Sunteți pe pagina 1din 7

1008 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO.

3, JUNE 2010

Multi Pixel Photon Counters (MPPC) as an


Alternative to APD in PET Applications
A. Nassalski, Member, IEEE, M. Moszynski, Fellow, IEEE, A. Syntfeld-Kazuch, Member, IEEE,
T. Szczesniak, Member, IEEE, Swiderski, Member, IEEE, D. Wolski, T. Batsch, and J. Baszak

AbstractThe performance of multi pixel photon counters count rate and capacitance of the MPPC detectors [5]. Unfor-
(MPPC) of 3 mm 2 3 mm size, with 14400 and 3600 pixels, were tunately, most of these features because of their contradictory
studied by means of the signal from a laser light pulser and
using the 3 mm 2 3 mm 2 20 mm LSO pixel scintillator. Special
nature cant be realized in one device.
attention was paid to measure number of fired pixels, generated Recently Hamamatsu has introduced on the market a
by the light of pulser and that of the LSO crystal, using a direct 3 3 mm MPPC with 3600 and 14400 pixels, better fitted
method of a comparison of the light peak position in the pulse to make tests with scintillators of the size assuring acceptable
height spectrum with that of the single photoelectron. The tests detection efficiency for gamma rays.
of the LSO crystal showed 1550 6 80 fired pixels per MeV in This paper will focus on two aspects in characterization of
the MPPC with 14400 pixels assuring a good linearity of the
response up to about 1 MeV energy of gamma rays absorbed in
the MPPC detectors, first using a laser diode pulser and in the
the LSO crystal. Energy resolution of 14.8% for 662 keV gamma second part, a study with LSO (Lu SiO :Ce) pixel crystal cou-
rays from 137 Cs source and a time resolution of about 850 ps for pled to the MPPC. A considerable effort has been made to mea-
511 keV annihilation quanta were limited by a rather low number sure the amount of detected light as estimated by the number of
of the fired pixels compared to the number of photoelectrons in fired pixels (comparable to the photoelectron number in other
photomultipliers.
detectors) and the pulse height resolution. In scintillation tests
Index TermsDetector linearity, multi pixel photon counters the measurements covered the determination of the energy res-
(MPPC), scintillation detection. olution, non-proportionality of the light yield and the time res-
olution for 511 keV annihilation quanta for both types of the
3 3 mm Hamamatsu MPPC detectors.
I. INTRODUCTION Effects relating to the cross-talk and afterpulses in the MPPC
were not studied in this work, because they were completely
ACING the challenge to introduce a solid state photode-
F tector into different fields of photon detection, from med-
ical to high energy physics applications, a lot of efforts have
masked by a high noise level.

been made to develop new multi pixel photon counters (MPPC). II. EXPERIMENTAL DETAILS
The MPPC is a solid state photodetector also known as the sil- All the studies were done using two Hamamatsu MPPC of
icon photomultiplier (SiPM). It is a photon-counting device con- 3 mm 3 mm size, S10362-33-025C and S10362-33-050C,
sisting of multiple avalanche photodiode pixels connected in consisting of 14400 and 3600 pixels, respectively. Two samples
parallel and operating in the Geiger mode. When photons enter of each device were tested. In the first part of the study, the
a pixel while it operates in Geiger mode, the pulse output from MPPCs were illuminated by a red laser diode type HL6501MG
the pixel is constant regardless of the number of photons. This from Opnext, Japan. In the second part of the study, a detector
means that each pixel provides only the information whether or consisting of a 3 mm 3 mm 20 mm LSO pixel crystal, cou-
not it received one or more photons. The output signal from the pled to both types of MPPC detectors, was analyzed. The LSO
MPPC is proportional to the number of excited pixels. Details crystal was coated with Teflon tape and coupled to the MPPC
of the operational characteristics of a SiPM can be found in the with silicone grease. The main MPPC parameters, according
literature [1][7]. to the manufacturer data, are summarized in Table I, and the
Nowadays the manufacturers efforts are focused on improve- main physical properties of the LSO scintillator are collected
ment of overall photon detection, effective area, amount of ac- in Table II.
tive pixels, detection efficiency as well as reduction of dark In all the measurements, the MPPC signal was sent to the
charge sensitive preamplifier [11] and then to the Ortec 460
Manuscript received October 18, 2008; revised April 02, 2009, July 23, 2009, Delay Line Amplifier, selected to reduce a shift of the baseline at
and October 08, 2009; accepted October 08, 2009. Date of current version June the amplifier output caused by a high rate of noise pulses in the
16, 2010.
A. Nassalski, M. Moszynski, A. Syntfeld-Kazuch, T. Szczesniak, . MPPC. A PC-based multichannel analyzer (MCA) Tukan 8 K
Swiderski, D. Wolski, and T. Batsch are with the Soltan Institute for Nuclear [12] recorded energy spectra. Peak positions and their full width
Studies, PL 05-400 Otwock-Swierk, Poland (e-mail: marek@ipj.gov.pl). at half maximum (FWHM) were obtained by applying Gaussian
J. Baszak is with Hamamatsu Photonics Deutschland GmbH, Polish Office,
PL 02-525 Warsaw, Poland (e-mail: jbaszak@hamamatsu.de). fits. In the analysis, the FWHM of the peaks was not corrected
Digital Object Identifier 10.1109/TNS.2010.2044586 for the saturation effect in the MPPC.

