Supplies & Consumables

In CSSD
Dr. Ashraf Abd Elkader El-Kelany
General Directorate, infection Prevention & Control
MOH - KSA
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
• Item Number: 124990034

Jan-2014 2
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Item Specifications:
• Test pack designed for daily monitoring of the
performance of pre-vacuum steam sterilizers
operating at 134°C / 3.5 ± 5s minutes and 121°C /
15 (± 5s) minutes.
• In addition to detecting air leaks, inadequate steam
penetration and vacuum pump failures, the test
should detect wet steam, superheated steam and
non-condensable gases problems.
Jan-2014 3
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Item Specifications:
• Test uses thermo-chromic ink formulation, free
from lead or any toxic materials.
• The indicator reagent should be uniformly
distributed on the test sheet to cover not less than
30% of the surface area of the sheet, with distance
between adjacent areas of indicator reagent should
not exceed 20 mm and sufficient strength to
withstand steam sterilization.
Jan-2014 4
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Item Specifications:
• Difference in reflectance density should not be less
than 0,3 between the substrate and either the
changed indicator or the unchanged indicator as
specified by the manufacturer.
• The back of the test sheet should include inquiries
about department, machine No., cycle No.,
operating temperature, operator, date, result, and
supervisor for quality assurance records.
Jan-2014 5
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Item Specifications:
• The outside of the test pack should show lot
number, expiration date and class 1 chemical
indicator spot.
• Conforms to EN ISO 11140-3,5 and EN285.
• Material safety data sheet (MSDS) should be
submitted.

Jan-2014 6
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Instructions For Use:
Always follow the manufacturer's instructions for use,
storage, expiration date and interpretation of results.
• Frequency of use: Bowie-Dick test pack should be
done daily for each pre-vacuum steam sterilizer.
• Timing: if the steam is turned off at night, the test
should be performed the first cycle of the day. If the
sterilizers stay on 24-hours a day, the test should be
performed the same time every day.
Jan-2014 7
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Instructions For Use:
• Placement: The recommended location is on the
bottom shelf over the drain line (coldest spot).
• The test should be run for 3 ½ to 4 minutes with no
dry time [Bowie-Dick (B&D) cycle].
• The Bowie-Dick test is done in an EMPTY load
(nothing else in the sterilizer) and no other tests
could be performed along with it.
Jan-2014 8
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Instructions For Use:
• At the end of the cycle, the door should be opened
and the cart with the pack removed (The cart should
not remain inside the sterilizer to cool down).
• After cooling, the pack should be opened and the
results reviewed:
1) A complete and even color change of the indicator
sheet = (PASS).
2) Any incomplete or inconclusive results = (FAIL).
Jan-2014 9
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Unused Bowie Dick Chemical Indicator sheet contained
within the air removal test pack.

Jan-2014 10
Dynamic Air Removal - DAR
(Bowie-Dick) test pack

(PASS) result

Jan-2014 11
Dynamic Air Removal - DAR
(Bowie-Dick) test pack

Typical air leak failure patterns

Jan-2014 12
Dynamic Air Removal - DAR
(Bowie-Dick) test pack

Typical non condensable gas failure patterns

Jan-2014 13
Dynamic Air Removal - DAR
(Bowie-Dick) test pack

Typical wet steam failure patterns

Jan-2014 14
Dynamic Air Removal - DAR
(Bowie-Dick) test pack

Typical superheated steam failure patterns

Jan-2014 15
Dynamic Air Removal - DAR
(Bowie-Dick) test pack
Instructions For Use:
• The failure result should be immediately reported to
the CSSD supervisor and the test repeated. If the
second test also fails, do not use the sterilizer and
call maintenance services.

Jan-2014 16
Transportation Gel Spray

• Item Number: 124990040

Jan-2014 17
Transportation Gel Spray

Item Specifications:
• Gel formula that clings to soiled instruments to
maintain the moisture during transport up to 72
hours.
• Dye-free formula to prevent stain of instruments or
equipment (soiled sharps are visible through the
gel).
• Ready-to-use bottle with spray applicator that
controls the area and intensity of spray.
Jan-2014 18
Transportation Gel Spray

Item Specifications:
• Contain surfactants and / or enzymes that initiate
the cleaning process of soiled instruments at the
point of use.
• Neutral PH, contains corrosion inhibitors with
broad compatibilities with materials used in the
manufacturing of reusable surgical instruments
(safe for aluminium, soft metals and gold handles).

