Documente Academic
Documente Profesional
Documente Cultură
techniques. It is essential to develop the required not been absorbed by the mixture flowing through the
methodology, operating procedures, and analytical techniques Venturi section.
to provide the industry with the full scope of services normally Pressure and temperature measurements are also
performed with a test separator. Sampling counts among these accurately measured for estimating the PVT properties at
essential services. line conditions.
Most experts in this domain are still debating how this The low-energy count rate is related to the composition of
should be done. Since a static-phase equilibrium that is close the fluidthereby responding to the water/liquid ratio
to what is expected in a test separator is never achieved inside (WLR)whereas the high-energy count rate is primarily
a multiphase flowmeter, samples from the gas stream or the related to the mixture density. As the effects of WLR and
oil/water stream are not representative from a compositional mixture density on the two attenuations are coupled, the
and phase behavior point of view. flowmeter solves the set of equations to compute the holdups
However these nonrepresentative samples may be of oil, gas, and water.
properly segregated in a static laboratory environment to
obtain representative sets of samples for recombination. This This is expressed mathematically as
methodology allows qualification of samples collected during
multiphase metering operations for PVT analyses. It thereby AoH AwH AgH o AmH
validates the full potential of these new well-testing L
techniques. Ao AwL AgL w = AmL , (1)
1 1 1 g 1
Phase-Fraction Computations
H ,L
Phase fractions ( o , w, g for oil, water, and gas respectively) N vac
where AoH,w, L, g ,m = ln are the attenuations at high and
can be computed using the new flowmeter technology. Fig. 1 N oH,w, L, g ,m
shows a cross section of the Venturi section and the dual- low energies of oil, water, gas, and mixtures, respectively, and
energy spectral gamma ray detector. H ,L
N vac ,o , w, g , m are the detected photons in vacuum, oil, water,
gas, and the mixture passing through the Venturi throat.
,o , w, g ) and solving
H ,L
By measuring nuclear attenuations ( N vac
Venturi
the above equations, phase fractions o, w, and g are
obtained.
When there is no slip between oil and water, the WLR is
w
Detector WLR = . ... (2)
o +w
Source
Graphical interpretation of the above equations is
presented on a compositional triangle (Fig. 2), where the
P apexes are the single-phase attenuations. An operating point
example is shown at 50% gas volume fraction (GVF) and
P Composition 50% WLR.
Meter
T
- Low energy attenuation
Flow -5 -4 -3 -2 -1 0
0 Vacuum
00
-0.1
Solution Triangle gas
-0.2
- High energy attenuation
Operating Point
Operating point -0.3
Fig 1: Cross Section of the Multiphase Flowmeter GVF = 50%
WLR = 50 % -0.4
-0.5
The Venturi section measures the total mass flow rate Constant GVF line (50%) -0.6
using differential pressure sensors, and the dual-energy Constant WLR line (50%) -0.7
spectral gamma ray detector is combined with a radioactive -0.8
oil
chemical source measures the oil, water, and gas fractions. -0.9
-1
The radioactive chemical source emits gamma rays at water
-1.1
numerous energy levels, and the attenuation of the gamma -1.2
rays by the fluid is measured at two different energy levels. A
scintillation detector combined with a photomultiplier located Fig. 2: Compositional Triangle
on the other side of the pipe detects the gamma rays that have
SPE 77405 PVT SAMPLING WITH MULTIPHASE FLOWMETERS THEORETICAL JUSTIFICATIONS AND FIELD LIMITATIONS 3
The corners of the solution triangle correspond to single compositional model of the fluid under consideration. This
phases. The constant GVF lines are parallel to the water/oil EOS model is used to account for phase-property variations
line and the constant WLR lines intersect the gas point. Fluid resulting from line flow condition fluctuations, to identify
properties define the position of the apexes of the gross changes in produced fluid composition or production
solution triangle. conditions, and to provide reliable estimates of standard
condition volumetric flow rates. To achieve these objectives,
Fluid Properties Required For Tool Calibration knowledge of the fluid composition from a previous PVT
To compute accurate volumetric flow rates of phases flowing analysis is necessary. The key to maintaining these predictions
through the flowmeter at a given pressure and temperature, during the anticipated variations in flow conditions during
phase densities, nuclear attenuations, and elemental longer-term production is building a reservoir fluid model
composition of water must be known or else measured or based on knowledge of the molecular composition and some
computed as accurately as possible. fluid properties.