0018-9499/$26.00 2010 IEEE


NASSALSKI et al.: MULTI PIXEL PHOTON COUNTERS (MPPC) AS AN ALTERNATIVE TO APD IN PET APPLICATIONS 1009

TABLE I
MAIN PARAMETERS OF THE HAMAMATSU MPPC USED IN THE STUDY

Fig. 1. Separated single photoelectrons obtained with the S10362-33-025C


MPPC detector.

TABLE II The rise time of the MPPC response to the laser pulse is de-
MAIN PROPERTIES OF LSO SCINTILLATOR fined by the shape of the light pulse, while its fall time is the
effect of the large capacitance of the device. This 320 pF capac-
itance, together with the 50 ohms load resistor at the input of
the scope forms a low pass filter with the loading time constant
of 16 ns. It is confirmed at the lower panel of Fig. 2 displaying
the MPPC output signal recorded with an additional 50 ohms
resistor at the anode of the MPPC. The fall time is shorter by
about factor of 2.
The reported shape of the MPPC output pulse is different than
that observed and predicted for MPPC devices with small capac-
itance. The small MPPCs characterize by about 10 times lower
capacitance and then, show two components of the fall of the
output signal [13], distinguished however, due to a much lower
loading integration time constant, below 2 ns. A direct compar-
ison of the response of 1 1 mm and 3 3 mm detectors is
III. RESULTS presented in [5], confirming results of Fig. 2.
In the next part of the study, tests of the linearity of the output
A. MPPC Performance signal were performed. Since each pixel detects only whether
A dark counts pulse height spectrum of the MPPC or not, one or more photons have entered the detector, the total
S10362-33-025C is shown in Fig. 1. The MPPC dark noise number of the fired pixels does not directly correspond to the
spectrum was measured at 69 V bias, corresponding to 1 V number of detected photons. If two or more photons are trig-
over-voltage, and at the temperature in the laboratory of about gering one pixel then the photon detection linearity degrades
20 C. A very high counting rate of dark noise pulses shifts because the number of incident photons becomes larger than the
up the base line at the output of the amplifier introducing number of fired pixels. This is reflected in (1), which presents
a positive offset in the spectrum that degrades pulse height the relation of the number of fired pixels as a function of the
resolution. Thus, the position of the single photoelectron peak number of detected photons for a given number of pixels in the
was corrected for this offset, which was determined from the MPPC [2]
calibration based on the single and double photoelectron peaks.
It was used further, as the reference, to measure the number (1)
of the fired pixels for the light flashes from the laser pulser or
LSO crystal. where
The MPPC response to the light pulse from the laser diode : number of excited pixels,
is presented in Fig. 2, as observed at a digital scope (Tektronix
TDS5034B). The laser was triggered by the short pulse of 3 ns : total number of pixels,
(FWHM) producing light pulse of 6 ns (FWHM), and monitored : number of incident photons,
by the very fast PMT (R5320), characterized by a rise time of
PDE: photon detection efficiency.
the single photoelectron pulse of 700 ps.
1010 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE 2010

Fig. 2. S10362-33-025C MPPC output pulse, the light pulse from laser, as seen by the fast PMT (R5320) and the electrical pulse used to trigger the laser. Upper
panel presents output signal on 50 ohms loading resistor at the input of a scope, while at the lower panel the second 50 ohms resistor was placed at the anode of
the MPPC. Note a faster fall time due to a shorter time constant at the MPPC output.

