Sunteți pe pagina 1din 7

I Review

Accessory Equipment Considerations


with Respect to MRI Compatibility
Elaine K. Keeler, PhD Francis X. Casey Hans Engels Elizabeth Lauder Catherine Anne Pirto
Theodore Reisker James Rogers Daniel J. Schaefer, PhD Thomas Tynes

The MR Section of The National Electrical ManufacturersAs- THIS DOCUMENT IS intended to address considera-
sociation (NEMA). in response to a request from the Food & tions with regard to the design and testing of third
Drug Administration (FDA),recently issued a position paper party equipment for use with a n MR system. It also
to address generic issues related to the compatibility of ac- provides information that may be of use to the end user
cessory equipment produced by third party equipment man-
ufacturers or MR equipment users and intended to be used
of that equipment. The information provided herein is
in conjunction with MR equipment or within the MR scan- generic with respect to MRI systems. Designers of third
ning room. The recommendations concern scanning acces- party equipment for use with MR systems, and users
sories, such as radiofrequency (RF) coils, patient monitoring of MR equipment building their own coils and acces-
equipment and injectors, as well as patient comfort acces- sories, should be aware that MR is a n evolving tech-
sories and positioning devices. The following issues related nology and specifications and performance
to safety performance are discussed (a)the interaction of characteristics of specific MR systems continually
the equipment with the MR scanner, (b) interactions of the change. Products from third party sources should be
MR scanner with the equipment, and (c)potential safety haz- tested under simulated or actual conditions of use. Be-
ards for patients and staff that can be posed by accessory cause upgrades to the MR system to achieve higher
equipment in the MR scan environment. The recommenda-
tions are based on combined input from NEMA member com- performance levels may affect the characteristics of the
panies who manufacture MR systems and MR accessories equipment, manufacturers of accessory devices are ad-
and are presented for consideration in the design of MR ac- vised that today's testing may not be sufficient to en-
cessory products and incorporation of these concepts into sure the safety and effectiveness of their device with
testing plans to ensure MR compatibility of third party de- tomorrow's MR system.
vices. Additional information with regard to system-specificre-
Index terms: MR compatibility - M K safety. MK equipment quirements should be obtained directly from the MR mar-
ufacturer.
JMRJ 1998: &:12-18
DeBnition of M R Compatibility
Abbreviations: CHT = cathode ray tube. DC = direct current. EMC = electromag- A device can be considered a s MR compatible if it can
netic compatil,ility. FDA = Fond & Drug Administration. LCD :. liquid crystal display. be used safely, a s indicated, within the MR scan room
LED = light r'mitting diode. NEMA = National Electncal Manufiicturers Association.
OE:M = original equipment m;iriulacturer. RF = radiolrequency. SAR = sperifir ab-
and during the scan procedure without adversely affect-
soiption rat<' T/R = transmitJrrceive. ing the performance of either the MR system or the device
itself. A device may be MR compatible for certain types of
uses but not for others. The device labeling must clearly
and explicitly define the conditions for safe and effective
use.

Scope
The document covers the following areas: (a)interac-
tion of the equipment with the MR system, (b)interaction
of the MR system with the equipment, (c) potential safety
hazards resulting from the use of the equipment in an
MR scanning environment, (d) special considerations for
surface coils, insert gradient coils, and cervical fixation
devices.
It provides guidance that is intended for use during the
initial design and subsequent installation of accessory
equipment obtained from third parties. Third party
equipment is that produced by someone other than an
original equipment manufacturer (OEM) of MR systems
From I he Maqnetic Resonance Section, Diagnostic Imaging and ThPrapy Systems Di- and intended for use a s an accessory in the MR environ-
vision. Na1ioii.il Electrical Manufacturers Association. 1300 Norlh 17th Street. Suite ment.
1847. K o s s l ~ nVA 22209. E-mail ekkQmr.pirker.com. Received September 19. 1996: Because the number of possible devices that could be
revision rrqiirsted May 20. 1997: accepted May 28. Presented a t the Workshop on
Advances in MR Safety and Compatibility. Macinan. VA. June $1-1 1. 1996. Address
useful accessories to MR scanners is very large and the
reprint requests t o E K.K. at Picker International. 595 Miner Road. Cleveland. OH number actually used in the MR environment grows
44143. daily, we have chosen those in most common use as ex-
ISMRM. l < l Y 4 amples to illustrate the variety of interactions that are of

