Sunteți pe pagina 1din 64

Corus

Gas volume converter

Installation guide V2.6


TABLE OF CONTENTS

1 INTRODUCTION .......................................................................................................................................................... 3
1.1 PRODUCT PRESENTATION .................................................................................................................................................................. 3
1.2 OPERATING PRINCIPLE ...................................................................................................................................................................... 3
1.3 MAIN CHARACTERISTICS.................................................................................................................................................................... 5
1.4 OVERALL DIMENSIONS (IN MM) .......................................................................................................................................................... 7
1.5 MAIN COMPONENTS ........................................................................................................................................................................ 7
2 SYNOPTIC OF THE PRODUCT ...................................................................................................................................... 8
2.1 HARDWARE ARCHITECTURE ............................................................................................................................................................... 8
2.2 METROLOGICAL SEALING SCHEME ....................................................................................................................................................... 8
2.3 CUSTOMER SEALING......................................................................................................................................................................... 9
2.4 METROLOGICAL LABEL .................................................................................................................................................................... 10
3 METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS .....................................................................................11
3.1 TEMPERATURE MEASUREMENT ......................................................................................................................................................... 11
3.2 PRESSURE MEASUREMENT ............................................................................................................................................................... 11
3.3 COMPRESSIBILITY RATIO CALCULATION ............................................................................................................................................... 12
3.4 CONVERSION FACTOR CALCULATION .................................................................................................................................................. 13
3.5 VOLUMES METERING ..................................................................................................................................................................... 13

4 NON-METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS............................................................................16


4.1 TAMPER INPUT ............................................................................................................................................................................. 16
4.2 COHERENCE LF INPUT .................................................................................................................................................................... 16
4.3 ON/OFF INPUTS ............................................................................................................................................................................ 16
4.4 ON/OFF OUTPUTS ......................................................................................................................................................................... 17
4.5 INSTANTANEOUS FLOWS ................................................................................................................................................................. 18
4.6 SECOND PRESSURE (OPTIONAL BOARD) ............................................................................................................................................... 18

5 COMMUNICATION ....................................................................................................................................................19
5.1 OPTICAL PORT .............................................................................................................................................................................. 19
5.2 RS232 PORT ............................................................................................................................................................................... 19
5.3 INTERNAL ATEX PSTN MODEM (OPTIONAL BOARD) ............................................................................................................................. 20
5.4 2-CHANNEL RS485 (OPTIONAL BOARD) ............................................................................................................................................ 20

6 INSTALLATION, COMMISSIONING AND MAINTENANCE ..........................................................................................22


6.1 FIXING ........................................................................................................................................................................................ 22
6.2 CONNECTORS AND TERMINALS ......................................................................................................................................................... 22
6.3 INSTALLATION OF THE SENSORS......................................................................................................................................................... 24
6.4 POWER SUPPLY ............................................................................................................................................................................. 25
6.5 PROGRAMMING............................................................................................................................................................................ 27
6.6 PRESSURE AND TEMPERATURE ADJUSTMENT ON SITE ............................................................................................................................. 28

7 MID: COMPLIANCE WITH WELMEC 7.2 GUIDE .........................................................................................................29


7.1 FIRMWARE STRUCTURE ................................................................................................................................................................... 29
7.2 FIRMWARE IDENTIFICATION ............................................................................................................................................................. 30
7.3 FIRMWARE PROTECTION AND DATA CONTROL (CRC) ............................................................................................................................. 30
7.4 DATA PROTECTION DATA ACCESS ................................................................................................................................................... 31
7.5 INDEX DISPLAY FORMAT .................................................................................................................................................................. 32
8 FUNCTIONAL DESCRIPTION .......................................................................................................................................33
8.1 DATABASE ................................................................................................................................................................................... 33
8.2 ALARMS ...................................................................................................................................................................................... 36
8.3 ALARMS RESET ............................................................................................................................................................................. 41

9 USER INTERFACE ........................................................................................................................................................42


9.1 KEYBOARD ................................................................................................................................................................................... 43

1
9.2 DISPLAY ...................................................................................................................................................................................... 43
10 SAFETY INSTRUCTIONS ..............................................................................................................................................53
10.1 GENERAL .................................................................................................................................................................................... 53
10.2 MARKING RELATIVE TO ATEX - ZONE ................................................................................................................................................ 53
10.3 SPECIFIC INSTRUCTIONS .................................................................................................................................................................. 53

11 MANUFACTURING .....................................................................................................................................................55

12 APPENDIX 1: INITIAL ATEX CERTIFICATE ...................................................................................................................56

13 APPENDIX 2: EC DECLARATION OF CONFORMITY ....................................................................................................59

THIS MANUAL INCLUDES DOCUMENT D0214888 AD


PROVIDED FOR MID CERTIFICATION

ADDITIONAL INFORMATION CONCERNS THE NON METROLOGICAL PARTS

2
1 INTRODUCTION

1.1 Product presentation

Corus is a type 1 gas volume converter that can be configured as a T, PT or PTZ version.
It is designed to calculate gas volumes under base conditions, based on a volume at metering conditions, a
temperature, a pressure and a compressibility factor. Its volume input is a low frequency one that can be
connected to any Ex compliant gas meter that provides such information.
Corus belongs to a new generation of volume converters that are designed to achieve a high level of
performance thanks to powerful electronic equipment. Based on the use of flash memory, its architecture
allows the download of a new firmware release through a laptop without board modification.
Corus is approved according the new European directive (MID) based on standard EN 12405-1/A1 and can
be used for commercial and fiscal gas transactions. According to the European regulation, it bears the CE
marking and complies with the following directives:

89/336/EEC (electromagnetic compatibility directive) EMC


94/9/EC (potentially explosive atmospheres directive) ATEX
2004/22/EC (metrology) (MID)

Corus provides the following functions:


acquisition and integration of the volume measured by the meter
measurement of gas temperature and gas pressure
calculation of the compressibility factor (PTZ version)
calculation of the conversion factor and of the volume in base conditions using a T, PT or PTZ conversion
calculation of the measured and converted flow rates
display of the measured and calculated data
management of alarms
management of a large database
multiple communication channels for local and remote operations
on/off inputs and outputs
optional slot for one ATEX additional board: Internal PSTN modem or double RS485 com ports board or
second pressure input board (P2).

1.2 Operating principle

The gas meter measures a volume of gas at metering conditions (Vm), under specific pressure (Pm) and
temperature (Tm) conditions. This volume is converted into a volume at base conditions by Corus according
to EN 12405:
Pm Tb Zb
Vb Vm C Vm
Pb Tm Zm

where :
Vm = Volume at measuring conditions
Vb = Converted volume at reference (base) conditions
Tm = Gas temperature at metering conditions

3
Tb = Reference (base) absolute temperature
Pm = Gas absolute pressure at metering conditions
Pb = Reference (base) absolute pressure
Zm = Gas compressibility factor at metering conditions
Zb = Gas compressibility factor at reference (base) conditions
C = Conversion factor

4
1.3 Main characteristics

Name CORUS
Manufacturer ITRON
Type Gas volume conversion device - Type 1
Model T, PT or PTZ
Metrology Conform to EN12405-1/A1
MID test certificate: T10323 by NMI (module B)
MID production approval by PTB (module D)
PTZ model : Z computed according :
S-GERG 88
AGANX19
AGANX19 mod.
Table of Z
AGA8 Gross method 2 (GM2)
AGA8 Detailed method (DM)
Ambient temperature range [-25C ; +55C]
Gas temperature range [- 30C ; +70C]
Reference conditions Pb : from 0,9 bar to 2 bar
Tb : from 0C to 40C
Protection class IP65 according EN60529
Power Supply Internal battery (16.5A.h) or external DC voltage
Internal power supply Limits : 2.6V DC < Vint < 3.7V DC
Nominal voltage : 3.3V DC
External power supply Limits : 6 V DC < Vext < 12 V DC
Nominal voltage : 10V DC
Pressure ranges Number : 3
Sensor : 0.9 10 bars abs.
Sensor : 3 30 bars abs.
Sensor : 7.2 80 bars abs.
Pressure sensor type External Piezo-resistive sensor
Temperature probe Platinum probe - PT1000 class A according EN60751
Metering input (volume) Low frequency type (2Hz max.)
Cables length LF volume : 0.8 m or 2.5 m

