Documente Academic
Documente Profesional
Documente Cultură
Radiology
756
first medical CT scanners were two-sec- angiography of lower extremity vessels
ESSENTIALS tion systems, such as the EMI (England) (12), thicker (eg, 2.5-mm) collimated sec-
head scanner, introduced in 1972, and tions had to be chosen to complete the
● Multi– detector row CT allows substan- the Siemens Siretom (Erlangen, Ger- scan within a reasonable time frame.
tial reduction in examination time for many), introduced in 1974. With the ad- Scan times were often too long to allow
standard protocols, coverage of ex- vent of whole-body fan-beam CT systems image acquisition during a purely arterial
Radiology
tended anatomic volumes, and, most for general radiology, two-section acqui- phase. For CT angiography of the circle
important, substantially increased lon- sition was no longer used. Apart from a of Willis, for instance, a scan range of
gitudinal resolution by means of re- dedicated two-section system for cardiac about 100 mm must be covered (14).
duced section width. applications, the Imatron C-100 (Ima- With four-section CT at a collimated sec-
tron, San Francisco, Calif), which was in- tion width of 1 mm, pitch of 1.5, and
troduced in 1984, the first step toward gantry rotation time of 0.5 second, this
● Near-isotropic spatial resolution in rou-
multisection acquisition in general radi- volume can be covered in about 9 sec-
tine examinations, which has been
ology was a two-section CT scanner in- onds, not fast enough to avoid venous
achieved with 16-section CT systems, troduced in 1993 (Elscint TWIN; Elscint, overlay, assuming a cerebral circulation
enables 3D renderings of diagnostic Haifa, Israel) (6). In 1998, several CT time of less than 5 seconds. (Note: The
quality and oblique MPRs and maxi- manufacturers introduced multi– detec- definition of pitch for multi– detector
mum intensity projections with resolu- tor row CT systems, which provided con- row CT is discussed later in this review.)
tion similar to that of the transverse siderable improvement in scanning As a next step, the introduction of an
images. speed and longitudinal resolution and eight– detector row CT system in 2000
better utilization of the available x-ray enabled shorter scan times but did not
● Scanning at narrow collimation does power (7–10). These systems typically of- yet provide improved longitudinal reso-
fered simultaneous acquisition of four lution (thinnest collimation, eight sec-
not markedly increase the radiation
sections at a gantry rotation time of 0.5 tions at 1.25 mm). The latter was
dose to the patient, as long as the ef-
second. achieved with the introduction of 16 –
fective milliampere-seconds level is kept Simultaneous acquisition of m sections detector row CT (15), which made possi-
constant. results in an m-fold increase in speed if all ble the routine acquisition of substantial
other parameters (eg, section thickness) anatomic volumes with isotropic submil-
● A key challenge for image reconstruc- are unchanged. This increased perfor- limeter spatial resolution and scan times
tion with multi– detector row CT is the mance of multi– detector row CT relative of less than 10 seconds for 300 mm of
cone angle of the measurement rays; to single-section CT allowed the optimi- coverage (Fig 1). While in-plane spatial
this requires novel reconstruction tech- zation of a variety of clinical protocols. resolution is not substantially improved,
niques such as 3D back projection, The examination time for standard pro- the two major advantages of fast multi–
AMPR, or weighted hyperplane recon- tocols could be substantially reduced, detector row CT are a true isotropic
which proved to be of immediate clinical through-plane resolution and a short ac-
struction.
benefit for the quick and comprehensive quisition time that enable high-quality
assessment of trauma patients and unco- examinations in severely debilitated and
● Z filtering makes it possible to recon- operative patients (11). Alternatively, the severely dyspneic patients (Fig 1).
