DIGITAL AMMETER

EPM-4D / EPM-4C/ EPM-4A / EPM-4AC


EPM-4D (Digital Ammeter With Demand) :EPM-4D is designed to measure RMS value of AC Setting of Low Setpoint Value (For EPM-4C and 4AC)
current which flows from the line and saving the instant and average demands. The current transformer ratio
can be set between 5/5A and 10000/5A by the buttons on the panel. Settings and demands remain stored in SET Press SET button for 3 sec.
the memory when the power supply is OFF, and the ammeter continues its function by this stored value when
the power supply is ON. Press UP or DOWN button until SP L is
EPM-4A (Digital Ammeter With Demand and Current Transformer): EPM-4A is designed displayed.
to measure RMS value of AC current from 0,07 to 200A via the current transformer connected on the device
via the 0.5m cable. Press SET button.
SET
EPM-4C (Digital Ammeter With Setpoint) :EPM-4C measures the RMS value of AC current which
flows from the line. If the measured AC current is over the high setpoint, or it is under the low setpoint, output Press UP or DOWN button until requested
2500
relay is activated at the adjusted time delay. Also, it includes all functions of EPM-4D. value is displayed. (If pressed
buttons continuously, displayed value is
EPM-4AC (Digital Ammeter With Setpoint and current transformer): EPM-4AC includes
all functions of EPM-4C, but it measures via current transformer connected to the device via the 0.5m cable
changed quickly). 2250
like EPM-4A SET Press SET button to save.
*Important Notice: The secondary value of current transformer must be limited at 5A. Setting of Hysteresis Value (For EPM-4C and 4AC)
Turn ( ) : Turn number of cable, which passes inside of CT-25 is selected in this menu. This number SET Press SET button for 3 sec.
can be selected between 1 and 10.
Demand Time ( ) : Device records measured demand and maximum demand values. Demand is
Press UP or DOWN button until HYS is
hys
the average current value measured in demand time. Maximum demand is the maximum average current value
measured in demand time. Maximum demand value remains stored, even if the power supply is off and demand
displayed.
value becomes 0 (zero).
Setpoint ( s p h ve sp ): When the AC current which flows through the device is over the high setpoint SET Press SET button.
value or it is under the low setpoint value, the output relay is activated in order to generate an error signal, at
Press UP or DOWN button until requested 5.000
the end of the delay time. If the AC current value is under the setting value, before the end of the delay time, value is displayed. (If pressed
the relay resets itself and no tripping occurs. buttons continuously, displayed value is
Latch Function ( ) : In case of overcurrent or undercurrent the output relay is activated. Latch changed quickly).
function is used to select the output relay operation mode. Either OF or On position may be selected.
6.000
At position OF: If the current is below the setting value, output relay releases SET Press SET button to save.
At position On: Even if the current is below the setting value, output relay remains activated and releases Setting of On Time Value (For EPM-4C and 4AC)
only by pressing the set button.
Instant Tripping ( ) : If the AC current is over the1.5 times of setting value or it is under the 0.5 SET Press SET button for 3 sec.
times of setting value, the output relay is activated without any delay time. This function is user-selectable.
Start Time Delay ( ): When the current is start flowing, the setpoint values are ignored for the Press UP or DOWN button until On t is
determined time. If auto start mode is activated, after the AC current is cutted, the device returns into the displayed.
starting state.
SET Press SET button.
Setting of Current Transformer Ratio (For EPM-4C and 4D)
Press UP or DOWN button until requested
value is displayed. (If pressed
SET Press SET button for 3 sec. buttons continuously, displayed value is
changed quickly).
Press UP or DOWN button until Ctr is
displayed. SET Press SET button to save.
Setting of Off Time Value (For EPM-4C and 4AC)
SET Press SET button.
5000 SET Press SET button for 3 sec.
Press UP or DOWN button until requested
value is displayed. (If pressed Press UP or DOWN button until OF t is
buttons continuously, displayed value is displayed.
changed quickly). 2500
SET Press SET button to save. SET Press SET button.
Press UP or DOWN button until requested
Setting of Turn Number (For EPM-4A and 4AC) value is displayed. (If pressed
buttons continuously, displayed value is
SET Press SET button for 3 sec. changed quickly).

