Sunteți pe pagina 1din 5

Estimation of Blood Glucose by Non Invasive Technique

Anshuman Sahana, Dr. A. Rathinam, Harshvardhan Nagelia, Abhijeet Ghodgaonkar


Electrical and Electronics Engineering
SRM University, Chennai

Abstract – Diabetes is a metabolic pathological increased frequency of testing and tighter control of
condition, which affects important organs of the glucose concentrations. Near-IR (NIR) is one of
human body if not diagnosed and treated on the most widely recommended optical region
correct time. Nowadays glucose measurements because of its high penetration in skins.
methods are invasive which generally involves
puncturing finger. These methods are painful Measurement of blood glucose in a painless and
and has difficulties of spreading infectious non-invasive manner involves passing a selected
diseases. Therefore, there is need of developing a beam of wavelength of near-infrared light through
non-invasive monitoring system which measures the human tissue. The suitable NIR band of 940nm
blood glucose without any pain. The proposed wavelength for glucose absorption is identified.
method is based on the direct effect of glucose on
the scattering properties of the organ. Glucose
decreases the mismatch in refractive index
[2] METHODOLOGY
between scatters and their surrounding media, In order to overcome the difficulties of invasive
leading to a small scattering coefficient and, blood glucose measurements the concept of non-
consequently, a shorter optical path. The invasive blood glucose measurements arises. Here
reduction in scattering is due to an increase in the non-invasive blood glucose device is developed
glucose concentration. As a result, with the using the optical techniques. The optical technique
growing concentration of glucose, fewer photons used here is the NIR due to its high penetration in
are absorbed and the light intensity increases. In skin. When a light ray interacts with human body
the present work, we have used PPG technique. tissue, it gets attenuated by scattering as well as by
An algorithm was developed from the PPG data absorption by the tissues. Beer- Lambert Law plays
for monitoring blood glucose. a major role in absorbance measurement which
Index Terms – Non-Invasive, Blood Glucose, states the absorbance of light through any solution is
Photo plethysmograph. in proportion with the concentration of the solution
and length path travelled by the light rays.

A. Photoplethysmography (PPG)
[1] INTRODUCTION
Most of the soft tissues will transmit both infrared
Diabetes is emerged as a major healthcare problem
and visible radiation to some extent, even though
in India. Today approximately 8.32% of global adult
human body is blurred to light transmission.
population is suffering from diabetes. Diabetes
Amount of light is transmitted through whole hand,
mellitus is a medical concern in which the body is
if the hand is put-off the end of light in the dark
not producing the quantity or the quality of insulin
room. The transmitted light is absorbed and
needed to maintain the normal blood glucose in the
reflected, and scattered by the tissues and blood. By
body. Insulin is the hormone in the human body that
placing the light source usually in the range of
enables glucose to enter the body cells to get energy.
700nm to 2500nm on one side of the skin and
The technologies available in the market for blood detector on the other side. Plethysmograph measures
glucose measurement are invasive. Invasive changes in volume. The only method to measure
methods are painful, time consuming, expensive and absolute changes in blood volume accurately in the
also there is a potential risk of spreading disease like extremities by using chamber-plethysmography.
Hepatitis and HIV. The noninvasive techniques The volume change can be converted into blood
offer several advantages of no exposure to sharp flow by using F=dV/dt. This method is also called
objects and biohazard materials, the absence of pain, as transmission mode Photoplethysmograpy.
in transmitted light intensity in accordance with
glucose concentration can be achieved. The
developed setup utilises the attenuation of light at
different wavelengths where absorption of light
varies with the sugar content in the blood.

Figure 1. Transmission mode photoplethysmograph

Figure 3. Schematic description of the influence of


glucose on light propagation (a) with less glucose
(b) with more glucose in the blood