struct images retrospectively with dif- scan range that could be covered within a Traditional CT applications have been
ferent section widths from the same certain time was extended by a factor of enhanced and strengthened by the remark-
raw CT data set, trading off, in this m, which is relevant for oncologic stag- able, although incremental, improvement
way, z-axis resolution and image noise. ing or for CT angiography with extended in scanner performance by the addition
coverage (eg, the lower extremities) (12). of more detector rows. Multi–detector row
The most important clinical benefit, CT also dramatically expanded into areas
however, proved to be the ability to scan previously considered beyond the scope
a given anatomic volume within a given of third-generation CT scanners that
longitudinal (z-axis) resolution along the scan time with substantially reduced sec- were based on the mechanical rotation of
patient axis is determined by the col- tion width at m times increased longitu- an x-ray tube and detectors, such as car-
limated section width and the spiral dinal resolution. Because of this, the goal diac imaging with the addition of elec-
interpolation algorithm. Use of a thick of isotropic resolution was within reach trocardiographic (ECG)-gating capabil-
collimation of 5– 8 mm results in a con- for many clinical applications. Examina- ity. With a gantry rotation time of 0.5
siderable mismatch between the longitu- tions of the entire thorax (13) or abdo- second and dedicated image-reconstruc-
dinal resolution and the in-plane resolu- men could now be routinely performed tion approaches, the temporal resolution
tion, which is 0.5– 0.7 mm, depending on with a 1.0- or 1.25-mm collimated section for acquisition of an image was improved
the reconstruction kernel. Thus, with sin- width. Despite these promising advances, to 250 msec and less (16,17), which
gle-section spiral CT, the ideal of isotro- clinical challenges and limitations re- proved to be sufficient for motion-free
pic resolution can only be achieved for mained for four-section CT systems. True imaging of the heart in the mid- to end-
very limited scan ranges (5). isotropic resolution for routine applica- diastolic phase when the patient had a
Strategies to achieve more substantial tions had not yet been achieved, because slow to moderate heart rate (ie, up to 65
volume coverage with improved longitu- the longitudinal resolution of about 1 beats per minute [18]). With four simul-
dinal resolution include the simulta- mm does not fully match the in-plane taneously acquired sections, coverage of
neous acquisition of more than one sec- resolution of about 0.5– 0.7 mm in a rou- the entire heart volume with thin sec-
tion at a time and a reduction in the tine examination of the chest or abdo- tions (ie, four sections at 1.0- or 1.25-mm
gantry rotation time. Interestingly, the men. For large volumes, such as for CT collimation) within a single breath hold
CURRENT TECHNIQUES
System Design
Detector design.—For clinical purposes,
different section widths must be avail-
able to adjust the optimum scan speed,
Figure 2. Illustration shows prepatient collimation of the x-ray longitudinal resolution, and image noise
beam to obtain different collimated section widths with a single– for each application. With a single– de-
detector row CT detector. FOV ⫽ field of view. tector row CT scanner, different colli-
mated section widths are obtained by
means of prepatient collimation of the
became feasible. This 1.0 –1.25-mm lon- lemma, mainly because of partial volume x-ray beam (Fig 2). For a very elementary
gitudinal resolution combined with the artifacts as a consequence of insufficient model of a two-section CT scanner (m ⫽
improved contrast resolution of modern longitudinal resolution (22). For patients 2, or two detector rows), Figure 2 demon-
CT systems enabled noninvasive depic- with a higher heart rate, careful selection strates how different section widths can
tion of the coronary arteries (19 –22). Ini- of separate reconstruction intervals for be obtained by means of prepatient col-
tial clinical studies demonstrated the po- different coronary arteries has been man- limation if the detector is separated mid-
tential of multi– detector row CT to not datory (25). It is almost impossible for way along the z-axis extent of the x-ray
only demonstrate but to some degree patients with manifest heart disease to beam. For m ⬎ 2, this simple design prin-
also characterize noncalcified and calci- comply with the breath-hold time of ciple must be replaced by more flexible
fied plaques in the coronary arteries on about 40 seconds required to cover the concepts requiring more than m detector
the basis of plaque CT attenuation entire heart volume (approximately 12 rows to simultaneously acquire m sec-
(22,23). cm) with four-section CT. The ongoing tions.
The limitations of four– and eight– de- technical refinement of multi– detector Different manufacturers of multi– de-
tector row CT systems, however, have so row CT, however, holds the promise of tector row CT scanners have introduced
far prevented the successful integration gradually overcoming some of these lim- different detector designs. In order to be
of CT coronary angiography into routine itations. The most important steps to- able to select different section widths, all
clinical algorithms: Stents or severely cal- ward this goal are gantry rotation times scanners combine several detector rows
cified arteries constitute a diagnostic di- faster than 0.5 second (26,27) for im- electronically to a smaller number of sec-
adjust both the functional form and the thick sections for initial viewing and re-
FWHM of the spiral SSP. Still, the cone cording and thin sections for detailed di-
angle is neglected. A representative ex- agnosis or as an input for advanced 3D
ample of a z-filter approach is the adap- postprocessing are routinely reconstructed.
