Sunteți pe pagina 1din 11

Scand J Clin Lab Invest 2000; 60: 677 ± 686

Validation of inductively coupled plasma atomic


emission spectrometry technique (ICP-AES) for
multi-element analysis of trace elements in
human serum
R . RA HI L -K HA ZE N, H . H EN R I KSE N, B . J . BO L ANN & R . J . UL VI K
Institute of Clinical Biochemistry, Haukeland University Hospital, Bergen, Norway

Rahil-Khazen R, Henriksen H, Bolann BJ, Ulvik RJ. Validation of inductively


coupled plasma atomic emission spectrometry technique (ICP-AES) for multi-
element analysis of trace elements in human serum. Scand J Clin Lab Invest
2000; 60: 677± 686.

The use of inductively coupled plasma atomic emission spectrometry (ICP-AES)


for the simultaneous determination of Al, B, Ba, Be, Cd, Co, Cr, Cu, Fe, Li,
Mn, Ni, Pb, Se, Sr and Zn in human serum in a clinical laboratory was
validated. Samples were digested and then analysed using yttrium as an internal
standard and a serum-matched calibration standard. The criteria used to assess
the analytical performance of the ICP-AES were detection and quanti® cation
limits, linearity, sensitivity, recovery, interference from alkali and acid, trueness
and precision. Detection limits were 0.002 ± 0.003 mmol/L for Mn, Sr, Ba, and
Cd; 0.014 ± 0.07 mmol/L for Co, Zn, Fe, Be, Li, Pb, Cu, Ni, and Cr; and 0.2 ±
0.9 mmol/L for B, Se, and Al. Trueness, as controlled by analysis of bovine
serum certi® ed reference material, was acceptable for Co, Cu, Se and Zn, while
Fe was 5.1% and Mn 6.2% below the lowest limit of the certi® ed material
interval. We conclude that ICP-AES can be used for multi-element analysis of B,
Ba, Cu, Fe, Li, Se, Sr and Zn in serum. Serum levels of Al, Be and Co were
below the detection limits while serum levels of Cd, Cr, Ni and Pb were below
the quanti® cation limits of the ICP-AES. These trace metals cannot be analysed
as routine by the ICP-AES. However, in cases of intoxication with elevated
serum concentrations mean recovery of 100¡10% was obtained at an addition
of 2.22 mmol/L for Al, 0.11 mmol/L for Be, 0.03 mmol/L for Co, 0.39 mmol/L for
Cr, 0.14 mmol/L for Ni, and 0.12 mmol/L for Pb.

Key words: Inductively coupled plasma atomic emission spectrometry; method


validation; trace element analysis; trace element serum levels

Rune J. Ulvik, Haukeland University Hospital, Institute of Clinical Biochemistry,


N-5021, Bergen, Norway. Tel. z47 5597 3149, fax. z47 5597 3115, e-mail.
rulv@haukeland.no

677
678 R. Rahil-Khazen et al.

