Purpose: The purpose of this study was to systematically review clinical studies on the incidence of
abutment screw loosening in single-implant restorations with different implant-abutment connection
geometries. Materials and Methods: The literature search was conducted using several electronic
databases. Specific terms were used for the database search, which spanned the years 1990 to
2006. The search was augmented by using the option of “related articles” as well as hand searching
of references and relevant journals. Relevant studies were selected according to predetermined inclu-
sion and exclusion criteria. Agreement between reviewers was determined by using Cohen’s kappa
coefficients. Three-year complication-free rates (survival proportions) were calculated with the aid of a
survival function, assuming constant failure rates. Summary estimates per group for complication-free
rate after 3 years (M-estimator) were calculated using Tukey’s biweight estimator. Results: The initial
database search yielded 1,526 relevant titles. After the subsequent filtering process, 27 studies were
finally selected. Interexaminer agreement ranged from good to perfect. The external-connection group
comprised 12 studies following 586 single-implant restorations for a mean follow-up time that ranged
from 3 to 5 years. The estimated percent of complication-free single-implant restorations after 3 years
was 97.3% (95% CI: 95.6–98.3). The internal connection group comprised 15 studies following 1,113
single-implant restorations for a mean follow-up time that ranged from 3 to 10 years. The estimated
percentage of complication-free single-implant restorations after 3 years was 97.6% (95% CI:
96.5–98.3). Conclusion: The results show that abutment screw loosening is a rare event in single-
implant restorations regardless of the geometry of implant-abutment connection, provided that proper
antirotational features and torque are employed. INT J ORAL MAXILLOFAC IMPLANTS 2008;23:681–690
Key words: implant complications, implant geometry, meta-analysis, screw loosening, single implants
he stability of the connection between different the reconstruction. This is especially true for single-
T implant parts is important for the overall success of tooth restorations, where a strong interlock between
the abutment and implant is necessary. Various mech-
anisms have been proposed to connect the dental
1Doctoral Candidate, Department of Fixed Prosthesis and Implant implant abutment to the implant body. Different sys-
Prosthodontics, School of Dentistry, Aristotle University of Thes- tems vary in connection geometry, materials, and over-
saloniki, Thessaloniki, Greece. all screw mechanics. The implant-abutment connec-
2Lecturer, Department of Fixed Prosthesis and Implant Prostho-
tion stability is also influenced by factors such as
dontics, School of Dentistry, Aristotle University of Thessaloniki, component fit, machining accuracy, saliva contamina-
Thessaloniki, Greece.
3Statistician, DCS Company, Thessaloniki, Greece. tion, and screw preload.1–8 Most in vitro studies,9–12
4Professor and Chair, Department of Fixed Prosthesis and with the exception of one,13 have demonstrated that
Implant Prosthodontics, School of Dentistry, Aristotle University internal connections are more mechanically stable
of Thessaloniki, Thessaloniki, Greece. than external flat (hex-type) connections. Loosening of
abutment screws, particularly with the “classic” exter-
Correspondence to: Dr Haralampos P. Petridis, Department of
Fixed Prosthesis and Implant Prosthodontics, School of Dentistry, nal-hex implant systems, has been a well-known tech-
Aristotle University of Thessaloniki, 54124 Thessaloniki, Greece. nical problem, occurring mainly during the first 2 years
Fax: +302310999575. E-mail: Lpetridi@dent.auth.gr after delivery of the crown.14–17 The stability of the
external implant-abutment connection was subse-
Presented as an oral presentation during the annual Interna-
tional Association for Dental Research-Continental European quently improved by altering the screw alloys and
Division (IADR-CED) meeting, which took place in Thessaloniki, their surfaces and applying proper torque values to
Greece, September 26–29, 2007. establish higher initial preloads.3,18–21
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 682
Theoharidou et al
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 683
Theoharidou et al
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 684
Theoharidou et al
All selected studies25–51 were published in the last able implant systems. Most of the external-connec-
12 years. The publication date of the external-con- tion implant systems included in the final studies
nection implant systems ranged from 1996 to 2004 were either 3i (Biomet 3i, Palm Beach Gardens, FL) or