Jan-2014 19
Transportation Gel Spray

Item Specifications:
• Rinses easily and completely, even when dried (no
need for special rinsing before manual or automated
processing - compatible with all types of automated
cleaning equipment).
• Size: 600ml - 1000ml
• Conforms to relevant ISO standards
• Material safety data sheet (MSDS) should be
Jan-2014
submitted. 20
Transportation Gel Spray

Instructions For Use:


• Simple one step procedure: spray used instruments
immediately in the operating room (after removal
of gross soil, emptying
fluids from reusable
containers, & removal
disposable items).

Jan-2014 21
Heavy duty gloves
(Large – Medium – Small)
Items Numbers:
• 124990043 - medium
• 124990068 - small
• 124990069 - large

Jan-2014 22
Heavy duty gloves
(Large – Medium – Small)
Item Specifications:
• Latex and Powder free
• Thick (13 Mil thickness)
• Long cuffs (more than 290 mm from the index
finger),
• Dark blue or yellow colour,
• Disposable, Non-sterile,
• Size: small – medium – large.
Jan-2014 23
Rapid-readout Biological Indicator

• Item Number: 124990044

Jan-2014 24
Rapid-readout Biological Indicator

•A Biological Indicator challenge test pack


containing Biological Indicator & a Class 5 or
Class 6 chemical Indicators.

Jan-2014 25
Rapid-readout Biological Indicator

Item Specifications:
• Rapid readout biological indicator (1292E):
with 3 hours initial fluorescent result, to monitor
121°C & 134°C vacuum assisted steam sterilization
cycles.
• Self contained ampoule type and compatible with
biological incubators provided by MOH.

Jan-2014 26
Rapid-readout Biological Indicator

Item Specifications:
• Comply with (ISO 11138 - Parts 1 & 3 /
ANSI/AAMI/ISO 11138 - Parts 1 & 3 / ISO 14161
- ANSI/AAMI/ISO 14161).
• Material safety data sheet (MSDS) should be
submitted.

Jan-2014 27
Rapid-readout Biological Indicator

Instructions For Use:


Always follow the BIs manufacturer's instructions
for storage, expiration date, how to use, cycles that
can be monitored, how to follow-up after sterilization
(incubation) and how to interpret results.
• Frequency of use: BIs must be used to monitor each
autoclave machine:
1) At least weekly; preferably daily.
2) For release of each load containing implantable
Jan-2014 devices. 28
Rapid-readout Biological Indicator

• Frequency of use (con.):


3) Ideally, for routine release of loads containing
non-implantable items.
4) After steam sterilizer installation, relocation,
malfunction, major repairs, and after sterilization
process failures. Each sterilization cycle type used
should be tested
5) When running a gravity displacement cycle since
this cycle is not routinely tested.
Jan-2014 29
Rapid-readout Biological Indicator

• Timing: In the first working load of the day after


Dynamic Air Removal test (Bowie-Dick test pack).
• Placement: Biological Indicator challenge test pack
must be placed in a fully loaded chamber, in the
area of the sterilizer chamber and load that is least
favorable for sterilization (usually in the front,
bottom section near the drain line). The sterilizer
manufacturer should identify the exact location of
this area.
Jan-2014 30
Rapid-readout Biological Indicator

• Incubation: unless there are special instructions for


activation and incubation by the manufacturer,
"rapid readout" Biological Indicator test providing a
initial result for steam sterilization cycles in three
hours (wrapped loads) and one hour for flash cycle
(unwrapped loads).
• Control: Control Biological Indicator vial is needed
to verify the pre-sterilization viability of the spores.
Note: The test BI vial must be of the same lot number as the
Jan-2014 control BI vial otherwise the test is considered not valid. 31
STERRAD® Systems Sterilant
Cassettes
Items Numbers:
• 124990045 Sterrad 100 S (Cassettes / 5 cycles)
• 124990046 Sterrad NX (Cassettes / 5 cycles)
• 124990060 Sterrad 200 S (Cassettes / 2 cycles)
• 124990081 Sterrad 100 NX (Cassettes / 5 cycles)

