AXIOM Luminos dRF

The 2-in-1 Solution Remote-Controlled Fluoroscopy


and Radiography System with Flat Detector (FD)

Technical Data
s
www.siemens.com/medical medical
AXIOM Luminos dRF
Remote-controlled fluoroscopy and radiography system with flat detector (FD)

AXIOM Luminos dRF is a highly versatile System highlights Image acquisition system:
radiography and fluoroscopy system with
Dynamic flat detector with excellent
dynamic flat detector (FD) technology. Patient table:
spatial and contrast resolution
The four main pillars of the AXIOM Modular construction adaptable to your 43 cm x 43 cm detector surface
concept are: individual requirements: 3 zoom levels for better detail
Easy handling Tilt range in head-down position recognition
up to 45 with adjustable table High-contrast image display with
Outstanding image quality
height optimal resolution
Maximum radiation reduction
up to 90 with adjustable table
Optimal networking height* Image processing system:
The integrated flat detector makes this Flat detector FLUOROSPOT Compact high-resolution
system the universal digital solution for Fully digital system with large digital x-ray system with up to
improved workflows during fluoroscopy 43 cm x 43 cm flat detector 2840 x 2880/12-bit matrix
and radioscopy examinations as well as Remote control console for system Standard operating modes from a single
optimized networking capabilities for control image to series with frame rates from
documentation and integration into 0.5 to 8 f/s in up to 1440 x 1440/12-bit
Tableside control in the exam room
the clinical environment. matrix
The innovative design of the AXIOM Options for the patient table: Pulsed fluoroscopy* CAREVISON
Luminos dRF offers a variable table and expanded CARE applications
Universal collimator with five remote-
height and excellent access to the Full DICOM 3 functionality for patient
controllable, semi-transparent wedge-
patient, even from the rear of the data acquisition* as well as for
shaped filters
system, for more comfort for both documentation and archiving*
user and patient. Remote-controlled compression cone
with park position outside of the Harmonization (DDO)
Thanks to the high load capacity of the
table and robust x-ray tube, the large radiation field DiamondView improves organ-specific
field of view on the FD and the excellent Tomography with six programs detail contrast
dose-reducing functions, the system Injector interface with radiation release
can be used for a wide range of patient X-ray generation:
synchronization
types. High-frequency generator with
Intercom system between the
For better utilization of the fluoroscopy CAREMATIC fully automatic dose rate
examination and control room
system, FD technology can fully unite control
digital fluoroscopy and digital radio- Trolley-mounted remote control console
Choice of 65 kW and 80 kW*
graphy in one room. in the examination room
Automatic exposure control
Table for remote control console
Dual-focus high-performance x-ray tube
Scan technology for composite image assembly with air cooling in the housing
display on the syngo Workplace High-performance focal spot for best
image quality
Graphite compound anode for short
tube cooling periods and high patient
throughput

Display:
48 cm monochrome display

Display carrier*:
Display trolley for 1 or 2 displays
Ceiling suspension for 1 or 2 monitors
with longitudinal and vertical travel as
well as swivel support
* Option
AXIOM Luminos dRF
Technical Data

Table and patient tabletop


Table tilt Motorized from + 90/ 45 or + 90/ 90*;
two different tilt speeds of 3/s and 6/s;
tilt time from 90 to 0: 17 s including soft start and stop)
Selectable automatic stop in horizontal position (0),
digital angle display for remote and tableside control
Table height 50 cm 100 cm continuously adjustable, approx. 4 cm/s
48 cm 98 cm for floor installation
Tabletop Dimensions: 210 cm x 80 cm;
radiolucent area: 193 cm x 53.5 cm;
flat, scratch-resistant surface, flat, accessory rails;
Al equivalent:
0.65 mm ( 0.1) Al at 100 kV/3.7 mm Al HLV)
Patient weight Up to 150 kg without limitations,
151 180 kg at slower table speed,
longitudinal movement limited to 40 cm;
181 230 kg, from center position, tilt allowed;
Because of CPR requirements, interventional examinations limited to
patients up to 180 kg (with additional 50 kg CPR equipment on centered table)
Longitudinal travel 160 cm motorized, 80 cm at both head and foot end;
speed approx. 6 cm/s (50 Hz) or approx. 7.2 cm/s (60 Hz)
Transverse travel 35 cm motorized, 17.5 cm to the left and right;
speed approx 4.5 cm/s
Footboard For head or foot end, height adjustable (3 positions),
can be lowered to 4 cm above the floor with the table in vertical position

