Documente Academic
Documente Profesional
Documente Cultură
1 INTRODUCTION .......................................................................................................................................................... 3
1.1 PRODUCT PRESENTATION .................................................................................................................................................................. 3
1.2 OPERATING PRINCIPLE ...................................................................................................................................................................... 3
1.3 MAIN CHARACTERISTICS.................................................................................................................................................................... 5
1.4 OVERALL DIMENSIONS (IN MM) .......................................................................................................................................................... 7
1.5 MAIN COMPONENTS ........................................................................................................................................................................ 7
2 SYNOPTIC OF THE PRODUCT ...................................................................................................................................... 8
2.1 HARDWARE ARCHITECTURE ............................................................................................................................................................... 8
2.2 METROLOGICAL SEALING SCHEME ....................................................................................................................................................... 8
2.3 CUSTOMER SEALING......................................................................................................................................................................... 9
2.4 METROLOGICAL LABEL .................................................................................................................................................................... 10
3 METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS .....................................................................................11
3.1 TEMPERATURE MEASUREMENT ......................................................................................................................................................... 11
3.2 PRESSURE MEASUREMENT ............................................................................................................................................................... 11
3.3 COMPRESSIBILITY RATIO CALCULATION ............................................................................................................................................... 12
3.4 CONVERSION FACTOR CALCULATION .................................................................................................................................................. 13
3.5 VOLUMES METERING ..................................................................................................................................................................... 13
5 COMMUNICATION ....................................................................................................................................................19
5.1 OPTICAL PORT .............................................................................................................................................................................. 19
5.2 RS232 PORT ............................................................................................................................................................................... 19
5.3 INTERNAL ATEX PSTN MODEM (OPTIONAL BOARD) ............................................................................................................................. 20
5.4 2-CHANNEL RS485 (OPTIONAL BOARD) ............................................................................................................................................ 20
1
9.2 DISPLAY ...................................................................................................................................................................................... 43
10 SAFETY INSTRUCTIONS ..............................................................................................................................................53
10.1 GENERAL .................................................................................................................................................................................... 53
10.2 MARKING RELATIVE TO ATEX - ZONE ................................................................................................................................................ 53
10.3 SPECIFIC INSTRUCTIONS .................................................................................................................................................................. 53
11 MANUFACTURING .....................................................................................................................................................55
2
1 INTRODUCTION
Corus is a type 1 gas volume converter that can be configured as a T, PT or PTZ version.
It is designed to calculate gas volumes under base conditions, based on a volume at metering conditions, a
temperature, a pressure and a compressibility factor. Its volume input is a low frequency one that can be
connected to any Ex compliant gas meter that provides such information.
Corus belongs to a new generation of volume converters that are designed to achieve a high level of
performance thanks to powerful electronic equipment. Based on the use of flash memory, its architecture
allows the download of a new firmware release through a laptop without board modification.
Corus is approved according the new European directive (MID) based on standard EN 12405-1/A1 and can
be used for commercial and fiscal gas transactions. According to the European regulation, it bears the CE
marking and complies with the following directives:
The gas meter measures a volume of gas at metering conditions (Vm), under specific pressure (Pm) and
temperature (Tm) conditions. This volume is converted into a volume at base conditions by Corus according
to EN 12405:
Pm Tb Zb
Vb Vm C Vm
Pb Tm Zm
where :
Vm = Volume at measuring conditions
Vb = Converted volume at reference (base) conditions
Tm = Gas temperature at metering conditions
3
Tb = Reference (base) absolute temperature
Pm = Gas absolute pressure at metering conditions
Pb = Reference (base) absolute pressure
Zm = Gas compressibility factor at metering conditions
Zb = Gas compressibility factor at reference (base) conditions
C = Conversion factor
4
1.3 Main characteristics
Name CORUS
Manufacturer ITRON
Type Gas volume conversion device - Type 1
Model T, PT or PTZ
Metrology Conform to EN12405-1/A1
MID test certificate: T10323 by NMI (module B)
MID production approval by PTB (module D)
PTZ model : Z computed according :
S-GERG 88
AGANX19
AGANX19 mod.
Table of Z
AGA8 Gross method 2 (GM2)
AGA8 Detailed method (DM)
Ambient temperature range [-25C ; +55C]
Gas temperature range [- 30C ; +70C]
Reference conditions Pb : from 0,9 bar to 2 bar
Tb : from 0C to 40C
Protection class IP65 according EN60529
Power Supply Internal battery (16.5A.h) or external DC voltage
Internal power supply Limits : 2.6V DC < Vint < 3.7V DC
Nominal voltage : 3.3V DC
External power supply Limits : 6 V DC < Vext < 12 V DC
Nominal voltage : 10V DC
Pressure ranges Number : 3
Sensor : 0.9 10 bars abs.
Sensor : 3 30 bars abs.
Sensor : 7.2 80 bars abs.
Pressure sensor type External Piezo-resistive sensor
Temperature probe Platinum probe - PT1000 class A according EN60751
Metering input (volume) Low frequency type (2Hz max.)
