Sunteți pe pagina 1din 32

R-Biopharm AG

Good ELISA Practice


Manual

R-Biopharm Innovative Lsungen fr die Lebensmittel- & Futtermittelanalytik


Good ELISA Practice Manual 2
3

Disclaimer

The Good ELISA Practice (GEP) manual represents a guideline for the establishment of reasonable framework
conditions and conditions of use, which shall be complied with, when using test-kits of R-Biopharm AG and
performing ELISA-analysis.

The GEP manual makes no claim to completeness, but describes only certain minimum standards. The
compliance with such minimum standards does not guarantee the achievement of correct analytical results,
but serves to increase the quality of assessment at ELISA-analysis. The manual applies in addition to the
respective detailed instructions of the test-kits. For the proper use of test-kits, the respective instruction
manual is decisive and shall take priority over this GEP manual.

The GEP manual is continuously up-dated. Please make sure that you have the newest version of the manual
prior to performing the ELISA-analysis. The manual can be retrieved, printed and downloaded under the
website www.r-biopharm.com/products/food-feed-analysis.

Introduction

The Good ELISA Practice (GEP) manual provides a comprehensive overview for both beginners and
advanced analysts in order to improve the quality of performed ELISA analysis.

The manual is divided into three chapters, which follow the work flow of an ELISA based analysis:

In the first chapter, basic knowledge about the test principle, ELISA components and required laboratory
equipment is given.

The second chapter describes the implementation of the analysis consisting of sample preparation and
ELISA procedure.

The third chapter explains the process of data evaluation from measurement to interpretation.

Additionally, FAQs are available on request.


Good ELISA Practice Manual 4

Inhalt

1 ELISA Basics 6
1.1 Antibody Antigen Detection 6
1.2 Analyte 6
1.3 ELISA Formats 6
1.3.1 Sandwich-ELISA 6
1.3.2 Competitive ELISA (formats) 7
1.4 ELISA components 7
1.4.1 Microtiter plate (MTP) 7
1.4.2 Conjugate and substrate 8
1.4.3 Standards (Calibrator) 8
1.4.4 Buffers 8
1.4.5 Stop-solution 8
1.4.6 Additional components 8
1.5 Lab equipment and its maintenance 9
1.5.1 Pipettes 9
1.5.2 Washer 9
1.5.3 ELISA reader 9
1.5.4 Automation 10
1.5.5 Additional equipment 10
1.5.6 General words of advice 10
1.6 Good laboratory practice (GLP) 10
1.7 Test kit labeling 10

2 Sample preparation and test implementation 11


2.1 Pipetting techniques 11
2.1.1 General pipetting instructions 11
2.1.2 Forward pipetting 12
2.1.3 Reverse pipetting 12
2.1.4 Pipetting of organic solvents 13
5

2.2 Handling of samples 14


2.2.1 Storage of unprepared samples 14
2.2.2 Sample drawing 14
2.2.3 Sample preparation 14
2.2.4 Use of frozen samples 15
2.2.5 Certified reference material 15
2.3 Preparation and handling of components 16
2.3.1 Storage of kits 16
2.3.2 First in first out 16
2.3.3 Pre-Warming 17
2.3.4 Temperature control 17
2.3.5 Avoiding of contamination and sample mix-up 18
2.3.6 General test handling 18
2.3.7 Time management for pipetting 19
2.3.8 Correct washing 20
2.3.9 Storage of unused components for further experiments 21
2.3.10 Interchange of reagents between tests and batches 21
2.3.11 Security references 21
2.4 Stopping and measuring of the ELISA 22
2.5 Parallel performance of tests 22

3 Data evaluation and interpretation of results 23


3.1 Determination of unknown samples by standard curve 23
3.2 Standard curve fittings 23
3.3 Standard curves of sandwich and competitive ELISAs 24
3.4 Spectrophotometer and software 25
3.5 Compliance of standard curve 26
3.6 Determination of analyte concentration 26
3.7 Measuring range and dilution factor 27
3.8 Units and dimensions 28
3.9 Limit of detection and quantification 28
3.10 Trueness and recovery 29
3.11 Specificity and cross reactivity 30
3.12 Interferences and matrix effects 31
Good ELISA Practice Manual 6

1 ELISA Basics

1.1 Antibody Antigen Detection


Antibodies are proteins produced in plasma cells of Performing an ELISA involves at least one specific
vertebrates as part of the adaptive immune system antibody for a particular antigen. A main principle
against structures (antigens) which are recognized is that one of these immunological components
as foreign to the body. is immobilized to a solid phase, the cavities of
a microtiter plate. The analyte from the sample
Antibodies bind to their antigens by a distinct interacts with the antibody-antigen system. This
pattern of ionic and hydrophobic interactions, interaction can be visualized by enzymes, linked
hydrogen bridge linkages and Van-der Waals to secondary antibodies or antigens, and indicate
forces. The interaction between antibody and its if an antibody-antigen binding has occurred.
antigen is selective and highly specific, similar to a An added substrate is converted by the coupled
lock and key. The Enzyme Linked Immunosorbent enzyme resulting in a change of colour, which can
Assay (ELISA) is based on this selective and specific be measured with a spectrophotometer.
antibody-antigen recognition. Many formats of
qualitative or quantitative ELISA tests have been
established, a selection of them are illustrated in
the following chapters.

1.2 Analyte
The claimed analyte of an ELISA could be (1) a proteins e.g. staphyloenterotoxins A, B, C, D, and
defined chemical substance e.g. aflatoxin B1, E, (5) a more or less defined group of proteins
(2) a group of defined chemical substances e.g. from a food commodity e.g. caseins (as a part of
aflatoxins B1, B2, G1 and G2, (3) a specific protein milk proteins), (6) a food commodity e.g. peanut
e.g. staphyloenterotoxin A, (4) a group of specific protein.

1.3 ELISA Formats


Currently the following three different systems
exist. In every system, the result is measured on
the basis of the optical density.

1.3.1 Sandwich-ELISA

In this method a specific antigen (analyte) in washing, substrate is added and the results can be
the sample binds to antibodies attached to the measured via the optical density (OD). The signal
solid phase of a microtiter plate. After washing, is proportional to the amount of antigen (analyte)
conjugated antibody is added which binds to a (Figure 1).
second epitope of the antigen (analyte). After
7

1.3.2 Competitive ELISA (formats)

There are two possible formats for a competitive analyte that is present, the smaller the OD value.
ELISA. The first system consists of a microtiterplate This method is suitable to measure samples with
where the antibody is bound to the surface. just one epitope as well as small analytes such as
The competition is between the analyte and an mycotoxins or antibiotics (Figure 2).
enzyme-analyte conjugate. The second system is
based on a microtiter plate with a fixed amount of For the second system (with an antibody bound
analyte which is bound to the surface. The analyte on the surface) a capture antibody is often used
from the sample is added and the competition (Figure 3). The advantage of this system is that
on antibody binding sites starts after adding the the reaction starts at the point of time the capture
antibody. After washing, substrate is added and antibody is filled in, so there will be nearly the
the measured OD value is inversely proportional same incubation time over the whole plate.
to the amount of analyte in the sample. The more

Fig. 1: Schematic structure of Fig. 2: Schematic structure of Fig. 3: Schematic structure of


sandwich ELISA competitive ELISA indirect competitive ELISA

1.4 ELISA components


As already mentioned above, several components (calibrators), substrate, stop-solution, and buffers.
are needed for an ELISA test system: Microtiter Furthermore, it is essential to use controls to be
plates (MTP), enzyme-conjugates, standards sure that the test procedure is working correctly.

