Documente Academic
Documente Profesional
Documente Cultură
CT Image Guidance at
Confirming Device Access of the Frontal Sinus?
Frank Robert Glatz III, MD
The Center for Sinus, Allergy, and Sleep Wellness • Email: bobg38514@hotmail.com
ABSTRACT INTRODUCTION
RESULTS
CONCLUSION
–1–
While it is important for a frontal sinus positioning XprESS Transillumination
technology to have a high rate of success, it is also important When attempting to access the frontal sinus with the
that a positioning technology can give the user an indication XprESS LoProfile with LED Light Fiber device, the light was
of confidence of the device position, which can be used activated and multiple attempts were performed as needed
to inform decision making such as where or whether to to try to position the tip of the device in the frontal sinus. If
surgically intervene with dissection or balloon dilation desired, the physician was also allowed to utilize the Light
on a particular frontal sinus. Seeker (Entellus Medical) tool and then follow-up with the
XprESS/LED Light Fiber device. With transillumination the
physician relied on observation of the light pattern on the
forehead, as well as tactile feel, endoscope cues, and baseline
METHODS
CT observations.
IGS Probe When the IGS probe was utilized as the second device in the
When attempting to access the frontal sinus with the IGS sinus, it was correctly placed in 14 out of 15 attempts (93.3%
probe, the physician made multiple attempts as needed to placement rate) for an overall placement rate of 90.6%.
try to position the tip of the device into the frontal sinus.
During placement of the IGS probe, the physician utilized
the feedback provided from the CT image, as well as tactile
feel and endoscopic cues.
–2–
Figure 3. IGS correctly placed. A) IGS image with probe tip in sinus. Confidence and Placement Prediction
B) Surgical view. C) Transnasal view. D) Intrasinus view from
superior aspect.
• IGS
A B
The treating physician indicated they were “very confident”
(confidence score of 4) in the positioning of the IGS probe
in the frontal sinus in 28 of 32 cannulation attempts, and
in 27 of the 28 times the doctor was very confident their
assessment of frontal sinus access was accurate (Table 1).
There was one instance in which the doctor was very
confident the probe was in the frontal sinus when in fact it
C D
was not. In the 4 instances in which the doctor was not very
confident (confidence score 0-3), the IGS probe was in the
frontal sinus 2 of the 4 times.
• Transillumination
The treating physician indicated they were very confident
Figure 4. Example of mispositioned IGS. A) Tip of probe is not within (rating of 4) in the positioning of the balloon device in the
ostium but behind cell wall (arrow). B) Corresponding IGS image. frontal sinus in 29 of 32 cannulation attempts, and in all 29
A B attempts with a very confident rating this was an accurate
assessment by the physician (Table 1). In the 3 instances
where the confidence was below a 4 rating level, the balloon
device accessed the frontal sinus in 1 of the 3 times.
Transillumination IGS
• Transillumination Confidence Not in Confidence Not in
Rating In Frontal Frontal Rating In Frontal Frontal
When the XprESS/LED Light Fiber was used as the first 4 14+15 = 29 0+0 = 0 4 13+14 = 27 0+1 = 1
device, it was correctly placed in 15 of 15 sinuses attempted 0-3 1+0 = 1 0+2 = 2 0-3 2+0 = 2 2+0 = 2
(100% success rate). See Figure 5 for an example of a
correctly placed XprESS balloon catheter.
Analysis of Sensitivity,
Confidence Rating Specificity,
In Frontal Positive Predictive Value,
Not in Frontal
When transillumination using the XprESS/LED Light Fiber and Negative
4 Predictive Value from a Physician
True Positive Confidence
False Positive
was utilized as the second device in the sinus, it was correctly Level of 40-3for Each of Transillumination
False Negative True Negative
and IGS
Positive predictive value = TP/(TP+FP)
placed in 15 of 17 sinuses (88.2% success rate), for an overall Negative predictive value = TN/(TN+FN)
Sensitivity = TP/(TP+FN)
placement rate of 93.8%. In 2 instances, the XprESS device When the placement confidence levels for each technol-
Specificity = TN/(TN+FP)
Transillumination IGS
was not positioned in the frontal sinus. See Figure 6 for an ogy are grouped between
Confidence
those with
Not in
“very high confidence”
Confidence Not in
example of a mispositioned XprESS device. Note that in (rating
Rating 4) and less confidence
In Frontal Frontal (rating
Rating0-3), the results of
In Frontal Frontal
4 14+15 = 29can0+0 Transillumination IGS0+1 = 1
this example, when the medial cell was retrospectively placement success be =analyzed
0 4 determine
to 13+14 = 27if placement
Positive
0-3 predictive
1+0 =value
1 0+2 = 229/29 = 100%0-3 2+0 =27/28
2 = 2+0
96%= 2
appreciated, the XprESS was subsequently placed correctly. confidence
Sensitivity can be a predictor
Transillumination ofIGSplacement
29/30 = 97% success = 93% 2).