I N TR OD UC T I ON instruments manufactured at the beginning of


the 1990s [4, 7, 8].
Up to now, atomic absorption spectrometry In addition to important, technical improve-
(AAS) has been the method of choice in clinical ments, the matrix problems and the effect of
laboratories for quanti® cation of trace elements using external and internal standards have been
in blood and biological tissues. However, there thoroughly addressed in the literature [9 ± 12].
has been an increased interest in evaluating the Up to now, ICP-AES has been primarily used
usefulness of atomic emission from high-energy in industrial rather than in medical laboratories.
sources like inductively coupled plasma (ICP). Considering the improved analytical perfor-
It remains to be seen what role this method will mance of modern instruments, ICP-AES may
attain in the laboratories in competition with be a potential alternative or supplement to AAS
AAS [1]. in clinical laboratories with interest in trace
Inductively coupled plasma (ICP) is an element analysis.
electrical conducting gaseous mixture composed Most of the published reports about analy-
of cations and electrons and is obtained when tical performance refer to the ICP-MS techni-
argon is converted to plasma in a high-energy que. We lack a recent and more complete
radio-frequency ® eld, giving rise to an excita- validation of the ICP-AES method useful for
tion temperature of 6500 ± 7000 K. ICP can be the clinical laboratory. Therefore, the aim of
used as a source for emission of speci® c spectral this study was to validate the ICP-AES method
lines from atoms and ions formed by metals to see if it may prove useful for analytical
which are introduced into the plasma in an purposes in a clinical chemical laboratory.
aerosol suspended in argon gas [2, 3]. The two
to three times higher temperature in the
analytical region of ICP gives a more complete
M AT ER I A LS AN D M ET H ODS
atomization and fewer chemical interferences
than in AAS.
Chemicals
In contrast to conventional ¯ ame or electro-
thermal AAS, the high-energy ICP-technique Multi-element standard solutions were pre-
allows simultaneous recording of a range of pared from Spectrascan single element stock
different spectra, which enables multi-element solutions (1000 mg/L) purchased from Tekno-
analysis of a single sample. Further, the linear lab, Drù bak, Norway. Calibration standards
concentration range is several-fold greater than were made by successive dilutions of the stock
for AAS. solution to obtain a ® nal concentration of
The two principal methods in current use for 2500 mg/L for Cu (39.3 mmol/L), Fe (44.8 mmol/
detection of spectra emitted from ICP, are mass L) and Zn (38.2 mmol/L); 500 mg/L for Se
spectrometry (MS) and atomic emission spec- (6.3 mmol/L); 300 mg/L for Al (11.1 mmol/L),
trometry (AES). ICP-MS is traditionally B (27.7 mmol/L) and Pb (1.4 mmol/L); 100 mg/L
regarded as a more versatile method than for Li (14.4 mmol/L), Ba (0.7 mmol/L), Be
ICP-AES. However, since ICP-MS is a much (11.1 mmol/L), Co (1.7 mmol/L), Mn
more complex and expensive technique than (1.8 mmol/L), Ni (1.7 mmol/L) and Sr
ICP-AES, it is of interest to ® nd out for which (1.1 mmol/L); and 50 mg/L for Cd (0.4 mmol/L)
analytical purposes ICP-AES may be an and Cr (1.0 mmol/L).
adequate alternative [4 ± 6]. As internal standard, yttrium was used at a
Since the introduction of the ® rst commercial concentration of 1000 mg/L (11.2 mmol/L). It
ICP-AES instruments at the beginning of the was measured at three different wavelengths
1970s, the technique has been signi® cantly 242.220, 324.228 and 417.754 nm to cover the
developed. The greatest progress has been wavelength range of the elements. Ultrapure
achieved in the last 4 ± 5 years by introducing nitric acid and suprapure hydrogen peroxide
the axial plasma torch, solid state detectors and were from Merck, Darmstadt, Germany.
the cyclone spray chamber to improve the Seronorm Trace Elements Serum (Nycomed
aerosol production. As a result, modern instru- AS, Oslo, Norway) was used as a commercial
ments have better long-term stability, higher control material.
resolution and lower detection limits than Bovine serum (SRM 1598) from the National
ICP-AES technique validation 679

Institute of Standards and Technology (NIST) Expansion Pty. Ltd., Australia was used. The
(Gaithersburg, USA) was used as a standard operating conditions are given in Table I.
reference material. Microwave digestion was carried out using
To imitate the concentrations found in serum, the Milestone microwave system (MLS 1200
suprapure sodium chloride, potassium chloride MEGA, Italy). The Milestone VAC-60 vacuum
and calcium carbonate from Merck, Darmstadt, module was used to evaporate the liquid in the
Germany and magnesium from Spectrascan samples.
were added to the calibration standards and
the blanks to give a ® nal concentration of
3200 mg/L (139.2 mmol/L) for sodium, 200 mg/
L (5.1 mmol/L) for potassium, 100 mg/L Microwave digestion and evaporation procedure
(2.5 mmol/L) for calcium and 20 mg/L Each microwave digestion procedure was
(0.8 mmol/L) for magnesium. carried out using six samples of 3 ml each. To
The acidity in the samples after the digestion each sample 2 ml nitric acid, 1 ml of hydrogen
and evaporation process was measured by peroxide and yttrium (11.2 mmol/L) was added.
titration against alkali with mixed indicators Although we used a sample volume of 3 ml
using the acidity test reagent kit Aquamerck 11 throughout this study, using 1.5 to 2 ml is also
108, Merck, Darmstadt, Germany. possible.
Laboratory glassware, plastic bottles, diges- After completion of the digestion procedure
tion vessels and covers were all soaked in 1.4 N using the program presented in Table II, the
solution of nitric acid overnight (or at least half vessels were cooled in cold water for 20 min.
an hour for the vessels and covers) and then They were then placed in the vacuum module
washed in de-ionized water supplied from a and evaporated (Table II) until a ® nal sample
Milli-Q system (Millipore Corporation, Bed- volume of 0.6 ± 1 ml was obtained. Thereafter,
ford, USA) with a resistance close to the content of each vessel was transferred to a
18 MVcm± 1. 12 ml polypropylene tube (Sarstedt, NuÈm-
brecht, Germany) and diluted with de-ionised
Instrumentation water to 3 ml before it was nebulized into the
ICP.
Trace element concentrations were measured Serum samples, control materials and the
using a programmable simultaneous plasma reference material were all treated in the same
emission spectrometer (IRIS/AP ICP-AES) way mentioned above.
from Thermo Jarell-Ash, Franklin, MA, USA The conditions described here and in Table II
with a ``charged injection device detector’’ and were the result of a systematic study where the
an axial viewing mode. A microconcentric ratio of sample, acid and peroxide, and the
nebulizer (Micromist nebulizer) from Glass microwave programs used in the digestion and
evaporation procedure, were modi® ed until the
sample appeared as a faint yellow, clear
TABLE I. Operating conditions.
solution.
Plasma frequency 27.12 Mhz
Rf Power 1150 watt
Torch gas ¯ ow 0.56 L/min
Auxiliary gas ¯ ow 0.5 L/min TABLE II. Microwave oven programs used.
Plasma viewing mode Axial
Nebulizer Micromist nebulizer Digestion program Evaporation program
Nebulizer gas ¯ ow 0.56 L/min
Spray chamber Cyclone Time Power Time Power
Nebulizer uptake 0.5 ml/min (min.) (W) (min.) (W)
Sample volume used 1.5 ml/2 repeats
Flush time 40 s 2 250 2 200
Purge time 90 s 1 0 3 300
Signal integration time 60 s High wavelength 5 250 7 400
10 s Low wavelength 5 400 6 450
Detector Charge-injection device 7 650
680 R. Rahil-Khazen et al.