Jan-2014 33
STERRAD® Systems Sterilant
Cassettes

STERRAD® STERRAD® STERRAD®


200 STERRAD® NX
100S 100NX
Jan-2014 34
STERRAD® Systems Sterilant
Cassettes
Instructions For Use:
• Always follow manufacturer's instructions of
STERRAD® systems sterilant cassettes about
storage, expiration date, how to use, and how to
dispose expired and used cassettes.

Jan-2014 35
Adhesive Tape
for H2O2 Sterilization Systems with chemical indicator

• Item Number: 124990047

Jan-2014 36
Adhesive Tape
for H2O2 Sterilization Systems with chemical indicator

Item Specifications:
• Non-cellulosic adhesive sealing tape with exposure
monitor for use in Hydrogen Peroxide Sterilization
Systems.
• Lead free, colour stable indicator ink that Provides
distinct colour changes when exposed to vaporized
hydrogen peroxide.
• 50 to 100 meters long × 1.5 – 2 cm wide.
Jan-2014 37
Adhesive Tape
for H2O2 Sterilization Systems with chemical indicator

Item Specifications:
• Meets ISO 11140-1 Class 1 requirements for
hydrogen peroxide vapour process indicators.
• Material safety data sheet (MSDS) should be
submitted.

Jan-2014 38
Chemical Indicator
for H2O2 Sterilization Systems
• Item Number: 124990048

Jan-2014 39
Chemical Indicator
for H2O2 Sterilization Systems
Item Specifications:
• An internal pack chemical indicator for use in H2O2
Sterilization Systems.
• Lead free, colour stable indicator ink that provides
distinct colour changes when exposed to vaporized
hydrogen peroxide. End point colour standard on
each strip to facilitate colour comparison.
• Size: 1.5 – 2 cm × 7.5 - 10 cm.
Jan-2014 40
Chemical Indicator
for H2O2 Sterilization Systems
Item Specifications:
• The product provides assurance that items inside a
pack have been exposed to sterilant, even with
relatively low hydrogen peroxide concentrations
(it does not indicate medical device sterilization).
• Meets ISO 11140-1 Class 1 requirements for
hydrogen peroxide vapour process indicators.
• Material safety data sheet (MSDS) should be
submitted.
Jan-2014 41
Chemical Indicator
for H2O2 Sterilization Systems
Instructions For Use:
Always follow Chemical Indicator manufacturer's
instructions for storage, expiration date, how to
use, and how to interpret results.
• Frequency of use: Chemical Indicators for low
temperature gas plasma must be routinely used
inside each package, or containment device.

Jan-2014 42
Biological Indicator
for H2O2 Sterilization Systems
• Item Number: 124990049

Jan-2014 43
Biological Indicator
for H2O2 Sterilization Systems
Item Specifications:
• Self-Contained BI (Polypropylene vial type)
contains Geobacillus stearothermophilus bacterial
spores with specific population range strip and an
ampoule containing growth media.

Jan-2014 44
Biological Indicator
for H2O2 Sterilization Systems
Item Specifications:
• Comply with ISO 11138 General Requirements.
Media viable count complies with ATCC 7953 with
min. value of 1 X 10 power 6 spore population.
• The bidder must submit certificates, which confirm
its conformity with the required specifications.

Jan-2014 45
Biological Indicator
for H2O2 Sterilization Systems
Instructions For Use:
Always follow the BI manufactures instructions for
storage, expiration date, how to use, how to follow
up after sterilization (incubation), how to interpret
results and how to dispose expired and used BIs.
• Frequency of use: Biological Indicators for Low
temperature gas plasma must be used to monitor gas
plasma sterilizer daily.