5
Pressure : 0.8 m or 2.5 m
Temperature : 0.8 m or 2.5 m
CE marking Conformity to :
- EMC directive (89/336/EEC)
- ATEX directive (94/9/EC)
- MID directive (2004/22/EC)
Electromagnetic class E2
Mechanical class M2
Humidity Designed for non-condensing humidity
ATEX classification II1G ia IIC T4
Enclosure material Poly-carbonate
Enclosure dimensions (cm) 22.2 x 14.5 x 8.6
Total weight (approx.) 1.5 kg
Display Graphic
Keyboard 5 keys
Communication ports Standard : RS232 & Optical
Option : RS485 2 channels or PSTN modem
Communication protocols IFLAG (based on IEC 62056-21 also know as EN61107)
Modbus RTU
Slave mode for both protocols
Nature of measured gas Fuel gases of the first and second family according EN437
Units of main data Pressure (absolute): bar, Psi, kPa
Temperature : C , F
Volume : m3, Cft
Firmware Compliant with WELMEC 7.2 guide according following modules:
- P : Embedded software
- I : Instrument specific requirements (converter)
- S : Software separation
--------------------------------------------------------------------
2 separate micro-controllers (C) :
M16C C (CPU board)
Metrological kernel
Non-metrological application
XEMICS C (I/O board)
Metrological firmware (analogue measurements)

6
1.4 Overall dimensions (in mm)

1.5 Main components

Programming Programming Battery Fuse Battery


Display Switch Connector Connector

A
m
0
5
2
EULAV

rab 8 9 6 .5 1 = P
C 5 . 3 1 = T
5 3 00 . 1 = 0Z/Z Customer
3
4 5 46 . 5 = C Switch
h /3 m 6 5 .4 5 5 1 = mQ
h / mN 6 9 .3 4 8 1 = bQ

Metrology Shield

1 2 3 4 5 6 1 2 3 125mA

1 2 3 4 5 6 7 8 9 10 1 2 3 4 5
1 2 3 4 5 6 1 2 3 4

Keyboard C
Programming SRAM Extension Slot
Connection Connector Memory (J3 Connector)

Flash
Memory

Optional board
Fixing

7
2 SYNOPTIC OF THE PRODUCT

Corus is delivered as an IP65 wall-mountable device designed in a tight polycarbonate enclosure.


A hinge on the left of the box makes any operation in the field very easy, providing a clear access to all
major parts of the product (terminals, switches, internal battery)
Corus features three electronic boards:
a display board, that includes the liquid crystal display
a 'CPU' board that holds the heart of the product (the microcontroller and its direct peripherals: RAM and
flash memories)
an 'I/O' board , that gathers all terminals, the pressure and temperature acquisition circuits, the LF and
tamper inputs, the digital inputs and outputs, the RS232 port and the internal battery.
Note: two components in Corus integrate firmware:
one micro-controller on the CPU board
one specific chip on the I/O board that is dedicated to pulse acquisition, pressure and temperature
measurement.

2.1 Hardware architecture

2.2 Metrological sealing scheme

The metrological sealing protects the access:


to all metrological components
to all connections to the sensors and to the meter
to the prog switch that is used to configure the access to metrological data (see 6.4)

8
The standard sealing is made up of two plates that cover respectively the CPU board and the metrological
part of the I/O board. These plates can be sealed as indicated below:

1: metrological plate protected by seal


2: metrological plate protected by seal
3: inviolable metrological label

2.3 Customer sealing

It is possible to place on the top-right corner and/or on the bottom-right corner of the box a non-
metrological seal (called customer seal) to prevent any access to the inside of the enclosure of Corus. Any
attempt to open the box lets permanent visible marks on the seal or on the enclosure.
This seal notably protects the access to:
the non-metrological interfaces
the customer switch that is used to configure the access to customer data (see 6.4)
the battery

1: customer seals

9
2.4 Metrological label

A label is provided on the top of the product for the metrological marking and the CE marking. This label is
inviolable: it cannot be unstuck without being destroyed. This label provides the following data:
product name
name of the manufacturer (Itron)
serial number of the device
manufacturing date
ambient temperature range
operating gas pressure range
gas temperature range
reference pressure
reference temperature
data related to MID metrological approval
data required by the potentially explosive atmospheres regulation
IP protection level
CE marking

Marking related to MID :

Manufacturing year

M08 0102 Notified body number (Module D)

T10323 Corus MID certificate number

Marking related to ATEX :

0081 II1G Ex ia IIC T4 - LCIE 03 ATEX 6165X

with internal optional PSTN modem:

0081 II 1/2 G Ex eiamb IIC T4 - LCIE 03 ATEX 6165X

10
3 METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS

3.1 Temperature measurement

Corus is provided with a four-wire Pt1000 temperature sensor that is in accordance with IEC/EN 60751
(1000resistance at 0C). Its four-wire mounting provides a high accuracy by eliminating the resistance of
the cable.
The temperature sensor has an IP67 protection level according to EN 60529 and is delivered with a cable
length of either 2.50m or 0.80m. Its operating range (gas temperature range) is [-30C ; +70C].
The temperature measurement period is programmable from 1 to 30 seconds. When the display is On, this
period is forced to 1 second.
Whenever the temperature measurement goes beyond the operating range defined by two minimal and
maximal temperature thresholds or whenever a temperature sensor failure is detected, a temperature
metrological alarm is generated (see 8.2.2).

3.2 Pressure measurement

3.2.1 T version

When used as a T converter, Corus uses a fixed pressure that is calculated according to the following
formula:


pressure 1.01325 bara gas.rel.press 1.202 104 bar/m altitude

where:
pressure = fixed pressure used for the T conversion
gas.rel.press = gas fixed relative pressure to be programmed into the device
altitude = altitude of the installation to be programmed into the device

3.2.2 PT and PTZ versions

When used as a PT or PTZ converter, Corus is provided with an external absolute piezo-resistive pressure
sensor providing a rangeability higher than 11. Three (3) operating pressure ranges are available:
0.9 bara to 10 bara
3 bara to 30 bara
7.2 bara to 80 bara
The sensor is enclosed into a stainless steel enclosure and has an IP66 protection level according to EN
60529. It can withstand an overpressure of 1.25 times its maximal pressure (12.5, 37.5 or 100 bara) for 30
minutes. It is equipped with a G1/4 (BSP) male terminal that allows its direct connection to the gas meter or
to the gas pipe. It is delivered with a cable length of either 2.50m or 0.80m.
All pressure sensors are delivered with their own characterization coefficients (12) that must be programmed
into the converter so that the accuracy required by EN 12405-1/A1 can be reached over the whole pressure
and temperature range.

11
The pressure measurement period is programmable from 1 to 30 seconds. When the display is On, this
period is forced to 1 second.
Whenever the pressure measurement goes beyond the operating range defined by two minimal and
maximal pressure thresholds or whenever a pressure sensor failure is detected, a pressure metrological
alarm is generated (see 8.2.1).

3.3 Compressibility ratio calculation

Corus can be programmed to display either Zb/Zm or Zm/Zb (this choice is a metrological parameter).

3.3.1 T and PT versions

When configured as a T or PT converter, Corus uses a constant value as compressibility factor. This value
(Zb/Zm or Zm/Zb fixed ratio) can be set by the user.

3.3.2 PTZ version

Corus calculates the compressibility factor:


every 1s to 30s (programmable period) if the display is off (same period than for P and T acquisition)
every second if the display is on
When configured as a PTZ converter, Corus computes the compressibility ratio of the gas using one of the
following formulas:
S-GERG88
AGANX19 standard
AGANX19 modified
AGA8 Gross method 2
AGA8 Detailed method
16 Coefficients (Table of Z)
According to the selected formula, the following data are necessary and must be programmed into the
device:

S-GERG AGANX19 AGA8 GM2 AGA8 Detailed Table of Z


(Std. or Mod.)
%CO2
%N2
%H2
HSV (Cv)
Specific gravity
Prd
Trd
Full composition
16 coefficients

12
According the selected formula, gas composition has to be compatible with the following ranges:
S-GERG: AGANX19 (Std or mod.): AGA8 (Gross method):
0% < %CO2 < 30% 0% < %CO2 < 30% 0% < %CO2 < 30%
0% < %H2 < 10% 0% < %N2 < 50% 0% < %N2 < 50%
0.55 < SG < 0.9 0.55 < SG < 0.9 0.55 < SG < 0.9
5.27 < Cv (kWh) < 13.33

AGA8 (Detailed method):


70% < CH4 < 100% 0% < C5H12 < 0,5% 0% < CO2 < 20%
0% < C2H6 < 10% 0% < C6 < 0,1% 0% < N2 < 20%
0% < C3H8 < 3,5% 0.55 < C7 < 0.05% 0% < H2 < 10%
0% < C4H10 < 1,5% 0.55 < C8+ < 0.05% 0% < CO < 3%

Table of Z:
The 16 coefficients are provided by Itron according the required gas.

3.4 Conversion factor calculation

Corus calculates the conversion factor:


every 1s to 30s (programmable period) if the display is off (same period than for P and T acquisition)
every second if the display is on
The calculation is done according to the formula defined in EN 12405-1/A1:
Pm Tb Zb
C
Pb Tm Zm
The reference pressure Pb and the reference temperature Tb must be programmed into the device.

