Sunteți pe pagina 1din 7

Polymer Testing 33 (2014) 138–144

Contents lists available at ScienceDirect

Polymer Testing
journal homepage: www.elsevier.com/locate/polytest

Test method

A novel method for the determination of steady-state torque


of polymer melts by HAAKE MiniLab
Cong Wang, Jing Wang, Chenyang Yu, Bingtian Wu, Ya Wang, Wenguang Li*
State Key Laboratory of Polymer Materials Engineering, Polymer Research Institute of Sichuan University, Chengdu, Sichuan 610065,
People’s Republic of China

a r t i c l e i n f o a b s t r a c t

Article history: HAAKE MiniLab is an apparatus specially designed for compounding polymer material and
Received 25 October 2013 on-line testing of rheological properties. For the first time, it was used to establish a dy-
Accepted 2 December 2013 namic speed test method for the rapid determination of steady-state torque of polymer
melts. The choices of sample feed quantity and screw rotation speed, as well as calibration
Keywords: for real torque, were carefully studied before torque measurements. The repeatability and
HAAKE MiniLab
reliability of torque data were also evaluated. Results showed that the torque could be
Steady-state torque
calibrated by subtracting the torque without samples. Also, a feed quantity of ca. 6 g with a
Dynamic speed test method
Repeatability
dynamic speed test range of 10–105 r/min was suitable for the determination of steady-
Reliability state torque of polyolefin samples. The new method was quick, effective and reliable to
correlate the steady-state torque with rotation speed. Therefore, MiniLab would be a very
useful tool in exploring and characterizing polymer flow behavior through its torque
measurements.
Ó 2013 Elsevier Ltd. All rights reserved.

1. Introduction real processing conditions. Therefore, many researchers


have used it to study the processing properties and mixing
In the polymer field, torque refers to the hindering force quality of polymers [5–12]. However, as the mixing of
met when polymer resins or compounds are plasticized polymer materials with the rotation of rotors generates
and mixed with rotation of screws or rotors. It has been heat from viscous dissipation, the pre-set temperature al-
used as an engineering indicator to learn the polymer ways changes during the mixing process, dependent upon
melting process for determining suitable processing con- material and experimental conditions such as mixing time
ditions, as well as to monitor industrial extrusion produc- and rotation rate. As a result, it is almost impossible to
tion for process quality control. Steady-state torque is the obtain a consistent torque for a polymer at a given tem-
measurement of the torque at equilibrium as the plasti- perature [13,14]. As for measuring the correlation of steady-
cized material is in a stable and homogeneous melt state. state torque with rotation rate at a given temperature, it is
For a pure polymer in a given process, the steady-state generally required to select several rotation rates and
torque is a characteristic of the polymer, which can be measure separately the variation of torque with time at
used as a rheological parameter to characterize the poly- each rotation rate. This is a time-consuming process with
mer flow behavior [1–4]. no certainty of reliable data. Only a few researchers have
The torque-rheometer is the most commonly used tool tried such experiments with no satisfactory results [15–17].
in polymer processing laboratory for torque measurements. For the above reasons, the application of torque-
It can simulate a mixer and the extrusion process in nearly rheometers for characterizing the properties of polymer
melts has been greatly restricted.
MiniLab Micro Compounder and Rheometer, devised by
* Corresponding author. Tel.: þ86 028 85460817; fax: þ86 028
85402465. HAAKE Company, Germany, is specially designed for the
E-mail address: wgli2007@126.com (W. Li). compounding of small volume (7 mL only) samples.

0142-9418/$ – see front matter Ó 2013 Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.polymertesting.2013.12.001
C. Wang et al. / Polymer Testing 33 (2014) 138–144 139

Simultaneously, the rheological properties can be recorded Table 1


to document structural changes. Its main features include Characteristics of materials used.

integrated torque and viscosity measurements, co- and Material Grade Melt flow index Supplier
counter rotation twin screws and an automatic bypass (g/10 min)
operation for circulation/extrusion with pneumatic HDPE 6070EA 7.2 (190  C/2.16 kg) Dushanzi Petrochemical,
feeding. Therefore, MiniLab is considered as a combination China
HDPE 60550 7.0 (190  C/2.16 kg) Lanzhou Petrochemical,
of a batch mixer, a twin-screw extruder and a rheometer.
China
Based on the unit’s specifications, we found that, in LDPE 18D 1.5 (190  C/2.16 kg) Daqing Petrochemical,
addition to small usage of samples, there are two distinct China
advantages in determining steady-state torque of polymer iPP T30S 2.0 (230  C/2.16 kg) Lanzhou Petrochemical,
melts by MiniLab over other torque-rheometers. First, China

