrane Caps All 805 and 810 series probes
080015 Calibration Sleeve All 820, 830, 840, 850, 860, 870, and 880 series probes
080016 Precision Calibration Beaker All 820, 860, 870, and 880 series probes
080018 Automatic Calibrating Beaker 087010
080042 Field Stirrer 084025, 084050, 08405X
080082 Overflow Funnel Set All 820, 860, 870, and 880 series probes
080086 BOD Stirring Accessory All 820, 860, 870, and 880 series probes
080160 BOD Stirring Accessory All 810 series probes
080150 Field Case Models 805 and 810 Meters
010018 Line Converter, 110 volt Models 820, 840 Meters
010026 Line Converter, 220 volt Models 820, 840 Meters
020125 Line Converter, 110 volt Models 805, 810 and 850 Meters
020130 Line Converter, 220 volt Models 805, 810 and 850 Meters
010020 Rechargeable Battery Model 820 Meter
010024 Rechargeable Replacement Battery Model 840 Meter
080030 Printer Interface Cable for GLP Printer Model 870 Meter
080031 Printer Interface Cable for 55PR40, 55PR49 Model 870 Meter
080032 Interface Cable for IBM compatible PC Model 870 Meter
0ACBL0 Interface Cable for IBM compatible PC Models 805, 810 and 850 Meters
0900A0 Printer, 110V Models 805, 810 and 850 Meters
0900A1 Printer, 110V Models 805, 810 and 850 Meters
55PR40 Printer (Seiko), 110V Model 870 Meter
55PR49 Printer (Seiko), 220V Model 870 Meter
0GLP02 Printer (GLP), 110V Model 870 Meter
0GLP03 Printer (GLP), 220V Model 870 Meter

Thermo Orion Thermo Orion Europe Thermo Orion Far East Thermo Orion Customer Support
500 Cummings Center 12-16 Sedgeway Business Park Room 904, Federal Building Toll Free: 800-225-1480
Beverly, MA 01915-6199 USA Witchford, Cambridgeshire 369 Lockhart Road www.thermoorion.com
Tel: 978-232-6000 England, CB6 2HY Wanchai, Hong Kong Dom. e-mail: domcs1@thermoorion.com
Dom. Fax: 978-232-6015 Tel: 44-1353-666111 Tel: 852-28360981 Intl. e-mail: intcs1@thermoorion.com
Intl. Fax: 978-232-6031 Fax: 44-1353-666001 Fax: 852-28345160

A Thermo Electron business


Formerly Orion Research, Inc.

241679-001 rev. A

S-ar putea să vă placă și