Sunteți pe pagina 1din 8

See

discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/229218341

Phosphate biosensor based on polyelectrolyte-


stabilized pyruvate oxidase

Article in Analytica Chimica Acta January 2001


DOI: 10.1016/S0003-2670(00)01204-6

CITATIONS READS

50 48

2 authors, including:

Nikos Chaniotakis
University of Crete
109 PUBLICATIONS 3,796 CITATIONS

SEE PROFILE

All content following this page was uploaded by Nikos Chaniotakis on 01 September 2014.

The user has requested enhancement of the downloaded file.


Analytica Chimica Acta 427 (2001) 271277

Phosphate biosensor based on polyelectrolyte-stabilized


pyruvate oxidase
Vasilis G. Gavalas, Nikolas A. Chaniotakis
Laboratory of Analytical Chemistry, Department of Chemistry, University of Crete,
71409 Iraklion, Crete, Greece
Received 27 June 2000; accepted 11 September 2000

Abstract
In this work the development of a pyruvate oxidase-based phosphate biosensor is illustrated. The use of polyelectrolyte
stabilized recombinant pyruvate oxidase in conjunction with a porous conductive carbon results in the development of a
simple, reproducible and stable phosphate biosensor. The polyelectrolyte diethylaminoethyl-dextran or DNA was used as
the enzyme stabilizer, and the resulting enzymepolyelectrolyte complexes were physically adsorbed into the transducer, a
highly porous and conductive carbon electrode, for the construction of the biosensor. The optimized biosensor exhibits high
operational (67% remaining activity after 220 h) and storage (49% remaining activity after 24 weeks) stability, and very good
sensor-to-sensor reproducibility. The optimized phosphate biosensor was used for the measurement of the phosphate ion
activity in serum. 2001 Elsevier Science B.V. All rights reserved.
Keywords: Recombinant pyruvate oxidase; Phosphate; Biosensor; Polyelectrolyte; Porous carbon electrode; Serum analysis

1. Introduction glucomutase and glucose-6-phosphate dehydrogenase


[9,10], maltose phosphorylase and glucose oxidase
The simple and accurate measurement of the ac- [11], co-immobilized with phosphatase and mura-
tivity of the orthophosphate anion is very important torase [12] have been employed with various degrees
in clinical, biological and environmental studies. A of success. Camman et al. [12] have reported the low-
sensor system (ISE or biosensor) for the sensitive and est detection limit (0.01 M) up until now, while the
selective measurement of this illusive anion is still a best operational stability (70% remaining activity after
challenge [1]. The development of biosensors sensitive 20 h) was reported by Coulet et al. [5] using nucleoside
to phosphate has attracted many researches over the phosphorylase and xanthine oxidase. Additionally the
past 25 years [2]. Amperometric biosensors have been best storage stability (66% remaining activity after 5
usually constructed based on multi-enzyme schemes: weeks dry storage at 20 C) was reported by Cosnier
alkaline phosphatase and glucose oxidase [2,3] or et al. [4] using alkaline phosphatase and polyphenol
polyphenol oxidase [4], nucleoside phosphorylase and oxidase. The simplest and most efficient approach is
xanthine oxidase [58], phosphorylase A, phospho- achieved with the use of pyruvate oxidase (POD), an
enzyme that catalyzes the following reaction
Corresponding author. Tel.: +30-81-393618;
Pyruvate + H2 PO4 + O2
fax: +30-81-393601.
E-mail address: nchan@chemistry.uch.gr (N.A. Chaniotakis). Acetylphosphate + CO2 + H2 O2 (1)

0003-2670/01/$ see front matter 2001 Elsevier Science B.V. All rights reserved.
PII: S 0 0 0 3 - 2 6 7 0 ( 0 0 ) 0 1 2 0 4 - 6
272 V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277