with half of the studies published after 2000. The Nobel Biocare (Nobel Biocare, Göteborg, Sweden),
publication date of the internal-connection implant which essentially have similar geometry for their
systems ranged from 1998 to 2006, with the majority respective regular platforms. The exception was 1
of the studies published after 2000. No study directly study that utilized Threadlock/Spline (Sulzer Calcitek,
compared the incidence of screw loosening between Carlsbad, CA) implants. The internal-connection sys-
internal- and external-connection implant systems. tems were Straumann (Straumann, Basel, Switzer-
Most of the studies were classified as category C land), Astra (Astra Tech, Mölndal, Sweden), Frialit (Fri-
according to the strength of the evidence, and most atec, Mannheim, Germany), Mac System (Cabon,
were implemented in a university setting. Milan, Italy), Bicon (Bicon, Boston, MA), and Ankylos
The studies included a total of 2,038 patients with (Degussa Hulls, Hanau, Germany). Due to the possible
an age range of 13 to 90 years. The demographics of effect of the variability of internal-connection geom-
the included studies are depicted in Tables 3 and 4. etry on the stability of the implant-abutment inter-
The studies reported on various commercially avail- face, the internal connection group of studies was
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 685
Theoharidou et al
further categorized into 3 categories: “Straumann,” The design of the study (prospective versus retro-
“Astra,” and “Other.” The information on implants and spective) did not have any effect on the outcome.
SIR is depicted in Tables 5 and 6. The estimated percentage of complication-free SIR
The external-connection group comprised 12 after 3 years was as follows: external connection/
studies following 586 SIR for a mean follow-up time prospective 97.2% (95% CI: 94.9–98.5), external con-
that ranged from 3 to 5 years (Table 7). The estimated nection/retrospective 97.1% (95% CI: 93.6–98.7),
percentage of complication-free SIR after 3 years was internal connection/prospective 97.6% (95% CI:
97.3% (95% CI: 95.6–98.3). 95.9–98.6), and internal connection/retrospective
The internal-connection group comprised 15 stud- 97.6% (95% CI: 95.8–98.6).
ies following 1,113 SIR for a mean follow-up time that
ranged from 3 to 10 years (Table 8). The estimated
percentage of complication-free SIR after 3 years was DISCUSSION
97.6% (95% CI: 96.5–98.3). The complication-free per-
centage of SIR for the 3 groups was as follows: Strau- Systematic reviews differ from other types of reviews
mann 98.2% (95% CI: 96.8–99.0), Astra 97.0% (95% CI: in that they adhere to a strict scientific design to
94.4–98.4), and Other 97.1% (95% CI: 94.7–98.5). make them more comprehensive to minimize the
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 686
Theoharidou et al
Table 8 Abutment Screw Loosening Rates of Internal-Connection Implants (per Implant Type and Total)
Estimated Estimated
Total Mean No. of minimum failure rate Estimated
Type of no. of follow-up failure time (per 100 complication-free:
Study Year implant implants time (y) (loosen) exposure (y) implant years) % after 3 y
Behneke et al39 2000 Straumann 19 5.4 1 103 0.97 97.1
Brägger et al44 2005 Straumann 65 10 2 650 0.31 99.1
Duncan et al43 2003 Straumann 34 3 0 102 0.00 100
Mericske-Stern et al45 2001 Straumann 106 4.2 0 445 0.00 100
Nedir et al51 2006 Straumann 155 3 2 465 0.43 98.7
Levine et al37 1999 Straumann 157 3.3 5 518 0.97 97.1
Total 536 10 2,283
Summary estimate (95% CI) 0.59 (0.33–1.07) 98.2 (96.8–99.0)
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 687
Theoharidou et al
chance of bias and to ensure their reliability. Rather authors’ intention to try to include as many clinical
than reflecting the views of the authors or being studies as possible in order to reach a stronger con-
based on only a (possibly biased) selection of the clusion. Had the cutoff point for mean follow-up time
published literature, they contain a comprehensive been set at 5 years, 17 more studies would have been
summary of the available evidence with strict inclu- excluded. Furthermore, it has been shown that most
sion and exclusion criteria. Methodologies for under- of mechanical problems take place in the first years
taking systematic reviews have been described.87,88 after abutment connection.14–17 The mean follow-up
The gold standard for systematic reviews is to study time for the included studies ranged between from 3
randomized clinical trials (RCT), which are the studies to 10 years. However the results do not provide safe
with the most robust design. Most of the studies conclusions regarding the long-term stability of vari-
included in this review were retrospective. They were ous implant-abutment connections.
