Sunteți pe pagina 1din 11

Dry Powder Inhaler Device Influence on Carrier Particle

Performance
MARTIN J. DONOVAN,1 SIN HYEN KIM,2 VENKATRAMANAN RAMAN,2 HUGH D. SMYTH1
1
Division of Pharmaceutics, College of Pharmacy, University of Texas at Austin, Austin, Texas 78712
2
Department of Aerospace Engineering and Engineering Mechanics, University of Texas at Austin, Austin, Texas 78712

Received 3 February 2011; revised 27 August 2011; accepted 28 October 2011


Published online 17 November 2011 in Wiley Online Library (wileyonlinelibrary.com). DOI 10.1002/jps.22824

ABSTRACT: Dry powder inhalers (DPIs) are distinguished from one another by their unique
device geometries, reflecting their distinct drug detachment mechanisms, which can be broadly
classified into either aerodynamic or mechanical-based detachment forces. Accordingly, pow-
der particles experience different aerodynamic and mechanical forces depending on the in-
haler. However, the influence of carrier particle physical properties on the performance of DPIs
with different dispersion mechanisms remains largely unexplored. Carrier particle trajectories
through two commercial DPIs were modeled with computational fluid dynamics (CFD) and
the results were compared with in vitro aerosol studies to assess the role of carrier particle
size and shape on inhaler performance. Two percent (w/w) binary blends of budesonide with
anhydrous and granulated lactose carriers ranging up to 300 :m were dispersed from both an
AerolizerR
and Handihaler R
through a cascade impactor at 60 L min1 . For the simulations, car-
rier particles were modeled as spherical monodisperse populations with small (32 :m), medium
(108 :m), and large (275 :m) particle diameters. CFD simulations revealed the average
number of carrier particleinhaler collisions increased with carrier particle size (2.34.0) in
the Aerolizer R
, reflecting the improved performance observed in vitro. Collisions within the

R
Handihaler , in contrast, were less frequent and generally independent of carrier particle size.
The results demonstrate that the aerodynamic behavior of carrier particles varies markedly
with both their physical properties and the inhalation device, significantly influencing the per-
formance of a dry powder inhaler formulation. 2011 Wiley Periodicals, Inc. and the American
Pharmacists Association J Pharm Sci 101:10971107, 2012
Keywords: excipients; physical characterization; pulmonary drug delivery; formulation;
morphology; particle size

INTRODUCTION perse the powdered formulation, and have evolved


considerably from their inception, incorporating a di-
Dry powder inhaler (DPI) formulations are typically
verse array of geometries, including vortex generators
binary mixtures comprising micronized active agent
and cyclones, to induce turbulence and particlein-
blended with a coarse carrier particle population that
haler collisions to maximize the detachment potential
constitutes the bulk of the formulation (>95%, w/w).1
of the available flow stream.4,5
When a patient inhales through a DPI, the energy de-
The use of computational fluid dynamics (CFD) to
rived from their inspiratory effort fluidizes the pow-
model the inhalation flow stream, pressure profiles,
dered dose, detaching a fraction of the drug from the
and particle trajectories within therapeutic inhalers
larger carrier particles and depositing it in the deep
has become increasingly prevalent, finding employ-
lung where it exerts its intended therapeutic effect.2
ment in the optimization of device parameters dur-
It is well recognized that DPIs possess distinct mecha-
ing development or, ex post facto, to elucidate the
nisms of powder dispersion.3 Passive DPIs rely solely
deagglomeration principle of a commercially avail-
on the patients inspiratory effort to fluidize and dis-
able inhaler.611 By coupling CFD simulations with
in vitro aerosol performance, information on the rel-
Correspondence to: Martin J. Donovan (Telephone: + 512-471-
4752; Fax: +512-471-7182; E-mail: mjdonovan@utexas.edu)
ative contributions of device and formulation can be
Journal of Pharmaceutical Sciences, Vol. 101, 10971107 (2012)
obtained. A notable example of this approach is a se-
2011 Wiley Periodicals, Inc. and the American Pharmacists Association ries of studies by Coates and colleagues610 on the

JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012 1097


1098 DONOVAN ET AL.

Aerolizer R
(Plastiape S.p.A., Italy) DPI, where the de- ceived. Analytical grade ethanol was supplied by
vice geometry (e.g., air inlet area, mouthpiece length) Sigma Chemical Company (St. Louis, MO). Sam-
and operating parameters (volumetric flow rate, cap- ples of anhydrous (SuperTab R
22AN), and granu-