12
concern. These include pulse oximeters, ECG monitors, through the patient interface connections (ECG leads or
respirators, anesthetic apparatus, contrast injectors, and pulse pickup) into the imaging area.
infusion pumps. It also includes scanning accessories To prevent this RF interference from degrading the MR
such as surface coils and patient positioning devices and image, appropriate measures should be used to provide
patient comfort systems such a s audio and video devices. RF shielding, such a s housing the display in a n RF tight
Although implanted metallic, electronic, or mechanical enclosure or locating the unit outside the RF room. In
devices a s well a s surgical intervention devices are not either case, the connections between the display unit and
explicitly included in this document, the principles stated the patient interface should pass through low pass filters
herein can be applied to such devices with appropriate that significantly attenuate (minimum of - 100 dB) fre-
attention being paid to the unique circumstances being quencies around the Larmor frequency of the MR system
set up by the devices proximity to specific tissue. (eg, 21.3 MHz for a 0.5T magnet, or 64 MHz for a 1.5-T
magnet). For patient contact devices, such a s ECG mon-
0 THE MR ENVIRONMENT itors, safety guidelines for patient isolation and leakage
The MR environment is the term used to describe the currents must be ensured per the requirements of IEC
general environment in the vicinity of the MR scanner (ie, 60 1- 1.2
Sensors and detectors such as those used in pulse
magnet). It includes the region from the center of the
oximetry equipment contain a n additional interference
magnet bore out to the 5-G line (.0005 T). Beyond the 5-G
line. the levels of magnetic field are generally considered source. Pulse oximeters measure the absorption ratio of
to be safe, although some effect may still be produced a red LED and infrared LED to determine the patients
on devices such as cathode ray tube (CRT) displays and pulse and blood saturated oxygen content. Units that cy-
image intensifiers. cle the LEDS on and off at a rate of several kilohertz can
be a source of radiated RF interference, which can be
picked up easily by the MR receive coil and result in de-
Characteristics of the M R Environment graded image quality, because the sensor is located very
The MR environment is characterized by the following: near the receiving coil of the MR system. Shielding of the
1. The static magnetic field (most commonly in the range sensor is recommended to prevent this noise from reach-
of .2to 2.0 T, although field strengths a s low a s .06 T ing the MR system.
and greater than 4 T are also in use) and the associ- Similar RF considerations should be made in the selec-
ated spatial field gradient. tion of patient audio and video equipment. The tape deck,
2. The rapidly changing magnetic fields (gradients)that tuner, or CD player for a patient audio system usually is
are used during imaging (- 1 Hz). located near the operators console and a speaker or
3. Radiofrequency (RF) pulses (on the order of tens to headphone set is located within the scan room. MR man-
hundreds of MHz). ufacturers require that signals between the audio equip-
ment and the speaker must penetrate the RF room
0 INTERACTION OF EQUIPMENT WITH THE MR through appropriate filters or make use of a nonelectri-
SYSTEM cally conducting medium, eg, fiber optics or infrared
External equipment can interact with the MR system transmission, to prevent RF interference outside the
either through the generation of R F interference or dis- room from being conducted into the MRI environment.
turbance of the homogeneity of the magnetic field. Either Specific information about these requirements can be ob-
of these may adversely affect scanner performance and tained from the Siting Guides, available from most OEM
reduce its diagnostic effectiveness. MR system manufacturers.
Simple solutions may not be available with video equip-
RF Interference Considerations ment such as patient TV. As with audio equipment, con-
In general, third party equipment should be designed trol signals must be filtered at the RF room wall.
to meet the requirements of IEC 601-1-2. This standard Additionally, consideration must be given to the interfer-
defines field levels for RF emissions and electromagnetic ing frequencies that could be generated by the TV display
compatibility (EMC).Related standards and tests can be itself. If a flat panel display is used, it should be properly
found in CISPR- 11 and IEC 80 1- 1 through 4. It should enclosed and RF should be shielded to prevent interfer-
be noted that none of the current standards are specifi- ence with the MR system.
cally concerned with MR compatibility and additional
steps actually are required. Some general guidance re- Magnetic Field Considerations
lated to design practices for MRI accessories is given be- Significant masses of ferromagnetic material in prox-
low. imity to the magnet can disturb the homogeneity of the
If RF shielding is inadequate, equipment that includes static MR magnetic field. Distorted images could result if
display screens and/or microprocessors, such a s patient the homogeneity of the main magnetic field is degraded
monitoring equipment, may produce RF interference that significantly (1). Respirators and anesthesia equipment,
affects image quality. Interference sources include micro- carts, or gas cylinders should be designed to use nonfer-
processor clocks and displays that can generate a wide romagnetic materials. Patient monitoring equipment lo-
range of frequencies. Light emitting diode (LED), liquid cated within the scan room also could have an effect on
crystal display (LCD),or CRT displays and indicators may the magnetic field homogeneity. Power supply transform-
appear visibly steady but are, in reality, being electrically ers or the frame of the unit could contain sufficient mag-
turned on and off at a high frequency a s a requirement netic material to result in a distortion of the field.
of the display technology or to conserve power. The re- Other conductive, but nonferromagnetic, materials
sulting clock and display interference can be radiated di- may exhibit induced RF currents that can lead to image
rectly into the RF room environment or conducted distortion. Artifacts caused by nonferromagnetic materi-

These documents can be obtained from the ANSI Sales Office. 1 I West 42nd Street, This document can be obtained from the ANSI Sales Oflice. 1 1 West 42nd Strert
New York. NY 10036. (212) 642-4900 New York. NY 10036. (212) 642-4900

