Sunteți pe pagina 1din 8

Ophthalmic Instruments from Carl Zeiss LNDY202

VISULAS YAG II plus


Intelligence and precision
in a standard-setting system
LNDY202

Peak performance for


enhanced photodisruption

The VISULAS-YAG-II-plus laser and optimized slit The compact corneal microscope for enhanced
lamp combine to create a safe and versatile system. ergonomic operation.
Superb performance characteristics include extremely Easy retrofitting with either the beam splitter or
high light efficiency and resolution. The system offers compact camera for individual needs.
a range of advanced features including: Choice of anterior or posterior offsets that provide
an ideal environment for visualization and treatment.
A lowered illuminating prism that allows treatment An articulated fixation lamp holder ensures flex-
at any position of the lamp housing and corneal micro- ibility and a wide operating area.
scope.
Precisely matched energy stepping to provide optimum
energy for every application.
LNDY202

A range of functions
for a variety of benefits

It simply can do more Certified safety

The VISULAS-YAG-II plus Laser provides a variety of Carl Zeiss complies with ISO 9001 guidelines and meets
benefits in a single, compact system: the stringent requirements of U.S. Good Manu-
facturing Practices. Plus, every system is backed by a
Low energy for enhanced optical breakdown 150-year history of Zeiss commitment to quality,
Comprehensive four-point aiming beam safety and performance.
Convenient modular design
LNDY202

Designed for
precision and confidence

The VISULAS YAG-II-plus Laser incorporates intel- The system was developed for use in both an office
ligent aiming beam patterns. Astigmatic distortions (2) practice environment and outpatient clinics, giving
are clearly visible in the otherwise symmetrical aiming physicians a choice of venues for enhanced utilization.
beam pattern (1). As a result, physicians have the Plus, the ergonomic, noiseless and open design
opportunity to adjust the energy before firing the keeps physicians in constant visual contact with their
laser. patients. Full-filter laser safety glasses or goggles
provide excellent protection and meet the guidelines
of DIN-58215.

1 2
LNDY202

Accurately matched -
optics and energy

Variable offset ensures


a consistently sharp image
for any application.

Positive focus shift - e.g. posterior capsulotomy

Negative focus shift - for anterior applications

Consistent development of
technology results in gentle
Mode structures treatment at lowest energy levels.

Multimode laser Fundamental mode laser Super Gaussian laser VISULAS YAG II plus

Optical breakdown at: approx. 8 mJ approx. 4 mJ approx. 2.5 mJ


LNDY202

Technical Data

Laser wavelength 1064 nm

Mode Super Gaussian

Optical breakdown typ. 2.5 mJ in air

Pulse duration < 4 ns (typ. 2 - 3 ns)

Max. laser energy single pulse, typically 10 mJ


double pulse, typically 23 mJ
triple pulse, typically 37 mJ

Energy attenuation 10 steps (3, 5, 8, 12, 16, 20, 30, 50, 70


and 100% transmission)

Pulse repetition frequency max. 2 Hz

Focus diameter 10 m in air (1/e2)

Angle of exit aperture 16

Aiming beam Laser diode 670 nm, power: 5 W - 150 W


4-spot focusing system

Offset variable 150 m


________________________________________________________________________
Electrical connection 115/230 V, 50 - 60 Hz, 140 W

Illumination 12V; 30W halogen lamp, adjustable

Magnification 5, 8, 12, 20, 32X through Galilean changer


with 10X eyepieces and tube f = 140 mm

Tube Straight tube f = 140 mm with 50 - 78 mm PD


adjustment, convergent tube available as option

Eyepieces 10X high-eyepoint eyepieces with 8 D


compensation of ametropia;
12.5X available as option

Slit adjustment Width: 0 - 14 mm, continuous


Length: in steps 1/3/5/9/14 mm

Dimensions (H x W x D) 623 x 400 x 350 mm, slit lamp with laser head
(24 x 16 x 14 inches)
128 x 165 x 315 mm, control console
(5 x 7 x 13 inches)

Weight 11 kg slit lamp with laser head (24 lbs.)


4.5 kg control console (10 lbs.)

Electrical supply 115/230 V, 50 - 60 Hz, 80 VA (slit lamp)


115/230 V, 50 - 60 Hz, 140 W (control console)

VISIBLE OR INVISIBLE LASER RADIATION


AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
CLASS 4 LASER PRODUCT AS PER EN 60825-1:94
Laser 1 1064 nm 50 mJ 4 ns
Laser 2 660-680 nm 150 W cw
LNDY202

Accessories
help you make the
most of your investment

Optional video equipment can be readily inte-


grated into the system for co-observation and
documentation.

The highly compact and mobile VISULAS-YAG-II


plus Laser can be quickly and easily transported
between practice locations.

Diverse high quality contact lenses are available.


Publication No: YPB Rev. A 10M 1098
LNDY202
Specifications subject to change.
Printed in U.S.A.

Humphrey Systems Toll-free: 1-877-HUMPHREY


5160 Hacienda Dr. (877-486-7473)
Dublin, CA 94568 U.S.A. Phone: 925-557-4651
E-mail: info@humphrey.com Fax: 925-557-4778
Web: www.humphrey.com

CMYK PMS Reflex Blue