B. Principle
The basic principle behind the blood glucose
measurement is that glucose in the blood absorbs
infrared light many times more strongly than the
Figure 2. AC & DC components of PPG signal
remaining skin tissue. About 15% is absorbed by
The AC and DC components of PPG signal obtained epidermis. In the presence of dark light as
from a biological tissue (fingertip) is shown above. background, skin tissue without blood they appear
The amplitude of constant voltage offset of the DC as dark lines so that full blood vessel reflects
component can be determined by the nature of the approximately 10 times less than remaining tissue.
material through which the light passes. The AC or Due to higher wavelength of transmitted light (more
pulsatile component, which is synchronous with the than 940 nm) most of the photons is transmitted to
heart rate, is depending on the arterial blood volume the other side of the finger segment and detected by
pulse. The shape of the AC signal is indicative of the photo detector. Another part of the infrared light
cardiac performance and vessel compliance. The
is reflected by the tissues and remaining components
amplitude of AC component has 1-2% of the DC
value. By using non-invasive device like of the blood. The PPG signal has been obtained in
Photoplethysmograph (PPG), one can monitor the accordance with the pulsatile component of the
heartbeat by means of differences in light absorption arterial blood flow. But this waveform contains both
during systole and diastole of blood pumping. Non- dc component (due to tissue, vein, artery) and ac
invasive measurement of blood glucose is done by component (due to pulsatile nature of arterial blood).
focusing a beam of light on the finger and analysing For getting glucose level in blood, there is a need to
the light after transmission through the finger focus more on the ac component of received PPG
segment. The absorbance of light by skin is waveform. This can be done with the help of the
depended upon various components in the bloodlike equation
water, haemoglobin, fat and glucose. But
absorbance of light by these components strongly
depends on the wavelength of light which is
ΔODλ = log [1+ (ΔIλ(ti)/Iλ(ti+1))
transmitted. At a particular wavelength, the bond Where, ΔODλ is the optical density difference
between the atoms of components vibrates and
between time ti and ti+1. ΔIλ (ti) is pulsatile
absorbs the light. The glucose content of the blood
component at ti and Iλ(ti+1) is the intensity of light at
can be calculated by measuring the infrared light
transmitted through the thin part of body segment ti+1.
(usually fingertip). With less Glucose concentration, Light transport theory describes light attenuation as:
the light ray travels long optical paths and more
scattering of light occurs and results in less peak. In
the case of more glucose present in the blood vessel, I=I0-μeffL
light requires less optical paths due to less scattering.
This results higher peaks in the output. An increase
Where, I is the reflected light intensity, I0 is the alterations. To improve accuracy of the sensor finger
incident light intensity and L is the length of optical cap is made so that other lights will not enter and the
path inside the tissue. Attenuation of light inside designed sensor will give accurate result. This
thetissue depends on the coefficient known as voltage variation is so feeble and additional signal
effective attenuation coefficient (μeff) which is conditioning stages are necessary to convert it into a
defined as, recognizable form by microcontroller Arduino Uno

μeff = √3μa(μa+μs)
R3. This signal conditioning part will consist of
filtering stage to filter out noise and amplification
stage. Next Arduino Uno R3 will convert amplified
The absorption coefficient μa is described as the analog voltage to digital value. Regression analysis
probability of absorption of photons inside tissue per
unit path length and is given by: is also done by microcontroller and predicted
glucose value subsequently displayed on computer.

μa = 2.303 ε C

Where, ε is the molar extinction coefficient and C is


the concentration of glucose contained in blood.

Figure 5. Filter and Amplifier circuit

As shown in the figure 4.7, operational amplifier IC


LM 324 having single supply quad operational
amplifier is selected for the project as it has low
input offset voltage (3 mV) and high CMRR (80
Figure 4. Block diagram of designed model dB). Gain of 101 and cut off frequency of 2.5 Hz is
designed for filter. For amplification of the sensor
The block diagram given below represents the
signal, non-inverting amplifier is used; having
general procedure to get the estimated glucose value
required voltage gain is around 101 and suitable
on the computer. The infrared rays are emitted on
higher input resistance, by using formula for gain.
the fingertip with the help of IR LED and the
absorbed radiations are detected using Photo Voltage Gain Av = 1 + (R4/R5)
detector. The signal is then sent to the signal Normal pulse rate of human being is varied between
conditioning circuit where it is passed through the 60 to 180 pulses per minute. Therefore, considered
High Pass filter (HPF). Signals are filtered at 2.3 Hz. low pass filter of cut off frequency 1 Hz to 3 Hz is
Then the next stage is Non-Inverting Amplifier designed to remove high frequency components and
where the signals are amplified. The amplified 50 Hz power line interference. High pass filter of cut
signals are then given as input to the off frequency 0.5 Hz is used to remove the baseline
Microcontroller. With further calculations, a look up drift or low frequency signals and it is given by:
table was formed. Further regression analysis is
performed on the measured data which will be Cut off Frequency Fc = 1 / (2π*R4*C2)
helpful in forming a linear equation which gives us
the value of estimated Glucose on the computer. The second stage is non-inverting amplifier with the
gain of 101. Output voltage is connected for the data
The experimental circuit is set up using
acquisition by interfacing the computer by Arduino
near Infrared (NIR) spectral range to measure the
Uno R3. The circuit is also simulated using Proteus
blood glucose. The data recorded show differences
software before it is actually implemented on the
of voltage value related to their blood-glucose
breadboard.
[3] RESULT AND DISCUSSION equation was found out using regression analysis
method:
Constructed system determines a method for the y = 927.89x + 19.811, where x is the value of optical
prediction of Blood-Glucose level for human using density measured from the voltage value obtained
non-invasive methods. By using the voltage value as and y is the Glucose Concentration in mg/dl.
X in regression equation, glucose concentration can
be predicted. Dataset on sample basis is formed
using prick method and voltage value from the
prototype and a look up table is formed. This is used
in forming the equation of the best fit curve which is
then used to estimate the glucose concentration. The
regression analysis is carried out using NFTOOL in
MATLAB and also with the help of Ms-Excel. The
correlation coefficient R is found out. The value of
R close to 1 depicts a more accurate result and its
value close to 0 depicts less accurate result.
Figure 6. Regression analysis using MS Excel
Table 1. Lookup table for glucose concentration
The data obtained from intensity variation study
shows that there exists a correlation between the
S.No Max Min Glucose
variation in voltage intensity and glucose level. The
Peak Peak (mg/dl)
polynomial equation relation between voltage value
(V) (V)
and glucose concentrations, glucose concentration
1. 2.825 2.307 82 to be measured by this method are predicted in the
2. 2.835 2.297 86 above table. Results are tabulated and statistical
analysis is done for validation of this estimation of
3. 2.639 1.955 107 the glucose concentration.