MiniLab has a special backflow channel integrated with the


extrusion housing for circulation, which ensures a precise
and stable operating temperature, thus signifying that the unit (2), extruder housing (3) and feeding device (5). The
variation of polymer melt state with temperature due to control system includes the manual operating panel (4) and
viscous dissipation can be negligible. Second, MiniLab can the application software (1). Both the dynamic speed
perform dynamic speed measurements in a selected speed measurement and the experimental data documentation
range by a pre-set test program. We will show that in this can only be performed through the software. The drive unit
way the correlation of steady-state torque with rotation rate offers motor motion and precisely controls experimental
can be quickly obtained through only a single test, making it conditions, such as testing temperature and rotation speed.
possible and convenient to explore the rheological proper- It also measures the motor torque from a mixer sensor. The
ties of polymer melts with significant time and cost savings. structure of the extruder housing with co-rotation twin
Despite the prominent characteristics of MiniLab as stated screws is displayed in Fig. 2 (the co-rotation mode was used
above, published research still relates to using it only as a in this study). The filled-in sample can be extruded in a
conventional mixer to prepare experimental samples and to circulation via an integrated backflow channel. The back-
compound expensive materials such as nano-composites, flow channel is constructed as a rheological slit capillary
bio-polymers and pharmaceuticals [18–27]. No attention with two pressure sensors. While the sample is extruding
has been paid to its potential powerful function in torque through the backflow channel, the rheological information
measurements. The reason may be that the reliability of about the viscosity of the sample can be obtained from the
rheological measurements by MiniLab has not been evalu- two pressure sensors. Lastly, the feeding device enables
ated. Moreover, no appropriate method has been established periodic sample feeding by the piston of a pneumatic
for the determination of steady-state torque of polymer melts. cylinder.
This situation was motivation for the current research.
This paper presents a novel method, termed dynamic speed 2.3. Dynamic speed test method
test method, for the determination of steady-state torque of
polymer melts by means of the MiniLab’s functions. We will There are two stages in the dynamic speed test method.
introduce how to obtain repeatable, reliable torque data as In the first stage, samples were dynamically plasticized and
well as the correlation of steady-state torque with rotation melted under a rotation speed range predetermined ac-
rate in the MiniLab test. The calibration of torque data, sample cording to the requirement of the experiment and the na-
loads and screw rotation speeds were carefully investigated ture of the material. The number of speed points was set to
as regards the repeatability of data. The comparison of the 5–20 with equal distance logarithmic intervals (rpm)
new method with the common fixed speed method for within the speed range. At each speed, the software pro-
steady-state torque measurements was also done to gram measured the torque of the material, and then
demonstrate the reliability of data. Our results show that the
dynamic speed test is a quick, effective and reliable method
for determining steady-state torque of polymer melts.

2. Experimental

2.1. Materials

All the samples in this study were carefully selected


from commercial resins. These were high-density poly-
ethylene (HDPE), low-density polyethylene (LDPE) and
isotactic polypropylene (iPP). The characteristics of the
materials were shown in Table 1.

2.2. Instrumental details

The HAAKE MiniLab, as illustrated in Fig. 1, is made up of Fig. 1. The constitution of HAAKE MiniLab: (1) and (4) control system, (2)
five functional elements: control system (1 and 4), drive drive unit, (3) extruder housing, (5) feeding device.
140 C. Wang et al. / Polymer Testing 33 (2014) 138–144

2.7. Study on repeatability of torque data

Two polyethylenes, HDPE (6070 EA) and LDPE (18D)


with a feed quantity of 6 g, were chosen as test samples.
Plasticizing and measuring were done with a dynamic
speed range of 10–105 r/min. The temperature was set at
190  C for 6070 EA and at 150  C for 18D.