The availability of fluid properties can be categorized However, if pressure and temperature variations result in
as follows: significant variations in line phase densities and compositions,
Well-test applications with no or very limited then one must continuously adjust the model parameters to
knowledge of reservoir fluid properties account for these variations. Models utilized to convert line
Well-test applications with limited knowledge of flow rates to standard condition flow rates must therefore be
reservoir fluid properties flexible enough to adapt these gross variations and recompute
Periodic monitoring applications with reservoir fluid the flow rates as necessary.
properties known from previous well tests.
Nuclear attenuations. Nuclear attenuations of oil, gas, and
Depending on the scenario, available fluid properties are water are either computed from the elemental analysis of each
either drawn directly from the PVT reports or converted as phase or measured in situ as the meter is filled with each
necessary. Often fluid parameters are not available before successive phase.
operations at the wellsite begin. If the elemental composition method is selected, the mass
The fluid properties required for flowmeter tool calibration attenuation of the mixture or compound is determined as the
and for quantification of the individual phase fractions are mass fraction-weighted average of the mass attenuations of its
phase densities and nuclear attenuations (both for high- and components or elements using the following relationship that
low-energy levels) of oil, water, and gas at line conditions applies to oil, water, gas, and solids:
H ,L H ,L
(flowing pressure and temperature).
o,w. g
= wi
i
(3)
Phase densities. Phase densities are computed from the i
densities of each individual phase at standard conditions using
atmospheric fluid samples and empirical correlations. On the other hand, if the in-situ measurement is selected,
Accurate handheld densitometers are used for that purpose. the mass attenuation is computed from
Gas density is measured using a densitometer on the gas H ,L
AoH,w, L, g
1 N H ,L
sample collected from the tool along with N2, H2S, and CO2 = ln Hvac, L = , . (4)
content using detector tubes for Z-factor corrections. o,w, g o,w, g d N o,w, g o,w, g d
Computation of densities of each phase at line conditions where d is the throat diameter.
is performed using published empirical correlations similar to
The attenuations of oil and water are determined using the
the ones used in traditional well-testing operations. Oil
meter itself. A small sample of oil or water is located in front
viscosity at standard conditions is measured using a handheld
of the measuring section and the acquisition software
viscometer. Ion analyses on water samples are performed
automatically performs the computations of the high- and low-
using a portable spectrophotometer. Extended analyses of
energy attenuations.
water samples may be required because certain ions have high
preponderance on the overall mass attenuation computations.
Flow Rate Computation Using a Simplified Model
If these ions exist in water and are not measured, a significant
To understand the factors affecting the sensitivities of the
discrepancy may be observed. Adhering to American
parameters of the multiphase well testing technology, a
Petroleum Institute Recommended Practice 45 will provide
simplified model is presented below:
coherent results.
The total mass flow rate (Q) is computed from the mixture
To convert volumes at line conditions to standard
density (m) and the pressure differential (p) between the
conditions and vice versa, a set of PVT relationships based on
inlet and the Venturi throat (minus head corrections).
empirical correlations or EOS, depending on the fluid type,
are again utilized.
On the other hand, for volatile oils and gas condensates, C (Re ) C ' 2
Q= d 2p m , ...(5)
phase densities at line conditions are computed using an EOS 1 4 4
4 S. JAYAWARDANE, B.C. THEUVENY SPE 77405
1
GVF
with liquid = WLR w + (1 WLR ) o .(9)
Theoretical
1
et al.15-16
Test Separator
(1.00) (0.50) - 0.50 1.00 1.50 2.00 2.50
qg = .. (11) F
g
and
Fig. 3: Hoffman-Crump-Hocott Plot on Separator Gas and Oil
Q Qg Samples Compared with Theoretical Equilibrium Ratios
qliquid = . .. (12)
liquid
Test separators are thought to be at static phase
equilibriumeven though the phases in reality are dynamic
Sampling From Traditional Test Separators
because of the assumption that over a long period of time the
Traditionally, samples are captured during well-testing
mass equilibration is reached between phases. That is, what
operations on the flowline where single-phase flowmeters for
enters into a separator is equal to what is removed and hence
oil, gas, and water are installed, generally the outlet of
within the separator the virtual equilibrium is achieved and
test separators.
phases remain at equilibrium.