Analysis of trace elements in human serum on the signal intensity of different elements, we
decided to use a calibration standard prepared
Serum samples were taken from human
in 4N HNO3.
serum pools gathered from patients in the
Fig. 1 shows how the signal intensity of the
hospital. The following 16 trace elements were
elements was suppressed in the presence of 4N
studied at the same time at their speci® c
HNO3 (St0). Taking the response of each
wavelengths given in Table III: Al, B, Ba, Be,
element prepared in 0.1N acid as 100%, a
Cd, Co, Cr, Cu, Fe, Li, Mn, Ni, Pb, Se, Sr, and
decrease in the intensities of the different
Zn. Some elements were studied at an addi-
analyte signals ranging from about 4% to
tional wavelength: Ba at 233.527, Cd at 226.502
more than 20% were seen in the St0 (black
and 228.802, Co at 228.616, Cu at 327.396, Fe
bars). Al, Li and Se were the elements most
at 239.562, Mn at 260.569, Sr at 421.552, and
affected by the acid.
Zn at 213.856 nm.
Seven different levels of trace elements in a
constant serum matrix were obtained by dilut- Use of matrix matched calibration standard
ing serum with added standard with the same
It is known that the alkali elements Na, K,
serum devoid of standard. The procedure was
Ca and Mg which are naturally present in
repeated on three different days. Seronorm was
serum, suppress the signal intensity of trace
used regularly as a control.
metals when analysed by the ICP-method [13 ±
16]. As shown in Fig. 1, when the multi-element
standard with 4N HNO3 (St0) was enriched
R ES UL TS
with Na, K, Ca and Mg (StCaNaMgK) at the
concentrations as described in Chemicals, the
Effect of acid on the signal intensity of the
signal intensity of all elements except Li, was
elements
depressed (dotted bars).
The multi-element calibration standard used
was originally prepared in 0.1N HNO3 acid.
Yttrium as internal standard
The measured acidity in the digested samples,
however, ranged between 3 and 5.5N. Based on By using yttrium as internal standard the
a thorough study of the effect of various relative intensity of the elements was obtained
concentrations of acid between 3 and 5.5 N by dividing their intensity at the respective

TABLE III. Detection limits and quanti® cation limits for the trace elements at their respective wavelengths.

Element Wavelength (nm) Detection Limits (mmol/L)a Quanti® cation Limits (mmol/L)b

Al 396.152 0.9 3.1


B 249.678 0.2 0.7
Ba 493.409 0.003 0.010
Be 313.107 0.024 0.080
Cd 214.438 0.003 0.010
Co 237.862 0.014 0.046
Cr 267.716 0.07 0.23
Cu 324.754 0.05 0.16
Fe 259.940 0.02 0.07
Li 670.784 0.04 0.12
Mn 257.610 0.002 0.005
Ni 231.604 0.05 0.16
Pb 220.353 0.038 0.126
Se 196.090 0.29 0.97
Sr 407.771 0.002 0.008
Zn 206.200 0.015 0.053
a
Calculated as 36SD of the concentration of a blank solution matched to the digested serum matrix with
respect to alkali elements and acid, and measu red 20 times.
b
Calculated as 106SD of the concentration of a blank solution matched to the digested serum matrix with
respect to alkali elements and acid, and measu red 20 times.
ICP-AES technique validation 681