Jan-2014
• Timing: In the first load of the day. 46
Biological Indicator
for H2O2 Sterilization Systems
Instructions For Use:
• Placement: Biological Indicator test pack must be
placed on the bottom shelf, back of gas plasma
sterilizer (spot of greatest challenge).
• Incubation: Unless there are special instructions for
incubation by the manufacturer, (for 24 hrs. At 55 –
60 °C) / Colour change in media indicates failure of
the BI test.
• Control: Control Biological Indicator vial is needed
Jan-2014 to verify the pre-sterilization viability of the spores. 47
Chemical Emulating Indicators
for Steam Sterilization

• Item Number: 124990050

Jan-2014 48
Chemical Emulating Indicators
for Steam Sterilization

•A Chemical Indicator challenge test pack


(containing a Class 6 emulating chemical Indicator
or a Class 5 integrating chemical Indicator).

Jan-2014 49
Chemical Emulating Indicators
for Steam Sterilization

Item Specifications:
• Steam sterilization cycle verification strip Class 6
chemical indicator that confirm exposure to certain
critical variables (temperature, exposure time and
saturated steam) for a specified sterilization cycle(s).
• Internal pack Indicator that can be used routinely in
the commonly used pre-vacuum cycles (Dual
Temperatures Indicator: 121 °C for 12 minutes - 134
°C for 4 minutes).
Jan-2014 50
Chemical Emulating Indicators
for Steam Sterilization

Item Specifications:
• Conform to EN 867-1 Class D and ISO 11140-1
(Class 6) standards.
• The indicator ink is non-toxic and free of lead, and
provides permanent colour change that indicates
specified sterilization parameters have been met,
with endpoint colour standard on each strip to
facilitate colour comparison.
• Material safety data sheet (MSDS) should be
Jan-2014 submitted. 51
Chemical Emulating Indicators
for Steam Sterilization

Instructions For Use:


Always follow Class 6 Chemical Indicator
manufacturer's instructions for storage, expiration
date, how to use, cycles it can be used to monitor
(e.g., Dual Temperature 121°c / 12 minutes - 134°c /
4 minutes) and how to interpret results.
• Frequency of use: Chemical Indicators must be
routinely used within each package, tray, or
containment device required to be steam sterilized.
Jan-2014 52
Chemical Emulating Indicators
for Steam Sterilization

Instructions For Use:


• Placement: Chemical Indicators must be placed in
the area of the package, tray, or containment device
considered to be least accessible to steam
penetration (spot of greatest challenge). For
containment device, the manufacture's written
instructions for placement of the Chemical Indicator
should be consulted.
Note: Chemical Indicator is not a substitute for
Jan-2014 Biological Indicator. 53
General Instrument Cleaning
Stainless Steel Brush
• Item Number: 124990053

Jan-2014 54
General Instrument Cleaning
Stainless Steel Brush
Item Specifications:
• Toothbrush-style, high quality brush.
• Disposable.
• Non-sterile.
• Stainless Steel bristles.

Jan-2014 55
General Instrument Cleaning
Stainless Steel Brush
Instructions For Use:
• This brush is ideal for cleaning rasps, files and
burrs.

Jan-2014 56
Washer Disinfector Soil Test

• Item Number: 124990054

Jan-2014 57
Washer Disinfector Soil Test

Item Specifications:
• Made of plastic or stainless steel substrate with
pattern of dried soil (pork free formula) to challenge
the cleaning efficacy of mechanical cleaning
equipment and detergents.
• Designed to use with every washer cycle, to detect
chemical and mechanical activity problems.
• Contains two sources of blood proteins in
proportions similar to human blood proteins, lipids
Jan-2014 and polysaccharides. 58
Washer Disinfector Soil Test

Item Specifications:
• Performance is equivalent to cleaning efficacy soil
tests for surgical instruments, anaesthetic
equipment, bowls, utensils, glassware, etc. as stated
in EN ISO 15883 and HTM 2030.
• Supplied with plastic or stainless steel holder which
provides a physical barrier similar to areas of a
surgical instrument not exposed to direct spray
action, and used to fix it in the washer tray.
Jan-2014 59
Washer Disinfector Soil Test

Item Specifications:
• Supplied with trouble shooting guide for problems
associated with the cleaning cycle.
• Material safety data sheet (MSDS) should be
submitted.

