Documente Academic
Documente Profesional
Documente Cultură
Service manual
Service Manual -
Philips MicroDose L30
Software 8.4
1007989 O 2 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preventive maintenance
Service prerequisites and material
Disassembling instructions
Maintenance and inspection
System adjustments
Service tool
Quality control
1007989 O 3 (322)
Service Manual Philips MicroDose L30, Software 8.4
Contents
1.
Introduction ....................................................................................................................... 9
1.1
Purpose ..................................................................................................................... 9
1.2
Scope ........................................................................................................................ 9
1.3
Intended readers ........................................................................................................ 9
1.4
Chapter overview ...................................................................................................... 9
1.5
References ................................................................................................................ 9
1.6
Symbols used in the document ............................................................................... 10
1.7
Service and maintenance ........................................................................................ 10
1.8
Definitions, acronyms and abbreviations ............................................................... 11
2.
System overview .............................................................................................................. 12
2.1
Mammography stand .............................................................................................. 12
2.2
Side cabinet ............................................................................................................. 13
2.3
Acquisition workstation .......................................................................................... 14
3.
Overview of modality sub-systems ................................................................................ 15
3.1
Overview and functionality of sub-systems ........................................................... 15
3.2
System mains power overview ............................................................................... 18
3.3
Performing an exposure .......................................................................................... 19
4.
Technical data ................................................................................................................. 21
5.
Starting up the system .................................................................................................... 22
5.1
Switch OFF all circuit breakers .............................................................................. 22
5.2
Initial start-up ......................................................................................................... 22
5.3
Fill up the detector cooler ....................................................................................... 23
5.4
Continue the start-up sequence ............................................................................... 25
6.
How to shut down the system......................................................................................... 26
6.1
Shutting down the modality .................................................................................... 26
6.2
Cutting power to the system ................................................................................... 26
7.
Service prerequisites and material ................................................................................ 28
7.1
Service toolbox ....................................................................................................... 28
7.2
Special software tools ............................................................................................. 31
7.3
Maintenance material ............................................................................................. 31
7.4
Fuses ....................................................................................................................... 32
8.
Disassembling instructions ............................................................................................. 34
8.1
X-ray prevent cover (1010576) .............................................................................. 35
8.2
Pre collimator (1010574) ........................................................................................ 39
8.3
Changing compression paddles .............................................................................. 40
8.4
Front cover (1010565/1020252) and Face protection shield (1010564) ................ 41
8.5
Display cover (1010547/1020253) ......................................................................... 42
8.6
C-arm covers (1010558, 1010559) ......................................................................... 43
8.7
Patient support (1010566/1020286) ....................................................................... 44
8.8
Opening and removing the doors of the main cabinet ............................................ 45
8.9
Removing the back plate of the main cabinet ......................................................... 46
8.10
Mounting service wheels for repositioning ............................................................ 47
8.11
Removing the top cover and side panels from detector cooler SMC ..................... 48
9.
Maintenance and inspection .......................................................................................... 49
1007989 O 4 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 5 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 6 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 7 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 8 (322)
Service Manual Philips MicroDose L30, Software 8.4
1. Introduction
1.1 Purpose
This manual contains instructions and procedures for regular maintenance and service tasks
regarding Philips MicroDose L30.
1.2 Scope
The information in this document applies to Philips MicroDose L30 for service and
maintenance purposes.
1.5 References
[1] Quality control procedures Philips MicroDose L30, Software 8.4
[2] AW System administrators guide Philips MicroDose L30, Software 8.4
[3] Technical Specification Philips MicroDose L30, Software 8.4
1007989 O 9 (322)
Service Manual Philips MicroDose L30, Software 8.4
Caution! Caution is used when danger for patient, personnel or system exists.
Exposure
The directions for service and maintenance described in this manual must be followed. No
other spare part than those listed in ref [5] may be used. If these conditions are not fulfilled
Philips cannot guarantee function and safety of the equipment and the warranty will not be
applied.
1007989 O 10 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 11 (322)
Service Manual Philips MicroDose L30, Software 8.4
2. System overview
2.1 Mammography stand
1007989 O 12 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 2 a) Overview of the air-cooled side cabinet and its main components.
b) Exterior of the water-cooled side cabinet.
1007989 O 13 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 14 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 15 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 16 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 17 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 18 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 19 (322)
Service Manual Philips MicroDose L30, Software 8.4
Exposure
When the Exposure button is pressed the
following sequence starts, see Figure 7.
A. The High Voltage Generator is set to
ready to output X-ray.
B. The scan movement is initiated by the
Stand Control System.
C. The X-ray is switched on (after a
scheduled delay), by the Stand Control
System activating a signal to the High
Voltage Generator.
The scan continues with no other
Figure 7 Sequence during exposure.
activities until the image field has been
reached.
D. Trig signals for detector readout are generated by the Stand Control System while the
scan is in the image field.
In the image field, the Detector Unit is detecting X-ray and pixel values are read out
and accumulated into image data by the Data Acquisition System.
E. The trig signals for detector readout are stopped and the X-ray is switched off at the
end of the scan.
F. The scan is finished and the stand is restored to stand-by state.
1007989 O 20 (322)
Service Manual Philips MicroDose L30, Software 8.4
4. Technical data
See ref [3], Technical Specification for data on e.g. measures, weight, radiation quality and
electrical data such as required type and rating of the external fuse.
1007989 O 21 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 22 (322)
Service Manual Philips MicroDose L30, Software 8.4
Always wear protective gloves and protective goggles when handling the
detector cooling fluid.
Caution!
Do not operate the detector cooler without enough cooling fluid in the
tank.
Illustrations
Figure 9 Front panel of Lauda detector cooler. Level indicator (1), Lid to tank (2), Display with
buttons (3), ON/OFF power switch (4).
1007989 O 23 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 10 Front panel of SMC detector cooler. Level indicator (1), Lid to tank (2), Run indicator (3),
Display for current temperature and alarm codes (4), Display for temperature set point (5).
The ON/OFF power switch is found on back of the detector cooler.
Procedure
1. Fold down the foldable rails.
2. If the SMC detector cooler is used, unscrew the four screws of the bracket holding the
detector cooler and lift off the bracket.
Pull out the detector cooler on the rails, see Figure 9 and Figure 10.
Unscrew the screws holding the metal shelter in place and remove the shelter. Note that the
shelter shall be remounted if moving the side cabinet to a new location.
Unscrew the lid on the front of the detector cooler.
Make sure that you have the correct type of water coolant mixture for the
Notice! detector cooling unit.
Use Detector cooling fluid (Part no 1013399).
Fill up the coolant system with fluid using a can with a funnel. Fill up to the indicated
maximum level shown by the level indicator on the front of the unit.
Switch ON the circuit breaker F10.
Check all hoses and connections for leakage. In case of leakage, immediately turn the cooler
OFF and attend to the fault.
Be prepared to turn off the detector cooler if the fluid level runs below the
Caution!
indicated minimum level.
If necessary, fill up the coolant system with more fluid until the fluid level stabilizes and all
air has disappeared from the system and the hoses. Fill up to the indicated maximum level.
Check that the temperature set point is set to 20C. If not, adjust the set point to 20C
according to the instructions in sections 10.5.1 or 10.6.4.
Screw the lid back on and tighten thoroughly.
Roll the detector cooler back into the side cabinet.
If the SMC detector cooler is used, screw the bracket holding the detector cooler back in
place.
Unscrew the four screws of the foldable rails, remove the rails and store them on site.
1007989 O 24 (322)
Service Manual Philips MicroDose L30, Software 8.4
Table 1 Fuses.
Order Component Serving Noticeable condition when turned on
1 F2 T1 Transformer A slight hum from the T1 Transformer.
1007989 O 25 (322)
Service Manual Philips MicroDose L30, Software 8.4
Please note that during a normal installation process shutting down the
Notice
system will not be necessary.
1007989 O 26 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preventive maintenance
Service prerequisites and material
Disassembling instructions
Maintenance and inspection
System adjustments
Service tool
Quality control
1007989 O 27 (322)
Service Manual Philips MicroDose L30, Software 8.4
Consumptions goods
q Electrical tape
q Duct tape (TESA)
q PTFE tape
q Double sided adhesive tape
q Fuses (See section 7.4)
q Lubrication
a. C-arm rotation centre
(Part no. 1013018)
b. C-arm height movement and compression movement screw
(Part no. 1013019)
c. Collimator and scan movement
(Part no: 1015566)
q Loctite 243 (for locking and sealing of threaded fasteners).
q Loctite 641
q Loctite 480
1007989 O 28 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 29 (322)
Service Manual Philips MicroDose L30, Software 8.4
The following items should also be included in the toolbox, but are supplied by Philips:
Phantoms
q Line pair phantom (Part no. 1010618)
q CDMAM phantom (Part no. 1013026)
q RMI phantom (Only US market. Part no. 1013027)
q PAS phantom (Only on certain markets. (Part no. 1014958)
QC kit
q PMMA plates 6*1 cm + 0.5*2 cm (Part no. 1013028)
q Aluminium filter (Part no. 1013029)
q Compression force measuring tool (Part no. 1013030)
q Fluorescent Paper (Part. No.1013032)
q Compression height calibration reference phantom (Part no. 1013031)
q Geometry phantom (Part no. 1012092)
q Transport bag for the QC kit (Part no. 1013034)
1
Only compatible with L30.
1007989 O 30 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 31 (322)
Service Manual Philips MicroDose L30, Software 8.4
7.4 Fuses
See Table 1 in Section 5.4 for fuses located in the side cabinet.
See Table 2 below for fuses located on the SCS-board.
Table 2 Fuses on SCS-board in main cabinet.
1 Fuse for C-motor 20A Very fast acting Littlefuse series 297 fuse
2 Fuse for Z-motor 20A Very fast acting Littlefuse series 297 fuse
7 Fuse for Aux-motor 20A Very fast acting Littlefuse series 297 fuse
8 Fuse for Scan-motor 20A Very fast acting Littlefuse series 297 fuse
1007989 O 32 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 33 (322)
Service Manual Philips MicroDose L30, Software 8.4
8. Disassembling instructions
This chapter contains instructions on how to dismantle and remove parts and covers that, when
removed, will allow service personnel access to other vital parts of the system.
1007989 O 34 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure - disassembling
1. Protect the patient support with (corrugated) cardboard or similar.
2. Dismount the X-ray prevent cover from the pre-collimator by loosen the Allen screw
securing the lock mechanism in the grey aluminum part a little bit. See A in the figure
below.
3. Move the X-ray prevent cover upwards by holding on to the grey aluminum part A.
4. Loosen (do not remove) the Allen screws B
and C that hold the upper part of the X-ray
prevent cover to the scan beam.
5. Un-hook the X-ray prevent cover from the three
Allen screws. (Depending on the construction of
the scan beam it might be necessary to remove
the two front screws C in order to remove the
X-ray prevent cover.)
1007989 O 35 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure - reassemble
1. Check that the two top screws of the X-ray prevent cover are not tightened in order to
allow adjustment. See B in Figure 13. (This only applies the first time the X-ray prevent
cover is mounted.)
2. Mount the X-ray prevent cover by lifting it up and hook it on the screws. See A in
Figure 13. (Depending on the construction of the C-arm, it might be necessary to remove
some of the screws A before positioning and securing the X-ray prevent cover.)
3. Fasten but do not secure the three Allen screws,
A. (I.e. the screws should be secured but allow
a certain amount of movement/adjustment.)
4. Move the X-ray prevent cover in its top end
position. Check if the magnet is able to hold the
X-ray prevent cover in its top end position.
5. If applicable, remove the plastic bag from the
collimator cradle.
6. Check that the collimator is free from dirt and
small metal particles. If necessary, use topz for
cleaning.
7. Move the interface to its lowest position by hand
(by holding on to the grey aluminum part, see
Figure 12 and D in Figure 14.
8. Attach the X-ray prevent cover to the interface.
See C in Figure 14.
9. Tighten the three Allen screws, A, at 2.9 Nm.
Figure 14 Attach the X-ray prevent cover
(D) to the interface (C).
1007989 O 36 (322)
Service Manual Philips MicroDose L30, Software 8.4
10. Tighten the Allen screw securing the lock mechanism in the grey aluminum part D.
11. Check that the C-arm and scan-arm are in vertical position.
12. Check that the two top screws C of the X-ray prevent cover are not tightened in order to
allow adjustment. (This only applies the first time the X-ray prevent cover is mounted.).
Figure 15 Test and adjust the X-ray prevent cover after mounting.
13. Move the interface A up and down, see Figure 15. Check that the two lowest parts B
of the X-ray shield are moving simultaneously and smooth.
14. Move the interface A to its lowest position (see Figure 15).
15. Secure the two top screws C at 2.9 Nm.
In order to avoid affecting the line-up of the X-ray prevent cover; do not
Notice
touch the X-ray prevent cover when securing the screws.
The X-ray prevent cover should now be positioned. This operation should only need to
be performed once. I.e. if the X-ray prevent cover is removed and re-mounted, the top
screws C should not need to be adjusted again.
16. Move the interface A up and down by hand.
a. Check that there is a play between the moving parts of the X-ray prevent cover
(see B with red marks in Figure 15).
b. Check that the moving parts B are moving simultaneously and smooth. If not,
try to adjust using the four screws D.
17. Check that the X-ray prevent cover does not press against the collimator. (Too much
pressure might affect the collimator alignment.)
18. Check that there is no built-in tension in the mount: De-attach the X-ray prevent cover
from the interface A and re-attach.
19. Secure the quick lock with the Allen screw E.
1007989 O 37 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure - Finalization
1. Verify that the X-ray prevent cover is mounted correctly by performing a Daily Quality
Control according to Quality control procedures, ref [1].
a. Check that there are no artefacts in the image.
b. Check that the X-ray prevent cover does not mask the image near the chest wall.
Solution for masked image:
i. Loosen the top screws (C in Figure 16). One on each side.
ii. Tilt the X-ray prevent cover a little bit forward.
iii. Repeat step 13 above to verify smooth movement.
iv. Perform a Daily Quality Control.
1007989 O 38 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure - disassembling
1. Remove the Allen screw securing both pre collimator and interface. See Figure 17 a.
2. Remove the pre collimator and interface. See Figure 17 b.
Procedure reassembling
Reassemble the pre collimator and interface in reverse order according to Figure 17. Tighten the
screw at 9.7 Nm.
1007989 O 39 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 40 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure - disassembling
1. Unsnap the face protection shield and
remove it, see Figure 19.
2. From underneath, disconnect the two cable
connectors that are attached to the
manoeuvre buttons on the Front cover.
3. Loosen the four hexagonal screws that hold
the front cover in place. See A in Figure
19. Do not remove the four screws just
loosen them. This will enable you to slide
the front cover slightly outward so that it
can be removed. (As the four holes in the
front cover are key hole-shaped.) Figure 19 Remove the face protection
shield.
Procedure - remounting
1. Slip the cover onto the four
screws. See A in Figure 20.
Slide the cover inwards until the
Front cover reaches the guides
B.
2. Secure the front cover by
fastening the four hexagonal
screws that hold the Front cover in
place.
3. Re-mount the two cable Figure 20 Dissemble front cover.
connectors.
4. Mount the Face protecting shield.
1007989 O 41 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 21 Lift off display cover. Left: Stand without protective metal plate. Right: Stand with protective
metal plate.
1. Unscrew the two Allen screws at the bottom of the display cover, see A in Figure 21.
2. Unhook the upper part of the display cover from the stand and lift off the display cover,
see B in Figure 21.
3. If the stand has a protective metal plate, unscrew the two Allen screws at the bottom of
the metal plate, see A in Figure 22.
4. If the stand has a protective metal plate, unscrew, unhook the upper part of the metal plate
from the stand and lift off the metal plate, see B in Figure 22.
Procedure - reassembling
1. If the stand has a protective metal plate, hook the upper part of the metal plate back onto
the stand.
2. If the stand has a protective metal plate, screw the two Allen screws at the bottom of the
metal plate.
3. Hook the upper part of the display cover back onto the stand.
4. Screw the two Allen screws at the bottom of the display cover.
1007989 O 42 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure - disassembling
Each side cover (left and right) is secured with four hexagonal screws and two cover nuts. These
screws are positioned as follows: one on the lower side of the C-arm, three on the back and two on
the top of the C-arm, see Figure 23 a.
1. If necessary, rotate the C-arm for easy access and remove the screws.
2. Remove the side covers from the C-arm; see Figure 23 b and c.
3. Return the C-arm to the zero degree position.
Procedure reassembling
1. Follow the instructions above in reverse order.
2. When reassembling the side covers, make sure that the chink between the left side cover
and patient support is the same as between the right side cover and patient support.
1007989 O 43 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Remove C-arm covers as described in section 8.6.
Rotate the C-arm to 180.
The upper part of the detector unit is very fragile. Make sure that the
Caution! patient support does not touch the detector during mounting/dismounting
of the patient support.
Procedure - disassembling
1. Disconnect the five cables connected to the patient support. Note the routing of the cables.
2. Remove the six Allen screws and washers. See Figure 25.
3. Carefully slide the patient
support off the C-arm. (Try to
pull out the lower part of the
patient support first in order to
reduce the risk of touching the
upper part of the detector when
the patient support comes loose.)
Procedure reassembling
1. Carefully slide the patient
support on to the C-arm. The
patient support should be placed
on the three nut bar-pins. Figure 25 Removing the six Allen screws securing the
patient support. (The last two screws on the
2. Tighten the six Allen screws with right side of the patient support are not
washers at 9.8 Nm. visible in this illustration.)
3. Connect the cables of the patient
support.
1007989 O 44 (322)
Service Manual Philips MicroDose L30, Software 8.4
Make sure that doors, covers and ground plate do not get any scratches or
Notice
other damages during disassembling and reassembling of the doors.
Procedure reassemble
Repeat the steps above in reverse order.
1007989 O 45 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure disassemble
1. Lift off the top cover of the main cabinet.
2. Loosen, do not remove, the two screws securing the back plate to the
main cabinet. Note how the protective earth cable marked BC is
mounted.
Figure 27 Back plate protective earth cable. (Back plate not mounted in the image.)
Make sure that doors, covers and ground plate do not get any scratches or
Notice
other damages during this operation.
Procedure - reassemble
1. Repeat the steps above in reverse order. Make sure that the protective earth cable is firmly
fastened to the back plate.
2. Measure the resistance between incoming earth in the side cabinet and the screw that
secures the protective earth cable to the back plate. The resistance must be less than 0.2
ohm.
Make sure that the protecting earth cable is correctly reconnected when
Caution!
the back plate is remounted on the main cabinet.
1007989 O 46 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 47 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 29 Mounting right front wheel. The left wheel is mounted in a similar way.
5. Continue to screw the four vertical bolts (alternate between front and rear wheels) until
the ground plate is in the desired height above the floor.
