Documente Academic
Documente Profesional
Documente Cultură
discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/229218341
CITATIONS READS
50 48
2 authors, including:
Nikos Chaniotakis
University of Crete
109 PUBLICATIONS 3,796 CITATIONS
SEE PROFILE
All content following this page was uploaded by Nikos Chaniotakis on 01 September 2014.
Abstract
In this work the development of a pyruvate oxidase-based phosphate biosensor is illustrated. The use of polyelectrolyte
stabilized recombinant pyruvate oxidase in conjunction with a porous conductive carbon results in the development of a
simple, reproducible and stable phosphate biosensor. The polyelectrolyte diethylaminoethyl-dextran or DNA was used as
the enzyme stabilizer, and the resulting enzymepolyelectrolyte complexes were physically adsorbed into the transducer, a
highly porous and conductive carbon electrode, for the construction of the biosensor. The optimized biosensor exhibits high
operational (67% remaining activity after 220 h) and storage (49% remaining activity after 24 weeks) stability, and very good
sensor-to-sensor reproducibility. The optimized phosphate biosensor was used for the measurement of the phosphate ion
activity in serum. 2001 Elsevier Science B.V. All rights reserved.
Keywords: Recombinant pyruvate oxidase; Phosphate; Biosensor; Polyelectrolyte; Porous carbon electrode; Serum analysis
0003-2670/01/$ see front matter 2001 Elsevier Science B.V. All rights reserved.
PII: S 0 0 0 3 - 2 6 7 0 ( 0 0 ) 0 1 2 0 4 - 6
272 V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277
the oxidation and the reduction part connected to a dc The flow injection system consisted of a wall-jet flow
power supply. The constant potential method at an op- cell, an injection valve with loop volume 400 l, while
erational potential above that required to oxidize all the solvent delivery was done using a peristaltic pump
interfering species, but bellow that for the oxidation of (No 7554-30, Cole-Palmer). The response is calcu-
H2 O is chosen. Each part has separate inlet and outlet lated as the peak height from the baseline.
of the flow sample. The inlet of the oxidation part is
connected to the injection valve so that the sample af-
ter the injection will pass only through this part of the 3. Results and discussion
cell. The outlet of the oxidation part (anode) of the cell
is connected to the wall-jet flow cell where the biosen- A detailed experiment in order to obtain the opti-
sor is placed. The inlet of the reducing part (cath- mum operational pH for the phosphate biosensor re-
ode) of the cell is connected to a reservoir containing veals that the pH profile of the sensor is bell shaped
0.1 M KCl while the outlet is connected to the wastes. (data not shown), with optimum value at around pH
The electrical contact of the two half-cells is 7.5. This is slightly shifted to basic pH compared with
achieved through a cellulose acetate membrane (12 the POD pH profile that has been shown to be at pH
16 kDa cut-off) placed between the upper side of the 7.0 [30]. Further experiments were performed using
vitreous carbon foam and the lower side of the plat- 10 mM HEPES buffer adjusted to pH 7.5.
inum mesh. The reactor where the oxidation takes
place is the lower part of the cell, which contains the 3.1. Effect of pyruvate concentration
vitreous carbon foam (6.3 cm 0.6 cm 0.5 cm). It
is composed of a polymethylmethacrylate (PMMA) From Eq. (1) it is clear that pyruvate, a co-substrate
plate (8.3 cm 4.0 cm 1.0 cm) where a hole in the in the enzymatic reaction, is required for the proper
dimensions of the foam is made. To assure homo- operation of the system, and the measurement of
geneous distribution of the potential throughout the phosphate. The optimum amount of pyruvate was
carbon without any IR drop, the electrical contact to determined by varying its concentration in the carrier
the power supply is achieved from the lower side of stream from 0.110 mM, while monitoring its effect
the carbon plate using carbon fabric. on the sensitivity and detection limit as determined
The cathode compartment (also made by PMMA from the calibration curve obtained for each pyruvate
8.3 cm 4.0 cm 1.0 cm) contains the platinum concentration. The results shown in Fig. 1 indicate
mesh (3.5 0.6). The outlet of this part of the cell that an increase in the pyruvate concentration up to
is designed so that the hydrogen bubbles generated 1.0 mM results in an increase in the sensitivity (linear
during the experiment will be quickly removed from
the platinum surface.