Volume 8 Number 1 JMRl 13


als result from eddy currents that can be generated in Accuracy of analog style gauges may be affected by the
the objects by gradient magnetic fields used for MRI that, MR magnet fringe field because the movement mecha-
in turn, disrupt the local magnetic field and distort the nism in some types of gauges makes use of a small mag-
image (2.3). net and a coil for operation. Digital gauges are unlikely
to be affected by the magnetic field, but a potential FW
0 INTERACTION OF MR SYSTEM WITH noise source may arise from the numeric displays. The
EQUIPMENT effect of the magnetic field on magnetic media, such as
Operation, reliability, and accuracy of accessory equip- digital tape or floppy disks, may cause damage or erasure
ment, if not designed correctly, can be affected by the MR of data, so devices used in the magnet room should not
system. These could have serious effects on the safety incorporate these types of media for operation or storage
and effectiveness of the device. Potential problems that of patient data.
must be addressed are interactions of the static magnetic Manufacturers of third party equipment should have
field, FW fields, and gradient fields with the accessory appropriate test data to demonstrate routine operation of
equipment. their equipment without compromising its effectiveness
at the maximum magnetic field level within which it may
Static Field Effects be placed. For equipment that is located immediately at
Magnets used for MR systems in routine clinical oper- the periphery of the magnet covers, the magnetic field
ation range in field strength from a s low as .06 T (600G) level can be as high a s (or in some cases higher than) the
to 2.0 T (20, 000 G). A number of scanners currently in central field. Within the magnet room, field levels can
clinical investigation operate in the range of 3 to 4.6 T. range from <.5mT to full system field strength. It is rec-
Superconductive, cryogenic magnets (>.5T) can only be ommended that equipment intended for use within the
turned off through a controlled ramp down or by a n magnet room be clearly labeled with the maximum field
expensive quench process. Lower field-resistive mag- level within it which it can operate effectively. Instruction
nets are easily powered down, whereas low field perma- must be provided regarding the effects on the equipment
nent magnets are always turned on. I t should be noted of exceeding that field limit. Critical devices, the failure
that small MR systems, designed for specific applica- of which could cause serious injury to a patient, must be
tions, such as extremity imaging, may have very well con- secured or permanently located in a safe position for op-
tained magnetic fields (eg, 5 G is within 6 inches of the eration.
magnet). These particular systems do not represent a
worst case, and the discussion that follows may not ap-
ply. The inherent differences in magnet types must be R F Field Effects
well understood before any MR compatible equipment is Physiologic measurements, using electrical transduc-
designed. These facts are basic to understanding how the ers or equipment, for example patient ECG and pulse ox-
safety and effectiveness of a device can be influenced by imetry equipment, may be affected by the RF transmit
the static field environment produced by the MR system. field. Because the sense leads for this equipment are sit-
Mention must be made of the serious hazard caused by uated within the imaging environment, R F energy from
the potential missile effect that exists when ferromag- the transmit RF field can easily couple into the leads and
netic materials are brought near the magnet. The strong the input amplifiers of the equipment causing either sat-
magnetic field generated by the MR system extends uration of the amplifier, which could result in momentary
throughout the scan room and, in some cases, even be- loss of the monitored signal, or permanent damage if ad-
yond. The attractive forces increase rapidly as one nears equate protection is not provided.
the magnet, producing a force that can displace even The amount of FW interference which is coupled into
large objects and pull them to the magnet. Objects such the leads can be controlled by appropriate design meas-
as oxygen bottles or equipment carts, when pulled to the ures (for example, by selecting materials that are used for
magnet, can produce costly damage to the MR equipment the ECG leads and by incorporating low pass filtering on
as well a s serious injury to patients and/or workers. De- the input of the amplifier circuitry).
vices should be scrutinized carefully to remove ferromag-
netic parts. These may work loose over time and become Gradient Field Effects
a serious problem. Equipment with cables or leads in the magnet bore or
Ferromagnetic objects within the magnetic field are attached to the patient, such a s ECG equipment, also can
also subject to forces that try to bring them into align- be affected by the gradient magnetic fields used in im-
ment with the magnetic field. These strong torque effects aging. When the ECG leads are connected to the patient,
can cause metallic objects to twist. For this reason, MRI there is a n electrical loop created by the leads, the pa-
is contraindicated for a patient with a n implanted ferro- tient, and the ECG conditioning amplifier. The pulsing
magnetic aneurysm clip or similar implant. Device com- gradient fields generate a circulating current within this
ponents subject to torque effects can break loose, loop, resulting in a voltage across the conditioning am-
causing the device to fail or possibly injure a patient or plifier. This voltage can be several times the magnitude
worker. of the ECG R wave and can result in saturation of the
Respirators, infusion pumps, contrast injectors, and ECG conditioning amplifier, incorrect triggering of the
anesthesia equipment may contain pumps, electric mo- ECG equipment, or both.
tors, or gauges that can be directly affected by the mag-
netic field. The magnets in motors may become saturated
by the fringe field of the MR magnet. Saturation can ei- 0 POTENTIAL PATIENT SAFETY HAZARDS
ther result in lack of motor operation, a slowing of the Accessory equipment design should take into account
operation, or a n increase in operating current of the mo- potential patient safety hazards. These include magnetic
tor, which could ultimately b u m the motor out. Similarly, components being attracted to the MR magnet, changes in
electronic circuitry may contain components that are af- ground leakage currents, and possible patient bums
fected by static magnetic fields, such a s transformers, caused by the heating of material placed within the im-
femte cores, relays, and switches. aging area.

