Documente Academic
Documente Profesional
Documente Cultură
edited by
Cost-Effective Teacher Harold H. Harris
University of MissouriSt. Louis
St. Louis, MO 63121
Division of Chemical Education www.JCE.DivCHED.org Vol. 86 No. 5 May 2009 Journal of Chemical Education 617
In the Laboratory
80
red
70
50
40
30
green blue
20
10
0
46000 48000 58000 60000
Intensity of Pixel
Figure 2. Image produced by a transparency scan of a microplate Figure 3. Results of image analysis by ImageJ of an elliptical selection
with ammonium standards (020 mg L1 N in NH4+) developed of a microwell on a microplate determining nitrate by Szechrome
according to Sims et al. (3). Parallax due to the sensor location on NAS. The red and green channels were more tightly grouped than
the vertical axis is apparent in the left- and rightmost columns and the blue channel, and the red and green channels were also more
suggests the use of ellipses rather than spheres to describe regions of sensitive to changes in nitrate concentration than the blue channel
interest within each microwell. (Shown in color on p 533.) (see Figure 4).
Photo, Long Beach, CA, $449), a 200 dpi, 48-bit (216 values per an ImageJ results spreadsheet. These values were then manually
primary color) digital image of the 96-microwell plates (available transferred to a standard electronic spreadsheet. After examining
from Fisher Scientific) was obtained. We built a simple opaque the images and their central tendencies for the effects of occa-
polycarbonate template that covers the entire scanner bed with sional bubbles and dust particles, we selected median values for
a cut-out to hold microplates in the same location (Figure 1) so red, green, and blue channels of the region of interest to calculate
that the scanned images could be cropped automatically to the absorbance values, as described in the introduction. The macro
size of the microplate with the software provided by the scanner is written as a text file, which permits simple editing of the fields
manufacturer. The scanned image was saved in TIFF format, to further customization.
a lossless format. An example of a scanned image is shown in For comparison, we measured absorbance values for each
Figure 2. microplate with a standard microplate reader (Bio-Tek Power-
The scanned microplate image was opened with public wave XS microplate scanner; Winooski, VT, $12,400) at the
domain image analysis software, ImageJ version 1.40 (8). This appropriate wavelength for each method, within five minutes
software splits all pixels within an image into their red, green, of the digital scan.
and blue components. To automate the image analysis process,
we developed a custom macro.1 This macro initially requires as Hazards
input the x and y coordinates (in pixels) of the top left micro
well, the number of rows and columns of microwells on the The use of a desktop scanner as a microplate reader presents
microplate, the number of pixels between them, and the size and no innate hazards. A method for analyzing phosphate is pre-
eccentricity of the elliptical region of interest for each microwell. sented in the online material with the following hazards: Ammo-
Use of a template to hold microplates in a repeatable location nium molybdate is an irritant to eyes, skin, and the respiratory
on the scanner ensures that these values do not need to be re system. Antimony potassium tartrate is toxic and can be fatal if
adjusted for subsequent scans. An elliptical region of interest was swallowed. It is also an irritant to eyes, skin, and the respiratory
chosen because the three-dimensional objects viewed from the system. Monopotassium phosphate is an irritant to eyes, skin,
central axis of the scanner created a slight overlap of the image of and the respiratory system. Sulfuric acid is corrosive and inhala-
walls and well bottoms. The magnitude of the overlap increased tion can produce damaging effects to the upper respiratory tract.
with distance from the central axis of the scanner bed (Figure2). Ingestion or contact with skin can cause severe burns. Contact
When conducting image analysis on the scanned images, one with eyes can cause severe burns or blindness.
could either analyze pixels in a smaller circle avoiding the image
of the walls or use an ellipse as the region of interest instead of a Results
complete circle; we chose the latter option because it allowed a
greater number of pixels to be analyzed per well without chang- Image analysis of the colorimetric reactions in the mi-
ing the scanner resolution. croplates produced tightly clustered red, green, and blue values
The macro was designed to export the mean, median, for each microwell (Figure 3), which allowed for collection of
and mode values of red, green, and blue color channels from single central values of each to be transformed into absorbance.
