Sunteți pe pagina 1din 5

[Downloaded free from http://www.ojoonline.org on Monday, March 21, 2016, IP: 146.185.28.

58]

Original Article

Comparison of Goldmann applanation tonometer, Tono-Pen


and noncontact tonometer in children
Usha Kaul Raina, Neha Rathie, Anika Gupta, Shantanu Kumar Gupta, Meenakshi Thakar
Department of Ophthalmology, Guru Nanak Eye Centre, Maulana Azad Medical College, NewDelhi, India

Background: To evaluate the agreement of Goldmann Results: The mean age was 13.373.51years. The
applanation tonometer(GAT) with Tono-Pen and mean IOPvalues recorded with NCT; Tono-Pen and
noncontact tonometer(NCT) for measurement of GAT were 14.38, 15.63, and 12.44mmHg, respectively.
intraocular pressure(IOP) in pediatric age group Both Tono-Pen and NCT recorded statistically higher
and to evaluate the correlation between central IOPvalues than the GAT(P=0.00) regardless of the
corneal thickness(CCT) and IOP measured with the CCT. The percentage increase of IOP measured over
tonometers used. GAT was 15.66% for NCT and 25.70% for Tono-Pen
which was also statistically significant. Acorrelation
Materials and Methods: IOP was measured in 200 was found between CCT and IOPvalues obtained with
eyes in a group of Indian children, aged between 8 and all the three tonometers.
18years using three different tonometers: NCT, the
Conclusion: IOP measurements on children vary
Tono-Pen and GAT. All IOP readings were made in the
significantly between instruments and correlations are
office settings by the same examiner. Readings obtained
affected by the corneal thickness. Further studies on
were compared between the instruments and with the
children are needed to determine which instrument is
CCT for each tonometer. Tonometer intermethod most appropriate and to derive a normative IOP scale
agreement was assessed by the BlandAltmann method. for the growing eye.
The relations of CCT with absolute IOPvalues and
intertonometer differences were analyzed by linear Keywords: Central corneal thickness, intraocular
regression. pressure in children, tonometers

Introduction cooperation.[3] In addition, there are various factors such as the


central corneal thickness(CCT), subjects seated or recumbent
Comprehensive eye care for children and adolescents requires position, eye movements or facial mask expressions, which
measurement of intraocular pressure(IOP). The World Glaucoma influence IOP measurement in children.[49]
Congress wrote that the question whether one tonometer is
superior to the others in pediatric patients is unresolved.[1] The The Tono-Pen tonometer was developed for use in patients
widely accepted international gold standard for IOP measurement who present with the sort of measurement problems that
in adults is the Goldmann applanation tonometer(GAT).[2] are often associated with children. The instrument is easy to
However, measuring IOP with conventional contact tonometers handle, portable, light weight, and does not require the use of
can be difficult in young patients owing to their lack of fluorescein.[10] Tonopen is useful for supine patients and in
Correspondence:
Dr.Anika Gupta, E187, New Rajendar Nagar, NewDelhi110060, India.
Email:anika574@gmail.com This is an open access article distributed under the terms of the Creative
Commons AttributionNonCommercialShareAlike 3.0 License, which allows
Access this article online others to remix, tweak, and build upon the work noncommercially, as long as the
author is credited and the new creations are licensed under the identical terms.
Quick Response Code:
Website:
For reprints contact: reprints@medknow.com
www.ojoonline.org

DOI: Cite this article as: Raina UK, Rathie N, Gupta A, Gupta SK, Thakar M.
10.4103/0974-620X.176096
Comparison of Goldmann applanation tonometer, Tono-Pen and noncontact
tonometer in children. Oman J Ophthalmol 2016;9:22-6.

