Sunteți pe pagina 1din 36

Maintenance Guide

FA-200/300/400
Fully Automatic Biochemistry Analyzer
CLINDIAG SYSTEMS B.V.B.A.
Contents
1 Basic Principle................................................................................................1
2 Operational Environment................................................................................1
3 Warnings and Precautions..............................................................................1
4 Qualifications and Tools..................................................................................1
4.1 MAINTENANCE STAFF QUALIFICATIONS........................................................1
4.2 KNOWLEDGE REQUIRED.............................................................................2
4.3 TOOLS.......................................................................................................2
5 Maintenance...................................................................................................2
5.1 DAILY MAINTENANCE..................................................................................2
5.1.1 Maintenance with Power ON.............................................................2
5.1.2 Maintenance with Power OFF...........................................................3
5.2 WEEKLY MAINTENANCE..............................................................................3
5.3 MONTHLY MAINTENANCE............................................................................3
5.4 YEARLY MAINTENANCE...............................................................................3
6 Basic Structure and Main Components..........................................................4
6.1 MAIN CONTROL PANEL...............................................................................4
6.1.1 Functions...........................................................................................4
6.1.2 Basic Structure..................................................................................4
6.2 MOTOR DRIVE MODULE..............................................................................4
6.2.1 Functions...........................................................................................4
6.2.2 Basic Structure..................................................................................5
6.3 POWER SYSTEM.........................................................................................7
6.4 ANALYTIC SYSTEM......................................................................................8
6.4.1 Optical System..................................................................................8
6.4.2 Heating System.................................................................................8
6.4.3 Test System.......................................................................................9
6.4.4 Replacements....................................................................................9
6.5 SAMPLING SYSTEM..................................................................................13
6.5.1 Functions.........................................................................................13
6.5.2 Installation of Reagent Plate and Sample Plate..............................13
6.6 TUBING SYSTEM.......................................................................................16
6.6.1 Functions.........................................................................................16
6.6.2 Structure of Tubing System.............................................................16
7 Replacement of Easily Damaged Parts........................................................17
7.1 REPLACEMENT OF LAMP...........................................................................17
7.2 DILUTER PISTON REPLACEMENT...............................................................18
7.3 PROBE REPLACEMENT.............................................................................19
7.4 CUVETTE REPLACEMENT..........................................................................20
7.5 FUSE REPLACEMENT................................................................................20
7.6 ADJUSTMENT OF PROBING SENSITIVITY....................................................21
7.7 ADJUSTMENT OF GAIN AND OFFSET......................................................21
7.7.1 Adjustment of OFFSET Value.........................................................21
7.7.2 Adjustment of GAIN Value...............................................................23
8 Troubleshooting............................................................................................23
9 Appendix: Wiring Diagram of FA Series.......................................................29
1 Basic Principle
Biochemistry analyzer bases its testing principle on Lambert Leer Law. By
mixing up reagents and patient serum samples in a fixed proportion and then
incubating the mixture in a cuvette at a certain temperature, the instrument
continuously measures light absorption of a certain wavelength and
automatically calculates concentration of a tested matter according to different
light absorption results.

2 Operational Environment
Normal operational environment requires:
-- Ambient Temperature: 10-30;
-- Relative Humidity: 30-70;
-- Air Pressure: 86.0hPa-106.0hPa;
-- Power Supply: 220V22V, 50Hz1 Hz
-- Avoid intense electromagnetic interference, mechanical vibrations,
corrosive gases and direct exposure to bright sunlight.
-- Reliable ground connection.

3 Warnings and Precautions


Note the Warnings & Precautions when performing installing and servicing:
-- Always make sure to cut off power supply before servicing to avoid getting
an electric shock by high voltage of 220V inside the instrument.
-- Do not drop any fluid onto circuit board or power supply when servicing
tubing system.
-- Make sure to cut off power supply before servicing circuits.
-- Avoid biology contamination when maintaining the instrument.
-- Make sure the instrument is properly grounded.

