Documente Academic
Documente Profesional
Documente Cultură
In a previous work, coupled-column L C (LC/LC) with direct large-volume sample injection for the analysis of
fl,-agonists in bovine urine was investigated, in view of its combination with thermospray (TSP) tandem mass
spectrometric (MS/MS) detection. In this work, the potential of both conventional L C and LC/LC coupled to
TSP-MS/MS for the rapid determination of P,-agonists in bovine urine was evaluated. I t was first established that,
in order to avoid disturbances in TSP-MS/MS detection due to changes in pressure and flow rate of the mobile
phase, LC/LC is preferable to conventional LC. Then, selecting clenbuterol and salbutamol as model compounds
for the class of P,-agonists, two single-residue LC/LC/TSP-MS/MS methods were developed, permitting the deter-
mination of the analytes in bovine urine samples with a limit of quantification of at least 0.1 ng, ml-', with a
sample throughput of 4 h-I. The methods were validated by analysing both spiked urine samples and samples
containing clenbuterol or salbutamol residues. Mean recoveries from urine samples spiked with clenbuterol a t levels
of 5 and 0.22 ng ml-' were 107% (n = 9, RSD = 7.4%) and 102% ( n = 6, RSD = 14%), respectively. A bovine
urine sample containing 1.28 ng ml-' clenbuterol residue (previously quantified by GC/MS) was re-analysed by
LC/LC/TSP-MS/MS on four different days (n = 3, each day) and yielded a mean concentration of 1.31 ng ml-',
with a reproducibility of 8.4%. Bovine urine samples spiked with salbutamol at levels of 10.8 and 0.55 ng ml-'
yielded mean recoveries of 101% (n = 3, RSD = 1%) and 89% (n = 3, RSD = 13%), respectively. In bovine urine
samples containing salbutamol residues, the concentrations of unconjugated salbutamol were found to be in the
range 8-57 ng ml-'.
common LC detectors on the one hand and by the diffi- potential of these hyphenated techniques in the rapid,
culties of interfacing LC with selective detection on-line determination o f B,-agonists in bovine urine
methods such as mass spectrometry (MS) on the other. was evaluated.
The development of effective LC/MS interfaces, such
as thermospray (TSP) and electrospray (ESI), has
opened up new possibilities in the application of
LC/MS to the analysis of j?,-agonists, and some EXPERIMENTAL
methods have been proposed for the determination of
residues of these drugs in bovine urine.32p34Blanchflo-
wer and Kennedy32 proposed a method based on LC/ Chemicals and reagents
TSP-MS for the assay of clenbuterol, while Debrauwer
and B o r i e ~have
~ ~ applied ESI-MS, after off-line LC -
Clenbuterol HCl, salbutamol- H,SO,, and terbutaline
separation, to the detection of clenbuterol residues.
Flow injection analysis (FIA) followed by TSP-tandem
-H,SO, (internal standard for salbutamol) were
obtained from Biomedica Foscama (Rome, Italy), Glaxo
mass spectrometry (MS/MS) has been used by van Wellcome (London, UK) and Pharmacia (Milan, Italy),
Rhijn et ~ 1 to .screen
~ j?,-agonists.
~ Recently, Doerge et respectively. A 1.00 mg ml-' stock solution of each
al. 3 s have proposed an LC/MS/MS method using an analyte was prepared in methanol and stored at
atmospheric pressure chemical ionization (APCI) inter- - 20 "C. Stock solutions were diluted in HPLC-grade
face for the screening of j?,-agonists in plasma extracts. water and kept in a refrigerator at about 4°C. A 5 pg
Tandem mass spectrometry yields very high detection -
ml- dilution of terbutaline H,S04 was prepared for
selectivity. By selecting appropriate parent ions in the
first mass analyser and by monitoring specific fragments
-
direct addition to urine samples. Clenbuterol-d, HCl,
obtained from Rijksinstituut voor Volksgezondheid en
(product ions) produced by collision of parent ions with Milieu (Bilthoven, The Netherlands) was used as an
an inert gas (e.g. argon) with the second mass analyser, internal standard for clenbuterol. Clenbuterol-d, was
the signal-to-noise ratio is increased considerably, thus dissolved in HPLC-gradc water at a concentration of 4
increasing the detectability of the analytes. Although pg m l '.~
MS/MS can be used for analytical purposes (with a Methanol (UV-grade), water (HPLC-grade) and tri-
direct probe or with FJA,34 for example), a purification ethylamine (p.a.) (99.5%) were obtained from Carlo
step prior to analysis is unavoidable, particularly in the Erba (Milan, Italy) and ammonium acetate (p.a.) (99%)
case of bovine urine samples containing non-volatile and formic acid (p.a.) (96%) from Aldrich (Milan, Italy).