SPE 77405 PVT SAMPLING WITH MULTIPHASE FLOWMETERS THEORETICAL JUSTIFICATIONS AND FIELD LIMITATIONS 5
Experience has shown that compositional equilibrium in a An oil sample (with some water and excess gas) is
dynamic test separator is never really achieved, as shown by collected at the liquid collector (Fig. 6) using the vertical
the Hoffmann-Crump-Hocott plot19-24 in Fig. 3. liquid collector.
As shown in the figure gas sample collected at the test
separator has lost heavy components (compared to theoretical
K-values predicted by EOS compositional model), which
clearly indicates that often it is difficult to reach a static
equilibrium in a dynamic test separator. This effect can further
deteriorate by condensation of heavy components on the wall
and valves of the gas sampling cylinder.
However, the industry has learned to live with these
uncertainties, and it will face new challenges with multiphase
well testing technology, sampling, and analysis-related issues.
result in a partial flash of the sample inside the contrary can be observed: Because of the flash that takes
sampling cylinder. place, the gas phase loses heavy and the liquid phase loses
light components.
Calculating total reservoir fluid composition, therefore,
must be done with utmost care because the prevailing phase
compositions and their relative quantities (GOR) may not be
considered as truly representative of an equilibrium flash
process that usually takes place in a traditional separator. It is
noteworthy that the total mass flow rate through the meter is
representative of the total rate of mass withdrawal from the
reservoir, even though volumetric relationships are governed
by the prevailing (changing) pressure, temperature, solution
and/or dissolution of one phase in another, flow regimes, slip
between phases, etc.
Fig. 6: Oil Sampling Point Conditioning of Multiphase Gas Samples. Flowmeter gas
samples usually contain excess liquid caused by liquid carry-
Wellsite Analysis over. This may be further accelerated due to condensation
Performing mini-PVT and/or gas chromatography analysis at owing to the temperature drop.
the wellsite on flowmeter samples is not recommended Therefore, the piston-type gas cylinders must be heated to
because obtaining phase-representative samples is more the line temperature and left for few hours for temperature
difficult. Segregation of phases can be done in the laboratory stabilization within the gas phase. The cylinder should be
prior to recombination, but it is impossible at the wellsite. agitated gently to obtain equilibrium between phases and help
Special methods must be developed to achieve that goal. vaporize any liquid that is stagnant at the bottom of the
cylinder. However, after agitation, sufficient time must be
PVT Analysis Procedures for Multiphase given to recondense the liquid phase if any excess liquid has
Flowmeter Samples been vaporized unintentionally due to violent agitation.
When flowmeter samples are collected to obtain fluid For PVT analyses, a portion of equilibrium gas at line
properties, PVT laboratory staff may not realize the temperature and pressure is isobarically displaced in the
differences between these samples and separator samples and laboratory into another cylinder before recombination.
may treat them as traditional samples. Certain properties are
required for flowmeter recalculations and the samples must be Conditioning of Multiphase Liquid Samples. Like gas
handled appropriately for physical recombination and samples, oil samples need conditioning because often they
subsequent PVT analyses. At a minimum, phase densities at contain free gas and free water at sampling (line) conditions as
line conditions are required. Optionally, compositional a result of various flow regimes, fluid type, densities,
analyses are needed if EOS modeling work is foreseen. viscosities, presence of emulsion, prevailing temperature and
pressure conditions, etc. Care must be taken while handling
Determine recombination ratios. The GOR that must be these samples, and they must not be treated the same as
used for the physical recombination work is the true GOR traditional separator samples collected during a well test.
measured by the flowmeter at line conditions, since this In some cases, oil samples may contain water and gas. If
reflects the computed phase fractions. Quite often, the gas any excess gas is sampled, the sampling cylinder must be
sample collected from a multiphase flowmeter, contains brought to line temperature and pressure and the excess gas
entrained liquid and similarly the liquid sample contains free and water removed. This can be achieved by transferring the
gas. Therefore, the composition of the gas phase is generally all the contents of the sampling cylinder into a visual
richer and that of the liquid is poorer than equilibrium PVT cell.
