Sunteți pe pagina 1din 18

Special Review

Radiology

Thomas G. Flohr, PhD Multi–Detector Row CT


Stefan Schaller, PhD
Karl Stierstorfer, PhD Systems and Image-
Herbert Bruder, PhD
Bernd M. Ohnesorge, PhD
U. Joseph Schoepf, MD
Reconstruction Techniques1
Published online before print
10.1148/radiol.2353040037 The introduction in 1998 of multi– detector row computed tomography (CT) by the
Radiology 2005; 235:756 –773 major CT vendors was a milestone with regard to increased scan speed, improved
z-axis spatial resolution, and better utilization of the available x-ray power. In this
Abbreviations:
AMPR ⫽ adaptive MPR review, the general technical principles of multi– detector row CT are reviewed as
ECG ⫽ electrocardiography they apply to the established four- and eight-section systems, the most recent
FWHM ⫽ full width at half maximum 16-section scanners, and future generations of multi– detector row CT systems.
MPR ⫽ multiplanar reformation Clinical examples are used to demonstrate both the potential and the limitations of
SSP ⫽ section-sensitivity profile
the different scanner types. When necessary, standard single-section CT is referred
3D ⫽ three-dimensional
to as a common basis and starting point for further developments. Another focus is
the increasingly important topic of patient radiation exposure, successful dose
1
From Siemens Medical Solutions, CT management, and strategies for dose reduction. Finally, the evolutionary steps from
Division, Forchheim, Germany (T.G.F.,
S.S., K.S., H.B., B.M.O.); Department traditional single-section spiral image-reconstruction algorithms to the most recent
of Diagnostic Radiology, Tübingen approaches toward multisection spiral reconstruction are traced.
University, Germany (T.G.F.); and De- © RSNA, 2005
partment of Radiology, Medical Uni-
versity of South Carolina, 169 Ashley
Ave, Charleston, SC 29425 (U.J.S.).
Received January 7, 2004; revision re- Supplemental material: radiology.rsnajnls.org/cgi/content/full/2353040037/DC1
quested March 9; revision received
April 26; accepted May 24. Address
correspondence to U.J.S. (e-mail:
schoepf@musc.edu).
U.J.S. is a medical consultant to Sie- Computed tomography (CT) was introduced in the early 1970s and has revolutionized the
mens Medical Solutions, CT Division, practice not only of diagnostic radiology but also of the whole field of medicine. CT was
Forchheim, Germany. the first technology to marry a computer to a medical imaging machine, the first to display
© RSNA, 2005 x-ray images as cross sections, and the first modality to herald a new era of digital imaging.
A glossary of terms used in this review is available online in Appendix E1 (radiology
.rsnajnls.org/cgi/content/full/2353040037/DC1).

EVOLUTION OF SPIRAL CT: FROM ONE SECTION TO 16

The introduction of spiral CT in the early 1990s constituted a fundamental evolutionary


step in the development and ongoing refinement of CT imaging techniques (1,2). For the
first time, volume data could be acquired without misregistration of anatomic detail.
Volume data became the basis for applications such as CT angiography (3), which has
revolutionized the noninvasive assessment of vascular disease. The ability to acquire
volume data also paved the way for the development of three-dimensional (3D) image-
processing techniques such as multiplanar reformation (MPR), maximum intensity pro-
jection, surface-shaded display, and volume-rendering techniques (4), which have become
a vital component of medical imaging today.
Ideally, volume data are of high spatial resolution and are isotropic in nature: Each
image data element (voxel) is of equal dimensions in all three spatial axes, and this forms
the basis for image display in arbitrarily oriented imaging planes. For most clinical
scenarios, however, single-section spiral CT with a 1-second gantry rotation is unable to
fulfill these requirements. To prevent motion artifacts and optimally utilize the contrast
agent bolus, body spiral CT examinations need to be completed within a certain time
frame of, ordinarily, one breath hold (25–30 seconds). If a large scan range such as the
entire thorax or abdomen (30 cm) has to be covered within a single breath hold, a thick
collimation of 5– 8 mm must be used. While the in-plane resolution of a CT image depends
on the system geometry and on the reconstruction kernel selected by the user, the

756
first medical CT scanners were two-sec- angiography of lower extremity vessels
ESSENTIALS tion systems, such as the EMI (England) (12), thicker (eg, 2.5-mm) collimated sec-
head scanner, introduced in 1972, and tions had to be chosen to complete the
● Multi– detector row CT allows substan- the Siemens Siretom (Erlangen, Ger- scan within a reasonable time frame.
tial reduction in examination time for many), introduced in 1974. With the ad- Scan times were often too long to allow
standard protocols, coverage of ex- vent of whole-body fan-beam CT systems image acquisition during a purely arterial
Radiology

tended anatomic volumes, and, most for general radiology, two-section acqui- phase. For CT angiography of the circle
important, substantially increased lon- sition was no longer used. Apart from a of Willis, for instance, a scan range of
gitudinal resolution by means of re- dedicated two-section system for cardiac about 100 mm must be covered (14).
duced section width. applications, the Imatron C-100 (Ima- With four-section CT at a collimated sec-
tron, San Francisco, Calif), which was in- tion width of 1 mm, pitch of 1.5, and
troduced in 1984, the first step toward gantry rotation time of 0.5 second, this
● Near-isotropic spatial resolution in rou-
multisection acquisition in general radi- volume can be covered in about 9 sec-
tine examinations, which has been
ology was a two-section CT scanner in- onds, not fast enough to avoid venous
achieved with 16-section CT systems, troduced in 1993 (Elscint TWIN; Elscint, overlay, assuming a cerebral circulation
enables 3D renderings of diagnostic Haifa, Israel) (6). In 1998, several CT time of less than 5 seconds. (Note: The
quality and oblique MPRs and maxi- manufacturers introduced multi– detec- definition of pitch for multi– detector
mum intensity projections with resolu- tor row CT systems, which provided con- row CT is discussed later in this review.)
tion similar to that of the transverse siderable improvement in scanning As a next step, the introduction of an
images. speed and longitudinal resolution and eight– detector row CT system in 2000
better utilization of the available x-ray enabled shorter scan times but did not
● Scanning at narrow collimation does power (7–10). These systems typically of- yet provide improved longitudinal reso-
fered simultaneous acquisition of four lution (thinnest collimation, eight sec-
not markedly increase the radiation
sections at a gantry rotation time of 0.5 tions at 1.25 mm). The latter was
dose to the patient, as long as the ef-
second. achieved with the introduction of 16 –
fective milliampere-seconds level is kept Simultaneous acquisition of m sections detector row CT (15), which made possi-
constant. results in an m-fold increase in speed if all ble the routine acquisition of substantial
other parameters (eg, section thickness) anatomic volumes with isotropic submil-
● A key challenge for image reconstruc- are unchanged. This increased perfor- limeter spatial resolution and scan times
tion with multi– detector row CT is the mance of multi– detector row CT relative of less than 10 seconds for 300 mm of
cone angle of the measurement rays; to single-section CT allowed the optimi- coverage (Fig 1). While in-plane spatial
this requires novel reconstruction tech- zation of a variety of clinical protocols. resolution is not substantially improved,
niques such as 3D back projection, The examination time for standard pro- the two major advantages of fast multi–
AMPR, or weighted hyperplane recon- tocols could be substantially reduced, detector row CT are a true isotropic
which proved to be of immediate clinical through-plane resolution and a short ac-
struction.
benefit for the quick and comprehensive quisition time that enable high-quality
assessment of trauma patients and unco- examinations in severely debilitated and
● Z filtering makes it possible to recon- operative patients (11). Alternatively, the severely dyspneic patients (Fig 1).
struct images retrospectively with dif- scan range that could be covered within a Traditional CT applications have been
ferent section widths from the same certain time was extended by a factor of enhanced and strengthened by the remark-
raw CT data set, trading off, in this m, which is relevant for oncologic stag- able, although incremental, improvement
way, z-axis resolution and image noise. ing or for CT angiography with extended in scanner performance by the addition
coverage (eg, the lower extremities) (12). of more detector rows. Multi–detector row
The most important clinical benefit, CT also dramatically expanded into areas
however, proved to be the ability to scan previously considered beyond the scope
a given anatomic volume within a given of third-generation CT scanners that
longitudinal (z-axis) resolution along the scan time with substantially reduced sec- were based on the mechanical rotation of
patient axis is determined by the col- tion width at m times increased longitu- an x-ray tube and detectors, such as car-
limated section width and the spiral dinal resolution. Because of this, the goal diac imaging with the addition of elec-
interpolation algorithm. Use of a thick of isotropic resolution was within reach trocardiographic (ECG)-gating capabil-
collimation of 5– 8 mm results in a con- for many clinical applications. Examina- ity. With a gantry rotation time of 0.5
siderable mismatch between the longitu- tions of the entire thorax (13) or abdo- second and dedicated image-reconstruc-
dinal resolution and the in-plane resolu- men could now be routinely performed tion approaches, the temporal resolution
tion, which is 0.5– 0.7 mm, depending on with a 1.0- or 1.25-mm collimated section for acquisition of an image was improved
the reconstruction kernel. Thus, with sin- width. Despite these promising advances, to 250 msec and less (16,17), which
gle-section spiral CT, the ideal of isotro- clinical challenges and limitations re- proved to be sufficient for motion-free
pic resolution can only be achieved for mained for four-section CT systems. True imaging of the heart in the mid- to end-
very limited scan ranges (5). isotropic resolution for routine applica- diastolic phase when the patient had a
Strategies to achieve more substantial tions had not yet been achieved, because slow to moderate heart rate (ie, up to 65
volume coverage with improved longitu- the longitudinal resolution of about 1 beats per minute [18]). With four simul-
dinal resolution include the simulta- mm does not fully match the in-plane taneously acquired sections, coverage of
neous acquisition of more than one sec- resolution of about 0.5– 0.7 mm in a rou- the entire heart volume with thin sec-
tion at a time and a reduction in the tine examination of the chest or abdo- tions (ie, four sections at 1.0- or 1.25-mm
gantry rotation time. Interestingly, the men. For large volumes, such as for CT collimation) within a single breath hold

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 757


proved temporal resolution and robust-
ness of use, 16-section submillimeter ac-
quisition for increased longitudinal
resolution and shorter breath-hold times,
and novel sophisticated approaches for
Radiology

image acquisition and reconstruction.


In this review, ECG-synchronized ex-
aminations of the heart and of the car-
diothoracic anatomy will be very suc-
cinctly discussed, since this topic has
been extensively reviewed elsewhere
(28). Similarly, advanced 3D postprocess-
ing techniques are omitted. In this arti-
cle, we will review the general technical
principles of multi– detector row CT as
they apply to the established four– and
eight– detector row systems, the more re-
cent 16 – detector row scanners, and gen-
erations of CT systems yet to come. On
the basis of the technologic description
of different scanner types and image-re-
Figure 1. Transverse sections (top) and coronal MPRs (bottom) from a thoracic examination construction approaches, we provide
illustrate clinical performance of CT. Left: single-section 8-mm-thick images. Middle: four-section practical “take-home points” to enable
1.25-mm-thick images. Right: 16 – detector row 0.75-mm-thick images. Differences in diagnostic better translation into daily clinical prac-
image quality are most obvious in the MPRs. With 16 – detector row images, the goal of isotropic tice of the technology and science re-
resolution in routine examinations has been reached. Single- and four-section images were
viewed here. Useful up-to-date informa-
synthesized from the 16-section CT data.
tion regarding multi– detector row CT is
also readily available on the Internet at,
for example, the UK Medicines and
Healthcare products Regulatory Agency
CT Web site (www.medical-devices.gov.uk)
or the Advanced Medical Imaging Labo-
ratory site (www.ctisus.org).