compounds (e.g. inorganic salts). Indeed: almost all the Mobile phases containing 0.1 .M ammonium acetate,
LCjMS and MS/MS procedures available in the liter- 0.01 M triethylamine and 0.17 M formic acid consisting
ature adopt off-line sample preparation procedures of methanol-water (30: 70, v/v) and methanol-water
based on solid-phase extraction (SPE)33or liquid-liquid (18:82, v/v) were used for clenbuterol and salbutamol
partitioning or without purification by back- analysis, respectively. Before use, a freshly prepared
e~traction.~~ mobile phase was filtered through a type HA 0.45 pm
Column-switching LC can easily provide automated filter (Millipore-Waters, Milford, MA, USA).
clean-up procedures. However, recent reviews36337
reveal that the number of applications of this technique
on-line with TSP-MS detection are limited. Precolumn
switching has been successfully applied as a phase Equipment
switching system (PSS), providing both purification and
solvent compatibility with the TSP i n t e r f a ~ e . ~ ~ . ~ ~ The LC/MS instrumentation, depicted schematically in
We have recently investigated method development Fig. 1, consisted of the following components: a
strategies as applied in pesticides residue analysis using Rheodyne (Cotati, CA, USA) model 7125 manual injec-
coupled-column LC (LC/LC)40-44 in automated single- tor, IV; a Valco (VICI, Schenkon, Switzerland) EC4W
residue (SRM) and multi-residue analysis (MRM) of P2- security switching valve controlled by the MS system,
agonists in urine samples.45 The major advantage of SV; two Rheodyne model 7010 switching valves, HV-A
LC/LC is the utilization of the separation power of the and HV-B, installed and controlled by a MUST device
first separation column. This feature is particularly from Spark Holland (Emmen, The Netherlands); a
favourable in the case of polar analytes with little C,, Waters Model 510 isocratic LC pump, IP; a Waters
retention, such as terbutaline or salbutamol. Compared Model 600 MS gradient LC pump equipped with a
with automated sample processing procedures, such as flow-rate stabilization option and a helium degassing
precolumn switching (PC/LC) and SPE/LC using system, GP; a Waters Model 441 UV detector set at
hydrophobic materials, LC/LC offers the advantages of 280 nm, UV; a Perkin-Elmer (Norwalk, CT, USA)
efficient clean-up and chromatographic analysis in com- Model 561 recorder; a Finnigan MAT (San Jose, CA,
bination with large-volume injection of aqueous USA) TSQ-700 tandem mass spectrometer, MS/MS,
samples. equipped with a Finnigan MAT TSP-2 thermospray
In this study, we investigated the potential of the interface ; and a cold trap containing liquid nitrogen,
combination of both conventional LC and of LC/LC CT.
with TSP-MS/MS in the analysis of two B,-agonists Two 50 x 5.6 mm I.D. columns packed with 3 pm
selected as model compounds : clenbuterol and salbuta- Microspher C,, particles from Chrompack (Bergen op
mol. Using the results of our previous LC work4' and Zoom, The Netherlands) were used as the first (C-1) and
earlier mass spectrometric experimental the second (C-2) columns. A Chrompack 10 x 3 mm I.D.
420 E. A. HOGENDOORN ET AL.
M
U
Figure 1 . Overview of the LC/MS system. I V = injection valve with 5 0 0 pl loop; SV =security valve; M U S T = two-valve switching device:
-
I P = isocratic LC pump; G P =gradient LC pump; UV UV detector; TSP-MS/MS -thermospray tandem mass spectrometer; CT -solvent
cold trap; C-1, C - 2 =first and second separation columns; M = mobile phase used on C-1 and C-2; W =waste.