tive axial interpolation algorithm (51) The thinnest available section width is
implemented in Siemens CT scanners, the collimated section width (1.0 mm for
which is illustrated in Figure E5 (radiology four sections at 1.0-mm collimation),
.rsnajnls.org/cgi/content/full/2353040037 which is created by using nonlinear spiral
/DC1). Another example is the “multislice weighting functions at the expense of in-
cone-beam tomography,” or MUSCOT, al- creased image noise and increased sus-
gorithm (52) used by Toshiba. Z filtering ceptibility to artifacts. Thus, as a take-
allows the system to trade off z-axis res- home point, the thinnest available
olution (the SSP) with image noise section should only be used for high-con-
(which directly correlates with required trast applications such as high-spatial-
dose). resolution lung imaging. For general pur-
With adaptive axial interpolation, the pose scanning, a 1.25-mm section width
spiral pitch is freely selectable in the for four-section CT at 1.0-mm collima-
range 0.5–2.0, and the same effective sec- tion (and 3.0-mm section width for four
tion width, which is defined as the sections at 2.5-mm collimation) is recom-
FWHM of the spiral SSP, is generated at mended as the most suitable trade-off
all pitch values (7,51,53). Therefore, lon- between longitudinal resolution, image
gitudinal resolution is independent of noise, and artifacts, in particular when thin
pitch, unlike single-section spiral CT and sections are reconstructed as an input for
multi– detector row CT that relies on 180° 3D postprocessing such as for MPR, max-
and 360° linear interpolation (51,54). Fig- imum intensity projection, or volume-
ure E6 (radiology.rsnajnls.org/cgi/content rendering techniques. For a 1.25-mm spi-
/full/2353040037/DC1) shows the SSPs of a ral section width reconstructed from four-
2-mm section (for four-section CT at 1-mm section CT at 1.0-mm collimation, 0.61–
collimation) and MPRs of a spiral z-axis 0.69 times the dose (depending only
resolution phantom for selected pitch val- slightly on spiral pitch) is required to
ues. As a consequence of the pitch-inde- maintain the image noise of a sequential
pendent spiral section width, the image scan at the same collimation (see refer-
noise for a fixed tube current (in milliam- ences 54,55). Unlike 180° and 360° mul-
peres) would decrease as pitch is decreased, tidetector linear interpolation, image
owing to the increasingly overlapping spi- noise is therefore practically indepen-
ral acquisition. Instead, the user selects an dent of pitch at constant dose.
“effective” milliampere-seconds value, and For a given collimation, such as four
the tube current is then automatically sections at 2.5 mm, image quality can be
adapted to the pitch of the spiral scan to optimized with regard to spiral artifacts
compensate for dose accumulation. The by lowering the pitch (56). Another
dose for fixed effective milliampere-sec- means to reduce spiral artifacts is to use
onds is independent of the spiral pitch and narrow collimation: A given section
equals the dose of a transverse scan with width (eg, 3.0 mm) can be obtained with
the same milliampere-seconds setting (see different collimations, in this case four
Appendix E2, radiology.rsnajnls.org/cgi sections at 1.0 mm and at 2.5 mm. For
/content/full/2353040037/DC1). optimum image quality, collimation that
Thus, as a take-home point, unlike in is narrow relative to the desired section
Figure 10. (a) SSP of 3-mm section for four-sec- single-section spiral CT a change in pitch width is preferable (51). Furthermore, a
tion CT at 1.0- and 2.5-mm collimation. (b) Images does not result in a change in dose to the more rectangular SSP can be established.
of thorax phantom with 3.0-mm section width patient. Accordingly, the use of a higher Figure 10a shows the SSPs of a 3.0-mm
obtained from four-section CT at 2.5 mm colli- pitch does not result in a dose saving, section for four-section CT at both 1.0-
mation and pitch of 0.75 (top) and at 1.0-mm
collimation and pitch of 1.75 (bottom). Despite
which is an important practical consider- and 2.5-mm collimation. Figure 10b
higher pitch, image acquired with 1.0-mm colli- ation with CT systems that rely on adap- shows 3.0-mm transverse sections of a
mation shows reduced artifacts at ribs. tive axial interpolation. thorax phantom scanned with four-sec-
The intrinsic resolution of a multi– de- tion CT at 2.5- and 1.0-mm collimation.
tector row spiral scan is determined by Despite the higher pitch, the 3.0-mm im-
the choice of collimation (eg, four sec- age obtained at 1.0-mm collimation
lation for each projection angle is no tions at 1.0 or 2.5 mm). Z filtering makes shows fewer artifacts. Similar to single-
longer restricted to the two rays closest to it possible to reconstruct images retro- section spiral CT, narrow collimation at
sion).
The major improvements of 16-section
CT, compared with established four-sec-
tion scanners, include improved temporal
resolution due to shorter gantry rotation
time, better spatial resolution owing to sub-
millimeter collimation, and considerably re-
duced scan acquisition times (26,27). The
time to cover the entire heart volume
(about 12 cm) with four-section CT at
1.0-mm collimation is about 40 seconds,
which is at the limit for a scan requiring
patient breath holding. ECG-gated CT of
the entire thorax or the aorta is not pos-
sible within reasonable scan durations.