2.8. Study on reliability of torque data

Fig. 2. The structure of extruder housing: (1) backflow channel, (2) pressure
HDPE (6070 EA) was chosen as test material with a
sensor, (3) conical twin-screw, (4) bypass valve, (5) temperature sensor, (6)
extrusion channel. usage of 6 g. The comparison of steady-state torque
determined by the common fixed speed method was made
with that measured by the new dynamic speed test
selected 20 data points that met the program’s required method. In the fixed speed method, the test samples were
stability criteria. Afterwards, the instrument would auto- mixed for 8 min at speeds of 30 r/min, 50 r/min, 80 r/min
matically switch to the next speed point until the operating and 100 r/min. The torque values from 5 min to 8 min were
program was over. In the second stage, the same program averaged to obtain the steady-state torques under different
ran again in the same way as in the first stage. Then, the rotation speeds. For the dynamic speed test method, after
measured data were recorded to correlate steady-state being plasticized under a dynamic speed range of 10–105 r/
torque with rotation speed for polymer melts at the test min, the sample was measured in the same speed range to
temperature. Generally speaking, it only took about 15 min acquire the steady-state torques. The experimental tem-
to finish the whole measurement with a speed range of 10– perature in both methods was kept at 150  C.
105 r/min.
3. Results and discussion
2.4. Torque measurements by MiniLab
3.1. Torque calibration without samples
There were two main steps for the determination of
According to the instruction manual for the HAAKE
steady-state torque of the material. In the first step, device
MiniLab, before beginning the measurement, motor torque
setup was carried out and the measuring procedure was
without material but with rotating screws must be cali-
defined using the software. When the temperature reached
brated to its torque zero point. It was found that the torque
the set point, torque measurements without samples were
calibration in the manual worked well for a fixed speed but
performed to obtain a relationship between torque and
not for a dynamic speed test. Some torque values were not
rotation speed for calibration purposes. In the second step,
always accurate over a wide rotation speed range. Hence in
the torque measurements were done after the sample was
this paper, we calibrated torque through subtraction on the
fed into the extruder and plasticized. The correlation of
basis of the following.
steady-state torque with rotation speed was achieved via
Fig. 3 depicts the torque plotted against rotation speed
calculating the data exported from the above two-step
without samples in the MiniLab. It can be seen that there
measurements.
is a linear relationship between torque and rotation speed
within a rather wide range of 5–205 r/min, which can be
2.5. Choice of feed quantity expressed as y ¼ 0.0018*x þ 0.019, where x is the rotation
speed, y is the corresponding torque value. It should be
To choose a suitable feed quantity for torque measure- noted that the slope or intercept of the linear line might
ments, iPP (T30S) and HDPE (60550) with a dosage of 4–7 g vary slightly with the time passed. However, our MiniLab
were used. The test was carried out under a dynamic speed instrument, after many years of usage, has consistently
range of 10–105 r/min and the temperature was fixed at demonstrated this linear relationship between torque
190  C. and rotation speed. The instrument’s performance
was not subject to atrophy that may apply to other
instruments.
2.6. Choice of rotation speed Based on the above finding, we proposed a subtraction
approach to calibrate the torque of polymer melts. That is,
HDPE (6070 EA) with 6 g was selected as the sample and the real torque value was obtained by subtracting the tor-
three experiments were conducted as follows. Two samples que value without samples from the torque values with
were separately plasticized for 5 min under a fixed speed of samples. For any dynamic torque measurement, it was
50 r/min and 100 r/min, respectively, and another under a necessary to preferentially establish the correlation of tor-
dynamic speed range of 10–105 r/min. All the samples were que with rotation speed without samples before beginning
measured in the dynamic speed range of 10–105 r/min. The the measurement. Thus, the real torque of polymer melts
temperature was set at 150  C for both plasticizing and could be obtained throughout the dynamic speed test
measurement. method.
C. Wang et al. / Polymer Testing 33 (2014) 138–144 141

Fig. 3. Relationship of torque-rotation speed without samples.