8.11 Removing the top cover and side panels from detector
cooler SMC
To remove the top cover, unscrew the two screws of the top cover and lift off the cover.
To remove the side panels, unscrew the six screws of each panel and lift off the panels.
1007989 O 48 (322)
Service Manual Philips MicroDose L30, Software 8.4
9.2.3 Cables
Thoroughly check that all cables and cable chains are correctly routed. Make sure that no cables
are jammed, pinched or is showing sign of wear.
Also check that no cables are obstructing the movement of the mechanical parts in the modality.
1007989 O 49 (322)
Service Manual Philips MicroDose L30, Software 8.4
9.2.8 Fans
Mammography stand
The fan unit is designed to force air into the stand and is located on the back of the mammography
stand.
1. Inspect that the fan unit (with two fans) is working properly.
2. Replace the filter of the fan unit (once each year).
Side cabinet
In case of an air-cooled side cabinet, check the top fan (mounted on the roof of the cabinet).
Procedure
1. Start SCS Calibration Tool.
2. Mount a compression force measurement tool in order to see at which force the
compression will be interrupted.
3. For projections 0 and 60, repeat steps 4 to 11 several times (3-4 tests) in order to get
good repeatability. The extreme values in each projection should be regarded, not the
average.
4. Reset SCS in calibration mode (Settings /Reset SCS Calibration Mode).
5. Set the desired projection (according to step 3).
6. Go to the Runtime tab in SCS Calibration Tool and select the motor Compression
(under Motor / Selected Motor).
7. Reset the dial.
8. Press button Goto and select New Position = 0 and Speed = 50. (The compression
will move down immediately after Speed has been entered.)
9. Wait until the stand enters emergency stop state.
10. Record the mechanical compression force limit shown on the compression force
measurement tool.
11. If the mechanical compression safety coupling is working properly, the compression shall
have been disengaged when the compression force exceeds 25 daN (+/- 2 daN).
1007989 O 50 (322)
Service Manual Philips MicroDose L30, Software 8.4
12. Reset the system (by pressing the Reset button on the stand) to make it available for
normal use again.
Procedure
1. Set the projection (C-arm angle) to 0.
2. Mount a compression force measurement tool
3. Place a soft phantom below the pad of the compression force measurement tool.
4. Perform a Reset (normal mode) to make sure that the stand is in normal mode.
5. Use SCS Calibration Tool to temporarily change the maximal force (Compression,
MaxForce) to 230N.
6. Reset the dial and compress until the stand emergency stops with error code 3-502-2.
7. Record the value shown on the display of the stand just before the stand enters emergency
stop. The value shall be 22-23 daN.
8. Record the value shown on the compression force measurement tool. The value shall be
the same as in step 7 ( 2 daN).
9. Reset the system (by pressing the Reset button on the stand) to make it available for
normal use again
Procedure
1. Remove the compression paddle.
2. Take the force measurement tool, and remove the bracket holding the dynamometer, then
reverse the dynamometer and remount it. Compress the dynamometer by hand and verify
that the motion is not obstructed by the mounting bracket.
3. Attach the force measurement tool to the modality. Ensure that the measurement pad now
points towards the collimator.
Place a thick soft material on the patient support to protect from the falling
Caution! dynamometer dial when emergency stop is activated.
1007989 O 51 (322)
Service Manual Philips MicroDose L30, Software 8.4
8. Go to tab Runtime in SCS Calibration Tool and select the motor Collimator (under Motor
/ Selected Motor).
9. Reset the dial on the force measurement tool.
10. Press button Goto and select New Position = 2300 and Speed = 1000. (The collimator
will move down immediately after Speed has been entered.)
11. Wait until the stand enters emergency stop state.
12. Record the force shown on the force measurement tool. (If the collision protection for the
collimator is working properly, the collimator shall stop and the stand enter emergency
stop when the force is less than 75N.)
13. If the test for any projection indicates a higher force than 75N, recalibrate the collision
protection for the collimator with SCS Calibration Tool.
14. Reset the system (by pressing the Reset button on the stand) to make it available for
normal use again.
15. Restore the force measurement tool to the original direction. Compress the dynamometer
by hand and verify that the motion is not obstructed by the mounting bracket.
Procedure
1. Set the projection (C-arm angle) to 0.
2. Move the height to 110 cm.
3. Apply the dynamometer from below so that it will hit the bottom of the patient support at
current height.
4. Move the height to 120 cm.
5. Set temporarily HighSpeed for C-arm Height to 10. (High speed will be 1 cm/s.)
6. Reset the dynamometer.
7. Lower the C-arm until the collision protection sets in. (The height will move up 1cm after
collision has occurred.)
8. Make a note of the measured value on the dynamometer. The value shall be less than
300N.
9. Restore the changed HighSpeed for C-arm height to its original value (50).
10. Reset the system (by pressing the Reset button on the stand) to make it available for
normal use again.
1007989 O 52 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Set the C-arm at a height of approximately 90cm.
2. Set the projection (C-arm angle) to +80.
3. Apply the dynamometer from below so that it will hit the C-arm at current position. Note:
The hit shall occur at a radius of approximately 60 cm from the centre (at the X-ray tube).
4. Set the projection (C-arm angle) to +70.
5. Set temporarily MaxSpeed for C-arm Rotation to 200. (Rotation speed will be 2/s.)
6. Reset the dynamometer.
7. Rotate the C-arm clockwise until the collision protection sets in.
8. Make a note of the measured value on the dynamometer. The value shall be
approximately 240N20N but absolutely not above 300N
9. Set the C-arm at a height of 90cm.
10. Restore the changed MaxSpeed for C-arm rotation to its original value (1000).
11. Set the projection (C-arm angle) to -80.
12. Apply the dynamometer from below so that it will hit the C-arm at current position. Note:
The hit shall occur at a radius of approximately 60 cm from the centre (at the X-ray tube).
13. Set the projection (C-arm angle) to -70.
14. Set temporarily MaxSpeed for C-arm Rotation to 200. (Rotation speed will be 2/s.)
15. Reset the dynamometer.
16. Rotate the C-arm anti-clockwise until the collision protection sets in.
17. Make a note of the measured value on the dynamometer. The value shall be
approximately 240N20N and absolutely not above 300N.
18. Restore the changed MaxSpeed for C-arm rotation to its original value (1000).
19. Reset the system (by pressing the Reset button on the stand) to make it available for
normal use again.
1007989 O 53 (322)
Service Manual Philips MicroDose L30, Software 8.4
Handling irregularities
If the left and right gap (mentioned above) differs:
Check if the compression paddle is mounted correctly
Check if the compression paddle is skewed
Check if the compression mechanics is skewed or damaged. If so, contact support.
1007989 O 54 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 31 Typical crack in hose. The hose must be replaced. Note that the appearance of the hose may be
different than illustrated here.
Cosmetic defects such as discolouration or yellowing are allowed. Smooth deformations without
sharp edges that are caused by contact with clamping are also allowed.
Procedure
1. Look for dust on the condenser, ventilation opening and front grille. Use compressed air
or a vacuum cleaner to remove dust. If necessary remove one or both side walls in order
to facilitate cleaning.
2. Check hoses and connections (chapter 9.2.16).
Always wear safety gloves and protective goggles when handling the water
Caution! coolant mixture.
Do not operate the detector cooler without enough cooling fluid in the tank.
1007989 O 55 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 32 Front panel of Lauda detector cooler. Level indicator (1), Lid to tank (2), Display with buttons
(3), ON/OFF power switch (4).
Figure 33 Back panel of Lauda detector cooler. Return hose (1), Outlet hose (2), Drain tap (3).
1007989 O 56 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 34 Front panel of SMC detector cooler. Parts indicated in illustration are: level indicator (1), lid to
tank (2), run indicator (3), display for current temperature and alarm codes (4), display for
temperature set point (5).
Figure 35 Back panel of SMC detector cooler. Return hose (1), Outlet hose (2), Drain plug (3), ON/OFF
power switch (4).
(Note: The illustration shows a transport hose connected.)
1007989 O 57 (322)
Service Manual Philips MicroDose L30, Software 8.4
4. If the foldable rails were removed for storage on site at installation, mount the rails back
on the rails in the side cabinet.
5. Pull out the detector cooler on the rails, see Figure 32and Figure 34 above.
6. If the Lauda detector cooler is used, place a vessel with a volume of at least 15 litres
below the tap on the back of the detector cooler. If the SMC cooler is used, place a vessel
with a volume of at least 8 litres and a height of maximum 160 mm below the rails.
7. Unscrew the tank lid on the front of the detector cooler.
8. Open the tap or plug on the back of the detector cooler and empty the cooling system.
9. Close the tap or plug.
Notice! Do not let the system remain empty longer than necessary.
Make sure that you have the correct type of detector cooling fluid.
Note!
Use Detector cooling fluid (Part no 1013399).
2. Fill up the coolant system with fluid. If the SMC detector cooler is used, fill up to the
indicated maximum level. If the Lauda detector cooler is used, then fill up to at least the
midpoint between the minimum and maximum level.
3. Switch ON the circuit breaker F10.
4. Check all hoses and connections for leakage. In case of leakage, immediately turn the
cooler OFF and attend to the fault.
Be prepared to turn off the detector cooler if the fluid level runs below the
Caution!
indicated minimum level.
5. If necessary, fill up the coolant system with more fluid until the fluid level stabilizes and
all air has disappeared from the system and the hoses. Fill up to the indicated maximum
level.
1007989 O 58 (322)
Service Manual Philips MicroDose L30, Software 8.4
6. Check that the temperature set point is set to 20C. If not, adjust the set point to 20C
according to the instructions in sections 10.5.1 or 10.6.4.
7. Screw the lid back on.
8. Roll the detector cooler back into the side cabinet.
9. Unscrew the four screws of the foldable rails, remove the rails and store them on site.
10. If the SMC detector cooler is used, screw the bracket holding the detector cooler back in
place.
1007989 O 59 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 60 (322)
Service Manual Philips MicroDose L30, Software 8.4
Required equipment
Approximately 5 cm2 of lubricant SKF LGMT2 (1013019).
Preparation
Clean the screw using a clean dry cloth.
Procedure
1. Apply grease on different places on the screw. A total volume of 5 cm2 of the lubricant is
suitable.
2. Run the nut in order to ensure that the grease is applied everywhere.
Required equipment
Scan screw: Grease KSS MSG No1 Lithium (1015566).
Collimator screw: Grease KSS MSG No1 Lithium (1015566).
Compression screw: SKF LGMT2 (1013019).
Preparation
Clean the screw using a clean dry cloth.
Procedure
1. Apply grease on different places on the screw.
2. Run the nut in order to ensure that the grease is applied everywhere.
1007989 O 61 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Open left and right side door.
Procedure
Visually check the oil levels of the internal worm-gear for Z-height and C-rotation. The oil level
should be up to at least the middle of the sight glass. If the oil level is lower, contact product
support.
The oil volume for the Z-height rotation worm gear is 15 cl and the oil volume for the C-rotation
worm gear is 65 cl.
Note Under normal circumstances the oil level should not change over time.
1007989 O 62 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 38 Main parts of purge kit: (1) Oil can, (2) Air valve, (3) Oil inlet, (4) Oil outlet.
Preparation
Let the system cool down for at least 30 minutes with the oil pump running before the air purge
procedure starts.
1007989 O 63 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 64 (322)
Service Manual Philips MicroDose L30, Software 8.4
b. Fill the outlet oil hose with oil by tilting the purge kit.
c. Disconnect the connectors of the outlet and inlet hoses. Make sure that no air is
introduced to the outlet oil hose.
Prepare system
1. If covers are mounted to the system, remove the face protection shield, the C-arm front
cover and the two C-arm side covers, see sections 8.4 and 8.6.
2. Open the side doors of the main cabinet.
3. Rotate the C-arm to +45 and keep this position for at least 10 minutes with the oil pump
running.
4. Remove the bracket holding the cables from the X-ray tube.
5. Secure the oil outlet hose of the X-ray tube with a cable strap.
6. Open SCS Calibration Tool and select Settings / Reset SCS - Calibration Mode (No start
movements).
7. Switch OFF the circuit breaker F13 to stop the oil pump.
8. Rotate the C-arm to -45 and lower it to the lowest position.
1007989 O 65 (322)
Service Manual Philips MicroDose L30, Software 8.4
3. Connect the purge kit to the connectors of the oil outlet hose of the X-ray tube.
Reset system
1. Disconnect the oil hoses of the purge kit from the connectors of the oil outlet hose of the
X-ray tube.
2. Unhook the air purge kit from the main cabinet.
3. Reconnect the outlet oil hose to the X-ray tube.
4. Remove the cable straps from the outlet hose of the X-ray tube.
5. Reattach the bracket holding the cables of the X-ray tube.
6. Switch ON the circuit breaker F13 to start the oil pump.
7. Reassemble the face protection shield, the C-arm front cover and the two C-arm side
covers, see sections 8.4 and 8.6.
8. Close the side doors of the main cabinet.
1007989 O 66 (322)
Service Manual Philips MicroDose L30, Software 8.4
9. In the SCS Calibration Tool, select Settings / Reset SCS Normal Mode.
10. If you are to purge several systems within a short period of time the oil can be reused for
up to five systems. When finished, dispose of the used oil according to local regulations.
1007989 O 67 (322)
Service Manual Philips MicroDose L30, Software 8.4
10.1.1 Preparations
1. Start the AW-computer.
2. Log in as MDM Administrator
3. Start the application SCS Calibration Tool (found under the menu Start / Sectra.)
4. Enter the password Mamea.
Procedure
1. Choose the Runtime tab.
2. Check that a value (~ 25000-40000) exists for Loadcell Z and that it changes more than
1000 units when the C-arm is pressed upwards/downwards hard, but without breaking the
covers.
3. Check that a value (~ 25000-40000) exists for Loadcell C and that it changes more than
1000 units when the C-arm is pressed right/left hard on the top of the C-arm, but without
breaking the covers.
1007989 O 68 (322)
Service Manual Philips MicroDose L30, Software 8.4
Figure 40 Buttons and signals in the SCS Calibration Tool user interface (tab Runtime).
1007989 O 69 (322)
Service Manual Philips MicroDose L30, Software 8.4
Signals
These signals are read from the SCS. For example in Figure 40 the signal DX_LIGHT is on which
means that the Dx indicator (light) is lit.
EM_STOP_SW1 Software emergency stop signal (for emergency
loop 1), used to force the SCS Unit into emergency
stop state from software.
OFF = Emergency stop active.
EM_STOP_SW2 Software emergency stop signal (for emergency
loop 2), used to force the SCS Unit into emergency
stop state from software.
OFF = Emergency stop active.
DX_LIGHT State of Dx-button lamp on patient support.
SIN_LIGHT State of Sin-button lamp on patient support.
GEN_SHUTDOWN OFF = The HVG is powered down
GEN_ENABLE ON = HVG enable, i.e. exposure possible
GEN_EXPOSURE ON = Exposure signal is on
DISPLAY_DIR N/A (Not used in L30)
HW_BOOT ON = Emulate a button press of SW4 (reset button
at the top of the SCS Unit)
WATCHDOG_SET Signal to reset the hardware watchdog of the SCS
WATCHDOG Service pulse from SCS software to the watchdog
that shows that the software is alive
Z_UNLOCK_BRAKE ON = Z-brake unlocked
C_UNLOCK_BRAKE ON = C-brake unlocked (Not used on L30)
24V_POWER ON = Power to motors is on (not height and
compression)
24V_POWER2 ON = Power to height and compression motor is
on
LDO_POWER_GOOD ON = Power to external interfaces (4.8V, e.g.
encoders) is ok
SW_RESET Software reset (not implemented)
COMPRESSION_PLATE_PRESEN ON = Compression paddle mounted (not
T implemented in mechanics)
STEREO_NEEDLE_SWITCH N/A
STEREO_SWITCH_AUX N/A
COLLIMATOR_PRESENT ON = Collimator mounted (not implemented in
mechanics)
STEREO_UNIT_PRESENT ON = Stereo unit mounted and powered up
CP_ID_1N
CP_ID_2N N/A
CP_ID_3N (Compression paddle identification not implemented
CP_ID_4N at present.)
CP_ID_5N
CP_ID_DATA N/A
(Data from compression ID circuit.)
TP8
N/A
TP9
(Used for testing.)
TP10
1007989 O 70 (322)
Service Manual Philips MicroDose L30, Software 8.4
Buttons
These buttons control the stand. For example in Figure 40 the buttons Side Dx and Light are
pressed. Thus the Dx-button on patient support is illuminated and the working light of the stand is
lit.
Exposure 1
Corresponds to the exposure buttons on the AW
Exposure 2
table and the optional exposure pedal.
Exposure 3
C Rotation Right Corresponds to the rotation right buttons on each
keypad on the C-arm.
C Rotation Left Corresponds to the rotation left buttons on each
keypad on the C-arm.
Height Up Corresponds to the rotation left buttons on each
keypad on the C-arm.
Height Down Corresponds to the C-arm up-buttons on each
keypad on the C-arm.
Compression Done Corresponds to the compression done buttons on
the lower keypads of the C-arm and the
compression pedals.
Compression Up Corresponds to the compression up buttons on the
compression pedals.
Compression Down Corresponds to the compression down buttons on
the compression pedals.
Side Sin Corresponds to the Sin (left) button on the patient
support.
Side Dx Corresponds to the Dx (right) button on the patient
support.
Emergency Corresponds to the first loop of emergency stop
switches.
Emergency (2) Corresponds to the second loop of emergency stop
switches.
Light Corresponds to the light buttons of the upper
keypads of the C-arm.
Button 1 Corresponds to the three buttons on the SCS board
Button 2 lower side. Button 1 is the button to the left. See
Button 3 section 20.4.2.
1007989 O 71 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 72 (322)
Service Manual Philips MicroDose L30, Software 8.4
10.1.5 Calibrations
Figure 41 shows the calibration tab in SCS Calibration Tool. The following sections describe how
to calibrate the SCS. Remember to make a backup of the configuration before performing any
calibrations, in case you need to roll back the changes.
Procedure
1. Choose Reset SCS Calibration Mode (under the
menu Settings). See Figure 42.
2. Choose the tab Calibration, see Figure 41.
3. It is recommended that the calibrations are performed
in the order Motor Currents, Height Position, C-
Rotation Position, Height Collision, Rotation
Collision, Scan, Collimator Height, Collimator
Collision, Compression Height, Compression Switch
and Compression Force.
4. It is, of course, possible to perform only one or more Figure 42 Reset SCS -
of the calibrations if necessary. E.g. if the collimator Calibration Mode
has been modified, calibrations Collimator Height and
Collimator Collision should be performed.
5. Follow the on-screen instructions. (See also more detailed instructions for each calibration
in the following sections.)
1007989 O 73 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Read the instructions in section 10.1.1.
Remove the display cover as described in section 8.5 in order to access the SCS.