4. 2.859 2.297 104


5. 2.776 2.366 76
6. 2.707 2.322 72
7. 2.835 2.297 83
8. 2.786 1.955 113
9. 3.226 2.199 133
Figure 7. Regression analysis using MATLAB ANN
10. 2.884 2.346 83
The voltages are found out with the help of
11. 2.835 2.199 107 Photoplethysmograph which is obtained with the
help of the Pulse Sensor (HRM-2511) put on the
12. 2.688 2.297 74 fingertip. The graph obtained is shown below.
13. 2.932 2.272 115
14. 3.201 2.297 125
15. 2.859 2.395 77
16. 3.030 2.356 101

By using above table, we can perform the


second order polynomial regression analysis. The
[4] CONCLUTION Communication, Artificial Intelligence in
Medicine 53 (2011) 127-138
Thus the prototype for the Non-Invasive
Measurement of Blood Glucose was designed and [5] D. X. Guo, Y.Z. Shang, R.Peng, S.S.Yong,
implemented using the reflection spectroscopic X.A.Wang, “Non-Invasive Blood Glucose
method. This device offers several advantages such Measurement Based on NIR Spectrums
as reduced pain while measuring Blood Glucose, and Double ANN Analysis”, Journal of
reduced risk of spreading of infection, Less Biosciences and Medicines, 2015, 3, 42-48
recurring cost over the conventional Invasive [6] G. D. Jindal, T.S. Ananthakrishnan, Rajesh
method for Blood Glucose Measurement. Kumar Jain, Vineet Sinha, A. R. Kini &
Determination of cardiac related pulsatile voltage Alaka K Deshpande, “Non-Invasive
changes and subsequently conducting regression Assessment of Blood Glucose by
analysis for the estimation of Blood glucose level Photoplethysmography”, IETE Journal of
were carried out successfully. In the prototype Research, Vol 54, No.3, May-June 2008,
developed, an IR LED of peak wavelength of 940 ISSN: 0377-2063
nm was chosen with fingertip as the body site for the
measurement of Blood Glucose. A low cost device [7] Vasanthadev Surya Kala S., Shankar K.
was designed and implemented on the printed circuit M., Shanthi Prince, “Blood Glucose
board and the results were validated using different Estimation by Non-Invasive Optical
statistical methods. The correlation coefficient Technique”, IJCTA, 9(15), 2016, PP.
between the cardiac related pulsatile voltage 7333-7338
intensity and the Glucose value obtained with the [8] Swathi Ramasahayam, Sri Haindavi K,
help of Prick method was found out to be 0.867. Bharat Kavala, Shubhajit Roy Chowdhury,
Therefore, we can conclude that there exists a “Non-Invasive Estimation of Blood
relationship between the predicted glucose and the Glucose using Near Infrared Spectroscopy
voltage value from the sensor. and Double Regression Analysis”, 2013
Seventh International Conference on
Sensing Technology, 978-1-4673-5221-
[5] REFERENCES 5/13, IEEE
[1] Komal Lawand, Shital Patil, Mahesh
Parihar, “Design and Development of
Infrared LED based Non-Invasive Blood
Glucometer”, IEEE INDICON 2015
1570186243, ISSN: 2325-9418
[2] Brince Paul, Melvin P Manuel, Zachariah
C Alex, “Design and Development of Non
Invasive Glucose Measurement System”,
ISPTS, 2012 1st International Symposium,
Pune, India, 7-10 March 2012
[3] Yamakoshi Kenichi, Yamakoshi Y, “Pulse
Glucometry: A new approach for Non
Invasive Blood Glucose Measurement
using Instantaneous Differential Near-
Infrared Spectrophotometry”, Journal of
Biomedical Optics, 11(5), 2006-09
[4] Enric Monte-Moreno, “Non-Invasive
Estimate of Blood Glucose and Blood
Pressure from a Photoplethysmograph by
means of Machine Learning Techniques”,
Department of Signal Theory and

S-ar putea să vă placă și