Cables length LF volume : 0.8 m or 2.5 m
5
Pressure : 0.8 m or 2.5 m
Temperature : 0.8 m or 2.5 m
CE marking Conformity to :
- EMC directive (89/336/EEC)
- ATEX directive (94/9/EC)
- MID directive (2004/22/EC)
Electromagnetic class E2
Mechanical class M2
Humidity Designed for non-condensing humidity
ATEX classification II1G ia IIC T4
Enclosure material Poly-carbonate
Enclosure dimensions (cm) 22.2 x 14.5 x 8.6
Total weight (approx.) 1.5 kg
Display Graphic
Keyboard 5 keys
Communication ports Standard : RS232 & Optical
Option : RS485 2 channels or PSTN modem
Communication protocols IFLAG (based on IEC 62056-21 also know as EN61107)
Modbus RTU
Slave mode for both protocols
Nature of measured gas Fuel gases of the first and second family according EN437
Units of main data Pressure (absolute): bar, Psi, kPa
Temperature : C , F
Volume : m3, Cft
Firmware Compliant with WELMEC 7.2 guide according following modules:
- P : Embedded software
- I : Instrument specific requirements (converter)
- S : Software separation
--------------------------------------------------------------------
2 separate micro-controllers (C) :
M16C C (CPU board)
Metrological kernel
Non-metrological application
XEMICS C (I/O board)
Metrological firmware (analogue measurements)
6
1.4 Overall dimensions (in mm)
A
m
0
5
2
EULAV
rab 8 9 6 .5 1 = P
C 5 . 3 1 = T
5 3 00 . 1 = 0Z/Z Customer
3
4 5 46 . 5 = C Switch
h /3 m 6 5 .4 5 5 1 = mQ
h / mN 6 9 .3 4 8 1 = bQ
Metrology Shield
1 2 3 4 5 6 1 2 3 125mA
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5
1 2 3 4 5 6 1 2 3 4
Keyboard C
Programming SRAM Extension Slot
Connection Connector Memory (J3 Connector)
Flash
Memory
Optional board
Fixing
7
2 SYNOPTIC OF THE PRODUCT
8
The standard sealing is made up of two plates that cover respectively the CPU board and the metrological
part of the I/O board. These plates can be sealed as indicated below:
It is possible to place on the top-right corner and/or on the bottom-right corner of the box a non-
metrological seal (called customer seal) to prevent any access to the inside of the enclosure of Corus. Any
attempt to open the box lets permanent visible marks on the seal or on the enclosure.
This seal notably protects the access to:
the non-metrological interfaces
the customer switch that is used to configure the access to customer data (see 6.4)
the battery
1: customer seals
9
2.4 Metrological label
A label is provided on the top of the product for the metrological marking and the CE marking. This label is
inviolable: it cannot be unstuck without being destroyed. This label provides the following data:
product name
name of the manufacturer (Itron)
serial number of the device
manufacturing date
ambient temperature range
operating gas pressure range
gas temperature range
reference pressure
reference temperature
data related to MID metrological approval
data required by the potentially explosive atmospheres regulation
IP protection level
CE marking
Manufacturing year
10
3 METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS
Corus is provided with a four-wire Pt1000 temperature sensor that is in accordance with IEC/EN 60751
(1000resistance at 0C). Its four-wire mounting provides a high accuracy by eliminating the resistance of
the cable.
The temperature sensor has an IP67 protection level according to EN 60529 and is delivered with a cable
length of either 2.50m or 0.80m. Its operating range (gas temperature range) is [-30C ; +70C].
The temperature measurement period is programmable from 1 to 30 seconds. When the display is On, this
period is forced to 1 second.
Whenever the temperature measurement goes beyond the operating range defined by two minimal and
maximal temperature thresholds or whenever a temperature sensor failure is detected, a temperature
metrological alarm is generated (see 8.2.2).
3.2.1 T version
When used as a T converter, Corus uses a fixed pressure that is calculated according to the following
formula:
pressure 1.01325 bara gas.rel.press 1.202 104 bar/m altitude
where:
pressure = fixed pressure used for the T conversion
gas.rel.press = gas fixed relative pressure to be programmed into the device
altitude = altitude of the installation to be programmed into the device
When used as a PT or PTZ converter, Corus is provided with an external absolute piezo-resistive pressure
sensor providing a rangeability higher than 11. Three (3) operating pressure ranges are available:
0.9 bara to 10 bara
3 bara to 30 bara
7.2 bara to 80 bara
The sensor is enclosed into a stainless steel enclosure and has an IP66 protection level according to EN
60529. It can withstand an overpressure of 1.25 times its maximal pressure (12.5, 37.5 or 100 bara) for 30
minutes. It is equipped with a G1/4 (BSP) male terminal that allows its direct connection to the gas meter or
to the gas pipe. It is delivered with a cable length of either 2.50m or 0.80m.
All pressure sensors are delivered with their own characterization coefficients (12) that must be programmed
into the converter so that the accuracy required by EN 12405-1/A1 can be reached over the whole pressure
and temperature range.
11
The pressure measurement period is programmable from 1 to 30 seconds. When the display is On, this
period is forced to 1 second.
Whenever the pressure measurement goes beyond the operating range defined by two minimal and
maximal pressure thresholds or whenever a pressure sensor failure is detected, a pressure metrological
alarm is generated (see 8.2.1).
Corus can be programmed to display either Zb/Zm or Zm/Zb (this choice is a metrological parameter).
When configured as a T or PT converter, Corus uses a constant value as compressibility factor. This value
(Zb/Zm or Zm/Zb fixed ratio) can be set by the user.
12
According the selected formula, gas composition has to be compatible with the following ranges:
S-GERG: AGANX19 (Std or mod.): AGA8 (Gross method):
0% < %CO2 < 30% 0% < %CO2 < 30% 0% < %CO2 < 30%
0% < %H2 < 10% 0% < %N2 < 50% 0% < %N2 < 50%
0.55 < SG < 0.9 0.55 < SG < 0.9 0.55 < SG < 0.9
5.27 < Cv (kWh) < 13.33
Table of Z:
The 16 coefficients are provided by Itron according the required gas.