Procedure
1. Click on the button Motor Currents (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of Motor Currents ended > is printed in
the log window. The string is followed by either [SUCCESS] or [FAILURE]
indicating whether the calibration was successful or not.
1007989 O 74 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Read the instructions in section 10.1.1.
Make sure that the upper front surface of the patient support is exactly one meter above the
floor, see Figure 43.
Procedure
1. Click on the button Height (in Figure 41).
2. Follow the instructions given by the program carefully.
3. The calibration is finished when < Calibration of Height ended > is printed in the log
window. The string is followed by either [SUCCESS] or [FAILURE] indicating
whether the calibration was successful or not.
4. Test Collision protection for the C-arm height movement (chapter 9.2.12).
1007989 O 75 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Read the instructions in section 10.1.1.
Make sure the C-arm is in angle zero (0) degrees. Use (for example) a water level.
Procedure
1. Click on the button C-Rotation (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of C-Rotation ended > is printed in the
log window. The string is followed by either [SUCCESS] or [FAILURE] indicating
whether the calibration was successful or not.
4. Test Collision protection for the C-arm rotation movement (chapter 9.2.13).
1007989 O 76 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
A Height calibration (10.1.7) shall have been performed before this calibration.
Procedure
1. Click on the button Height Collision (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of Height Collision ended > is printed in
the log window. The string is followed by either [SUCCESS] or [FAILURE]
indicating whether the calibration was successful or not.
4. Test Collision protection for the C-arm height movement (chapter 9.2.12).
Preparation
A Height calibration (10.1.7) and a C-Rotation calibration (10.1.8) shall have been performed
before this calibration.
Procedure
1. Click on the button C-Rotation Collision (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of C-Rotation Collision ended > is
printed in the log window. The string is followed by either [SUCCESS] or
[FAILURE] indicating whether the calibration was successful or not.
4. Test Collision protection for the C-arm rotation movement (chapter 9.2.13).
1007989 O 77 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Read the instructions in section 10.1.1.
Remove the compression paddle.
Procedure
1. Click on the button Collimator Height (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of Collimator Height ended > is printed
in the log window. The string is followed by either [SUCCESS] or [FAILURE]
indicating whether the calibration was successful or not.
Preparation
Read the instructions in section 10.1.1.
Remove the compression paddle.
Procedure
1. Click on the button Collimator Collision (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of Collimator Collision ended > is
printed in the log window. The string is followed by either [SUCCESS] or
[FAILURE] indicating whether the calibration was successful or not.
4. Test Collision protection for the collimator movement (section 9.2.11).
1007989 O 78 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Read the instructions in section 10.1.1.
Procedure
1. Click on the button Scan (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of Scan ended > is printed in the log
window. The string is followed by either [SUCCESS] or [FAILURE] indicating
whether the calibration was successful or not.
Preparation
Read the instructions in section 10.1.1.
Remove the compression paddle.
Procedure
1. Click on the button Compression Height (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when < Calibration of Compression Height ended > is
printed in the log window. The string is followed by either [SUCCESS] or
[FAILURE] indicating whether the calibration was successful or not.
4. Calibrate each of the compression paddles (chapter 11.3.8).
1007989 O 79 (322)
Service Manual Philips MicroDose L30, Software 8.4
Preparation
Read the instructions in section 10.1.1.
Remove the C-arm covers to be able to adjust the compression switch.
Procedure
1. Click on the button Compression Switch (see Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The calibration is finished when the text < Calibration of Compression Force Switch
ended > is printed in the log window. The string is followed by either [SUCCESS] or
[FAILURE] indicating whether the calibration was successful or not.
Preparation
Read the instructions in section 10.1.1.
Attach the compression force measurement tool to the machine.
Procedure
1. Click on the button Compression Force (in Figure 41).
2. Follow the on-screen instructions given by the program carefully.
3. The system runs a calibration which is finished when the text
< Calibration of Compression Force ended > is shown in the log window. The string is
followed by either [SUCCESS] or [FAILURE] indicating whether the calibration was
successful or not.
1007989 O 80 (322)
Service Manual Philips MicroDose L30, Software 8.4
Notice Note that the motor temperature is not monitored by the system.
Preparation
The axis (motor) for which to perform the axis burn in must be calibrated.
Procedure
1. Click on the Axis Burn In button in the Burn In group.
2. Choose the desired axis by entering a value between 0 and 4. Press Enter.
3. Enter the number of test cycles. Press Enter. (20 cycles is the default value. One cycle =
back and forth.)
1007989 O 81 (322)
Service Manual Philips MicroDose L30, Software 8.4
5. Decide whether to save data to log file. If Yes, choose file name to save log data to.
6. If Yes, choose the file name to save log data to and which data to log.
8. The operation is finished when < Axis burn in test cycle finished > is printed in the log-
in window.
Do not run the motors so hot that they cannot be touched. Preferably use
Caution!
default values for number of test cycles and waiting time.
1007989 O 82 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Choose the tab Calibration and press Test Cycle Burn In, see Figure 51. Test Cycle Burn
In takes approximately 1 hour 40 minutes.
1007989 O 83 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Choose the tab Calibration and press Exam Test Cycle, see the figure below. Exam Test
Cycle takes approximately 15 minutes.
Procedure
Start the test by clicking the C-Rot Collision Test button. Make sure that the stand can rotate
freely.
1007989 O 84 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Choose the tab Parameters. Make sure Automatic update (in the lower right corner) is
checked.
1007989 O 85 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
Follow the instructions in section 10.2.1 where the group C_Rot should be chosen followed by the
parameter MaxSpeed.
1000 (= 10/second) is the default value of MaxSpeed. (MaxSpeed is in 1/100 /second.)
Note: Do not change MaxSpeed to a value larger than the default value (1000).
Procedure
Follow the instructions in section 10.2.1 where the group Height should be chosen followed by the
parameter HighSpeed and/or LowSpeed.
50 and 25 mm/s respectively are the default values.
Procedure
Follow the instructions in section 10.2.1, choose the group Compression followed by the
parameter HomePosition.
The default value shown in SCS Calibration Tool is 1410, representing the compression paddle
distance from patient support in 0.1 mm increments, and corresponds to 141 mm.
Adjust the value for desired new home position.
By measuring check the gap between high edge compression paddle and the face shield is not less
than 72 mm, compression paddle returned to home position as in normal operation.
1007989 O 86 (322)
Service Manual Philips MicroDose L30, Software 8.4
2. Press again.
5. Press or .
No other parameters should be set or altered in the detector cooler.
It is not possible to define max and min alarm set points on the Lauda detector cooler.
1007989 O 87 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 88 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Start the Service Tool, choose the tab Service and click the Field limiter scan button.
2. Follow the on-screen instructions.
3. The result is shown by the upper right colour indicator boxes in the Field Limiter
dialog, see Figure 55. If the middle main indicator is green, the field limiter is good. If the
indicator is red, the field limiter needs to be adjusted, see 10.7.2.
Figure 55 Field limiter scan. The over-all result is shown by the upper right color indicator
boxes. For the main indicator to be green, the gradient (from gray to black) created by
the field limiter must be located within the red continuous lines throughout the whole
image height. When adjusting the field limiter, the gradient should be located within
the dashed lines that appear when checking Adjustment, see 10.7.2.
1007989 O 89 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Loosen the four screws, see Figure 56. Then tighten them again, the front screws first.
Always tighten the front screw before the back screw. Otherwise the
Note
software will have trouble guiding the adjustment.
Figure 56 Location of the four screws securing the field limiter under the X-ray tube.
2. Press Exposure in the Field Limiter dialog (or press File limiter scan in Service Tool).
Follow the on-screen instructions. Loosen the four screws (or two screws if only one side
of the field limiter needs to be adjusted). See Figure 56.
3. Loosen the four screws (or two screws if only one side of the field limiter needs to be
adjusted). See Figure 56.
1007989 O 90 (322)
Service Manual Philips MicroDose L30, Software 8.4
4. Adjust the field limiter according to the instructions in the Field limiter dialog using the
two adjustment screws on each side. See Figure 57.
5. Tighten the four screws (see Figure 56), the front screws first.
Always tighten the front screw before the rear screw. Otherwise the
Note
software will have trouble guiding the adjustment.
6. Press Exposure in the Field Limiter dialog. Follow the on-screen instructions.
7. If the main indicator is red, repeat from step 3 until the indicator is green.
8. If the tube has two filaments and it is desired to adjust the field limiter with respect to
both filaments (optional), perform the instruction in section 10.7.3. Otherwise, continue
with step 9.
9. Reassemble the C-arm covers.
10. Re-assemble the X-ray prevent cover.
11. Perform Tests after maintenance or service according to chapter 14.
1007989 O 91 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 92 (322)
Service Manual Philips MicroDose L30, Software 8.4
5. Subtract 3 mm from the measured value. Round to the closest integer below. This number
is the compression height limit for the standard and the high edge compression paddles.
1007989 O 93 (322)
Service Manual Philips MicroDose L30, Software 8.4
The number NNN shall be replaced by the determined compression height limit with the
following considerations:
If the determined compression height limit is below 100 mm, NNN shall be set to
100.
If a file with the same content is already present, modify the existing file.
If the determined compression height limit is 104 mm or above, do not create any
file. If a file with this content is already present, delete the file.
Examples:
If the measured distance between the patient support and the collimator is 105.8
mm, the compression height limit is 102 mm and NNN shall be set to 102.
If the measured distance is below 104 mm, NNN shall be set to 100.
If the measured distance is above 107 mm, there shall be no file present with this
content.
7. Abort the Warm up tube procedure by pressing Cancel in the Warm up tube procedure
dialog. Press Yes to acknowledge aborting.
8. Save the file and restart the CCS. This can be done by pressing Reboot CCS on the
Service tab.
9. Open the file C:\Program Files\Sectra\CCS\Log\
actual_config.log and search for the parameter compHeightLimit. Make sure that the
values for the standard and high edge paddles is in accordance with the local
configuration created in step 7.
1007989 O 94 (322)
Service Manual Philips MicroDose L30, Software 8.4
1007989 O 95 (322)
Service Manual Philips MicroDose L30, Software 8.4
When?
Once a week, or if Quick calibration fails.
When?
A Quick calibration should be performed when a line artefact occurs. See example in section
18.2.1.
When?
When four hours have passed since the last exposure was performed, the AWS will recommend
warming up the X-ray tube using this procedure.
1007989 O 96 (322)
Service Manual Philips MicroDose L30, Software 8.4
When?
If Full calibration or Quick calibration fails because of previous problems which have now been
resolved, this step may be required. The channels which are masked because of the history
function are shown in the Channel mask viewer as Check history, see section 11.8.
1007989 O 97 (322)
Service Manual Philips MicroDose L30, Software 8.4
Asic calibration trims the amplifier settings for the ASICs in the detector. The calibration is
performed without X-ray exposure.
Handling irregularities
See section 20.3 for information regarding handling failed calibrations.
1007989 O 98 (322)
Service Manual Philips MicroDose L30, Software 8.4
Procedure
1. Press Emergency stop and
lower the collimator by hand.
2. Mount the lower part of the
ACA mechanics, see Figure
61. The lower part is attached
to the pre collimator assembly
by two Allen screws.
3. Loosen (do not remove) the
three screws holding the
collimator to its base, see
Figure 62. In what follows,
we refer to these as the Figure 61 The upper ACA mechanics is mounted on the so
alignment screws. You should called interface position. The lower ACA
now be able to move the mechanics is attached to the pre collimator
collimator in all directions. assembly. (The X-ray prevent cover is not
4. Tighten the alignment screws visible in this illustration.)
to 1.2 Nm or by hand until
there is no vertical play
when you pull the
collimator front up and
down. The collimator
should however still be
possible to move by
hand from left to right.
5. Mount the upper part(s)
of the ACA mechanics.
You may need to move
the collimator left/right
in order for the parts to
fit in each other. Apply
the springs (see A in
Figure 61) and check
that the collimator
position can be adjusted Figure 62 The three alignment screws of the collimator.
by turning the knobs.
6. Adjust both screw micrometers to the starting position:
10 mm for blue ACA (revision C or later).
13 mm for silver ACA (revision B or earlier).
7. Connect the power cable and the computer cable.
1007989 O 99 (322)
Service Manual Philips MicroDose L30, Software 8.4
8. Release the emergency stop and perform a reset. If the X-ray prevent cover is mounted,
make sure that it is attached to the collimator interface during the reset.
Make sure that no part of the ACA mechanics collide with anything during
Notice the Reset. Otherwise the homing of the collimator movement will be
incorrect and the alignment useless.
9. Click on System Configuration (see Figure 59) in the Service Tool, under the tab Service.
Check that the Detector S/N is correct. If it is not correct, enter the correct one and click
OK.
10. Click Collimator Alignment (see Figure 59) in the Service Tool, under the tab Service to
start ACA.
11. Click OK. The ACA motor starts the homing procedure, and the collimator can be seen to
move slowly in the left/right direction.
As the screws are tightened, you must not move the motor! Therefore, at
Notice
this stage, be sure not to check Screws are made loose.
16. If the tube has two filaments and it is desired to align the pre-collimator with respect to
both filaments (optional), perform the instruction in section 11.3.3. Otherwise, continue
with step 17.
17. When the alignment is good enough (all indicators are green), click Done.
18. Disconnect and remove the ACA mechanics.
19. Perform a Geometry calibration. See section 11.3.5.
Handling irregularities
See section 20.3.1 for an alignment trouble shooting guide.
Figure 65 Verify alignment. If the indicators are green, the alignment is good.
Figure 66 Verify Alignment dialog will appear if full collimator alignment was performed with
L30 SW 8.1 or earlier.
2. Verify that the Collimator height um is set to a positive number (it might be for example
235000). Click Exposure and expose.
3. Reduce the Collimator height um by 7000 um (following the example above, we should
set the height to 228000). Click Exposure and expose.
4. The green field (asics that are still counting at the same rate) should be properly centred
inside the blue fields (asics that count less). See example of correct alignment in Figure 67
(left). If the collimator is not aligned, see an example in Figure 67 (right), a full collimator
alignment has to be performed.
Handling irregularities
See section 20.3 for information regarding handling failed calibrations.
Procedure
1. Click on Geometry calibration in the Service Tool,
under the Service tab.
2. Place the Geometry and image field phantom on the
patient support. Compress. Press Compression Done
and hold down the exposure button.
3. A progress bar is shown and the current step is
indicated. Figure 68 Geometry phantom.
(Part no. 1012092)
4. The user is informed whether the calibration was
successful or not.
5. The status field is updated.
6. It is advisable to check the channel mask after this
calibration, especially if the calibration was unsuccessful.
Handling irregularities
If the Geometry calibration fails:
1. Verify that the phantom is correctly mounted on the patient support. The front guiding
edge of the phantom should touch the front end of the patient support.
2. Run Verify alignment.
3. If the alignment is correct, check if the detector touches the patient support. This can be
performed
a. by checking the Scan speed in Measurements in the Service tool. If the scan speed
diagram shows unexpected peaks, the detector might touch the patient support.
b. or by removing the side covers of the C-arm. A visual check can performed
verifying that there is free space between the detector and patient support.
If no problems are found performing the above described actions, contact product support. See
section 20.3 for more information regarding failed calibrations.
Skew calibration measures and compensates for the deflection of the detector as an effect of
gravity.
Procedure
1. Click on Skew calibration in the Service Tool, under the Service tab.
2. Rotate the C-arm to -90 degrees. Place the Geometry and image field phantom on the
patient support. Compress gently, press Compression Done and hold down the exposure
button.
3. A progress bar is shown and the current step is indicated.
4. The user is informed whether the calibration was successful or not.
Handling irregularities
Contact product support if the calibration fails,
Procedure
1. Click AEC calibration in the Service Tool, under the Service tab.
2. The user is informed whether the calibration was successful or not.
Handling irregularities
AEC calibration is a part of Full Calibration that is performed on regular basis. However if an
AEC calibration would fail, probable explanations might be that the X-ray tube is new and has to
be burned in by performing a number of exposures before output is ok. A second explanation
might be that the X-ray tube is old and need to be replaced.
See section 20.3 for information regarding handling failed calibrations.
Procedure
1. Click Compression paddles in the Service Tool, under the Service tab.
2. All configured compression paddles are shown to the upper left. The text (calibrated) is
added after the names of the paddles that are already calibrated. See Figure 80.
3. Mark the compression paddle you want to calibrate.
4. Place the compression height calibration phantom on the patient support.
5. Enter the True height of the phantom.
6. Compress to a certain force between 50 and 200 N. Click Measure. The measured values
are displayed in the plot.
7. Compress to a different force between 50 N and 200 N. Click Measure. Repeat this until
there are at least five points in the plot.
8. Click Accept. The text (calibrated) is added after the name of the compression paddle.
9. Mark another compression paddle to calibrate or close the window.
It is also possible to test a calibrated compression paddle, by checking the Test box. Then the
height should equal the real height of the object, regardless of the compression force. A calibrated
compression paddle can be rejected (i.e. marked as not calibrated) by pressing Reject.
Outgoing control
After the compression paddles have been calibrated:
Compress all the way to the patient support. Check that the display shows 0 mm to 2 mm
and no compression force.
Handling irregularities
1. Check if the compression force measurement is correct.
2. The calibration might fail if the values are too spread. Repeat the procedure until it is
successful.
See section 20.3 for more information regarding handling failed calibrations.
11.4 Measurements
Under the Service tab, various measurements can be performed by clicking Measurements.
The peaks in the middle part of Figure 70 show the measured frequencies in the X-ray output
variation. The dominating peaks should be the anode rotation frequency around 130-180 Hz (and
multiples). Within this interval the position of the peak can change depending on the temperature
of the tube, being at higher frequency when cold and lower when hot. Since the ripple peak
reflects the anode speed it will also be at higher frequency if the time between Compression Done
and Exposure button pressed, is longer. Peaks in 50/60 Hz (and multiples) might be caused by
electrical disturbances in the system.
It is also possible to view the ripple from the latest exposure by clicking Get latest exp. In this
case, for a normal scan with varying trig times, no frequency plot will be shown.
If the RMS/mean value is above 1.5% it could indicate that the ripple is too high, and visual study
of images should be performed.
Notice This tool should only be run upon direct instructions from product support!
It is not only possible to see which parts that have been masked, but also when the masking took
place, e.g. during which calibration step. This information is found in the top left corner under the
displayed bits panel. Below follow a brief explanation of the different parts:
CM Part Description
Elec ASICs which are electronically turned off in the CCS configuration.
Asic Channels masked during ASIC calibration.
Alignment Channels masked during collimator alignment.
Geometry Sensors masked during geometry calibration.
Flatfield Masking by the Full calibration. Deviating channels are masked.
Check These are channels that passed the Full calibration but found to be deviating
during Quick calibration. This mask is reset during full calibration.
