Documente Academic
Documente Profesional
Documente Cultură
58]
Original Article
Background: To evaluate the agreement of Goldmann Results: The mean age was 13.373.51years. The
applanation tonometer(GAT) with Tono-Pen and mean IOPvalues recorded with NCT; Tono-Pen and
noncontact tonometer(NCT) for measurement of GAT were 14.38, 15.63, and 12.44mmHg, respectively.
intraocular pressure(IOP) in pediatric age group Both Tono-Pen and NCT recorded statistically higher
and to evaluate the correlation between central IOPvalues than the GAT(P=0.00) regardless of the
corneal thickness(CCT) and IOP measured with the CCT. The percentage increase of IOP measured over
tonometers used. GAT was 15.66% for NCT and 25.70% for Tono-Pen
which was also statistically significant. Acorrelation
Materials and Methods: IOP was measured in 200 was found between CCT and IOPvalues obtained with
eyes in a group of Indian children, aged between 8 and all the three tonometers.
18years using three different tonometers: NCT, the
Conclusion: IOP measurements on children vary
Tono-Pen and GAT. All IOP readings were made in the
significantly between instruments and correlations are
office settings by the same examiner. Readings obtained
affected by the corneal thickness. Further studies on
were compared between the instruments and with the
children are needed to determine which instrument is
CCT for each tonometer. Tonometer intermethod most appropriate and to derive a normative IOP scale
agreement was assessed by the BlandAltmann method. for the growing eye.
The relations of CCT with absolute IOPvalues and
intertonometer differences were analyzed by linear Keywords: Central corneal thickness, intraocular
regression. pressure in children, tonometers
DOI: Cite this article as: Raina UK, Rathie N, Gupta A, Gupta SK, Thakar M.
10.4103/0974-620X.176096
Comparison of Goldmann applanation tonometer, Tono-Pen and noncontact
tonometer in children. Oman J Ophthalmol 2016;9:22-6.
the presence of corneal pathology.[11] A significant advantage anesthetic(proparacaine hydrochloride) and fluorescein dye in
of noncontact tonometry(NCT) is the elimination of potential the eye. All measurements were done in the afternoon between 2
hazards associated with all contact tonometers viz., corneal and 3 pm in sitting position and by the same examiner to avoid any
abrasion, reaction to topical anesthetic or flourescein and spread interobserver variation. To avoid any intrainstrument variability,
of infection.[12] However, with NCT, the subject should be able the same tonometers were always used. GAT was calibrated each
to fixate on the target and hence its use is limited in children week, and Tono-Pen calibration was carried out each day prior to
with poor fixation, nystagmus, and inability to see the target. Eyes IOP recording.
with corneal surface irregularities are also not suitable for use of
NCT because the instrument is designed for clear and smooth The mean value of three consecutive measurements by each
corneas.[13] tonometer was used for final analysis. For CCT also, an average
of three successful readings was used. Intermethod agreement
Although accurate and consistent IOP measurements are difficult was assessed, and the influence of CCT on the IOPs measured
to obtain, they are of extreme clinical importance when trying to was evaluated.
diagnose and manage pediatric glaucoma.[14] Moreover, there have
been few studies comparing the accuracy of various tonometers Statistical analysis
conducted in children.[1,3,15,16] The present study compared IOP The data thus obtained was analyzed using SPSS 17 (SPSS Statistics
readings by GAT, Tono-Pen, and NCT to determine which type is for Windows, Version 17.0. Chicago: SPSS Inc.). The difference
best tolerated in clinical practices with children and which is most between the two means was observed by the paired ttest, and a
reliable. In addition, the influence of CCT on the IOPs measured P<0.05 was considered statistically significant. Agreement between
with each tonometer was evaluated. the methods was observed by estimating the limits of agreement and
assessed using Bland and Altmann plots.[17] Correlation between
Materials and Methods the CCT and the IOP measured by various instruments was studied
using Pearsons/Spearmans rank correlation coefficient.
An observational study was performed at Guru Nanak Eye Centre,
Maulana Azad Medical College, NewDelhi from 2012 to 2013. Results
Two hundred eyes of 100 pediatric were included in the study
who presented to either the Outpatient Department or Pediatric Results of 200 eyes of 100 healthy subjects(47 boys and 53
Ophthalmology Clinic. Children included in the study were girls) were included in the study. Mean age of the subjects was
between 8 and 18years of age with the ability to have CCT and 13.373.51(range: 818years).