Linearity
The multi-element calibration standard in 4N
HNO3 and containing Na, Ca, Mg and K (i.e.
StCaNaMgK) was tested for linearity up to a
concentration of 18.5 mmol/L for Al, 46.2 mmol/
L for B, 7.3 mmol/L for Ba, 11.1 mmol/L
for Be, 0.9 mmol/L for Cd, 1.7 mmol/L for
Co, 1.9 mmol/L for Cr, 78.7 mmol/L for
Cu, 89.5 mmol/L for Fe, 72.0 mmol/L for Li,
1.8 mmol/L for Mn, 1.7 mmol/L for Ni,
2.4 mmol/L for Pb, 6.3 mmol/L for Se,
1.1 mmol/L for Sr, and 76.5 mmol/L for Zn.
Linear regression analysis gave regression
coef® cient R240.998 for all the elements in
the concentration ranges mentioned above.
A series of seven serum samples with a
constant matrix, but different levels of elements,
was prepared by diluting a serum spiked with
elements with the same serum without added
elements. The concentration range of each
element is presented in Table IV.
The concentrations of the added elements
were plotted against the measured values.
Regression coef® cients (R2) between 0.973 and
0.999 were obtained, with the highest coef® -
FIG. 1. Intensity and relative intensity of the differ- cients 40.999 for B, Ba, Be, Co, Mn and Sr,
ent elements in St0 (a multi-element calibration stan- and the lowest values 0.973 for Cu (Table IV).
dard prepared in 4N HNO3) and StCaNaMgK (a
multi-element calibration standard prepared in 4N
HNO3 with added alkalis) taking 0.1N HNO3
multi-element standard as 100%. Intensities of St0 Trueness
and StCaNaMgK are shown as and respectively;
relative intensities to yttrium of St0 and The results of analysis of a human serum
StCaNaMgK are shown as and respectively. control (Seronorm) and a bovine serum stan-
dard reference material (NIST) with known
concentrations of trace elements are given in
wavelength by the intensity of yttrium. The
Table V.
relative intensity measured for some selected
Of the seven concentration levels tested, the
elements in the multi-element standards St0 and
minimum level for each element that gave a
StCaNaMgK is shown in Fig. 1 as grey bars and
mean recovery of 100¡10% or better of three
oblique striped bars respectively. The relative
repeats, is presented in Table VI. Similar results
intensity of each element prepared in the 0.1N
were obtained at higher concentration levels,
HNO3 multi-element standard was taken as but with even better precision.
100%. The inherent content of trace elements in
serum was calculated by the standard addition
method [17] and compared with measured
Detection and quanti® cation limits values (Table VI). In the standard addition
method, increasing concentrations of an added
The detection limits and the quanti® cation element were plotted against their relative
limits were calculated as 36SD and 106SD, intensities measured by the ICP-AES at the
respectively, of the concentration of a blank levels of addition. Linear regression analysis
solution matched to the digested serum matrix was performed, and by extrapolating the
with respect to alkali elements and acid, and regression line to the x-axis the inherent
measured 20 times (Table III). concentration of the element in the serum was
682 R. Rahil-Khazen et al.

TABLE IV. Linearity was tested by diluting serum samples supplemented with a multi-element standard with
increasing amounts of serum samples devoid of the standard. Linear regression analysis in the speci® ed con-
centration ranges measured on three different days with the slopes, intercepts and the regression coef® cient
(R2) is given.

Elem Concentration Range mmol/L R2 (Lowest value of 3 repeats) Slope (Range) Intercept (Range) mmol/L

Al 0± 5.56 0.997 1.04 ± 1.10 ± 0.02 ± 1.08


B 0± 13.87 0.999 1.05 ± 1.08 1.93 ± 2.05
Ba 0± 1.46 0.999 0.96 ± 1.00 0.51 ± 0.53
Be 0± 2.77 0.999 0.99 ± 1.04 ± 0.02 ± 0.00
Cd 0± 0.22 0.984 0.97 ± 1.00 0.00 ± 0.01
Co 0± 0.85 0.999 1.00 ± 1.04 0.00 ± 0.01
Cr 0± 0.96 0.994 0.95 ± 1.02 0.13 ± 0.23
Cu 0± 7.87 0.973 0.91 ± 1.00 20.60 ± 21.06
Fe 0± 8.95 0.996 1.02 ± 1.09 15.88 ± 16.48
Li 0± 7.20 0.995 0.99 ± 1.06 4.68 ± 4.94
Mn 0± 0.46 0.999 0.96 ± 1.00 0.02 ± 0.04
Ni 0± 0.85 0.980 0.83 ± 0.93 0.03 ± 0.10
Pb 0± 0.48 0.991 0.96 ± 0.96 ± 0.08 ± ± 0.04
Se 0± 3.8 0.996 1.03 ± 1.06 1.01 ± 1.03
Sr 0± 0.29 0.999 0.99 ± 1.01 0.27 ± 0.27
Zn 0± 7.65 0.987 0.97 ± 1.04 11.67 ± 12.42