compositions if excess phase is not removed. However, in
certain areas where a localized pressure drop may exist, the Compositional Analysis of Gas. Compositional analyses by
gas chromatography can be carried out on a conditioned
SPE 77405 PVT SAMPLING WITH MULTIPHASE FLOWMETERS THEORETICAL JUSTIFICATIONS AND FIELD LIMITATIONS 7
sample after flashing the gas through a cold trap. The liner Separator Versus Flowmeter Samples: Case 1
between the gas cylinder and the chromatographs injector A field case study was conducted using a traditional separator
port must be heated to a temperature slightly above the and the flowmeter to validate the hypothesis of conditioning
sampling (line) temperature to avoid liquid condensation in samples and to compare flowmeter and separator samples. The
the liner.
compositional analyses were very close (within 2%). Other
PVT properties such as oil formation volume factor and oil
Compositional Analysis of Oil. Oil compositional analysis is density at line conditions were comparable with 2% to 3% and
conducted by flashing the oil sample to atmospheric pressure
were in line with the EOS predictions for given
as previously described. Water that may be present in oil can phase compositions.
be separated either by centrifugation or by using Karl-Fischer
titration and correcting for water volumes. The composition of the excess free gas (gas cap)
If an emulsion exists, centrifugation or adding a de- segregated from the oil (Table 1) was also very close to that
emulsifier may help. However, adding a foreign compound of the line gas sample collected, indicating that our phase
may alter the fluid composition. Therefore, care should be conditioning procedure was correct.
taken to select a chemical that does not affect the hydrocarbon
composition. If, however, a peak for the de-emulsifier is Table 1: Comparison of Gas Compositions
identified, its area must be disregarded in computations. Separator Flowmeter Gas Cap
Gas Conditioned from
Measurement of Density at Line Conditions Gas Flowmeter
Liquid
When true densities of each phase (oil, gas, and water) at line
conditions are required, PVT analysis can be performed on the Mole % Mole % Mole %
recombined sample as outlined below. Flow rates can be
recalculated using true phase densities and rerunning the N2+CO2 0.38 0.46 0.44
computation program.
C1C4 99.44 99.41 99.35
Pycnometer Method. Gas or oil density can be measured at
C5C7 0.15 0.12 0.15
line conditions with a reasonable accuracy by transferring a
known volume of gas or live oil at line conditions to a high- C8C10+ 0.03 0.01 0.06
pressure pycnometer using an accurate volumetric pump. The
pycnometer is weighed before and after the transfer, and
density is calculated. If water is present in the oil, the density
measured will indicate this, and flashing the sample will Just as above, the composition of the excess free liquid (oil
provide water volume. leg) segregated from the gas was in agreement with the
composition of the flowmeter liquid sample collected.
Using a Visual Cell. Oil volume at line conditions is Separator liquid and flowmeter-conditioned liquid were
measured using the cathetometer mounted to the visual cell. comparable, as shown in Table 2 below.
Then a portion of oil is flashed to atmospheric conditions
through a cool trap, and mass of all fractions recovered is
measured, as well as the flash gas composition. Flash gas is Table 2: Comparisons of Liquid Compositions
analyzed with gas chromatography, and the flash liquid mass Separator Liquid Flowmeter-
and the density are measured. Oil density at line conditions is Conditioned
computed using this flash technique combined with Liquid
mass balance.
Mole % Mole %
Using a Blind Cell. When a pycnometer or a high-pressure N2+CO2 0.05 0.05
visual cell is not available, a blind cell and positive
displacement pump can be used to accomplish the task. C1C4 17.11 17.80
A known volume of oil is transferred into a PVT cell using
a positive displacement pump and then pressurized to line C5C7 12.98 13.47
pressure and heated to line temperature. The new oil volume
C8C10 17.89 17.73
resulting from expansion and contraction during each of the
above processes is recorded. C11C12+ 51.97 50.95
The oil sample is then flashed from the blind cell as
described earlier and oil density is computed.