22 2016 Oman Ophthalmic Society|Published by Wolters KluwerMedknow


[Downloaded free from http://www.ojoonline.org on Monday, March 21, 2016, IP: 146.185.28.58]

Raina, etal.: Tonometry methods in children

the presence of corneal pathology.[11] A significant advantage anesthetic(proparacaine hydrochloride) and fluorescein dye in
of noncontact tonometry(NCT) is the elimination of potential the eye. All measurements were done in the afternoon between 2
hazards associated with all contact tonometers viz., corneal and 3 pm in sitting position and by the same examiner to avoid any
abrasion, reaction to topical anesthetic or flourescein and spread interobserver variation. To avoid any intrainstrument variability,
of infection.[12] However, with NCT, the subject should be able the same tonometers were always used. GAT was calibrated each
to fixate on the target and hence its use is limited in children week, and Tono-Pen calibration was carried out each day prior to
with poor fixation, nystagmus, and inability to see the target. Eyes IOP recording.
with corneal surface irregularities are also not suitable for use of
NCT because the instrument is designed for clear and smooth The mean value of three consecutive measurements by each
corneas.[13] tonometer was used for final analysis. For CCT also, an average
of three successful readings was used. Intermethod agreement
Although accurate and consistent IOP measurements are difficult was assessed, and the influence of CCT on the IOPs measured
to obtain, they are of extreme clinical importance when trying to was evaluated.
diagnose and manage pediatric glaucoma.[14] Moreover, there have
been few studies comparing the accuracy of various tonometers Statistical analysis
conducted in children.[1,3,15,16] The present study compared IOP The data thus obtained was analyzed using SPSS 17 (SPSS Statistics
readings by GAT, Tono-Pen, and NCT to determine which type is for Windows, Version 17.0. Chicago: SPSS Inc.). The difference
best tolerated in clinical practices with children and which is most between the two means was observed by the paired ttest, and a
reliable. In addition, the influence of CCT on the IOPs measured P<0.05 was considered statistically significant. Agreement between
with each tonometer was evaluated. the methods was observed by estimating the limits of agreement and
assessed using Bland and Altmann plots.[17] Correlation between
Materials and Methods the CCT and the IOP measured by various instruments was studied
using Pearsons/Spearmans rank correlation coefficient.
An observational study was performed at Guru Nanak Eye Centre,
Maulana Azad Medical College, NewDelhi from 2012 to 2013. Results
Two hundred eyes of 100 pediatric were included in the study
who presented to either the Outpatient Department or Pediatric Results of 200 eyes of 100 healthy subjects(47 boys and 53
Ophthalmology Clinic. Children included in the study were girls) were included in the study. Mean age of the subjects was
between 8 and 18years of age with the ability to have CCT and 13.373.51(range: 818years).
IOP measurements taken in the office. Children with any anterior
segment dysgenesis, cataract, active inflammation, glaucoma, and Mean IOP recorded was 14.382.35mmHg by NCT,
those with any gross ocular pathology such as microphthalmos, 15.632.78mmHg by Tono-Pen and 12.461.44mmHg by
aniridia, and coloboma were excluded from the study. The study GAT. There was statistically significant difference between the
was reviewed and approved by the Institutes Ethics Committee. IOP measured by the three instruments regardless of the corneal
thickness(P=0.00). The percentage increase over GAT was
A written informed consent was obtained from parents or 15.66% for NCT and 25.70% for Tono-Pen. On an average NCT
guardians of all the children. All children underwent a detailed overestimated IOP readings by GAT by 1.964.16mmHg while
ophthalmologic examination including visual acuity using Tono-Pen overestimated GAT by 3.25.18mmHg within the
Snellens chart, cycloplegic refraction with 2% homatropine, slit IOP range studied. The difference in IOP measurement was more
lamp examination, fundus examination, keratometry, and axial in the higher IOP range when the average IOP was more than
length measurement. CCT was measured using Pac scan plus 14mmHg.
digital biometric ruler(Pacscan 300 AP, Sonomed, Lac Success,
NewYork). The pachymeter probe was placed on the center of the There was good intermethod agreement, i.e.,between
cornea, and three consecutive central corneal pachymetry values GAT and Tono-Pen and GAT and NCT with 95% limits of
were recorded. The mean of the readings also was calculated and agreement falling between 1.98 to 8.38 and 2.2 to 6.12,
recorded. respectively[Figures1 and 2].

