SPECIFICATIONS

for
Diagnostic Ultrasound System

MODEL: ALOKA LISENDO 880

1
LISENDO 880

Recently, Ultrasound Diagnostic Systems is needed for


aging society in cardiac application.
LISENDO 880 is used in critical care medical center.
Because it is available to adaptive for various request.

1. Pure Image
It is performed new optical system “Pure Symphonic
Architecture”. It is available to display high quality image.
It is available to receive high quality signal by using
optical sector probe for cardiology. And it is used new
Tx/Rx focusing method ”eFocusing” and then it is
available to display clear image from near field to far field.
In display, it is implemented OLED Monitor (high contrast
ratio), it is supplied cleared image for each medical case.

2. Seamless Workflow
It is needed to measure for each scene on cardiology. It
is implemented many kinds of automated system
(Measure, Search) through the examination.
It is expected to examine for short time by using
automated system.

3. Your Application
It is available to solve for any kinds of scene for
cardiology. Vector Flow Mapping is available to grasp
hemodynamics, Dual Gate Doppler & R-R navigation is
available to detect corrected time phase. And if it is not
available to detect regurgitation point, it is useful to use
3D imaging with matrix probe.

2
Scanning Method (Triplex mode)
Electronic Convex B(eFLOW) and D simultaneous real-time display
Electronic Linear (Triplex mode)
Electronic Phased Array Sector B and B(Color Flow) simultaneous real-time display
(Dual Flow)
Operating Modes B and B(Power Doppler) simultaneous real-time display
・ B-mode (Dual CF)
・ BiPlane-mode B and B(eFLOW) simultaneous real-time display (Dual
・ M-mode Flow)
・ D: Spectral Doppler mode (PW, CW, HPRF-PW) Dynamic Slow-motion Display
・ Dual Gate Doppler mode (Real-time image/Slow-motion image, side by side
・ Color Flow mode display)
・ Power Doppler mode (Directional Power Doppler) Panoramic View *
・ eFLOW mode (Directional eFLOW) TDI (Tissue Doppler Imaging)
・ Cardiac 3D mode BiPlane mode(Color)
(3D Zoom , Active 3D , Wide Angle 3D) 3D Zoom mode(Color)
Active 3D mode(Color)

Image Display Modes Wide Angle 3D mode(Color)

B: gray-scale imaging
* Option: SOP-LISENDO880-1
Dual B
Quad B
Beam former
M
Transmission
B and M
CPWG (Compound Pulse Wave Generator)
D: Spectral Doppler (PW, HPRF PW, and CW)
Programmable waveform transmission
B and D
Reception
B(Color Flow)
12-bit A/D converter (4096 gray levels)
B(Power Doppler)
A/D Sampling Frequency: 40[MHz]
B(eFLOW)
Parallel processing: Up to 20 directions
Dual B(Color Flow)
Transmission/Reception method of Harmonics
Quad B(Color Flow)
Reception frequency: 3.2-18.0 MHz*1
Dual B(Power Doppler)
*1 18MHz can be achieved under below conditions.
Quad B(Power Doppler)
・Probe: L64
Dual B(eFLOW)
・Frequency ”High”
Quad B(eFLOW)
・Frequency Information ”Receive”
M(Color Flow)
Transmission/Reception method of pulse modulation
M(Power Doppler)
– Chirp Pulse modulation
M(eFLOW)
Compound Broadband Pulse transmission
B(Color Flow) and M (Color Flow)
B(Power Doppler) and M(Power Doppler)
Tissue Adaptive Technology
B(eFLOW) and M (eFLOW)
Sound speed adjustment: 26 steps
B(Color Flow) and D
Automated sound speed adjustment
B(Color Flow) and Dual
B(Power Doppler) and D
Focusing
B(Power Doppler) and Dual
Transmission: Multi-stage transmission focus of up
B(eFLOW) and D
to 4 stages out of 16 stages (probe dependent)
B(eFLOW) and Dual
Reception: PixelFocus
B(Color Flow) and D simultaneous real-time display
eFocusing
(Triplex mode)
B(Power Doppler) and D simultaneous real-time display
3
System Dynamic Range mode but B mode.
320dB Adaptive Imaging
: HI REZ function 8 steps / HI REZ Plus
System Processing Channels Adaptive imaging is the technique to generate optimum
7,072,000 Channels image with changing filter characteristic depending on
input signal.
Frame rate This technique which installed this ultrasound diagnostic
Max. More than 882 frames/sec* scanner reduces speckle noise on B-mode images and
* Depends on probes and various settings provides uniformly and high contrast images.
* Gain and Dynamic range can be changed after freezing
B-mode
Border Clear Filter
Display Gray Scale: 256 levels This technique is expanded gray scale so it is reduced
Scanning area: noise and emphasized border.
100% to 25%, 1 degree step Acoustic Noise Reduction : 8 steps for 2 types.
Line Density : 8 steps This technique is reduced acoustic noise
HI Frame Rate(Multi Parallel Processing) (Based on signal level)
Zoom Low Echo Reduction
HI zoom (real-time image): This technique is compressed low gradient echo level.
Max. Up to 0.5 cm display image Auto-optimizer (Optimization of gain, sound velocity
The maximum magnification is 8. adjustment)
PAN zoom (real time and frozen image):
Max. Up to 0.5 cm display image M-mode:
The maximum magnification is 8. Sweep method: Moving bar
Depth range selections: Sweep speed:
0.75/1.0/1.5/2.0/2.5/3.0/3.5/4.0/4.5/5.0, 1 cm
300.0, 200.0, 133.3, 100.0, 66.7, 50, 40.0 mm/sec
intervals from 5.0 to 24 cm , and 2cm intervals from
Gain: B-gain ±30 dB
24cm to 40cm (probe dependent)
Longitudinal and lateral inversion Dynamic Range: 40 dB-90 dB