CURRENT TECHNIQUES
System Design
Detector design.—For clinical purposes,
different section widths must be avail-
able to adjust the optimum scan speed,
Figure 2. Illustration shows prepatient collimation of the x-ray longitudinal resolution, and image noise
beam to obtain different collimated section widths with a single– for each application. With a single– de-
detector row CT detector. FOV ⫽ field of view. tector row CT scanner, different colli-
mated section widths are obtained by
means of prepatient collimation of the
became feasible. This 1.0 –1.25-mm lon- lemma, mainly because of partial volume x-ray beam (Fig 2). For a very elementary
gitudinal resolution combined with the artifacts as a consequence of insufficient model of a two-section CT scanner (m ⫽
improved contrast resolution of modern longitudinal resolution (22). For patients 2, or two detector rows), Figure 2 demon-
CT systems enabled noninvasive depic- with a higher heart rate, careful selection strates how different section widths can
tion of the coronary arteries (19 –22). Ini- of separate reconstruction intervals for be obtained by means of prepatient col-
tial clinical studies demonstrated the po- different coronary arteries has been man- limation if the detector is separated mid-
tential of multi– detector row CT to not datory (25). It is almost impossible for way along the z-axis extent of the x-ray
only demonstrate but to some degree patients with manifest heart disease to beam. For m ⬎ 2, this simple design prin-
also characterize noncalcified and calci- comply with the breath-hold time of ciple must be replaced by more flexible
fied plaques in the coronary arteries on about 40 seconds required to cover the concepts requiring more than m detector
the basis of plaque CT attenuation entire heart volume (approximately 12 rows to simultaneously acquire m sec-
(22,23). cm) with four-section CT. The ongoing tions.
The limitations of four– and eight– de- technical refinement of multi– detector Different manufacturers of multi– de-
tector row CT systems, however, have so row CT, however, holds the promise of tector row CT scanners have introduced
far prevented the successful integration gradually overcoming some of these lim- different detector designs. In order to be
of CT coronary angiography into routine itations. The most important steps to- able to select different section widths, all
clinical algorithms: Stents or severely cal- ward this goal are gantry rotation times scanners combine several detector rows
cified arteries constitute a diagnostic di- faster than 0.5 second (26,27) for im- electronically to a smaller number of sec-

758 䡠 Radiology 䡠 June 2005 Flohr et al


tions according to the selected beam col- mens), uses 24 detector rows (15). The 16
limation and the desired section width. central rows define 0.75-mm collimated
For established four-section CT sys- section widths at the isocenter, and the
tems, two detector types are commonly four outer rows on both sides define
used. The fixed-array detector consists of 1.5-mm collimated section widths (Fig
detector elements with equal sizes in the 3c). The total coverage in the longitudi-
Radiology

longitudinal direction. A representative nal direction is 24 mm at the isocenter.


example of this scanner type, the Light- By means of appropriate combination of
speed scanner (GE Medical Systems, Mil- the signals of the individual detector
waukee, Wis), has 16 detector rows, each rows, either 12 or 16 sections with 0.75-
of them defining a 1.25-mm collimated or 1.5-mm collimated section width can
section width in the center of rotation be acquired simultaneously. The Light-
(8,10,29). The total coverage in the lon- speed 16 scanner (GE Medical Systems)
gitudinal direction is 20 mm at the iso- uses a similar design: It provides 16 sec-
center; owing to geometric magnifica- tions with either 0.625- or 1.25-mm col-
tion, the actual detector is about twice as limated section width. The total coverage
wide. By means of prepatient collimation in the longitudinal direction is 20 mm at
and combination of the signals of the the isocenter. Yet another design, which
individual detector rows, the following is implemented in the Aquilion scanner
section widths (measured at the iso- (Toshiba, Tokyo, Japan), can provide 16
center) can be realized: four sections at sections with either 0.5-, 1.0-, or 2.0-mm
1.25 mm, 2.5 mm, 3.75 mm, and 5.0 mm collimated section width, with a total
(Fig 3a). The same detector design is used coverage of 32 mm at the isocenter.
for the eight-section version of this sys-
tem and provides eight sections at 1.25-
Radiation Dose
and 2.5-mm collimated section widths.
A different approach uses an adaptive- Radiation dose and dose efficiency.—Ra-
array detector design, which comprises diation exposure to the patient at CT and
detector rows with different sizes in the the resulting potential radiation hazard
longitudinal direction. Scanners of this have recently gained considerable atten-
type, the Mx8000 four-section scanner tion in both the public and the scientific
(Philips Medical Systems, Best, the Neth- literature (30,31). Typical values for the
erlands) and the Somatom Sensation 4 effective patient dose for selected CT pro-
scanner (Siemens), have eight detector tocols are 1–2 mSv for a head CT, 5–7
rows (7,9). Their widths in the longitudi- mSv for a chest CT, and 8 –11 mSv for
nal direction range from 1 to 5 mm at the abdominal and pelvic CT (32,33). This
isocenter and allow the following colli- radiation exposure must be appreciated
mated section widths: two sections at 0.5 in the context of the average annual
mm, four at 1.0 mm, four at 2.5 mm, four background radiation, which is 2–5 mSv
at 5.0 mm, two at 8.0 mm, and two at (3.6 mSv in the United States). Despite
10.0 mm (Fig 3b). the undisputed clinical benefits, multi-
The selection of the collimated section section CT scanning is often considered
width determines the intrinsic longitudi- to require increased patient dose com- Figure 3. Illustrations show examples of
nal resolution of a scan. In a “step-and- pared with the dose from single-section (a) fixed-array and (b, c) adaptive-array detec-
shoot” sequential mode, any multiple of CT. Indeed, a certain increase in radia- tors used in commercially available four- and
the collimated width of one detector sec- tion dose is unavoidable owing to the 16-section CT systems.
tion can be obtained by adding the de- underlying physical principles.
tector signals during image reconstruc- In the x-ray tube of a CT scanner, a
tion. In a spiral mode, the effective small area on the anode plate, the focal- With single-section CT, the entire trap-
section width, which is usually defined as spot, emits x-rays that penetrate the pa- ezoidal dose profile can contribute to the
the full width at half maximum (FWHM) tient and are registered by the detector. A detector signal, and the collimated sec-
of the spiral section-sensitivity profile collimator between the x-ray tube and tion width is determined as the FWHM of
(SSP), is adjusted independently in the the patient, the prepatient collimator, is this trapezoid. The relative dose utiliza-
spiral interpolation process during image used to shape the beam and to establish tion of a single-section CT system can
reconstruction. Hence, from the same the dose profile. In general, the colli- therefore be close to 100%. In most cases
data set, both narrow sections for high- mated dose profile is a trapezoid in the with multi– detector row CT, only the
spatial-resolution detail or for 3D post- longitudinal direction. In the umbral re- plateau region of the dose profile is used
processing and wide sections for better gion (ie, plateau region of the trapezoid), to ensure an equal signal level for all de-
contrast resolution or quick review and x-rays emitted from the entire area of the tector elements. The penumbral region is
filming may be derived. focal spot illuminate the detector. In the then discarded, either by a postpatient
Sixteen-section CT systems usually penumbral regions, only a part of the collimator or by the intrinsic self-colli-
have adaptive-array detectors. A repre- focal spot illuminates the detector, while mation of the multisection detector, and
sentative example for this scanner type, the prepatient collimator blocks off other represents “wasted” dose. The relative
the Somatom Sensation 16 scanner (Sie- parts. contribution of the penumbral region in-

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 759


ing. Dose reduction can be achieved by
reductions in the milliampere-seconds
and voltage settings. Most CT manufac-
turers provide dedicated pediatric proto-
cols with, for example, milliampere-sec-
Radiology

onds and voltage settings adjusted


according to the weight of the child.
Another means to reduce radiation
dose is to adapt the x-ray tube voltage to
the intended application. In contrast
Figure 4. Dose profiles for four- and 16-section CT systems with
identical collimated width of one detector (Det.) section. The relative
agent– enhanced studies such as CT an-
contribution of the penumbral region, which represents wasted dose, giography, the contrast-to-noise ratio for
decreases with increasing number of simultaneously acquired sec- fixed patient dose increases with decreas-
tions. ing x-ray tube voltage. As a consequence,
to obtain the desired contrast-to-noise ra-
tio, the patient dose can be reduced by
pending on the scanner type. With 16- choosing a lower voltage setting. The po-
section CT systems and submillimeter tential for dose saving is more substantial
collimation, dose utilization can be im- for patients with a smaller diameter. This
proved to 84%, again depending on scan- can be demonstrated, for example, by
ner type (25). Some multi– detector row means of phantom measurements of
CT systems offer special implementa- small tubes filled with diluted contrast
tions of even more dose-efficient modes agent embedded in acrylic plastic phan-
that use a portion of the penumbral re- toms with different diameters (38). The
gion. iodine contrast-to-noise ratio at constant
A clinically appropriate measure for radiation dose for various voltage settings
dose is the weighted CT dose index, or is shown in Figure E2 (radiology.rsnajnls
CTDIw (34), which uses the absorbed .org/cgi/content/full/2353040037/DC1) as
dose in a polymerized methyl methacry- a function of the phantom diameter.
late (acrylic plastic) phantom as an ap- Compared with a standard scan at 120 kV
proximation of the dose delivered to a in a 32-cm-diameter phantom (corre-
cross section of the patient’s anatomy sponding to that for an average adult),
(see Appendix E2, radiology.rsnajnls.org/cgi the same contrast-to-noise ratio is ob-
/content/full/2353040037/DC1). Figure E1 tained with 0.49 times the dose (1.3
(radiology.rsnajnls.org/cgi/content/full times the milliampere-seconds setting)
/2353040037DC1) shows CTDIw at 120 kV for 80 kV and 0.69 times the dose (1.1
for the 32 cm body phantom as a func- times the milliampere-seconds) for 100
tion of the total collimated width of the kV. Thus, ideally, 80 kV should be used
detector for a four-section CT system and for CT angiography in order to reduce
a 16-section CT system with a similar sys- patient dose.
tem geometry. The CTDIw for 16-section Clinical studies (38) have confirmed
CT at 0.75-mm collimation is 7.8 mGy/ these findings and demonstrated a po-
100 mAs, whereas the CTDIw for four- tential for dose reduction of about 50%
Figure 5. Transverse (top) and coronal max- section CT at 1.0-mm collimation is 9 when 80 kV is used for CT angiography
imum intensity projection (bottom; 5-mm slab mGy/100 mAs. Thus, different from the instead of 120 kV. In reality, however,
thickness) thoracic CT images in a patient with transition from single-section CT to the maximum x-ray tube current avail-
pulmonary embolism. Scans were acquired 4-section CT systems, a further increase able at 80 kV is generally not sufficient to
with 16-section scanner at 100 kV and 120
in radiation exposure with the more scan bigger patients, which limits the
mAs. Effective patient dose was 2.3 mSv, 25%
less than for the standard 120 kV protocol. widespread availability of 16-section CT routine application of this approach.
(Images courtesy of Peter Herzog, MD, Klini- systems is not to be expected. Therefore, use of 100 kV appears to be a
kum Grosshadern, Munich, Germany.) Concepts for radiation dose reduction.— suitable compromise and the method of
The most important factor for reducing choice for CT angiography. Figure 5
radiation exposure is an adaptation of shows pulmonary CT angiographic im-
creases with decreasing section width, the dose to the patient’s size and weight ages of a patient with pulmonary embo-
and it decreases with increasing number (35–37). lism; the scan was performed on a 16-
of simultaneously acquired images. This As a general rule for the practicing ra- section scanner at 100 kV and 120 mAs,
is demonstrated in Figure 4, which com- diologist, the dose necessary to maintain and the effective patient dose for this
pares the “minimum width” dose profiles constant image noise has to be doubled if scan was 2.3 mSv, 25% less than that for
for a four-section CT system and a corre- the patient diameter is increased by 4 cm. the standard 120-kV protocol. Authors of
sponding 16-section CT system with Correspondingly, for a patient diameter recent study (39) recommended 100 kV
equal collimated width of one detector that is 4 cm smaller than average, half as the standard mode for thoracic and
section. Correspondingly, the relative the standard dose is sufficient to main- abdominal CT angiography and report
dose utilization with four-section 1-mm- tain adequate image quality. This is of dose savings of 30% without loss of diag-
collimation CT is 70% or less (10), de- particular importance in pediatric imag- nostic information.