14 JMRl JanuarylFebruary 1998


Leakage Currents terials may produce a serious hazard. Equipment in-
Consideration must be made when using filters to sup- tended for placement close to the MR magnet should
press RF interference from ECG and pulse oximetry mon- contain a minimum of ferromagnetic material. Manufac-
itoring equipment. The input amplifier for this type of turers should understand the need to minimize or elim-
equipment should, by design, be isolated from earth inate the use of magnetic material in their products to
ground to prevent the flow of unacceptable levels of 60- avoid equipment being attracted to the magnet. The force
Hz currents within the patient. If RF filters are placed can be significant and could result in injury of personnel
directly in series with the ECG leads, there is the poten- trying to restrain the equipment or being trapped be-
tial for increasing the 60-Hz currents above acceptable tween the equipment and the magnet. Additional precau-
levels because the RF filters are referenced to the RF tions should be taken during the course of product
room and power supply grounds. To prevent this prob- design to prevent the inclusion of small components that
lem, a separate battery or isolated powered preamplifier could be pulled loose and attracted to the magnet bore.
should be located within the RF room for direct connec- For example, connector covers that are ferromagnetic
tion to the ECG leads. The output signals can then be must be attached to the equipment by a chain. Remova-
filtered safely when exiting the R F room. ble equipment covers should use captive hardware
(screws and fasteners). Battery-powered equipment may
Heating seem to be nonmagnetic, but it is very important to use
ECG leads and pulse pickups placed within the RF batteries that are totally nonmagnetic. Battery-powered
transmit field can provide a path for flow of significant equipment should be tested at its intended location and
levels of RF currents through the patient or along the maximum field strength to ensure that there is no sig-
shield of the cable to ground. If this RF current is not nificant attraction.
properly limited, patient b u m s can result at the ECG
electrode location or a t body parts that are located next 0 SPECIAL PURPOSE RF COILS
to a n insufficiently insulated RF shield. The method of The design of special purpose coils and associated pa-
cable routing within the bore of the magnet can be im- tient positioning devices for use in MRI requires special
portant in reducing these effects. In addition, use of con- attention. Guidance, based on the collective experience
ductive leads can be replaced by alternate technology, of MR manufacturers, is given below to allow avoidance
such a s fiber optic, carbon fiber, or high impedance of safety, image quality, and reliability issues (private
leads, minimizing the possibility of bums. communications). It is hoped that this will be helpful for
Equipment that is of primary concern for RF heating is those producing RF coils for their own use (research) or
that which can be positioned directly in the RF transmit as after-market MR accessories.
field. A device must not have the potential to become hot
enough to cause injury during scanning. This includes Magnetic Effects
ECG electrodes and leads, pulse pickup and pulse ox- All components that are used in RF coils should be pre-
imetry sensors, as well a s metallic components, connect- screened to ensure that they do not affect image quality.
ors, cables, and surface coils (4-8). Electronic components should be specified a s nonmag-
There is very little research available indicating accept- netic. However, care must be taken because, apparently,
able levels of RF currents, which can safely flow within a even nonmagnetic components may have some magnetic
patient at MRI frequencies. Measurements made during content, for example, housings and leads with nickel
electrosurgery indicate that RF current flow within the flash. Brass bolts, nuts, and screws used to fasten the
ECG leads must be limited to less than 100 mA to prevent coil onto the housing can contain up to a 1% contami-
RF bums of the patient (9).To provide the same level of nation with iron or nickel. The associated magnetic dis-
current limitation during MRI, it has been found that tortions are most severe in gradient-recalled echoes with
graphite ECG leads must be used during MR studies. Ad- a high flip angle. By affixing sample components to the
ditionally, it also may be prudent to locate supplementary exterior of a n imaging phantom and running such a pulse
RF impedance in the leads near their connection to the sequence, penetration of any artifacts into the imaging
ECG electrode pads. Metal conductor ECG lead wires are region of interest can be determined. A simple procedure
definitely unacceptable for this purpose. for evaluating image distortion can be found in National
Pulse pickup and pulse oximetq sensors provide two Electrical Manufacturers Association (NEMA)Standards
possible areas of concern. The first results from the ground- Publication No. M S 2, Determination of Two-Dimensional
ing of the shield incorporated into the sensor cable. If this GeometricDistortion in Diagnostic MagneticResonancelm-
cable is placed too close to the RF transmit field, significant ages." An extensive discussion of the techniques for eval-
RF currents can flow down the shield. The RF currents can uation of various materials with respect to the distortion
result in thermal heating of the cable and body parts lo- of MR images has been done by Schenck (1). The per-
cated next to the shield and may result in bums. formance specification for image distortion should be ob-
A second pulse pickup consideration is coupling of the tained from the MR manufacturer.
transmitted RF field with the LEDs and photo detectors
used in pulse sensors. It is possible for the LEDs and
photo diodes to rectify the transmitted RF and dissipate Proton Signal
a large amount of power in a short period of time. This Special purpose coils are normally housed in plastic
can result in localized heating of the sensor and has been packaging. Materials used range from Plexiglas to ure-
reported to cause bums in some cases. Not only must thane. Care must be taken to ensure that the materials,
shielding be provided to prevent noise generated by the including adhesives, do not generate a n MR signal, be-
sensor from degrading the image but also to prevent cou- cause, depending on field of view and phase-encoding di-
pling of the RF transmit field with the sensor. rection, this signal may end up in the region of interest
and produce image artifacts. Signal arises from protons
Magnetic Components
As previously stated, forces causing displacement
("missile effect") or torque effects on ferromagnetic ma- i This standard can be obtained from NEMA Customer Semces. 1301) 841-3200.

