Tintometer®-Group

Spectrophotometer
Tintometer®-Group

Spectrophotometer

For water and


waste water testing
330 - 900 nm

Highlights
Applications ■ 330 bis 900 nm
N Waste Water ■ Interface RS232

N Drinking Water ■ Large illuminated


display
N Industrial Process Water ■ Touch-sensitive
N Scientific & Research film keypad with
logical layout
N Governmental and
Private Laboratories ■ Use of round vials and
rectangular cells of
different sizes without
78 adapter
■ 35 user-specific methods
■ Fast, easy lamp change
■ Update via Internet
Regularly updated information: www.tintometer.com General Catalogue September 2008
The SpectroDirect is a modern single-beam Self-test Choice of language
spectrophotometer with an excellent price/perfor-
mance ratio that is specifically designed for water After it is switched on, the SpectroDirect automa- The user prompt in the display can be switched
testing. tically performs a self-test – beginning with a func- to German, English, French, Italian, Spanish or
tion test of the stepper motor and the halogen Portugese. If further languages are available they
The instrument is equipped with a wide range of lamp, followed by an optics test. For this purpose, can be updated via internet.
pre-programmed methods based on the proven the unit has a built-in didymium glass filter. This
range of Lovibond® tube tests, tablet reagents, filter checks the correct wavelength setting. If the
liquid reagents and powder reagents (Vario wavelengths are incorrect, the optical system is N.I.S.T. Traceability
Powder Packs). automatically adjusted during the self-test. This spectrophotometer can be calibrated using a
Secondary Standard Filter Set (order code 711160)
Optics which is N.I.S.T. traceable. Furthermore the
Maintenance

Photometry
The SpectroDirect is a single-beam spectral instrument may be calibrated for each method in
photometer (see illustration). Thanks to the design of the SpectroDirect, the only a "user calibration mode” with N.I.S.T. traceable
maintenance that is required is replacement of standards.
The light source is a tungsten halogen lamp with the light source. The lamp is situated at the back
flash function. The lamp is switched on only of the photometer in an easily accessible position.
momentarily during of the measurement process1), Changing the lamp is fast and simple and does not Functions
so there is no need for a warm-up period. The require any tools. The positioning of the assembly ■ Pre-programmed Lovibond® methods
SpectroDirect is ready to perform a self-test as ensures optimum focusing of the halogen lamp.
soon as it is switched on. ■ Absorption

The light passes through an entry slot to the


Power supply ■ Transmission
monochromator, where it is split into spectral The required input voltage is 12 V. ■ Spectral data recording
ranges. The monochromator is a holographically The SpectroDirect is connected to an external ■ User calibration (polynomials)
produced, transparent grating. The movable mir- power pack as standard. Battery operation is also ■ Concentration (linear)
ror ensures that light of the desired wavelength possible by using an external energy station (see
is focused automatically so that it passes through accessories).
the exit slot, into the sample chamber and there-
fore through the water sample. The light that is
not absorbed by the sample travels to the silicon
photodiode detector. This signal is then evaluated
by a microprocessor and shown as a result in the
display.
1) (Exception: permanent light is used for a wavelength scan).

Multifunctional
sample chamber
Round vials measuring 16 mm and 24 mm in dia-
meter and rectangular cells with pathlengths from
10 to 50 mm may be used without an adapter.
Only the 10 mm cell will be fixed by a little holder
that must inserted into the sample chamber (see
picture).

New methods
Test methods are continuously updated to suit
market requirements.
You can find updates for new methods and
additional languages on our website at
www.tintometer.com.

79

General Catalogue September 2008


Tintometer®-Group

Countdown function 1) 6)

With some methods, after adding the indicator


to the water sample, the user has to wait for a
predefined colour reaction time. This time interval
is shown in the display. The remaining time is dis- 2)
played continuously. An alarm sounds during the
last 10 seconds of the time periode. Measurement
then starts automatically, and the result is shown 4)
in the display. The countdown function can be
switched off to allow rapid processing of a series
of samples.
3)
Differentiation of results 5)