760 䡠 Radiology 䡠 June 2005 Flohr et al


An approach that is finding increased
implementation in clinical practice is an-
atomic tube current modulation. With
this technique, tube output is adapted to
the patient geometry during each rota-
tion of the scanner to compensate for
Radiology

strongly varying x-ray attenuation in


asymmetric body regions such as the
shoulders and pelvis. The variation of the
tube output is either predefined by
means of an analysis of a localizer scan
(topogram, scout view) or is determined
online by evaluating the signal from a
detector row. With this technique, dose
can be reduced by 15%–35% without de-
grading image quality, depending on the
body region (40,41). Figure 6. Automatic exposure control. Lateral topogram (scout
view) for thoracoabdominal CT in a 6-year-old child is shown with
In more sophisticated approaches,
automatically adapted milliampere-seconds value as function (curve)
tube output is modified according to the of z-axis position during spiral CT. Although the standard adult
patient geometry not only during each protocol with 165 mAs was used, the average milliampere-seconds
rotation but also in the longitudinal di- value throughout the scan was adjusted to 38 mAs owing to auto-
rection (automatic exposure control), to matic exposure control. (Image courtesy of Michael Lell, MD, Univer-
maintain adequate dose when moving to sity of Erlangen, Germany.)
different body regions (eg, from thorax to
abdomen). In one implementation, the
attenuation for each body region of a the spiral scan, the output of the x-ray tion width can be increased. As an ex-
“standard-sized” patient is stored in the tube is modulated according to the pa- ample, a scan with four sections at
control computer. This attenuation cor- tient’s ECG trace. It is kept at its nominal 1.0-mm collimation provides either four
responds to the milliampere-seconds set- value during a user-defined phase of the images with 1.0-mm section width, two
ting of the standard protocol. If the ac- cardiac cycle—in general, the mid- to images with 2.0-mm section width, or
tual attenuation of the patient deviates end-diastolic phase. During the rest of one image with 4.0-mm section width.
from the “standard” attenuation, the the cardiac cycle, the tube output is typ- The option to realize a wider section by
tube output is adapted correspondingly. ically reduced to 20% of the nominal val- summing several thin sections is benefi-
Figure 6 shows the variation of the milli- ues, although it is not switched off en- cial for examinations that require narrow
ampere-seconds output for a CT scan of tirely, to allow image reconstruction collimation to prevent partial volume ar-
the chest and abdomen in a 6-year-old throughout the entire cardiac cycle. tifacts and low image noise to allow de-
child. Although the standard protocol Thus, although the signal-to-noise ratio tection of low-contrast details (eg, neuro-
with 165 mAs was used—which would is decreased at certain phases of the car- logic examinations of posterior fossa or
have resulted in substantially higher ra- diac cycle, the low-dose images are still cervical spine). In the head, partial vol-
diation dose than necessary in a standard sufficient for evaluation of functional pa- ume artifacts typically manifest as dark
mode of operation—the average milliam- rameters such as ejection fraction, should streaks or areas of hypoattenuation and
pere-seconds value throughout the scan this kind of information be desired. are due to a nonlinear effect that has
was adjusted to 38 mAs by means of au- been described in reference 45. Figure 7
tomatic exposure control. Automatic ad- SEQUENTIAL SCANS AND shows an example of a patient who un-
aptation of tube current to patient size IMAGE-RECONSTRUCTION derwent follow-up CT after surgical re-
prevents both over- and underirradia- TECHNIQUES moval of a pituitary tumor. From the
tion, considerably simplifies the clinical same scan data—four sections at 1.0-mm
workflow for the technician, and elimi- With the advent of multi– detector row collimation— both 4.0-mm-thick images
nates the need for look-up tables of pa- CT, sequential “step-and-shoot” scan- with a standard head kernel for soft-tis-
tient weight and size for adjusting the ning has remained in use for only a few sue evaluation and 1.0-mm-thick images
milliampere-seconds settings. clinical applications, such as head CT, with a bone kernel were reconstructed.
Radiation dose for ECG-synchronized high-spatial-resolution lung CT, perfu- For best image quality, the posterior fossa
CT for cardiac applications has been a sion CT, and interventional applications. should be scanned with a collimated sec-
topic of considerable controversy. Recent A detailed theoretical description to pre- tion width not larger than 1.25 mm,
studies (32,33) based on four-section CT dict the performance of multi– detector whereas wider collimation can be used in
systems find an effective patient dose of row CT in sequential mode can be found the supratentorial region (46).
roughly 1 mSv for ECG-triggered calcium in reference 44.
scoring with 3-mm section width and The number of images acquired during SPIRAL SCANS AND
roughly 10 mSv for ECG-gated CT an- a sequential scan corresponds to the IMAGE-RECONSTRUCTION
giography of the coronary arteries with number of active detector sections. By TECHNIQUES
1.0- or 1.25-mm section width. Radiation adding the detector signals of the indi-
dose in ECG-gated spiral CT can be re- vidual sections during image recon- Spiral scanning is the method of choice
duced by 30%–50% with use of ECG-con- struction, the number of images per for the majority of all multi– detector row
trolled dose modulation (42,43). During scan can be reduced, and the image sec- CT examinations and requires more at-

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 761


the x-ray tube and detector have ex-
changed positions. This is the so-called
complementary ray. In spiral CT, z-axis
resolution is determined not only by the
collimated beam width (as in sequential
Radiology

scanning) but also by the effective sec-


tion width, which is established in the
spiral interpolation process. Usually, the
effective section width is defined as the
FWHM of the SSP. Effective section
width increases with increasing pitch for
both 360° and 180° linear interpolation,
and longitudinal resolution degrades (Fig
E3, radiology.rsnajnls.org/cgi/content/full/
Figure 7. Clinical performance of four-section CT in sequential scan 2353040037/DC1). This is a consequence
mode. Follow-up images in a patient after surgical removal of pitu-
of the increasing longitudinal distance of
itary tumor. Left: 4-mm-thick image with standard head kernel for
soft-tissue evaluation. Right: 1-mm-thick image with bone kernel for the projections used for spiral interpola-
bone evaluation. Both images were generated from the same scan tion. With 180° linear interpolation, the
data (four sections at 1-mm collimation). effective sections width equals the colli-
mated section width at a pitch of 1, but
effective section width equals 1.27 times
nition holds for both single-section and the collimated width at a pitch of 2, so
multi– detector row CT. It shows whether that a collimated 5-mm-thick section is
data acquisition occurs with gaps (p ⬎ 1) an actual 6.4-mm-thick section at a pitch
or with overlap (p ⬍ 1) in the longitudi- of 2. The image noise in single-section
nal direction. With 16 sections at spiral CT is independent of the pitch
0.75-mm collimation and a table-feed of if the tube current (in milliamperes) is
18 mm per rotation, the pitch is p ⫽ left unchanged, and patient dose de-
18/(16 ⫻ 0.75) ⫽ 18/12 ⫽ 1.5. With four creases with increasing pitch (see Appen-
sections at 1.0-mm collimation and a ta- dix E2, radiology.rsnajnls.org/cgi/content/
ble-feed of 6 mm per rotation, the pitch full/2353040037/DC1).
again is p ⫽ 6/(4 ⫻ 1) ⫽ 6/4 ⫽ 1.5. In the Single-section spiral CT is based almost
early days of four-section CT, the term exclusively on 180° linear interpolation,
detector pitch had been additionally intro- owing to the narrower SSP of this algo-
duced, which accounts for the width of a rithm, despite its increased susceptibility
single section in the denominator. For to artifacts and increased image noise.
the sake of clarity and uniformity, the For the same milliampere-seconds set-
detector pitch should no longer be used. ting, image noise is about 15% higher
Figure 8. Transverse sections of anthropo-
morphic thorax phantom from 16-section CT than that in sequential CT mode. Spiral
at 0.75-mm collimation and pitch of 1. Images Short Review of Single-Section artifacts gradually increase as pitch is in-
were reconstructed with adaptive multiplanar Spiral CT Reconstruction creased. Spiral artifacts typically manifest
reconstruction (AMPR; Siemens), with 1.0-mm as hyper- or hypoattenuating “windmill”
(top) and 3.0-mm (bottom) section width. Spi- Spiral CT requires an interpolation of structures surrounding z-axis inhomoge-
ral interpolation artifacts are reduced with sec- the acquired measurement data in the
tions widths that are thick relative to collima- neous high-contrast objects (eg, bones),
longitudinal (through-plane) direction to which rotate when scrolling through a
tion. In clinical practice, best image quality for
estimate a complete CT data set at the
a desired section width is obtained by acquir- stack of images. Spiral artifacts are caused
ing narrow-collimation data. desired plane of reconstruction. The
by the spiral interpolation process and
most commonly used single-section spi-
can also be observed on multi– detector
ral interpolation schemes are the 360°
row CT images (see Fig 8). With single-
and 180° linear interpolation methods.
section CT, scanning at a higher pitch is
tention than does sequential multi– de- The 360° linear interpolation method
often used to reduce patient dose at the
tector row CT because it is conceptually exploits the 360° periodicity of the pro-
jection data (1,2). For each projection an- expense of section broadening—if the
more demanding.
gle, a linear interpolation is performed collimation is kept constant—and in-
between those two projections on either creased spiral artifacts. For CT angio-
Definition of Spiral Pitch graphic applications in particular, it is
side of the image plane that are posi-
An important parameter for character- tioned closest to the image plane and are more favorable to scan a given volume in
izing a spiral CT scan is the pitch. Accord- 360° apart (ie, are measured in subse- a given time by using narrow collimation
ing to International Electrotechnical Com- quent rotations). The 180° linear interpo- at a high pitch rather than wider collima-
mission specifications (34), the pitch (p) is lation technique makes use of the fact tion at a low pitch. The motivation for
given by p ⫽ TF/W, where TF is the table that for each measurement ray, an inter- increasing pitch and reducing collima-
feed per rotation, and W is the total polation partner is already available after tion is to improve longitudinal resolu-
width of the collimated beam. This defi- approximately half a rotation (47), when tion by narrowing the SSP (48).

762 䡠 Radiology 䡠 June 2005 Flohr et al


The Cone-Angle Problem in
Multi–Detector Row CT
Two-dimensional image-reconstruction
approaches used in commercially avail-
able single-section CT scanners require
Radiology