Rotation by 90 degrees (probe dependent) AGC: 8 steps

Gain*: 80 dB range Acoustic Noise Reduction

Echo enhancement: 8steps Low Echo Reduction

Texture: 2 steps(Smooth/Sharp), not available in FAM (Free Angular M-mode)

M-mode
Enhance / Smoothing : 8 steps for each other Spectral Doppler:
Persistence: 8steps, including Off Display: Power spectrum

(Available to change type for adaptive frame rate) Real-time Doppler Auto Trace

TGC (time gain control) : Doppler methods:

Gain versus depth curve control: 8 slide controls – PW (Pulsed Wave) Doppler

TGC curve memory function – HPRF (High Pulse Repetition Frequency)


LGC (lateral gain control): PW Doppler
Gain versus angle curve control: 8 sectors (Sector – Dual Gate Doppler
probes only) – CW (Continuous Wave) Doppler *
Dynamic range: 40-90 dB Reference frequencies (probe dependent):
Gray map : 10 types, changeable after frozen Maximum 3 frequencies
AGC: 8 steps PW: 1.5, 1.9, 2.0, 2.1, 2.5, 3.0, 3.3, 3.5, 3.8,
View gamma: 4types 4.0, 4.4, 5.2, 6.0, 6.3 MHz
Spatial Compound Imaging CW: 1.7, 1.8, 2.0, 2.1, 2.5, 3.0, 3.3, 3.8, 5.0 MHz
(by selected linear and convex sector probes)
Analysis rate:
Trapezoidal scan (possible by selected linear probes)
PW: 0.05 to 40 kHz
B steer function : This function enables you to use
oblique function not only Color Flow/ Power Doppler CW: 1.1 to 40 kHz

4
Max. velocity range: ±127 levels for velocity (red and blue)
PW/HPRF:±1.26 cm/sec to ±802.08 cm/sec 64 levels for variance (green)
CW:±25.07 cm/sec to ±802.08 cm/sec – Color Inversion: Normal, Invert
Base line shift: Adjustable in real-time and frozen – Smoothing: 5 steps, adjustable in real-time and
Steerable CW Doppler: probe dependent frozen
Steered linear scanning: Max. ±30 degrees – Wall filter: 6 steps
changeable at 5 degrees interval – Persistence (Color) : 8 steps
Auto angle Correction – Wall Motion Reduction: Off + 3 steps
Spectrum inversion – Packet Size : 3 levels
Angle correction: Available up to 80 degrees – Base line shift(Color):
(Pre-settable, changeable after freezing) Up to double velocity (±127 steps)
Auto angle correction, Warning function available – Color coding: 15 kinds
Sample volume size for PW Doppler: – TGC Enhancement : 2 kinds
0.5 – 20 mm, changeable in 0.5mm, 1.0 mm step ・Power Doppler Mode
Wall motion filter: 12steps, 1/16 of PRF is Max. – Display patterns: Power Doppler, Directional
Doppler gain: 60dB variable, adjustable in real-time and Power Doppler
frozen.
– Gradation: 256 levels
Echo Enhancement(CW) : 3 steps
– Color coding 15 kinds
Dynamic Range: 40-90dB 1dB steps
– Non-display of B/W image: Possible in ROI
(Changeable after freezing)
– Smoothing: 5 levels
Auto-Optimizer
・ High Resolution Power Doppler(eFLOW) Mode
(Optimization of gain, velocity range and baseline
One of the Color Flow imaging functions that can
shift )
display blood flow information in a high spatial and
Audio output: 2 channels
temporal resolution.
– Display patterns:
Color Doppler Mode eFLOW, Directional eFLOW
Color area size: Variable from 100% to 25%
– Pulse repetition frequency:
continuously
0.03 to 19.8 kHz
・ Steered linear scanning: – Gradation: 256 levels (±127 levels for directional)
Max. ±30 degrees *, changeable at 5 degrees interval – Color coding: 15 kinds
* Probe dependent
– Non-display of B/W image: Possible in ROI
Line density: Up to 8 steps – Advanced wall motion reduction
(Changeable setting independently with B mode) – Smoothing: 5 levels
HI Frame Rate(Color)(Multi Parallel Processing) TDI(Tissue Doppler Imaging)
Auto-Optimizer (Optimization of gain) – Function Available