Volume 8 Number 1 JMRl - 15


in the plastic molecules. Although the signal decay, char- Care must be taken with all passively decoupled coils to
acterized by T2 relaxation time, usually is very short, ensure that components are correctly rated to withstand
with progress in MR pulse sequence development, TEs RF currents induced from the body coil RF field. For ex-
have become shorter and shorter and signal from the coil ample, if the diode chosen cannot handle the current in
housing may be visible in the image. The same is true of the decoupling circuit during the transmit pulse, the diode
molded foam pads that are used for patient positioning. will blow and may cause an open circuit. The result will
Any materials that are to be used should be evaluated be that the coil is no longer decoupled from the body coil,
with the shortest TE pulse sequence used in clinical prac- potentially affecting performance of both coils. If an induc-
tice to determine whether signal from the material could tor in the tank circuit does not have the proper gauge wire
cause image problems. to handle the current flowing through it, it may become
very hot, in which case the housing of the coil can melt or
Gradient Eddy Currents even catch fire. In the extreme case, the inductor will un-
Every coil or cervical fixation device that contains a solder itself, creating coupling with the body coil.
Special purpose coils must be designed to work with
continuous loop or copper plane that is not interrupted,
pulse sequences with low flip angles. If not correctly de-
for example, by capacitors, will create gradient eddy cur-
rents, adversely affecting the rise and fall times of the signed, the decoupling circuits may not operate because
gradient pulses, thus potentially degrading image qual- the associated transmit power may be insufficient to bias
the diodes. Slice profile also may be affected because the
ity.
amplitude of the side lobes of the RF pulse is insufficient
to bias the decoupling diodes.
Cable Resonances Care must be taken with the design of actively decou-
Special purpose coils have a cable that is grounded pled coils to ensure protection of devices if the surface
where the cable connects to the MR equipment. Interac- coil is present inside the body coil but is not plugged in.
tion with the RF electric field of the body coil can create In this case, the diodes are not DC biased. The conse-
high voltages on the cable shield. If these are not properly quences depend on the type of diode used, but diode fail-
taken care of via balanced/unbalanced circuits and a de- ure is a potential outcome.
vice that prevents the cable from looping, the signal-to-
noise ratio of the surface coil can be adversely affected 0 INSERT GRADIENT COILS
and, in case of a cable loop, arcing can occur between the Removable gradient coils, designed to be inserted into
cables, which may be dangerous to the patient. the magnet bore, can be used to produce very steep gra-
dient magnetic fields for rapid imaging techniques. Re-
System Concerns search groups or manufacturers designing these
Improper surface coil design can affect the performance accessory devices may find the following information
and reliability of the MR system. The output channel of helpful.
special purpose coils should be protected against over-
voltage with a diode network or damage to the input stage Torque
of the system preamp may result. A gradient coil may or may not be torque compensated.
Transmit/receive (T/R) coils exhibit mutual coupling In a torque-compensated coil, torque on one side of the
between the body coil and the T/R coil, which is a func- coil is of equal but opposite sign to that on the other side
tion of relative size and proximity. T/R special purpose of the coil. An uncompensated coil in a DC magnetic field
coils do not contain any decoupling and, therefore, mu- will rotate when currents are applied to the transverse
tual coupling may be severe, in which case the body coil gradients (xand y for a DC field in the z direction) and
may become detuned and mismatched, causing power re- will tend to shift when current is applied to the parallel
flection from the body coil to its interface circuitry. De- gradient (z gradient). A compensated coil may do the
pending on the amount of power, the transmit/receive same if a fault condition occurs where the coil is damaged
switch in the coil interface may be damaged. internally (eg, a short). The resulting torque and forces
T/R special purpose coils not designed and tested in depend on the amplitude of the gradient pulse and/or the
conjunction with particular MR systems may cause the gradient field strength and on the amplitude of the DC
power optimization phase of the scan to fail, resulting in magnetic field. It should be noted that torque also can
incorrect setting of the flip angle. There is particular con- occur when the gradient coil is located outside of the sys-
cern with T/R coils of small dimensions compared with tem, where the magnetic field is nonuniform and nonlin-
the system standard coils. For example, the system may ear. Although there have been no reported incidents of
optimize for a 270" flip angle or even a 450" flip instead patient injury related to the use of insert gradient coils,
of 90". With the incorrect flip angle, image quality (con- the torque and forces can be of sufficient intensity to cre-
trast and uniformity) will be affected. In addition, esti- ate a safety concern. For this reason, all insert gradient
mation of specific absorption rate (SAR) levels for RF coils should be surrounded with a frame such that the
exposure may be incorrect, a potential safety hazard that coil is essentially locked in place. The frame may be de-
may result in localized tissue heating. T/R special purpose signed to withstand the high forces and torque created.
coils should contain an attenuator or other means for en- It also should be capable of withstanding high accelera-
suring that the system flip angle is correctly set. tion forces because the torque usually is created within
Receive-only special purpose coils are decoupled from 100 ps, depending on the rise time of the gradient.
the body coil during the transmit pulse via a n internal Again, severe torques also will occur if the operator
electronic circuit. Electrically decoupled coils use either uses the coil in any area outside the high homogeneity
passive decoupling, in which the decoupling is achieved sphere of the magnet (ie, in an area where the DC mag-
via a parallel tank that contains fast switching diodes netic field is inhomogeneous). This situation should be
switched by the induced RF voltage during the transmit prevented by mechanical or electrical (microswitch)
pulse, or active decoupling, in which the diodes are ac- means so that the coil can only be used in the isocenter
tively biased via a direct current (DC) control pulse. of the magnet.