all measurement rays that contribute to


an image to run in a plane perpendicular
to the patient’s longitudinal axis. In mul-
ti– detector row CT systems, this require-
ment is violated. Figure 9 shows the ge-
ometry of a four-section scanner: The
measurement rays are tilted by the so-
called cone angle with respect to the cen- Figure 9. Diagram shows geometry of four-section CT scanner dem-
ter plane. The cone angle is largest for the onstrating the cone-angle problem: Measurement rays are tilted by
sections at the outer edges of the detec- the so-called cone angle with respect to the center plane. Left and
right: Two view angles from sequential scan that are shifted by 180°
tor, and it increases as the number of
so that positions of x-ray tube and detector are interchanged. With
detector rows increases, if their width is single-section CT, identical measurement values would be acquired.
kept constant. As a first approximation, With multi– detector row CT, different measurement values are ac-
the cone angle is neglected in multi– de- quired. SFOV ⫽ scan field of view.
tector row CT reconstruction approaches:
The measurement rays are treated as if
they traveled perpendicular to the z-axis,
and modified two-dimensional image-re- the isocenter, direct and complementary proximately 1.27 times the collimated
construction algorithms are used. The rays interleave in the z-axis direction for section width, and a 1.25-mm collimated
data are then inconsistent, however, and selected pitch values. This way, the dis- section width results in a 1.5–1.6-mm ef-
produce cone-beam artifacts at high-con- tance between measured samples is sub- fective section width.
trast objects such as bones. It has been stantially reduced and equals half the When comparing dose and image noise
demonstrated that cone-beam artifacts collimated section width, which results for different pitch values, the widening
can be tolerated if the maximum number in the desired narrow SSPs. Appropriate of the SSP has to be taken into account. To
of simultaneously acquired sections does pitch values are 0.75 for four-section obtain the same image noise as in a se-
not markedly exceed four (49). As a con- scanning (29) and 0.5625 or 0.9375 for quential scan with the same collimated
sequence, the image-reconstruction ap- 16-section scanning (50). The 180° and section width, 0.73–1.68 times the dose
proaches of all commercially available 360° multidetector linear interpolation (depending on spiral pitch) is required,
four-section CT systems and of some sys- approaches are schematically illustrated with the lowest dose at the highest pitch
tems with even more sections neglect the in Figure E4 (radiology.rsnajnls.org/cgi/content (see reference 50). Some manufacturers
cone angle of the measurement rays. /full/2353040037/DC1) for the example of a provide a semiautomatic adaptation of the
four-section CT scanner. milliampere value to keep the image noise
In general, scanners that rely on 180° constant if the pitch is changed. In clinical
MULTI–DETECTOR ROW SPIRAL or 360° multidetector linear interpola- practice, therefore, it is permissible to as-
CT RECONSTRUCTION tion techniques and extensions thereof sume that scanners offering discrete op-
APPROACHES THAT NEGLECT provide selected discrete pitch values to timized pitch values based on 180° and
CONE-BEAM GEOMETRY the user, such as 0.75 and 1.5 for four- 360° multidetector linear interpolation
section scanning (29) or 0.5625, 0.9375, techniques are comparable to single-sec-
Multi–Detector Row 180° and 360°
1.375, and 1.75 for 16-section scanning tion CT systems in some core aspects: At
Linear Interpolation
(50). These pitch values are intended to high pitch, the section widens and the
The 360° and 180° linear interpolation provide optimized sampling schemes in longitudinal resolution degrades; at low
single-section spiral reconstruction ap- the longitudinal direction and, hence, pitch, the narrowest possible SSP (compa-
proaches can be extended to multi– de- optimized image quality. rable to that of 180° single-section linear
tector row spiral scanning in a straight- The user has to be aware of pitch-de- interpolation at pitch of 1) can be ob-
forward way (29,50,51). Both 360° and pendent effective section widths. For tained, but a higher dose is necessary to
180° multidetector linear interpolation low-pitch scanning (pitch of 0.75 for four maintain the signal-to-noise ratio. Thus,
methods are characterized by a projec- sections and 0.5625 or 0.9375 for 16 sec- as a take-home point, when one selects
tion-wise linear interpolation between tions), the effective section width ap- the scan protocol for a particular applica-
two rays on either side of the image proximates the collimated section width; tion, scanning at low pitch optimizes im-
plane. The cone angle of the measure- for a 1.25-mm collimated section width, age quality and longitudinal resolution
ment rays is not taken into account. In the resulting effective section width re- at a given collimation but at the expense
the 360° linear interpolation spiral recon- mains 1.25 mm. The narrow SSP, how- of increased patient dose. To reduce pa-
struction approach, rays measured either ever, is achieved by using 180° multide- tient dose, either milliampere settings
at the same projection angle by different tector linear interpolation reconstruction should be reduced at low pitch values or
detector rows or in consecutive rotations with conjugate interpolation at the price high pitch values should be chosen.
of the scanner (ie, 360° apart) are used for of increased image noise (29,50). For
spiral interpolation. In the 180° spiral re- high-pitch scanning (pitch of 1.5 for four Z-Filter Approaches
construction approach, both direct and sections and 1.375 or 1.75 for 16 sec- In a z-filter multi– detector row spiral
complementary rays are considered. At tions), the effective section width is ap- reconstruction (51,52), the spiral interpo-

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 763


the image plane. Instead, all direct and spectively with different section widths
complementary rays within a selectable from the same raw CT data set. Only sec-
distance from the image plane contribute tion widths equal to or larger than the
to the image. The weighting function for section width of one active detector row
the rays is selectable, which allows one to can be obtained. In many cases, both
Radiology

adjust both the functional form and the thick sections for initial viewing and re-
FWHM of the spiral SSP. Still, the cone cording and thin sections for detailed di-
angle is neglected. A representative ex- agnosis or as an input for advanced 3D
ample of a z-filter approach is the adap- postprocessing are routinely reconstructed.
tive axial interpolation algorithm (51) The thinnest available section width is
implemented in Siemens CT scanners, the collimated section width (1.0 mm for
which is illustrated in Figure E5 (radiology four sections at 1.0-mm collimation),
.rsnajnls.org/cgi/content/full/2353040037 which is created by using nonlinear spiral
/DC1). Another example is the “multislice weighting functions at the expense of in-
cone-beam tomography,” or MUSCOT, al- creased image noise and increased sus-
gorithm (52) used by Toshiba. Z filtering ceptibility to artifacts. Thus, as a take-
allows the system to trade off z-axis res- home point, the thinnest available
olution (the SSP) with image noise section should only be used for high-con-
(which directly correlates with required trast applications such as high-spatial-
dose). resolution lung imaging. For general pur-
With adaptive axial interpolation, the pose scanning, a 1.25-mm section width
spiral pitch is freely selectable in the for four-section CT at 1.0-mm collima-
range 0.5–2.0, and the same effective sec- tion (and 3.0-mm section width for four
tion width, which is defined as the sections at 2.5-mm collimation) is recom-
FWHM of the spiral SSP, is generated at mended as the most suitable trade-off
all pitch values (7,51,53). Therefore, lon- between longitudinal resolution, image
gitudinal resolution is independent of noise, and artifacts, in particular when thin
pitch, unlike single-section spiral CT and sections are reconstructed as an input for
multi– detector row CT that relies on 180° 3D postprocessing such as for MPR, max-
and 360° linear interpolation (51,54). Fig- imum intensity projection, or volume-
ure E6 (radiology.rsnajnls.org/cgi/content rendering techniques. For a 1.25-mm spi-
/full/2353040037/DC1) shows the SSPs of a ral section width reconstructed from four-
2-mm section (for four-section CT at 1-mm section CT at 1.0-mm collimation, 0.61–
collimation) and MPRs of a spiral z-axis 0.69 times the dose (depending only
resolution phantom for selected pitch val- slightly on spiral pitch) is required to
ues. As a consequence of the pitch-inde- maintain the image noise of a sequential
pendent spiral section width, the image scan at the same collimation (see refer-
noise for a fixed tube current (in milliam- ences 54,55). Unlike 180° and 360° mul-
peres) would decrease as pitch is decreased, tidetector linear interpolation, image
owing to the increasingly overlapping spi- noise is therefore practically indepen-
ral acquisition. Instead, the user selects an dent of pitch at constant dose.
“effective” milliampere-seconds value, and For a given collimation, such as four
the tube current is then automatically sections at 2.5 mm, image quality can be
adapted to the pitch of the spiral scan to optimized with regard to spiral artifacts
compensate for dose accumulation. The by lowering the pitch (56). Another
dose for fixed effective milliampere-sec- means to reduce spiral artifacts is to use
onds is independent of the spiral pitch and narrow collimation: A given section
equals the dose of a transverse scan with width (eg, 3.0 mm) can be obtained with
the same milliampere-seconds setting (see different collimations, in this case four
Appendix E2, radiology.rsnajnls.org/cgi sections at 1.0 mm and at 2.5 mm. For
/content/full/2353040037/DC1). optimum image quality, collimation that
Thus, as a take-home point, unlike in is narrow relative to the desired section
Figure 10. (a) SSP of 3-mm section for four-sec- single-section spiral CT a change in pitch width is preferable (51). Furthermore, a
tion CT at 1.0- and 2.5-mm collimation. (b) Images does not result in a change in dose to the more rectangular SSP can be established.
of thorax phantom with 3.0-mm section width patient. Accordingly, the use of a higher Figure 10a shows the SSPs of a 3.0-mm
obtained from four-section CT at 2.5 mm colli- pitch does not result in a dose saving, section for four-section CT at both 1.0-
mation and pitch of 0.75 (top) and at 1.0-mm
collimation and pitch of 1.75 (bottom). Despite
which is an important practical consider- and 2.5-mm collimation. Figure 10b
higher pitch, image acquired with 1.0-mm colli- ation with CT systems that rely on adap- shows 3.0-mm transverse sections of a
mation shows reduced artifacts at ribs. tive axial interpolation. thorax phantom scanned with four-sec-
The intrinsic resolution of a multi– de- tion CT at 2.5- and 1.0-mm collimation.
tector row spiral scan is determined by Despite the higher pitch, the 3.0-mm im-
the choice of collimation (eg, four sec- age obtained at 1.0-mm collimation
lation for each projection angle is no tions at 1.0 or 2.5 mm). Z filtering makes shows fewer artifacts. Similar to single-
longer restricted to the two rays closest to it possible to reconstruct images retro- section spiral CT, narrow collimation at

764 䡠 Radiology 䡠 June 2005 Flohr et al


high pitch is preferable to wide collima-
tion at low pitch for artifact reduction.
Except for a minor dose increase due to
the different relative contributions of the
penumbral zones of the dose profile,
scanning at narrow collimation does not
Radiology

result in higher radiation dose to the pa-


tient as long as the effective milliampere-
seconds level is kept constant. Narrow-
collimation scanning should, therefore,
be the protocol of choice for all applica-
tions that require 3D postprocessing as
part of the clinical evaluation. In the
clinical treatment of uncooperative or
Figure 11. Transverse head examination performed with four-sec-
trauma patients or for protocols such as tion CT system. Comparison of sequential image obtained at two-
routine oncologic staging, the use of section CT at 8.00-mm section width (left) and spiral image (right;
wider collimation can be considered. The 8.0-mm section width from four-section CT at 1.0-mm collimation)
best suppression of spiral artifacts is in same patient. Image quality can be considered equivalent owing to
achieved by using both narrow collima- narrow-collimation z-filter reconstruction. (Image courtesy of Roland
Brüning, MD, Klinikum Grosshadern, Munich, Germany.)
tion (relative to the desired section
width) and reduced spiral pitch.
In general, more challenging clinical
protocols, such as CT of the spine and of
introduced for sequential scanning, to near the image plane—that is, measure-
the skull base, are reliant on a combina-
multisection spiral scanning (59,60). ment rays running in or very close to the
tion of narrow collimation and low
With this approach, the measurement image plane are available. These condi-
pitch. When multi– detector row spiral
rays are back projected into a 3D vol- tions need to be fulfilled for a standard
CT of the head is performed with narrow
ume along the lines of measurement, two-dimensional reconstruction. In a fi-
collimation, low pitch, and z-filter recon-
accounting in this way for their cone- nal z-axis reformation step, the tradi-
struction of wider sections, the results are
beam geometry. Three-dimensional back tional transverse images are calculated by
equivalent to those of traditional sequen-
projection is computationally demand- interpolating between the tilted original
tial CT. Figure 11 shows an example of a
ing and requires dedicated hardware to image planes.
head scan performed with a four-section
achieve acceptable image-reconstruction Advanced single-slice rebinning en-
CT system in which a sequential image
times. Other manufacturers use varia- counters its limitations when the spiral
(two-section CT at 8 mm) and a spiral
tions and extensions of nutating-section
image (8-mm section width from four- pitch is reduced to make use of the over-
algorithms (61– 66) for image reconstruc-
section CT at 1-mm collimation) are lapping spiral acquisition and the result-
tion. These algorithms split the 3D recon-
compared in the same patient. ing dose accumulation. The AMPR algo-
struction task into a series of conven-
Some manufacturers who use a z-filter rithm (67,68) addresses this problem:
tional two-dimensional reconstructions
approach do not provide completely free Instead of all available data being used
on tilted intermediate image planes, in
selection of the spiral pitch but recom- for a single image, the data are distrib-
this way benefiting from established and
mend a selection of fixed pitch values (eg, uted to several partial images on double-
very fast two-dimensional reconstruction
pitch of 0.625, 0.75, 0.875, 1.125, 1.25, oblique image planes, which are individ-
techniques. Representative examples are
1.375 and 1.5 for four-section CT with the ually adapted to the spiral path and fan
AMPR (Siemens) (67,68) and the weighted
MUSCOT algorithm [52]) that are aimed at out like the pages of a book (Fig 13, left).
hyperplane reconstruction (proposed by
optimizing the z-axis sampling scheme To ensure full dose utilization the num-
GE Medical Systems) (69,70) techniques.
and reducing spiral artifacts. ber of partial images (“pages” in the
AMPR Method book), as well as the length of the data
interval per image, depend on the spiral
MULTI–DETECTOR ROW SPIRAL The AMPR approach (67,68) is an ex-
pitch. The final transverse (or arbitrarily
RECONSTRUCTION tension and generalization of the “ad-
oriented) images are calculated by means
APPROACHES THAT ACCOUNT vanced single-slice rebinning” (63,64)
of z-axis interpolation between the tilted
FOR CONE-BEAM GEOMETRY method. AMPR allows free selection of
partial image planes (Fig 13, right). The
the spiral pitch with optimized dose uti-
Overview of Cone-Beam shape and the width of the z-axis inter-
lization, which is beneficial for medical
Reconstruction Algorithms polation functions are selectable. Differ-
applications. With advanced single-slice
For CT scanners with 16 or more detec- rebinning, a partial scan interval (about ent SSPs and different section widths can
tor rows, modified reconstruction ap- 240° of scan data) is used for image re- therefore be adjusted, so that z-axis reso-
proaches that account for the cone-beam construction. The image planes are no lution (SSP) can be traded off with image
geometry of the measurement rays have longer perpendicular to the patient axis; noise. The spiral pitch is freely selectable
to be considered. Some manufacturers instead, they are tilted to match the spi- and the section width—and consequently
(Toshiba, Philips) have extended the ral path of the focal spot; see Figure 12 for the z-axis resolution—are independent of
Feldkamp algorithm (57,58), an ap- a 16-section scanner at a pitch of 1.5. For the pitch. The concept of effective milliam-
proximate 3D convolution back-projec- every view angle in this partial scan in- pere-seconds and automatic adaptation of
tion reconstruction that was originally terval, the focal spot is positioned in or the tube current to the pitch also apply to