Manual Channels manually masked in the Raw data viewer.
Flatfield These are channels that have been disabled in recent Full calibrations.
history
Check history These are channels that have been disabled in recent Quick calibrations.
CM Part Description
High bits At present not in use.
(unknown)
The measurements table is an extension of the X-ray tube status in the Tools menu in the AWS
application. The ESAK value corresponds to the current tube output.
History
Channel rate history shows the detected
rate of X-ray photons over time,
measured over all functioning channels in
the detector. The curves gradually
decrease over time due to loss of
efficiency in the X-ray tube, and increase
abruptly when the X-ray tube is replaced.
Channel yield history shows the fraction
of functioning channels in the detector
over time. Large variations in this curve
may be caused by detector instabilities.
ESAK History
ESAK conversion history shows the
ESAK conversion factor which is used to
translate detected photon rate into dose
rate. This value is usually constant over
time, but may be affected by software
updates in conjunction with changes to
reference methods for dose measurement.
ESAK per mAs history shows the tube
output over time. The last point
corresponds with the ESAK value in the
measurements table. To measure relative
changes in tube output, the Channel rate
history graph is more appropriate, as it is
not affected by changes to the ESAK
conversion factor. Figure 84 ESAK History view in the AEC Status dialog.
11.10 Settings
11.10.1 Dose configuration
The dose configuration can be changed by clicking Dose configuration under the Service tab in
the Service Tool. Two dose configurations (C100, and C120) can be found via the drop down
menu, see Figure 85 below.
The dose configuration may only be changed after consultation with the
Notice
product management!
11.10.3 Compression
There are two compression modes: Default and
Phantom. Default mode is for normal use, while
Phantom mode is for compression of phantoms in
the AWS. In the Service Tool, Phantom mode is
used everywhere except for this dialog, where the
currently selected mode is used.
For each compression mode, the high and low
speeds, the first stop and the keep compression
speed may be adjusted. The high speed is the
speed without compression and the low speed is
with compression. The keep compression speed is
Figure 87 Settings for compression.
used in Keep compression mode. The upper limit for the
keep compression speed is 25 mm/s.
If changing speeds for Phantom mode, verify that the mechanical emergency stop will not be
triggered when compressing against a stiff object.
The first stop may be set which is the compression force when the paddle stops. The compression
pedal must be pressed again for higher pressure.
It is also possible to select force unit for the display at the stand to deca Newton (daN) or pounds
(lbs).
It is possible to acquire image data in three ways with the Service Tool. Still without X-ray, Still
with X-ray, and Scan with X-ray. See Figure 91.
In the Image viewer it is possible to select how to view the data and thereby rebuild the image.
When choosing Build and Get (in the Image Viewer) the dialog in Figure 94 appears.
Either raw data from single lines can be fetched, or a built image. It is possible to select which bits
in the channel mask that will affect the retrieved data. (To affect an image the bits should be
changed before building the image.)
The data can be retrieved as time normalized, flat fielded or ripple cancelled. When getting an
image the user can select to retrieve e.g. the image, image with median filter (as in AWS but
without IPK filtering), image overlap count. The data will be shown in the image viewer, and
more data can easily be retrieved by using the Build and Get button in the image viewer.
Note: When scanning with X-ray an image is automatically built and shown (ripple cancelled
image without median filter).
Selecting To Raw Data Viewer from point in the Left button drop list allows for clicking in the
image to view data from individual lines of the detector that contributes to this region of the
image.
3. Remove the four Allen screws on the upper right and left side of the collimator
curtain, i.e. two on each side. See Figure 96 a).
4. Remove the Allen screw securing the interface. See Figure 96 b).
5. Remove the interface together with the pre-collimator. See Figure 96 c).
Figure 96 a) Removing upper right Allen screws for the collimator curtain.
b) Allen screw securing the interface.
c) Removing Allen screw and interface.
6. Remove the two Allen screws securing the lower bracket for the X-ray prevent cover,
see Figure 97 a.
7. Lift up the interface pin in order to facilitate the removal of the collimator curtain, see
Figure 97 b.
8. Carefully remove the collimator curtain, see Figure 97 c.
9. Remove the collimator curtain bracket (A), mounted on the C-arm, by loosen the two
Allen screws (B). See Figure 98.
Figure 98 Mounting/removing the upper collimator bracket, A, by tighten/loosen the two Allen
screws B.
Notice Do not re-use the old collimator curtain bracket on the new collimator
curtain.
Do not re-use the old collimator curtain bracket on the new collimator
Notice
curtain.
2. Un-hook the wire clip on the back of the collimator curtain cassette (on the horizontal curtain).
See part (C) in Figure 99.
3. Loosen the four small nuts and remove the two support pins on the back of the collimator
curtain cassette. See part (D) and (E) respectively in Figure 99.
Figure 100 Mounting/removing the upper collimator curtain bracket, (A), by tighten/loosen
the two Allen screws (B).
2. Use a 2.5 mm Allen key (A) or a thickness gauge to verify a 2.5 mm distance between
the scan beam (B) and the upper collimator curtain bracket (C). See Figure 101.
Figure 101 Using a 2.5 mm Allen key (A) for adjusting the 2.5 mm distance between scan
beam (B) and upper collimator curtain bracket (C).
3. Attach collimator curtain assembly on the interface pin. See Figure 102 a.
Figure 102 a) Attaching the collimator curtain assembly to the interface pin.
b) Securing the collimator curtain to the upper collimator curtain bracket.
4. Hook on the lower part of the collimator curtain to the C-arm, see Figure 103.
Figure 103 Attaching the lower part of the collimator curtain to the C-arm.
5. Attach the collimator curtain to the upper collimator curtain bracket using four Allen
screws (two on each side). See Figure 102 b.
6. Re-mount the support pins using four small nuts according to (D) and (E) in Figure
99.
7. Re-mount the wire-clip. The wire-clip must be pushed down as much as possible in
the curtain and behind the lower support pin, as shown in Figure 104.
Figure 104 Correct mounting of wire-clip. Pushed down as much as possible in the curtain
and behind the horizontal pin.
8. Check that there is approximately 1 mm play between collimator curtain and
collimator motor (bracket). See Figure 105 a.
9. Remove the compression paddle and lower the collimator to its lowest position.
Move the compression wagon up and down and verify that there is a play between the
collimator curtain and the scan beam. See Figure 105 b.
10. Re-mount the lower bracket of the X-ray prevent cover. See Figure 97 a.
11. Re-mount the interface. See Figure 96 c and b respectively.The Allen screw for the
interface should be secured with 9.7 Nm.
Related drawings
See electrical drawing 1007140 in the service binder.
Preparations
Remove the C-arm front cover and the left and right C-arm side covers, as described in
sections 8.4 and 8.6.
Perform air purge and oil level adjustment of the X-ray system according to section
9.2.23. Leave the purge kit hooked onto the stand and the covers removed.
Move the C-arm to a vertical position (0 angle) and lower the C-arm. This to make it
easier to reach and change the X-ray tube, which is very heavy.
Power down the system and shut off the fixed main switch according to sections 6.1
and 6.2. This enables manual movement of the collimator wagon and makes it possible
to disconnect cables without risk of harm to the system or personnel.
The cooling hoses of the X-ray tube can easily slip into the scan arm when
Note disconnected from the X-ray tube. Secure the hoses using e.g. cable straps
before disconnecting the hoses.
3. Note how the two cooling hoses of the X-ray tube unit are placed in order to place the
hoses of the new unit the same way. See Figure 108 and Figure 109. Disconnect the
hoses.
When disconnecting the X-ray tube high voltage cable always connect
Caution! the cable contact pins to the stands chassis ground. This to prevent
electrical discharges.
The X-ray tube unit is very heavy. Use a stable chair or a ladder with
Note
suitable height to access the X-ray tube unit.
7. Wrap a broad band around the X-ray tube unit in order to get a firm grip.
8. Remove the X-ray tube unit by carefully lifting it upwards.
7. Connect the high voltage cable to the back of the X-ray tube. See Figure 108. Be sure
to give excessive grease time to flow out from the contact under pressure before you
turn the contact.
8. Connect the three protective ground cables to the protective ground screw, see (F) in
Figure 108. I.e. two ground cables of the filament cable (black and green, yellow
respectively) and the uncoated ground cable of the high voltage cable.
9. Reattach the two cover brackets by tighten the four Allen screws. See (E) in Figure
108.
Figure 109 C-arm seen from back. Cable bracket (G) and routing of cables and hoses.
Location of PE-screw (F).
It is important that the tube exposure statistics are reported correctly, and
Notice that the tube exposure counter is reset for the new tube, according to the
procedure above.
6. In the System configuration dialog, enter the serial number of the new X-ray tube.
7. Make sure that filament 0 is selected by looking in C:\Program
Files\Sectra\CCS\local_config for a file with the contents:
HVG {
# Filament selection (0 or 1).
filament = 1
}
If Brake anode is checked, i.e. the anode is braked after each exposure,
Notice wait at least 30 seconds after completed exposure before pressing
Compression Done.
11. Verify the collimator alignment (section 11.3.4). Align the collimator if needed,
(section 11.3.2) and perform a Geometry calibration (section 11.3.5).
Tip: If you need to align the collimator, try to make the alignment work for both
filaments and you can save time on the service visit when you change filaments. See
instruction in 11.3.3. Remember to change back to filament = 0 when you are finished
and verify that it is ok.
12. Verify the position of the field limiter (section 10.7.1). Align the field limiter if
needed.
Tip: If you need to adjust the field limiter, try to make the adjustment work for both
filaments and you can save time on the service visit when you change filaments. See
instruction in section 11.3.3. Remember to change back to 0 when you are finished
and verify that it is ok.
13. If collimator alignment and geometry calibration have been done, check the Image
Field and missed tissue by the chest wall, see Quality control procedures [1].
14. Perform a ripple test (see section 11.4.1).
15. Verify that there is no radiation leakage (see section 15.6).
16. Perform a Full calibration, see Quality control procedures [1].
17. According to Quality control procedures [1], check:
Tube output
HVL
Tube Voltage
Air Kerma reproducibility
The tube output will decrease in the beginning, depending on how much
Notice
the Modality is used.
Figure 110 Using the SCS Calibration tool for moving the scan arm during X-ray tube
exchange.
4. In the Tools menu on the AW, uncheck the Brake anode-setting in to limit the strain
on the high speed starter and anode assembly. Perform three exposures on minimum
mAs, at each available kV, i.e. totally 3x5 = 15 exposures. Change the Brake anode
setting back to normal.
If Brake anode is checked, i.e. the anode is braked after each exposure,
Notice wait at least 30 seconds after completed exposure before pressing
Compression Done.
5. Verify the collimator alignment (section 11.3.4), align the collimator if needed
(section 11.3.2) and perform a geometry calibration (section 11.3.5).
6. Verify the position of the field limiter (section 10.7.1), adjust the field limiter if
needed (section 10.7.2).
7. If collimator alignment and geometry calibration have been done, check Image Field
and missed tissue by the chest wall according to Quality control procedures [1].
8. Perform a ripple test (section 20.1).
9. Verify that there is no radiation leakage (section 15.6).
10. Perform a Full calibration, see Quality control procedures [1].
11. According to Quality control procedures [1] check:
Tube output and Air Kerma reproducibility
HVL
Preparation
Raise the C-arm to highest possible position (in order to access HVG easier).
Take a close look at the markings of the cables and note how the cables are
Notice routed, for easy reattachment.
Be careful not to drop any screws into the generator unit.
When disconnecting the high voltage cable, always connect the cable
contact pins to the stands chassis ground. This to prevent electrical
discharges.
Caution! The high voltage cable and filament cable are very fragile. Be careful
when handling.
When disconnecting TB3, be careful when pressing the screw driver into
the connection terminal, to avoid any damage.
4. Loosen upper screw and the two lower screws. See Figure 111.
5. Carefully rotate the HVG clockwise (while sliding it along the back beam) until it is
outside the main cabinet.
6. Lift HVG carefully and lay it down.
7. Carefully loosen cables attached to the bottom side.
8. Remove upper and lower plate to be re-used on new HVG. See Figure 112.
Figure 111 High voltage cable (J1), filament cable, (J2), flow control cable (TB3) and the earth
cables.
Figure 112 Attaching chassis mounting plates to the HVG and cable connections on the bottom side
of the HVG. HV Generator Signal Cable (JB1), HVG three phase power cable (TB1),
single phase mains power cable (TB2) and protective earth (PE).
5. Connect the filament cable and the flow control cable on top of the generator.
6. Carefully clean the contact of the high voltage cable. Put a thin layer of grease on the
receptacles body, a thin layer on receptacles tip and a large amount on the receptacles
45 degrees edge. See Figure 107. The grease must be Varian medical systems
vapour-proofing compound or Wacker Silicon paste P4. Connect the high voltage
cable to the high voltage generator.
7. Check that no cables are squeezed or strained. If the cable bracket was loosened,
remember to tighten it again.
8. Secure the cables with tie wraps.
9. Close left door on the stand.
If Brake anode is checked, i.e. the anode is braked after each exposure,
Notice wait at least 30 seconds after completed exposure before pressing
Compression Done.
Preparations
All disconnecting and connecting of oil hoses to the tube should be done with tubes
(and systems) that are close to room temperature. If the tube is used hard just before the
change it should be left at least 10 min with the heat exchanger on before disconnecting
the oil hoses.
Shut down the system and turn OFF the fixed mains circuit breaker.
Pull out the detector cooler on the rails. See section 13.30.
If there is not enough space to fully open the side or back door it might be
necessary to remove the roof of the side cabinet in order to replace the
Notice
cooling unit. On that occasion the cable to the electric fan located under
the roof must be disconnected.
2. Disconnect the two self-shutting couplings mounted on the two oil hoses leading from
the cooling unit. Note how the hoses are placed in the cabinet.
3. Remove the self-locking nylon cable tie around the expansion vessel.
4. Reach underneath the cooling unit and unscrew the four Allen screws that hold the
unit into place.
5. Lift out the cooling unit from the side cabinet.
Related drawings
See electrical drawing 1007058 in the service binder.
Preparation
Note which firmware version is used.
Open left and right side doors.
Remove the Display cover (chapter 8.5).
Procedure Calibration
1. Perform all Calibrations except Scan Length, Collimator Height, Compression Height
and Compression Switch, described in chapter 10.1.
2. Perform a Full Calibration according to Quality Control Procedures [1].
Preparation
Check which firmware that is installed (found in
the About-window in the AWS).
Shut down the system by pressing the OFF-button
on the stand. Wait for the system to shut down.
Open right side door.
Procedure Tests
1. Perform Tests after Service and Maintenance (chapter 14).
Tools
Screw driver
2.5, 4, 5 mm Allen key
Torque screw driver with 2.5 and 4 mm Allen bits
10 mm fixed key
Big side cutters (cutting pliers)
Hose Clamp pliers
Bucket or bowl (minimum 5 litres)
Pair of scissors
Notice Never touch the upper part, side panels or the electronics of the detector.
3. Aerate the cooling hoses connected to the detector. Slightly bend one of the cooling
hoses until air is sucked into the system between the hose and the hose connection of
the detector. When no more air is sucked into the hose do the same procedure on the
other hose connection. There should not be any visible coolant in the bend of the
cooling hoses at the top position of the C-arm.
Make sure that the cooling hoses connected to the detector unit are
Caution!
properly drained. If not almost 5 litres of coolant will be depleted.
Even though the connection hoses are empty there is still coolant inside the
Notice
detector.
9. Loosen the four nuts securing the detector unit. (Tip! Move the scan arm manually in
order to access the nuts more easily.)
10. Pull out the detector half way on the four guides.
11. Remove the power cable from the detector, see (B) in Figure 116.
12. Remove the detector by holding it horizontally until the remaining coolant inside the
detector can be poured into the bucket.
Figure 116 Tube bracket (A) and location of power cable (B).
To be able to lift the detector out of the box, loosen only the four
Notice
peripheral bolts. The spare part detector may contain coolant.
1. Protect the two screws (A) with electrical tape (B) according to Figure 117. The tape
should be applied next to the metal edge but not on the flat side (C).
Figure 117 Protect two screws (A) with electrical tape (B).
2. Mount the four studs on the new detector unit. Tighten an expended nut on a stud to
be able to use a torque wrench with 10 mm fixed key to tighten the stud with 3 Nm. If
there is no torque wrench available, tighten the studs until they reach the bottom of
the screw holes.
3. Push the detector half way on the four studs through the scan arm.
4. Connect the power cable.
5. Secure the detector with the four nuts (6.1 Nm).
6. Tape the chink between the detector unit and the scan arm using copper tape. Apply
tape on all sides of the detector unit, as shown in Figure 118.
a. Do not apply tape on the plane part of the scan arm shown in (C).
b. Do not cover the screw holes of the tube brackets.
c. Do not apply tape over the serial number.
d. Cut and fold the copper tape behind the connection pipe on each side, as
shown in (A). (Tip! Cut out a U-shaped piece of tape to fit around the
connection pipes.)
e. Fold the copper tape around the edge, as shown in (B).
f. Make sure that there is no visible chink between the detector and scan arm.
Figure 118 Applying copper tape over the joint between detector unit and scan arm. The image shows
the end result (with tube brackets mounted) when the C-arm is in 0.
7. Connect the fibre cable and re-mount the plastic lid. To protect the fibre cable
connection, move the protection plug from the spare part detector to the replaced
detector.
8. Slip on hose clamps on the hoses and connect the hoses to the connection pipes of the
detector. Secure the hose clamps with the squeezed part pointing (135) downwards
using a hose clamp pliers.
9. Mount the tube brackets. One on each side. Use the shim washers.
10. Power up the system (chapter 5.1, 5.2) and check for any leakages.
11. If necessary fill up the detector cooling unit (chapter 5.3).
12. Continue the start-up sequence (chapter 5.4).
13. Rotate the C-arm to 180 and remount the patient support (chapter 8.7).
14. Manually check the end stops for the scan movement (chapter 15.2) and check that
the detector does not touch the patient support from inside. There should be 0.3 1.3
mm of space between the front of the detector and the patient support.
10. Verify the Image field and X-ray field agreement according to Quality control
procedures [1].
11. Verify missed tissue at chest wall according to Quality control procedures [1].
12. Update the Service & maintenance tab in C&S database with among other the
information from steps 7 to 9, and attach the data files delivered with the spare part
detector. (Tip: Compress the data files to one file.)
Calibration
When the scan motor is replaced, the following calibrations need to be performed
Calibration of the scan in SCS Calibration Tool
Check scan speed
Full calibration
Outgoing Control
Test motor currents of the scan movement motor according to chapter 15.4.3 and 15.5.2.