8 S. JAYAWARDANE, B.C. THEUVENY SPE 77405
Physical recombination using appropriate GORs yield The objective was to evaluate the flowmeter sampling
fluids that are comparable, as shown in Table 3. procedures by comparing PVT analyses performed on both
types of samples, without conditioning the flowmeter samples
Table 3: Comparison of Recombined Fluid as was done in Case1.
Compositions
Components Recombined Recombined Deviation Comparison of Results
Separator Flowmeter- in Both sets of pressurized oil and gas samples were analyzed in
Conditioned the laboratory without conditioning the samples. Obtained
Fluids Fluids Reservoir results are shown in Table 4 and 5.
GOR=640
GOR=589 scf/STB Fluids
scf/STB Table. 4: Fluid Composition from Separator Samples
Components Separator Samples
Mole% Mole% Mole% Separator Separator Reservoir
Liquid Gas Fluid A
N2+CO2 0.23 0.27 0.04 Mole % Mole % Mole %
N2 0.04 0.24 0.14
C1C4 61.16 63.01 1.85
CO2 0.21 0.49 0.35
C5C7 6.11 6.08 -0.03 C1 19.98 79.54 50.67
C2 11.15 12.68 11.94
C8C10 8.33 7.92 -0.41
C3 10.14 5.10 7.54
C11C12+ 24.17 22.72 -1.45 iC4 1.46 0.41 0.92
nC4 5.56 1.05 3.24
iC5 1.86 0.16 0.98
The recombined fluid compositions differ because of the nC5 3.17 0.22 1.65
GORs used. As shown earlier, phase compositions were C6 3.89 0.05 1.91
much closer. C7+ 42.54 0.06 20.66
Oil and Gas Gravity Measurements at the Wellsite. During Table. 5: Fluid Composition from Flowmeter
this field test, number of atmospheric oil and gas samples Samples Without Conditioning
were taken at the flowmeter and at the separator for gravity Components Flowmeter Samples Deviation
measurement on various wells and at chokes sizes.
Liquid Gas Reservoir Between
At the flowmeter, samples were collected from the gas and Fluid B Fluids
liquid sampling points; at the separator they were collected A and B
from the top of the gas outlet and from the oil sight glass. Mole % Mole % Mole % Mole%
The maximum oil gravity deviation between the flowmeter N2 0.15 0.57 0.37 0.23
and the separator samples was 0.5% as measured with a CO2 0.22 0.40 0.31 -0.04
hydrometer. The maximum gas gravity deviation between the C1 20.64 82.21 53.01 2.34
samples was 1%, measured with a orifice meter. This leads to C2 11.30 11.53 11.42 -0.52
the conclusion that oil and gas gravity measurements are C3 10.86 3.91 7.21 -0.33
comparable in the flowmeter and separator samples, but that
iC4 1.58 0.29 0.90 -0.02
instruments limit accuracy.
nC4 6.02 0.73 3.24 0.00
PVT Laboratory Measurements. During this field test, iC5 2.76 0.13 1.38 0.40
pressurized oil and gas cylinders were taken at the flowmeter nC5 3.81 0.14 1.88 0.23
and at the separator on two choke sizes. C6 5.82 0.04 2.78 0.87
At the flowmeter, several 10-L gas-sampling cylinders C7+ 36.84 0.05 17.50 -3.16
were collected from the gas sampling point and 600-cm3 oil-
sampling cylinders were taken from the liquid sampling point. Total 100.00 100.00 100.00
At the separator, 10-L gas sampling cylinders were collected Molar ratio 0.4743 0.5257 1.0000
from the top of the gas outlet and 600-cm3 oil-sampling
cylinders were collected from the oil sight glass.
SPE 77405 PVT SAMPLING WITH MULTIPHASE FLOWMETERS THEORETICAL JUSTIFICATIONS AND FIELD LIMITATIONS 9
As it can be seen from the fluid compositions for flowmeter Table 8: Sensitivity to Oil Line Temperature
and separator, they do not agree as well as in Case 1, where Black Oil
C7+ and methane content discrepancies are lower. PVT Rec Rec Fluid 7 Rec Fluid
Property Fluid 1 (Liq. 8
No temp.+10F) (Liq.