4 Qualifications and Tools

4.1 Maintenance Staff Qualifications


Service and maintenance shall be done by Clindiag after-sale engineers or
authorized professionals.

4.2 Knowledge Required


-- Basic knowledge on analog circuitry
-- Basic knowledge on digital circuit
-- Good command of performance and test methods to basic component

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 1 of 32


-- Basic knowledge on computer
-- Knowledge on service safety

4.3 Tools
General Tools: 1 medium-sized Philips screwdriver, 1 small Philips
screwdriver, 1 medium-sized slot-head screwdriver, 1 small slot-head
screwdriver, 1 pair of needle nose pliers, 15019 wrench, 1 set inner hexagon
spanners, 1 150W electric iron, 1 pack of soldering tin and 1 digital multimeter
Specific Tools: 1 set 5.5 mm and 7 mm non-adjustable wrenches, 1 set 8 mm
and 10 mm non-adjustable wrenches, 1 digital thermometer with an accuracy
of 0.1C

5 Maintenance

5.1 Daily Maintenance


Daily maintenance incorporates maintenances with instrument power on and
off, and should be done every day.

5.1.1 Maintenance with Power ON


-- Turn on power switch and the computer.
-- Check distilled water, waste fluid and pipes for any signs of abnormity.
-- Open operation interface.
-- Run the device.
-- Click Device maintenance icon to flush cuvettes.
-- Click Check cuvette blank icon.
-- Fill water before beginning cuvette blank check (check cuvette blank for
three times and save blank values).
-- Pump out water or perform preset maintenance procedure.

5.1.2 Maintenance with Power OFF


-- Click Check cuvette blank icon.
-- Fill water.
-- Click on Device maintenance icon.
-- Wash needle for 4 times.
-- Turn off the instrument or perform preset maintenance procedure.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 2 of 32


5.2 Weekly Maintenance
-- Perform maintenance once a week.
-- Clean all reagent bottles and water buckets.
-- Free instrument surface from dust and dirt.

5.3 Monthly Maintenance


-- Check and soak clean cuvettes with cleaning solution. Replace cuvettes if
necessary.
-- Apply lubricants (Vaseline and lubricating oil) to metal rods of sampling
arm and cleaning arm.
Note
Do not apply too many lubricants to avoid abnormal operation of mechanical
components.

5.4 Yearly Maintenance


Yearly maintenance shall be done by Clindiag professional engineers or
trained engineers.
-- Remove dust from components inside the instrument, such as fans and
sensors.
-- Check belt operation for each component. Adjust belt tightness if
necessary.
-- Check tubing system. If there are any aged pipes, replace them.
-- Replace syringe piston.
-- Adjust AD voltage value and OFFSET value on main control panel.
-- Backup database files and empty tables starting by S in database.

6 Basic Structure and Main Components

6.1 Main Control Panel

6.1.1 Functions
Main control panel, core component in the instrument, is to executive orders
when connected to computers. By controlling operation of each component, it
collects and amplifies signals for AD conversion, which then will be sent to
computer for further processing.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 3 of 32


6.1.2 Basic Structure
Main control panel is mainly composed of digital signal processor, complex
programmable logic device, memory chip, AD conversion chip, AD
amplification chip, RS232 interface, parallel interface of reaction plate and
RS485 parallel bus. The basic structure of main control panel is shown as
Figure 6-1.

Figure 6-1 Basic structure of main control panel

6.2 Motor Drive Module

6.2.1 Functions
Motor drive module receives orders from main control panel to collect sensor
signals, drive motor and control solenoids.

6.2.2 Basic Structure


Motor drive module consists of a motor bracket, bus plate, motor board, high
power motor board, the motor board on reaction plate, 5V power board and a
liquid level detecting board. The basic structure of motor drive module is
shown as Figure 6-2.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 4 of 32


Figure 6-2 Basic structure of motor drive module
Motor bracket is a mechanical component used for holding motor boards. It
is designed with fans to radiate heats on motor boards. As shown in Figure 6-
2.
Bus plate is designed with slots and connectors for connecting motor boards,
motors and optical couplers (As shown in Figure 6-3).