The photon detection efficiency, PDE, is the product of the MPPCs of 14400 and 3600 pixels show non-linear behavior.
quantum efficiency (QE), the fill factor or geometrical factor However it is clear that the range of linearity is more extended
(GF) and the probability that an incoming photon trigger a in case of the MPPC with a higher number of active pixels.
breakdown (Pt) [2] In the next experiment the MPPC was illuminated by light
pulses from the laser of different intensities and then their pulse
(2) height spectra were stored at MCA. For each position of the
peak, the corresponding number of the fired pixels was deter-
The number of detected photons is described, in turn, by
mined in relation to the single photoelectron peak and its re-
(3) spective FWHM, expressed in the number of the fired pixels
was obtained from the Gaussian fit. The results are presented in
Fig. 3 presents the measured number of the fired pixels versus Fig. 4, where the plot of the measured FWHM versus the square
the number of detected photons, calculated, following (1) and root of the measured number of fired pixels is presented. The
(3). It reflects the deviation from the linear dependence repre- nonlinear response of MPPC is reflected by a deviation of the
sented by the solid line, calculated for the infinite number of FWHM of the continuous line calculated from Poisson statistic,
pixels in the MPPC. Error bars are within the point size. Both for a perfectly linear photodetector. In the case that two or more
NASSALSKI et al.: MULTI PIXEL PHOTON COUNTERS (MPPC) AS AN ALTERNATIVE TO APD IN PET APPLICATIONS 1011

Fig. 3. Number of excited pixels versus the number of detected photons, for Fig. 5. Energy spectra of singe photoelectron and cesium source ( Cs), mea-
both 14400 and 3600 pixels Hamamatsu MPPC, error bars are within the point sured for LSO crystal with 14400 pixels Hamamatsu MPPC.
size.

TABLE III
RESULTS OF LSO GAMMA-RAY SPECTROMETRY MEASUREMENTS

the method used in the scintillation detectors based on the PMT


light readout [14].
Fig. 4. FWHM of the light peaks versus the square root of the fired pixels. In this method the number of photoelectrons is measured
directly by comparing the position of the full energy peak of
-rays detected in the crystal to the position of the single photo-
electron peak. The observed positions of the single and double
photons are triggering one pixel, the photon detection linearity
photoelectrons was used to correct for the offset introduced into
lowers because the number of incident photons is larger than the
spectrum by a high counting rate of noise pulses.
number of excited pixels. Fig. 4 reflects clearly an influence of
Table III summarizes the performance of the detectors when
nonlinear response of MPPC for larger light signals.
coupled to an LSO scintillator in comparison to the Pho-
Fig. 3 presents in fact, the illustration of the nonlinear be-
tonis XP2020Q PMT and the Hamamatsu S8550 APD array
havior of the MPPC depending on the number of pixels since it
[15]. Since the pixel size in the S8550 APD array is only 1.8
is based on the theoretical (1). In turn, Fig. 4 confirms a good
mm 1.8 mm, the tests were done with a 2 mm 2 mm 15
accuracy of the measured number of fired pixels, well related to
mm LSO crystal.
the width (FWHM) of the light pulser peak.
In the third column, the number of photoelectrons is shown
for the tested 3 3 20 mm LSO crystal. Note a lower number
B. Gamma-Ray Spectrometry
of photoelectrons for the 14400 pixel detector in comparison
In this part of the study, the MPPC signal was sent to the to the 3600 pixels detector, also confirmed by a poorer energy
charge sensitive preamplifier [11] and then to the Ortec 460 resolution, collected in the fourth column. It confirms a lower
Delay Line Amplifier operated with 0.25 s integration time fill factor quoted for this MPPC by the manufacturer.
constant. The energy spectra of the single and double photo- To compare the MPPC performance two types of photode-
electrons and gamma rays of a Cs source are presented in tector were selected: a commonly used Photonis photomultiplier
Fig. 5. It allows measuring the number of the fired pixels by XP2020Q, characterized by a photocathode blue sensitivity of
1012 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE 2010

