Sunteți pe pagina 1din 2

INSTRUMENTATION AND CONTROL Process Connection: Process connection fittings includes all

TOPIC: TEMPERATURE standard fittings used with thermocouples (i.e.


compression, welded, spring-loaded, etc.)
TEMPERATURE measure of the intensity or degree of
hotness in a body. Wire configuration: RTDs are available in 2,3 and 4 wire
HEAT measure of the quantity of heat energy present in configuration. 3 wire configurations are the most common
a body. for industrial applications. Teflon and fiber glass are the
standard wire insulation materials. Teflon is moisture
TEMPERATURE SCALE resistant and can be used up to 400 F. Fiberglass can be
FAHRENHEIT (Gabriel Fahrenheit, 1714) used up to 1000 F.
Melting Point: 32
Boiling Point: 212 Cold end termination: RTDs can terminate on the cold
Absolute Zero: -460 ends with plugs, bare wires, terminal heads and any of the
CELSIUS (Anders Celsius, 1742) reference junctions common to thermocouples.
Melting Point: 0
Boiling Point: 100 Transmitter: Isolate, amplify, filter noise, linearize, and
Absolute Zero: 273 convert the input signal from the sensor then send
KELVIN (Lord kelvin, 1848) (transmit) a standardized output signal to the control
Often used by scientist because its temperature device.
reading begins at absolute zero and due to the fact
that this scale is proportional to the amount of DIAGRAMS
heat energy found in an object.
Melting Point: 273 Two wire Configuration
Boiling Point: 373 The simplest resistance-thermometer
configuration uses two wires. It is only used when high
TEMPERATURE MEASUREMENTS accuracy is not required, as the resistance of the connecting
NON-ELECTRICAL METHOD wire is added to that of the sensor, leading to errors of
ELECTRICAL METHOD measurement. This applies equality to balanced bridge and
RADIATION METHOD fixed bridge systems.

DEVICES USED IN MEASURING TEMPERATURE Three wire Configuration


Thermocouple In order to minimize the effects of the lead
Thermistors resistance, a three wire configuration can be used. In this
Resistance Temperature Detector (RTD) method the two leads to the sensor are on adjoining arms.
Pyrometer There is a lead resistance in each arm of the bridge, so that
Langmuir Probes (for electron temperature of a the resistance is cancelled out if the two lead resistance are
plasma) accurately the same.
Infrared
Four-wire Configuration
The four-wire resistance thermometer
INSTRUMENTATION AND CONTROL EQUIPMENT
configuration increases the accuracy and reliability of the
Resistance Thermometer
resistance being measured: the resistance error due to lead
A device used to determine the temperature by
wire resistance is zero. In the diagram above a standard
measuring the resistance of pure electrical wire. It works
two-terminal RTD is used with another pair of wires to
by changing resistance with a change in temperature in a
form an additional loop that cancels out the lead
repeatable manner. It is also called Resistance Temperature
resistance.
Detector (RTD).
MEASUREMENT
COMPONENTS
A resistance thermometer or RTD operates on the
measurement principle that a materials electrical resistance
changes with temperature. The relationship between an
RTDs resistance and the surrounding temperature is highly
predictable, allowing for accurate and consistent
temperature measurement. By supplying an RTD with a
constant current and measuring the resulting voltage drop
across the resistor, the RTDs resistance can be calculated
and the temperature can be determined.
Resistance Element: This is the actual temperature setting
MEASUREMENT AND TESTING ERRORS
portion of the RTD commonly platinum. Elements range
The measurement and testing errors on the resistance
in length from 1/8 to 3. There are many options. The
thermometer are due to the following sources/error
standard temperature coefficient is an alpha of .00385 and
sources:
the standard resistance is 100 ohms at 0 C.
Lead Wire Resistance
Outside Diameter: The most common outside diameter is Insulation Resistance
in the US or 6mm (.236) for non-US application. Corrosion and Contamination
However, outside diameters range from .063 to .500. Shock and Vibration
Self-Heating
Tubing Material: Stainless steel is commonly used for
assemblies up to 500 F.