calculated. A representative plot for the stan- Mn at a level three times above the quanti® ca-
dard addition procedure is presented in Fig. 2, tion limit, and to a lesser extent Li had
taking Mn as an example. relatively high CV. B, Ba, Cu, Fe, Sr and Zn
had CV55%, while Se had a CV510%.
The day-to-day precision was done by
Precision measuring 21 samples taken from the same
The data from the within-run and the day-to- serum pool, but treated separately and mea-
day precision studies showed as expected, that sured on 21 different days. Thus the variation
the coef® cient of variation (CV) was highest for re¯ ects both sample treatment and measure-
elements at concentrations close to their detec- ment. One outlier 43 SD from the mean was
tion limits, such as Cd and Cr (Table VII). Also removed.

TABLE V. Analysis of human serum control material and NIST bovine serum reference material.

Control material Reference material

Recommended Measured mean Certi® eda Measured


Element valuea (range) mmol/L (range)b mmol/L mmol/L mmol/L

Al 2.5 (2.1 ± 2.9) 2.5 (2.0 ± 2.8) 0.14¡0.03c ND


Cd ± ± 0.0008¡0.0001c ND
Co 0.001 ± 0.006c ND 0.022¡0.003 0.022
Cr 0.02 ± 0.03c ND 0.003¡0.002c ND
Cu 20.5 (18.7 ± 21.7) 20.2 (19.4 ± 21.5) 11.7¡0.6 12.1
Fe 19.7 (19.2 ± 19.9) 21.3 (20.2 ± 21.9) 46.98¡1.84 42.85
Mn 0.11 (0.09 ± 0.13) 0.09 (0.07 ± 0.13) 0.070¡0.006 0.060
Ni 0.04 ± 0.06c ND ± ±
Se 1.01 (0.89 ± 1.17) 1.01 (0.96 ± 1.09) 0.55¡0.05 0.52
Zn 22.6 (20.6 ± 25.5) 21.0 (20.0 ± 22.3) 14.0¡0.9 14.2
a
Determined mostly by Electrothermal Atomic Absorption Spectrometry and F lame Atomic Absorption
Spectrometry (see discussion).
b
M ean and range of 8 repeats, for Se mean and range for 6 repeats are given.
c
Values are below ICP-AES detection limits.
ICP-AES technique validation 683

TABLE VI. Pooled serum levels in (mmol/L) calculated by the standard addition method vs. those measured
and percentage recovery in spiked serum at the level of element addition selected. The mean and range of
three repeats done on three different days are given.

Serum level (mmol/L)

Standard addition Measured Mean Amount Added % Mean Recovery


Element method Mean (Range) (Range) mmol/L (Range)
a
Al (0.02 ± 1.26) 2.22 (101 ± 111)
B (1.93 ± 2.08) (2.13 ± 2.25) 1.39 (91 ± 99)
Ba (0.52 ± 0.55) (0.51 ± 0.52) 0.06 (99 ± 116)
a
Be (0.003 ± 0.019) 0.11 (100 ± 104)
Cd (0.004 ± 0.006) (0.003 ± 0.004)bc 0.01 (78 ± 97)
a
Co (0.002 ± 0.014) 0.03 (86 ± 97)
Cr (0.12 ± 0.26) (0.12 ± 0.20)c 0.39 (95 ± 103)
Cu (18.9 ± 22.6) (20.4 ± 21.1) 1.97 (88 ± 106)
Fe (15.4 ± 16.3) (15.8 ± 16.6) 0.90 (73 ± 116)
Li (4.4 ± 5.0) (4.7 ± 4.8) 2.88 (106 ± 113)
Mn (0.029 ± 0.046) (0.024 ± 0.044) 0.05 (95 ± 97)
Ni (0.002 ± 0.104) (0.019 ± 0.085)bc 0.14 (78 ± 105)
c
Pb (0.043 ± 0.082) 0.12 (92 ± 110)
Se (0.96 ± 1.06) (1.02 ± 1.15) 0.63 (78 ± 104)
Sr (0.269 ± 0.272) (0.269 ± 0.270) 0.03 (93 ± 110)
Zn (11.7 ± 12.8) (11.3 ± 12.4) 1.27 (106 ± 126)
a
Values are below ICP-AES detection limits.
b
Values are near ICP-AES detection limits.
c
Values are below ICP-AES quantiŽ cation limits.