TABLE IV
LIGHT OUTPUT OF THE LSO PIXEL CRYSTAL AND PHOTON DETECTION
EFFICIENCY OF MPPC PHOTODETECTORS

10.1 A/lmF (where F stands for filtered blue sensitivity) [16]


and a Hamamatsu S8550 APD array, previously tested by the
author in [15]. Note the observed by a factor of two and four
higher photoelectron and electron-hole pair numbers for PMT
and APD, respectively, than that of the MPPC and significantly Fig. 6. Non-proportionality characteristics for LSO crystal measured with
Hamamatsu MPPC 14400 and 3600 pixels detector in comparison to the
better energy resolution. XP2020Q PMT.
The measured photoelectron number with the XP2020Q
PMT allows for estimation of the photon detection efficiency
of the tested MPPC devices for the LSO pixel crystal. First
the light output of the LSO pixel was calculated assuming the
integral quantum efficiency of the XP2020Q, following [14],
see Table IV. For the MPPCs the procedure was reversed and
the PDE was calculated based on the measured light output of
the LSO crystal.
The estimated PDEs of about 10% is comparable to that mea-
sured in [3] for SiPM of comparable size of pixels. It could be
improved by a larger over-voltage, however, the bias voltages of
both MPPC was very close to those recommended by the man-
ufacturer, see Table I, corresponding to 1 V of over-voltage. In
the case of the MPPC 050C, the estimated PDE could be af-
fected partly by a nonlinear response for 662 keV peak in LSO,
see Fig. 7 below. A low number of the fired pixels in the MPPCs
is also confirmed by a poor energy resolution, see Table III.
In the Fig. 6, the non-proportionality characteristics of the
LSO pixel crystal, as measured with the MPPC, with 14400 and Fig. 7. Energy resolution (FWHM) characteristics for LSO crystal measured
3600 pixels, in comparison to the XP2020Q PMT, are presented. with Hamamatsu MPPC 14400 and 3600 pixels detector in comparison to the
It confirms that the LSO light signal is within the linear response XP2020Q PMT.
of the MPPC up to about 1 MeV gamma ray energy in case
of the 14400 pixels detector. The bending down of the curve
at about 1 MeV is common for the MPPC and XP2020 PMT, C. Time Resolution Study
characterized by the excellent linearity of the current pulse up To check the usefulness of the MPPC detector for PET in-
to 100 mA. This effect is characteristic for the pixel crystals strumentation, the coincidence time resolution of the tested de-
and it is related to a reduced light output of high energy gamma tectors for 511 keV annihilation quanta was measured using the
rays, probably because of a larger light attenuation. In case of the same LSO pixel crystal. As the reference detector, a truncated
3600 pixels MPPC detector, the lower number of active pixels conical BaF crystal 25 mm in diameter and 15 mm high was
limits the linearity of the detector, below 500 keV gamma ray used. The crystal was coupled to the Photonis XP20Y0 QDA
energy. PMT equipped with a screening grid at the anode. Its time reso-
Fig. 7 presents the energy resolution measured with the LSO lution was equal to ps for 511 keV, as reported in [17].
crystal as a function of incident -rays energy. The curves The timing measurements were performed with a Na -source.
are linear in a double logarithmic scale. The MPPC detectors The energy windows in the side channels of the detectors were
present the same behavior of the energy resolution in the whole set at the 511 keV full energy peaks.
range of energies. The energy resolution measured with the In the time setup, the Ortec 935 four channels Constant Frac-
PMT is superior in the whole measured energy range. It is tion Discriminator (CFD) was used. Note that the CFD was used
evidently associated with a larger number of photoelectrons in the leading-edge mode with a short circuit of shaping delay
produced at the PMT photocathode than that of the MPPC. and its threshold was set just above the noise level. The output
NASSALSKI et al.: MULTI PIXEL PHOTON COUNTERS (MPPC) AS AN ALTERNATIVE TO APD IN PET APPLICATIONS 1013

Fig. 9. LSO light pulse at the output of MPPC with 3600 pixels, as observed
at a digital scope (Tektronix TDS5034B).
Fig. 8. Time resolution spectra for LSO crystal measured with Hamamatsu
MPPC 14400 and 3600 pixels detector in comparison to the XP2020Q PMT.