Corus manages two volume indexes and two under alarm volume counters:
an unconverted index
a converted index
an under alarm unconverted counter
a total converted counter
The two indexes are available in the 'INDEX' menu of the display, whereas the under alarm unconverted
counter and the total converted counter can be found in the 'ALARM' menu (see 9.2).
13
3.5.1 Main LF input and unconverted volumes
Corus features a low frequency (LF) input that can work up to 2 Hertz and that is used to meter the
unconverted volumes.
The pulses sent by the meter on this input (called main LF input or LF1) are balanced by the input pulse
weight and are added to the unconverted index. The input pulse weight can be chosen among the following
values:
0.001 m3/pulse
0.01 m3/pulse
0.1 m3/pulse
1 m3/pulse
10 m3/pulse
100 m3/pulse
The LF input can be associated to any gas meter that outputs LF signal or to any device that provides an
electrically compliant interface:
dry contact
static relay
open collector or open drain output
For its connection to the meter, it shares the same terminals (J3 and J4) and the same cable as the
coherence LF input (see 4.1) and as the tamper input (see 4.2). Corus can be delivered with several
different LF cables, in order to comply with all existing types of meter connections. All cables are available
with length of 0.80m or 2.50m.
After every conversion, Corus updates its converted volumes: the converted index and the total converted
counter are increased by the unconverted volume metered since the last conversion balanced by the
conversion factor.
14
normal operation operation under alarm
under alarm
Stopped Incremented
unconverted counter
Incremented with back-up
total converted counter Incremented
values
metering
15
4 NON-METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS
Corus manages an on/off tamper input that allows to detect any magnetic violation attempt on meters that
are equipped with an anti-tampering reed contact. This input shares the same terminals (J3 and J4) and the
same cable as the main LF input.
Tamper input can be set to one of the following modes:
disabled
normally open (NO)
normally closed (NC)
When the input is enabled and not in its normal state, the device generates a tamper alarm that is not
metrological and that does therefore not impact the management of the unconverted and converted
indexes.
Corus features a second LF input called coherence LF input or LF2. This input shares the same terminals
(J3 and J4) and the same cable as the main LF input.
It allows checking of good operation of the metering function and of the pulse emitter of the meter. If this
function is activated, Corus constantly compares the number of pulses metered on both LF inputs and
generates a coherence alarm as soon as the spread between both inputs exceeds a programmable threshold
within a programmable period of time. The coherence LF input and the coherence alarm are not
metrological: they have no effect on the unconverted and converted indexes.
Corus features two on/off inputs that can be connected to different types of intrinsically safe interfaces:
reed contact
static relay
open collector or open drain output
The field of use of these inputs includes the connection to a safety valve to a delta pressure filter status or
to a station door opening contact.
On/Off inputs can be set to one of the following modes:
disabled
normally open (NO)
normally closed (NC)
When any on/off input is enabled and not in its normal state, the device generates a non-metrological
alarm.
16
4.4 On/off outputs
When configured as unconverted or converted volume or total converted counter pulse outputs, on/off
outputs behave like the LF outputs of a meter:
they are individually characterized by a pulse weight that can be configured to any value greater than or
equal to the input pulse weight (see 3.5.1) among 0.001 m3/pulse, 0.01 m3/pulse, 0.1 m3/pulse, 1
m3/pulse, 10 m3/pulse, 100 m3/pulse
they generate a pulse every time the corresponding (unconverted or converted) index gets increased by a
value equal to the selected output pulse weight.
The closing time of the volume pulse outputs can be programmed (default value is 250ms). Maximal
frequency of outputs is 2Hz.
Base volume outputs are automatically disabled as long as any metrological alarm is active.
Each alarm can be individually configured to activate the alarm outputs when triggered. The activation
duration of the alarm outputs can be selected. The alarm outputs can remain active:
either as long as any alarm is active (standard mode)
or as long as any alarm is active with a duration limit of 30 minutes (limited mode)
The On/Off outputs of CORUS can be configured as 4/20mA outputs. This option requires the use of an
external frequency to current converter (model KFU8-UFC-EX1D from Pepperl&Fuchs) and an external
power supply for Corus.
The analogue measurement to be transmitted can be selected among:
Pressure
Temperature
Unconverted flow rate
Converted flow rate
Only one of the above analogue measurements can be transmitted per output.
A more detailed installation guide can be provided for this function.
17
4.4.4 1Hz clock signal
When selected, the Corus provides permanently a 1Hz square signal (50% duty-cycle) on the corresponding
output. Caution: It is recommended to use this function with an external power supply or during a short
period if the Corus is powered by its internal battery only.
When selected, the Corus provides on the selected output a pulse (duration programmable from 1 to 60
seconds) at each completion of a database period (programmable among Interval, Hourly, Daily, Monthly)
Corus calculates every second the gas instantaneous unconverted and converted flows:
the unconverted flow is derived from the incoming main LF signal
the converted flow is calculated from the unconverted flow balanced by the conversion factor
Theses flows are indicative, not metrological. If the function is activated, every time they are updated, the
instantaneous flows are compared respectively to:
a minimal and maximal unconverted flow threshold
a minimal and maximal converted flow threshold
When a flow rate leaves the range defined by the corresponding threshold, a flow alarm gets active.
18
5 COMMUNICATION
Corus provides two (2) communication ports that can be used simultaneously:
an optical port
a RS232 port
Optionally, a board can be plugged on the CPU board slot. Following communication boards (ATEX) are
available :
an internal PSTN modem
a 2-channel RS485 board
For more details, please refer to the corresponding operating manuals for each optional board.
Additional board operation with communication ports (modem or RS485) can operate simultaneously.