・ Color Flow Mode Manual


– Display patterns: It is applicable to browse instruction manual on this
Velocity (derived from mean Doppler frequency system.
shift), Velocity + variance, Variance ・ On-board operation manual

– Max. velocity range:±0.63cm/sec to


Cine Memory
±458.33cm/sec
・ Cine search and loop display (in B mode):
– Reference frequency: (Probe dependent)
ECG time phase display is possible
1.9, 2.0, 2.1, 2.5, 3.2, 3.3, 3.5, 4.0, 4.4, 5.2,
・ Capacity
6.0, 6.3 MHz
B mode: Max. 63,500 frames.
– Pulse repetition frequency:
M and D modes: Max. approx. 900 seconds.
0.03 to 19.8 kHz
– Gradation: Data Management

5
1. Image data Procedure ID, Accession, Study ID, Study
1-1. Format Description Referring physician, Reporting Phys,
Multiple-frame (moving) image Sonographer
DICOM (Raw, MJPEG) * Conforms to DICOM 3.0 standard
PC Format(AVI) (AVI CODEC, WMV, MJPEG, 4. Data storage
MSMP4V2) ・ Main unit hard disk
Single-frame (still) image
Capacity: Approx. 500GB
DICOM (Non-compressed, RLE, RGB(Plane/Pixel),
* About 72,000 still images are storable.
JPEG)
PC Format (Tiff, Bmp, JPEG) ・ USB memory