D I SC US SI O N procedure is required when the concentration of


the elements to be analysed is so low that
A major problem with the ICP technique is the dilution will give a concentration below the
tendency of organic matrix constituents to detection limit of the instrument. Since we
interfere with nebulization and formation of wanted to validate the ICP-AES instrument for
aerosol from the sample material in the a range of trace metals including those at very
instrument. In general, these problems are low concentrations in serum, we had to perform
overcome by extensive dilution or acid digestion acid digestion of serum before analysis.
of the sample. The more cumbersome digestion First it was necessary to de® ne the optimal
conditions for digestion of serum and modi® ca-
tion of the calibration standard to match the
serum sample with respect to the content of
alkali elements and acid. The program for
digestion and evaporation presented in Table II
is the result of a systematic examination to ® nd
the optimal pre-analytical conditions. With
20 min cooling of the sample after digestion,
the pre-analytical procedure was completed in
about 1 h.
Residual inorganic matrix constituents in the
digested sample prone to give spectral inter-
ference in the ICP, were acid and the alkali
FIG. 2. Representative plot of the standard addi- elements Na, K, Ca and Mg inherent in serum.
tion procedure demonstrating the relative intensity
measured in the serum matrix. The ® gure shows the
Fig. 1 shows that the suppression of the signal
regression line and the extrapolation to the x-axis intensity caused by acid and alkali elements
giving the serum level as calculated by the standard could be suf® ciently compensated by the use of
addition method taking Mn as an example. yttrium as an internal standard for some
684 R. Rahil-Khazen et al.

TABLE VII. Within-run and day-to-day precision measured in the serum pool.

Element Within-run CV(%)a Day-to-day CV(%)b Mean serum pool level (mmol/L)

B 2.0 4.5 2.3


Ba 0.6 1.5 0.67
Cd 23.7 18.0 0.01d
Cr 7.8c 39.6 0.09d
Cu 0.6 2.3 19.68
Fe 0.5 1.9 17.85
Li 12.9 9.0 1.54
Mn 14.7 26.6 0.01
Se 9.3 9.1 1.62
Sr 2.3 4.6 0.27
Zn 0.3 1.3 14.5
a
Based on 20 repetitive measurements of a digested serum samp le.
b
Based on 20 measurements over 20-day period (one outlier 43SD was removed).
c
Based on 10 repetitive measu rements
d
Level below quantiŽ cation limit of the ICP-AES

elements, but not for all. The elements that were were measured at a level varying from 30 to 245
best compensated by the internal standard were times above the quanti® cation limit with Sr in
in falling order Zn4Se4Cd4Cr4Ba. The the lower and Fe in the upper end of the range.
remaining elements were not adequately com- The concentrations of B and Mn were measured
pensated by the internal standard, therefore the about three and seven times above the quanti-
use of the matrix matched calibration standard ® cation limit respectively, while the concentra-
was necessary. tion of Se was determined very close to its
Use of a matrix matched calibration standard quanti® cation limit.
excludes the simultaneous determination of the The linearity was satisfying for all elements
alkali elements Na, K, Ca and Mg. This is not a with slopes and detection limits that indicated
serious objection, however, since the alkali acceptable sensitivity. Further, recoveries of
elements are usually measured by other analy- 100¡10% were obtained by the addition of
tical techniques in the clinical laboratory. rather small amounts of metal (Table VI).
Our detection limits for Cu, Fe, Zn, Li, Mn Neither linearity nor recovery was affected by
and Sr (Table III) are generally lower than or the inherent serum concentration of trace
comparable to the values in other studies [18 ± metals, which differed substantially between
24]. For Mn especially, they are comparable to the metals. The analytical trueness was also
those of graphite furnace atomic absorption controlled by analysing the Seronorm human
spectrometry [25 ± 28] and lower than other serum control material with recommended
ICP-AES techniques [29]. values for some of the trace elements, and a
As demonstrated in Table VI, 12 out of 16 bovine serum from NIST with certi® ed values
trace elements could be simultaneously quanti- (Table V). These were the only commercial
® ed in serum above their detection limits, control and reference materials available.
covering a concentration range with a 5200 Except for Fe, none of the elements in the
fold difference between the lowest and highest control and reference material had been ana-
level represented by Cd and Cu, respectively. lysed by ICP-AES. The most frequent method
The serum concentration of Al, Be and Co as used was electrothermal atomic absorption
determined by the standard addition method, spectrometry (EAAS). Some elements were
was below the detection limit, while Pb was analysed by ¯ ame atomic absorption spectro-
close to the detection limit. The concentration metry (FAAS); and particularly in the NIST-
of Cd, Cr and Ni were measured 2 ± 3 fold material, elements were analysed by several
below the quanti® cation limit, at the level of the other methods like mass spectrometry and
detection limit for Cd and Ni, and about 2-fold neutron activation analysis in addition to
above the detection limit for Cr. The concen- AAS [30]. Although the elements in the control
trations in serum of Ba, Cu, Fe, Li, Sr and Zn and reference materials were analysed by other
ICP-AES technique validation 685