RS232, internal PSTN modem and RS485 ports can communicate with both protocols IFLAG and Modbus
RTU (both in slave mode). IFLAG is always activated and is used also for the communication on the optical
port. Modbus protocol can be activated separately for each of above mentioned ports. Both protocols can
be active on the same port. Corus is able to identify the protocol that the master is using to communicate.
Corus provides an asynchronous serial optical port that is compliant with the optical head used with existing
Itron gas products.
This port can be used for serial asynchronous (start-stop) half-duplex communications. It operates at 1200
baud, 2400 baud, 4800 baud, 9600 baud.
A disc made into a magnetic material is fixed under the cover and allows the mechanical fixing of the optical
head. The presence of an optical head is magnetically detected. This allows the optical port to be by default
in a power-down mode and to wake up automatically when necessary. In case of failure, the magnetic
detection can be overridden by turning on the user interface.
19
J6.1 TxD
J6.2 RxD
J6.3 DTR
J6.4 Supply+
J6.5 Supply-
The RS232 port can be used for serial asynchronous (start-stop) full-duplex communications. It operates at
300 baud, 1200 baud, 2400 baud, 4800 baud, 9600 baud, 19200 baud. The RS232 port is by default in a
power-down mode. When the connected device initiates a communication, it has to take into account a
wake up time of a few milliseconds.
The intrinsic safety interface for the RS232 of Corus, proposed by Itron, is the ISB+. For more details see the
manual of this product.
Corus internal modem board is ATEX certified (approved according to 94/9/EC directive related to
introduction of material in explosive atmosphere area).
Corus equipped with its internal modem board is approved as category 2 equipment and is then suitable for
use in zone 1.
The main characteristics of the internal modem board are:
The RS485 board can be used in hazardous area and up to 4 Corus can be connected to the same bus (for
RTU, modem, gas analyzer). The maximal distance between the modem (or RTU) in safe area and Corus,
depends on the number of Corus (1 to 4) and the electrical characteristics of the cable (linear L, C) and is, in
any case, 200m at the most.
The user can select the speed and other communication parameters for each of the two ports, including the
choice of the protocol (Modbus, I-Flag).
The board can be used when Corus is powered by battery or external power supply.
Main characteristics:
ATEX internal board with two independent RS485 ports
ATEX compatibility with a maximum of 4 Corus connected on the same RS485 bus and a maximum
distance of 200m with the IS interface
Connection on the optional slot (J3 CPU board)
20
Two wires passive RS485 ports half-duplex
Simultaneous communication on all ports without restrictions
Configurable speed up to 38400 bauds
Communication protocol depending on the application
The two ports are passive meaning that they must be powered by the intrinsic safety interface that will be
connected on Corus. Each port is isolated separately and can be connected to two different types of devices
(modems, RTUs, PCs, etc).
Suggested intrinsic safety interfaces are:
ISB+ (Itron) for both data signals and port powering
or
the barrier Z961 (from Pepperl &Fuchs) for data signals
the BXNE31 (from Georgin) power supply for powering the port.
A separate power supply and barrier must be used for the second port.
21
6 INSTALLATION, COMMISSIONING AND MAINTENANCE
6.1 Fixing
Corus enclosure features four lateral holes (diameter 5.2mm) for wall-mounting or for mounting on special
plate. Product shall be installed in a place protected from direct exposure to sunlight and rain.
Following the installation procedure, Corus has to be carefully closed using the four screws placed on each
corner. All cable glands have moreover to be tightened enough. These two actions are mandatory to
achieve a correct water tightness of the enclosure.
P sensor
RS232
Optional input :
J1 J3
T sensor Meter
Brown
Shield
Shield
Shield
White
White
Grey
Red
Red
1 2 3 4 5 6 1 2 3 J5 J6
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5
1 2 3 4 5 6 1 2 3 4
IN 1 -
IN 2 -
OUT 1 -
OUT 2 -
EXT -
TxD
RxD
DTR
+6V
+0V
IN 1+
IN 2+
OUT 1+
OUT 2+
EXT+
Yellow
Yellow
Brown
White
Green
White
Green
Pink
Grey
Pink
22
6.2.1 Pressure and temperature sensor
6.2.2 LF cable
J3.1 Tamper + (brown) J4.1 LF1 + (yellow)
J3.2 Tamper (gray) J4.2 LF1 (white)
J3.3 shield (LF cable) J4.3 LF2 + (pink)
J4.4 LF2 (green)
J6.1 TxD
J6.2 RxD
J6.3 DTR
J6.4 Supply+
J6.5 Supply-
23
6.3 Installation of the sensors
The temperature sensor has to be placed into a protective thermowell at a point of the pipe where the
measured temperature is representative of the gas temperature through the meter.
The generally accepted installation conditions are:
distance from the meter: less than 1 meter
position: downstream for turbines
immersion : 1/3 to 2/3 of the internal diameter of the pipe
The presence of a second thermowell located close to the first one is recommended to check Corus
temperature measurement, using a reference thermometer, without having to remove the probe.
The pressure sensor connector has a G1/4 (BSP) male terminal. It must be preferentially screwed to the "Pr"
reference pressure tap of the meter:
either directly (through a mechanical adapter if necessary)
or through a 3-way valve (which will avoid any modification of the installation during the periodical check
of the sensor)
If a failure of the pressure sensor makes it mandatory to change it, the pressure coefficients have to be
reprogrammed with the ones delivered with the new sensor. If Corus is in operation during the change, a
pressure alarm will be generated.
The pressure sensor features two (2) holes allowing the sealing the sensor with the meter installation.
Important notice:
- The pressure sensor has a gland for its cable which shall not be screwed or unscrewed by the user. Any
additional screwing can damage the wires of the cable.
- When connecting the sensor on the meter, the cable of the sensor shall not be twisted.