1-2. Image acquisition mode ・ USB HDD

・ Real-time multi-frame image acquisition (Raw, Image) ・ CD-R


Aquiring both RAW and Image at the same time ・ DVD-RAM
Post ECG: Max. 10cardiac cycles (R-R) ・ Network interface(DICOM format): 1Gbps
Pre ECG: Max. 10 cardiac cycles (R-R) ,BASE/TX, (automatically switched)
Post Time: Max. 90 seconds 5. DICOM network communication*1
Pre Time: Max. 16 seconds Conformity to DICOM service class:
Manual: Ultrasound image storage SCU
Raw data: Max. 150 seconds Ultrasound multi-image storage SCU
Image data: Max. 180 seconds Storage media FSC/FSR
・ Cine loop high-speed data transfer (Raw, Image) Print management SCU
It is possible to selectively store data of arbitrary Modality worklist management SCU
section in the Cine Memory. (For details, please refer to the DICOM Conformance
Statement issued by Hitachi, Ltd.)
・ Simultaneous output to multiple media
Modality performed procedure step (MPPS) SCU
It is possible to output still image data to multiple of
Storage: Possible to store patient information directly to
storage media include network and printers at the DICOM file server
touch of a button. Print: Possible to printout images with DICOM
1-3 Image data management tool compatible printer directly
Image viewer Work list management: Retrieval of patient and
・Compatible with DICOM and PC-format images reservation information from hospital information
・ Simultaneous display of stored and real-time system (HIS)
images is possible (Compare mode) NOTE: The HIS needs to be compatible with DICOM standard
・Thumbnail display of stored images (1-36 images) supplement 10. The HIS network and the DICOM network
・Check mark is put on a transferred image need to be linked.
・Image zoom, rotation, inversion Router setting: possible
・1:1 replay (main unit HDD or DICOM storage data) Compatible with SR (Structured Report) for OB,
・ DVD-RAM cardiology, vascular and abdominal measurements*2
・ CD-R Query/Retrieve*3
・ USB memory Compatible with ED(Evidence Documents)
・ USB HDD Profile
・Re-storing to media, transfer IHE (Integrated Healthcare Enterprise)
・Adjustment is possible on the reconstructed SWF (Scheduled Work Flow)
Raw-data image (gain, dynamic range, gamma PIR (Patient Information Reconciliation)
curve type, and color coding in Color flow mapping *1 Option: SOP-LISENDO880-10
mode) *2 Option: SOP-LISENDO880-21
2. Measurement data *3 Option: SOP-LISNEDO880-59
It is possible to store measurement data in the main unit 6. Security measures
hard disk User authentication function is available.
3. Patient data
3 Types of user authority can be set.
Displayed information*
It is possible to set whether password is required or
Patient information
ID (up to 64 characters), Name (up to 64 not at the start of operation.
characters, including middle name), Birthday, Sex, 7.Teaching file can be made.
Age, Height, Weight, Occupation
Study information
6
Measurements and Analysis: Coronary flow
TDI PW
・ Basic measurements
B(Flow)/D mode
On B-mode image
PISA measurements
Distance, Dist-trace, Area/Circum, Volume, Hip J Angle,
B TDI mode
Histogram, Angle, B.Index
Asynchrony analysis for CRT
On M-mode image
Length, Time, Heart Rate, M.VEL, M.Index ・ Vascular analysis
On spectral Doppler Carotid artery:
CCA (common carotid artery)
D.VEL, ACCEL, RI, Time, P1/2T, Heart Rate, D.Caliper,
ICA (internal carotid artery)
D.Index (Caliper), D.Index (Trace), Mean.VEL., PI, ECA (external carotid artery)
D.Trace, Steno Flow, Regurg Flow, Real-time Doppler BIFUR (Bifurcation of carotid artery)
auto trace VERT (Vertebral artery)
On B/D mode % Stenosis area
% Stenosis diameter
Blood Flow
IMT (Intima-media thickness) Measurements
On B(Flow) mode
Automated IMT measurements*
Flow Profile*
* Option: SOP-ARIETTA70-38
* Option: SOP-LISENDO880-7
Measurements of arteries in extremities:
Lower extremity artery flow
Application measurements
Upper extremity artery flow
・ Cardiac analysis
Stenotic rate:
B mode
% Stenosis area
LV Volume measurements
% Stenosis diameter
Area-length, BP-ellipse, Simpson (Disc), Modified
Transit time of Vessel Flow measurements(TVF)
Simpson*, Bullet, Pombo**, Teichholz**, Gibson**
* Option: SOP-ARIETTA70-47
*: Automatic heart cavity trace is possible.
Measurements of veins in extremities:
(3-point or full automated method)
Lower extremity venous flow
**: Automatic measurement is possible.
Upper extremity venous flow
Valve area measurements (AVA, MVA)
Trans-cranial blood flow measurements
LA/AO
Ratio
・Report Functions
Right ventricle measurements
– Cardiac function report
LV myocardial mass
※It is possible to add below function for report.
LA / RA Volume measurements
・It is possible to edit measurement values by using
Area-length, Simpson*
Worksheet.
*: Automatic heart cavity trace is possible.
・It is possible to input comment automatically.
(3-point or full automated method)
・It is possible to convert PDF.
FAC measurements*
SOP-LISENDO880-70 is required.
*: Automatic heart cavity trace is possible. (3-point)
– Vascular report
IVC (inferior vena cava) measurements
– IMT (Intima-Media Thickness) report
M mode
Pombo (wall), Teichholz (wall), Gibson (wall)
It is possible to recall past measurement reports.
Mitral valve measurements
Examination data history can be plotted on the report.
LA/AO measurements*
Direct printout of each report is possible with an optional PC
*: Automatic measurement is possible.
printer.
Tricuspid valve measurements
Output of measurement values in CSV file is possible.
Pulmonary valve measurements
IVC (inferior vena cava) measurements ・ Hot Key function: It is possible to assign measuring
TAPSE measurements functions to the alphabet keys on the keyboard
Doppler mode ・ Measurements on VCR playback image: Possible
LVOT (left ventricle outflow tract) flow (manual calibration)
RVOT (right ventricle outflow tract) flow ・ User’s calculation
Trans-mitral flow
30 equations can be set for each application
Regurgitant flow (AR, PR, MR, TR)
User-assignable terms: 60 words possible
Volumetric flow(MR)
Stenotic flow (AS, PS, MS, TS) ・ Font size of measurement result: Possible to change in 3
Pulmonary vein flow kinds

7
Physiological Signal Display
Displayed information: ECG, PCG*1, Pulse *2wave,
breathing waveform
ECG synchronized display: Available for one phase
Detect regular pulse from arrhythmia(RRp/RRpp)
Display position: Continuously variable (both in B and M
modes)
Bar graph display for breathing waveform
3 type electrocardiogram(Ⅰ、Ⅱ、Ⅲ)
Automated detection end diastole and end systole
phase.
・ Automated split as end diastole(left) and end
systole(right)
*1 Option: MA-300 (Not available in EU)
*2 Option: TY-307A (Not available in EU)

Dual Gate Doppler (Probe dependent)


This is a function which displays Doppler Spectrums of
two different sample points simultaneously.
Supported combinations are PW/PW, TDI/TDI, and
PW/TDI.*
This function is beneficial for functional evaluations of
heart failures and phase analysis. For example, E/e'
measurements can be performed in a single scan.
*
:Automated sample gate setting for each gate.