Sensitivity Analysis temp. temp.-
drop 10F)
A sensitivity analysis was conducted on a typical black oil and o
a gas condensate to evaluate the impact on PVT properties, if From pb at 200 F 1426 1449 1523
Differential (psia)
the liquid sample had a slightly different pressure or Liberation Bob 1.439 1.475 1.489
temperature than that of the gas sample. This objective was set Rsb(scf/STB) 632.6 677.0 702.7
due to the sampling point configurations on the flowmeter. From LC
o (lbm/ft3) 48.957 48.840 49.070
These sensitivity analyses were performed with a Separation LC
g (lbm/ft3) 3.06 3.279 3.319
Test
compositional model using 3-parameter Peng-Robinson
EOS.26
Tables 6 and 7 below show few PVT properties (pb - Table 9: Sensitivity to Oil Line Temperature
bubblepoint pressure, Bob formation volume factor at Gas Condensate
PVT Rec Rec Fluid 9 Rec
bubblepoint, Rsb Solution gas/oil ratio at bubblepoint) of Property Fluid 2 (Liq. Fluid10(Liq.
recombined fluids obtained by mixing gas sample with oil No temp.+10F) temp.-
sample having no pressure drop and also samples having a temp. 10F)
drop
pressure drop of 15- or 30-psia. o
From pd at 180 F 3140 2919 2815
Constant (psia)
Volume
Table 6: Sensitivity to Oil Line PressureBlack Oil
Depletion
PVT Rec Rec Fluid 3 Rec Fluid 4 LC
Property Fluid 1 (Liq. (Liq.
From o (lbm/ft3) 38.936 39.149 38.695
Separation LC
g (lbm/ft3) 3.998 3.912 34.091
No pressure - pressure -
Test
pressure 15 psia) 30 psia)
drop
From pb at 200 F
o
1426 1468 1452 As we can observe from the above tables, gas condensates
Differential (psia) are more sensitive to both the pressure change and the
Liberation Bob 1.439 1.479 1.477 temperature change in the liquid sampling line, than the
Rsb 632.6 684.4 679.1
(scf/STB) black oils.
LC
From Sep. o (lbm/ft3) 48.957 48.999 49.042
Test LC
g (lbm/ft3) 3.06 3.250 3.193 Conclusions
New challenges have been encountered in multiphase well
testing, sampling and PVT analysis. Contrary to general
opinion, there are no insurmountable technical hurdles to
Table 7: Sensitivity to Oil Line PressureGas
collecting a set of gas and oil samples from the multiphase
condensate
PVT Property Rec Rec Fluid 5 Rec Fluid 6
flowmeter and developing a rigorous treatment for the fluid
Fluid 2 (Liq. (Liq. property setup. Paying close attention to fluid property
No pressure - pressure - measurements and modeling issues is important to
pressure 15 psia) 30 psia) obtain reliable phase fraction computations using multiphase
drop
o flowmeter.
From pd at 180 F 3140 2870 2872
Constant (psia) The key to success in the collection of multiphase samples
Volume is the ability of the flowmeter technology to split completely
Depletion between the meter setup parameter (empty pipe count and
LC
From o (lbm/ft3) 38.936 39.019 39.103 static calibration of the pressure, differential pressure, and
Separation LC
g (lbm/ft3) 3.998 3.928 3.858
Test temperature transducers) and the fluid properties (basic PVT
properties and nuclear attenuations of oil, gas, and water).
Particular sampling procedures are required to deal with
The same analysis was performed on these samples to specific issues related to safety, environmental protection, and
evaluate the temperature effect. Tables 8 and 9 show PVT sample quality. Gas and oil samples collected from multiphase
properties of these recombined fluids obtained by mixing gas flowmeter contain some fraction of the other phase and water.
sample with oil sample having no temperature change and also Therefore samples must be conditioned, i.e. excess phases
samples having a temperature variation of +10 oF or 10 oF. isobarically removed after reaching the phase equilibrium in
the laboratory using correct PVT procedures. These
segregated phases can be physically recombined using
10 S. JAYAWARDANE, B.C. THEUVENY SPE 77405