16 JMRl January/February 1998


Acoustic Noise Because insert gradients usually are much farther
An insert gradient coil is much closer to the ears of the away from the cryostat screen, the eddy currents may be
patient than a whole body gradient coil. For this reason, negligible. Therefore, upon insertion of the local gradient,
ear protection should be required for the patient. This the eddy current compensation may be changed via hard-
should be indicated on the insert coil housing in clear ware or software means, if required.
view of the operator (10-17). The insert gradient coil will typically present a different
load to the amplifier than the body gradient coil and,
thus, a load-compensating network may be required
dB/d t
when switching between the whole body and insert gra-
Because of the low inductances and fast rise times,
dient coils. Failure to provide proper compensation may
small insert gradient coils may produce dB/dt levels that
exceed recommended safety limits. Insert gradient man- result in damaged coils or gradient amplifiers.
ufacturers should calculate the maximum dB/dt values
and limit a patients exposure either by software (limiting Magnetism/Proton Signal
the gradient waveforms allowed peak and slew values) or No magnetic parts should be included in any device
by hardware (limiting the gradient amplifiers output).Al- that is inserted in the bore of the magnet, not only for
ternatively, the housing of the gradient coil could be de- safety reasons but also to maintain a highly homogene-
signed so that sufficient distance exists between the ous DC magnetic field.
localized dB/dt peaks at the coils edge and the patient. Gradient coil material in the vicinity of the RF coil
should not give off any proton signal. This signal could
fold back into the image and be mistaken for pathology.
Heating
Insert gradient coils often become quite hot because of
the high current densities. If the temperature becomes CERVICAL FIXATION DEVICES
too high, the gradient coil material may lose its physical Some special concerns are presented by cervical fixa-
strength, which again is potentially dangerous due to the tion devices or halos. The design of certain types of ha-
high torques and forces. These limitations should be well los include a continuous metallic ring with no
understood and the data provided by the coil manufac- interruptions of a capacitive nature. In this case, there is
turer should describe operating conditions that cause the a n interaction with the RF coils and with the gradient
temperature of the coil to rise above 40C. In addition, coils. The halo will couple with the RF field from transmit
there are limits to the temperature of parts that can be a s well a s receive coils. Although the ring is not resonant,
in contact with the patient. RF currents will be induced during the transmit pulse
Because of these dangers, methods of eliminating haz- but with a lower amplitude than in a resonant ring. De-
ards due to excess heating must be provided, either by pending on the amplitude of the mutual inductive cou-
providing a cooling mechanism (air, water, or both) or by pling, these currents can create RF magnetic fields that
providing hardware feedback to the MRI system that lim- locally exceed SAR limits in the patients tissue. The halo
its the output of the gradient amplifiers to safe levels. will couple to the gradient coil and adversely affect the
In either case, thermocouple or heat-sensitive snap rise time of the gradient pulses that will lead to phase
switches should be mounted at critical spots of the gra- errors in the image. Each halo device should be evaluated
dient coil so the gradient amplifiers may be switched off carefully to determine whether diagnostic images can be
when a certain temperature is exceeded. produced despite these imaging artifacts. Although there
The material being used for the actual gradient coil is no published evidence to document this, there is con-
should be of a high flame rating (UL-5V) and should cern that if metallic screws are used to secure the halo
maintain its strength at high temperatures. to the patients skull, a possibility exists that the screws
may heat up during scanning due to a buildup of F W cur-
rent density and may cause injury to the patient. Tests
Interface should be designed and performed to verify that heating
The gradient coil should be connected to the system via during scans does not occur (18-20).
cables that can handle the applied current amplitudes
and duty cycles. Cables for forward and return currents
should be self canceled or secured such that the cables CONCLUSION
do not move in the magnetic field upon application of a It is the hope of NEMA member companies that this
gradient pulse. document will be helpful in the design and selection of
A connector interface box should be mounted on the accessory MR equipment. MR is a continuously evolving
system with different connectors for the body gradient and technology with new developments and clinical applica-
the insert gradient. This box should be interlocked in such tions being introduced at a rapid rate. We anticipate that
a way that no one can touch live parts without switching this document will be updated periodically to keep pace
off the gradient amplifiers first. The connectors should with this rate of development.
only fit one way, such that reversing the gradient field is
impossible. References
1. Schenck JF. Review article: role of magnetic susceptibility in
Eddy Currents/Load Compensation MRI. J Med Phys 1996: 23(6);815-850.
When using a n unshielded body gradient coil, eddy 2. Bellon EM, Haacke EM, Coleman PE, el al. MR artifacts: a re-
view. Magn Reson Imaging 1984; 2:41-52.
currents are created in the cryostat screen and other 3. Pusey E. Lukin RB, Brown RKJ, et al. Magnetic resonance
nearby metal parts. These eddy currents increase the rise imaging artifacts: mechanism and clinical significance. Radi-
time of the coil and have undesired effects on the phase ographics 1986: 6:891-911.
of the NMR signal. An eddy current compensation circuit 4. Fisher DM. Lilt W, Cote CJ. Use of oximetry during MR imag-
ing of pediatric patients. Radiology 199 1: 178:891-892.
in the gradient amplifier typically is used to correct for 5. Kana1 E, Shellock FG. Patient monitoring during clinical MR
this effect. imaging. Radiology 1992: 185:623-629.