8
Optional Functions ・Linear velocity distribution display
(Local-line Velocity Distribution)
PC printer ・Linear velocity profile display(Velocity Profile)
It is possible to printout OB/GYN, cardiology, PV, small ・Flow volume (Flow Display)
parts and urology report screens including ultrasound ・Flow volume curve (Time Flow Curve Display)
images directly with an external PC printer. ・Reliability Mark
*1 Option:SOP-LISENDO880-64 is necessary
Cardiac 3D (4D) *2:Option:SOP-FLISENDO880-77 is necessary
Cardiac 3D(4D) function provides the capability to display ※It is available to connect only phased array probe
Cardiology 3D image and to display dual image(reference (Refer mode list)
image and biplane image) simultaneously by using matrix
probe(TTE , TEE) 2D Tissue Tracking (2DTT)
2DTT enables the evaluations of regional wall motion
・Scanning rate: Up to 999 volumes/sec abnormalities within your region of interest, such as the
・Biplane Mode(Color)
myocardium and valves automatically by speckle tracking
・3D Zoom Mode(Color)
method on a B-mode image. With only minimal angle
・Active 3D Mode(Color)
・Wide Angle 3D Mode(Color) dependency, analysis form various cross sections have
・Live 1,2,3,4,5,6 beat 3D volume imaging become possible. Multiple analyses, including wall
・Live 1,2,3,4,5,6,7 beat 3D color volume imaging thickness, various strains and rotation angle can be
・It is possible to display 3 arbitrary sections simultaneously performed.
・MPR (Multi-planar Reconstruction) ・Analysis
・Surgeon’s View Display
- Free use
・2 click cropping operation
Distance and angular change between 2 points of
・Display Layout(Single , Dual , Quad , MSI)
・Function to change 3D Imaging Parameter your choice can be analyzed.
・Change MPR View to operate by drag - SAX: Parasternal short-axis view
・Density(Both Lateral and Elevation) ・Circumferential Strain and Strain Rate
・Performance mode(Function to control of rising ・Radial Strain and Strain Rate
temperature for 3D TEE surface part)
・Angle and Angle Rate
・Function to rotate(Tilt) plane operation(same as probe
・Twist (With Overlay function)
operation)
*Probe dependent -APEX: Apical long-axis view
・Apex-S (Strain)
Vector Flow Mapping Analysis( VFM) *1 ・Longitudinal Strain and Strain Rate
VFM detects velocity components perpendicular to the ・Transverse Strain and Strain Rate
beam, which couldn’t be done with the conventional color ・Global Strain and Strain Rate
Doppler technique. Direction and velocity of local blood flow ・Apex-V (Volume)
can be displayed as vectors, independent of the angle ・Volume for cavity
between the probe and the flow. VFM is expected to be ・Single and Biplane EF
effective for analyzing hemodynamics which could not be (With Overlay Function)
evaluated before. ・View
・The blood flow velocity distribution in the scanning section
-SAX
・Flow volume cutting across the specified lines on the
・Basal SAX
scanning section
・Mid SAX
・Flow line display and flow volume profile
・Apical SAX
-APEX-S, -V
・The blood flow velocity distribution in the scanning section
・2ch, 3ch, 4ch (Inversion view included)
(2D Velocity Distribution)
・Vorticity display (Vorticity Line,Vorticity) ・Graph Type
・Stream line display (Stream Line Display) -Line Graph
・Energy Loss -Color Graph
・Wall Shear Stress -Line & Color Graph
・Relative Pressure*2
-Bull’s eye (with Overlay function)
・Energy Loss Profile
・Measurements
・Circulation(Profile of Vortex)
-Point to Point
9
-Time to Peak Time, strain
M-mode
-3 point: One-third point of diastole duration
Velocity trace
-Distance Velocity, time, acceleration, ratio, velocity
difference
Stress Echo analysis TDI-Myocardial Thickness (wall thickness)
Image display modes in which image acquisition is Distance, time, velocity
possible: B, Each Flow modes Velocity Profile
 Image acquisition methods: Velocity, distance
ECG synchronized acquisition CSV output of analyzed data is possible. CSV is a file
 Compatible frame rate: Up to 75 Hz format that can be converted into Excel file directly.
 Recalled screen
Playback speed: Selectable CHI (Contrast Harmonic Imaging)*
Image allocation: Possible Contrast agent generates abundant second harmonics
Scoring: Possible when disrupted, which eases detection by Harmonic Echo.
Automatic registration: On/Off Setting for low-pressure type contrast agent is also
 Protocol: Skip view function is available.
incorporated.
Exercise stress protocols:
・ Wide-band Contrast Harmonic Imaging (WbC)
- Exercise Stress Echo
- Treadmill Exercise The wideband pulse inversion method enables retrieval
- Bicycle Exercise of echoes from the ultrasound contrast agent across a
Pharmacological stress protocols: broad spectrum and display them in CHI mode at a high
- DSE level of sensitivity.
- High-Dose DSE
・Tissue Reduction Contrast Harmonic Imaging (TrC)
- Low-Dose DSE
The Amplitude Modulation method makes it possible to
- Arbutamine
- Dipyridamole reduce signals from tissue to get clearer CHI mode
User's protocol: images.
The user can make a protocol within 8 views X 12 ・Monitor mode
stages in 1 exam. In the Monitor mode, images are available with a low
Full disclosure (Multi acquisition): 270 seconds sound pressure during the intermission of high sound
 Scoring screen
pressure transmission.
Playback speed: Selectable
・ Motion-compensated Accumulation Imaging
Comparison between different stages in the same view
is possible It is possible to display running of fine blood vessels by
Image playback range is selectable accumulating contrast echo information.
Systolic image acquisition ・Flash
Bull’s eye display (16 or 17 segmentation selectable) This mode sends high acoustic pressure transmissions at
 Report screen set intervals to destroy the ultrasound contrast
Display format
agent.
Chart/Stage overview/View overview
* Option: SOP-LISENDO880-44 (Probe dependent)