The SpectroDirect allows differentiated tests for 1) Tungsten halogen lamp


certain methods. With the Chlorine method, for
example, differentiated measurement is possible 2) Monochromator
for free, combined and total chlorine. 3) Movable mirror
4) Sample chamber
Functions 5) Silicon photodiode
The SpectroDirect is ideal for routine laboratory 6) Microprocessor unit
Printer/PC connection use and is equipped with additional functions for
On the back of the SpectroDirect photometer, user-specific applications. One example is the crea-
there is an RS232 interface with a 9-pin D-Sub tion of a user-defined method for a routine check. Polynomials
connector for connection of a PC or a printer with With the help of an external mathematical pro-
serial interface (see accessories). gram, the corresponding polynomial is created
Printing data from data pairs (concentration/absorption). A
known polynomial may also be used. 25 order po-
Every result is printed with date, time, reg. no, lynomials (y = A+Bx+Cx2 +Dx3 +EX4 + FX5) can be
code no., measuring range and method number. stored together with user-specific parameters such
as wavelength, measuring range, unit and number
Storing data of decimals.
You can store results of programmed and user-
specific methods (polynomials) in a memory with Concentration
a capacity of 1000 data sets. Alongside the result, This function can be used to measure 2 to 14
the data sets contain information on method, date known standards. On the basis of the concentra-
and time of the test. tions/absorption pairs obtained, the photometer
will calculate a linear interpolation between the
User prompt measured points. Up to 10 methods can be stored
Spectral data
The user prompt is a convenient and easy to A wavelength scan is performed over the user- for further sample measurements.
understand feature that guides the user step by defined interval between 330 and 900 nm.
step all the way through to the test result.
The display shows the graph of the spectrum; if
Zero calibration and measurements the user presses a key, the display also shows a
The user chooses the desired method either data list with the corresponding maximum and
from the method list in alphabetical order or by minimum absorption levels. For example:
entering a numerical code. If desired additional
information like the required vial, the reagent
Absorption/Transmission
type and the measuring range can be displayed Using this function, the operator can, for
using the functional keys. The date and time are example, carry out measurement of standards
shown in the display by pressing the "clock key". with different concentrations using the user-
The SpectroDirect automatically selects the correct selected wavelength in order to obtain the data
wavelength. pairs required for a polynomial. Result output is in
Abs and % Transmission.
Zero calibration is performed with the water
sample by pressing the ZERO key.
A characteristic coloration develops when you
add the indicator to the water sample. Press the
[Test] key to initiate the measurement (which starts
either immediately or after the time required for
80 colour development).

Regularly updated information: www.tintometer.com General Catalogue September 2008


Technical data Accessories Code Lovibond® SpectroDirect
Wavelength range: 330 to 900 nm Replacement halogen lamp 71 10 00 Spectrophotometer 330 - 900 nm
complete with power supply unit (100-240 V, 50-
Magnetical pin (for updates) 19 80 16 87 60 Hz), two batteries for keeping data and serial
Photometric range: -0.3 to 2.5 Abs
Connection cable to a PC 19 81 97 cable for connection to a PC (D9F-D9M).
Spectral bandwidth: 10 nm Connection to a 12 V plug 71 10 40 Order code: 71 20 00

Wavelength ±2 nm Case for transport 71 20 50


accuracy: Power supply unit 71 10 90 Lovibond® SpectroDirect
100-240 V / 50-60 Hz Spectrophotometer 330 - 900 nm
Wavelength ±1 nm in case as above, with energy station, replacement
reproducibility: Universal adapter for sockets 19 20 65 lamp, 12 round vials with lid (height 48 mm,
Secondary standard set 71 11 60 ø 24 mm),10 round vials with lid (height 90 mm,
Light source: Pre-adjusted tungsten ø 16 mm), W 100 rectangular cell (50 mm path
Power station, 71 10 50
halogen lamp length), W 100 rectangular cell (10 mm path
230 V / 50 Hz
with cable for connection length), plastic stirring rod, but without reagents.
Monochromator: Holographic grating
Order code: 71 20 05

Photometry
Detector: Silicon photodiode 12 round vials with lid 19 76 20 We would be pleased to quote a ready to use
Height 48 mm, 24 mm Ø spectrophotometer unit for the parameters and
Multifunctional Round vials required accessories.
sample chamber for: 24 and 16 mm Ø, 5 round vials with lid 19 76 29
Rectangular cells Height 48 mm, 24 mm Ø
10-50 mm 12 round vials with lid 19 76 65
Height 90 mm, 16 mm Ø
Display: Backlit LCD
graphic display W 100, rectangular cell 60 10 40
optical glass OG,
Language options: German, English, 10 mm path length
French, Italian, Spanish, W 100, rectangular cell 60 10 70
Portugese optical glass OG,
Storage capacity: 1000 test data sets 50 mm path length
W 110, rectangular cell 66 11 30
Serial interface: RS232 Quartz-UV-glass,
10 mm path length
Dimensions: 270 x 275 x 150 mm
(L x W x H)
Paper printer DPN 2335 19 80 75
Weight: approx. 3.2 kg with power pack (230 V, 50 Hz)
connection cable and one
Power supply Input: 100 - 240 V ~ paper roll
unit: 1.0 A 50 - 60 Hz
Output: 12 V .... 30 W
Setup for Arsenic-test complet 37 05 00
CE: EN 50082-1 Standard - Spare parts:
1992 Edition-Immunity
Generic Standard Erlenmeyer flask 37 05 01
EN 55022 Standard Glass stopper 37 05 02
Class - 1994 Edition Absorption tube 37 05 03
Emission Generic
Standard W 100, cell, Optical-Glass-OG, 60 10 50
EN 5081-1 Standard 20 mm path length
1992 Edition –Emission
Generic Standard