to the rise time of the output pulse of the MPPC of about 20 ns


TABLE V
TIME RESOLUTION OF THE LSO SCINTILLATOR [16]. It is confirmed in Fig. 9, below, which presents the LSO
light pulse at the output of the MPPC 050C with 3600 pixels.
The rise time of 20.6 ns, seen in Fig. 9, confirms well an
earlier prediction. Thus, a 1 mm 1 mm MPPC is preferable
in this respect, due to 9 times lower capacitance. In the tested
MPPC, the speed of the LSO pulse could be largely improved by
a drastic reduction of the load resistor, below 10 ohms [17]. Both
tested MPPCs give comparable time resolution, which follows
measured numbers of fired pixels. It is well reflected in the same
values of normalized time resolution.
The strength of SiMPs in the fast timing is their excellent time
resolution for 1 photoelectron of about 100 ps [2]. However,
for slow decaying scintillators, like LSO, the time resolution is
more governed by the decay of the light pulse than the time jitter
signals from the CFD channels provided start and stop input sig- of photodetector [17]. Thus, the high number of photoelectrons
nals for the Ortec 567 Time-to-Amplitude Converter (TAC). The is more important than a low time jitter [19]. A low time jitter of
TAC output was recorded by a PC-based multichannel analyzer MPPC will be reflected in timing with fast light pulses of plastic
Tukan 8 K [12]. The output signals were gated by the full energy scintillators or Cherenkov radiation.
peaks of 511 keV -rays. The time spectra are shown in Fig. 8 However, the MPPCs are superior in comparison to the APD
and the results of the time resolution measurements are given in array. The measured time resolution is more than twice better
Table V. than in case of APD arrays. It is the result of dark noise of the
In the third column the number of photoelectrons (fired pixels APD array which limits mainly the time resolution.
for MPPC) corresponding to the 511 keV peaks are listed. The The presented timing study of the tested detectors, confirmed
time resolution corrected for the contribution of the reference a future possibility to introduce the proposed detectors in PET
detector (128 ps) is presented in the fourth column. The last applications.
column shows the time resolution of the tested detectors nor-
malized to the number of photoelectrons. It is not done for the IV. CONCLUSION
APD, since the measured time resolution is mainly determined Multi pixel photon counters are novel type of solid state
by the signal-to-noise ratio at the input of a fast discriminator. photodetectors with significant advantages such as a high gain,
The best time resolution was measured with the XP2020Q fast timing, compactness and insensitivity to magnetic field. All
PMT, three times better that those measured with the MPPCs. It of them make MPPC very attractive in application to nuclear
follows, first of all, a factor two larger photoelectron number ob- medicine and high energy physics.
tained with the PMT. However, the normalized time resolution The present studies showed that the new Hamamatsu
to the photoelectron number, presented in the last column, re- S10362-33-025C/050C MPPC could be easily applied in nu-
flects also a higher speed of the PMT in comparison to the tested clear medicine, in particular in detectors for positron emission
MPPCs. This is the effect of a large time constant of 16 ns due to tomography, as an alternative to APD arrays. A large number of
the 50 ohms load resistor and the MPPC capacitance of 320 pF. pixels in the detector assure a good linearity of the response to
For the LSO light pulse, with the decay time of 40 ns, it leads the light from LSO pixel crystals and a good energy resolution.
1014 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE 2010