Jan-2014 60
Washer Disinfector Soil Test

Instructions For Use:


Always follow the manufacturer's instructions for
storage, expiration date, how to use, and how to
interpret results.
• Frequency of use: must be used to monitor each
washer disinfector machine.
1) At least weekly (preferably daily) for routine use.
2) After washer disinfector machine installation,
Jan-2014
malfunction, or major repairs. 61
Washer Disinfector Soil Test

Instructions For Use:


• Placement: must be placed in a fully loaded cycle,
in the least favorable areas for cleaning. The exact
location of these areas is usually identified by
machine manufacturer, usually indicators are used
on each rack level and the middle rack is considered
the most challenging to clean.

Jan-2014 62
Enzymatic Detergent
For Instruments Cleaning
• Item Number: 124990055

Jan-2014 63
Enzymatic Detergent
For Instruments Cleaning
Item Specifications:
• Concentrated formula low dose formula (1-5 ml/L),
effective on blood, fat, mucous and other soils
(specifically designed for use on hard-to-clean soils).
• Broad range of water quality and type (dissolves
completely in hard or soft water) & easy to rinse.
• Low foaming.
• Suitable for pre-soak and manual cleaning
Jan-2014
applications after adequate dilution. 64
Enzymatic Detergent
For Instruments Cleaning
Item Specifications:
• Broad-spectrum compatibilities with materials used
in the manufacture of reusable surgical instruments
& medical devices.
• Enhanced corrosion inhibitor properties to protect
instruments from the harmful effects of water and
water impurities.
• Biodegradable, EPA Registered.
Jan-2014 65
Enzymatic Detergent
For Instruments Cleaning
Item Specifications:
• Must be compatible with all manufactures’
ultrasonic washers, washer/disinfectors and other
automatic mechanical washer applications used by
MOH (The bidder must provide letter to confirm
compatibility).
• Conforms to relevant ISO standards.
• Material safety data sheet (MSDS) should be
submitted.
Jan-2014 66
Enzymatic Detergent
For Instruments Cleaning
Instructions For Use:
Always follow the manufacturer's instructions for
storage, expiration date, how to mix or prepare,
how to use, recommended water temperature,
safety precautions, and how to dispose used
solutions.

Jan-2014 67
Alkaline Detergent
For Instruments Cleaning
• Item Number: 124990056

Jan-2014 68
Alkaline Detergent
For Instruments Cleaning
Item Specifications:
• Concentrated low dose formula (1-5 ml/L), with
advanced surfactant activity to remove protein,
blood, fat, grease and oil.
• Broad range of water quality and type (dissolves
completely in hard or soft water) & easy to rinse
• Low foaming.
• Suitable for manual cleaning applications after
Jan-2014
adequate dilution. 69
Alkaline Detergent
For Instruments Cleaning
Item Specifications:
• Broad-spectrum compatibilities with materials used
in the manufacture of reusable surgical instruments
& medical devices.
• Enhanced corrosion inhibitor properties to protect
instruments from the harmful effects of water and
water impurities.
• Biodegradable, EPA Registered.
Jan-2014 70
Alkaline Detergent
For Instruments Cleaning
Item Specifications:
• Must be compatible with all manufactures’
ultrasonic washers, washer/disinfectors and other
automatic mechanical washer applications used by
MOH (The bidder must provide letter to confirm
compatibility).
• Conforms to relevant ISO standards.
• Material safety data sheet (MSDS) should be
submitted.
Jan-2014 71
Alkaline Detergent
For Instruments Cleaning
Instructions For Use:
Always follow the manufacturer's instructions for
storage, expiration date, how to mix or prepare,
how to use, recommended water temperature,
safety precautions, and how to dispose used
solutions.

Jan-2014 72
High Level Disinfectant (OPA)

• Item Number: 124990058

Jan-2014 73
High Level Disinfectant (OPA)

Item Specifications:
• Orthophthaldhyde solutions 0.55% for HLD of
endoscopes.
• Compatible with all flexible endoscopes.
• Supplied with required efficacy testing strips.
• CE and/or EPA Registered.
• Conforms to relevant ISO standards.
• Material safety data sheet (MSDS) should be
submitted.
Jan-2014 74
High Level Disinfectant (OPA)

Instructions For Use:


Always follow the manufacturer's instructions for
storage, expiration date, how to use the testing
strips, how to use the solution, safety precautions,
and how to dispose used solution.