Tools
2 mm and 3 mm Allen keys
Fixed keys 10 mm and 13 mm
Procedure
1. Loosen the four Allen screws of the axis coupling (A), see Figure 120.
2. Slide the axis coupling as near to the scan motor as possible.
3. Remove the M8 nut (B) using 10 mm and 13 mm fixed keys.
4. Remove the four MC6S M4x12 screws (C).
5. Lift off the old scan movement ball screw with nut.
Tools
1.5, 2 mm and 3 mm Allen key, Torx T9
Fixed keys 12 mm and 13 mm (and/or spanner)
90 Torque wrench with 3 mm Allen bits
0 Torque wrench (15 Nm)
Torque screw driver with 1.5 mm Allen bits
Protective gloves
Procedure
1. Place the scan movement ball screw (A) with nut (B), see Figure 121.
2. Tighten the M8 nut (C) on the scan movement screw (A) at 15 Nm.
3. Centre the scan coupling (D) so that equal lengths of the motor axis and scan
movement screw are inside the scan coupling.
4. Tighten the four M2 Allen screws. (E) at 0.35 Nm.
5. Mount the four MC6S M4x12 screws (F). Do not tighten the screws yet.
6. Rotate the scan screw (A) by hand until the nut (B) is as close as possibly to the
motor bracket (G).
7. Tighten the four MC6S M4x12 (F) at 2,9 Nm.
8. Lubricate the scan screw with 1015566 Grease KSS MSG No.1 Lithium.
Parts
1015558 Scan movement nut holder.
Tools
1.5 mm and 3 mm Allen keys.
90 Torque wrench with 3 mm Allen bits
Torque screw driver with 1.5 mm Allen bits
Preparation
Remove C-arm covers
Rotate C-arm to 180
System in emergency stop
Figure 122 Replace scan movement nut holder that has been upgraded in field.
5. Loosen the two Allen screws (G) securing the readhead (H) and the two cable clips
(I) securing the readhead cable to the scan movement nut holder (E).
Figure 123 Remove scan movement nut holder from linear guide and scan movement base.
6. Remove the eight Allen screws (J) securing the scan movement nut holder (E) to the
linear guide wagons (K).
7. Lift up the scan movement nut holder (E) and the scan movement base (L) in order to
access the screw (M). See Figure 123.
8. Remove the screw (M) securing the scan movement nut holder (E) to the scan
movement base (L).
9. Lift the scan movement nut holder (E) off the scan movement base (L).
Parts
1015555 Scan movement nut holder.
Tools
1.5 mm and 3 mm Allen keys.
90 Torque wrench with 3 mm Allen bits
Torque screw driver with 1.5 mm Allen bits
Preparation
Remove C-arm covers
Rotate C-arm to 180.
System in emergency Stop
Procedure
Follow Procedure in 13.14.1.
The only difference is that the spacer (C in Figure 122) is not used on systems delivered as
S1. I.e systems delivered with 1015555.
Outgoing control
Follow Outgoing Control in 13.14.1.
Parts
1015559 Scan motor bearing assembly. Field
upgrade.
1010588 Motor Scan Movement
1014247 Grease KSS MSG No.1 Lithium
Tools
1009403 Scan motor fixture.
1.5, 2.5 and 3 mm Allen keys
90 Torque wrench with 3 mm Allen bits
Torque screw driver with 1.5 mm Allen bits
12 mm fixed key.
Preparation
Remove C-arm covers
Rotate C-arm to 180.
System in emergency stop
Procedure remove old motor bearing assembly (i.e. motor bracket with bearing)
1. Disconnect the DC-cable (A) of the scan motor (white quick coupling).
2. Loosen the two Allen screws (B) securing the motor axis coupling (C) to the scan
screw (I).
3. Loosen the nut (D).
4. Remove the three Allen screws (E) securing the motor bracket to the C-arm.
5. Remove the motor bracket (F) from the C-arm.
Figure 124 Replace scan motor bearing assembly (i.e. motor bracket and bearing) for systems
upgraded to S1 in field.
Procedure mount new motor bearing assembly (i.e. motor bracket with bearing)
1. Remove the motor axis coupling (C) from the old motor.
2. Mount the new motor on the new scan motor bearing assembly (i.e. motor bracket
with bearing) using the six Allen screws (G).
3. Pre-mount the motor axis coupling (C) on the new motor. Do not tighten the two
Allen screws (H).
4. Carefully place the motor bracket on the C-arm and push the scan screw (I) through
the bearing while pre-mounting the nut (D) on the scan screw.
5. Mount the motor bracket to the C-arm using the three Allen screws (E). Fasten, but do
not tighten, the screws.
6. Mount the scan motor fixture (J) on the linear guide and secure the fixture to the
motor bracket using the two Allen screws (K).
7. Tighten the nut (D) at 24 Nm.
8. Secure the four screws (B) and (H) at 0.35 Nm.
9. Adjust the position of the motor bracket (F).
10. Tighten the screw (G) in order to align the motor bracket to the linear guide.
11. Secure the motor bracket to the C-arm using the three Allen screws (D) at 5.7 Nm.
12. Remove the scan motor fixture (I) by loosen (K) and (G).
13. Lubricate the scan screw with 1014247 Grease KSS MSG No.1 Lithium.
14. Connect the DC-cable (A) to the white quick coupling.
Parts
1015556 Scan motor bearing
assembly.
1010588 Motor Scan
Movement
1014247 Grease KSS MSG
No.1 Lithium
Tools
1.5 mm and 3 mm Allen keys
90 Torque wrench with 3 mm
Allen bits
Torque screw driver with
1.5 mm Allen bits
Vernier Calliper Figure 125 Scan motor bearing assembly (1015556).
Preparation
Remove C-arm covers
Rotate C-arm to 180.
System in emergency stop
Procedure
1. In order to re-create the physical position of the end stop after replacing the scan
motor bearing assembly, do the following:
a. Use a pen to mark the position of the scan motor bearing assembly on the C-
arm. See red line (A) in Figure 125. This mark will be used later when mounting
and positioning the spare part on the C-arm.
b. Measure the distance (B) on the existing scan motor bearing assembly.
c. Measure the corresponding distance on the spare part. Make sure that the spare
part has the same measure. Adjust the screw on the spare part if necessary.
2. Follow Procedure in chapter 13.15.1.The only difference for (1015556) is that the
scan motor fixture is not needed. Systems delivered as S1 have namely a longer linear
guide mounted on the C-arm and a corresponding spline (C) in the scan motor bearing
assembly.
Tools
1015565 Readhead setting jig 0.8 mm
2.5 mm Allen key
Preparation
Remove C-arm front cover as described in section 8.4.
Remove C-arm side covers as described in section 8.6.
Rotate the C-arm to 180.
Identify if the system is field up-graded from S0 to S1. A field-upgraded system does
not have a spline in the motor bracket. (See images on motor brackets in chapter
13.15.1 and 13.15.2.)
Procedure
1. Disconnect the encoder reader head cable (A) attached to J23 at the circuit board on
the C-arm. See Figure 126.
2. Note how the encoder cable is routed from J23 all the way to the encoder.
3. Note the location of the cable straps securing the encoder cable (and in some places
other cables as well).
It is important that the new encoder reader head and encoder cable are
Notice routed and secured exactly the same way as the old reader head and old
cable.
Figure 126 Replacing scan movement encoder. (The right image shows cable routing of a field-
upgraded system.)
Outgoing Control
1. Check function of scan movement read head (chapter 15.3).
2. Perform a Scan calibration using the SCS calibration tool
3. Perform a Geometry calibration (chapter 11.3.5).
4. Perform a Full Calibration according to the User manual 1007994.
5. Verify the resolution with a line pair phantom according to Quality Control
Procedures [1].
6. Perform an Image field and X-ray field agreement test according to Quality Control
Procedures [1].
Tools
1015564 Scale applicator including a 1mm Allen key
2.5 mm Allen key (for temporarily removing the encoder)
Alcohol (For cleaning the surface before mounting the new encoder scale.)
Preparation
Remove C-arm front cover as described in section 8.4.
Remove C-arm side covers as described in section 8.6.
Rotate the C-arm to 180.
Figure 127 How to position scale applicator and how to route the scale through the scale applicator.
2. Remove the splitter screw from the scale applicator. See Figure 127.
3. Move the scan mechanics to the end stop at the scan motor bracket.
4. Separate the backing paper from the first 40 mm of scale and feed it into the scale
applicator as shown in Figure 127.
5. Remount the splitter screw (A) so that the backing paper pass around the splitter
screw. See Figure 127.
6. Push the scale carefully through the scale applicator to mark indicating the position of
the original encoder scale. Make sure that the encoder scale does not stick to the
mounting surface until it is in position.
7. Attach the end of the scale to the mounting surface with light finger pressure.
Procedure
1. Disconnect the cable of the desired compression pedal and route the cable through the
slot on the lower part of the back panel.
2. In a similar manner route and connect the cable of the new compression pedal to the
same D-sub connector.
3. Close the door of the stand.
4. Test the function of the new compression pedal (up, down, compression done).
Figure 130 Foot switch connectors on terminal plate (X12 and X13).
Procedure - disassembling
1. Disconnect the encoder cable of the compression motor. See 1 in Figure 131.
2. Disconnect the power cable of the compression motor. See 2.
3. Loosen the three Allen screws using a 2.5 mm Allen key.
Procedure - assembling
1. Mount the compression motor with the encoder aligned as shown in Figure 131.
2. Tighten the three 2.5 mm Allen screws at 1.2 Nm using a torque tool.
3. Connect the power cable of the compression motor, see 2.
4. Connect the encoder cable of the compression motor, see 1.
5. Check that the compression motor can move easily between its end positions.
Tools
Needle Nose pliers
Procedure
When attaching the new spring the compression force mass must be held
Notice
up in order to stretch the spring as little as possible.
1. Use needle nose pliers (A) when mounting the spring and hold the spring by its loop
(B) and never by the coil. Use only the tips of the pliers, otherwise the loop may get
flattened. See Figure 132.
Procedure disassembling
1. Disconnect the encoder cable from J17 (A).
2. Disconnect the DC cable (B).
3. Cut the cable strap (C) securing the DC and encoder cables to the C-arm.
4. Cut the cable strap (D) securing the DC and encoder cables to the cable tie base.
5. Carefully route the DC and encoder cables out from the C-arm.
6. Loosen the upper Allen screw (E) of the axis coupling securing the collimator motor
axis.
7. Loosen the six Allen screws (F) securing the motor to the chassis and carefully
remove the motor.
Procedure - reassembling
1. Mount the encoder cable on the new collimator motor. See Figure 134.
2. Secure the encoder cable to the collimator motor using cable straps. See Figure 134.
3. Insert the motor axis of the collimator motor into the axis coupling.
4. Tighten the six Allen screws (F in Figure 134.) of the collimator motor.
5. Tighten the Allen screw (E) of the axis coupling,.
6. Route and connect the DC cable (B) of the collimator motor. See Figure 134.
7. Route and connect the encoder cable (A) to J17 on the circuit board. See Figure
134 for routing.
8. Secure the DC and encoder cables to the C-arm using a cable strap (C).
9. Secure the DC and encoder cables to the cable tie base using a cable strap (D).
10. Remount Front cover, Face protection shield and covers.
Calibrations
1. Collimator height (chapter 10.1.9)
2. Collimator collision (chapter 10.1.12)
Tools
Allen keys 3 and 4 mm
Cutting Pliers
Preparation
Remove the Front cover, Face protection shield and covers. See Section 8 for
instructions.
Press the emergency stop button.
Parts
Part.no Name Qty
1015272 Collimator Mov. upgr. kit L30 1
1015538 Cable strap 75 mm (B) 2
1000776 Cable strap (C) 1
1007258* Encoder cable 1
* A part of 1015272.
Tools
2.5, 3 and 4 mm Allen keys/bits
Preparation
Move the collimator mechanics just under its uppermost position.
Procedure
Notice Do not use parts like washers and screws from old assembly.
1. Mount the encoder cable (A) and the two cable straps (B) on the encoder. Cut the
ends of the straps after tightening.
2. Mount the cable strap (C) on the encoder. Cut the end of the strap after tightening.
3. Mount the preassembled Collimator movement upgrade kit on the scan arm using
the four Allen screws (D). Tool: 4 mm Allen key.
4. Mount the connector bracket on the pre collimator bracket using the four Allen
screws with washers (E). Tool: 3 mm Allen key.
5. Loosen each and one of the four screws of (D), (E) and (F) a little bit. About
revolution, i.e. 90 degrees counter clockwise. (The reason is that the parts should be
secured, but allow a certain amount of movement/adjustment in the following
operations.)
Tools
90 Torque wrench , with 4 mm Allen bits.
4 mm Allen key
Procedure
1. Move the collimator movement mechanics down and up 3-5 times.
(To its lowest position and up just under its top end position.)
This action will adjust the position of the motor bracket.
2. Tighten the two Allen screws at 5.7 Nm (A). (I.e. located diagonal)
3. Move the collimator movement mechanics down and up 3-5 times. The movement
should be smooth throughout the whole movement.
4. Tighten the two Allen screws at 5.7 Nm (B). (I.e. located diagonal)
5. Repeat step 3.
Outgoing control
Move the collimator movement mechanics down and up 3-5 times. (To its lowest position and
up just under its top end position.) This action will adjust the position of the motor bracket.
Figure 138 Adjusting the position of the collimator screw in the connector bracket.
It is very important that the mechanics does not reach the top end position
Notice
during the movement.
Procedure
1. Move the collimator movement mechanics up just under its top end position.
2. Tighten the two Allen screws (A) at 1.2 Nm. (I.e. located diagonal)
3. Move the collimator movement mechanics down and up 3-5 times. The movement
should be smooth throughout the whole movement.
4. Tighten the two Allen screws (B) at 1.2 Nm. (I.e. located diagonal)
5. Repeat step 3.
Outgoing control
Smooth movement of the collimator movement mechanics. If not, loosen the screws a little bit
and repeat this section again.
Parts
Part.no Name Qty
1014247 Grease KSS MSG No.1 Lithium 0.05 1
Tools
90 Torque wrench with 3 mm Allen bits.
3 mm Allen key
Protective gloves
It is very important that the mechanics does not reach the top end position
Notice
during the movement.
Procedure
1. Move the collimator movement
mechanics up just under its top end
position.
2. Tighten the two upper Allen screws at
2.9 Nm (A).
3. Lubricate the whole screw using
1014247 Grease KSS MSG No.1
Lithium 0.05. Use protective gloves.
4. Move the collimator movement
mechanics down and up 3-5 times. The
movement should be smooth throughout
the whole movement.
5. Tighten the two lower Allen screws at
2.9 Nm (B). Figure 139 Adjusting the position of the connector
bracket.
6. Repeat step 4.
Calibrations
1. Collimator height (chapter 10.1.9)
2. Collimator collision (chapter 10.1.12)
3. Automatic Collimator Alignment, ACA (chapter 11.3.2).
4. Full Calibration according to Quality Control Procedures [1].
Preparation
Shut down the system and turn off the main circuit breaker.
Open left and right door of the stand.
Additional material
Two metal clamps.
Procedure disassembling
1. Remove the two clamps mounted inside the left main beam. See Figure 143.
2. Mount the clamps2 under the housing of the C-arm. See Figure 140.
Figure 140 Mounting clamps on (a) the left and (b) right side of the housing.
Warning!
It is extremely important that the two clamps are mounted under the C-arm
height/rotation-unit before removing the height motor. This to prevent the
housing and C-arm to fall headlong to the floor when the motor is
removed.
3. Measure and mark the cables before disconnecting from the connector. It is
recommended to use a multimeter for the measurements.
2
If the metal clamps are not available, place e.g. a wood board between the C-arm housing and the
lower horizontal beam to prevent the C-arm from falling down when the motor is removed.
Procedure reassembling
1. Insert the new motor. (The motor needs to be
rotated so that the pin and gear will mesh.)
2. Secure the motor using the four Allen screws.
See Figure 142.
3. Route and connect the power and decoder
cables. See Figure 144.
4. Remount the upper part of the wire sensor. See
Figure 141.
5. Double check that all parts are properly
connected and secured.
6. Remove the clamps. Figure 142 Removing the height motor.
Circuit diagram
Figure 144 Circuit diagram for height motor and C-arm rotation motor.
Preparation
Rotate the C-arm upside-down (i.e. lowest possible centre of
gravity in order to prevent undesired movements of the C-arm
when the motor is removed).
Shut down the system and turn of the main circuit breaker.
Open left and right door of the stand.
Procedure disassembling
1. Disconnect the pre-assembled power cable of the motor.
2. Unscrew the four Allen screws (MC6S 8.8 M6x16) securing the motor to the housing.
Use extension tool.
3. Remove the motor.
Procedure reassembling
The new motor is inserted in reverse order.
Calibration
Perform a calibration of the C-rotation (chapter 10.1.8).
Outgoing Control
Perform Tests after maintenance or service (chapter 14).
Tools
3 mm Allen key with rounded head.
Torque wrench with 3 mm Allen bits.
Preparation
Push the Emergency Stop button.
Open the right side door of the mammography stand.
Procedure disassembling
1. While holding on to the upper part of the wire (A), remove the Allen screw (B)
securing the wire to the motor housing (C). Save the Allen screw.
2. Let the wire sensor house (D) slowly rewind the wire.
3. Disconnect the cable (E).
4. Remove the two 4 mm Allen screws (F) securing the wire sensor house (D) to the
frame. Lift off the wire sensor.
Procedure reassembling
1. Mount the new wire sensor to the frame using the two 4 mm Allen screws (F). Secure
the Allen screws at 2.9 Nm.
2. Connect the cable (E).
3. Secure the upper part of the wire to the motor housing at 2.9 Nm.
Calibration
Perform a height calibration (chapter 10.1.7).
Outgoing Control
Perform Tests after maintenance or service (chapter 14).
Tools
2.5 and 3 mm Allen key with rounded head.
Preparation
Rotate the C-arm to +45, and run the C-arm to lowest position downward. This is the
ideal position for mounting of the wire sensor.
Push the emergency stop button.
Open both side doors of the mammography stand
Prepare the new wire sensor according to Figure 146.
Be careful not to expose the cable for excessive strain at the connection to
Notice
the wire sensor housing..
Procedure disassembling
5. Remove the Allen screw (A) securing the wire to the cable rotation pipe (B). Save the
Allen screw.
6. Let the wire sensor slowly rewind the wire.
7. Disconnect the cable connector (E), see Figure 148.
8. Remove the screws (D) holding the cable clamp (C) and the wire sensor. Lift of the
wire sensor.
Procedure reassembling
1. Mount the new wire sensor with mounting plate (F) and the cable clamp (C) to the
motor housing (G) using the two 4 mm Allen screws with flange (D).
2. Route the cable trough the cable clamp (C) such as it is positioned under the wire
sensor as indicated in Figure 148.
3. Secure the Allen screws (D) to the motor housing (G).
4. Connect the cable connector (E).
5. Route the wire around the cable rotation pipe (B). Ensure that the wire is positioned
within the groove. See Figure 147.