range 0.051.0 mM phosphate). Further increase in creased beyond the level of 0.25%, w/v the stabilizing
the pyruvate concentration results in lower sensitivi- effect is eliminated and the sensor loses its sensitivity
ties. Even though this decrease of the sensitivity was even faster than the control. The reduced biosensors
also observed by Kubo et al. [14], the large negative stability at higher amounts of deae-dextran has been
slope was not expected. Based on this observation, the previously observed with other biosensors [2224] and
detection limit of the sensor might also be influenced has been attributed to the reduction of enzyme activity
by the background pyruvate concentration. When the at high polyelectrolyte concentrations. Additionally, it
pyruvate concentration is increased from 0.1 up to is shown that DNA has a smaller stabilizing effect on
0.5 mM, there is no influence on the lower detection the sensor. When 0.1%, w/v DNA is used the initial
limit. When the concentration of pyruvate is increased sensors sensitivity is smaller than the control, and the
from 0.5 to 10 mM the lower detection limit also in- stability of the sensor is slightly reduced. Increasing
crease linearly. The fact that the sensitivities obtained the concentration to 0.25%, w/v, the stability of the
with 5 or 10 mM pyruvate were similar indicates that sensor is improved over that of the control, but to a
at these pyruvate concentrations the rate limiting sub- lesser extent from the deae-dextran stabilized sensor.
strate is oxygen. From these results the 1 mM pyruvate Based on these results sensors 1, 2 and 6 were
was chosen to be used for all further experiments. chosen for further evaluation of their operational sta-
bility. For this experiment the sensors were incorpo-
3.2. Effect of polyelectrolyte rated in a flow system and were continuously po-
larized at +600 mV versus Ag/AgCl. The flow rate
The effect of the amount and type of polyelectrolyte was 0.65 ml/min, the temperature was held constant
on the phosphate biosensor was then investigated. Sen- at 24.0 0.1 C and the working solution was 10 mM
sors were prepared from solutions containing vari- HEPES, 1 mM pyruvate and 1 mM phosphate adjusted
ous amounts of deae-dextran or DNA. Deae-dextran to pH 7.5. Under these conditions (buffer stream con-
was varied from 0.1 to 0.5%, w/v, while two different taining the required substrates, pyruvate and phos-
amounts of DNA were tested (0.1 and 0.25%, w/v). phate), there is a continuous enzymatic catalysis. The
Higher amount of DNA resulted in the formation of a system was calibrated daily, and the sensitivity was
suspension, preventing the construction of the sensor. used as a measure to evaluate the activity of the sen-
To monitor the effect of the amount and type of poly- sor. For the calibration procedure, the phosphate was
electrolyte, the sensors were periodically tested. Be- removed from the buffer stream and after equilibration
tween the test times, all sensors were stored at 4 C in different concentrations of phosphate were injected.
buffer solution (10 mM HEPES, pH 7.5). Table 1 sum- Table 2 shows the remaining activity of the three
marizes the results of the sensitivity and the remaining sensors under the continuous operation test. It is inter-
activity of the sensors, 1 month after their construction. esting to note that the operational stability of all the
From this data it is clear that both polyelectrolytes in- sensors monitored is better than any such sensors pre-
crease the stability of the biosensor. It is interesting to sented so far in the literature (70% remaining activity
note that when the amount of deae-dextran used is in- after 20 h [5]). Moreover, the operational stability of
Table 1
Effect of polyelectrolyte type and concentration on the sensitivity and remain activity of the phosphate biosensor
Polyelectrolyte Concentration (%, w/v) Sensitivitya (A/mM) Remaining activitya (%)
Table 2
sensitivity is 10.3 0.2 A/mM and the response
Operational stability of the 1, 2 and 6 phosphate biosensors
is linear to phosphate concentrations from 50 to
Hours polarized Remaining activity (%) 1250 M. The limit of detection for phosphate based
1 2 6 on a signal-to-noise ratio of three is 4.8 M. The re-
20 77 114 104 sponse time varies between 20 and 40 s, depending on
40 72 106 94 the concentration of the analyte. The sensor-to-sensor
60 72 97 89 reproducibility based on the sensitivities is 2.2%
80 69 86 84
100 65 79 79
RSD, while based on the response to 1 mM phos-
120 61 73 72 phate it is 7.6% RSD. These results indicate that the
140 59 71 69 method developed enables the construction of very
220 51 67 56 reproducible sensors. The storage stability of the
phosphate biosensor was then examined, using the
the sensor, 2 is exceptional, with 67% remaining ac- calibration curve of the fresh sensors as the reference
tivity after 220 h. DNA stabilized sensor, 6 has 56% curve. These studies indicate that there is a 84 5%
and the control sensor has 50% remaining activity. residual activity after 2 months of dry storage at 4 C,
while after 6 months the residual activity is reduced to
49 4%.