POD requires the presence of thiamine pyrophos- 2. Experimental


phate (TPP), flavine adenine dinucleotide (FAD) and
Mg(II) for the catalysis to take place. These cofactors 2.1. Materials-reagents
make the system not very selective, while POD-based
biosensors are shown to suffer form low operational Recombinant pyruvate oxidase (POD) from Lac-
and storage stability [13,14]. The absence of phos- tobacillus plantarum (EC 1.2.3.3, 1.66 U/mg), avail-
phate anions from the working and storage buffer that able from Roche Diagnostics GmbH (Germany), was
is crucial for the enzyme stability conduces to these kindly provided by Dr. Spohn (University of Halle).
problems [13]. Diethylaminoethyl-dextran (deae-dextran) (D-9885)
In this work, the construction of stable phosphate and DNA sodium salt (D-1626) were obtained from
biosensor based on recombinant pyruvate oxidase Sigma. 4-(2-Hydroxyethyl)piperazine-1-ethanesulfo-
stabilized with polyelectrolyte is presented. Pyruvate nic acid (HEPES), biochemica grade was purchased
oxidase from Lactobacillus plantarum contains the from Fluka.
cofactors TPP, FAD and Mg(II) bound in the active The porous carbon was obtained from BF HIRM
center [15] and is genetically engineered to enhance (Austria). Vitreous carbon foam (bulk density
its stability [1618]. Bergmann et al. [19] have uti- 0.05 g/cm3 , porosity 96.5%), carbon fabric (weight/
lized this enzyme in conjunction with horseradish m2 200 g, warp and weft yarn 200 Tex, plain
peroxidase for the construction of a bienzyme mod- weave) and platinum mesh (wire diameter 0.06 mm,
ified carbon paste biosensor for the measurement of wires/inch 82 82) were purchased from Goodfellow
pyruvate. Cambridge Ltd. In all experiments nano-pure water
In this work, the stabilized pyruvate oxidase is (18 M, EASYpure model D7033, Barnstead) was
immobilized by physical adsorption into a highly used. All other reagents used were of analytical grade.
porous and conductive carbon electrode. This mate-
rial has been used with great success for the con- 2.2. Biosensor construction
struction of highly stable and reproducible glucose
and lactate biosensors [2224]. The polyelectrolytes A rod (2.1 mm diameter, 4.0 mm height) of the
diethylaminoethyl-dextran and DNA are employed porous carbon was cleaned in a sonicated ethanol bath
during the immobilization procedure to stabilize POD for 10 min and then in water bath for 10 min, followed
via a cage formation around the POD. The use of by oven drying at 150 C for 30 min. Pyruvate oxidase
polyelectrolytes such as diethylaminoethyl-dextran was dissolved (to a final concentration 100 U/ml) in
or polyethylenimine (positively charged macro- 10 mM HEPES buffer (pH 7.5) containing the desired
molecules) have been proved to stabilize enzymes amount of the stabilizer. The enzyme was allowed to
[1927] in various environments. It has been shown react with the polyelectrolyte for 20 min at +4 C. Af-
that these polyelectrolytes form a cage-like stable mi- ter this, the cleaned and dried carbon rod was placed
croenvironment around the enzyme via electrostatic into the enzyme solution for 20 h at +4 C. The car-
interactions. Recently, DNA (a negatively charged bon rod was then removed from the solution, washed
macromolecule) was shown to enhance the stability of thoroughly with buffer, and placed in a testing holder.
certain biosensors [28]. Here, the effects of the type It is estimated, based on the amount of solution ad-
and the amount of stabilizer used during the construc- sorbed by the carbon, that 1 unit per carbon rod is im-
tion of the POD-based biosensor on the characteristics mobilized. The electrical contact was achieved from
of the sensor are evaluated, and an optimized sys- the backside through a platinum wire. When not used
tem is constructed. The optimized biosensor is then the biosensors were kept at +4 C in HEPES buffer.
incorporated in a flow system for the measurement
of phosphate levels in serum. The usually occurring 2.3. Pre-oxidizing cell
serum interferences from the easily electro-oxidazible
species are eliminated using an efficient preoxidizing The pre-oxidizing cell used for the blood measure-
cell [29], allowing for the fast and accurate measure- ments was slightly modified from the one previously
ments of phosphate in this complicated matrix. described [29]. The cell consists of two separate parts,
V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277 273