classified in category C according to the strength of The results of this review showed that abutment
evidence and clearly indicate a lower quality of data screw loosening is a rare event in SIR regardless of the
compared with prospective studies. 24 This clearly implant-abutment connection geometry, provided
points out the need for clinical studies with better that proper antirotational features and torque are
designs. No RCTs existed which directly compared employed. More than 97% of SIR studied maintained
the incidence of abutment screw loosening between a stable implant-abutment connection after 3 years
external and internal connection implant systems. of service. This is in agreement with results from in
The effect of study design (prospective versus retro- vitro studies,18–21 which have demonstrated stable
spective) was not significant; therefore, the results abutment screw joints for internal-connection
could be pooled together. implants as well as for external-connection implants
Two reviewers were used to ensure that tasks such with improved screw materials and preload. This
as selection of studies for inclusion and data extrac- study did not look into other types of mechanical
tion could be performed independently, increasing complications, such as screw fracture, where implant-
the chance that errors were detected. The interexam- abutment connection geometry might play a role.
iner agreement ranged from good to perfect agree- The results of this systematic review are in agreement
ment. Database search was also augmented by hand with the results of a previous systematic review92 that
searching to minimize selection bias.89 examined the incidence of biological and technical
The exclusion of papers in languages other than complications in implant therapy reported in
English may have led to the omission of some papers. prospective longitudinal studies of at least 5 years.
This is problematic for 2 reasons: (1) the precision of That study found a very low incidence of complica-
summary estimates could be reduced if a substantial tions affecting implant-connection components after
number of additional studies published in other lan- 5 years of clinical service. Another systematic review91
guages existed; (2) bias may have been introduced if of single implants stated that maintenance treat-
the results of studies published in English differed sys- ments were quite common with approximately 1 out
tematically from those published in other languages. of 5 single-tooth implants in need of maintenance,
However, a recent empirical study found little effect of varying from abutment screw retightening to crown
the inclusion/exclusion of trials published in language remake. The study did not look specifically at the inci-
other than English on combined effect estimates in dence of abutment screw loosening.
meta-analyses of RCTs.90 Moreover, it is difficult to have No attempt was made to distinguish between
access to non-English journals all over the world, and it anterior and posterior SIR, even though one could
is hard to establish the features of the peer-review assume that the vector of force is different. The rea-
processes of these journals. When these non-English sons were that most of the studies did not specify
papers are selected, based on their abstracts, the con- the number of posterior versus anterior SIR and that
tents must be translated. This includes the risk of inter- the incidence of abutment screw loosening reported
pretation problems.91 was extremely low. The same holds true for the possi-
The majority of the studies included in this review ble effect of restorative implant platform diameter
were conducted in a university setting. The results and method of crown retention. In the majority of
may therefore not be applicable to all practice set- included studies, implants with a regular-diameter
tings. The papers originating from private practice restorative platform and cement-retained restora-
settings showed a tendency for a slightly higher inci- tions were used. The method of reporting did not
dence of abutment screw loosening, especially permit the statistical analysis of the possible effect of
regarding internal-connection implant systems. the aforementioned clinical factors on the incidence
The follow-up time chosen was set at a minimum of abutment screw loosening.
of 3 years which could be considered small. It was the
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 688
Theoharidou et al
CONCLUSION 19. Drago CJ. A Clinical study of the efficacy of gold-tite square
abutment screws in cement-retained implant restorations. Int
J Oral Maxillofac Implants 2003;18:273–278.