3.5 Volumes metering

Corus manages two volume indexes and two under alarm volume counters:
an unconverted index
a converted index
an under alarm unconverted counter
a total converted counter
The two indexes are available in the 'INDEX' menu of the display, whereas the under alarm unconverted
counter and the total converted counter can be found in the 'ALARM' menu (see 9.2).

13
3.5.1 Main LF input and unconverted volumes

Corus features a low frequency (LF) input that can work up to 2 Hertz and that is used to meter the
unconverted volumes.
The pulses sent by the meter on this input (called main LF input or LF1) are balanced by the input pulse
weight and are added to the unconverted index. The input pulse weight can be chosen among the following
values:
0.001 m3/pulse
0.01 m3/pulse
0.1 m3/pulse
1 m3/pulse
10 m3/pulse
100 m3/pulse
The LF input can be associated to any gas meter that outputs LF signal or to any device that provides an
electrically compliant interface:
dry contact
static relay
open collector or open drain output
For its connection to the meter, it shares the same terminals (J3 and J4) and the same cable as the
coherence LF input (see 4.1) and as the tamper input (see 4.2). Corus can be delivered with several
different LF cables, in order to comply with all existing types of meter connections. All cables are available
with length of 0.80m or 2.50m.

3.5.2 Converted volumes

After every conversion, Corus updates its converted volumes: the converted index and the total converted
counter are increased by the unconverted volume metered since the last conversion balanced by the
conversion factor.

3.5.3 Under alarm operation

As long as a metrological alarm is in progress, volumes are managed as specified in EN 12405-1/A1:


the unconverted main index continues to be incremented
the under alarm unconverted counter is incremented
metering on the converted main index is stopped
the total converted counter is incremented during the alarm using a conversion factor based on pressure
and temperature back-up values, if this function is activated with programmed values.
Twelve monthly temperatures and one pressure back-up values are available. They can be either:
programmed by the user
automatically calculated by Corus from stored data (using the database)

14
normal operation operation under alarm

unconverted index Incremented Incremented

converted index Incremented Stopped

under alarm
Stopped Incremented
unconverted counter
Incremented with back-up
total converted counter Incremented
values

metering

15
4 NON-METROLOGICAL MEASUREMENTS AND INPUTS/OUTPUTS

4.1 Tamper input

Corus manages an on/off tamper input that allows to detect any magnetic violation attempt on meters that
are equipped with an anti-tampering reed contact. This input shares the same terminals (J3 and J4) and the
same cable as the main LF input.
Tamper input can be set to one of the following modes:
disabled
normally open (NO)
normally closed (NC)
When the input is enabled and not in its normal state, the device generates a tamper alarm that is not
metrological and that does therefore not impact the management of the unconverted and converted
indexes.

4.2 Coherence LF input

Corus features a second LF input called coherence LF input or LF2. This input shares the same terminals
(J3 and J4) and the same cable as the main LF input.
It allows checking of good operation of the metering function and of the pulse emitter of the meter. If this
function is activated, Corus constantly compares the number of pulses metered on both LF inputs and
generates a coherence alarm as soon as the spread between both inputs exceeds a programmable threshold
within a programmable period of time. The coherence LF input and the coherence alarm are not
metrological: they have no effect on the unconverted and converted indexes.

4.3 On/off inputs

Corus features two on/off inputs that can be connected to different types of intrinsically safe interfaces:
reed contact
static relay
open collector or open drain output
The field of use of these inputs includes the connection to a safety valve to a delta pressure filter status or
to a station door opening contact.
On/Off inputs can be set to one of the following modes:
disabled
normally open (NO)
normally closed (NC)
When any on/off input is enabled and not in its normal state, the device generates a non-metrological
alarm.

16
4.4 On/off outputs

Corus features two on/off outputs that can be each configured:


as an unconverted volume pulse output
as a converted volume pulse output
as a total converted counter pulse output
as an alarm output (standard mode)
as an alarm output (limited mode)
as a 4/20 mA output (with external F/I converter)
1Hz clock signal
End of period signal

4.4.1 Volume pulse output mode

When configured as unconverted or converted volume or total converted counter pulse outputs, on/off
outputs behave like the LF outputs of a meter:
they are individually characterized by a pulse weight that can be configured to any value greater than or
equal to the input pulse weight (see 3.5.1) among 0.001 m3/pulse, 0.01 m3/pulse, 0.1 m3/pulse, 1
m3/pulse, 10 m3/pulse, 100 m3/pulse
they generate a pulse every time the corresponding (unconverted or converted) index gets increased by a
value equal to the selected output pulse weight.
The closing time of the volume pulse outputs can be programmed (default value is 250ms). Maximal
frequency of outputs is 2Hz.
Base volume outputs are automatically disabled as long as any metrological alarm is active.

4.4.2 Alarm output mode

Each alarm can be individually configured to activate the alarm outputs when triggered. The activation
duration of the alarm outputs can be selected. The alarm outputs can remain active:
either as long as any alarm is active (standard mode)
or as long as any alarm is active with a duration limit of 30 minutes (limited mode)

4.4.3 4-20mA mode

The On/Off outputs of CORUS can be configured as 4/20mA outputs. This option requires the use of an
external frequency to current converter (model KFU8-UFC-EX1D from Pepperl&Fuchs) and an external
power supply for Corus.
The analogue measurement to be transmitted can be selected among:
Pressure
Temperature
Unconverted flow rate
Converted flow rate
Only one of the above analogue measurements can be transmitted per output.
A more detailed installation guide can be provided for this function.

17
4.4.4 1Hz clock signal

When selected, the Corus provides permanently a 1Hz square signal (50% duty-cycle) on the corresponding
output. Caution: It is recommended to use this function with an external power supply or during a short
period if the Corus is powered by its internal battery only.

4.4.5 End of period signal

When selected, the Corus provides on the selected output a pulse (duration programmable from 1 to 60
seconds) at each completion of a database period (programmable among Interval, Hourly, Daily, Monthly)

4.5 Instantaneous flows

Corus calculates every second the gas instantaneous unconverted and converted flows:
the unconverted flow is derived from the incoming main LF signal
the converted flow is calculated from the unconverted flow balanced by the conversion factor
Theses flows are indicative, not metrological. If the function is activated, every time they are updated, the
instantaneous flows are compared respectively to:
a minimal and maximal unconverted flow threshold
a minimal and maximal converted flow threshold
When a flow rate leaves the range defined by the corresponding threshold, a flow alarm gets active.

4.6 Second pressure (optional board)

A Second pressure board can be used in hazardous area.


This pressure measurement is not metrological.
The sensors managed by this board are:
CORUS sensor 0.9-10 bara
CORUS sensor 3-30 bara
CORUS sensor 7.2-80 bara
Other compatible gauge or absolute sensors compliant with the ATEX parameters of this board.

18
5 COMMUNICATION

Corus provides two (2) communication ports that can be used simultaneously:
an optical port
a RS232 port

Optionally, a board can be plugged on the CPU board slot. Following communication boards (ATEX) are
available :
an internal PSTN modem
a 2-channel RS485 board

For more details, please refer to the corresponding operating manuals for each optional board.
Additional board operation with communication ports (modem or RS485) can operate simultaneously.

RS232, internal PSTN modem and RS485 ports can communicate with both protocols IFLAG and Modbus
RTU (both in slave mode). IFLAG is always activated and is used also for the communication on the optical
port. Modbus protocol can be activated separately for each of above mentioned ports. Both protocols can
be active on the same port. Corus is able to identify the protocol that the master is using to communicate.

5.1 Optical port

Corus provides an asynchronous serial optical port that is compliant with the optical head used with existing
Itron gas products.
This port can be used for serial asynchronous (start-stop) half-duplex communications. It operates at 1200
baud, 2400 baud, 4800 baud, 9600 baud.
A disc made into a magnetic material is fixed under the cover and allows the mechanical fixing of the optical
head. The presence of an optical head is magnetically detected. This allows the optical port to be by default
in a power-down mode and to wake up automatically when necessary. In case of failure, the magnetic
detection can be overridden by turning on the user interface.

5.2 RS232 port

Corus features an insulated RS232 port that can be used for:


local communication with a laptop through an intrinsic safe isolation
remote communication through an external modem (either through an IS associated modem or through
an IS barrier connected to a standard modem)
The RS232 port provides the following signals:
transmitted data (TxD)
received data (RxD)
data terminal ready (DTR)
As it is insulated, it needs to be externally supplied by a voltage between 6V and 12V. The terminals (J6) that
correspond to the serial port are located in the customer area:

19
J6.1 TxD
J6.2 RxD
J6.3 DTR
J6.4 Supply+
J6.5 Supply-

The RS232 port can be used for serial asynchronous (start-stop) full-duplex communications. It operates at
300 baud, 1200 baud, 2400 baud, 4800 baud, 9600 baud, 19200 baud. The RS232 port is by default in a
power-down mode. When the connected device initiates a communication, it has to take into account a
wake up time of a few milliseconds.
The intrinsic safety interface for the RS232 of Corus, proposed by Itron, is the ISB+. For more details see the
manual of this product.