Jan-2014 75
Surface Cleaning & Disinfecting Spray

• Item Number: 124990059

Jan-2014 76
Surface Cleaning & Disinfecting Spray

Item Specifications:
• Ready to use quaternary ammonium compounds
based detergent disinfectant (to clean & disinfect
hard environmental surfaces).
• Bactericidal, virucidal including (hepatitis B,
hepatitis C and HIV) and fungicidal.
• Compatible with wide range of materials including
stainless steel, aluminum, chrome, plastics, vinyl,
acrylic, and glass.
Jan-2014 77
Surface Cleaning & Disinfecting Spray

Item Specifications:
• Phenols and chlorine free
• Leaves no residue (rinsing is not necessary).
• Conforms to OSHA Guidelines, EPA registered, or
meets European certification standards (CE
marked)
• Bottle 500-1000 ml
• Material safety data sheet (MSDS) should be
submitted.
Jan-2014 78
Surface Cleaning & Disinfecting Spray

Instructions For Use:


• Apply the foam disinfectant detergent directly to
the area to be treated or preferably on a non-woven
cloth.
• Spread the product out well.
• Leave to take effect. Rinsing is unnecessary unless
the medical device is likely to come into contact
with patient mucosa.
Jan-2014
• Contact time: minimum 5 minutes. 79
Sterilization Roll
for H2O2 Sterilization Systems
• Item Number:
124990061
124990062
124990063

Jan-2014 80
Sterilization Roll
for H2O2 Sterilization Systems
Item Specifications:
• H2O2 Sterilization Systems wrapping flat roll, One
Tyvek material face (porous material) and One
laminated polyester and polypropylene face
(transparent film).
• Sizes: 150 mm x 70 m - 350 mm x 70 m - 500 mm
x 70 m.
• Conform to ISO 11607-1 and EN 868- 1,3,5.
Jan-2014
• Shelf life should be not less than 2 years. 81
Sterilization Roll
for H2O2 Sterilization Systems
Item Specifications:
• Process H2O2 exposure indicator: should be printed
at distance not more than 15 cm, and each
individual indicator should not be less than 100
mm² area. Indicator should not be affected by the
sealing procedure.
• The roll should not be printed on any surface that
comes into direct contact with the items to be
packed.
Jan-2014 82
Sterilization Roll
for H2O2 Sterilization Systems
Item Specifications:
The following information should be printed on the roll:
1) The words “Do not use if the sterile barrier is
damaged” or other equivalent phrase,
2) Trade name or Manufacture’s name,
3) Lot number,
4) Nominal dimensions and/or identification code,
5) Peel direction (direction which ensures least fiber
tear).
Jan-2014 83
Sterilization Package Labels Gun
& Labels Roll
• Item Number:
124990064
124990065

Jan-2014 84
Sterilization Package Labels Gun

Item Specifications:
• Printing applicator that applies the labels to
packages and record sheets, charts or envelopes for
identification and recall purposes.
• Lightweight, with 2 lines printing system.
• Durable, and easy to use.
• Reliable, with high quality printout on the labels.
• Charging capacity: sterilization load labels roll
(1000 - 1200 labels / roll), with label size not less
Jan-2014 than 15 mm x 28 mm. 85
Sterilization Package Labels Roll

Item Specifications:
• Strong label adhesive keeps label withstand steam
sterilization, eliminating the need for relabeling.
• The Label provide 2 lines of information with up to
10 characters per line
(one line for load and sterilizer numbers and the second
line for sterilization Gregorian dating).
• Labels / roll: 1000 - 1200, with label size not less
that: 15 mm x 28 mm.
Jan-2014 86
Sterilization Package Labels Roll

Item Specifications:
• Labels comes printed with red area "sterilized" in-
between the 2 line and the statement "sterile unless
the package is opened or damaged“
• Compatible with label gun (printing applicator -
item code 124990064).
• Conforms to relevant international standards.