R
sule size) were varied to assess their influence on lated (SuperTab 30GR) lactose were obtained from
in vitro aerosol performance. DFE Pharma (formerly DMV-Fonterra, Princeton,
Drug detachment from carrier particles is believed NJ). Size three gelatin capsules were provided by
to proceed through two primary mechanisms: (1) aero- CapsugelR
(Greenwood, SC).
dynamic, or fluid-based, detachment arising from the
Fractionation of Lactose Carrier Particles
direct interaction of the flow stream with drug parti-
cles located on the carrier surface; and (2) mechani- Samples of each lactose batch were fractionated on an
cal, or impaction-based, detachment due to collisions Autosiever vibrating sieve shaker (Gilson Company
between carrier particles and the inhaler walls dur- Inc., Lewic Center, OH) with a sieve intensity, or am-
ing particle transit through the device.3,12 Although plitude, setting of 40 for 5 min through the following
the diverse internal geometries of inhalers will vary sieves: 300, 250, 212, 180, 150, 125, 90, 75, 63, 45, and
the relative contributions between aerodynamic and 32 :m. Following the initial fractionation, the lactose
mechanical detachment forces, it has been reported carriers were again sieved to obtain narrow particle
that the magnitude of impact-based events may ex- size distributions.
ceed those of flow-based events, and could potentially
Preparation of Budesonide/Lactose Binary Blends
be the dominating factor in drug detachment from
carrier particles.3,1315 Budesonide and lactose were mixed in a ratio of
Previous research in our laboratory has demon- 1:50 (w/w) via geometric dilution to obtain 500 mg
strated that large lactose carrier particles (>180 :m) of a 2% binary blend. The formulations were blended
can improve drug deposition in vitro for both an- with a Turbula R
orbital mixer (Glen Mills, Clifton,
hydrous and granular lactose populations, although New Jersey) for 40 min at 46 rpm. Samples were
the improvement with carrier particle size was more stored in a desiccator at least 5 days prior to use.
pronounced in formulations with granulated carrier Blend uniformity was determined by randomly se-
particles.16 It is speculated that the surface rough- lecting eight 20-mg samples from each mixture, and
ness of the granulated particles shifted the drug de- assessing the drug content in the powder. Formula-
tachment mechanism from flow-based detachment for tions were considered well blended and ready for use
carrier particles with minimal surface roughness, to if the coefficient of variation (% CV) between the sam-
impaction-based detachment forces as carrier parti- ples for a given blend was below 5%.
cle size and roughness increased. The aerosol perfor-
Scanning Electron Microscopy
mance improvement was thought to result from an
increased incidence of carrier particleinhaler colli- The size and morphology of carrier particles were
sions, given the higher inertia of larger carrier par- visually assessed by scanning electron microscopy
ticles, which inhibit their ability to travel with the (SEM; Supra 40VP, Zeiss, Germany). Prior to SEM,
flow stream through the inhaler. However, it is recog- approximately 20 nm of platinum was deposited onto
nized that the aerodynamic behavior of carrier par- the particle surfaces via sputter coating.
ticles can vary significantly based on the inhaler
Surface Area Analysis
through which a formulation is dispersed, and im-
proved performance may be observed if the carrier Specific surface areas of the lactose carrier parti-
particle design matches the predominant mecha- cle populations were determined via nitrogen ad-
nism(s) of detachment induced by specific inhaler sorption with a single-point BET method using a
types. Accordingly, the specific aim of this study was Monosorb R
surface area analyzer (Quantachrome,
to assess the role of the DPI design on aerosol per- Boynton Beach, FL).
formance as the size and shape of the carrier particle
Density of Lactose Carriers
population are varied. CFD simulations of carrier par-
ticle trajectories were coupled with in vitro drug de- The true densities of the lactose carrier particles were
position studies to investigate the combined influence determined with a helium multipycnometer (Quan-
of device and carrier on formulation performance. tachrome; Boynton Beach, FL).
In Vitro Drug Deposition
MATERIALS AND METHODS About 20 (1) mg of powder was loaded into size 3
gelatin capsules and dispersed through an Aerolizer
R
Materials 
R
(Plastiape S.p.A.) and Handihaler (Boehringer
Micronized budesonide (EP) was purchased from Ingelheim Inc., Ridgefield, CT) DPI into a next gen-
Spectrum Chemicals (Gardena, CA) and used as re- eration cascade impactor (NGI; MSP Corporation,

JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012 DOI 10.1002/jps
DPI INFLUENCE ON CARRIER PARTICLE PERFORMANCE 1099

Shoreview, MN) at a volumetric flow rate of 60 L assumed to be 10% for all the cases. The flow equa-
min1 actuated for 4-s intervals. To prevent parti- tions were discretized using a first-order numerical
cle re-entrainment, the NGI stages were precoated scheme. By using a large number of computational
with a 1% (w/v) solution of silicon oil in hexane. Prior volumes, the dissipative errors associated with a first-
to each actuation, 15 mL of EtOH was added to the order scheme were minimized. Extensive grid conver-
preseparator and collected following powder disper- gence studies demonstrated that the mesh used here
sion from each capsule. Drug depositing in the cap- provides mesh-converged results. The RANS equa-
sule, inhaler, mouthpiece adaptor, and induction port tions were solved until a steady state solution was
were collected by rinsing each component with 10 mL reached. After this step, discrete particles correspond-
of EtOH, whereas the NGI stages were each rinsed ing to the drug carrier particles were introduced in
with 5 mL. Drug content was assessed via ultraviolet- the flow. The particles were evolved using a Stoke-
visible absorption spectroscopy at 244 nm. The emit- sian drag law, with nonspherical corrections imposed
ted fraction (EF) was calculated as the ratio of the to account for particle shape effects [Fluent Inc. (2006)
drug mass depositing in the mouthpiece, induction Centerra Resource Park, Lebanon, New Hampshire).
port, preseparator, and impactor stages over the cu- The flow rate is high enough for gravitational forces
mulative mass of drug collected following actuation to be negligible. The initial locations of the particles
(total drug deposited in the capsule, inhaler, mouth- were different for the two configurations (as shown
piece, induction port, preseparator, and stages). The in Fig. 2). These locations were based on the assump-
fine particle fraction (FPF) of each dose was the ra- tion that the capsule was perforated from one side,
tio of the drug mass depositing on stages 3 through 8 and the particles leave the capsule with identical ve-
(corresponding to an aerodynamic diameter less than locities. Because the mass loading of the particles in
4.46 :m) of the impactor over the emitted dose. The relation to the local fluid mass is very small, the effect
respirable fraction (RF) was the ratio of the drug mass of the particles on the flow field is neglected. Three
deposited on stages 38 over the entire dose recovered different particle diameters were simulated for each
following each actuation. inhaler. The particles were assumed to be a monodis-
perse population of uniform spheres with diameters
of 32, 108, and 275 :m. In each simulation, roughly
Computational Fluid Dynamics
1000 particles were initiated with zero velocity. Be-
Computational fluid dynamics analysis was employed cause the mass loading is small, collisions amongst
to study the flow field and assess carrier particle tra- the particles were neglected. The number of colli-
jectories within the DPIs during actuation. The geom- sions that each particle experienced with different
etry of the AerolizerR
was generously provided as a sections of the DPIs was tracked as the particles trav-
CAD file by the manufacturer (Plastiape S.p.A.); the eled through the inhaler. In order to ensure statistical
computational mesh consisted of approximately 3.7 convergence of the results, different particle numbers
million unstructured computational volumes. By com- were considered. It was found that using 500 parti-
parison, the Handihaler R
is composed of relatively cles was sufficient to obtain converged statistics for
simpler internal geometric features, which were mod- the average number of collisions that the particles
eled from measurements obtained with a caliper. The undergo with the inhaler walls.
Handihaler R
geometry was modeled using 3 million
unstructured computational volumes. The computa-
tional mesh, in each case, was clustered near the walls Statistics
to provide better resolution of the near-wall turbu- Statistical significance between performance values
lence. For instance, the y+ value, which is a normal- was determined with one-way analysis of variance
ized distance used to characterize the near-wall reso- with post hoc tests between groups according to the
lution with respect to the boundary layer, was roughly Bonferroni method (P < 0.05).
15. For both the configurations, the incoming mass
flow rate was set at 60 L min1 . A size 3 capsule was
placed in the capsule chamber. The flow inside the in- RESULTS AND DISCUSSION
haler is turbulent, composed of a range of length and
Physical Characterization of Carrier Particle Populations
time scales. To ensure computational tractability, this
turbulent flow field is modeled using the Reynolds- The double-sieving technique yielded narrow par-
averaged Navier Stokes (RANS) approach.17 In this ticle size distributions (Fig. 1). Although "-lactose
work, the commercial solver FLUENT R
(ANSYS, Inc., monohydrate is typically used as the carrier particle
Canonsburg, Pennsylvania) is used to solve the RANS population in binary dry powder formulations, other
equations. The turbulence properties of the flow field grades including anhydrous and granulated lactose
are described using the shear-stress transport- based have been studied, affording the opportunity to as-
k-T model. The turbulence intensity at the inflow was sess the influence of carrier morphology on aerosol

