Documente Academic
Documente Profesional
Documente Cultură
*
Patent pending
The following experiment verifies that arsine Improved Sensitivity
exposure is in fact minimized. The Hach
apparatus was compared to a currently available The larger sample size, combined with the unique
test kit using the system shown below. design of the cap for Hachs reaction vessel,
focuses the arsine gas on a very small surface.
System for Checking Arsine The generated arsine must react with the test strip
Emissions before it can exit the reaction vessel. This allows
for a calibrated range of 0 to 500 g/L (ppb).
Threaded Cap Draeger The calibration scale has distinct steps at 0, 10,
Tube
30, 50, 70, 300, and 500 g/L.
Sulfide Interference
PVC Cylinder 20
20
Unfortunately, during the reduction of arsenic to
Vacuum arsine gas, sulfides are also reduced to form
Pump at
20 Inches hydrogen sulfide (H2 S). Hydrogen sulfide reacts
Hg with the mercuric bromide test paper in the same
way as the arsine gas, causing an interference.
Current methods for removing sulfide interference
AsH3 Gen. entail passing the arsine gas stream through a
Vial
scrubber to remove the hydrogen sulfide. These
Air In
scrubbers are usually made of cotton that has been
soaked in lead acetate solution (zinc and copper
The Draeger-Tube1 detects arsine gas in air. have also been used but have been found to be
The distance that a purple color travels up the less efficient). The sulfide reacts with the lead on
tube quantifies the amount of arsine present. the cotton to form solid lead sulfide, thereby
Samples containing 500 g/L arsenic were tested removing it from the arsine gas stream.
using both methods. The Hach method uses a 50
mL sample and the competitor uses a 5 mL There are some drawbacks to this method. First,
sample. The vacuum pump was set at 20 inches the seal must be tight enough to obligate the
of Hg and allowed to run continually for the full passage of all of the gas through the scrubber.
30 minutes of each test. The results of the Then the rate of gas evolution must be controlled
experiment are shown here. to allow adequate contact time for all of the
sulfide to react. Lastly, the operator must handle
and dispose of hazardous lead acetate and lead
Results of Arsine Exposure Test sulfide.
Apparatus mg/L AsH3 after 30 min.
Competitor with no 0.8 ppm Hachs new method eliminates the need to rely on
test strip lead acetate to remove sulfide interference. The
Competitor with test 0.3 ppm Hach method employs the addition of three unit-
strip dose reagent pillows to convert the sulfide present
Hach with no test > 3.0 ppm in the sample to a form that will not interfere in
strip the test.
Hach with test strip 0.0 ppm
First, sodium phosphate dibasic and potassium
monopersulfate (Oxone2 ) are added to the
Even though it generates more arsine gas overall,
sample. This results in the formation of strong
the new Hach apparatus minimizes the release of oxidizing conditions. Sulfides are oxidized to
arsine gas to the atmosphere, where it might pose
sulfate in which form they no longer interfere.
a hazard.
Next, a mixture of disodium and tetrasodium
EDTA is added to the sample to remove any
residual Oxone and to prevent its interference in
1
Draeger-Tube is a trademark of Draegerwerk
2
AG Oxone is a registered trademark of DuPont
the subsequent reduction and evolution of arsine total arsenic, including organic arsenic, with this
gas. method, modify the procedure as follows:
Tests were run to determine the levels at which Collect the sample in a glass beaker. Add the
sulfide interfered. A sulfide concentration was disodium phosphate and Oxone. Place the
said to interfere if the color developed in the beaker in a boiling water bath for 30 minutes.
arsenic test exceeded more than of the step Remove the beaker from the water bath and
between 0 and 10 g/L arsenic on the color transfer the contents to the reaction vessel. Allow
chart, or approximately 5 g/L arsenic. the sample to cool to room temperature.
Complete the procedure as for inorganic arsenic.
Sodium sulfide standards were prepared and
their concentrations were verified using method Results for Various Organic Arsenics
# 8131 (Methylene Blue), the Hach DR-4000 Compound 500 g/L Std 10 g/L Std
Spectrophotometer, and the instrument-stored Cacodylic acid 500 10
program for sulfide. When using the kit (Dimethyl
reagents, it was found that as much as 5 mg/L S2- arsenic acid)
could be tolerated before the acceptance criterion C2 H8 AsO2 Na
of 5 g/L arsenic equivalence was exceeded. o-Arsanilic acid 500 10
H2 NC6 H4 AsO3 H2
Interferences p-Arsanilic acid 500 10
H2 NC6 H4 AsO3 H2
Various substances likely to interfere with the Arsenazo III 500 10
method were evaluated using the same C22 H18 N4 O14 S2 A
acceptance criterion (5g/L As). s2
The following were found to interfere at The results on all organic arsenics tested showed
concentrations greater than those listed. 100% recovery at both 10 and 500 g/L As.
Ion or Substance Concentration
2-
Sulfide (S ) 5 mg/L Testing Samples of Bottled Water
Selenium (Se) 1 mg/L
Antimony (Sb) 0.25 mg/L Samples of bottled water from a variety of sources
Tellurium (Te) Likely to interfere, but were tested, and then spiked and retested. The
not tested samples showed an initial arsenic level that
closely matched levels by the Natural Resource
The following did not interfere at the levels Defense Council (NRDC) reported for these
tested. products.
Ion or Substance Level Tested Bottled Water Results
Hardness 1000 mg/L CaCO3 Source Initial With 10 NRDC
Alkalinity 1000 mg/L CaCO3 Detected g/L Reported
Iron (Fe) 10 mg/L (g/L) Spike Value
Temperature 10 to 40 degrees C Michigan 0 10 Not Done
France 0 10 0