PRESENTORS: ABARCA & SENTIN-EE5


ELECTRIC SET-UP representing the absolute zero point, whereas zero degrees
In an electric set-up of a four-wire RTD bridge. If Celsius represents the freezing point of water. Absolute
wires are matched properly in terms of length and cross temperature must be specified in Kelvins calculation, but
section area, then their impedance effects will cancel for the difference between two temperatures, Kelvin and
because each wire is in opposite position. So that the degrees Celsius are interchangeable. Temperatures
dummy wire acts as a sense lead to measure the voltage specified in Kelvin are never negative, and may be
drop across the RTD resistance and carries no current. In translated to and from degrees Celsius by subtracting and
this circuit, the output voltage is directly proportional to adding the absolute temperature of the freezing point of
the temperature. So, we need one calibration equation to water, 273.15K (alternatively, absolute zero in degrees
find the temperature. Celsius at -273.15C):

MAGNETIC SET-UP 0C = 273.15K


The resistance thermometer is used as a detecting 0K = -273.15C
element and is mounted by a built-in magnet to an object.
The temperature of a place, where a magnet is absorbed TELE-METERING
of refrigerators and motors can be measured. The resistance thermometer can give output for
remote indication and monitoring with the use of a signal
MEASUREMENT SYSTEM converter that can be connected to the remote indicator
Resistance thermometers are supplied with 1,2 or and SCADA system.
sometimes with 3 measuring resistors. The measuring point Temperature sensors are used to collect
is isolated. The connection between resistance temperature constantly and to send it to the
thermometer and measuring instrument is mostly of the 2- microcontroller which is interfaced with the PC. If the
wire type. To obtain accurate measurements, the 3-wire temperature exceeds the predefined temperature limits,
technique is used. If highly accurate measurements are then the microcontroller sends to the relay driver which in
desired, the 4-wire technique with constant current and turn activates the relay to turn the lamps on or off. If this
high-resistance voltage pick-up are used. system fails, then an AV alarm generates from the PC using
high-limit and low-limit features.
CONDITIONING Hence, the mechanism of industries can be
Resistance thermometer must be brought into controlled more efficiently and safely by using SCADA,
contact with the medium to be measured in the best which is more economical and time saving technology.
possible manner. To avoid thermal conduction errors, the
immersion depth should be: CONTROL CONFIGURATIONS
In fluids 6.8 times greater All RTDs usually come in a red and black or red
In gasses 10.15 times greater and white wire-color combination. The red wire is the
Than the protective tube diameter. If only very short excitation wire and the black or white wire are ground
insertion lengths are possible, special designs should be wires. If you are not sure which wire are connected to
applied. It is often helpful to install in a pipe bend, which side of the resistive element, you can use a digital
whereby the protective tube must be positioned against multimeter (DMM) to measure the resistance between the
the flow of the medium. leads. If there is close to 0-ohm resistance, then the lead is
attached to the same node. If the resistance is close to the
LOGGING AND RECORDING nominal gage resistance (100 ohms) then the wire is in the
Temperature logging and recording can be done opposite side of the resistive element.
in two different ways. The Analog recording and the RTD is a passive measurement device, therefore, it
Digital Recording. must be supplied with an excitation current.
In analog recording the transmitted spikes per
minute are converted to a voltage by counting circuit. This Connect the red RTD lead to the excitation positive.
voltage is recorded on a pen-and-ink strip chart recorded Place a jumper from the excitation positive pin to the
as the temperature trace. In Digital recording, the spikes channel positive on the data acquisition device. Connect
per minute from the logging cable are counted digitally at the black (white) RTD lead to the excitation negative.
the surface, and the resulting count rate is converted to a Place a jumper from the excitation negative to the channel
temperature trace by the computers program. negative on the data acquisition device.

DATA ACQUISITION AND RETRIEVAL OF DIGITAL ALGORITHMS


DATA
The data from RTD reader is transmitted via serial
port to the computer using a LabVIEW virtual instrument
application. The data that have been collected are directed
to a specified path file of the application and are being Calibration equation. Commented [C1]: If we know the values of Vs and Vo,
saved at Microsoft excel. we can find Rg and then we can find the temperature
value using calibration equation.
MEASUREMENT AND CONTROL OF NON-ELECTRICAL Get the expression for Rg.;
QUANTITIES Commented [C2]: If R3=Rg; then Vo=0 and the
All metals increase in resistance with temperature, bridge is balanced. We can now solve the equation 3 to
so their resistance change can be used to measure It assumes; RL=0 get the expression for Rg.
temperature. The resistance of metals is almost directly Suppose, if RL is present
proportional to absolute temperature which is a measure
Commented [C3]: Suppose, if RL is present in a
of the average density of heat in an object in relation to
situation, then the expression Rg becomes,
the absolute zero point at which all molecular motion
stops. So,
Absolute temperature is measured in Kelvin, a unit there is an error in the RTD resistance value because of the
of equal size to degrees Celsius, but with zero kelvin RL resistance.
PRESENTORS: ABARCA & SENTIN-EE5

S-ar putea să vă placă și