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 765


down to 0.6 mm in diameter can be re-
solved, the MPRs are relatively free of
geometric distortions, and the spatial in-
tegrity of the 3D image is maintained.
Multi– detector row spiral CT with
Radiology

AMPR is characterized by the same key


properties as adaptive axial interpola-
tion, which can be directly derived from
information in the section on z-filter re-
construction presented earlier in this re-
view. Thus, all recommendations regard-
ing selection of collimation and pitch
that were discussed there also apply for
AMPR. In particular, a change in pitch
does not result in a change in radiation
exposure to the patient, and the use of
higher pitch does not result in dose sav-
Figure 12. Left: Schematic 3D illustration of “advanced single-slice rebinning” approach for ing. Narrow collimation should be used
16-section CT system at pitch of 1.5. Left: Curved line represents spiral path of the focal spot.
whenever possible. With 16-section 0.75-
Intermediate image plane is indicated by gradient-shaded rectangle and is no longer perpendic-
ular to patient axis; instead, it is tilted to match spiral path of the focal spot. Right: Projection mm-collimation CT, the thinnest avail-
onto a plane containing the z-axis, where the spiral path is represented as a sinusoidal line. A able reconstruction section width of 0.75
partial scan interval (about 240°) is used for image reconstruction. For all view angles, focal spot mm is created by using nonlinear weight-
is close to the image plane. ing functions at the z-axis image-refor-
mation step, at the expense of increased
image noise and increased susceptibility
to artifacts. As a take-home point, this
approach again should only be used for
high-contrast applications such as high-
spatial-resolution lung imaging. When
thin sections are reconstructed as input
for 3D postprocessing such as MPR, max-
imum intensity projection, or volume-
rendering techniques, a 1.0-mm section
width is recommended as the most suit-
able trade-off between longitudinal reso-
lution, image noise, and artifacts.
Figure 13. Illustration of AMPR approach. Left: First, multisection
spiral CT data are used to reconstruct several partial images on dou-
ble-oblique image planes, which are individually adapted to the spiral Weighted Hyperplane
path. Partial images fan out like pages of a book. Right: Second, final Reconstruction
images with full dose utilization are calculated with z-axis interpola-
tion between tilted partial image planes. The weighted hyperplane reconstruc-
tion method, which has been described
elsewhere (69,70), uses concepts related
AMPR (see Appendix E2, radiology.rsnajnls medical CT systems with up to 64 detec- to AMPR but is derived differently. Simi-
.org/cgi/content/full/2353040037/DC1). tor rows (71). lar to AMPR, 3D reconstruction is split
With the AMPR approach, sufficient The remaining artifacts in Figure 14 are into a series of two-dimensional recon-
image quality is obtained for all pitch spiral interpolation artifacts (windmill structions. Instead of reconstruction of
values between 0.5 and 1.5 (68). Figure artifacts), not cone-beam artifacts. Wind- traditional transverse sections, convex
14 shows transverse sections and MPRs of mill artifacts are not related to the cone- hyperplanes are proposed as the region
an anthropomorphic thorax phantom. beam geometry and result from the finite of reconstruction. The increasing spiral
Scan data for 16 sections at 0.75-mm col- width of the detector rows, which require overlap with decreasing pitch is handled
limation and pitch of 1 were recon- interpolation between the rows for image by introducing subsets of detector rows,
structed with 1-mm section width with z reconstruction. Hence, windmill artifacts which are sufficient to reconstruct an im-
filtering, the AMPR algorithm, and 3D occur independent of the reconstruction age at a given pitch value. At pitch of
back projection. Neglecting the cone an- approach. They are exaggerated in the 0.5625 with a 16-section scanner, the
gle leads to artifacts at high-contrast ob- mathematic phantom shown (Fig 14) data collected by detector rows one to
jects and geometric distortions, particu- and can be reduced by decreasing the nine form a complete projection data set.
larly in MPRs (Fig 14, top). Both AMPR pitch and/or increasing the reconstruc- Similarly, projections from detector rows
and 3D back projection restore the spatial tion section width relative to the collima- two to 10 can be used to reconstruct an-
integrity of the high-contrast objects, re- tion (Fig 8). Figure 15 shows MPRs of a other image at the same z-axis position.
duce cone-beam artifacts, and are fully z-axis resolution phantom scanned with Projections from detector rows three to
equivalent for 16-section scanning. Re- 16-section CT at 0.75-mm collimation 11 yield a third image and so on. In a
cent studies have demonstrated the ade- and pitches of 0.75, 1.0, 1.25, and 1.5. way, these “subimages” are related to the
quacy of extended versions of AMPR for Independent of the pitch, all cylinders “book pages” of AMPR. The final image is

766 䡠 Radiology 䡠 June 2005 Flohr et al


Radiology

Figure 14. Transverse sections (left) and sagittal MPRs (right) of


anthropomorphic thorax phantom. Scan data for 16-section CT at
0.75-mm collimation and pitch of 1 were reconstructed with 1.0-mm
section width and z filtering that neglected the cone angle of mea-
surement rays (top), with AMPR algorithm (middle), and with 3D
back projection (bottom). Neglecting cone angle leads to artifacts at
high-contrast objects, particularly in MPRs (top). Both AMPR (mid-
dle) and 3D back-projection (bottom) images reduce cone-beam arti-
facts and are fully equivalent for 16-section CT.

Figure 15. MPRs of z-axis resolution phantom at isocenter, scanned


based on a weighted average of the sub- with 16-section CT at 0.75-mm collimation and pitches of 1.5, 1.25, 1.0,
images. In the article by Hsieh et al (70), and 0.75 (MPR section width, 0.75 mm; increment, 0.4 mm). Phantom
consists of polymerized methyl methacrylate plate with rows of cylin-
good image quality was demonstrated for
drical holes (diameters of 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.2, 1.5, 2.0, and 3.0
a 16-section CT system (Lightspeed 16; mm) aligned in the longitudinal direction. Independent of pitch, all
GE Medical Systems) with which the cylinders down to 0.6 mm in diameter can be resolved.
weighted hyperplane reconstruction ap-
proach was used. By performing parame-
ter optimizations, an optimal balance sponds to the number of active detector allow retrospective selection of the data
among various system performance pa- sections. A partial scan data interval is ac- segments used for image reconstruction.
rameters, such as noise, artifacts, and quired with a predefined temporal offset Only scan data acquired in a predefined
SSPs, can be achieved (72). relative to the R waves of the patient’s ECG cardiac phase, usually the diastolic phase,
trace, which can be either relative (as a are used for image reconstruction
ECG-SYNCHRONIZED SCAN AND certain percentage of the R-R interval) or (16,17,74,75). The data segments con-
IMAGE-RECONSTRUCTION absolute (in milliseconds) and either for- tributing to an image begin with a user-
TECHNIQUES ward or reverse (17). Some 16-section CT defined offset relative to the onset of the
systems offer gantry rotation times shorter R waves, similar to ECG-triggered se-
One of the most exciting new applications than 0.5 second (eg, 0.42, 0.40, or 0.37 quential scanning. Image reconstruction
of multi– detector row CT is the ability to second). In this case, temporal resolution generally consists of two steps: multi– de-
image the heart and the cardiothoracic can be as good as 0.21, 0.20, or 0.185 sec- tector row spiral interpolation to com-
anatomy without motion artifacts. For ond (26,27). pensate for the continuous table move-
ECG-synchronized scanning of the cardio- With retrospective ECG gating, the ment and to obtain scan data at the
thoracic anatomy, either ECG-triggered se- heart volume is covered continuously by desired image z-axis position, followed
quential scanning or ECG-gated spiral a spiral scan. The basic concepts for ECG- by a partial scan reconstruction of the
scanning can be used. In ECG-triggered se- gated spiral imaging, such as single-seg- transverse data segments. The temporal
quential scanning, the heart volume is cov- ment and multisegment reconstruction, resolution of an image can be improved
ered by subsequent transverse scans with a had already been developed in 1998 (73). up to trot/(2N) by using scan data of N
step-and-shoot technique. For each trans- The patient’s ECG signal is recorded at subsequent heart cycles for image forma-
verse scan, the number of images corre- the same time the CT data are acquired to tion in a so-called multisegment recon-

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 767


encounters its limitations in patients with
arrhythmia. To maintain the benefits of
ECG-gated spiral CT but reduce patient
dose, ECG-controlled dose modulation has
been developed (42,43) (see earlier discus-
Radiology

sion).
The major improvements of 16-section
CT, compared with established four-sec-
tion scanners, include improved temporal
resolution due to shorter gantry rotation
time, better spatial resolution owing to sub-
millimeter collimation, and considerably re-
duced scan acquisition times (26,27). The
time to cover the entire heart volume
(about 12 cm) with four-section CT at
1.0-mm collimation is about 40 seconds,
which is at the limit for a scan requiring
patient breath holding. ECG-gated CT of
the entire thorax or the aorta is not pos-
sible within reasonable scan durations.
For a 16-section CT system, the time to
cover the entire heart volume with sub-
millimeter collimation is about 15 sec-
onds. With 16-section CT, coverage of
the entire thorax (30 cm) can be com-
pleted in about 38 seconds at 0.75-mm
Figure 16. Clinical performance of ECG-gated 16-section CT collimation and in about 19 seconds at
(0.75-mm collimation, 0.42-second gantry rotation) of entire thorax. 1.5-mm collimation. ECG-gated exami-
Coronal volume-rendered reconstruction shows left internal mam- nations of extended cardiothoracic anat-
mary bypass graft (LIMA) to left anterior descending coronary artery omy became feasible with 16-section CT,
and saphenous venous bypass graft (SVG) to right coronary artery. which lends itself to a spectrum of appli-
Native right internal mammary artery (RIMA) is also visible.
cations where suppression of cardiac
pulsation is desired. Typical diagnostic
pitfalls caused by transmitted cardiac
struction mode (16,73–77), where trot is different approach, single-segment par- pulsation can be avoided, such as an ar-
the gantry rotation time of the CT scan- tial-scan images are prospectively recon- tifactual intimal flap resembling dissec-
ner. With increased N, better temporal structed as baseline images, followed by tion in the ascending aorta (79). Suppres-
resolution is achieved but at the expense retrospective two-segment reconstruc- sion of cardiac pulsation improves the
of slower volume coverage: Increased N tion for improved temporal resolution in assessment of paracardiac lung segments
and slower patient heart rate require a patients with a higher heart rate. Yet an- and allows confident exclusion of small
reduction in spiral pitch. other approach is prospective adjustment peripheral pulmonary emboli in segmen-
Multisegment approaches rely on a com- of the gantry rotation time to the heart tal and subsegmental arteries (80). In rou-
plete periodicity of the heart motion, and rate of the patient to obtain an optimized tine thoracic studies, which are not syn-
these approaches encounter their limita- temporal resolution for a multisegment chronized to the patient’s ECG signal,
tions in patients with arrhythmia or a reconstruction. Again, this approach re- cardiac motion usually precludes the assess-
heart rate that changes during scan acqui- quires a stable and predictable heart rate ment of coronary bypass grafts. Figure 16
sition. Multisegment reconstruction may during scan acquisition. shows an example of an ECG-gated scan
improve image quality in selected cases, Prospective ECG triggering combined of the entire thorax for a patient with
but the reliability of good-quality image with sequential step-and-shoot acquisition bypass grafts; this scan was acquired with
acquisitions with N-segment reconstruc- of transverse sections has the benefit of 16-section CT at 0.75-mm collimation
tion is compromised with increases in N. smaller patient dose than that of ECG- and 0.42-second gantry rotation.
In general, clinical practice suggests gated spiral scanning, because scan data
the use of one segment at lower heart are acquired only during the desired heart
rates and two or more (N ⱖ 2) segments phases. However, this technique does not APPLICATIONS
at higher heart rates. Use of single-seg- provide continuous volume coverage with
ment versus multisegment reconstruc- overlapping sections, and misregistration Clinical applications benefit from multi–
tion is integrated in the data acquisition of anatomic details cannot be avoided. Fur- detector row CT technology in several
process in a variety of ways, depending thermore, reconstruction of images in dif- ways: (a) shorter scan time (important for
on the scanner type. One approach con- ferent phases of the cardiac cycle for func- trauma patients and pediatric patients, CT
sists of automatic division of the partial- tional evaluation is not possible. Since angiography), (b) extended scan range (im-
scan data segment into one or two sub- ECG-triggered sequential scanning de- portant for CT angiography, combined
segments, depending on the patient’s pends on a reliable prediction of the pa- chest-abdomen scans such as in oncologic
heart rate during acquisition (“adaptive tient’s next R-R interval by using the mean staging), and (c) improved longitudinal res-
cardio volume” algorithm [74]). With a of the preceding R-R intervals, the method olution (beneficial for all reconstructions,