5.3 Internal ATEX PSTN modem (optional board)

Corus internal modem board is ATEX certified (approved according to 94/9/EC directive related to
introduction of material in explosive atmosphere area).
Corus equipped with its internal modem board is approved as category 2 equipment and is then suitable for
use in zone 1.
The main characteristics of the internal modem board are:

ATEX internal PSTN Modem for Corus


Connection to the optional slot (J3 CPU board)
Chip Modem V32 bis (max.14400 bauds)
Communication protocol depending on the application
The modem can be used when Corus is powered by battery or external power supply.

5.4 2-Channel RS485 (optional board)

The RS485 board can be used in hazardous area and up to 4 Corus can be connected to the same bus (for
RTU, modem, gas analyzer). The maximal distance between the modem (or RTU) in safe area and Corus,
depends on the number of Corus (1 to 4) and the electrical characteristics of the cable (linear L, C) and is, in
any case, 200m at the most.
The user can select the speed and other communication parameters for each of the two ports, including the
choice of the protocol (Modbus, I-Flag).
The board can be used when Corus is powered by battery or external power supply.

Main characteristics:
ATEX internal board with two independent RS485 ports
ATEX compatibility with a maximum of 4 Corus connected on the same RS485 bus and a maximum
distance of 200m with the IS interface
Connection on the optional slot (J3 CPU board)

20
Two wires passive RS485 ports half-duplex
Simultaneous communication on all ports without restrictions
Configurable speed up to 38400 bauds
Communication protocol depending on the application

The two ports are passive meaning that they must be powered by the intrinsic safety interface that will be
connected on Corus. Each port is isolated separately and can be connected to two different types of devices
(modems, RTUs, PCs, etc).
Suggested intrinsic safety interfaces are:
ISB+ (Itron) for both data signals and port powering

or
the barrier Z961 (from Pepperl &Fuchs) for data signals
the BXNE31 (from Georgin) power supply for powering the port.
A separate power supply and barrier must be used for the second port.

21
6 INSTALLATION, COMMISSIONING AND MAINTENANCE

6.1 Fixing

Corus enclosure features four lateral holes (diameter 5.2mm) for wall-mounting or for mounting on special
plate. Product shall be installed in a place protected from direct exposure to sunlight and rain.
Following the installation procedure, Corus has to be carefully closed using the four screws placed on each
corner. All cable glands have moreover to be tightened enough. These two actions are mandatory to
achieve a correct water tightness of the enclosure.

Important notice: Cable glands must be screwed exclusively by hand.

6.2 Connectors and terminals

On/Off Pulse output #1

External power supply (opt.)


On/Off Pulse output #2
PSTN line or RS485 bus or P2 sensor

LF and tamper cable


T sensor

P sensor

RS232
Optional input :

J1 J3
T sensor Meter
Brown

Shield
Shield

Shield
White

White

Grey
Red

Red

1 2 3 4 5 6 1 2 3 J5 J6

1 2 3 4 5 6 7 8 9 10 1 2 3 4 5
1 2 3 4 5 6 1 2 3 4
IN 1 -

IN 2 -

OUT 1 -

OUT 2 -

EXT -

TxD
RxD
DTR
+6V
+0V
IN 1+

IN 2+

OUT 1+

OUT 2+

EXT+
Yellow

Yellow
Brown
White
Green

White

Green
Pink

Grey

Pink

P sensor Meter On / Off On / Off Ext. RS 232


J2 J4 Inputs Outputs Pow.

22
6.2.1 Pressure and temperature sensor

J1.1 shield (T sensor) J2.1 pink (P sensor)


J1.2 red (T sensor) J2.2 yellow (P sensor)
J1.3 white (T sensor) J2.3 gray (P sensor)
J1.4 red (T sensor) J2.4 brown (P sensor)
J1.5 white (T sensor) J2.5 white (P sensor)
J1.6 shield (P sensor) J2.6 green (P sensor)

6.2.2 LF cable
J3.1 Tamper + (brown) J4.1 LF1 + (yellow)
J3.2 Tamper (gray) J4.2 LF1 (white)
J3.3 shield (LF cable) J4.3 LF2 + (pink)
J4.4 LF2 (green)

6.2.3 On/off inputs, on/off outputs, external power supply

J5.1 on/off in1+


J5.2 on/off in1-
J5.3 on/off in2+
J5.4 on/off in2-
J5.5 on/off out1+
J5.6 on/off out1-
J5.7 on/off out2+
J5.8 on/off out2-
J5.9 ext.supply+
J5.10 ext.supply-

6.2.4 RS232 interface

J6.1 TxD
J6.2 RxD
J6.3 DTR
J6.4 Supply+
J6.5 Supply-

23
6.3 Installation of the sensors

6.3.1 Temperature sensor

The temperature sensor has to be placed into a protective thermowell at a point of the pipe where the
measured temperature is representative of the gas temperature through the meter.
The generally accepted installation conditions are:
distance from the meter: less than 1 meter
position: downstream for turbines
immersion : 1/3 to 2/3 of the internal diameter of the pipe
The presence of a second thermowell located close to the first one is recommended to check Corus
temperature measurement, using a reference thermometer, without having to remove the probe.

6.3.2 Pressure sensor

The pressure sensor connector has a G1/4 (BSP) male terminal. It must be preferentially screwed to the "Pr"
reference pressure tap of the meter:
either directly (through a mechanical adapter if necessary)
or through a 3-way valve (which will avoid any modification of the installation during the periodical check
of the sensor)
If a failure of the pressure sensor makes it mandatory to change it, the pressure coefficients have to be
reprogrammed with the ones delivered with the new sensor. If Corus is in operation during the change, a
pressure alarm will be generated.
The pressure sensor features two (2) holes allowing the sealing the sensor with the meter installation.

Important notice:
- The pressure sensor has a gland for its cable which shall not be screwed or unscrewed by the user. Any
additional screwing can damage the wires of the cable.
- When connecting the sensor on the meter, the cable of the sensor shall not be twisted.
Twisting the cable can damage the wires of the sensor permanently.
Sensors shall be mounted either by using an adaptor which allows the sensor to be screwed without
turning its body, either by dismounting the sensor from Corus, screwing it to the meter (by turning the
body of the sensor together with the cable), then reconnecting it to the Corus.

6.3.3 LF cable

The LF cable provided with the CORUS has to be connected directly to the LF output plug of the gas meter.
The various LF cable (Binder plug ) available are :
Type 1 : Type 2 : Type 3 : Type 4 :
Pins 1-4 : LF1 Pins 4-6 : LF1 Pins 1-2 : LF1 Pins 1-4 : LF1
Pins 2-5 : LF2 Pins 3-5 : LF2 Pins 5-6 : LF2 Pins 2-5 : LF2
Pins 3-6 : Strap Pins 1-2 : AT Pins 3-4 : AT Pins 3-6 : AT

Each LF cable binder plug features two (2) holes allowing sealing the CORUS cable to the meter installation.

24
6.4 Power supply

Corus can be supplied either:


by its internal battery
by an external power supply

6.4.1 Internal battery

Corus is delivered with a 16.5A.h D-size lithium battery that provides autonomy for 5 years in average
conditions. This battery includes all protections required by intrinsic safety and cannot be changed for a
different type.
When the battery reaches the end of its life, it has to be replaced according to the following procedure that
ensures a non-stop operation of the product:
do not provoke short-circuits on the new battery : it would be definitively damaged
check the voltage of the new battery : it should be above 3.6V
open the cover of Corus box
connect the new battery on the free battery connector (J7 or J8)
disconnect the old battery and remove it from the battery support
put the new battery into the battery support
reprogram the battery life to its nominal value (1825 days)
close the enclosure and check Corus

6.4.2 External power supply

Corus can also be supplied by an external intrinsic safety associated module from:
a 220 VAC mains
a 24 VDC mains
The corresponding module has to be connected to the right hand side of the terminal J5 (J5.9 and J5.10).
The converter must then be configured in the 'external power supply' mode. In this mode, the internal
battery has to remain plugged in case of external power cut.
For safety reasons, it is strictly forbidden to use other modules than those specified by 'Itron'.

6.4.3 Autonomy

The table below shows indicative (see notes) autonomy of CORUS in years when using the battery mode,
according the 3 main parameters: Acquisition period, Retransmissions, Communication.