81

General Catalogue September 2008


Tintometer®-Group

Parameter, Ranges, Reagents


Test λa) / nm Method Cuvette Range Display

Alkalinity-m 615 Acid/Indicator 1, 2 24 mm ø 5 - 200 mg/l CaCO3


Alkalinity-p 551 Acid/Indicator 1, 2 24 mm ø 5 - 300 mg/l CaCO3


Aluminium VARIO 530 Eriochrome cyanine R 2 24 mm ø 0.01 - 0.25 mg/l Al




Aluminium 535 Eriochrome cyanine R 2 24 mm ø 0.01 - 0.3 mg/l Al


Ammonia 676 Indophenole blue 1, 2 24 mm ø 0.02 - 1 mg/l N



Ammonia VARIO 655 Salicylate 2 24 mm ø 0.01 - 0.8 mg/l N




Ammonia VARIO LR 655 Salicylate 2 16 mm ø 0.02 - 2.5 mg/l N





Ammonia VARIO HR 655 Salicylate 2 16 mm ø 1 - 50 mg/l N



Arsenic (III, IV) 507 Silberdiethyldithiocarbamate1 20 mm o 0.02 - 0.6 mg/l As
Boron 430 Azomethine 24 mm ø 0.1 - 2 mg/l B 3



50 mm o 0.05 - 1 mg/l

Bromine 510 DPD 5 10 mm o 0.1 - 3 mg/l Br
24 mm ø 0.05 - 6.5 mg/l
Bromine 510 DPD 5 24 mm ø 0.1 - 4.5 mg/l Br

Cadmium (Cd2+) 525 Cadion 16 mm ø 0.025 - 0.75 mg/l Cd

MSDS (Material Safety Data Sheets): www.tintometer.com


Legend
1
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlamm- Untersuchung
2
Standard Methods for the Examination of Water and Wastewater, 18th Edition; 1992
3
Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart; 1989
4
Photometrische Analyse, Lange/Vejdelek, Verlag Chemie; 1980
82 5
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond‚
o Rectangular cell, ø Round vial
#
including stirring rod, 10 cm

Regularly updated information: www.tintometer.com General Catalogue September 2008


Reagent Form of reagent/Quantity Order-No.

ALKA-M-PHOTOMETER Tablet / 100 51 32 10


ALKA-P-PHOTOMETER Tablet / 100 51 32 30
VARIO Aluminum ECR/F20 Powder Pack / 100
VARIO Aluminum Hexamine/F20 Powder Pack / 100
VARIO Aluminum Masking Reagt. Liquid reagent / 25 ml
Set 53 50 00

Photometry
ALUMINIUM No. 1 Tablette / 100 51 54 60
ALUMINIUM No. 2 Tablette / 100 51 54 70
Combi pack# each 100 51 76 01
ALUMINIUM No.1 / No.2 each 250 51 76 02
AMMONIA No. 1 Tablet / 100 51 25 80
AMMONIA No. 2 Tablet / 100 51 25 90
Combi pack# each 100 51 76 11
AMMONIA No.1 / No.2 each 250 51 76 12
VARIO Ammonia Salicylate F10 Powder Pack / 100 53 13 80
VARIO Ammonia Cyanurate F10 Powder Pack / 100 53 13 70
Set, each 200 53 55 00
VARIO Ammonia Salicylate F5 Powder Pack / 50
VARIO Ammonia Cyanurate F5 Powder Pack / 50
VARIO Am Diluent Reagent Low Range Tube test / 50
VARIO Deionised water (for Zero) Bottle, 100 ml
Set 53 56 00
VARIO Ammonia Salicylate F5 Powder Pack / 50
VARIO Ammonia Cyanurate F5 Powder Pack / 50
VARIO Am Diluent Reagent High Range Tube test / 50
VARIO Deionised water (for Zero) Bottle, 100 ml
Set 53 56 50
for chemicals see manual, reagents at specialized chemistry dealer
BORON No. 1 Tablet / 100 51 57 90
BORON No. 2 Tablet / 100 51 58 00
Combi pack# each 100 51 76 81
BORON No.1 / No.2 each 200 51 76 82
DPD No. 1 Tablet / 100 51 10 60
DPD No. 1 HIGH CALCIUM f) Tablet / 100 51 57 40

DPD 1 Buffer-Solution Liquid reagent / 15 ml 47 10 10


DPD 1 Reagent-Solution Liquid reagent / 15 ml 47 10 20
Spectroquant® 1.14834.0001 Tube test / 25 42 07 50

Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 83
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents

General Catalogue September 2008


Tintometer®-Group

Parameter, Ranges, Reagents


Test λa) / nm Method Cuvette Range Display

50 mm o 0.02 - 0.5 mg/l



Chlorine b) 510 DPD 1, 2 10 mm o 0.1 - 6 mg/l Cl2
24 mm ø 0.02 - 3 mg/l


Chlorine b) 510 DPD 1, 2 24 mm ø 0.02 - 3 mg/l Cl2

Chlorine VARIO b) 510 DPD 1, 2 24 mm ø 0.01 - 2 mg/l Cl2

Chlorine HR 470 Potassium iodide 5 16 mm ø 5 - 200 mg/l Cl2



Chlorine Dioxide 510 DPD 1,2 50 mm o 0.04 - 1 mg/l ClO2
24 mm ø 0.5 - 2.5 mg/l



Chlorine Dioxide 510 DPD 1,2 24 mm ø 0.5 - 2.5 mg/l ClO2

Chloride 450 Silver nitrate/turbidity 24 mm ø 0.5 - 25 mg/l Cl

Chloride 455 Iron (III)-thiocyanate 4 24 mm ø 5 - 60 mg/l Cl

Chromium d)* 542 1,5-Diphenylcarbohydrazide1,2 50 mm o 0.005 - 0.5 mg/l Cr
16 mm ø 0.02 - 2 mg/l
COD vario LR* 420 Dichromate/H2SO4 1 16 mm ø 0 - 150 mg/l COD
(ISO 15705:2003-01)
COD vario MR* 620 Dichromate/H2SO4 1 16 mm ø 0 - 1500 mg/l COD
(ISO 15705:2003-01)
COD vario HR* 620 Dichromate/H2SO4 1 16 mm ø 0 - 15000 mg/l COD
Copper 559
b)
Biquinoline 50 mm o 0.05 - 1 mg/l Cu 4
24 mm ø 0.5 - 5 mg/l


Copper VARIO 560 Bicinchoninate 24 mm ø 0.05 - 5 mg/l Cu
Cyanide 585 Pyridine-barbituric acid 1 50 mm o 0.005 - 0.2 mg/l CN
24 mm ø 0.01 - 0.5 mg/l
DEHA 562 PPST 3 24 mm ø 0.02 - 0.5 mg/l DEHA

MSDS (Material Safety Data Sheets): www.tintometer.com


Legend
1
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlamm- Untersuchung
2
Standard Methods for the Examination of Water and Wastewater, 18th Edition; 1992
3
Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart; 1989
4
Photometrische Analyse, Lange/Vejdelek, Verlag Chemie; 1980
84 5
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond‚
o Rectangular cell, ø Round vial
#
including stirring rod, 10 cm

Regularly updated information: www.tintometer.com General Catalogue September 2008


Reagent Form of reagent/Quantity Order-No.

DPD No. 1 Tablet / 100 51 10 60


DPD No. 1 HIGH CALCIUM f) Tablet / 100 51 57 40
DPD No. 3 Tablet / 100 51 10 80
Combi pack# each 100 51 77 11
DPD No.1 / No.3 each 250 51 77 12
DPD 1 Buffer-Solution Liquid reagent / 15 ml 47 10 10

Photometry
DPD 1 Reagent-Solution Liquid reagent / 15 ml 47 10 20
DPD 3 Solution Liquid reagent / 15 ml 47 10 30
VARIO Chlorine FREE-DPD/F10 Powder Pack / 100 53 01 00
VARIO Chlorine TOTAL-DPD/F10 Powder Pack / 100 53 01 20
CHLORINE HR (PI) Tablet / 100 51 30 00
ACIDIFYING GP Tablet / 100 51 54 80
Combi pack# each 100 51 77 21
CHLORINE HR (KI)/ACIDIFYING GP each 250 51 77 22
DPD No. 1 Tablet / 100 51 10 60
DPD No. 1 HIGH CALCIUM f) Tablet / 100 51 57 40
GLYCINE (additionally for ClO2 in presence of chlorine) Tablet / 100
Combi pack# each 100 51 77 31
DPD No.1 / GLYCINE each 250 51 77 32
DPD 1 Buffer-Solution Liquid reagent / 15 ml 47 10 10
DPD 1 Reagent-Solution Liquid reagent / 15 ml 47 10 20
CHLORIDE T1 Tablet / 100 51 59 10
CHLORIDE T2 Tablet / 100 51 59 20
RT (Chlorid-51 / Chlorid-52) Reagent test (Liquid reagent) 2 41 90 31
approx. 50-75 tests
"PERSULF. RTG FOR CR", Powder Pack / 100 53 73 00
Chromium Hexavalent Powder Pack /100 53 73 10
COD VARIO 0 - 150 mg/l Tube test / 25 2 42 07 20

COD VARIO 0 - 1500 mg/l Tube test / 25 2 42 07 21

COD VARIO 0 - 15000 mg/l Tube test / 25 2 42 07 22


COPPER No. 1 Tablet / 100 51 35 50
COPPER No. 2 Tablet / 100 51 35 60
Combi pack# each 100 51 76 91
COPPER No.1 / No.2 each 250 51 76 92
VARIO Cu 1 F10 Powder Pack / 100 53 03 00
RT (Cyanid-11 / Cyanid-12 / Reagent test (Powder, 2 41 88 75
Cyanid-13) Liquid reagent) / 200 Tests
DEHA Solution Liquid reagent / 100 ml 46 11 81
DEHA Tablet / 100 51 32 20

Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 85
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents

General Catalogue September 2008


Tintometer®-Group

Parameter, Ranges, Reagents


Test λa) / nm Method Cuvette Range Display

DEHA 562 PPST 3 24 mm ø 0.02 - 0.5 mg/l DEHA





Fluoride 580 SPADNS 2 24 mm ø 0.05 - 1.5 mg/l F

Formaldehyde 585 H2SO4/Chromotropic acid 10 mm o 1 - 5 mg/l HCHO
50 mm o 0.02 - 1 mg/l
Formaldehyde 575 H2SO4/Chromotropic acid 16 mm ø 0.1 - 5 mg/l HCHO
Hardness-total 571 Metallphthalein 3 24 mm ø 2 - 50 mg/l CaCO3
Hardness-total HR 571 Metallphthalein 3 24 mm ø 20 - 500 mg/l CaCO3
Hazen (Pt-Co units, APHA) 455 Direct reading 2 50 mm o 0 - 500 mg Pt/l Pt-Co units
Hydrazine 455 Dimethylaminobenzaldehyde 3 24 mm ø 0.05 - 0.5 mg/l N2H4

Hydrazine 455 Dimethylaminobenzaldehyde 3 24 mm ø 0.005 - 0.6 mg/l N2H4
Hydrogenperoxide 510 DPD/Catalyst 5 50 mm o 0.01 - 0.5 mg/l H2O2
24 mm ø 0.03 - 1.5 mg/l
Iodine 510 DPD5 24 mm ø 0.05 - 3.6 mg/l I
Iron (ferrous Fe2+, 50 mm o 0.01 - 0.5 mg/l
ferric Fe 3+) soluble
562 PPST 3 10 mm o 0.1 - 1 mg/l Fe
24 mm ø 0.1 - 1 mg/l
Iron (ferrous Fe2+) 50 mm o 0.01 - 0.5 mg/l
soluble 562 PPST 3 10 mm o 0.1 - 1 mg/l Fe
24 mm ø 0.1 - 1 mg/l
Iron VARIO (ferrous Fe2+, 510 1, 10- Phenantroline 1, 2 24 mm ø 0.1 - 3 mg/l Fe
(ferric Fe3+) soluble
Iron, total 590 TPTZ 24 mm ø 0.1 - 1.8 mg/l Fe
Lead (Pb2+) 520 4-(2-Pyridylazo)-resorcin 10 mm o 0.1 - 5 mg/l Pb
Lead (Pb2+) 515 4-(2-Pyridylazo)-resorcin 16 mm ø 0.1 - 5 mg/l Pb
Manganese 450 Formaldoxime 1 24 mm ø 0.2 - 4 mg/l Mn




Manganese VARIO LR 558 PAN 24 mm ø 0.01 - 0.7 mg/l Mn




Manganese VARIO HR 525 Periodate oxidation2 24 mm ø 0.1 - 18 mg/l Mn

MSDS (Material Safety Data Sheets): www.tintometer.com


Legend
1
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlamm- Untersuchung
2
Standard Methods for the Examination of Water and Wastewater, 18th Edition; 1992
3
Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart; 1989
4
Photometrische Analyse, Lange/Vejdelek, Verlag Chemie; 1980
86 5
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond‚
o Rectangular cell, ø Round vial
#
including stirring rod, 10 cm

Regularly updated information: www.tintometer.com General Catalogue September 2008


Reagent Form of reagent/Quantity Order-No.

OXYSCAV 1 RGT Powder Pack / 200


VARIO DEHA 2 RGT Solution / 100 ml

Set 53 60 00
SPADNS-Reagent Liquid reagent / 250 ml 46 74 81
Fluoride-Standard c) Liquid / 30 ml 20 56 30
e)
Spectroquant® 1.14678.0001 Reagent test / approx. 75-100 Tests 42 07 51

Photometry

Spectroquant® 1.14500.0001 e) Tube test / 25 42 07 52
HARDCHECK P Tablet / 100 51 56 60
HARDCHECK P Tablet / 100 51 56 60
--- --- ---
Hydrazine Test Powder Powder / 30 g 46 29 10
Spoon 38 49 30
VARIO Hydra 2 Liquid reagent / 100 ml 53 12 00
HYDROGENPEROXIDE LR Tablet / 100 51 23 80