X-ray tube stand


SID 115 cm, 150 cm, motorized adjustment at approx. 5 cm/s
Oblique projection Max. 40; motorized adjustment of pivot point height 10 to 300 mm above
the tabletop (= parallax correction of the object); digital angle display for
remote and tableside control
Tube rotation Motorized from + 90 to 180
Tabletop-detector distance Min. 7.0 cm
Scatter radiation grid Stationary, Pb 15:1, 80 lines/cm, f0 = 125 cm, comfortable distance between
grid and table front
Image receptor movement Max. 113 cm, (remote and tableside control),
precision-adjustable up to 7 cm/s max.
Central beam height above floor 77 cm to 182 cm with system set up (+ 90);
57 cm to 162 cm at max. 90 (only with 90 version)
Central beam distance to end of table Min. 38 cm (head end) with orthogonal beam path

4 * Option
Flat detector 43 cm x 43 cm
Amorphous silicon flat detector based on indirect conversion technology
High-resolution 2840 x 2880 matrix with 148 m pixel size and 16-bit digitization depth
High-performance fiber-optic connection to digital imaging system
Removable grid
Input fields Overview Zoom 1 Zoom 2 Zoom 3
(active field) 42.0 cm x 42.6 cm 30 cm x 30 cm 22 cm x 22 cm 15 cm x 15 cm
Material aSi with CsI scintillator
Pixel size 148 m
Spatial resolution (Nyquist frequency) 3.4 Lp/mm
Maximum acquisition speed Up to 8 f/s (fluoro 15 f/s)
Matrix Up to 2840 x 2880 pixels
Digitization depth 16 bits
Detective quantum efficiency (DQE) > 65% (at 0 Lp/mm)
Modulation depth > 63% (at 1 Lp/mm)
Modulation depth at the Nyquist frequency Approx. 13%
Weight 25 kg
Five selectable solid-state measurement fields; measurement fields are adjusted according to collimation depending on
the zoom format selected.

Compression system*
Compression system Remote-controlled, radiolucent cone, retractable, park position outside the
radiation field.
Automatic safety switch-off for system movements starting at 50 N
Compression > 50 N Limited, controlled system movements
Compression force 5 N to max. 155 N (or 80 N in Japan), continuously adjustable
Force display Digital display (LCD) in increments from 1 to 15 (1 level approx. 10 N)

Primary collimator
Inherent filtration 1 mm Al at 80 kV
Distance from source to collimator flange 80 mm
Cu prefilter (1 HLV mm Al) 0.1 mm (3.5 mm), 0.2 mm (7.1 mm), 0.3 mm (10.8 mm), motorized and
positionable, copper filter for DFR configurable in organ program, display on
monitor and LCD of collimator
Collimator rotation Up to a maximum of 45, detent at 0
Profile rails Two levels for inserting additional filters, cones or collimators
Collimator with wedge-shaped filters* Two asymmetrical semi-transparent dual wedge-shaped filters at 6 and 14;
one symmetrical semi-transparent dual wedge-shaped filter at 8;
free motorized positioning in the radiation field

* Option 5
AXIOM Luminos dRF
Technical Data

Dose area product measurement / CAREMAX*


Measurement chamber Ionization chamber, built into the primary collimator
Display of measurement values a) on the console;
b) on the live image monitor (CAREWATCH)
Measurement values a) and b) dose area product
b) skin entrance dose, standardized at 30 cm from table, can be displayed in
mGy or as a percentage of a configurable dose limit value or the incident
dose rate during fluoroscopy in accordance with IEC

Tomography*
Principle Linear tomo technique (planigraphy) with 6 programs in all system positions
with detector; automatic positioning of object and image receptor within the
work area of the stand
Planigraphic angle Tomo Time Stand work area Object scan range
8 0.4 s / 0.8 s 62.5 cm 198.0 cm
20 0.6 s / 1.2 s 48.0 cm 192.5 cm
40 1.2 s / 2.5 s 17.0 cm 177.0 cm
Tomo angle at 300 mm tomo height
SID 115 cm
Tomo height 10 mm to 300 mm above tabletop, adjustable in mm (speed 5 cm/s),
height selectable in mm, remote or tableside control, display in mm on the
main display of the remote control console

Operation
Operating console For system operation from the control room and optional connection to a
second operating console for the examination room, mounted on an
optional trolley.
Complete operation via modular control elements for patient table and tube
stand movements as well as compression.
Integrated touchscreen control panel for operating generator and imaging
system, display of current system state.
Footswitch for fluoroscopy and acquisition, optional second footswitch for
examination room
Tableside control panel at the system With control elements for table and tube stand movement and collimator
setting.