3.2. Effects of feed quantity on torque measurements

The torque, by definition, is obviously dependent on the


amount of material used in a measurement. For the Mini-
Lab, the feed quantity needs to be suitable for obtaining a
consistent measuring torque value.
Fig. 4 outlines the effects of feed quantity on the steady-
state torque of T30S (a) and 60550 (b). It can be seen from
Fig. 4(a) that the torque of T30S significantly increased with
feed quantity varying from 4 g to 6 g at given rotation
speeds. When continuing to add up to 7 g, the steady-state
torque leveled off. It should be noted that the torque started
to fall with increased feed quantity from 6 g to 7 g at a
rotation speed of 105 r/min. The most probable explanation
of this trend was that the PP chains degraded with rotation
at such a high speed. Fig. 4. The relationship between steady-state torque and feed quantity to
As shown in Fig. 4(b) for 60550, it was also found that T30S (a) and 60550 (b).
the torque increased with the increased feed quantity from
4 g to 6 g. However, unlike T30S, the steady-state torque of
60550 was constant when the feed quantity was between repeatable and valuable data from torque measurements in
6 g–7 g, even at a high rotation speed of 105 r/min. This the MiniLab be obtained.
may be due to the better chain stability of HDPE than that of
PP. 3.3. Effects of rotation speed on torque measurements
The above results indicated that loads of ca. 6 g were
suitable for obtaining consistent torque values in the Steady-state torque depends, to an extent, on both
MiniLab. This is reasonable if the following calculation is plasticizing and measuring conditions, especially on the
considered. As bulk densities of PP and HDPE are about screw rotation speed. If samples are plasticized or
0.91 g/cm3 and 0.96 g/cm3, respectively, the volume of 6 g measured at a very high rotation speed, polymer chains can
sample could correspond to 6.6 mL for PP and 6.3 mL for be easily broken down, leading to a low torque value. If
HDPE at room temperature. When PP or HDPE is molten, samples are measured at a very low rotation speed, the
the polymer volume will expand somewhat. Since MiniLab torque may be too small to be accurately determined.
has a cavity of 7 mL volume, the feeding quantity of ca. 6 g Hence, there is an optimum rotation speed range for torque
polyolefin samples should completely fill the cavity, which measurements, in which the measured torque can accu-
leads to the maximum torque as expected. Therefore, the rately represent the rheological character of the material.
experimental results are in agreement with the theoretical Fig. 5 gives the steady-state torques of 6070 EA plasti-
calculations. cized under three different plasticizing conditions. It was
It should be mentioned that filling with 4–5.5 g of ma- observed that the torque-rotation speed curve of 6070 EA
terial, as suggested in the operating manual of MiniLab, was plasticized at a dynamic speed range of 10–105 r/min was
not suitable for torque measurements. Since the steady- in accordance with that plasticized at a fixed speed of 50 r/
state torque of polymer melts is strongly related to feed min, revealing that two plasticizing conditions could
quantity, it is important to determine the optimum quan- generate almost identical polymer melts with homoge-
tity for any unknown material. Only by doing that, can neous and stable states. However, for the sample
142 C. Wang et al. / Polymer Testing 33 (2014) 138–144

Fig. 5. The steady-state torque of 6070 EA plasticized under different rota- Fig. 7. Torque-speed curves of 18D measured in three consecutives.
tion speed.

105 r/min). Therefore, the data repeatability was very good


plasticized at a fixed speed of 100 r/min, the torque- for the same samples tested at different times. The results
rotation speed curve dramatically declined with also indicated that the working state of MiniLab was very
increasing speed as compared to that of other two samples. stable.
This trend indicates that polymer chains had been greatly Fig. 7 shows the relationship between torque and rota-
degraded under that plasticizing condition. tion speed of 18D measured on three consecutive tests after
Thus, from the above results, a dynamic speed range of being plasticized at 150  C. The torque curve in the first test
10–105 r/min was chosen for torque measurements. This was almost completely overlaid with that in the second
speed range was appropriate enough for the preparation of test. In the third test, it started to decline slightly, probably
samples with identical melt states. Therefore, the true due to degradation of LDPE chains with time at high tem-
torque values can be obtained within a relatively wide perature. The results demonstrated that the data repeat-
speed range. ability was also very good for the sample tested in the first
two replicates after the sample was plasticized.
3.4. Evaluating the repeatability of torque data In general, if test temperature and dynamic speed range
are well set up to minimize the degradation of polymer
Repeatability of measured data must be a primary cri- chains, we can always get repeatable torque measurements
terion of how well an instrument works. Fig. 6 shows the for any sample by MiniLab.
torque-rotation speed curves of 6070 EA measured at
different times (on March 23, May 17 and June 6 of 2012, 3.5. Evaluating the reliability of torque data
respectively). It can be seen that, although there was a
small deviation of data at the speeds of above 80 r/min, the Data reliability is critical for a measuring method.
highest standard deviation was only 2.15% (at a speed of Although the measurements were obtained quickly, the