768 䡠 Radiology 䡠 June 2005 Flohr et al


particularly when 3D postprocessing is part
of the clinical protocol).
Most protocols even benefit from a com-
bination of all of these advantages. The
near isotropic spatial resolution in routine
examinations enables 3D renderings of di-
Radiology

agnostic quality and oblique MPRs and


maximum intensity projections of a reso-
lution similar to that of the transverse im-
ages. The availability of multi– detector
row CT technology has already begun to
change the traditional perception of CT
imaging. In CT, a distinction is tradition-
ally made between longitudinal and in-
plane resolution. This distinction is based
mainly on historical reasons. Before the in-
troduction of spiral CT, longitudinal reso-
lution was determined by section collima-
tion alone, while the convolution kernel
determined in-plane resolution. With spi-
ral CT, collimation is no longer the only
factor used to determine longitudinal res-
olution; the spiral interpolation function
Figure 17. Clinical performance of 16-section CT (0.75-mm collima-
also comes into play. This has been a first tion, 0.5-second gantry rotation). Coronal volume-rendered reconstruc-
step toward decoupling the image section tion shows occlusion of left common iliac artery (arrow). (Image cour-
width from the beam width as determined tesy of Axel Küttner, MD, University of Tübingen, Germany.)
by the collimation. Multi– detector row CT
now allows reconstruction of arbitrary sec-
tion widths from a given collimation by
using z-filter techniques, as long as the de- pitch of 1.5 requires only 9 seconds for a patients with limited ability to cooperate
sired section width is not smaller than the scan range of about 300 mm (with table (11,83). Meanwhile the use of multi– detec-
collimation. The potential to trade off z-axis feed of 36 mm/sec). For the first time, true tor row CT for a combined diagnosis of
resolution and image noise for the same data arterial phase imaging of the entire carotid pulmonary embolism and deep venous
set is the most important benefit of z-filter artery with high spatial resolution can be thrombosis has been clinically established
reconstruction. In many applications, data performed. Clinical practice indicates the (83). Both a native and a contrast-en-
acquisition with narrow collimation is rec- potential of 16-section CT angiography to hanced scan of the thorax can be obtained
ommended independently of the section replace conventional interventional an- within the same breath hold for matching
width desired for primary viewing. giography in the evaluation of carotid ar- of both image volumes as a basis for inves-
The distinction between longitudinal tery stenosis (81). Evaluation of the su- tigational applications such as lung perfu-
and in-plane resolution will gradually be- praaortic vessels with 16-section CT is sion imaging.
come a historical curiosity, and the tradi- particularly useful in emergency situations, Sixteen-section CT enables whole body
tional transverse section will loose its clin- since CT allows a quick diagnosis with op- angiographic studies with submillimeter
ical importance. In its place, interactive timized patient access. resolution in a single breath hold. Also,
viewing and manipulation of isotropic vol- For patients suspected of having isch- 16-section CT yields the same morpho-
ume images will become commonplace, emic stroke, both the status of the vessels logic information as invasive angiography
with only the key sections or views in ar- supplying the brain and the location of the (84,85). CT angiography of the chest and
bitrary directions recorded and stored. intracranial occlusion can be assessed dur- abdomen with submillimeter collimation
Spiral scanning with 16 submillimeter ing the same examination (82). Brain per- can be completed in about 17 seconds for a
sections, in particular, represents a break- fusion CT can be performed by using the scan range of 600 mm (Fig 17). When true
through on the way to true isotropic res- same modality, with the goal of differenti- isotropic resolution is not required, the
olution for routine clinical applications. ating irreversibly damaged brain tissue use of 16-section CT at 1.25- or 1.5-mm
Improved longitudinal resolution is com- from reversibly impaired tissue at risk. The collimation enables even shorter exami-
bined with considerably reduced scan combined use of nonenhanced CT, perfu- nation times or extended scan ranges (eg,
times, which facilitate examinations in sion CT, and CT angiography may rapidly for oncologic screening, trauma cases, or
uncooperative patients and reduce the provide comprehensive information re- CT angiography). Whole-body 16-sec-
amount of contrast material needed (al- garding the extent of ischemic damage in tion CT angiography with 1500-mm scan
though optimized contrast material pro- patients with acute stroke (46). range, 1.5-mm collimation, 0.5-second
tocols are also required). Scan acquisition of the entire thorax rotation time, and pitch of 1.25 (table
Furthermore, new clinical applications (350 mm) with submillimeter collimation feed, 60 mm/sec) can be completed in
are evolving as a result of the increased can now be performed in approximately only 26 seconds.
speed of volume scanning. CT angiogra- 11 seconds. Owing to the short breath- ECG-gated cardiac scanning benefits
phy of the carotid arteries and the circle of hold time, central and peripheral pulmo- from both improved temporal resolution
Willis with 16 sections at 0.75-mm colli- nary embolism can be reliably and accu- and improved spatial resolution. Detec-
mation, 0.5-second rotation time, and rately diagnosed even in severely dyspneic tion and characterization of coronary

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 769


stance, opens a wide spectrum of appli-
cations ranging from oncologic staging
to comprehensive cardiac examinations.
The clinical potential of these scanners is
currently being evaluated (91). Recon-
Radiology

struction of the CT images in a sufficient


field of view without truncation of ana-
tomic structures (eg, arms) is a prerequi-
site for adequate attenuation correction
of the PET images. An enlarged field of
view of up to 70 cm can be realized by
Figure 18. Clinical performance of ECG-gated 16-section coronary
extrapolating from the measured CT
CT angiography (0.75-mm collimation, 0.37-second gantry rotation). data. Pertinent algorithms can be found
Images of patient after insertion of Y stent into bifurcation of left in, for example, reference 92. Figure 19
main coronary artery into left anterior descending and left circumflex shows MPRs from CT images in a 46-year-
coronary arteries. Mean heart rate of the patient during examination old man with renal cancer who had un-
was 67 beats per minute. Left: Y stent is shown in 3D volume- dergone nephrectomy and chemother-
rendered reconstruction. Right: Multiplanar reformations of left an-
terior descending and circumflex arteries demonstrate stent patency
apy, with PET images superimposed.
with sufficient diagnostic quality to obviate invasive coronary angiog- Areas with increased metabolism are en-
raphy. (Image courtesy of Filippo Cademartiri, MD, Thorax Center hanced, and a metastatic mediastinal
Rotterdam, the Netherlands.) lymph node can be identified, which
supports the notion of PET as adding a
“new contrast agent” to CT.
Systems that combine CT and single-
plaque, even in the presence of severe by administering ␤-blockers to such pa- photon emission computed tomography
calcifications, greatly benefits from the tients. Improved temporal resolution is are another promising modality. Poten-
increased robustness of the technology. desirable in the future to prevent the tial applications are currently being in-
Sixteen-section CT allows assessment of need for heart rate control. Increased vestigated and range from the localiza-
small, peripheral coronary segments that, gantry rotation speed, rather than multi- tion of parathyroid lesions (93) and
until now, could not be evaluated. In a segment reconstruction, appears to be heterotopic splenic tissue (94) to detec-
recent study (86) in which coronary CT an- preferable for robust clinical perfor- tion of recurrent nasopharyngeal carci-
giography with a 16-section system was mance. Obviously, substantial develop- nomas (95) to imaging of aortic prosthe-
investigated in 59 patients, 86% specific- ment efforts are needed to account for sis infection (96).
ity and 95% sensitivity were demon- the notable increase in mechanical forces CT virtual simulation is gaining increas-
strated for identification of significant (about 17g for 0.42-second rotation, ing importance with a more widespread
coronary artery stenosis. None of the ⬎33g for 0.3-second rotation) and in- adoption in 3D conformal and intensity-
patients had to be excluded, unlike in creased data transmission rates. A rota- modulated radiation therapy. With gen-
previous studies that were based on less- tion time of less than 0.2 second (me- eral-purpose CT systems that have a gan-
advanced scanner technology. Other in- chanical force ⬎ 75g), which is required try opening with a typical diameter of 70
vestigators have reported similar results to provide a temporal resolution of less cm, some patients (eg, women with
(87). Early clinical experience with 0.37- than 100 msec independent of heart rate, breast cancer) cannot always be scanned
second gantry rotation indicates im- appears to be beyond today’s mechanical in the treatment position. Such applica-
proved image quality due to reduced limits. An alternative to further increases tions, along with interventional proce-
cardiac motion and increased clinical in rotation speed is to reconsider the dures and trauma protocols, will be facil-
robustness at higher heart rates, which scanner concept with multiple tubes and itated by CT systems with a larger bore
thereby potentially reduce the number multiple detectors that had already been (97). Recently, concepts have been intro-
of patients who require heart rate con- described in the early years of CT (88,89). duced for four- and 16-section CT scan-
trol (Fig 18). Owing to its ease of use and its wide- ners with a bore diameter of up to 85 cm
spread availability, general-purpose CT and a reconstruction field of up to 82 cm,
FUTURE DIRECTION OF continues to evolve into the most widely owing to image reconstruction based on
MULTI–DETECTOR ROW CT used diagnostic modality for routine ex- data extrapolation. These systems will
aminations, especially in emergency probably gain considerable importance
Sixteen-section CT, which has become situations or for oncologic staging. CT in the near future, in particular with re-
widely available, enables truly isotropic primarily provides morphologic infor- gard to the dramatically increasing num-
submillimeter imaging for virtually any mation; in combination with other mo- ber of severely obese patients in the
application. In the case of cardiac imag- dalities, however, functional and meta- Western countries.
ing, 16-section CT sets today’s bench- bolic information can also be obtained For general purpose CT, we will witness
mark in spatial resolution for noninva- (90). Therefore, combined systems for a moderate increase in the number of
sive coronary artery imaging. Motion obtaining comprehensive structural and simultaneously acquired sections in the
artifacts in patients with a higher heart functional diagnoses will gain increasing near future. A new generation of CT systems
rate remain the most important chal- importance in the near future. with 32, 40 and—in combination with re-
lenge for multi– detector row coronary The combination of state-of-the-art mul- fined z-axis sampling techniques—64 simul-
CT angiography, although diagnostic im- ti– detector row CT with positron emission taneously acquired sections are currently
age quality can be achieved in most cases tomographic (PET) scanners, for in- being introduced. However in contrast to

770 䡠 Radiology 䡠 June 2005 Flohr et al


Radiology

Figure 20. Prototype CT system incorporates cesium iodide flat-


panel detector into a standard CT gantry.
Figure 19. Clinical performance of PET/CT. Sagittal (left) and coro-
nal (right) MPRs from CT data in patient with renal cancer, with PET
images superimposed. Areas with enhanced metabolism show more
avid tracer accumulation. Metastatic mediastinal lymph node (ar-
rows) can be identified.