1.4.1 Microtiter plate (MTP)

The MTP (96 or 48 wells), is the basis for the of these plates. Without this activation no or only
analysis. In every well the antibody or antigen minor binding of antibodies or antigens occurs. In
(depends on the format) is bound to the surface. every well the antigen-antibody reaction and the
A common plate material is polystyrene though conjugate-substrate reaction takes place. Lastly the
other materials can be used. The material is reaction is stopped with a specific stop-solution
activated by - or -radiation by the manufacturer and the optical density is measured.
Good ELISA Practice Manual 8

1.4.2 Conjugate and Substrate

Antibodies or analytes linked to an enzyme are chromogenic mix which reacts with the conjugate
called conjugates. The linked enzyme converts its enzyme. The result is a coloured solution of which
specific substrate into a colored bluish product. the optical density can be measured.
The substrate is usually a hydrogen peroxide/
QUALITY ASSURANCE CERTIFICATE

RIDASCREEN Gliadin
Art. No.: R7001 Lot: 14383 Expiry: 2014-12
1.4.3 Standards (Calibrator) R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications

Standards

All quantitative ELISA systems are calibrated by 2.4


Standard curve
Std.
Conc.(ppb)
n
mean
CV(%)

the use of standards. Therefore, the samples with 2.2

Std1 8
2.0
0.00 0.058

unknown concentrations and a set of standards 1.8


9.3

Std2 8
5.00 0.294
1.6

with known concentrations are analyzed in parallel A 5.7


b
s 1.4 Std3 8
o 10.00 0.635

on one plate. The result will be a calibration curve


r 4.2
b 1.2
a Std4 8
n 1.0 20.00 1.094
c

(with the associated mathematic formula) built out


2.8
e
0.8 Std5 8
40.00 1.683
0.6 2.2

of the measured OD values and the concentrations 0.4


Std6
80.00
8
2.366
2.1

of the standard (Figure 4). Based on this, the 0.2

0.0
5.00 10.00 20.00 40.00 80.00

analyte concentration in the sample can be Concentration (ppb)

calculated. Fig. 4: Calibration curve

Lot No. Expiry


Microwell plate 15043 2014-12
Standards 11353 2015-04
Conjugate 11353 2015-04
Buffer1 12173 2015-03

1.4.4 Buffers
Substrate/Chromogen 15183 2015-10
Stop solution 15183 2018-04
Washing buffer 11243 2015-11

Please note:

All ELISA systems contain components of biological also has an influence on test kit performance. For
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of the
curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8.
Indication of instability or deterioration of reagents.
origin. For long term storage of these components the convenience
sign.: Edda Rohm
of test kit users, these buffer areDate: 2013-09-19
and for proper function during the testing included ready-to-use or as concentrates. They
Quality Assurance Representative

Remark: This document has been created electronically and is therefore valid without a signature.

procedure, pH-value and ionic strength needs to are also used for sample preparation and washing
be constant. Often the kind of buffer component procedures of the microtiter plates.
The R-Biopharm group is DIN EN ISO 9001 certified.
www.r-biopharm.com

1.4.5 Stop-solution

The stop-solution terminates the enzyme- the reaction the colour will change from blue to
conjugate-reaction. In most cases, sulphuric yellow and will remain stable until measurement
acid at low concentrations is used. By stopping within 10 min.

1.4.6 Additional components

For most systems a positive control is This gives the opportunity to control your test
recommended, for example a spiking solution. system.
9

1.5 Lab equipment and its maintenance


Depending on the requirements of the test computer to calculate concentrations. However,
system, various specialized laboratory equipment not all of the equipment mentioned here is
is needed for the different steps e.g. pipettes, needed for all test systems. In every case regular
equipment for plate washing, incubator for maintenance and calibration is required for
constant temperature, ELISA reader and a machine and lab equipment which are used.

1.5.1 Pipettes

A pipette is used to transfer a precisely defined Multistepper pipettes (with the possibility to
volume of a solution to e.g. the wells of a pipette multiple times of a specific volume); used
microtiter plate. Pipettes are crucial for results normally for addition of antibody or conjugate
with a high precision. Therefore, special attention solutions
should be set on maintenance by experts and Multi-channel pipettes (with the possibility to
proper pipetting technique. Different kinds of pipette the solution in 8 or 12 cavities at the
pipettes exist: same time); used normally for washing steps or
Single-channel pipettes with fixed volumes e.g. addition of antibody or conjugate solutions
50 l Bottle top dispenser (used normally for washing
Single-channel pipettes (e.g. with variable steps)
volumes between 10 and 100 l); used normally Fully automatic machines (all pipetting and
for samples and standards incubation steps are done automatically)

1.5.2 Washer

After every incubation step (except the incubation step when performing an ELISA to obtain results
with the substrate) the ELISA plate has to be with high precision.
washed with washing buffer. Washing is a crucial

1.5.3 ELISA reader

The ELISA reader is a spectrophotometer which Regular maintenance by experts is essential for
allows to measure the optical density (OD) and exact results.
to calculate the concentration of your sample.
Good ELISA Practice Manual 10

1.5.4 Automation

One possibility for working with an ELISA is the use kind of automation are the ChemWell and the
of an automated system which allows you to test GEMINI. For further information please contact:
your samples without any manual steps. Therefore, sales@r-biopharm.de.
the automated system has to be validated and
calibrated for your test system. Possibilities for that

1.5.5 Additional equipment 1.5.6 General words of advice

In some tests an incubator is required to To ensure a high precision of the results, the
guarantee a stable temperature while the test is equipment should be calibrated regularly. Please
running. Sometimes, a seal or a protecting plate ask the manufacturer for the calibration interval
cover are necessary to prevent evaporation or and include this in quality control plans.
contamination.

1.6 Good laboratory practice (GLP)


Dependent on the toxicities and contagiousness However, it is necessary to study the material
of the used materials, different levels of protective safety data sheets (MSDSs) carefully for all
actions are necessary to guarantee health and chemicals substances used. The MSDS contains
safety of the user. However, basic protective information about the dangers when working
clothing is already required to avoid contamination with a particular substance, required protective
of the samples, which would lead to incorrect measures, as well as required actions in case of
results. The following equipment is a minimum emergencies. The MSDS are available on request
requirement for every lab: for each product. Please contact:
sales@r-biopharm.de.
Lab coat Gloves
Eye protection Fume hood
(occasionally)

1.7 Test kit labeling


To ensure the correct handling and storage of your The expiry date is particularly important, as the
components please read the Instructions for Use specified behavior can only be guaranteed until
(IFU) or the test kit insert. On all components the then. After expiration the ingredients can degrade
follow information is stated (if applicable): and the results can decrease in accuracy.

Product name Lot number


Article number Storage temperature
Name of the Concentration
component
11

2 Sample preparation and test


implementation

2.1 Pipetting techniques


Accurate and precise pipetting is crucial in ELISA The two pipetting techniques used for ELISA are
analytics, particularly at high sensitivity levels forward pipetting (standard pipetting) and reverse
where a small mistake in pipetting can induce pipetting. The forward pipetting technique is
large differences in the final test results. Be recommended for aqueous solutions. Using this
consistent during pipetting and do not change technique, some liquids may induce bubbles or
your technique while pipetting an assay. Be foam during pipetting. As an alternative, reverse
prepared before starting pipetting e.g. put all pipetting lowers this risk and is recommended
standards in a row and arrange all samples in a for liquids with higher viscosity. It requires more
consistent way to allow uniform pipetting. Ensure liquid volume (dead volume) and is more error-
that enough pipette tips are prepared and a waste prone at high volume transfers. Familiarize yourself
container is in place. with pipetting techniques before you start the
experiments.

2.1.1 General pipetting instructions

If you use an adjustable pipette please adjust it Always check the adjusted volume directly before
to the desired volume before you start pipetting. starting to pipette (Figure 5).

Forward Reverse
Ready position

First stop

Second stop

Fig. 5: Schematic illustration of forward and reverse pipetting techniques when using a single channel pipette (piston pipette).
Blue blocks indicate the steps where the pipette needs to be placed in liquid to aspirate. Grey blocks indicate the target wells.
Good ELISA Practice Manual 12

2.1.2 Forward pipetting 2.1.3 Reverse pipetting

1
1 Put a new tip on your pipette and check for a 1 Put a new tip on your pipette and check for a
1
firm fit. firm fit.

2
1 Press the operation button to the first stop. 2
1 Press the operation button to the second stop.