When securing the wire to the cable rotation pipe, be careful not to
Notice damage the aluminium threading. Use Loctite to prevent the Allen screw
from loosening.
6. Secure the upper part of the wire using the Allen screw (A). Apply a small amount of
Loctite 243 to prevent the screw from loosening.
Calibration
Check the wire position in the cable rotation pipe by moving the C-arm angle
Perform a C-rotation calibration (chapter 10.1.8).
Outgoing Control
Perform Tests after maintenance or service (chapter 14).
Figure 149 Replacing the fan unit. (The image shows main cabinet without back plate. It is
not necessary to remove the back plate in order to replace the fan unit.)
Procedure - disassembling
1. Mount the fan unit on the two screws.
2. Tighten the screws.
3. Connect the power cable of the fan unit.
Tools
19 mm fixed key or spanner
Hose clamp pliers
Side cutter
Additional material
PTFE tape
Hose clamps
Always wear safety gloves and protective goggles when handling the water
coolant mixture.
Caution!
Do not operate the detector cooler without enough cooling fluid in the
tank.
Procedure disassembling
1. Close the ballofix couplings of the hoses located near the main cabinet.
2. If the foldable rails were removed for storage on site at installation, mount the rails
back on the rails in the side cabinet.
3. If the SMC detector cooler is used, unscrew the four screws of the bracket holding the
detector cooler and lift off the bracket.
4. Pull out the detector cooler on the rails, see Figure 150.
Figure 150 Detector cooler pulled out on the rails in work position.
5. Tap the detector cooler from cooling fluid, see section 9.2.18. This must be done in
order to prevent leakage during transport.
6. Disconnect the fluid hoses from the connection pipes on the back of the detector
cooler.
7. If the Lauda detector cooler is used: Disconnect the power cable of the detector cooler
from F10 and the X6-PE terminal. Note how the cables are connected and routed. If
the SMC detector cooler is used, disconnect the power cable from the back of the
detector cooler.
8. Gently pull out the cable from the cable guides.
9. Roll the detector cooler back into the side cabinet.
10. Angle the rails and roll the detector cooler down the rails, see Figure 151.
11. Place the detector cooler in the packing of the new spare part detector cooler. Since
the cooler might leak if it tilted, make sure that the top of the packing is marked up
in order to prevent leakage.
Procedure reassembling
1. If the available mains voltage is not 230 V, make sure that the mains transformer T1
is configured for the available voltage, see System Installation Manual [4].
2. Roll the new detector cooler on the angled rails into the side cabinet, see Figure 151.
Position the rails into work position and roll the detector cooler out on the rails, see
Figure 150.
3. If a Lauda detector cooler is replaced with a SMC detector cooler, put the hose
clamps around the fluid hoses and connect the fluid hoses to the detector cooler. For
the Lauda detector cooler, the outlet connection pipe (A) should be mounted at 7
oclock in order to enable free operation of the tap (B), see Figure 152. Use a pair of
hose clamp pliers to tighten the hose clamps.
Figure 152 Mounting the connection pipes of the Lauda detector cooler.
4. If the Lauda detector cooler is used, route the power cable of the detector unit inside
the side cabinet and connect the cable to F10 and the X6-PE terminal. If the SMC
detector cooler is used, connect the power cable to the back of the detector cooler.
Note: Verify that the ground connection is maximum 0.1 .
5. Roll the detector cooler into the side cabinet.
6. Unscrew the four screws of the foldable rails, remove the rails and store them on site.
7. If the SMC detector cooler is used, screw the bracket holding the detector cooler back
in place.
8. Open the ballofix couplings.
9. Fill the detector cooler with cooling fluid and start the system, see chapter 5.
10. Check that the temperature set point is set to 20.0C. If not, adjust the set point to
20.0C according to the instructions in sections 10.5.1 (for Lauda) or 10.6.2 (for
SMC).
11. Perform tests after maintenance and service, see chapter 14.
Procedure
1. If the detector cooler has side panels, unscrew the two screws in the upper corners of
each side panel, disconnect the yellow/green ground cables and lift off the side
panels.
2. Unscrew the four screws holding the detector cooler top cover in place. Disconnect
the yellow/green ground cable and lift of the top cover.
3. Note the positions of each of the cables connected to the old control board, see
illustration and table below.
Connector Connected Position Cable
component colour
X1 Display 1 White
2 Brown
3 Green
4 Yellow
X2 Transformer 1 Brown
2 Black
3 Blue
4 -
X3 On/off switch 1 Violet
2 -
3 Black
4 -
X4 On/off switch 1 Blue
2 Brown
Mains cable 3 Blue
4 Brown
X5 Pump 1 Brown
2 Blue
3 -
4 -
X6 Compressor 1 Brown
2 Black
3 Grey
4 -
4. Disconnect each of the cables connected to the control board by pressing down the
corresponding locking tab with a flat screwdriver or similar and pulling out the cable.
5. Unscrew the four screws in the corners of the old control board and lift off the old
control board. Dispose the old control board according to local regulations.
6. Position the new control board and screw the four screws in the corners of the control
board.
7. Connect each of the cables to the new control board. Make sure to connect each of the
cables to the same position as on the old control board. Carefully pull the cable to
make sure that it is properly connected.
8. Re-connect the yellow/green ground cable of the detector cooler top cover, and screw
the top cover back on. Note that no side panels should be mounted.
Preparation
Shut down the system and turn off the main circuit breaker, F2.
Turn of both incoming and outgoing water.
Figure 153 Mounting the water cooling unit to the front door of the side cabinet.
Check all hoses and connections for leakage. (In case of leakage,
Notice
immediately turn the cooler OFF and solve the problem.)
5. Connect electrical cables from the water cooling unit to:Terminal X6-PE and fuse
F14.
6. Turn on the main circuit breaker, F2 and power up the system.
7. Perform tests after maintenance and service. See chapter 14.
Preparation
Power down the system and switch off the fixed main switch.
Open the side doors of the main cabinet.
Remove the display cover, see chapter 8.5.
Remove the SCS board and box, see steps 1-4 in chapter 13.8.
Figure 154 Screws securing the power supplies to the frame. Four screws for the upper
power supply (A), and four screws for the lower power supply (B).
When the lower power supply unit (B) has been replaced:
1. Verify the PSU output voltage 48V to the detector unit and 24V and 5V (2%) on the
K13 connector.
2. Compare the status of the LEDs of the SCS with Explanation of LEDs on the SCS
unit found in section 20.4, i.e. look for potential errors related to the power supply
and/or miss fitted cables.
Preparation
1. Power down the system and switch off the fixed main switch.
2. Open the right side door of the main cabinet.
Procedure
1. Follow instructions for ESD-protection (chapter 13.1).
Disconnect the cables to the old board and remove the board.
Mount the new board and connect the cables.
Remount Front cover, Face protection shield and covers.
Perform Tests after Maintenance and service (chapter 14).
Preparation
1. Shut down the system (chapter 6.1).
2. Cut the power to the system (chapter 6.2).
3. Open the front door of the side cabinet.
Procedure
1. Disconnect the desired relay by carefully pulling the relay out from the quick
coupling.
2. Set the correct time on the new relay.
3. 120 seconds for shut down relay (A).
4. 20 seconds for power up relay (B).
5. Mount the new relay.
6. Start up the system (chapter 5.1, 5.2 and 5.4). Check the operation of the start-up
relay by timing the relay after pressing the ON-button on the main cabinet. (The small
green LED of the power-up relay starts flashing. After 20 seconds the LED should be
lit, i.e. fixed light.)
7. Close the rear door of the side cabinet.
8. Perform tests after maintenance or service (chapter 14).
Preparation
1. Shut down the system (chapter 6.1).
2. Cut the power to the system (chapter 6.2).
3. Open the rear door of the side cabinet.
Procedure
4. Disconnect the two wires under the thermostat.
5. Remove the thermostat.
6. Set the temperature to 50C on the new thermostat.
7. Mount the new thermostat.
8. Re-connect the two wires.
9. Close the rear door of the side cabinet.
10. Start up the system (chapter 5.1, 5.2 and 5.4).
11. Perform tests after maintenance or service (chapter 14).
Procedure
1. Turn off the stand and cooler by tripping the F10 and F15 fuses in the side cabinet.
2. Roll the detector cooler onto the rails.
3. Remove the side panels of the cooler by unscrewing the two screws. Make sure you
have enough room on both sides of the cooler to be able to work in a good manner.
Note: The side panels are not to be reassembled anymore store on site if possible or
throw away according to local recycling guidelines!
Figure 162 Electrical connection of fan within the starter box of compressor
5. Open the cable connection box positioned at the compressor. To open the box, open
the click-lock at the bottom of the box using a flat screwdriver, see figure below.
7. Important: Before removing the old motor assembly, mark the position of the motor
holder on both sides using a permanent marker or similar. This will help out a lot in
the alignment process.
Figure 163 The Fan blade is fixed with a central screw and positioned by a catching mount.
11. Mount the fan blade onto the new motor.
12. Mount the new motor into the motor holder.
13. Mount the motor back in the cooler. Position the motor using the marks made in step
7.
14. Align the fan assembly in the cooler by spinning the fan blades by hand and make
sure that no blades touch the condenser metal opening. Reposition if necessary until a
good position is found, that is, no blades are touching the condenser opening.
15. Connect the new motor wires to the correct terminals in the connection box; use
correct crimp terminals if needed.
16. Close the connection box.
17. Ensure that the fan motor is running and that the air is sucked into the cooler through
the condenser when restoring F10. Also, pay attention to any noise as described in
step 14.
18. If step 17 is unsuccessful go through the steps above again to locate the fault.
Verification
19. Restore F15 and let the system start up normally.
20. Perform a Full Calibration when the detector has completed its warm up period
(15 minutes).
21. Roll the detector cooler, without its side panels, back into the side cabinet.
22. Perform Tests after maintenance and service according to the Service Manual for
respective modality.
23. Report any unwanted behaviour to product support.
Expected result
Tests after maintenance and service are completed successfully and the system operates
normally for at least 30 min.
Typically this would release the fibre optic transceiver but the mechanics
Notice
may prevent the locking mechanism to open.
2. Grab the handle on the fibre optic transceiver and pull it straight out while using
gentle force. If it is not removed easily, use a screwdriver to push the locking
mechanism downwards in order to release the fibre optic transceiver. A clicking
sound should be heard. Then remove the optic transceiver.
Be careful when doing this so that the detector or the fibre optic
Caution
transceiver is not damaged.
Lock the fibre optic transceiver by folding the handle backwards until it makes a clicking
sound.
Preparations
Power down the system and switch off the main switch.
Close the ballofix couplings near the main cabinet before emptying the detector cooler,
see section 9.2.18.
Remove the top cover and the right side panel, see section 8.11.
Illustration
Figure 164 Detailed view of temperature sensor PT1 and PT1 connector. Parts indicated in the
illustration are: PT1 connector (A), mainboard (B), temperature sensor (C), threads (D)
and grommet (E).
Do not apply the PTFE tape to the first thread. Otherwise, the PTFE tape
will spread to the circulating circuit and cause the product to fail. To
Tip
prevent dust from getting caught in the threads, check there is no dust on
the threads before the sensor is mounted.
Preparations
Power down the system and switch off the main switch.
Close the ballofix couplings near the main cabinet before emptying the detector cooler,
see section 9.2.18.
Remove the top cover and the right side panel, see section 8.11.
Illustration
Figure 165 Detailed view of temperature sensor PT2 and PT2 connector. Parts indicated in the
illustration are: PT2 connector (A), mainboard (B), temperature sensor (C), threads (D)
and grommet (E).
Do not apply the PTFE tape to the first thread. Otherwise, the PTFE tape
will spread to the circulating circuit and cause the product to fail. To
Tip
prevent dust from getting caught in the threads, check there is no dust
stuck to the threads before the screw is mounted.
Interval
Replace the pump at least after 20 000 hours of operation.
Preparations
Power down the system and switch off the main switch.
Close the ballofix couplings near the main cabinet before emptying the detector cooler,
see section 9.2.18.
Remove the top cover and the right side panel, see section 8.11.
Illustration
Figure 166 Detailed view of pump. Parts indicated in the illustration are: four screws holding the
support panel (A), support panel (B), connector of pump cable (C), pump cable (D), power
board (E), ground cable (F), fitting A (G), clip (H), fitting B (I), screws holding the pump
(J), grommet (K).
Interval
Replace the fan at least after 20 000 hours of operation.
Preparations
Power down the system and switch off the main switch.
Remove the top cover, the left side panel and the right side panel, see section 8.11.
Illustration
Figure 167 Detailed view of fan. Parts indicated in the illustration are: fan motor cable (A),
compressor cable connector (B), ground cable (C), capacitor (D), screws holding the fan
(E), fan (F).
Preparations
Power down the system and switch off the main switch.
Close the ballofix couplings near the main cabinet before emptying the detector cooler,
see section 9.2.18.
Remove the top cover and the left side panel, see section 8.11.
Illustration
Figure 169 Detailed view of level switch. Parts indicated in the illustration are: level switch pin LS1
(A), mainboard (B), nut (C), tank lid (D), grommet (E).
Always wear protective gloves when in contact with the detector cooling
Caution
fluid.
1. Disconnect the level switch pin LS1 (A) from the mainboard (B) using the level
switch tool SMC.
2. Unscrew the nut (C) using a 21 mm fixed key or a spanner.
3. Unscrew the tank lid (D) and pull out the level switch from the tank.
Mount the top cover and the left side panel of the detector cooler back in place.
Screw the tank lid back on.
Open the ballofix couplings near the main cabinet.
Start up the system, see chapter 5.
Preparations
Power down the system and switch off the main switch.
Close the ballofix couplings near the main cabinet before emptying the detector cooler,
see section 9.2.18.
Remove the top cover and the right side panel, see section 8.11.
Illustration
Figure 171 Detailed view of pressure sensor. Parts indicated in the illustration are: Connector of
pressure sensor cable (A), pressure sensor (B), threads (C).
2. Make an exposure:
a._________________
a. Start the Acquisition workstation
software
b._________________
b. Open an examination and create a
test patient
c._________________
c. Make an exposure of a
suitable test phantom.
d._________________
d. Check that the exposure
indicators work correctly:
green light at compression
ready
yellow light during exposure
e._________________
e. Verify that the image quality is
acceptable and check that the X-
ray prevent cover does not mask
the X-ray beam near the chest
wall.*
7. Vehicle installation:
a. Check the bolts securing the
following units to the vehicle
chassis: mammography stand, side
cabinet, lead glass and AW-table
b. Check for leakage of cooling fluid
from the detector cooler.
Tools
Vernier calliper (or ruler)
1015473 L-shaped piece of cardboard (A)
Fine pen, preferably a pencil
Figure 172 Measure the deviation of the collimator screw relative the scan arm.
Procedure
1. Move the collimator movement mechanics to its top end position (B), see Figure 172.
Hold the cardboard against the end of the collimator screw (D) and the inner edges (E) of the
cardboard against the scan arm (F).
Mark the lateral position of the collimator screw on the cardboard, both parallel (G) and
perpendicular (H) to the scan arm. (Figure 172 B illustrates the collimator screw when the
collimator mechanics is in its top end position).
Move the collimator movement mechanics to its lowest position (C).
Repeat step 3.
Measure the distance between the two marks on the cardboard parallel to the scan arm using a
vernier calliper (or ruler).
Repeat step 6 for the lines perpendicular to the scan arm.
Expected results
The end of the collimator screw must not deviate more than 2 mm in any direction.
Procedure
1. Manually move the scan beam towards its two end stops, (A) and (B) respectively and
verify that there are no obstacles along the way. E.g. that
a. the detector touches the patient support before the end stop (C)
b. the scan screw mechanics reaches the motor bracket before the end stop. (D)
Perform trouble shooting if step 1 fails.
c. If end stop A fails try to add skims (1008310) behind the tube bracket.
d. If end stop B fails, try first to add skims (1008310) behind the tube bracket
and if necessary move the motor bracket according to chapter 20.6.1.
Procedure
1. Move the scan movement mechanics back and forth. Check that the green LED (A) of
the readhead is lit throughout the whole movement.
See trouble shooting (chapter 20.6.2) if the test fails.
Procedure
1. Start the Service Tool.
Choose the tab Service and click on Measurements.
Click on Scan speed.
Click Yes (in order to perform a new scan).
Choose min scan speed.
Note the values of deviation, overshoot and max speed variation fast/slow (A in Figure 175)
respectively and compare the values with the reference values (B in Figure 175).
Tip: Use the table in Figure 176 for logging and evaluating values.
Click OK to close the window and Cancel (i.e. not saving results) to return to the tab
Measurements.
Repeat step 3-7 for max scan speed.
Repeat step 3-8 for the C-arm in -90 and +90.
Outgoing Control
The values must not exceed the reference values, see (B) in Figure 175.
Handling irregularities
If the scan speed test fails, check motor currents according to chapter 15.4.1.
If the currents are too high it might indicate:
That there is dirt on the screw. If so clean the screw.
That the detector unit touches the patient support.
That the scan movement parts are not mounted properly.
Outgoing control
Max 1.7 A when moving the collimator up from patient support (i.e. 2300 => 27) in 0.
Max 1.5 A when moving the collimator towards patient support (i.e. 27=>2300) in 90.
Outgoing control
Continuous motor current during movement: 4.5 A or less. Must never exceed 5.5 A.
Outgoing control
Continuous motor current during movement: 2 A or less. Must never exceed 3.1 A.
Procedure
1. Choose the tab Runtime in the SCS Calibration tool.
Select the desired motor to move. See (A) in Figure 177.
Press Goto (B).
Enter the desired new position of the motor. (This will be the starting position when logging
the motor current.)
Enter the desired Speed. (The movement will start immediately after Speed has been entered.)
Procedure
6. Choose the Runtime tab in the SCS Calibration tool.
7. Select the desired motor to move (A).
8. Press Goto (B).
9. Enter the desired new position of the motor.
10. Enter the desired Speed. (The movement will start immediately after speed has been
entered.)
11. Choose Stop log (C).
12. Choose View Log Graph(s) (D).
Compare the Graph with the corresponding reference image found in chapter 15.4.8.
Home = 27
Down = 2300
Table 4 Example of table for evaluating "idle" currents of the scan movement motor.
Figure 186 How to check "idle" currents of the compression movement motor.
Table 5 Example of table for evaluating "idle" currents of the scan movement motor
Compression motor current < 2 A (OK/Fail)
Position Test #1 * Position Test #2 *
C-arm angle 0 1480 0 1480
0
* Repetition needed due to fluctuating measurements/values.
15.6Radiation leakage
Tools
Fluorescent paper (1008279)
Procedure
1. Place fluorescent paper on the patient support. Make sure that the piece of paper
partly covers the front of the patient support.