the oxidation and the reduction part connected to a dc The flow injection system consisted of a wall-jet flow
power supply. The constant potential method at an op- cell, an injection valve with loop volume 400 l, while
erational potential above that required to oxidize all the solvent delivery was done using a peristaltic pump
interfering species, but bellow that for the oxidation of (No 7554-30, Cole-Palmer). The response is calcu-
H2 O is chosen. Each part has separate inlet and outlet lated as the peak height from the baseline.
of the flow sample. The inlet of the oxidation part is
connected to the injection valve so that the sample af-
ter the injection will pass only through this part of the 3. Results and discussion
cell. The outlet of the oxidation part (anode) of the cell
is connected to the wall-jet flow cell where the biosen- A detailed experiment in order to obtain the opti-
sor is placed. The inlet of the reducing part (cath- mum operational pH for the phosphate biosensor re-
ode) of the cell is connected to a reservoir containing veals that the pH profile of the sensor is bell shaped
0.1 M KCl while the outlet is connected to the wastes. (data not shown), with optimum value at around pH
The electrical contact of the two half-cells is 7.5. This is slightly shifted to basic pH compared with
achieved through a cellulose acetate membrane (12 the POD pH profile that has been shown to be at pH
16 kDa cut-off) placed between the upper side of the 7.0 [30]. Further experiments were performed using
vitreous carbon foam and the lower side of the plat- 10 mM HEPES buffer adjusted to pH 7.5.
inum mesh. The reactor where the oxidation takes
place is the lower part of the cell, which contains the 3.1. Effect of pyruvate concentration
vitreous carbon foam (6.3 cm 0.6 cm 0.5 cm). It
is composed of a polymethylmethacrylate (PMMA) From Eq. (1) it is clear that pyruvate, a co-substrate
plate (8.3 cm 4.0 cm 1.0 cm) where a hole in the in the enzymatic reaction, is required for the proper
dimensions of the foam is made. To assure homo- operation of the system, and the measurement of
geneous distribution of the potential throughout the phosphate. The optimum amount of pyruvate was
carbon without any IR drop, the electrical contact to determined by varying its concentration in the carrier
the power supply is achieved from the lower side of stream from 0.110 mM, while monitoring its effect
the carbon plate using carbon fabric. on the sensitivity and detection limit as determined
The cathode compartment (also made by PMMA from the calibration curve obtained for each pyruvate
8.3 cm 4.0 cm 1.0 cm) contains the platinum concentration. The results shown in Fig. 1 indicate
mesh (3.5 0.6). The outlet of this part of the cell that an increase in the pyruvate concentration up to
is designed so that the hydrogen bubbles generated 1.0 mM results in an increase in the sensitivity (linear
during the experiment will be quickly removed from
the platinum surface.

2.4. Electrochemical measurements

In all experiments a three electrode Metrohm 641


VA-Detector, a silver/silver chloride double junction
reference electrode (Model 90-02, ORION Res. Inc.),
and a platinum counter electrode (CORNING, cat. no.
476060) were used. The working potential was set at
+800 mV in all experiments, except from the oper-
ational stability where the sensors were polarized at
+600 mV versus Ag/AgCl. The signal was recorded
via a personal computer equipped with a 16-bit A/D
converter and controlled with software written in BA- Fig. 1. Effect of pyruvate concentration, incorporated in the HEPES
SIC. Temperature control at 24.00.1 C was achieved buffer, on the sensitivity () and the detection limit () of the
with a circulating bath (Model 9105, PolyScience). phosphate biosensor.
274 V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277