A slow output pulse of MPPC for the LSO crystal, associated [9] C. L. Melcher and J. S. Schweitzer, Cerium doped lutetium oxy-
with a high capacitance of the devices, limits the time resolu- orthosilicate: A fast efficient new scintillator, IEEE Trans. Nucl. Sci.,
vol. 39, pp. 502505, 1992.
tion; however, it is still superior in comparison to APDs. [10] M. Kapusta, P. Szupryczynski, C. Melcher, M. Moszynski, M. Bal-
A considerable effort was done to determine the number of cerzyk, A. A. Carey, W. Czarnacki, A. Spurrier, and A. Syntfeld,
fired pixels and its relation to the pulse height and time reso- Non-proportionality and thermoluminescence of LSO:Ce, IEEE
lutions. The estimated photon detection efficiency for the light Trans. Nucl. Sci., vol. 52, no. 4, pp. 10981104, Aug. 2005.
[11] M. Kapusta, P. Crespo, D. Wolski, M. Moszynski, and W. Enghardt,
from LSO crystal showed a value of about 10%, typical for sil- Hamamatsu S8550 APD arrays for high-resolution scintillator ma-
icon photomultipliers. trices readout, Nucl. Instrum. Meth. A, vol. 504, no. 13, pp. 139142,
May 2003.
REFERENCES [12] Z. Guzik, S. Borsuk, K. Traczyk, and M. Plominski, TUKANAn
[1] P. Buzhan et al., Silicon photomultipliers and its possible applica- 8 K pulse height analyzer and multi-channel scaler with PCI or USB
tions, Nucl. Instrum. Meth. A, vol. 504, no. 13, pp. 4852, May 2003. interface, IEEE Trans. Nucl. Sci., vol. 55, no. 1, pp. 231235, Feb.
[2] D. Renker, Geiger-mode avalanche photodiodes, history, properties 2006.
and problems, Nucl. Instrum. Meth. A, vol. 567, no. 1, pp. 4856, Nov. [13] F. Corsi et al., Electrical characterization of silicon photo-multiplier
2006. detectors for optimal front-end design, in Proc. IEEE Nucl. Sci. Symp.
[3] G. Llos et al., Novel silicon photomultipliers for PET applications, Conf. Rec., 2006, pp. 12761280, N30-222.
IEEE Trans. Nucl. Sci., vol. 55, no. 3, pp. 877881, Jun. 2008. [14] M. Moszynski, M. Kapusta, M. Mayhugh, D. Wolski, and S. O. Flyckt,
[4] G. Llos et al., Silicon photomultiplier and SiPM matrices as pho- Absolute light output of scintillators, IEEE Trans. Nucl. Sci., vol. 44,
todetectors in nuclear medicine, in Proc. Nucl. Sci. Symp. Conf. Rec., no. 3, pp. 10521061, Jun. 1997.
2007, pp. 32203223, N14-4. [15] A. Nassalski, M. Moszynski, A. Syntfeld-Kazuch, . Swiderski, T.
[5] C. Piemonte, M. Boscardin, G. F. Dalla Betta, M. Melchiorri, N. Zorzi, Szczesniak, D. Wolski, and T. Batsch, Application of Hamamatsu
R. Battiston, A. Del Guerra, and G. Llosa, Recent developments on sil- S8550 APD array to the common PET/CT detector, IEEE Trans. Nucl.
icon photomultipliers produced at FBK-irst, in Proc. Nucl. Sci. Symp. Sci., vol. 55, no. 5, Oct. 2008.
Conf. Rec., 2007, pp. 20892092, N41-2. [16] S.-O. Flyckt and C. Marmonier, Eds., Photomultiplier Tubes Principles
[6] D. P. McElroy, V. Saveliev, A. Reznik, and J. A. Rowlands, Evaluation and Applications, Sep. 2002, Photonis, Editors.
of silicon photomultipliers: A promising new detector for MR compat- [17] M. Moszynski, M. Kapusta, A. Nassalski, T. Szczesniak, D. Wolski,
ible PET, Nucl. Instrum. Meth. A, vol. 571, no. 12, pp. 106109, Feb. L. Eriksson, and C. L. Melcher, New prospects for time-of-flight PET
2007. with LSO scintillators, IEEE Trans. Nucl. Sci., vol. 53, no. 5, pp.
[7] A. N. Otte, J. Barral, B. Dolgoshein, J. Hose, S. Klemin, E. Lorenz, R. 24842488, Oct. 2006.
Mirzoyan, E. Popova, and M. Teshima, A test of silicon photomulti- [18] S. Seifert, D. R. Schaart, H. T. van Dam, J. Huizenga, R. Vinke, P.
pliers as readout for PET, Nucl. Instrum. Meth. A, vol. 545, no. 3, pp. Dendooven, H. Lohner, and F. J. Beekman, A high bandwidth pream-
705715, Jun. 2005. plifier for SiPM-based TOF PET scintillation detectors, in Proc. Nucl.
[8] Q. Xie, R. G. Wagner, G. Drake, P. DeLurgio, Y. Dong, C. T. Chen, and Sci. Symp. Conf. Rec., 2008, pp. 16161619, NM1-2.
C. M. Kao, Performance evaluation of multi-pixel photon counters for [19] T. Szczesniak, M. Moszynski, . Swiderski, A. Nassalski, P. Lavoute,
PET imaging, in Proc. Nucl. Sci. Symp. Conf. Rec., 2007, pp. 969974, and M. Kapusta, Fast photomultipliers for TOF PET, IEEE Trans.
N18-1. Nucl. Sci., vol. 56, no. 1, pp. 173181, Feb. 2009.

S-ar putea să vă placă și