The results show that abutment screw loosening is a
20. Tan KB, Nicholls JI. Implant-abutment screw joint preload of 7
rare event in SIR regardless of the geometry of hex-top abutment systems. Int J Oral Maxillofac Implants
implant-abutment connection, provided that proper 2001;16:367–377.
antirotational features and torque are employed. 21. Siamos G, Winkler S, Boberick KG. Relationship between
implant preload and screw loosening on implant-supported
prostheses. J Oral Implantol 2002;28:67–73.
22. Millennium Research Group. US markets for dental implants:
REFERENCES Executive summary. Implant Dent 2003;12:108–111.
23. Millennium Research Group. European markets for dental
1. Patterson EA, Johns RB. Theoretical analysis of the fatigue life implants and final abutments 2004: Executive summary.
of fixture screws in osseointegrated dental implants. Int J Oral Implant Dent 2004;13:193–196.
Maxillofac Implants 1992;7:26–34. 24. Jokstad A, Brägger U, Brunski JB, Carr A, Naert I, Wennerberg A.
2. Binon PP, McHugh JM. The effect of eliminating implant/abut- Quality of dental implants. Int Dent J 2003;53:409–443.
ment rotational misfit on screw joint stability. Int J Prostho- 25. Wannfors K, Smedberg J-I. A prospective clinical evaluation of
dont 1996;9:511–519. different single-tooth restorations designs on osseointe-
3. Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints grated implants. A 3-year follow-up of Brånemark Implants.
for single crowns supported by osseointegrated implants. Int Clin Oral Implants Res 1999;10:453–458.
J Oral Maxillofac Implants 1992;7:353–359. 26. Bambini F, Lo Muzio L, Procaccini M. Retrospective analysis of
4. Norton MR. Assessment of cold welding of the internal conical the influence of abutment structure design on the success of
interface of two commercially available implant systems. J implant unit. A 3-year controlled follow-up study. Clin Oral
Prosthet Dent 1999;81:159–166. Implants Res 2001;12:319–324.
5. Schwarz MS. Mechanical complications of dental implants. 27. Cho S-C, Paula N, Elian N, Tarnow D. Screw loosening for stan-
Clin Oral Implants Res 2000;11:156–158. dard and wide diameter implants in partially edentulous
6. Nikopoulou F. Saliva and dental implants. Implant Dent 2006; cases: 3- to 7-year longitudinal data. Implant Dent 2004;13:
15:372–374. 245–250.
7. Khraisat A, Abu-Hammad O, Al-Kayed AM, Dar-odeh N. Stabil- 28. Andersson B, Ödman P, Lindvall AM, Brånemark PI. Cemented
ity of the implant/abutment joint in a single-tooth external- single crowns on osseointegrated implants after 5 years:
hexagon implant system: Clinical and mechanical review. Clin Results from a prospective study on CeraOne. Int J Prostho-
Implant Dent Relat Res 2004;6:222–229. dont 1998;11:212–218.
8. McGlumphy EA. Implant screw mechanics. Dent Clin North 29. Parein AM, Eckert SE, Wollan PC, Keller EE. Implant reconstruc-
Am 1998;42:71–89. tion in the posterior mandible: A long-term prospective study.