3
1 Pipette tips from some manufacturers need to 1 Put the pipette tip approx. 1 cm deep into the
3
be flushed before aspiration and dispensing liquid. Slowly release the operating button to
of the appropriate liquid. Please check the ready position and wait until the desired liquid
according manual. In case of any doubt, flush volume has been aspirated. Ensure that not
the tip before pipetting. bubbles or foam occurs in the pipette.

4
1 Put the pipette tip approx. 1 cm deep into the 1 Remove excessive liquid from the outside of the
4
liquid. Slowly release the operation button to tip.
the ready position and wait until the desired
liquid volume has been aspirated. Ensure that 1 Dispense the liquid into the desired well by
5
no bubbles or foam occurs in the pipette. pressing the operation button to the first stop.
Ensure that no liquid remains on the outside of
5
1 Remove excessive liquid from the outside of the the tip.
tip by touching the test tube with the tip.
1 For repetitive liquid pipetting, press the
6
1 Dispense the liquid into the desired well by
6 operation button to the first stop and repeat
pressing the operation button to the second steps 3 - 5.
stop.
1 Remove the tip to waste.
7
1 Remove the tip to waste.
7
13

2.1.4 Pipetting of organic solvents

Organic solvents show high vapour pressures Fig. 6: Pipetting of organic solvents with pipettes using the
which can affect precise pipetting. The use of air displacement technique. Particular care should be taken
to prevent evaporation into and leaking out of the tip (A).
pipettes with air displacement technique to transfer Flushing the pipette before liquid transfer helps to transfer the
organic solvents may lead to evaporation of the correct volume (B).

solvent or a leaking out of the tip (Figure 6). For the


pipetting of organic solvents we recommend:

1
1 multistep pipettes which use the positive
displacement technique

1
2 serological pipettes for larger sample volumes,
since the graduation allows for the pipetting of
exact volumes

1
3 bottle top dispensers (A) (B)
1
4 pipettes designed especially for the handling of
organic solvents

If single channel pipettes are used to transfer


organic solvents, the pipette tip and the air inside
the pipette needs to be saturated with organic
solvent vapour before pipetting the desired
volume. For this aspirate and dispense the organic
solvent at least 3 times before the desired volume
is transferred. Use appropriate quality control
procedures to monitor the correctness of these
kinds of pipetting steps.
Good ELISA Practice Manual 14

2.2 Handling of samples

2.2.1 Storage of unprepared samples

Unwanted contamination of samples can influence Generally, samples should be prepared and tested
test results significantly. If there are signs of immediately whenever possible. If storage cant
unwanted contamination or spoilage do not use be avoided check for optimal storage conditions
the sample and request a new one. Store the and analyze them as soon as possible. All samples
sample according to the test kit manufacturer or need to be correctly labelled and sealed to avoid
according to the best scientific knowledge. evaporation or dry-out. Inappropriate storage
conditions may influence later analysis and alter
test results.

2.2.2 Sample drawing

The drawing of a representative sample is a be homogenized as much as possible. Inform a


crucial step. For some parameters, legislations customer beforehand about a minimum sample
apply on how to take a representative sample amount. State in the test report that the result
(e.g. mycotoxins). Please follow these rules refers to the sample as it was sent.
precisely. If there are no rules, samples should

2.2.3 Sample preparation

For sample preparation please follow the Depending on the parameter to be tested,
instructions for use provided with the test kit. instructions for use may contain information on
Changes or variations may lead to incorrect how prepared samples can be stored for later
test results. Make sure to use only suitable and analysis. Please follow these instructions carefully
maintained equipment for sample preparation. For or prepare samples directly before analysis.
any related questions please contact R-Biopharm AG.
15

2.2.4 Use of frozen samples

Before further use, samples must be completely should be avoided since it may denature proteins.
thawed. Thawing of frozen samples should A further sample preparation of liquid samples
be performed at 4C or at room temperature, (e.g. in case of milk) may be necessary before
dependent on analyte stability. analysis. Please check the instructions for use for
further information.
In case of unprepared, non-liquid samples
check the instructions for use for further sample Depending on the sample, freezing and thawing
preparation and homogenization. can lead to crystallization or coagulation. Avoid
freeze-thaw cycles wherever possible, since it can
Liquid samples must be thoroughly mixed change sample integrity and alter test results. If
before they can be used for analysis. To achieve a possible, aliquot liquid samples before storage at
homogeneous sample carefully vortex or invert 20C to avoid freeze-thaw cycles.
the sample. Foam formation or intensive mixing

2.2.5 Certified reference material

Certified reference materials (CRM) are naturally Fig. 7: Certified reference materials and standard solutions are
essential quality control tools.
contaminated, homogeneous matrices whose
analyte content has been exactly and reliably
determined (Figure 7). The regular use of CRMs
is recommended to perform internal quality
controls. This allows checking for the trueness and
precision of experimental procedures and for the
testing of handling skills. If no reference material
is available, the use of matrices spiked with
defined analyte concentrations is recommended.
RBiopharm can help you to find suitable certified
reference materials and spiking solutions.
Good ELISA Practice Manual 16

2.3 Preparation and handling of components

2.3.1 Storage of kits

The expiry date printed on the outer label applies Fig. 8: Storage conditions and expiry date are printed on
the outer label of each ELISA test kit.
to all reagents contained in the kit and is also valid
after first use. To maintain the shelf life, store the
kit at conditions noted on the outside label of the
package (Figure 8).

Most kits have to be stored in a dry place and at


4C. Freezing of components needs to be avoided.
If you use a refrigerator for storage, make sure the
kits are not stored close to the back wall to avoid
freezing. Improper storage of kits or components
may lead to damage. Freezing of test components
may decrease test performance and lead to invalid
test results or as an extreme- non-functional test
kits. Indicators for false storage are decreased or
non-detectable optical density and the alteration
of test results.

2.3.2 First in first out

The expiry date of the kit is labelled on the outer The exposure of the ELISA kit to cold/warm cycles
label of the kit package. At least until this date the should be kept as low as possible. We recommend
kit will perform within specifications. Additionally that samples should be collected. Testing of higher
every kit component has its own expiry date which sample numbers at once reduces the expenditure
is identical or even exceeds the expiry date of the of time per sample. Rather test higher sample
test kit. numbers at once than test only a few samples
consecutively. Please check the instructions for use
If you have more than one kit to hand it is for relevant limitations.
recommended to use the first in first out
principle. This means the kit with the shortest
expiry date on the outer label should be used first.
We recommend indicating the date of the first
use on the outer label to avoid a mix-up of this
principle.
17

2.3.3 Pre-Warming

All reagents need to be at room temperature Fig. 9: Bring all test kit components to room temperature
before use and perform the test at 20 - 25 C (68 - 77 F).
before they can be used in the test. Take all
components out of the kit package before use
and allow them to reach room temperature
(20 - 25 C). Larger bottles and higher 20 - 25 C/
volumes may require more time to reach room 68 - 77 F
temperature. Check the temperature of the
components in any case of doubt (Figure 9).

After use it is recommended that all components


are put back into the kit package to avoid
the unintended mixture of components with
components from other kits or lots. Return the
kit back to the recommended storage conditions
as soon as possible (see outer kit label for storage
conditions).

2.3.4 Temperature control

ELISA tests are sensitive to temperature should be prevented from direct exposure to
fluctuations. Therefore try to stabilize and sunlight and ventilations. Cold lab ware and cold
control laboratory conditions. This includes the benches may also influence the temperature. It
temperature during photometric analysis. Perform is helpful to isolate the microtiter plate from the
ELISA tests between 20 and 25 C and avoid bench surface by performing the test on a suitable
conditions that are able to drastically change the underlay. A cheap and easy solution is the use of
temperature or increase evaporation. ELISA tests paper towels.
Good ELISA Practice Manual 18

2.3.5 Avoiding of contamination and sample mix-up

A clean and reproducible way of working is crucial Fig. 10: Carefully label all used containers and document
your labelling.
for optimal results in food and feed analysis. A very
common source of contamination is insufficiently
cleaned re-usable lab ware. To avoid this, it is
highly recommend to use solvent-resistant single-
use lab ware. If this is not possible, re-usable
lab ware should be laboratory sterile and free of
contamination. We recommend using a laboratory
dishwasher or equivalent. Use quality control
blank samples to check for contamination.