3.3. Characterization of the phosphate biosensor
A batch of the most stable sensors (2) was then 3.4. Application to serum analysis
prepared. Three of these sensors were used to study
their analytical characteristics, and the remaining were A key factor for the application of an amperomet-
freeze dried and stored at 4 C for storage stability ric sensor in real life samples is its selectivity towards
measurements. At the end of each month one sensor the easily electro-oxidazible species. In the case of
was rehydrated for 20 h and then its residual activity serum, ascorbic acid, uric acid, acetaminophen and
was calculated from the sensitivity obtained from the many other compounds can interfere. The porous car-
calibration curve. bon used here for the construction of the phosphate
In Fig. 2, the initial average calibration curve of sensor has shown to be sensitive to these species (be-
the three freshly prepared sensors is presented. The tween 10 and 13 A/mM), preventing its application
to serum analysis. For this reason a pre-oxidizing cell
[29] was incorporated in the FIA system, to oxidize
these species before they reach the electrode surface.
Initial experiments indicated the need for the buffer
to be mixed with the sample after the pre-oxidation
step. When the buffer was introduced prior to the cell
the response of the sensor was not reproducible. The
final experimental set-up is shown in Fig. 3. The con-
centration of pyruvate in blood varies considerably
(70150 M) [31]. Unless there is a high and constant
concentration of pyruvate in the buffer stream, this
fluctuation will introduce an error to the biosensors
response. Therefore, the concentration of pyruvate in
the buffer stream was raised from 1 to 10 mM. At the
same time, the concentration of HEPES in the buffer
stream was also raised in order to increase its buffer
capacity. Thus, the buffer stream was 50 mM HEPES
Fig. 2. Calibration curve of the phosphate biosensor in 10 mM
HEPES, 1 mM pyruvate buffer adjusted at pH 7.5. The insert and 10 mM pyruvate adjusted to pH 7.5 at a flow rate
presents the linear part of the curve. The working potential was of 0.65 ml/min and the flow rate of the sample was
+800 mV vs. Ag/AgCl. also 0.65 ml/min.
276 V.G. Gavalas, N.A. Chaniotakis / Analytica Chimica Acta 427 (2001) 271277
Fig. 3. Diagram of the flow injection system. The buffer consists of 50 mM HEPES and 10 mM pyruvate adjust to pH 7.5. The sample
passes through the pre-oxidizing cell where the interfering species are electrooxidized, then is mixed with the buffer and reach the biosensor
to complete the analysis.
The analytical characteristics of the phosphate for ascorbate, acetaminophen and urate, respectively.
biosensor incorporated in the flow injection system When the cell current is on, all sensitivities are re-
are as follows. The sensitivity is 445 9 nA/mM, the duced to 0.26, 0.30 and 0.51 A/mM, respectively.
response is linear from 0.5 to 10 mM, while the de- With these sensitivities the detection limit of the sys-
tection limit is less than 0.3 mM. In Fig. 4, a typical tem for these species is 362, 853 and 816 M for
recording of phosphate injections is presented. ascorbate, acetaminophen and urate, respectively, val-
Prior to real sample analyses, the efficiency of the ues that are much higher than their normal concentra-
cell in the elimination of interferents was then investi- tion in serum.
gated. With the cell current off, the phosphate biosen- The above experimental set-up was then used to de-
sor exhibits sensitivities 1.72, 2.27 and 1.63 A/mM termine the phosphate level in serum. Three samples
of serum were analyzed and the results were com-
pared with those obtained with ion-chromatography.
The samples were injected in the system without di-
lution and the calculated concentration was found to
be 1.03 0.06 mM. The results from the chromato-
graphic analysis were 0.99 0.02 mM, very close to
the results obtained with the biosensor.
4. Conclusions
procedure results in a very good sensor-to-sensor [11] S. Huwel, L. Haalck, N. Conrath, F. Spener, Enz. Microb.
reproducibility. The use of the sensor in an FIA sys- Technol. 21 (1997) 413.
tem equipped with a pre-oxidizing cell allows for [12] N. Conrath, B. Grundig, S. Huwel, K. Cammann, Anal. Chim.