9. Merz BR, Hunenbart S, Belser UC. Mechanics of the implant- J Prosthet Dent 1997;78:34–42.
abutment connection: An 8-degree taper compared to a butt- 30. Scholander S. A Retrospective evaluation of 259 single-tooth
joint connection. Int J Oral Maxillofac Implants 2000;15: replacements by the use of Brånemark implants. Int J Prostho-
519–526. dont 1999;12:483–491.
10. Balfour A, O’Brien GR. Comparative study of antirotational sin- 31. Scheller H, Urgell PJ, Kultje C, et al. A 5-year multicenter retro-
gle tooth abutments. J Prosthet Dent 1995;73:36–43. spective study on implant supported single crown restora-
11. Norton MR. An in-vitro evaluation of the strength of an inter- tions. Int J Oral Maxillofac Implants 1998;13:212–218.
nal conical interface compared to a butt joint interface in 32. Gibbard LL, Zarb GA. A 5-year prospective study of implant-
implant design. Clin Oral Implant Res 1997;8:290–298. supported single-tooth replacements. J Can Dent Assoc 2002;
12. Weiss EI, Kozak D, Gross MD. Effect of repeated closures on 68:110–116.
opening torque values in seven abutment-implant systems. J 33. Balshi TJ, Hernandez RE, Pryszlak MC, Rangert B. A comparative
Prosthet Dent 2000;84:194–199. study of one implant versus two replacing a single molar. Int J
13. Piermatti J, Yousef H, Luke A, Mahevich R, Weiner S. An in vitro Oral Maxillofac Implants 1996;11:372–378.
analysis of implant screw torque loss with external hex and 34. Andersson B, Taylor A, Lang BR, et al. Alumina ceramic implant
internal connection implant systems. Implant Dent 2006;15: abutments used for single-tooth replacement: A prospective
427–432. 1- to 3- year multicenter study. Int J Prosthodont 2001;14:
14. Jemt T, Pettersson P. A 3-year follow-up study on single 432–438.
implant treatment. J Dent 1993;21:203–208. 35. Vigolo P, Givani A, Majzoub Z, Cordioli G. Cemented versus
15. Ekfeldt A, Carlsson GE, Borjesson G. Clinical evaluation of sin- screw-retained implant-supported single-tooth crowns: A 4-
gle-tooth restorations supported by osseointegrated year prospective clinical study. Int J Oral Maxillofac Implants
implants: A retrospective study. Int J Oral Maxillofac Implants 2004;19:260–265.
1994;9:179–183. 36. Glauser R, Sailer I, Wohlwend A, Studer S, Schibli M, Scharer P.
16. Engquist B, Nilson H, Astrand P. Single-tooth replacement by Experimental zirconia abutments for implant-supported sin-
osseointegrated Branemark implants. A retrospective study of gle-tooth restorations in esthetically demanding regions: 4-
82 implants. Clin Oral Implants Res 1995;6:238–245. year results of a prospective clinical study. Int J Prosthodont
17. Henry PJ, Laney WR, Jemt T, et al. Osseointegrated implants for 2004;17:285–290.
single-tooth replacement: A prospective 5-year multicenter 37. Levine LR, Clem SD III, Wilson GT, Higginboth F, Solnit G. Multi-
study. Int J Oral Maxillofac Implants 1996;11:450–455. center retrospective analysis of the ITI implant system used
18. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment for single-tooth replacements: Results of loading for 2 or more
screw rotations and preloads for four different screw materials years. Int J Oral Maxillofac Implants 1999;14:516–520.
and surfaces. J Prosthet Dent 2001;86:24–32.
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 689
Theoharidou et al
38. Palmer RM, Palmer PJ, Smith BJ. A 5-year prospective study of 56. Lindquist LW, Carlsson GE, Jemt T. A prospective 15-year fol-
Astra single-tooth implants. Clin Oral Implants Res 2000;11: low-up study of mandibular prostheses supported by
179–182. osseointegrated implants. Clin Oral Implants Res 1996;7:
39. Behneke A, Behneke N, d’Hoedt B. The longitudinal clinical 329–336.
effectiveness of ITI solid screw implants in partially edentu- 57. Priest G. Single-tooth implants and their role in preserving
lous patients: A 5-year follow-up report. Int J Oral Maxillofac remaining teeth: A 10-year survival study. Int J Oral Maxillofac
Implants 2000;15:633–645. Implants 1999;14:181–188.