6 * Option
FLUOROSPOT Compact Digital Fluoro-Radiography System
Image acquisition system Intel compatible microprocessor with PCI bus architecture, min. 2 GHz,
1 GB RAM, SCSI drive, USB 2.0 and interface cards for the detector/x-ray
system, Windows XP-based operating system
Acquisition memory 10,000 images on hard disk in 1k x 1k matrix and
2,000 images in 2840 x 2880 matrix
Frame rates Fluoroscopy:
Pulsed: 15 f/s
1k x 1k/12-bit matrix
Pulsed* (CAREVISION):
7.5 or 3 f/s: 1k x 1k/12-bit matrix
DSA* and series exposures:
1k x 1k or 1440 x 1440 matrix: 0.5, 1, 2, 4 or 8 f/s
Digital radiography DR images up to 2840 x 2880 pixels/12-bit
Image display Aspect ratio 5:4, corresponding to 1280 x 1024 matrix,
1k x 1k image content
Image processing Edge enhancement filters, black/white image reversal, windowing for
contrast/brightness, electronic shuttering, roaming, vertical and horizontal
image reversal, 2 x zoom (full size), 2 x electronic magnifying glass (EMG),
right/left indication, single image digital subtraction (SDS) with pixel shift
DDO (Harmonization) Digital Density Optimization (DDO), preprogrammed and/or during
postprocessing, to compensate for undesirable density differences in the
image, including online DDO for fluoroscopy
DiamondView (only in radiography mode) DiamondView is a multiscale method, i.e., filter size and strength are
weighted differently and are adapted to the overall image content.
DiamondView* expands the signal processing capabilities of the dynamic
range and improves organ-specific detail contrast (soft tissue and bones)
Quantification Angle/distance measurement, automatic or manual distance calibration
Text/graphic functions Text: marking, annotation, image comments, R/L marking
Graphics: quantification with angle/distance measurement
Organ programs X-ray parameters: e.g., kV automatic or fixed, dose, frame rate,
The following can be set: frame rate reduction, focal spot, tube focus, automatic prefiltration,
FL program, automatic exposure with selection of measurement chamber
Imaging processing parameters: e.g., window values, edge enhancement,
positive/negative image display, harmonization factor (DDO), DiamondView
Automatic functions: yes/no: e.g., auto-window, auto-shutter, auto Cu filter
Background functionality Imaging functions such as DICOM Send/Print, CD-R or DVD-R/DVD+R*
burning are performed in background mode
Patient directory Input of, e.g., patient name, identification number, date of birth,
examination number, order number, physician, organ. Input via keyboard or
directly via DICOM Worklist*

* Option 7
AXIOM Luminos dRF
Technical Data

Options for FLUOROSPOT Compact


Reference image display 1k x 1k matrix: Storage and direct opening of up to 16 reference images on
the second display
DSA functionality Online DSA with pixel shift, remasking, roadmap, peak opacification for
iodine contrast (MaxOp) and CO2 contrast (MinOp), display of anatomical
background (landmark) from 0% to 100%, summation of mask and filling
images for contrast enhancement
Fluoro loop incl. storage expansion Storage and display of dynamic fluoroscopy sequences,
to 30,000 images maximum storage duration depends on the pulse frequency,
e.g., 15 f/s approx. 30 s, 7.5 f/s approx. 60 s and 3 f/s approx. 150 s
Stenosis quantification Quantification program for geometric and densitometric values
Scan technology for display of long images
Security package For enhanced user administration, including user authentication
and data protection

Archiving
DVD recorder for fluoro recording Direct output of fluoroscopy and acquisition series on DVD recorder
CD/DVD recorder DVD drive for digital image storage on CD-R, DVD+R or DVD-R for offline
data exchange in DICOM 3, TIFF and AVI format
Printer connection For paper copies from Level 2 PostScript printer; not suitable for diagnosis

8
DICOM 3 network interfaces
DICOM Send/St C (basic configuration)
Network interface in DICOM 3 standard for DICOM 3-compatible image communication network
Confirmation from the image archive (St C = Storage Commitment)
DICOM Print
For connection to a network laser camera or network printer
DICOM Query/Retrieve*
Retrieve images from a patient image archive (PACS)
DICOM Get Worklist/MPPS*
Get Worklist function for importing patient data from a data management system (RIS/HIS). XRF, CR and DX worklist
entries supported, configurable.
Modality Performed Procedure Step (MPPS); send examination statistics and dose information to a data management
system