DOI 10.1002/jps JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012
1100 DONOVAN ET AL.

Figure 1. SEM micrographs of uncoated (a) 3245 :m anhydrous lactose, (b) 90125 :m
anhydrous lactose, (c) 250300 :m anhydrous lactose, (d) 3245 :m granulated lactose,
(e) 90125 :m granulated lactose, and (f) 250300 :m granulated lactose sieve fractions. Scale
bars denote 200 :m.

performance.16,1820 The disparity in surface rough- The resulting turbulent flow field induces the capsule
ness between the anhydrous and granulated lactose to rotate and rattle with high frequency during in-
particles increases with particle size, as evidenced by halation, assisting in ejecting and dispersing the pow-
the diminishing specific surface areas of the anhy- dered dose through the perforations in the capsule
drous particles relative to their granulated counter- wall.7 It is noted that although the motion of the cap-
parts with increasing carrier sieve fraction (Table 1). sule itself is not simulated in this work, it may affect
Surface areas of larger granulated lactose did not di- particle release and transport in the fluid flow, and it
minish appreciably as particle sizes were increased has been speculated that the emitted powder parti-
(as would be expected for perfectly smooth spherical cles may collide with the capsule, though a previous
particles). Instead, the increasing roughness (due to investigation concluded that impactions between the
granulation) causes a relatively minimal decline in powder particles and the spinning capsule do not sig-
surface area as the size of the particles within the nificantly contribute to dispersion performance in the
powder is increased. The true densities of all carrier Aerolizer R 7
. However, those studies were performed
particle size fractions in both grades of lactose ranged on pure spray-dried mannitol, and the inclusion of
between 1.54 and 1.58 g/cm3 . carrier particles may potentially alter this result and
is a topic that merits future study.
CFD Analysis
Internal geometries of the Handihaler R
and
R
Aerolizer are illustrated in Figure 2. The flow
stream enters the Aerolizer R
via two tangential in-
lets found on opposite sides of the capsule chamber.9

Table 1. Specific Surface Areas (SSA) of Anhydrous and


Granulated Lactose Carrier Particles by Sieve Fraction

Carrier Sieve Anhydrous Granulated


Fraction (:m) (m2 g1 ) (m2 g1 )
<32 0.94 1.87
3245 0.48 1.71
4563 0.38 1.62
6375 0.43 1.64
7590 0.37 1.52
90125 0.41 1.56
125150 0.38 1.81
150180 0.38 1.73
180212 0.40 1.44 Figure 2. Schematic views of the Handihaler R
(left) and

R
212250 0.42 1.67 Aerolizer dry powder inhalers. The bold arrow denotes the
250300 0.36 1.46 particle launch location for each inhaler.

JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012 DOI 10.1002/jps
DPI INFLUENCE ON CARRIER PARTICLE PERFORMANCE 1101

A key feature of the internal geometry of the


Aerolizer R
is a grid dividing the capsule chamber and
the mouthpiece to prevent capsule egress from the de-
vice during inhalation. In addition, it is noted that the
grid has been demonstrated to straighten the flow and
suppress turbulence downstream of its location.6 On
the contrary, the mechanism of the Handihaler R
is
based on the sudden expansion of the inhalation flow
stream as it passes from the inlet into the larger cap-
sule chamber. The abrupt opening from the smaller
inlet chamber to the capsule chamber causes the
boundary layer of the flow stream to separate at the
corner of the expansion, creating an annular region
where a portion of the flow recirculates, resulting in
a pressure loss in this region.21 The incoming flow
stream pushes the capsule toward the grid while the
low pressure region simultaneously attracts the cap-
sule toward the inlet, causing the capsule to spin and
vibrate in the chamber.11 Similar to the Aerolizer R
,