Fig. 6. Torque-speed curves of 6070 EA measured at different times. Fig. 8. Torque measurements of 6070 EA under fixed speed.
C. Wang et al. / Polymer Testing 33 (2014) 138–144 143

Fig. 10 presents the comparison of the steady-state


torques determined by the two methods. It can be seen
that, except at high rotation speed where the torque
measured by the dynamic method was slightly higher than
that determined by the fixed speed method (due to small
degradation of polymer chains), all of other torque data
results were in agreement with each other. This revealed
that these two methods for steady-state torque measure-
ments worked very well in MiniLab. However, the dynamic
speed method had advantages over the common fixed
speed method because of time and cost savings, as well as
achieving accurate data and minimum degradation of
polymer chains.

4. Conclusions

We proposed a novel method for determination of


Fig. 9. Torque measurements of 6070 EA under dynamic speed.
steady-state torque of polymer melts using the features of
the MiniLab instrument. The torque was calibrated by
reliability of the torque data was not compromised. Fig. 8 subtracting that without material. The optimum feed
exhibits the variation of torque with time for 6070 EA quantity of ca. 6 g and a dynamic speed range of 10–105 r/
under the fixed speed test method. It can be seen from min were determined for torque measurements of poly-
Fig. 8 that, after feeding the sample, the torque values olefin samples. The data repeatability was evaluated by
under each fixed speed from 30 r/min to 100 r/min all measuring the same samples separately at different times
changed with time in the first 5 min. Thereafter, they or with sample replicates. The data reliability was
reached equilibrium values up to 8 min. As expected, the confirmed by comparing the new method with the com-
steady-state torques increased with increase of rotation mon fixed speed method for steady-state torque mea-
speed. It is important to note that, if the sample is added surements. The results showed that the dynamic speed test
into the extruder cavity all at once, the feed path becomes was a direct, effective and rapid method for the determi-
obstructed. Thus, the sample was divided into two portions, nation of steady-state torque of polymer melts by MiniLab.
resulting in two peaks on the graph. The torque data obtained were repeatable and reliable.
The dynamic speed test method includes plasticizing Moreover, the new method had a remarkable time and cost
and measuring stages. After the plasticizing stage, the savings as compared to the conventional method. The
sample becomes a homogeneous and stable melt. Thus, the correlation of steady-state torque with rotation speed was
torque of a sample quickly reaches an equilibrium value generally considered as a rheological curve for polymer
when the rotation speed changes from one speed to melts. This promises great potential in academic research
another. Fig. 9 describes the variation of torque with time in and industrial applications.
the measuring stage, by which the steady-state torques
were directly measured with 10 separate rotation speeds.
Please note that only one sample was used in the method References
and the steady-state torques could be obtained for 10
[1] J.E. Goodrich, R.S. Porter, A rheological interpretation of torque-
different rotation speeds within no more than 15 min. rheometer data, Polymer Engineering and Science 7 (1967) 45–51.
[2] L.L. Blyler, J.H. Daane, An analysis of brabender torque rheometer
data, Polymer Engineering and Science 7 (1967) 178–181.
[3] G.C.N. Lee, J.R. Purdon, Brabender viscometry: I. Conversion of
Brabender curves to instron flow curves, Polymer Engineering and
Science 9 (1969) 360–364.
[4] J.G. Mallette, R.R. Soberanis, Evaluation of rheological properties of
non-newtonian fluids in internal mixers: an alternative method
based on the power law model, Polymer Engineering and Science 38
(1998) 1436–1442.
[5] E. Freire, O. Bianchi, E.E.C. Minteiro, et al., Processability of PVDF/
PMMA blends studied by torque rheometry, Materials Science and
Engineering: C 29 (2009) 657–661.
[6] X. Wang, X.Y. Chen, W.C. Yu, et al., Applications of rheological
torque-time curves to the study of thermooxidative degradation of
polypropylene powder, Journal of Applied Polymer Science 105
(2007) 1316–1330.
[7] O. Breuer, H.B. Chen, B. Lin, et al., Simulation and visualization of
flow in a new miniature mixer for multiphase polymer systems,
Journal of Applied Polymer Science 97 (2005) 136–142.
[8] V. Nassehi, M.H.R. Glloreishy, Modeling of mixing in internal mixers
with long blade tips, Advances in Polymer Technology 20 (2001)
132–145.
[9] L. Stradins, T.A. Osswald, Predicting the effect of viscosity ratios on
the mixing of polymer blends using the boundary element method,
Fig. 10. Comparison of two methods. Polymer Engineering and Science 36 (1996) 979–984.
144 C. Wang et al. / Polymer Testing 33 (2014) 138–144