Reagents should be handled with calibrated


devices, clean pipettes and containers. Only take
the amount of reagent needed and do not put
reagents back into the container once it has been
removed. Use separate containers and pipette tips
for every reagent to prevent cross contamination.
Make sure to label all containers correctly
(Figure 10). Exchange your single use lab ware
as often as needed.

2.3.6 General test handling

Before you start, read the instructions for use To obtain optimal results, test samples and control
enclosed in the kit. Prepare all extraction solutions samples (standards, reference samples) have to
and buffers according to these instructions and be available in the same diluent. Strictly following
follow the described procedures to obtain optimal the sample preparation protocol ensures this. The
results. To allow an unobstructed test procedure pipetting onto the plate needs to be performed
it is helpful to prepare a pipetting scheme before quickly and without interruptions at every step of
you start with your experiment. the test procedure. For the transfer of samples we
recommend single channel pipettes, or, if a pre-
Depending on the ELISA, antibody and conjugate dilution plate is used, a multi-channel pipette. For
solution may need to be diluted prior to use. the pipetting of antibody and enzyme solutions,
These dilutions should be prepared directly multistep pipettes are the best option.
before use and shall not be stored for further use.
Contaminated or incorrectly stored conjugate Between handling steps a dry-out of the wells has
solutions may have a reduced enzyme activity or to be prevented.
may induce a background signal.
19

2.3.7 Time management for pipetting

Pipetting of standards and samples is general the instructions for use. To obtain comparable results
most time consuming and laborious step when in all wells, the incubation time of each single well
performing an ELISA. In general it is advised needs to be identical.
to keep the pipetting technique as constant as
possible and the absolute pipetting time to a To achieve this, start the clock after you have
minimum. This is extremely important when pipetted a component starting a reaction
the last component is added and the analytical (standards or samples, antibody solution,
reaction starts (e.g. addition of antibody in a conjugate solution) into the last well. In case of
competitive ELISA when sample and conjugate substrate/chromogen solution, start the clock
is already added). To avoid an optical density before you pipette the solution into the first well.
(OD) shift from the first to the last well, it is Stop the substrate/chromogen reaction after the
recommended to use multichannel or stepper defined time by adding stop solution in the same
pipettes for components which are added to all order substrate/chromogen solution has been
wells such as antibody, conjugate, substrate/ added. It is very important to meet the incubation
chromogen or stop solution. However, it is not times noted in the instructions for use.
recommended to pipette in a rush as mistakes
such as to pipetting into wrong wells, etc. may The activity of the chromogen may be influenced
occur. by light. Protect the chromogen from light and
store it in the brown flask in which it is delivered.
It is very important that all samples are handled To stop the reaction use the stop solution
in a comparable way. Strictly follow the pipetting delivered with the kit.
order and the incubation times noted in the
Good ELISA Practice Manual 20

2.3.8 Correct washing

Washing is a crucial step to remove all unbound Automated washing systems are not available at
components that might influence reactions or all testing sites. Due to this we highly recommend
lead to false results. For washing only use washing the use of the relatively cheap and easy to handle
buffers recommended in the instruction for use. bottle top dispensers with an 8 - or 12 -fold
Many kits contain washing buffer salts or solutions manifold for your washing procedure. Other
that can be used to prepare ready to use washing manual washing techniques like washing bottles
buffers. This removes the necessity to weigh may not allow to treat every well exactly the same
reagents to prepare your own buffers. For stability and should not be applied.
and storage information on the individual washing
buffers delivered with the kits, see the instructions Washing steps should be performed fast but
for use. As all other components, washing buffers efficiently and accurate. Make sure that the time
need to be at room temperature before use. Follow between addition of washing buffer to the first
the specific recommendation of the instructions and the last well is as short as possible. This
for use regarding the number of washing steps. ensures that wells do not dry-out and it minimizes
differences in incubation times. Despite fast
At the end of the incubation steps pour the liquid working speed pay attention to accuracy. Spillover
out of the wells and tap the microtiter plate holder of liquid from one well to another needs to be
vigorously upside down on absorbent paper to avoided. In case of any doubt add an additional
ensure the complete removal of the liquid from washing step.
the well. All liquid has been successfully removed
Fig. 11: Bottle-top dispenser plate washers with manifolds are
when no signs of liquid remains on the paper an accurate and affordable option for washing steps.
towel. Most ELISA tests require 250 - 300 l of
washing buffer per washing step and well. Add
the washing buffer firmly and remove the liquid by
pouring out and tapping. Repeat the washing step
3 - 5 times (see instructions for use). For washings
steps we recommend to use a bottle-top dispenser
with manifold (Figure 11).

If you use such an automatic washer, service,


maintain and use the device according to the
manufacturers instructions. Ensure that the device
does not show salt residues that may prevent
liquid flow. Buffers, tubes, manifolds and washing
needles need to be kept free from contamination
of microorganisms. Keep your device clean and if
necessary remove contaminated parts. In case of
any doubt, use the manual washing procedure or
contact the manufacturer of the washer.
21

2.3.9 Storage of unused components for further experiments

Microtiter plates are delivered in a re-selable bag important for components such as standard
with a pouch containing a desiccant (Figure 12). solutions that may contain organic solvents with a
In case not all wells of the plate are needed, store high vapour pressure. We recommend putting all
the rest of the wells in this bag. Put the wells and components back into the kit package for storage.
the microtiter plate together with the desiccant Store all components upright and under the
into the bag and close it. Close all flasks and make indicated conditions (Figure 13).
sure to screw the lids on firmly. This is especially

Fig. 12: Unneeded plate strips should be stored together Fig. 13: Until further use all components should be stored in
with the desiccant in the re-sealable pouch in which they are the kit package in an upright position.
delivered.

2.3.10 Interchange of reagents between tests and batches

The components of each lot are thoroughly allowed. An exchange of components of kits
adjusted to deliver ELISA kits that show the optimal with the same product number is possible if the
performance. The exchange of one or more of lot number of the kit is identical. However, we
these components between different lots will recommend using only components delivered
change the performance of the tests and is not with the particular kit.

2.3.11 Security references

ELISA test kits may contain hazardous substances. laboratory security precautions. While performing
For information on hazards contained in the the test procedure use laboratory gloves, wear
substances take note of the warnings on the labels a lab coat, do not eat, drink, smoke and keep
of the components and refer to the appropriate all components away from sources of ignition.
material safety data sheets (MSDS). Generally Disposal of waste may differ from country to
handle all components with care and take all usual country. Please refer to local disposal rules.
Good ELISA Practice Manual 22

2.4 Stopping and measuring of the ELISA


Most ELISAs are measured at a wavelenght of Fig. 14: Addition of stop solution causes a color change from
blue to yellow.
450 nm. The correct wavelength for reading can
be found in the test kit manual under point 3. test
principle. At the end of the test implementation,
stop solution which contains sulfuric acid is added
to each microtiter well. The acid denatures all
proteins including the antibodies and thus stops
the reaction (Figure 14).

Nevertheless it is recommended to read the


microtiter plate directly after addition of stop
solution or at least within the time stated in the
instructions for use. A large delay may still cause a
shift of the absolute values measured.

2.5 Parallel performance of tests


If several ELISA tests are performed in parallel, Automation of parallel analysis is possible by
extraordinary care should be taken. Label microtiter using biochemistry analyzers such as ChemWell
plates and reagents properly to avoid mistakes and (Awareness Technology Inc.) or Gemini analyzer
mixing of reagents between assays. Use a separate (Stratec Biomedical AG). Please contact us for
lab timer for each microtiter plate (Figure15). Take a list with tests, which are already validated on
care, that handling steps of different assays do not biochemistry analyzers.
overlap, e.g. washing of one microtiter plate when
a second one has to be stopped. Schedule your
pipetting, washing and OD reading activities before
starting multiplex analysis.

