Operation of AP Monitor

Tsuyoshi Okamoto

1) Power on the Polymate II and connect it to the PC that is installed with AP Monitor.
2) Stat the AP Monitor.
3) Make and save the "Montage" based on your recording.
4) Set the event names on tabs below the window.
Ex.:
A) F1:Start
B) F2:Stop
C) F3:Break
D) F4:Restart
5) Fill in the Subject information.
6) Select your Montage.
7) Set the time bin to 10sec.
8) Click the IMPD CHECK.
9) Place the electrodes on the head for EEG, below eyes for EOG, near the wrists for ECG of
subject. See IMPD CHECK. The impedance of each electrode should be lower than 20kΩ.
10) Check the waveform of ECG.
11) Instruct the subject to close eyes and check the alpha wave of EEG.
12) When something is wrong with the waveforms, see Montage and IMPD CHECK and reset the
suspect electrodes (E, R, other electrodes).

13) Click the button of Start recording.


Confirm the message of "Start recording in Polymate".
Don't forget the operation!

14) Start the experiment.


15) After the experiment, click the button of Stop recording.