IOP measurements taken in the office. Children with any anterior
segment dysgenesis, cataract, active inflammation, glaucoma, and Mean IOP recorded was 14.382.35mmHg by NCT,
those with any gross ocular pathology such as microphthalmos, 15.632.78mmHg by Tono-Pen and 12.461.44mmHg by
aniridia, and coloboma were excluded from the study. The study GAT. There was statistically significant difference between the
was reviewed and approved by the Institutes Ethics Committee. IOP measured by the three instruments regardless of the corneal
thickness(P=0.00). The percentage increase over GAT was
A written informed consent was obtained from parents or 15.66% for NCT and 25.70% for Tono-Pen. On an average NCT
guardians of all the children. All children underwent a detailed overestimated IOP readings by GAT by 1.964.16mmHg while
ophthalmologic examination including visual acuity using Tono-Pen overestimated GAT by 3.25.18mmHg within the
Snellens chart, cycloplegic refraction with 2% homatropine, slit IOP range studied. The difference in IOP measurement was more
lamp examination, fundus examination, keratometry, and axial in the higher IOP range when the average IOP was more than
length measurement. CCT was measured using Pac scan plus 14mmHg.
digital biometric ruler(Pacscan 300 AP, Sonomed, Lac Success,
NewYork). The pachymeter probe was placed on the center of the There was good intermethod agreement, i.e.,between
cornea, and three consecutive central corneal pachymetry values GAT and Tono-Pen and GAT and NCT with 95% limits of
were recorded. The mean of the readings also was calculated and agreement falling between 1.98 to 8.38 and 2.2 to 6.12,
recorded. respectively[Figures1 and 2].
IOP measurements were done using noncontact tonometer(NCT Mean CCT was 539.6530.48m(range: 464604m).
10, Shin Nippon, Japan), Tono-Pen(Medtronic Ophthalmic Thicker corneas were associated with higher IOP with all the
Incorporation, Jacksonville, Florida, USA) and GAT(HaagStreit three tonometers used. CCT was found to have a significant
AG, Bern, Switzerland) in that sequence, at an interval of 15min, correlation with IOP measured with Goldmann and NCT while
right eye followed by left. For NCT and Tono-Pen readings, an Tono-Pen readings varied less with a change in corneal thickness.
average of three successive measurements was taken. Before For 100m change in CCT, IOP changed by 1.03mmHg with
the measurement of GAT, the drum was reset to approximately GAT, 2.41mmHg with NCT and only 0.81mmHg with Tono-
10mmHg after each reading, and the biprism was cleaned with Pen[Figures35]. Of the other factors studied such as axial length,
3% hydrogen peroxide. IOP was recorded after instilling topical average keratometry values, and spherical equivalent, none was
14 8
12
10 6 6.12
8 8.38
4
6
4
2
2
0
8 10 12 14 16 18 20 0
-2 8 10 12 14 16 18 20
-4 -1.98
-2
-2.2
-6
-8 -4
Mean IOP (mmHg) Mean IOP (mmHg)
Figure 1: BlandAltman plot for Goldmann applanation tonometer and Tono-Pen Figure 2: BlandAltman plot for Goldmann applanation tonometer and noncontact
tonometry
19 25
23
17
21
15
19
GAT (mm Hg)
13
17
NCT (mm Hg)
11 15
13
9
11
7
9
5
450 470 490 510 530 550 570 590 610 630 7
CCT (m)
5
Figure 3: Correlation of Goldmann applanation tonometer and central corneal thickness
450 470 490 510 530 550 570 590 610 630
CCT (m)
found to have a significant correlation with IOP measured with
Figure 4: Correlation of noncontact tonometry and central corneal thickness
the three instruments.
Conflicts of interest in a sitting and recumbent position of the patientsA clinical study on 251
eyes. Klin Monbl Augenheilkd 2002;219:78590.
There are no conflicts of interest.
11. Frenkel RE, Hong YJ, Shin DH. Comparison of the TonoPen to the
Goldmann applanation tonometer. Arch Ophthalmol 1988;106:7503.
References 12. BuscemiM, CapoferriC, GaravagliaA, NassiveraC, NucciP. Noncontact
tonometry in children. Optom Vis Sci 1991;68:4614.