Fig. 15: Parallel performance of several ELISA tests must be carefully scheduled and organized.
23

3 Data evaluation and interpretation


of results

3.1 Determination of unknown samples by standard curve


The concentration of the analyte in an unknown used to generate a standard curve covering the
sample can be determined by comparing the concentration range of interest. There are also
measured signal of the sample with the signal of ELISAs which use single calibration technology
standards containing known concentrations of (SC), where a single standard is used to check
the analyte. In ELISAs usually 5 - 7 standards are compliance with a deposited standard curve.

3.2 Standard curve fittings


QUALITY ASSURANCE CERTIFICATE
Depending
Q U A L Ion
T Y Athe
S S U assay,
RANCE Cthe
E R Tstandard E L I Tcurve
I F I CQAUT A Y A S S Uis
R A N C E C E R T I linear
F I C A T E regression, logit-log, cubic spline and
QUALITY ASSURANCE CERTIFICATE
RIDASCREEN Gliadin
RIDASCREEN Histamin (enzymatic) RIDASCREEN FAST DON
calculated by different curve fittings: 4 parameter 2 order polynomial
nd
(Figure 16)FAST DON
RIDASCREEN
QUALITY ASSURANCE CERTIFICATE
Art. No.: R5901 Lot: 13504 Expiry: 2015-10
Art. No.: R1605 Lot: 12184 Expiry: 2014-11 Art. No.: R7001 Lot: 14383 Expiry: 2014-12
Art. No.: R5901 Lot: 13504 Expiry: 2015-10 Pag
R-Biopharm
R-Biopharm AG, Darmstadt, Germany certifies that this AG, approved
batch has been Darmstadt, Germany
by the Quality certifies that this batch has been approved by theRIDASCREEN
Quality Gliadin
Assurance Department and conforms with specificationsAssurance Department and conforms withR-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Qualit
specifications
Assurance Department and conforms with specifications
Linear regression Logit-log Cubic spline Art. 4 parameter
Datasheet
No.: R7001 Lot: 14383 Expiry: 2014-12
Standards Standards Standards
Standard curve Standard curve Standard curve R-Biopharm AG,
Standards
3.5 Std. n Std.Darmstadt,
n Germany
Standardcertifies
curve that this batch has been approved by the Qua
2.0 2.4 Std. n
Conc.(ppb) mean
CV(%) Assurance Department and conforms with specifications
Conc.(mg/L) mean Conc.(ppm) mean 3.5 Std. n
Conc.(ppm) mean
3.0 CV(%) CV(%) B/B0
1.8 2.2 3.0 CV(%) B/B0

RIDASCREEN Calprotectin / One-Point


2.5 Std1 8

Standard curve Calibration and 4 Param


Std1 2 2.0 Std1 4 2.5 Standards
A 0.00 0.009 0.00 0.058 Std1 4
1.6 0.00 1.937
2.0 9.3 Std.
0.00 n
1.937
b 7.9 0.2 100.0 2.4
2.0
L 1.8 Conc.(ppb)
0.2 mean
100.0
s 1.5 Std2 L Std2 8 CV(%)
1.4 o 2 Std2 4
1.55.00 0.294
o 1.00 0.106 1.6 0.222 1.328 o 2.2 Std2 4
g 5.7
1.0 0.0 A 2.4 conc.
68.6 of calprotectin 0.222
Std1calculated
1.328
68.6 values
r b g 1.0 8
1.2 i Calprotectin mean Standard Curve 2.4
b 0.5 Std3 2
s 1.4
Std3 and
4 corresponding OD i
2.0 Std3 8 0.00 0.058
t o 10.00
0.5 0.635 9.34-parameter logisti
a 5.00 0.484 r 0.666 0.862values of mean 3.5
t3.500
1.8 4.2
Std3 4
1.0 0.0 0.6
n
b 1.2 2.0 44.5
standard curve 0.0
0.666
Std2 model
0.862
8 (4PL)
a 3.0
3.000 Std4 8 2.0
5.00
44.5
0.294
c -0.5 Std4 2 n 1.0 Std4 4 1.620.00 1.094 5.7
A 0,0
OD [450 nm; Ref. 620 nm]

0.8 A
e
10.00 0.938 c 2.00 mg/kg
0.422 stool OD -0.5
2.5 2.8
b 2.500
Std4 4
-1.0 0.6 e 3.9 21.8 2.00
Std3 0.422
8
0.8 0 0,010 s 1.4 -1.0
Std5 8 3.9 21.8
0.6
-1.5 Std5 2 Std5 4
2.0
o 2.000
40.00 1.683
10.00
B 0.635
1,5
15.00 1.399 0.6 6.00
19,5
0.204
0,091 r
b 1.2 -1.5
2.2
4.2
Std5 4
1.5
a 1.500
0.4 -2.0 0.9 4.7 10.5
33 0,169 n 1.0 Std6
-2.0 8
Std4
6.00
4.7
20.00
C 8
0.204
10.5
1.094
5,4
0.4
1.080.00
c 1.000
-2.5 Std6
20.00
2
1.844
56 0,370 e 2.1
-2.5
2.366 2.8
D 3,6
0.2 0.8
-3.0 0.5 0.2
95 0,733 0.5
0.500
-3.0
Std5
40.00
8
1.683
0.0 -3.5 0.0 160 1,188 0.6
0.0
0.000 2.2
0.00 5.00 10.00 15.00 20.00 0.222 0.666 2.00 6.00 5.00 10.00 20.00 40.00 80.00 -3.5
275 2,015 1 0.222 10
0.666 100
2.00 1000
6.00 Std6 8
Concentration (mg/L) 0.4
Concentration (ppm) Concentration (ppb) [mg Calprotectin/kg stool] 80.00 2.366
470 2,704 Concentration (ppm) 2.1
0.2
800 3,113
X-axis
Corr.Coeff.: 0.9999 Corr.Coeff.: 0.9983
slope = 0.0918 50% inhibition = 0.519 0.0 Corr.Coeff.: 0.9983
5.00
50% inhibition10.00
= 0.519 20.00 40.00 80.00

Concentration Concentration (logarithmic) Concentration (logarithmic)


Lot No.
Concentration
Expiry
(logarithmic) Concentration (ppb)
Lot No. Expiry Lot No. Expiry
Microwell plate 15043 2014-12 Lot No. Expiry
Y-axis Microwell plate
Standards
14463
12084
2014-11
2015-01
MTP M
Standards
15464
12494
2017-04
2016-05
Standards
Conjugate
11353 2015-04
MTP M 15464 2017-04
Negative control12173
11353 2015-04
Buffer1 13084 2015-01 Conjugate 11484 2015-10 Standards 12494 2016-05
OD Enzyme Solution
Spike Solution
13084
12124
Logit of concentration
2015-01
2015-02
Antibody
Red Chromogen Pro
11484
14413
B/B0 in %
2016-04
2016-03
Buffer1
Substrate/Chromogen
15183 OD
2015-03
2015-10
Conjugate
Antibody
11484
11484
2015-10
2016-04
Stop solution
15183 2018-04 Lot No. Expiry
Stop solution
Washing buffer salt
14114
051M8211
2019-02
2016-06
Negative control
Washing buffer 11243Diluent 2015-11 3 QC specification:
Red Chromogen Pro
Microwell plate
Stop solution
14413
15043
14114
2016-03
2014-12
2019-02
OD < 0,05
The relationship In Logit-log is used to Cubic spline is a numeric 4 parameter is a nonlinear Standards
Washing buffer salt
Conjugate
11353
051M8211
11353
2015-04
2016-06
2015-04

plots forPlease note: function that is piecewise- Please note:


Buffer1 12173 2015-03
R-Biopharm between
AG, Darmstadt,variable
Germany certifiesy and that this batch Please
has been linearize
note: approved by the curved
Quality regression model with Substrate/Chromogen
Stop solution
15183
15183
2015-10
2018-04
Assurance Department and conforms with specifications
variable x is linear. Thus, The absorbance further analysis
for the standards by linear
may decrease duringThe shelf life ofdefined
the absorbance forkit.
the theThe
bythe3shape
standards
general polynomial
may decrease
of the 4 parameters:
during the shelf life of the kit. the TheWashing
general shape of
buffer 11243 2015-11

curve will remain similar, while the slope might change curve will
slightly. remain similar,
Furthermore refer to Calibrator
while
product slope
leaflet might
8. change slightly. Furthermore
The absorbance refer
for thetostandards
product leaflet 8.
may decrease during the shelf life of the kit. The general shape o
simple linear regression is regression models. functions with high degree note: bottom of the curve
Indication of instability or deterioration of reagents. Indication of instability or deterioration of reagents. curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8
sign.: Edda Rohm Date: 2014-04-29 Please Indication of instability or deterioration of reagents.
The target value of the Calibrator Calibrator has been determined
applied using the formula: sign.: N. Stork-Heininger
Quality Assurance Representative
sign.: Edda Rohm of smoothness
Quality Assurance Representative to be:
at the
Date: 2014-12-10
The absorbance
fortop
sign.: N. Stork-Heininger
of the Date:curve
2013-09-19
the standards may decrease during the shelf life of the kit. The general
OD 1, 188shap
Date: 20

Quality Assurance Representative
y = mx + b
Remark: This document has been created electronically and is therefore valid without a signature.
connections of the curve will
Quality
Indication
remain
Assurance
of
EC50
similar, while
Representative
instability or
the slope
deterioration of
might change
reagents.
slightly. Furthermore refer to product leafle
The rangeand(target value without 3 SD) should be within: OD 0,651 1
polynomial pieces (knots). Remark: This document
Remark:
Remark: This document has been created electronically and is therefore valid This document
without has been created
a signature. electronically is therefore valid a signature.
slope atcreated
has been theelectronically
inflectionand is therefore valid without a signature.
sign.: Edda Rohm Date: 201
Quality Assurance Representative
point of the curve =EC50

The R-Biopharm group is DIN EN ISO 9001 certified.


Positive controlRemark: This document has been created electronically and is therefore valid without a signature.
The R-Biopharm group is DIN EN ISO 9001 certified. The R-Biopharm group is DIN EN ISO 9001 certified.
The R-Biopharm group is DIN EN ISO 9001 certified.
www.r-biopharm.com www.r-biopharm.com www.r-biopharm.com
The target value of the Positive control Control + has been www.r-bioph

145,1 mg/kg
determined to be:
Fig. 16: Overview about standard curve fittings The R-Biopharm group is DIN EN ISO 9001 certified.
The range (target value 3 SD) should be within: 113,3 176,9
www.r-biopharm

Low positive control


The Low positive control can optionally be tested.
The target value of the Low positive control Low control + has
54,4 mg/kg
been determined to be:
The range (target value 3 SD) should be within: 41,5 67,3 m
Good ELISA Practice Manual Q U A L I T Y A S S U R A N C E C E R T I F I C24
ATE

RIDASCREEN Gliadin
Art. No.: R7001 Lot: 14383 Expiry: 2014-12

3.3 Standard curves of sandwich and competitive ELISAs R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications

Standards
The correct algorithm for the respective ELISA 2.4
Standard curve
Std. n
Conc.(ppb) mean

is depicted on the certificate of analysis and is 2.2


CV(%)

Std1 8
preset in the RIDASOFT Win.net-Software after 2.0
0.00
9.3
0.058

1.8
choosing the method. Depending on assay format 1.6
Std2
5.00
8
0.294
A 5.7

(sandwich or competitive) the standard curve is b


s
o
1.4 Std3
10.00
8
0.635
r 4.2
calculated differently. b
a
1.2
Std4 8
n 1.0 20.00 1.094
c 2.8
e
0.8 Std5 8

In sandwich ELISAs the concentration of the


40.00 1.683
0.6 2.2

Std6 8
standards is plotted on the horizontal xaxis, while 0.4
80.00
2.1
2.366

0.2
QUALITY ASSURANCE CERTIFICATE
OD is plotted on the vertical y-axis (figure 2). 0.0
5.00 10.00 20.00 40.00 80.00

Sometimes B/Bmax, a percentage value is plotted RIDASCREEN


Concentration (ppb)
Chloramphenicol
on the vertical y-axis. For B/Bmax calculation, the Fig. 17: Standard Art.
curveNo.:
of anR1505
ELISA inLot:
sandwich
11084format,
Expiry: 2015-04
standard with the highest analyte concentration is where absorbance is plotted over concentration.
R-Biopharm AG, Darmstadt, Germany certifies thatLotthis
No. batch has been
Expiryapproved by the Qualit
set as 100 % and B/Bmax of the further standards Assurance Department and conforms
Microwell plate
Standards
15043 with specifications
2014-12
11353 2015-04
Conjugate 11353 2015-04
is calculated by dividing the absorbance of the Buffer1
Substrate/Chromogen
12173
15183
2015-03
2015-10 Standards
Standard
Stop solution curve 15183 2018-04
respective standard by the absorbance of the 100 Washing buffer 11243
Std.
2015-11
Conc.(ppt)
CV(%)
n
mean
B/B0

highest standard. The result is multiplied by 100 to Please note:


90
Std1 8
0.00 2.302
obtain percentage units (Figure 17). 80
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of t
2.3 100.0

curve will remain70similar, while the slope might change slightly. Furthermore refer
Std2 to product
8 leaflet 8.
25.00 2.068
Indication of Ainstability
b
or deterioration of reagents. 2.8 89.8
s 60
For competitive ELISAs, the concentration of the o Std3 8
sign.: Edda Rohm
r 50.00 Date:
1.751 2013-09-
Quality Assuranceb Representative 3.1 76.1
a 50
standards is plotted on the horizontal x-axis, n
Remark: This document
c has been created electronically and is therefore valid without a signature.
Std4
100.00
8
1.348
e 40 3.0 58.6
while B/B0, a percentage value is plotted on the (%) Std5 8
30
250.00 0.719
vertical y-axis. For B/B0 calculation standard 1 20
3.4 31.2

Std6 8

which contains no analyte is set as 100 % and B/B0 The R-Biopharm group10is DIN EN ISO 9001 certified.
750.00
4.2
0.306
13.3
www.r-biopharm.com
of the further standards is calculated by dividing 0
25.00 50.00 100.00 250.00 750.00

the absorbance of the respective standard by the Concentration (ppt)

absorbance of standard 1. The result is multiplied Fig. 18: Standard curve of an competitive ELISA, where B/B0
50% inhibition = 132.2

with 100 to obtain percentage units (Figure 18). is plotted over concentration
Lot No. Expiry
Microwell plate 15313 2015-04
Standards 12034 2015-09
Conjugate 14064 2015-07
Buffer1 12044 2015-12
Red Chromogen Pro 11113 2015-05
Stop solution 11373 2018-08
Washing buffer salt 051M8211 2016-06

Please note:

The absorbance for the standards may decrease during the shelf life of the kit. The general shape
the curve will remain similar, while the slope might change slightly. Furthermore refer to product lea
8. Indication of instability or deterioration of reagents.

sign.: Edda Rohm Date: 2014-0


Quality Assurance Representative

Remark: This document has been created electronically and is therefore valid without a signature.

The R-Biopharm group is DIN EN ISO 9001 certified.


www.r-biopharm.co
25

3.4 Spectrophotometer and Software


The optical density is read by a microtiter plate
spectrophotometer at a certain wavelength.
There are many different spectrophotometer from
different manufacturers available. R-Biopharm
offers spectrophotometers and a software,
called RIDASOFT Win.net, which is tailor-made
for the analysis of ELISAs from R-Biopharm
(Figure 19, 20). For the use of the software a
manual (Art. No. R9996) is available on request. If
a spectrophotometer is already available, please
do not hesitate to contact us, to check if your
spectrophotometer can be used in conjunction
with the software and our assays.