TDI analysis
Contrast Echo analysis*
Tissue Doppler Imaging (TDI) analysis is an
・ Image Subtraction
echocardiographic technique employing the Doppler
principle to measure the velocity of myocardial segments Fixed Reference: Subtraction of reference frame from all
and other cardiac structures. frames
Strain information analysis is also available. Any 2 Frame: Subtraction between 2 selected frames
B-mode Display modes: All images, arbitrary images
Temporal Velocity Profile
Velocity, time, acceleration, ratio ・ Time-Intensity Curve display
Regional Velocity Profile - Intensity level graphing of average, median,
Velocity, distance max and mode in the ROI.
TDI-Myocardial Thickness (Wall thickness)
- Standard deviation (SD) graphing of intensity in
Distance, time, velocity
the ROI .
Strain rate
- Accumulation imaging : Possible
Time, strain rate
Strain ・ Inflow Time Mapping

10
The inflow time of contrast medium can be indicated by
color map on the multi images recorded by contrast echo
inspection.
-Measurement
・Point to point
・SD
・Raising time
・Fitting curve*
・Distance
・Inflow Time Mapping Measurement
*: Parameters A, B and Beta can be calculated.
Series: Graphic display in frame sequence or time
sequence
By Group: Graphic display with the time of one sequence
of intermittent acquisition as the horizontal scale
(Graphs of multiple sequences are overlapped.)
Display mode: Image, Graph
ROI type: Square, Draw, Arc, and Circle
CSV output of analyzed data is possible.
* Option: SOP-LISENDO880-44 (Probe dependent)

EyeballEF*
The EyeballEF function provides the capability to examine
the cardiac volume and ejection
fraction in a short period of time. By using the EyeballEF
function, you can check the reference values such as an EF
value in real time and check the measurement results
immediately after Freeze.
* Probe dependent

11
possible.
* Option: SOP-LISENDO880-1
Optional Analysis Functions
Protocol Assistant*
eTRACKING (Echo Tracking) *
This function provides the capability to guide study by
It is possible to precisely measure displacement of blood
displaying protocol list (view name , mode , measure ).
vessel to obtain indices of stiffness of the vessels such as
It is useful to improve everyday study.
pressure-strain elastic modulus (Ep), stiffness
・Available to registered protocol : 128 kinds
parameter (β), arterial compliance (AC), one-point pulse
・Suspend protocol function
wave velocity (PWV), and augmentation index (AI).
・Available to edit protocol on the system
Blood-pressure gauge is necessary for this examination.
・Available to import or export like preset
* Option: SOP-LISENDO880-11

* Option: SOP-LISENDO880-79
FMD (Flow Mediated Dilatation) analysis*
It is possible to continually record and plot the vessel
diameter of the whole processes from baseline through
occlusion and vasodilatation to recovery.
* Option: SOP-LISENDO880-16 (includes eTRACKING software.)
Optional Arm holder MP-AH0001, MP-AH0001-1 and Probe
Fixing unit MP-PH0001 for FMD examination are available.

WI (Wave Intensity)*
The heart and the arterial system are acting, constantly
interfering with each other through forward traveling
waves and reflected waves. WI is calculated based on
changes in blood pressure and blood flow speed
obtained at an arbitrary point in a circulatory system. WI
is a new indicator of blood flow dynamics, which is
expected to help pave the way for analysis of the
interference between the heart and the arterial system.
The analysis comprises of contraction and dilatation
characteristics, influence of reflected waves from
peripherals, and an index related to time.
* Option: SOP-LISENDO880-34 (includes eTRACKING software.)
Blood-pressure gauge is necessary for this examination.