Figure 6-3 Bus plate


Motor board drives motors of sampling arm and sample rotor and controls
solenoid on/off. Its address code can be set as required (1-16) (as shown in
Figure 6-4).
There are four rows of contact pins below single chip of motor board, which
are called 8421 codes. The codes are arranged in a sequence of 1, 2, 4, and
8 from outside to inside and can be connected with shorting stubs. The
address code shown in Figure 6-5 is 6.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 5 of 32


Figure 6-5 Address code

High power motor board is to control operation of diluter and solenoids


on/off. It has 4 fixed address codes: 16, 17, 18 and 41 (as shown in Figure 6-
5).

Figure 6-5 High power motor board


Reaction plate motor board is to control operation of reaction plate. It is
under direct control of main control panel by means of parallel serial ports (as
shown in Figure 6-6).

Figure 6-6 Reaction plate motor board


5V power board is a module to power optical coupler and motor boards (as

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 6 of 32


shown in Figure 6-7).
Liquid level detecting board controls level signals to adjust liquid level
detecting sensitivity.

Liquid Level Detecting Board 5V Power Board


Figure 6-7 Liquid level detecting board and power board

6.3 Power System


Power system consists of a main power switch, a wave filter, an isolating
transformer, a switching power supply and electric wires. The main function of
the system is to supply required voltages to instrument and to control
instrument on/off. Circuit diagram of power system is shown as Figure 6-8.

Figure 6-8 Circuit diagram of power system

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 7 of 32


6.4 Analytic System
Analytic system is essential to the instrument.

6.4.1 Optical System


Optical system is mainly composed of a halogen lamp, an optical fiber and a
filter.
Structure chart of optical system is shown as Figure 6-9.

Figure 6-9 Optical system

6.4.2 Heating System


Heating system consists of a heating ring, a temperature meter, a power
control board and a temperature sensor. Structure chart of heating system is
shown as Figure 6-10.

Figure 6-10 Heating system

6.4.3 Test System


Test system contains a test board, signal cables and a main control panel.
Circuit diagram of test board is shown as Figure 6-11.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 8 of 32


Figure 6-11 Circuit diagram of test board

6.4.4 Replacements

6.4.4.1 Code Wheel Adjustment


Tighten jackscrews on code wheel of reaction plate after adjustment.
Press code wheel, shaft plate and shaft cap into place to avoid up and down
movement of shaft plate (as shown in Figure 12).

Code Wheel

Axis

Jackscrews

Figure 6-12 Code wheel

Tighten screws and nuts after adjusting heating ring to prevent it from
rubbing with spinning reaction plate.

Retaining Screw

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 9 of 32


Cleaning Head

Tray

Optical Fiber

Heating Ring

Figure 6-13 Heating ring

6.4.4.2 Optical Fiber


-- Adjust optical fiber copper pillar until facula diameter is close to 2.5mm.
-- Optical fibers are sequentially addressed by natural numbers. No. 1 is for
use of 340 nm test channel, No. 2 for 405 nm test channel and others for the
rest of test channels.
-- Tighten optical fiber head with M3*3 jackscrews. Make sure the screw
heads are lower than surface of heating ring to assure smooth rotation of
reaction plate.
Compress optical fiber with trim strip. Check and confirm optical fiber wont
rub with pivoting axis of reaction plate (as shown in Figure 14).

