Spectrophotometer
Tintometer®-Group
Spectrophotometer
Highlights
Applications ■ 330 bis 900 nm
N Waste Water ■ Interface RS232
Photometry
The SpectroDirect is a single-beam spectral instrument may be calibrated for each method in
photometer (see illustration). Thanks to the design of the SpectroDirect, the only a "user calibration mode” with N.I.S.T. traceable
maintenance that is required is replacement of standards.
The light source is a tungsten halogen lamp with the light source. The lamp is situated at the back
flash function. The lamp is switched on only of the photometer in an easily accessible position.
momentarily during of the measurement process1), Changing the lamp is fast and simple and does not Functions
so there is no need for a warm-up period. The require any tools. The positioning of the assembly ■ Pre-programmed Lovibond® methods
SpectroDirect is ready to perform a self-test as ensures optimum focusing of the halogen lamp.
soon as it is switched on. ■ Absorption
Multifunctional
sample chamber
Round vials measuring 16 mm and 24 mm in dia-
meter and rectangular cells with pathlengths from
10 to 50 mm may be used without an adapter.
Only the 10 mm cell will be fixed by a little holder
that must inserted into the sample chamber (see
picture).
New methods
Test methods are continuously updated to suit
market requirements.
You can find updates for new methods and
additional languages on our website at
www.tintometer.com.
79
Countdown function 1) 6)
Photometry
Detector: Silicon photodiode 12 round vials with lid 19 76 20 We would be pleased to quote a ready to use
Height 48 mm, 24 mm Ø spectrophotometer unit for the parameters and
Multifunctional Round vials required accessories.
sample chamber for: 24 and 16 mm Ø, 5 round vials with lid 19 76 29
Rectangular cells Height 48 mm, 24 mm Ø
10-50 mm 12 round vials with lid 19 76 65
Height 90 mm, 16 mm Ø
Display: Backlit LCD
graphic display W 100, rectangular cell 60 10 40
optical glass OG,
Language options: German, English, 10 mm path length
French, Italian, Spanish, W 100, rectangular cell 60 10 70
Portugese optical glass OG,
Storage capacity: 1000 test data sets 50 mm path length
W 110, rectangular cell 66 11 30
Serial interface: RS232 Quartz-UV-glass,
10 mm path length
Dimensions: 270 x 275 x 150 mm
(L x W x H)
Paper printer DPN 2335 19 80 75
Weight: approx. 3.2 kg with power pack (230 V, 50 Hz)
connection cable and one
Power supply Input: 100 - 240 V ~ paper roll
unit: 1.0 A 50 - 60 Hz
Output: 12 V .... 30 W
Setup for Arsenic-test complet 37 05 00
CE: EN 50082-1 Standard - Spare parts:
1992 Edition-Immunity
Generic Standard Erlenmeyer flask 37 05 01
EN 55022 Standard Glass stopper 37 05 02
Class - 1994 Edition Absorption tube 37 05 03
Emission Generic
Standard W 100, cell, Optical-Glass-OG, 60 10 50
EN 5081-1 Standard 20 mm path length
1992 Edition –Emission
Generic Standard
81
Photometry
ALUMINIUM No. 1 Tablette / 100 51 54 60
ALUMINIUM No. 2 Tablette / 100 51 54 70
Combi pack# each 100 51 76 01
ALUMINIUM No.1 / No.2 each 250 51 76 02
AMMONIA No. 1 Tablet / 100 51 25 80
AMMONIA No. 2 Tablet / 100 51 25 90
Combi pack# each 100 51 76 11
AMMONIA No.1 / No.2 each 250 51 76 12
VARIO Ammonia Salicylate F10 Powder Pack / 100 53 13 80
VARIO Ammonia Cyanurate F10 Powder Pack / 100 53 13 70
Set, each 200 53 55 00