Automated IMT Measurement*


It is possible to automatically extract max IMT, min IMT
and mean IMT by simply setting ROI (region of interest) on
a long-axis view of the vessel. In addition, the thicknesses
at 3 points, i.e. , the point at max IMT, and the points at
1cm on the right and left of the max IMT, can be
automatically detected and averaged.
* Option: SOP-LISENDO880-38

Panoramic View*
It is possible to display an image of an extensive range of
the body by moving the probe. An area wider than the
scanning width of the probe can be displayed.
Panoramic View with flow, Power Doppler, and eFLOW are

12
General Specifications Input
・ Y/C: 1 channel
Acoustic Power – Network
・ 0 to 100%, 5% step ・ LAN (Wired, Wireless)
– Others
Preset Function
Audio (L/R): 2 channel (Output 1, Input 1)
・ 100 kinds (Max. 10 kinds per each probe)
・ Preset contents storable in USB memory
Viewing Monitor
・ Q.S.S.(Quick Scanning Selector)
・ 22 inch OLED display
Image modifying parameters (e.g. Gain, frequency,
Resolution: WXGA(1600 x 900)
depth) of your choice can be registered. (Up to 4 sets
・ Tilt and swivel are possible.
per preset)
・ Height adjustment and swivel together with operation
These parameters can immediately be registered and
panel
selected by touch panel during examinations.
Safety Regulation
Characters and graphic displays
・ IEC 60601-1 Ed.3.0: 2005
Character input area:
Class I, Type BF
ID*, name, age, sex, retained text
*: Can be corrected after exam Environmental Requirements
Input is possible with virtual keyboard on LCD panel In Operation
Automatic Annotation Labeling: ・ Temperature: +10 to +40 degrees C
800 words (User registration is possible.) ・ Relative Humidity: 30 to 75%
Body mark: (non condensing)
38 kinds are available per each region. ・ Atmospheric pressure: 700 to 1060 hPa
8 regions+1 user are able to register. ・ Altitude: Up to 3000m
Body mark editor to create user’s body mark: Available In Storage/transportation
Probe mark: 4 kinds ・ Temperature: -10 to +50 degrees C
Display position: changeable (0 to +50 degrees C for mechanical probes)
Fetal body mark: rotatable ・ Relative Humidity: 10 to 90%
Assist line display (Probe dependent) (non condensing)
・ Atmospheric pressure: 700 to 1060 hPa
Menu control
・ 10.4-inch color TFT LCD touch panel Power Requirement
・ 100 to 120/ 200 to 240V ±10%, 50 or 60 Hz,
Active Probe Ports
Max. 1300 VA (with optional recorders connected)
・ For electronic scanning probes: 6 (4 active, 2 parking)
Max. 900 VA (main unit only)
・ For independent probes*: 1
*Option: EU-9187 Dimensions
・ 55 cm (W) × 90 cm (D) ×122 – 169.5cm (H)
Input/Output Signals
Weight
– Data Input/Output
・ 150 kg (main unit only)
・ USB2.0 : 9 channel
(Main unit 6+ Operation Panel 3)
・ USB3.0 : 1 channel
– Digital Video Input/Output
・ DVI-D digital : 2 channel(Output 1, Input1)
Resolution: Full HD(1600x900)
– Analog Video Input/Output
Output
Color composite (BNC): 1 channel
Y/C: 1 channels

13
System Configuration
LISENDO 880 Main unit (includes a viewing monitor )IPF-2201

VFM software
IND probe connecting unit Picture in Picture software
SOP-LISENDO880-64 SOP-LISENDO880-63
EU-9187
Relative Pressure software
Panoramic view software HD Video Recorder*2
SOP-LISENDO880-77
SOP-LISENDO880-1 HVO-550MD / FHD
(With DVD drive)
eTRACKING software *PM-LI880-H003 is necessary
Automated IMT measurement software
SOP-LISENDO880-11
SOP-LISENDO880-38

WI software Monochrome Printer*2


(eTRACKING is included) Flow profile measurement software
UP-X898MD (Sony)
SOP-LISENDO880-34 SOP-LISENDO880-7 P95DW (Mitsubishi)

Color Printer*2
TVF measurement software
FMD software
(eTRACKING is included)
UP-D25MD (Sony)
SOP-LISENDO880-47 CP30DW (Mitsubishi)
SOP-LISENDO880-16
Foot Switch
Probe Fixing Unit DICOM network communication
software MP-2345B
MP-PH0001 MP-2819 (3-point)
Arm holder SOP-LISENDO880-10
MP-AH0001
MP-AH0001-1 DICOM SR software Jelly warmer
JW-3000U
SOP-LISENDO880-21
(Mounting kit is required)
Contrast Harmonic Imaging software
DICOM QR software MP-FX-AVA-2-R or
SOP-LISENDO880-44 MP-FX-AVA-2-L
SOP-LISENDO880-59
Advanced Cardiac Report software
McAfee embedded control2 software
SOP-LISENDO880-70
SOP-LISENDO880-69

Protocol Assistant software

SOP-LISENDO880-79

*1 All items are necessary to perform.