Twisting the cable can damage the wires of the sensor permanently.
Sensors shall be mounted either by using an adaptor which allows the sensor to be screwed without
turning its body, either by dismounting the sensor from Corus, screwing it to the meter (by turning the
body of the sensor together with the cable), then reconnecting it to the Corus.
6.3.3 LF cable
The LF cable provided with the CORUS has to be connected directly to the LF output plug of the gas meter.
The various LF cable (Binder plug ) available are :
Type 1 : Type 2 : Type 3 : Type 4 :
Pins 1-4 : LF1 Pins 4-6 : LF1 Pins 1-2 : LF1 Pins 1-4 : LF1
Pins 2-5 : LF2 Pins 3-5 : LF2 Pins 5-6 : LF2 Pins 2-5 : LF2
Pins 3-6 : Strap Pins 1-2 : AT Pins 3-4 : AT Pins 3-6 : AT
Each LF cable binder plug features two (2) holes allowing sealing the CORUS cable to the meter installation.
24
6.4 Power supply
Corus is delivered with a 16.5A.h D-size lithium battery that provides autonomy for 5 years in average
conditions. This battery includes all protections required by intrinsic safety and cannot be changed for a
different type.
When the battery reaches the end of its life, it has to be replaced according to the following procedure that
ensures a non-stop operation of the product:
do not provoke short-circuits on the new battery : it would be definitively damaged
check the voltage of the new battery : it should be above 3.6V
open the cover of Corus box
connect the new battery on the free battery connector (J7 or J8)
disconnect the old battery and remove it from the battery support
put the new battery into the battery support
reprogram the battery life to its nominal value (1825 days)
close the enclosure and check Corus
Corus can also be supplied by an external intrinsic safety associated module from:
a 220 VAC mains
a 24 VDC mains
The corresponding module has to be connected to the right hand side of the terminal J5 (J5.9 and J5.10).
The converter must then be configured in the 'external power supply' mode. In this mode, the internal
battery has to remain plugged in case of external power cut.
For safety reasons, it is strictly forbidden to use other modules than those specified by 'Itron'.
6.4.3 Autonomy
The table below shows indicative (see notes) autonomy of CORUS in years when using the battery mode,
according the 3 main parameters: Acquisition period, Retransmissions, Communication.
25
Retrans No Retrans. F = 0,1 Hz F = 0,5 Hz F = 1 Hz F = 2 Hz Comm.
Acq. Period Duration
1 second 1 1 1 0,9 0,8 15min./ month
1 1 1 0,9 0,8 15min./ week
5 seconds 3,7 3,6 3,2 2,7 2,1 15min./ month
3,7 3,6 3,1 2,7 2,1 15min./ week
10 seconds 5,7 5,4 4,4 3,6 2,7 15min./ month
5,6 5,3 4,4 3,6 2,6 15min./ week
15 seconds 6,9 6,4 5,1 4,1 2,9 15min./ month
6,7 6,3 5,0 4,0 2,9 15min./ week
20 seconds 7,7 7,1 5,6 4,3 3,0 15min./ month
7,5 6,9 5,4 4,2 3,0 15min./ week
30 seconds 8,7 8,0 6,1 4,7 3,2 15min./ month
8,4 7,8 5,9 4,6 3,1 15min./ week
The same table can be given in number of days. The user must set the starting value of the counter of
battery life time into Corus according to the corresponding conditions of use. By default the value
programmed is 1825 days, meaning 5 years.
26
Note 1:
The values provided in this table correspond to typical values in average conditions of use. They are
calculated with a 30% margin to take in account the possible variations of operating conditions that could
affect the consumption of the product (temperature, humidity, self-discharge of the battery)
Note 2:
The frequency of retransmissions indicated in the table corresponds to average frequency on both outputs,
when both channels are configured for pulses.
Note 3:
In case of activation of a digital output as ALARM output, the consumption of the device can be impacted,
especially in Standard mode for which the output is activated as long as the alarm is active. Then, it is
recommended to use preferably the limited mode or to use CORUS with an external power supply module
if Standard mode is required.
Note 4:
The values provided in this table correspond to a Corus without any optional board.
Note 5:
The use of AGA8 detailed formula reduces the battery life time, due to increased calculation time necessary
for this formula.
6.5 Programming
To program/modify parameters protected by the programming switch, the following procedure has to be
followed:
Open CORUS
Remove the seal of the white metrological cover, placed on the left part of the I/O board
Put the programming switch On
Program the required parameters (Keyboard or communication)
Put the programming switch Off
Place the metrological cover and seal it
Close CORUS
To program/modify parameters protected by the Client switch, the following procedure has to be
followed:
Open CORUS
Put the Client switch On
Program the required parameters (keyboard or communication)
Put the Client switch Off
Close CORUS
27
6.6 Pressure and temperature adjustment on site
It is possible to adjust the pressure and temperature values through a linearization function available
within Wincor software tool.
P = (a x P) + b
T = (c x T) + d
The coefficients a, b, c, d can be introduced in Corus, after an on-site verification, performed on 2 different
operating measurement values.
The use of this function is subject to national regulations. Refer to your local authorities before using it.
28
7 MID: COMPLIANCE WITH WELMEC 7.2 GUIDE
Corus complies with MID directive (2004/22/EC) and consequently to the corresponding harmonized
standard, EN12405-1/A1. Additionally, the internal firmware of the product complies with the WELMEC 7.2
guide. This document provides guidance for software-equipped measuring instruments within the scope of
the MID directive.
Corus complies with the following requirement sets of this guide:
P : Embedded software
I : Instrument specific requirements (converter)
S : Software separation
Corus firmware is split in to 2 separate microcontrollers but must be considered functionally as based on 3
parts:
F1: One metrological firmware dedicated to analogue measurements on the I/O board (metrol).