Fig. 19: Biotek ELx800 microtiter plate spectrophotometer.

Fig. 20: Screenshots of RIDASOFT Win.net:


plate layout, standard curve, results
Good ELISA Practice Manual 26

3.5 Compliance of standard curve


To check compliance of your standard curve please refer to the specific document Acceptance
and its parameters (e.g. OD of standard criteria for RIDASCREEN standard curves, which
1, 50 %-dose, B/B0-values, Coefficent of Variation is available on request for each product group.
QUALITY ASSURANCE CERTIFICATE
(CV) of standard duplicates and shape of curve)
RIDASCREEN Chloramphenicol
Art. No.: R1505 Lot: 11084 Expiry: 2015-04

3.6 Determination of analyte concentration R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications

Standards

The software creates the standards curve and 100


Standard curve
Std.
Conc.(ppt)
n
mean
CV(%) B/B0

calculates the concentration of an analyte in an 90


Std1 8
0.00 2.302
unknown sample as shown in figure 21. Outside 80 2.3 100.0

Std2 8
70
the measurement range the software will calculate A
b
s
Measured
25.00
2.8
2.068
89.8
60 Std3 8
o
no values. r
b
50
50.00
3.1
1.751
76.1
a
n Std4 8
c 100.00 1.348
e 40 3.0 58.6
(%) Std5 8
30
250.00 0.719
3.4 31.2
20 Std6 8
contentration 750.00 0.306
10 obtained 4.2 13.3

through
0
25.00 50.00 100.00 250.00 750.00

Concentration (ppt)

Fig. 21: Figure 21:50%


Determination
inhibition = 132.2
of analyte concentration
from unknown sample through standard curve
Lot No. Expiry
Microwell plate 15313 2015-04
Standards 12034 2015-09
Conjugate 14064 2015-07
Buffer1 12044 2015-12
Red Chromogen Pro 11113 2015-05
Stop solution 11373 2018-08
Washing buffer salt 051M8211 2016-06

Please note:

The absorbance for the standards may decrease during the shelf life of the kit. The general shape of
the curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet
8. Indication of instability or deterioration of reagents.

sign.: Edda Rohm Date: 2014-02-17


Quality Assurance Representative

Remark: This document has been created electronically and is therefore valid without a signature.

The R-Biopharm group is DIN EN ISO 9001 certified.


www.r-biopharm.com
27

3.7 Measuring range and dilution factor


If an absorbance or a B/B0 value is obtained The measuring range of the ELISA is determined
which is below or above the standard curve, by the concentration of the standards, and by the
the RIDASOFT Win.net software gives a results dilution factor of the sample preparation method.
< (below) or > (above) standard range. The For example, if milk is diluted 1:4 (1+3) before
software provides the option to extrapolate the applying to the microtiter well, the dilution factor
concentration value. In general, these extrapolated is 4. This means, that the results read from the
values are only estimates and are not reliable. The standard curve have to be multiplied by 4 to obtain
further the sample is below the lowest or above the correct concentration of the analyte in the
the highest standard, the bigger the uncertainty sample. If the analyte in a sample is concentrated
of the calculated concentration is. It is advised, by a factor of 2 during sample preparation, e.g.
to dilute samples which are above the largest by evaporation or column clean-up, the dilution
QUALITY ASSURANCE CERTIFICATE QUALITY ASSURANCE CERTIFICATE
standard and to repeat the analysis until the result factor is 0.5. This means, that the results read from
RIDASCREEN
is within the concentration rangeChloramphenicol
of the standard the standard curve have RIDASCREEN
to be multipliedChloramphenicol by 0.5 to
Art. No.: R1505 Lot: 11084 Expiry: 2015-04 Art. No.: R1505 Lot: 11084 Expiry: 2015-04
curve. obtain the correct concentration of the analyte in
the sample
R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications (FigureAssurance
22, 23).
R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Department and conforms with specifications

Standards Standards
Standard curve Standard curve
Std. n Std. n
100 100
Conc.(ppt) mean Conc.(ppt) mean
CV(%) B/B0 CV(%) B/B0
90 90
Std1 8 Std1 8
0.00 2.302 0.00 2.302
80 2.3 100.0 80 2.3 100.0

Std2 8 Std2 8
70 70
A 25.00 2.068 A 25.00 2.068
b
s
Measured 2.8 89.8 b
s
Measured 2.8 89.8
60 Std3 8 60 Std3 8
o o
r 50.00 1.751 r 50.00 1.751
b 3.1 76.1 b 3.1 76.1
a 50 a 50
n Std4 8 n Std4 8
c 100.00 1.348 c 100.00 1.348
e 40 3.0 58.6 e 40 3.0 58.6
(%) Std5 8 (%) Std5 8
30 30
250.00 0.719 250.00 0.719
3.4 31.2 3.4 31.2
20 20
contentration Std6 8 contentration Std6 8
750.00 0.306 750.00 0.306
10
obtained 4.2 13.3
10
obtained 4.2 13.3
through through
0 0
25.00 50.00 100.00 250.00 750.00 25.00 50.00 100.00 250.00 750.00

Concentration (ppt) Concentration (ppt)

Fig. 22: A milk sample was= 132.2


50% inhibition diluted 1:4 before applying to the Fig. 23: A meat sample was
50% inhibition concentrated by factor 2 during
= 132.2

microtiter well. The results have to multiplied by 4 to obtain sample preparation 1. The result has to be multiplied with 0.5
the correct concentration: The concentrationLot15313
Microwell plate
of
No.
the analyte
Expiry
2015-04
in to obtain the correct concentration:
Microwell plate
The concentration
Lot No.
15313
of the
Expiry
2015-04
the sample is 100 ng/kgStandards
x 4 = 400 ng/kg. 12034
Conjugate 14064
2015-09
2015-07
analyte in the sample isStandards
100 ng/kg x 0.5 = 5012034
Conjugate
ng/kg. 2015-09
14064 2015-07
Buffer1 12044 2015-12 Buffer1 12044 2015-12
Red Chromogen Pro 11113 2015-05 Red Chromogen Pro 11113 2015-05
Stop solution 11373 2018-08 Stop solution 11373 2018-08
Washing buffer salt 051M8211 2016-06 Washing buffer salt 051M8211 2016-06

Please note: Please note:

The absorbance for the standards may decrease during the shelf life of the kit. The general shape of The absorbance for the standards may decrease during the shelf life of the kit. The general shape of
the curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet the curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet
8. Indication of instability or deterioration of reagents. 8. Indication of instability or deterioration of reagents.

sign.: Edda Rohm Date: 2014-02-17


sign.: Edda Rohm Date: 2014-02-17
Quality Assurance Representative Quality Assurance Representative

Remark: This document has been created electronically and is therefore valid without a signature. Remark: This document has been created electronically and is therefore valid without a signature.

The R-Biopharm group is DIN EN ISO 9001 certified. The R-Biopharm group is DIN EN ISO 9001 certified.
www.r-biopharm.com www.r-biopharm.com
Good ELISA Practice Manual 28

3.8 Units and dimensions


Concentrations are sometimes not expressed in SI SI-units decimal parts per symbol
units, but in miscellaneous dimensionless quantity
g/kg 10-3 parts per mille
part per annotation (Figure 24). We recommend
using SI units. mg/kg 10-6 parts per million ppm

g/kg 10-9 parts per billion ppb

ng/kg 10-12 parts per trillion ppt

pg/kg 10-15 parts per quadrillion ppq

Fig. 24: Overview about SI-unit and the miscellaneous


dimensionless quantities part per annotation

3.9 Limit of detection and quantification


The Limit of Detection (LOD) is the lowest Results below the LOD indicate that a sample is
concentration of an analyte which can be clearly negative or that the concentration of the analyte(s)
distinguished from blank sample readings. The is below the LOD.
LOD is determined experimentally by measuring
the concentration of at least 20 blank matrix Results, which are above the LOD and below the
samples and then calculated by the formula: Mean LOQ are qualitative (negative/positive) results only.
concentration of blank samples + 3-fold standard This means, that the sample contains the analyte,
deviation of the concentrations of blank samples. but the exact amount cant be quantified, as value
is below the LOQ.
The Limit of Quantification (LOQ) is the lowest
concentration of an analyte which can be
detected quantitatively. The LOQ is determined 80
70 LOQ; 69

experimentally by measuring the concentration 60


concentration

of at least 20 blank matrix samples and then 50

calculated by the formula: Mean concentration of 40


LOD; 53
30
blank samples + (most often but not necessarily 20 MeanBlank; 18

9-fold standard deviation of the concentrations of 10

blank samples (Figure 25). 0


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
samples

It is recommended to verify these values by Fig. 25: 20 blank samples were analyzed. Mean of blank
samples = 157 ng/kg with a standard deviation of 92,4 ng/kg
spiking experiment with your sample matrix and LOD = 157 ng/kg + 3 x 92,4 ng/kg = 434 ng/kg
laboratory equipment. LOQ =157 ng/kg + 9 x 92,4 ng/kg = 711 ng/kg
29