*2 Optional Mounting Kit is necessary.

14
OPTIONAL PROBES

Electronic convex sector probes


Application Frequency Scanning angle
Model range (MHz) (degrees) Optional accessories
(description)
Abdominal C35 8.0~2.0 70 -

Abdominal C251 5.0~1.0 70


Abdominal C252 6.0~1.0 70
Intraoperative C42K 10.0~4.0 65 Puncture Adapter
MP-2783, MP-2458

Electronic linear probes


Application Frequency range Scanning width
Model (MHz) (mm) Optional accessories
(description)
Small Organ L34 7.0~3.0 38 Coupler attachment
EZU-PA3C1H
Other Attachment
EZU-PA3C1
EZU-PA3C2
Peripheral Vessel L441 12.0~2.0 38 Elastography Stabilizer
MP-2804
Small Organ L64 18.0~5.0 38 Needle guide bracket
EZU-PA7L3

Electronic phased array sector probes


Frequency range Scanning angle
Application (description) Model Optional accessories
(MHz) (degrees)
Cardiac Adult S121 5.0~1.0 90 -
Cardiac Pediatric S31 9.0~2.0 90 -
Neonatal Cardiology S42 14.0~3.0 90 -
Cardiac, TEE S3ESL1 9.0~2.0 90 -
Cardiac Adult, TEE S3ESEL 8.0~2.0 90 -

Matrix 3D(4D) Probes*


Frequency range Scanning angle
Application (description) Model Optional accessories
(MHz) (degrees)
Cardiac Adult, TEE MXS2ESLL 10.0~1.0 90 -

Independent CW Doppler Probes*


* Independent probe connection unit EU-9187 is necessary.
Application Model Frequency range (MHz) Optional accessories

Cardiac Adult UST-2265-2 2.0 -


Cardiac Adult UST-2266-5 5.0 -

15
Dual Gate


Doppler
Linear

Trapezoid


CW



-

Compound


Cardiac 3D


Dual Gate

-
-



CW


Doppler
Dual Gate BiPlane


-
-





Compound ET/FMD/WI Doppler






Dual Gate Doppler

-
-
-
CW






eFLOW eFLOW






CW




eFLOW






Panoramic View Panoramic View






-

eFLOW

-
-
-
TDI






TDI TDI


TDI


-
-
-
-
-
-

CHI (High)

-
-
-
-
-
CHI (High) CHI (High) CHI (High)

-
-
-

-
-
-
-

CHI (Mid)

-
-
-
-
-
CHI (Mid) CHI (Mid) CHI (Mid)

-
-
-

-
-
-
-

CHI (Low)


-
-
-
-
CHI (Low)

-
-
-
CHI (Low) CHI (Low)

-
-
-
-

HdTHI


-
-
-
-
HdTHI

-
-
-
HdTHI HdTHI







-
Probes and available functions

WbT


-
-
-
-
WbT

-
-
-
WbT WbT



-
-

-
FmT




-
-
FmT

-
-
-
FmT ○ FmT


-
-
-
HI Framerate(CF)


-
-
-
-
HI Framerate(CF)

-
-
HI Framerate(CF) HI Framerate(CF)

Matrix 3D (4D)/ Independent




-
-
-
-
HI Framerate(B)


-
-
-
-
HI Framerate(B)


-
HI Framerate(B) HI Framerate(B)


-
-
-
-

-
eFocusing


-
-
-
-
eFocusing

-
-
-
eFocusing eFocusing

Phased array sector




-
-

Function
Function

UST-2265-2
UST-2266-5
Function

MXS2ESLL
Function

S3ESL1
S3ESEL
Convex sector

S121
S31
S42
L441
C42K
C251
C252

L34

L64
C35

Probe

Probe

Probe
Probe

16
The specifications are subject to change without notice.

The standard components and optional items differ depending on the country.
Not all products are marketed in all countries.
Please contact your local Hitachi distributors for details.

ALOKA, LISENDO, ARIETTA, HdTHI, HI REZ are registered trademarks or trademarks of Hitachi Ltd. in Japan
and other countries.

McAfee is a registered trademark or trademark of McAfee, Inc. and/or its affiliates in the United States and
other countries.

DICOM is a registered trademark of the National Electrical Manufacturers Association in United States for its
standards publications relating to digital communications of medical information.

2-16-1, Higashi-Ueno, Taito-ku, Tokyo, Japan www.hitachi.com/healthcare


SPH-LISEND880-V14-E01 Printed in Japan, 2017-06