25
Retrans No Retrans. F = 0,1 Hz F = 0,5 Hz F = 1 Hz F = 2 Hz Comm.
Acq. Period Duration
1 second 1 1 1 0,9 0,8 15min./ month
1 1 1 0,9 0,8 15min./ week
5 seconds 3,7 3,6 3,2 2,7 2,1 15min./ month
3,7 3,6 3,1 2,7 2,1 15min./ week
10 seconds 5,7 5,4 4,4 3,6 2,7 15min./ month
5,6 5,3 4,4 3,6 2,6 15min./ week
15 seconds 6,9 6,4 5,1 4,1 2,9 15min./ month
6,7 6,3 5,0 4,0 2,9 15min./ week
20 seconds 7,7 7,1 5,6 4,3 3,0 15min./ month
7,5 6,9 5,4 4,2 3,0 15min./ week
30 seconds 8,7 8,0 6,1 4,7 3,2 15min./ month
8,4 7,8 5,9 4,6 3,1 15min./ week
The same table can be given in number of days. The user must set the starting value of the counter of
battery life time into Corus according to the corresponding conditions of use. By default the value
programmed is 1825 days, meaning 5 years.

Retrans No Retrans. F = 0,1 Hz F = 0,5 Hz F = 1 Hz F = 2 Hz Comm.


Acq. Period Duration
1 second 365 365 365 329 292 15min./ month

365 365 365 329 292 15min./ week


5 seconds 1351 1314 1168 986 767 15min./ month

1351 1314 1132 986 767 15min./ week


10 seconds 2081 1971 1606 1314 986 15min./ month
2044 1935 1606 1314 949 15min./ week
15 seconds 2519 2336 1862 1497 1059 15min./ month

2446 2300 1825 1460 1059 15min./ week


20 seconds 2811 2592 2044 1570 1095 15min./ month

2738 2519 1971 1533 1095 15min./ week


30 seconds 3176 2920 2227 1716 1168 15min./ month

3066 2847 2154 1679 1132 15min./ week

26
Note 1:
The values provided in this table correspond to typical values in average conditions of use. They are
calculated with a 30% margin to take in account the possible variations of operating conditions that could
affect the consumption of the product (temperature, humidity, self-discharge of the battery)

Note 2:
The frequency of retransmissions indicated in the table corresponds to average frequency on both outputs,
when both channels are configured for pulses.

Note 3:
In case of activation of a digital output as ALARM output, the consumption of the device can be impacted,
especially in Standard mode for which the output is activated as long as the alarm is active. Then, it is
recommended to use preferably the limited mode or to use CORUS with an external power supply module
if Standard mode is required.

Note 4:
The values provided in this table correspond to a Corus without any optional board.

Note 5:
The use of AGA8 detailed formula reduces the battery life time, due to increased calculation time necessary
for this formula.

6.5 Programming

To program/modify parameters protected by the programming switch, the following procedure has to be
followed:
Open CORUS
Remove the seal of the white metrological cover, placed on the left part of the I/O board
Put the programming switch On
Program the required parameters (Keyboard or communication)
Put the programming switch Off
Place the metrological cover and seal it
Close CORUS
To program/modify parameters protected by the Client switch, the following procedure has to be
followed:
Open CORUS
Put the Client switch On
Program the required parameters (keyboard or communication)
Put the Client switch Off
Close CORUS

27
6.6 Pressure and temperature adjustment on site

It is possible to adjust the pressure and temperature values through a linearization function available
within Wincor software tool.

P = (a x P) + b
T = (c x T) + d
The coefficients a, b, c, d can be introduced in Corus, after an on-site verification, performed on 2 different
operating measurement values.

P(theoretical_1) => P(measured_1)


=> Determination of a and b
P(theoretical_2) => P(measured_2)

T(theoretical_1) => T(measured_1) => Determination of c and d


T(theoretical_2) => T(measured_2)

The use of this function is subject to national regulations. Refer to your local authorities before using it.

28
7 MID: COMPLIANCE WITH WELMEC 7.2 GUIDE

Corus complies with MID directive (2004/22/EC) and consequently to the corresponding harmonized
standard, EN12405-1/A1. Additionally, the internal firmware of the product complies with the WELMEC 7.2
guide. This document provides guidance for software-equipped measuring instruments within the scope of
the MID directive.
Corus complies with the following requirement sets of this guide:
P : Embedded software
I : Instrument specific requirements (converter)
S : Software separation

7.1 Firmware structure

Corus firmware is split in to 2 separate microcontrollers but must be considered functionally as based on 3
parts:
F1: One metrological firmware dedicated to analogue measurements on the I/O board (metrol).
F2: One metrological firmware (kernel) that controls the heart of the product and ensures the integrity
of all metrological functions.
F3: One non-metrological firmware managing all non-metrological functions (called application ).
F2 and F3 are included into the same microcontroller on the CPU board.

C1 C2

F2 firmware (kernel) F1 firmware (metrol.)

F3 firmware
(application)
CPU board I/O board

So, the part of firmware which contains all legally relevant functions is clearly separated from other parts of
firmware. The Kernel manages the non-legally relevant firmware tasks and the priority is given to the
metrological tasks. Application firmware can be updated without breaking the metrology seals.

The Kernel integrates:


Metrological functions.
Management and protection of the metrological parameters.
Management and protection of the measurement data.

29
Protected interface with other parts of the firmware.
Backup of the metrological parameters and measurement data.
Verification of the kernel, parameters and data integrity.
Display of metrological measurements and parameters.

7.2 Firmware identification

It is possible to get information (version, date) related to the 3 different firmware versions in the first INDEX
sub-menu.
INDEX VALUE ALARM DATA CONFIG

06/09/2008
14:00:00
Kernel : Kr1.00 13/04/07
Metrol : Met1.00 19/02/03
Appli : Ver5.00 12/08/08

7.3 Firmware protection and data control (CRC)

To ensure the integrity of metrological firmware and associated parameters, Corus manages 4 CRC (cyclic
redundancy code):

Metrological firmware CRC check (Metrol and Kernel) :


Corus checks every hour the CRC of the 2 metrological firmware parts (Metrol and Kernel) and displays
CRC
the CRC value. If the result is not OK, a specific icon ( ) is displayed.
Note: when this screen is active, it is possible to proceed to a live CRC calculation check by pressing the
OK key for 2 seconds.
The upload of the Kernel and Metrol firmware is protected by the programming switch. Then, no change
of the metrological firmware is possible without breaking the metrological seal.
In case of error, a CRC alarm is generated and the corresponding event is provided to the application for
storage.

Parameters CRC check :


Corus manages a CRC related to all metrological parameters (P & T thresholds, input pulse weight). This
CRC is checked every 10 minutes or is recalculated automatically when the value of such parameters is
modified on request.
In case of error, a CRC alarm is generated and the corresponding event is provided to the application for
storage.

30
Index CRC check :
Corus manages a CRC related to index & counters (converted and unconverted).
This CRC is checked every second, corresponding to the refreshment period of the index.
In case of error, a CRC alarm is generated and the corresponding event is provided to the application for
storage.

INDEX VALUE ALARM DATA CONFIG

Kernel CRC : 0CAFE


Metrol CRC : 310EF
Param CRC : FE47
Index CRC : OK
Prog. switch : On
Cust. switch : Off
Alarm P: Off Alarm T: Off

7.4 Data protection Data access


All parameters and index access can be protected by a hardware and/or software protection, except main
index (unconverted and converted), which are protected by hardware programming switch in any case.

The read access is free and the write access can be protected by different means.
The access level of each parameter is defined in a table called Table protect.
The table protect itself is protected by the programming switch, meaning that it is not possible to modify
an access right without breaking the metrological seal of the product.

Hardware protection:
Programming switch
This switch has to be turned ON before any change on parameters declared as protected by
this switch. Access to this switch is protected by metrological seal and then prevent un-
authorized access to metrological data
Client switch
This switch has to be turned ON before any change on parameters declared as protected by
this switch. Access to this switch is protected by client seal (if any).

Software protection:
Software protection modes are approved according MID requirements. This includes flexible management
of password/keys/user groups for reading and programming.
For more information please contact your local Itron office.

31
7.5 Index display format

In order to comply with the MID requirement stating that the display shall have a sufficient number of
digits to ensure that both unconverted and converted index incremented during 8000 hours at maximum
conditions (Qmax, Cmax) do not return to its initial value, Corus display format is as follows, depending
dynamically on the index value itself.

Value (m3) Format


Max < 99999999
00000000.000
99999999 < Max < 999999999
000000000.00
999999999 < Max < 9999999999
0000000000.0
9999999999 < Max < 99999999999
00000000000
99999999999 < Max < 999999999999
000000000000

Note :

The raw and base indexes are displayed on the first screen.
This screen is updated every second.