27/29(Table
Specificity
Confidence Not in 2/2 =Confidence
100% 2/3 = 67%Not in
Table 2. Prediction
Rating
value definitions
In Frontal Frontal Rating In Frontal Frontal
Figure 5. XprESS LoProfile with LED Light Fiber correctly placed. Negative predictive value 2/3 = 67% 2/4 = 50%
4
Confidence 14+15
Rating= 29 0+0 =In0 Frontal 4 13+14Not
= 27 0+1 = 1
in Frontal
A) Light illuminating the sinus cavity. B) XprESS tip in sinus.
0-3 4 1+0 = 1 0+2 True
= 2 Positive 0-3 2+0False
= 2 Positive
2+0 = 2
A B 0-3 False Negative True Negative
Positive predictive value = TP/(TP+FP)
Negative predictive value = TN/(TN+FN)
Sensitivity = TP/(TP+FN)
Confidence Rating In Frontal Not in Frontal
Specificity = TN/(TN+FP)
4 True Positive False Positive
0-3 False Negative True Negative
Positive predictive value = TP/(TP+FP)
The results of using the confidence measure as a positive
Negative predictive value = TN/(TN+FN) Transillumination IGS
Figure 6. Example of mispositioned XprESS and then repositioned predictor
Sensitivity of placement
Positive=predictive
TP/(TP+FN)
Specificity = TN/(TN+FP)
value success
29/29are shown
= 100% in Table
27/283.
= 96%
correctly. A) Transillumination of medial frontal cell. B) Transnasal view. Sensitivity 29/30 = 97% 27/29 = 93%
C) XprESS tip seen in medial cell (arrow). D) CT scan showing medial Table 3. Prediction value results 2/2 = 100%
Specificity 2/3 = 67%
frontal cell. E) XprESS repositioned in frontal sinus. F) XprESS tip Negative predictive value 2/3 = 67% 2/4 = 50%
correctly positioned in frontal sinus. Transillumination IGS
D E F
Due to a small number of instances (seven) in which the
doctor was not “very confident” in device access of the frontal
sinus, and due to the low incidence of failing to cannulate the
frontal sinus (five), it isn’t appropriate to measure the
–3–
specificity or negative predictive value of the two
CONCLUSIONS
technologies. The most important function of IGS and
transillumination as used in this study was to indicate to
the treating physician if the device was correctly located in Using appropriate procedural techniques and a malleable-
the frontal sinus. There was only one instance in which a tipped lighted balloon device, cannulation of the frontal
technology provided information to the doctor that recess/frontal ostium can be achieved successfully and
incorrectly indicated the probe/device was in the frontal repeatedly. In a head-to-head comparison, light confirmation
sinus and this was with IGS. Overall, this study suggests that was as accurate as IGS confirmation when used to position
transillumination is as sensitive and predictive as IGS in the device tip in the frontal sinus.
confirming device location in the frontal sinus. The results
indicate that when the study device was not accurately With both technologies, the positive predictive value of
positioned within the frontal sinus, the treating physician device placement in the frontal sinus is very high.
was not very confident based on light confirmation/IGS.
Study Limitations
The study was set up to allow both treatment devices (IGS REFERENCES
Probe and XprESS with LED Light Fiber) to access each
frontal sinus. A limitation of this study is that the second 1
Sindwani R, Metson R. Image-guided frontal sinus surgery. Otolaryngol
Clin North Am. 2005 Jun; 38(3): 461-71.
access of a sinus may have been influenced by the accuracy 2
Citardi MJ, Batra PS. Intraoperative surgical navigation for endoscopic
of the initial access. sinus surgery: rationale and indications. Curr Opin Otolaryngol Head
Neck Surg. 2007 Feb; 15(1): 23-7.
3
Leventhal D, Heffelfinger R, Rosen M. Using image guidance tracking
during balloon catheter dilation of sinus ostia. Otolaryngol Head Neck
Surg. 2007 Aug; 137(2): 341-2.
4
Gould J, Alexander I, Tomkin E, Brodner D. In-office, multisinus balloon
dilation: 1-year outcomes from a prospective, multicenter, open label
trial. Am J Rhinol Allergy. 2014 Mar-Apr; 28 (2): 156-163.
5
Brodner D, Alexander I, Chandler S, Cutler J, Saigal K. Accuracy of
transnasal cannulation and dilation of the maxillary ostium in cadavers
with intact uncinates. Am J Rhinol Allergy. 2013; 27(1): 58-61.
6
Friedman M, Wilson M. Illumination guided balloon sinuplasty.
Laryngoscope. 2009 Jul; 119(7): 1399-402.
866-620-7615
www.EntellusMedical.com
–4–