40. Norton M. Biologic and mechanical stability of single-tooth 58. Levine LR, Clem D, Beagle J, et al. Multicenter retrospective
implants: 4- to 7-year follow-up. Clin Implant Dent Relat Res analysis of the solid screw ITI implant for posterior single
2001;3:214–219. tooth replacements. Int J Oral Maxillofac Implants 2002;17:
41. Mangano C, Bartolucci GE. Single tooth replacement by morse 550–556.
taper connection implants: A retrospective study of 80 59. Schropp L, Kostopoulos L, Wenzel A, Isidor F. Clinical and radio-
implants. Int J Oral Maxillofac Implants 2001;16:675–680. graphic performance of delayed-immediate single-tooth
42. Krennmair G, Schmidinger S, Waldenberger O. Single tooth implant placement associated with peri-implant bone
replacement with the Frialit -2 system: A retrospective clinical defects. A 2-year prospective, controlled, randomized follow
analysis of 146 implants. Int J Oral Maxillofac Implants 2002; up report. J Clin Periodontol 2005;32:480–487.
17:78–85. 60. Andersson L, Emami-Kristiansen Z, Högström J. Single-tooth
43. Duncan PJ, Nazarova E, Vogiatzi T, Taylor DT. Prosthodontic implant treatment in the anterior region of the maxilla for
complications in a prospective clinical trial of single-stage treatment of tooth loss after trauma: A retrospective clinical
implants at 36 months. Int J Oral Maxillofac Implants 2003; and interview study. Dental Traumatology 2003;19:126–131.
18:561–565. 61. Vermylen K, Collaert B, Linde U, Bjorn A-L, De Bruyn H. Patient
44. Brägger U, Karoussis I, Persson R, Pjetursson B, Salvi G, Lang NP. satisfaction and quality of single-tooth restorations. Clin. Oral
Technical and biological complications/ failures with single Implants Res 2003;14:119–124.
crowns and fixed partial dentures on implants: A 10-year 62. Karlsson U, Gotfredsen J, Olsson C. Single-tooth replacement
prospective cohort study. Clin Oral Implants Res 2005;16: by osseointegrated Astra Tech dental implants: A 2-year
326–334. report. Int J Prosthodont 1997;10:318–324.
45. Mericske-Stern R, Crutter L, Rosch R, Mericske E. Clinical evalu- 63. Jemt T, Bergendal B, Arvidsson K, et al. Laser-welded titanium
ation and prosthetic complications of single tooth replace- frameworks supported by implants in the edentulous maxilla:
ments by non-submerged implants. Clin Oral Implants Res A 2-year prospective multicenter study. Int J Prosthodont
2001;12:309–318. 1998;11:551–557.
46. Muftu A, Chapman RJ. Replacing posterior teeth with free- 64. Wie H. Registration of localization, occlusion and occluding
standing implants: Four-year prosthodontic results of a materials for failing screw joints in the Brånemark implant sys-
prospective study. J Am Dent Assoc 1998;129:1097–1102. tem. Clin Oral Implants Res 1995;6:47–53.
