Documente Academic
Documente Profesional
Documente Cultură
Abstract – Diabetes is a metabolic pathological increased frequency of testing and tighter control of
condition, which affects important organs of the glucose concentrations. Near-IR (NIR) is one of
human body if not diagnosed and treated on the most widely recommended optical region
correct time. Nowadays glucose measurements because of its high penetration in skins.
methods are invasive which generally involves
puncturing finger. These methods are painful Measurement of blood glucose in a painless and
and has difficulties of spreading infectious non-invasive manner involves passing a selected
diseases. Therefore, there is need of developing a beam of wavelength of near-infrared light through
non-invasive monitoring system which measures the human tissue. The suitable NIR band of 940nm
blood glucose without any pain. The proposed wavelength for glucose absorption is identified.
method is based on the direct effect of glucose on
the scattering properties of the organ. Glucose
decreases the mismatch in refractive index
[2] METHODOLOGY
between scatters and their surrounding media, In order to overcome the difficulties of invasive
leading to a small scattering coefficient and, blood glucose measurements the concept of non-
consequently, a shorter optical path. The invasive blood glucose measurements arises. Here
reduction in scattering is due to an increase in the non-invasive blood glucose device is developed
glucose concentration. As a result, with the using the optical techniques. The optical technique
growing concentration of glucose, fewer photons used here is the NIR due to its high penetration in
are absorbed and the light intensity increases. In skin. When a light ray interacts with human body
the present work, we have used PPG technique. tissue, it gets attenuated by scattering as well as by
An algorithm was developed from the PPG data absorption by the tissues. Beer- Lambert Law plays
for monitoring blood glucose. a major role in absorbance measurement which
Index Terms – Non-Invasive, Blood Glucose, states the absorbance of light through any solution is
Photo plethysmograph. in proportion with the concentration of the solution
and length path travelled by the light rays.
A. Photoplethysmography (PPG)
[1] INTRODUCTION
Most of the soft tissues will transmit both infrared
Diabetes is emerged as a major healthcare problem
and visible radiation to some extent, even though
in India. Today approximately 8.32% of global adult
human body is blurred to light transmission.
population is suffering from diabetes. Diabetes
Amount of light is transmitted through whole hand,
mellitus is a medical concern in which the body is
if the hand is put-off the end of light in the dark
not producing the quantity or the quality of insulin
room. The transmitted light is absorbed and
needed to maintain the normal blood glucose in the
reflected, and scattered by the tissues and blood. By
body. Insulin is the hormone in the human body that
placing the light source usually in the range of
enables glucose to enter the body cells to get energy.
700nm to 2500nm on one side of the skin and
The technologies available in the market for blood detector on the other side. Plethysmograph measures
glucose measurement are invasive. Invasive changes in volume. The only method to measure
methods are painful, time consuming, expensive and absolute changes in blood volume accurately in the
also there is a potential risk of spreading disease like extremities by using chamber-plethysmography.
Hepatitis and HIV. The noninvasive techniques The volume change can be converted into blood
offer several advantages of no exposure to sharp flow by using F=dV/dt. This method is also called
objects and biohazard materials, the absence of pain, as transmission mode Photoplethysmograpy.
in transmitted light intensity in accordance with
glucose concentration can be achieved. The
developed setup utilises the attenuation of light at
different wavelengths where absorption of light
varies with the sugar content in the blood.
B. Principle
The basic principle behind the blood glucose
measurement is that glucose in the blood absorbs
infrared light many times more strongly than the
Figure 2. AC & DC components of PPG signal
remaining skin tissue. About 15% is absorbed by
The AC and DC components of PPG signal obtained epidermis. In the presence of dark light as
from a biological tissue (fingertip) is shown above. background, skin tissue without blood they appear
The amplitude of constant voltage offset of the DC as dark lines so that full blood vessel reflects
component can be determined by the nature of the approximately 10 times less than remaining tissue.
material through which the light passes. The AC or Due to higher wavelength of transmitted light (more
pulsatile component, which is synchronous with the than 940 nm) most of the photons is transmitted to
heart rate, is depending on the arterial blood volume the other side of the finger segment and detected by
pulse. The shape of the AC signal is indicative of the photo detector. Another part of the infrared light
cardiac performance and vessel compliance. The
is reflected by the tissues and remaining components
amplitude of AC component has 1-2% of the DC
value. By using non-invasive device like of the blood. The PPG signal has been obtained in
Photoplethysmograph (PPG), one can monitor the accordance with the pulsatile component of the
heartbeat by means of differences in light absorption arterial blood flow. But this waveform contains both
during systole and diastole of blood pumping. Non- dc component (due to tissue, vein, artery) and ac
invasive measurement of blood glucose is done by component (due to pulsatile nature of arterial blood).
focusing a beam of light on the finger and analysing For getting glucose level in blood, there is a need to
the light after transmission through the finger focus more on the ac component of received PPG
segment. The absorbance of light by skin is waveform. This can be done with the help of the
depended upon various components in the bloodlike equation
water, haemoglobin, fat and glucose. But
absorbance of light by these components strongly
depends on the wavelength of light which is
ΔODλ = log [1+ (ΔIλ(ti)/Iλ(ti+1))
transmitted. At a particular wavelength, the bond Where, ΔODλ is the optical density difference
between the atoms of components vibrates and
between time ti and ti+1. ΔIλ (ti) is pulsatile
absorbs the light. The glucose content of the blood
component at ti and Iλ(ti+1) is the intensity of light at
can be calculated by measuring the infrared light
transmitted through the thin part of body segment ti+1.
(usually fingertip). With less Glucose concentration, Light transport theory describes light attenuation as:
the light ray travels long optical paths and more
scattering of light occurs and results in less peak. In
the case of more glucose present in the blood vessel, I=I0-μeffL
light requires less optical paths due to less scattering.
This results higher peaks in the output. An increase
Where, I is the reflected light intensity, I0 is the alterations. To improve accuracy of the sensor finger
incident light intensity and L is the length of optical cap is made so that other lights will not enter and the
path inside the tissue. Attenuation of light inside designed sensor will give accurate result. This
thetissue depends on the coefficient known as voltage variation is so feeble and additional signal
effective attenuation coefficient (μeff) which is conditioning stages are necessary to convert it into a
defined as, recognizable form by microcontroller Arduino Uno
μeff = √3μa(μa+μs)
R3. This signal conditioning part will consist of
filtering stage to filter out noise and amplification
stage. Next Arduino Uno R3 will convert amplified
The absorption coefficient μa is described as the analog voltage to digital value. Regression analysis
probability of absorption of photons inside tissue per
unit path length and is given by: is also done by microcontroller and predicted
glucose value subsequently displayed on computer.
μa = 2.303 ε C