CARE
CAREMATIC
CAREMATIC automatic x-ray control system for fully automatic calculation and optimization of exposure data based on
fluoroscopic values
CAREFILTER
Three-level adaptive CU prefiltration (CAREFILTER) for reducing patient dose; organ programming, selection automatically
controlled according to the absorption of the patient (autofilter)
Filter levels 0.1/0.2/0.3 mm Cu
CAREVISION*
Pulsed fluoroscopy with additional reduced pulse frequencies of 3.0 and 7.5 P/s
Adjustment of the pulse frequency to the respective requirements of the application for a significant reduction of
radiation exposure, especially with time-intensive interventional procedures
CAREPROFILE*
Radiation-free primary collimation and positioning of semi-transparent wedge-shaped filters in LIH image (last image
hold) via graphic display on the system display
Automatic selection of graphic displays during control of primary collimation and filter
CAREPOSITION*
Radiation-free positioning of the object and/or target system via graphic display of the central beam and the image edges
in the LIH image on the image monitor as orientation points (only in combination with CAREPROFILE)
CAREMAX**/*
Electronics unit with KermaX, a measurement chamber integrated into the collimator housing for acquisition of dose area
product and/or standardized patient dose
Displayed on the generator display and image system monitor (CAREWATCH)
Different displays can be configured for fluoroscopy and for fluoro pause:
During fluoro: Skin dose level
During fluoro pause: Accumulated skin dose or area dose product or percentage of a configurable
dose limit value (amount of fluoro and radiography)

** Option
** Mandatory in IEC countries 9
AXIOM Luminos dRF
Technical Data

19 Monochrome display DSB 1906


Diagonal screen size 19 (48 cm) display
Matrix 1280 x 1024 pixels
Brightness Standard, 450 cd/m2, max. 600 cd/m2, light sensor for automatic alteration,
standard image angle 80
Dimensions (housing) 464 mm x 425 mm x 113 mm (H x W x D), without attachment adapter
Power consumption 55 W
Weight 6.5 kg (incl. attachment adapter)

Display trolley*
Suitable for one or two monitors with a diagonal screen size of 46 cm each; two radiation ON indicators
Display tilt range + 15/ 10
Weight with two displays Approx. 50 kg (110 lbs)

Display ceiling suspension DCS-F*


Ceiling suspension system for one or two monitors with a diagonal screen size of 19; two radiation ON indicators
and buttons for electromechanical brakes
Length of longitudinal rails 425 cm
Travel range of ceiling-mounted carriage 325 cm
Vertical lift (height adjustment) 96.5 cm (for room height between 3.00 and 3.56 m)
Rotation range of the ceiling-mounted Maximum 300 10;
support to the rail axis for installation in increments of 30, infinitely adjustable
Rotation range of displays Maximum 330 10, for installation in increments of 30,
infinitely adjustable
Weight including rails With two 19 monitors: approx. 200 kg (441 lbs.)

10 * Option
OPTITOP 150/40/80 HC-100 x-ray tube
Nominal voltage 150 kV (IEC 613)
Focal spot 0.6 1.0 (IEC 60336)
Nominal power 40 kW 80 kW (IEC 60613)
Cold output 52 kW 103 kW
Anode angle 12
Anode heat storage capacity 580,000 J (783,000 HU) (IEC 613)
Anode drive 180 Hz/9,000 U/min (50 Hz line)
Total filtration 2.5 mm Al (IEC 60601-1-3)
Weight Approx. 26 kg

POLYDOROS F generator
POLYDOROS F HF generator with fluoro control and automatic exposure control
IONTOMAT PN
Power 65 kW (650 mA at 100 kV) or 80 kW* (800 mA at 100 kV)
according to IEC 60601-2-7
Exposure voltage 40 kV to 150 kV
Generator versions 65 kW with connection for 1 tube;
optional: 80 kW;
optional: Connection for 2 tubes