47. Romanos EG, Nentwig HG. Single molar replacement with a 65. Lekholm U, van Steenberghe D, Herrmann I, et al. Osseointe-
progressive thread design implant system: A retrospective grated implants in the treatment of partially edentulous jaws:
clinical report. Int J Oral Maxillofac Implants 2000;15:831–836. A prospective 5-year multicenter study. Int J Oral Maxillofac
48. Norton MR. Multiple single-tooth implant restorations in the Implants 1994;9:627–635.
posterior jaws: Maintenance of marginal bone levels with ref- 66. Lekholm U, Gunne J, Henry P, et al. Survival of the Brånemark
erence to the implant-abutment microgap. Int J Oral Maxillo- Implant in partially edentulous jaws: A 10-year prospective
fac Implants 2006;21:777–784. multicenter study. Int J Oral Maxillofac Implants 1999;14:
49. Gotfredsen K. A 5-year prospective study of single-tooth 639–645.
replacements supported by the Astra Tech implant: A pilot 67. Jemt T, Henry P, Linden B, Naert I, Weber H, Bergström C. A
study. Clin Implant Dent Relat Res 2004;6:1–9. comparison of laser welded titanium and conventional cast
50. Wennström JL, Ekestubbe A, Grondahl K, Karlsson S, Lindhe J. frameworks supported by implants in the partially edentulous
Implant supported single-tooth restorations: A 5-year jaw: A 3-year prospective multicenter study. Int J Prosthodont
prospective study. J Clin Periodontol 2005;32:567–574. 2000;13:282–288.
51. Nedir R, Bischof M, Szmukler-Moncler S, Belser CU, Samson J. 68. Simon LR. Single implant-supported molar and premolar
Prosthetic complications with dental implants: From an up- crowns: A 10-year retrospective clinical report. J Prosthet Dent
to-8-year experience in private practice. Int J Oral Maxillofac 2003;90:517–521.
Implants 2006;21:919–928. 69. Kourtis GS, Sotiriadou S, Voliotis S, Challas A. Private practice
52. Levine LA, Clem DS III, Wilson TG, Higginbottom F, Saunders results of dental implants. Part I: Survival and evaluation of risk
SL. A Multicenter retrospective analysis of the ITI Implant sys- factors. Part II: Surgical and prosthetic complications. Implant
tem used for single tooth replacements: Preliminary results at Dent 2004;4:373–382.
6 or more months of loading. Int J Oral Maxillofac Implants 70. Haas R, Mensdorff-Pouilly N, Mailath G, Watzek G. Brånemark
1997;12:237–242. single tooth implants: A preliminary report of 76 implants. J
53. McMillan AS, Allen PF, Bin Ismail I. A retrospective multicenter Prosthet Dent 1995;73:274–279.
evaluation of single tooth implant experience at three centers 71. Polizzi G, Fabbro S, Furri M, Herrmann I, Squarzoni S. Clinical
in the United Kingdom. J Prosthet Dent 1998;79:410–414. application of narrow Brånemark system implants for single-
54. Dhanrajani PJ, Al Rafee MA. Single-tooth implant restorations: tooth restorations.Int J Oral Maxillofac Implants 1999;14:496–503.
A retrospective study. Implant Dent 2005;14:125–130. 72. De Boever AL, Keersmaekers K, Vanmaele G, Kerschbaum K,
55. Ericsson I, Nilson H, Lindh T, Nilner K, Randow K. Immediate Theuniers G, De Boever JA. Prosthetic complications in fixed
functional loading of Brånemark single tooth implants. An 18 endosseous implant-borne reconstructions after an observations
months’ clinical pilot follow-up study. Clin Oral Implants Res period of at least 40 months. J Oral Rehabil 2006;33: 833–839.
2000;11:26–33.
COPYRIGHT © 2007 BY QUINTESSENCE PUBLISHING CO, INC. PRINTING OF THIS DOCUMENT IS RESTRICTED TO PERSONAL USE ONLY. NO PART OF THIS
ARTICLE MAY BE REPRODUCED OR TRANSMITTED IN ANY FORM WITHOUT WRITTEN PERMISSION FROM THE PUBLISHER
Theoharidou.qxd 7/17/08 4:31 PM Page 690
Theoharidou et al
73. Scurria S M, Morgan ZV, Guckes DA, Li S, Koch G. Prognostic 83. Wennström JL, Ekestubbe A, Grondahl K, Karlsson S, Lindhe J.
variables associated with implant failure: A retrospective Oral rehabilitation with implant supported fixed partial den-
effectiveness study. Int J Oral Maxillofac Implants 1998;13: tures in periodontitis susceptible subjects. A 5-year prospec-
400–406. tive study. J Clin Periodontol 2004;31:713–724.