VARIO Ammonia Salicylate F5 Powder Pack / 50
VARIO Ammonia Cyanurate F5 Powder Pack / 50
VARIO Am Diluent Reagent Low Range Tube test / 50
VARIO Deionised water (for Zero) Bottle, 100 ml
Set 53 56 00
VARIO Ammonia Salicylate F5 Powder Pack / 50
VARIO Ammonia Cyanurate F5 Powder Pack / 50
VARIO Am Diluent Reagent High Range Tube test / 50
VARIO Deionised water (for Zero) Bottle, 100 ml
Set 53 56 50
for chemicals see manual, reagents at specialized chemistry dealer
BORON No. 1 Tablet / 100 51 57 90
BORON No. 2 Tablet / 100 51 58 00
Combi pack# each 100 51 76 81
BORON No.1 / No.2 each 200 51 76 82
DPD No. 1 Tablet / 100 51 10 60
DPD No. 1 HIGH CALCIUM f) Tablet / 100 51 57 40
Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 83
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents
Chlorine b) 510 DPD 1, 2 24 mm ø 0.02 - 3 mg/l Cl2
Chlorine VARIO b) 510 DPD 1, 2 24 mm ø 0.01 - 2 mg/l Cl2
Chlorine HR 470 Potassium iodide 5 16 mm ø 5 - 200 mg/l Cl2
Chlorine Dioxide 510 DPD 1,2 50 mm o 0.04 - 1 mg/l ClO2
24 mm ø 0.5 - 2.5 mg/l
Chlorine Dioxide 510 DPD 1,2 24 mm ø 0.5 - 2.5 mg/l ClO2
Chloride 450 Silver nitrate/turbidity 24 mm ø 0.5 - 25 mg/l Cl
Chloride 455 Iron (III)-thiocyanate 4 24 mm ø 5 - 60 mg/l Cl
Chromium d)* 542 1,5-Diphenylcarbohydrazide1,2 50 mm o 0.005 - 0.5 mg/l Cr
16 mm ø 0.02 - 2 mg/l
COD vario LR* 420 Dichromate/H2SO4 1 16 mm ø 0 - 150 mg/l COD
(ISO 15705:2003-01)
COD vario MR* 620 Dichromate/H2SO4 1 16 mm ø 0 - 1500 mg/l COD
(ISO 15705:2003-01)
COD vario HR* 620 Dichromate/H2SO4 1 16 mm ø 0 - 15000 mg/l COD
Copper 559
b)
Biquinoline 50 mm o 0.05 - 1 mg/l Cu 4
24 mm ø 0.5 - 5 mg/l
Copper VARIO 560 Bicinchoninate 24 mm ø 0.05 - 5 mg/l Cu
Cyanide 585 Pyridine-barbituric acid 1 50 mm o 0.005 - 0.2 mg/l CN
24 mm ø 0.01 - 0.5 mg/l
DEHA 562 PPST 3 24 mm ø 0.02 - 0.5 mg/l DEHA
Photometry
DPD 1 Reagent-Solution Liquid reagent / 15 ml 47 10 20
DPD 3 Solution Liquid reagent / 15 ml 47 10 30
VARIO Chlorine FREE-DPD/F10 Powder Pack / 100 53 01 00
VARIO Chlorine TOTAL-DPD/F10 Powder Pack / 100 53 01 20
CHLORINE HR (PI) Tablet / 100 51 30 00
ACIDIFYING GP Tablet / 100 51 54 80
Combi pack# each 100 51 77 21
CHLORINE HR (KI)/ACIDIFYING GP each 250 51 77 22
DPD No. 1 Tablet / 100 51 10 60
DPD No. 1 HIGH CALCIUM f) Tablet / 100 51 57 40
GLYCINE (additionally for ClO2 in presence of chlorine) Tablet / 100
Combi pack# each 100 51 77 31
DPD No.1 / GLYCINE each 250 51 77 32
DPD 1 Buffer-Solution Liquid reagent / 15 ml 47 10 10
DPD 1 Reagent-Solution Liquid reagent / 15 ml 47 10 20
CHLORIDE T1 Tablet / 100 51 59 10
CHLORIDE T2 Tablet / 100 51 59 20
RT (Chlorid-51 / Chlorid-52) Reagent test (Liquid reagent) 2 41 90 31
approx. 50-75 tests
"PERSULF. RTG FOR CR", Powder Pack / 100 53 73 00
Chromium Hexavalent Powder Pack /100 53 73 10
COD VARIO 0 - 150 mg/l Tube test / 25 2 42 07 20
Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 85
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents
Set 53 60 00
SPADNS-Reagent Liquid reagent / 250 ml 46 74 81
Fluoride-Standard c) Liquid / 30 ml 20 56 30
e)
Spectroquant® 1.14678.0001 Reagent test / approx. 75-100 Tests 42 07 51
Photometry
Spectroquant® 1.14500.0001 e) Tube test / 25 42 07 52
HARDCHECK P Tablet / 100 51 56 60
HARDCHECK P Tablet / 100 51 56 60
--- --- ---
Hydrazine Test Powder Powder / 30 g 46 29 10
Spoon 38 49 30