F2: One metrological firmware (kernel) that controls the heart of the product and ensures the integrity
of all metrological functions.
F3: One non-metrological firmware managing all non-metrological functions (called application ).
F2 and F3 are included into the same microcontroller on the CPU board.
C1 C2
F3 firmware
(application)
CPU board I/O board
So, the part of firmware which contains all legally relevant functions is clearly separated from other parts of
firmware. The Kernel manages the non-legally relevant firmware tasks and the priority is given to the
metrological tasks. Application firmware can be updated without breaking the metrology seals.
29
Protected interface with other parts of the firmware.
Backup of the metrological parameters and measurement data.
Verification of the kernel, parameters and data integrity.
Display of metrological measurements and parameters.
It is possible to get information (version, date) related to the 3 different firmware versions in the first INDEX
sub-menu.
INDEX VALUE ALARM DATA CONFIG
06/09/2008
14:00:00
Kernel : Kr1.00 13/04/07
Metrol : Met1.00 19/02/03
Appli : Ver5.00 12/08/08
To ensure the integrity of metrological firmware and associated parameters, Corus manages 4 CRC (cyclic
redundancy code):
30
Index CRC check :
Corus manages a CRC related to index & counters (converted and unconverted).
This CRC is checked every second, corresponding to the refreshment period of the index.
In case of error, a CRC alarm is generated and the corresponding event is provided to the application for
storage.
The read access is free and the write access can be protected by different means.
The access level of each parameter is defined in a table called Table protect.
The table protect itself is protected by the programming switch, meaning that it is not possible to modify
an access right without breaking the metrological seal of the product.
Hardware protection:
Programming switch
This switch has to be turned ON before any change on parameters declared as protected by
this switch. Access to this switch is protected by metrological seal and then prevent un-
authorized access to metrological data
Client switch
This switch has to be turned ON before any change on parameters declared as protected by
this switch. Access to this switch is protected by client seal (if any).
Software protection:
Software protection modes are approved according MID requirements. This includes flexible management
of password/keys/user groups for reading and programming.
For more information please contact your local Itron office.
31
7.5 Index display format
In order to comply with the MID requirement stating that the display shall have a sufficient number of
digits to ensure that both unconverted and converted index incremented during 8000 hours at maximum
conditions (Qmax, Cmax) do not return to its initial value, Corus display format is as follows, depending
dynamically on the index value itself.
Note :
The raw and base indexes are displayed on the first screen.
This screen is updated every second.
Unconverted index :
00012345.000m3
Converted index :
00054321.100Nm3
32
8 FUNCTIONAL DESCRIPTION
Note: The legally relevant firmware (Kernel & Metrol) is then approved under MID/ Welmec 7.2.
Through the application of the S requirements set, the rest of the firmware (Appli.) has no influence on
metrology and can be adapted to various markets requirements (e.g : database, non-metrological alarms,
communication protocol, display of non-metrological items, digital I/Os management, etc)
8.1 Database
The hourly log contains up to 1440 records (2 months data) that include:
the date and time
the hourly unconverted consumption
the hourly converted consumption
the hourly under alarm unconverted consumption
the hourly total converted consumption
the value of unconverted index (end of period)
the value of converted index (end of period)
the value of under alarm unconverted counter (end of period)
the value of total converted counter (end of period)
the minimum, maximum and average gas temperature
the minimum, maximum and average gas pressure
the minimum, maximum and average second gas pressure (if P2 optional board)
the minimum, maximum and average unconverted flow rate
the minimum, maximum and average converted flow rate
the status of the converter during the hour (presence of metrological alarms, Date/Time change)
The daily log contains up to 124 records (4 months data) that include:
the date and time
the daily unconverted consumption
the daily converted consumption
33
the daily under alarm unconverted consumption
the daily total converted consumption
the value of unconverted index (end of period)
the value of converted index (end of period)
the value of under alarm unconverted counter (end of period)
the value of total converted counter (end of period)
the minimum, maximum and average gas temperature
the minimum, maximum and average gas pressure
the minimum, maximum and average second gas pressure (if P2 optional board)
the minimum, maximum and average unconverted flow rate
the minimum, maximum and average converted flow rate
the status of the converter during the day (presence of metrological alarms, Date/Time change)
The user can introduce a "gas hour" in Corus. All daily consumptions are stamped from the gas hour to the
gas hour of the next day.
The user can introduce a "Gas day" in Corus. All monthly consumptions are stamped from the gas day (at
the gas hour) to the gas day of the next month (at the gas hour).
The user can select the items to store in calendar logs (Hourly, Daily, Monthly) for each record. However
some items will always be recorded and the user cant unselect them. By reducing the number of items, the
number of database records is increased.
The fixed part of the databases includes:
For the hourly and daily logs
the unconverted consumption
the converted consumption
the under alarm unconverted consumption
34
the total converted consumption
date/time and status
Its to be noted that the user can select to store only the integer part of index values to save space in
database.
Corus logs the occurrence of all main events (alarms, reset, switches change, programming).
For each event, the following data are available:
Date and time of the event
Nature of the event
35
Specific value (for some events only)
Status
The events log has a capacity of 800 events and cannot be reset. It is circular: when it is full, every new
event replaces the oldest one.
8.2 Alarms
Metrological part of Corus manages alarms as described below.
Others types of alarms depend on the application (non-metrological alarms) :
T low threshold
temp.
metrological alarms
T high threshold
T sensor failure
P low threshold
pressure
P high threshold
P sensor failure
CRC 'Parameters'
CRC 'Index'
Metrol task
36
The rest of the alarms that are managed are:
T low threshold
pressure temp.