For a 16-section CT system, the time to
cover the entire heart volume with sub-
millimeter collimation is about 15 sec-
onds. With 16-section CT, coverage of
the entire thorax (30 cm) can be com-
pleted in about 38 seconds at 0.75-mm
Figure 16. Clinical performance of ECG-gated 16-section CT collimation and in about 19 seconds at
(0.75-mm collimation, 0.42-second gantry rotation) of entire thorax. 1.5-mm collimation. ECG-gated exami-
Coronal volume-rendered reconstruction shows left internal mam- nations of extended cardiothoracic anat-
mary bypass graft (LIMA) to left anterior descending coronary artery omy became feasible with 16-section CT,
and saphenous venous bypass graft (SVG) to right coronary artery. which lends itself to a spectrum of appli-
Native right internal mammary artery (RIMA) is also visible.
cations where suppression of cardiac
pulsation is desired. Typical diagnostic
pitfalls caused by transmitted cardiac
struction mode (16,73–77), where trot is different approach, single-segment par- pulsation can be avoided, such as an ar-
the gantry rotation time of the CT scan- tial-scan images are prospectively recon- tifactual intimal flap resembling dissec-
ner. With increased N, better temporal structed as baseline images, followed by tion in the ascending aorta (79). Suppres-
resolution is achieved but at the expense retrospective two-segment reconstruc- sion of cardiac pulsation improves the
of slower volume coverage: Increased N tion for improved temporal resolution in assessment of paracardiac lung segments
and slower patient heart rate require a patients with a higher heart rate. Yet an- and allows confident exclusion of small
reduction in spiral pitch. other approach is prospective adjustment peripheral pulmonary emboli in segmen-
Multisegment approaches rely on a com- of the gantry rotation time to the heart tal and subsegmental arteries (80). In rou-
plete periodicity of the heart motion, and rate of the patient to obtain an optimized tine thoracic studies, which are not syn-
these approaches encounter their limita- temporal resolution for a multisegment chronized to the patient’s ECG signal,
tions in patients with arrhythmia or a reconstruction. Again, this approach re- cardiac motion usually precludes the assess-
heart rate that changes during scan acqui- quires a stable and predictable heart rate ment of coronary bypass grafts. Figure 16
sition. Multisegment reconstruction may during scan acquisition. shows an example of an ECG-gated scan
improve image quality in selected cases, Prospective ECG triggering combined of the entire thorax for a patient with
but the reliability of good-quality image with sequential step-and-shoot acquisition bypass grafts; this scan was acquired with
acquisitions with N-segment reconstruc- of transverse sections has the benefit of 16-section CT at 0.75-mm collimation
tion is compromised with increases in N. smaller patient dose than that of ECG- and 0.42-second gantry rotation.
In general, clinical practice suggests gated spiral scanning, because scan data
the use of one segment at lower heart are acquired only during the desired heart
rates and two or more (N ⱖ 2) segments phases. However, this technique does not APPLICATIONS
at higher heart rates. Use of single-seg- provide continuous volume coverage with
ment versus multisegment reconstruc- overlapping sections, and misregistration Clinical applications benefit from multi–
tion is integrated in the data acquisition of anatomic details cannot be avoided. Fur- detector row CT technology in several
process in a variety of ways, depending thermore, reconstruction of images in dif- ways: (a) shorter scan time (important for
on the scanner type. One approach con- ferent phases of the cardiac cycle for func- trauma patients and pediatric patients, CT
sists of automatic division of the partial- tional evaluation is not possible. Since angiography), (b) extended scan range (im-
scan data segment into one or two sub- ECG-triggered sequential scanning de- portant for CT angiography, combined
segments, depending on the patient’s pends on a reliable prediction of the pa- chest-abdomen scans such as in oncologic
heart rate during acquisition (“adaptive tient’s next R-R interval by using the mean staging), and (c) improved longitudinal res-
cardio volume” algorithm [74]). With a of the preceding R-R intervals, the method olution (beneficial for all reconstructions,
spectively ECG-gated multi-slice spiral CT. voltages for protocol optimization in pediat-
1. Kalender W, Seissler W, Klotz E, Vock P. Spi- Circulation 2000; 102:2823–2828. ric CT (abstr). Radiology 2001; 221(P):366.
ral volumetric CT with single-breath-hold 20. Becker C, Knez A, Ohnesorge B, Schöpf U, 39. Wintersperger B, Becker CR, Schaller S, Rist
technique, continuous transport, and con- Reiser M. Imaging of non calcified coronary C, Jakobs TF, Reiser MF. Low dose 16-detec-
tinuous scanner rotation. Radiology 1990; plaques using helical CT with retrospective tor-row CT angiography of the abdominal
176:181–183. EKG gating. AJR Am J Roentgenol 2000; 175: vessels with improved vessel enhancement
2. Crawford CR, King KF. Computed tomogra- 423– 424. (abstr). In: Radiological Society of North
phy scanning with simultaneous patient 21. Knez A, Becker C, Leber A, Ohnesorge B, America scientific assembly and annual
translation. Med Phys 1990; 17:967–982. Reiser M, Haberl R. Non-invasive assessment meeting program. Oak Brook, Ill: Radiologi-
3. Rubin GD, Dake MD, Semba CP. Current sta- of coronary artery stenoses with multidetec- cal Society of North America, 2003; 541–542.
tus of three-dimensional spiral CT scanning tor helical computed tomography. Circula-
for imaging the vasculature. Radiol Clin 40. Kalender WA, Wolf H, Suess C. Dose reduc-
tion 2000; 101:e221– e222. tion in CT by anatomically adapted tube cur-
North Am 1995; 33:51–70. 22. Nieman K, Oudkerk M, Rensing BJ, et al.