Copper Pillar
Heating Ring
Jackscrew
Optical Fiber
Sheath Strip

Figure 6-14
Light beams are gathered at the center of cuvette after installing optical fiber.
Use a white cardboard with a width of 5mm to observe light paths. Higher
facula requires more reagents while lower facula may result in incorrect test
results.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 10 of 32


Figure 6-15

6.4.4.3 Optical Coupler Replacement


Reaction plate is designed with a long optical coupler and a short optical
coupler. The long coupler is used to reset reaction plate, while the short one is
a computing coupler for controlling AD sampling. If the short coupler is faulty,
cleaning head will collide with reaction plate when cleaning and the
absorbency could be 0.
-- Code wheel is placed in the center between the two couplers. Make sure it
doesnt rub with the couplers while spinning.
-- Make sure the holder stands in parallel with tangent of code wheel when
installing the two optical couplers.
-- After resetting reaction plate, indicators of the two optical couplers come
on at the same time. If indicators dont come on all at once, reaction results
could be shown in a jagged curve or cleaning needle may collide with reaction
plate.

Figure 6-16

Short Optical
Coupler
Long Optical
Coupler

Figure 6-17

6.4.4.3 Filter Replacement


Check filter. If filter is severely damaged after long time operation or filter is
mouldy due to moist operational environment, replace filter.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 11 of 32


-- First remove the faulty filter.
-- Install filter cover with the arrow facing outward before installing photocell
cover. Compress filter to prevent it from becoming loose. Do not touch surface
of filter with bare hands or tools. Do not clean filter surface with water or any
organic solutions.
-- Install filter cover onto heating ring and align it with inner surface of heating
ring. Failure to do so may hinder reaction plate rotation or cause damage to
cuvettes. Tighten filter with inner hexagon jackscrews. Keep screw heads
below surface of heating ring. Failure to do so may hinder reaction plate
rotation.
Installation schematic diagram is shown as Figure 6-18.

Photocell Cover
Heating Ring
Filter Jackscrew

Filter Cover

Figure 6-18 Installation of filter

6.5 Sampling System

6.5.1 Functions
Sampling system is to automatically add required reagents and samples
during testing process.
Sampling system consists of a reagent plate, a sample plate, a sampling
arm, a sampling needle, a diluter, a mixer, pipes, etc.

6.5.2 Installation of Reagent Plate and Sample Plate

6.5.2.1 Motor Replacement


-- Loosen four crews that secure motor bracket to keep synchronous belt
loose.
-- Remove broken motor from motor bracket.
-- Install the belt onto synchronous pulley and tighten the screws on motor.
-- Pull hard on synchronous belt and tighten the screws on motor bracket.
-- Adjust synchronous pulley to prevent belt from rubbing with code wheel or
synchronous pulley. Tighten jackscrews on synchronous pulley.
Schematic diagrams of motor replacement are shown as Figure 6-19.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 12 of 32


Motor Bracket Motor Synchronous Belt
Retaining Screw Retaining Screw Pulley
Figure 6-19 Replacement of motor

6.5.2.2 Belt Adjustment and Replacement


Belt will become loose or frayed after long time operation. Adjust its tightness
or replace with a new belt if necessary.
-- Loosen four screws on motor bracket to adjust belt tightness. Secure four
screws on motor bracket after adjustment.
-- When replacing belt, loosen four screws on motor bracket to take out
optical coupler under code wheel. Then remove aged or broken belt.
-- Replace with a new belt and pull hard on it to tighten belt. Install four
screws onto motor bracket and position optical coupler to its original place.
-- Check and confirm belt is installed correctly and doesnt rub with code
wheel.
Schematic diagrams of belt replacement are shown as Figure 6-20.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 13 of 32


Optical Coupler
Code Belt
Figure 6-20 Replacement of belt
Wheel

6.5.2.4 Peltier Cooler Replacement


-- Remove faulty Peltier cooler.
-- Evenly apply thermally conductive silicone onto contact surface of reagent
plate, surfaces of Peltier cooler and radiator.
-- Before installing radiator, place Peltier cooler surface against metal surface
of reagent plate and cooling pad against heating surface (as shown in Figure
6-21).
-- Apply power to instrument to test refrigeration effect. Check and confirm
temperature of cooling plate is lowered by more than 10 in 10 minutes.
As shown in Figure 6-21. The upper side is cooling surface when Peltier
cooler is placed with the black wire on your left.