2. Turn off the light in the room.
3. Perform a Still with x-ray exposure from Service Tool and at the same time look at
the fluorescent paper. Check that only the pattern from the collimator slits is visible,
that is, that no radiation except for the primarily radiation down through the X-ray
prevent cover is emitted (see Figure 187).
4. If there are other lit areas or spots, start trouble shooting for radiation leakage.
Figure 187 Fluorescent paper on the patient support, Still with x-ray exposure. No radiation leakage
can be seen, only the pattern from the collimator slits.
Outgoing control
No radiation leakages from the X-ray prevent cover.
Figure 188 SCS with JTAG connector and reset button marked.
Procedure
1. Stop the CCS service
2. Start the Calibration tool.
3. Log in with password Mamea.
4. Click on the Parameters tab. (If not already selected.)
5. Make sure that Automatic update is checked.
6. Check that the text "SCS connected." is shown in the lower left corner.
7. Verify that the parameter values are not only zero. (In case of connection problem, try
again from step 1.)
8. Choose Save to file.
9. Give the file a filename that contains the current date, e.g.
SCS_parameters_20071107.txt.
10. Close the SCS Calibration tool.
11. Start the CCS service.
Procedure
1. Start the Calibration tool.
2. Log in with password Mamea.
3. Click on the Parameters tab (if not already selected).
4. Make sure that Automatic update is checked.
5. Check that the text "SCS connected." is shown in the lower left corner.
6. Select Advance mode in the Settings menu.
7. Choose Load from file...
8. Browse and choose the desired file.
9. In the Events window, check that the correct file has been loaded and that the
parameters have been set and verified.
10. Choose Store permanently.
11. Under Settings, store current calibration for the Load Cell Amplifiers (LCA). Press
Yes when asked to store current calibration.
12. In the Events window, check that the parameters have been stored permanently.
13. Close the SCS Calibration tool.
14. Reset the system by pressing the reset button on the right side of the main cabinet.
17.2IPB
17.2.1 Installing IPB firmware
The IPB JTAG1 interface is used to program the IPB firmware into the flash memories
located on IPB. Three flash memories are used for storing IPB firmware.
Preparation
Note the firmware version in case of re-installation of the same firmware.
In case of firmware upgrade, the SCS and HVG units or CCS software may also need
an upgrade. Refer to the software release information for a complete list of the
firmware/software that must be installed at the same time.
Procedure
1. Connect the Xilinx Platform Cable USB to the computer and to the JTAG 1
connector of IPB.
Outgoing control
Perform a Daily Quality Control, according to Quality Control Procedures [1].
Preparation
In order to access the JTAG connector of the DU, the left C-arm cover must be
removed as described in section 8.6.
Note the firmware version in case of re-installation of the same firmware.
In case of firmware upgrade, the SCS and HVG units or CCS software may also need
an upgrade. Refer to the software release information for a complete list of the
firmware/software that must be installed at the same time.
Procedure
1. Connect the Xilinx Platform Cable USB to the computer and to the JTAG
connector of the DU.
Repeat steps 14 16 for the other two Flash devices you have assigned files to.
Disconnect the Xilinx Platform Cable USB from the JTAG connector of DU.
The new firmware is now loaded into Flash memory on DU. To make DU use the new
firmware switch off the modality and switch it on. You can do this immediately or
perform other updates directly and switch the modality off and on later.
Outgoing control
Perform a Daily Quality Control, according to ref [1], Quality Control Procedures.
17.4SCS
17.4.1 Installing SCS firmware
Preparation
Note the firmware version in case of re-installation of the same firmware.
In case of firmware upgrade, the IPB and HVG units may also need an upgrade. Refer
to the software release information for a total list of the firmware that must be installed
at the same time.
Perform backup of the SCS parameters. See section 16.3.
To avoid problems with a bug in the Xilinx Virtex-4 circuit (FPGA) on the
SCS board, the second flash memory (device #2) must always be
Notice
programmed before the first flash memory (device #1) is programmed, see
step 11 and step 14 in the following Procedure.
Procedure
1. Connect the Xilinx Platform Cable USB to the computer and to the JTAG
connector of SCS. See A Figure 193.
Figure 193 SCS with (A) JTAG connector and (B) reset button marked.
Start iMPACT and press Cancel in the Project dialog that pops up.
17.5HVG
17.5.1 Required equipment
1 computer with Windows (preferably Windows XP) and usb-port (if using a laptop it
should be connected to power to prevent it from shutting off during programming)
1 Spellman HVG Programming Kit (programming cable, usb-RS485 converter and
power adaptor)
Extension cord for powering the programming cable and computer (if the generator is
not very near a power outlet)
1 flashlight
1 non-conductive thin object, e.g. a mechanical pencil made of plastic, a small plastic
screwdriver or a small Allen key with tape on
Software: Spellman Mamea GUI (USB version)
Software: Spellman Firmware Upload GUI (USB version)
Software: Installation CD for the usb-RS485 converter
Software: DSP and/or FPGA firmware file for update (DSP-file starts always with M
e.g. M0045005.S2 and FPGA-file starts with P e.g. P0009009.S2).
It can be chosen to program only DSP, only FPGA or both but when
programming both the generator have to be restarted in between
according to the following procedure. Note that the generator
Caution!
settings HAS TO BE RESTORED after any programming as
described in section 17.5.3. Neglecting to do this before connecting
the generator back to the system can burn the filament on the tube!
Procedure
1. Switch off F1 and F12 in the side cabinet (three phase main and auxiliary power to
the HVG).
4. Connect the programming cable to the HVG. On L30 this is easiest done by removing
1007160 Cable communication SCS-HVG from the SCS K2:HVG (if needed the
SCS K2:AWP can be temporarily removed to easier reach the SCS-HVG cable).
Connect the programming cable to the HVG-SCS cable, see Figure 195.
Connect the power adaptor (5 VDC) to a power outlet, 230 VAC main.
Switch on the single phase power to the generator.
Check that the connection works and that it is possible to read out parameters by following
the next few steps. Start the Mamea GUI. Choose the COM-port found in step 3 above.
Open the Help menu and click Version. Check the firmware version. It should be the same as
in the preparation step.
Click the Configure Filament Stand By-button in the upper left, see Figure 196.
Figure 196 Mamea GUI, main window. Configure Filament Stand by marked in the upper
left.
Click on Get standby status, see Figure 197. You should read 2 A and Filament Standby is
ON.
Figure 197 Mamea GUI, after pressing Configure Filament Stand by. Get Standby Status is
marked
Close the Mamea GUI.
Switch off the single phase power to the generator (fuse F12).
Set SW1:1 in the generator to Open (programming mode) see Figure 198 and Figure 199
b.
Figure 198 LED and SW1 on the HVG. SW1 can be reached through the hole with the
LEDs
Figure 199 SW1 settings in different modes. SW1:8 are open or closed depending on X-ray
tube (open for RAD70B and closed for RAD70).
Switch on the single phase power to the generator, (fuse F12).
Start the Firmware Upload GUI (USB version) and make sure that the correct COM-port is
chosen; this can be done either in the Port setting in the GUI or by changing the first number
in the Commcfg-file next to the GUI-file with was created the first time you choose settings
for the GUI on your computer.
The other settings should be as according to Figure 200. Click Save after altering the settings.
You should now see the main window of the GUI according to Figure 201. The Altera FPGA
and TI DSP buttons are disabled out at this point.
Click Load File to choose firmware file.
For programming the FPGA, select the file with name starting with a P e.g. P0009009.S2
and click Open. A popup confirming that the file has been loaded into the application properly
appears, click OK.
When returning to the Firmware Upload GUI the Altera FPGA button is activated. Click on it
and confirm with Yes to start the programming.
Make sure that the computer does not go into any kind of sleep or energy
Notice
saving mode during the programming.
Check that all the LEDs on the HVG are lit during the programming. If they are not lit
during the programming, the device is not properly programmed and the procedure
has to be re-done.
If the LEDs are not lit during FPGA-programming, please check that
correct COM-port is set and that all cables are connected correctly! Do
Notice
not shut of the generator! The LEDs should go out when the programming
has been completed.
It takes about one minute for the DSP to clear its memory before programming starts. When
programming the DSP, the LEDs continue to look normal (big and small green LED shining
and one small red LED blinking, compared to FPGA where all LEDs shine). Wait until the
programming is complete and a window with Programming OK appears before moving on
to the next step.
When the programming is completed and the progress window has closed. Click on the OK
button on the confirmation that programming is OK. Close the Firmware Upload GUI by
clicking on the X in the upper right corner.
Switch off the single phase power to the generator (fuse F12).
Set SW1:1 in the generator to Closed.
Also this firmware has to be transferred inside the generator. Switch on the single phase
power to the generator (fuse F12) and leave it on for 20 seconds, then switch it off again.
If Brake anode is checked, i.e. the anode is braked after each exposure,
Notice wait at least 30 seconds after completed exposure before pressing
Compression Done.
Perform if the procedure done directly after a reprogramming: Perform a Full calibration, see
Quality control procedures [1].
Perform if the procedure done directly after a reprogramming: According to Quality control
procedures [1], check:
Tube output
Tube Voltage
Air Kerma reproducibility
Perform a Daily Quality Control. If this looks OK, the system is ready to use.
Procedure
1. Switch off fuses F1 and F12.
Configure SW1 (see Figure 198 and Figure 199 b.) depending on the X-ray tube type:
If the tube is of RAD70-type: Set SW1:8 to Closed.
If the tube is of RAD70B-type (spare part no 1012515): Set SW1:8 to Open.
Set SW1:2 to Open, Restore mode (see Figure 199c).
Switch on fuse F12 and wait 10 seconds.
Switch off fuse F12.
Set SW1:2 to Closed, Normal mode.
Switch on fuses F1 and F12.
Note X-ray tubes item and serial number and the switch setting in the modalitys Service
history record.
Solution
A quick calibration, which
identifies and disables instable
channels, should eliminate these
artefacts.
Solution
Figure 204 Loss of X-ray output.
Investigate the log-files to see what has occurred.
Check all connections and/or replace the broken
18.3.2 Ripple
Description
This type of artefact appears when the surface of the
rotating anode contains some kind of unevenness. Due to
the rotation of the anode, the image will contain some
kind of pattern depending on the severity of the
unevenness. The artefact is intermittent and more often
visible for quick scans and in the back of the images.
Solution
If the software version is below 8.1 P2 the system should
be upgraded. If the ripple is disturbing in normal images,
the X-ray tube needs to be replaced. Figure 205 Close-up of ripple
artefact. (Scan
See also section 11.4.1 for instructions on how to direction left-right.)
perform a ripple test in order to detect and estimate the
severity of ripple.
Solution
Perform Quick calibration to mask the sensor. (The masking
performed by the Quick Calibration will be removed upon
next Full Calibration.) If the problem remains, manual Figure 206 Detector burst.
masking need to be performed, see section 18.6. (Scan direction
top-town.)
18.3.4 Unstable X-ray output
Description
This type of artefact can appear when the
preheat time is set too short or if the tube
or generator have regulation problems. It
can also appear because of uneven scan
speed or mechanical problems.
Solution
Check the preheat time settings.
Check Scan speed.
Check mechanical parts involved in
scan motion.
Solution
Lift up X-ray prevent cover according
to section 8.1 in order to see if any
loose objects are visible inside the
collimator. If so, remove the object(s),
otherwise contact product support.
Solution
Wipe clean the patient support, the compression
paddle and the phantom (if used). If the dots
remain, there could be something stuck in one
of the above mentioned parts. Remove one part
at a time and make exposures to locate the
defect. Please contact product support.
Solution
If the phenomenon is seen over the entire image and in all images the blurriness is probably
not due to patients moving. Instead, check for strange sounds or and/or abnormal scan arm
movement in trying to locate the problem. Moreover, contact product support.
Also, when reporting an image quality problem try to remember the following:
Get as much information about the problem as possible, e.g. answer the questions
above.
Attach examples of the artefact (often screenshots will do).
Create a backup file of the channel mask by clicking on Backup CCS Mask to File
under Tools Channel mask viewer in the Service Tool. See section 11.8. Attach the
created file (ChannelMask.Array2D) with the report.
18.6Manual masking
Line artefacts that reoccur several times in the same place indicate instable detector parts that
should be masked permanently. There are three ways to accomplish this: entering the instable
ASICs in CCS local configuration or masking via the Raw data viewer.
Figure 211 Accessing the Raw Data Viewer with the latest image.
The normal procedure of masking channels in the Raw Data Viewer is to scroll through the
image line by line, see 6 in Figure 212. Areas are masked/unmasked by clicking and
dragging with the mouse.
Manually masked areas are marked dark red, see 3. Areas that have been automatically
masked by Quick Calibration are marked light red, see 4.
The latter ones can serve as a help when finding and permanently masking problem areas.
1. Sets the function of the left mouse Check box for showing masked channels.
button.
Check box for showing the background
Area in the image (black) where the image (i.e. the composite image). Would
chosen line does not provide image data. appear in the area marked 2 in the figure
above when activated.
Manually masked image area (dark red)
for the chosen line. Button for rebuilding the image in order to
preview the result of masking/unmasking
Automatically masked image area (light
channels.
red) for the chosen line.
Save the image mask in order to be used
The magnification of the image window.
for future image acquisitions.
Chosen line (0-20) or Image (i.e. the
Close the Raw Data Viewer without
composite image).
changing the channel mask.
Function of left mouse button
Channel Information = Information about channel and ASIC is shown when clicking in
the image
Mask Channels = Channel masked when clicking and dragging in the image
Unmask Channels = Channel unmasked when clicking and dragging in the image
After masking and unmasking in the raw data viewer, an AEC calibration
Notice
has to be performed.
20. Troubleshooting
20.1X-ray stability test (Ripple test)
A ripple measurement is performed by both using the Service Tool and visually study images
in detail. See chapter 11.4.1.
20.2Communication errors
See section 11.4.5 on page 110 for information about communication errors.
20.3Failed calibrations
The result of calibrations can be found in the file
/ServiceTool/log/CalibrationLog.txt or by clicking on ST Log under the tab
Normal in the Service Tool.
Figure 213 Access to the ST Log via the GUI of the Service Tool.
Example:
-----------------------------------------------------
Wed Oct 24 14:00:28 2007: Geometry calibration failed
-----------------------------------------------------
Scan geometry error before calibration = 25.04um
hs = 673.926mm
zs = -138.424mm gamma = 0.000867062radians
Scan geometry error = 41.9343um (<= 30um) FAIL!
Scan start = 7584um (>= 0um)
Scan stop = 268000um (<= 276000um)
Text in parenthesis shows the required result. The calibration will fail if the result of the
calibration does not meet the requirements.
The calibration in the example above failed due to insufficient precision in scan geometry.
The reason for this is that the starting point of the ACA movement is not
far enough out to the left. This problem can be solved by applying
washers (0.5 mm 1 mm) between the lower part of the ACA mechanics
and the collimator base according to the pictures.
This will shift the motion of the collimator to the left. (It also has effect to
minimize the play of the motor drive, but this is less efficient.) Note that
temporary loss of power to the ACA means that the alignment process
needs to be restarted, so that the motor homing process is started once
again. Loss of power could otherwise lead to a behaviour similar to the
one described here.
Turning the It may happen that turning the knobs does not move the collimator in the
knobs does not expected way. If this is the case, make sure that the springs behind the
have any effect knobs are able to apply force at all desirable knob positions. Adjust the
nuts behind the springs if necessary. If the springs do not apply any force,
the effect of turning the knobs will be highly unpredictable.
Problems It may happen during the first or last (fixation) step that even though the
during last or pictures appears to be normal, it is still impossible to adjust the collimator
first step: bad to a position that is accepted. This can happen if the detector has a
margin geometrical deformation or if a sensor is not properly aligned under the
slit of the post-collimator. If the left/right margin during the first steps is
close to default limit 70 m (below 80 m, say) it is likely that you will
run into problems when trying to fix the collimator with margin above 70
m. In this case, make a local configuration of the Service Tool: (1) Close
the Service Tool. (2) Place the ACAReducedLRMargin.def in :\Program
Files\Sectra\ServiceTool\local_config. The file shall have the following
content:
AutoCollimAlign {
DataAnalyzer {
limitMarginLR = 40
}
}
(3) Start the Service Tool and run Collimator alignment. When the
alignment is completed, check with Verify Alignment that the left/right
margin is above 0 m in 90 and 90, with the x-ray cover mounted. If
the margin is below 0 m, please contact product support.
Figure 214 Explanation of LEDS and push buttons on the SCS unit.
20.4.1 LEDs
On top of the SCS unit there are 15 LEDs. All motor-drivers have a diagnosis LED and if the
voltage supplying a motor is below 21V, the LED lights up to indicate Error, i.e. red light.
This might also be used to locate blown fuses.
There is also a diagnosis signal to indicate that the 24V is off, this will be used during upstart
to test that the emergency stop works properly.
LED # Description Indication
1 Z-OFF
2 C-ROT OFF Diagnostics LEDs.
3 AUX OFF Green light = voltage ok
4 Collimator OFF Red light = blown fuse or voltage below 21 V
(e.g. in emergency stop)
5 Scan OFF
6 Compression OFF
7 3,3 V Green light = voltage ok
8 5V Green light = voltage ok
Green light = voltage ok
9 24 OFF Red light = Loss of power supply for motors (e.g. in
emergency stop)
10
11 FPGA status See table below on FPGA LEDs.
12
Green light shows HVG tunnelling activity and is lit
13 HVG status
when receiving and sending data from/to the HVG.
13 HVG status Green light = HVG initialized correctly
This LED is switched on in the beginning of the main
loop and switched off at the end of the main loop. If
14 Software active
the LED is lit continuously or not lit at all this would
indicate that the software is not running.
Green light = n/a
Red light = Any of the emergency stop buttons have
15 Emergency stop button
been pressed
Not illuminated = Normal operation
FPGA LEDs
10 11 12 Active motor
OFF OFF OFF No motor active
OFF OFF ON C-arm rotation motor active
OFF ON OFF Compression motor active
OFF ON ON C-arm height motor active
ON OFF OFF Scan motor active
ON OFF ON Collimator motor active
ON ON OFF Aux motor active
ON ON ON N/A
20.4.2 Push-buttons
SCS features three pushbuttons for use in testing, debugging etcetera and one pushbutton for
manually resetting the board, only the reset button will be reachable from outside the
enclosure.
20.5Drawings
Drawings can be found under the corresponding tabs in the service binder delivered with the
system.
For technical data and mechanical overviews of the different Philips MicroDose
configurations, water/air cooled and American/Asian/European, see:
1. Chapter 2 in this document and
2. Technical specification, ref [3].
20.6Scan movement
20.6.1 Move scan motor bracket
This action might be needed if the test of scan movement end stops (chapter 15.2) fails.