DPD No. 1 Tablet / 100 51 10 60



IRON LR Tablet / 100 51 53 70


IRON (II) LR Tablet / 100 51 54 20

VARIO Ferro F10 Powder Pack /100 53 05 60

Vario Iron TPTZ F10 Powder Pack / 100 53 05 50


Spectroquant® 1.09717.0001 e) Reagent test / 50 Tests 42 07 53
Spectroquant® 1.14833.0001 e) Tube test / 25 42 07 54
MANGANESE LR 1 Tablet / 100 51 60 80
MANGANESE LR 2 Tablet / 100 51 60 90
Combi pack# each 100 51 76 21
MANGANESE LR 1 / LR 2 each 250 51 76 22
VARIO Ascorbic Acid Powder Pack / 100
VARIO Alkaline-Cyanide Liquid reagent / 50 ml
VARIO PAN Indicator Liquid reagent / 50 ml
VARIO Rochelle Salt Solutionh) 30 ml
Set 53 50 90
VARIO Manganese Citrate Buffer F10 Powder Pack / 100
VARIO Sodiumperiodate F10 Powder Pack / 100
Set 53 51 00

Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 87
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents

General Catalogue September 2008


Tintometer®-Group

Parameter, Ranges, Reagents


Test λa) / nm Method Cuvette Range Display

Molybdate 366 Thioglycolate 4 24 mm ø 1 - 30 mg/l MoO4






Molybdate VARIO 420 Mercaptoacetic acid 24 mm ø 0.5 - 66 mg/l MoO4


Nickel 443 Dimethylglyoxime 2, 3 50 mm o 0.02 - 1 mg/l Ni
24 mm ø 0.2 - 7 mg/l
Nitrate VARIO 410 Chromotropic acid 16 mm ø 1 - 30 mg/l N



Nitrate 340 2,6-Dimethylphenole 2,3 16 mm ø 0.5 - 14 mg/l N

Nitrite 545 NED 2,3 24 mm ø 0.01 - 0.5 mg/l N
Nitrite 545 Sulfanilic acid, Naphthyl-amine 16 mm ø 1 0.03 - 0.6 mg/l N
0.3 - 3 mg/l
Nitrite VARIO 507 Diazotation 24 mm ø 0.01 - 0.3 mg/l N
Nitrogen-total* 340 2,6-Dimethylphenole 2,3 16 mm ø 0.5 - 14 mg/l N
5 - 140 mg/l



Nitrogen VARIO, total LR * 410 Persulphate digestion method 16 mm ø 0.5 - 25 mg/l N






Nitrogen VARIO, total HR * 410 Persulphate digestion method 16 mm ø 5 - 150 mg/l N





Ozone 510 DPD/Glycine 5 50 mm o 0.02 - 0.5 mg/l O3
24 mm ø 0.02 - 1 mg/l


MSDS (Material Safety Data Sheets): www.tintometer.com


Legend
1
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlamm- Untersuchung
2
Standard Methods for the Examination of Water and Wastewater, 18th Edition; 1992
3
Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart; 1989
4
Photometrische Analyse, Lange/Vejdelek, Verlag Chemie; 1980
88 5
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond‚
o Rectangular cell, ø Round vial
#
including stirring rod, 10 cm

Regularly updated information: www.tintometer.com General Catalogue September 2008


Reagent Form of reagent/Quantity Order-No.

MOLYBDATE HR No. 1 Tablet / 100 51 30 60


MOLYBDATE HR No. 2 Tablet / 100 51 30 70
Combi pack# each 100 51 76 31
MOLYBDATE No.1 HR / No.2 HR each 250 51 76 32
VARIO Molybdenum HR1 F10 Powder Pack / 100
VARIO Molybdenum HR2 F10 Powder Pack / 100

Photometry
VARIO Molybdenum HR3 F10 Powder Pack / 100
Set 53 53 00
RT (Nickel-51, Nickel-52) Reagent test (Powder, 2 41 90 33
Liquid reagent) / 50 Tests
VARIO Nitrate Chromotropic Powder Pack / 50
VARIO Nitra X Reagent tube Reaktionsküvette / 50
VARIO Deionised water (for Zero) Bottle 100 ml
Set 53 55 80
TT (Nitrat-111) Tube test 2 42 07 02
(Liquid reagent) / 24
NITRITE LR Tablet / 100 51 23 10
TT (Nitrit-101) Tube test (Powder) / 24 2 41 90 18

VARIO Nitri 3 Powder Pack / 100 53 09 80


TT (Digestion reagent, Tube test (Powder, 2 42 07 03
Compensation- reagent, Nitrat-111) Liquid reagent) / 24
VARIO TN HYDROX. LR tubes Digestion tubes / 50
VARIO PERSULFATE reagent Powder Pack / 50
VARIO TN reagent A Powder Pack / 50
VARIO TN reagent B Powder Pack / 50
VARIO TN ACID LR/HR tubes Reaktion tubes / 50
VARIO Deionised water (for Zero) Bottle 100 ml
Set 53 55 50
VARIO TN HYDROX. HR tubes Digestion tubes / 50
VARIO PERSULFATE reagent Powder Pack / 50
VARIO TN reagent A Powder Pack / 50
VARIO TN reagent B Powder Pack / 50
VARIO TN ACID LR/HR tubes Reaktion tubes / 50
VARIO Deionised water (for Zero) Bottle 100 ml
Set 53 55 60
DPD No. 1 Tablet / 100 51 10 60
DPD No. 1 HIGH CALCIUM f) Tablet / 100 51 57 40
DPD No. 3 Tablet / 100 51 10 80
Combi pack# each 100 51 77 11
DPD No.1 / No.3 each 250 51 77 12
DPD GLYCINE g) Tablet / 100 51 21 70

Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 89
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents

General Catalogue September 2008


Tintometer®-Group

Parameter, Ranges, Reagents


Test λa) / nm Method Cuvette Range Display

Phenols 507 4-Aminoantipyrine 2 24 mm ø 0.1 - 5 mg/l C6H5OH



Phosphate-total (PMB)* 690 Phosphomolybdic acid/ 16 mm ø 0.07 - 3 mg/l P
Ascorbic acid 2 0.2 - 10 mg/l PO4
Phosphate-total (PMB)* 690 Phosphomolybdic acid/ 16 mm ø 1.5 - 20 mg/l P
Ascorbic acid 2 5 - 60 mg/l PO4
Phosphate LR-ortho 710 Ammonia molybdate 2, 3 24 mm ø 0.05 - 4 mg/l PO4



Phosphate HR, ortho 470 Vanadomolybdate 2 24 mm ø 1 - 80 mg/l PO4



Phosphate VARIO, ortho 890 Ascorbic acid 2 24 mm ø 0.06 -2.5 mg/l PO4

Phosphate VARIO, ortho 890 Ascorbic acid 2 16 mm ø 0.06 - 5 mg/l PO4


Phosphate-ortho (VM) 438 Vanadomolybdate 2 16 mm ø 3 - 60 mg/l PO4
Acid digestion 16 mm ø acid hydrolyzable:
Phosphate VARIO* Ascorbic acid2 ^
0.02 - 1.6 mg/l P =
P
acid hydrolyzable 660 0.06 - 5 mg/l PO4 P
and total Acid-/Persulphate digestion 16 mm ø total:
Ascorbic acid2 ^
0.02 - 1.1 mg/l P =
P
0.06 - 3.5 mg/l PO4 P



Phosphate VARIO, total* 660 Acid-/Persulphate digestion 16 mm ø ^

0.02 - 1.1 mg/l P = P
Ascorbic acid2 0.06 - 3.5 mg/l PO4 P




pH value 558 Phenol red 5 24 mm ø 6.5 - 8.4 pH
pH value 558 Phenol red 5 24 mm ø 6.5 - 8.4 pH
Potassium 730 Tetraphenylborate, Turbidity 4 24 mm ø 1 - 10 mg/l K


Silica 820 Silicomolybdate 1,2 24 mm ø 0.05 - 3 mg/l SiO2


MSDS (Material Safety Data Sheets): www.tintometer.com


Legend
1
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlamm- Untersuchung
2
Standard Methods for the Examination of Water and Wastewater, 18th Edition; 1992
3
Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart; 1989
4
Photometrische Analyse, Lange/Vejdelek, Verlag Chemie; 1980
90 5
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond‚
o Rectangular cell, ø Round vial
#
including stirring rod, 10 cm

Regularly updated information: www.tintometer.com General Catalogue September 2008


Reagent Form of reagent/Quantity Order-No.

PHENOLE No. 1 Tablet / 100 51 59 50


PHENOLE No. 2 Tablet / 100 51 59 60
TT (Phosphat-101 , Tube test (Powder, 2 41 90 19
Phosphat- 102, Phosphat-103) Liquid -reagent) / 24
TT (Phosphat-101 , Tube test (Powder, 2 42 07 00
Phosphat-102, Phosphat-103) Liquid -reagent) / 24