Metal surface Peltier Cooler Cooling Pad


Radiator
Figure 6-21

6.6 Tubing System

6.6.1 Functions
Tubing system automatically cleans sampling needle during sampling
process and cleans cuvetes after analyzing samples.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 14 of 32


6.6.2 Structure of Tubing System
Tubing system is composed of an inlet pump, draining pump, a 3-way
solenoid, a 2-way solenoid and a cleaning sink, etc.
Properly connect and arrange pipes. Make sure the following phenomenon
doesnt exist:
-- A 90 degree bend
-- Flattened pipes
-- U-shaped draining pipes in cleaning sink
Extra pipes wound around instrument case.

Pipe connection diagram is shown as Figure 6-22.

Figure 6-22 Pipe connection

7 Replacement of Easily Damaged Parts


Regularly maintain, adjust or replace components to ensure instrument
normal operation.
Note
Maintenance, adjustment or replacement of components shall be done by
professional engineers or users trained with professional knowledge.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 15 of 32


7.1 Replacement of Lamp
Make sure that the lamp doesnt work or has worked for 2 years before
replacing. Use the following steps for replacement:
-- Switching off the instrument for 15 min.
-- Open instrument back plate and disconnect power adaptor of lamp.
-- Unscrew four screws on lampshade. Remove lampshade and the ceramic
socket.
-- Unscrew the screws that secure lamp before removing it.
-- Reassemble lamp ceramic socket and tighten lamp with screws.
-- Secure lampshade with screws.
-- Connect power adaptor of lamp.
-- Close instrument back plate.
Schematic diagrams for lamp replacement are shown as Figure 6-23.

Lampshade Lamp Ceramic socket Shell


Figure 6-23 Lamp replacement

Note
-- Check and confirm the instrument is switched off when replacing lamp.
Replacing lamp during running state may cause damage to new lamp.
-- Do not replace lamp immediately after the instrument is switched off.
Failure to do so may get you burnt by its extremely hot surface.

7.2 Diluter Piston Replacement


Use the following steps to replace diluters piston:
-- Access Device Run\Device maintenance and click Device reset.
-- Use a tool to open window of diluter.
-- Loosen nuts at the end of diluters piston (A shown in Figure 6-24).
-- Unscrew the screws retaining diluter (B shown in Figure 6-24). Remove
glass tube.
-- Release piston from glass tube. Clean new piston with an alcohol pad and
apply seal grease at 1/2 of piston end before gently putting it into glass tube.
-- Carefully adjust and tighten screws that secure diluter to make piston fall
naturally at the center of glass tube. Schematic diagram of diluters piston
installation is shown as 6-25.
-- Along the direction of tube glass, pull piston downwards 3mm.
-- Tighten screws on both sides of piston.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 16 of 32


-- Close diluter window and secure it with screws.

Figure 6-24

Correct
Figure 6-25 Wrong

Note
-- Do not scratch piston when installing. Failure to do so may decrease
accuracy of test results caused by pipe leakage.
-- Be sure to install piston at the center of glass tube. Failure to do so may
reduce service life of piston or cause damage to glass tube.
-- Apply seal grease to piston every three month.

7.3 Probe Replacement


Use the same steps to replace sampling needle and reagent needle.
-- First open the cap on the needle. Then use pliers to snip the string that
fixes needle.
-- Unscrew the screws that secure string and the forcing screws that secure
needle.
-- Remove probe, then remove the hose on it.
-- Replace with a new probe and put the hose back on. Structure chart of
hose is shown as Figure 6-26.
-- Fix needle with forcing screws. Fix wires with strip. As shown in Figure 6-
27.
-- Put the cap back on needle.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 17 of 32


Figure 6-26

Figure 6-27
Note
-- Make sure the probe is vertical.
-- Retaining screws M24 of wire 1 should not be too long to avoid
interference with sample needle.

7.4 Cuvette Replacement


If blank absorbency is higher than 0.02A, which is caused by worn or dirty
cuvette, replace cuvette.
Note

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 18 of 32


-- Do not touch front and back sides of cuvette.
-- Keep upper surfaces of cuvettes at the same horizontal level. Failure to do
so may cause water to remain in cuvettes, which may result in incorrect test
results.
-- Be sure to use cuvettes of the same lot.