SET Press SET button to save.


Press UP or DOWN button until trn is
displayed. Setting of Start Time Value (For EPM-4C and 4AC)
SET Press SET button for 3 sec.
SET Press SET button.
0001 Press UP or DOWN button until Strt is
Press UP or DOWN button until requested displayed.
value is displayed.
0010 SET Press SET button.
SET Press SET button to save. Press UP or DOWN button until requested
value is displayed. (If pressed
Setting of Demand Time (For EPM-4A, 4C, 4D and 4AC) buttons continuously, displayed value is
changed quickly).
SET Press SET button for 3 sec. SET Press SET button to save.
Setting of Auto Start Mode (For EPM-4C and 4AC)
Press UP or DOWN button until dE t is
displayed. SET Press SET button for 3 sec.

SET Press SET button. 0015 Press UP or DOWN button until AutO is
displayed.
Press UP or DOWN button until requested
value is displayed. (If pressed
buttons continuously, displayed value is 0020 SET Press SET button.

SET
changed quickly).
Press SET button to save. Press UP or DOWN button until requested
o
value is displayed.
Setting of High Setpoint Value (For EPM-4C and 4AC)
SET Press SET button to save.
oF
SET Press SET button for 3 sec.
Setting the Instant Tripping Function (For EPM-4C and 4AC)
Press UP or DOWN button until SP H is
displayed. sp h SET Press SET button for 3 sec.

SET Press SET button. 2500 Press UP or DOWN button until trP is
displayed.
Press UP or DOWN button until requested
value is displayed. (If pressed
buttons continuously, displayed value is 2750 SET Press SET button.
changed quickly). o
SET Press SET button to save. Press UP or DOWN button until
requested value is displayed.
oF
SET Press SET button to save..
DIGITAL AMMETER
EPM-4D / EPM-4C/ EPM-4A / EPM-4AC
Setting of Latch Function (For EPM-4C and 4AC) Technical Data
SET Press SET button for 3 sec. Operating Voltage (Un),*
Operating Range (DU) : Please look at labels on the device.
Press UP or DOWN button until LAtC is Operating Frequency (f) : 45-65 Hz
displayed.
Accuracy : 1% 1 digit [(%10-%100) full scale]
Measuring Input (Iin) : 0.05-5.5 A (for EPM-4D, 4C )
SET Press SET button.
o Measuring Range
0,07 -200 A (for EPM-4A, 4AC)
: 0,05 -10000 A with Ct
Press UP or DOWN button until requested
value is displayed. (for EPM-4D, EPM-4C )
oF 0,07-200 A (for EPM-4A, 4AC)
SET Press SET button to save. Current Transformer Ratio (Ct) : 5...10000 / 5A ( for EPM-4D, 4C)
Adjustment Contact Situation (For EPM-4C and 4AC) Power Consumption(Pcons) : <4 VA
Burden : <1 VA
SET Press SET button for 3 sec.
Output Contact : 5A, 250V, 1250VA (Resistive)
Press UP or DOWN button until Out is (for EPM-4C and 4AC)
displayed. Ins. Tripping : >1.5 x SP H or <0.5 x SP L
(for EPM-4C and 4AC)
SET Press SET button.
c Hysteresis
Delay Times
: 0-0.5 x Full scale (for EPM-4C, 4AC)
: 0.0 - 999.9 sec. (for EPM-4C, 4AC)
Press UP or DOWN button until requested
value is displayed. Demand Time (Average) :1-60 min. (for EPM-4A, 4C, 4D and 4AC )
o Enclosure : Non-flammable
SET Press SET button to save. Equipment Protection : Double Insulation ( ),
Measuring Category III
Escaping the Set Menu (For EPM-4A, 4C, 4AC and 4D) Ambient Temperature
O
: -5 C; +50 C
O