T high threshold
Warning thersholds
P low threshold
P high threshold
pressure
P2 low threshold
second
P2 high threshold
P2 sensor failure
tamper input
on/off input #1
on/off input #2
conv.
vol.
37
8.2.1 Pressure alarms
38
8.2.3 Kernel alarms
The kernel firmware manages 5 alarms related to the control of firmware and metrological parameters
integrity:
Alarm CRC Metrol : active if the CRC of the Metrol firmware is not correct
Alarm CRC Kernel : active if the CRC of the Kernel firmware is not correct
Alarm CRC Param : active if the CRC of the metrological parameters is not correct
Alarm CRC Index : active if the CRC of the Index is not correct
Alarm Metrol. task : active if the kernel metrological task duration is higher than 1 second. In any case,
after 1 second, priority is given to the kernel and the corresponding application task is stopped.
39
8.2.7 Conversion factor alarm
If this function is activated, Corus generates this alarm when the computed conversion factor gets out of a
programmed range [Cmin ; Cmax]. This alarm is not metrological and is dedicated to the control of the
whole metering chain.
Corus manages two conversion factor alarms:
conversion factor low threshold alarm
conversion factor high threshold alarm
When a conversion factor alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the conversion factor alarm is set into the 'ALARM' menu and is stamped in the events log
Corus sets this alarm when the converted consumption during a given interval is higher than a programmed
threshold. The interval can be selected among:
the current database interval
the ongoing hour
the ongoing day
the ongoing month.
When the alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the alarm is set into the 'ALARM' menu and is stamped in the events log
Every 24 hours, the internal battery counter is decreased of one day. When the counter reaches 10% of the
nominal battery life (about 6 month (182 days) for a 5 years battery life), the battery alarm is generated.
40
When changing the battery, the battery counter has to be re-programmed so as to acknowledge the battery
alarm.
When the battery alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the alarm is set into the 'ALARM' menu and is stamped in the events log
The battery alarm is not metrological, meaning that when the battery alarm occurs, Corus keeps working
until the battery is fully discharged.
When Corus is supplied through external power, it is able to detect a mains cut.
Then, this alarm is set as long as the external supply is not present.
During the mains cut, the internal backup battery will ensure the normal operation of Corus and the battery
counter will be decreased as described in the "battery alarm" chapter.
When such an alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for alarm
output
The alarm is set into the 'ALARM' menu and is stamped in the events log
Through keyboard :
Go to the menu ALARM
Select reset
Select all alarms
Press OK to confirm
Through software
Go to the Reset alarms menu of the configuration tool
Select type of alarm to be reset : Metrologic (P or T) and/or Non metrologic
Click on Prog button
Note 1: it is not possible to reset an active alarm, except the Coherence alarm and the Consumption
threshold alarm.
41
9 USER INTERFACE
Corus features a graphic liquid crystal display and a five-key keyboard. The five keys allow scrolling through
the various menus as described below.
The interface architecture is based on five main menus:
INDEX, VALUE and ALARM menus are always displayed. DATA and CONFIG menus can be displayed or not
according configuration.
INDEX VALUE ALARM DATA CONFIG
the inner bar graph of the battery icon is set according to the remaining battery life
time ratio in battery and external power mode (if there is a mains cut).
it blinks when the battery alarm is active
the key icon:
is blinking when the shunt of optical port detection is activated.
is off otherwise
42
9.1 Keyboard
The left/right arrow keys allow scrolling through the menus. Items can be selected using the 'OK' key. When
the display is off, any press on a key puts the display on. Access to the DATA and CONFIG menus can be
restricted according the application and the display configuration which is selected. After 2 minutes with no
action on the keyboard, the display switches off. When pressing a key, the display automatically starts on
the INDEX menu.
When the display is on, the acquisition of pressure, temperature and the calculation of compressibility and
conversion factor is performed every second.
9.2 Display
Below is shown the display sequence managed directly by the Kernel. Content and access to other(s)
menu(s) depend on the application and the display configuration which is selected.
43
Kernel Stand alone
OK OK
OK OK
Serial N: SC08000001234
-----------------
P Serial N: KB08060123 Non-Metrological
Pmin : 0.90000 bar
Pmax: 10.00000 bar Add
-----------------
T Serial N: CB05080546
OK
Test LCD
44
display test
When associated to an application, this menu (and associated sub-menus) can be enhanced with other(s)
data.
This menu displays the instantaneous values that are measured or calculated by Corus:
pressure
temperature
compressibility factor
conversion factor
battery life
When associated to an application, this menu can be enhanced with others data (flow-rates, second
pressure P2,), display of a graph,
The content of this menu depends on the application and the associated display configuration which is
selected. As an example, this menu usually allows to:
get information related to active alarms
get information related to memorized alarms
get information related to under alarm counters
reset alarms
Active alarms
The active alarms submenu displays the list of all active alarms. Using the up/down arrows and the OK keys,
it is possible to get more details about a given active alarm (Start time-stamping)
Memorized alarms
The memorized alarms submenu displays the list of all memorized alarms. Using the up/down arrows and
the OK keys, it is then possible to get more details about a given memorized alarm (Start and End time-
stamping)
This submenu allows to reset all memorized alarms and/or the under alarm counters (according to the
status of the prog switch for metrological alarms)
45
9.2.5 'DATA' menu
The content of this menu depends on the application and the associated display configuration which is
selected. For each active log, it is possible to consult the content of each record stamped in the
corresponding log of the product.