RP-guard column was inserted before each separation Quantification of clen buterol or salbutamol in bovine
column. A flow rate of 1.3 ml/min of the appropriate urine samples was performed by comparing the
mobile phase (see above) was used with both C-1 and clenbuterol/clenbuterol-d, or salbutamol/terbutaline
c-2. TSP-MS/MS response peak-area ratio obtained for the
The TSP interface was operated in the buffer ioniza- sample with that obtained for the injection of standards.
tion mode (filament and discharge off). The temperature In the case of clenbuterol, a calibration graph was pre-
of the source block was 200°C and the vaporizer tem- pared by randomly injecting standards of clenbuterol at
perature was 105 and 103 "C for clenbuterol and salbu- concentrations of 0.2, 1, 5, 10 and 20 ng m l - ' (each
tamol analysis, respectively. The repeller was set to 60 V point in triplicate). After each day of analysis both
in both cases. columns were washed with 20 ml of methanol, while the
MS/MS analysis was performed in the selected reac- TSP interface was washed with methanol (5 ml) fol-
tion monitoring, precursor ion mode. The reactions lowed by water (5 ml).
monitored were m/z 277 203 (clenbuterol), m/z
-+
TS P Repeller (V) 60 60
Vaporizer temperature ("C) 105 103
Source block temperature ("C) 200 200
MS/MS Collision energy (eV) -25 15
Collision gas (argon) pressure (mTorr) 1.7 1.2
Reactions monitored (rn/z) 277 --t 203 240 + 1 6 6
283 -+ 203 226 --t 1 5 2
LC/LC/TSP-MS/MS ANALYSIS OF P,-AGONISTS IN URINE 42 1
-
emphasized that columns packed with 3 pm C,, par-
ticles are particularly appropriate in this type of
analysis. With a flow rate of 1.3 ml min-' (selected for
adequate TSP interfacing), this type of column behaves
similarly to a high-speed column, reducing both the
elution volume and retention time of the anayte.
TSP-MS/MS conditions were determined by FIA (20
p1 loop injections) as mentioned above. Under these
0 4 8 12 min conditions, the limit of quantification (signal-to-noise
Figure 2. UV monitoring of the mobile phase ( M ) flowing on ratio (S/N) = 10) was about 10 pg of clenbuterol,
C - I , illustrating the sample sequence of the clenbuterol analysis Taking into account the increase in peak volume after
using LC/LC/TSP-MS/MS. A =injection of sample on C-1; A-
6 =clean-up time on C - I ; B-C = transfer time (C-I on-line with
chromatographic analysis, we calculated that a sample
C-2); C-A = TSP-MS/MS analysis of clenbuterol (C-1 off-line injection volume of 0.2 ml would be sufficient to detect
with C-2). clenbuterol at the required level of 0.1 ng ml-'.
However, to increase the robustness of the procedure, a
sample injection volume of 0.5 ml was selected.
both SRMs by FIA of clenbuterol and salbutamol in The column switching conditions, that is, the clean-
the respective mobile phases, increasing VT from 90 to up time on C-1 and the transfer time on the analyte-
120°C (in 5°C steps) and measuring absolute peak containing fraction from C-1 to C-2, are easily
areas. At the beginning of each session, the optimum determined by injecting 0.5 ml of a standard solution
VT was checked by FIA and adjusted if needed. An without column switching and monitoring the signal by
adjustment of 2 "C with respect to the previous value UV detection (see Fig. 1). It was our intention to
was sometimes required. perform a cleaning of C-1 with methanol after the
analysis of urine as in our previous As shown in
Fig. 1, the valve switching system was connected so that
Set-up of MS/MS conditions the gradient pump could perform this step after the
analysis of a sample. However, because of the high
Combined loss of water and methylpropene from the selectivity of the detection mode, we investigated the
protonated molecular ion to yield an ion of mass possibility of omitting this step. Experiments indicated
[MH -~741 is the most important fragmentation
+
that, when analysing a standard solution after each
pathway in the collisionaIly activated dissociation urine sample, a C-1 washing step was superfluous,
spectra of B,-agonists containing a tert-butylamino allowing a reduction in analysis time. In addition, accu-
group (e.g. clenbuterol, salbutamol and rate settings of the clean-up and transfer time were also
t e r b ~ t a l i n e ) . We
~ ~ ,therefore
~~ optimized the collision unnecessary, making the system very robust to the
gas (argon) pressure and collision energy to yield the changes in retention time expected because of contami-
maximum abundance of the [MH - 741 product ion +
nation of C-1 after repeated injections in the absence of
for each analyte. Selecting a collision gas pressure in the a washing step. The analytical strategy that was finally
range 1.2-1.7 mTorr, the optimum collision energy was adopted, consisting of analysis of a sample followed by
found to be -25 eV for clenbuterol and - 15 eV for analysis of standard, is illustrated in Fig. 2, which shows
salbutamol. In the case of salbutamol, higher collision the UV signal monitored on C-1. Under these condi-
energies resulted in a reduction in the abundance of the tions the total time of analysis is only 7 min for the
sample plus additional 7 min for the standard.
