Auto Refractometer: Provides widest possible range of refraction and
keratometry measurements and objective testing capability. A fast and
reliable method to obtain the precise diagnostic data you need.
• If Pilot lamp does not light, • Plug power cable into • If fuse blows when
Check to see if power cable machine body. POWER switch is
is un-plugged. pressed on, contact
customer support.
• If Power Save function is on, return to normal status.
2. Printing issue.
• Paper comes out without printing: Check to see if the paper roll direction is correct.
• Paper does not come out: Is “PAPER END” displayed on monitor screen? If so, replenish paper.
Note: Always use Topcon paper (part #42330 9003B)
• When taking measurement, ensure patient is not blinking or moving. If the wrong measurement
still appears, objective lens may need to be cleaned.
KR-8000 Page 1 of 5
KR-8000
4. Instructions for cleaning the objective lens.
5. Change the tissue frequently as the procedure outline in step #4 is repeated. This will insure
complete cleaning.
KR-8000 Page 2 of 5
KR-8000
5. If measurement is still incorrect, check the following menu settings.
1. STEP = .25
2. VD = 13.75
To access the DIOPT. SHIFT menu turn on the Power Switch and simultaneously hold down the
Menu button until you hear the beep sound.
KR-8000 Page 3 of 5
KR-8000
6. Unit will not move side to side.
KR-8000 Page 4 of 5
KR-8000
6. Unit will not move side to side. (continued)
• Make sure that the track wheels line up and that the brass pin is situated over the slot, while
placing head back onto base.
• Re-install the track covers and plug electrical cord back in.
• Mode settings must be changed by pressing the “Mode” button on the bottom left
• KRT mode only = Green dot (door is up)
• Refractor or R/K mode = Farmhouse
KR-8000 Page 5 of 5