7.5 Fuse Replacement


Use the following steps to replace fuse.
-- Extra fuses are provided in a package attacked with the instrument. The
instrument uses an 8A fuse, while water heating system uses a 4A fuse.
-- Switch off the instrument and unplug power cord.
-- Unplug power cord from power socket on the instrument and remove fuse
base inside the socket with a tool.
-- Remove blown fuse, put the new fuse into fuse base and reset the base
back to its original place.
-- Connect power plug.

7.6 Adjustment of Probing Sensitivity


Use the following steps to adjust probing sensitivity:
Open back plate of instrument with a tool. Follow Figure 6-28 and the
flashing indicator to locate corresponding circuit board. Each probing control
board is designed with a row of binary dial plate. A higher value corresponds
to a more sensitive detecting board. Schematic diagram is shown as Figure 6-
29.

Figure 6-28

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 19 of 32


Sensitivity: 4
1248
Figure 6-29

7.7 Adjustment of GAIN and OFFSET

7.7.1 Adjustment of OFFSET Value


Check OFFSET value when unusual reaction curve, abnormal absorbency
value or extremely high absorbance of 2.2, or poor repeatability occurs.
-- After resetting the device, access Check cuvette blank interface and
click Real time check. As shown in Figure 6-30.
-- Replace cuvettes (starting with No. 67 under 340 for FA-400, starting with
No. 24 under 340 for FA-200, starting with No. 8 under 340 for FA-300) with
black cups. As shown in Figure 6-31.
-- Observe real-time voltage and adjust potentiometer on the right of main
control panel to required value of 20-50.

Figure 6-30

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 20 of 32


Figure 6-31

7.7.2 Adjustment of GAIN Value


After resetting instrument, access Check cuvette blank interface and click
Real time check. As shown in Figure 6-40.
Click on Fill Water and wait until it finishes.
Observe real-time voltage. Adjust potentiometers on the left of main control
panel to keep voltage value around 5,400.

8 Troubleshooting
The troubleshooting is listed as Table 8-1.
Table 8-1 Troubleshooting
Problem Cause Solution
The instrument Wrong connection of Check and confirm power
doesnt work after power cord cord is correctly connected.
starting up. (Power Main program not started Turn off the instrument for 5
indicator doesnt min. then restart it.
come on.) Blown fuse Check the fuse. If fuse is
blown, replace it.
No power with AC power Check if there is power from
outlet AC power outlet.
Liquid leakage Damaged pipes Turn off power switch and
Detached joints wipe off fluids on pipes.
Diluter leakage Check for any signs of loose
Blocked 3-way valve joints, damaged pipes, leaky
Pump leaking pump or bubble case.
Leakage in bubble unit

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 21 of 32


On-line failure or the Incorrect COM port Reset COM1 port for the
instrument doesnt MAINCOM=0COM1 instrument.
work. (Power RS232 data cable isnt Check data cable
indicator comes properly connected or connections.
on.) poor contact in internal
wiring of data cable
Test channel isnt set Recheck settings of test
correctly. channel under Biochemistry
Assay setup\Device.
COM port of the computer Replace COM port of the
fails to work. computer.

Detached signal wire of Check and confirm motor


motor control control wires are properly
connected.
Main program not started Switch off the instrument for
or main control board 5 min. then restart it.
faulty
Missing hardware Backup hardware
parameters parameters or restore data
with Hardware parameter
backup. Password: 999.
Two windows are running Close a window.
at the same time.
Incorrect selection of 12 Select a right channel.
channel under Hardware
Parameter Setup
Cant suck in Blocked sampling needle Use an acupuncture pin to
samples clear blockage and perform
maintenance to sampling
needle.
Diluter doesnt work. Check diluter, motor,
address code of diluter as
well as its wiring connection.
Leaky diluter Replace diluters piston.

