Documente Academic
Documente Profesional
Documente Cultură
INSTRUMENTS
Bulletin 2700-E
Model
Classification Measuring Range (Accuracy) Page
Number
0.1 1 10 100V
2707
Galvanometer 4
2708
2752 (0.3%) 7
Double bridge
2769 6
Standard
2792A 8
resistor
279301 (0.01%)
10
279303 (0.05%)
2791
270710 2708
2768
275597 276910
182226128mm 2kg 182226128mm 2.6kg
(7-1/494-3/4 4.4lbs) (7-1/494-3/4 5.7lbs)
2752
2753
2754
Accuracy 40% higher than our previous models! The 2972A Series of standard resistors are metal foil resistors,
Temperature coefficient down 2/3! while the previous 2972 Series were winding resistors. The
error range of the 2972A Series is much smaller than those of
Main Features the previous 2972 Series, as demonstrated by the resistance
Traced to the national standard for high accuracy; test temperature coefficient curves (Graph 1). Precision temperature
(calibrated) accuracy of 5 ppm control equipment such as oil baths, are not needed for
Resistance temperature coefficient measurement and calibration. The 2972A can do these in an air
Excellent temperature characteristics in the range of 0 -50C; environment.
resistance temperature coefficient less than 2.5 ppm/C At development sites, the 2972A serves many purposes,
A variety of models ranging from precision measurement to calibration of
Eight models with nominal resistance values ranging between equipment. The main body of the 2972A employs the same
0.001 and 10 k plastic case as that of the previous series, and the 2972A has
Precision temperature control equipment, such as an oil bath, a damage-resistant and easy-to-connect terminal block. The
not needed for calibration due to marked improvement in 2972A can be even more effective in precision measurement
resistance temperature coefficient when it is combined with our potentiometer and double bridge.
Included document: Test certificate
2792A Series
100
0
35 40
0 5 10 15 20 25 30
-100
Error (ppm)
-200
-300
2792 Series
-400
OPERATING PRECAUTIONS
Temperature Current
To maintain a high level of measurement accuracy, the When electric current flows through the Standard Resistor, its
Standard Resistor must be used in rooms with a constant resistance value changes due to the generated heat. To ensure
ambient temperature or placed in a thermostatic chamber. accurate measurement, use the Standard Resistor below the
Measure the ambient temperature accurately and perform maximum allowable current, which is one of the standard test
compensation according to the temperature coefficient. conditions. When the applied current exceeds the maximum
Ensure that the instrument is not subjected to rapid changes allowable current, the resistance could change, or the internal
in temperature during storage as changes in resistance or circuit could be broken.
deterioration of the insulator may result. (Instrument can be
operated stably for over two hours.) Thermal emf and Contact Resistance
To minimize the effects of thermal emf, connect the circuit
Temperature coefficients and resistance with copper wire, and take the average of two readings obtained
The resistance at temperature t is expressed by the following by reversing the direction of the current flowing through the
equation: current terminals. Incomplete contact between the lead wire and
Rt = R23 {1 + 23(t -23) + (t -23)2} the terminals causes current change due to contact resistance,
where and temperature errors due to heat generation.
Rt : Resistance value at tC
R23 : Resistance value at 23C Mechanical Shock and Vibration
23: 1st temperature coefficient at 23C Mechanical shock and vibration may cause distortion of the
: 2nd temperature coefficient at 0 to 50C resistive elements, which results in instability and changes in the
resistance value.
279301
10
2786 2791
11
YOKOGAWA METERS & INSTRUMENTS CORPORATION Tachihi Bld. No.2, 6-1-3 Sakaecho, Tachikawa-shi, Tokyo, 190-8586 Japan
International Sales Dept. Phone: +81-42-534-1413 Facsimile: +81-42-534-1426 Represented by:
Subject to change without notice. [Ed: 03/b] Printed in Japan: Aug. 2011(C)/1,000(KP)