methods, there was an acceptable agreement Of the 16 elements that were studied, we
between the given intervals and measured values found that the ICP-AES can be used for multi-
(Table V). Co, Cu, Se and Zn were measured element analysis of the following elements in
within the given NIST-interval, while Fe was serum: B, Ba, Cu, Fe, Li, Se, Sr, and Zn. The
5.1% and Mn 6.2% below the lowest limit of the precision of the Mn analysis was not as good as
NIST interval. desired, but given the uncertainty of the
The within-run coef® cient of variation was analysis, also determination of Mn can be
between 10% and 15% for Mn and Li, and recommended.
about 24% for Cd. The other elements had The concentrations of Al, Be and Co in
variation coef® cients below 10% (Table VII). serum were below the detection limits and the
Cd and Mn had high day-to-day coef® cients of serum levels of Cd, Cr, Ni, and Pb were below
variation, but the highest value was found for the quanti® cation limits of the ICP-AES.
Cr. Particularly low variation coef® cients were Therefore, these trace metals can not be
obtained for Cu and Zn which are frequently analysed as routine by the ICP-AES. However,
analysed trace metals in medicine. Also, Se had in cases of intoxication with elevated serum
coef® cients of variation below 10% both within- concentrations above the quanti® cation limits,
run and from day-to day. In general, and not and by using the standard addition method, it is
unexpected the elements at the highest concen- even possible to get an idea of the serum levels
tration levels also had the lowest variation of these metals by using the ICP-AES.
coef® cients. Precision for Cu, Fe and Zn was
high in comparison to other studies [18, 22], but
lower than expected for Mn as compared to
A CK NO WL E DG EM E NT S
graphite furnace atomic absorption spectro-
metry [26, 27] at the low levels studied.
The authors thank Professor Rolf Isrenn and
By using the standard addition method (Fig.
Ms Torill Standal for providing technical help,
2) it is possible to estimate the concentration of
and the Section of Special Analysis in the
trace metals in serum at levels below the
Laboratory of Clinical Biochemistry, Hauke-
detection limit. This was done for Al, Be, Cd,
land University Hospital, for their assistance.
Co and Pb that were below or very close to the
detection limit. For the 9 elements which are
present in serum at concentrations above the
quanti® cation limit, there was a fairly good R EF E RE NC ES
correlation between the results obtained by the
standard addition method and measured values 1 Milne DB. Trace elements. In: Burtis CA, Ash-
for Ba, Cu, Fe, Li, Sr and Zn (Table VI). For B wood ER, editors. Tietz text book of clinical
chemistry, 3rd ed. Philadelphia: W. B. Saunders;
and Se measured values were on average about
1999. p. 1029 ± 55
8% and 7% higher than those estimated from 2 Houk RS. Mass spectrometry of inductively
the standard addition method; while for Mn the coupled plasmas. Anal Chem 1986; 58: 97A ±
measured value was 8% lower than the value 105A.
estimated from the standard addition method. 3 Schramel P. Atomic emission spectrometry. In:
Herber RFM, Stoeppler M, editors. Trace element
Thus, the standard addition method can be
analysis in biological specimens. Amsterdam:
used to analyse trace metals at levels below their Elsevier Science B. V.; 1994. p. 91 ± 127
detection limits. Also, it can be used to control 4 Broekaert JAC. Trends of development in plasma
results that appear between the detection and techniques for atomic spectrometric analysis: a
quanti® cation limits. review. Mikrochimic Acta 1995; 120: 21 ± 38.
5 Dean JR. Comparison of techniques. In: Ando
In this study we have validated the following DJ, editor. Atomic absorption and plasma spec-
criteria to assess the analytical performance and troscopy, 2nd ed. Chichester: John Wiley; 1997. p.
usefulness of the ICP-AES method: detection 155 ± 9
and quanti® cation limits, linearity, sensitivity, 6 Rommers P, Boumans P. ICP-AES versus (LA-)
recovery, interference from alkali and acid, ICP-MS: competition or a happy marriage? A
view supported by current data. Fresenius J Anal
trueness and precision. Acceptable speci® city Chem 1996; 355: 763 ± 70.
emerges from the selection of speci® c spectral 7 Marshall J, Chenery S, Evans EH, Fisher A.
lines for the elements. Atomic spectrometry update--atomic emission
686 R. Rahil-Khazen et al.