Sampling needle comes Adjust height of sampling


into contact with cuvette needle.
bottom.
Sampling needle cant Lower liquid level sensitivity
sense liquid level properly. of sampling needle and
check its connecting wires.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 22 of 32


Cracked pipes or Check pipes for any
detached joints detached joints or cracked
pipes.
Broken solenoids or Use 12V voltage to test
solenoids arent under solenoids.
control.
Sampling needle Disconnected wires or Check and confirm wires of
comes into contact poor contact in liquid level liquid level sensor are
with cuvette bottom sensor properly connected.
or sampling needle
Higher or lower sensitivity Adjust sensitivity of liquid
cant be lowered into
of liquid level detecting level detecting board (8-13).
cuvette
board

Incorrect descending Adjust descending height of


height of needle reagent needle to 3mm
above cuvette bottom, and
sampling needle to 3mm
above bottoms of cuvette
and serum cup.
Broken motor control Use the program to test
board motor board.

Distorted sample cup Replace with a new sample


cup.
Liquid drop remains Air leakage caused by Check pipes for any
on tip of needle after cracked or detached pipes leakage.
cleaning

Blocked reagent needle Rinse needle or use an


acupuncture pin to clear
blockage.

Broken solenoids Use the program to test


motor board.
Leaky diluter Replace diluters piston or
tube glass. Apply lubricants
on piston.

Sample needle cant touch Adjust height of sample


cuvette bottom. needle or liquid level
detecting sensibility (8-13).

Bubbles in pipes Rinse with needle for 5


times.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 23 of 32


Dripping cleaning Short needle wont stop Remove pipes for cleaning.
needle dripping after cleaning, Then apply reverse
and one-way valve pressure to close one-way
doesnt stay closed. valve.

Short needle wont stop Adjust descending depth to


dripping after cleaning, allow cleaning block of long
and needle is lowered needle to contact cup
down to cup bottom. bottom.

Long and short needles of Adjust long needles to


cleaning head wont stop the same horizontal line.
dripping, uneven long Keep Cleaning block
needles 1mm longer than the 7
long needles.

3-way solenoid is not Open solenoid to clean


tightly closed if short valve.
needle wont stop dripping
during device reset or
needle cleaning.

Unsecured cleaning block Hold cleaning block with


glue.

Improperly positioned Adjust cleaning block to a


cleaning block vertical position. Adjust
position of cleaning head or
holder to put cleaning block
at the center of cuvette.

Incorrect results Dirt cuvette Replace with a clean


cuvette.
Facula isnt at the center Adjust reaction plate to
of cuvette center facula and keep
facula 1.5 - 2mm above
cuvette bottom.
Cuvette voltage isnt Readjust GAIN and
within normal range. OFFSET values.

Cracked or detached Check pipes for any


pipes abnormity.

Leaky diluter Replace glass tube or


piston. Do not forget to
apply seal grease on piston.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 24 of 32


sampling needle Adjust position and height of
position or height sampling needle.
incorrect
Invalid reagents or Replace reagents or control
control materials materials.

Broken solenoids Use the program to test


solenoids.

Liquid level sensing Adjust sensitivity of liquid


function failure level detecting board.

Incorrect test parameters Check and reset parameters


when it is necessary.

Unstable test voltage Check ground and voltage


and absorbency stabilization. Check if the
following phenomenon
exists, test board is loose,
filter is affected with damp,
test board isnt installed
vertically with optical path,
light leakage, loose optical
path or facula beyond the
range of22.5.
Blocked suction needle Rinse needle or use an
acupuncture pin to rule out
blockage.
Incorrect temperature Recheck or readjust
in reaction plate temperature control of
reaction plate.
Lose RS232 data cable Properly connect data cable.