INDEX VALUE ALARM DATA CONFIG

Unconverted index :
00012345.000m3
Converted index :
00054321.100Nm3

32
8 FUNCTIONAL DESCRIPTION

Note: The legally relevant firmware (Kernel & Metrol) is then approved under MID/ Welmec 7.2.
Through the application of the S requirements set, the rest of the firmware (Appli.) has no influence on
metrology and can be adapted to various markets requirements (e.g : database, non-metrological alarms,
communication protocol, display of non-metrological items, digital I/Os management, etc)

8.1 Database

Corus manages six logs :


an hourly log
a daily log
a monthly log
an interval log
an events log
a parameters log

8.1.1 Hourly Log

The hourly log contains up to 1440 records (2 months data) that include:
the date and time
the hourly unconverted consumption
the hourly converted consumption
the hourly under alarm unconverted consumption
the hourly total converted consumption
the value of unconverted index (end of period)
the value of converted index (end of period)
the value of under alarm unconverted counter (end of period)
the value of total converted counter (end of period)
the minimum, maximum and average gas temperature
the minimum, maximum and average gas pressure
the minimum, maximum and average second gas pressure (if P2 optional board)
the minimum, maximum and average unconverted flow rate
the minimum, maximum and average converted flow rate
the status of the converter during the hour (presence of metrological alarms, Date/Time change)

8.1.2 Daily Log

The daily log contains up to 124 records (4 months data) that include:
the date and time
the daily unconverted consumption
the daily converted consumption

33
the daily under alarm unconverted consumption
the daily total converted consumption
the value of unconverted index (end of period)
the value of converted index (end of period)
the value of under alarm unconverted counter (end of period)
the value of total converted counter (end of period)
the minimum, maximum and average gas temperature
the minimum, maximum and average gas pressure
the minimum, maximum and average second gas pressure (if P2 optional board)
the minimum, maximum and average unconverted flow rate
the minimum, maximum and average converted flow rate
the status of the converter during the day (presence of metrological alarms, Date/Time change)
The user can introduce a "gas hour" in Corus. All daily consumptions are stamped from the gas hour to the
gas hour of the next day.

8.1.3 Monthly Log

The daily log contains up to 24 records (2 years data) that include:


the date and time
the monthly unconverted consumption
the monthly converted consumption
the monthly under alarm unconverted consumption
the monthly total converted consumption
the minimum, maximum and average gas temperature
the minimum, maximum and average gas pressure
the minimum, maximum and average second gas pressure (if P2 optional board)
the minimum, maximum and average unconverted flow rate
the minimum, maximum and average converted flow rate
the status of the converter during the month (presence of metrological alarms, Date/Time change)
the value of the main unconverted index at the end of the month
the value of the main converted index at the end of the month
the value of the unconverted counter at the end of the month
the value of the converted counter at the end of the month
the maximum unconverted consumption during the month (interval) with corresponding date/time
the maximum converted consumption during the month (interval) with corresponding date/time

The user can introduce a "Gas day" in Corus. All monthly consumptions are stamped from the gas day (at
the gas hour) to the gas day of the next month (at the gas hour).
The user can select the items to store in calendar logs (Hourly, Daily, Monthly) for each record. However
some items will always be recorded and the user cant unselect them. By reducing the number of items, the
number of database records is increased.
The fixed part of the databases includes:
For the hourly and daily logs
the unconverted consumption
the converted consumption
the under alarm unconverted consumption

34
the total converted consumption
date/time and status

For the monthly log


date/time and status
the monthly unconverted consumption
the monthly converted consumption
the monthly under alarm unconverted consumption
the monthly total converted consumption
the value of the main unconverted index at the end of the month
the value of the main converted index at the end of the month

Its to be noted that the user can select to store only the integer part of index values to save space in
database.

8.1.4 Interval Log

This log is also made up of two parts:


a fixed one
a configurable one
The configurable part allows the user to build his own database structure according to his needs.
The fixed part includes the following data:
the date and time
the interval unconverted consumption
the interval converted consumption
the interval under alarm unconverted consumption
the interval total converted consumption
a status of the converter (presence of metrologic alarms, Date/Time change)
In the configurable part, the user can select the following fields:
the minimum, maximum and average gas temperature
the minimum, maximum and average gas pressure
the minimum, maximum and average unconverted flow rate
the minimum, maximum and average converted flow rate
the value of unconverted index (end of period)
the value of converted index (end of period)
the value of under alarm unconverted counter (end of period)
the value of total converted counter (end of period)
the minimum, maximum and average second gas pressure (if P2 optional board)
The value of the interval is programmable among 1, 2, 3, 4, 5, 10, 15, 20, 30 and 60 minutes. The capacity of
the interval log extends from 3100 (all optional fields selected) to 5900 (no optional field selected).

8.1.5 Events Log

Corus logs the occurrence of all main events (alarms, reset, switches change, programming).
For each event, the following data are available:
Date and time of the event
Nature of the event

35
Specific value (for some events only)
Status
The events log has a capacity of 800 events and cannot be reset. It is circular: when it is full, every new
event replaces the oldest one.

8.1.6 Parameters Log

Corus logs the modifications of the following key parameters:


temperature specific data: alarm thresholds,
pressure specific data: alarm thresholds, coefficients,
compressibility ratio specific data: gas composition, calorific value,
conversion specific data : formula,
volume specific data: index, input pulse weight,
database specific or related data: interval duration, gas hour and day
In case of modification of one of the above parameters, the following information is stored:
date and time
nature of the parameter
former value of the parameter
new value of the parameter
values of the unconverted and converted indexes
The parameters log has a capacity of 200 modifications and cannot be reset. It is circular: when it is full,
every new modification replaces the oldest one.

8.2 Alarms
Metrological part of Corus manages alarms as described below.
Others types of alarms depend on the application (non-metrological alarms) :

T low threshold
temp.
metrological alarms

T high threshold
T sensor failure

P low threshold
pressure

P high threshold
P sensor failure

CRC 'Metrol' firmware


kernel alarms

CRC 'Kernel' firmware


kernel

CRC 'Parameters'
CRC 'Index'
Metrol task

36
The rest of the alarms that are managed are:

T low threshold

pressure temp.
T high threshold
Warning thersholds

P low threshold
P high threshold
pressure

P2 low threshold
second

P2 high threshold
P2 sensor failure

pulse inputs coherence


inputs

tamper input
on/off input #1
on/off input #2
conv.

C factor low threshold


C factor high threshold
Other alarms

vol.

base cons. high threshold

unconv. flow low threshold


unconv. flow high threshold
flow

base flow low threshold


base flow high threshold
supply

remaining battery life alarm


external power supply

37
8.2.1 Pressure alarms

Pressure alarms occur:


when the gas pressure is out of the operating range [Pmin ; Pmax]
when the pressure sensor is out of order.
Corus manages three pressure alarms:
a pressure low threshold alarm
a pressure high threshold alarm
a pressure sensor failure alarm
These alarms are automatically disabled when a T conversion is used.
When a pressure alarm occurs:
the alarm icon and the 'P' icon are displayed (blinking while the alarm is active)
the incrementing of the main converted index is stopped
if the alarm is due to a pressure sensor failure :
pressure, compressibility and conversion factor are set to "not-available" on the display
the total converted counter is incremented based on the back-up pressure value (if existing)
the on/off outputs activated as alarm outputs are switched 'on' , if the corresponding alarms were
configured for the alarm output
the corresponding alarm is set into the 'ALARM' menu and is stamped in the events log

8.2.2 Temperature alarms

Temperature alarms occur:


when the gas temperature is out of the operating range [Tmin ; Tmax]
when the temperature sensor is out of order.
Corus manages three temperature alarms:
a temperature low threshold alarm
a temperature high threshold alarm
a temperature sensor failure alarm
When a temperature alarm occurs:
the alarm icon and the 'T' icon are displayed (blinking while alarm is active)
the incrementing of the main converted index is stopped
if the alarm is due to a temperature sensor failure :
temperature, compressibility and conversion factor are set to "not-available" on the display
the total converted counter is incremented based on the back-up temperature value (if existing)
the on/off outputs activated as alarm outputs are switched 'on' , if the corresponding alarms were
configured for the alarm output
the corresponding alarm is set into the 'ALARM' menu and is stamped in the events log

38
8.2.3 Kernel alarms
The kernel firmware manages 5 alarms related to the control of firmware and metrological parameters
integrity:

Alarm CRC Metrol : active if the CRC of the Metrol firmware is not correct
Alarm CRC Kernel : active if the CRC of the Kernel firmware is not correct
Alarm CRC Param : active if the CRC of the metrological parameters is not correct
Alarm CRC Index : active if the CRC of the Index is not correct
Alarm Metrol. task : active if the kernel metrological task duration is higher than 1 second. In any case,
after 1 second, priority is given to the kernel and the corresponding application task is stopped.

8.2.4 Warning threshold alarms


Warning for pressure, temperature and second pressure input have no impact to the operation of Corus and
their thresholds have no connection with thresholds for metrological P and T alarms. Warnings are set to
monitor the functional operation of the station and therefore have no connection with metrology.

8.2.5 LF coherence alarm

This alarm can be enabled or disabled by programming.