VARIO Hydra 2 Liquid reagent / 100 ml 53 12 00
HYDROGENPEROXIDE LR Tablet / 100 51 23 80
IRON (II) LR Tablet / 100 51 54 20
Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 87
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents
Photometry
VARIO Molybdenum HR3 F10 Powder Pack / 100
Set 53 53 00
RT (Nickel-51, Nickel-52) Reagent test (Powder, 2 41 90 33
Liquid reagent) / 50 Tests
VARIO Nitrate Chromotropic Powder Pack / 50
VARIO Nitra X Reagent tube Reaktionsküvette / 50
VARIO Deionised water (for Zero) Bottle 100 ml
Set 53 55 80
TT (Nitrat-111) Tube test 2 42 07 02
(Liquid reagent) / 24
NITRITE LR Tablet / 100 51 23 10
TT (Nitrit-101) Tube test (Powder) / 24 2 41 90 18
Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 89
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents
pH value 558 Phenol red 5 24 mm ø 6.5 - 8.4 pH
pH value 558 Phenol red 5 24 mm ø 6.5 - 8.4 pH
Potassium 730 Tetraphenylborate, Turbidity 4 24 mm ø 1 - 10 mg/l K
Silica 820 Silicomolybdate 1,2 24 mm ø 0.05 - 3 mg/l SiO2
Photometry
PHOSPHATE No. 1 LR Tablet / 100 51 30 40
PHOSPHATE No. 2 LR Tablet / 100 51 30 50
Combi pack# each 100 51 76 51
PHOSPHATE No.1 LR / No.2 LR each 200 51 76 52
PHOSPHATE No. 1 HR Tablette / 100 51 58 10
PHOSPHATE No. 2 HR Tablette / 100 51 58 20
Combi pack# each 100 51 76 61
PHOSPHATE No.1 HR / No.2 HR each 200 51 76 62
VARIO PHOS3 F10 Powder Pack / 100 53 15 50
VARIO Diluent Vial Tubes / 50
VARIO PHOS3 F10 Powder Pack / 50
VARIO Deionised water (for Zero) Bottle 100 ml
Set 53 52 00
TT Tube test / 24 2 42 07 01
Tubes Acid reagent / 50
VARIO Phosphate RGT, 10 ml Powder Pack / 50
VARIO Deionised water (for Zero) Bottle 100 ml
1N NaOH Bottle / 100 ml
1.54N NaOH Bottle / 100 ml
VARIO POTASSIUM PERSULFATE Powder Pack / 50
Set 53 52 50
Tubes Acid reagent / 50
VARIO Phosphate RGT, 10 ml Powder Pack / 50
VARIO Deionised water (for Zero) Bottle 100 ml
1.54N NaOH Bottle / 100 ml
VARIO POTASSIUM PERSULFATE Powder Pack / 50
Set 53 52 10
PHENOLRED/PHOTOMETER Tablet / 100 51 17 70
PHENOLRED Solution Liquid reagent / 15 ml 47 10 40
POTASSIUM T Tablette / 100 51 56 70
SILICA No. 1 Tablet / 100 51 31 30
SILICA No. 2 Tablet / 100 51 31 40
Combi pack# each 100 51 76 71
SILICA No.1 / No.2 each 200 51 76 72
SILICA PR (in presence of phosphate) Tablet / 100 51 31 50
Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 91
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents
Photometry
Set 53 57 00
Legend
a)
Wavelength, b) Determination of free, combined and total is possible.
c)
User must re-calibrate the photometer with each new batch of SPADNS-reagent. Therefore the Fluoride-standard is needed
d)
Determination of total chromium, Cr(VI) and Cr(III) is possible
e)
Please order from your Merck local dealer. Spectroquant® is a Merck KGaA Trademark.
f)
alternative reagent, used instead of DPD No.1 where turbidity in the water sample caused by high concentration of calcium and/or high conductivity
g)
additionally required for determination of chlorine dioxide / ozone in the presence of chlorine 93
h)
additionally required for samples with hardness values above 300 mg/l CaCO3
* Reactor necessary for COD (150°C), TOC (120°C), total -chromium, -nitrogen, -phosphate (100°C)
RT Reagent test: Includes complete set of chemicals necessary for this method
TT Tube test: Includes vials and reagents