Fluid remains in cuvette Adjust cuvette position and


caused by improperly height of cleaning head to
positioned cuvette or make it horizontal.
incorrect height of
cleaning head
Code wheel rubs with Readjust code wheel to a
optical coupler center position. Clean
optical coupler with alcohol
or replace with a new optical
coupler if necessary.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 25 of 32


Poor contact in test Reconnect test lines.
lines

Unused detecting Connect Pin 2 and Pin 3 on


channels not jumpered main control board with
jumper block.
Loose computing Tighten computing optical
optical coupler coupler

Insufficient cleaning water Increase water inlet time to


in cuvette ensure the cuvette is half full
for complete cleaning.
Voltage is or less Lamp failure Replace lamp if necessary.
than 0 during blank
check Loose connecting wires of Check lamp connecting
lamp wires.
Abnormal voltage to lamp Check or replace power
supply if necessary.
Wrong signal connecting Check signal connecting
wires wires.
Main control panel failure Replace main control panel
if necessary.
Filter affected with damp Replace the filter.

Incorrectly connected Check connecting wires or


RS232 wires replace them if necessary.
Reagents and water Blocked suction needle Rinse needle or use an
cant be sucked in or acupuncture pin to clear
delivered blockage.
Leaky diluter Replace glass tube or
piston.
Liquid level sensing Adjust sensitivity of liquid
function failure level detecting board.

Corresponding solenoids Use the program to test


failure solenoids.

Detached pipes Check pipes for any


abnormity.

Mechanical Hardware files read-only Remove the read-only attrib


parameters cant be ute.
saved

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 26 of 32


Blue or red color Red indicates that Replace cuvettes when
during blank check detecting voltage is detecting voltage is less
beyond normal range of than 30,000. Adjust
30,00062,000. detecting voltage when it is
more than 62,000.

Blue indicates that Clean or replace


blank absorbency is cuvettes with smaller
beyond normal range gap if necessary.
and cuvettes gap is
too large.

Incorrect Incorrect wavelength Select a right wavelength.


absorbency

Disordered characters in Open the file and empty its


Cupblank contents. Do blank check
again and save the values.

9 Appendix: Wiring Diagram of FA Series


-- Wiring diagram of FA-400 Motor Drive System is shown below.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 27 of 32


MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 28 of 32
Where
-- M1: Motor Board of Reaction Plate
-- M15: Washing Arm
-- M16: Diluter I
-- M17: Diluter II
-- M18: Diluter III
-- M6: Reagent Arm I
-- M7: Reagent Arm I
-- M9: Reagent Arm II
-- M10: Reagent Arm II
-- M12: Sampling Arm
-- M13: Sampling Arm
-- M5: Sample Plate
-- M3: Reagent Plate II
-- M4: Reagent Plate I
-- M33: Mixing Arm I
-- M34: Mixing Arm I
-- M35: Mixing Arm II
-- M36: Mixing Arm II
-- M37: Electrolyte
-- M38: Electrolyte
-- M41:Diluter

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 29 of 32


-- Wiring diagram of FA-300 Motor Drive System is shown below

Where,
-- M12: Mixing Arm
-- M13: Mixing Arm
-- M16: Reagent A Diluter
-- M17: Reagent B Diluter
-- M1: Reaction Plate
-- M4: Reagent Plate
-- M5: Sample Plate
-- M15: Cleaning Rod
-- M6: Reagent Needle I
-- M7: Reagent Needle I
-- M9: Reagent Needle II
-- M10: Reagent Needle II
-- S1: Reagent Needle Liquid Level
-- S2: Sample Needle Liquid Level

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 30 of 32


-- Wiring diagram of FA-200 Motor Drive System is shown below.

Where,
-- M1: Reaction Plate
-- M4: Reagent Plate
-- M5: Sample Plate
-- M15: Cleaning rod
-- M6: Reagent Needle
-- M7: Reagent Needle
-- M16: Diluter
-- M9: Mixing Arm
-- M10: Mixing Arm

END

CLINDIAG SYSTEMS B.V.B.A.

MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 31 of 32


MG 02 17/08/2009 FA-200/300/400 Fully Automatic Biochemistry Analyzer Page 32 of 32

S-ar putea să vă placă și