IOP measurements were done using noncontact tonometer(NCT Mean CCT was 539.6530.48m(range: 464604m).
10, Shin Nippon, Japan), Tono-Pen(Medtronic Ophthalmic Thicker corneas were associated with higher IOP with all the
Incorporation, Jacksonville, Florida, USA) and GAT(HaagStreit three tonometers used. CCT was found to have a significant
AG, Bern, Switzerland) in that sequence, at an interval of 15min, correlation with IOP measured with Goldmann and NCT while
right eye followed by left. For NCT and Tono-Pen readings, an Tono-Pen readings varied less with a change in corneal thickness.
average of three successive measurements was taken. Before For 100m change in CCT, IOP changed by 1.03mmHg with
the measurement of GAT, the drum was reset to approximately GAT, 2.41mmHg with NCT and only 0.81mmHg with Tono-
10mmHg after each reading, and the biprism was cleaned with Pen[Figures35]. Of the other factors studied such as axial length,
3% hydrogen peroxide. IOP was recorded after instilling topical average keratometry values, and spherical equivalent, none was

Oman Journal of Ophthalmology, Vol. 9, No. 1, 2016 23


[Downloaded free from http://www.ojoonline.org on Monday, March 21, 2016, IP: 146.185.28.58]

Raina, etal.: Tonometry methods in children

14 8

12

IOP difference (NCT-GAT) in mmHg


IOP difference (Tonopen-GAT) in mmHg

10 6 6.12

8 8.38
4
6

4
2
2

0
8 10 12 14 16 18 20 0
-2 8 10 12 14 16 18 20

-4 -1.98
-2
-2.2
-6

-8 -4
Mean IOP (mmHg) Mean IOP (mmHg)

Figure 1: BlandAltman plot for Goldmann applanation tonometer and Tono-Pen Figure 2: BlandAltman plot for Goldmann applanation tonometer and noncontact
tonometry

19 25

23
17

21
15
19
GAT (mm Hg)

13
17
NCT (mm Hg)

11 15

13
9

11
7
9

5
450 470 490 510 530 550 570 590 610 630 7
CCT (m)
5
Figure 3: Correlation of Goldmann applanation tonometer and central corneal thickness
450 470 490 510 530 550 570 590 610 630
CCT (m)
found to have a significant correlation with IOP measured with
Figure 4: Correlation of noncontact tonometry and central corneal thickness
the three instruments.

Studies in the past have determined the agreement between different


Discussion tonometers. Minckler etal.[18] compared GAT with the first generation
Tono-Pen(Tonopen1) in adults and found similar readings in the
For many years, applanation tonometry was considered to set
IOP measured by the two methods. They reported that Tonopen1
a standard against which other instruments were measured.
Measuring IOP in children is more challenging because it requires underestimated the IOP in eyes with higher IOP and overestimated
certain capabilities from the instrument and some degree of the IOP in eyes with lower IOP. Frenkel etal. also compared Tono-Pen
experience from the user. The instrument must provide quick, and GAT in adults and concluded that the Tonopen measures IOP in a
accurate readings. For these reasons, the standard Goldmann manner that corresponds well to the GAT in the 1120mmHg interval,
tonometer, which is mounted on a slit lamp, is not suitable for and fairly well in the 410mmHg and 2130mmHg interval.[11]
children. However, recent reviews that have compared a number
of instruments suggest that depending on the population and the Bradfield etal.[1] in their study in children and adolescents found
application, different instruments may be preferred.[1,3,15,16] that in an office setting, Tono-Pen measurements tended to be

24 Oman Journal of Ophthalmology, Vol. 9, No. 1, 2016


[Downloaded free from http://www.ojoonline.org on Monday, March 21, 2016, IP: 146.185.28.58]

Raina, etal.: Tonometry methods in children

25 IOP measurement is influenced by the CCT.[19] IOP measurements


obtained with GAT tended to be overestimated in eyes with
23 thick corneas and underestimated in eyes with thin corneas in
children.[20] Bradfield etal. found that the GAT measured IOP was
21 1.9mmHg higher for every 100 um increase in CCT in children.[21]
This accords with the results of a prior small pediatric study by
19
Muir etal.[22] They suggested that this small difference in the
Tonopen (mm Hg)

17 effect of CCT may be secondary to the large corneal surface area


contacted by the GAT device. Whereas, Tono-Pen measurements
15 have been found to be less dependent than GAT on the biochemical
properties of cornea.[19] Recently, Feng et al.[16] also found the
13
rebound and noncontact tonometry to overestimate IOP relative
11
to GAT for thicker CCT.