4. Napel S, Rubin GD, Jeffrey RB. STS-MIP: a rent modulation. II. Phantom measure-
Coronary angiography with multi-slice com- ments. Med Phys 1999; 26:2248 –2253.
new reconstruction technique for CT of the puted tomography. Lancet 2001; 357:599 –
chest. J Comput Assist Tomogr 1993; 17:832– 41. Greess H, Wolf H, Baum U, et al. Dose reduc-
603. tion in computed tomography by attenua-
838. 23. Schroeder S, Kopp A, Baumbach A, et al.
5. Kalender W. Thin-section three-dimensional tion-based on-line modulation of the tube
Noninvasive detection and evaluation of current: evaluation of six anatomical re-
spiral CT: is isotropic imaging possible? (ed-
atherosclerotic coronary plaques with multi- gions. Eur Radiol 2000; 10:391–394.
itorial). Radiology 1995; 197:578 –580.
slice computed tomography. J Am Coll Car- 42. Jakobs TF, Becker CR, Ohnesorge B, et al.
6. Liang Y, Kruger RA. Dual-slice spiral versus
diol 2001; 37:1430 –1435. Multislice helical CT of the heart with retro-
single-slice spiral scanning: comparison of
24. Schroeder S, Flohr T, Kopp AF, et al. Accuracy spective ECG gating: reduction of radiation
the physical performance of two computed
of density measurements within plaques lo- exposure by ECG-controlled tube current
tomography scanners. Med Phys 1996; 23:
cated in artificial coronary arteries by x-ray modulation. Eur Radiol 2002; 12:1081–1086.
205–220.
multislice CT: results of a phantom study. 43. Poll L, Cohnen M, Brachten S, Ewen K, Mod-
7. Klingenbeck-Regn K, Schaller S, Flohr T,
J Comput Assist Tomogr 2001; 25:900 –906. der U. Dose reduction in multidetector row
Ohnesorge B, Kopp AF, Baum U. Subsecond
25. Kopp A, Schröder S, Küttner A, et al. Coro- CT of the heart by use of ECG-controlled
multi-slice computed tomography: basics
nary arteries: retrospectively ECG-gated tube current modulation (“ECG pulsing”):
and applications. Eur J Radiol 1999; 31:110 –
multi– detector row CT angiography with
124. phantom measurements. Rofo Fortschr Geb
selective optimization of the image recon-
8. Hu H, He HD, Foley WD, Fox SH. Four mul- Rontgenstr Neuen Bildgeb Verfahr 2002;
struction window. Radiology 2001; 221:
tidetector-row helical CT: image quality and 174:1500 –1505.
volume coverage speed. Radiology 2000; 683– 688.
44. Hsieh J. Investigation of the slice sensitivity
215:55– 62. 26. Flohr T, Bruder H, Stierstorfer K, Simon J,
profile for step-and-shoot mode multi-slice
9. Ohnesorge B, Flohr T, Schaller S, et al. Tech- Schaller S, Ohnesorge B. New technical de-
computed tomography. Med Phys 2001; 28:
nische grundlagen und anwendungen der velopments in multislice CT. II. Sub-millime-
491–500.
mehrschicht-CT. Radiologe 1999; 39:923– ter 16-slice scanning and increased gantry
45. Joseph PM, Spital RD. The exponential edge-
931. rotation speed for cardiac imaging. Rofo
gradient effect in x-ray computed tomogra-
10. McCollough CH, Zink FE. Performance eval- Fortschr Geb Rontgenstr Neuen Bildgeb Ver-
phy. Phys Med Biol 1981; 26:473– 487.
uation of a multidetector row CT system. fahr 2002; 174:1022–1027.
27. Flohr T, Schoepf UJ, Kuettner A, et al. Ad- 46. Tomandl BF, Klotz E, Handschu R, et al.
Med Phys 1999; 26:2223–2230. Comprehensive imaging of ischemic stroke
11. Remy-Jardin M, Tillie-Leblond I, Szapiro D, vances in cardiac imaging with 16-section
CT-systems. Acad Radiol 2003; 10:386 – 401. with multisection CT. RadioGraphics 2003;
et al. CT angiography of pulmonary embo- 23:565–592.
lism in patients with underlying respiratory 28. Schoepf UJ, Becker CR, Ohnesorge BM, Yucel
EK. CT of coronary artery disease. Radiology 47. Polacin A, Kalender WA, Marchal G. Evalua-
disease: impact of multislice CT on image tion of section sensitivity profiles and image
quality and negative predictive values. Eur 2004; 232:18 –37.