range 0.051.0 mM phosphate). Further increase in creased beyond the level of 0.25%, w/v the stabilizing
the pyruvate concentration results in lower sensitivi- effect is eliminated and the sensor loses its sensitivity
ties. Even though this decrease of the sensitivity was even faster than the control. The reduced biosensors
also observed by Kubo et al. [14], the large negative stability at higher amounts of deae-dextran has been
slope was not expected. Based on this observation, the previously observed with other biosensors [2224] and
detection limit of the sensor might also be influenced has been attributed to the reduction of enzyme activity
by the background pyruvate concentration. When the at high polyelectrolyte concentrations. Additionally, it
pyruvate concentration is increased from 0.1 up to is shown that DNA has a smaller stabilizing effect on
0.5 mM, there is no influence on the lower detection the sensor. When 0.1%, w/v DNA is used the initial
limit. When the concentration of pyruvate is increased sensors sensitivity is smaller than the control, and the
from 0.5 to 10 mM the lower detection limit also in- stability of the sensor is slightly reduced. Increasing
crease linearly. The fact that the sensitivities obtained the concentration to 0.25%, w/v, the stability of the
with 5 or 10 mM pyruvate were similar indicates that sensor is improved over that of the control, but to a
at these pyruvate concentrations the rate limiting sub- lesser extent from the deae-dextran stabilized sensor.
strate is oxygen. From these results the 1 mM pyruvate Based on these results sensors 1, 2 and 6 were
was chosen to be used for all further experiments. chosen for further evaluation of their operational sta-
bility. For this experiment the sensors were incorpo-
3.2. Effect of polyelectrolyte rated in a flow system and were continuously po-
larized at +600 mV versus Ag/AgCl. The flow rate
The effect of the amount and type of polyelectrolyte was 0.65 ml/min, the temperature was held constant
on the phosphate biosensor was then investigated. Sen- at 24.0 0.1 C and the working solution was 10 mM
sors were prepared from solutions containing vari- HEPES, 1 mM pyruvate and 1 mM phosphate adjusted
ous amounts of deae-dextran or DNA. Deae-dextran to pH 7.5. Under these conditions (buffer stream con-
was varied from 0.1 to 0.5%, w/v, while two different taining the required substrates, pyruvate and phos-
amounts of DNA were tested (0.1 and 0.25%, w/v). phate), there is a continuous enzymatic catalysis. The
Higher amount of DNA resulted in the formation of a system was calibrated daily, and the sensitivity was
suspension, preventing the construction of the sensor. used as a measure to evaluate the activity of the sen-
To monitor the effect of the amount and type of poly- sor. For the calibration procedure, the phosphate was
electrolyte, the sensors were periodically tested. Be- removed from the buffer stream and after equilibration
tween the test times, all sensors were stored at 4 C in different concentrations of phosphate were injected.
buffer solution (10 mM HEPES, pH 7.5). Table 1 sum- Table 2 shows the remaining activity of the three
marizes the results of the sensitivity and the remaining sensors under the continuous operation test. It is inter-
activity of the sensors, 1 month after their construction. esting to note that the operational stability of all the
From this data it is clear that both polyelectrolytes in- sensors monitored is better than any such sensors pre-
crease the stability of the biosensor. It is interesting to sented so far in the literature (70% remaining activity
note that when the amount of deae-dextran used is in- after 20 h [5]). Moreover, the operational stability of

Table 1
Effect of polyelectrolyte type and concentration on the sensitivity and remain activity of the phosphate biosensor
Polyelectrolyte Concentration (%, w/v) Sensitivitya (A/mM) Remaining activitya (%)

1 5.56 0.06 51.8


2 Deae-dextran 0.10 7.94 0.12 77.1
3 Deae-dextran 0.25 5.88 0.14 68.0
4 Deae-dextran 0.50 5.31 0.08 43.4
5 DNA 0.10 3.00 0.03 49.6
6 DNA 0.25 6.15 0.12 57.5
a One month after construction.
V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277 275

Table 2
sensitivity is 10.3 0.2 A/mM and the response
Operational stability of the 1, 2 and 6 phosphate biosensors
is linear to phosphate concentrations from 50 to
Hours polarized Remaining activity (%) 1250 M. The limit of detection for phosphate based
1 2 6 on a signal-to-noise ratio of three is 4.8 M. The re-
20 77 114 104 sponse time varies between 20 and 40 s, depending on
40 72 106 94 the concentration of the analyte. The sensor-to-sensor
60 72 97 89 reproducibility based on the sensitivities is 2.2%
80 69 86 84
100 65 79 79
RSD, while based on the response to 1 mM phos-
120 61 73 72 phate it is 7.6% RSD. These results indicate that the
140 59 71 69 method developed enables the construction of very
220 51 67 56 reproducible sensors. The storage stability of the
phosphate biosensor was then examined, using the
the sensor, 2 is exceptional, with 67% remaining ac- calibration curve of the fresh sensors as the reference
tivity after 220 h. DNA stabilized sensor, 6 has 56% curve. These studies indicate that there is a 84 5%
and the control sensor has 50% remaining activity. residual activity after 2 months of dry storage at 4 C,
while after 6 months the residual activity is reduced to
49 4%.
3.3. Characterization of the phosphate biosensor