The linearity of the final procedure (see Experimental)
_ _ _ ~ ___
was tested by randomly analysing standards of clenbu-
Table 2. Clenbuterol recoveries from bovine urine terol with concentrations in the range 0.2-20 ng m1-I.
Spiked level No of Mean recovery RSD
The calibration plot obtained (y = 0 . 1 6 2 ~- 0.0189)
Day (ngim1) experiments We) ("/I showed good linearity over a two-decades range
1 0.22 3 106 9.5 (r2 = 0.9993). However, further experimental work
3 0.22 3 97 16.8 revealed that the precision of analysis of urine samples
1 5.3 3 107 7.0 containing less than 1.0 ng ml-' was improved by
3 5.3 6 108 7.8 applying a one-point calibration, using a standard con-
taining a concentration of clenbuterol close to that in
Overall recovery (YO) 105.5
10.4
the sample.
Repeatability (%)
Reproducibility (YO) 10.2 The method was validated by analysing blank
samples of bovine urine spiked with clenbuterol at
LC/LC/TSP-MS/MS ANALYSIS OF b,-AGONISTS IN URINE 423
levels of 0.22 and 5.3 ng ml-*, on two different days. tification limit of 0.1 ng ml-' (S/N = 10) can be easily
The results, given in Table 2, show that both the repea- achieved.
tability and reproducibility were satisfactory. The availability of a bovine urine sample in which a
Figure 3 shows the chromatograms obtained for the concentration of clenbuterol residue of 1.28 ng ml- '
LC/LC/TSP-MS/MS analysis of a blank bovine urine had been previously measured by GC/MS4 gave the
sample (A), and for the same urine sample spiked at a opportunity to investigate the performance of the
level of 0.22 ng ml- of clenbuterol (B). As can be seen, method on a real sample. In order to obtain informa-
the developed procedure is highly selective, and a quan- tion about both repeatability and reproducibility, the
El04
p.000
4001
zoo
A A
m/z 277+203
5:38
296114
100 -
3 588,lO1 - Et05
2.961
50-
mh 283-203
I " . - . -
5:40
3165'17
...__
301 1,613
100-
i
3.165
50-
=+05
mh 283+203
100-
4.758
50-
mh 277-203
C
5:40
281544
3 2 1 9,10 4
100- -6t05
2.845
50-
mh 283-203
, . . . ' I . . . I . . . . , ...,.... . .,.... , , . . .
Figure 3. Ion chromatograms obtained for the LC/LC/TSP-MS/MS analysis of clenbuterol in bovine urine samples. (A) Blank urine
sample; (B) blank urine sample spiked with clenbuterol at a level of 0.22 ng ml-' ; (C) bovine urine sample containing a clenbuterol residue
of 1.31 ng ml-'. The MS/MS transitions monitored for each analyte are indicated. Peaks: 1 -clenbuterol; 2 =internal standard.
424 E. A. HOGENDOORN ET AL.
1 1
2:36
955623
6341,501
9 .5 5 6
Figure 4. Ion chromatograms obtained for the LC/LC/TSP-MS/MS analysis of salbutamol in a bovine urine sample containing a residue Of
7.9 ng ml-' of unconjugated salbutamol. The MS/MS transitions monitored for each analyte are indicated. Peaks: 1 -salbutamol:
2 = internal standard.
LC/LC/TSP-MS/MS ANALYSIS OF B2-AGONISTS IN URINE 425
neither interferences in analyte detection nor deleterious samples of bovine urine. The two SRMs developed for
chromatographic effects were observed during the pro- clenbuterol and salbutamol demonstrate that the
cessing of about 40 samples per day. approach can be easily adapted to different p,-agonists
The method was validated by the analysis of blank and other analytes amenable to TSP-MS/MS. In addi-
bovine urine spiked with salbutamol at two different tion, the high sample throughput yields a favourable
levels, 0.55 and 10.8 ng rn1-I. The results, given in Table cost-benefit ratio for the expensive equipment used.