29. Hu H. Multi-slice helical CT: Scan and recon- noise in spiral CT. Radiology 1992; 185:29 –
Radiol 2002; 12:1971–1978. 35.
12. Rubin GD, Schmidt AJ, Logan LJ, Sofilos MC. struction. Med Phys 1999; 26:5–18.
30. Brenner D, Elliston C, Hall E, Berdon W. 48. Rubin GD, Napel S. Increased scan pitch for
Multi– detector row CT angiography of lower vascular and thoracic spiral CT. Radiology
extremity arterial inflow and runoff: initial Estimated risks of radiation-induced fatal
cancer from pediatric CT. AJR Am J Roentge- 1995; 197:316 –317.
experience. Radiology 2001; 221:146 –158.
nol 2001; 176:289 –296. 49. Saito Y, Suzuki T. Evaluation of the perfor-
13. Schoepf UJ, Bruening RD, Hong C, et al. Mul-
31. Nickoloff E, Alderson P. Radiation exposure mance of multidetector row CT system in
tislice helical CT of focal and diffuse lung
to patients from CT: reality, public percep- non-helical scanning (abstr). Radiology 1998;
disease: comprehensive diagnosis with re-
tion, and policy. AJR Am J Roentgenol 2001; 209(P):578.
construction of contiguous and high-resolu-
177:285–287. 50. Hsieh J. Analytical models for multi-slice he-
tion CT sections from a single thin-collima-
tion scan. AJR Am J Roentgenol 2001; 177: 32. McCollough C. Patient dose in cardiac com- lical CT performance parameters. Med Phys
179 –184. puted tomography. Herz 2003; 28:1– 6. 2003; 30:169 –178.
14. Villablanca JP, Lahan R, Hooshi P, et al. De- 33. Morin R, Gerber T, McCollough C. Radiation 51. Schaller S, Flohr T, Klingenbeck K, Krause J,
tection and characterization of very small dose in computed tomography of the heart. Fuchs T, Kalender WA. Spiral interpolation
cerebral aneurysms by using 2D and 3D he- Circulation 2003; 107:917–922. algorithm for multi-slice spiral CT. I. Theory.
lical CT angiography. AJNR Am J Neurora- 34. International Electrotechnical Commission IEEE Trans Med Imaging 2000; 19:822– 834.
diol 2002; 23:1187–1198. 60601–2– 44. Amendment 1: medical electri- 52. Taguchi K, Aradate H. Algorithm for image
15. Flohr T, Stierstorfer K, Bruder H, Simon J, cal equipment, part 2– 44: particular require- reconstruction in multi-slice helical CT. Med
Schaller S. New technical developments in ments for the safety of x-ray equipment for Phys 1998; 25:550 –561.
multislice CT. I. Approaching isotropic reso- computed tomography. Geneva, Switzer- 53. Flohr T, Klingenbeck-Regn K, Ohnesorge B,
lution with sub-mm 16-slice scanning. Rofo land: International Electrotechnical Com- Schaller S. Multislice scanning with the So-
Fortschr Geb Rontgenstr Neuen Bildgeb Ver- mission, 2002. matom Volume Zoom: an optimized design
fahr 2002; 174:839 – 845. 35. Donnelly LF, Emery KH, Brody AS, et al. Min- for volume scanning. In: Reiser MF, Taka-
16. Kachelriess M, Ulzheimer S, Kalender W. imizing radiation dose for pediatric body ap- hashi M, Modic M, Bruening R, eds. Multi-
ECG-correlated image reconstruction from plications of single-detector helical CT: strat- slice CT: medical radiology— diagnostic im-
subsecond multi-slice spiral CT scans of the egies at a large children’s hospital. AJR Am J aging and radiation oncology. Berlin,
heart. Med Phys 2000; 27:1881–1902. Roentgenol 2001; 176:303–306. Germany: Springer-Verlag, 2001.
