Documente Academic
Documente Profesional
Documente Cultură
Time Temperature
The specified time temperature curve is the same for both standards and
is based on the international standard ISO 834.
Measuring Devices
The UK Furnaces Use the 1.5mm Inconel sheathed thermocouple and the
American Standards uses the bare wired thermocouple sheathed in a
12mm metal tube.
Failure Criteria
Insulation
Integrity
Both standards use sustained flaming for a period of 10 seconds and the
cotton pad test.
In the BS test the specimens have been tested to positive pressure within
the furnace for over 30 years, whereas in the US positive pressure has
only been introduced in the last few years.
Direction of exposure
BS476 states that any specimen must be tested from its worse side, to
establish the maximum performance.
Additional testing requirement
There are no additional tests in the UK but it is usual to test a sample until
it fails both the insulation and integrity criteria. This enables the full fire
performance of the product to be established.
In the US, the hose stream test has been established as a way of
evaluating the robustness of the sample at the end of the fire test. It is not
intended to evaluate the any condition, which the sample will experience
in use ie in a fire fighting scenario.
Conclusions
Now that the USA test methods have adopted positive pressure, testing of
their products the testing regime between the two standards is similar.
However, there will be many products the testing regime between the two
standards is similar. Therefore, there will be many products, which have
been tested to the old negative pressure test (USA Test), so care will be
needed when reviewing the different tests as the UK test is the better.
The hose stream test is used in America, but not in any of the European
countries.