The user can program a number of pulses (Np) and duration (Dp). If the difference between the number of
pulses on input LF1 and input LF2 is higher or equal to Np during a period Dp expressed in days (see 4.2),
Corus generates an alarm.
This alarm is not metrological, meaning that it does not impact the management of the unconverted and
converted indexes.
When the alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the LF coherence alarm was configured
for the alarm output
the LF coherence alarm is set into the 'ALARM' menu and is stamped in the events log

8.2.6 Tamper alarm, on/off1 alarm, on/off2 alarm

These alarms can be enabled or disabled by programming.


When enabled, the tamper input and the on/off inputs can be set as normally open or normally closed (see
4.1). Then, when Corus detects that an input is not in its normal state, the corresponding alarm is
generated.
When such an alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for alarm
output
the alarm is set into the 'ALARM' menu and is stamped in the events log

39
8.2.7 Conversion factor alarm

If this function is activated, Corus generates this alarm when the computed conversion factor gets out of a
programmed range [Cmin ; Cmax]. This alarm is not metrological and is dedicated to the control of the
whole metering chain.
Corus manages two conversion factor alarms:
conversion factor low threshold alarm
conversion factor high threshold alarm
When a conversion factor alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the conversion factor alarm is set into the 'ALARM' menu and is stamped in the events log

8.2.8 Converted consumption alarm

Corus sets this alarm when the converted consumption during a given interval is higher than a programmed
threshold. The interval can be selected among:
the current database interval
the ongoing hour
the ongoing day
the ongoing month.
When the alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the alarm is set into the 'ALARM' menu and is stamped in the events log

8.2.9 Flow-rate alarms

Corus manages four flow alarms:


unconverted flow low threshold alarm
unconverted flow high threshold alarm
converted flow low threshold alarm
converted flow high threshold alarm
Every time the unconverted and converted instantaneous flows are calculated, their values are compared to
four thresholds that are programmable by the user. If the function is activated, an alarm among the above
list is set, if the corresponding flow-rate gets higher or lower than the threshold.
When such an alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the alarm is set into the 'ALARM' menu and is stamped in the events log

8.2.10 Battery alarm

Every 24 hours, the internal battery counter is decreased of one day. When the counter reaches 10% of the
nominal battery life (about 6 month (182 days) for a 5 years battery life), the battery alarm is generated.

40
When changing the battery, the battery counter has to be re-programmed so as to acknowledge the battery
alarm.
When the battery alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for the alarm
output
the alarm is set into the 'ALARM' menu and is stamped in the events log
The battery alarm is not metrological, meaning that when the battery alarm occurs, Corus keeps working
until the battery is fully discharged.

8.2.11 External power supply alarm

When Corus is supplied through external power, it is able to detect a mains cut.
Then, this alarm is set as long as the external supply is not present.
During the mains cut, the internal backup battery will ensure the normal operation of Corus and the battery
counter will be decreased as described in the "battery alarm" chapter.
When such an alarm occurs:
the alarm icon is displayed
the on/off outputs activated as alarm outputs are switched 'on', if the alarm was configured for alarm
output
The alarm is set into the 'ALARM' menu and is stamped in the events log

8.3 Alarms Reset


The alarms can be reset either via keyboard, or via software. This action is allowed according to data access
rights that have been programmed in CORUS (see 7.4).

Through keyboard :
Go to the menu ALARM
Select reset
Select all alarms
Press OK to confirm

Through software
Go to the Reset alarms menu of the configuration tool
Select type of alarm to be reset : Metrologic (P or T) and/or Non metrologic
Click on Prog button

Note 1: it is not possible to reset an active alarm, except the Coherence alarm and the Consumption
threshold alarm.

41
9 USER INTERFACE

Corus features a graphic liquid crystal display and a five-key keyboard. The five keys allow scrolling through
the various menus as described below.
The interface architecture is based on five main menus:
INDEX, VALUE and ALARM menus are always displayed. DATA and CONFIG menus can be displayed or not
according configuration.
INDEX VALUE ALARM DATA CONFIG

Also, Corus manages the following icons:

the turbine icon is refreshed every second and is:


on if any pulse has been metered since the last screen refresh
off otherwise
the bell icon:
blinks when any alarm is active
is on when any alarm is memorized but none is active
is off when all alarms are off (neither active nor memorized)
the pressure icon:
blinks when any pressure alarm is active
is on when any pressure alarm is memorized but none is active
is off when all pressure alarms are off (neither active nor memorized)
the temperature icon:
blinks when any temperature alarm is active
is on when any temperature alarm is memorized but none is active
is off when all temperature alarms are off (neither active nor memorized)
the phone icon is:
on when any communication is ongoing
off otherwise
the plug icon:
is off when CORUS is used in battery operating mode
is on when CORUS is used in external supply mode and main power is present
blinks when CORUS is used in external supply mode and main power is not
present

the inner bar graph of the battery icon is set according to the remaining battery life
time ratio in battery and external power mode (if there is a mains cut).
it blinks when the battery alarm is active
the key icon:
is blinking when the shunt of optical port detection is activated.
is off otherwise

42
9.1 Keyboard

The left/right arrow keys allow scrolling through the menus. Items can be selected using the 'OK' key. When
the display is off, any press on a key puts the display on. Access to the DATA and CONFIG menus can be
restricted according the application and the display configuration which is selected. After 2 minutes with no
action on the keyboard, the display switches off. When pressing a key, the display automatically starts on
the INDEX menu.
When the display is on, the acquisition of pressure, temperature and the calculation of compressibility and
conversion factor is performed every second.

9.2 Display

Below is shown the display sequence managed directly by the Kernel. Content and access to other(s)
menu(s) depend on the application and the display configuration which is selected.

43
Kernel Stand alone

INDEX VALUE ALARM DATA CONFIG


DATA INDEX VALUE ALARM DATA CONFIG

Unconverted index : P = 15. 698 bar


T = -13. 5 C
00012345.000m3 Zb/Z = 1. 0035 Non-Metrological
Converted index : C = 5. 6454 Add
Bat = 250 days
00054321.100Nm3 Non-Metrological

OK OK

INDEX VALUE ALARM DATA CONFIG INDEX


INDEX VALUE
VALUE ALARM
ALARM DATA
DATA CONFIG
CONFIG

Kernel CRC : 0CAFE


Metrol CRC : 310EF 13/04/07
06/09/2008
Param CRC : FE47 OK 07:18:00
14:00:00 Non-Metrological
Index CRC : OK
Prog. switch : On Kernel : Kr1.00
V1.00 13/04/07
13/04/07 Add
Metrol : Met1.00
V1.00 08/07/02
Cust. switch : Off 2s Appli : Ver5.00
19/02/03
Ver3.00 25/12/07
12/08/08
Alarm P: Off Alarm T : Off

OK OK

INDEX VALUE ALARM DATA CONFIG

Serial N: SC08000001234
-----------------
P Serial N: KB08060123 Non-Metrological
Pmin : 0.90000 bar
Pmax: 10.00000 bar Add
-----------------
T Serial N: CB05080546

OK

Test LCD

9.2.1 'INDEX' menu

This menu provides access to the following items:


display of unconverted and converted index
adjustment of the contrast of the display.
display of date and time
display firmware versions and associated dates
display of the 4 CRCs managed by the kernel
status of metrological alarms (P and T)
status of Prog. And Client switches
display of main parameters (sensors serial numbers, P and T thresholds, type of conversion, input pulse
weight,.)

44
display test
When associated to an application, this menu (and associated sub-menus) can be enhanced with other(s)
data.

9.2.2 'VALUE' menu

This menu displays the instantaneous values that are measured or calculated by Corus:
pressure
temperature
compressibility factor
conversion factor
battery life
When associated to an application, this menu can be enhanced with others data (flow-rates, second
pressure P2,), display of a graph,

9.2.3 'ALARM' menu

The content of this menu depends on the application and the associated display configuration which is
selected. As an example, this menu usually allows to:
get information related to active alarms
get information related to memorized alarms
get information related to under alarm counters
reset alarms

Active alarms
The active alarms submenu displays the list of all active alarms. Using the up/down arrows and the OK keys,
it is possible to get more details about a given active alarm (Start time-stamping)

Memorized alarms
The memorized alarms submenu displays the list of all memorized alarms. Using the up/down arrows and
the OK keys, it is then possible to get more details about a given memorized alarm (Start and End time-
stamping)

Under alarm counters


This submenu displays the under alarm unconverted counter and the total converted counter.

9.2.4 Alarms reset

This submenu allows to reset all memorized alarms and/or the under alarm counters (according to the
status of the prog switch for metrological alarms)

45
9.2.5 'DATA' menu

The content of this menu depends on the application and the associated display configuration which is
selected. For each active log, it is possible to consult the content of each record stamped in the
corresponding log of the product.