4, show acceptable recovery and repeatability at both Owing to the high selectivity of the detection mode,
levels. The limit of quantification (S/N = 10) was esti- the ability of LC/LC to yield an efficient clean-up is
mated to be 0.1 ng ml-'. hardly utilized. In the configuration used here, the main
The LC/LC/TSP-MS/MS method was also used for features of LC/LC are to provide (after a simple on-line
the analysis of three bovine urine samples which were clean-up step) (i) a very stable flow during the intro-
found to be positive for b,-agonists by enzyme-linked duction of the analyte into the TSP-MS/MS system and
immunosorbent assay.' The presence of salbutamol was (ii) the ability to perform large-volume sample injection
confirmed by GC/MS. The results of this experiment, and efficient chromatographic elution of polar analytes.
given in Table 5, show that the current methodology is All these features are favourable to sensitivity.
able to determine quantitatively unconjugated salbuta- Useful applications may be expected in the on-line
mol in bovine urine samples. Figure 4 shows the ion analysis of serum samples by applying an internal
chromatograms obtained for the LC/LC/TSP-MS/MS surface reversed-phase (ISRP) column as the first
analysis of a bovine urine sample (code A-10; see Table column in the proposed methodology. This interesting
5) in which a concentration of 7.9 ng ml-' of uncon- possibility and the potential of LC/LC/TSP-MS/MS in
jugated salbutamol was measured. The potential of the multi-residue analysis of b,-agonists will be a subject
described instrumental configuration in the direct for future research.
analysis of salbutamol conjugates was not investigated.
Acknowledgement
CONCLUSIONS
This work was carried out with financial support from the European
Commission (DG XII. Science and Technology, Division XII-H-l), in
the framework of the Human Capital and Mobility Programme, as
Our results show that the LC/LC/TSP-MS/MS part of the network project Hyphenation Analytical Chemistry for
hyphenated system is a very efficient approach for the Environmental and Public Health Research in the European Union
rapid, selective and sensitive analysis of p,-agonists in (Grant No. ERBCHRXCT930274).
REFERENCES
1 . J. P. Hanrahan, Befa-Agonisrs and Their €ffecfs on Animal 18. M. C. Dumasia and E. Houghton, J . Chromafogr. 564, 503
Growfh and Carcass Qualify. Elsevier Applied Science, (1991).
Barking (1987). 19. L. Leyssens, C Driessen, A. Jacobs, J. Czech and J. Raus, J.
2. J. F. Martinez-Navarro, Lancet 336,1311 (1991). Chromatogr. 564,51 5 (1991 ).
3. C. Puke, D. Lamaison, G. Keck, C. Bostvironnols, J. Nicolas 20. L. A. van Ginkel, R . W. Stephany and H. J. van Rossum, J.
and J. Descotes, Vet. Hum. Toxicol. 33,480(1 988). Assoc. Off.Anal. Chem. 75,554 (1992).
4. A. Polettini. A. Groppi. M. C. Ricossa and M. Montagna, Biol. 21. W. J . Blanchflower, S. A. Hewitt, A. Cannavan, C. T. Elliott
Mass Spectrom. 22.457 (1993). and D. G. Kennedy, Biol. Mass Specfrom. 22.326 (1993).
5. Muscling in on clenbuterol (editorial), Lancet 340, 403 22. T. Visser, M. J. Vredenbregt. A. P. J. M. de Jong, L. A. van
(1992). Ginkel, H. J. van Rossum and R . W. Stephany, Anal. Chim.
6. Infernal Report of Acfivifies. Games of the XXV Olympiad of Acfa 275,205 (1993).
Barcelona '92. Report IMIM/FT/DOPING/JJ00/1/1.2, lnstitut 23. M.-P. Montrade, B. Le Bizec, F. Monteau, B. Siliart and F.
Municipal d'lnvestigacio Medica. Barcelona (1992). Andre, Anal. Chim.Acta 275,253 (1993).
7. I. Yamamoto and K. Iwala, J. lmrnunoassay 3,155 (1982). 24. J. A. van Rhijn. W. A. Traag and H. H. Heskamp, J . Chromafo-
8. M . Paleologo Oriundi, G. Giacomini, F. Ballaben, F. Berti, F. gr. 619,243 (1993).
Benedetti, R . Bagnati and E. Bastiani, Food Agric. lmmunol. 4, 25. L. A. van Ginke1.J. Chromatogr. 564,363 (1991).
73 (1992). 26. A. Polettini, M. Montagna, J. Segura and X de la Torre, J .