a 424-pixel ellipsoid at the center of each microplate well into The minimum requirement for a transparency scanner image
618 Journal of Chemical Education Vol. 86 No. 5 May 2009 www.JCE.DivCHED.org Division of Chemical Education
In the Laboratory
A B
0.8
0.07
0.7
red channel red channel
green channel 0.06 green channel
0.6
blue channel blue channel
0.05
Absorbance
Absorbance
0.5
0.04
0.4
0.03
0.3
0.02
0.2
0.1 0.01
0.0 0.00
0 5 10 15 20 0.0 0.2 0.4 0.6 0.8 1.0
0.18 0.9
red channel
0.16 green channel 0.8
blue channel
0.14 0.7
Absorbance
Absorbance
0.12 0.6
red channel
0.10 0.5 green channel
0.08 0.4
blue channel
0.06 0.3
0.04 0.2
0.02 0.1
0.00 0.0
0 5 10 15 20 0.00 0.05 0.10 0.15 0.20 0.25
as a substitute for traditional absorbance measurements is the are found here to be quite usable with the red primary owing to
ability to produce calibration curves using at least one of either the broad absorbance peak of the colored complex. Somewhat
the red, green, or blue channels for established colorimetric more difficult to explain to students is the inverse nature of the
reactions. For ammonium, bromide, nitrate, and phosphate color measured. For example, for ammonium ion analysis, the
ions, this clearly occurs as seen in Figure 4. In some cases, such yellow blank is the result of the addition of red and green pri-
as phosphate, all three primaries were correlated with analyte maries. As the absorption of the red wavelengths increases with
concentration and the line with steepest slope should be chosen increasing ammonium concentrations, the remaining mixture
as the most sensitive. In other cases, one or more of the primaries appears greener.
are relatively unresponsive to the analyte concentration, but in For these colorimetric reactions, the use of a scanner oper-
all cases at least one of the three primaries was correlated with ating in transparency mode may be favorably compared to the
analyte concentration. The primary color that is most sensitive dedicated microplate reader (Figure 5), sometimes matching
for each analytical method generally follows the standard wave- the exceptional dynamic range of the microplate reader but at
length chosen for analysis by spectrophotometry. For example, a much lower cost.
the red primary (~630 nm) is close to the wavelength usually
used for ammonium by phenate, 667 nm, and green (~530 Conclusions
nm) is close to that used for bromide analysis by phenol red,
575 nm. Interestingly, the phosphate measurements, usually The results of this study indicate that digital image
read at a wavelength of 850 nm, that is, in the infrared range, analysis can replace a microplate reader for several microscale
Division of Chemical Education www.JCE.DivCHED.org Vol. 86 No. 5 May 2009 Journal of Chemical Education 619
In the Laboratory
A B
bromide (green channel, 575 nm) phosphate (red channel, 850 nm)
0.8 ammonium (red channel, 667 nm) 0.14 nitrate (red channel, 570 nm)
0.7 0.12
Absorbance (Scanner)
Absorbance (Scanner)
0.6
0.10
y = 0.009 + 0.204x y = 0.002 + 0.226x
0.5 r 2 = 0.997 r 2 = 0.999
0.08
0.4
y = 0.028 + 0.151x
0.06
0.3
r 2 = 0.999
0.1 0.02
0.0 0.00
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7
Figure 5. Relationship between absorbance of the red, green, or blue channel compared to absorbance as determined by microplate reader
at the appropriate wavelength for (A) ammonium by the phenate method and bromide by phenol red and (B) nitrate with Szechrome NAS
and phosphate by ascorbic acid.
620 Journal of Chemical Education Vol. 86 No. 5 May 2009 www.JCE.DivCHED.org Division of Chemical Education