Acta 309 (1995) 47.
the measurement of the activity of phosphate in dif- [13] N. Gajovic, K. Habermuller, A. Warsinke, W. Schuhmann,
ficult samples such as those of human serum with F.W. Scheller, Electroanalysis 11 (1999) 1377.
results in good agreement with those obtained with [14] I. Kubo, M. Inagawa, T. Sugawara, Y. Arikawa, I. Karube,
ion-chromatography. Anal. Lett. 24 (1991) 1711.
[15] Y. Muller, G. Schumacher, R. Rudolph, G. Schulz, J. Mol.
Biol. 237 (1994) 315.
[16] E. Kopetzki, K. Lehnert, P. Buckel, Clin. Chem. 40 (1994)
Acknowledgements 688.
[17] B. Risse, G. Stempfer, R. Rudolph, H. Mollering, R. Jaenicke,
Prot. Sci. 1 (1992) 1699.
We would like to thank, Dr. U. Spohn (University of
[18] B. Risse, G. Stempfer, R. Rudolph, G. Schumacher, R.
Hall) for supplying the enzyme, G. Kouvarakis (ECPL, Jaenicke, Prot. Sci. 1 (1992) 1710.
University of Crete) for the ion-chromatography mea- [19] W. Bergmann, R. Rudolph, U. Spohn, Anal. Chim. Acta 394
surements and the Greek Secretary General of Re- (1999) 233.
search and Development for financial support through [20] A. Appleton, T.D. Gibson, J.R. Woodward, Sens. Act. B 43
(1997) 65.
the YPER program (97 YPER-219).
[21] T.D. Gibson, J.R. Woodward, in: P.G. Eldman, J. Wang (Eds.),
Biosensors and Chemical Sensors, ACS Books, 1992, p. 40.
[22] V.G. Gavalas, N.A. Chaniotakis, T.D. Gibson, Biosens.
References Bioelectron. 13 (1998) 1205.
[23] V.G. Gavalas, N.A. Chaniotakis, Anal. Chim. Acta 404 (2000)
[1] S.O. Engblom, Biosens. Bioelectron. 13 (1998) 981. 67.
[2] G.G. Guilbault, M. Nanjo, Anal. Chim. Acta 78 (1975) 69. [24] V.G. Gavalas, N.A. Chaniotakis, Mikrochim. Acta, submitted
[3] Y. Su, M. Mascini, Anal. Lett. 28 (1995) 1359. for publication.
[4] S. Cosnier, C. Gondran, J.-C. Watelet, W. De Giovani, R.P.M. [25] U. Spohn, D. Narasaiah, L. Gorton, Electroanalysis 8 (1996)
Furriel, F.A. Leone, Anal. Chem. 70 (1998) 3952. 507.
[5] E.M. dUrso, P.R. Coulet, Anal. Chim. Acta 239 (1990) 1. [26] H. Liden, J. Volc, G. Marko-Varga, L. Gorton, Electroanalysis
[6] E.M. dUrso, P.R. Coulet, Anal. Chim. Acta 281 (1993) 535. 10 (1998) 223.
[7] S.D. Haemmerli, A.A. Suleiman, G.G. Guilbault, Anal. [27] M. Tessema, E. Csoregi, T. Ruzgas, G. Kenausis, T. Solomon,
Biochem. 191 (1990) 106. L. Gorton, Anal. Chem. 69 (1997) 4039.
[8] U. Wollenberger, F. Schubert, F.W. Scheller, Sens. Act. B 7 [28] P. Dantoni, S.H.P. Serrano, A.M.O. Brett, I.G.R. Gutz, Anal.
(1992) 412. Chim. Acta 366 (1998) 137.
[9] J.J. Fernandez, J.R. Lopez, X. Correig, I. Katakis, Sens. Act. [29] V.G. Gavalas, M.G. Fouskaki, N.A. Chaniotakis, Anal. Lett.
B 47 (1998) 13. 33 (2000) 2391.
[10] J. Parellada, A. Narvaez, M.A. Lopez, E. Dominguez, J.J. [30] Roche Molecular Biochemicals 1999/2000.
Fernandez, V. Pavlov, I. Katakis, Anal. Chim. Acta 362 (1998) [31] R.K. Murray, D.K. Granner, P.A. Mayes, V.W. Rodwell,
47. Harpers Biochemistry, Appleton & Lange, 1996, p. 829.