9. M . J. Sauer, R . J . H. Pickett and A. L. MacKenzie, Anal. Chim. Mass Specfrom. 31,47 (1995).
Acta 275, 195 (1 993). 27. B. Diquet. L. Doare and P. Simon, J . Chromafogr. 336, 415
10. C. T. Elliott, J. D. G. McEvoy, W. J. McCaughey, D.H. Shortt (1 984).
and S. R. H. Crooks,Analyst 118,447(1993). 28. C. Eddins, J. Harnann and K. Johnson, J. Chromafogr.Sci. 23,
1 1 . J. G. Leferink, I. Wagernaker-Engels, R. A. A. Maes, H. 308 (1985).
Lamont, R . Pauwels and M. van der Straeten, J . Chromatogr. 29. H. H. D. Meyer, L. Rinke and I. Diirsch. J . Chrometogr. 564,
143,299 (1977). 551 (1991).
12. R . A. Clare, 0 . S. Davies and T. A. Baillie. Biomed. Environ. 30. H. Hooijerink. R . Schilt, W. Haasnoot and D. Courtheyn, J.
Mass Spectrom. 6,31 (1979). Pharm. Biomed. Anal. 9,485 (1991).
13. H. J. Forster. K. L. Rominger, E. Ecker, H. Peil and A. Wittrock, 31. D. Courtheyn, C. Desaever and R . Verhe, J. Chromafogr. 564.
Biomed. Environ. Mass Spectrom. 17.417 (1988). 537 (1991).
14. J. Schmid and A. Bucheler, Biorned. Environ. Mass Spectrom. 32. W. J. Blanchflower and D. G. Kennedy, Biomed. Environ.
17,415(1988). Mass Specfrom. 18,935 (1989).
15. J. Girault, P. Gobin and J. B. Fourtillan, Biomed. Environ. 33. L. Debrauwer and G. Bories. Anal. Chim. Acre 275, 331
Mass Spectrom. 19,80 (1990). (1 993).
16. J. Girault and J. B. Fourtillan, J . Chromafogr. 518,41 (1990). 34. J. A. van Rhijn, M. O'Keeffe. H. H. Heskamp and S.Collins, J.
17. A. Polettini. M. C. Ricossa, A. Groppi and M . Montagna, J. Chromafogr.A 712,67(1995).
Chromatogr. 564,529 (1 991 ). 35. D. R. Doerge, S. Bajic, L. R. Blamkenship, S . W. Preece and
426 E. A. HOGENDOORN ET AL.
M .I. Churchwell, J. Mass Specfrom. 30, 91 1 (1995). U. A. Th. Brinkman and P. van Zoonen, J. Chromatogr. Sci.
36. P. Arpino, Mass Spectrom. Rev. 9, 631 (1 990). 31,433 (1993).
37. P. Arpino, Mass Spectrom. Rev. 11, 3 (1992). 43. S.M. Gort, E. A. Hogendoorn, E. Dijkman, P. van Zoonen and
38. E. R . Verhey, H. J. E. M. Reeuwijk, G. F. LaVos, W. M. A. R. Hoogerbrugge, Chromarographia 42, 17 (1996).
Niessen, U. R . Tjaden and J. van der Greef, Biomed. Fnviron. 44. E. A. Hogendoorn and P. van Zoonen, J. Chromatogr. A 703,
Mass Specfrom. 16, 393 (1988). 149 (1995).
39. A. Walhagen, L.-E. Edholm, E. E. M . Heeremans, R . A. M. van 45. A. Polettini, M . Montagna, E. A. Hogendoorn, E. Dijkman, P.
der Hoeven, W. M. A. Niessen, U. R. Tjaden and J. van der van Zoonen and L. van Ginkel, J . Chromatogr. A 695, 19
Greef. J . Chromatogr. 474, 257 (1 989). (1995).
40. E. A. Hogendoorn, U. A. Th. Brinkman and P. van Zoonen, J. 46. A. Polettini, in Doping e Scienze Medico-Forensi, Proceedings
Chromatogr. 644, 307 (1 993). of r'Vl Convegno Nazionale Gruppo Tossicologi Forensi
41. E. A. Hogendoorn, R. Hoogerbrugge, P. van Zoonen, C. E. S.1.M.LA.". Pavia, September 1993, edited by M. Montagna,
Goewie and P. J. Schoenmakers, J . Chromafogr. 552, 113 0. Crippa and P. Maternini, p. 226. EDIMES, Pavia (1993).
(1 991 ). 47. W. Genuit and H. van Binsbergen, J . Chromatogr. 474, 145
42. E. A. Hogendoorn, R. Hoogerbrugge, E. Dijkman, S. M. Gort, (1989).