74. Bher M, Lang R, Leibrock A, Rosentritt M, Handel G. Complica- 84. Schwartz-Arad D, Samet N, Samet N. Single tooth replacement
tion rate with prosthodontic reconstructions on ITI and IMZ of missing molars: A retrospective study of 78 implants. J Peri-
dental implants. Clin Oral Implants Res 1998;9:51–58. odontol 1999;70:449–454.
75. Wee AG, McGlumphy EA. Prosthodontic complications of 85. Wyatt CC, Zarb GA. Treatment outcomes of patients with
Spline dental implants. Implant Dent 2003;12:151–159. implant-supported fixed partial prostheses. Int J Oral Maxillo-
76. Balshi TJ, Wolfinger GJ. Two-implant-supported single molar fac Implants 1998;13:204–211.
replacement: Interdental space requirements and comparison 86. Carlson B, Carlsson GE. Prosthodontic complications in
to alternative options. Int J Periodontics Restorative Dent osseointegrated dental implant treatment. Int J Oral Maxillo-
1997;17:426–435. fac Implants 1994;9:90–94.
77. Fartash B, Arvidson K. Long-term evaluation of single crystal 87. Higgins JPT, Green S (eds). Cochrane Handbook for Systematic
sapphire implants as abutments in fixed prosthodontics. Clin Reviews of Interventions 4.2.6 [updated September 2006]. In:
Oral Implants Res 1997;8:58–67. The Cochrane Library, Issue 4, 2006. Chichester, UK: John Wiley
78. Moberg L-E, Kondell P-A, Kullman L, Heimdalh A, Gynther WG. & Sons, 2006.
Evaluation of single tooth restorations on ITI dental implants. 88. Needleman IG. A guide to systematic reviews. J Clin Periodon-
A prospective study of 29 patients. Clin Oral Implants Res tol 2002;29(suppl 3):6–9.
1999;10:45–53. 89. Hopewell S, Clarke M, Lefebvre C, Scherer R. Hand searching
79. Weigl P. New prosthetic restorative features of the Ankylos versus electronic searching to identify reports of randomized
implant system. J Oral Implantol 2004;30:178–188. trials. Cochrane Database of Systematic Reviews 2007, Issue 2.
80. Döring K, Eisenmann E, Stiller M. Functional and esthetic con- Art. No.: MR000001.pub2
siderations for single-tooth Ankylos implant crowns: 8 years of 90. Egger M, Jüni P, Bartlett C, Holenstein F, Sterne J. How impor-
clinical performance. J Oral Implantol 2004;30:198–209. tant are comprehensive literature searches and the assess-
81. Cooper L, Felton AD, Kugelberg FC, et al. A multicenter 12- ment of trial quality in systematic reviews? Empirical study.
month evaluation of single-tooth implants restored 3 weeks Health Technol Assess 2003;7:1–76.
after 1-stage surgery. Int J Oral Maxillofac Implants 2001;16: 91. Creugers NHJ, Kreulen CM, Snoek PA, de Kanter RJAM. A sys-
182–192. tematic review of single-tooth restorations supported by
82. Norton MR. A short-term clinical evaluation of immediately implants. J Dent 2000;28:209–217.
restored maxillary TiO-blast single-tooth implants. Int J Oral 92. Berglundh T, Persson L, Klinge B. A systematic review of the
Maxillofac Implants 2004;19:274–281. incidence of biological and technical complications in implant
dentistry reported in prospective longitudinal studies of at
least 5 years. J Clin Periodontol 2002;29(suppl 3):197–212.