17. Ohnesorge B, Flohr T, Becker C, et al. Cardiac 36. Frush DP, Soden B, Frush KS, Lowry C. Im- 54. Fuchs T, Krause J, Schaller S, Flohr T,
imaging by means of electrocardiographi- proved pediatric multidetector body CT us- Kalender WA. Spiral interpolation algo-
cally gated multisection spiral CT: initial ex- ing a size-based color-coded format. AJR rithms for multislice spiral CT. II. Measure-
perience. Radiology 2000; 217:564 –571. Am J Roentgenol 2002; 178:721–726. ment and evaluation of slice sensitivity pro-
cological patients: influence of table support 73. Kachelriess M, Kalender W. Electrocardio- raphy and multiplanar reconstruction. Cir-
and detector configuration on image quality gram-correlated image reconstruction from culation 2003; 107:664 – 666.
and radiation exposure. Rofo Fortschr Geb subsecond spiral computed tomography 88. Robb R, Ritman E. High speed synchronous
Rontgenstr Neuen Bildgeb Verfahr 2001; scans of the heart. Med Phys 1998; 25:2417– volume computed tomography of the heart.
173:373–378. 2431. Radiology 1979; 133:655– 661.
57. Feldkamp LA, Davis LC, Kress JW. Practical 74. Flohr T, Ohnesorge B. Heart-rate adaptive 89. Ritman E, Kinsey J, Robb R, Gilbert B, Harris
cone-beam algorithm. J Opt Soc Am A 1984; optimization of spatial and temporal resolu- L, Wood E. Three-dimensional imaging of
1:612– 619. tion for ECG-gated multi-slice spiral CT of heart, lungs, and circulation. Science 1980;
58. Grass M, Köhler T, Proksa R. 3D cone-beam the heart. J Comput Assist Tomogr 2001; 25: 210:273–280.
CT reconstruction for circular trajectories. 907–923.
90. Townsend DW, Cherry SR. Combining anat-
Phys Med Biol 2000; 45:329 –347. 75. Taguchi K, Anno H. High temporal resolu-
omy and function: the path of true image
59. Wang G, Lin T, Cheng P. A general cone- tion for multi-slice helical computed tomog-
fusion. Eur Radiol 2001; 11:1968 –1974.
beam reconstruction algorithm. IEEE Trans raphy. Med Phys 2000; 27:861– 872.
91. Namdar M, Kaufmann P, Hany T, von Schul-
Med Imaging 1993; 12:486 – 496. 76. Bruder H, Schaller S, Ohnesorge B, Mertel-
60. Schaller S. Practical image reconstruction for meier T. High temporal resolution volume thess G. Combined CT-angiogram and PET
cone-beam computed tomography (disserta- heart imaging with multirow computed to- perfusion imaging for assessment of CAD in
tion). Nürnberg, Germany: Erlangen Univer- mography. In: Hanson KM, ed. Proceedings a novel PET/CT: a pilot feasibility study (ab-
sity; 1998. of SPIE: medical imaging 1999 —image pro- str). Eur Radiol 2003; 13(suppl):165.
61. Turbell H, Danielsson PE. An improved PI- cessing. Vol 3661. Bellingham, Wash: Inter- 92. Ohnesorge B, Flohr T, Schwarz K, Heiken JP,
method for reconstruction from helical cone national Society for Optical Engineering, Bae KT. Efficient correction for CT image ar-
beam projections. Presented at the IEEE Med- 1999; 420 – 432. tifacts caused by objects extending outside
ical Imaging Conference, Seattle, Wash, Oc- 77. Pan T, Shen Y. New multi-sector reconstruc- the scan field of view. Med Phys 2000; 27:
tober 24 –30, 1999. tion for cardiac ct. Presented at IEEE Medical 39 – 46.
62. Proksa R, Koehler T, Grass M, Timmer J. The Imaging Conference, Lyon, France, October 93. Kaczirek K, Prager G, Kienast O, et al. Com-
n-PI method for helical cone-beam CT. IEEE 16 –20, 2000. bined transmission and (99m)Tc-sestamibi
Trans Med Imaging 2000; 19:848 – 863. 78. Cesmeli E, Edic M, Iatrou M, Pfoh A. A novel emission tomography for localization of me-
63. Larson G, Ruth C, Crawford C. Nutating slice reconstruction algorithm for multiphasic diastinal parathyroid glands. Nuklearmedi-
CT image reconstruction. U.S. patent appli- cardiac imaging using multislice CT. In: An- zin 2003; 42:220 –223.
cation WO 98/44847; filed April 8, 1998. tonuk LE, Yaffe MJ, eds. Proceedings of SPIE: 94. Horger M, Eschmann SM, Lengerke C, Claus-
64. Kachelriess M, Schaller S, Kalender WA. Ad- medical imaging 2001—physics of medical sen CD, Pfannenberg C, Bares R. Improved
vanced single-slice rebinning in cone-beam imaging. Vol 4320. Bellingham, Wash: Inter- detection of splenosis in patients with
spiral CT. Med Phys 2000; 27:754 –772. national Society for Optical Engineering; haematological disorders: the role of com-
65. Bruder H, Kachelriess M, Schaller S, Stierstor- 645– 654. bined transmission-emission tomography.
fer K, Flohr T. Single-slice rebinning recon- 79. Loubeyre P, Angelie E, Grozel F, Abidi H, Eur J Nucl Med Mol Imaging 2003; 30:316 –
struction in spiral cone-beam computed to- Minh VA. Spiral CT Artifact that simulates 319.
