Documente Academic
Documente Profesional
Documente Cultură
available, the most important of which are shortly exclusively refer to this electrode) until the use of
Reference electrodes and their usage discussed below: mercury was banned from more and more
laboratories. The electrode potential is +0.241 V
Silver/Silver Chloride (Ag/AgCl in saturated KCl) vs. SHE at 25°C. Compared to the Ag/AgCl
Introduction This is probably the most widely used reference electrode, this electrode has the disadvantage that
electrode, since the use of mercury became less it cannot be used above 50°C due to instability of
A reference electrode has a stable and well popular. This electrode consists of a Ag wire in the Hg2Cl2.
defined electrochemical potential (at constant contact with AgCl in a saturated KCl solution. This
temperature) against which the applied or results in an electrode potential of +0.197 Volt vs Mercury/mercury sulphate (Hg/Hg2SO4 in 0.5 M
measured potentials in an electrochemical cell are SHE at 25° C. H2SO4)
referred. A good reference electrode is therefore This reference electrode is used in some cases
non-polarizable, in other words, the potential of where the use of chloride ions is not desirable. The
such an electrode will remain stable upon passage electrode potential of this system is +0.680 Volt vs
of a small current. One could also say that the SHE
impedance of an ideal reference electrode is zero.
In practice the non-polarizability only holds at small Mercury/mercury oxide (Hg/HgO in 1 M NaOH)
currents, and therefore the input impedance of the Used in alkaline solutions only. The electrode
reference electrode input on a potentiostat should potential of this electrode is +0.140 Volt vs. SHE
be as high as possible.
AUTOLAB ELECTROCHEMICAL INSTRUMENTS ARE DEVELOPED AND PRODUCED BY ECO CHEMIE B.V. IN THE NETHERLANDS Appl038-2
www.ecochemie.nl.
AUTOLAB APPLICATION NOTE
reference when measuring in organic solvents. Consequences for AC measurements Stability issues on capacitive cells
The ferrocene-ferricinium redox couple for
example can be used for this purpose. The redox In the Autolab instruments the reference electrode Potentiostats use a so-called negative feedback
potential measured for this couple can then be input typically has a capacitance of 8 pF. The system to keep the desired potential between WE
used as an extra reference point. impedance of the reference electrode in and RE stable. As soon as the there is a change in
combination with the input capacitance, results in a the desired potential in the positive direction, the
Influence of the reference electrode on RC low pass filter. A typical value for a Metrohm feedback system ensures a change in the negative
Ag/AgCl reference electrode is 1 kOhm, but when direction, hence the term negative feedback.
potentiostat stability using the double junction electrode with LiCl in Positive feedback however may arise as soon as
Ethanol as the outer electrolyte, the impedance of the phase shift becomes 180°, for example due to
Although the interest of the user usually lies in the the reference electrode may be as high as 100 a phase shift contribution of the feedback system
behaviour of the working electrode, the reference kOhm. The latter case in combination with a 8 pF itself. If positive feedback occurs, the potentiostat
electrode can have considerable influence on the input capacitance results in an RC filter with a 800 will go into oscillation. Under normal
stability of a potentiostat. ns time constant. This filter will result in a phase circumstances this will not happen, but the
As stated above an ideal reference electrode has shift of -27º at 100 kHz. (see figure below) As the presence of a reference electrode with high
zero impedance. In practice however, due to the effect is especially noticeable at higher impedance will result in an additional phase shift
use of ceramic (or Vycor) frits or salt bridges, the frequencies, for impedance measurements it is (added to the one of the feedback system), which
impedance maybe considerably higher (up to 20 recommended to use a platinum wire coupled to can in turn lead to instability.
kOhm or more). In order to avoid stability problems the reference electrode through a capacitor The Autolab potentiostats have been tested on
it is thus advisable to take good care of the (typically 0.1 µF- 1 µF). The high frequency signal capacitances ranging from 10 pF to 3000 F, both
reference electrode: will bypass the reference electrode, whereas the on the High Stability and High Speed settings, this
1. Make sure that the reference electrode dc component will still pass through the reference poses no stability problems. However, as soon as
compartments are filled with electrolyte electrode. the reference electrode impedance reaches values
solution (i.e. KCl in case of a Ag/AgCl higher than 30 kOhm the system becomes
electrode) unstable in the High Speed mode.
2. Avoid Luggin capillaries that are very It is therefore recommended to use low impedance
narrow at the end reference electrodes in combination with the High
3. Avoid the use of extra (high impedance) Stability setting when possible.
frits between the cell and the reference
electrode when possible
4. Only use the High Speed setting of the
Autolab when needed (i.e. when signals
with frequencies higher than 12.5 kHz are
involved).
AUTOLAB ELECTROCHEMICAL INSTRUMENTS ARE DEVELOPED AND PRODUCED BY ECO CHEMIE B.V. IN THE NETHERLANDS Appl038-2
www.ecochemie.nl.