9.2.6 'CONFIG' menu


The content of this menu depends on the application and the associated display configuration which is
selected. This menu allows the user to program the main parameters of Corus, allowing a basic
commissioning without laptop:
Authorization to modify a parameters value through keyboard is related to the access rights (see 7.4)

Communication : Modbus protocol activation and communication parameters for each com port
Volume : Input pulse weight, unconverted Index, converted Index
Metrology : Pbase, Tbase, Gas composition (according formula)
Date / Time : Current date / time
Supply : Mode (battery, external) , autonomy
Digital Output 1 : Activation/Deactivation , type (pulse, alarm,4/20mA) , parameters
Digital Output 2 : Activation/Deactivation , type (pulse, alarm,4/20mA) , parameters

46
INDEX MENU

Index menu
INDEX VALUE ALARM DATA CONFIG

contrast+
Unconverted index :
00012345.000m3 config
menu
Date / Time
Versions
value
menu
Converted index :
00054321.100Nm3 contrast-

Date / Time and Versions menu


INDEX VALUE ALARM DATA CONFIG

06/09/2008
14:00:00 config
menu
Status value
menu
Main metrology
parameters
(press 2s )
Kernel : Kr1.00 13/04/07
Metrol : Met1.00 19/02/03
Appli : Ver5.00 12/08/08

Status menu

INDEX VALUE ALARM DATA CONFIG

Add. Board None


Conversion type SGERG
On/off1 Normal config Index value
config LCD Test value
menu (press 2 s) menu
On/off2 Normal menu menu menu

Tamper Alarm
Prog. switch On
Cust. switch On

Display test (progressing test)

Index
menu

47
VALUE MENU

INDEX VALUE ALARM DATA CONFIG


previous
P = 15.698 bar item
T = -13.5 C
Zb/Z = 1.0035 Index s how Alarm
C = 5.6454 menu Graph menu

Qm = 152.5 m3/h
Qb = 153.9 Nm3/h next
item
Bat = 250 days
Value menu

Graph
INDEX VALUE ALARM DATA CONFIG
previous item
4.2 graph

Index Value Alarm


menu menu menu

4.0 next item


graph
P (bar) 4.025

Instantaneous
value

48
ALARM MENU

Alarm menu
INDEX VALUE ALARM DATA CONFIG

Alarms:
02active
02 active alarms
alarms
00 memorized alarms

Under alarm counters


Reset

List of alarms Reset Under alarm counters


INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG

Pressure sensor failure Reset: Under alarm counters:


Tamper All alarms Unconverted :
Under alarm counters 00000012.000 m3
Converted :
00000018.567 Nm3
Active alarms 1/2

previous upper
alarm item

back to see details back to back to reset back to


alarms for selected alarms alarms selected alarms
menu alarm menu menu item menu

next lower
alarm item

Details of selected alarm Confirmation screen

INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG

Pressure sensor failure Warning:

start: 14/08/2002 16:03:41 All alarms


end : --/--/---- --:--:-- will be reset.
Press OK to
Active alarms 1/2 confirm.

back to
reset
menu

back to confirm back to


reset reset reset
menu menu

back to
reset
menu

INDEX VALUE ALARM DATA CONFIG

Alarms reset :
Completed

49
DATA MENU

Data menu

INDEX VALUE ALARM DATA CONFIG


upper
Hourly log menu
item
Daily log
Monthly log alarms enter config
Interval log menu selected
submenu menu
Event log
Parameters log lower
menu
item

Date selection
INDEX VALUE ALARM DATA CONFIG
previous
14/08/2002 date
13/08/2002
12/08/2002 back to
database
display
records
back to
database
11/08/2002 menu on date menu
10/08/2002
09/08/2002 next
date
Hourly log 1/60
Hour selection

INDEX VALUE ALARM DATA CONFIG previous


record
16:00 > Now
15:00 > 16:00
back to choose back to
14:00 > 15:00 date selected date
selection record selection
13:00 > 14:00
12:00 > 13:00 next
11:00 > 12:00 record
14/08/2002 2/24
Data

INDEX VALUE ALARM DATA CONFIG


scroll up
Time : 60 mn
UV : 152.000 m3 back to back to
CV : 165.395 Nm3 record
selection
record
selection
UV (Al) : 0.000 m3
CV (Al) : 165.395 Nm3 scroll down
Hourly 16/08/2002
21:00 > 22:00 1/18

50
CONFIG MENU

Config menu
INDEX VALUE ALARM DATA CONFIG

Communication
Volume
Metrology
Date/Time
Supply
Output 1
Output 2

Communication

OK for
modification

Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys

Volume
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG

Pulse weight : 1.000 m3 Current value


UV Index : 000143.000 m3 UV Index : 000143.000 m3
CV Index : 000165.321Nm3 OK for

000185.000 validation

Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys
Metrology
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG

Current value
Pb : 1.01325 bar Gas rel. density : 0.57000
Tb : 273.15 K OK for

Gas rel. density : 0.57000


C02 : 1.8 % 0.57000 validation
N2 : 0.6%

Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys
Date/Time
INDEX VALUE ALARM DATA CONFIG
INDEX VALUE ALARM DATA CONFIG
Current value
Date: 07/06/2003 07:43:00 Date : 07/06/2003 07:43:00
07/ 06/2003 OK for

validation
07:43:00

OK Selection of the digit with left and right keys


Adjustment of value with up and down keys
Supply
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG

Battery : 1794 days Current value


Mode : Battery Mode : Battery
OK for

Battery validation

Selection with up and down keys + OK Selection with up and down keys

51
Output 1 or 2
INDEX VALUE ALARM DATA CONFIG INDEX VALUE ALARM DATA CONFIG
Mode : Unconv. Pulse Current value
Pulse Weight : 1.000 m3 Pulse Weight : 1 m3
Time closed : 250 ms OK for


1.000 m3 validation


Selection with up and down keys + OK Selection of the digit with left and right keys
Adjustment of value with up and down keys

52
10 SAFETY INSTRUCTIONS

10.1 General

Corus gas volume converter is approved according to 94/9/CE directive (ATEX) relative to introduction of
material in explosive atmosphere area.
The respect of this directive is mandatory (from July, 1st 2003) and is included in the 'CE' marking.
Thus, 'CE' marking of Corus implies now conformity with the following directives:
94/9/EC (ATEX), 89/336/EEC (EMC) and 2004/22/EC (MID).

10.2 Marking relative to ATEX - Zone

CORUS is approved as category 1 equipment and is thus suitable for use in zone 0 (areas with permanent
presence of gas).

The corresponding marking is: 0081 LCIE 03 ATEX 6165X II1G Ex ia IIC T4
The ambient temperature range for operation is: -25C; +55C

10.3 Specific Instructions

To avoid erroneous operations, it is recommended to read the whole operating manual before
commissioning.
Corus is suitable for being used with gases belonging to IIA, IIB and 'IIC' classification gases, but :
Use with hydrogen is forbidden. (Hydrogen implies specific tightness elements).
Use with oxygen is strictly forbidden. (Oxygen is not covered by ATEX)
The connection of Corus to any other instrument or device must be done in accordance to the electrical
parameters indicated on the certificate.
It has to be noted that the directive 1999/92/EEC ("Installation / Users") has also to be respected for the
commissioning of the device.
When Corus is used in the external power supply mode, the supply has to be provided through an
intrinsically safe module with electrical compatible parameters; Use of any other type of module is
strictly forbidden. In this mode, the main battery remains plugged inside Corus in case of external mains
failure.
When using the battery mode, it is allowed to change the battery on site, by plugging the new battery
on the free connector (J7 or J8 of the I/O board) and then, removing the old battery. Its strictly
forbidden to leave Corus with two batteries permanently plugged in hazardous area.
Caution: The battery of Corus includes all protections required for intrinsic safety. Thus, it is strictly
forbidden to replace the battery unit by any battery other than the one specified by Itron (reference:
LS33600 from Saft , SL2780 from Sonnenschein or TD5930 from Tadiran)
Battery warning : The battery of Corus contains Lithium :
Fire, explosion and severe burn hazard

53
Do not recharge, short-circuit, crush, disassemble, heat above 100C (212F), incinerate or
expose contents to water
Do not solder directly to the cell
Respect storage recommendations (-30C ; +60C)
To prevent any damage, the following operations are recommended to limit electrostatic discharges on
the enclosure of Corus :

- wipe only with a wet cleaning rag

- the person carrying out the installation can discharge himself/herself by touching the potential
equalization line

54
11 MANUFACTURING

Corus is manufactured by :

Itron GMBH
Hardeckstrasse 2
76185 Karlsruhe
Germany

The site has ISO9001, ISO14401 certifications as well as certifications to produce Corus with ATEX and MID
markings.

55
12 APPENDIX 1: INITIAL ATEX CERTIFICATE

56
57
58
13 APPENDIX 2: EC DECLARATION OF CONFORMITY

59
60
61
62

S-ar putea să vă placă și