Volume 8 Number 1 JMRl 17


6. Shellock FG, Myers SM, Kimble K. Monitoring heart rate and 8. Food & Drug Administration. Guidance for content and review of a
oxygen saturation during MRI with a fiber-optic pulse oximeter. magnetic resonance diagnostic device 5 10(k) application: safety
Am J Roentgenol 1991; 158:663-664. parameter action levels. Rockville. MD: Center for Devlces and Radi-
7. Shellock FG. Slimp G. Severe b u m of the finger caused by us- ological Health Report, 1988.
ing a pulse oximeter during MRl (letter).Am J Roentgenol 1989: 9. Kagetsu N J . Litt AW. Important considerations in measurement of
153:1105. attractive force on metallic implants in MR imagers. Radiolo~g1991;
8. Shellock FG. Kanal E. B u m s associated with the use of mon- 179:505-508.
itoring equipment during MR procedures. JMRl 1995: 6 2 7 1 . 10. Hubbard A. Markowitz R, Kimmel B. Kroger M. Bartko M. Sedation
9. DeRosd J F , Gadsby PD. Radio-frequency heating under RF for pediatric patients undergoing CT and MRI. J Comput Assist Tom-
electrodes. Med Instrum 1979; 13(5):273-276. ogr 1992; 16:333-336.
10. Brummett RE, Talbot JM, Charuhas P. Potential hearing loss 11. Kanal E, Shellock FG. MR imaging of patients with intracranial an-
resulting from MR imaging. Radiology 1988; 169:539-540. eurysm clips. Radiology 1993: 187:612-614.
11. Goldman AM, Gossman WE, Friedllander PC. Reduction of 12. Kanal E. An overview of electromagnetic safety considerations as-
sound levels with antinoise in MR imaging. Radiology 1989: sociated with magnetic resonance imaging. Ann N Y Acad Sci 1992;
1731549-550. 649:204-224.
12. Hurwitz R, Lane SR. BeU RA, Brant-Zawadzh MN. Acoustic 13. Lewin JS. Duerk JL, Jain VR. et al. Needle localization in MR-
analysis of gradient-coil noise in MR imaging. Radiology 1989: guided biopsy and aspiration: effects of field strength, sequence de-
173:545-548. sign, and magnetic field orientation. Am J Roentgenol 1996: 166:
13. Shellock FG. Morisoli SM, Ziarti M. Measurement of acoustic 1337-1345.
noise during MR imagmg: evaluation of six Worst case pulse se- 14. Lufiin R, Jordan S. Lylyck P. et al. MR imaging with topographic
quences. Radiology 1994; 191:91-93. EEG electrod% in place. Am J Neuroradiol 1988: 9:953-954.
14. McJury M, Blug A, Joerger C, Condon B, Wyper D. Acoustic 15. McGowan JE. Erenberg A. Mechanical ventilation of the neonate
noise levels during magnetic resonance imaging scanning a t during magnetic resonance imaging. Magn Reson Imaging 1989: 7:
1.5T. f3r J Radiol 1994: 67:413-415. 145- 148.
15. McJury M J . Acoustic noise levels generated during high field 16. Moscatel M, Shellock FG. Monsoli S. Biopsy needles and devices:
MR imaging. Clin Radiol 1995: 30:331-334. assessment of ferromagnetism and artifacts during exposure to a 1.5
16. Miller LE, Keller AM. Regulation of occupational noise. In: Har- Tesla RM system. JMRI 1995: 5:369-372.
ris CM, ed. Handbook of noise control. New York: McGraw-Hill, 17. Person B. ed. Potential health hazards and safety aspects of clinical
1979: 2. NMR examinations. Semin Biomed Appl NMR. Lund Radiation Physics
17. Melnick W. Hearing loss from noise exposure. In: H a m s CM, Department, 1984.
ed. Handbook of noise control. New York: McGraw-Hill, 1979: 18. Reilly JP. Cardiac sensitivity to electncal stimulation. Report MT.
1-16. Washington, DC: Food & Drug Administration, 1989: 89-101.