R
the Handihaler also contains a mesh between the
capsule chamber and the mouthpiece. However, this
grid is made of thinner wires and its effect on the flow Figure 3. Contour of velocity magnitude inside
field is comparably less significant than the grid lo- Handihaler
R
(left) and Aerolizer
R
(right).
cated in the Aerolizer R
, as the blockage represented
by the wires is a very small fraction of the total sur-
face area. (increasing with carrier particle size), once they are
Figure 3 depicts the velocity magnitude of the launched on this trajectory the fluid flow is unable to
flow stream in the inhalers at 60 L min1 . In the significantly alter the particle trajectory. The initial
Handihaler R
, the maximum flow speed is found up- acceleration around the capsule introduces the colli-
stream of the capsule with very high shear flow sional movement, pushing the particles downstream
around the capsule itself. In contrast, the Aerolizer R toward the inhaler exit.
exhibits a much more complex flow pattern, as the Table 2 shows the influence of particle diameter
incoming flow generates a swirling motion inside the on the number of collisions with the mouthpiece wall
capsule chamber. The grid between the capsule cham- experienced by the particles. It should be noted that
ber and the mouthpiece constricts the flow, and con- the fluid flow corresponds to particles with zero diam-
sequently the velocity magnitude is very high when eter, or zero Stokes number. Increasing the particle
the fluid exits the capsule chamber through the grid diameter increases the Stokes number and causes a
spacing. The grid also regulates the swirling flow gen- departure of the particle trajectory from the fluid tra-
erated by the capsule chamber.6 jectory. As the particles become larger, their response
Carrier particle trajectories were simulated to changes in the fluid flow is slower. Consequently,
through the DPIs to quantify the frequency of their larger particles can possess ballistic trajectories and
collisions with the internal geometry of the inhaler undergo more collisions, and increasing the particle
(Table 2). It was expected that the distinct internal diameter increases the number of collisions. It is seen
geometries of the inhalers would yield distinct par- that the Aerolizer R
exhibits greater sensitivity to par-
ticle trajectories. As seen in Figure 4, the fluid flow ticle size relative to the Handihaler R
. This is primar-
inside the Aerolizer R
actively promotes particle colli- ily attributed to the fact that the Handihaler R
does
sions with the inhaler wall. The tangential air inlets not introduce a swirling motion inside the mouth-
introduce a swirl component to the flow stream that piece, which causes the particles to have a longer res-
is imparted to the powder as they exit the capsule, idence time (or trajectory length) inside the mouth-
causing particles to swirl through the mouthpiece. piece.
In contrast to the Aerolizer R
, carrier particles
R
within the Handihaler experience markedly fewer
In Vitro Aerosol Performance
collisions with the inhaler (Fig. 5). The wake region
downstream of the cavity causes the particles to ac- In vitro drug dispersion profiles for all experimen-
celerate and travel at an angle to the vertical direc- tal formulations are presented in Tables 3 and 4.
tion, directing the particle trajectory toward the in- The percentage of the nominal dose emitted from
haler wall. As the particles have significant inertia the Handihaler R
exceeded that emitted from the

DOI 10.1002/jps JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012
1102 DONOVAN ET AL.

Table 2. Average Number of Collisions Between a Carrier Particle and the Inhaler As It Exits the
Device During Actuation from the Aerolizer
R
and Handihaler
R
at 60 L min1

Aerolizer
R
Handihaler
R

Carrier Particle Sieve Fraction (:m) 32 (:m) 107.5 (:m) 275 (:m) 32 (:m) 107.5 (:m) 275 (:m)
Inhalercarrier collisions 2.3 3.6 4.0 1.7 2.2 2.3

cating that although for anhydrous formulations the


Handihaler R
is more effective at emitting drug from
the device, the Aerolizer R
demonstrates a greater ten-
dency for detaching drug from the carrier particles.
In contrast, granulated carrier particle formulations
dispersed from the Handihaler R
outperformed those
from the Aerolizer for all but the largest carrier par-
R

ticle size fractions.


From the drug dispersion profiles, it is observed
that the increased number of collisions between the
carrier particles and inhaler does not directly trans-
late into improved aerosol performance, as measured
by RF. For a given dry powder formulation, RF val-
ues reflect the combined EF and FPF, revealing an
Figure 4. Carrier particle trajectory inside the Aerolizer
R
inhalers potential to both emit the drug from the de-
1
at 60 L min (From left, dparticle = 32, 108, and 275 :m).
vice and detach it from the carrier, such that it can
travel to the deep lung.18 By itself, EF is a metric
Aerolizer R
, yielding over 75% EF for all carrier parti- describing the inhalers ability to emit the drug from
cle sizes between both lactose grades. For anhydrous the device and deliver it to the patient, but reveals
lactose formulations dispersed from the Handihaler R
, nothing as to whether the drug is detached from the
the enhanced EF did not correspond to improved carrier, and drug that remains adhered to the coarse
drug deposition compared with the Aerolizer R
, as carrier particles will deposit in the mouth, throat and
RF values were generally comparable between de- upper airways. Accordingly, FPF is a better measure
vices across most carrier particle size fractions. The of an inhalers drug detachment potential, providing
disparity between EF and RF is reconciled by the the fraction of detached drug exiting the device rela-
higher FPF values obtained from the Aerolizer R
, indi- tive to the total amount that leaves the inhaler. A high
EF coupled with a high FPF will generate excellent
RF values, but opposing EF and FPF performance
(e.g., high EF and low FPF) will only yield moderate
respirable fractions.
Although significant differences in emitted dose
were observed between the devices, this did not
yield a consistent disparity in overall performance,
as the relatively high EF from Handihaler R
was off-

R
set by the Aerolizer s enhanced tendency to emit
drug particles detached from the carriers. To pro-
vide an approximate measure of the relative drug
detachment potential of the DPIs, the ratio of FPF
values from the Aerolizer R
over the Handihaler R

(FPFAerolizer /FPFHandihaler ) against the carrier particle


size fraction were plotted (Fig. 6). Values >1 indi-
cate the Aerolizer R
may be more effective at drug
detachment while a ratio less than unity gives the
advantage to the Handihaler R
. It must be noted
that detached drug particles may be agglomerates
of primary drug particles, which due to their large
Figure 5. Carrier particle trajectory inside the size will deposit in the throat and conducting air-
HandihalerR
at 60 L min1 (From left, dparticle = 32, ways. This is especially relevant for the corticosteroid
108, and 275 :m). budesonide, which has been demonstrated to be a

JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012 DOI 10.1002/jps
DPI INFLUENCE ON CARRIER PARTICLE PERFORMANCE 1103

Table 3. Emitted Fraction (EF), Fine Particle Fraction (FPF), and Respirable Fraction (RF) for 2% (w/w) Budesonide Formulations with
Anhydrous Lactose Carrier Particles Characterized In Vitro from the Aerolizer
R
and Handihaler R
Dry Powder Inhalers at 60 L min1