the transition from single-section to four-


and 16-section CT, clinical performance
will improve only incrementally with
further increases in the number of detec-
tor rows. The achievable clinical benefit
will have to be carefully considered in the
light of the necessary technical efforts Figure 21. Volume-rendered display of stationary heart specimen
and the cost. Clinical progress can more scanned with flat-panel CT prototype with cesium iodide detector
likely be expected from further improve- shown in Figure 20. The 0.25-mm3 isotropic resolution enables ex-
quisite delineation of small side branches of contrast material–filled
ments in spatial resolution rather than
coronary artery tree. LAD ⫽ left anterior descending coronary artery,
from an increase in the volume-coverage RCA ⫽ right main coronary artery.
speed. In clinical reality, the latter has
only rarely been a limiting factor since
the introduction of 16-section CT. As
soon as all relevant examinations can be to contrast resolution and fast data read- trast objects such as joints, the inner ear,
performed in a comfortable breath hold out. A scanner with 256 0.5-mm detector or contrast material–filled vessel speci-
of not more than 10 seconds, a further elements has been proposed by one man- mens (98,99).
increase in the number of sections will ufacturer and appears to be conceptually Figure 20 shows a prototype set-up,
not provide a substantial clinical benefit. promising, but this system is still in the where a flat-panel detector was incorpo-
At this point, a qualitative enhance- prototype stage. Prototype systems by rated into a standard CT gantry (Soma-
ment of CT that allows new clinical ap- other vendors use cesium iodide–amor- tom Sensation 16; Siemens). The detector
plications may again bring substantial phous silicon flat-panel detector technol- covers a 25 ⫻ 25 ⫻ 18 cm scan field of
clinical progress with, for example, the ogy that was originally used for conven- view, and the pixel size is 0.25 ⫻ 0.25
introduction of area detectors large tional angiography, which is limited in mm, both measured at the center of ro-
enough to cover entire organs such as the terms of low contrast resolution and im- tation. Figure 21 shows volume render-
heart, kidneys, or brain in one sequential aging speed. Owing to the intrinsic slow ings of a heart specimen (80 kV, 20 mA,
scan (approximate scan range, 120 mm). signal decay of flat-panel detectors, rota- 20-second gantry rotation) that demon-
With these systems, dynamic volume tion times of at least 20 seconds are strate excellent spatial resolution, which
scanning would become feasible, which needed to acquire a sufficient number of enables visualization of even very small
would open up a whole spectrum of new projections (ⱖ600 projections). The spatial side branches of the coronary artery tree.
applications such as functional or vol- resolution of such systems is excellent, The combination of area detectors that
ume perfusion studies. though, because of the small detector provide sufficient image quality with fast
Area-detector technology is currently pixel size. Excessive dose requirements to gantry rotation speed will be a promising
under development, but no commer- date, however, preclude the examination technical concept for medical CT sys-
cially available system so far fulfills the of larger objects. Initial experimental re- tems. The vast spectrum of potential ap-
requirements of medical CT with regard sults are thus limited to small high-con- plications may bring about another

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 771


quantum leap in the evolution of medi- 18. Hong C, Becker CR, Huber A, et al. ECG- 37. Wildberger JE, Mahnken AH, Schmitz-Rode
cal CT imaging; however such systems gated reconstructed multi– detector row CT T, et al. Individually adapted examination
coronary angiography: effect of varying trig- protocols for reduction of radiation exposure
will probably not be available in the near ger delay on image quality. Radiology 2001; in chest CT. Invest Radiol 2001; 36:604 – 611.
future. 220:712–717. 38. Schaller S, Niethammer MU, Chen X, Klotz
19. Achenbach S, Ulzheimer S, Baum U, et al. E, Wildberger JE, Flohr T. Comparison of sig-
Noninvasive coronary angiography by retro- nal-to-noise and dose values at different tube
References
Radiology

spectively ECG-gated multi-slice spiral CT. voltages for protocol optimization in pediat-
1. Kalender W, Seissler W, Klotz E, Vock P. Spi- Circulation 2000; 102:2823–2828. ric CT (abstr). Radiology 2001; 221(P):366.
ral volumetric CT with single-breath-hold 20. Becker C, Knez A, Ohnesorge B, Schöpf U, 39. Wintersperger B, Becker CR, Schaller S, Rist
technique, continuous transport, and con- Reiser M. Imaging of non calcified coronary C, Jakobs TF, Reiser MF. Low dose 16-detec-
tinuous scanner rotation. Radiology 1990; plaques using helical CT with retrospective tor-row CT angiography of the abdominal
176:181–183. EKG gating. AJR Am J Roentgenol 2000; 175: vessels with improved vessel enhancement
2. Crawford CR, King KF. Computed tomogra- 423– 424. (abstr). In: Radiological Society of North
phy scanning with simultaneous patient 21. Knez A, Becker C, Leber A, Ohnesorge B, America scientific assembly and annual
translation. Med Phys 1990; 17:967–982. Reiser M, Haberl R. Non-invasive assessment meeting program. Oak Brook, Ill: Radiologi-
3. Rubin GD, Dake MD, Semba CP. Current sta- of coronary artery stenoses with multidetec- cal Society of North America, 2003; 541–542.
tus of three-dimensional spiral CT scanning tor helical computed tomography. Circula-
for imaging the vasculature. Radiol Clin 40. Kalender WA, Wolf H, Suess C. Dose reduc-
tion 2000; 101:e221– e222. tion in CT by anatomically adapted tube cur-
North Am 1995; 33:51–70. 22. Nieman K, Oudkerk M, Rensing BJ, et al.
4. Napel S, Rubin GD, Jeffrey RB. STS-MIP: a rent modulation. II. Phantom measure-
Coronary angiography with multi-slice com- ments. Med Phys 1999; 26:2248 –2253.
new reconstruction technique for CT of the puted tomography. Lancet 2001; 357:599 –
chest. J Comput Assist Tomogr 1993; 17:832– 41. Greess H, Wolf H, Baum U, et al. Dose reduc-
603. tion in computed tomography by attenua-
838. 23. Schroeder S, Kopp A, Baumbach A, et al.
5. Kalender W. Thin-section three-dimensional tion-based on-line modulation of the tube
Noninvasive detection and evaluation of current: evaluation of six anatomical re-
spiral CT: is isotropic imaging possible? (ed-
atherosclerotic coronary plaques with multi- gions. Eur Radiol 2000; 10:391–394.
itorial). Radiology 1995; 197:578 –580.
slice computed tomography. J Am Coll Car- 42. Jakobs TF, Becker CR, Ohnesorge B, et al.
6. Liang Y, Kruger RA. Dual-slice spiral versus
diol 2001; 37:1430 –1435. Multislice helical CT of the heart with retro-
single-slice spiral scanning: comparison of
24. Schroeder S, Flohr T, Kopp AF, et al. Accuracy spective ECG gating: reduction of radiation
the physical performance of two computed
of density measurements within plaques lo- exposure by ECG-controlled tube current
tomography scanners. Med Phys 1996; 23:
cated in artificial coronary arteries by x-ray modulation. Eur Radiol 2002; 12:1081–1086.
205–220.
multislice CT: results of a phantom study. 43. Poll L, Cohnen M, Brachten S, Ewen K, Mod-
7. Klingenbeck-Regn K, Schaller S, Flohr T,
J Comput Assist Tomogr 2001; 25:900 –906. der U. Dose reduction in multidetector row
Ohnesorge B, Kopp AF, Baum U. Subsecond
25. Kopp A, Schröder S, Küttner A, et al. Coro- CT of the heart by use of ECG-controlled
multi-slice computed tomography: basics
nary arteries: retrospectively ECG-gated tube current modulation (“ECG pulsing”):
and applications. Eur J Radiol 1999; 31:110 –
multi– detector row CT angiography with
124. phantom measurements. Rofo Fortschr Geb
selective optimization of the image recon-
8. Hu H, He HD, Foley WD, Fox SH. Four mul- Rontgenstr Neuen Bildgeb Verfahr 2002;
struction window. Radiology 2001; 221:
tidetector-row helical CT: image quality and 174:1500 –1505.
volume coverage speed. Radiology 2000; 683– 688.
44. Hsieh J. Investigation of the slice sensitivity
215:55– 62. 26. Flohr T, Bruder H, Stierstorfer K, Simon J,
profile for step-and-shoot mode multi-slice
9. Ohnesorge B, Flohr T, Schaller S, et al. Tech- Schaller S, Ohnesorge B. New technical de-
computed tomography. Med Phys 2001; 28:
nische grundlagen und anwendungen der velopments in multislice CT. II. Sub-millime-
491–500.
mehrschicht-CT. Radiologe 1999; 39:923– ter 16-slice scanning and increased gantry
45. Joseph PM, Spital RD. The exponential edge-
931. rotation speed for cardiac imaging. Rofo
gradient effect in x-ray computed tomogra-
10. McCollough CH, Zink FE. Performance eval- Fortschr Geb Rontgenstr Neuen Bildgeb Ver-
phy. Phys Med Biol 1981; 26:473– 487.
uation of a multidetector row CT system. fahr 2002; 174:1022–1027.
27. Flohr T, Schoepf UJ, Kuettner A, et al. Ad- 46. Tomandl BF, Klotz E, Handschu R, et al.
Med Phys 1999; 26:2223–2230. Comprehensive imaging of ischemic stroke
11. Remy-Jardin M, Tillie-Leblond I, Szapiro D, vances in cardiac imaging with 16-section
CT-systems. Acad Radiol 2003; 10:386 – 401. with multisection CT. RadioGraphics 2003;
et al. CT angiography of pulmonary embo- 23:565–592.
lism in patients with underlying respiratory 28. Schoepf UJ, Becker CR, Ohnesorge BM, Yucel
EK. CT of coronary artery disease. Radiology 47. Polacin A, Kalender WA, Marchal G. Evalua-
disease: impact of multislice CT on image tion of section sensitivity profiles and image
quality and negative predictive values. Eur 2004; 232:18 –37.
29. Hu H. Multi-slice helical CT: Scan and recon- noise in spiral CT. Radiology 1992; 185:29 –
Radiol 2002; 12:1971–1978. 35.
12. Rubin GD, Schmidt AJ, Logan LJ, Sofilos MC. struction. Med Phys 1999; 26:5–18.
30. Brenner D, Elliston C, Hall E, Berdon W. 48. Rubin GD, Napel S. Increased scan pitch for
Multi– detector row CT angiography of lower vascular and thoracic spiral CT. Radiology
extremity arterial inflow and runoff: initial Estimated risks of radiation-induced fatal
cancer from pediatric CT. AJR Am J Roentge- 1995; 197:316 –317.
experience. Radiology 2001; 221:146 –158.
nol 2001; 176:289 –296. 49. Saito Y, Suzuki T. Evaluation of the perfor-
13. Schoepf UJ, Bruening RD, Hong C, et al. Mul-
31. Nickoloff E, Alderson P. Radiation exposure mance of multidetector row CT system in
tislice helical CT of focal and diffuse lung
to patients from CT: reality, public percep- non-helical scanning (abstr). Radiology 1998;
disease: comprehensive diagnosis with re-
tion, and policy. AJR Am J Roentgenol 2001; 209(P):578.
construction of contiguous and high-resolu-
177:285–287. 50. Hsieh J. Analytical models for multi-slice he-
tion CT sections from a single thin-collima-
tion scan. AJR Am J Roentgenol 2001; 177: 32. McCollough C. Patient dose in cardiac com- lical CT performance parameters. Med Phys
179 –184. puted tomography. Herz 2003; 28:1– 6. 2003; 30:169 –178.
14. Villablanca JP, Lahan R, Hooshi P, et al. De- 33. Morin R, Gerber T, McCollough C. Radiation 51. Schaller S, Flohr T, Klingenbeck K, Krause J,
tection and characterization of very small dose in computed tomography of the heart. Fuchs T, Kalender WA. Spiral interpolation
cerebral aneurysms by using 2D and 3D he- Circulation 2003; 107:917–922. algorithm for multi-slice spiral CT. I. Theory.
lical CT angiography. AJNR Am J Neurora- 34. International Electrotechnical Commission IEEE Trans Med Imaging 2000; 19:822– 834.
diol 2002; 23:1187–1198. 60601–2– 44. Amendment 1: medical electri- 52. Taguchi K, Aradate H. Algorithm for image
15. Flohr T, Stierstorfer K, Bruder H, Simon J, cal equipment, part 2– 44: particular require- reconstruction in multi-slice helical CT. Med
Schaller S. New technical developments in ments for the safety of x-ray equipment for Phys 1998; 25:550 –561.
multislice CT. I. Approaching isotropic reso- computed tomography. Geneva, Switzer- 53. Flohr T, Klingenbeck-Regn K, Ohnesorge B,
lution with sub-mm 16-slice scanning. Rofo land: International Electrotechnical Com- Schaller S. Multislice scanning with the So-
Fortschr Geb Rontgenstr Neuen Bildgeb Ver- mission, 2002. matom Volume Zoom: an optimized design
fahr 2002; 174:839 – 845. 35. Donnelly LF, Emery KH, Brody AS, et al. Min- for volume scanning. In: Reiser MF, Taka-
16. Kachelriess M, Ulzheimer S, Kalender W. imizing radiation dose for pediatric body ap- hashi M, Modic M, Bruening R, eds. Multi-
ECG-correlated image reconstruction from plications of single-detector helical CT: strat- slice CT: medical radiology— diagnostic im-
subsecond multi-slice spiral CT scans of the egies at a large children’s hospital. AJR Am J aging and radiation oncology. Berlin,
heart. Med Phys 2000; 27:1881–1902. Roentgenol 2001; 176:303–306. Germany: Springer-Verlag, 2001.
17. Ohnesorge B, Flohr T, Becker C, et al. Cardiac 36. Frush DP, Soden B, Frush KS, Lowry C. Im- 54. Fuchs T, Krause J, Schaller S, Flohr T,
imaging by means of electrocardiographi- proved pediatric multidetector body CT us- Kalender WA. Spiral interpolation algo-
cally gated multisection spiral CT: initial ex- ing a size-based color-coded format. AJR rithms for multislice spiral CT. II. Measure-
perience. Radiology 2000; 217:564 –571. Am J Roentgenol 2002; 178:721–726. ment and evaluation of slice sensitivity pro-