mography. IEEE Trans Med Imaging 2000; aortic dissection: image reconstruction with 95. Tai CJ, Shian YC, Wang IJ, Ho YJ, Ho ST, Kao
19:873– 887. use of 180° and 360° linear-interpolation al- CH. Detection of recurrent or residual naso-
66. Kohler T, Proksa R, Bontus C, Grass M, Tim- gorithms. Radiology 1997; 205:153–157. pharyngeal carcinomas after radiotherapy
mer J. Artifact analysis of approximate heli- 80. Schoepf U, Helmberger T, Holzknecht N, et with technetium-99m tetrofosmin single
cal cone-beam CT reconstruction algo- al. Segmental and subsegmental pulmonary photon emission computed tomography and
rithms. Med Phys 2002; 29:51– 64. arteries: evaluation with electron-beam CT comparison with computed tomography: a
67. Schaller S, Stierstorfer K, Bruder H, Kachel- versus spiral CT. Radiology 2000; 214:433– preliminary study. Cancer Invest 2003; 21:
riess M, Flohr T. Novel approximate ap- 439. 536 –541.
proach for high-quality image reconstruc- 81. Lell M, Wildberger J, Heuschmid M, et al. 96. Hofmann A, Zettinig G, Wachter S, Kurtaran
tion in helical cone beam CT at arbitrary CT-angiography of the carotid artery: first A, Kainberger F, Dudczak R. Imaging of aortic
pitch. In: Sonka M, Hanson KM, eds. Pro- results with a novel 16-slice-spiral-CT scan-
prosthesis infection with a combined
ceedings of SPIE: medical imaging 2001— ner. Rofo Fortschr Geb Rontgenstr Neuen
SPECT/CT device. Eur J Nucl Med Mol Imag-
image processing. Vol 4322. Bellingham, Bildgeb Verfahr 2002; 174:1165–1169.
ing 2002; 29:836.
Wash: International Society for Optical En- 82. Ertl-Wagner B, Hoffmann RT, Brüning R,
97. Garcia-Ramirez JL, Mutic S, Dempsey JF, Low
gineering, 2001; 113–127. Dichgans M, Reiser MF. Supraaortale gefäss-
68. Flohr T, Stierstorfer K, Bruder H, Simon J, DA, Purdy JA. Performance evaluation of an
diagnostik mit dem 16-zeilen-multidetektor-
Polacin A, Schaller S. Image reconstruction spiral-CT: untersuchungsprotokoll und erste 85-cm-bore x-ray computed tomography
and image quality evaluation for a 16-slice erfahrungen. Radiologe 2002; 42:728 –732. scanner designed for radiation oncology and
CT scanner. Med Phys 2003; 30:832– 845. 83. Schoepf UJ, Becker CR, Hofmann LK, et al. comparison with current diagnostic CT scan-
69. Hsieh J, Toth TL, Simoni P, Grekowicz B, Multislice CT angiography. Eur Radiol 2003; ners. Int J Radiat Oncol Biol Phys 2002; 52:
Slack CC, Seidenschnur GE. A generalized 13:1946 –1961. 1123–1131.
helical reconstruction algorithm for multide- 84. Wintersperger B, Helmberger T, Herzog P, et 98. Nikolaou K, Flohr T, Stierstorfer K, Becker
tector row CT (abstr). Radiology 2001; al. New abdominal CT angiography protocol CR, Reiser MF. Flat panel computed tomog-
221(P):217. on a 16 detector-row CT scanner. Radiologe raphy of human ex vivo heart and bone
70. Hsieh J, Grekowicz B, Simoni P, et al. Con- 2002; 42:722–727. specimens: initial experience. Eur Radiol
volution reconstruction algorithm for multi- 85. Wintersperger B, Herzog P, Jakobs T, Reiser 2005; 15:329 –333.
slice helical CT. In: Sonka M, Fitzpatrick MJ, M, Becker C. Initial experience with the 99. Knollmann F, Pfoh A. Image in cardiovascu-
eds. Proceedings of SPIE: medical imaging clinical use of a 16 detector row CT system. lar medicine: coronary artery imaging with
2003—image processing. Vol 5032. Belling- Crit Rev Comput Tomogr 2002; 43:283– flat-panel computed tomography. Circula-
ham, Wash: International Society for Optical 316. tion 2003; 107:1209.
Engineering, 2003; 716 –723. 86. Nieman K, Cademartiri F, Lemos PA, Raaij-
71. Stierstorfer K, Flohr T, Bruder H. Segmented makers R, Pattynama PM, de Feyter PJ. Reli-