Aerolizer
R
Handihaler
R

Carrier Particle Sieve Fraction (:m) EF FPF RF EF FPF RF


<32 63.5 (1.6) 29.1 (1.4) 18.4 (0.7) 86.4 (1.1) 18.5 (0.7) 15.9 (0.4)
3245 64.5 (1.8) 21.3 (0.8) 13.7 (0.3) 79.0 (8.0) 15.5 (0.7) 12.2 (1.1)
4563 63.1 (2.2) 23.1 (1.8) 14.6 (1.4) 76.3 (1.2) 17.1 (1.2) 13.0 (0.9)
6375 68.9 (1.8) 15.1 (0.4) 10.4 (0.3) 79.2 (2.5) 13.7 (0.3) 10.8 (0.2)
7590 71.8 (2.4) 15.2 (0.3) 10.9 (0.5) 77.1 (1.3) 14.4 (0.7) 11.1 (0.4)
90125 62.3 (1.9) 13.7 (0.6) 8.5 (0.4) 80.0 (8.1) 15.9 (0.8) 12.4 (1.2)
125150 62.7 (2.1) 15.6 (1.4) 9.8 (1.2) 81.1 (4.7) 9.9 (0.5) 8.0 (0.8)
150180 62.3 (2.0) 12.1 (0.8) 7.5 (0.3) 82.3 (1.9) 9.5 (1.6) 7.8 (1.3)
180212 61.8 (1.9) 14.1 (1.1) 8.7 (0.5) 79.9 (1.8) 10.1 (0.2) 8.1 (0.2)
212250 63.4 (2.3) 15.5 (0.8) 9.9 (0.6) 82.3 (0.8) 8.7 (1.9) 7.1 (1.5)
250300 61.6 (1.7) 15.0 (1.1) 9.3 (0.9) 80.2 (0.4) 9.8 (0.3) 8.1 (0.6)

Values are given as the mean of N = 3 replicates, and values within parentheses represent the standard deviation for N = 3 replicates.

relatively cohesive drug.22,23 However, as the same Influence of Device on Performance


formulation was used for each size fraction within Because of the marked variation in the internal ge-
a lactose grade, the extent of drug agglomeration ometries of the two inhalers, differences in aerosol
was presumed constant between samples, with the performance were expected between the Aerolizer R

inhaler as the variable. For the anhydrous carrier R


and Handihaler . Regarding device resistance, the
formulations, all but a single size fraction (90125 Aerolizer R
and Handihaler R
reside on opposing ends
:m) exceeded 1 (Fig. 6), indicating the ability of of the spectrum, as the Aerolizer R
has less than half
the Aerolizer R
to emit drug particles detached from the resistance [0.072 (cmH2 O) /(L min1 )] of the
0.5
the carrier outperforms the Handihaler R
. Conversely, Handihaler R
[0.158 (cmH2 O)0.5 /(L min1 )].9,24 This
most granulated carriers yielded better drug detach- increased resistance of the Handihaler R
arises from
ment from the Handihaler R
(ratios were less than 1). the narrow inlet tube at the base of the capsule cham-
However, at carrier particle sizes greater than 7590 ber (Fig. 2).11 It is noted that the flow rate selected
and 90125 :m fractions for the granulated and an- for this study (60 L min1 ) is much higher than that
hydrous carriers, respectively, the FPF ratio began typically generated through the Handihaler R
, but is
to increase, indicating an improvement in drug de- readily attainable by > 90% of adult patients through
tachment potential of the Aerolizer R
relative to the an Aerolizer R 2426
. The relationship between inhaler

R
Handihaler for large carrier particle formulations. It resistance (R), volumetric flow rate (Q), and pressure
is speculated that the higher frequency of carrierin- drop (P) across a device is (P)0.5 = QR.27 For a
haler collisions within the AerolizerR
as carrier size is fixed flow rate, a higher resistance device will pro-
increased may account for the observed improvement duce a greater pressure drop, which has been found
in drug detachment. In these larger carrier formu- to correlate with improved aerosol performance.2830
lations, impaction-based forces may be a significant In addition, the overall energy passing through an
mechanism of drug detachment as discussed below. inhaler can be approximated from the product of the

Table 4. Emitted Fraction (EF), Fine Particle Fraction (FPF), and Respirable Fraction (RF) for 2% (w/w) Budesonide Formulations with
Granulated Lactose Carrier Particles Characterized In Vitro from the Aerolizer
R
and Handihaler R
Dry Powder Inhalers at 60 L min1

Aerolizer
R
Handihaler
R

Carrier Particle Sieve Fraction (:m) EF FPF RF EF FPF RF


<32 57.0 (1.4) 18.5 (1.2) 11.1 (1.6) 76.9 (1.6) 17.1 (1.4) 13.1 (0.8)
3245 69.1 (3.7) 11.1 (0.6) 7.7 (0.1) 82.0 (0.5) 15.4 (1.0) 12.6 (0.8)
4563 74.4 (4.1) 8.9 (1.2) 6.6 (0.6) 82.1 (1.9) 12.8 (0.2) 10.5 (0.3)
6375 73.8 (3.7) 12.6 (1.4) 9.3 (0.6) 78.4 (2.7) 16.2 (0.7) 12.6 (0.2)
7590 73.9 (6.6) 9.7 (2.1) 7.2 (0.8) 78.7 (1.9) 19.4 (2.4) 15.3 (1.6)
90125 77.8 (1.8) 11.6 (0.9) 9.0 (0.5) 86.1 (2.0) 17.9 (0.8) 15.3 (0.4)
125150 71.3 (4.2) 18.9 (1.5) 13.4 (0.3) 84.3 (3.4) 20.5 (2.5) 17.2 (1.7)
150180 73.8 (2.5) 14.2 (1.3) 10.5 (0.7) 87.3 (1.4) 15.2 (1.5) 13.3 (1.1)
180212 72.6 (3.2) 20.8 (1.1) 15.1 (0.8) 87.1 (1.6) 16.1 (1.4) 14.0 (1.1)
212250 68.7 (5.4) 20.2 (1.5) 13.8 (0.7) 84.3 (1.1) 15.4 (2.8) 13.1 (2.3)
250300 68.5 (4.5) 20.1 (1.5) 14.5 (0.5) 85.5 (0.6) 18.4 (1.0) 15.7 (1.0)

Values are given as the mean of N = 3 replicates, and values within parentheses represent the standard deviation for N = 3 replicates.