9 In this study also, thicker corneas were associated with higher


IOP recordings with all three tonometers used. CCT was found
7 to have a significant correlation with IOP measured with GAT
and NCT while Tono-Pen readings varied less with a change in
5
450 470 490 510 530 550 570 590 610 630 corneal thickness. These findings are probably because Tono-Pen
CCT (m) applanates a smaller area of cornea compared to GAT and hence,
its recordings are less influenced by the CCT.
Figure 5: Correlation of Tono-Pen and central corneal thickness

To the best of our knowledge, this is the first study of IOP


slightly lower than GAT when IOP was<11mmHg and slightly
measurements on healthy Indian children. It contributes to the
higher than GAT when IOP was>11mmHg. In normal children,
literature on IOP in the young eye and provides a comparative study
average differences between IOP measured by Tono-Pen and GAT
to compare the commonly used tonometers in children along with
were small, although there was substantial testretest variability.
the correlation between CCT and IOP measurements. However, our
Bordon et al.[3] compared the accuracy of IOP measurement
study has a few limitations. First, we enrolled only normal subjects
in pediatric patients afflicted with retinopathy of prematurity
and hence it might not be appropriate to extrapolate this data in
using three tonometers: Perkins, Schiotz, and Tono-Pen. Agood
children with or suspected of glaucoma. Second, in this study,
correlation was found between Tono-Pen and Perkins tonometers,
IOP was measured sequentially with NCT, Tono-Pen, and GAT.
whereas the Schiotz measurements were significantly higher.
As noted by Feng etal., this might contribute to the higher mean
Hence, they suggested that Tono-Pen with a higher correlation
IOPs recorded with NCT and Tono-Pen because after repeated
coefficient can be used reliably to assess IOP in pediatric patients.
IOP measurements the child may feel more comfortable with the
Kageyama etal.[15] compared the ICare rebound tonometer with
NCT in healthy children and concluded that IOP measurements procedure. The application of topial anesthetic for GAT may also
performed using ICare are better tolerated in the pediatric contribute to less anxiety and increased cooperation by the child.
population, as compared with measurements using NCT,
especially in children below the age of 6years. Arecent study by Conclusion
Feng etal. compared rebound tonometry, NCT, and GAT and their
relationships to CCT in children. They suggested that all three Each of the three tonometers evaluated in this study can be
tonometers would be clinically acceptable for pediatric patients routinely used in the clinical setting for IOP measurement in
with IOP within the normal range of values, but children with children. However, eyes with high IOP measurements must be
IOPs within a suspicious range should be reevaluated carefully.[16] carefully evaluated for the effect of CCT and other factors which
might influence IOP readings in children. Tonopen may be used
In our prospective study, we compared GAT, Tono-Pen for children in the presence of corneal pathology, as it is least
and NCT in pediatric eyes. Findings from this study show affected by variations in corneal thickness and NCT may be used
that both Tono-Pen and NCT provided statistically higher in the presence of corneal or other ocular infection when a contact
IOPvalues(15.632.78mmHg and 14.382.35mmHg, procedure is contraindicated. However, further studies for IOP
respectively) than GAT(12.461.44mmHg) regardless of the measurements on large populations of children, with and without
corneal thickness. The difference in the IOP recordings with eye disease or conditions that may affect IOP, from a variety of
different tonometers was more for higher IOP range. This is countries and ethnic groups is required, to find an ideal and
probably because a lesser amount of force is required to applanate accurate tonometer applicable for every age group.
a softer eye and vice versa. This difference is probably exaggerated
in the case of Tono-Pen as its principle combines both indentation Financial support and sponsorship
and applanation. Nil.

Oman Journal of Ophthalmology, Vol. 9, No. 1, 2016 25


[Downloaded free from http://www.ojoonline.org on Monday, March 21, 2016, IP: 146.185.28.58]