Jan-2014 87
Cannullated Instrument Cleaning
Brush
• Item Number: 124990066

Jan-2014 88
Cannullated Instrument Cleaning
Brush
Item Specifications:
• Cannullated instruments cleaning high quality
brush.
• Stainless steel handle & white medical grade nylon
bristles.
• Whole brush length not less than 35 cm / Bristle
end: 20 - 25 mm in diameter.
• Non-sterile, Disposable.
Jan-2014 89
Large Instruments Cleaning Brush

• Item Number: 124990067

Jan-2014 90
Large Instruments Cleaning Brush

Item Specifications:
• Large instruments cleaning high quality brush.
• Brush head: 16 – 20 cm length x 25 – 40 mm width,
with white medical grade nylon bristles (20 -25 mm
length).
• Plastic handle 15 – 20 cm

Jan-2014 91
General Instrument Cleaning
Nylon Brush
• Item Number: 124990070

Jan-2014 92
General Instrument Cleaning
Nylon Brush
Item Specifications:
• Toothbrush-style, high quality brush.
• Disposable.
• Non-sterile.
• White medical grade soft nylon bristles.

Jan-2014 93
Ultrasonic Cleaner Indicator

• Item Number: 124990072

Jan-2014 94
Ultrasonic Cleaner Indicator

Item Specifications:
• Self contained design (e.g. vial) that does not
contaminate the bath.
• Can be visually checked for colour change (+ve
result), which indicates sufficient Ultrasonic
energy).

Jan-2014 95
Ultrasonic Cleaner Indicator

Instructions For Use:


• Ultrasonic cleaner indicator is placed in the
ultrasonic basket between the items of the load.

Jan-2014 96
High level disinfectant
(Peracetic Acid or Hydrogen Peroxide)
• Item Number: 124990073

Jan-2014 97
High level disinfectant
(Peracetic Acid or Hydrogen Peroxide)
Item Specifications:
• Active ingredient Peracetic acid or Hydrogen
Peroxide (or both) with concentrations approved for
high level disinfection of endoscopes and thermo-
sensitive equipment and medical devices.
• Safe, easy-to-use, and environmentally friendly.
• It can be re-used for up to 14 days, and then
disposed safely into sanitary sewers.
Jan-2014
• Compatible with all flexible scopes manufactures. 98
High level disinfectant
(Peracetic Acid or Hydrogen Peroxide)
Item Specifications:
• Supplied with required efficacy testing strips.
• Must be compatible with Automated Endoscope
Processors provided by MOH.
• CE and/or EPA Registered
• Conforms to relevant ISO standards
• Material safety data sheet (MSDS) should be
submitted.
Jan-2014 99
High level disinfectant
(Peracetic Acid or Hydrogen Peroxide)
Instructions For Use:
Always follow the manufacturer's instructions for
storage, expiration date, how to use the testing
strips, how to use the solution, safety precautions,
and how to dispose used solution.
• Pour the activated solution (pour the activator in the
5L can), and cover the tank.
• Contact time: 10 or 30 min according to the
requested activity.
Jan-2014 100
High level disinfectant
(Peracetic Acid or Hydrogen Peroxide)
Instructions For Use:
• For endoscopic equipment: make solution circulate
inside channels.
• Rinse thoroughly the medical device with sterile or
filtered water (0.2 μm)
• For endoscopic equipment: rise the internal and
external parts of the medical device.
• Dry with a single-use towel.
• Keep the medical device as aseptically as possible
Jan-2014 up to use it again. 101
Scope Valves Basket

• Item Number: 124990075

Jan-2014 102
Scope Valves Basket

Item Specifications:
• Light weight fenestrated single use basket with
click lock catch to keep the scope valves inside
secured.
• Simple twist loop to attach the basket to the scope
during automated reprocessing.

Jan-2014 103
Cannullated Instrument Cleaning
Brush
• Item Number: 124990076

Jan-2014 104
Cannullated Instrument Cleaning
Brush
Item Specifications:
• Tube style high quality brush
• Soft medical grade nylon bristles embedded in
twisted stainless steel wire with suitable handle for
good manipulation.
• Whole brush length is 45 – 50 cm / Bristle end: 4 -
5 mm in diameter.