DOI 10.1002/jps JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012
1104 DONOVAN ET AL.

Figure 6. Ratio of the FPF values obtained from the


Aerolizer
R
over the Handihaler R
for both anhydrous and
granulated lactose formulations.

pressure drop across the device, volumetric flow rate,


and actuation time.31 Thus, although the volumetric
flow rate is the same through both inhalers (60 L
min1 ), a greater pressure drop, and hence a higher
energy value is calculated for the Handihaler R
, which
is predicted to improve performance from this DPI.
However, despite the device differences, RF values
exhibited minimal dependence on the inhaler for
the anhydrous carrier particles, varying instead as
a function of carrier particle size, with larger carrier Figure 7. Respirable fraction (RF) values of (a) anhydrous
particle diameters diminishing overall performance and (b) granulated lactose carrier particles dispersed from
(Fig. 7). By comparison, granulated lactose formu- the Aerolizer
R
and Handihaler R
at 60 L min1 . Values are
lations showed some device dependency, notably in given as the mean of N = 3 replicates, with error bars rep-
the smaller carrier particle size fractions, with both resenting the standard deviation for N = 3 replicates.
inhalers demonstrating comparable performance as
carrier size increased.
dient develops across the capsule chamber, pulling
For anhydrous lactose formulations, the relatively
drug particles primarily from the lower capsule per-
flat surfaces of the carriers permit direct interac-
foration located nearest the corner of the flow stream
tion between the drug particles and inhalation flow
expansion.11 Accordingly, the relatively high energy
stream. The overall dispersion performance between
input through this device at the studied flow rate,
the devices is similar across all anhydrous carrier
coupled with the Handihaler R
s mechanism-of-action,
size fractions, indicating the comparable ability of the
may enable drug within the surface asperities of the
DPIs to detach drug readily accessible to the air flow.
granulated particles to be pulled-off of the carriers,
However, in granulated lactose formulations, where
allowing performance to be relatively independent
direct interaction between the drug and flow stream
of surface roughness, and thus carrier particle size,
is inhibited by the surface roughness of the carri-
given that the surface roughness of granulated carri-
ers, the Handihaler R
exhibited higher RF values up
ers increased with particle diameter.
to the largest size fractions, when it is speculated
that increasingly larger mechanical forces allowed
Interparticle Collisions
the Aerolizer R
to improve its detachment potential
(as indicated by the increasing FPF ratio), and attain In addition to carrier particles colliding with the in-
comparable RF values. The improved performance haler walls, interparticle collisions between carriers
at smaller carrier sizes of granulated lactose in the during actuation may provide an additional source
Handihaler R
may be attributed to the greater energy of mechanical detachment forces. For a fixed mass
input through this device (as mentioned above), which of powder, increasing the diameter of the carrier
has been found to correlate positively with disper- population diminishes the overall number of carrier
sion performance.31 Furthermore, a previous study particles in the dose, such that formulations with
on the Handihaler R
revealed that a pressure gra- the largest lactose sizes contain a fraction of the

JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012 DOI 10.1002/jps
DPI INFLUENCE ON CARRIER PARTICLE PERFORMANCE 1105

number of carrier particles relative to the small-


est carrier size ranges. Previous studies have re-
ported that lactose fluidizes from capsules via two
distinct mechanisms dependent upon the flowability,
and hence the cohesiveness, of the powder.3234 Pow-
ders with good flowability are entrained by way of an
erosion mechanism, where the powder is fluidized
by layers and exits the inhaler gradually as a contin-
uous stream.33,34 By contrast, for poorly flowing pow-
ders fluidization occurs by the fracture mechanism
whereby the powder fluidizes as multiple agglomer-
ates, and in extreme cases may fluidize as a single
aggregated powder plug.
It has been proposed that smaller carrier particle
size fractions, particularly those with a significant
concentration of lactose fines, improve aerosol perfor-
mance by coupling the fracture entrainment mecha-
nism with an extensive number of carrier particles,
thereby generating a high density aerosol cloud that
promotes collisions between lactose carriers.33,34 In
the present study, only anhydrous formulations ex-
hibited performance consistent with this mechanism,
but performance from granulated carriers was incom-
patible with this theory, as the <32 :m carrier popula-
tions did not yield RF values significantly higher than
the larger carriers. However, variables apart from car- Figure 8. Mass median aerodynamic of budesonide par-
rier particle size can affect the extent to which inter- ticles following dispersion of 2% (w/w) formulations with
particle collisions may occur. For instance, the DPI (a) anhydrous lactose, and (b) granulated lactose car-
employed may influence interparticle collisions, as rier particles at 60 L min1 from both the Aerolizer R

devices with more complex internal geometries may and Handihaler R


DPIs. Values are given as the mean of
enhance the frequency and/or magnitude of interpar- N = 3 replicates, with error bars representing the standard
deviation for N = 3 replicates.
ticle collisions relative to inhalers with simpler inter-
nal geometries. In addition, the cross-sectional area
of the capsule perforation may likewise affect inter-
particle collisions, as smaller openings can prevent
the powder from exiting the capsule as agglomerated significantly between inhalers for both anhydrous
plugs, thus limiting the density of the aerosol cloud. and granulated lactose formulations (Fig. 8). Fur-
Thus, it is noted that the inhalers employed in these thermore, although the MMAD from the Aerolizer R