3.10 Trueness and recovery


Trueness means the closeness of agreement If samples were measured in areas where the
between the average value obtained from a large standard curve is relatively flat, the accuracy
series of test results and an accepted reference of the results mayQ U A be
L I Tlow
Y A especially
S S U R A N Cat E high
CERTIFICATE

value. Trueness can only be established by means concentrations, because


small differences in Fast
OD or
RIDASCREEN Clenbuterol
of certified reference material (CRM). Trueness is B/B0 may cause Art.large differences
No.: R1701C Lot: in concentration
16213 Expiry: 2015-04
calculated by dividing the measured concentration (Figure 26).
R-Biopharm AG, Darmstadt, Germany certifies that this batch has been approved by the Quality
Assurance Department and conforms with specifications
of the CRM by the assigned concentration of the
CRM. The result is multiplied by 100 to obtain a Standards
Standard curve
percentage unit. The absolute difference between 100 Std.
Conc.(ppt)
n
mean
CV(%) B/B0

both values is called bias. 90

Std1 8
A 80 0.00 1.642
b 6.5 100.0

If no CRM is available, e.g. for antibiotics in food


s 70 Std2 8
o 100.00 1.327
2.0 80.8
r
of animal origin, instead of trueness, the recovery b
60
Std3 8
a 300.00 0.964
can be determined. Recovery is determined n
50 1.7 58.7

c Std4 8
experimentally by measuring samples which are e
40
900.00
3.2
0.414
25.2

spiked prior to sample preparation. Recovery is


30
Std5 8
(%)
2700.00 0.173
20 5.3 10.5
calculated by dividing the measured concentration Std6 8
10 8100.00 0.098
of the spiked sample by the spiking concentration. 5.1 6.0

0
The result is multiplied by 100 to obtain a 100.00 300.00 900.00 2700.00 8100.00

Concentration (ppt)
percentage unit. Please note, that spiking on
Fig. 26: In flat areas of the standard curve small differences
the surface of a matrix may be not identical to in B/B0 produces large differences in concentration: 5 % B/B0
50% inhibition = 399.1

naturally contaminated samples, which can lead to gives an 3 times higher results as 10 % B/B0.
Lot No. Expiry
deviations in results between spiked and naturally MTP K
Standards
14113
11193
2015-08
2015-04
Conjugate 14193 2015-04
incurred samples. Antibody 16193 2015-04
Buffer1 12173 2015-04
Red Chromogen Pro 11113 2015-05
Stop solution 11163 2018-03

Recovery or trueness is given in the instructions


Please note:
for use and represents the mean of the measured
The absorbance for the standards may decrease during the shelf life of the kit. The general shape of the
samples The recovery of single samples can vary in curve will remain similar, while the slope might change slightly. Furthermore refer to product leaflet 8.
Indication of instability or deterioration of reagents.
accordance to the precision of the assay. sign.: N. Stork-Heininger Date: 2014-0
Quality Assurance Representative

Remark: This document has been created electronically and is therefore valid without a signature.

The R-Biopharm group is DIN EN ISO 9001 certified.


www.r-biopharm.c
Good ELISA Practice Manual 30

3.11 Specificity and cross reactivity


Antibodies are highly specific to a single analyte, the analyte or cross reactive substance in a suitable
but sometimes they can also bind other molecules concentration series (Figure 27). After calculation
with different affinities. of the 50 %-dose of the analyte or cross reactive
substance, the specificity or cross reactivity is
The specificity or cross reactivity is determined by calculated as follows:
the measurement of a standard curve consisting of

50 %dose of analyte or cross reactive substance


Specificity or cross reactivity = x 100 %
50 %dose of standard substance

If the binding of the antibody affects a substance


100 standard
that is included in the scope of the method, this is 90
substance

called specificity of the method. If this binding is 80


cross
related to unwanted substances that are not within 70 reactive
substance
60
the scope of the method it is called cross reactivity.
B/B0 [%]

50

40

1st Example: The scope of an ELISA describes that 30

20
the system was developed to quantify aflatoxin
10
M1 in milk. Therefore, the specificity for aflatoxin 0
20 200
M1 is 100 %. Furthermore it is stated that a cross- concentration of standards [g/kg]

reactivity of 10 % to aflatoxin M2 exists.


Fig. 27: The 50%-dose of the standard curve of the standard
substance (blue) is more sensitive (170 g/kg) as the 50%-dose
of the standard curve of the cross reactive substance (210 g/kg).
Due to the principle of an ELISA system, the
The cross reactivity of the cross reactive substance is therefore:
antibody is not able to discriminate between the (170 g/kg / 210 g/kg) x 100 % = 81 %
specificity to aflatoxin M1 and the cross-reactivity
related to aflatoxin M2. If both are present in an quantification is only possible if only one analyte
unknown sample, the result is the sum of both or one cross-reacting substance is present in the
substances. But, due to the low cross-reactivity of sample.
10 %, a 10-fold higher concentration of aflatoxin
M2 (compared to M1) is necessary to resulting In this case, the concentration can be calculated as
in a signal comparable to aflatoxin M1. An exact follows:

measured concentration
Concentration of analyte or cross reactive substance = specificity or cross reactivity of substance in %
100 %
31

2nd example: The scope of a Tetracyclin ELISA higher than specified threshold should be verified
describes that the assay is specific for the by a confirmatory method e.g. LC-MS/MS.
determination of tetracycline, chlortetracycline,
rolitetracycline and demeclocycline in different In case of an unknown matrix and/or another
matrices. The calibrator material is tetracycline specific analyte which are not included in the
with a specificity of 100 %. The analysis of scope of the method, the user must determine the
a chlortetracycline containing milk sample Limit of Detection and the Recovery of the specific
resulted in a concentration of 10 g/L. Since the analyte in the particular sample matrix. Please note
specificity of chlortetracycline is 70 % in this that the specificities and cross reactivitys were
system, the concentration is calculated to 14 g experimentally determined in the buffer system
chlortetracycline per L of milk. only, as it is very time consuming and laborious
to determine every specificity or cross reactivity
Nevertheless, ELISA test are used as screening of any analyte or cross reactive substance in every
methods. Any positive results or concentration matrix.

3.12 Interferences and matrix effects


In general, food and feed are considered as Consequently, it cannot be excluded that natural
highly difficult sample matrices, due to their components or any contaminant of a sample
complex composition. For example, besides its interferes with the reagents used during sample
natural ingredients, (carbohydrates, proteins and preparation or in the ELISA test and cause so called
water) honey may contain lots of environmental matrix-effects.
contaminants such as heavy metals, radioactive
isotopes, organic pollutants, pesticides,
pathogenic bacteria and their toxins and
genetically modified organisms.
R-Biopharm AG

R-Biopharm AG
An der neuen Bergstrae 17
64297 Darmstadt, Germany
Phone: +49 (0) 61 51 - 81 02-0
Fax: +49 (0) 61 51 - 81 02-40
E-mail: info@r-biopharm.de
www.r-biopharm.com
07/2015

S-ar putea să vă placă și