Jan-2014 105
Stain, Spot and Rust Remover

• Item Number: 124990077

Jan-2014 106
Stain, Spot and Rust Remover

Item Specifications:
• Phosphate-free formula that Restores and maintains
the protective passive layer integrity of stainless-
steel surfaces, resulting in a surface that is more
resistant to corrosion and pitting
• Effective in removing stains, rust and hard water
scale deposits from a stainless steel surfaces to
protect surgical instrument and extends its life.

Jan-2014
• Easy to rinse leaving residue-free surface. 107
Stain, Spot and Rust Remover

Instructions For Use:


• Wear suitable protective equipment before using the
product: gloves, glasses and mask.
• Soak the instruments in a 5% dilution in warm water.
• Soaking time: 1 to 2 hours (every 30 minutes check
the surface state of the instruments).
• When the instruments are visibly renewed, remove
them from the solution.
• Rinse carefully and dry.
Jan-2014 • Renew the soaking bath once a day at least. 108
Initial Cleaner
For Channelled Scopes
• Item Number: 124990078

Uno-Flush

Jan-2014 109
Initial Cleaner
For Channelled Scopes
Item Specifications:
• Ready to use solution (400 - 500 ml) with scope
shaped sponge (solution & sponge in one
container),
• Designed for initial flush (immediately post
procedure) of Channelled scopes.
• Non Enzymatic formulation based on quaternary
ammonium compounds.
• Contains surfactants, polymers, anticorrosive, and
Jan-2014 antifoaming agents. 110
Initial Cleaner
For Channelled Scopes
Item Specifications:
• Does not contain aldehydes or chlorine generating
components.
• Nontoxic and safe to handle, flush away and 100%
bio-degradable formulation.
• Bactericidal including mycobacteria, virucidal
(hepatitis B, hepatitis C and HIV) and fungicidal.
• CE mark and/or FDA approved & EPA Registered.
• Material safety data sheet (MSDS) should be
Jan-2014 submitted. 111
Channelled Scopes Cleaning Brush

• Item Number: 124990080

Jan-2014 112
Channelled Scopes Cleaning Brush

Item Specifications:
• Long channelled scopes high quality cleaning brush.
• White medical grade soft nylon bristle tip attached to
stainless steel spring coil wire for hard-to-turn
corners.
• Beaded tip at bristle end and loop handle at the other
end.
• Latex-free.
Jan-2014
• Disposable, Non-sterile. 113
Channelled Scopes Cleaning Brush

Item Specifications:
• Brush Diameter: Tapered 4 - 6 mm
• Bristle Area Length: 5 - 10 mm
• Overall Length: 230 - 250 cm

Jan-2014 114
Sponge For Initial Cleaning of Rigid
Scopes & Probes
• Item Number: 124990082

Uno-Wipe

Jan-2014 116
Sponge For Initial Cleaning of Rigid
Scopes & Probes
Item Specifications:
• Ready to use impregnated scope shaped sponge.
• Designed for initial wiping (immediately post
procedure) of NON-Channelled rigid scopes such as
ENT, Laryngoscopes or Ultrasound Probes.
• Impregnated in Non Enzymatic formulation based
on quaternary ammonium compounds.
• Contains surfactants, polymers, anticorrosive, and
Jan-2014
antifoaming agents. 117
Sponge For Initial Cleaning of Rigid
Scopes & Probes
Item Specifications:
• Does not contain aldehydes or chlorine generating
components.
• Nontoxic and safe to handle, flush away and 100%
bio-degradable formulation.
• Bactericidal including mycobacteria, virucidal
(hepatitis B, hepatitis C and HIV) and fungicidal.
• CE mark and/or FDA approved & EPA Registered
• Material safety data sheet (MSDS) should be
Jan-2014 submitted. 118
Crape Tray Liner

• Item Number: 124990083

Jan-2014 119
Crape Tray Liner

Item Specifications:
• Made of cellulose fibers 80g / m2 to be used under
the surgical tray to absorb any moisture after the
sterilization cycle.
• Size: 50 x 30 cm. Pack of 500 pieces.
• Compatible with ISO 6588, ISO 536, ISO 9197-1,
ISO 9198. and DIN 58953-6 1987 agar plate test.

Jan-2014 120