spectrometry. J Anal At Spectrom 1998; 13: by inductively coupled argon plasma atomic
107R ± 30R. emission spectrometry. Anal Chim Acta 1996;
8 Mermet JM, Poussel E. ICP emission spectro- 327: 301 ± 6.
meters: 1995 analytical ® gures of merit. Appl 20 Lim HB, Han MS, Lee KJ. Determination of trace
Spectrosc 1995; 49: 12A ± 8A. elements in human serum by inductively coupled
9 Chudinov EG, Ostroukhova II, Varvanina GV. plasma atomic emission spectrometry with ¯ ow
Acid effects in ICP-AES. Fresenius Z Anal Chem injection. Anal Chim Acta 1996; 320: 185 ± 9.
1989; 335: 25 ± 33. 21 McLeod CW, Worsfold PJ, Cox AG. Simulta-
10 De Boer JLM, Velterop M. Empirical procedure neous multi-element analysis of blood serum by
for the reduction of mixed-matrix effects in ¯ ow injection--inductively coupled plasma
inductively coupled plasma atomic emission spec- atomic--emission spectrometry. Analyst 1984;
trometry using an internal standard and propor- 109: 327 ± 32.
tional correction. Fresenius J Anal Chem 1996; 22 Melton LA, Tracy ML, Moller G. Screening trace
356: 362 ± 70. elements and electrolytes in serum by inductively-
11 Jason LI. Improving analytical precision by coupled plasma emission spectrometry. Clin Chem
utilizing intrinsic internal standards for determin- 1990; 36: 247 ± 50.
ing minor constituents by inductively coupled 23 Mauras Y, Allain P. Automatic determination of
plasma atomic emission spectrometry. J Anal At aluminum in biological samples by inductively
Spectrom 1996; 11: 683 ± 7. coupled plasma emission spectrometry. Anal
12 Marshall J, Rodgers G, Campbell WC. Myers -- Chem 1985; 57: 1706 ± 9.
Tracy signal compensation in inductively coupled 24 Nixon DE, Moyer TP, Johnson P, McCall JT,
plasma atomic emission spectromtery with high Ness AB, Fjerstad WH, Wehde MB. Routine
dissolved solids solutions. J Anal At Spectrom measurement of calcium, magnesium, copper, zinc,
1988; 3: 241 ± 4. and iron in urine and serum by inductively
13 Brenner I, Zischka M, Maichin B, Knapp G. Ca coupled plasma emission spectroscopy. Clin
and Na interference effects in an axially viewed Chem 1986; 32: 1660 ± 5.
ICP using low and high aerosol loadings. J Anal 25 Halls DJ. Manganese. In: Herber RFM, Stoeppler
At Spectrom 1998; 13: 1257 ± 64. M, editors. Trace element analysis in biological
14 Dubuisson C, Poussel E, Todoli JL, Mermet JM. specimens. Amsterdam: Elsevier Science B. V.;
Effect of sodium during the aerosol transport and 1994. p. 385 ± 401
® ltering in inductively coupled plasma atomic 26 Baruthio F, Guillard O, Arnaud J, Pierre F,
emission spectrometry. Spectrochim Acta Part B Zawislak R. Determination of manganese in
1998; 53: 593 ± 600. biological materials by electrothermal atomic
15 KovacÏic N, BudicÏ B, Hudnik V. Matrix effects absorption spectrometry: a review. Clin Chem
from magnesium and lithium in inductively 1988; 34: 227 ± 34.
coupled plasma atomic emission spectrometry. J 27 Subramanian KS, Meranger JC. Graphite furnace
Anal At Spectrom 1989; 4: 33 ± 7. atomic absorption spectrometry with nitric acid
16 Tripkovic MR, Holclajtner-Antunovic ID. Study deproteinization for determination of manganese
of the matrix effect of easily and non-easily in human plasma. Anal Chem 1985; 57: 2478 ± 81.
ionizable elements in an inductively coupled 28 Slavin W. Graphite furnace AAS for biological
argon plasma. J Anal At Spectrom 1993; 8: materials. Sci Total Environ 1988; 71: 17 ± 35.
349 ± 57. 29 BussieÁre L, Dumont J, Hubert J. Direct determin-
17 Dean JR. Methodology in trace element analysis. tation of calcium, copper, iron, magnesium,
In: Ando DJ, editor. Atomic absorption and manganese and zinc in amniotic ¯ uid samples
plasma spectroscopy, 2nd ed. Chichester: John using inductively coupled plasma-atomic emission
Wiley; 1997. p. 1 ± 20 spectrometry. Anal Chim Acta 1989; 224: 73 ± 81.
18 Chappuis P, Poupon J, Rousselet F. A sequential 30 National Institute of Standards & Technology.
and simple determination of zinc, copper and Certi® cate of analysis. Standard Reference Mate-
aluminium in blood samples by inductively rial 1598 inorganic constituents in bovine serum.
coupled plasma atomic emission spectrometry. Gaithersburg: NIST, 1990
Clin Chim Acta 1992; 206: 155 ± 65.
19 Le¯ on P, Plaquet R, Rose F, Hennon G, Ledeme
N. Rapid determination of lithium in human Received: 30 May 2000
serum and urine, at physiological concentrations, Accepted: 31 August 2000

S-ar putea să vă placă și