Certipur Reference Material

Platinum Cobalt Color Reference Solution (HAZEN 500)


According to DIN EN ISO 7887, ASTM D 1209-05 and DIN EN ISO 6271-1

1.00246.0250 Lot No.: sample

Composition:
Potassium hexachloroplatinate and Cobalt chloride hexahydrate in hydrochloric acid.
The solution was manufactured gravimetrically under controlled environmental
conditions. The manufacturing, testing and distribution of this reference material meet
the requirements of the quality management norm DIN EN ISO 9001:2008.

Material Handling and Storage:


Use immediately after decanting and discard after use. Store at 15 25 C. Protect
from light.

Specification of the absorbance tolerance limits according to ASTM D 1209-05:

Wavelength [nm] Absorbance Absorbance


(Accuracy 0.2 nm) (tolerance) of this standard
430 0.110 to 0.120
455 0.130 to 0.145
480 0.105 to 0.120
510 0.055 to 0.065

Path length: 10 0.01 mm


Type of cell: Rectangular Quartz cell, 10 0.01 mm
Reference: empty beam path
Baseline: deionised water
Measuring temp: 20 1 C
Slit width: 1 nm 0.2 nm

Measurement Uncertainty:
The reported uncertainty represents the expanded uncertainty (U = ku) for a
coverage probability of 95% (k=2).
Traceability and instrument qualification:
The HAZEN 500 Standard is traceable to NIST SRM 930e and has been analysed on
a high-performance UV/VIS/NIR spectrophotometer within the spectral range 190 -
800 nm (monochromatic). The spectrophotometer is regularly checked for accuracy
and reproducibility of absorbance and wavelength as well as for linearity, baseline
drift, stray light and spectral resolution power. The relevant parameters for this
standard are checked on the day of lot certification.

Absorbance: NIST SRM 930 + 2930 or SRM 1930 and double aperture method
Merck Cat. No. 108166 (holmium oxide solution), emission lines of
Wavelength:
D2-, Hg- and Ar lamps
Merck Cat. No. 108164 (KCl standard, 200 nm), Merck Cat. No.
Stray light: 108163 (NaI standard, 220 nm) and Merck Cat. No. 108161
(NaNO2 standard, 370 nm)
Spectral
Half width value of D2 emission lines for checking the effective
resolution
spectral slit width.
power:

Date of release:

Minimum shelf life: 18 month

......................................
(responsible laboratory manager quality control)