A batch of the most stable sensors (2) was then 3.4. Application to serum analysis
prepared. Three of these sensors were used to study
their analytical characteristics, and the remaining were A key factor for the application of an amperomet-
freeze dried and stored at 4 C for storage stability ric sensor in real life samples is its selectivity towards
measurements. At the end of each month one sensor the easily electro-oxidazible species. In the case of
was rehydrated for 20 h and then its residual activity serum, ascorbic acid, uric acid, acetaminophen and
was calculated from the sensitivity obtained from the many other compounds can interfere. The porous car-
calibration curve. bon used here for the construction of the phosphate
In Fig. 2, the initial average calibration curve of sensor has shown to be sensitive to these species (be-
the three freshly prepared sensors is presented. The tween 10 and 13 A/mM), preventing its application
to serum analysis. For this reason a pre-oxidizing cell
[29] was incorporated in the FIA system, to oxidize
these species before they reach the electrode surface.
Initial experiments indicated the need for the buffer
to be mixed with the sample after the pre-oxidation
step. When the buffer was introduced prior to the cell
the response of the sensor was not reproducible. The
final experimental set-up is shown in Fig. 3. The con-
centration of pyruvate in blood varies considerably
(70150 M) [31]. Unless there is a high and constant
concentration of pyruvate in the buffer stream, this
fluctuation will introduce an error to the biosensors
response. Therefore, the concentration of pyruvate in
the buffer stream was raised from 1 to 10 mM. At the
same time, the concentration of HEPES in the buffer
stream was also raised in order to increase its buffer
capacity. Thus, the buffer stream was 50 mM HEPES
Fig. 2. Calibration curve of the phosphate biosensor in 10 mM
HEPES, 1 mM pyruvate buffer adjusted at pH 7.5. The insert and 10 mM pyruvate adjusted to pH 7.5 at a flow rate
presents the linear part of the curve. The working potential was of 0.65 ml/min and the flow rate of the sample was
+800 mV vs. Ag/AgCl. also 0.65 ml/min.
276 V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277

Fig. 3. Diagram of the flow injection system. The buffer consists of 50 mM HEPES and 10 mM pyruvate adjust to pH 7.5. The sample
passes through the pre-oxidizing cell where the interfering species are electrooxidized, then is mixed with the buffer and reach the biosensor
to complete the analysis.

The analytical characteristics of the phosphate for ascorbate, acetaminophen and urate, respectively.
biosensor incorporated in the flow injection system When the cell current is on, all sensitivities are re-
are as follows. The sensitivity is 445 9 nA/mM, the duced to 0.26, 0.30 and 0.51 A/mM, respectively.
response is linear from 0.5 to 10 mM, while the de- With these sensitivities the detection limit of the sys-
tection limit is less than 0.3 mM. In Fig. 4, a typical tem for these species is 362, 853 and 816 M for
recording of phosphate injections is presented. ascorbate, acetaminophen and urate, respectively, val-
Prior to real sample analyses, the efficiency of the ues that are much higher than their normal concentra-
cell in the elimination of interferents was then investi- tion in serum.
gated. With the cell current off, the phosphate biosen- The above experimental set-up was then used to de-
sor exhibits sensitivities 1.72, 2.27 and 1.63 A/mM termine the phosphate level in serum. Three samples
of serum were analyzed and the results were com-
pared with those obtained with ion-chromatography.
The samples were injected in the system without di-
lution and the calculated concentration was found to
be 1.03 0.06 mM. The results from the chromato-
graphic analysis were 0.99 0.02 mM, very close to
the results obtained with the biosensor.

4. Conclusions

The development and evaluation of a phosphate


biosensor based on stabilized with polyelectrolytes
recombinant pyruvate oxidase adsorbed into a porous
carbon has been presented. Both electrolytes used
Fig. 4. Recording of the response of the biosensor incorporated in were shown to have a positive effect on both the oper-
the flow injection system shown in Fig. 3 to phosphate injections
ational and storage stability of the resulting biosensor,
(2.5, 5.0, 7.5 and 10 mM phosphate, respectively). The sensor is
polarized at +800 mV vs. Ag/AgCl. The flow rate is 0.65 ml/min with 0.1%, w/v deae-dextran being the optimum. The
and the buffer consist of 50 mM HEPES and 10 mM pyruvate phosphate biosensor exhibits exceptional operational
adjusted to pH 7.5. and storage stability while the simple construction
V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277 277

