Documente Academic
Documente Profesional
Documente Cultură
Lecturer
Medical Engineering Technology Section
UniKL BMI
Sem 1P 2015
Cardiac Defibrillators
Ventricular fibrillation is life-threatening
Ventricular fibrillation (v-fib for short) is the most
serious cardiac rhythm
1. AC Defibrillator
2. DC Defibrillator
Defibrillators
Before 1960 were AC model
1- Lown
2- Monopulse
4- Trapezoidal wave
Lown
- The current will rise very rapidly to about 20 A under the influence of
slightly less than 3 kV .
- The waveform then decays back to zero within 5ms and then produces
a smaller negative pulse also about 5ms.
Lown wave form defibrillator
Simple Block Diagram - Defibrillator
The energy stored in the capacitor is proportional to
the square of the voltage between its plates.
o The amount of energy typically stored in the capacitor
of a defibrillator, so that it can be later delivered to
the patient, ranges from 50 to 400 joules.
All of this energy does not get into the patient.
o Some is lost in the internal resistance of the defibrillator
circuit, RD and some is wasted in the paddle—skin
resistance, RE .
• To calculate how much of this
energy gets to the patient,
resistance RT, consider the
equivalent circuit.
– The four resistors in this
circuit are in series.
That is, the capacitor stores energy, WA, which develops a
voltage, V, across its metal plates.
V2
WA C
2
RT
WT WD
RD 2 RE RT
Monopulse is a modified Lown waveform and commonly
found in certain portable defibrillator. It is created by the same
circuit of Lown but without inductor L.
Tapered delay wave form , a lower amplitude 1.2 kV and longer
duration 15ms to a chive the energy level. It is created by
placing two L–C sections
Trapeziodal low voltage / long duration ( 800 V : 500 V & 20ms)
Placement of Electrode/Paddle
Waveforms and its importance
Monophasic or
Biphasic
Monophasic waveform.
Defibrillators with this type of waveform deliver current in one
polarity and were the first to be introduced. They can be
further categorized by the rate at which the current pulse
decreases to zero. If the monophasic waveform falls to zero
gradually, the term damped sinusoidal is used. If the waveform
falls instantaneously, the term truncated exponential is used
(figure 1).
The damped sinusoidal monophasic waveforms have been the
mainstay of external defibrillation for over three decades
Biphasic waveform
For the external defibrillation are usually metal discs about 3-5
cm in diameter and are attached to highly insulated handles.
Button
External Paddle
For internal defibrillation when the chest is open, large spoon-
shaped electrodes are used.
Electrodes used in
cardiac defibrillation (a)
A spoon-shaped internal
electrode that is applied
directly to the heart. (b)
A paddle-type electrode
that is applied against the
anterior chest wall.
Internal Paddle
Types of defibrillator
different shocks).
External Manual Defibrillator
2. Manual internal defibrillator
They are virtually identical to the external version, except that the
charge is delivered through internal paddles in direct contact with
the heart. These are almost exclusively found in operating
theatres, where the chest is likely to be open, or can be opened
quickly by a surgeon.
Manual internal defibrillator
3. Automated or Advisory external defibrillator (AED)
Here, the leads are placed in the standard position on the chest,
and the defibrillator paddles or adhesive electrodes are placed
appropriately.
The EGG from the patient is amplified by the EGG unit and
presented to the QRS detector.
Schematic Diagram of Cardioversion
This is the point in time that the heart normally depolarizes and
delivering the defibrillation pulse at that time should not cause
the heart to fibrillate.