Preparation
Rotate the C-arm to 180
Stand in Emergency stop mode
Procedure
1. Loosen (do not remove) the three Allen screws (A) securing the motor bracket to the
C-arm. See Figure 216.
Mount the scan motor fixture (B) on the linear guide and secure the fixture to the motor
bracket using the two Allen screws (C). The fixture will ensure that the motor bracket remains
parallel to the linear guide.)
Adjust the position of the motor bracket.
Tighten the screw (D).
Secure the motor bracket to the C-arm using the three Allen screws (A) at 5.7 Nm.
Remove the scan motor fixture (B).
Preparation
Rotate the C-arm to 180
Stand in Emergency stop mode
Tools
1015462 Readhead setting jig 0.8mm
Procedure
1. Align the scan wagon to the two holes (A).
Loosen the two screws (B) securing the readhead
Insert the readhead spacer (C) under the readhead (D) for a rough alignment.
Tighten the two screws (B) a little bit to allow adjustment.
Remove the readhead spacer (C).
Move the scan mechanics back and forth observing the LED of the readhead. The LED should
be green along the whole movement. If the LED turns red the position of the readhead need to
be adjusted.
When the LED is green through the whole movement, secure the readhead to the scan wagon
with the two screws (B).
Figure 217 Adjust the position of the scan movement readhead using the readhead spacer (1015462).
20.7Collimator movement
20.7.1 If the collimator screw is deviating more than 2 mm
Solution
Try to loosen the mechanics and remount according to the appropriate sub-chapters in chapter
13.24.
20.8Compression movement
20.8.1 If Compression Force calibration fails
Solution
1. Run the calibration again. Make sure that all compression measurements are
performed directly after compressing to the desired force level, if over-compressed,
decompress fully before recompressing.
If it still fails, follow the Motion smoothness guide.
Also, when reporting an X-ray system problem try to remember the following:
Get as much information about the problem as possible, e.g. answer the questions
above.
Attach CCS-log and possibly examples of the problem (could be screenshots).
Explain what troubleshooting has been done so far.
Attach serial no of tube, serial no and revision of generator, firmware version of
generator.
New tubes can arc the first weeks, especially in the morning. This most often levels out and
arcs will be fewer and fewer. Doing a couple of low kV, low mAs exposures in the mornings
can reduce the arcing.
5. Click the plus sign next to ActiveErrors to open the collapsed list, see Figure 222.
6. Click the Update tree to get the current status, see Figure 223.
Figure 223 Click on Update tree to show the current values in the main window.
Use the Error codes and solutions, section 20.13, to get advice on how to handle the
possible faults listed under Active Errors.
Make sure that the potentiometer works and that the SCS measures it correctly by:
c. Use the stand buttons to rotate the C-arm to 0. Measure the angle manually.
d. Take a note of the C-Rotation potentiometer value (in mV) from the runtime
page in SCT (the stand must be physically at 0).
e. Measure the voltage between terminals 1 and 2 (red probe on terminal 2) on
the potentiometer with a multimeter (see Figure 224). If the measured voltage
(with the multimeter) is negative, move the black probe from terminal 1 to
terminal 3. Take a note of the measured voltage.
f. Rotate the C-arm to 45. Measure the angle manually.
g. Take a note of the C-Rotation potentiometer value (in mV) from the runtime
page in SCT (the stand must be physically at 45).
h. Measure the voltage between terminals 1 and 2 (red probe on terminal 2) on
the potentiometer with a multimeter (Figure 224). If the measured voltage
(with the multimeter) is negative, move the black probe from terminal 1 to
terminal 3. Take a note of the measured voltage.
i. Control the measurements:
i. The voltage measured with the multimeter must have changed with at
least 250mV. If no change has occurred there is a problem with the
potentiometer. Replace the potentiometer and perform a rotation
calibration.
ii. The voltage measured by calibration tool must have changed with at
least 250mV. If no change has occurred there is a problem with
either the cables or the SCS Unit. Make sure that the cables are not
damaged and that they are correctly mounted. If no problem with the
cables can be found, the problem probably lies with the SCS Unit.
Contact Product support.
Perform a C-rotation calibration.
Perform a test of the C-arm rotation potentiometer from SCS Calibration Tool. The test is
initiated by selecting Test Potentiometers Check C-arm potentiometer from the menu.
On the graph presented, the range shall be 2 and the absolute value of the offset 1.
If the problem remains, contact product support.
20.11.3 Scan
1. Set the projection to 0.
Perform a calibration mode reset with no start movement in SCS Calibration Tool.
Note down MaxPosition parameter for the Scan motor for later use.
Push the scan arm all the way to the left hand end stop. (Use the arrow button in the SCT if
the scan arm is too hard to move by hand, but be careful!)
Note down the encoder value of the scan motor from the runtime page in SCT.
Push the scan arm all the way to the right hand end stop.
The new encoder value should have changed by more than [MaxPosition] units (the
parameter previously noted down), if it has not then there is a problem with the scan encoder
signal.
If the previous step was OK, push the scan arm all the way back to the left hand end stop.
The new encoder value should be within 300units of the original encoder value at the left
hand end stop. If it is not, then there is a problem with the scan encoder signal.
If you have come this far without a problem then it is probably not a problem with the
encoder signal.
20.11.4 Collimator
1. Set the projection to 0 and remove the compression paddle.
Perform a calibration mode reset with no start movement in SCS Calibration Tool.
This test is designed for a Collimator with an EncResolution parameter (encoder units/cm) of
4000. All the calculations hence forwards use 4000 but if the EncResolution of your particular
stand differs then recalculate the values with that particular EncResolution.
Note down MaxPosition parameter for the Collimator motor for later use.
Push the collimator all the way to the upper end stop.
Note down the encoder value of the collimator motor from the runtime page in SCT.
Push the collimator all the way to the bottom end stop.
The new encoder value should have changed by more than ([MaxPosition] -
[EncResolution]/10) = ([MaxPosition] 400) units, if it has not then there is a problem with
the collimator encoder signal.
If the previous step was OK, push the collimator all the way back to the upper end stop.
The new encoder value should be within ([EncResolution]/10) = 400units of the original
encoder value at the upper end stop. If it is not, then there is a problem with the collimator
encoder signal.
If you have come this far without a problem then it is probably not a problem with the
encoder signal.
20.11.5 Compression
1. Set the projection to 0 and remove the compression paddle.
Perform a calibration mode reset with no start movement in SCS Calibration Tool.
This test is designed for a Compression with an EncResolution parameter (encoder
units/cm) of 5000. All the calculations hence forwards use 5000 but if the
EncResolution of your particular stand differs then recalculate the values with that
particular EncResolution.
Note down MaxPosition parameter for the Compression motor for later use.
Push the compression arm all the way to the upper end stop.
Note down the encoder value of the compression motor from the runtime page in SCT.
Push the compression arm all the way to the bottom end stop.
The new encoder value should have changed by more than ([MaxPosition] -
[EncResolution]/10) = ([MaxPosition] 500) units, if it has not then there is a
problem with the compression encoder signal.
If the previous step was OK, push the compression arm all the way back to the upper end
stop.
The new encoder value should be within ([EncResolution]/10) = 500units of the original
encoder value at the upper end stop. If it is not, then there is a problem with the
compression encoder signal.
If you have come this far without a problem then it is probably not a problem with the
encoder signal.
If the stand is in emergency stop (the display shows: STOP) even though all emergency stop
switches have been deactivated then there is a high chance that one of the cables or one of the
emergency stop buttons have broken. The diagram shown in Figure 226 can be used to help
find the cause of the problem. One can either have the power connected to the SCS Unit and
have all the cables connected and try to find where the 24V disappears in order to narrow
down the search. Another way is to disconnect the cables and measure on the cables and find
which of the cables that has no continuity.
EM-loop 1
EM-loop 2
SCS Unit
Pin 101
Pin 103
24V
AW
K2
P Pin 104
Pin 102
Pin A3
Pin A5
Left Side
On Stand Pin A6
Pin A4
K7
Pin B1
Pin B3
Right Side
On Stand Pin B4
Pin B2
Pin A1
Pin A3
Compression
K16
Action Description
0 Guidance message: An instruction to help the user to perform the correct action.
1 Notice: An important but not critical message.
2 Reset of the system is required to continue.
3 A full calibration is required to continue.
4 Recommendation to contact product support, due to a configuration error or non-
recoverable fault.
5 A DQC (daily quality control) is required to continue.
6 Reset of the system and DQC is required to continue.
7 A DQC is recommended before continuing.
8 A Full calibration is recommended before continuing.
9 A Warm-up procedure is recommended before continuing.
Error Each module has a list of definitions of errors. The error code is passed on to and
collected by the CCS.
The CCS is translating the error message and writes an error message in plain text in the
error-log.
The error code X-X-X is passed on to the AWS by the CCS. The AWS has its own lists of
translations, which are displayed for the user. The AWS translations can be edited.
For the error codes starting with a 4, trouble shooting help can be found in chapter 20.9.
The HVG has detected an 1. Reset the system, try low kV, low mAs
open Thermal Switch in exposures.
the x-ray tube for more 2. Check connector on generator side, it
than 500 ms via pin 5 and breaks easily. Verify that all pins and
7 in the filament cable, contacts in both male and female
this can be due to: connector are on the same height. If
connector is ok, try using a cable known to
Arc (thermal switch
be fault free
signalling disturbed by
4-45-2 Tube temp too high. 3. Check that the oil cooler is functional and
the arc)
running and that fuse F13 in side cabinet
Filament cable or
is on. If the tube is hot you should be able
connector broken
to feel a temperature difference between
Tube is too warm
the two oil hoses going into the tube when
because of limited oil
the cooler is functional. Try to restart the
flow
oil cooler by switching F13.
Temp switch in tube 4. Let the tube cool down and then measure
broken if the temp switch in the tube is still open.
HVG has detected an 1. Reset the system, try low kV, low mAs
open Flow Switch in the exposures.
heat exchanger for more 2. Verify that crimps are securely tightened
than 2 seconds via TB3 in connector
on the generator, this can 3. Verify that crimps are intact and securely
be due to: tightened
Arc (flow switch 4. Check that the oil cooler is functional and
signalling disturbed by running and that fuse F13 in side cabinet
the arc) is on. If the tube is hot you should be able
4-46-2 Tube oil cooler flow low to feel a temperature difference between
Cable to TB3 on top of
the two oil hoses going into the tube. Try
generator fits poorly
to restart the oil cooler by switching F13.
Poor connection in
Adjust hoses so they are quite loose
X5:13 or X5:14
where the connections are located and
No oil flow because of
that there is as few sharp bends as
strain on the hoses possible. Make sure that there is no strain
quick connections when the C-arm is rotated.
Air in oil cooling 5. Leave C-arm upright overnight, rotate C-
system arm to -60 degrees
Exposure parameters Verify that kV, mA and exposure time is within
4-47-2 Wrong exposure settings
invalid valid ranges
The HVG has detected a
DC Rail Voltage less than
445Vdc, this can be due
to: 1. Check that three phase cable is securely
Mains power cable not tightened at both
Dc Rail (Rectified 3-
4-48-2 tightly fitted or 2. Check that the fuse F1 is on. Measure that
phase below limit)
Fuse in side cabinet there is incoming power to the generator.
off. 3. Improve the input power situation.
Incoming power is
lower than required in
the specification.
1. Keep doing exposures, preferably on low
Arc in the tube
mAs, low kV at start.
Arc in high voltage
2. Clean connectors and apply clean high
connector on HVG or
voltage grease.
4-49-2 Tube Arc. tube
3. Leave C-arm upright overnight, rotate
Air in tube housing (oil C-arm to -45 degrees.
level low or air 4. If problem persists consider changing the
bubbles) X-ray tube.
The HVG detected a
filament current over:
RAD70
5 A for more than 50 ms
or
4.4 A for more than 300
ms
RAD70B
6 A for more than 100 ms 1. Verify that the preheat time is correctly
or configured.
5.5 A for more than 300 2. Restore the generator parameters by
4-50-2 Filament overcurrent ms switching SW1:2 according to section
17.5.2.
this can be due to: 3. Consider changing the tube.
4. Consider changing the HVG.
Preheat time set
wrongly.
Preheat current value
in the generator set to
high
X-ray tube faulty and
needing too much
filament current to give
the right tube current.
HVG faulty.
Data acquisition
subsystem error: Incompatible functionality
Command not in CCS and IPB, Might be
6-0-4 Contact product support.
supported. Please incompatible versions of
contact product the CCS and IPB.
support.
Data acquisition
subsystem error:
6-1-4 Command failed. Internal IPB error. Contact product support.
Please contact product
support.
Data acquisition
subsystem error: DU IPB has no contact with
6-2-4 not responding. Please the DU. Check cable Contact product support.
contact product between IPB and DU.
support.
Data acquisition
Incompatible functionality
subsystem error:
in CCS and IPB. Might be
6-8-4 Unknown command. Contact product support.
incompatible versions of
Please contact product
the CCS and IPB.
support.
Data acquisition
Parameter from CCS to
subsystem error: Invalid
IPB is invalid. Might be
6-9-4 parameter. Please Contact product support.
incompatible versions of
contact product
the CCS and IPB.
support.
Data acquisition
Incompatible functionality
subsystem error: Invalid
in CCS and IPB. Might be
6-10-4 command format. Contact product support.
incompatible versions of
Please contact product
the CCS and IPB.
support.
Data acquisition
IPB has no contact with
subsystem error: DU Temporary solution: Try again.
6-100-4 the DU. Check cable
timeout. Please contact Contact product support.
between IPB and DU.
product support.
Data acquisition
subsystem error: DU
Packet error from IPB to
communication error Temporary solution: Try again.
6-101-4 DU. Check cable between
IPB to DB. Please Contact product support.
IPB and DU.
contact product
support.
Data acquisition
subsystem error: DU
Packet error from DU to
communication error Temporary solution: Try again.
6-102-4 IPB. Check cable between
DB to IPB. Please Contact product support.
IPB and DU.
contact product
support.
Data acquisition
Communication error
subsystem error: DU
between DU and IPB.
communication error
6-103-4 Might be incompatible Contact product support.
UNKNOWN. Please
versions of the DB, LB
contact product
and/or IPB.
support.
Packet errors in the DU
Data acquisition (problem communicating)
subsystem error: DU within the detector
communication error between DB and LB). Temporary solution: Try again.
6-200-4
DB to LB 0. Please If the problem is Contact product support.
contact product temporary: disturbance.
support. If the problem remains:
Contact product support.
AL04 Temperature of cooling Check the ambient temperature and the Continue Temperature sensor
fluid at fluid outlet temperature of the cooling fluid. PT1
AL05 High pressure at cooling 1. Ensure that the cooling fluid flow is Stop Temperature sensor
fluid inlet 5 l/min or more. PT2
2. Reduce the ambient temperature or
heat load.
AL06 High fluid pressure at Check the hoses for obstructions. Stop Temperature sensor
cooling fluid outlet PT1
AL07 Abnormal pump Restart the detector cooler and check Stop Pressure sensor,
operation the pump operation. Pump
AL08 Rise in fluid pressure at Check the hoses for obstructions. Continue Pressure sensor
fluid outlet
AL09 Drop in fluid pressure at 1. Restart the detector cooler and Continue Pressure sensor,
fluid outlet check that the pump is operating. Pump
2. Check the fluid level in the tank.
AL10 High compressor intake Check the temperature of the cooling Stop N/A
temperature fluid returning to the detector cooler.
If remaining, contact product support.
AL11 Low compressor intake 1. Check the flow of the circulating fluid. Stop Pump
temperature 2. Make sure that the circulating fluid in
AL12 Low super heat the evaporator is not frozen. Stop Pump
temperature 3. Use a 15% ethylene glycol aqueous
solution if operating with a set
temperature lower than 10C.
AL13 High compressor Reduce the ambient temperature or heat Stop Dustproof filter,
pressure load. Fan
AL15 Pressure drop on the Replace the detector cooler. Stop Detector cooler
high pressure side of the
refrigerant circuit
AL16 Pressure rise on the low Reduce ambient temperature or heat Stop Dustproof filter,
pressure side of the load. Fan,
refrigerant circuit If remaining, replace the detector cooler. Detector cooler
AL17 Pressure drop on the Check the flow of the cooling fluid. Stop Pump
low pressure side of the
refrigerant circuit
AL18 Compressor overload 1. Leave the detector cooler for 10 min. Stop Detector cooler
2. Restart the detector cooler.
3. If remaining, replace the detector
cooler.
AL19 Communication error Contact product support. Continue N/A
AL20 Memory error Stop N/A
AL21 DC line fuse Stop N/A
3
The default operation setting. Indicates if the operation of the detector cooler stops or continues after
the alarm has occurred.
4
Spare parts that might need to be replaced.
3 4
Code Cause Action Operation Spare parts
AL22 Temperature sensor Replace temperature sensor PT1. Stop Temperature sensor
failure at cooling fluid PT1
outlet
AL23 Temperature sensor Replace temperature sensor PT2. Stop Temperature sensor
failure at cooling fluid PT2
inlet
AL24 Compressor Replace the detector cooler. Stop Detector cooler
temperature sensor
failure
AL25 Pressure sensor failure Replace pressure sensor. Stop Pressure sensor
at cooling fluid outlet
AL26 Pressure sensor failure Replace the detector cooler. Stop Detector cooler
at cooling fluid inlet
AL27 Compressor pressure Replace the detector cooler. Stop Detector cooler
sensor failure
AL28 Operating time for the Replace the pump and reset the Continue Pump
pump has increased operating time.
20 000 h.
AL29 Operating time for the Replace the fan and reset the operating Continue Fan
fan has increased time.
20 000 h.
AL30 Operating time for the Replace the detector cooler. Continue Detector cooler
compressor has
increased 50 000 h.
Comments:
5
The serial number of the stand can be found (1) in documentation delivered with the system, (2) on a
label on the door of the side cabinet or (3) in the About box found in the application.
If the agreement with customer is that Philips should perform the semi-annual and annual
quality control procedures, also perform the testing quality control procedures, described in
the Quality control procedures [1].
Objects affecting
Calibration SCS X-ray focus
Hardware DU (T b lli t t )
Status SCS
ASIC
Event calibration
Comp paddle Comp
calibration paddle
Collimator
alignment
Detector Objects in the
startup beam path
(Fixed added filter, Scan geometry
i )
Flatfield Flatfield*
timeout
Detector
Flatfield geometry
check
AEC calibration*
System
ready
www.philips.com/healthcare
healthcare@philips.com This Medical Device meets the provisions of
the transposition of the Medical Device
Directive 93/42/EEC within the country of
Manufacturers address the EU Authorized Representative for the
device concerned.
Philips Digital Mammography
Copyrights and all other proprietary rights in
Sweden AB any software and related documentation
(Software) made available to you rest
Smidesvgen 5 exclusively with Philips or its licensors. No