1. BradfieldYS, KaminskiBM, RepkaMX, Melia M; Pediatric Eye Disease 13. ForbesM, Pico G Jr., GrolmanB. Anoncontact applanation tonometer.
Investigator Group, DavittBV, JohnsonDA, etal. Comparison of TonoPen Description and clinical evaluation. Arch Ophthalmol 1974;91:13440.
and Goldmann applanation tonometers for measurement of intraocular 14. IesterM, MermoudA, AchacheF, RoyS. New TonoPen XL: Comparison
pressure in healthy children. JAAPOS 2012;16:2428. with the Goldmann tonometer. Eye(Lond) 2001;15(Pt 1):528.
2. KassMA. Standardizing the measurement of intraocular pressure 15. KageyamaM, HirookaK, BabaT, ShiragaF. Comparison of ICare rebound
for clinical research. Guidelines from the eye care technology forum. tonometer with noncontact tonometer in healthy children. JGlaucoma
Ophthalmology 1996;103:1835. 2011;20:636.
3. BordonAF, KatsumiO, HiroseT. Tonometry in pediatric patients: A 16. Feng CS, Jin KW, Yi K, Choi DG. Comparison of intraocular pressure
comparative study among TonoPen, Perkins, and Schitz tonometers. measurements obtained by rebound, noncontact, and Goldmann
JPediatr Ophthalmol Strabismus 1995;32:3737. applanation tonometry in children. Am J Ophthalmol 2015;160:93743.e1.
4. MermoudA, BaerveldtG, MincklerDS, LeeMB, RaoNA. Intraocular 17. BlandJM, AltmanDG. Statistical methods for assessing agreement
pressure in Lewis rats. Invest Ophthalmol Vis Sci 1994;35:245560. between two methods of clinical measurement. Lancet 1986;1:30710.
5. MermoudA, BaerveldtG, MincklerDS, LeeMB, RaoNA. Measurement 18. MincklerDS, BaerveldtG, HeuerDK, QuillenThomasB, WalonkerAF,
of rabbit intraocular pressure with the TonoPen. Ophthalmologica WeinerJ. Clinical evaluation of the oculab TonoPen. Am J Ophthalmol
1995;209:2757. 1987;104:16873.
6. AzuaraBlancoA, BhojaniTK, SarhanAR, PillaiCT, DuaHS. TonoPen 19. TonnuPA, HoT, NewsonT, El SheikhA, SharmaK, WhiteE, etal. The
determination of intraocular pressure in patients with band keratopathy or influence of central corneal thickness and age on intraocular pressure
glued cornea. Br J Ophthalmol 1998;82:6346. measured by pneumotonometry, noncontact tonometry, the TonoPenXL,
7. MokKH, WongCS, LeeVW. TonoPen tonometer and corneal thickness. and Goldmann applanation tonometry. Br J Ophthalmol 2005;89:8514.
Eye(Lond) 1999;13(Pt 1):357. 20. ResendeGM, LupinacciAP, rieta CE, CostaVP. Central corneal
8. Mnage MJ, KaufmanPL, CroftMA, LandaySP. Intraocular pressure thickness and intraocular pressure in children undergoing congenital
measurement after penetrating keratoplasty: Minified Goldmann cataract surgery: A prospective, longitudinal study. Br J Ophthalmol
applanation tonometer, pneumatonometer, and TonoPen versus 2012;96:11904.
manometry. Br J Ophthalmol 1994;78:6716. 21. Pediatric Eye Disease Investigator Group, Bradfield YS, Melia BM,
9. CoptRP, ThomasR, MermoudA. Corneal thickness in ocular hypertension, RepkaMX, KaminskiBM, DavittBV, etal. Central corneal thickness in
primary openangle glaucoma, and normal tension glaucoma. Arch children. Arch Ophthalmol 2011;129:11328.
Ophthalmol 1999;117:146. 22. MuirKW, DuncanL, EnyediLB, StinnettSS, FreedmanSF. Central
10. ViestenzA, LausenB, Jnemann AM, MardinCY. Comparison of precision corneal thickness in children: Stability over time. Am J Ophthalmol
of the TonoPen XL with the Goldmann and Draeger applanation tonometer 2006;141:9557.