If you do not request to change any Degree of Protection : IP 40 (Front Panel)


adjustment in the menu; Wire Thickness : 2.5mm2 (for terminal block)
Press UP or DOWN button until ESC is 16 mm2 (Measuring Input Wire Thickness
displayed.
for EPM-4A, 4AC)
SET Press SET button.
Installations : Flush mounting with rear terminals
Displayed Average Demand (For EPM-4C, 4AC, 4D and 4A ) Dimensions : Type PR 18, PR 19, PR 20
Weight : 0.28 kg (for PR 18)
Press and hold the DOWNbutton to see
the Average demand. 0.30 kg (for PR 19)
0.25 kg (for PR 20)
Displayed Peak Value (For EPM-4C, 4AC, 4D and 4A) 0.31 kg (for EPM-4A and 4AC in PR 18)

Press and hold the UP button to see Packaging Information


the peak value
Pcs per Package : 16 Pcs (for PR 18)
Deleting the Demand and Peak Value Package Weight : 4.5 kg
(For EPM-4C, 4AC, 4D and 4A) Pcs per Package : 16 Pcs (for EPM-4A and 4AC in PR 18)
Package Weight : 5 kg
Press UP and DOWN buttons at the same time Pcs per Package : 12 Pcs (for PR 19)
Package Weight : 3.6 kg
PRECAUTIONS FOR INSTALLATION AND SAFE USE Pcs per Package : 20 Pcs (for PR 20)
Failure to follow those instructions will result in death or serious injury. Package Weight : 5 kg
Disconnect all power before working on equipment.
* Please check the device label for proper value.
When the device is connected to the network, do not remove the front panel. *Different supply voltages are adjustable upon request.
Do not try to clean the device with solvent or the like. Only clean the device with a dried
Dimensions
cloth.
Verify correct terminal connections when wiring.
96mm

Electrical equipment should be serviced only by your compedent seller. No responsibility


72mm

48mm

is assured by the manufacturer or any of its subsidiaries for any consequences arising out
of the use of this material.
Only for rack panel mounting. 72mm 96mm 96mm
TYPE PR 18(72x72) TYPE PR 19(96x96) TYPE PR 20(48x96)
Connection Diagrams
The Area Measurements on The Control Panel
Wall Wall Wall
EPM-4A EPM-4AC
79.3mm
82.5mm 70mm
75mm 63mm
4 3 2 1 6 5 4 3 2 1 54 mm
42mm
65.5mm

90mm

Output
43mm
1A

Contact
1A

1,5m 1,5m

L L
N N TYPE PR 18 TYPE PR 19 TYPE PR 20

Measuring Measuring
Input (Iin) Input (Iin)
50mm 50mm 50mm

Control Panel Cut Out


92mm
68mm

43mm

EPM-4D EPM-4C
68mm 92mm 90,5mm

4 3 2 1 6 5 4 3 2 1 TYPE PR 18(72x72) TYPE PR 19(96x96) TYPE PR 20(48x96)

Output
Measuring l
in lin
Factory Settings :
1A

Contact
1A

Input
k l k l
L Measuring L Ctr = 0005 SP L = 0.250 (EPM-4C) Strt = 006.0
K L Input K L
N trn = 0001 HYS = 4.000 (EPM-4AC) AutO = OF
Warning : dE t = 0015 HYS = 0.100 (EPM-4C) trP = OF
SP H = 160.0 (EPM-4AC) On t = 010.0 LAtC = OF
a) A switch or circuit breaker must be connected between the network SP L = 10.00 (EPM-4AC) OF t = 010.0 Out = nO
and the auxiliary supply input of device. SP H = 4.000 (EPM-4C)
b) Connected switch or circuit breaker must be in close proximity to the device.
c) Connected switch or circuit breaker must be marked as the disconnecting
device for the equipment.
d) The type of the used fuse must be FF type and the current of the used fuse
must be 1A.
e) No need of a vantilator in the installation area

A3359/Rev-2