studies may not be ideal for promoting interparticle displayed an overall decreasing trend with increas-
interactions, as the swirling particle trajectories in ing carrier particle size for both lactose populations,
the capsule chamber of the Aerolizer R
may be off- MMAD values obtained from the Handihaler R
were
set by the small cross-sectional area of the capsule generally independent of the formulation. The dis-
perforations, whereas the larger diameter capsule parity between the MMAD values for formulations
openings from the Handihaler R
are coupled with an dispersed through the inhalers were greatest at the
internal geometry that may not induce extensive in- smaller carrier particle formulations, and eventually
terparticle collisions relative to the Aerolizer R
. Ac- converged on a mutual value for the largest carrier
cordingly, although interparticle collisions may influ- size fractions. In our previous study, it was postulated
ence aerosol performance under certain conditions, that the reduction in MMAD obtained from larger car-
this detachment mechanism was not supported by rier particles dispersed through the Aerolizer R
may
the data for both lactose grades, and thus CFD sim- be attributed to the increased detachment of smaller
ulations capturing carriercarrier interactions were drug particles from the carrier surface.16 As the col-
deemed beyond the scope of the present study. lision force between a particle and inhaler is depen-
dent on the mass, and thus the volume, of the carrier
Mass Median Aerodynamic Diameter
particle, the momentum generated from a collision in-
In contrast to the respirable fraction, the mass me- creases with the cube of the carrier particle diameter,
dian aerodynamic diameter (MMAD) of drug deposit- assuming a relatively constant particle velocity (as
ing on the stages of the cascade impactor varies volume diameter3 ).2 Thus, larger carrier particles

DOI 10.1002/jps JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012
1106 DONOVAN ET AL.

can generate strong mechanical detachment forces, rier diameter. In contrast, the internal geometry of
which can potentially exceed fluid-based forces.3,13,14 the Aerolizer R
is capable of generating a higher
These strong mechanical forces may dislodge drug number of particleinhaler collisions, with the fre-
particles resistant to fluid-based detachment, in- quency of the collisions markedly increasing with car-
cluding small aggregates and primary drug parti- rier diameter. Accordingly, aerosol performance from
cles located within carrier particle surface asperities, this DPI exhibited a dependence on the carrier par-
inhibiting their interaction with the flow stream. Ac- ticle size fraction, which was not observed in the
cordingly, the size of drug particles depositing in the Handihaler R
to a similar extent. In addition, per-
impactor would be expected to decrease with larger formance was also significantly influenced by carrier
carrier particle populations, as observed in both the particle morphology, and although granulated lac-
anhydrous and granulated formulations dispersed tose is not commonly employed in DPI formulations,
from the Aerolizer R
. Comparing MMAD values ob- the use of this lactose grade afforded greater insight
tained from the Aerolizer R
with the corresponding into distinctions in DPI performance than relatively
RF values, the carrier particle size fraction when smooth surfaced anhydrous lactose. In conclusion, the
the MMAD markedly declined matched the carrier results of this study suggest that matching the physi-
size when aerosol performance exhibited a local im- cal properties of the carrier population to the predom-
provement; this corresponded to the 180212 :m size inant detachment mechanism of the DPI may signifi-
fraction for anhydrous carriers, and 90125 :m for cantly influence aerosol performance.
granulated lactose (p < 0.05). Conversely, MMAD
values in the Handihaler R
were relatively consis-
tent between carriers for both lactose grades, dis- ACKNOWLEDGMENTS
playing no general trend with performance, although The authors would like to express their gratitude to
the MMAD from formulations dispersed through the Prof. Hak Kim Chan and Mr. Mauro Citterio for their
Handihaler R
was consistently lower than that dis- assistance in providing the Aerolizer
R
CAD file. The
persed from the Aerolizer R
, with the exception of the authors also wish to acknowledge Stephen Marek for
largest carrier sizes (>180 :m). obtaining the SEM images.
In combination with the particle trajectory sim-
ulations, the aerosol performance data suggest the
presence of an impaction-based mechanism of drug REFERENCES
detachment in the Aerolizer R
whereby the relatively 1. Hickey AJEd. 2004. Pharmaceutical inhalation aerosol tech-
numerous and high energy collisions between large nology. 2nd ed. New York City, New York: Marcel Dekker, Inc.
carriers and the inhaler wall generate a force suf- 2. Finlay WH. 2001. The Mechanics of inhaled pharmaceutical
ficient in both magnitude and direction to detach aerosols. London, UK: Academic Press.
smaller drug particles immune to detachment by 3. de Boer A, Hagedoorn P, Gjaltema D, Goede J, Frijlink H.
2003. Air classifier technology (ACT) in dry powder inhalation
the flow stream. As discussed in detail elsewhere, Part 1. Introduction of a novel force distribution concept (FDC)
this favors large carrier particles with extensive sur- explaining the performance of a basic air classifier on adhesive
face roughness, and accordingly the improvement mixtures. Int J Pharm 260(2):187200.
in aerosol performance with larger carrier popu- 4. Frijlink HW, De Boer AH. 2004. Dry powder inhalers for
lations is more pronounced in granulated lactose pulmonary drug delivery. Expert Opin Drug Deliv 1(1):6786.
5. Islam N, Gladki E. 2008. Dry powder inhalers (DPIs)a
formulations.16 However, it is noted that while the review of device reliability and innovation. Int J Pharm
present studies were performed at a fixed flow rate, 360(12):111.
the velocity, and hence momentum, of the carrier par- 6. Coates MS, Fletcher DF, Chan H-K, Raper JA. 2004. Effect of
ticles may significantly affect performance, particu- design on the performance of a dry powder inhaler using com-
larly for larger carrier populations. Accordingly, stud- putational fluid dynamics. Part 1: Grid structure and mouth-
piece length. J Pharm Sci 93(11):28632876.
ies exploring the effectiveness of mechanical detach- 7. Coates MS, Fletcher DF, Chan H-K, Raper JA. 2005. The
ment forces at variable flow rates are presently under role of capsule on the performance of a dry powder inhaler
investigation in our lab. using computational and experimental analyses. Pharm Res
22(6):923932.
8. Coates MS, Chan H-K, Fletcher DF, Raper JA. 2005. Influ-
CONCLUSION ence of air flow on the performance of a dry powder inhaler
using computational and experimental analyses. Pharm Res
Coupling the CFD particle trajectory simulations 22(9):14451453.
with the in vitro results, it is concluded that 9. Coates MS, Chan H-K, Fletcher DF, Raper JA. 2006. Effect
impaction-based forces are not a significant mecha- of design on the performance of a dry powder inhaler using
nism of detachment in the Handihaler R
, as reflected computational fluid dynamics. Part 2: Air inlet size. J Pharm
Sci 95(6):13821392.
by both the absence of improved performance at the 10. Coates MS, Chan H-K, Fletcher DF, Chiou H. 2007. Influence
large carrier fractions and the minimal increase in of mouthpiece geometry on the aerosol delivery performance
simulated carrier particleinhaler collisions with car- of a dry powder inhaler. Pharm Res 24(8):14501456.

JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012 DOI 10.1002/jps
DPI INFLUENCE ON CARRIER PARTICLE PERFORMANCE 1107

11. Wachtel H, Ertunc O, Koksoy C, Delgado A. 2008. Aerody- 24. Alshowair R, Tarsin W, Assi K, Pearson S, Chrystyn H.
namic optimization of Handihaler and respimat: The roles of 2007. Can all patients with COPD use the correct inhala-
computational fluid dynamics and flow visualization. Respir tion flow with all inhalers and does training help? Respir Med
Drug Deliv 1:165174. 101(11):23952401.
12. Voss A, Finlay WH. 2002. Deagglomeration of dry powder phar- 25. Chodosh S, Flanders JS, Kesten S, Serby CW, Hochrainer
maceutical aerosols. Int J Pharm 248:3950. D, Witek TJ. 2001. Effective delivery of particles with the
13. Nichols SC, Wynn E. 2008. New approaches to optimizing dis- HandiHaler R
dry powder inhalation system over a range of
persion in dry powder inhalersDispersion force mapping and chronic obstructive pulmonary disease severity. J Aerosol Med
adhesion measurements. Respir Drug Deliv 1:175184. 14(3):309315.
14. Wong W, Fletcher DF, Traini D, Chan H-K, Crapper J, Young 26. Bronsky EA, Grossman J, Henis MJ, Gallo PP, Yegen U,
PM. 2010. Particle aerosolisation and break-up in dry powder Cioppa GD, Kottakis J, Mehra S. 2004. Inspiratory flow rates
inhalers 1: Evaluation and modelling of Venturi effects for and volumes with the Aerolizer dry powder inhaler in asth-
agglomerated systems. Pharm Res 27(7):13671376. matic children and adults. Curr Med Res Opin 20(2):131
15. de Boer A, Hagedoorn P, Gjaltema D, Goede J, Kussendrager 137.
K, Frijlink H. 2003. Air classifier technology (ACT) in dry 27. Clark AR, Hollingworth AM. 1993. The relationship between
powder inhalation Part 2. The effect of lactose carrier surface powder inhaler resistance and peak inspiratory conditions in
properties on the drug-to-carrier interaction in adhesive mix- healthy volunteersimplications for in vitro testing. J Aerosol
tures for inhalation. Int J Pharm 260(2):201216. Med 6(2):99110.
16. Donovan MJ, Smyth HDC. 2010. Influence of size and sur- 28. Mendes P, Pinto J, Sousa J. 2007. A non-dimensional func-
face roughness of large lactose carrier particles in dry powder tional relationship for the fine particle fraction produced by
inhaler formulations. Int J Pharm 402(12):19. dry powder inhalers. J Aerosol Sci 38(6):612624.
17. Pope SB. 2000. Turbulent flows. Cambridge University Press 29. Srichana T, Martin GP, Marriott C. 1998. Dry powder in-
New York, NY, USA halers: The influence of device resistance and powder formu-
18. Kawashima Y, Serigano T, Hino T, Yamamoto H, Takeuchi lation on drug and lactose deposition in vitro. Eur J Pharm Sci
H. 1998. Effect of surface morphology of carrier lactose on 7(1):7380.
dry powder inhalation property of pranlukast hydrate. Int J 30. Louey M, Vanoort M, Hickey A. 2006. Standardized entrain-
Pharm 172:179188. ment tubes for the evaluation of pharmaceutical dry powder
19. Larhrib H, Zeng XM, Martin GP, Marriott C, Pritchard J. dispersion. J Aerosol Sci 37(11):15201531.
1999. The use of different grades of lactose as a carrier for 31. Chew NYK, Chan HK, Bagster DF, Mukhraiya J. 2002. Char-
aerosolised salbutamol sulphate. Int J Pharm 191:114. acterization of pharmaceutical powder inhalers: Estimation of
20. Pitchayajittipong C, Price R, Shur J, Kaerger JS, Edge S. 2010. energy input for powder dispersion and effect of capsule device
Characterisation and functionality of inhalation anhydrous configuration. Aerosol Sci 33:9991008.
lactose. Int J Pharm 390(2):134141. 32. Tuley R, Shrimpton J, Jones M, Price R, Palmer M, Prime D.
21. Gupta V, Gupta SK. 1984. Fluid mechanics and its applica- 2008. Experimental observations of dry powder inhaler dose
tions. New Delhi, India: New Age International. fluidisation. Int J Pharm 358(12):238247.
22. Begat P, Morton DAV, Staniforth JN, Price R. 2004. Dry pow- 33. Watling CP, Elliott JA, Cameron RE. 2010. Entrainment of
der inhaler formulations I: Direct quantification by atomic lactose inhalation powders: A study using laser diffraction.
force microscopy. Pharm Res 21(9):15911597. Eur J Pharm Sci 40(4):352358.
23. Saleem I, Smyth H, Telko M. 2008. Prediction of dry pow- 34. Shur J, Harris H, Jones MD, Kaerger JS, Price R. 2008. The
der inhaler formulation performance from surface energet- role of fines in the modification of the fluidization and dis-
ics and blending dynamics. Drug Dev Ind Pharm 34(9):1002 persion mechanism within dry powder inhaler formulations.
1010. Pharm Res 25(7):16311640.

DOI 10.1002/jps JOURNAL OF PHARMACEUTICAL SCIENCES, VOL. 101, NO. 3, MARCH 2012

S-ar putea să vă placă și