* Option 11
AXIOM Luminos dRF
Technical Data

Installation data
Line voltage connection to the rail axis 3/N/PE 400/440/480 V 10% at 50 Hz or 60 Hz; 440/480 V ( 10%)
optional line voltage connection is available
Internal line impedance according to planning guides
Power consumption (system) Max. 145 kVA for exposures with POLYDOROS 80
Room height Without ceiling-mounted support: min. 3.2 m
With ceiling-mounted support: min. 3.3 m
(no restriction of system movement);
from 2.5 m to 3.2 or 3.3 m with automatic movement restriction
through collision protection
Ambient conditions (operation) Temperature range: + 10C to + 35C
Relative humidity: 15% to 7% below dew point
Barometric pressure 700 kPa to 1060 kPa
Weight Patient table, complete: approx. 1,600 kg
Remote control console: approx. 5 kg
Generator cabinet: approx. 290 kg
FLUOROSPOT container: approx. 80 kg
Trolley-mounted operating console*: approx. 40 kg

12 * Option
Standard accessories
Description
R/F footswitch in the control room
Side hand rail (rear)
Removable footboard, height-adjustable (3 positions)
Handgrip, angled (front)
Shoulder supports (1 pair)
Head-end hand rail
Note: See Accessories for Fluoroscopy brochure for a complete listing of accessory options

13
AXIOM Luminos dRF
Technical Data

System expansion: Ceiling stand*


VERTIX Solitaire Ceiling stand without positive beam limitation
X-ray tube OPTITOP 150/40/80 HC-100
Focus 0.6 40 kW
Focus 1.0 80 kW
Anode heat storage capacity 580 kJ (783 kHU)

POLYDOROS IT80
High-frequency generator with automatic exposure control
Generator power 80 kW (650 mA at 100 kV) according to IEC 60601-2-7
Exposure voltage 40 kV to 150 kV
Shortest exposure time 1 ms
Vertical lift 150 cm (manual)
mAs area (mAs) 0.5 800 (depending on acquisition mode)
Travel range 354 cm longitudinal, 222 cm diagonal (manual) 1 cm
Tube rotation around vertical axis: + 154, 182, detents every 90
around horizontal axis: 120, detent at 0, 90
Collimator Full field light localizer, manually adjustable filters (0, 0.1, 0.2, 0.3 mm Cu),
45 rotatable with 0 detent
Collision guard Release for table movements only in park position of 3D TOP
Extension* For room heights over 3.3 m

Bucky wall stand: VERTIX PRO / TOP*


VERTIX PRO Bucky wall stand, manual or with ACSS*
together with the first flat x-ray tube at + 90 (= vertical system);
movable Bucky tray
VERTIX TOP Bucky wall stand, manual, tiltable* Buck tray* + 90; 20
Vertical lift 152 cm
Central beam height 38 cm to 190 cm 2 cm above floor level
Cassettes Cassettes according to DIH and ANSI standards in the following formats:
13 cm x 18 cm to 35 cm x 43 cm, portrait and landscape
Cassette loading Left or right (specify when ordering)
Grid Pb 13/70; f0 180 cm or f0 150 cm or f0 115 cm, reciprocating
Object-to-film distance 4.0 cm
Weight VERTIX PRO: approx. 160 kg;
VERTIX TOP: approx. 205 kg

14 * Option
Sample room layouts (mm)

Room layout with upper body radiation protection

5250
Room layout with VERTIX Solitaire 3700

5400

Room height 2.5 m


with restriction
Room height 3.2 m
without restriction

15
The information in this document contains Contact Address
general descriptions of the technical options Siemens AG, Medical Solutions
available and may not always apply in Angiography, Fluoroscopic and
individual cases. Radiographic Systems
The required features should therefore be Siemensstr. 1, D-91301 Forchheim
specified in each individual case at completion Germany
of contract. Telephone ++49 9191 18-0
Siemens reserves the right to modify the design
and specifications contained herein without
prior notice. Please contact your local Siemens
sales representative for the most current
information.
Original images always lose a certain amount
of detail when reproduced.
In the interest of complying with legal require-
ments concerning the environmental
compatibility of our products (protection of
natural resources and waste conservation),
we recycle certain components.
Using the same extensive quality assurance
measures as for factory-new components,
we guarantee the quality of these recycled
components.

08.2007 Siemens AG
Order No. A91AX-40004-11T2-7600
Printed in Germany
AX CRM MC 08072.0

Siemens AG Headquarters
Wittelsbacherplatz 2 Siemens AG, Medical Solutions
D-80333 Muenchen Henkestr. 127, D-91052 Erlangen
Germany Germany
Telephone ++49 9131 84-0
www.siemens.com/medical

www.siemens.com/medical

S-ar putea să vă placă și