[10] T. Avalosse, M.J. Crochet, Finite-element simulation of mixing: 2. [20] S. Migneault, A. Koubaa, F. Erchiqui, et al., Effect of fiber length on
Three-dimensional flow through a kenics mixer, AIChE Journal 43 processing and properties of extruded wood-fiber/HDPE compos-
(1997) 588–597. ites, Journal of Applied Polymer Science 110 (2008) 1085–1092.
[11] H.E. Burch, C.E. Scott, Effect of viscosity ratio on structure evolution [21] Y.J. Xu, W.J. Brittain, C.C. Xue, et al., Effect of clay type on
in miscible polymer blends, Polymer Engineering and Science 42 morphology and thermal stability of PMMA-clay nanocomposites
(2001) 7313–7325. prepared by heterocoagulation method, Polymer 45 (2004) 3735–
[12] C.H. Yao, Z.I. Manas, R. Regalia, et al., Distributive mixing in variable 3746.
intermeshing clearance mixers: simulation and experiments, Rub- [22] A. Walther, K. Matussek, A.H.E. Müller, Engineering nanostructured
ber Chemistry and Technology 71 (1998) 690–708. polymer blends with controlled nanoparticle location using Janus
[13] A. Marquez, J. Quijano, M. Gaulin, A calibration technique to eval- particles, ACS Nano 2 (2008) 1167–1178.
uate the power-law parameters of polymer melts using a torque- [23] S.C. Wong, A. Baji, Fracture strength and adhesive strength of
rheometer, Polymer Engineering and Science 36 (1996) 2556–2563. hydroxyapatite-filled polycaprolactone, Journal of Materials Sci-
[14] B.J. Cheng, C.X. Zhou, W. Yu, et al., Evaluation of rheological pa- ence: Materials in Medicine 19 (2008) 929–936.
rameters of polymer melts in torque rheometers, Polymer Testing [24] Y. Sabba, E.L. Thomas, High-concentration dispersion of single-wall
20 (2001) 811–818. carbon nanotubes, Macromolecules 37 (2004) 4815–4820.
[15] T. Weber, M.G. Oliveira, et al., Processability of revulcanizable SBR [25] O. Valentino, M. Sarno, N.G. Rainone, et al., Influence of the polymer
compositions, Polymer Bulletin 61 (2008) 217–224. structure and nanotube concentration on the conductivity and
[16] S.C. Liu, L.N. Lu, Melt rheological properties of reactive compatibi- rheological properties of polyethylene/CNT composites, Physica E:
lized HDPE/PET blends, Journal of Applied Polymer Science 108 Low-dimensional Systems and Nanostructures 40 (2008) 2440–
(2008) 3559–3564. 2445.
[17] A.Q. Zhang, L.S. Wang, Y.Y. Zhou, A study on rheological properties [26] A. Loiseau, J.F. Tassin, Model nanocomposites based on laponite and
of carbon black extended powdered SBR using a torque rheometer, poly(ethylene oxide): preparation and rheology, Macromolecules 39
Polymer Testing 22 (2003) 133–141. (2006) 9185–9191.
[18] L. Adragna, F. Couenne, P. Cassagnau, et al., Modeling of the complex [27] W.E. Dondero, R.E. Gorga, Morphological and mechanical properties
mixing process in internal mixers, Industrial & Engineering Chem- of carbon nanotube/polymer composites via melt compounding,
istry Research 46 (2007) 7328–7339. Journal of Polymer Science Part B: Polymer Physics 44 (2006) 864–
[19] W. Yang, Z.Y. Liu, G.F. Shan, et al., Study on the melt flow behavior of 878.
glass bead filled polypropylene, Polymer Testing 24 (2005) 490–
497.

S-ar putea să vă placă și