procedure results in a very good sensor-to-sensor [11] S. Huwel, L. Haalck, N. Conrath, F. Spener, Enz. Microb.
reproducibility. The use of the sensor in an FIA sys- Technol. 21 (1997) 413.
tem equipped with a pre-oxidizing cell allows for [12] N. Conrath, B. Grundig, S. Huwel, K. Cammann, Anal. Chim.
Acta 309 (1995) 47.
the measurement of the activity of phosphate in dif- [13] N. Gajovic, K. Habermuller, A. Warsinke, W. Schuhmann,
ficult samples such as those of human serum with F.W. Scheller, Electroanalysis 11 (1999) 1377.
results in good agreement with those obtained with [14] I. Kubo, M. Inagawa, T. Sugawara, Y. Arikawa, I. Karube,
ion-chromatography. Anal. Lett. 24 (1991) 1711.
[15] Y. Muller, G. Schumacher, R. Rudolph, G. Schulz, J. Mol.
Biol. 237 (1994) 315.
[16] E. Kopetzki, K. Lehnert, P. Buckel, Clin. Chem. 40 (1994)
Acknowledgements 688.
[17] B. Risse, G. Stempfer, R. Rudolph, H. Mollering, R. Jaenicke,
Prot. Sci. 1 (1992) 1699.
We would like to thank, Dr. U. Spohn (University of
[18] B. Risse, G. Stempfer, R. Rudolph, G. Schumacher, R.
Hall) for supplying the enzyme, G. Kouvarakis (ECPL, Jaenicke, Prot. Sci. 1 (1992) 1710.
University of Crete) for the ion-chromatography mea- [19] W. Bergmann, R. Rudolph, U. Spohn, Anal. Chim. Acta 394
surements and the Greek Secretary General of Re- (1999) 233.
search and Development for financial support through [20] A. Appleton, T.D. Gibson, J.R. Woodward, Sens. Act. B 43
(1997) 65.
the YPER program (97 YPER-219).
[21] T.D. Gibson, J.R. Woodward, in: P.G. Eldman, J. Wang (Eds.),
Biosensors and Chemical Sensors, ACS Books, 1992, p. 40.
[22] V.G. Gavalas, N.A. Chaniotakis, T.D. Gibson, Biosens.
References Bioelectron. 13 (1998) 1205.
[23] V.G. Gavalas, N.A. Chaniotakis, Anal. Chim. Acta 404 (2000)
[1] S.O. Engblom, Biosens. Bioelectron. 13 (1998) 981. 67.
[2] G.G. Guilbault, M. Nanjo, Anal. Chim. Acta 78 (1975) 69. [24] V.G. Gavalas, N.A. Chaniotakis, Mikrochim. Acta, submitted
[3] Y. Su, M. Mascini, Anal. Lett. 28 (1995) 1359. for publication.
[4] S. Cosnier, C. Gondran, J.-C. Watelet, W. De Giovani, R.P.M. [25] U. Spohn, D. Narasaiah, L. Gorton, Electroanalysis 8 (1996)
Furriel, F.A. Leone, Anal. Chem. 70 (1998) 3952. 507.
[5] E.M. dUrso, P.R. Coulet, Anal. Chim. Acta 239 (1990) 1. [26] H. Liden, J. Volc, G. Marko-Varga, L. Gorton, Electroanalysis
[6] E.M. dUrso, P.R. Coulet, Anal. Chim. Acta 281 (1993) 535. 10 (1998) 223.
[7] S.D. Haemmerli, A.A. Suleiman, G.G. Guilbault, Anal. [27] M. Tessema, E. Csoregi, T. Ruzgas, G. Kenausis, T. Solomon,
Biochem. 191 (1990) 106. L. Gorton, Anal. Chem. 69 (1997) 4039.
[8] U. Wollenberger, F. Schubert, F.W. Scheller, Sens. Act. B 7 [28] P. Dantoni, S.H.P. Serrano, A.M.O. Brett, I.G.R. Gutz, Anal.
(1992) 412. Chim. Acta 366 (1998) 137.
[9] J.J. Fernandez, J.R. Lopez, X. Correig, I. Katakis, Sens. Act. [29] V.G. Gavalas, M.G. Fouskaki, N.A. Chaniotakis, Anal. Lett.
B 47 (1998) 13. 33 (2000) 2391.
[10] J. Parellada, A. Narvaez, M.A. Lopez, E. Dominguez, J.J. [30] Roche Molecular Biochemicals 1999/2000.
Fernandez, V. Pavlov, I. Katakis, Anal. Chim. Acta 362 (1998) [31] R.K. Murray, D.K. Granner, P.A. Mayes, V.W. Rodwell,
47. Harpers Biochemistry, Appleton & Lange, 1996, p. 829.

View publication stats

S-ar putea să vă placă și