Raina, etal.: Tonometry methods in children

Conflicts of interest in a sitting and recumbent position of the patientsA clinical study on 251
eyes. Klin Monbl Augenheilkd 2002;219:78590.
There are no conflicts of interest.
11. Frenkel RE, Hong YJ, Shin DH. Comparison of the TonoPen to the
Goldmann applanation tonometer. Arch Ophthalmol 1988;106:7503.
References 12. BuscemiM, CapoferriC, GaravagliaA, NassiveraC, NucciP. Noncontact
tonometry in children. Optom Vis Sci 1991;68:4614.
1. BradfieldYS, KaminskiBM, RepkaMX, Melia M; Pediatric Eye Disease 13. ForbesM, Pico G Jr., GrolmanB. Anoncontact applanation tonometer.
Investigator Group, DavittBV, JohnsonDA, etal. Comparison of TonoPen Description and clinical evaluation. Arch Ophthalmol 1974;91:13440.
and Goldmann applanation tonometers for measurement of intraocular 14. IesterM, MermoudA, AchacheF, RoyS. New TonoPen XL: Comparison
pressure in healthy children. JAAPOS 2012;16:2428. with the Goldmann tonometer. Eye(Lond) 2001;15(Pt 1):528.
2. KassMA. Standardizing the measurement of intraocular pressure 15. KageyamaM, HirookaK, BabaT, ShiragaF. Comparison of ICare rebound
for clinical research. Guidelines from the eye care technology forum. tonometer with noncontact tonometer in healthy children. JGlaucoma
Ophthalmology 1996;103:1835. 2011;20:636.
3. BordonAF, KatsumiO, HiroseT. Tonometry in pediatric patients: A 16. Feng CS, Jin KW, Yi K, Choi DG. Comparison of intraocular pressure
comparative study among TonoPen, Perkins, and Schitz tonometers. measurements obtained by rebound, noncontact, and Goldmann
JPediatr Ophthalmol Strabismus 1995;32:3737. applanation tonometry in children. Am J Ophthalmol 2015;160:93743.e1.
4. MermoudA, BaerveldtG, MincklerDS, LeeMB, RaoNA. Intraocular 17. BlandJM, AltmanDG. Statistical methods for assessing agreement
pressure in Lewis rats. Invest Ophthalmol Vis Sci 1994;35:245560. between two methods of clinical measurement. Lancet 1986;1:30710.
5. MermoudA, BaerveldtG, MincklerDS, LeeMB, RaoNA. Measurement 18. MincklerDS, BaerveldtG, HeuerDK, QuillenThomasB, WalonkerAF,
of rabbit intraocular pressure with the TonoPen. Ophthalmologica WeinerJ. Clinical evaluation of the oculab TonoPen. Am J Ophthalmol
1995;209:2757. 1987;104:16873.
6. AzuaraBlancoA, BhojaniTK, SarhanAR, PillaiCT, DuaHS. TonoPen 19. TonnuPA, HoT, NewsonT, El SheikhA, SharmaK, WhiteE, etal. The
determination of intraocular pressure in patients with band keratopathy or influence of central corneal thickness and age on intraocular pressure
glued cornea. Br J Ophthalmol 1998;82:6346. measured by pneumotonometry, noncontact tonometry, the TonoPenXL,
7. MokKH, WongCS, LeeVW. TonoPen tonometer and corneal thickness. and Goldmann applanation tonometry. Br J Ophthalmol 2005;89:8514.
Eye(Lond) 1999;13(Pt 1):357. 20. ResendeGM, LupinacciAP, rieta CE, CostaVP. Central corneal
8. Mnage MJ, KaufmanPL, CroftMA, LandaySP. Intraocular pressure thickness and intraocular pressure in children undergoing congenital
measurement after penetrating keratoplasty: Minified Goldmann cataract surgery: A prospective, longitudinal study. Br J Ophthalmol
applanation tonometer, pneumatonometer, and TonoPen versus 2012;96:11904.
manometry. Br J Ophthalmol 1994;78:6716. 21. Pediatric Eye Disease Investigator Group, Bradfield YS, Melia BM,
9. CoptRP, ThomasR, MermoudA. Corneal thickness in ocular hypertension, RepkaMX, KaminskiBM, DavittBV, etal. Central corneal thickness in
primary openangle glaucoma, and normal tension glaucoma. Arch children. Arch Ophthalmol 2011;129:11328.
Ophthalmol 1999;117:146. 22. MuirKW, DuncanL, EnyediLB, StinnettSS, FreedmanSF. Central
10. ViestenzA, LausenB, Jnemann AM, MardinCY. Comparison of precision corneal thickness in children: Stability over time. Am J Ophthalmol
of the TonoPen XL with the Goldmann and Draeger applanation tonometer 2006;141:9557.

26 Oman Journal of Ophthalmology, Vol. 9, No. 1, 2016

S-ar putea să vă placă și