Communication : Modbus protocol activation and communication parameters for each com port
Volume : Input pulse weight, unconverted Index, converted Index
Metrology : Pbase, Tbase, Gas composition (according formula)
Date / Time : Current date / time
Supply : Mode (battery, external) , autonomy
Digital Output 1 : Activation/Deactivation , type (pulse, alarm,4/20mA) , parameters
Digital Output 2 : Activation/Deactivation , type (pulse, alarm,4/20mA) , parameters
46
INDEX MENU
Index menu
INDEX VALUE ALARM DATA CONFIG
contrast+
Unconverted index :
00012345.000m3 config
menu
Date / Time
Versions
value
menu
Converted index :
00054321.100Nm3 contrast-
06/09/2008
14:00:00 config
menu
Status value
menu
Main metrology
parameters
(press 2s )
Kernel : Kr1.00 13/04/07
Metrol : Met1.00 19/02/03
Appli : Ver5.00 12/08/08
Status menu
Tamper Alarm
Prog. switch On
Cust. switch On
Index
menu
47
VALUE MENU
Qm = 152.5 m3/h
Qb = 153.9 Nm3/h next
item
Bat = 250 days
Value menu
Graph
INDEX VALUE ALARM DATA CONFIG
previous item
4.2 graph
Instantaneous
value
48
ALARM MENU
Alarm menu
INDEX VALUE ALARM DATA CONFIG
Alarms:
02active
02 active alarms
alarms
00 memorized alarms
previous upper
alarm item
next lower
alarm item
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG
back to
reset
menu
back to
reset
menu
Alarms reset :
Completed
49
DATA MENU
Data menu
Date selection
INDEX VALUE ALARM DATA CONFIG
previous
14/08/2002 date
13/08/2002
12/08/2002 back to
database
display
records
back to
database
11/08/2002 menu on date menu
10/08/2002
09/08/2002 next
date
Hourly log 1/60
Hour selection
50
CONFIG MENU
Config menu
INDEX VALUE ALARM DATA CONFIG
Communication
Volume
Metrology
Date/Time
Supply
Output 1
Output 2
Communication
OK for
modification
Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys
Volume
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG
000185.000 validation
Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys
Metrology
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG
Current value
Pb : 1.01325 bar Gas rel. density : 0.57000
Tb : 273.15 K OK for
Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys
Date/Time
INDEX VALUE ALARM DATA CONFIG
INDEX VALUE ALARM DATA CONFIG
Current value
Date: 07/06/2003 07:43:00 Date : 07/06/2003 07:43:00
07/ 06/2003 OK for
validation
07:43:00
Battery validation
Selection with up and down keys + OK Selection with up and down keys
51
Output 1 or 2
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG
Mode : Unconv. Pulse Current value
Pulse Weight : 1.000 m3 Pulse Weight : 1 m3
Time closed : 250 ms OK for
1.000 m3 validation
Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys
52
10 SAFETY INSTRUCTIONS
10.1 General
Corus gas volume converter is approved according to 94/9/CE directive (ATEX) relative to introduction of
material in explosive atmosphere area.
The respect of this directive is mandatory (from July, 1st 2003) and is included in the 'CE' marking.
Thus, 'CE' marking of Corus implies now conformity with the following directives:
94/9/EC (ATEX), 89/336/EEC (EMC) and 2004/22/EC (MID).
CORUS is approved as category 1 equipment and is thus suitable for use in zone 0 (areas with permanent
presence of gas).
The corresponding marking is: 0081 LCIE 03 ATEX 6165X II1G Ex ia IIC T4
The ambient temperature range for operation is: -25C; +55C
To avoid erroneous operations, it is recommended to read the whole operating manual before
commissioning.
Corus is suitable for being used with gases belonging to IIA, IIB and 'IIC' classification gases, but :
Use with hydrogen is forbidden. (Hydrogen implies specific tightness elements).
Use with oxygen is strictly forbidden. (Oxygen is not covered by ATEX)
The connection of Corus to any other instrument or device must be done in accordance to the electrical
parameters indicated on the certificate.
It has to be noted that the directive 1999/92/EEC ("Installation / Users") has also to be respected for the
commissioning of the device.
When Corus is used in the external power supply mode, the supply has to be provided through an
intrinsically safe module with electrical compatible parameters; Use of any other type of module is
strictly forbidden. In this mode, the main battery remains plugged inside Corus in case of external mains
failure.
When using the battery mode, it is allowed to change the battery on site, by plugging the new battery
on the free connector (J7 or J8 of the I/O board) and then, removing the old battery. Its strictly
forbidden to leave Corus with two batteries permanently plugged in hazardous area.
Caution: The battery of Corus includes all protections required for intrinsic safety. Thus, it is strictly
forbidden to replace the battery unit by any battery other than the one specified by Itron (reference:
LS33600 from Saft , SL2780 from Sonnenschein or TD5930 from Tadiran)
Battery warning : The battery of Corus contains Lithium :
Fire, explosion and severe burn hazard
53
Do not recharge, short-circuit, crush, disassemble, heat above 100C (212F), incinerate or
expose contents to water
Do not solder directly to the cell
Respect storage recommendations (-30C ; +60C)
To prevent any damage, the following operations are recommended to limit electrostatic discharges on
the enclosure of Corus :
- the person carrying out the installation can discharge himself/herself by touching the potential
equalization line
54
11 MANUFACTURING
Corus is manufactured by :
Itron GMBH
Hardeckstrasse 2
76185 Karlsruhe
Germany
The site has ISO9001, ISO14401 certifications as well as certifications to produce Corus with ATEX and MID
markings.
55
12 APPENDIX 1: INITIAL ATEX CERTIFICATE
56
57
58
13 APPENDIX 2: EC DECLARATION OF CONFORMITY
59
60
61
62