772 䡠 Radiology 䡠 June 2005 Flohr et al


files and noise at a clinical multislice system. multiple plane reconstruction: a novel ap- able noninvasive coronary angiography with
IEEE Trans Med Imaging 2000; 19:835– 847. proximate reconstruction scheme for multi- fast submillimeter multislice spiral com-
55. Schaller S, Ohnesorge B, Flohr T, Klingen- slice spiral CT. Phys Med Biol 2002; 47:2571– puted tomography. Circulation 2002; 106:
beck-Regn K. Dose in multislice spiral CT 2581. 2051–2054.
(abstr). Radiology 1999; 213(P):284. 72. Toth TL, Simoni P. A quantitative measure of 87. Ropers D, Baum U, Pohle K, et al. Detection
56. Wessling J, Fischbach R, Ludwig K, et al. CT helical artifact (abstr). Radiology 2001; of coronary artery stenoses with thin-slice
Multi-slice spiral CT of the abdomen in on- 221(P):218. multi-detector row spiral computed tomog-
Radiology

cological patients: influence of table support 73. Kachelriess M, Kalender W. Electrocardio- raphy and multiplanar reconstruction. Cir-
and detector configuration on image quality gram-correlated image reconstruction from culation 2003; 107:664 – 666.
and radiation exposure. Rofo Fortschr Geb subsecond spiral computed tomography 88. Robb R, Ritman E. High speed synchronous
Rontgenstr Neuen Bildgeb Verfahr 2001; scans of the heart. Med Phys 1998; 25:2417– volume computed tomography of the heart.
173:373–378. 2431. Radiology 1979; 133:655– 661.
57. Feldkamp LA, Davis LC, Kress JW. Practical 74. Flohr T, Ohnesorge B. Heart-rate adaptive 89. Ritman E, Kinsey J, Robb R, Gilbert B, Harris
cone-beam algorithm. J Opt Soc Am A 1984; optimization of spatial and temporal resolu- L, Wood E. Three-dimensional imaging of
1:612– 619. tion for ECG-gated multi-slice spiral CT of heart, lungs, and circulation. Science 1980;
58. Grass M, Köhler T, Proksa R. 3D cone-beam the heart. J Comput Assist Tomogr 2001; 25: 210:273–280.
CT reconstruction for circular trajectories. 907–923.
90. Townsend DW, Cherry SR. Combining anat-
Phys Med Biol 2000; 45:329 –347. 75. Taguchi K, Anno H. High temporal resolu-
omy and function: the path of true image
59. Wang G, Lin T, Cheng P. A general cone- tion for multi-slice helical computed tomog-
fusion. Eur Radiol 2001; 11:1968 –1974.
beam reconstruction algorithm. IEEE Trans raphy. Med Phys 2000; 27:861– 872.
91. Namdar M, Kaufmann P, Hany T, von Schul-
Med Imaging 1993; 12:486 – 496. 76. Bruder H, Schaller S, Ohnesorge B, Mertel-
60. Schaller S. Practical image reconstruction for meier T. High temporal resolution volume thess G. Combined CT-angiogram and PET
cone-beam computed tomography (disserta- heart imaging with multirow computed to- perfusion imaging for assessment of CAD in
tion). Nürnberg, Germany: Erlangen Univer- mography. In: Hanson KM, ed. Proceedings a novel PET/CT: a pilot feasibility study (ab-
sity; 1998. of SPIE: medical imaging 1999 —image pro- str). Eur Radiol 2003; 13(suppl):165.
61. Turbell H, Danielsson PE. An improved PI- cessing. Vol 3661. Bellingham, Wash: Inter- 92. Ohnesorge B, Flohr T, Schwarz K, Heiken JP,
method for reconstruction from helical cone national Society for Optical Engineering, Bae KT. Efficient correction for CT image ar-
beam projections. Presented at the IEEE Med- 1999; 420 – 432. tifacts caused by objects extending outside
ical Imaging Conference, Seattle, Wash, Oc- 77. Pan T, Shen Y. New multi-sector reconstruc- the scan field of view. Med Phys 2000; 27:
tober 24 –30, 1999. tion for cardiac ct. Presented at IEEE Medical 39 – 46.
62. Proksa R, Koehler T, Grass M, Timmer J. The Imaging Conference, Lyon, France, October 93. Kaczirek K, Prager G, Kienast O, et al. Com-
n-PI method for helical cone-beam CT. IEEE 16 –20, 2000. bined transmission and (99m)Tc-sestamibi
Trans Med Imaging 2000; 19:848 – 863. 78. Cesmeli E, Edic M, Iatrou M, Pfoh A. A novel emission tomography for localization of me-
63. Larson G, Ruth C, Crawford C. Nutating slice reconstruction algorithm for multiphasic diastinal parathyroid glands. Nuklearmedi-
CT image reconstruction. U.S. patent appli- cardiac imaging using multislice CT. In: An- zin 2003; 42:220 –223.
cation WO 98/44847; filed April 8, 1998. tonuk LE, Yaffe MJ, eds. Proceedings of SPIE: 94. Horger M, Eschmann SM, Lengerke C, Claus-
64. Kachelriess M, Schaller S, Kalender WA. Ad- medical imaging 2001—physics of medical sen CD, Pfannenberg C, Bares R. Improved
vanced single-slice rebinning in cone-beam imaging. Vol 4320. Bellingham, Wash: Inter- detection of splenosis in patients with
spiral CT. Med Phys 2000; 27:754 –772. national Society for Optical Engineering; haematological disorders: the role of com-
65. Bruder H, Kachelriess M, Schaller S, Stierstor- 645– 654. bined transmission-emission tomography.
fer K, Flohr T. Single-slice rebinning recon- 79. Loubeyre P, Angelie E, Grozel F, Abidi H, Eur J Nucl Med Mol Imaging 2003; 30:316 –
struction in spiral cone-beam computed to- Minh VA. Spiral CT Artifact that simulates 319.
mography. IEEE Trans Med Imaging 2000; aortic dissection: image reconstruction with 95. Tai CJ, Shian YC, Wang IJ, Ho YJ, Ho ST, Kao
19:873– 887. use of 180° and 360° linear-interpolation al- CH. Detection of recurrent or residual naso-
66. Kohler T, Proksa R, Bontus C, Grass M, Tim- gorithms. Radiology 1997; 205:153–157. pharyngeal carcinomas after radiotherapy
mer J. Artifact analysis of approximate heli- 80. Schoepf U, Helmberger T, Holzknecht N, et with technetium-99m tetrofosmin single
cal cone-beam CT reconstruction algo- al. Segmental and subsegmental pulmonary photon emission computed tomography and
rithms. Med Phys 2002; 29:51– 64. arteries: evaluation with electron-beam CT comparison with computed tomography: a
67. Schaller S, Stierstorfer K, Bruder H, Kachel- versus spiral CT. Radiology 2000; 214:433– preliminary study. Cancer Invest 2003; 21:
riess M, Flohr T. Novel approximate ap- 439. 536 –541.
proach for high-quality image reconstruc- 81. Lell M, Wildberger J, Heuschmid M, et al. 96. Hofmann A, Zettinig G, Wachter S, Kurtaran
tion in helical cone beam CT at arbitrary CT-angiography of the carotid artery: first A, Kainberger F, Dudczak R. Imaging of aortic
pitch. In: Sonka M, Hanson KM, eds. Pro- results with a novel 16-slice-spiral-CT scan-
prosthesis infection with a combined
ceedings of SPIE: medical imaging 2001— ner. Rofo Fortschr Geb Rontgenstr Neuen
SPECT/CT device. Eur J Nucl Med Mol Imag-
image processing. Vol 4322. Bellingham, Bildgeb Verfahr 2002; 174:1165–1169.
ing 2002; 29:836.
Wash: International Society for Optical En- 82. Ertl-Wagner B, Hoffmann RT, Brüning R,
97. Garcia-Ramirez JL, Mutic S, Dempsey JF, Low
gineering, 2001; 113–127. Dichgans M, Reiser MF. Supraaortale gefäss-
68. Flohr T, Stierstorfer K, Bruder H, Simon J, DA, Purdy JA. Performance evaluation of an
diagnostik mit dem 16-zeilen-multidetektor-
Polacin A, Schaller S. Image reconstruction spiral-CT: untersuchungsprotokoll und erste 85-cm-bore x-ray computed tomography
and image quality evaluation for a 16-slice erfahrungen. Radiologe 2002; 42:728 –732. scanner designed for radiation oncology and
CT scanner. Med Phys 2003; 30:832– 845. 83. Schoepf UJ, Becker CR, Hofmann LK, et al. comparison with current diagnostic CT scan-
69. Hsieh J, Toth TL, Simoni P, Grekowicz B, Multislice CT angiography. Eur Radiol 2003; ners. Int J Radiat Oncol Biol Phys 2002; 52:
Slack CC, Seidenschnur GE. A generalized 13:1946 –1961. 1123–1131.
helical reconstruction algorithm for multide- 84. Wintersperger B, Helmberger T, Herzog P, et 98. Nikolaou K, Flohr T, Stierstorfer K, Becker
tector row CT (abstr). Radiology 2001; al. New abdominal CT angiography protocol CR, Reiser MF. Flat panel computed tomog-
221(P):217. on a 16 detector-row CT scanner. Radiologe raphy of human ex vivo heart and bone
70. Hsieh J, Grekowicz B, Simoni P, et al. Con- 2002; 42:722–727. specimens: initial experience. Eur Radiol
volution reconstruction algorithm for multi- 85. Wintersperger B, Herzog P, Jakobs T, Reiser 2005; 15:329 –333.
slice helical CT. In: Sonka M, Fitzpatrick MJ, M, Becker C. Initial experience with the 99. Knollmann F, Pfoh A. Image in cardiovascu-
eds. Proceedings of SPIE: medical imaging clinical use of a 16 detector row CT system. lar medicine: coronary artery imaging with
2003—image processing. Vol 5032. Belling- Crit Rev Comput Tomogr 2002; 43:283– flat-panel computed tomography. Circula-
ham, Wash: International Society for Optical 316. tion 2003; 107:1209.
Engineering, 2003; 716 –723. 86. Nieman K, Cademartiri F, Lemos PA, Raaij-
71. Stierstorfer K, Flohr T, Bruder H. Segmented makers R, Pattynama PM, de Feyter PJ. Reli-

Volume 235 䡠 Number 3 Multi–Detector Row CT and Image Reconstruction 䡠 773

S-ar putea să vă placă și