Photometry
PHOSPHATE No. 1 LR Tablet / 100 51 30 40
PHOSPHATE No. 2 LR Tablet / 100 51 30 50
Combi pack# each 100 51 76 51
PHOSPHATE No.1 LR / No.2 LR each 200 51 76 52
PHOSPHATE No. 1 HR Tablette / 100 51 58 10
PHOSPHATE No. 2 HR Tablette / 100 51 58 20
Combi pack# each 100 51 76 61
PHOSPHATE No.1 HR / No.2 HR each 200 51 76 62
VARIO PHOS3 F10 Powder Pack / 100 53 15 50
VARIO Diluent Vial Tubes / 50
VARIO PHOS3 F10 Powder Pack / 50
VARIO Deionised water (for Zero) Bottle 100 ml
Set 53 52 00
TT Tube test / 24 2 42 07 01
Tubes Acid reagent / 50
VARIO Phosphate RGT, 10 ml Powder Pack / 50
VARIO Deionised water (for Zero) Bottle 100 ml
1N NaOH Bottle / 100 ml
1.54N NaOH Bottle / 100 ml
VARIO POTASSIUM PERSULFATE Powder Pack / 50
Set 53 52 50
Tubes Acid reagent / 50
VARIO Phosphate RGT, 10 ml Powder Pack / 50
VARIO Deionised water (for Zero) Bottle 100 ml
1.54N NaOH Bottle / 100 ml
VARIO POTASSIUM PERSULFATE Powder Pack / 50
Set 53 52 10
PHENOLRED/PHOTOMETER Tablet / 100 51 17 70
PHENOLRED Solution Liquid reagent / 15 ml 47 10 40
POTASSIUM T Tablette / 100 51 56 70
SILICA No. 1 Tablet / 100 51 31 30
SILICA No. 2 Tablet / 100 51 31 40
Combi pack# each 100 51 76 71
SILICA No.1 / No.2 each 200 51 76 72
SILICA PR (in presence of phosphate) Tablet / 100 51 31 50

Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 91
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents

General Catalogue September 2008


Tintometer®-Group

Parameter, Ranges, Reagents


Test λa) / nm Method Cuvette Range Display


Silica LR VARIO 815 Heteropoly blue2 24 mm ø 0.1 - 1.6 mg/l SiO2




Silicia HR VARIO 452 Silicomolybdate 24 mm ø 1 - 100 mg/l SiO2


436
Spectral Direct Reading 1
525 50 mm o 0 - 50 m-1 ---
Absorption-coefficient ISO 7887:1994
620
Sulphate VARIO 450 Barium sulphate, Turbidity 2 24 mm ø 2 - 100 mg/l SO4
Sulphide 668 DPD/Catalyst 24 mm ø 0.04 - 0.5 mg/l S 3,4

Sulphite LR 405 DTNB 10 mm o 0.1 - 10 mg/l SO3
24 mm ø 0.05 - 4 mg/l
Surfactants (anionic) 653 Methylene blue 1 16 mm ø 0.05 - 2 mg/l MBAS
TOC* 596 H2SO4/ Indicator 16 mm ø 50 - 800 mg/l TOC

Turbidity 860 Attenuated Radiation Method 50 mm o 5 - 500 NTU


Urea 610 Urease / Indophenole 24 mm ø 0.1 - 2 mg/l Urea






Zinc 616 Zincon 3 24 mm ø 0.02 - 0.5 mg/l Zn

MSDS (Material Safety Data Sheets): www.tintometer.com


Legend
1
Deutsche Einheitsverfahren zur Wasser-, Abwasser- und Schlamm- Untersuchung
2
Standard Methods for the Examination of Water and Wastewater, 18th Edition; 1992
3
Photometrische Analysenverfahren, Schwedt, Wissenschaftliche Verlagsgesellschaft mbH, Stuttgart; 1989
4
Photometrische Analyse, Lange/Vejdelek, Verlag Chemie; 1980
92 5
Colorimetric Chemical Analytical Methods, 9th Edition, Lovibond‚
o Rectangular cell, ø Round vial
#
including stirring rod, 10 cm

Regularly updated information: www.tintometer.com General Catalogue September 2008


Reagent Form of reagent/Quantity Order-No.
VARIO Amino Acid F10 Powder Pack / 100
VARIO Citric Acid F10 Powder Pack / 100
VARIO Molybdate 3 Reagent solution Liquid reagent / 50 ml
Set 53 56 90
VARIO Silica HR Molybdate F10 Powder Pack / 100
VARIO Silica HR Acid Rgt F10 Powder Pack / 100
VARIO Silica HR Citric Acid F10 Powder Pack / 100

Photometry
Set 53 57 00

--- --- ---

VARIO Sulpha 4 Powder Pack / 100 53 21 50


SULFIDE No. 1 Tablet / 100 50 29 30
SULFIDE No. 2 Tablet / 100 50 29 40
SULFITE LR Tablet / 100 51 80 20

Spectroquant® 1.14697.0001 e) Tube test / 25 42 07 55
Spectroquant® 1.14879.0001 e) Tube test / 25 42 07 56
Aluminium screwcaps (6 pc.) 42 07 57
--- --- ---
UREA Reagent 1 Liquid reagent / 15 ml 45 93 00
UREA Reagent 2 Liquid reagent / 10 ml 45 94 00
AMMONIA No. 1 Tablet / 100 51 25 80
AMMONIA No. 2 Tablet / 100 51 25 90
Combi pack# each 100 51 76 11
AMMONIA No.1 / No.2 each 250 51 76 12
COPPER/ZINC LR Tablet / 100 51 26 20
EDTA (in case of zinc in presence of chlorine) Tablet / 100 51 23 90
DECHLOR Tablet / 100 51 23 50

Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 93
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents

General Catalogue September 2008