18. Ballock FC. Hajed PC, Byme TP, et al. The quality of magnetic 19. Reilly JP. Peripheral nerve stimulation by induced electric currents:
resonance imaging, as affected by the composition of the halo exposure to time-varying magnetic fields. Med Biol Eng Comput
orthosis. J Bone Joint Surg 1989: 71A:431-443. 1989; 27:lOl-110.
19. Clayman DA, Murakami ME, Vines FS. Compatibility of cer- 20. Schaefer D J . Safety aspects of magnetic resonance imaging. In:
vical spine braces with MR imaging: a study of nine nonferrous Wehrli FW,Shaw D. Kneeland JB, eds. Biomedical magnetic reso-
devices. Am J Neuroradiol 1990: 11:385390. nance imaging 1988: 13:553-578.
20. Shellock FG, Slimp G. Halo vest for cervical spine fixation dur- 21. Schaefer DJ. Dosimetry and effects of MR exposure to RF and
ing MR imaging. Am J Roentgenol 1990; 154:631-632. switched magnetic fields. Ann N Y Acad Sci 1992: 649225-236.
22. Selden H. DeChateau P. Ekman G, et al. Circulatory monitoring of
children during anesthesia in low-field magnetic resonance imaging.
Acta Anesthesiol 1990: 34:41-43.
Suggested Readings 23. Shellock F, Kanal E. Magnetic resonance: bioeffects, safety and pa-
1. Occupational Safety and Health Administration. Occupationalnoise tient management. New York: Raven Press, 1994.
exposure. 29CFR Part 1910.95. Washington, DC: Occupational 24. Shellock FG, Kanal E. Magnetic resonance: bioeffects. safety. and
Safety and Health Administration, Department of Labor, 1997. patient management, 2nd ed. New York: Raven Press, 1996.
2. Adair ER. Berglund LG. On the thermoregulatoly consequences of 25. Shellock FG. Pocket guide to MR procedures and metallic objects:
NMR imaging. Magn Reson Imaging 1986: (4):321-333. update 1994. New York Raven Press, 1995.
3. Athey TW. Current FDA guidance for MR patient exposure and con- 26. National Electrical Manufacturers Association. Determination of
siderations for the future. Ann N Y Acad Sci 1992: 649242-257. two-dimensional geometric distortion in diagnostic magnetic reso-
4. Barnett GH, Roper MID. Johnson AK. Physiological support and nance images. Standards Publication No. MS 2. Rosslyn. VA:
monitoring of critically ill patients during magnetic resonance imag- National Electrical Manufacturers Association. 1990.
ing. J Neurosurg 1988: 68:246-250. 27. National Electrical Manufacturers Association. Characterization of
5. Boesigner P. Buchli R. Saner M. Meier D. An overview of electro- special purpose coils for diagnostic magnetic resonance images.
magnetic safety considerations associated with magnetic resonance Standards Publication No. M S 6. Rosslyn, VA: National Electrical
imaging. Ann N Y Acad Sci 1992: 649: 160-165. Manufacturers Association, 1991.
6. Budinger TF. Nuclear magnetic resonance technologies:health and 28. National Electrical Manufacturers Association. Characterization of
safety. Ann N Y Acad Sci 1992: 649:l-19. the specific absorption rate for magnetic resonance imaging systems.
7. Dunn V~ Coffman CE. McGowan JE. Ehrardt JC. Mechanical ven- Standards Publication/No. M S 8. Rosslyn. VA: National Electrical
tilation during magnetic resonance imaging. Magn Reson Imaging Manufacturers Association, 1993.
1985; 3:169-172. 29. Villa M, Mustarelli P. Caprotti M. Mini review: biological effects of
magnetic fields. Life Sci 1991: 49:85-92.

18 JMRl - January/February 1998

S-ar putea să vă placă și